Service Report Service Report Perbaikan MRI Hitachi 0,4tesal New

The document is a service report from a medical equipment distributor. It details repairs made to an MRI machine for a customer, including inspecting and replacing a faulty brain coil component. Testing showed the coil was then functioning properly.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
57 views2 pages

Service Report Service Report Perbaikan MRI Hitachi 0,4tesal New

The document is a service report from a medical equipment distributor. It details repairs made to an MRI machine for a customer, including inspecting and replacing a faulty brain coil component. Testing showed the coil was then functioning properly.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

PT.

ALISHA JAYA MEDIKA


Medical and Hospital Equipment Distributor

Service Report
CustomerDetails
5067 – RSUD.dr.Moch.Ansari Saleh
Customer Name :
Jl.Brig Jend.Hasan Basri No,1 Alalak Utara Kec Banjarmasin
Equipment Location :
Utara.Kalimantan Selatan 70125
SONo :
Contact Name : Mr.Imam
Contact Phone : +628115003580
Initial Problem Description : Coil brain does not work issue
Contract Class : Not Service Contract
Report Type : Repaired separe part
Request No. : 23-111333

Contract End Day : --

EquipmentDetails
Initial Equipment Status : Ready
Product Description : MRI Aperto Lucent 0,4 Tesla
System ID : X421
Serial No. : 1519
Customer Eq# : --

Activity

Activity # : 2-20240324
Activity Type : Repaired acceccorise coil brain large
Reported Problem : Appear pop up error code 22038 “Singnal underflow”( Stict
Frequency Search )
Identified Problem : Acceccorise coil MRI
Solution : Check and inspect electricity from hospital OK
- General maintenance
- All check connector and port coil from table
- Test frequency coill from receiver unit not good
- Repaired and replacement component coil brain
- Reinsert connection all cable table to gantry
- Reinsert connection cable gantry to cabinet console
- General cleaning all unit
Test Procedure: : Turn On unit and test
- Creat new patien from pc control
- Connect coil brain to table port
- Insert penthom No.3
- Start scano Axial,Sagital and coronal OK
- Scano complete
- Coil brain functional OK
Test Passed: : Yes
Close Date: : 26 Mar 2024
Field Engineer: : Sukrisdianto
Office : Jalan Raden Inten II AG 14 No. 1 C Duren Sawit
Kec. Duren Sawit Jakarta Timur 13440
Telp. (021) 4800535 Page1of2
E-mail : [email protected]
Labor
DATE TYPE START TIME - END TIME HOURS
24 Mar 2024 Repaired 15:00:00 - 22:00:00 7.00
25 Mar 2024 Maintenance & 11:00:00 - :23:00:00 12.00
repaired
26 Mar 2024 Functional test on 11:00:00 – 13:00:00 2:00
patien

Materials
PCB Receiver

26-Mar-202410:00

26-Mar-2024 10:00 26-Mar-2024 10:00


CustomerSignature1

Office : Jalan Raden Inten II AG 14 No. 1 C Duren Sawit


Kec. Duren Sawit Jakarta Timur 13440
Telp. (021) 4800535
E-mail : [email protected]