ID CID F en
ID CID F en
Catalog No.00000
Bulletin No. 2340
ID-C/ID-F
High-performance
ABS Digimatic
Indicator
New-generation ID series making
measurement operations smoother and
enhancing production quality
Bidirectional serial communication that helps increase work efficiency
Meeting the need for more precise A wide range of support functions
measurements for smoother measurement work
2
The first Mitutoyo measuring tools to support bidirectional serial communication.
Dramatically improve work efficiency by connecting and linking with a PC.
The ID-C/ID-F units are Mitutoyo’s first measuring tools to support bidirectional serial communication.* They can be easily
connected and linked with a PC via a USB input tool, etc., and in addition to conventional measurement data collection, they
also enable control and setting of the ID-C/ID-F units, collection of gauge information, and other operations to be performed
in batch from the PC. This contributes to drastic improvement in work efficiency.
*Achieved through I/F compatible with an original bidirectional serial communication specification (Digimatic S1). See P.6 for details.
An optional cable and measurement data input unit are required for bidirectional serial communication. USB-ITPAK V3.0 must be installed on the PC used for communication.
Improved work efficiency thanks to excellent readability Setting of frequently used functions for easy operation Improved measurement
work efficiency
Large screen and analog bar Three large buttons
Simple calculation
The units have large screens that can display Ease of use is greatly
various information in an easy-to-read manner. improved by three large
function
They also have an analog bar, convenient for buttons. You can freely The result of the spindle
observing subtle movements such as the approach set any frequently used movement value multiplied
to tolerance. functions to these buttons. by the calculation coefficient
can be displayed in real
Previous model Addition of New model time. This reduces the work
analog bar
of measuring with a jig or
Parameter setting mode similar tool.
• Counting direction switching
• Tolerance judgment function setting • Resolution switching
ID-C ID-C • Calculation function setting • Function lock setting f(x) = Ax
Improved analog bar S witching between ABS length measurement system
(presetting) and INC length measurement system (zero setting) f(x): Displayed value
Character size Power ON/OFF x: Spindle movement value
enlarged about • Data output (when connected to an external device)
1.5 times A: Selected value
ID-F ID-F • Data hold (when not connected to an external device)
3
ID-C Series – Battery Operated
543-720B 543-722B
543-730B 543-737B
ID-C0525NXB ID-C0525ENXB ID-C0550NXB ID-C1050ENXB
4
SPECIFICATIONS
Inch/Metric ISO/JIS type (Ø8 mm stem, M2.5 x 0.45 contact point) ASME/ANSI /AGD type (Ø3/8"stem, 4-48 UNF contact point)
Order No. Maximum permissible error MPE*1 Measuring Net mass(g)
Inter-
Range Resolution Hysteresis Repeatability force Flat
w/ lug Flat back MPEE*3 face W/lug
MPEH MPER MPL (N) back
543-701 543-701B 1.5 or less d1/d2/S1 175 165
543-702 543-702B 0.5 in/ 1.5 or less d1/d2/S1 195 165
543-706 *2
543-706B *2 12.7 mm ±0.00012 in/ 0.4 to 0.7 d1/d2/S1 170 160
0.00002/0.00005/
543-707*2 543-707B*2 0.0001/0.0005 in 0.003 mm 0.00008 in/ 0.00008 in/ 0.4 to 0.7 d1/d2/S1 190 160
— 543-721B 1 in/ 0.0005/ 0.001/0.01 mm 0.002 mm 0.002 mm 1.8 or less d1/d2/S1 — 195
25.4 mm (selectable)
— 543-722B 1.8 or less d1/d2/S1 — 195
— 543-731B 2 in/ ±0.0002 in/ 2.3 or less d1/d2/S1 — 260
— 543-732B 50.8 mm 0.005 mm 2.3 or less d1/d2/S1 — 260
543-711 543-711B 0.9 or less d1/S1 170 160
543-712 543-712B 0.5 in/ 0.9 or less d1/S1 190 160
543-716*2 543-716B*2 12.7 mm ±0.001 in/ 0.2 to 0.5 d1/S1 165 155
543-717*2 543-717B*2 0.0005 in/ 0.02 mm 0.001 in/ 0.0005 in/ 0.2 to 0.5 d1/S1 185 155
— 543-726B 1 in/ 0.01 mm 0.02 mm 0.01 mm 1.8 or less d1/S1 — 190
— 543-727B 25.4 mm 1.8 or less d1/S1 — 190
— 543-736B 2 in/ ±0.0015 in/ 2.3 or less d1/S1 — 245
— 543-737B 50.8 mm 0.04 mm 2.3 or less d1/S1 — 245
*1 These values apply at 20 °C.
