UPS Interface Box Configuration

The document provides information about an interface box used to connect digital and relay input/output modules to a CAN bus. The interface box includes power supply, digital input, relay output modules, and connection terminals. It describes the module configurations and indicators, and provides instructions for setting the CAN bus speed and node ID during initial configuration.

Uploaded by

Nicolae Arpentii
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
86 views3 pages

UPS Interface Box Configuration

The document provides information about an interface box used to connect digital and relay input/output modules to a CAN bus. The interface box includes power supply, digital input, relay output modules, and connection terminals. It describes the module configurations and indicators, and provides instructions for setting the CAN bus speed and node ID during initial configuration.

Uploaded by

Nicolae Arpentii
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

UP S Interface Box

DI1/DI2 - Digital Input modules- opto isolated inputs


SM1 DI1 DI2 RO1 END
to CAN bus.
STOP A C A C A C
RUN B D B D RO1 - Relay Output module-relay output signals
TX
OVERFLOW
from CAN bus.
RX
I/O SM1 - Power supply module supplies necessary
voltages to other modules on the bus.
Bus connection is accomplished with hidden data
contacts on the sides of the modules.
END – A blank module to keep the other modules
squeezed together and keep bus contacts in contact.
During boot and also when CAN communication is
not present, the TX/RX LEDs will blink alternately.
When CAN communication is enabled/restored, the
TX/RX LEDs will blank and the I/O LED will flash
intermittently then turn on solid green.

Configuration procedure:
NORMAL UPS Power NORMAL • Switch OFF mod. pwr sup (unplug J7-24V to CANopen module)
• Set all Dip switches (SW) on OFF position
DI1-A=ON DI1-A=ON DI1 DI2 RO1 • Supply power to the module (+24V)
DI1-B=OFF DI1-B=OFF • Set the CAN bus speed: set SW 1 ON and SW2 ON for 250K. Stop &
DI1-C=OFF DI1-C=OFF Run LEDS blink/are lit.
DI1-D=ON DI1-D=ON • Save speed setting by setting SW8 ON. Blinking Led is lit.
DI2-A=ON DI2-A=ON UPS Power • Switch OFF the module power supply (+24V)
DI2-B=OFF DI2-B=ON • Set all switches (SW) on OFF position
DI2-C=OFF DI2-C=ON DI1 DI2 RO1 • Configure node Id = 8 by setting SW4 to ON
DI2-D=OFF DI2-D=ON • Supply power to the module (+24V). The node Id will be read by the
RO1-A=OFF RO1-A=ON system at each power on/module boot
RO1-C=OFF RO1-C=OFF
Revision DATE PAGE DRAWN BY
REV 0 01/31/2007 1 OF 3 Fillet
UPS Interface Box
J1
50
Power Supply
Show
25 J10 J5A CanOPEN Fieldbus Coupler
C1 28260 43
18 35 CAN L COM1 6 SM1
RT BUS/
CAN BUS
36 CAN H COM1 2 Field
Bus I/F
UL-PDB
49, 50 GND 1 5
J2 7
18
43
C2 28259 25 2 6
50
CAN
Electronics

J3 J6 CR1 F1 J7 3 7
13
28262

+24V 1 1 1
ETH 4 CAN 24V 2

DS1 5V
Bus
5V 4 8
R2 Modules
4.7K 24V DI1
3 3
14 GND 2 6 CAN 0V 4 DI2
RO1 750-
XA
24VDC 1
612
XA
Supply 2 24
J2 CE PDB
24 11
24V RO1
INVERTER DI1 A C
ON 11 A C B D PDBR7 C3
B D 1 5
SNMP R 8 8 1 5 NO1 NO2
BD C2
DI1 DI2
STOP C8 R7
2 6 C4
X-RAY GEN
OPERATION 120vac A C ON UTILITY
9 9 2 6 C1 C2
A
165.100.10.5 24v 24v

16 16
3 7
LOAD ON 3 7 NC1 NC2
PDB CE
INVERTER C6
15 24v 24v
15
C8
X-RAY GEN
28257

4 8 ON UPS
UPS 2 2 4 8
B D DI2 J4 TB1
UPS BATTERY
DI3 DI4
A C BNC
B D 14 42
FAULT 3 750-
39
3 750- 1 5 517
432
C8
DI1 DI2
PW-SW-CMD R7
13 41
4 4 TB3-3
2 6 38 24Vac
RECTIFIER A C

28258
24v 24v R1 C1
MAIN
FAILURE 5 5
18 18 10 39
36
3 7
24v 24v
GENERAL SWITCH STATUS C6 C4
ALARM
19 19 11 40
37
4 8 GENERATOR
21 21 B D
DI3 DI4 ON UPS
RECTIFIER
ON 22 22 750-
432

EMERGENCY
OFF TO UPS
30
30
X2-1
EPO X2-2 28256 R3
R3
31
31

Revision DATE PAGE DRAWN BY


REV 0 01/31/2007 2 OF 3 Fillet
UPS Interface Box
Element Description Definition Related SUIF Errors**
DS1 Green LED ON: +24 UPS Aux Power Supply
STOP Red LED ON: The bus coupler/node is in the STOP state
TX OVERFLOW Red LED CAN Transmitter is full-Flashes when no CAN signal (ie on boot up)
RX OVERFLOW Red LED CAN Receiver buffer is full-Flashes when no CAN signal
I/O RED/GRN/ORG Inidcation of node operation-flashes when node is being accessed
ON=NORMAL 410007 Unexp STOP operation
DI1-A Stop Operation OFF=UPS OFF 3 min timer start received while powered on main
OFF=Load on Inverter (NORMAL)
410001
DI1-B Load on Inverter ON=Load on Bypass (I.e.during bootup)
DI1-C Not used Not Used
ON=Battery OK (NORMAL)
410013 Battery Test Failure
DI1-D UPS Battery Fault OFF=Battery Fault
ON=UPS OK (NORMAL) 410012 UPS error. Check UPS
DI2-A General Alarm OFF=UPS Alarm (check front panel) error log.
Lets DL know that PDB has switched to UPS power
410017 UPS Pwr switch failure on
OFF=Jedi on Utility (NORMAL)
switch status
DI2-B PDB Switch Status ON=Jedi on UPS
OFF=Utility power OK (NORMAL) 410012 UPS error. Check UPS
DI2-C Rectifier Main Failure ON=Utility power lost, switch to battery power error log.
OFF=Rectifier ON & synced to Utility power (NORMAL)
ON=Recifier OFF--On Battery Power OR error OR No Error
DI2-D Rectifier On rectifier in process of syncing (blinking rect icon) to line power.

Command to PDB to switch the power from utility mains to UPS and visa-versa
OFF=Jedi on Utility (NORMAL)
RO1-A PDB PW_SW_CMD ON=Jedi on UPS
Cmd from DL to start inverter upon Innova boot up.
ON=DL starting inverter-ON for approx 4 seconds at bootup
RO1-C Inverter ON CMD OFF=NORMAL

** Related Errors were induced one at a time while in an exam and on utility power. Signals were forced into the 'false'/non-
normal state either by opening the circuit or closing the circuit.

Revision DATE PAGE DRAWN BY


REV 0 01/31/2007 3 OF 3 Fillet