Medical-Grade Software Development. How To Build Medical-Device Products That Meet The Requirements of IEC 62304 and ISO 13485-Routledge. Taylor & Francis Group (2024)

Uploaded by

rh1987
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (3 votes)
1K views367 pages

Medical-Grade Software Development. How To Build Medical-Device Products That Meet The Requirements of IEC 62304 and ISO 13485-Routledge. Taylor & Francis Group (2024)

Uploaded by

rh1987
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 367

Medical-Grade Software

Development
This book is a practical guide to meeting IEC 62304 software-development
requirements within the context of an ISO 13485 quality management system
(QMS). The book proves this can be done with a minimum amount of friction,
overlap, and back-and-forth between development stages. It essentially shows
you how you should shape your medical-software development processes to fit
in with the QMS processes in the smartest and leanest way possible.
By following the advice in this book, you can reuse processes from your
QMS, ensure your product-realization processes meet the requirements for
medical-software development, and marry all the requirements together
using tried and tested solutions into one efficient system. The expertise of
the authors here goes beyond just the experiences of one real-world proj-
ect as they tap into over 30 years of experience and countless software and
software-assessment projects to distill their advice.
The book takes a hands-on approach by first teaching you the top 25
lessons to know before starting to develop a process for medical-software
development. It then walks you through the expectations placed on the key
aspects of such a process by the key standards. The book progresses from
an overview of both standards and the general requirements involved to a
detailed discussion of the expected stages from software development and
maintenance to risk management, configuration management, and problem
resolution. The book provides insightful advice on how the requirements of
the IEC 62304 software-development life cycle can be married with an ISO
13485 QMS, how the development of the technical file should be organized,
and how to address conformity assessment, the daily after-approval, and the
recent trends that will affect the industry in the coming years.
The book is modeled after the IEC 62304 standard and adopts its clause
structure in the numbering of sections for easy reference. The book does not
attempt to replicate either standard. For the ISO 13485 standard, it recites the
necessary requirements succinctly. For IEC 62304, the discussion is in-depth and
also addresses the impact of ISO 13485 on the requirements discussed. In this
way, the book drills into both standards to expose the core of each requirement
and shape these into a practical, cohesive workflow for developing, maintaining,
and improving a Lean software development pipeline.
Medical-Grade Software
Development
How to Build Medical-Device
Products That Meet the Requirements
of IEC 62304 and ISO 13485

Ilkka Juuso
Ilpo Pöyhönen
First published 2024
by Routledge
605 Third Avenue, New York, NY 10158

and by Routledge
4 Park Square, Milton Park, Abingdon, Oxon, OX14 4RN

Routledge is an imprint of the Taylor & Francis Group, an informa business

© 2024 Ilkka Juuso & Ilpo Pöyhönen

The right of Ilkka Juuso & Ilpo Pöyhönen to be identified as author of this work has been asserted by them in
accordance with sections 77 and 78 of the Copyright, Designs and Patents Act 1988.

All rights reserved. No part of this book may be reprinted or reproduced or utilised in any form or by any electronic,
mechanical, or other means, now known or hereafter invented, including photocopying and recording, or in any
information storage or retrieval system, without permission in writing from the publishers.

Trademark notice: Product or corporate names may be trademarks or registered trademarks, and are used only for
identification and explanation without intent to infringe.

ISBN: 9781032593180 (hbk)


ISBN: 9781032594729 (pbk)
ISBN: 9781003454830 (ebk)

DOI: 10.4324/9781003454830

Typeset in Garamond
by Deanta Global Publishing Services, Chennai, India
Dedicated to encounters of the best kind
– and learning from it all.
Contents

List of Figures and Tables...................................................................... xiv


Preface.......................................................................................................xix
Acknowledgments................................................................................. xxiv
About the Authors................................................................................. xxvi
1 What to Know before Getting Started..........................................1
1.1 There Is No One Correct Answer to Anything..................................2
1.2 Why Do We Need a Standard at All?.................................................5
1.3 What Is Medical-Grade Software?.......................................................8
1.4 Software Is Classified by the Safety Concerns Associated with It...10
1.5 Software Is Made of Units, Items, and Systems...............................12
1.6 Software Development Has a Life Cycle..........................................14
1.7 Everything Is a Process, and Those Consist of Activities and
Tasks..................................................................................................16
1.8 IEC 62304 Does Not Insist on a Specific Process Structure............18
1.9 IEC 62304 Does Not Impose a Documentation Set.........................19
1.10 Applicable Regulations as Part of the Requirements.......................20
1.11 Building on the Quagmire That Surrounds IEC 62304....................22
1.12 Retaining Trust at Each Step of the Way Is Perhaps the Most
Critical of Tasks.................................................................................24
1.13 The Relationship between Code and Documentation is at the
Core of the Marriage, Even after a Shotgun-Wedding.....................26
1.14 Evaluation, Review, Verification, and Validation..............................27
1.15 How Many Reviews Is Enough?........................................................31
1.16 How Much Testing Is Enough?.........................................................34
1.17 Fail and Pass......................................................................................38
1.18 One Step Forward, Two Steps Back.................................................41
1.19 Continuous Improvement..................................................................43

 vii
viii ◾ Contents

1.20 What Is SOUP?...................................................................................45


1.21 Configuration Items and Controlled Changes..................................46
1.22 Legacy Software Has a Dedicated Pathway Through IEC 62304....48
1.23 Software Always Fails, and Other Software Risk Beliefs.................49
1.24 Certification Bodies and the Notified Body.....................................51
1.25 Safety Is Relative................................................................................53
2 ISO 13485 as the Backbone of It All..........................................55
2.1 The Whole of the Quality Management System..............................56
2.2 The Process for Product Realization.................................................57
3 IEC 62304 as The Flesh around The Bones................................61
3.1 The Past, Present, and Future...........................................................62
3.1.1 Times of the Primordial Ooze................................................62
3.1.2 Modern Times.........................................................................65
3.1.3 The Future.............................................................................. 67
3.2 Contents.............................................................................................69
3.2.1 Foreword.................................................................................69
3.2.2 Introduction............................................................................70
3.2.3 Clause 1: Scope.......................................................................73
3.2.4 Clause 2: Normative References.............................................76
3.2.5 Clause 3: Terms and Definitions............................................77
3.2.6 Clause 4: General Requirements............................................79
3.2.7 Clause 5: Software Development Process..............................80
3.2.8 Clause 6: Software Maintenance Process...............................80
3.2.9 Clause 7: Software Risk-Management Process.......................82
3.2.10 Clause 8: Software Configuration Management Process.......82
3.2.11 Clause 9: Software Problem Resolution Process....................83
3.2.12 Annex A – Rationale for the Requirements of This
Standard..................................................................................84
3.2.13 Annex B – Guidance on the Provisions of This Standard....85
3.2.14 Annex C – Relationship to Other Standards..........................85
3.2.15 Annex D – Implementation....................................................88
4 General Requirements................................................................89
4.1 Quality Management Systems...........................................................90
4.2 Risk Management..............................................................................91
4.3 Software Safety Classification...........................................................92
4.3.1 Classification Formula.............................................................93
4.3.2 What Is Harm?........................................................................94
4.3.3 Accept That All Software Will Fail.........................................95
Contents  ◾ ix

4.3.4 Use of External Risk Controls................................................96


4.3.5 Reducing Both Probability and Severity................................97
4.3.6 Recording and Inheriting the Classification...........................98
4.3.7 Final Remarks.........................................................................99
4.4 Legacy Software..............................................................................100
4.4.1 Risk-Management Activities..................................................100
4.4.2 Gap Analysis......................................................................... 101
4.4.3 Gap Closure Activities..........................................................103
4.4.4 Rationale for Use of Legacy Software..................................104
5 Software Development..............................................................105
5.1 Development Planning....................................................................108
5.1.1 Expectations from ISO 13485...............................................109
5.1.2 Expectations from IEC 62304............................................... 111
5.1.3 Suggested Synthesis.............................................................. 116
5.2 Requirements Analysis....................................................................124
5.2.1 Expectations from ISO 13485...............................................126
5.2.2 Expectations from IEC 62304...............................................127
5.2.3 Suggested Synthesis..............................................................130
5.3 Architectural Design........................................................................132
5.3.1 Expectations from ISO 13485...............................................133
5.3.2 Expectations from IEC 62304...............................................134
5.3.3 Suggested Synthesis..............................................................137
5.4 Detailed Design...............................................................................138
5.4.1 Expectations from ISO 13485...............................................139
5.4.2 Expectations from IEC 62304............................................... 140
5.4.3 Suggested Synthesis.............................................................. 141
5.5 Unit Implementation and Verification............................................ 141
5.5.1 Expectations from ISO 13485............................................... 142
5.5.2 Expectations from IEC 62304............................................... 143
5.5.3 Suggested Synthesis.............................................................. 145
5.6 Integration and Integration Testing.................................................146
5.6.1 Expectations from ISO 13485............................................... 147
5.6.2 Expectations from IEC 62304...............................................148
5.6.3 Suggested Synthesis.............................................................. 151
5.7 Software-System Testing.................................................................. 152
5.7.1 Expectations from ISO 13485............................................... 153
5.7.2 Expectations from IEC 62304............................................... 154
5.7.3 Suggested Synthesis.............................................................. 156
x ◾ Contents

5.8 Release............................................................................................. 156


5.8.1 Expectations from ISO 13485............................................... 158
5.8.2 Expectations from IEC 62304............................................... 162
5.8.3 Suggested Synthesis.............................................................. 165
5.9 The Parts Left Out by IEC 62304.................................................... 170
5.9.1 Conducting Reviews in D&D Stages.................................... 170
5.9.2 D&D Verification.................................................................. 170
5.9.3 D&D Validation..................................................................... 172
6 Software Maintenance.............................................................. 175
6.1 Software Maintenance Plan............................................................. 176
6.1.1 Expectations from ISO 13485............................................... 177
6.1.2 Expectations from IEC 62304............................................... 179
6.1.3 Suggested Synthesis.............................................................. 179
6.2 Problem and Modification Analysis................................................180
6.2.1 Expectations from ISO 13485...............................................182
6.2.2 Expectations from IEC 62304...............................................182
6.2.3 Suggested Synthesis..............................................................184
6.3 Modification Implementation.......................................................... 185
6.3.1 Expectations from ISO 13485...............................................186
6.3.2 Expectations from IEC 62304...............................................186
6.3.3 Suggested Synthesis..............................................................187
6.4 The Parts Left Out By IEC 62304....................................................188
7 Risk Management.....................................................................189
7.1 Analysis of Software Contributing to Hazardous Situations.......... 194
7.1.1 Expectations from ISO 13485............................................... 197
7.1.2 Expectations from IEC 62304............................................... 197
7.1.3 Suggested Synthesis..............................................................199
7.2 Risk-Control Measures.....................................................................199
7.2.1 Expectations from ISO 13485...............................................199
7.2.2 Expectations from IEC 62304...............................................199
7.2.3 Suggested Synthesis..............................................................200
7.3 Verification of Risk-Control Measures............................................200
7.3.1 Expectations from ISO 13485...............................................200
7.3.2 Expectations from IEC 62304...............................................200
7.3.3 Suggested Synthesis..............................................................201
7.4 Risk Management of Software Changes.........................................201
7.4.1 Expectations from ISO 13485...............................................201
Contents  ◾ xi

7.4.2 Expectations from IEC 62304...............................................202


7.4.3 Suggested Synthesis..............................................................202
7.5 The Parts Left Out by IEC 62304....................................................202
8 Configuration Management......................................................205
8.1 Identification of Configuration Items..............................................207
8.1.1 Expectations from ISO 13485...............................................208
8.1.2 Expectations from IEC 62304...............................................208
8.1.3 Suggested Synthesis..............................................................209
8.2 Change Control (Incl. Verification)................................................. 215
8.2.1 Expectations from ISO 13485............................................... 215
8.2.2 Expectations from IEC 62304............................................... 216
8.2.3 Suggested Synthesis.............................................................. 217
8.3 History of Controlled Items............................................................ 218
8.3.1 Expectations from ISO 13485............................................... 219
8.3.2 Expectations from IEC 62304............................................... 219
8.3.3 Suggested Synthesis.............................................................. 219
8.4 The Parts Left Out by IEC 62304....................................................220
9 Problem Resolution..................................................................221
9.1 Prepare Problem Reports................................................................222
9.1.1 Expectations from ISO 13485...............................................224
9.1.2 Expectations from IEC 62304...............................................225
9.1.3 Suggested Synthesis..............................................................225
9.2 Investigate the Problem...................................................................226
9.2.1 Expectations from ISO 13485...............................................227
9.2.2 Expectations from IEC 62304...............................................227
9.2.3 Suggested Synthesis..............................................................228
9.3 Advise Relevant Parties...................................................................229
9.3.1 Expectations from ISO 13485...............................................230
9.3.2 Expectations from IEC 62304...............................................230
9.3.3 Suggested Synthesis..............................................................230
9.4 Use Change-Control Processes....................................................... 231
9.4.1 Expectations from ISO 13485...............................................232
9.4.2 Expectations from IEC 62304...............................................232
9.4.3 Suggested Synthesis..............................................................232
9.5 Maintain Records.............................................................................233
9.5.1 Expectations from ISO 13485...............................................234
9.5.2 Expectations from IEC 62304...............................................234
9.5.3 Suggested Synthesis..............................................................234
xii ◾ Contents

9.6 Analyze Problems for Trends..........................................................235


9.6.1 Expectations from ISO 13485...............................................235
9.6.2 Expectations from IEC 62304...............................................236
9.6.3 Suggested Synthesis..............................................................236
9.7 Verify Software Problem Resolution...............................................237
9.7.1 Expectations from ISO 13485...............................................238
9.7.2 Expectations from IEC 62304...............................................239
9.7.3 Suggested Synthesis..............................................................240
9.8 Test Documentation Contents.........................................................241
9.8.1 Expectations from ISO 13485...............................................241
9.8.2 Expectations from IEC 62304...............................................242
9.8.3 Suggested Synthesis..............................................................242
9.9 The Parts Left Out by IEC 62304....................................................243
10 Integration with Your QMS......................................................245
10.1 Write Your Product Realization Processes in the QMS..................247
10.2 Develop Your Planning Documentation According to Your
QMS............................................................................................... 250
10.3 Execute on Your Plans as Expected by Your Arrangements.........250
10.4 Execute on Your Post Release Activities......................................... 251
10.5 Keep Your QMS in Good Shape..................................................... 251
11 Technical Documentation......................................................... 252
11.1 What Is Technical Documentation Anyway?..................................253
11.2 Process Documents.........................................................................256
11.3 Development Records.....................................................................256
11.4 The Audit Package..........................................................................259
12 Seeing into the Future..............................................................260
12.1 Agile Software Development.......................................................... 261
12.1.1 What Is Agile Anyway?.........................................................262
12.1.2 Development Life Cycle in a Regulated Environment.........265
12.2 IEC 82304-1 on Health Software..................................................... 267
12.2.1 Product Requirements..........................................................269
12.2.2 Software Development Process............................................271
12.2.3 Validation..............................................................................271
12.2.4 Product Documentation.......................................................272
12.2.5 Post-Market Activities............................................................275
12.3 Segregation in Software Architecture.............................................276
12.4 Risk Management Influenced by FMEA.........................................280
Contents  ◾ xiii

12.5 Usability Engineering......................................................................284


12.6 Cybersecurity...................................................................................290
12.6.1 Cybersecurity Over Product Life Cycle................................291
12.6.2 Cybersecurity Across Jurisdictions.......................................293
12.6.3 IEC 81001-5-1........................................................................294
12.6.4 ISO/IEC 27001.......................................................................295
12.7 Artificial Intelligence.......................................................................296
12.8 Cloud Computing............................................................................300
12.9 Edge Intelligence.............................................................................302
13 Conformity Assessment............................................................304
13.1 Requirements Placed on the Assessor............................................305
13.2 The Assessment...............................................................................308
13.3 Typical Shortcomings in Assessments............................................ 310
13.4 What Happens Afterward?.............................................................. 313
14 Regulatory Approval................................................................. 315
14.1 Benefits of the IEC 62304 Test Report............................................ 315
14.2 ISO 13485 Certification................................................................... 317
14.3 Medical-Device Conformity Assessment........................................ 318
14.4 Changes........................................................................................... 319
15 Business as Usual..................................................................... 321
15.1 The Everyday Processes to Now Run.............................................322
15.2 The Future of IEC 62304.................................................................324
15.3 The Future of ISO 13485.................................................................325
15.4 The Joy of Compliance...................................................................325
16 Conclusions...............................................................................327
References....................................................................................... 331
Index...............................................................................................332
List of Figures and Tables

Figures

Figure 1.1 The sweet spot is wider than just a single hole to hit.................3
Figure 1.2 Standards are like maps to the otherwise uncharted world.......6
Figure 1.3 The definition of “medical software” is a slippery concept
to nail down.........................................................................................................8
Figure 1.4 It’s not quite as easy as in that nursery rhyme, but
knowing your ABCs is required........................................................................11
Figure 1.5 System begets an item, an item begets a unit............................12
Figure 1.6 Not to put too fine a point on it all, but a life cycle is just
that..................................................................................................................... 15
Figure 1.7 Everything is a process...............................................................17
Figure 1.8 The pieces of the puzzle are up to you.....................................18
Figure 1.9 Is a document by any other name not still a document?..........19
Figure 1.10 Applicable regulations are a foundational part of your
requirements......................................................................................................20
Figure 1.11 The IEC 62304 has reached a point of stable uncertainty.......23
Figure 1.12 Steps to trust may be different in height, but the action is
always the same.................................................................................................24
Figure 1.13 Code and documentation should have a symbiotic
relationship.........................................................................................................26

xiv
List of Figures and Tables  ◾ xv

Figure 1.14 Verification and validation have similar goals, but


different points of view.....................................................................................28
Figure 1.15 “Review vertigo” may lead to more fatigue than attention
to detail..............................................................................................................32
Figure 1.16 Testing is required to trust that it all works.............................35
Figure 1.17 To pass or to fail, that is the Shakespearean question in
software..............................................................................................................39
Figure 1.18 Maintaining forward momentum may at times be difficult....42
Figure 1.19 Your path into QMS should have a rising crescendo..............44
Figure 1.20 SOUP or not, you’ll need to know what’s in it........................45
Figure 1.21 Configuration covers both versions of your product and
what is in it........................................................................................................46
Figure 1.22 The IEC 62304 has a charted path to conformance for
old software.......................................................................................................48
Figure 1.23 You don’t have to be Tom Cruise to find your “Porsche”
in a creek if you ignore the signs.....................................................................50
Figure 1.24 Know that not all certificates carry equal value......................51
Figure 1.25 Marshmallows on cobra fangs might increase your sense
of safety, but will it be safe enough?................................................................53
Figure 3.1 The collateral and particular standards of IEC 60601-1.............63
Figure 3.2 Scope of IEC 60601-1-4 on software..........................................64
Figure 3.3 Demonstration of compliance with standards IEC 60601-1
Clause 14, IEC 62304, and IEC 82304-1............................................................66
Figure 3.4 Word cloud for the Introduction.................................................74
Figure 3.5 Word cloud for Clause 1..............................................................76
Figure 3.6 Word cloud for Clause 2.............................................................77
Figure 3.7 Word cloud for Clause 3.............................................................79
Figure 3.8 Word cloud for Clause 4.............................................................80
Figure 3.9 Word cloud for Clause 5.............................................................81
Figure 3.10 Word cloud for Clause 6...........................................................82
xvi ◾ List of Figures and Tables

Figure 3.11 Word cloud for Clause 7............................................................83


Figure 3.12 Word cloud for Clause 8...........................................................83
Figure 3.13 Word cloud for Clause 9...........................................................84
Figure 3.14 Word cloud for Annex A...........................................................85
Figure 3.15 Word cloud for Annex B...........................................................86
Figure 3.16 Word cloud for Annex C...........................................................87
Figure 3.17 Word cloud for Annex D..........................................................88
Figure 4.1 Key to safety classification..........................................................93
Figure 4.2 The use of P1 and P2 in a risk matrix.......................................97
Figure 5.1 Both standards look to provide bookends and bookmarks
for your work...................................................................................................108
Figure 5.2 During planning your horizons are at their widest.................109
Figure 5.3 Requirements may be a bit of an octopus sushi to
wrangle, but they should never be treated as an alphabet soup...................125
Figure 5.4 Floppies were once the modern equivalent to paper
scrolls, but now they too may be relegated to ancient ruins on our
architectural map.............................................................................................133
Figure 5.5 Detailed design has us zooming into the architectural map
we previously created......................................................................................138
Figure 5.6 After all that designing, this is where we move to first
doing and then checking the results.............................................................. 142
Figure 5.7 This is that “E pluribus unum” moment: out of many
becomes one.................................................................................................... 147
Figure 5.8 Testing should lead to that “all systems are go” feeling.......... 153
Figure 5.9 When the software is ready for the limelight, it’s time............ 158
Figure 6.1 Software maintenance must be on your roadmap................... 177
Figure 6.2 My dog ate my software........................................................... 181
Figure 6.3 Problem-resolution process may be akin to field surgery....... 185
Figure 7.1 Standalone networks, private networks, and the cloud
each come with different risks associated with them....................................196
List of Figures and Tables  ◾ xvii

Figure 11.1 Example structure for technical documentation.................... 255


Figure 11.2 Technical documentation in DIR and DOR...........................257
Figure 11.3 Technical documentation in DTR...........................................257
Figure 12.1 Segregation breakdown due to memory access....................278
Figure 12.2 Hazards as a gateway to risk management...........................281
Figure 12.3 RPN risk scores scaled to emphasize the role of severity.....283
Figure 12.4 Deriving use specifications and task descriptions from
the intended use..............................................................................................287
Figure 12.5 The difference between black-box and white-box AI...........298

Tables

Table 2.1 A Rough Mapping between ISO 13485 Clause 7 and IEC
62304..................................................................................................................59
Table 5.1 Topics of Software Development................................................106
Table 5.2 Alignment of ISO 13485 and IEC 62304 Clause 5..................... 107
Table 5.3 Synthesis of Top-Level Requirements......................................... 117
Table 5.4 Synthesis of Top-Level Requirements......................................... 131
Table 5.5 Synthesis of Top-Level Requirements.........................................137
Table 5.6 Synthesis of Top-Level Requirements......................................... 141
Table 5.7 Synthesis of Top-Level Requirements......................................... 146
Table 5.8 Synthesis of Top-Level Requirements......................................... 152
Table 5.9 Synthesis of Top-Level Requirements......................................... 157
Table 5.10 Synthesis of Top-Level Requirements.......................................166
Table 6.1 Topics of Software Maintenance................................................. 176
Table 6.2 Synthesis of Top-Level Requirements.........................................180
Table 6.3 Synthesis of Top-Level Requirements.........................................184
Table 6.4 Synthesis of Top-Level Requirements.........................................187
xviii ◾ List of Figures and Tables

Table 7.1 Topics of Risk Management........................................................ 192


Table 8.1 Topics of Configuration Management.........................................207
Table 8.2 Synthesis of Top-Level Requirements.........................................209
Table 8.3 Synthesis of Top-Level Requirements......................................... 218
Table 8.4 Synthesis of Top-Level Requirements......................................... 219
Table 9.1 Topics of Problem Resolution.....................................................223
Table 9.2 Synthesis of Top-Level Requirements.........................................226
Table 9.3 Synthesis of Top-Level Requirements.........................................228
Table 9.4 Synthesis of Top-Level Requirements......................................... 231
Table 9.5 Synthesis of Top-Level Requirements.........................................232
Table 9.6 Synthesis of Top-Level Requirements.........................................234
Table 9.7 Synthesis of Top-Level Requirements.........................................236
Table 9.8 Synthesis of Top-Level Requirements.........................................240
Table 9.9 Synthesis of Top-Level Requirements.........................................243
Table 10.1 Basic Software Development Life Cycle...................................248
Table 13.1 Excerpt From IEC 60601-1 Clause 14.6 Test Report.................309
Table 13.2 Excerpt From IEC 62304 Clause 5.4 Test Report......................309
Preface

