User Verification of Abbott Alinity HQ Hematology
User Verification of Abbott Alinity HQ Hematology
Research Article
background (limit of blank [LoB]), and linearity throughout the operation, patient samples are sequentially pipetted into these blocks
expected range of results [4–10]. However, the assessment of to enable high throughput. Here, for a more comprehensive assess-
ment, the precision of each block was evaluated separately. In addition
analytical performance on different levels is usually not
to calibration and quality control (QC), carry-over should also be
detailed, and interpretation is the responsibility of the labora- assessed on the two incubation blocks [12].
tory professionals. It is up to the laboratory specialist or labo- The CELL-DYN Ruby (Abbott Laboratories, Diagnostic Division,
ratory procedures to decide which standard to follow or which Abbott Park, IL, USA) provides 22 blood count parameters, including a
verification limit to use [10]. five-part WBC differential. The system utilizes the Multiangle Polarized
Scatter Separation technology and laser flow cytometry. The CELL-DYN
This study aimed to verify the performance characteris-
Ruby also provides an integral reticulocyte analysis and nuclear optical
tics of the Alinity HQ hematology analyzer compared with the
count [13].
previously used Abbott Cell Dyn Ruby in a routine laboratory
setting according to the CLSI H26-A2, CLSI EP15-A3, CLSI EP09,
CLSI EP17, and ICSH guidelines [4–8]. We analyzed precision Performance evaluation
(short-term and long-term), accuracy (method comparison
and estimation of bias as a measure of trueness), confirmation Precision studies: The precision study was performed in two stages:
of a LoB, and carry-over to offer guidance on the verification short-term precision and long-term precision, as recommended by the
ICSH and CLSI guidelines, EP15-A3 and H26-A2, respectively [4–6].
of the automated hematology analyzer.
Imprecision was reported in terms of CV%. All precision studies were
performed for both incubation blocks.
Abbott Cell Dyn Ruby analyzer. The results were then evaluated using Carry-over: Carry-over was assessed by analyzing three pairs of samples
Passing–Bablok regression analysis and Bland–Altman plots. To [4, 5]. The data yield from a high-concentration sample followed by a low
ensure a comprehensive comparison of various CBC parameters, we concentration could be evaluated by running sample A (high sample)
utilized a minimum of 120 patient samples. These samples were spe- three times (A1, A2, and A3), followed by sample B (low sample) three
cifically selected from diverse departments, including but not limited times (B1, B2, and B3). The percentage of carry-over for RBC, PLT, and
to normal, hematological, oncological, adult, and pediatric units. Due WBC parameters was calculated as follows:
to inherent workload and cost constraints, we maintained the sample
Carryover% = [(B1 − B3)/(A3 − B3)] × 100
size at around 120 for method comparison according to the CLSI EP09
guideline [7].
During the study, IQC materials from Abbott Diagnostics (IL, USA) Statistical evaluation: All statistical analyses were performed using R
were assessed every 8 h, and monthly proficiency testing (PT) was statistical software 4.3.1. In this study, we utilized up-to-date imprecision
conducted using RANDOX quality controls (County Antrim, United and bias targets sourced from the biological variation (BV) due to its
Kingdom) to ensure the analytical performance of the Abbott Cell Dyn more rigorous criteria instead of adhering to the clinical decision levels
Ruby analyzer. All necessary corrective actions were performed outlined in CLSI H26-A2. The precision study conducted with patient
according to the laboratory policy for IQC results, and all results were samples was evaluated according to BV imprecision limits sourced from
evaluated as acceptable for proficiency testing. the EFLM Biological Variation database [14]. Repeatability and within-
laboratory precision results for IQC materials were evaluated according
to the manufacturer’s claims. Finally, long-term precision results were
Estimation of the bias: The estimation of bias was performed using PT
assessed using the BV imprecision data. All limits related to precision
samples in bias studies, as they are prioritized over IQC samples
and accuracy studies are given in Supplementary Table 1. The method
according to the CLSI guideline. However, the study was designed to
comparison results were not normally distributed, and Spearman cor-
span three days due to constraints with the PT samples, allowing for
relation coefficients were thus calculated for method comparisons. A
only three replicates in the left incubation block. In accordance with
degree of agreement between the same parameters analyzed with the
CLSI EP15-A3 [6], four RANDOX quality control PT samples were used to
two hematology analyzers was evaluated using the non-parametric
evaluate whether the results obtained by the laboratory were within
Passing–Bablok regression method. The mean percentage differences
the verification limits. Grand means x were determined, and standard
were calculated from the Bland–Altman plots and were used to assess
error of means (sex) were calculated using the sr and swl values
the agreement between the methods. The minimum allowable bias from
obtained from the precision study. Peer results from the PT were used
the BV database was used as the limit of agreement [14].
