(기기분석) Ch.15 Molecular Luminescence Spectrometry

The document discusses principles of fluorescence and phosphorescence spectroscopy including theory of excited states, factors affecting emission, and applications. It covers concepts such as Jablonski diagrams, Stokes shift, quantum yield, and effects of temperature, pH, concentration and solvent on fluorescence intensity.

Uploaded by

kmscom1204
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
59 views55 pages

(기기분석) Ch.15 Molecular Luminescence Spectrometry

The document discusses principles of fluorescence and phosphorescence spectroscopy including theory of excited states, factors affecting emission, and applications. It covers concepts such as Jablonski diagrams, Stokes shift, quantum yield, and effects of temperature, pH, concentration and solvent on fluorescence intensity.

Uploaded by

kmscom1204
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 55

Principles of

Instrumental Analysis

Youngsoo Kim
Molecular Luminescence Spectrometry

 Theory of Fluorescence and Phosphorescence

 Instruments for measuring Fluorescence and Phosphorescence

 Applications of Photoluminescence Methods

 Chemiluminescence
Molecular Luminescence Spectrometry

광/발광(photoluminescence)
형광 : 전자 스핀 유지, 짧은 수명
인광 : 전자 스핀 변화, 긴 수명
들뜸 복사선보다 긴 파장

화학/발광(chemiluminescence)
화학 반응 중에 형성된 들뜬 화학종에 의한 복사선의 방출에 기반

발광 방법의 특징
무기 및 유기 화학종을 정량적으로 측정
고유한 감도 : 흡수 분광법에서 발생하는 것보다 낮은 검출 한계
흡수법에서보다 훨씬 큰 선형 농도 범위
Theory of Fluorescence and
Phosphorescence
Resonance fluorescence
: the absorbed radiation is reemitted without a change in
frequency

For Na,
: excitation 3s → 3p: absorption at 589.6 nm and 589.0 nm
: relaxation 3p → 3s: emission at 589.6 nm and 589.0 nm

Stoke shift
: molecular fluorescence bands center at wavelength
longer than the resonance line
→ shift toward longer wavelength
Excited states producing fluorescence and
phosphorescence
Electron spin

Pauli Exclusion Principle


- No two electrons in an atom can have the same set of four quantum number
- No more than two electrons occupy an orbitals and furthermore the two have opposed
spin state
- spin paring: diamagnetic
- free radicals, which contain unpaired electrons: paramagnetic
Excited states producing fluorescence and
phosphorescence
Singlet and triplet excited states

Singlet state
: A molecular electronic state in which all electron spins are pared
: No splitting of electronic energy levels occurs when the molecule is exposed to the magnetic field

Excited singlet state


: the spin of the promoted electrons is still paired with the ground-state electrons
: Diamagnetic

Excited triplet state


: the spins of the two electrons have become unpaired, thus parallel (paramagnetic)
: the excited triplet state is less energetic than the corresponding excited singlet state
Excited states producing fluorescence and
phosphorescence
Excited singlet vs. Excited triplet

Energy: excited singlet > excited triplet

Excitation probability: singlet ↔ singlet > singlet ↔ triplet

Absorption: singlet ↔ singlet > singlet ↔ triplet

Average lifetime: excited triplet (10-4s ~ several s) > excited singlet (10-8)

Phosphorescence emission: excited singlet → excited triplet → ground state


Energy-level diagrams for
photoluminescent molecules
Jablonski diagram
Energy-level diagrams for
photoluminescent molecules
Rates of absorption and emission
Rate of photon absorption: 10-14 ~ 10-15 s

Rate of emission
Fluorescence emission: 10-5 ~ 10-10 s, (singlet-singlet transition)
Phosphorescence emission : 10-4 ~ 10 s, (triplet-singlet transition)
Energy-level diagrams for
photoluminescent molecules
Deactivation processes
An excited molecule can return to its ground state by a combination of several
mechanistic steps (AKA, Relaxation or recombination process)

: Radiative emission process: fluorescence and phosphorescence


: non-radiative emission process: involves vibration relaxation, internal conversion,
external conversion

- most favored route to the ground state is the one that minimizes the lifetime of the
excited state
Energy-level diagrams for
photoluminescent molecules
1) Vibrational relaxation

- Collisions between molecules of the excited species and those of the solvent
→ rapid energy transfer
→ minuscule increase in temperature of solvent

