Technical Data 3870

The document provides technical specifications for autoclaves, including: - Operating conditions are limited to temperatures of 5-40°C, humidity below 85%, and altitude below 2000 meters. - Electrical specifications include power requirements of 3300W or 6000W depending on the model, at voltages of 230V or 400V. - Water quality guidelines specify limits for various contaminants in feed water and condensate to protect the autoclave. - Autoclave dimensions and weights are provided for two models. Utilities required are specified. - Construction materials include stainless steel for the chamber, door and outer cabinet. Insulation is fiberglass reinforced.

Uploaded by

fqalves
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
152 views18 pages

Technical Data 3870

The document provides technical specifications for autoclaves, including: - Operating conditions are limited to temperatures of 5-40°C, humidity below 85%, and altitude below 2000 meters. - Electrical specifications include power requirements of 3300W or 6000W depending on the model, at voltages of 230V or 400V. - Water quality guidelines specify limits for various contaminants in feed water and condensate to protect the autoclave. - Autoclave dimensions and weights are provided for two models. Utilities required are specified. - Construction materials include stainless steel for the chamber, door and outer cabinet. Insulation is fiberglass reinforced.

Uploaded by

fqalves
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 18

3.

Technical Data
Storage Conditions

The packed or unpacked autoclave shall be retained in indoor


conditions!

Operating Conditions

only.
Only autoclavable materials shall be used.
The environment shall not exceed an ambient temperature range of
5ºC (41ºF)-40ºC (104ºF) and a relative humidity of 85% respectively.
The operation altitude shall not be over 2000 meters (6561 feet)
(ambient pressure shall not be lower than 80 kPa (11.6 psi)).
The autoclave shall not be used in a manner not specified in this
manual!
Do not use the autoclave in the presence of dangerous gases.
The packed or unpacked device shall be ions.
Operate the autoclave only in the manner specified in the manual. If
the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.

Caution!
Waste water should be brought into the public net in accordance
with the local rules or requirements i.e. only non-hazardous
liquids shall be disposed in public sewage!

Page 11
Directives and Standards

Every autoclave meets the provisions of the following Directives and is in


compliance with the following Standards:

ISO 9001: Quality Management System


ISO 14001: Environmental Management System

ISO 17025: General requirements for the competence of testing


and calibration autoclaves
ASME Code Section I and section VIII. Div. I
PED 2014/68/EU
Chinese Regulations Special Equipment Licensing Office
IEC 61010-1 / UL 61010-1: Safety requirements for electrical equipment for
measurement, control, and laboratory use Part 1:
General requirements
IEC 61010-2-040: Safety requirements for electrical equipment for
measurement, control, and laboratory use Part 2-
040: Requirements for sterilizers and washer-
disinfectors used to treat medical materials
EN 61326-1: EMC Requirements for Electrical Equipment

Environment Emission Information

A. Peak sound level generated by the sterilizer is « 70 / dBA with a back-


sound level of 60 dBA.
B. Total heat transmitted by the sterilizer is < 600 W/h.

Page 12
Electrical Data

230V (WR only) configuration

Property Value: 3850 Value: 3870


Total Power 3300W 3300W
Voltage 1Ph, 230VAC 1Ph, 230VAC
Amperage 15A 15A
Protection against electrical
Class I (IEC 61010-1)
shock
Mains supply fluctuation +/- 10%
Degree of protection by
IP31
enclosure

400V configuration

Property Value: 3850 Value: 3870


Total Power 6000W (3*2000W) 6000W (3*2000W)
Voltage 3Ph, 400VAC 3Ph, 400VAC
Amperage 9A 9A
Protection against electrical
Class I (IEC 61010-1)
shock
Mains supply fluctuation +/- 10%
Degree of protection by
IP31
enclosure

Page 13
Water Quality

The distilled or mineral-free water supply shall be according to the table


below:
Suggested maximum limits of contaminants in water for steam
sterilization per EN13060

Substance Feed Water Condensate


Evaporate residue
Silicate (SiO2)
Iron 0.1mg/l
Cadmium mg/l
Lead
Rest of heavy metals
except iron,
cadmium, lead
Chloride (Cl)
Phosphate
Conductivity (at
15 µs/cm m
20°C)
pH value 5 to 7.5 5 to 7
Hardness
Appearance Colorless, clean, without sediments
Note: The condensate is produced from steam taken from the empty
sterilizer chamber.

Compliance with the above data should be tested in accordance with


acknowledged analytical methods, by an authorized laboratory.
Attention:
The use of water for autoclaves that do not comply with the table
above may have severe impact on the working life of the sterilizer
and can invalidate the manufact
Use only deionized water, having a maximum conductivity of 15
µs/cm. Conductivity greater than 15 µs/cm may cause failures.

Tap water supply


The range of hardness value 0.7-2.0 mmol/l (70- 200 mg/l CaCO3)
The use of soft water is strictly forbidden!
Please consult a water specialist!