*2 Low measuring force
*3 Error of indication for the total measuring range
Metric
Order No. Maximum permissible error MPE*1 (mm) Measuring Net mass(g)
Range Resolution Inter-
Hysteresis Repeatability force Flat
w/ lug Flat back (mm) (mm) MPEE*3 face W/lug
MPEH MPER MPL (N) back
543-700 543-700B 1.5 or less d1/d2/S1 175 165
*2 *2
12.7
543-705 543-705B 0.003 0.4 to 0.7 d1/d2/S1 170 160
0.0005/0.001/0.01(selectable) 0.002 0.002
— 543-720B 25.4 1.8 or less d1/d2/S1 — 195
— 543-730B 50.8 0.005 2.3 or less d1/d2/S1 — 260
543-710 543-710B 0.9 or less d1/S1 170 160
12.7
543-715*2 543-715B*2 0.02 0.2 to 0.5 d1/S1 165 155
0.01 0.02 0.01
— 543-725B 25.4 1.8 or less d1/S1 — 190
— 543-735B 50.8 0.04 2.3 or less d1/S1 — 245
*1 These values apply at 20 °C.
*2 Low measuring force
*3 Error of indication for the total measuring range
Common Specifications
12.7 mm/0.5 in models Low measuring force models*1 25.4 mm/1 in, 50.8 mm/2 in models
Display 7 segments height: 11.0 mm, Analog bar (±20 scale)
Display rotation 330 °
Protection level*2 Equivalent to IP-42
0.0005 mm models: Plunger downward only
Possible plunger direction All directions Up to direction in which plunger is horizontal
0.01 mm models: Up to direction in which plunger is horizontal
Power supply Lithium metal battery CR2032 (1pc.)
Battery life*3 Approx. 2.5 years (normal use), Approx. 2,700 hours(continuous use)
Detection method Electrostatic capacitance type absolute linear encoder
Response speed No limit
Errors, Alarms Various setting errors, Sensor error, Display overflow, etc.
Operating temperature 0 to 40 °C
Storage temperature -10 to 60 °C
*1: See the order numbers with an asterisk 2 (*2) in the table above.
*2: Protection level (IP=International Protection) is based on IEC 60529/DIN40050 Part 1/JIS D0207, C0920. The levels shown are valid for factory conditions only.
*3: When the data processor is not connected. Battery life depends on use of the indicator. Use the above value as a guide.
5
DIMENSIONS
.5"/12.7 mm range models
ASME/ANSI/AGD Type ISO/JIS Type
Unit: mm
Unit: in
25.4mm=1”
50.1
1/4 5
ø2.32
ø2.13
ø59
ø54
ø1
ø6
/4
.5
1.74
46.5
2.00
54.4
0 0
ø0.375 -0.0012 ø8 -0.009
0.79 0.58
17
ø0.19 ø4.8
21.2
0.52
13.3
1/4
7.3
Ball contact (ø0.12) 0.30 Ball contact (ø3) 7.6
21BZB005 901312
(thread: No. 4-48UNF) (thread: M2.5×0.45)
Note: Products with an Order No. suffixed "B" have a flat back, and other models have a center-lug back.