Medical software has a lot to answer for. It is an industry based on decades


of experience built up from the development of software in industries
ranging from aviation and military to mobile applications and the piece of
embedded binary DNA running your toaster. Software is in all these cases
the logic or even the intelligence animating the physical devices, giving them
an artificial brain or a rudimentary nerve cord of sorts. The structure and
rigor involved in the development of software on these disparate domains,
however, varies greatly. After all, the different application areas for software
have vastly different characteristics. A mid-air software update to a jetliner
on a transatlantic route is unacceptable and could cause the plane to crash.
A forced mid-surgery operating system update in the hospital operating
room will cause headlines across the world, but it does happen. An update
to a popular mobile application or game on the other hand is expected and
is often a sign of a well-maintained and evolving app.
Medical software has evolved out of this diverse pedigree. Some medical
software today is developed by individual hobbyists, others by established
medical-device companies, and yet others by big companies moving into
health and wellbeing from other domains of business. The division between
what is considered a simple health application and what is hardcore medi-
cal software is at times a judgment call, and occasionally similar, almost
identical, software products can sit on opposite sides of this fence. At the
same time, it is universally acknowledged that most advances in medical
devices, or most new advanced features built into those devices, will require
the use of software. Artificial intelligence (AI) is the popular umbrella term
often used when talking about the unknowable next generation of medical
devices. Some of the features covered by this umbrella will sit firmly in the
domain of artificial intelligence, others will be content with applying more
traditional machine learning technologies (ML), and yet others just invoke

xix
xx ◾ Preface

these terms generously to claim a marketing stake in the field. AI/ML will
certainly usher in new generations of medical devices that perform increas-
ingly advanced tasks on behalf of their human masters, and – even if we are
not quite there yet – many eyes and much hope is placed on autonomous
medical software.
The development of medical software in businesses unfolds in different
ways, following different development models from waterfall to agile, but
all of these will have to be linked to international standards and regula-
tory requirements before the resulting software can be sold and used. The
IEC 62304 standard on software life-cycle management and the ISO 13485
standard on quality management are two of the cornerstones of medical-
software development today. Most medical-device companies must abide
by both to meet the expectations placed on them by their customers and
regulators. These are not the only two standards you will perhaps want to
consider – ISO 14971 risk management, IEC 62366-1 usability, and ISO 14155
clinical investigation will also inform your software operations in all likeli-
hood – but these two are the foundational standards that will inform much
of the fabric of your business in medical software.
Combining the IEC 62304 and ISO 13485 standards in a practical and
practically extensible manner will be your second question to answer,
right after, “How do I implement the correct quality management system
for my company”? Appropriately, this book builds on the earlier book
Developing an ISO 13485-Certified Quality Management System – An
Implementation Guide for the Medical-Device Industry written by Dr
Juuso (Routledge 2022). That earlier book focused on setting up a lean
and powerful QMS at a medical-device company. The focus here in this
book is on how to marry quality management and software development
together in a practical way for producing medical software – whether that
is software for use inside medical devices, software as a medical device,
or something in between. If you are building a company from scratch, we
recommend you first read the above-mentioned QMS book and then con-
tinue to the software-specific discussion with this book. If your interest
is primarily in software development, and you won’t be responsible for
the rest of your QMS, you can jump ahead to here – the QMS book offers
good context for the work here, but it may not be a must-read for you
here and now. We also expect this book to be a handy tool for software
suppliers who don’t develop medical-device software themselves, but who
are asked to meet some subset of the requirements by their B2B custom-
ers who do.
Preface  ◾ xxi

The discussion here adopts the structure of ISO 13485 quality man-
agement as its backbone and then builds on this to construct a practical
framework for medical-grade software development. ISO 13485 is the big,
beautiful standard your medical-device operations will be measured against,
and against which you will ultimately want to retain your status and any
possible official certification status. IEC 62304, by comparison, is a vastly
popular supporting standard in medical-software development and later
maintenance. It is, however, plagued by a strained, drawn-out revision pro-
cess and has had to contend with gathering storm clouds since 2006. IEC
62304 is now technically stable for the next few years, but its future is con-
tentious and may see major revision taking place. For this reason, it makes
all the sense in the world to build your software-development efforts based
on the bedrock offered by ISO 13485 and then add to this as IEC 62304
suggests.
This book was borne out of a wish to use ISO 13485 quality manage-
ment as a diving board into medical-grade software development and to find
answers to the grave questions posed to IEC 62304 over the past decade and
a half. Furthermore, we wanted to provide practical solutions to developing
medical software and the associated documentation in a smart, lean, and
organically developer-friendly way.
Our whole philosophy in writing this book is to build a sturdy, but light-
weight process for medical-software development. A process that ensures
you meet your requirements in a lean, timely, straightforward, and safe man-
ner. The process is based on our own experiences in designing, analyzing,
and assessing medical-software development processes for medical-device
companies. This work has been informed by the standard, the decades-long
revision attempts around it, and both the Medical Device Regulation of the
EU and the regulation in the US.
The specifics of the IEC 62304 standard will change, they may even
take on a completely new shape and structure, but the needs for medical-
software development will not change in any hurry. The need to elicit
requirements, work efficiently towards delivering a product, meet require-
ments with the product, and support the product over its entire life cycle
will remain constant. The need to maintain trust in what the software does
all through its life cycle will be the biggest requirement of all – this should
be your North Star or the big neon-colored buzzing sign hanging above
your goals on the office walls. That trust is essential for your development
work to proceed predictably and achieve the goals set for it, but it will also
be the overriding requirement for turning potential users into actual users
xxii ◾ Preface

and allowing them – and your regulators – to rely on your software. This
is particularly the case if your software product has an artificial intelligence
flavor to it. At a recent FDA event attracting exceptionally comprehensive
participation from patients, physicians, industry, and academia the consensus
was that transparency is an increasingly important topic to consider in the
development of medical devices. Providing all of the development data and
all of the information on the development work done will never be possible,
nor would it be useful for the end users. However, providing the right infor-
mation, the right amount of information, and presenting all this information
in an actionable form will be crucial throughout the development and sub-
sequent use of any advanced medical device. This is true both within and
without your software-development pipeline.
In the coming sections, we will align our radio antennas at the distant
twin planets of ISO 13485 and IEC 62304 to figure out an intelligent way
of welding the two worlds together. To ensure a good contact is made, just
like in the opening scene of that film by Robert Zemeckis from 1997, we
will apply both a long history of collected wisdom from regulated software
development and a keen ear for sifting through and making sense of recent
developments to navigate our path.
Simply put, the goal of this book is to extend the framework set by the
ISO 13485 standard on quality management in a way that readily supports
the development of software-based medical devices while at the same time
adding as little overhead as possible. The aim is to perform all the key
activities expected by both the quality management and the software-devel-
opment standard but to eliminate all unnecessary friction and any duplica-
tion of effort between the two domains. To think twice and do once.
The lofty goal here is to take a deep dive into IEC 62304 software life-
cycle processes, observe the key processes defined therein, identify any
overlap with ISO 13485, and then design a process model that merges the
two worlds as safely and efficiently as possible. The underlying premise is
that the two standards are fundamentally compatible and have the same
goal of ensuring safe and high-performing software products reach the
markets. Unfortunately, the two standards have not been written with any
real plug-n-play compatibility in mind, which means that forging a unified
process meeting both standards requires some careful thought – especially if
the intention is to also at the same time use a software-development model
that is more agile than gravity-bound.
The good news here is that this sort of coupling is in fact routinely
done in the medical-device business. We know that it can be done even if
Preface  ◾ xxiii

no convenient ready-made models exist to lay out the lines for the align-
ment. We know that some attempts in the past have been less success-
ful than others. This is where this book steps in, to provide you with a
thought-out practical example and an adaptation to tweak to your particu-
lar needs.
In writing the discussion in the following sections we have attempted to
distill both standards to their cores, expose the underlying requirements,
provide practical real-world examples where we thought they would help,
and forge a practical, low-friction model for conducting modern software
development in a regulated space and under industry-standard quality man-
agement. This book will give you the guidelines, sighting angles, and instru-
ments to successfully dock your software-development activities into the
rest of your organization’s quality management system. Think of the Apollo
space program, the lunar landing module, and the command module, but
rest assured that no duct tape or underwear containers will be needed here
to secure the connection between the modules.
The quality management system itself is described in detail in our previ-
ous QMS book (Juuso 2022). The emphasis in this present book will be on
extending that quality-management system to also cover software-devel-
opment activities in a smart way. The base QMS you will be working from
here could also be something different. What matters here is that you want
to fuse quality management and software-development activities together in
an efficient and safe way. The goal is to focus most on what truly matters,
to achieve everything you need to while not having to repeat yourself or
spread your resources thin.
The following sections of this book will introduce you to the ISO 13485
standard on quality management, the IEC 62304 standard on software life-
cycle processes, and a way of adopting both for agile software development
in a regulated industry. The discussion is infused with our own experience
in the field and the emerging trends that shape the future of both standards.
Our aim is to write the description within these covers in a way that is
easily approachable to all interested audiences from customers and manage-
ment to individual developers, programmers, and suppliers. We hope that
the discussion and examples provided in the text will highlight synergies,
illuminate possible issues, and allow for the informed design of unified
processes across all of the involved activities. The discussion herein is based
on our decades of experience in the design, inspection, certification, and
everyday running of software-development activities in both industry and
academia.
Acknowledgments

After the incredibly warm and encouraging reception to our first book on
ISO 13485 quality management for the medical-device industry (Juuso 2022),
it was with profound gratitude and an unquenchable thirst for knowledge
that we embarked on this second book project. This time we dove into
understanding, and making sense of, the required second layer of the puzzle
for the development of software-based medical devices: the IEC 62304 stan-
dard for software-development life cycles.
We wish to thank all of the people from around the world who have
provided their thoughts and reviews of the first book. All of your feedback
is greatly appreciated and has, in part, informed the writing of this second
book.
This time, too, the people who took time from their busy schedules and
provided us with invaluable feedback on the evolving manuscript all deserve
our gratitude. Jussi Ala-Kurikka, Tuomas Granlund, and Päivi Turta all gave
us in-depth notes on our text and made our scribblings undeniably better.
Sandra Liede and Nina Vartiainen, too, conducted a welcome sanity check
of the key lessons we wished to depart here. The comments you all gave
sharpened our focus, improved our output and added insight we would
have perhaps otherwise neglected to include. In the same vein, we also wish
to thank our respective organizations, Kasve Consulting and SGS Fimko, for
their supportive attitude toward our work here.
Finally, we acknowledge with immeasurable gratitude the understanding
and love of our families. Writing this book has not taken place without some
weekends and nights spent on Zoom calls and in front of the computer. All
of this has, regrettably, been time stolen away from our families, but perhaps
the invigorating effects of the collaboration have carried over to family life
also. We also take pleasure in the fact that some of the hundreds of hours of
conversation have taken place over pints of beer or even in the glow of the

xxiv
Acknowledgments  ◾ xxv

occasional fireplace transmitted over a Zoom line. As a quest to consolidate


past experiences and learn more about the topics involved this has been an
outstanding experience for us. For that, we make no apologies but promise
to act better in the future.
As a result of this joint book project, we have both learned a great deal,
realized we knew quite a bit beforehand, and come out with a wealth of
considered ideas on how to best combine the many standards involved in
developing medical software. We hope that we have been able to write this
book in a way that passes on some of those ideas and gives rise to new,
leaner, and to-the-point solutions at each of your organizations.
About the Authors

Ilkka Juuso has 20-plus years of experience working on multidisciplinary


R&D projects in both industry and academia. He is one of the founders of
the medical-device startup Cerenion, the Principal Regulatory Engineer at
the medical-device quality consultancy Kasve, and a post-doctoral researcher
at the University of Oulu in Finland.
His main interests are international regulatory affairs, standardization, and
healthcare business development. He has successfully led the development
of an ISO 13485, ISO 14971, and IEC 62304 compliant quality management
system (QMS) from the ground up, its subsequent day-to-day operation, and
certification by a notified body. He has had a key role in the launch of a
CE-marked Class IIb medical device based on artificial intelligence.
He has repeatedly served as a committee member and the head of
the Finnish national delegation in key committees of the International
Organization for Standardization (ISO). He is a member of the manage-
ment team for CEN TC 251 on health informatics, and secretary for its WGII
on technology and applications. He is an active member of the Finnish
Standards Association (SFS) and the industry forum Healthtech Finland. He
is also the author of the book Developing an ISO 13485-Certified Quality
Management System (Routledge 2022).

Ilpo Pöyhönen has 30-plus years of experience working on medical-device


research, development, testing, and safety and performance evaluation
including in the context of an accredited certification body. During that time,
he has performed approximately 200 software evaluations according to IEC
60601-1-4, IEC 60601-1 Clause 14, IEC 62304, and IEC 82304-1. His particular
areas of interest in this work have been the role of programmable database
systems and the development of test equipment for diverse needs. The work
has taken him across the globe and even to the edge of space.

xxvi
About the Authors  ◾ xxvii

Today his main interests are international regulatory affairs, standardiza-


tion, and the intelligent control of medical-device design processes to con-
tinuously meet the requirements imposed by, for example, cyber security,
usability engineering, risk management, and agile development models. The
use of emerging technologies, such as artificial intelligence, also holds spe-
cial appeal to him.
He has been active in research initiatives that have, for example, exam-
ined the software-development documentation required in a regulated envi-
ronment, the impact of risk management, the performance of risk analysis
itself as part of the software development life-cycle, and the reliability factors
involved in the supply of complex software systems.
He is a long-time committee member of SFS/SR301 on healthcare IT and
a sought-after lecturer on topics such as medical-device software, risk man-
agement, usability, mHealth apps, and cloud services in the context of medi-
cal devices.
Chapter 1

WHAT TO KNOW BEFORE


GETTING STARTED

The journey’s start is looming just beyond this section, but before we
embark, there are a few key things worth pointing out. This section
will list some of the most valued lessons we have learned over the
years, and hopefully provide you with welcome insight on the path
to designing the most streamlined and thought-out development
pipeline for medical software.
In the army, new recruits are first broken down to clam chowder and then
rebuilt in the image of super-soldiers once everyone is on the same page.
Something similar happens with the gluten in flour when making noodles.
The purpose of this section is not to break you down or brainwash you into
adopting standards and our views on these two particular standards, but to
arm you with the arguments you may need to convince yourself and others
on the merits of combining IEC 62304 and ISO 13485 in the development of
medical-grade software products.
In addition to the lessons set out here you will find a host of other fun-
damental lessons mostly drawn from implementing and working with an
ISO 13485 quality-management system in our earlier book Developing an
ISO 13485-Certified Quality Management System (Juuso 2022). The discus-
sion in that book will be particularly useful to you if you are now figuring
out the relationship between quality management and software develop-
ment, or even writing the standard operating procedures for software prod-
uct realization yourself. If you are interested in just understanding software

DOI: 10.4324/9781003454830-1 1
2 ◾ Medical-Grade Software Development

development itself, without consideration of the surrounding quality man-


agement context, the discussion in this present book will likely suffice.
This book will concentrate on the IEC 62304 standard that acts as the
often-quoted backbone to the development and maintenance of software
in the context of medical devices. The discussion in the following sections
will distill the essential concepts, requirements, and solutions of the stan-
dard and fit these within the wider context set by the ISO 13485 standard
on quality management. Both standards are de facto expectations in medical
devices for much of the world, and countless organizations need to adopt
and adapt the two standards into their own operations. The way to combine
the standards into a lean and frictionless setup for your organization is the
goal here. Step one is to understand both standards, and the first step in that
is to dispel a few easy misunderstandings. This section will do just that to
ensure we are on an even keel and get off to a good start.

1.1 There Is No One Correct Answer to Anything


This may sound like it pulls the rug from under this whole book, but
it really doesn’t. Understanding this one simple lesson will let you
begin to see the matrix of intentions and goals behind the stan-
dards and all regulations and to begin crafting your operations
on top of that network. Don’t get stuck looking for that one needle,
there is more than one in the haystack. In other words, the sweet
spot is wider than just one hole to hit on a golf course, as Figure 1.1
illustrates.
Not all roads lead to Rome, but more than one does. This all demonstrates
that you can adopt a standard in more ways than one. To any question you
find in the standard you can bet there is more than a single equally accept-
able answer. Use this realization to save you from looking for that nonexis-
tent single needle in a haystack, and to give you a sense of calm in knowing
your own pockets when constructing your processes according to the stan-
dards. Don’t use it as a call to only do as you will, though.
After you have answered a few such questions you will have started to
make design choices about your overall approach that probably affect your
later answers as much as anything you then read in the standards. You must
understand the standards and meet the requirements given in them, but how
exactly you go about doing this, which answers out of the pool of possible
What to Know before Getting Started ◾ 3

Figure 1.1 The sweet spot is wider than just a single hole to hit.

answers you choose for the implementation of your processes depends on


the context given by your specific environment. The one universal charac-
teristic of any resulting good development model is that it ensures you meet
all the requirements and create the required deliverables consistently, and
not only as an afterthought to how you actually want to work every day.
Otherwise, you will not be using your resources, whatever they may be, in
any optimal way, and you won’t be well-equipped to deal with any future
changes to the environment. Doing something extra not required by the
standards is not a horrible state of affairs if your resources allow it and your
peace of mind requires it, but doing one thing and then duct-taping some-
thing else together based on that is not lean, it’s not smart, and it probably
isn’t safe either.
Trying to create an ideal showcase, a perfect development process with
perfect documents to show your auditors, is not a noble goal either – par-
ticularly if your showcase does not correspond with the back office for any
reason. If the showcase you have created does not lead to a lean, efficient,
and safe software development process, or if it doesn’t jibe with your daily
operations, it is just a plain bad idea. Perfection only lasts in fairy tales – the
grind of the everyday will call that “happily ever after” into serious question
sooner or later if you’re not economical and strategic in your designs.
4 ◾ Medical-Grade Software Development

Another occasional sin here is to write abstract, lofty processes thinking


that you don’t want to accidentally require doing more than the standards
call for and thus should stick to the language of the standards as closely
as possible. This is a fine theory, but one which may fail to translate the
requirements onto a practical level. As a result, you may end up having to
reinvent the wheel every time you then invoke a process, which will surely
waste resources and might even have you coming up with some oddly dif-
ferent wheels over time. In practice, you’ll want to find a good, concrete
level of defining processes so that they are actionable, safe, and to the point.
Knowing why the processes are run is the key to finding this level. Most
everyone knows the story of Icarus, the winged boy from Greek mythol-
ogy who flew too close to the sun, but not everyone remembers that he
was actually instructed to fly high enough to not dip into the sea and low
enough not to melt his artificial wings in the sun. There is both a ceiling
and a floor for flying in that story, and the same is true for constructing your
QMS. This is also a shortcoming with some of the lesser-known standards
out there: some are much too theoretical for practical application and oth-
ers much too micromanaging to negotiate in the real world. The good news
here is that the standards discussed in this book are neither.
This book highlights the design choices you will be faced with and pro-
vides you with suggestions and possible answers based on what has worked
in the past, but the answers you ultimately choose will be up to you. Lesson
one, therefore, is that just because something is standard it isn’t necessar-
ily uniform, or all cut from the same mold. That famous Apple advert called
1984 was not aimed against standards but against unquestioned monotony.
Standards are our best chance of ensuring pieces fit together and can be
both inspected and understood from the outside in, but the way those
pieces are constructed can be as varied as life itself. In fact, DNA too can
be thought of as a standard, but the life encoded by that code could not be
thought of as monotonous.
So, how do you recognize the wrong answers within the pool of the
many possible answers? How do you split the wheat from the chaff, in other
words? Any answer that lets you dismiss some requirement that you don’t
quite understand is probably going to be a fake answer akin to burying a
timebomb under the foundations of your house or leaving a leaking water
pipe alone in an area prone to sinkholes. On the other hand, an answer that
lets you be a little lazy may not be a horrible starting point if you under-
stand the requirements and are confident that your solution achieves what is
needed efficiently, reliably, and in harmony with the rest of your operations.
What to Know before Getting Started ◾ 5

The standards call for a wealth of aspects to be considered throughout your


operations, but never will they insist on a separate, single record being cre-
ated in a vacuum and for the sake of itself. You must know how the stan-
dard relates to your operations, but the standards are not your project plans.
The IEC 62304 takes this to one extreme by stating that you should map
your processes to the requirements of the standard, but you don’t have to
adopt the processes as such.
In general, one of the best excuses for developing software in the first
place is that it lets you instruct something once and have it performed for
you over and over again without raising much more than a fingertip if even
that. Just like in developing software where copy-pasting algorithms all over
your code, working on too many things at once, or creating new separate
functions for each new identified requirement, the same is true for look-
ing to comply with the processes and deliverables described in a standard.
Think of any standard as a new comb to use in sifting through your own
operations and identifying the points of common interest to shape your
answers. Some standards you can use as inspiration and others you must
follow closely enough to reach compliance. No standard is a project plan for
a business, though. At best a standard could be thought of as the inflatable
raft that you breathe life into to cross the waters to where you want to go.
Really the choice between the possible implementations answering to any
particular requirement from any standard comes down to you feeling confi-
dent you know all the touchpoints around the issue, and that it does not feel
like you are brushing anything under the rug. Only realize that the person
coming to look under the rug at the end will be your customer, your certifi-
cation body, or your regulatory authorities – and that lives may be at stake.

1.2 Why Do We Need a Standard at All?


If there is no one correct answer, then why do we need a standard
at all? We’ve heard this asked every once in a while, and we’ve
even heard senior consultants argue a version of this occasion-
ally. Standards, though, allow us to plot a course, as Figure 1.2
illustrates.
The premise is that manufacturers will do the right thing anyway, and a
micromanaging standard just gets in the way of that and detracts from mak-
ing a better product. To a certain extent this is true, splitting attention and
6 ◾ Medical-Grade Software Development

Figure 1.2 Standards are like maps to the otherwise uncharted world.

distracting anyone from doing their work is certainly to be frowned upon


– or squished mercilessly like an unwanted bug that creeps into your code.
Doing more different things will, without fail, mean getting less of each
thing. Having to do something some way that feels stupid, silly, or just cum-
bersome is never a goal.
The problem is that “anyway”. It is arrogant to the point of extreme to
suggest that, if left to our own devices, every manufacturer would consider
all the same poignant points, solve any design choices in a compatible way,
and be able to explain what they have done using language and concepts
readily accepted by the users, partners, and regulators out there. All of this
is why standards are such practical tools: not only will they help ensure you
have covered all of your bases, but that everyone else will speak the same
language and have the same fundamental expectations. From a conformity
assessment point of view, standards make it possible to formulate require-
ments in a universal way that then lends itself to a more straightforward
assessment. In other words, we need standards to meet common expecta-
tions and to achieve a good level of interoperability everywhere where it
matters.
The use of IEC 62304 is optional, but in practice, you should be pre-
pared for long discussions and endless surprises if you opt out. Use of the
standard is expected for any software touching on or taking the form of a
What to Know before Getting Started ◾ 7

medical device. IEC 62304 is, therefore, all but required. As for some other
key standards, the same arguments apply. IEC 62304 bootstraps a few addi-
tional standards into the mix. It strongly implicates the ISO 13485 standard
for the quality management of your overall operations, even if it stops short
of requiring it (see Section 4.1). The present version of IEC 62304 takes an
even stronger stance on risk management: it casts your feet in a vat of con-
crete and throws you into the pool of ISO 14971 risk management whether
you feel like it or not (see Section 4.2). Think of the opening scene of Billy
Bathgate, the new-found clarity on life Bruce Willis gains with his new
footwear and embrace your new standardization overlords as friends. Or be
prepared to argue your case for individuality like Mel Gibson did as William
Wallace in Braveheart.
All joking aside, standards may seem like something extra to do, but
they should not be that. Standards should help you inspect your processes,
prioritize between aspects of the processes, assign resources, and improve
on the processes to better achieve your goals – all the while making them
instantly comprehensible to your stakeholders. There are tens of thousands
of standards out there, not all of which you should ever adopt, but the select
few featured in this book really are industry standards that you should give
serious consideration to. Here our focus is primarily on the two standards,
IEC 62304 and ISO 13485, which we see forming the foundations for much
of the medical-software development work today. We will also touch on a
few others occasionally but treat these references as beyond the scope of
our discussion here.
Following standards is not mandatory, but it may be in your interest to
do so. Following a standard will increase your likelihood of considering all
the poignant angles and developing a solution that is compatible with other
stakeholders. At a recent industry event, a representative of a competent
authority went as far as to imply that standards can be met selectively. This
is certainly true if you don’t plan on reaching conformity with the standard,
but instead only apply it as a source of inspiration for your operations. If
conformity is your goal, however, pay careful attention to how each stan-
dard is allowed to be sampled, what it itself and your applicable regulations
say on the matter, and what this may mean for your other related operations.
If you are in the EU, of special interest here are the harmonized versions of
each standard and the special Z-annexes made for them, if these are avail-
able. The Z-annexes of the standards will be valuable tools for you as you
look to meet regulatory requirements in the EU by utilizing the standards.
Also know that another concept, the so-called “state of the art” (see Juuso
8 ◾ Medical-Grade Software Development

2022), will be your friend if that harmonization process is lagging behind


and you want to use a newer, but already commonly accepted version of
a standard instead of the archaic version which may still be the latest ver-
sion officially harmonized with the EU regulations. Don’t do so on a whim
though, but rather after careful thought as the European system is intended
to use harmonized standards.

