Product# : G3272B, G3273B
Product Description: Agilent 7500cx, 7500cs Series ICP-MS
Site Preparation Specification
Purpose of Procedure You must use Site Preparation Manual (PN G3270-90206) for
To assure that the installation of essential detail information. Please follow the site preparation instructions
instruments and systems can be provided in the manual. This document is only an overview of the main
completed successfully by careful requirements.
preparation and evaluation of the Tick Boxes
installation site and by ensuring the Procedure Checklist
availability of appropriate utilities,
consumables and supplies.
Dimensions and Weight
Customer Responsibilities
Customers should ensure that all Agilent 7500 Series ICP-MS:
necessary operating supplies;
consumables and usage dependent items Weight: 175 kg 386 lbs
required for the successful installation of Width: 110 cm 43.3 in
instruments and systems are available. Depth: 60 cm (72cm with optional duct)
Installation sites should be prepared in 23.6 in (28.3in with optional duct)
accordance with the following Height: 64 cm (83cm with optional duct)
specifications.
25.2 in (32.7in with optional duct)
The Agilent 7500 Series ICP-MS, its data
system and accessories will be delivered At least 60 cm (2 ft) on all sides of the instrument
to your site in large and small shipping must be kept clear for maintenance access.
containers. The largest container is
approximately Power Consumption
102 cm (40.2 in) High
122cm (48.0 in) Wide Agilent 7500 Series ICP-MS:
85 cm (33.5 in) Deep Single Phase 200-240VAC 30 amps 47-63 Hz,
230 kg (506 lbs) Weight
The containers will be delivered in a
Rush current 150A 15 millisecond
large truck. You must furnish a forklift,
or other suitable lifting device, and make Plug: Nema L6-30P, Cable length: 5 meter, if this
arrangements to unload the truck and plug type is not compatible with your electrical
transport the containers to your site. All outlet, the customer must provide a safe electrical
doorways, hallways, floors and elevators connection to the ICP-MS. A certified electrician
must be able to accommodate the largest, provided by the customer must perform this.
heaviest container. Do not open any of
the shipping containers unless a Stability:
representative of Agilent Technologies is
present.
Sags or surges: < 3 Cycles duration time
< +15% of mean voltage
Important Information Transients: Low Energy <1micro second
If you have problems in providing any of @ 1KV
the following, please contact your local High Energy <50 micro
Agilent Technologies office for seconds @ 1KV
assistance. Assistance with user specific
applications may be provided but should RFI susceptibility: 3V per meter
be contracted separately. Users of the Permitted Magnetic field 1.0 Gauss (1.0x10-4
instrument should be present throughout Tesla)
the installation and familiarization
otherwise important operational,
maintenance and safety information may Isolated, noise free ground
be missed.
Computer System, Autosampler & Modem:
Five(5) 100 – 127, 200-240V power outlets, total
combined current = 15 Amps.
Issued: May 2007 – Rev. 2.0 PN# G3270-90231 TAD Product Support Page 1 of 6
Product# : G3272B, G3273B
Product Description: Agilent 7500cx, 7500cs Series ICP-MS
Site Preparation Specification
Tick Boxes
Heat Exchanger:
One(1) 100 - 120V, 50/60Hz or 200 – 240V, 50Hz,
10 Amp power supply
Plug: Standard appliance connector for country of
purchase.
Water Recirculating Chiller:
One(1) 200 - 230V, 12 Amp 50/60Hz power
supply.
Plug: Standard appliance connector for country of
purchase (Neslab M75)
Exhaust and Venting Requirements
Agilent 7500cx:
> 4.0 m3/min <6.0 m3/min
(>12.5 ft/s <18.7 ft/s)
Agilent 7500cs (and 7500cx with optional duct) :
> 7.0 m3/min <8.0 m3/min
(>21.7 ft/s <24.9 ft/s)
This flow must be maintained when connected to the
instrument. The backpressure of the instrument is
approximately 140Pa.
Stability of Exhaust Venting requirement: +/- 10%
This will remove approximately 2000 watts (6,824 Btu / hour)
The ventilation ducting must have an ID of 150mm (5.9
inches)
Ducting between the instrument and lab extraction system is
supplied by the customer.
Environmental Conditions:
Temperature: 15-30°C (59-86°F) <2 °C/h change
and total change should be < 5°C
Rel. Humidity: 20 - 80% non-condensing,
non corrosive atmosphere
Altitude: up to 2000 meter
Agilent Technologies recommends you minimize the
airborne particle density by producing a better than
class 10,000 laboratories, and for ultimate
performance better than class 100 is recommended.
Issued: May 2007 – Rev. 2.0 PN# G3270-90231 TAD Product Support Page 2 of 6
Product# : G3272B, G3273B
Product Description: Agilent 7500cx, 7500cs Series ICP-MS
Site Preparation Specification
Tick Boxes
Heat Dissipation
Chiller in the room with the Agilent 7500 :
50Hz 3,960 watts
(13,512 Btu / hour)
60Hz 4,260 watts
(14,536 Btu / hour)
Chiller in another room :
1050 watts
(3,583 Btu / hour)
Heater Exchanger in the room with the Agilent 7500 :
50Hz 3050 watts
(10,407 Btu / hour)
60Hz 3050 watts
(10,407 Btu / hour)
Heat Exchanger in another room :
1050 watts
(3,583 Btu / hour)
Note :
The ambient temperature around the heat exchanger must
not exceed 30deg. C. for normal ICP-MS operation.