as the target value (TV), and the standard error of the target value
(serm) was calculated using peer-group standard deviation and number
of peers. Using sex and serm combined, standard error (sec) of mean and
TV were calculated. The multiplier factor (m) was calculated for four Results
samples, and verification intervals (VI) were determined. The calcu-
lation steps are given in the following equations:where m is multipli-
cation factor; PT is proficiency testing; SD is standard deviation; sec is
Precision study results
combined standard error; serm is standard error of the target value; sex
is standard error of means; TV is target value, and VI is verification Short-term precision
interval.
Repeatability and within-laboratory imprecision
Based on the repeatability results (Table 1) regarding patient
samples, it was observed that for the RBC measurements, the
CV% values for all levels in both incubation blocks were
consistently lower than the optimal analytical allowable
imprecision (CVA) of 0.7 %, which was calculated in accor-
dance with BV data. For the PLT measurements, the CV% for
Evaluation of limit of blank: The LoB is determined by analyzing mul- the low level was less than the minimum CVA determined via
tiple sample replicates with no analyte present and then computing the BV data at 5.7 %. Moreover, the CV values were under the
average result and the standard deviation. The determination of the LoB
optimum CVA of 1.9 % for the medium and high levels in the
for WBCs, RBCs, and platelets (PLTs) was conducted in accordance with
the CLSI EP17 guideline, utilizing 60 blank samples and employing the two incubation blocks. Finally, the WBC measurements
following formula [8]: showed that the CV values for all levels in both incubation
LoB = M B + C p × SDb
blocks remained generally below the minimum CVA as
1.645 defined by BV data. An exception was seen in the left block
cp =
readings, where the low-level CV value slightly exceeded the
1 − (4(B−K)
1
)
minimum CVA threshold.
where LoB is limit of blank; MB is mean of blank samples; SDb is standard The results for the evaluation of repeatability CV (CVr)
deviation of blank samples; Cp is multiplier; B is total number of blank and within-laboratory CV (CVwl) values obtained according
results; and K is number of blank samples. to manufacturer claims are given in Table 2 and Figure 1. As
4 Bayraktar and Topcu: User verification of Alinity HQ
Table : Repeatability study results conducted with patient samples. Table 2 shows, for both blocks, the CVwl values in level 1
and level 2 MCV measurements are above the verification
Test Level Left block Right block limit calculated according to the CV values given by the
Mean CV, % Mean CV, % manufacturer. Moreover, CVwl values in level 1 and level 2
WBC, × /L Low . . . . MCHC QC measurements of the left block were above the
Normal . . . compliance limit calculated according to the CV values
High . . . . given by the manufacturer. All other measurement results
RBC, ×/L Low . . . . were within calculated limits.
Normal . . . .
High . . . .
PLT, ×/L Low . . , . Long-term precision
Normal . .
High . .
When the long-term precision data were analyzed, the CV values
CV, coefficient of variation; WBC, white blood cell. obtained for the basophil count at all levels and for eosinophil
Table : Repeatability and within-laboratory study results conducted with internal quality control (IQC) materials provided by the manufacturer.
CVr, CVwl, Mean Actual Limit Actual Limit Mean Actual Limit Actual Limit
% % CVr, % CVr, % CVwl, % CVwl, % CVr, % CVr, % CVwl, % CVwl, %
WBC, ×/L Level . . . . . . . . . . . .
Level . . . . . . . . . . . .
Level . . . . . . . . . . .
Neutrophils, Level . . . . . . . . . .
×/L Level . . . . . . . . . .
Level . . . . . . . . . .
Lymphocytes, Level . . . . . . . . . . . .
×/L Level . . . . . . . . . .
Level . . . . . . . . . .
Monocytes, Level . . . . . . . . . . . .
×/L Level . . . . . . . . . .
Level . . . . . . . . .
Basophils, ×/ Level . . . . . . . .
L Level . . . . . .
Level . . . . . . . . . .
Eosinophils, Level . . . . . . . .
×/L Level . . . . . . . . . . . .
Level . . . . . . . . . .
RBC, ×/L Level . . . . . . . . . . . .
Level . . . . . . . . . . . .
Level . . . . . . . . . . . .
Hemoglobin, Level . . . . . . . . . . . .
g/dL Level . . . . . . . . . .