1.3 What Is Medical-Grade Software?


Software used to meet a medical need is naturally subject to stricter
requirements than your run-of-the-mill Angry Birds app, but how
do you know when a piece of software should be considered medi-
cal in nature? There are multiple dimensions to defining the con-
cept, as Figure 1.3 illustrates.

Figure 1.3 The definition of “medical software” is a slippery concept to nail down.
What to Know before Getting Started ◾ 9

This is a big question, one almost on par with, “what is the meaning of life”?
The line of demarcation between medical software and other software is
the equivalent of a modern-day battle line separating two belligerent camps.
It is the entrenched line between what is a medical device subject to strict
regulation and what is a mere app or device for some wellbeing applica-
tion. It is the line between what is expected in the regulated domain of
medical devices and what just about anyone can publish in an app store by
themselves.
Even the above is a gross simplification. The fact of the matter is that what
qualifies as medical software is not governed by standards, but by regulations
instead. Those regulations will change from one geographic jurisdiction to the
next and will even differ based on how the manufacturer has worded their
intended use and marketing message for the software. In Europe, the Medical
Device Coordination Group (MDCG) acting under the European Commission
has put out helpful guidance on determining what software qualifies as a
medical device (e.g., MDCG 2019-11) leaving out, for example, electronic
health-record systems that store and display the entered data without lay-
ers of machine-generated analytics in between. Similarly, the Medical Device
Regulation (MDR) itself, and particularly its Article 2, affects what software
should be considered medical and what not. In the US, the FDA has made a
wealth of guidance and thinking available on these matters via its website.
IEC 62304 takes a pragmatic stance on the matter – no doubt sheltered
by the years that elapsed between its minting and the present debate. The
standard defines medical-device software as something developed for the
purpose of either being incorporated into a medical device or acting as one
itself. The former could be called Software in a Medical Device (SiMD) and
the latter Software as a Medical Device (SaMD) in present parlance. It has
occasionally been argued that IEC 62304 doesn’t readily apply to standalone
software, and this is reflected in some of the goals of the standard (e.g., the
suggestion of synchronizing risk-management activities over both the soft-
ware and its host device or performing additional validation activities on top
of the work the standard instructs), but this does not have to be the case.
The standard today is routinely used to assess different types of software
from firmware to software as a medical device (SaMD). It is noteworthy that
even an assessment of a hardware device based on IEC 60601-1 requires the
use of IEC 62304 for any embedded software components to reach its goal
(see Clause 14 of IEC 60601-1).
Another part of the debate is whether medical devices are included in
the definition of health devices. Most experts would agree that these are
10 ◾ Medical-Grade Software Development

two opposite ends of the same swimming pool, but they would be hesi-
tant to say that wading the shallow end would set you up for surviving the
dive boards at the other end. Similarly, you might look silly and be chroni-
cally out of breath if you geared up for the deep end and spent your time
at the shallow end. Both medical software and health software may have
some aspect approaching a medical purpose or saying something about an
aspect of health but treating them the same is problematic to say the least.
The attempt to do so appears to also be one of the major factors behind the
delay of the new version of IEC 62304.
Following the same hardcore model of software development expected of
the highest-risk class of medical software when developing a low-risk health
app is not the best use of your resources – and requiring manufacturers to
do so will stop many development projects dead in their tracks. The loss
will not just be monetary, but a tangible lack in the availability of solutions
and the benefits they could bring to the health of every one of us. Viewing
your software-development operations through the lenses provided by IEC
62304, ISO 13485, and even ISO 14971 risk management will, however,
show you where the paths diverge and what you need to do when moving
between the ends of the swimming pool.
IEC 62304 makes a distinction between “software” and “medical device
software” in that the latter is software – or a software system, to be exact
– intended to be used in a medical device or as a medical device (Clause
3.12). This is compatible with what ISO 13485 considers a medical device
(its Clause 3.11). Not all software, even inside a medical device, has to be
medical-device software, though. This all depends on what your product
is, how you design the software architecture, and how you implement the
features corresponding to your intended use. The software that implements
your intended use definitely is that, but some supporting software may not
be. This much is also hinted at by the MDCG 2019-11 guidance.

1.4 Software Is Classified by the Safety


Concerns Associated with It
Tom Hanks had to put a leash on Hooch in that film to prevent wan-
ton havoc to his apartment, and any time you give Schwarzenegger
a big gun you can bet there will be gratuitous destruction. Both
Hooch and Arnie should thus come with a warning, an inflatable
sumo suit, or just some instructions for careful use. The software
What to Know before Getting Started ◾ 11

safety classification instructed by IEC 62304 is that added caution,


as illustrated in Figure 1.4.
In the previous lesson we started to lift the lid of the Pandora’s box of what
software goes into your medical device. The standard offers a three-level
safety classification you apply to the software as a whole, but also to the
various subsystems underneath. Think of this as the different colored high-
visibility vests given to your software components based on what level of
risks or safety concerns are attached to them. It’s not a license to goof off or
a radio beacon to stay clear, but it is a tool to help you allocate the neces-
sary attention and resources to where they are needed.
The ABC classification, from low risk (A) to high risk (C), is a fundamen-
tal feature of the IEC 62304 standard that changes how the requirements in
Clauses 5 through 9 of the standard apply to your software. In other words,
your understanding of at least 58% of the contents of the standard may be
affected by the safety profile of your software. In practice the figure here is
lower as many requirements are shared between the classes, but occasion-
ally the differences are significant.
The classification is not a quality stamp on your software, class A soft-
ware is not better than class C, but it does allow you to focus more of your
resources on the development of class C software where risks and potential
harms may be greater. The classification is addressed in detail in Section 4.3

Figure 1.4 It’s not quite as easy as in that nursery rhyme, but knowing your ABCs is
required.
12 ◾ Medical-Grade Software Development

of this book, but it is worth being aware of this undercurrent right from the
start.
Note also that the FDA has its own old three-class Level of Concern (LoC)
rating and a new two-class Documentation Level rating to categorize the
risk of software as a medical device. Neither categorization is directly related
to the ABC classification of the standard, although both are compatible.
Instead, if your market area covers the US, you must follow the guidance
given by the FDA, answer a set of specific questions listed in the guidance,
and arrive at a LoC rating for your software. Also noteworthy is that not only
is the LoC categorization different from the ABC classification given by the
IEC 62304, but it is also separate from any medical-device classification (e.g.,
Class I–III). The good news is that all three are compatible, even if not iden-
tical as mechanisms.

1.5 Software Is Made of Units, Items, and Systems


Here we start to lower our diving bubble, like Jacques Cousteau or
Jules Verne might have done, into the IEC 62304 standard by look-
ing at how it defines the construction of software. On the way down
to the core of the constellation of a software system we’ll encounter
items and units, as illustrated by Figure 1.5.
The standard takes an engineer’s approach to understanding concepts by
dissecting them into smaller and smaller elements until you feel like you
can’t continue anymore. The French philosopher René Descartes (1596–1650)
would certainly be proud of this drive to reduce complex objects into collec-
tions of simpler interconnected objects. Intuitively, too, this makes sense for
understanding the inner workings of software. On a first dive, though, the
sea of interlocking terms may seem bewildering. For this reason, this book

Figure 1.5 System begets an item, an item begets a unit.


What to Know before Getting Started ◾ 13

will introduce you to the essential terms and concepts as we progress, not
all at once, but when you first need them.
The first set of terms to know are software unit, software item, and
software system. These are all terms used when dissecting software and
will be covered in detail as part of the terms and definitions in Section
3.2.5. For now, it suffices to know that what we typically discuss as a piece
of software is actually a software system in the standard’s parlance. This is
the entirety of the thing you are looking at as your software, whether that
is the whole of your medical device (e.g., software as a medical device) or
just software running inside (e.g., software in a medical device). This can
then be dissected into software items on a level of granularity that best
suits the conversation or the design and development (D&D) of the piece
of software. There can be several nested levels of software items, each
more refined in granularity than the previous. Ultimately, when an item
can’t be broken down into any more composite items it may be called a
software unit. In other words, units build up items, items build up soft-
ware systems.
One straightforward way of thinking about the granularity of software
items is that a software unit corresponds to a class in an object-oriented
programming language, a software item to a library (perhaps containing
frontend or backend components), and the software system would then
be the whole software itself. This is, however, not the only way to set up
granularity. It might be cheating to say that your software can’t be dissected
into any items, to say that in your unique case, your software system is
your software unit, but many other definitions of depth may be acceptable.
If your software comprises one long legacy algorithm with a completely
flat structure to it – without even a single GOTO instruction inside it – this
might really be the case, but for most modern software that is not going to
be a defensible position. Creating superfluous structure in your hierarchy
will not be practical, but at the same time it may also be questionable if
you would actually save any effort by squishing your hierarchy unnecessar-
ily here.
The slightly mad thing in the scheme is that all these various granular-
ity levels can be called software items. In other words, a software system
can be thought of as a software item, a group of nested software items can
be thought of as a software item, and even a software unit can be thought
of as a software item. The term is thus used quite liberally to refer to some
identifiable software object. Don’t worry, this only sounds mad but is quite
manageable in practice. In a way, this is like the fourth Matrix film where
14 ◾ Medical-Grade Software Development

we have already had three installments of the story but are now told that it
was all a game, led to a video game, and can all be discussed on video via
talking about the game. It’s nuts, but it’s also certifiable and we can live with
it all.
The above amorphism is, to some extent, a relic of the evolution of the
standard. In the IEEE 610.12:1990 standard, which forms a part of the ances-
try of the IEC 62304 standard, software architecture is described using the
terms system, component, unit, and module. The system there is said to con-
sist of parts which may be hardware or software, and which may be further
subdivided into other parts. The terms used for the parts are, for example,
module, component, and unit – all of which are noted to occasionally be
used interchangeably and in varying order. IEC 62304 thus has a broader
understanding of what a system is (e.g., it can include not just software and
hardware, but even the user themselves), a more focused concept of a soft-
ware system (just software), and the same fluidity towards what parts a soft-
ware actually consists of (all of which are now called software items). The
most striking difference between parent and child here is that IEC 62304
continues to think of a software unit as a logical block that is not subdi-
vided, but it does not directly speak of the testability of this block – it does
speak of unit verification later (see Section 5.5). Understanding this legacy
here is above and beyond what is expected to develop software today, but
it does demonstrate that the relative amorphism here was not accidentally
introduced by IEC 62304.

1.6 Software Development Has a Life Cycle


The Fast & Furious saga may keep on ticking and racking up those
version numbers, but it doesn’t really evolve in any discernible
way. The frequent updates are a sign of a well-oiled machine that
knows its intended use and formula. Similarly, developing software
must be followed with the appropriate steps in its later life cycle, as
depicted in Figure 1.6, to meet all the expectations of the users, cus-
tomers, and regulators.
IEC 62304 prides itself on providing a model for the software-development
life cycle. But what is a life cycle model anyway? Is software supposed to
be conceived, be born, grow up, produce offspring, retire, and ultimately
fade away? Will the software then be reincarnated as something else? Are
the medical-device labels “single-use” and “use by date” then the most
What to Know before Getting Started ◾ 15

Figure 1.6 Not to put too fine a point on it all, but a life cycle is just that.

vicious marks of all to be placed on such living software? Until we are faced
with autonomous, self-aware software systems we won’t have to get that
philosophical. Software for now is a thing, it doesn’t have life goals to meet,
it doesn’t have an Aristotelean drive to do anything by itself. Software is not
alive until HAL 9000 greets us as Dave. The life cycle is just a way of mod-
eling the path from conception to a releasable piece of software, and the
steps along that path.
In software development, as in life, the usual stages of existence can,
however, be characterized as a path from gestation to launch and then
constant improvement during use until decommissioning. Both the IEC
62304 and the ISO 13485 standards intend to ensure you know what you
are launching out into the world, and then make sure you don’t forget about
it once it is out there. You are even expected to make arrangements for the
funeral service before the clock is started.
The IEC 62304 defines the term software development life cycle model,
but it doesn’t define a term for the entire existence of software. Clause 3.24
defines the arc of a development life cycle from requirements to release. In
this conceptual structure, you model the following:

◾ Identify the processes (incl. activities and tasks) involved


◾ Describe the sequence of the identified activities and tasks, and any
dependencies between them
◾ Identify the objectives of the activities and tasks, i.e., the desired deliv-
erables (i.e., the required result, output, or documentation of an activ-
ity as per Clause 3.6) and set the milestones for verification of their
completeness
16 ◾ Medical-Grade Software Development

In the definition above it is striking that the later maintenance of software


is not included in the life cycle and no separate entry for the software-
maintenance life cycle is made in the standard. Rest assured that require-
ments for maintenance are placed on software though, and these are
exacted through, for example, Clause 6 of the standard on maintenance
(see Section 6) and the ISO 13485 standard. In no case is it enough to just
build the software and let it float off to the sunset – unless we are talking
about a piece of non-medical software like the E.T. Atari game, in which
case no users are involved in the first place. Software deployed directly
to a landfill notwithstanding, the life cycle is a broader concept than just
development.
The ideal development processes may have a life cycle model, but
increasingly the development phase is followed by an equally important
maintenance phase. Software is no longer just built and reused, but instead,
it evolves throughout its true life cycle. This is particularly true in the case of
any methods resembling artificial intelligence built into your software. Here
the regulators will be looking at the long-term life cycle of your medical
device.
The seeming short-sightedness of the life cycle as defined in the IEC
62304 standard is somewhat baffling, but the afterlife awaiting the software
after launch is an increasingly weighty topic today. In fact, the future we
envision for artificial intelligence in software, and other medical devices that
may one day self-evolve out there on the field once launched, is placing
some pressure on the concept of the life cycle to evolve. The FDA appears
to be the frontrunner in addressing this new post-launch evolution with
initiatives such as its Pre-Cert program set to enable some evolution to occur
within predetermined limits once a device is already in use.

1.7 Everything Is a Process, and Those


Consist of Activities and Tasks
A process is the ultimate Lego block in any standard. The process
gives structure to how you move from an input to an output in a
way that achieves both some value-adding work in between and
compatibility with the other Lego blocks of your system. A process,
then, can be thought of as consisting of activities, and these consist-
ing of tasks as Figure 1.7 illustrates.
What to Know before Getting Started ◾ 17

Figure 1.7 Everything is a process.

In another exercise in terminological reductionism, the IEC 62304 stan-


dard sees all operations as fundamentally consisting of processes to run.
These are in effect the chunky large nautical shipping containers for your
work. IEC 62304 is not alone in taking this view of processes as Lego
building blocks, the ISO 13485 standard too considers value-adding work
from an input toward a desired output to be a process. Processes are then
stacked to fill your cargo ship, or in this case, instruct your quality man-
agement system (QMS) and your software development pipeline. The IEC
62304 standard goes on to instruct that all processes consist of activities,
and these in turn consist of tasks. The terms process, activity, and task are
addressed in more detail in Section 3.2, but for now, it’s enough to be on
the lookout for this triplet of terms and know that they too form a Russian
doll of sorts.
There is one small exception to the above rule of thumb: instead of a
defined process, you might occasionally have documented arrangements
to instruct the handling of some activity. ISO 13485 mostly speaks of
defined processes and procedures, but on a few select occasions it shies
away from requiring a process and instead accepts a set of documented
arrangements. These occasions are contamination control (Clause 6.4.2)
and communication with customers (Clause 7.2.3). The standard also
speaks of the importance of adherence to documented arrangements in
product realization as part of D&D reviews (Clause 7.3.5), D&D verifica-
tion (Clause 7.3.6), D&D validation (Clause 7.3.6), and the monitoring and
measurement of the product (Clause 8.2.6) – and it requires these arrange-
ments to be satisfactorily complete before product release and service
delivery take place (Clause 8.2.6). Such arrangements are also to be a part
of your internal auditing activities (Clause 8.2.4). In practice, the difference
18 ◾ Medical-Grade Software Development

between a process and an arrangement may or may not be significant for


your operations.

1.8 IEC 62304 Does Not Insist on a


Specific Process Structure
As illustrated in Figure 1.8, processes slot together in some meaning-
ful way as instructed by your QMS and the IEC 62304 standard.
That creates the system in a quality-management system or the pipe-
line for your software development activities. Here you should take
your ISO 13485 QMS as the backbone and ensure your instructed
processes also meet the requirements of IEC 62304.
The introduction of the standard (see Section 3.2) makes it clear that the
standard assumes all work to take place within both a quality management
system and a risk management system (see Sections 4.1 and 4.2, respec-
tively). Within this context, it then defines a fundamental four-process frame-
work for the safe design and maintenance of medical-device software. The
four processes it presents are:

◾ Development (see Section 5)


◾ Maintenance (see Section 6)
◾ Configuration management (see Section 8)
◾ Problem resolution (see Section 9)

The IEC 62304 standard does not force this specific model of processes on
its users but instead expects the user to map the requirements it does make
on any of the processes (incl. activities and tasks) to the life-cycle model
they have adopted. This allows for great optimization in fitting the standard
and the real world together, but it will also be a major topic of discussion
when comparing the operation of any two manufacturers – not to mention a
major undercurrent of any conformity assessment involving the standard.

Figure 1.8 The pieces of the puzzle are up to you.


What to Know before Getting Started ◾ 19

1.9 IEC 62304 Does Not Impose a Documentation Set


The standard is secure enough in its requirements that it doesn’t
much mind if you meet those requirements using a novel set of doc-
uments. As long as you can map the requirements of the standard to
your documents and processes you may christen them however you
like. The naming convention of your document is thus up to you, as
Figure 1.9 illustrates.
The IEC 62304 standard requires that tasks are documented, but the shape
of this documentation is left to the user of the standard. No names, formats,
or explicit content is specified for the documentation to be produced in the-
ory. This is a significant design feature of the standard and one which you
may miss if you skip learning about the introduction of the standard (see
Section 3.2). The standard does discuss a few essential documents, such as
the software development plan and the software-maintenance plan, in detail,
but it is happy to find the same information in a document by another name
or as part of some other documents. Depart from the suggested optional
document names at your own peril, though. At the very least you will be
the recipient of that famous Clint Eastwood squint from his westerns if you
decide that, like the man, your documents need no names.
You should also realize that each review talked about by IEC 62304, or by
ISO 13485, must be documented somewhere and thus should be represented

Figure 1.9 Is a document by any other name not still a document?


20 ◾ Medical-Grade Software Development

by a report or other record in your documentation set. Similarly, the regula-


tions applicable to you place distinct expectations on what questions your
documentation set should answer, e.g., via Annex II of the EU MDR. How
you meet these requirements is up to you, but as time is money – more
specifically all time including your auditors’ time is your money – it pays
to meet the requirements as predictably and straightforwardly as possible.
It’s not even about money, not really, it’s about knowing what it is you are
doing, retaining that information also for later use, and keeping the pool of
documentation tidy and ready for use over the years to come.

1.10 Applicable Regulations as Part of the Requirements


Any time you see the term “applicable regulations” in a standard
you should feel Judge Dredd peeking over your shoulder and see
Stallone’s muscular jaw protrude into your field of vision to force
you to think carefully about what this actually means. Standards
are a big part of meeting regulatory requirements the world over,
but they can’t give you a free pass regarding the laws and regula-
tions – especially not the local laws across the globe. Figure 1.10
illustrates the stack of expectations placed on you.

Figure 1.10 Applicable regulations are a foundational part of your requirements.


What to Know before Getting Started ◾ 21

Regulations place a great many specific requirements on your work, but


the most fundamental one of these is that someone from the outside must
be able to trust that you have attempted to create a good piece of soft-
ware that is useful for someone out there – and should that someone have
issues with your software you will address those to the satisfaction of the
regulators.
Regulations don’t generally add any specific requirements to your soft-
ware-development operations that you wouldn’t already come across in the
standards. Safety and efficacy of the produced device is a major concern that
cuts through everything, but for the most part, the magnification lens here is
on the final product itself and not the process leading to it. Regulations can
set some very detailed requirements related to, for example, device classi-
fication, device registration, and reporting, but these too do not necessarily
concern your core development process. Identifying the relevant regulatory
areas of concern may be aided by regulatory guidance given on the type of
device you intend to produce or some general safety requirements as are,
for example, laid out in the General Safety and Performance Requirements
(GSPR) of the EU MDR. Understanding your applicable regulations is some-
thing required even by the standards, and an area where understanding
your requirements will enable you to fashion the smartest software-develop-
ment pipeline, or finetune the pipeline we will be discussing in this book.
Expect the following to be topics on how regulations affect your work
around software products (i.e., your processes):

◾ Device classification
◾ Device clearance
◾ Device registration
◾ Device lifetime
◾ Document and record retention
◾ Vigilance
◾ Post-market monitoring

In general, IEC 62304 has enjoyed a status as a harmonized standard in


the EU (originally harmonized according to EU MDD) and as a recognized
consensus standard in the US. The ISO 13485 standard is not only a fully
harmonized standard in the EU (both in terms of MDD and MDR) but also
presently fast tracked for use in the US by published guidance from the
FDA. The other key standards, including ISO 14971 and IEC 82304-1 appear
to be progressing along the same path.
22 ◾ Medical-Grade Software Development

In addition to regional authorities, you may also want to keep an eye


on the International Medical Device Regulators Forum (IMDRF) and the
guidance they publish. We recently had the opportunity to take part in an
IMDRF symposium organized by the European Commission and were pleas-
antly surprised by the practical, common-sense, and no-nonsense attitude
that cut through many of the presentations given by regulators from around
the world. The takeaway message from the week was that the medical-
device world is increasingly harmonized and connected – this means syner-
gies in both the assessment of medical devices and across the surveillance
systems built for them around the world.
Marrying regulatory requirements with the development of medical-grade
software (i.e., medical-device software) is a topic discussed throughout this
book, but especially in Sections 10 through 13.

1.11 Building on the Quagmire That Surrounds IEC 62304


Rome was built on a swamp, but it ruled much of the known world
for over a thousand years. St. Petersburg, Russia, was built over
a swamp too, and the beginnings of Washington D.C. weren’t too
spectacular either when the founding fathers got started with it. All
this is to say that the IEC 62304 may not be stable for a millennium,
it may not be all that we would like it to be today, but it is the code
much of software development in a medical context is expected to
follow. You can build on it, but you will do well to consider how you
build on it. Figure 1.11 illustrates this setup for IEC 62304.
Nothing about the revision of IEC 62304 has been speedy, so let’s not call it
quicksand, but the fact of the matter is that it is an aging standard last given
a partial facelift in 2015. It is nonetheless a backbone for developing medical
software and vast numbers of developers, regulators, and certification bodies
rely on it.
IEC 62304 is a bit like the legendary film actor Morgan Freeman: you
know that he is getting on, but every time you hear his deep soothing voice
you feel like a baby and take it as the voice of God. The standard has the
same function. It is not perfect, it is not even up to date, but it is the best
thing we have for understanding how software development is required to
take place – and how it is currently taking place – in medical-software com-
panies. To discard the standard would be to leave a great many developers
What to Know before Getting Started ◾ 23

Figure 1.11 The IEC 62304 has reached a point of stable uncertainty.

without a compass, and to miss out on providing them all with a migration
path to something more modern. The future of the standard looks uncertain
and may be subject to even drastic moves, but all the fundamental needs are
described in the present version of the standard.
The ISO 13485 standard by comparison is a de facto global standard in
the wider context of operating in the medical devices space. The develop-
ment of medical-grade software should fit within this wider context. This
standard devotes much of its column space to product realization, which in
the form of Clause 7 of that standard takes up 34% of the entire standard.
This standard is a global superhighway between different jurisdictions and
stakeholders, and while not all those intersections have yet been christened,
it appears to be our best hope for a truly international approach to medical
devices.
Consequently, our suggestion is to start with ISO 13485 to build your
overall approach to quality management and then to make sure IEC 62304
requirements are met within that grander context. This way you at the
very least know how your operations match with the expectations of both
standards, and how the two are aligned in terms of software develop-
ment. Even if the wildest ideas put out there come about, and IEC 62304
is indeed fused with ISO 13485 in some way, you will be in a good place
to revise your processes using ISO 13485 as your backbone. This book
will help you to see the linkages and possible disconnects between the
standards.
24 ◾ Medical-Grade Software Development

1.12 Retaining Trust at Each Step of the Way


Is Perhaps the Most Critical of Tasks
In movies anytime you hear the question, “Do you trust me?” you
can bet that something risky is about to happen. This happens in
Blade Runner, National Treasure, Aladdin, and countless other
films where a momentary act of blind faith will end up saving
the day. Blind faith does not match well with the development,
adoption, and use of medical-grade software. Instead, evidence is
always needed in one form or another as Figure 1.12 illustrates.
In case you don’t have the budget for getting Morgan Freeman or James Earl
Jones to sweet-talk both you and your customers into living with your soft-
ware, being able to demonstrably meet expectations will be a central topic
of everything you do. This applies to you trusting the rest of your develop-
ment team, your regulators maintaining trust in you, and your customers
knowing that they are getting what they want all throughout purchasing and
using your product.

