Healicom 14P Idrar Analizor
Healicom 14P Idrar Analizor
Operating Manual
1.1 Introduction
Intended use: the semi-auto urine analyzer is used in medical institutions
for semi-quantitative detection of biochemical composition in canine and
feline urine through reading and evaluating the matched test strips
specified by our company. Urinalysis includes the following items:
leucocytes (LEU), nitrite (NIT), urobilinogen (UBG), protein (PRO), potential
of hydrogen (pH), blood (BLD), specific gravity (SG), ketones (KET),
bilirubine (BIL), glucose (GLU), vitamin C (VC), calcium (Ca), creatinine (Cr)
and microalbumin (MA).
Main features:
Small and exquisite with stylish design and easy to carry.
Easy operation, measurement completing by pressing one button.
Built-in rechargeable lithium battery can support you to finish the
detection at anytime.
Bluetooth Transmission is supported.
Compatible with urine test strips of 11 to 14 test items.
Detachable test strips holder, easy to clean and reduce pollution.
1.2 Contraindications
None
1.4 Components
1.4.1 Front View
Power switch| Direction button| Setup button|
Function of Power switch ( ) :
Display screen
Press and hold this button for 3 seconds to
turn the analyzer on or off.
At any display: press this button to quit the
currentdisplay and return
to the main display.
Function of Direction (<, >) button:
Measure button
At main display: Press or to review the previous results forward
or backward one by one.
NOTE:
Some models of analyzer have this MicroUSB connector.
2 3
1.4.4 Display
unattended when using AC power supply.
Do not leave the equipment
placed. Keep it from falling, strong
5 67 The anaylzer should be appropriately mechanical
damage.
children or other
vibration, pets, pests,
88/B8/eDBB 0 Do not drop anything or put
it into the m e a s u r e m e n t entry of
the analyzer,
6. USB icon(displays when starting US8 function) Donot place anything on thetop of theanalyzer.
7. Battery status indicator
Menu, function code area
NOTE
8. The pictures and interfaces in this manual are for reference only.
This manual is based on the maximum configuration and therefore some
contents may not apply to your product. If you have any questions, please
1.5 Safety Information contact us.
WARNING
Before using the analyzer, you need access to professional medical 1.6 Symbols
Some symbols may not appear
knowledge. on your equipment.
equipment; any improper operation
may
the
Follow the instructions to use
5
2 Installation
NOTE:
Please save the packaging materials for future transport or storage use.
Keep the packing material out of children's reach. Dispose of the packaging
material, observing the applicable waste control regulations
The system may be contaminated by microorganism during transport, storage and
use. Verify the packaging, especially the packaging for the single use accessories,
is intact. In case of any damage, contact the carrier or our company immediately.
2.3 Connect the Power supply guidance. It must only be used by persons who have
received adequate
The analyzer can be powered by AC or built-in rechargeable Li-ion battery. operations of the analyzer by unauthorized or
training in its use. Any
untrained person are prohibited.
Power Requirements:
Main unit: DC 5V, 2A; AC 100V-240V+10%, 50/60Hz 3.1 Power On/Off
1. Before you start to make measurements, check the analyzer for any
NOTE: mechanical damage and make sure that accessories are properly
Use only the power adapter supplied or specified by Accuracy. connected.
Plug the adapter into the AC power source. If you run the analyzer
on
2.
2.4 Powered by Battery battery power, ensure that the battery is sufficiently charged.
The analyzer can be powered by built-in rechargeable Li-ion battery. If the 3. Press the button O for 3 seconds, "Beep" "Beep" will heard and the
AC power supply is suddenly off, the analyzer can be powered by battery. display screen of the analyzer will light on, self-test screen is displayed,
The battery will be charged when the analyzer is connected to AC power the transmission teeth starts moving.
supply. 4. The analyzer displays the main display after finishing the self-test.
5. Press and hold button for 3 seconds to turn off the analyzer after
On-screen battery symbols indicate the battery status as follows: finishing the measurement.
Indicates that the batteries work correctly. The solid portion
represents the current charge level of the batteries in proportion WARNING
to their maximum charge level. When the above indicator is Do not use the analyzer ifit is mechanically damaged or appears abnormal.
flashing, it indicates that the battery is low. Please charge it in Contact our company or our service personnel.
time.
The analyzer will not turn on when battery is lowif battery power used only
Indicates that the battery with full capacity.
3.2 Starting
NOTE: The self-test display is shown as right 22/c9/20 S 0
Do charge the battery separately with other charger.
not figure.
the Note: The figures shown in this manual are for
charge
t is
battery that
with power adapter supplied or specified by Accuracy
reference only. Please refer to the actual display
recommended you charge the battery within 2 months to ensure proper
use of the
analyzer long period no use of it.
if
Ensure that the battery has enough capacity in measurement; lower capaciy
lead to inaccurate results.