Cooling Water Requirements
Cooling capacity 1500 watts (5,118 Btu/hr) min
Flow >5 liters/min (1.32 US gallons)
Pressure 200 - 350kPa (29 - 50 psi)
Temperatures 15 - 40°C at the water inlet of the
ICP-MS.
Conductivity 50 ~ 150µS at the chiller reservoir
Softness Less than 0.100 grams/ liter
(6 grains/ US gallon)
Connections One hose, 10 meters (32.8 ft) long
ID=12mm (7/16 inch)
Male PT 3/8 inch
fittings on Agilent 7500 end. Male
1/2 inch fittings supplied with the
chiller or heat exchanger unit.
Maximum length of tubing can be 5 m without any
elevation
Issued: May 2007 – Rev. 2.0 PN# G3270-90231 TAD Product Support Page 3 of 6
Product# : G3272B, G3273B
Product Description: Agilent 7500cx, 7500cs Series ICP-MS
Site Preparation Specification
Gas Requirements Tick Boxes
Gas Argon
Purity UCE 99.99%
Pressure regulation 700kPa (100 psi) +3.5%
Flow 18 - 20 L/min
The gas regulators should be within 5 meters (16
feet) of the instrument.
Additional Gasses (Configuration Specific)
Gas Helium
Purity UCE 99.999%
Pressure regulation 50kPa (7.25 psi) +3.5%
Maximum flow 7 ml/min
Gas Hydrogen
Purity UCE 99.999%
Pressure regulation 50kPa (7.25 psi) +3.5%
Maximum flow 10 ml/min
Gas 20% Oxygen, 80% Argon
Purity UCE 99.999%
Pressure regulation 350kPa (50 psi) +3.5%
Maximum flow 1.0 L/min
Gas Xenon
Purity UCE 99.999%
Pressure regulation 50kPa (7.25 psi) +3.5%
Maximum flow 1.0 ml/min
Gas 10% Ammonia, 90% Helium
Purity UCE 99.999%
Pressure regulation 50kPa (7.25 psi) +3.5%
Maximum flow 10 ml/min
We suggest the upper limit of the regulators for
reaction gasses are 100 – 200kPa (14 – 28 psi)
Essential Requirements for Cell Gas Installation:
For optimum instrument performance, it is critical to
have stable high purity cell gas, with minimal dead
volume. The reaction cell gas supply and regulators
(H2 and He) must be located within 3 meters (10
feet) of the instrument. The following parts are
essential for installation.
Issued: May 2007 – Rev. 2.0 PN# G3270-90231 TAD Product Support Page 4 of 6
Product# : G3272B, G3273B
Product Description: Agilent 7500cx, 7500cs Series ICP-MS
Site Preparation Specification
Tick Boxes
Agilent Part Number Description Quantity
Gas Purification kit for ORS
• 8710-1709 (Tubing cutter)
• 5182-9706 (2 Position Base plate)
G3269A • 5182-9705 (Triple filter cartridge) x2 1
• G3270-65035 (1/8" Pre-cleaned stainless steel tubing,
6m, Fittings)
• 5182-0821 (Wall mount bracket)
• 5182-3423 (Spare O-ring, 8/pk, 4ea. of 2 sizes)
Description Quantity Note
Two stage regulator, must be fitted with stainless steel
diaphragms, 100 – 200kPa (14 – 28 psi) 1 These regulators must be ordered in
For use with: He the country of use. Gas cylinder –
Two stage regulator, must be fitted with stainless steel regulator connections vary throughout
diaphragms, 100 – 200kPa (14 – 28 psi) 1 the world and are not compatible.
For use with: H2
Communications
One(1) voice phone in close proximity to the
instrument so that an operator and support person
can talk while using the instrument.
Network Connection to Company LAN
(not needed for instrument operation)
If you need to connect the ICP-MS computer to your company
LAN, we only recommend connection via a switching HUB. A
small 10/100 6 port switch can be obtained from many sources at
low cost.
When connecting via a switching hub your IT department must
reserve two fixed (static) IP addresses that share the same subnet
mask and are dedicated to the ICPMS.
Caution
The Agilent ICP-MS system does not support the use of a
computer with two network interface cards. It has been observed
that the use of two NIC cards in an XP system frequently results in
crosstalk between the cards. This crosstalk can cause
communication failures.
Issued: May 2007 – Rev. 2.0 PN# G3270-90231 TAD Product Support Page 5 of 6
Product# : G3272B, G3273B
Product Description: Agilent 7500cx, 7500cs Series ICP-MS
Site Preparation Specification
Bench Requirements Tick Boxes
If the bench contains the Agilent 7500 and Agilent ASX500 Series
autosampler only, it should be capable of supporting 180Kgs
(396lbs) with an area of no less than 190cm x 63cm (6ft x 2ft).
Height should be approximately 75cm (2.5ft) for the standard
rough pump hoses.
Bench for computer and printer, requires an additional
45Kgs (100 lbs) of capacity and 120cm (4ft) of space.
Note when planning the bench layout, the communication cable
between Auto-sampler and computer is 2.0 meters long and the
communication cable between computer and instrument is 3.0
meters long, connection is located at the rear of the instrument on
the left hand side.
_____________________________________
Company Name
_____________________________________
Customer Signature/Date
_____________________________________
Engineer Signature/Date
______________________________________
Sales Order Number
Issued: May 2007 – Rev. 2.0 PN# G3270-90231 TAD Product Support Page 6 of 6