Figure 1.12 Steps to trust may be different in height, but the action is always the
same.
What to Know before Getting Started ◾ 25

At a recent FDA event on the adoption of artificial intelligence in the


context of healthcare, the special topic of building in transparency came up
as a particularly poignant goal. It will never be possible to provide all the
development data, documentation, and source code of your software to all
its potential users, have them take the required time and apply the required
acumen to assess the whole, and after that make the decision to purchase
and use your software. This would be a pipedream of the highest order,
and while easily required, never actually implementable in real life. Instead,
you have to be able to obtain and provide enough information on what is
happening inside the box for the person outside the box to stay confident
they know all they want to know and that it is all taking place as they
want.
Transparency is needed for trust in and adoption of the results of the
box, but we also need that information to be made available in an action-
able form. Overwhelming the user, any user, with the quantity or complex-
ity of information is not helping anyone. We need the right information, the
right amount of information, and that information to be presented in a read-
ily actionable form.
The need for transparency and appropriately timed information cuts
through the whole software-development cycle. The classic waterfall
model of software development meets the need of obtaining reviewed and
approved information about what has been done, which is ultimately what
everyone wants: to be able to trust the thing you are handed. It does not,
however, provide for a good real-time view into the development process
of software or account for accumulated knowledge and increased certainty
over time. Partly for this reason, other more agile methodologies are often
used in software development today and can in fact provide for an even
greater level of transparency during development. The threaded and iterative
nature of agile development will cause some head-scratching, particularly
during the development of the processes themselves, but when done right,
agile can be a win from all points of view.
The argument can be made that “retaining trust” is actually the whole
point of the existence of standards. Standards and standards-based pro-
cesses provide us with the necessary yardsticks to understand what we are
seeing, analyze how that compares to our expectations, and act accordingly.
Standards, then, are a useful tool for building in transparency both during
the software-development phase and the adoption, use, and maintenance of
that software.
26 ◾ Medical-Grade Software Development

1.13 The Relationship between Code and


Documentation is at the Core of the
Marriage, Even after a Shotgun-Wedding
Only bureaucrats care about documentation, right? Wrong. Your
regulators rely on documentation and even your customers will
need some practical documentation to be able to adopt, trust, and
use your software. Documentation is not something extra someone
else creates for your software, it should be the shadow that your
software brings with it every time it enters a room as illustrated by
Figure 1.13. That said, superfluous, unnecessarily verbose, and self-
centered documentation is something to beware.
The most basic way of checking whether your trust in someone is warranted
or not is to ask them to explain their actions and compare that with what
you know of their actions. Columbo, Hercule Poirot, Miss Marple, and just
about every Tom, Dick, and Harry P.I. out there rely on this. This is also the
way certification of those actions takes place: it’s primarily a meta-level exer-
cise based on what you document, both in terms of documents and records,
with a supporting live component inspecting what you are doing. Similarly,
the relationship between your software code and software documentation
is mortally entwined: they are intended to be two sides of the same coin. If
this is the case, great. If not, you may have a problem.
When you talk with some programmers, they almost instantly give off
airs of hating documentation. Documentation is something almost dirty,
demeaning, and certainly unnecessary to the beautiful code they could

Figure 1.13 Code and documentation should have a symbiotic relationship.


What to Know before Getting Started ◾ 27

be writing. The cliché is that the more virtuoso a programmer is the less
they want to explain how the magic in their code works. It should be self-
evident. No one wants to plan, comment, review, or especially document
the work they have done; the work should speak for itself. Only the work
does not speak. It does not answer the questions regulators, B2B customers,
and potential users have on their minds. For this reason, it is safe to assume
that at least 50% of the software-development work in a regulated context
will consist of work on more than just the code itself. Without the code the
documentation does not matter, but without the documentation the code
does not matter either – and if the code does not match the documentation
you are in a world of hurt.
We too are programmers at heart – or were once before the matrix of it
all swallowed us up and did its best to turn us into grim-faced suits with
earpieces. Programmers – like us – often have strong opinions about how
the code should look, how it should be composed to run as efficiently as
possible, and especially on what is unnecessary to it all. This occasionally
leads to some head-scratching and heated conversations when later deci-
phering code to explain it or perhaps reuse it in the next project. The use
of a good programming style guide may seem superfluous or even blasphe-
mous to some programmers, but it is a surprisingly powerful tool for meet-
ing regulatory requirements and avoiding silly mistakes (or common defects
as IEC 62304 might call these). A style guide coupled with consistent peer
reviews is a great start to developing quality software, and as a solution one
of the easiest to sell most programmers on. In a peer review, the previously
defined style guide now acts as review criteria for the assessment and works
to improve consistency and quality – and it does so in a way that can be
demonstrated to match with the company’s criteria.
This book will help you triage your documentation needs. Based on the
discussion between these covers you’ll be well-equipped to make decisions
on what documents, reviews, and checks are truly worthwhile and which
ones are just lengthening the to-do list.

1.14 Evaluation, Review, Verification, and Validation


This is the holiest of the trinities introduced by either of the two
standards, at least outside of the seeming Bermuda Triangle of the
patient, the user, and the customer – all of which may be the same
entity or multiple different ones. Both standards place expectations
28 ◾ Medical-Grade Software Development

on the ways you check up on performed work, and the negotiation


of these requirements together holds the promise of a lean pro-
cess. Figure 1.14 illustrates the complementary views offered by the
standards.
The two standards, IEC 62304 and ISO 13485, look at the same thing – soft-
ware in this case – from two different angles and want to ensure it all makes
sense according to their points of view. IEC 62304 logically looks at the for-
est and the trees that make up the forest. ISO 13485 equally logically looks
at whether your forest meets the requirements you had for a forest and the
expectations your customer has for a forest. This inspection is to take place
via testing on many levels as IEC 62304 sees it, but also for the purposes of
verification (what you wanted) and validation (what your customer needed)
as ISO 13485 expects. Luckily, both standards are in agreement on what is
meant by the term review.
As a result, you can combine the requirements from the two standards
in a great number of different ways. This is not unlike a Rubik’s Cube
where the colored squares somehow represent the requirements of each
standard. You may decide to look at one aspect, or side, at first, but you
will ultimately need to figure out what arrangement of reviewing, testing,
verification, and validation makes the most sense for the entirety of your
operations.
The terms evaluation, review, verification, and validation have distinct
characteristics, but they are also sometimes used interchangeably or with
an undefined overlap. For the purposes of developing medical software, the

Figure 1.14 Verification and validation have similar goals, but different points of
view.
What to Know before Getting Started ◾ 29

definitions may come from either standard, the ancestral ISO 9000 behind
ISO 13485, or in the absence of a definition in any of these standards, from
the generally accepted dictionary definition. Based on this lineage the fol-
lowing key terms are defined:

◾ Evaluation
This term is not defined by ISO 9000 or ISO 13485, but it is defined
by IEC 62304 (Clause 3.7) as the systematic determination of the extent
to which an entity meets its specified criteria. The ISO 9000, too, uses
the term in several contexts (e.g., Clauses 2.3.3.4, 2.3.4.4, and 3.11.9) to
indicate comparison against criteria. Thus, it is fair to conclude that an
evaluation requires previously set criteria before it can take place.
◾ Review
This term is not defined by ISO 9000, ISO 13485, or IEC 62304. Thus,
the standard dictionary definition of assessing something with the inten-
tion of acting on any discovered shortcomings applies here. In other
words, activities to assess verification or validation may also be said to
be reviews. In contrast to an evaluation, no previously set criteria may
be needed in a review.
◾ Verification
This term is not defined by ISO 13485, but it is defined in detail by
ISO 9000 which is thus the definition also required by ISO 13485. In
ISO 9000 Clause 3.8.12, the definition of verification is given as con-
firmation through the provision of objective evidence (i.e., evidence
on the existence or verity of something) that specified requirements
have been fulfilled. This evidence is noted to come from, for example,
inspection, measurement, testing, alternative calculations, or a review of
documents. Note also that the term “specified requirement” is tied to a
requirement stated in the “documented information” by ISO 9000 (see
Juuso 2022) although the term “requirement” would have a much wider
definition. Further notes in ISO 9000 explain that verification is some-
times referred to as a qualification process and that the word “verified”
is used to designate the corresponding status. IEC 62304 Clause 3.33
agrees with the definition made by ISO 9000, reiterates the above note
on the use of the term, and chooses to explain that in D&D the pro-
cess of verification concerns the examination of the result of an activity
for conformity with the requirements for that activity. The reference to
D&D is an oddity in the sense that this is the only time the body of IEC
62304 uses the term, while in ISO 13485 it is very common.
30 ◾ Medical-Grade Software Development

◾ Validation
This term is not defined by ISO 13485 or IEC 62304, but it is defined
in detail by ISO 9000 which is thus the definition also required by ISO
13485. The definition of validation is given as confirmation through
the provision of objective evidence that the requirements for a spe-
cific intended use or application have been fulfilled. Note here that the
term “specified requirement” is not used, but instead, the generic term
requirement is preferred allowing for a much wider base of analysis.
A note to the definition offers, for example, test results, performing
alternative calculations, and reviewing documents as sources of that
evidence. As with verification above, the word “validated” is said to be
used to designate the corresponding status. Interestingly, the use condi-
tions for validation are here allowed to be real or simulated.

Keep the above definitions in mind if you find yourself wondering what
a review is, what additional requirements are imposed for verification and
validation, and how the three are related. Remember that all reviews should
meet the requirements stated in ISO 13485 (see Section 5.9.1), but additional
requirements from both standards apply to your body of verification and
validation work (see Sections 5.9.2 and 5.9.3).
The most foolproof way of meeting all the requirements is to sandwich
the activities by first seeing to IEC 62304 reviewing, testing, and verifica-
tion requirements in a way that is forward-compatible with ISO 13485
expectations, and then consulting ISO 13485 to perform a summative
assessment and ensure nothing is left behind. In this book, we adopt this
model and run these activities first according to IEC 62304 and then look
at the output review, verification, and validation before moving ahead with
the release.
Alternatively, you could take the verification, and, at least to some extent,
the validation requirements ISO 13485 has and infuse them into the testing
activities of IEC 62304. Clause 7.3.2c of ISO 13485 seems to suggest this by
requiring that your D&D planning documents the verification, validation,
and design transfer activities appropriate at each D&D stage. This might lead
to the most high-performance process where the output was the closest to
being released with the minimum amount of after-the-fact checks needed on
top of the IEC 62304 software-development process. Most likely, though, you
would still end up with partially fused verification activities and a somewhat
separate validation activity – especially as validation is required to be per-
formed using a representative product (e.g., initial production units) and not
What to Know before Getting Started ◾ 31

some incremental development candidate still evolving on several fronts.


Both verification and validation activities are also required to observe any
needed connections to other devices.
These activities are discussed in more detail throughout the book, but
especially in Section 5.9.

1.15 How Many Reviews Is Enough?


Are you an effective team? That is the question posed to Tom Cruise
and Victoria Riseborough every morning by their boss in the film
Oblivion. This might be the leanest example of checking up on previ-
ous work and getting a feeling of the comfort and confidence level
of your employees. The medical-device standards require you to ask
a few more questions on the outcome of the work done during each
stage, whether that is on the level of planning your development
project or reporting on its outcomes. The number of reviews is often
a case for some deep thought as Figure 1.15 illustrates.
A central feature in all standards, and all planned activity really, is that the
plan for doing something must be approved before that something is done.
Coming up with a plan once you’ve already done something is self-evidently
silly and you will be fooling yourself if you think you have done more than
recorded the path you happened to take. This does frequently happen out
there, though. The Plan-Do-Check-Act model invented by W. E. Deming
in the 1950s – although he emphasized studying instead of checking – is
invoked in the ISO 13485 standard as generally accepted good practice: you
plan, you do, you check whether the actions and outcomes meet your plan,
and then act accordingly. This should be your default mode of addressing
just about anything in your business as time allows. But how does this trans-
late to medical-software development?
ISO 13485 and IEC 62304 set some expectations for the activities you
need to carry out. It may occasionally be debatable whether these activities
need to be carried out according to a predefined process or a plan, or just
performed according to your general arrangements, but it is never acceptable
to not take responsibility for the results or somehow check on the output
produced. This applies equally to the requirements you set at the start of
a development project and the outsourced components perhaps going into
your product through your assembly line. Here we are talking of a key type
of evidence you will need to run your business and keep churning out the
32 ◾ Medical-Grade Software Development

Figure 1.15 “Review vertigo” may lead to more fatigue than attention to detail.

products your revenue depends on, but we are also talking about a key type
of evidence you will need to show to your auditors whether those are cer-
tification bodies, notified bodies, or your B2B customers auditing you as a
supplier (see Juuso 2022).
The question of how many reviews is necessary is very much up to you.
Performing no reviews is not acceptable in any context other than a fai-
rytale. Performing one review might be theoretically possible as long as you
lock your plan documents in some airtight way and are happy taking a risk
that your later actions then either meet or fall short of these plans and any
acceptance criteria defined therein. In practice, this would be foolhardy and
might fall apart as a solution the moment you would need to adjust your
plans. On the other hand, performing a gazillion reviews is not feasible
or probably beneficial either. So, the question remains, how many reviews
might be optimal?
What to Know before Getting Started ◾ 33

In this book, we look at the stages of software development in-depth


starting from Section 5. The stages therein will provide you with a solid
framework for structuring your activities and identifying the plans that need
to be approved before moving on to the next stage. Within the stage too
there are activities that you will want to plan before executing them. In
some cases, it will be enough to refer to your standard operating procedures
(SOPs) and acknowledge that these will be followed in your activities. In
other cases, you will want to address project-specific aspects in planning
some activity, and thus modify your general procedure within the limits your
SOPs allow.
In general, the minimum level of specific review steps to address in your
development model is three. These are:

◾ Design Input Review (DIR)


This is where you assess the status of your inputs, especially the
software-system requirements you have formulated based on all the
information presented to you and available to you. Here you will have
to listen to your customer, the applicable regulatory environment, and
anything else you should know before getting started with the develop-
ment project. The direct harm from fumbling with requirements will be
that you waste resources in your project, but the ultimate harm may be
that you misunderstand some customer requirements and create a less-
than-perfect or potentially dangerous product. The DIR is discussed in
more detail in Section 5.2.
◾ Design Output Review (DOR)
This is where you assess the output of your design and development
work. The objective here is primarily to check that your outputs and
deliverables (incl. documentation) are as you planned. This review sits
conveniently at the place where IEC 62304 is mostly finished with your
product-development work, although software maintenance, risk man-
agement, and problem resolution will, for example, always continue to
be topical. The DOR is discussed in more detail in Section 5.8
◾ Design Transfer Review (DTR)
The last of the three big reviews is mostly a topic for ISO 13485. Here
you must ensure that you are happy moving your software from devel-
opment to production. The DTR is discussed in more detail in Section
5.8. This review may sometimes be called the Design Readiness Review
(DRR) or even the Transfer Readiness Review (TRR), but the intention is
the same.
34 ◾ Medical-Grade Software Development

The above-listed reviews are what you are expected to have a good
handle on. It might not be uncommon to have half a dozen or even close
to a dozen different reviews marked in your development model. The
standards do not require any specific number of reviews or even specific
names for the reviews, but they do want to see you plan and assess your
work.
The specifics of what deliverables are reviewed in what review, or espe-
cially what lower-level reviews these major-ticket reviews may be based on,
will vary. By going through the individual development stages later in this
book you will be able to fashion your optimal process and frame that dis-
cussion in the terms expected by outside stakeholders.
A poignant question related to the big reviews is, “When should they be
redone”? The simple answer is when the reviews may need to be updated
for one reason or another. If the DIR defines input requirements but identi-
fies some set of requirements as cumulating over the development project,
or the risk-management file as iterated over time, does it then need to be
redone if a new requirement is added or the matrix of identified risks is iter-
ated as more is learned over time? What if a software change leads to a clar-
ification of the wording of some label implicated in risk control? The answer
is that the reviews should be revised any time they no longer accurately
reflect the software-development work. If on the other hand, the change-
management process arrives at a conclusion that the change does not affect
the reviews, these probably should not be revised. The change-management
process taking effect after the release of the product, if not before, should
already cover the necessary reviews and thus be enough without redoing
DIR, DOR, and DTR in a simple case.

1.16 How Much Testing Is Enough?


This is a question that almost invariably comes up at one point or
another during the examination of any medical-device development
path. Less testing naturally equals savings in the short term. If you’re
not careful, however, this time and money may be borrowed from
the future, and the interest to pay may be steep when it needs to be
paid. The purpose of testing is not to tick a box, but to have built up
all the necessary confidence to either tick or not tick that box. This is
not unlike walking out in front of a live audience after testing your
standup skills on your friends, as Figure 1.16 attempts to point out.
What to Know before Getting Started ◾ 35

Figure 1.16 Testing is required to trust that it all works.

The answer to the question of what is enough is infuriatingly simple: it


depends on your product, its requirements, and how comprehensively
your tests look at meeting those requirements. If there are any significant
omissions or gaps in any of this, it will be difficult to make up for any
shortcomings in one area with the extraneous testing you may have done
in some other areas – so it is even possible to do both too much and too
little at the same time. Before you throw this book at the wall, let that sink
in. It means that you get to decide what needs to be tested for you to feel
confident that all is as planned and expected. Now, before you get too
arrogant you should remember that the requirements also include all the
relevant expectations from regulations and standards, in addition to the
features you and your customer jotted down on that napkin. Remember,
too, that your regulatory authorities need to then also agree with your rea-
soning here.
It is thus possible to approach testing from two opposite angles that
despite their differing points of view both have the goal of producing a
better, safer product that meets the expectations. Firstly, IEC 62304 looks
in-depth at the requirements and the software items built to those require-
ments, starting from units all the way up to the integrated software system.
Secondly, ISO 13485, and regulations in general, look at your whole set of
requirements (incl. applicable regulations and customer expectations) and
36 ◾ Medical-Grade Software Development

your evidence for meeting those requirements with your final device. The
requirements for testing from these two standards can be summarized as
follows.
Starting from the point of view of the finished product, ISO 13485
expects the following:

◾ You have, in due time, defined what is required for the testing of the
product and its acceptance criteria
◾ You have completed that testing prior to releasing the product and you
now maintain traceable records to provide the related evidence
◾ The records include the identification of the testing equipment and, at
least in the case of an implantable device, the identities of the testers.
For the purposes of traceability and reproducibility it will make sense to
retain a bit more information than required here, but the choice will be
yours on what information to retain for these purposes

From the point of view of building that product, IEC 62304 expects the
following:

◾ The software requirements are expressed in a form amenable to the


creation of test criteria and the performance of tests (see Section 5.2)
◾ Testing is done to verify the implementation of software units unless
this verification is performed by other means (see Section 5.5)
◾ Software-integration testing (see Section 5.6) and software-system test-
ing (see Section 5.7) are done either separately or together
◾ After testing, all requirements have been tested or otherwise verified

Note that IEC 62304 expects even legacy software to have, as the bare
minimum of deliverables identified by its gap analysis, a set of software-
system testing records (see Section 4.4.2). See the referred sections for notes
on how the software safety classification may influence the above-listed
expectations.
These two standards are hard to avoid if you are reading this book and
working on medical software. Beyond the two you must take a long hard
look at your product to know what you need to test. Knowing if you are
developing code for a pacemaker, an electronic health-record system, or a
fitness tracker will affect your consideration here. So too will the user popu-
lation, the use environment, the intended use, and everything else you think
you know about your product and its future use. These characteristics will
What to Know before Getting Started ◾ 37

also affect your interpretations of the two standards already mentioned, of


course.
Usability (see Section 12.5) is often a big-ticket item here and it may feed
into your clinical validation activities sooner or later. The main standard for
usability is IEC 62366. Another usual suspect is electrical safety based on the
IEC 60601-1 family of standards. A recent important extension of safety is
cybersecurity, which does not yet have a convenient single standard to refer
to (although see Section 12.6) but is on everyone’s lips from the regulatory
authorities to the users. External testing services are available for each of
these and much more.
For basic testing work all that is required is that you roll up your sleeves
and get to work. You will still want to ensure people are not approving their
own work, but mostly this is an in-house issue. For some other types of
testing, such as cybersecurity testing, the involvement of an outside body is
strongly implied. It is also worth pointing out that anytime your regulatory
authorities see you not clear of the height they expect, whatever the topic
of the shortcoming, you will be required to do more work to try the jump
again or stay on the field. At some point they may lose faith in your abilities
and either ask you to start again or ask you to bring in an external expert
to assess the status. An example of this would be external testing against
the IEC 62304 standard which may not be needed if your processes and
documentation obviously meet the requirements but will be required if they
don’t.
Whether or not you are compelled to use external testing providers, you
will still benefit greatly from having an outside expert come in to assess
your processes in a lower-stress context than an actual device inspection
and provide you helpful advice on how to both improve and streamline
your processes. If you chose this expert wisely, by sticking to accredited
testing providers, this work will be time away directly from your ultimate
regulatory audits. If your choice is hard to justify, your notified body may
throw out the testing work, possibly audit the testing body, redo testing in-
house, or require you to have it redone. All of this will cost time and money.
Thus, testing is primarily an exercise in building up the required con-
fidence in your own product. Once you are confident, you are done. The
notified body will then test your confidence by asking about the testing you
have done and see if it all holds water. This too will be a test, but it will not
be a part of your testing activities per se.
As a final anecdote on why testing is such an important and often trou-
blesome area, consider the fictional requirement of, “The device must read
38 ◾ Medical-Grade Software Development

input from a keyboard”. Here if the user types in “ABCD” and the program
accidentally reverses that while reading it into “DCBA” the requirement is
still met. No user would be happy with that outcome, but technically the
test and its ill-conceived test criteria could be satisfied. Think about this
when you work on defining tests and don’t just go for the quickest answer.
Defining tests is not just a chore to do – and fixing badly worded require-
ments discovered during the definition of tests will cause headaches of
its own.
Testing is often a source of problems when performing IEC 62304 assess-
ments. Defining requirements that are stated within this realm and that
don’t verge on the abstract or the superfluous, will help keep the later test-
ing work manageable. Occasionally it happens that requirements are later
discovered to be untestable in practice – and occasionally it even happens
that this minor detail has purportedly not prevented the conducting of tests
which are then well-documented despite not being possible. It goes without
saying that both are occasions where conformity with the standard is called
into serious question. Make sure you approach tolerances and acceptance
criteria with all care – and do so before you conduct the testing itself.
Testing is usually a planned activity where the testing, test cases, and
acceptance criteria are set during the planning stage. The key question
is often not how many tests are performed, but by whom the tests are
performed. Letting programmers test their own code is a natural part of
development work, but in terms of conducting final testing they may be
too lenient towards their own work or too familiar with the correct use
of their software to tease out any issues. In recognition of this, the more
hardware-oriented standard IEC 60601-1 Clause 14.10 even goes as far as
to request documentation of the appropriate level of independence for
the personnel performing verification activities (which may include test-
ing). IEC 62304 does not make a similar requirement, but nonetheless
using testers independent from the developers is known to produce good
results.

1.17 Fail and Pass


An IEC 62304 conformity assessment is probably on your road-
map to releasing medical-grade software. You'll want to pass the
assessment rather than fail it, which are the two choices shown in
Figure 1.17. No one sets out to fail their conformity assessment, of
What to Know before Getting Started ◾ 39

Figure 1.17 To pass or to fail, that is the Shakespearean question in software.

course, but understanding that there are some requirements where


shortcomings will cause an immediate failure, and others where a
failure may just need to be addressed in due time, may help you
pass the assessment in a more straightforward fashion. This les-
son points out a few of the “ fail fast” requirements built into the
standard.
In the IEC 62304 conformity assessment, your software-development activi-
ties are evaluated against each requirement of the standard, one at a time.
The status of each requirement is then evaluated as a verdict using three
possible values.

◾ Not applicable (N/A)


Meaning that the requirement is not applicable to your operations.
A requirement may not be applicable to your case based on the soft-
ware safety classification (e.g., the requirement only applies to classes B
and C and your software is class A) or because it has been labeled “as
appropriate” and you have a documented rationale for why it doesn’t
apply (e.g., your software does not provide any user interface and can’t
therefore meet Clause 5.2.2f requirements; see Section 5.2). Note that
the rationale for the applicability of the requirements will be recorded
in the results/remarks section of the official test report.
40 ◾ Medical-Grade Software Development

◾ Pass (P)
Meaning that your operations meet the requirement. For example, a
verification plan is expected for the software, and the manufacturer has
provided a plan that also meets all the content requirements for such a
plan. The requirement will thus be judged as passed.
◾ Fail (F)
Meaning that your operations do not meet the requirement. For
example, a verification plan is expected for the software, but no such
plan is found.

In practice, the most challenging part of the assessment is figuring out if


the manufacturer meets or does not meet a requirement that applies to
them. This watershed point is occasionally the source of some passionate
debate. This is also an area where the standard falls a little short as it does
not provide any guidelines or acceptance criteria for when a requirement
may be judged as passed and when it must be judged as failed. Instead, the
standard describes the ideal situation and then relies on the expertise of the
assessor to determine what is close enough in the real world. As a result,
the acceptance criteria may occasionally be subjective and dependent on
the assessor.
There are, however, some requirements where a failure will cause
an immediate stop to the conformity assessment. These fail-fast require-
ments are mostly addressed in Clause 4 of the standard (see Section 4) and
address fundamental expectations of the standard. The top examples here
are:

◾ Software development work must take place within an ISO 13485 qual-
ity management system
◾ Software development work must incorporate ISO 14971 risk
management

A few typical lower-level fails that are also easy to spot might be:

◾ One of the key planning documents required by the Software


Development Plan does not exist (see Section 5.1)
◾ No acceptance criteria have been defined for software testing (see
Section 5.1)
◾ The software risk-management activities do not cover the use of SOUP
components although these are used by the software (see Section 7.1)
What to Know before Getting Started ◾ 41

The above examples are somewhat Boolean in nature – or Shakespearean


as in to be or not to be. The previous shortcomings may be simple to spot,
but that is not the case with most other requirements in a real-world context.
Examples of shortcomings that will be harder to spot and to assess might
include:

◾ The verification of requirements does not adequately address the disam-


biguation of requirements, their mutually contradictory nature, or their
readiness to be tested (see Section 5.2)
◾ The safety and performance of the medical device have not been ade-
quately tested (see Section 5.7)
◾ The risk management of the medical device does not cover all relevant
risks (see Section 7)

The standard also covers some challenging requirements that often benefit
from a synchronization of expectations between the manufacturer and the
assessor prior to the assessment. Examples here include the handling of
SOUP components based on the software safety class, erroneous specifica-
tion of functionality as a potential cause of harm (Clause 7.1.2a; see Section
7.1), and what it means to reverse adverse trends during software problem
resolution (Clause 9.7b; see Section 9.7).
As mentioned earlier, the standard does not contain nor is it accompanied
by guidance on what solutions may be considered as meeting the require-
ments of the standard. The non-binding Annex B of the standard does
attempt to provide some further friendly guidance on the requirements and
what they actually mean. The description in the appendix is brief, non-bind-
ing, and void of sufficient examples to truly foster a uniform understand-
ing of the requirements across all stakeholders. Hopefully we will at some
not-too-distant stage see a good technical report or other guidance docu-
ment introduced to provide some practical framework for acceptance criteria
and a set of actionable examples to assist in the assessment of each set of
requirements.

1.18 One Step Forward, Two Steps Back


The standards are very good about requiring that you always keep
everything up to date, but how do you accomplish this without con-
stantly going back to update documents for every small thing? One
42 ◾ Medical-Grade Software Development

Figure 1.18 Maintaining forward momentum may at times be difficult.

step forward and two steps back is not a good model for a process,
as Figure 1.18 illustrates.
In our earlier book focusing on ISO 13485 quality management (Juuso
2022) we noted that updating documents to obtain new timestamps is just
all kinds of silly. Similarly, updating plans to keep track of progress and the
produced deliverables is counterintuitive – except for when your plans really
do change. The software-development life cycle sketched out by IEC 62304,
too, is littered with such back-and-forth connections between deliverables
and development stages. As a result, a development process written merely
to comply with the standards may be far from lean if you are not careful. If
you take an analytical approach to what the underlying goals of each activ-
ity are and how best to ensure those goals are met, you will find that the
standards do lend themselves to the design of lean processes.
Arriving at lean processes that move ahead reliably and predictably is no
small feat. The historical and organizational baggage of your organization
will weigh in heavily here, but so too will any best practices and practical
conventions you have developed over the years. If you are starting anew or
looking to overhaul your processes to something leaner the following design
goals may help you get there.

◾ Take a long hard look at the processes you are accustomed to, and how
these compare to what is expected by the standards and regulations.
Don’t just create something totally new that is so revolutionary no one
will want to learn it or follow it. A QMS is not a folder of aspirations that
you bring out for the auditor to marvel at, it should be your optimal way
of life: your way of doing things so that you meet all the requirements.
◾ Define your generic, overall processes in the SOPs and refer to them in
any plans you need. This should include your basic approach to testing,
What to Know before Getting Started ◾ 43

verification, and validation, too. Having one central location to refer to


will simplify the maintenance of links with any related QMS processes
(incl. staff training and customer communication), but it will also sim-
plify the actual running of those processes.
◾ Add project-specific details in a software development plan. This should
include how this project is expected to be different from the generic
project discussed by your SOP. Your project must still fit within the
parameters allowed by your SOP, of course.
◾ Don’t jump ahead to micromanage things. Instead, record important
things to address during a later stage and iron those details out at the
appropriate time. An example here might be to attempt defining execut-
able tests for all the requirements when you are not yet sure of your
final set of requirements – instead verify your requirements as expected
by the standards (see Section 5.2) and then later define the tests before
testing takes place.
◾ Avoid copy-paste. If you find yourself entering the same information in
several documents, consider using references instead and take a long
pause to figure out when the particular piece of information or instruc-
tion should actually be chiseled in stone and made available.
◾ Don’t update documents just for the sake of new timestamps. Just don’t
do it. It makes no sense and causes more issues downstream as people
will have to figure out what has changed and how it affects their work.

This book will help you fashion a streamlined process for your software-
development activities. The discussion from Section 5 onward will increas-
ingly discuss the requirements of the two key standards, ISO 13485 and IEC
62304, and guide you toward making the right choices for your organization.
The ultimate expectation of any industrial reading of the standards is that
by following the map they provide you will be sailing ahead towards your
destination in a smart, efficient, and predictable way making at least discern-
ible constant progress. Not having tailwinds to carry you all the way may be
expected, but rowing in the wrong direction, or backpaddling just to record
your recent progress, does not make much sense.

1.19 Continuous Improvement
On the previous note of maintaining forward momentum, another
perhaps related essential concept in the standards is the notion of
44 ◾ Medical-Grade Software Development

assessing where you are and then moving ahead in the right direc-
tion to improve the status quo. A QMS should be getting better over
time, as Figure 1.19 illustrates.
Years ago, in the heyday of the American automotive industry, a high-up
executive was quoted as saying that to have more than one feature which
sets you apart from your competitors is wasteful. The implication here was
that to excel in more than one aspect would be superfluous as the cus-
tomer would already buy your product based on the first aspect. This notion
appears drenched in hubris and is, today, antithetical to what modern qual-
ity management and standards like ISO 13485 want to see.
It does not make sense to constantly adjust processes, but to not measure
your processes somehow and to not know both where you are and where
you want to be is foolish. The father of modern quality management, W. E.
Deming, pointed as much out when coming up with his 14 points to guide
company management through quality management. You should, however,
always be on the lookout for areas of possible quality improvement, and the
ways you could then implement to achieve those improvements. This vigi-
lance and improvement should be continuous and over time lead to vastly
superior processes.
The ISO 13485 standard in particular wants to see your whole organiza-
tion working to improve upon the status quo. Small steps applied over a

Figure 1.19 Your path into QMS should have a rising crescendo.
What to Know before Getting Started ◾ 45

long time will get you to a much better state of affairs over time than a few
dramatic and perhaps ill-conceived actions. This applies to the management
of your organization, the management of quality throughout your organi-
zation, and it also applies to how you can approach the development and
fine-tuning of your software-development processes. What you have now
does not have to be perfect, but it does have to be good enough and worth
improving.

1.20 What Is SOUP?
Soup du jour? SOUP is a special concept in software development
and one which is a little more than just the chowder served at your
local establishment today. SOUP and soup have more in common
than just the letters, as Figure 1.20 illustrates.
Admit it, you instantly thought of that Andy Warhol painting of Campbell
soup? That’s not a terrible analogy here, as it turns out. Much like the
painted can of tomato soup in the picture, which you can only imagine con-
tains tomato soup inside, you have to trust that the contents of some piece
of outsourced and possibly expired software does what it was intended
to. That’s what the standard understands with the acronym “SOUP” which
stands for Software of Unknown Provenance.
SOUP is software that is generally available, developed for some other
purpose than going specifically into your product, or the development

Figure 1.20 SOUP or not, you'll need to know what’s in it.


46 ◾ Medical-Grade Software Development

process of which you might not have all the adequate records. The whole
software system itself can’t be SOUP, but it can be what is known as “legacy
software” with somewhat similar repercussions (see Section 1.22).
SOUP is a fundamental concept to software development and one which
you may already be familiar with. The topic will be brought up frequently in
the discussion in the following sections.

1.21 Configuration Items and Controlled Changes


It is intuitively only appropriate that something painstakingly
designed, reviewed, and approved can’t be changed in any hap-
hazard way thereafter. If changes were not addressed with rigor, the
best of devices could be undone over time and at the very least their
pedigree could be called into question. Part of being able to control
change is to have mechanism for measuring and labelling it, as
illustrated in Figure 1.21.
Having periods of blackout, or uncharted history, is unacceptable from an
auditing point of view – unless we are talking about legacy software (see
next lesson) – but the blackouts would also make it more difficult to prog-
ress with the design of your products later on with any degree of certainty.

Figure 1.21 Configuration covers both versions of your product and what is in it.
What to Know before Getting Started ◾ 47

The first step to controlling changes is to identify the states across


changes, i.e., the versions of the item changed. This is done through
what is called configuration items by IEC 62304. The process for identify-
ing such items and controlling changes to them is discussed in detail in
Section 8. The choice of what items to control is on one hand simple: you
want to control all the parts of your medical-device software. On the other
hand, it may not be quite that simple as the documentation of that soft-
ware and the data used by the software may also need to be controlled. In
general, any changes that may affect the conformity of your device with its
requirements need to be controlled regardless of if these changes apply to
the source code, documentation, models, data, or databases related to your
software. See Section 8.1 for more discussion on the use of configuration
items.
While on the topic of configuration management, it is pertinent to discuss
the related topic of traceability, which is a core concern in medical devices.
Regulations and the ISO 13485 standard, too, place requirements on how
you enable the identification of medical devices, and how you trace discov-
ered or otherwise investigated issues to their potential sources. Thus, trace-
ability is something you are required to do, but done right it can actually be
a competitive advantage for your organization: something which allows you
to investigate issues and answer design questions in a smarter, quicker way
during development, too. In this vein, and the part of most interest to IEC
62304 in this context, is traceability between various D&D deliverables or
items, such as:

◾ System requirements, software requirements, software-system testing,


and risk-control measures implemented in software
◾ Software items and hazardous situations B C
◾ Software items and risk software causes B C
◾ Risk software causes and risk controls implemented in software B C
◾ Risk controls implemented in software and their verification B C

In the above, superscript lettering denotes the software safety classes (i.e.
A, B, and C) the traceability requirement applies to. Also note that, as per
IEC 62304, the term software item is to be understood as software system,
software item, and software unit as is appropriate in each context.
The D&D documentation (incl. systems used to provide traceability)
must enable the above traceability as required by ISO 13485 and IEC 62304.
Additionally, it should be noted that risk management as per ISO 14971 has
48 ◾ Medical-Grade Software Development

further requirements on traceability between risk-management activities and


deliverables (see Section 4.1.4.5).

1.22 Legacy Software Has a Dedicated


Pathway Through IEC 62304
The term “legacy” perhaps conjures up images of dynasties and
foundations keeping alive the heritage of the tycoons who estab-
lished them decades ago, but in the standards, it refers to something
which is not easily discarded even if it isn’t the focus of intense
interest today. Doing away with such software might shutter some
devices and services users still rely on but updating everything to
modern standards would require an oversize investment on the part
of the manufacturer. To keep such software humming on in a reli-
able way the standard provides a special pathway, as illustrated by
Figure 1.22.
If that can of tomato soup was painted by you yourself or is by now only
stocked by your private cellars, it may more accurately be called legacy
software. The concepts of SOUP and legacy software are distinctly different

Figure 1.22 The IEC 62304 has a charted path to conformance for old software.
What to Know before Getting Started ◾ 49

but nonetheless related. Neither necessarily meets the full IEC 62304
requirements.
Legacy software is a term used for software built earlier that we still want
to use but that has not been developed according to IEC 62304. The 2015
amendment of the standard addressed just such legacy software and pro-
vided a new abbreviated pathway to achieving conformity with the standard.
Just like the software safety classification mentioned above (see Section 4.3
for an in-depth discussion), classification as legacy software means that how
the various clauses of the standard apply to the software is changed. This
matter is addressed in detail in Section 4.4, but for now, it’s enough to know
that legacy software has a special pathway of its own through the standard.
The point of legacy software is not that you are too lazy to upgrade your
code, or that you want to squeeze out the last drops of value from an aging
piece of software. That would be a dubious proposition, perhaps similar
in outcome to what happened when a good concept from the 1960s – a
low-ground-clearance airplane that could be loaded and unloaded from the
ground without any special equipment – was set to go out on the pastures
but was instead given one last tweak – in this case to accommodate larger
modern fuel-efficient engines: the whole veered off its goal and crashed
to the ground. The point of legacy software is to keep something available
that you know works, not tweak it unnecessarily, and not rob the users of
the benefit they have been enjoying. Legacy software is not a great plan for
the future, but it may be a good way of keeping around something already
widely used.
As a curiosity, also note that what the EU MDR defines as a legacy
device is again different to what IEC 62304 understands by the term. In
EU MDR parlance, a legacy device is a device put on the market before
the MDR legislation took effect, and that remains on the market after the
date of application of MDR (EU 2017/745). If this sounds foreign to you,
forget about it, as Anthony Soprano would say. If you are only now start-
ing work on developing and releasing medical-device software this does not
affect you.

1.23 Software Always Fails, and Other


Software Risk Beliefs
Risk management is the topic of hundreds if not thousands of
books written on management in general, and a good number
50 ◾ Medical-Grade Software Development

on medical devices in particular. For our discussion in this book


understanding the basics of risk management is enough. If you hap-
pen to be familiar with the ISO 14971 standard on risk management
for medical devices all the better. Figure 1.23 provides a filmic refer-
ence for risky business.
Risk management is a complex topic best left to the ISO 14971 standard, and
the countless books written on the subject, to explain. The IEC 62304 dab-
bles in risk management with good intentions, but not always successfully.
Your ISO 13485 quality-management system most likely already contains
processes for risk management so you should ensure these are observed and
reused here as much as possible. The purpose of IEC 62304 is to make sure
all relevant software-specific aspects are considered in your risk manage-
ment operations.
The two most common beliefs when it comes to how ISO 14971 risk
management is different for medical-device software than it is for other
types of devices are that a) software always fails, and b) you can only
use external, non-software risk controls to mitigate software risks. Neither
is true, but as ideas to shake you awake to the importance of risk man-
agement with software both are valuable. These topics are further dis-
cussed in Section 4.3 when assigning software safety classifications and
in Section 7 in the context of the full scope of your risk management
activities.

Figure 1.23 You don’t have to be Tom Cruise to find your “Porsche” in a creek if you
ignore the signs.
What to Know before Getting Started ◾ 51

For now, it is worth knowing about the most essential terms related to
risk management. These are: hazardous situation, hazard, harm, and risk.
The term “hazardous situation” (Clause 3.35) refers to circumstances wherein
people, property, or the environment are exposed to hazards. A “hazard”,
then, is a potential source of harm (Clause 3.9), and “harm”, in general, is
some physical injury and/or damage to the health of people, property, or
the environment (Clause 3.8). Finally, risk is the combination of the severity
and probability of that harm (Clause 3.16). These terms will be discussed in
much more detail in Section 7, and even before as appropriate.

1.24 Certification Bodies and the Notified Body


The concept of a notified body is a purely European concoction,
but the various types of other testing and certification bodies you
may engage to assess your operations according to a standard such
as IEC 62304 is similar the world over. Figure 1.24 illustrates how
certificates differ.
You may choose to ask a service provider to assess your software life-cycle
activities and deliverables for conformity with a standard such as IEC 62304.
No service provider will be able to issue you a certificate on IEC 62304 (see
Section 3.2.1), but the results of the assessment will enable you to convince
your stakeholders your pipeline is in great shape, and short of that, obtain
expert advice on where to improve your operations. You may also be
required by your notified body to have the assessment performed.
Your choice of service provider to employ here is wide and ranges from
hired consultants and unaccredited certification bodies to accredited cer-
tification bodies also acknowledged by your notified body. As remarked
on elsewhere in this book, the choice here is up to you, but it may mean
the difference between having your notified body accept the assessment

Figure 1.24 Know that not all certificates carry equal value.
52 ◾ Medical-Grade Software Development

as is, wanting to audit your assessor themselves, or demanding that it is all


redone. On the other hand, after a successful IEC 62304 conformity assess-
ment, your notified body will be able to mark some questions off its list and
concentrate on the remaining assessment preceding the clearance of your
product. The certification body thus concentrates on, perhaps, a single stan-
dard like the IEC 62304 whereas the notified body must also observe other
applicable standards and regulations and addresses, for example, clinical
evaluation in much more detail.
In general, a certification body (CB) is an organization that is accredited
to provide testing and certification services according to some specific stan-
dards in a particular domain of industry. The accreditation will be granted
under some specific accreditation scheme, such as ILAC (https://ilac​.org)
or IECEE (https://iecee​.org). Under the scheme, the national certification
body (NCB) will license testing laboratories (CBTL) to conduct assessments
under its authority and it may then issue certificates on passed conformity
assessments.
No certificates on IEC 62304 can be issued under the CB scheme. Instead,
under the IECEE CB scheme it is possible to certify a medical device based
on the IEC 60601-1 standard and include a mention of IEC 62304 under the
applicable collateral and particular standards via IEC 60601-1 Clause 14. On
the certificate the IEC 62304 standard can, at most, be listed under the sec-
tion for additional information. In absence of a certificate the final document
from an IEC 62304 assessment is the test report, which you can then show
to your stakeholders and the notified body.
Similarly, it may be possible to include IEC 62304 assessment as part of
an overall ISO 13485 assessment or even to obtain a certificate of some sort
on such conformity. In this case, the ISO 13485 certificate may include a
statement that the design and development activities of the QMS adhere to
IEC 62304 requirements for software development. The weight of such a
certificate may be debatable and will depend greatly on the credibility of
the issuing organization. A certificate you obtain from your uncle or from a
traveling snake oil salesman may not be worth much, but an optional third-
party certificate may carry at least some partial weight during your overall
conformity assessment. This is a matter you may want to discuss with your
certification body.
When in doubt, consider obtaining the services of an accredited test-
ing and certification provider that is part of either the ILAC or the IECEE
schemes. This way the verdict of the assessment will also be accepted
by third parties issued in the CB scheme. Conformity assessment and
What to Know before Getting Started ◾ 53

regulatory approval are discussed in more detail in Sections 13 and 14,


respectively.

1.25 Safety Is Relative
What do Butch Cassidy and the Sundance Kid, on the one hand,
and the square-glassed pensioner in the animated film “Up!”, on
the other, have in common? Both go flying off to adventure, but it is
probably safe to say that they have wildly differing expectations of
what is to come. Figure 1.25 shows one way to mitigate dangers.
Rather mind-blowingly, safety is defined as freedom from unacceptable risk
by IEC 62304 in its Clause 3.21, where the standard leans on the ISO 14971
standard for risk management. This makes sense, of course, but it means
that a skydiver jumping out of a plane with an unchecked parachute and
a granny sitting inside her locked house might both be considered safe

Figure 1.25 Marshmallows on cobra fangs might increase your sense of safety, but
will it be safe enough?
54 ◾ Medical-Grade Software Development

provided that the expectation of acceptability is adjusted separately for each.


The needs for insurance, then, should be the same for both people, but
somehow the price, premiums, and deductibles involved probably are not
equal in real life.
This realization lends new credence to the long-running debate on how
the acceptability of risk is defined under applicable regulations – whether
something mitigated to an acceptable level is also mitigated as low as pos-
sible, or vice versa. The debate is not just about semantics as it now rep-
resents the difference between setting acceptability at an appropriate level
versus a devil-may-care level. If, for example, the criminal being asked if he
felt lucky by Dirty Harry standing above him with a Magnum 44 had set the
acceptable risk of miscalculating the spent ammo at ± 6 bullets he might
have felt completely safe to gamble with his life. If he cared about his life
this would have been a high-risk proposition. The reason that the definition
of acceptability matters so much is not just that a higher risk gets you closer
to danger, but that an unacceptable definition of acceptability may have
you in significant constant danger. Let this thought linger as we move over
to look at the two essential standards, ISO 13485 and IEC 62304, and their
combined requirements next.
A minor note here is that IEC 62304 actually talks of both safety and
SAFETY. The former is solely used in the context of the software safety clas-
sification while the latter refers to the above definition based on freedom
from unacceptable harm. Technically safety, spelled with lowercase letters,
refers to the common dictionary definition of safety but this is only used in
the context of the software safety classification, and as the assignment for-
mula for that classification is concerned with freedom from harm and injury
(see Section 4.3), the difference between safety and SAFETY is academic.
Chapter 2

ISO 13485 AS THE


BACKBONE OF IT ALL

Quality management will form the backbone of all your operations,


including those centered around your software development and
product-realization activities. This section will introduce you to the
international ISO 13485 standard that forms the basis of just about
all work on medical devices. Here is where you’ll start putting
together the T-800 endoskeleton of your model for medical-grade
software development. In the next section we will then glue on the
exoskeleton parts from IEC 62304 software development.
By now ISO 13485 is an old, trusted friend to much of the medical-device
world. We have followed it to the bottom of the rabbit hole so deep and
gone without competing forms of oxygen for so long that it is hard to imag-
ine adopting another guide to navigate the global medical-device business.
Maybe this could be called succumbing to the Stockholm syndrome, falling
in love with your captor, or maybe it just means many of us see the stan-
dard as the logical expansion of the Helsinki declaration to the running of a
medical-device business where human subjects are involved both within and
without the factory walls. Either way, the standard is very much a part of the
“way of life” for this area of business.
The standard is not perfect – see the earlier QMS book written by
Dr Ilkka Juuso on the standard (Juuso 2022) for more discussion – but
it is a compelling approximation of perfection. The ISO 13485 standard
forms the backbone of so many critical initiatives in the field that you
may consider it bedrock here. The Medical Device Single Audit Program

DOI: 10.4324/9781003454830-2 55
56 ◾ Medical-Grade Software Development

(MDSAP) alone has anointed the standard as the messiah of the global
medical-device business, but also the US FDA and the European Union
are increasingly hitching their wagons to the standard – cautiously, but
purposefully. At a recent event, Dr Jeff Shuren, the director of the FDA’s
Center for Devices and Radiological Health (CDRH) put on a very posi-
tive spin suggesting American companies consider moving ahead with
the transition to the standard now. European companies are already there
with the advent of the EU MDR. This is all great news for a global medi-
cal-device ecosystem, where harmonization and the compatibility of regu-
latory solutions enable faster, safer, and more affordable manufacturing of
medical devices. The companies will benefit from this reduced complexity
and overhead, but so too will the patients, doctors, and other healthcare
customers in the form of new innovations reaching users quicker and
cheaper across the globe.
There may be future disruptions to this standard too as regulatory expec-
tations evolve as a result of new more advanced and autonomous software
products appearing on the scene, but nothing in the past decade has seri-
ously challenged the contents, shape, or role of the standard. Ripping the
standard apart and putting it back together now would be an act of inter-
national terrorism on the medical-device industry and the physicians and
patients relying on medical devices and medical software around the world.
Instead, expect accompanying guidance on the application of the standard
to new relevant subfields of medical devices.

