Radiographic Technique
Radiographic Technique
After the end of the discussion, the student should be able to:
Define radiographic technique.
Discuss milliampere second (mAs) and kilovolt peak (kVp) in relation to x-ray beam
quantity and quality.
Describe characteristics of the imaging system that affect x-ray beam quantity and
quality.
List the four patient factors and explain their effects on radiographic technique.
Identify four image-quality factors and explain how they influence the characteristics of
a radiograph.
RADIOGRAPHIC TECHNIQUE
Has more effect than any other factor on image receptor exposure because it affects
beam quality and, to a lesser degree, influences beam quantity.
With increasing kVp, more x-rays are emitted, and they have higher energy and
greater penetrability. Unfortunately, because they have higher energy, they also
interact more by Compton effect and produce more scatter radiation, which results
in reduced image contrast.
The kVp selected helps to determines the number of x-rays in the image-forming
beam, and hence the resulting average optical density (OD).
Finally, and perhaps most important, the kVp controls the scale of contrast on the
finished radiograph because as kVp increases, less differential absorption occurs.
Therefore, high kVp results in reduced image contrast.
MILLIAMPERES
The mA selected determines the number of x-rays produced and therefore the
radiation quantity.
As more electrons flow through the x-ray tube, more x-rays are produced.
A change in mA does not change the kinetic energy of electrons flowing from
cathode to anode. It simply changes the number of electrons. Consequently, the
energy of the x-rays produced is not changed; only the number is changed.
Often, x-ray imaging systems are identified by the maximum x-ray tube current
possible. Inexpensive radiographic imaging systems designed for private physicians’
offices normally have a maximum capacity of 600 mA. Interventional radiology
imaging systems may have a capacity of 1500 mA.
EXPOSURE TIME
Radiographic exposure times usually are kept as short as possible. The purpose is not to
minimize patient radiation dose but rather to minimize motion blur that can occur because
of patient motion.
Producing a diagnostic image requires a certain radiation exposure of the image receptor.
Therefore, when exposure time is reduced, the mA must be increased proportionately to
provide the required x-ray intensity.
An easy way to identify an x-ray imaging system as single phase, three phase, or high
frequency is to note the shortest exposure time possible. Single-phase imaging systems
cannot produce an exposure time less than 1/2 cycle or its equivalent 8 ms (10 ms on 50-Hz
generators). Three-phase and high-frequency generators normally can provide an exposure as
short as 1 ms. mA and exposure time (in seconds) are usually combined and used as mAs.
Indeed, many x-ray consoles do not allow the separate selection of mA and exposure time
and permit only mAs selection.
EXPOSURE TIME
Conic filters, either concave or convex, find application in digital fluoroscopy, in which the image
receptor, the image intensifier tube, is round.
COMPENSATING FILTERS
The radiologic technologist cannot select the type of high-voltage generator to be used for a
given examination. That choice is fixed by the type of x-ray imaging system that is used. Still, it
is important to understand.
How the various high-voltage generators affect radiographic technique and patient dose.
Three basic types of high-voltage generators are available: single phase, three phase, and high
frequency.
The radiation quantity and quality produced in the x-ray tube are influenced by the type of
high-voltage generator that is used.
A half-wave–rectified generator has 100% voltage ripple. During exposure with a half-wave–
rectified generator, x-rays are produced and emitted only half the time. During each negative
half-cycle, no x-rays are emitted.
HIGH-VOLTAGE GENERATION
A half-wave–rectified generator has 100% voltage ripple. During exposure with a half-
wave–rectified generator, x-rays are produced and emitted only half the time. During
each negative half-cycle, no x-rays are emitted.
Half-wave rectification is used rarely today. Some mobile and dental x-ray imaging
systems are half-wave rectified. The voltage waveform for full-wave rectification is
identical to that for half-wave rectification except there is no dead time. During
exposure, x-rays are emitted continually as pulses. Consequently, the required
exposure time for full-wave rectification is only half that for half-wave rectification.
Three-phase power comes in two principal forms: 6 pulse or 12 pulse. The difference
is determined by the manner in which the high-voltage step-up transformer is
engineered.
HIGH-VOLTAGE GENERATION
The difference between the two forms is minor but does cause a detectable change
in x-ray quantity and quality. Three-phase power is more efficient than single-phase
power. More x-rays are produced for a given mAs setting, and the average energy of
those x-rays is higher. The x-radiation emitted is nearly constant rather than pulsed.
High-frequency generators were developed in the early 1980s and are increasingly
used. The voltage waveform is nearly constant, with less than 1% ripple.
At present, high-frequency generators are used increasingly with dedicated
mammography systems, computed tomography (CT) systems, and mobile x-ray
imaging systems. It is likely that most high-voltage generators of the future will be of
the high-frequency type regardless of the required power levels.
CHARACTERISTICS OF THE VARIOUS TYPES OF HIGH-VOLTAGE
GENERATORS
EQUIVALENT TECHNIQUE
(kVp/mAs)
Generator Type Percentage Ripple Relative Quantity Chest Abdomen
Half wave 100 100 120/20* 74/40*
Full wave 100 200 120/20 74/40
3 phase, 6 pulse 14 260 115/6 72/34
3 phase, 12 pulse 4 280 115/4 72/30
High frequency <1 300 112/3 70/24
*The milliampere second value equals that for a full-wave generator; exposure time is doubled. kVp, kilovolt peak;
mAs, milliampere seconds.
PATIENT FACTORS
The patient’s size, shape, and physical condition greatly influence the required
radiographic technique.
The general size and shape of a patient is called body habitus; four such states have
been described. Sthenic—meaning “strong, active”—patients are average patients.
Hyposthenic patients are thin but healthy appearing; these patients require less
radiographic technique. Hypersthenic patients are big in frame and usually
overweight. Asthenic patients are small, frail, sometimes emaciated, and often elderly.
Recognition of body habitus is essential to radiographic technique selection. After this
has been established, the thickness and composition of the anatomy being examined
must be determined.
THICKNESS
kVp 80 80 80 80 80 80 80 80
Patient 16 18 20 22 24 26 28 30
thickness
(cm)
mAs 12 15 22 30 45 60 90 120
VARIABLE KILOVOLT PEAK TECHNIQUE FOR AN
ANTEROPOSTERIOR PELVIS EXAMINATION
The type of pathology, its size, and its composition influence radiographic technique.
In this case, the patient examination request form and previous images may be of
some help. The radiologic technologist should not hesitate to seek more information
from the referring physician, the radiologist, or the patient regarding the suspected
pathology.
Some pathology is destructive, causing the tissue to be more radiolucent. Other
pathology can constructively increase mass density or composition, causing the tissue
to be more radiopaque.
These changes can have a direct influence on the attenuation of the x-ray beam by
the affected tissue, and ultimately the selection of the technical x-ray exposure
factors.
CLASSIFYING PATHOLOGY
The fourth image-quality factor is distortion, the misrepresentation of object size and shape on the
radiograph.
Because of the position of the x-ray tube, the anatomical part, and the image receptor, the final image
may misrepresent the object.
Poor alignment of the image receptor or the x-ray tube can result in elongation of the image.
Elongation means that the anatomical part of interest appears bigger than normal.
Poor alignment of the anatomical part may result in foreshortening of the image. Foreshortening
means that the anatomical part appears smaller than normal.
Distortion can be minimized through proper alignment of the tube, the anatomical part, and the image
receptor. This alignment is fundamentally important for patient positioning.
REFERENCE: