The C-5000 Source Sampling System
The C-5000 Source Sampling System
SYSTEM
OPERATING MANUAL
Version 2
January 2008
ENVIRONMENTAL SUPPLY COMPANY
708 E. Club Blvd. Durham, North Carolina 27704 800-STACKS-5 (782-2575) (919) 956-9688 FAX: (919) 682-0333
TABLE OF CONTENTS
1 INTRODUCTION ............................................................................................................... 5
1.1 Operational Overview of the System ..................................................................... 5
2 MAINTENANCE................................................................................................................. 6
2.1 OPERATIONAL OVERVIEW OF THE SYSTEM ..................................................... 6
2.1.1 Maintenance Access ........................................................................................... 6
2.1.2 Pump ................................................................................................................... 6
2.1.3 Thermometer....................................................................................................... 8
2.1.4 Quick Connects................................................................................................... 8
2.1.5 Dual Inclined Manometer..................................................................................... 8
2.1.6 Pitot Tube Lines .................................................................................................. 8
2.1.7 Null Switch .......................................................................................................... 8
2.1.8 Maintenance Access ........................................................................................... 9
2.1.9 Maintenance Access ........................................................................................... 9
2.1.10 Vacuum System .................................................................................................. 9
2.2 SAMPLE CASE .......................................................................................................10
2.2.1 Heated Filter Compartment ...............................................................................10
2.2.2 Impinger (Cold) Box ..........................................................................................10
2.2.3 Impinger Glassware ..........................................................................................11
2.2.4 Filter Assembly Pressure Drop Check...............................................................11
2.3 PROBE ASSEMBLY ...............................................................................................11
2.3.1 The Probe Sheath and Probe Liner ...................................................................11
2.1.9 Probe Test .........................................................................................................12
2.3.3 Probe Cleaning..................................................................................................12
2.3.4 S-Type Pitot Tube Maintenance ........................................................................12
2.3.4 Nozzle................................................................................................................12
2.4 UMBILICAL CORD and adapter ............................................................................12
2.4.1 Quick Connects.................................................................................................13
2.4.2 Umbilical Adapter Thermocouple Check ...........................................................13
2.4.3 Umbilical Adapter Check Valve .........................................................................13
2.4.4 Sample and Pitot Lines .....................................................................................13
1. INTRODUCTION
The Environmental Supply Company C-5000 Source Sampling System is engineered for
optimal performance of isokinetic source sampling procedures. The C-5000 Source
Sampler performs under severe operating conditions encountered by stack sampling
professionals to satisfy the requirements as specified by the Environmental Protection
Agency in the Office of Air Programs Publication No. APTD-0576 and for legal and scientific
documentation of source particulate sampling methods. The system employs an array of
features designed in response to the suggestions and requests of stack sampling
professionals for accuracy, reliability and ease of use in operation, maintenance and
calibration.
2. MAINTENANCE
2.1 CONTROL CONSOLE
Periodic inspection of the C-5000 Source Sampler console components is essential to
maintain proper operation of the system.
Regular maintenance will insure a long system
lifetime.
2.1.2 Pump
A Gast fiber vane vacuum pump draws the stack gases through the sample train. This
rotary pump is a precision instrument with .002 clearance at the top and .0025"--.0035" total
at the ends of the rotor. The vanes self-adjust with wear and should last 5,000 --15,000
hours depending upon operating conditions. It is designed for pumping dry air and should
be protected against dirt and excessive moisture. With proper treatment and regular
lubrication this pump requires little additional maintenance to insure a long working life.
Regular Pump Maintenance
There is a small hole in the top of the rear panel for regular inspection of the knockout jar.
This jar should be emptied when accumulated oil interferes with the normal flow of air to the
dry gas meter.
Frequent removal and cleaning of the oiler jar is required when the oil level is low or the oil is
dirty. An inspection port and removable panel are located on the right rear side of the unit.
Unscrew the oiler jar and fill with clean oil as recommended in the table below. Use the
correct amount and type of oil; inadequate lubrication is far more damaging than excessive
lubrication. For installations below freezing, dilute oil with up to 25% kerosene. Do Not
Reuse Oil.
setting the clearance so that all points on the circumference of rotor will clear. Total end
clearance for both sides of rotor may vary from .0025" to .0035".