2.1 The Whole of the Quality Management System


The ISO 13485 standard covers the entirety of your medical-device opera-
tions via a comprehensive quality management system. The whole system
is aimed at giving structure to your operations in a way that enables you to
develop, release, and support medical-device products reliably.
The topics covered by the standard are:

◾ QMS documentation (incl. both documents and records)


◾ Corrective and preventive actions (i.e., CAPA)
◾ Monitoring and improvement of the QMS
◾ Infrastructure
◾ Human resources (incl. qualifications and training)
◾ Suppliers and distributors
ISO 13485 as the Backbone of It All ◾ 57

◾ Auditing (incl. internal audits, external audits, and the management


review)
◾ Communication, marketing, and sales
◾ Risk management
◾ Product realization

The topics touched on by the standard, but not covered in any detail,
include the following:

◾ Clinical evidence
◾ Regulatory affairs (incl. product registration)
◾ Post-market surveillance

All these topics are discussed in our previous book on ISO 13485 quality
management (Juuso 2022). In this present book, we are mostly interested
in how software development can slot into the wider ISO 13485 quality-
management context. The topics listed above are thus only addressed to the
degree necessary here. For a more in-depth discussion of the topics please
refer to our earlier book.

2.2 The Process for Product Realization


In the ISO 13485 standard, matters related to product realization (incl. devel-
opment and maintenance issues) are chiefly a topic for Clause 7 on product
realization. This massive clause comprises roughly 34% of the whole stan-
dard and could even be considered a standard within the standard. The
clause is intended to preside over all your product-realization activities. It
instructs subprocesses such as planning of product realization, customer-
related processes, design and development (D&D), purchasing, and produc-
tion and service provision. In this context, design and development includes
planning, inputs, outputs, reviews, verification, validation, transfer, control of
changes, and the file set. It is worth noting that the ISO 13485 standard does
let the user of the standard exclude all or some subset of the requirements
in Clause 7 if they have a valid rationale for doing so. It is thus possible that
an organization certified to ISO 13485 quality management does not actually
implement any of the requirements regarding the development or manu-
facturing of devices. An ISO 13485-certified quality management system
may mean different things to different organizations, but the elements and
58 ◾ Medical-Grade Software Development

processes they do implement should be compatible on a high level following


the standard.
In addition to Clause 7, other clauses of the standard may relate to
product-realization processes. This will often be the case as the purpose
of the organization will likely be linked to the medical device it some-
how manufactures or makes available to its customers. Typical examples
of requirements covered elsewhere in the ISO 13485 standard but con-
nected with activities you might undertake in the context of software
development include tool validations, supplier management, and marketing
communication.
Clause 7 is the most readily overlapping area of common interest
between the ISO 13485 and IEC 62304 standards. The overlap will be ana-
lyzed in-depth in the sections to come, but here it is useful to already set up
the general hierarchy of the court these standards keep. The patient is natu-
rally the king in this regal setting, but it is the ISO 13485 standard that sets
up many of the departments, professions, and pastimes of the castle. IEC
62304 makes use of this setting and goes into more detail on how the engi-
neers, blacksmiths, and other tradesmen might then go about their business
creating the deliverables asked of them. In other words, ISO 13485 provides
the environment the work takes place in and many of the supporting ser-
vices the work relies on (e.g., customer communication, purchasing, and the
final provision of the product or service), and IEC 62304 adds the nuts and
the bolts expected in between for software while relying on what the ISO
13485 has already said. Remember that having a QMS is not optional if you
want to use the IEC 62304.
A rough sketch of the areas of common interest between the two stan-
dards, or the overlap of sorts between them, is represented in Table 2.1
below. The table is a simplified overview of the general lay of the land, and
practice-related requirements, both big and small, are served out peppered
all over each of the standards. An empty cell in the table does not therefore
indicate that the standard would not have any wishes on the particular topic
only that it doesn’t dedicate real estate to discussing it. The discussion in the
coming sections will address all of the topics in good time.
ISO 13485 sees product realization as consisting of design and develop-
ment stages, the execution of which is planned in advance and supported
by a range of other processes related to the customer, purchasing, and prod-
uct provision. IEC 62304 shares the notion of planning and complements the
structure of design and development stages but is less concerned with any
supporting processes.
ISO 13485 as the Backbone of It All ◾ 59

Table 2.1 A Rough Mapping between ISO 13485 Clause 7 and IEC 62304
Topic ISO 13485 IEC 62304 See
Supporting 7.2 Customer-related - Juuso
processes processes 2022
7.4 Purchasing
7.5 Production and
service provision
7.6 Control of monitoring
and measuring
equipment
Planning 7.1 Planning of product 5.1 Development 5.1
realization planning
7.3.2 D&D planning 5.1
Inputs 7.3.3 D&D inputs 5.2 Requirements 5.2
analysis
Design and - 5.3 Architectural design 5.3
implementation
5.4 Detailed design 5.4
5.5 Unit implementation 5.5
and verification
5.6 Integration and 5.6
integration testing
5.7 Software system 5.7
testing
Outputs 7.3.4 D&D outputs 5.9.1
Review 7.3.5 D&D review 5.9.1
Verification 7.3.6 D&D verification 5.9.2
Validation 7.3.7 D&D validation 5.9.3
Transfer 7.3.8 D&D transfer 5.8 Release 5.8
Documentation 7.3.10 D&D files 11
Configuration 7.5.8 Identification 8 Configuration 8
management management
7.5.9 Traceability 8.1 Configuration
identification
(Continued)
60 ◾ Medical-Grade Software Development

Table 2.1 (Continued) A Rough Mapping between ISO 13485 Clause 7 and IEC
62304
Topic ISO 13485 IEC 62304 See
Maintenance 7.3.9 Control of D&D 6.1 Establish software 6.1
changes maintenance plan
6.2 Problem and 6.2
modification analysis
6.3 Modification 6.3
implementation
8.2 Change control 8.2
9 Problem resolution 9

The design and development stages defined by ISO 13485 are planning,
inputs, outputs, review, verification, validation, and control of changes.
The standard also has a clause on the files used in these stages. In the list
of stages, it is notable that a gaping hole exists between the input and the
output stages. This is where IEC 62304 comes in to instruct what it means to
be between inputs and outputs and what work is needed to proceed from
inputs to outputs. The match here is complementary, if not quite plug-n-play.
In general, ISO 13485 provides the larger framework within which IEC
62304 can sit. Often the former provides the process definitions the lat-
ter can use to create the documentation and (other) deliverables required.
IEC 62304 adds necessary information to processes and documentation
(such as the software development plan) to capture the needs of software
development.
Note that Table 2.1 essentially matches with the rough mapping given
in Annexes C and D of IEC 62304 (see Sections 3.2.14 and 3.2.15). We have
here opted to place system testing as part of D&D instead of as something
only looking at its outputs (although that too is a defensible interpretation)
and similarly taken the view that reviews, verification, and even to some
extent validation activities can – and often are – fused into the development
life cycle of software. The marriage of requirements from IEC 62304 and ISO
13485 is covered in-depth in Sections 5 to 9 of this book.
The ISO 13485 standard itself is addressed in great detail in the ear-
lier book written by Dr. Juuso, Developing an ISO 13485-Certified Quality
Management System (2022). The compatibility between ISO 13485 and IEC
62304 requirements is the topic of this book and will be addressed in detail
in the coming sections.
Chapter 3

IEC 62304 AS THE FLESH


AROUND THE BONES

If ISO 13485 quality management is the backbone of all medical-


device businesses, IEC 62304 provides the exoskeleton around this
to give you the necessary rigor for medical-grade software develop-
ment. This section introduces you to the standard, and together with
the following sections, helps you negotiate together all the pieces to
shape your software-development activities in a way that fits snug-
gly with the expectations of quality management and regulatory
requirements in the domain of medical devices.
As standards go, the IEC 62304 is a brief, but weighty piece of literature.
The standard is not the prettiest one, nor does it play the nicest with the
other kids in the standardization playground. In fact, the IEC 62304 has had
a fairly troubled life being bogged down with revision woes for much of
its existence. The standard is, however, widely used in the development of
medical devices, and a de facto requirement as an extension to ISO 13485
quality management for any medical-device company engaged in the devel-
opment of software. Your certification body will expect to see evidence of
how you have considered the requirements posed by the IEC 62304 stan-
dard and how you have met these in your products and operations if any of
these mention software. Depending on what level of billing you give soft-
ware, whether your product is software or contains software, you may find
these expectations need to be met via an in-house assessment or an external
assessment by a third party the regulatory body can trust.

DOI: 10.4324/9781003454830-3 61
62 ◾ Medical-Grade Software Development

Bruised and battered, the standard continues to give us the yardstick


against which all software development and maintenance operations are
measured. That alone is reason enough to take it seriously.

3.1 The Past, Present, and Future


To understand the future, we shall begin by looking briefly at the past and
the present. This will help us grasp the tectonic movements in the develop-
ment of software that have reverberated through the industry in the past
decades. Later in Section 12 we will point the juiced-up DeLorean even fur-
ther into the future and reach into our crystal ball to discuss the phenomena
shaping the near future of medical-software development in more detail.

3.1.1 Times of the Primordial Ooze


The rise of the integrated circuit, that ubiquitous brain-on-a-chip Lego-block
of the electronic world, some time in the 1980s increased interest in and
the development of software as part of electrical medical devices. At this
time the circuits, also known as microchips, became commonplace in the
design of physical electrical devices. The combination of the microchip and
the software that it could run unquestionably brought about greater compu-
tational capabilities for the devices but also enabled some completely new
functionalities for the devices. This small revolution led to greater emphasis
being placed on the criticality of software to the safe operation of devices.
Subsequently, work began on developing requirements for not just the
devices themselves but also for the software embedded in the devices. At
the same time, the manufactured devices themselves graduated from purely
electronics-based devices, such as an RCA controller-based sound-level mea-
surement device in the 1970s, to more modern Intel 8088-based architec-
tures for patient monitoring devices around 1985.
The safety of electrical medical devices themselves has been regulated
since 1977 by the first version of the international standard IEC 60601-1.
This did not, however, address software in any way. The second edition
published in 1988 and amended in 1995 cautiously mentioned software and
the, at the time, upcoming collateral standard IEC 60601-1-4 specifically on
software.
In 1996, the first true fruits of this work on software-specific safety
requirements saw the light of day as the IEC 60601-1-4 standard was
IEC 62304 as the Flesh around the Bones ◾ 63

published as part of the IEC 60601-1 series of standards. Finally, in 2006 hot
on the heels of the third edition of IEC 60601-1 came the first version of IEC
62304 which has since become a yardstick for medical software. All these
standards were published by the International Electrotechnical Committee,
that is still, among other standards, responsible for the IEC 62304 stan-
dard today. This series of standards created the foundations for assessing
the safety of medical-device software. The original series of five standards
addressed general requirements, the system, EMC, radiation, and software.
This set has since been extended with a number of collateral and particular
standards as shown in Figure 3.1. Two standards of the original set, system,
and software, have since been withdrawn as is evident from the missing col-
lateral standards 1-1 and 1-4 in the figure.
Part 1 of the series contains general requirements for basic safety and
essential performance of medical electrical equipment. The other parts,
called collateral or particular standards, supplement or modify the Part 1
requirements for particular types of devices. Part 1-1 and Part 1-4 on soft-
ware have since been withdrawn, but Parts 1, 1-2 (EMC) and 1-3 (radiation)
remain active in revised form. Also, additional parts have since been pub-
lished on usability (Part 1-6), alarms (Part 1-8), environmentally conscious
design (Part 1-9), physiologic closed-loop controllers (Part 1-10), home
healthcare (Part 1-11), and emergency medical-service environments (Part
1-12). These standards may be of interest to you if your software product
touches on any of these domains, but they are beyond our discussion here.

Figure 3.1 The collateral and particular standards of IEC 60601-1.


64 ◾ Medical-Grade Software Development

As a package of standards, the IEC 60601-1 is still one of the most expensive
items you may find yourself buying.
It is noteworthy that the scope of the standard was from the outset
tightly coupled with the safety of electrical medical devices and the software
embedded in those electrical medical devices. This was also the case in the
collateral standard Part 1-4 on software, where the scope of the standard
(its Clause 1.201) looked at the safety of programmable medical electrical
systems (PEMS), which it saw as being either equipment or systems both
­containing programmable electronic subsystems (PESS) as illustrated in
Figure 3.2.
Technological advances in the 1990s led to increased networking between
devices and the utilization of server infrastructure to greatly boost the com-
putational capabilities of devices. This development thus started to move the
focus from a physical electrical device to the networks now used to store
data, analyze it, and reuse it in later diagnosis.
Part 1-4 of the IEC 60601 series of standards, the collateral standard
on software, was subsequently updated in the year 2000, but it was soon
admitted that a new standard was needed to address the use of software in
healthcare even in cases where no physical device was implicated. At the
time, Europe was living with the Medical Device Directive (93/42/EEC), and
the view started to emerge that software could be a medical device by itself
as understood in the regulation.

Figure 3.2 Scope of IEC 60601-1-4 on software.


IEC 62304 as the Flesh around the Bones ◾ 65

Finally, in 2006 a new standard designated as IEC 62304 saw the light of
day. This standard was also quickly harmonized according to the Medical
Device Directive (93/42/EEC). The adoption of the then-new standard took
some time and was not without its opponents. Manufacturers were perhaps
already accustomed to requirements elicitation, for example, but the new
standard raised the level of the ask via setting more expectations on what is
to be covered in Clause 5.2.2. Another point of contention was traceability,
which caused considerable head-scratching then and still remains a common
point of failure in operations today. Risk management, too, became a much
more involved process with the introduction of the requirements in Clause 7.
The IEC 62304 standard brought assessment organizations an entirely
new assessment method, and one they then had to get to grips with in order
to start using the standard. Prior to the standard, assessment of software
had used IEC 60601-1 Clause 14, but due to the limitation of that standard
to physical medical electrical equipment standalone software was effectively
left out. This was also true of IEC 60601-1-4 which in practice addressed
embedded software only. The IEC 60601-1 family of standards did represent
an improvement over prior practices of assessing so-called single fault con-
ditions where individual fault conditions were simulated to study how the
device coped with or recovered from these faults. An early example here
is given by IEC 60601-2-34 on invasive blood pressure measurement where
testing required the use of a specific testing algorithm whereby an errone-
ous timer state was created in the software and the subsequent safety of the
hardware was then assessed.
The standard was updated in 2015 when the most significant change
was the addition of a pathway for legacy software, the so-called “Legacy
Concept” (see Section 4.4). This established an alternative route to demon-
strate safety and compliance in the case of products that had already been
on the market for a long period of time.
The scope of the standard is discussed in detail in Section 3.2.3.

3.1.2 Modern Times
The IEC 62304 standard established a means to evaluate software even
when that software was not a part of an electrical medical device. The
standard was a much-needed addition to the toolbox of the medical-device
industry, which had in fact for some time already utilized server-based soft-
ware for the storage and analysis of data in healthcare applications. The new
standard thus caught up with the development of technology.
66 ◾ Medical-Grade Software Development

After IEC 62304, a new much-needed pathway was in place to address


software independently of a device, as depicted in Figure 3.3. The inde-
pendence is not absolute, though, as the standard does remind the reader
to also observe the host device somehow (see Section 3.2.3), but this is left
for other standards to cover. Note how in the figure the paths of embedded
software and software in its own right converge before compliance is dem-
onstrated. Embedded software may be assessed as part of its host device,
provided that IEC 62304 considerations are observed. Similarly, the assess-
ment of standalone software must observe, in some way, what the host
device is.
Today, when IEC 62304 is used in the context of assessment against the
modern General Safety and Performance Requirements (GSPR, Annex I) of
the EU Medical Device Regulation (EU 2017/745), a new step is needed. As
technology, and the world too in general, have evolved since the last amend-
ment to IEC 62304 in 2015 new requirements have been introduced for, e.g.,
validation (EU MDR GSPR 17.2), cybersecurity (EU MDR GSPR 17.4), usability
(EU MDR GSPR 5) and the documentation accompanying a device (EU MDR
GSPR 22.x). Subsequently, the adoption of another standard, IEC 82304-1,
may be beneficial to the manufacturer.
The details of the IEC 82304-1 standard, too, are largely beyond our dis-
cussion here (although see Section 12.2). Nonetheless, this is a particularly
practical standard for use with health software that is designed to operate on
general computing platforms and intended to be placed on the market with-
out dedicated hardware. In keeping with modern thinking, the standard also
covers software validation, installation, maintenance, and disposal. It should

Figure 3.3 Demonstration of compliance with standards IEC 60601-1 Clause 14, IEC
62304, and IEC 82304-1.
IEC 62304 as the Flesh around the Bones ◾ 67

be noted that the standard speaks of health software which is a term that
may or may not be identical to medical-device software (see Section 12.2).
At the very least, the use of the standard may help guide the development
of health software and provide practical tips for the development of medical-
device software.
The IEC 62304 continues to be adopted for use with medical-device soft-
ware today. It provides a practical framework for structuring and analyzing
the software-development work of manufacturers both within and without
the organizations. As is one of the key intentions behind any standard, it
enhances compatibility between expectations and compatibility between
products that need to somehow link up whether during development or
once available as finished products. It is not a perfect standard, though, and
it is getting quite old.

3.1.3 The Future
The IEC 62304 standard provides a framework for the assessment of soft-
ware safety and conformity to requirements. The elephant in the room, or
the 800-pound gorilla cohabiting your small office cubicle, is that the latest
amendment to the standard is from 2015 and by now woefully out of date.
It is possible to apply IEC 82304-1 (see above) as a patch to wrangle the
IEC 62304 to the present day. This may be akin to what the Tyne Daly char-
acter attempts to do in the third installment of the Dirty Harry franchise –
drag a Neanderthal to the 1970s – but it may be what you want to do.
This would, however, not remove all uncertainties. The following are
some of the top concerns we hope are addressed when the IEC 62304 is
finally revised:

◾ The requirements in the standard are written on a very high level that
then leads to the possibility of significant differences in the acceptance
criteria employed by different auditors. Better guidance and examples
on the use of the standard in conformity assessments would thus be
welcome and could be provided, e.g., as appendices to the standard or
as a separate technical report.
◾ The handling of cybersecurity should be greatly improved. The current
passing reference in Clause 5.2.2 during requirements analysis is not
adequate or on par with modern medical-device regulation. The future
version of the standard should address the requirements for both the
implementation and the management of cybersecurity.
68 ◾ Medical-Grade Software Development

◾ The safe use of software will often be impacted by user interaction,


user interfaces, and other usability considerations. The future version of
the standard should thus also address these requirements in a practical
way.
◾ Work on revising the standard has been bogged down in issues
for the better part of its existence. One issue appears to have been
the scope of the standard, which appears to be in limbo between
strictly regulated medical devices and the more loosely defined health
devices. It is our understanding that the standard is applicable to the
assessment of software placed on the market as, or as part of, medical
devices under the EU Medical Device Regulation. The uncertainty in
the appropriate scope should be cleared up, perhaps by defining the
scope as applying to medical devices and being optionally applicable
to health devices.
◾ The standard relies heavily on an ABC software safety classification to
levy requirements based on the criticality of the software under con-
sideration (see Section 4.3). This is a good feature of the standard, but
the assignment of the safety classification could do with more concrete
guidance and examples on how each of the three classes is assigned. In
other words, how the classification can be more accurately and uni-
formly applied and assessed by the users of the standard.
◾ Similar to the previous point, more guidance should also be provided
on the requirement to rely only on risk-control measures external to
the software (see Section 4.3). This requirement is not clear, particularly
since a medical device may consist of several software systems each
perhaps providing risk-control measures over the others.

In practice, we feel that the IEC 62304 standard should be accompanied by


a technical report to provide the necessary guidance and examples needed
to bring the standard to the present day. To make up for lost ground, we
feel that a roadmap should be developed to gradually bring the standard
in line with modern expectations both now and perhaps in the future too.
This roadmap needs to address the increasing complexity of software, the
management and implementation of cybersecurity, the use of various cloud-
based architectures, and the evolving expectations of general safety and
performance for software-based products. Some of these topics will also be
covered later in Section 12 of this book.
In the next section we will dive into the contents of the present version
of the standard, and also comment on the changes we feel are warranted.
IEC 62304 as the Flesh around the Bones ◾ 69

3.2 Contents
The standard is roughly 40 pages of clauses and 47 pages of appendices in
its original 2006 form with an additional fatly spaced 51 pages of changes
made by the 2015 amendment. The appendices are there as guidance, to
enhance your understanding and provide suggestions on the interpretation
of the standard.
The business end of the standard consists of a foreword, an introduction,
and the clauses themselves. The clauses are where the actual requirements
of the standard are given, split into nine topics, and administered through
the safety classification of your software and the safety classifications of the
software items it consists of according to your design. Also, legacy software
has its own path through the standard, as discussed later in Section 4.4.
The nine main clauses are broken down into numbered subclauses,
which together give you the coordinate system often quoted when discuss-
ing the requirements imposed by the standard. We will go through each
clause in later sections of this book, but the following provides the first
bird’s-eye overview of what to expect in the standard. This is the “lay of the
land” you will be working with as you shape or reshape your own processes
to comply with the standard as leanly and as smartly as possible. Note also
that the numbering of Sections 4 through 9 of this book match with the
clause numbers of the standard for easy reference.

3.2.1 Foreword
This is the foreword to the standard by the International Electrotechnical
Committee (IEC), which has ownership of the IEC 62304 standard. The
foreword provides an overview of how IEC standards are developed and
expresses the goal of the standards to represent, as nearly as possible, an
international consensus of opinion on any topic addressed.
The foreword explains that the IEC does not provide any marking
mechanism to indicate its approval of, for example, devices manufactured
according to its standards. This is why you don’t get to apply an IEC 62304
certificate to your processes, but you may engage the services of several
different types of consultants and certification bodies to check your product
– and by extension your product-realization processes – for conformity with
the standard. The outcome of such a check-up is a test report containing
an assessment of how you meet the requirements of the standard, but not a
definitive certificate per se. The foreword also forewarns you that no liability
70 ◾ Medical-Grade Software Development

can be placed on the IEC or its staff. The conformity assessment and test
report are discussed in more detail in Section 13.
The foreword then describes how the standard uses formatting to high-
light requirements and definitions (printed using Roman type), informa-
tive notes (in small text), and terms (in small capitals) throughout the body
of the standard. Finally, an asterisk in front of a section title indicates that
Appendix B of the standard may be consulted for more guidance. The
important detail to notice here is that any notes given by the standard are
there in a non-binding informative role, as hopefully helpful comments that
don’t have the stature of requirements in the standard.