00:00:60
9
1. Preparing test stripP
If self-test error, the error code (ERR 002) 29/09/20 IS 0& urine test strip with the urine sample,
will display on screen, the analyzer issueda Use pipette to flood the
a
moistened. Do not bend the
the test pads are
long "Beep" alarm sound. making sure that all
strip.
Press button t o return to the main
display, and restart the analyzer. Please
contact our service personnel if the error
ERR 002 still exits.
Using Direction buttons ( , >) to view the results forward or backward. NOTE:
Use a clean urine cup and collect the fresh urine than 2 hours) for
(not more
testing.
3.3 Urine Analysis T h e bottom
CAUTION:
of the strip should be in full contact with the end of the notch.
Only remove the excess urine on sides
Please protective gloves to prevent contamination during operation.
T h e urine
of the strip.
wear
too high or too low. sample viscosity ensures that the strip is in close contact with the notch.
Do not use the analyzer if the ambient temperature measurement accuracy will If the
strip and the notch are dry, the test strip may slide and result in inaccurate
the
not the analyzer in direct sunlight place; measurements.
Do use
10 11
Starting t h e m e a s u r e m e n t
2 press the 22/09/20 Is 0 The measurement displays after 29/09/2o 15 0
a
Place the strip well, and e)
S6 finishing the test as shown on right MOCOCS
Measure button (), the measure
6:22:36
12 13
3.5 Review Measurements 2. Upload measurements to PC or inteligert terminas va Situeoc
At main display,press the Direction button (<, >) to view the previous (Optional)
results one by one. Press and hold the Direction button ( , >) to view the 1) At main display, press Setup button ()eeærhe seuc seen
results quickly. 2) Using direction button ( <, >) to seiect meru 'SET SU. s s o u n
At review measurements display below.
Press button to return to the main display.
Press Measure button ( ) to start another test.
NOTE:
U s e only the Us8 catle supeied or specfied by cur coman
09:4 : 9 When set the Us8 connecion, press irection u m ( ? iges a
3) then press Measure button d o e UsS comecteon
Press Measure button (),* flashes.
4) Press When set the Bhuetoath fanction, press directiorn u n ) iges ani
Direction
button ( <), lights on the screen. Press Measure
button () again to confirm the USB connection.
then press Measure buton
( dose
Suetmh camectic
Any questions about Usa and Biuetocth semins piease cavae c s g a
5) Press button to return tomain display. service personne
6) Connect the analyzer to PC with USB cable;run the software in PC, you
can upload the required measurements to PC after finishing the
connection settings.
15
14
3.7 Delete Measurements 3. Press Measure button (), the year numeric flashes.
1 At main display, press Setup button ( ) to enter the setup screen. 4. Using Direction button ( , >) tosetyear numeric, and then press
2 Using Direction button(<, >) to select menu "DEL DAT", as shown Measure button ( ) to confirm, the cursor skip to month numeric,
below the month numeric flashes.
5. Refer to step 4 to set the month, date, hour, minute and second in
dEL
turn.
6. Press Power button () to confirm the setting and return to the main
display.
Press Measure button (O), the screen flashes, and sound "Beep", SEE SEM
NOTE:
f the deletion is failed, the screen displays "FAIL"
If no results in analyzer, the screen displays "null".
Operate deletion with caution because this operation will delete all IS:S0:20
measurements stored in analyzer.
3. Press Measure button ) to choose the detection item, for example
3.8 Date and Time Setting LEU
1 At main display, press Setup button (O) to enter the setup screen. 22109120 16 0 a
2. SEE 000
Using Direction button (K,>) to select menu "SET TIE", as shown
below. LEU
209/20 1s 0.<
SEE
IS:S0:25
4 Using Direction button ( <, >) to adjust sensitivity value, and then
press Power button ( C) to confirm the setting and return to the main
ISS022
17
16
display. 4 Care and Cleaning
NOTE:
tis
recommended tousethe sensitivity adjustment function when the full
quality control fluid test data is accumulated. Use only the substances approved by us and methods listed in this chapter
Any questions about sensitivity adjustment please contact Kongsung or service to clean or disinfect your equipment. Warranty does not cover damage
personnel caused by unapproved substances or methods.
Standard reference
pad
18 19
Wipe the holder notch using a cotton swad dampened with akohol 5Warranty and Service
and remave the dirt in noteh.
5.1 Warranty
The analyzer is covered by a comprehensive warranty for 12 months from
the date of production. However, damage occurring under the following
conditions shall not be covered by this warranty
Man-made damage or damage caused by improper use.
5 Install the hoider into analyzer aitercieaning Damage caused by mishandling during shipment.
Damage caused by uncontrollable natural factors such as arthquake,
fire or war.