- Lifetime of vibrationally excited molecules (10-12) < average lifetime of electronically


excited state (10-5 ~ 10-10 s)
→ fluorescence always involves a transition from the lowest vibrational level of an
excited electronic state
→ stoke shift
Energy-level diagrams for
photoluminescent molecules
2) Internal conversion
- 복사선의 방출 없이 분자를 낮은 에너지 전자 상태로 만드는 분자 간 과정

- 같은 다중도의 두 상태 사이의 교차(단일 - 단일, 삼중 - 삼중)

- 전자 에너지 준위가 비슷하여 진동 에너지 준위가 서로 겹치게 될 때 효율적

들뜸
λ1 : S0 → S1
λ2 : S0 → S2

방출 : λ3
S2 → S1 : λ3
S1 → S0 : λ3
형광 파장 : 들뜸 파장과 무관
Energy-level diagrams for
photoluminescent molecules
Quinine
들뜸 파장 : 250 nm, 350 nm
최대 방출 파장 : 450 nm
Energy-level diagrams for
photoluminescent molecules
2-1) Predissociation
Occurs when an electron moves from a higher electronic
state to an upper vibrational level of a lower electronic
state in which the vibrational energy is enough to cause
rupture of a bond

vs.

Dissociation
the absorbed radiation directly excites the electron of a chromophore to a sufficiently
high vibrational level to cause rupture of the chromophoric bond.
→ No internal conversion
Energy-level diagrams for
photoluminescent molecules
3) External conversion
- 들뜬 분자와 용매 또는 다른 용질 사이의 상호작용 및 에너지 전달을 포함하는 과정

- 증거 : 대부분의 화학종의 형광 강도에 대한 현저한 용매 효과


낮은 온도, 높은 점성 조건에서는 용매와의 충돌이 줄어들게 되어 형광의
세기가 증가한다.
4) Intersystem crossing
- 다중성이 다른 전자 상태 사이에서의 교차
- S1 → T1
- 두 상태의 진동 준위가 겹칠 때 쉽게 발생
- 중원자 효과 : 아이오딘, 브로민 등 무거운 원자가 있는 분자에서 주로 발생

5) Phosphorescence
- 계간 교차 → 내부 전환, 외부 전환, 인광
- 삼중항 → 단일항 전이 : 낮은 가능성, 긴 수명
- 복사선 조사가 끝난 후에도 지속
Variables affecting fluorescence and
phosphorescence
분자 구조와 화학적 환경
- 발광 여부
- 발광 세기

양자 수득률
- 양자 효율(quantum efficiency)
- 발광하는 분자의 수와 들뜬 분자의 총 수와의 비율, 𝜙𝜙

𝑘𝑘𝑓𝑓
𝜙𝜙 =
𝑘𝑘𝑓𝑓 + 𝑘𝑘𝑖𝑖 + 𝑘𝑘𝑒𝑒𝑒𝑒 + 𝑘𝑘𝑖𝑖𝑖𝑖 + 𝑘𝑘𝑝𝑝𝑝𝑝 + 𝑘𝑘𝑑𝑑

k : 비활성화 과정의 속도상수


f : 형광, i : 계간 교차, ec : 외부 전환, ic : 내부
전환, pd : 선해리, d : 해리
Variables affecting fluorescence and
phosphorescence
Transition types in fluorescence
250 nm 이하의 UV 흡수
- 높은 에너지(200 nm = 140 kcal/mol)
→ 선해리나 해리에 의해서 들뜬 상태의 비활성화

- σ*→ σ 전이에 의한 형광은 드물게 관찰

- 형광은 가장 일반적으로 첫 번째 들뜬 전자 상태의 최저 진동 준위로부터 전자


바닥 상태의 진동 준위 중 하나로 전이됨에 따라 발생

- 𝜋𝜋∗→ n 혹은 𝜋𝜋∗ → 𝜋𝜋 전이 중에서 에너지가 덜한 것에 의해 복사선이 생성


Variables affecting fluorescence and
phosphorescence
Quantum efficiency and transition type

양자 효율 : π*→ π > π*→ n

몰흡광 계수 : π → π* > n → π*

평균 수명 : π* → π (10-7~10-9 s) < π*→ n (10-5~10-7 s)

들뜬 상태는 짧은 평균 수명을 나타내며 형광과 경쟁하는 비활성화 과정이 덜


일어나기 때문에 형광은 더 일반적으로 π, π* 상태와 연관
Variables affecting fluorescence and
phosphorescence
Fluorescence and structure
Strong and useful fluorescence

- 낮은 에너지의 π → π* 전이를 하는 방향족 작용기를 가지고 있는 화합물

- 대부분의 방향족 탄화수소는 형광 방출

- 양자 효율은 고리 수와 축합의 정도에 따라 증가

몰흡광 계수 증가, 들뜬 상태의 수명 감소


→ 형광 발생
Variables affecting fluorescence and
phosphorescence
벤젠 고리 + 치환체
- 흡수 λmax의 이동 → 형광 파장 이동
- 할로젠 치환 : 형광의 세기 감소(중원자 효과)
Effect of structural rigidity in
fluorescence
단단한 구조를 가진 분자에서 높은 형광을 나타냄
- fluorene vs. biphenyl : CH2에 의해 구조의 경직성이 증가

양자
효율

>
8-
HQ
Lack of rigidity in a molecules ↓
→ enhanced internal conversion rate (kic)
→ consequent increase in the probability for non-radiative deactivation
Temperature and solvent effect

The quantum efficiency of fluorescence in most molecules:


- decreases with increasing temperature
→ increased frequency of collision at elevated temperatures improves the probability
for deactivation by external conversion

- decreases in solvent viscosity


→ increases probability of external conversion, resulting in decreasing in quantum
efficiency

The fluorescence of a molecules


- decreased by solvents containing heavy atoms or other solute with heavy
atoms in their structure (ex, carbon tetrabromide, ethyliodide, etc)
Effect of pH on fluorescence

The fluorescence of a aromatic compound with acidic or basic ring substituents


is usually pH dependent
- the emission intensity are likely to be different for the protonated and unprotonated
forms of the compound
→ differing number of resonance species that are associated with the acidic and basic
forms of the molecules

→ Additional resonance forms leads to a more stable first excited state, resulting in
consequence of fluorescence in the UV region
Effect of pH on fluorescence

The fluorescence of a aromatic compound with acidic or basic ring substituents


is usually pH dependent
- the fluorescence of certain compounds as a function of pH have been used for the
detection of end-point in acid-base titration
ex) the fluorescence of 1-naphtol-4-sulfonic acid is found in the UV region
→ phenolate by the addition of base, emission band shift to the visible wavelength

In the presence of dissolved oxygen reduces the intensity of fluorescence in


solution
- photochemically induced oxidation of the fluorescing species
Effect of concentration on fluorescence
intensity
The power of fluorescence emission F
𝐹𝐹 = 𝜙𝜙𝑓𝑓 𝐾𝐾 ′′ 𝑃𝑃0 − 𝑃𝑃 = 𝐾𝐾 ′ 𝑃𝑃0 − 𝑃𝑃
P0: power of incident beam, P is the power after traversing a length b,
ϕf is quantum efficiency, K'' is constant

F to the concentration of c of fluorescing species, rewrite Beer’s law in the form


𝑃𝑃
= 10−𝜀𝜀𝑏𝑏𝑏𝑏
𝑃𝑃0
𝐹𝐹 = 𝐾𝐾 ′ 𝑃𝑃0 1 − 10−𝜀𝜀𝑏𝑏𝑏𝑏
2 3
2.303𝜀𝜀𝑏𝑏𝑏𝑏 2.303𝜀𝜀𝑏𝑏𝑏𝑏
𝐹𝐹 = 𝐾𝐾 ′ 𝑃𝑃0 2.303𝜀𝜀𝑏𝑏𝑏𝑏 − + +⋯
2! 3!

Under lower concentrated conditions, 2.303εbc = A < 0.05


𝐹𝐹 = 2.303𝐾𝐾 ′ 𝜀𝜀𝑏𝑏𝑏𝑏𝑃𝑃0
𝐹𝐹 = 𝐾𝐾𝐾𝐾 = 2.303𝐾𝐾 ′ 𝜀𝜀𝑏𝑏𝑏𝑏𝑃𝑃0 = 2.303𝜙𝜙𝑓𝑓 𝐾𝐾 ′′ 𝜀𝜀𝑏𝑏𝑏𝑏𝑃𝑃0 (at constant 𝑃𝑃0 )