3.2.2 Introduction
This is the introduction that walks you into the world of the standard and
gently sets up the scene for all that is to come. In so doing, the section is
roughly equivalent to Clause 0 in ISO 13485. In both standards this introduc-
tory text comprises roughly 8–9% of the entire body of the standards and
is more impactful than just thanking you for selecting this particular stan-
dard to read. In fact, the introduction packs a punch in setting up what you
should expect from the standard, and how the writers of the standard see
you making use of it.
The introduction begins by establishing software as an often-integral part
of medical devices and then sets out to provide a framework of life-cycle
processes to ensure the safe design and maintenance of such software. It
grounds the assessment of software safety and effectiveness in the intentions
for its use, and the evidence of those intentions being fulfilled without what
could be considered unacceptable risks. Note that the terms “intended use”,
“intended purpose”, and “clinical evidence” are not used here, but the impli-
cations are similar.
The description makes it clear that the standard assumes all work to take
place within both a quality management system (covered in more detail
in Clause 4.1; see Section 4.1) and a risk management system (covered in
more detail in Clause 4.2; see Section 4.2). Within this context, the standard
intends to define a four-process framework for the safe design and mainte-
nance of medical-device software. The four processes presented are:

◾ “Development” (Clause 5; see Section 5)


◾ “Maintenance” (Clause 6; see Section 6)
◾ “Configuration management” (Clause 8; see Section 8)
◾ “Problem resolution” (Clause 9; see Section 9)
IEC 62304 as the Flesh around the Bones ◾ 71

In case you are wondering what happens in Clause 7 which is missing from
the above list, that is dedicated to risk management and is a topic we will
address in Section 7. Note that despite introducing four distinct processes
here, the standard still sets itself up overall as providing a framework of
life-cycle processes for the safe design and maintenance of medical-device
software – thereby subjugating configuration management and problem
resolution to these grander goals.
The introduction further explains that all these processes are to be under-
stood as consisting of activities, and the activities then consist of tasks. Each
of these Lego blocks is given a definition in Clause 3 of the standard.

◾ Process
Defined as a set of interrelated or interacting activities that trans-
form inputs into outputs (Clause 3.14). The standard appears to
expect that a process always consists of more than one activity –
although it would be tough to argue this was a required character-
istic. The term is sourced from ISO 9000, and thus both IEC 62304
and ISO 13485 use the same source for the definition of the term, but
due to workflow reasons happen to refer to different versions of the
same standard (2000 versus 2015). The difference does not appear
significant.
◾ Activity
Defined as a set of interrelated or interacting tasks (Clause 3.1). This
may also address the use of resources, as suggested by a bafflingly
placed note to Clause 3.14. The term is not sourced from outside the
standard, but instead defined here from scratch. Interestingly, ISO
9000:2015 does also define the term and what is more puts a distinct
project-management slant on it to capture the smallest identified object
of work in a project. Thus, a minor point of contestation exists here
between activity as understood in IEC 62304 (a set of tasks) and ISO
13485 (smallest packet of work in a project). In practice, the difference
does not appear significant.
◾ Task
Defined as a single piece of work (Clause 3.31). This is the practical
atom of activity in a process or an activity as the standard sees it. The
term is not defined in ISO 9000, and thus any use of it in ISO 13485 is
according to some commonly accepted dictionary definition. The dic-
tionary definitions, (e.g., a piece of work to be done or undertaken),
match well with the definition adopted by IEC 62304.
72 ◾ Medical-Grade Software Development

As noted above, some minor differences may be observed in how the above
three concepts are used by the two standards, the aging IEC 62304, and the
ISO 13485. The differences are slight and do not prevent a practical arrange-
ment to be negotiated between the two standards. If altercation was the
goal, some contestation could be found in how the granularities of some
terms work across the two standards, but as even atoms can be split mul-
tiple times in laboratory conditions this seems hardly practical. What matters
is that you know how you want to use processes, activities, and tasks – and
that you are consistent in their use.
Importantly, no specific model of processes, activities, or tasks is forced
on the user of the standard, but instead the user is expected to map these
elements to their own chosen life-cycle model. This also means that the user
is expected to map the four key processes of the IEC 62304 standard to their
own processes. This mapping will then be a major topic of discussion when
comparing any two manufacturers and the processes they may have for soft-
ware development – not to mention a major undercurrent of any conformity
assessment based on the standard.
No organizational structure or division of work is imposed on the user,
either. Instead, it is only expected that the processes, activities, and tasks
defined by the standard are completed if conformity with it is claimed.
ISO 13485 will add the necessary organizational overhead here (see Juuso
2022).
Similarly, the IEC 62304 standard requires that tasks are documented, but
the shape of this documentation is left to the user. No names, formats, or
explicit content is specified for the documentation to be produced. These
are both significant design features of the standard and ones that you may
miss if you skip learning about the introduction of the standard. In practice,
you should aim to match the process and document names of the standard
if you are starting fresh. This way retaining a mapping between your pro-
cesses and documents, and those expected by the standard will be as easy
as possible. However, you don’t have to do so.
Importantly, the current version of IEC 62304 makes a normative (bind-
ing) reference to ISO 14971 risk management, which it thus requires every-
one to use along with whatever further software-specific requirements it lays
out for risk management (see Section 7). The intention here is to make some
further requirements in the standard to address the nature of software and
to ensure that the risk management for the software and any device using
that software occurs in some synchronized manner. It is noteworthy that
the reference both here and later in Clause 2 is made without specifying a
IEC 62304 as the Flesh around the Bones ◾ 73

version of the risk-management standard, and thus the reference is always


to the latest version of the standard whatever that happens to be (e.g., ISO
14971:2019).
The introduction also clarifies the use of two essential concepts which
show up repeatedly in the standard.

◾ “Establish”
Clarified as defined, documented, and implemented. Note that in
comparison to ISO 13485 (which defined “documented” as something
that is established, implemented, and maintained) here the sentiment is
the same, but the clarification does not explicitly call out maintenance.
◾ “as appropriate”
Clarified as appropriate in the context of a process, activity, or a task
unless excluded providing a documented rationale. Note that here the
clarification is somewhat looser than in ISO 13485 which states that if
a requirement is necessary for meeting product requirements, regula-
tory compliance, carrying out corrective actions, or managing risks,
it can’t be excluded. Similarly, the standard here contents itself with a
documented rationale instead of an acceptable rationale, and limits the
context to a process, an activity, or a task. The difference between the
two standards is noticeable, but in practice hard to argue. The key take-
away is that something qualified with “as appropriate” is not something
you can just ignore because it doesn’t seem relevant.

Finally, the introduction adopts the common definitions of shall (it is


required), should (it is recommended) and may (it is permissible) into the
standard. These are defined as you would expect, but if you are not accus-
tomed to these concepts in a professional setting, make sure you use them
as defined here. Note that the term can, used to indicate a possibility by the
ISO 13485, is not mentioned here. Figure 3.4 shows a word cloud of what to
expect in the introduction.

3.2.3 Clause 1: Scope
Clause 1 represents 6% of the standard and sets the scope for its applica-
tion. Clause 1.1 reiterates the intention of the standard to provide life-cycle
requirements for medical-device software by defining a common framework
for the life-cycle processes expected. The framework consists of a set of pro-
cesses, activities, and tasks as already set out by the introduction.
74 ◾ Medical-Grade Software Development

Figure 3.4 Word cloud for the Introduction.

Clause 1.2 defines the intended field of application for the standard. The
clause was rewritten completely for the 2015 amendment and now instructs
that the standard applies to the development and maintenance of medical-
device software regardless of:

◾ Nature of the software: The standard applies whether the software is


a medical device itself or embedded in one. A non-binding note further
suggests that requirements regarding software of unknown provenance
(SOUP) also apply and offers Clause 8.1.2 as a point of further reading.
Note though that an entire medical-device software system in itself can’t
be a SOUP (see a non-binding note to Clause 3.29)
◾ Mode of execution: The standard applies whether the software is
executed on a processor directly or via an interpreter
◾ Type of storage: The standard applies regardless of the type of persis-
tent storage used for the software (e.g., flash memory, hard disk, optical
disc)
◾ Type of delivery: The standard applies regardless of the method of
delivery used for the software (e.g., via e-mail, optical disc, flash mem-
ory). Note that this presumably also covers cloud solutions, and the
method itself is not considered to be a medical device by the standard

Clause 1.2 also points out that the standard does not cover software vali-
dation or final release, not even in cases where the software is a medical
device itself. A non-binding note to the standard further elaborates that
IEC 62304 as the Flesh around the Bones ◾ 75

validation and other system-level development activities may be found in


product standards such as IEC 60601-1 and IEC 82304-1 (see Section 12.2).
Clause 1.3 introduces the requirement that the standard is applied
together with other relevant standards when developing a medical device.
The list of relevant standards, and their mapping to this standard, is covered
in detail in Appendix C of the standard, but this should not be considered
exhaustive or up to date.
Clause 1.4 discusses how compliance with this standard may be claimed.
Compliance is defined as the implementation of the processes, activities, and
tasks defined in the standard as these are applicable to your software safety
class. The essential concept of a software safety class is thus first introduced
by this clause in passing, but with the momentous effect that not all the
requirements of this standard need to be fulfilled if your software safety
class is not implicated by the requirement. The concept itself will be covered
a little later in Clause 4.3.
The evaluation of compliance is determined via two actions:

◾ Inspection of your documentation (incl. your risk management file)


◾ Assessment of your processes, activities, and tasks

A non-binding note to the clause further clarifies that the assessment may
be carried out by internal or external audit, although the use of the former
option has in practice been a contested issue during regulatory assess-
ments of the final medical device in the EU. It should also be noted that
the standard appears to use the term inspection for the first action, the term
assessment for the second action, and the term determination for the overall
evaluation of compliance. This distinction here is less-than-optimal as taken
together with the note on “assessment” it appears to suggest that only the
second action could be performed internally or externally. In practice the
whole IEC 62304 evaluation may be performed internally or externally, but
the outcome is still only as valuable as the body conducting the final regu-
latory assessment of the medical device deems it. Using a reputable certifi-
cation body is recommended here. See Section 13 for more discussion on
conformity assessment.
Further non-binding notes to the clause acknowledge the need for flex-
ibility in assessing the methods of process implementation and the perform-
ing of activities and tasks and call for documented justification in cases
where a requirement marked as as appropriate by the standard is not ful-
filled. Finally, a non-binding note offers Clause 4.4 (see Section 4.4) as a
76 ◾ Medical-Grade Software Development

Figure 3.5 Word cloud for Clause 1.

further reference in the case of legacy software, and another non-binding


note comments that conformance may mean the same as compliance (e.g.,
in the case of ISO/IEC 12207).
The whole of Clause 1.4 is somewhat unusual in the sense that ISO
13485, for example, does not explicitly instruct how compliance is to be
addressed. The clause may be dropped from future editions of the standard,
but some mechanism of placing stricter requirements on software associated
with a higher level of concern (e.g., software for vital signs monitoring) as
opposed to a lower level of concern (e.g., mobile applications for personal
sleep monitoring) will no doubt endure. Figure 3.5 shows a word cloud of
what to expect in the clause.

3.2.4 Clause 2: Normative References


Clause 2 (less than 1% of the standard) makes one normative (binding) ref-
erence to the ISO 14971 standard on the application of risk management to
medical devices. Note that the reference is undated implying that the latest
version of the standard is implicated. In the case of ISO 14971 this means
a version of the risk-management standard that is a decade newer than the
version available when the IEC 62304 standard was amended in 2015. In
any case, the ISO 14971 standard is to be understood as indispensable and
explicitly required for the application of IEC 62304 for the time being.
A later edition of IEC 62304 may remove the normative reference to
ISO 14971 as new types of software, perhaps also including mobile appli-
cations and other health software, appear alongside traditional full-on
IEC 62304 as the Flesh around the Bones ◾ 77

Figure 3.6 Word cloud for Clause 2.

medical-device software. The requirement to apply hardcore ISO 14971 risk


management to all types of software has been one of the many points of
debate for the next revision of IEC 62304. We see ISO 14971 as the recom-
mended approach to risk management in the context of medical devices but
recognize the need to enable other lower-concern users to begin adopting
elements of the IEC 62304 without the immediate issue of implementing full
ISO 14971 risk management. Note also that in ISO 13485 a mention of a risk-
based approach does not automatically always refer to ISO 14971 risk man-
agement (see Juuso 2022). Figure 3.6 shows a word cloud of what to expect
in the clause – this time the view is not only easy to interpret but also quite
conclusive.

3.2.5 Clause 3: Terms and Definitions


Clause 3 comprises approximately 16% of the standard, all of which is real
estate it uses to define important terms used throughout the rest of the
standard. The amended version of the standard introduces 39 terms, out of
which five terms are new to the 2015 edition. One term (software product)
was rescinded in the amendment. In total, 12 terms are defined from scratch
while the remaining 27 terms are based on definitions made elsewhere (e.g.,
ISO 14971, ISO 13485, IEC 60601-1, ISO 9000, ISO/IEC 12207). Out of all the
terms now defined, the following group of four terms, presented from top to
bottom, is perhaps the most essential set of concepts for understanding the
clauses of the standard.

◾ System
An integrated composite (collection) of processes, hardware, software,
facilities, and people that acts to satisfy a need or objective (Clause
3.30).
◾ Software system
An integrated collection of software items organized to accomplish
a function or set of functions (Clause 3.27). This is the highest of three
levels decomposing software. Note that medical device software (Clause
3.12) is then defined as a software system developed for the purpose of
78 ◾ Medical-Grade Software Development

a) being incorporated into a medical device or b) intended to be used


as a medical device.
◾ Software item (incl. SOUP)
An identifiable part of a computer program (Clause 3.29). This is the
midlevel of three levels decomposing software. On a related note, if that
software item is a) developed elsewhere for a purpose other than being
incorporated into the software discussed here, and it is generally avail-
able, or is b) developed previously and adequate records of the devel-
opment are not available, then it is considered software of unknown
provenance (SOUP). Note that the entire software system can’t be
claimed as SOUP.
◾ Software unit
The lowest level of software which is not subdivided into further
parts. This is the lowest of the three levels of decomposing software.
See Clause 3.28.

These four terms thus comprise four nested levels for software decomposi-
tion as the standard understands it, or the “Russian doll” of software and
where it is used. Note that in a somewhat unhelpful move, the 2015 amend-
ment offers source code, object code, control code, control data, or a col-
lection of these as an example of a computer program, but does so in the
definition of a software item. It should be emphasized that the examples
listed apply equally as the representations of an identifiable part of a com-
puter program whether that is a software unit, a software item, or a soft-
ware system. They are not examples of a software item per se. In an equally
unhelpful manner each of these – software unit, software item, and software
system – may, according to the standard, also be called a “software item” by
themselves. This chain of definitions is a veritable Gordian knot, an intrac-
table problem much like the explanation behind the acronym PHP which is
short for PHP Hypertext Pre-processor. Someone out there is laughing their
head off, no doubt, but for a beginner trying to grasp IEC 62304 it suffices to
remember that units build up items, items build software systems, and these
are used in a wider system also including people. Note that the granularity
here is left up to the manufacturer to define (note to Clause 3.28).
Also apparent from reading the list of terms defined is that the standard
sees itself as extending ISO 14971 risk management for the special applica-
tion area of software. It is thus not surprising that a total of 13 terms have
been imported from the 2007 version of the ISO 14971 standard. These
terms are: “harm”, “hazard”, “hazardous situation”, “manufacturer”, “residual
IEC 62304 as the Flesh around the Bones ◾ 79

risk”, “risk”, “risk analysis”, “risk control”, “risk estimation”, “risk manage-
ment”, “risk management file” and “safety”. The sourcing of these terms from
the risk-management standard is a welcome choice promoting compatibility
between the standards, and boding well for the adoption of the later ISO
14971 editions.
By comparison only one term, “medical device”, has been imported from
the 2003 version of ISO 13485. The definition of what is and what is not a
medical device is, of course, a far bigger question than just a term definition
and is something better left to the regulations in each jurisdiction. A further
seven terms – “activity”, “process”, “release”, “risk analysis”, “system”, “trace-
ability”, and “verification” – are silently inherited by ISO 13485 from ISO
9000:2015 and shared by the two standards here. Figure 3.7 shows a word
cloud of what to expect in the clause.

3.2.6 Clause 4: General Requirements


Clause 4 (11% of the standard) addresses four fundamental topics: the qual-
ity management system (Clause 4.1), risk management (Clause 4.2), software
safety classification (Clause 4.3) and legacy software (Clause 4.4). The aim of
the discussion here is firstly to set up the foundations for developing soft-
ware by building on quality and risk management, and secondly to set you
off in the right direction for using the standard by instructing both the soft-
ware safety classification, used in interpreting the standard, and the special

Figure 3.7 Word cloud for Clause 3.


80 ◾ Medical-Grade Software Development

Figure 3.8 Word cloud for Clause 4.

pathway for legacy software, should that be your situation. The clause and
its subclauses will be covered in detail in Section 4 of this book. Figure 3.8
shows a word cloud of what to expect in the clause.

3.2.7 Clause 5: Software Development Process


Clause 5 is the heart of the standard. It comprises roughly 36% of the whole
standard and dives deep into how your software development is to unfold.
This is the clause that speaks to your software-development life cycle, and
that must match with your software-development process. The clause covers
development planning, requirements analysis, architectural design, detailed
design, unit implementation and verification, integration and integration
testing, software-system testing, and release. The clause and its subclauses
will be covered in detail in Section 5 of this book. Figure 3.9 shows a word
cloud of what to expect in the clause.

3.2.8 Clause 6: Software Maintenance Process


Clause 6 is the yin to the yang represented by the preceding clause of soft-
ware development. In other words, this is the Nick Nolte to your 48-hour
tour of the town with Eddie Murphy, or what you need to do after all the
IEC 62304 as the Flesh around the Bones ◾ 81

Figure 3.9 Word cloud for Clause 5.

frolicking of developing a piece of software. The clause comprises roughly


6% of the entire standard and covers topics such as maintenance planning,
problem and modification analysis, and modification implementation. The
clause may be easy to overlook during the initial excitement of software
development, but you will need to address these issues before you get an
82 ◾ Medical-Grade Software Development

Figure 3.10 Word cloud for Clause 6.

approving nod from your notified body, should you need it, and before you
release your product. The clause and its subclauses will be covered in detail
in Section 6 of this book. Figure 3.10 shows a word cloud of what to expect
in the clause.

3.2.9 Clause 7: Software Risk-Management Process


Clause 7 dives deep into the question of how ISO 14971 risk management
should take place in the context of software medical devices. The clause
comprises approximately 7% of the standard and covers topics such as anal-
ysis of software contributing to hazardous situations, risk-control measures
and their verification, and risk management of software changes. After a
series of fairly sequential clauses from 1 to 6, Clause 7 is the first clause that
folds into the previous discussion and wants to see risks managed through-
out the life cycle in a helpful way. The clause and its subclauses will be
covered in detail in Section 7 of this book. Figure 3.11 shows a word cloud
of what to expect in the clause.

3.2.10 Clause 8: Software Configuration Management Process


Clause 8 (approx. 5% of the standard) has the important task of introduc-
ing the concept of configurations (e.g., version-identified pieces of software
used in your product, and the version-identified product itself) and changes
to them. The requirements here look at your overall approach to configu-
rations, changes to configurations (incl. the software behind the version
number), and retaining an adequate history for the controlled items. The
IEC 62304 as the Flesh around the Bones ◾ 83

Figure 3.11 Word cloud for Clause 7.

Figure 3.12 Word cloud for Clause 8.


clause and its subclauses will be covered in detail in Section 8 of this book.
Figure 3.12 shows a word cloud of what to expect in the clause.

3.2.11 Clause 9: Software Problem Resolution Process


Clause 9 (approx. 5% of the standard) has the honor of closing out the
standard by instructing problem resolution regarding your medical-device
84 ◾ Medical-Grade Software Development

software. The clause covers topics such as problem reports, problem inves-
tigations, advising relevant parties, using your change-control process (from
Clause 8), maintaining adequate records, analyzing problems for trends, veri-
fying problem resolutions, and test documentation. The clause and its sub-
clauses will be covered in detail in Section 9 of this book. Figure 3.13 shows
a word cloud of what to expect in the clause.

3.2.12 Annex A – Rationale for the


Requirements of This Standard
The first of four informative-only annexes, Annex A, provides a brief ratio-
nale for the existence of the standard as a means of imposing an improved
level of control over the development of safe medical software. The notion
here is that the mere testing of software is not adequate to ensure its safety
in operation. Instead, the requirements of the standard are to be met in
designing the processes for software development and maintenance as is
appropriate for the risks associated with the type of software in question.
Adherence to these processes is even positioned as the primary requirement
of the entire standard by this annex. As prime directives go, this one really
should not catch you off guard like that secret fourth directive did Alex
Murphy towards the end of Robocop.
The 2015 amendment of the standard only made minor changes to the
annex. In addition to an updated table providing a non-normative overview
of which requirements map to which software safety classes, the new text
made four fairly cosmetic changes to the text. The changes made are not
worth commenting on here in any great detail, but if you find yourself in a

Figure 3.13 Word cloud for Clause 9.


IEC 62304 as the Flesh around the Bones ◾ 85

Figure 3.14 Word cloud for Annex A.

heated argument over whether software can contribute to hazards (old text)
or hazardous situations (new text), or whether the standard thinks it is easy
(old text) or just possible (new text) to assure class A software safety through
overall medical-device design activities and life-cycle controls, check here
for any discrepancies between you and your jousting partner. It is, however,
worth noting that the annex, too, expects such hazards to be a part of your
overall risk analysis activities.
Figure 3.14 shows a word cloud of what to expect in the annex.

3.2.13 Annex B – Guidance on the Provisions of This Standard


Annex B offers roughly 20 pages of commentary on the clauses of the
standard.
Figure 3.14 shows a word cloud of what to expect in the annex. The com-
mentary is not normative, as it does not place requirements on the adoption
of the standard, but it does speak to the intentions of its writers which may
or may not correspond with the views of medical device regulators.
The 2015 amendment of the standard made mostly terminological
changes to the annex (e.g., changing software product to “medical device
software”, “hazard” to “hazardous situation” in many places) and introduced
two pages of discussion on legacy devices. The contents of the annex are
discussed later in this book as needed, but for now, Figure 3.15 provides a
succinct overview of the topics discussed in the annex.

3.2.14 Annex C – Relationship to Other Standards


Annex C plots the place of the IEC 62304 in the 2015 world of standards.
Here the foundation for developing medical software is laid out as consisting
86 ◾ Medical-Grade Software Development

Figure 3.15 Word cloud for Annex B.

of the big management standards such as ISO 13485 for quality management
and ISO 14971 for risk management. On top of these you are expected to
employ safety standards as is appropriate for the type of devices you manu-
facture. Chief among the usual suspects here is IEC 60601-1 for hardware
and IEC 62304 for software. This much has not changed in the decade since
the standard was published. A good number of other potentially relevant
IEC 62304 as the Flesh around the Bones ◾ 87

standards have been published (e.g., see Section 12), and newer revised ver-
sions of many standards have been published since, but the basic concept
of fitting standards together and the biggest of the building blocks have
remained the same.

Figure 3.16 Word cloud for Annex C.


88 ◾ Medical-Grade Software Development

Figure 3.17 Word cloud for Annex D.


The 2015 amendment updated the references from 2006 but did not
alter the approach to using IEC 62304 in any significant way in Annex C.
Figure 3.16 provides a succinct overview of the topics discussed in the
annex.