Overheating: Dirty or clogged filter felts can be responsible for inability of the pump to build
up pressure or produce vacuum, and can lead to overheating. It is not unusual for exhaust
air temperature to reach 230°F when running continuously. If this temperature is exceeded
or is reached during intermittent operation, remove the filter felts and wash in solvent. If
overheating is not repaired by this procedure, stop operation immediately and send the unit
to the manufacturer for repairs. When corresponding or ordering spare parts, please give
complete model and/or serial number.
2.1.3 Thermometer
Check the dry-gas meter thermometer against room temperature using a standard mercury-
in-glass thermometer.
Keep quick connects clean by wiping with a clean rag and lubricating with a drop of
penetrating oil.
Make sure by a visual check that the Pitot and orifice manometer lines are free of fluid and
void of leaks, especially around the fluid-level zeroing controls and drain screws. Clean the
front of the manometer with a clean rag and the back with compressed air (the device can
be removed from the plumbing panel and wiped clean). Follow recommendations on the
manometer instruction plate for further cleaning directions.
Open the manometer ports 1.5 turns counterclockwise from closed position. Make sure that
the manometer lines are connected and the manometer is level. Check the fluid level and, if
necessary, add fluid by removing the screw on the left side. The manometer is zeroed when
the oil meniscus and the reflected image at zero are aligned, and the fluid-level plunger
(zeroing control) can travel between 1/4” and 1/2” inward.
Note: When shipping, disconnect the manometer lines and close the manometer ports by
turning clockwise. If the manometer unit ever becomes inverted (as may happen during
shipping), be sure the floating check valves, located under the manometer ports, return to
their normal position before use.
Check the Pitot tube line quick connects for obstructions by blowing gently, the Pitot tube
gauge should respond.
With all electrical switches turned off, plug the power cord into a 115-volt outlet. Turn the
power switch to the “ON” position. Power up is indicated when the digital temperature
controllers show a readout or, if analog temperature controllers are used on your system,
the temperature readout displays. Turn on the “NULL” switch; the solenoids should make an
audible “click”. If the clicking sound does not occur, check the power source, power cable,
power connections, and solenoid valve.
The “NULL” switch on the front panel of the console operates the solenoid valve assembly
for obtaining a zero reading during a test without interrupting sample collection. The normal
position is “OFF” while reading ∆H. In the “on” position the orifice lines are blocked to the
manometer.
Turn off all switches except the power and pump switches. Carefully adjust the Main-
(Coarse) and Fine-Adjust valves. The manometer should respond to valve controls. If no
movement is observed, check the solenoid valves or the orifice gauge system.
2.1.8 Timer
The timer is located next to the pump switch on the console electrical panel so that the
operator can easily turn on the timer along with the pump to measure elapsed time.
There are two types of solid-state temperature controllers available for probe and filter heat
regulation. Your system will have either two analog (dial type) controllers or two
programmable digital controllers. Each style of temperature controller employs an indicator
light to show when the controller is powering its 10 Amp solid-state relay, indicating heating
to setpoint. When setpoint is reached, the controller will cycle off and repeat the on-off cycle
in response to thermocouple feedback.
The analog controllers are operated by turning the knob to the desired setpoint. Setting or
resetting the temperature for the programmable, digital controller is outlined below. For
more details on the operation of the digital controller please refer to the manual included
with your console. NOTE: If your console is equipped with digital controllers, the switchable
temperature readouts for the probe and filter are disabled, so that the temperatures can be
read directly from the readout on the digital controller.
RESETTING THE TEMPERATURE CONTROLLER (FUJI)
1. Press “SV” gently to observe the temperature setpoint.
2. Press the “^” key of the digit to be set: 100 digit (left), 10 digit (middle), unit digit
(right). Repeatedly press each key until the desired digit is obtained.
3. When the desired temperature setpoint is keyed in properly, press the “ENT”
key to accept the setpoint. The flashing will stop.
4. Press “PV” to observe the current temperature.
EXAMPLE: Set Temperature to 250°°F
1. Press the 10 digit “^” key (middle) five times. Press the 100 digit “^” key (left) two
times.
2. Press ”ENT” to accept the 250°F setpoint. Press “SV” to check the setpoint
value. Press “PV” for the current temperature.
Insert a plugged 1/2" male quick connect into the sample inlet port on the front of the
console. Turn the Pump switch “ON”, the Coarse-Adjust valve “ON” and close the Fine-
Adjust valve. The vacuum gauge should read about 25 inches of mercury when ambient
barometric pressure is 30 inches of mercury. Check the leakage rate using the reading on
the dry gas meter and a watch.
