Service Report Perbaikan X-Ray Stationary Rumah Sakit Kurnia Serang

The document is a service report for repairing an X-ray machine. The machine was stuck and not working properly due to unstable electricity supply and grounding issues. The engineer performed various checks, replaced fuses, and tested/calibrated the machine. It was then functioning properly for patient examinations. Recommendations included regular maintenance and calibration.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
36 views2 pages

Service Report Perbaikan X-Ray Stationary Rumah Sakit Kurnia Serang

The document is a service report for repairing an X-ray machine. The machine was stuck and not working properly due to unstable electricity supply and grounding issues. The engineer performed various checks, replaced fuses, and tested/calibrated the machine. It was then functioning properly for patient examinations. Recommendations included regular maintenance and calibration.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Service Request Type: Service Request Number:

Service Report Corrective Repair 2022-09-06/0-22

Customer Details

Customer Name: Rs.Kurnia Serang Location: Serang - Indonesia


PO #: Contact Phone: +6285772766091
Contact Nome: Mr.Radji Email Address:
Initial Problem Description:

Equipment Details
System ID: Initial Equipment Status: System Down
Serial Number: HR0104-0495 Entitlement: --
Product Description: X-ray Stationary
Universal 300mA

Activity Description

Activity #: Activity Close Dote:


Activity Type: Field Support Engineer Name: SUKRISDAINTO
Tube Slice Reading: EMEREAL_D 125(Varian)
Problem Description: Console stuck and can’t work properly

Couse/ Diagnosis: Sep- 06/2022


- No maintenance On schedule
- Unstable supply electricity from hospital
- grounding function is not good

Resolution/ Service Performed: Sep -06/2022


- Turn On unit Stuck
- Turn off unit and All Check
- Check and inspect Electricity supply from Hospital
- Check and inspect All fuse from cabinet unit OK
- Check and inspect power supply unit OK
- Reinsert power supply in cabinet generator OK
- Turn On unit OK
- Test All functional keypad from console Unit OK
- Test and Set Kv40,mAs 5,1 On preparation Indicator X-ray not ready
- Test and Set Kv60 and all mAs on preparation Indicator X-ray not ready
- Turn Of unit and all check connection OK
- Trun On unit
- Adjustment Kv,mAs and calibration OK
- Test and set Kv 40,50,60,70,80,90,100 and all mAs OK
- Calibration result Passed
- Function test for patient examination OK
Recommendation
Maintenance on schedule
Calibration on schedule
Install switch Breaker In cabinet unit for safety transformer and power supply
Replace collimator unit and readjustment collimation

Page No.1 of 1
Service Request Type:
Service Report Request Number:
Corrective Repair
20220908

Date Time Hours Labour Type


Sep 06, 2022 16:00-18:30 Zona travel
& Repaired

Ports Used Details


Port Number Description
0tY Asset Serial No

Expense Tracker
Expense Dote Type Comments
0tY

Engineer Signoture: Customer Signoture:


06-Sep-2022 16:00
SUKRISDIANTO

Page No.2 of 2