Level . . . . . . . .
Hematocrit, % Level . . . . . . . . .
Level . . . . . . . . . . . .
Level . . . . . . . . . . . .
MCV, fL Level . . . . . . . . . .
Level . . . . . . . . .
Level . . . . . . . . . .
MCH, pg Level . . . . . . . . . . . .
Level . . . . . . . . . . . .
Level . . . . . . . . . . .
MCHC, g/L Level . . . . . . . . . . .
Level . . . . . . . . . . .
Level . . . . . . . . . . . .
Bayraktar and Topcu: User verification of Alinity HQ 5
Table : (continued)
CVr, CVwl, Mean Actual Limit Actual Limit Mean Actual Limit Actual Limit
% % CVr, % CVr, % CVwl, % CVwl, % CVr, % CVr, % CVwl, % CVwl, %
RDW-SD, fL Level . . . . . . . , . .
Level . . . . . . . . . .
Level . . . . . . . . . . . .
PLT, ×/L Level . . . . . . . . . .
Level . . . . . . . .
Level . , . . . . . .
MPV, fL Level . . . . . . . . . . . .
Level . . . . . . . . . . . .
Level . . . . . . . . . . . .
CV, coefficient of variation; CVr, repeatability CV; CVwl, with-in laboratory CV.
Figure 1: Internal quality control (IQC) precision study results. Black line represents actual coefficient of variation (CV) values. CV% values are plotted on
the x-axis, with repeatability and within-lab CVs depicted on the y-axis. Various colors distinguish different QC sample levels. Solid bars represent
manufacturer claims, while transparent bars indicate extended verification limits obtained from the precision study. A black bar highlights acquired CV%
values. Black line represents actual CV values. Green, pink and blue bars represent level 1, level 2 and level 3 QC materials respectively. Dark colors indicate
manufacturer claims and transparent colors indicate limits calculated according to the CLSI EP15A3. CV, coefficient of variation; CVr, repeatability CV;
CVwl, within-laboratory CV.
and monocyte counts at level 1 were above the minimum CVA Accuracy study results
value determined according to BV data. Similarly, the MCH,
MCHC, and RDW-SD parameters also exhibited values above Method comparison
the minimum permitted CVA for all levels and blocks. In
terms of MCV values, the level-1 CV for both blocks and level-2 In the patient result comparison, the correlation coefficients
CV for the right block were higher than CVA. In addition, varied from 0.46 to 1.00, as shown in Table 4. According to
level three CV values for RBC and hematocrit measurements the Passing–Bablok regression analysis results, the slope
were above the minimum permitted CVA (Table 3). All other confidence interval included 1, and the intercept confidence
measurement results were found to follow the BV claims. interval included 0 only for the eosinophil and monocyte
6 Bayraktar and Topcu: User verification of Alinity HQ
Table : Intermediate precision for days using manufacturer internal quality control (IQC) materials.
Low level Normal level High level Low level Normal level High level
Mean CV, % Mean CV, % Mean CV, % Mean CV, % Mean CV, % Mean CV, %
WBC, × /L . . . . . . . . . . . .
Neutrophils, ×/L . . . . . . . . . . . .
Monocytes, ×/L . . . . . . . . . . . .
Lymphocytes, ×/L . . . . . . . . . . . .
Basophils, ×/L . . . . . . . . . . . .
Eosinophils, ×/L . . . . . . . . . . . .
RBC, ×/L . . . . . . . . . . . .
Hemoglobin, g/dL . . . . . . . . . . . .
Hematocrit, % . . . . . . . . . . . .
MCV, fL . . . . . . . . . . . .
MCH, pg . . . . . . . . . . . .
MCHC, g/L . . . . . . . . . . . .
RDW-SD, fL . . . . . . . . . . . .
PLT, ×/L . . . . . . . . . . . .
MPV, fL . . . . . . . . . . . .
tests (Figure 2). The confidence intervals for the other tests were analyzed, a concentration-dependent pattern was
were extremely close to the desired intervals, except for the observed in the MCHC, MCH, and MPV parameters, including
MCHC and basophil tests. When the Bland–Altman plots the calculation shown in Figure 3.
WBC, ×/L .–. .–. . . (.–.) . (.–.) . (.) .
Neutrophils, .–. .–. . . (.–.) . (.–.) . (.) .
×/L
Monocytes, .–. .–. . . (.–.) −. (−. to .) −. (−.) .
×/L
Lymphocytes, .–. .–. . . (.–.) . (−. to .) . (.) .