Thus, the F vs. c should be linear at low concentrations


Quenching of Fluorescence

Quenching: usually refers to nonradiative energy transfer from excited species to other
molecules

Dynamic quenching (collision quenching)


: requires contact between the excited species and the quenching agent (Q)
: occurs as rapidly as the collision partner can diffuse together
: the rate depends strongly on temperature and viscosity
: the quencher concentration must be high enough that there is a high probability of a
collision between the excited species and the quencher during the lifetime of the
excited state
𝑘𝑘𝑒𝑒𝑒𝑒 = 𝑘𝑘𝑞𝑞 𝑄𝑄
𝑘𝑘𝑓𝑓 𝑘𝑘𝑓𝑓
𝜙𝜙𝑓𝑓0 = (1) 𝜙𝜙𝑓𝑓0 = (2)
𝑘𝑘𝑓𝑓 + 𝑘𝑘𝑖𝑖 + 𝑘𝑘𝑖𝑖𝑖𝑖 𝑘𝑘𝑓𝑓 + 𝑘𝑘𝑖𝑖 + 𝑘𝑘𝑖𝑖𝑖𝑖 + 𝑘𝑘𝑞𝑞 𝑄𝑄

𝜙𝜙𝑓𝑓0 𝐹𝐹0
= 1 + 𝐾𝐾𝑞𝑞 [𝑄𝑄] = 1 + 𝐾𝐾𝑞𝑞 [𝑄𝑄] Stern-Volmer equation
𝜙𝜙𝑓𝑓 𝐹𝐹
𝐾𝐾𝑞𝑞 = 𝑘𝑘𝑞𝑞 ⁄ 𝑘𝑘𝑓𝑓 + 𝑘𝑘𝑖𝑖 + 𝑘𝑘𝑖𝑖𝑖𝑖

Dynamic quenching reduces both the fluorescence quantum yield and fluorescence lifetime
Quenching of Fluorescence

Static quenching
: the quencher and the ground state fluorophore form a complex (dark complex)
: the decrease in fluorescence intensity can also be described by the Stern-Volmer eq.
Kq is now the formation of constant of dark complex
: the lifetime of fluorescence is not affected by the static quenching

Förster quenching
: energy transfer occurs without collision between molecules
: dipole-dipole coupling (interaction) between the excited fluorophore and quencher
account for the transfer
: dependence of the degree of quenching on the quencher concentration is complicated
not follow the Stern-Volmer behavior
Emission and Excitation spectra

들뜸 스펙트럼
- 들뜸 파장이 변화되는 동안 고정 파장에서 발광 세기 측정
→ 세기 vs. 파장
- 흡수 스펙트럼과 동일

전체 발광 스펙트럼 (방출 스펙트럼)


- 들뜸 파장 고정 → 각 파장에서 방출 세기 측정
- 파장 : 들뜸 < 형광 < 인광
- 형광 파장과 인광 파장의 차이 = 삼중항과 단일항의 에너지 차이와 유사
Emission and Excitation spectra

Tryptophan
Instruments for measuring Fluorescence
and Phosphorescence
Instruments for measuring Fluorescence
and Phosphorescence
기기의 구성 요소

광발광 측정 기기의 구성 요소
- 자외선-가시선 광도계 또는 분광 광도계와 유사
- 겹살 광학을 사용

들뜸 파장 선택기
- 들뜸 파장만 통과
- 형광 방출 파장은 제거

광원-시료-검출기
- 직각, 산란과 광원의 영향 최소화

방출 파장 선택기 : 형광 파장을 선택
Instruments for measuring Fluorescence
and Phosphorescence
형광계와 분광계의 구성
[광원]
- 감도는 광원의 복사선 세기 P0에 정비례한다 F ∝ Po
- 흡수 측정보다 세기가 더 큰 광원 사용

램프 청색 발광 다이오드(LED)
- 저압 수은 증기 램프 - 450~475 nm 파장 방출
- 형광을 들뜨게 하는 유용한 선들을 생성 - 특정 형광체의 들뜸에 사용
- 흡수 또는 간섭 필터를 사용해서 파장
선택

고압 제논 아크 램프 레이저
- 300 ∼1300 nm의 연속 복사선 광원 - 질소 기체 또는 Nd:YAG 레이저
- 크로마토그래피, 전기이동의
검출기에 사용
Instruments for measuring Fluorescence
and Phosphorescence
[필터와 단색화 장치]
• 형광계 : 필터
• 분광형광계 : 회절발 단색화 장치