Page 14
3.1.1 Reverse Osmosis
A Reverse Osmosis (RO) system may be used to improve the quality
of the water used to generate steam in the autoclave chamber.
In RO, the water is forced through a semi-penetrable membrane, which
filters out contaminants to a high degree of efficiency. In deionization
(DI) ions and charged particles are removed either by electric fields or
by ion exchange in resin beds.
Although the RO cannot normally attain the degree of purity possible
with the DI methods, it is more than adequate for the feed water
intended for clean-steam generators.
Moreover, the RO has several advantages:
1. RO is cheaper to install and to run than DI.
2. RO removes particulate matter, organic molecules, and pyrogens that
DI cannot remove.
3. RO water is less corrosive to steel and copper than DI water.
4. RO maintenance requirements are less demanding than those of the
DI units.
Therefore, the use of mineral free water will contribute to better
performance and longer life of the autoclave.

Page 15
Specifications

Model
3850 3870
Property
Overall Height 900mm 1000mm
dimensions
Width 730mm 730mm
(standard
configuration) Length 540mm 540mm
Overall Height 1080mm 1080mm
dimensions
Width 730mm 730mm
(WR
configuration) Length 7300mm 7300mm
Maximum Width 800mm 800mm
dimensions
(door open) Height 1400mm 1500mm
Net weight 108kg 113kg
Weight per support area According to overall weight and floor loading
(max. load) requirements
Shipping volume 0.9 m3 0.9 m3
Shipping weight 136kg 141kg
length 900mm 900mm
Shipping
width 730mm 730mm
dimensions
height 1350mm 1350mm
Max. Allowable Working 2.8 bar
2.8 bar
pressure (MAWP)
diameter 380mm 380mm
Chamber
Depth 490mm 690mm
Chamber Volume 65L 85L

Page 16
Construction

Chamber and door material Stainless steel


Outer Cabinet Stainless Steel
Chamber insulation Fiberglass with reinforced material
Water reservoir Stainless Steel

Utilities

3.1.2 Electrical utility

Switch Recommended
Power
box Circuit Breaker
208-220V (3 Ph + Earthing) No 20A
400V (3 Ph + Neutral +
No 16A
Earthing)
230V 1ph No 16A
Frequency 50/60Hz

3.1.3 Other Utilities

Compressed Air (C,


1/2", 3 Bar (44-58 psi)
cooling models only)
Tap water 1/2", 2-6 Bar (29-44 psi)
Mineral free water 1/2", 2-3 Bar (29-44 psi)
2" Minimum Withstanding
Drain
temp. of 80°C (176°F)

Attention:
A switch or circuit-breaker must be included in the building
installation. This switch or circuit-breaker shall be in close
proximity to the equipment, within easy reach of the operator; and
marked as the disconnecting device for the equipment.

Page 17
The electrical net must be protected with a current leakage safety
relay.
The electrical network must comply with local rules or
regulations.
Verify that there is an easy access to the main power switch, to
the water cut-off valve and to the current leakage safety relay.
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
All water connections to autoclave must be performed through
"BACK FLOW PREVENTION SYSTEM" only, as per IEC 61770.

Page 18
Stickers Description

Symbol Meaning Part Number Location


Caution! Hot LAB048-0058 Near the safety
steam. valve

Protective earth LAB048-0020 Near the earthing


(Ground) screw

Caution! Hot LAB048-0023 On the rear cover


surface. above the water
strainer
Press & Hold LAB048-0461 On the door right
(Automatic edge
door only)
Displayed = 10
sec.
Do not Touch LAB048-0462 On the door frame
Microswitches behind the door
please Contact cover
Tuttnauer
Service
Do not Touch LAB048-0463 On the door frame
Microswitches behind the door
please Contact cover
Tuttnauer
Service
ON On-Off LAB048-0018 Near the Power
switch

OFF

Page 19
Front View: Standard Configuration

12

No. Description No. Description


1 Front leg 8 Display
2 Rear leg 9 Keyboard
3 Left service door 10 Printer cover
4 Left service door grip 11 Paper slot and paper cutter
5 Left service door lock 12 USB socket
6 Door cover 13 Main switch
7 Pressure gauge 14 Printer

Page 20
Front View: What is special for WR Configuration

No. Description
1 Glass tube indicator of the reservoir water level
2 Inlet for manual water reservoir filling

Front View: What is Special for Fan Configuration

There is a configuration equipped with a fan that allows to shorten the


cooling operation. The fan is located on the door of the autoclave.

Fan

Page 21
Rear View: Standard Configuration

No. Description No. Description


1 Drain outlet 8 Safety Valve
2 Mineral free water inlet 9 Circuit breaker
3 Compressed air inlet 10 Ground connector
4 Cavitation port 11 City (tap) water inlet for cooling coil
Power supply electrical
5 12 Air pressure gauge
cord
6 Cut-off 13 Air pressure regulator
7 Fan grill

Page 22
Rear View: What is special for WR Configuration

No. Description
1 Water reservoir drain outlet
2 Water reservoir overflow
3 Wheel

Page 23
Overall Dimensions: 3850ELVC-D mm

Page 24
Overall Dimensions (3850ELV-WR-D Configuration), mm)

Page 25
Overall Dimensions: 3870ELVC-F-D, mm

Page 26
Overall Dimensions: 3870ELV-PRC-D, mm

Page 27
Overall Dimensions: 3870ELVC- WR-D Configuration, mm

Page 28