×/L
Basophils, ×/ .–. .–. . . (.–.) −. (−. to −.) . (.) .
L
Eosinophils, .–. .–. . . (.–.) . (−. to .) . (.) .
×/L
Erythrocytes, .–. .–. . . (.–.) . (.–.) . (.) .
×/L
Hemoglobin, .–. .–. . . (.–.) . (.–.) . (.) .
g/dL
Hematocrit, % .–. .–. . . (.–.) . (−. to .) . (.) .
MCV, fL .– .– . . (.–.) −. (−. −. (−.) .
to −.)
MCH, pg .–. .–. . . (.–.) . (.–.) −. (−.) .
MCHC, g/L .–. .–. . . (.–.) . (.–.) −. (−.) .
RDW-SD, fL .–. .–. . . (.–.) . (.–.) −. (−.) .
PLT, ×/L .– .– . . (.–.) −. (−. to .) −. (−.) .
MPV, fL .–. .–. . . (.–.) . (.–.) −. (−.) .
CI, confidence interval; Diff., difference. aThe minimum allowable bias from BV database.
Bayraktar and Topcu: User verification of Alinity HQ 7
Figure 2: Passing–Bablok regression analysis. Identity lines (y=x) are dashed green, confidence intervals are claret and regression lines are blue.
Figure 3: Bland–Altman plots for method comparison study. y=0 lines are dashed green, mean differences (%) are blue, confidence intervals are solid
gray. Second y axis represents CIs. Green area indicates the limit of agreement (minimum allowable bias 24.17 %).
8 Bayraktar and Topcu: User verification of Alinity HQ
The results of all parameters were within the verification The LoB was calculated as 0.086 × 109/L for WBCs,
ranges, except for the low-level QC results of PLT, as shown 0.002 × 1012/L for RBCs, and 0.331 × 109/L for PLTs, below the
in Table 5. manufacturer’s claims, as shown in Supplementary Table 2.
m, multiplication factor; sec, combined standard error; serm, standard error of the target value; sex, standard error of means. aGrand mean must be between
verification interval.
Bayraktar and Topcu: User verification of Alinity HQ 9
target value, which had a 95 % probability of true difference. Acknowledgments: We would like to thank the staff mem-
For all parameters, the estimated bias in this study was bers of hematology.
within the Vis, except for the low-level QC results for PLT Research ethics: The study was approved by the Ethics
count. Committee of the Başkent University with the decision
According to the relevant guidelines and previous numbered KA21/333, and dated October 14, 2022.
studies, a LoB check of the hematology analyzer is impor- Informed consent: None declared.
tant. The manufacturer defines the limit of background Author contributions: The authors have has accepted
electronic noise; for optimal conditions, the background responsibility for the entire content of this manuscript and
counts should be zero. This is especially important in body approved its submission.
fluids with extremely low cell counts, such as cerebrospinal Competing interests: The authors state no conflict of
fluid [4, 5, 17]. Here, the LoB was calculated as 0.086 × 109/L interest.
for WBCs, 0.002 × 1012/L for RBCs, and 0.331 × 109/L for PLTs, Research funding: None declared.
with all below the manufacturer’s claims. Data availability: Not applicable.
The carry-over values in this analytical performance
study were better than those recommended by the Alinity
HQ manufacturer; the carry-over for WBC and PLT must be
1 % and for RBC≤0.5 %. These data showed that there was no References
transfer of material from one sample to the next, and the
influence of contamination was close to zero. Many pub- 1. Bruegel M, George TI, Feng B, Allen TR, Bracco D, Zahniser DJ, et al.
Multicenter evaluation of the cobas m 511 integrated hematology
lished studies recommend additional washing between one
analyzer. Int J Lab Hematol 2018;40:672–82.
sample and the next to avoid a background count carry-over 2. Woo KS, Jeong IH, An GD, Lim HH, Han JY. Performance evaluation of
[18–20]. new Abbott Alinity HQ hematology analyzer. Int J Lab Hematol 2019;41:
A key strength of our study is that while the CLSI H26 657–63.
guideline specifies targets only for WBC, RBC, hemoglobin, 3. Clinical Laboratory Standards Institute. H20-A2: reference Leukocyte
(WBC) differential count (proportional) and evaluation of instrumental
hematocrit, and PLT parameters, we expanded the perfor-
methods, 2nd ed. Wayne, PA: Clinical Laboratory Standards Institute;
mance evaluation to include both repeatability and bias for
2007.
