UP S Interface Box
DI1/DI2 - Digital Input modules- opto isolated inputs
SM1 DI1 DI2 RO1 END
to CAN bus.
STOP A C A C A C
RUN B D B D RO1 - Relay Output module-relay output signals
TX
OVERFLOW
from CAN bus.
RX
I/O SM1 - Power supply module supplies necessary
voltages to other modules on the bus.
Bus connection is accomplished with hidden data
contacts on the sides of the modules.
END – A blank module to keep the other modules
squeezed together and keep bus contacts in contact.
During boot and also when CAN communication is
not present, the TX/RX LEDs will blink alternately.
When CAN communication is enabled/restored, the
TX/RX LEDs will blank and the I/O LED will flash
intermittently then turn on solid green.
Configuration procedure:
NORMAL UPS Power NORMAL • Switch OFF mod. pwr sup (unplug J7-24V to CANopen module)
• Set all Dip switches (SW) on OFF position
DI1-A=ON DI1-A=ON DI1 DI2 RO1 • Supply power to the module (+24V)
DI1-B=OFF DI1-B=OFF • Set the CAN bus speed: set SW 1 ON and SW2 ON for 250K. Stop &
DI1-C=OFF DI1-C=OFF Run LEDS blink/are lit.
DI1-D=ON DI1-D=ON • Save speed setting by setting SW8 ON. Blinking Led is lit.
DI2-A=ON DI2-A=ON UPS Power • Switch OFF the module power supply (+24V)
DI2-B=OFF DI2-B=ON • Set all switches (SW) on OFF position
DI2-C=OFF DI2-C=ON DI1 DI2 RO1 • Configure node Id = 8 by setting SW4 to ON
DI2-D=OFF DI2-D=ON • Supply power to the module (+24V). The node Id will be read by the
RO1-A=OFF RO1-A=ON system at each power on/module boot
RO1-C=OFF RO1-C=OFF
Revision DATE PAGE DRAWN BY
REV 0 01/31/2007 1 OF 3 Fillet
UPS Interface Box
J1
50
Power Supply
Show
25 J10 J5A CanOPEN Fieldbus Coupler
C1 28260 43
18 35 CAN L COM1 6 SM1
RT BUS/
CAN BUS
36 CAN H COM1 2 Field
Bus I/F
UL-PDB
49, 50 GND 1 5
J2 7
18
43
C2 28259 25 2 6
50
CAN
Electronics
J3 J6 CR1 F1 J7 3 7
13
28262
+24V 1 1 1
ETH 4 CAN 24V 2
DS1 5V
Bus
5V 4 8
R2 Modules
4.7K 24V DI1
3 3
14 GND 2 6 CAN 0V 4 DI2
RO1 750-
XA
24VDC 1
612
XA
Supply 2 24
J2 CE PDB
24 11
24V RO1
INVERTER DI1 A C
ON 11 A C B D PDBR7 C3
B D 1 5
SNMP R 8 8 1 5 NO1 NO2
BD C2
DI1 DI2
STOP C8 R7
2 6 C4
X-RAY GEN
OPERATION 120vac A C ON UTILITY
9 9 2 6 C1 C2
A
165.100.10.5 24v 24v
16 16
3 7
LOAD ON 3 7 NC1 NC2
PDB CE
INVERTER C6
15 24v 24v
15
C8
X-RAY GEN
28257
4 8 ON UPS
UPS 2 2 4 8
B D DI2 J4 TB1
UPS BATTERY
DI3 DI4
A C BNC
B D 14 42
FAULT 3 750-
39
3 750- 1 5 517
432
C8
DI1 DI2
PW-SW-CMD R7
13 41
4 4 TB3-3
2 6 38 24Vac
RECTIFIER A C
28258
24v 24v R1 C1
MAIN
FAILURE 5 5
18 18 10 39
36
3 7
24v 24v
GENERAL SWITCH STATUS C6 C4
ALARM
19 19 11 40
37
4 8 GENERATOR
21 21 B D
DI3 DI4 ON UPS
RECTIFIER
ON 22 22 750-
432
EMERGENCY
OFF TO UPS
30
30
X2-1
EPO X2-2 28256 R3
R3
31
31
Revision DATE PAGE DRAWN BY
REV 0 01/31/2007 2 OF 3 Fillet
UPS Interface Box
Element Description Definition Related SUIF Errors**
DS1 Green LED ON: +24 UPS Aux Power Supply
STOP Red LED ON: The bus coupler/node is in the STOP state
TX OVERFLOW Red LED CAN Transmitter is full-Flashes when no CAN signal (ie on boot up)
RX OVERFLOW Red LED CAN Receiver buffer is full-Flashes when no CAN signal
I/O RED/GRN/ORG Inidcation of node operation-flashes when node is being accessed
ON=NORMAL 410007 Unexp STOP operation
DI1-A Stop Operation OFF=UPS OFF 3 min timer start received while powered on main
OFF=Load on Inverter (NORMAL)
410001
DI1-B Load on Inverter ON=Load on Bypass (I.e.during bootup)
DI1-C Not used Not Used
ON=Battery OK (NORMAL)
410013 Battery Test Failure
DI1-D UPS Battery Fault OFF=Battery Fault
ON=UPS OK (NORMAL) 410012 UPS error. Check UPS
DI2-A General Alarm OFF=UPS Alarm (check front panel) error log.
Lets DL know that PDB has switched to UPS power
410017 UPS Pwr switch failure on
OFF=Jedi on Utility (NORMAL)
switch status
DI2-B PDB Switch Status ON=Jedi on UPS
OFF=Utility power OK (NORMAL) 410012 UPS error. Check UPS
DI2-C Rectifier Main Failure ON=Utility power lost, switch to battery power error log.
OFF=Rectifier ON & synced to Utility power (NORMAL)
ON=Recifier OFF--On Battery Power OR error OR No Error
DI2-D Rectifier On rectifier in process of syncing (blinking rect icon) to line power.
Command to PDB to switch the power from utility mains to UPS and visa-versa
OFF=Jedi on Utility (NORMAL)
RO1-A PDB PW_SW_CMD ON=Jedi on UPS
Cmd from DL to start inverter upon Innova boot up.
ON=DL starting inverter-ON for approx 4 seconds at bootup
RO1-C Inverter ON CMD OFF=NORMAL
** Related Errors were induced one at a time while in an exam and on utility power. Signals were forced into the 'false'/non-
normal state either by opening the circuit or closing the circuit.
Revision DATE PAGE DRAWN BY
REV 0 01/31/2007 3 OF 3 Fillet