Regulations require a leak rate in the train less than .02 cfm, however the pump and
console should not be taken to the field with a leak rate greater than .01 cfm. If this rate is
exceeded: 1) close the coarse valve and check again; 2) check the oiler and inlet filter jars
for tightness, broken threads, or bad gaskets; and 3) check fittings. Continued leaking is
indication of the need for repair or replacement.
Fill the impinger box with water and check for leaks. Leaks may be repaired with fiberglass
and epoxy resin or filled with silicone rubber.
Connect the filter holder assembly directly to the umbilical adapter and run the umbilical
cable to the console for a pressure-drop check. Turn the pump on and adjust flow on the
orifice gauge to a reading of 5.0 inches Hg. If the vacuum gauge on the console reads
higher than two inches of mercury, the frit needs cleaning and rechecking.
The probe nozzle is sealed with a front and rear ferrule and a union nut to the union body
that fastens to a permanently welded union nut on the sample entry end of the probe
sheath. The sample entry end of the probe liner inserts into this union body and is secured
with an inverted rear ferrule and Viton O-ring. The end of the probe liner should be
positioned flush with the front end of the welded union nut when inserting in the union body.
Detach the union assembly from the probe sheath welded nut before attempting to
remove the probe liner. Finger tightening is sufficient to reattach the union body assembly
(with inverted ferrule and Viton O-ring) to the welded nut. Do not use a wrench for this
procedure. However the probe nozzle-to-union body attachment requires two wrenches for
proper tightening. We suggest using mineral fiber string as an alternative to the Viton O-ring
when stack temperatures over 500°F are expected.
The liner must be removed from the probe sheath to inspect the heating mantle or replace
the liner. Remove the nozzle and union from the welded nut and gently slide the liner out of
the sheath. To reassemble, slide the probe liner inside the probe sheath and guide the end
of the liner through the hole of the welded union nut until it extends flush to the end of the
nut.
Examine the electrical components of the probe for any visible damage prior to following
test procedures.
1. Plug the probe into the sample case and set the probe heat controller to 250°. The
probe should become warm to the touch over its entire length in a few minutes. If the
probe does not heat, check the probe heat control for loose connections, the umbilical
cord and the sample case.
2. Use an ohmmeter to measure the resistance between leads and also to ground.
Deviations from proper value indicate faulty wiring.
Clean the probe liner internally by brushing, first using tap water, then distilled, deionized
water, and finally using acetone. Perform a final rinse with acetone and allow to air dry.
Inspect visually for cleanliness and repeat the procedure if necessary.
Clean the stainless steel union, nut ferrules, Viton O-ring, and nozzle by scrubbing. Rinse
with distilled, deionized water, and then with acetone and allow to air dry with all open ends
of probes and nozzles covered with serum caps or saran wrap to prevent contamination.
Remove the quick connects from the Pitot tube and wipe them clean. A drop of penetrating
oil helps keep them in good working condition.
Use compressed air to blow out the Pitot tube. Both the Pitot tube and sheath should be
wiped or washed clean with water and air-dried. Inspect the Pitot tube openings for damage
and misalignment, and, if necessary, repair and recalibrate the tube.
2.3.5 Nozzle
Inspect the knife-edge of the nozzle for damage. If damaged, repair and recalibrate the
nozzle (see recalibration section). Clean the nozzle by scrubbing, then rinse with distilled,
deionized water, and finally with acetone. Allow to air dry with the open ends covered with
serum caps to avoid contamination or damage to the knife-edge.
Ideally there should be a selection of nozzles ranging from 1/8” to 1/2” in diameter in 1/16”
increments with two or three sets of each size. A number reflecting the exact diameter in
1/32” increments should be etched near the inlet end of each of the nozzles (e.g. #8
indicates a 1/4” nozzle, #9 indicates a 9/32” nozzle).
The umbilical cord consists of one 1/2” sample (vacuum) line, two 3/8” Pitot lines, one 1/4”
orsat line, one power line with Amphenol connectors, and five thermocouple lines. Lines of
like dimension and purpose are numbered at each end to avoid cross connections and
assist in proper hookup of the umbilical couplings.
Wipe the vacuum line and Pitot tube quick connects and apply a drop of penetrating oil on
each to keep them in good working condition.