Environment in which the analyzer is used does not meet the
requirements indicated in this manual.
Damage caused by use of an unspecified power supply or any other
abnormality in the power supply.
T u m the analyrer of Damage caused as a result of maintenance performed by personnel
not authorized by the manufacturer.
NOTE: Malfunction of the analyzer whose serial number is not legible
Keeo the woriing enironment and the analyzer suface dean and do not
damage the sareen
enough
Malfunction not caused by the analyzer itself.
The aicohol gasoline, berzene organic sohet cieaning are forbidden to use,
these tests wmake the anayzer deformation or pait removed and affect the
periormane r aPpearance To protect your rights, please keep the evidence, such as invoices,
Cean the tet strip hoider after usg eey daay receipts and so on well and fill in the warranty card and post it to the
Do nct cortaminate ar damage the coior bioci when cdeaning the strip holder. manufacturers. The manufacturer will repair or replace the analyzer for
Any service of this analyzer can only be performed by an authorized senvie free during its warranty period and any damage caused by non-human
engineer. Do not maintain and disassembly the analyaer. f amy quaity problems
factors
ocaur, piease cal our astoner senie
To keep the srip hoider deaning remove the urine residue in time, the
cress-cortamination wl affect the acnracy of the test resuts 5.2 Service
f any questions in equipment operation, please contact the manufacturer
or local agency.
20 21
6 Accessory 7 Troubleshooting
NOTE: Some common troubles will occur during the operation of analyzer. Follow
Use the special strips provided by Accuracy in order to ensure the accuracy of the the below description to solve the troubles.
test results.
Use only the data cable and power adapter supplied or specified by Accuracy.
Strip is single use, do not reuse. Symptoms Potential Causes Solution
The analyzer Battery is low and no Check the battery, and charge the
Accessory name/Model Quantity Remarks cannot turn AC power is battery if the battery capacity is
Power adapter Stanadard configuration on. connected. empty. If the analyzer cannot turn
pcs
on, contact our service personnel.
Test strip 1 bottle Single use
Press the Test strip holder is Replace a new strip holder.
USB data cable 1 pcs Stanadard configuration Measure damaged.
Bluetooth thermal printer 1 pcs Optional button after The strip holder is Disassembly the holder and clean
loading strip installed in wrong it, check its transmission teeth
well, the and then reinstall the holder to
place or there is
analyzer object obstruction. analyzer. Restart the analyzer.
cannot start Contact our service personnel if
measure the trouble still exits.
after 60
Analyzer is damaged. Replace a new analyzer.
seconds
Battery is low and no Connect to the AC power supply
AC power is to charge battery and make a new
connected. measurement.
22 23
S Electromagnetic Compatibility A Product Specifications
24 25
Throughput 240 tests/hour in fast mode, 60 tests/hour in 8.0 8.0 8.0
normal mode 8.5 8.5 8.5
Interfaces Micro USB interface: charge the built-in battery Blood Ocells/ul Ocells/ul
or connect the data cable for data transmission. (BLD) 10cells/uL 10cells/u
25cells/uL 25cells/ul +1
Display LCD display
+2
80cells/u 80cells/ul
Button 5 buttons +3
250cells/uL 250cells/ul
Power supply DC 5Vt10%: Specific gravity 1.000 1.000 1.000
AC 100V-240V+10%. 20VA, 50/60Hz (SG) 1.005 1.005 1.005
1.010 1.010 1.010
26 27
2.5mmol/L 2.5mmol/L B Symbols and Abbreviation
5.0mmol/L 5.0mmol/l +2
10.0mmolL 10.0mmol/L +3 B.1 Units
Creatinine 0.9mmol/L 0.9mmol/L Abbreviation English
(C) 4,4mmol/L 4.4mmol/L +1 ul(uL) microliter
17.7mmol/L 17.7mmol/L +2 Lumol micromole
26.5mmol/L 26.5mmol/ A
ampere
Wicroalbumin Omg/L Omg/L C centigrade
(1MA) Cm centimeter
30mg/L 30mg/L
cells
80mg/L 80mg/L +2 cells
150mg/L 150mg/L +3
dl(dL) deciliter
gram
h
hour
Hz |hertz
L liter
milligram
mi
milliliter
mm
millimeteer
mmol millimole
second
V
volt
|W watt
B.2 Symbols
Symbols English
negative
percent
per; divide; or
to
positive
greater than or equal to
copyright
28 29
B.3 Terms
English Abbr. Full name
AC Alternating current
Ca Calcium
Cr Creatinine
DC Direct current
TEMP Temperature
UBG Urobilinogen
PRO Protein
pH Potential of Hydrogen
BLD Blood
SG Specific Gravity
KET Ketones
BIL Bilirubine
GLU Glucose
VC Vitamin C
30