[변환기]
• 광전 증배관
• 전하-이동 장치

[시료 용기와 용기칸]


• 유리 또는 실리카로 만든 원통형 및 직사각형 시료 용기
• 저용량 마이크로 시료 용기
Instruments for measuring Fluorescence
and Phosphorescence
기기 설계

[형광계]
• 수은 램프 + 필터 + 광전증배관
Instruments for measuring Fluorescence
and Phosphorescence
[분광형광계]
• 회절발 단색화 장치 + 광전증배관
Instruments for measuring Fluorescence
and Phosphorescence
[배열 검출기 기반 분광형광계]
Instruments for measuring Fluorescence
and Phosphorescence
[광섬유 형광센서]
• 광섬유 탐침
- 광원 및 검출기에서 멀리 떨어져 있는 위치에서 형광 측정
- 레이저 광원의 복사선이 광섬유를 통과하여 시료 용액에서 형광
- 형광 방출은 동일한 광섬유를 통해 다시 측정을 위해 검출기로 이동
- 형광 지시약 물질을 광섬유의 끝부분에 고정하여 비형광 분석 물질 측정
- 형광의 소광을 기반

Spectrometer
Power
cable
Laser

Optical fiber

Power supply

Sample holder
Instruments for measuring Fluorescence
and Phosphorescence
[인광계]
• 시료를 번갈아 조사하고 적절한 시간 지연 후에 인광 강도를 측정하는 장치
• 경질 매질 내에서 액체 질소 온도에서 수행
Instruments for measuring Fluorescence
and Phosphorescence
보정 및 보상 기법

광원의 보상
- 기준 광전증배관을 통해 광원의 세기를 모니터
- 시료의 발광 신호 대 기준 검출기로부터의 신호의 비율
- 광원 세기의 변동 및 들뜸을 보상

보정된 들뜸 스펙트럼
- 양자 계수기 사용
- 전산화된 분광광도계

보정된 방출스펙트럼
- 교정된 광원을 사용
- 방출 단색화장치 및 변환기에 대한 교정
계수 결정
- 적절한 보정 계수를 곱해서 스펙트럼 보정
Instruments for measuring Fluorescence
and Phosphorescence
기기의 표준화

- 안정한 형광체 표준 용액으로 수행


- 일반적인 시약은 약 10-5 M 농도의 quinine sulfate 표준 용액
- 350 nm에서 복사선에 의해 들뜨며 450 nm의 복사선을 방출
Instruments for measuring Fluorescence
and Phosphorescence
Applications of Photoluminescence
Methods
형광법과 인광법
- 가장 감도가 좋은 분석 기술 중 하나
- 향상된 감도는 광원의 세기에 직접 비례하는 농도와 관련
- 광발광 세기는 P0에 독립적으로 측정
- 흡광도 측정은 P0와 P 모두를 평가
- 흡수 방법보다 더 넓은 선형 동적 범위
- 분광광도법보다 낮은 정밀도와 정확도
Applications of Photoluminescence
Methods
무기물의 형광법 정량

직접법
- 형광성 킬레이트의 형성 후 방출을 측정

간접법
- 측정되는 물질의 소광 (quenching) 작용으로 인한 형광 방출의 감소를 측정
- 음이온 검출에 사용
Applications of Photoluminescence
Methods
형광 킬레이트를 형성하는 양이온
형광 킬레이트를 형성하는 전이 금속 이온의 수를 제한하
는 요인 (형광 X)
- 이온의 대부분은 삼중항 상태로의 계간 교차 속도를
증가시키는 상자기성
- 밀집된 에너지 준위 : 내부 전환에 의한 비활성화 증가

전이금속이 아닌 양이온 (형광 O)


- 내부 전환에 의한 비활성/인광 가능성 낮음
- 무색의 형광 킬레이트 생성

[형광용 시약]
- 금속 이온과 킬레이트 형성이 가능한 2개
이상의 주개 기능기가 있는 방향족 구조
Applications of Photoluminescence
Methods
Applications of Photoluminescence
Methods
유기 및 생화학 화학종에 대한 방법
식품, 의약품, 임상 시료 및 천연물의 분석
- adrenaline, morphine, penicillin, phenobarbital, procaine, reserpine, lysergic acid
diethylamide (LSD) 등