Check the impinger temperature thermocouple (“DRYER” connections and settings on the
console) on the umbilical adapter against a mercury-in-glass thermometer.
Disconnect the joint at the check valve and add a drop of lubricating oil on the valve.
Test the sample line for leaks by placing a plugged male quick connect into the inlet,
connecting the line to the console and checking for leakage as before. If there is leakage, all
connections should first be checked, and then the tubing. If the leak is not evident by
inspection, slightly pressurize the line and check for leaks using soapy water.
Connect one end of the Pitot tube line to a 36-inch U-tube mercury manometer or a similar
device. Pull a vacuum of at least ten inches, seal the tubing at the pump end, and check for
leaks by noting the mercury manometer. Do the same for the other side. If there is leakage,
check the tubing by following the same procedure as for the vacuum line.
Pitot tube acceptability is in the range of Cp = 0.85 ± 0.03 in both directions, since the
nomograph used with the sampling train is designed for this number. If the Cp is not in this
range, the Pitot tube was probably not calibrated properly. Be sure that the Pitot tube is
square with the duct and pointing directly into the gas stream. A special nomograph is
available to correct for variations in Cp. (catalog # M5-SRN)
It is important to take your sample data from a site that fits the following criteria. Existing
sample sites are often chosen based on inadequate information. If necessary, relocate the
sampling site to acquire acceptable data.
A sampling site should be chosen with a minimum of at least two diameters downstream
and 0.5 diameters upstream from any expansion, contraction, bend, obstruction, visible
flame, or exit, and requiring 48 sampling points within the duct. The optimum sampling
location is at least eight diameters downstream and two upstream, requiring only 12
sampling points. For rectangular cross sections the equivalent diameter is determined by
using the following equation:
equivalent diameter =2
(length)(width )
length + width
The chart below is provided to assist in determining the number of sampling points required
for a sampling site. First measure the distance from the chosen sampling location to the
nearest upstream and downstream disturbances. Using the upstream and downstream
scales, two numbers will be indicated. Select the higher of the two numbers. Then round off
the number for a round stack to the nearest multiple of four. For a square or rectangle stack,
divide the cross section into as many equal rectangular areas as sample points, such that
the ratio of the length to width of each elemental area is between one and two.
0
2 3 4 5 6 7 8 9 10
Number of Duct Diameters Downstream*
(Distance B)
For circular stacks, locate the sampling points on at least two diameters according to the
diagram below. Make sure that the traverse axes divide the stack into equal parts. For
rectangular stacks, locate the points at the centroid of each equal area as shown below.
Another factor to be considered in selecting the sampling site is that the distance from the
probe to the bottom of the sample case is about 13 inches. Although the equipment was
designed to fit into 2 1/2-inch holes, it has been found that 3-inch holes allow easier
entrance and removal without nicking the nozzle or picking up deposited dust.
The ESC Sample Case is designed for use with most any monorail system. The ESC
Suspension Rail System is both lightweight and flexible and permits rapid attachment of the
sample case to the suspension rail trolley by use of snap hooks. After hanging the sample
case on the suspension rail it may be freely rotated so that the probe can be attached by
the operator while standing on a platform very close to the stack. The sample case is then
pushed out to its farthest position by holding the probe prior to insertion into the stack port.
A slide box is probably the simplest support system of all. However it can only be used
conveniently when sampling very near the rooftop. A typical arrangement would be a
plywood table supported by any suitable weight bearing legs (cinder blocks or laterally
braced 2x4 assemblies) that will elevate the sample case and probe to be aligned with the
sample port. Many variations are possible depending on the operator and materials
available.
Once the sampling site has been prepared to accommodate all necessary apparatus, the
sample train can be assembled. The sample case should be properly fitted with the
impingers and glass connectors. Fill two of the impingers with 100 ml of distilled water and
one with 175 grams of silica gel. Place tarred filter in the filter holder and record the filter
weight on the data sheet. Environmental Supply Company supplies type "A" fiberglass
filters packaged for this purpose.
Tighten the phenolic screw joint around the filter holder to prevent leakage around the frit
gasket and position the assembly in the heated compartment of the sample case. Position
the umbilical adapter onto the last impinger arm while sliding the umbilical adapter bracket
into the support block mounted on the side of the impinger box. Tighten the thumbscrew to
hold the adapter firmly in place. The sample case can now be moved to the sample site.
