8 T E C H N I C A L PA R A M E T E R S
Name BTL-4000 Smart, BTL-4000 Premium
Models see Chapter 8.1
Operating conditions
ambient temperature +10 °C to +30 °C
relative humidity 30 % to 75 %
atmospheric pressure 800 hPa to 1,060 hPa
position vertical – on legs
type of operation continuous
Transport and storage conditions
ambient temperature -10 °C to +55 °C
relative humidity 10 % to 85 %
atmospheric pressure 650 hPa to 1,100 hPa
position any
other conditions transport only in the supplied packaging
Power supply external power adaptor
maximum input 60 W / 90 W
supply voltage 24 V, DC
protection class II for devices with adaptor I for devices with adaptor
SA160D-24U-M HPU101-108
device on/off button on the front panel, marked on/off
Internal chemical sources
battery lithium battery CR2430
Power adaptors BTL-4000 Adaptor BTL-4000 Adaptor
Model: SA160D-24U-M Model: HPU101-108
supply voltage ~ 100 V to 240 V AC
frequency 50 Hz to 60 Hz
cable length 150 cm 190 cm
rated voltage / current / input power 24 V / 2.5 A / 60 W 24 V / 4.16 Amax / 90 W
protection class II I (Caution: the equipment must always
be connected to the protective ground)
fuse internal
IP20 - protects against penetration of solid objects of a diameter of
covering grade
12.5 mm and more.
Reserve battery Model:
of 12.5 258-BATTPACK
mm and more.
Design
weight – device only max. 3 kg (6.1 lb)
weight – including packing and accessories max. 6 kg (13.2 lb)
dimensions (W x H x D) 380 mm x 190 mm x 260 mm (15“ x 7.5“ x 10.2“)
packing dimensions (W x H x D) 435 mm x 365 mm x 335 mm (17.1“ x 14.4“ x 13.2“)
covering grade IP20 – protects against penetration of solid objects of a diameter of
12.5 mm and more.
Display elements
display – BTL-4000 Smart LCD colour, 480 x 272, diagonal 10.9 cm (4.3“)
display – BTL-4000 Premium LCD colour, 800 x 480, diagonal 17.8 cm (7“)
Classification
applied parts of type BF
class according to MDD 93/42/EEC IIb
Therapy time
for electro and laser therapies 0 to 100 minutes
for ultrasound therapies 0 to 30 minutes
step of setting 1 second
accuracy of therapy time 5 % of the set value
Accuracy of time values 5 seconds per 1 day
BTL-4000 SMART/PREMIUM - USER‘S MANUAL I PAGE 45 OF 72