ø2.32
ø59
2.67
70
3.30
84.8
0 0
ø0.375-0.0012 ø8 -0.009
0.68
19.5
40.8
1.02
21BZB005 901312
26
7.3
6
2"/50.8 mm range models
ASME/ANSI/AGD Type ISO/JIS Type
Unit: in ø0.76 1.31 0.43 Unit: mm ø19.2 33.2 11
25.4mm=1”
85.2
3.35
ø2.32
ø59
97.3
3.74
110.6
4.32
0
0
ø0.375 -0.0012 ø8 -0.009
0.68
19.5
65.3
2.62
2.05
52
Comparison of functions
ID-C Series ID-F Series
Preset ✓ ✓
Zero set ✓ ✓
Peak detection (Max, Min, TIR) ✓ ✓
Unit system switching*1 ✓ ✓
Counting direction switching ✓ ✓
Resolution selecting ✓*2 ✓
Tolerance judgment ✓ ✓
Simple calculation ✓ ✓
Analog bar display ON/OFF ✓ ✓
Analog bar scale selecting ✓ ✓
Key customize ✓ ✓
Function lock ✓ ✓
Calibration schedule warning function ✓ ✓
Auto OFF ✓ —
Reset all settings ✓ ✓
*1: in/mm models only
*2: Except 0.01 mm/0.0005 in models
7
ID-F Series – AC Powered
SPECIFICATIONS
Inch/Metric ISO/JIS type (Ø8 mm stem, M2.5 x 0.45 contact point) ASME/ANSI /AGD type (Ø3/8"stem, 4-48 UNF contact point)
Maximum permissible error
MPE*2 Measuring
Inter- Mass
Order No.*1 Range Resolution force
*3 Hysteresis Repeatability MPL (N) face (g)
MPEE
MPEH MPER
±0.0001 in/
543-852A 1 in/25.4 mm 1.8 or less d1/d2/S1 240
0.0025 mm
0.005/0.001/0.01 mm ±0.00016 in/ 0.00008 in/ 0.00008 in/
543-854A 2.3 or less d1/d2/S1 330
0.00002/0.00005/0.0001/0.0005/0.001 in (selectable) 0.004 mm 0.002 mm 0.002 mm
2 in/50.8 mm
±0.00012 in/
543-858A 2.3 or less d1/d2/S1 330
0.003 mm
*1 To denote your AC power cable add the following suffixes to the order No.: A for UL/CSA, D for CEE, DC for CCC, E for BS, K for KC, No suffix is required for JIS/100 V
*2 These values apply at 20 °C. *3 Error of indication for the total measuring range
Metric
Maximum permissible error
MPE*2 (mm) Measuring
Range Resolution Inter- Mass
Order No.*1 force
(mm) (mm) Hysteresis Repeatability MPL (N) face (g)
MPE E*3
MPEH MPER
543-851A 25.4 0.0025 1.8 or less d1/d2/S1 240
543-853A 50.8 0.0005/0.001/0.01 (selectable) 0.004 0.002 0.002 d1/d2/S1 330
2.3 or less
543-857A 50.7 0.003 d1/d2/S1 330
Common Specifications
Display 7 segments height: 11.0 mm, Analog bar (±20 scale) Response speed No limit
Display rotation 330 ° Errors, Alarms Various setting errors, Sensor error, Display overflow, etc.
Protection level*1 Equivalent to IP-40 Output d1, d2
Possible plunger direction Up to direction in which plunger is horizontal I/O S1
Power supply AC adapter (DC 5.9 V) Operating temperature 0 to 40 °C
Detection method Electrostatic capacitance type absolute linear encoder Storage temperature -10 to 60 °C
*1: Protection level (IP=International Protection) is based on IEC 60529/DIN40050 Part 1/JIS D0207, C0920. The levels shown are valid for factory conditions only.