After selecting a suitable probe length and nozzle size, insert the probe liner carefully into
the sheath and connect the nozzle using the union connections. Cover the nozzle and
probe liner with serum caps or saran wrap. Mark the probe with heat-resistant tape or by
some other method to denote the proper distance into the stack or duct for each sampling
point. Move the probe to the sampling site.
and align with glassware inlet (cyclone bypass). Tighten the 1/4 - 20 wing nut to secure the
probe to the support bracket. For ball joint probe and glassware, clamp the probe liner ball
to the socket on the inlet glassware. For screw joint probe liner and glassware tighten the
screw cap with ring seal to the inlet glassware. Fit the filter inlet arm to the cyclone bypass
and secure.
For Vertical Up or Down Traverses--For these applications it is recommended that a
vertical probe support adapter (ESC catalog # M5-VPS) be used to hold the probe in proper
position. The assembly is the same as above except that an additional elbow must be
secured between the probe liner exit end and the glassware inlet. The ESC Compact Filter
Box (catalog # M5-CFB) may be used as an alternative to the vertical probe support
adapter. This lightweight heated filter box is designed to operate in any position and/or in
tightly constrained areas.
The sample run plan should consider the number of traverse points and the sampling time
at each point. A good rule of thumb to follow is to collect a weight of particulate matter equal
to about 20 percent of the filter weight. The upper limit is usually set by the pressure drop
across the filter and the amount of condensate the impingers can hold. Generally the length
of sampling time at each traverse point is 2 to 10 minutes.
During the sample traverse, the probe is moved from point to point without turning the pump
off except when changing to a different sampling port. The ∆P should be monitored and
adjustments made on the orifice gauge with the aid of the stack calculation slide rule.
Besides the regular time interval recordings, a set of readings should be recorded when the
∆P changes by more than 20 percent.
3.13 SHUTDOWN
When testing has been completed, close the sample valve, remove the probe from the
stack, and take a final set of readings. It is possible to turn off the pump rather than close
the sample valve. However, using this procedure may occasionally cause the contents of
the first impinger to back up into the filter unless a check valve is installed in the umbilical
adapter (such as ESC Model M5-UA). Turn off the Filter Heat and Probe Heat Controls and
remove the probe from the sampling port. Cover the nozzle tip as soon as possible to avoid
contamination or loss of sample. Disconnect the probe from the sample case support
bracket and cyclone bypass inlet glassware. Cover both the end of the probe and the
glassware inlet (cyclone bypass). Disconnect the umbilical cord from the sample case. The
probe and sample case are then ready for clean up and analysis.
(Cleanup kit, Model M5-CK contains each of the above in a steel carrying case.)
When the shutdown process has been completed, move the collection train from the test
site to the sample recovery area to minimize the loss of collected sample or gain of
extraneous particulate matter. Set aside a portion of the acetone used in the sample
recovery as a blank for analysis. Place the samples in container as follows:
Container #1: Place the water contents from the first three impingers in this
container and seal.
Container #2: Remove the filter from its holder, place in this container, and seal.
Container #3: In this container, place loose particulate matter and acetone
washings from the filter half of the filter holder, cyclone and flask,
probe liner, and probe nozzle. (This is commonly referred to as
the front half.) Cleanup is best accomplished with a simultaneous
acetone-wash bottle rinse and brushing into the glass funnel
seated in the mouth of the container. A good reagent grade of
acetone with less than .001 percent residue, stored in glass
containers, gives excellent results. Each piece should be cleaned
thoroughly until all glassware is clear. The probe nozzle should be
removed from the union body, both openings covered, and set
aside. The probe liner should now be acetone-wash bottle rinsed
and brushed until all particulate matter is removed. The nozzle is
now ready for cleaning. Smaller nozzle sizes may present a
difficult clean-up task. Each nozzle cleaned should at least rinse
clear and some attempt should be made to brush the inner nozzle
walls.
Container #4: Transfer the silica gel from the fourth impinger to the plastic bottle
and seal. Use a rubber policeman as an aid in removing silica gel
from the impinger.
NOTE: At no time should the clean-up procedure be undertaken with soiled hands, or with a
dirt film on any glassware or the probe. If needed, clean all exterior surfaces before the
clean-up procedure is followed.