WBC subgroups and RBC indices. Another strength was the 4. Briggs C, Culp N, Davis B, d’Onofrio G, Zini G, Machin SJ. ICSH guidelines
innovative precision plot, which succinctly represents CV for the evaluation of blood cell analysers including those used for
values, facilitating the evaluation of obtained CVs and differential leucocyte and reticulocyte counting. Int J Lab Hematol 2014;
enabling users to quickly identify any out-of-range values. 36:613–27.
5. Clinical and Laboratory Standards Institute (CLSI). H26-A2: validation,
The one limitation of the study is that bias and LoB
verification and quality assurance of automated hematology analyzers,
studies were performed in a single incubation block due to 2nd ed. Wayne (PA): CLSI document; 2010.
the limited amount of both patient and EQC samples. How- 6. Clinical and Laboratory Standards Institute (CLSI). EP15-A3: user
ever, the precision studies performed in both incubation verification of precision and estimation of bias; approved guideline, 3rd
blocks enabled us to demonstrate the agreement between ed. Wayne (PA): CLSI Document; 2014.
7. Clinical and Laboratory Standards Institute (CLSI). EP09c:
them.
measurement procedure comparison and bias estimation using
patient samples, 3rd ed. Wayne (PA): CLSI Document; 2018.
8. Clinical and Laboratory Standards Institute (CLSI). EP17-A2: evaluation
Conclusions of detection capability for clinical laboratory measurement procedures,
2nd ed. Wayne (PA): CLSI Document; 2012.
In conclusion, despite some measurements not meeting ex- 9. ISO. Medical laboratories – particular requirements for quality and
competence ISO 15189. Geneva: ISO; 2012.
pected requirements, the performance evaluation and
10. Vis JY, Huisman A. Verification and quality control of routine
comparative analyses of the automated Alinity HQ hematol- hematology analyzers. Int J Lab Hematol 2016;38:100–9.
ogy system indicated an acceptable analytical performance. 11. Van der Beken Y, Van Dalem A, Van Moer G, Segers E, Damiaens S,
Furthermore, the laboratory results were comparable to Hoffmann J, et al. Performance evaluation of the prototype
those of the established laboratory practice analyzers. Alinity Abbott Alinity hq hematology analyzerInt. J Lab Hematol 2019;41:
448–55.
HQ can be useful for all laboratories, mainly those with a large
12. Lakos G. Calibration and QC guidance for the Abbott Alinity HQ
number of normal and abnormal samples. Our research also
hematology analyzer. In: Diagnostics. Santa Clara, CA: Abbott; 2018;
showed that conducting long-term precision for hematology 1–7 pp.
analyzers can be challenging compared with when dealing 13. Lehto T, Hedberg P. Performance evaluation of Abbott CELL-DYN Ruby
with chemistry analyzers. for routine use. Int J Lab Hematol 2008;30:400–7.
Bayraktar and Topcu: User verification of Alinity HQ 11
14. Aarsand AK, Fernandez-Calle P, Webster C, Coskun A, Gonzales-Lao E, 18. Nakul-Aquaronne D, Sudaka-Sammarcelli I, Ferrero- Vacher C, Starck B,
Diaz-Garzon J, et al. The EFLM biological variation database. https:// Bayle J. Evaluation of the Sysmex Xe-2100 hematology analyzer in
biologicalvariation.eu/ [Accessed August 2023]. hospital use. J Clin Lab Anal 2003;17:113–23.
15. Khodaiji S, Mansukhani D. Performance evaluation of fully automated 19. Maciel TS, Comar SR, Beltrame MP. Performance evaluation of the
5-Part differential hematology analyzer ELite 580 (ERBA). Indian J Sysmex® XE-2100D automated hematologyanalyzer. J Bras Patol Med
Hematol Blood Transfus 2022;38:99–105. Lab 2014;50:26–35.
16. Becker PH, Fenneteau O, Da Costa L. Performance evaluation of the 20. Kaplan SS, Johnson K, Wolfe N, Brown W, Keeney M, Gray-Statchuk L,
Sysmex XN-1000 hematology analyzer in assessment of the white blood et al. Performance characteristics of the Coulter LH 500 hematology
cell count differential inpediatric specimens. Int J Lab Hematol 2016;38: analyzer. Lab Hematol 2004;10:76–87.
54–63.
17. Verbrugge SE, Huisman A. Verification and standardization of blood cell
counters for routine clinical laboratory tests. Clin Lab Med 2015;35: Supplementary Material: This article contains supplementary material
183–96. (https://doi.org/10.1515/tjb-2023-0185).