8
DIMENSIONS
1"/25.4 mm range models
ASME/ANSI/AGD Type ISO/JIS Type
Unit: in
Unit: mm
25.4mm=1”
ø 0.76 1.50 0.43 ø19.2 38.1 11
57.7
2.27
ø 2.32
ø59
1.38 35
2.67
70
84.8
3.30
ø 8 -0.009
0
0.68
19.5
ø0.375 -0.0012
0
ø4.8
Ball contact (ø0.12) ø 0.19 Ball contact (ø3)
40.8
1.66
1.02
21BZB005 901312
26
7.3
(thread: No. 4-48UNF) (thread: M2.5×0.45)
1/4
85.2
ø 2.32
ø59
1.38 35
97.3
3.74
4.32
110.6
0.68
0
ø 0.375 -0.0012
0 ø 8 -0.009
19.5
ø4.8
2.62
2.05
ø 0.19
65.3
21BZB005 901312
(thread: No. 4-48UNF) (thread: M2.5×0.45)
7.3
9
Related options
OPTIONS
Measurement data input unit U-WAVE
Can be used for both ID-C/ID-F with SPC cable. If using U-WAVE, please note that it does not support bidirectional serial communication.
No.02AZG021
Connection unit U-WAVE-T dedicated connection cable for
(For ID-C series 12.7 mm type only) foot switch
No.06AGL011 No.02AZF670
USB dongle
Measurement data collection software USB-ITPAK
V3.0 can be downloaded from our website. The above
dongle is required to use the full functions.
No.06AGL021 No.06AGR543
Measurement data collection software
SPC Connection cable (2 m)
USB-ITPAK V3.0
Many other options are also available. For details, please visit our website. https://www.mitutoyo.co.jp
10
SOFTWARE Reduces the time and effort needed for inspection work
Measurement data collection software
USB-ITPAK V3.0
USB-ITPAK is useful software to create procedures when inputting measurement data into Excel sheets. The latest version allows the
user to perform batch power-on for ID-C/ID-F units, batch power-off at the end of measurement, batch zero setting and presetting, data
acquisition instruction from a PC, etc.
Equipped with an automatic sorting function for sorting input measurement data [Easy input mode]
This function can be implemented even if the measuring instrument
does not support bidirectional serial communication. After setting, With normal input With automatic sorting function
measurement values are automatically sorted into an Excel sheet as (Entered into column A only.) (Once entered into column A, similar data is automatically classified.)
needed.
Only the number of measurement
items is preset.(Example: number of
measurement items = 3)
D: 10 mm
H: 20 mm
W: 30 mm
Symbols: ✓
1 : Can be used only when connected with USB-ITPAK V3.0 and ID-C/ID-F;
USB-ITPAK V2.1/V3.0 Function comparison table ✓ : Can be used; — : Cannot be used
ITPAK
11
© 2022 Mitutoyo America Corporation
Coordinate Measuring Machines Vision Measuring Systems Form Measurement Optical Measuring
Whatever your challenges are, Mitutoyo
supports you from start to finish.
Note: All information regarding our products, and in particular the illustrations, drawings, dimensional and performance M3 Solution Centers:
data contained in this printed matter as well as other technical data are to be regarded as approximate average values. We
therefore reserve the right to make changes to the corresponding designs. The stated standards, similar technical regulations, Aurora, Illinois (Headquarters)
descriptions and illustrations of the products were valid at the time of printing. In addition, the latest applicable version of our
General Trading Conditions will apply. Only quotations submitted by ourselves may be regarded as definitive. Specifications Boston, Massachusetts
are subject to change without notice.
Charlotte, North Carolina
Mitutoyo products are subject to US Export Administration Regulations (EAR). Re-export or relocation of our products may
require prior approval by an appropriate governing authority. Cincinnati, Ohio
Trademarks and Registrations Detroit, Michigan
Designations used by companies to distinguish their products are often claimed as trademarks. In all instances where Mitutoyo
America Corporation is aware of a claim, the product names appear in initial capital or all capital letters. The appropriate Los Angeles, California
companies should be contacted for more complete trademark and registration information. Seattle, Washington
Houston, Texas