3.15 ANALYSIS
Record the data required on a data sheet. Handle each sample container as follows:
Container #1: Measure the volume of water. If an analysis of the back half, i.e.
back half of the filter holder contents of the impingers and the
glass connectors is required, refer to the state laws that govern
the clean-up of the back half for various sources. If no back half
extraction is required, discard the water after measuring the
volume.
Container #2: Transfer the filter and any loose particulate matter from the
sample container to a tarred glass-weighing dish, desiccate, and
dry to a constant weight. Report results to the nearest 0.5 mg.
Container #3: Transfer the acetone washings to a tarred beaker and evaporate
to dryness at ambient temperature and pressure. Desiccate and
dry to a constant weight. Report results to the nearest 0.5 mg.
Container #4: Weigh the spent silica gel and report to the nearest gram.
3.16 CALCULATIONS
Average dry gas meter temperature and average orifice pressure drop. See data sheet
(appendix).
Correct the sample volume measured by the dry gas meter to standard conditions (70°F,
29.92 inches Hg) by using the equation below:
∆H ∆H
Pbar + Pbar +
Tstd
Vmstd =Vm 13.6 =17.65 oR Vm 13.6
Tm Pstd in.Hg Tm
Where:
Vmstd = Volume of gas sample through the dry gas meter at standard conditions,
cu. ft.
Vm = Volume of gas sample through the dry gas meter (meter conditions), cu.
ft.
Tstd = Absolute temperature at standard conditions 530°R.
Tm = Absolute temperature at meter (°F + 460), °R.
Pbar = Barometric pressure at the orifice meter, in. Hg
∆H = Average pressure drop across the orifice meter, in. H2O
13.6 = Specific gravity of mercury
Pstd = Absolute pressure at standard conditions, 29.92 inches Hg
3.16.3 Volume of Water Vapor
ρH 2O RTstd cu. ft.
Vwstd =V 1c =0.0472 V 1c
MH 2O Pstd ml
Where:
Vwstd = Volume of water vapor in the gas sample (standard conditions), cu. ft.
V1c = Total volume of liquid collected in impingers and silica gel
ρH2O = Density of water, 1g/ml
MH2O = Molecular weight of water, 18 lb./lb.-mole
R = Ideal gas constant, 21.83 in. Hg-cu. ft./lb.-mole-°R
Tstd = Absolute temperature at standard conditions, 530°R
Pstd = Absolute pressure at standard conditions, 29.92 in. Hg
3.16.4 Moisture Content
Vwstd
Bwo =
Vmstd +Vwstd
Where:
Bwo = Proportion by volume of water vapor in the gas stream, dimensionless
Vwstd = Volume of water in gas sample (standard conditions), cu. ft.
Vmstd = Volume of gas sample through the dry gas meter (standard conditions),
cu. ft.
3.16.5 Stack Gas Velocity
(Ts )avg . 1 / 2
(Vs )avg .=KpCp ( ∆p avg.
PsMs
Where:
(Vs) avg. = Stack gas velocity, feet per second (f.p.s.)
1/ 2
lb.
Kp = 85.48ft. /sec.
lb. mole−° R
Cp = Pitot tube coefficient, dimensionless
(Ts) avg. = Average absolute stack gas temperature, °R
( ∆p ) avg. = Average velocity head of stack gas, in. H2O
Ps = Absolute stack gas pressure, in. Hg
Ms = Molecular weight of stack gas (wet basis), lb./lb.-mole
Md = Dry molecular weight of stack gas, (28.8 for air)
Bwo = Proportion by volume of water vapor in the gas stream
3.16.6 Volumetric Flow Rate
Tstd Ps
Qs=60(1− Bwo )VsA
(Ts )avg. Pstd
Where:
Qs = Volumetric flow rate, dry basis, standard conditions, ft3/min.
A = Cross-sectional area of stack, ft2
Tstd = Absolute temperature at standard conditions, 530°R
Pstd = Absolute pressure at standard conditions, 29.92 in. Hg
3.16.7 Total Particulate Weight
Determine the total particulate catch from the sum of the weights from the analysis.
3.16.8 Concentration in gr./s.c.f.
C’s = 0.0154
gr. Mn
mg.Vmstd
Where:
C’s = Concentration of particulate matter in stack gas, gr./s.c.f., dry basis
Ms = Total amount of particulate matter collected, mg.
Vmstd = Volume of gas sample through dry gas meter (standard conditions), cu.
ft.
3.16.9 Emission Rate in Pounds/Hour
lbs. min .