인광법
- 인광과 형광은 상호 보완적
- 무거운 원자를 포함하는 화합물은 강한 인광
- 중원자가 없는 동일한 화합물은 인광보다 형광을
나타내는 경향
- 낮은 온도가 필요
- 일반적인 낮은 정밀도
Applications of Photoluminescence
Methods
수명측정
형광 수명 측정
- 충돌 비활성화 과정, 에너지 전달 속도 및 들뜬 상
태 반응에 대한 정보 제공
- 발광 화학종을 함유한 혼합물의 분석에서 추가적
인 선택성 제공
- 근래에는 형광 시간 척도(10-5 ∼10-9 초)에서 발광
붕괴 속도를 일상적으로 측정

시간 영역 측정
- 펄스 광원을 사용하고 형광의 시간 종속적인 붕괴
를 측정

주파수 영역 방법
- Sine 곡선 변조 광원을 사용하여 시료를 들뜨게 하
고 , 들뜸 파형에 대한 형광 방출의 위상 이동은 수명
에 대한 정보를 제공
Applications of Photoluminescence
Methods
형광영상법

형광 분광기와 광학 현미경의 결합
- 단일 세포 내 형광 물질 영상 생성
- 생체 분자의 고유 형광 이용
- 형광 지시계 이용

세포 내 Ca2+ 관찰
- 소뇌 중 Purkinje 뉴런 + Fura-2 지시약
- 형광의 예리한 변화 관찰
- 칼슘 의존성 전기적 사건이 일어나는
장소와 시간 결정 가능
Chemiluminescence

화학 발광 현상
- 화학반응에 의해 전자적으로 들뜬 화학종이 바닥 상태로 되돌아가면서 빛을
방출할 때 생성

A + B → C* + D
C* → C + hv (C*는 화학종 C의 들뜬 상태)

생체 발광
- 생물학적 시스템에서 일어나는 화학 발광
- 반딧불, 바다 팬지, 특정 해파리, 박테리아, 원생 동물 및 갑각류 등
Chemiluminescence
Chemiluminescence

복사선의 세기 ICL

- 초당 방출되는 광자 수
- 화학 반응의 속도(d[C]/dt)와 화학 발광 양자 수율 𝜙𝜙CL(반응한 분자당 광자 수)
에 의존

𝑑𝑑 𝐶𝐶 𝑑𝑑 𝐶𝐶
𝐼𝐼𝐶𝐶𝐶𝐶 = 𝜙𝜙𝐶𝐶𝐶𝐶 = 𝜙𝜙𝐸𝐸𝐸𝐸 𝜙𝜙𝐸𝐸𝐸𝐸
𝑑𝑑𝑑𝑑 𝑑𝑑𝑑𝑑

- 일반적으로 𝜙𝜙𝐶𝐶𝐶𝐶 은 0.01 ~ 0.2


Chemiluminescence

화학 발광의 측정
기기
- 반응 용기 + 광전증배관
- 광원 : 분석물질과 시약 간의 화학 반응
- 파장 선택기 불필요

정량 분석
- 일정 시간동안 신호를 적분 → 표준물의 신호와 비교
- 봉우리 높이의 비교
Chemiluminescence

[기체 분석]
- 오존, 질소 산화물, 유황 화합물 분석
- NO 측정
NO + O3 → NO2* + O2
NO2* → NO2 + hv (λ : 600 ~ 2800 nm)
- 질소 산화물 측정 : 열분해 반응 이용
NO2 ⇌ NO + O

- 대기 오존 모니터링
: 분석 물질이 활성화된 실리카 젤 표면에 흡착된 염료인 Rhodamine B와 반응
할 때 생성되는 발광을 기반

- 대기 황 화합물의 측정
: 시료는 수소 불꽃으로 연소되어 황 이합체를 생성하며, 이 이합체는 분해되어
빛을 방출
4H2 + 2SO2 = S2* + 4H2O
S2* → S2 + hv (λ : 384, 394 nm)
Chemiluminescence

[액상에서 무기 화학종의 분석]


- 루미놀 + 강염기
- O2, H2O2, 하이포아염소산 등의 산화제와 반응
- 촉매 필요

+ hv
Chemiluminescence

[유기 화학종의 분석]

- 효소 반응의 생성물인 H2O2 분석


uricase
- Uric acid + O2 allantoin + H2O2