G=0.000143 60 C ' sQs
gr . hr.
Where:
G = Stack emission rate, lbs./min.
C’s = Concentration of particulate matter in stack gas, gr./s.c.f., dry basis
Qs = Volumetric flow rate, dry basis, standard conditions, ft3/min.
3.16.10 Percent Isokinetic
Where:
I = Percent of isokinetic sampling
V1c = Total volume of liquid collected in impingers and silica gel, ml.
ρH2O = Density of water, 1g. /ml.
Mh2O = Molecular weight of water, 18 lb./lb.-mole.
Vm = Volume of gas sample through the dry gas meter (meter conditions), cu.
ft.
Tm = Absolute average dry gas meter temperature, °R
Pbar = Barometric pressure at sampling site, in. Hg
∆H = Average pressure drop across the orifice, in. H2O
Ts = Absolute average stack gas temperature, °R
Q = Total sampling time, min.
Vs = Stack gas velocity calculated by ft./sec.
Ps = Absolute stack gas pressure, in. Hg
An = Cross-sectional area of nozzle, ft2
3.16.11 Acceptable Isokinetic Sampling Results
If 90% ≤ I ≤ 110%, the results are acceptable; otherwise, reject the results and repeat the
test.
3.17 ORSAT SAMPLE SYSTEM (OPTION -O)
Remove stack moisture using a chiller or dryer located at the sampling site (on the stack).
Connect the 1/4” quick-connect from the umbilical cable to the 1/4” orsat line quick-connect
located on the probe sheath. Connect the other end of the umbilical orsat line to the 1/4”
quick-connect labeled “ORSAT IN” on the front panel of the console. The “ORSAT IN” port
is immediately next to the 3/8” “PITOT +” quick-connect. Connect a length of 1/4” tygon
tubing and a 1/4” quick connect (not included in orsat option) to the “ORSAT OUT” port and
prepare to connect the tygon tubing to a Tedlar bag. Activate the orsat pump using the
“ORSAT ON” switch on the front electrical panel. Close the orsat control valve on the orsat
flow meter located on the front panel prior to attempting to capture a sample.
After the probe assembly is inserted in the gas stream, let the system purge for 2 to 3
minutes before connecting the tygon tube to the Tedlar bag valve. Be sure that the bag is
evacuated before use. After purging, adjust the flowrate control valve to a rate of 1.0 to 1.5
scfh (0.50 - 0.71 lpm). Shut off the valve each time the probe is removed from the gas
stream.
3 3
SOL. SOL.
1 2 ∆P ∆H ∆P ∆H
1 2
TC
TC
GAS METER
(shown here in outline only)
TO
IN
EXHAUST MA VACUUM
HOSE GAUGE
ORSAT
OUT IN OUT IN
ORSAT
PUMP PUMP
SAMPLE PITOT PRESSURE VAC
INLET
PRESSURE LINE
GAS LINE
THERMOCOUPLE
ELAPSED TIMER
To T/C Switch
“TC READOUT”
ORSAT
+ EMI FILTER
- NULL
3 4 5 6 3 4 5 6
FILTER PROBE
TEMP TEMP
CONTROLLER CONTROLLER
1 8 7 POWER
To T/C SWITCH 2 1 8 7 2 5 AMP
“TC READOUT” FILTER
To T/C -
“FILTER” +
AC
To T/C -
FILTER
“PROBE” + 10 AMP
PROBE
10 AMP POWER
PROBE
ON/OFF
1/L1 2/T1 1/L1 2/T1
FILTER PROBE
SS RELAY SS RELAY
25 AMP 25 AMP
AUX /PUMP
D A 15 AMP
4/-A2 3/A1+ 4/-A2 3/A1+
4-PIN
AMPHENOL
TC READOUT
11 - 21
STACK - - +
DGM OUT PUMP AUX
12 - 22
- - + 18 - 28
- - +
+ + + + +
PROBE DGM IN - - - - -
13 - 23 17 - 27
- - + - - +
FILTER AUX
14 - 24 16 - 26
- - + - - +
Page 27
DRYER
15 - 25
- - +
T/C SWITCH To T/C SWITCH To T/CSWITCH To T/C SWITCH To T/C SWITCH To T/C SWITCH
7-STATION “AUX” “DRYER” “FILTER” “PROBE” “STACK” C-5000 ELECTRICAL DIAGRAM