3.2.15 Annex D – Implementation
Annex D is a brief two-page dip into how manufacturers can adopt the IEC
62304 standard in practice. The gist of the discussion is that medical-soft-
ware development is required to take place within the context of a quality
management system (see Section 4.1). This QMS does not, however, always
necessarily need to be certified in order for this standard to be adopted. If
a QMS or even a subset of it in the form of some relevant processes (e.g.,
regulation of development, verification, validation) exists before the adop-
tion of the standard, a gap analysis of some sort should be performed to see
how that QMS should be revised or whether the required processes should
be separately described. Maintaining separate processes may get to be more
trouble than it’s worth in the long term so our advice is to read this book,
and then marry the quality management system and software-development
requirements together in a practical way.
The annex also notes, in full agreement with ISO 13485, that the manu-
facturer retains responsibility for any outsourced software development. Note
though that the annex serves this sobering assertion with the provision that
the suppliers do not have a documented QMS of their own in place. ISO
13485 is not so lenient.
Finally, the annex offers up a sketch of a checklist for small manufacturers
without a certified QMS. This is, in practice, a table of contents for the stan-
dard with a checkbox and space for defining actions for following the manu-
facturer’s own assessment of how well the requirements of the clause are met
at the moment. The annex was not updated by the 2015 amendment. Figure
3.17 provides a succinct overview of the topics discussed in the annex.
Chapter 4

GENERAL REQUIREMENTS

This is where you are still infatuated with the new-car smell of your
fancy new ride but are silently slapped with a hefty insurance bill
for the privilege of actually driving it. Or in cinematic terms, you
are riding high after a dream wedding oblivious to the fact that your
spouse is a covert operative for some world government. In any case,
“general requirements” is a soft way of saying heads up this will
hurt. This is where you get to learn about the heavy lifting awaiting
you before that Schwarzenegger physique is looking at you from the
mirror. The good news is that you may already have a QMS and a
risk-management system in place, but even then, this clause may
have news for you regarding the software safety classification and
handling of legacy software.
General requirements are the topic for the entirety of Clause 4 in IEC 62304.
The clause comprises a deceptively slender 4% sliver of the standard. The
four main topics under discussion are:

◾ Quality management system (Clause 4.1)


◾ Risk management (Clause 4.2)
◾ Software safety classification (Clause 4.3)
◾ Legacy software (Clause 4.4)

Clause 4 could just as well be called “prerequisites”, “foundational require-


ments”, or “the forever plan to the perfect software pipeline”, but it is not.
Whatever the title of the clause, the amount of heavy lifting to meet its
requirements is formidable. The good news here is that your QMS will

DOI: 10.4324/9781003454830-4 89
90 ◾ Medical-Grade Software Development

already meet much of what is required. In other words, your quality- and
risk-management processes, as well as their instructed outputs, will form an
essential part of the documentation assessed in any IEC 62304 or ISO 13485
conformity assessment (see Section 11).
In the following, we will address each of the four subclauses in detail.
Note that the numbering of the following subsections corresponds to the
clause numbers in IEC 62304.

4.1 Quality Management Systems


Clause 4.1 of IEC 62304 requires that the manufacturer is able to dem-
onstrate their ability to consistently meet both customer and applicable
regulatory requirements. The implied requirement to implement a quality
management system (QMS) to do so is stated in the heading of the sub-
clause and in the non-binding notes to the clause. Quality management and
quality management systems (QMS) are the central topics of the ISO 13485
standard, and in practice best left for a discussion of that standard (see
Section 2).
The IEC 62304 standard, much like the ISO 13485 standard too, leaves
the door open to adopting other alternative national standards and regula-
tions on quality management systems. How you obtain the evidence for the
required demonstration is largely up to you, but you must have the evidence
and it should be based on the QMS you run. In practice, choosing some-
thing other than ISO 13485 as your bridgehead into QMS here will be a
Dunkirk waiting to happen, or a dodoist exercise in escapism. Don’t do it
unless you well and truly know the desert island you want to inhabit.
A further note on the clause offers ISO/IEC 90003 on guidelines for the
application of ISO 9001 to computer software as further reading on the inter-
section between quality management and software. Note, however, that the
standard in question has undergone several revisions since the publishing
of the IEC 62304 standard and that ISO 9001 too has been revised since the
ISO 13485 standard was forked from it. If your intention is to get certified to
ISO 13485 you should not lose sight of that goal here.
In practice, implementing the IEC 62304 standard as part of your organi-
zation’s overall QMS will lead to some software-specific requirements and a
process or two described as standard operating procedures (SOPs) in your
QMS (see Section 10). These requirements will be discussed in the coming
sections of this book.
General Requirements ◾ 91

4.2 Risk Management
The topic of risk management is addressed in Clause 4.2 of the IEC 62304
standard swiftly and mercilessly: you must adopt ISO 14971 risk manage-
ment. As elsewhere in the standard, except for the term definitions in Clause
3, the reference to ISO 14971 is made without specifying a version of the
standard which is to be interpreted as meaning the latest version of the stan-
dard published today. This requirement, and indeed the whole clause itself,
is only 12 words long, but the requirement is a profound one – at least for
now.
In later as-of-yet unpublished revisions of the IEC 62304 attempts have
been made to rework this clause in a way that is compatible with ISO 14971
but does not necessarily require the adoption of that standard. If these
changes came about, the general relationship between the two standards
would remain, but lower-risk health or wellness software could more read-
ily adopt IEC 62304 life-cycle processes without the need to also adopt ISO
14971 risk management. This would hopefully bring welcome rigor to the
development of non-medical software, and bring them within a more fea-
sible jumping distance if they then someday need to make the transition to
medical software.
The issue of risk management has been widely reported as one of the
biggest flashpoints in revising the IEC 62304 standard in the past years as
some experts see it as an absolute must and others as too much to ask of
every class of software. Time will tell where the chips fall on this matter, but
in general, adopting one standard should not really require the adoption of
also another standard – you claim conformity with the standards you meet
and not the standards those standards may link to in one way or another. If
you wish to adopt both IEC 62304 and ISO 14971 you may of course always
do so, and in general, this is likely to be a smart move, but bundling the
two standards and asking manufacturers to take it or leave it seems like a
bridge too far.
In practice, implementing the IEC 62304 standard as part of your orga-
nization’s overall QMS and risk-management activities will lead to some
software-specific requirements in your product-related risk-management
activities. These requirements may affect your standard operating procedures
for risk management and product realization, SOP-8 and SOP-10 as dis-
cussed in our QMS book (Juuso 2022).
As a final practical note on reacting to new versions of any standard, you
will want to apply consideration and restraint in adopting any new version.
92 ◾ Medical-Grade Software Development

Adopting a new version is usually not just a copy-replace operation in MS


Word. You will want to ensure you know what has changed in the new
version, how this affects your operations, and how you will go about imple-
menting the necessary changes before you contemplate pulling the trigger.
You won’t have all the time in the world to ponder the change but don’t
jump in blindfolded either.
In the IEC 62304 standard, product-specific risk management is addressed
in greater detail in Clause 7 (see Section 7). Clause 7, as you will later
see, makes heavy use of something called a software safety classification
(explained in the next section) to attach more requirements on higher-risk
software and allow lower-risk software to be developed more speedily. It is,
however, important to note that Clause 4.2 here expects to see risk manage-
ment take place according to ISO 14971 for all classes of software regardless
of their software safety classification.

4.3 Software Safety Classification


Clause 4.3 of the standard is dedicated to the essential topic of software
safety classification, or how to address different levels of concern in software.
Think of this classification as setting up the fault lines in the tectonic plates
that make up the standard and as your key to finding the right combination
of activities that unlocks the vault of medical software development for you.
The ABC classification described here will cut through the implementa-
tion of Clauses 5 through 9 of the standard based on the risk profile of the
software you are developing. The idea here is much the same as in the risk-
based approach employed in ISO 13485 quality management: resources are
allocated to, and more scrutiny is placed on, areas where risks are greater.
As a result, low-risk software and software components may be classified as
belonging to class A and thus dealt with in a simpler fashion than their high-
risk counterparts which are classified into class C.
The formula for assigning the safety classification is quite straightforward
but figuring out the harms and risks associated with the device, and tracking
these down to individual software items sooner or later, will require effort.
In the following, we will first look at the basic formula for assigning soft-
ware safety classifications and then look at some fundamental concepts and
interpretations that feed into arriving at the correct safety classification. Later
in Section 7 we will also look at how some of these same stipulations bleed
over to the actual risk management of your device.
General Requirements ◾ 93

4.3.1 Classification Formula
The 2015 amendment offers a helpful workflow to assign a safety classifica-
tion to software or a software component. This workflow can be summa-
rized as the following four rules:

◾ All software is class C unless otherwise successfully argued. In


other words, you can’t, for example, forget to perform the classification or
lose documentation and say that the software was something else than C.
◾ If software failure can’t lead to a hazardous situation, or if
that failure doesn’t result in unacceptable risk after mitigation
the software is class A. In the latter case, the risk-control measures
(mitigations) are to be external to, or independent from, the software
in question, and you must evaluate their effectiveness. The definition of
acceptable risk is for you to argue beforehand and in conformance to
ISO 14971 (see that standard’s Clause 4.2 on management responsibility
in particular).
◾ Otherwise, if the severity of the possible harm is non-serious
the class is B.
◾ Otherwise, if the severity of the possible harm is serious the
class is C.

The above rules may be condensed into the following simple decision dia-
gram (see Figure 4.1) that leads to the assignment of the ABC safety classifi-
cation after answering a maximum of three questions. Remember that until
these questions are answered the classification is C, that all software may
be expected to fail eventually, and that only external mitigation (risk-control
measures) may be considered here. Also remember that, as explained below,
any child components will inherit the safety class of their parent unless oth-
erwise successfully argued.

Figure 4.1 Key to safety classification.


94 ◾ Medical-Grade Software Development

In practice, when assigning a software safety classification, the manu-


facturer must consider the possible ways the operation of the software
may become flawed, what failures may be caused by its design, and also
the results of the risk-management activities concerning the product. Note
that here harm is only considered from the point of view of the patient,
user, and other humans (unlike what is defined for harm in ISO 14971; see
below). As noted above, the default software safety classification is C (cor-
responding to the highest risk) until a lower classification has been assigned
based on risk-management activities and a valid rationale. To use the above
classification formula, we will need to discuss the definition of harm and the
sources of harm. This is addressed next.

4.3.2 What Is Harm?
It is noteworthy that when assigning a software safety classification, the
present version of IEC 62304 only considers risk from the point of view of
harm to the patient, operator, or other people (Clause 4.3a). Harm to equip-
ment or the environment is not part of the prescribed picture here even
though these are a part of the definition of harm borrowed from ISO 14971
in Clause 3.8. In this setting, harm is considered as resulting from a hazard-
ous situation the software can contribute to in a worst-case scenario. The
implication here is that the software safety classification is based on some-
thing less than the full ISO 14971 risk matrix you may have built – and only
looks at the harms related to people.
In the above, the term “hazardous situation” (Clause 3.35) refers to cir-
cumstances wherein people, property, or the environment are exposed to
hazards. A “hazard”, then, is a potential source of harm (Clause 3.9), and
“harm”, in general, is some physical injury and/or damage to the health of
people, property, or the environment (Clause 3.8). Note that here the defi-
nition of harm is now wider than that used in the context of the software
safety classification where only harm to people was discussed. Subsequently,
software malfunctioning and causing a device to constantly overheat and
thus shorten its lifetime might not be a factor in assigning the software
safety classification, but if the decreased lifetime or device overheating
caused harm to the patient it would feature in the risk management of the
software. Similarly, for following green-coding practices and zero emissions
goals the risk might be worth tackling even if no direct patient harm was
involved. In practice, the incongruity here in the competing definitions of
harm is appreciable but not earth-shattering.
General Requirements ◾ 95

Finally, the seriousness of harm is given a definition in Clause 3.22,


where a serious injury is defined as an injury or illness that is a) life-threat-
ening, b) results in permanent impairment of a body function or permanent
damage to a body structure, or c) necessitates medical or surgical interven-
tion to prevent such impairment or damage. Again, note that due to Clause
4.3a only harm to people should be considered while assigning a software
safety class, but in practice, you will want to consider all types of harm in
some adequate way in your overall risk-management operations.
Similarly, if interpreted strictly, the body of Clause 4.3a actually speaks
of software contributing to an undesired outcome, not just a software fail-
ure contributing to it. The new flowchart added to the 2015 amendment
in Figure 3, however, clearly speaks of software failure as the trigger. The
wider focus appears to be reflected by regulatory expectations that your risk
management also covers harm resulting from the correct use of the software
within its intended use, as well as some foreseeable misuse of the software,
in addition to harm caused by software failures.

4.3.3 Accept That All Software Will Fail


In discussing the assignment of the software safety classification, the IEC
62304 standard takes a rather pessimistic view that sooner or later all soft-
ware will fail. That statement might conjure up images of the charmingly
depressed robot Marvin from The Hitchhiker’s Guide to the Galaxy, but here
the point appears to be a solemn realization that you should not expect
software to behave exactly as you programmed it. It may all be zeros and
ones behind that code you wrote, but every once in a while, something
unexpected will happen. It would be similarly appropriate to adopt the
stance that eventually all hardware will fail too, but here the goal is to not
get lulled into believing that software is perfect.
Accordingly, Clause 4.3a of the standard is adamant that in all cases the
probability of a software failure occurring is set at 100%, in other words, all
software will eventually fail. This does not, however, mean that all failures
lead to harm. Many organizations have thus opted to split probability into
two separate values, a P1 probability for a software failure (this is always
100%) and a P2 probability of that failure leading to harm (0–100%). Anyone
familiar with probability mathematics will perhaps scoff at this solution as
it does not change the outcome of the calculation, but the approach has
proven practical in the real world. Note that while IEC 62304 discusses the
matter in the limited context of software safety classification assignment,
96 ◾ Medical-Grade Software Development

this approach is compatible with the instructions given in Annex C of ISO


14971:2019 concerning the actual risk-management activities (see Section 7).

4.3.4 Use of External Risk Controls


Clause 4.3a suggests that software with an initial safety classification of
higher than class A may implement external risk controls to achieve a lower
safety classification. Here, Clause 4.3a is adamant that only risk-control mea-
sures (mitigations) external to the software in question may be considered
when assigning a safety classification to software (i.e., software system) or a
software component (i.e., software item, software unit). These may involve
redesigning the system architecture around the software component being
classified via, for example, the following measures:

◾ Hardware measures
◾ Independent software system measures
◾ Healthcare procedures
◾ Other means to minimize the ability of the software to contribute to a
hazardous situation

Note that independent software system measures are in fact considered


external to the software in question. In other words, it is incorrect to
assume that no risk-control measures could be implemented in software
or that these would somehow not count in your software safety classifi-
cation assignment or indeed the subsequent risk-management activities.
Independence, however, is a key property here: if a failure in one place can
lead to a failure elsewhere these two locations can’t be considered indepen-
dent. The rationale for segregation thus becomes equally important. Think
of the bulkheads on the Titanic getting overflown by water, one by one as
the ship tilted and sank, each flooded section weighing down the next and
pulling it under. Think of that auxiliary power system that brought back the
Terminator to finish that final fight after the glow in his eyes had already
been extinguished. Neither system was totally independent of the compo-
nent that failed in the first place, but the latter was independent enough
to save the day. Another example here might be a software watchdog in a
patient monitor that has been intended to ensure the software continues to
run in a responsive manner. However, an overly aggressive timeout set-
ting in the watchdog might cause the monitor to reboot unnecessarily, and
thereby perhaps miss a critical event in the vital signs of the patient. In
General Requirements ◾ 97

this case, the mitigation could thus cause the very harm it was intended to
prevent.
Finally, a non-binding note to Figure 3 in Clause 4.3a offers commentary
on risk-control measures implemented within a software by saying that the
software may fail, this may then contribute to a hazardous situation, and the
harm arising out of the failure may then also include the very harm the risk
control measure was designed to prevent. This realization, too, harks back
across a century to the above example provided by the Titanic.

4.3.5 Reducing Both Probability and Severity


A non-binding note to Figure 3 in Clause 4.3a clarifies that risk-control
measures may reduce both the probability that the software failure results in
harm and/or the severity of that harm. This too is a fundamental realization
in addressing the risks involved in the use of your medical software, and
one which is frequently asked about when discussing the risk management
of medical devices.
Note that when using the P1xP2 approach introduced above it is not
possible to lower the P1 probability as this is always set to 100% (or 1.0,
depending on your notation) to reflect the assertion that software will
always fail. It is thus left to us to consider reducing P2 via suitable means.
Figure 4.2 below shows an example.
In the figure, the probability of occurrence of harm is controlled via
mitigations such as including fault diagnostics. As a result of the risk control
introduced the resulting risk is lowered to an acceptable level (not shown

Figure 4.2 The use of P1 and P2 in a risk matrix.


98 ◾ Medical-Grade Software Development

in the figure). See Section 12.4 for more discussion of the Failure Mode and
Effects Analysis (FMEA) methodology.

4.3.6 Recording and Inheriting the Classification


Clause 4.3d instructs that a child component (be that a software item of a
software system, or a sublevel item underneath another item) will inherit the
safety classification of the parent unless the manufacturer documents a ratio-
nale saying otherwise. The rationale must also explain segregation between
the components so that they may be judged separate for classification. The
resulting safety classification for each such component (i.e., software item)
must be documented somewhere in your documentation if it differs from
that of the parent component (Clause 4.3e). Remember that according to
Clause 3.25 all levels of the software decomposition hierarchy – software
system, software item, and software unit – may be considered a software
item. Thus, the basic approach here should be to document the safety classi-
fication of each component if it is seen to be different from that of its par-
ent. Remember that you must also provide a valid rationale for this decision,
along with why the segregation between the items can be relied upon.
Similarly, the resulting safety classification for each software system
must be recorded in the risk-management file (see Section 10). Clause 4.3g
reiterates that this classification is always C until another software class is
assigned. Note that the safety classification of software items other than the
software system does not have to be in the risk management file per se
(Clause 4.3e). In practice, you will want to record the classification and the
logic for it somewhere, but this may, for example, be in the architectural
design and not in a risk-management document. Beware of repeating your-
self in several documents, though, as this will cause you gray hairs if the
classification of some component then needs to be revised and this leads to
editing an array of documents.
Finally, Clause 4.3f states that when evaluating the compliance of a group
of software items to this standard, the highest-classified item in the group is
taken as a yardstick – the implication being that the high-risk components
can’t hide behind the backs of the lower-risk items, not that low-risk items
are suddenly treated as high-risk items. The requirements placed on the
software (i.e., processes and tasks required) according to that highest safety
classification will then be observed over the entire group. Alternatively, a
rationale for using a lower classification may be documented in the risk
management file.
General Requirements ◾ 99

4.3.7 Final Remarks
Overall, the ABC software safety classification is a great concept, but enthu-
siasm for it has been dampened by its equation by developers with not so
much the level of concern, but the amount of documentation required for
the software. The classification has jokingly been referred to as the ABC
classification on the amount of documentation.
Some organizations have concluded that it’s easier to document everything
based on class C and thus avoid the need to keep track of any classifica-
tion peculiarities or train developers to know two or three different path-
ways through the processes. This approach is perhaps further justified by the
expansion of some requirements from higher classes down to class A software
items in the 2015 amendment. Using a one-size-fits-all approach here may also
make it more straightforward to adapt to any upcoming changes in safety clas-
sification due to new versions of the standard, but this may not always make
sense or be the most optimal use of your resources. Flattening the classifica-
tion also flies in the face of the general risk-based approach advocated by
ISO 13485 in prioritizing your use of resources to make the maximum impact
where it truly matters. As a mechanism to define requirements for different
types of software the safety classification is, however, a sound approach and
one which you can navigate based on your specific circumstances.
Also, remember that the standard does expect a software safety classifica-
tion to be defined for both the entire software system and for the individual
software items it consists of. This classification is to be appropriate based on
the risks associated with the software, and it is generally expected that a cli-
nician has been involved in the determination. If sufficient clinical expertise
is already involved on the risk-management level, and this software safety
classification is based on that activity, it may be debatable whether a clini-
cal expert is separately needed here. In any case, the hook into your risk-
management activities does mean that a mere statement of “this feels like a
class B software” is not enough. Not having a classification, or not having an
adequate rationale for it is a sure way to step on a mine during auditing.
In practice, you’ll need to ensure that you have mapped your SOPs and
product documentation to the expectations of the IEC 62304 standard,
which it has laid out using the ABC classification. This book will help you
accomplish this goal by discussing each area of the development process in
turn. In terms of making the smallest incision into your product realization
SOP, you could write that you only use class C from the classification but
then you must be prepared to swim at the deep end of the standard. This is
100 ◾ Medical-Grade Software Development

equivalent to going yachting on the USS Missouri, if not the USS Arizona –
not to make light of a dreadful incident. This may not be optimal for using
your resources strategically.
We expect Clause 4.3 to be clarified in later editions of the IEC 62304
standard. The scope of the application for the assertion of P1 to always be
100% is not entirely clear or helpful as it is occasionally perceived to also
apply to Clause 7 and the overall risk management of the device even if this
is not necessarily the intention of Clause 4.3 here.

4.4 Legacy Software
Clause 4.4 was added in the 2015 amendment to provide a standardized
way for addressing legacy software despite differences in their development
model and any trivial deficiencies in their documentation. Think of this as
the Danny de Vito subclause to Clause 4, or the good cop in the duo, even
if that slightly inverts the roles Schwarzenegger and de Vito play in their
joint ventures. The aim of this subclause is, in any case, to bring you up
to speed even if the circumstances surrounding your early life don’t quite
match up with those of your more fortunate sibling. You can overcome
issues in your genes and upbringing if you are smart about it all, just like in
that 1988 action comedy by Ivan Reitman.
In essence, the clause provides an alternative streamlined pathway for
legacy software to achieve compliance with the IEC 62304 without meet-
ing all the requirements in its Clauses 5 through 9 (stated in Clause 4.4.1).
Instead, the manufacturer of such software must perform a gap analysis
against certain clauses of the standard, and then act to close the identified
gaps. The term legacy software is given an unequivocal definition in Clause
3.36 as medical-device software legally placed on the market, and still mar-
keted today, but lacking sufficient objective evidence that it was developed
in compliance with the current version of IEC 62304.
The discussion in the clause is divided into four topics: risk management-
activities, gap analysis, gap-closure activities, and rationale for use of legacy
software. These topics will be discussed next.

4.4.1 Risk-Management Activities
Building on from Clause 4.2, Clause 4.4.2 makes two further stipulations.
Firstly, Clause 4.4.2a requires that the manufacturer assesses any feedback
General Requirements ◾ 101

and post-production information, both from within and without their orga-
nization, regarding incidents and near-incidents. Secondly, Clause 4.4.2b
requires that the manufacturer manages risks associated with the continued
use of the software. The latter involves performing risk management activi-
ties that consider the following aspects:

◾ The integration of the software in the overall medical-device


architecture
◾ The continuing validity of risk controls implemented as part of the soft-
ware (incl. controls for each potential cause for the software contribut-
ing to a hazardous situation)
◾ The identification of hazardous situations (incl. the identification of
potential causes for the software’s contribution to these).

In other words, you must ensure that you are keeping up to date on how
your software is performing by receiving feedback, assessing other post-mar-
ket information, keeping abreast of any changes to the host device (if any),
and ensuring your risk mitigation continues to be appropriate. The implica-
tion here is that you must meet the requirements both when bringing legacy
software to the market, and over the long term of its use. The clause does
not speak of the decommissioning of the software or its host device, but that
too may be a subject to consider.

4.4.2 Gap Analysis
Clause 4.4 on legacy devices is not meant to give you a free ticket to comply
with the standard without worrying about the other 79% of the standard. To
analyze the gap between your existing status and the expectations of the
standard you as the manufacturer are required to perform a gap analysis.
Gap analysis, in case you are not yet familiar with it, is a popular term
for just about any situation where you need to look at where you are, where
you need to be, and what steps you need to take to get there. It is probably
one of the very first terms you’ll come across when first getting into medical
devices and as an answer to a question it is about as satisfying as admitting
you have a problem. A gap analysis is a careful thought you must put in to
assess the situation and figure out what needs to be done.
Clause 4.4.3 of the standard provides a helpful limitation on the gap
analysis needed here by instructing you to look at how your available deliv-
erables meet the requirements placed on them by the following clauses:
102 ◾ Medical-Grade Software Development

◾ Clause 5.2 on Software requirements analysis


◾ Clause 5.3 on Software architectural design
◾ Clause 5.7 on Software system testing
◾ Clause 7 on Software risk management

In the above, when the standard speaks of available deliverables don’t


be fooled into thinking that some deliverable that you don’t have but is
expected by the standard would not be a problem (although see Clause
4.4.3b below). You must meet the requirements of the standard, not the
other way around. You are, of course, expected to utilize the ABC safety
classification of your software when analyzing the requirements. The listed
clauses themselves will be covered in detail in the following sections of this
book (see Sections 5 and 7).
Together the above four clauses make up 21% of the standard, which
will represent a sizable reduction in the list of requirements to fit into your
brain’s working memory. Do note, however, that the whole clause on risk
management is included on the list even though it’s featured as a separate
requirement in Clause 4.4.2 above. Similarly, Clause 6 requirements on
maintenance apply should any changes be required on the legacy software,
and Clause 9 requirements on problem resolution may also be applicable
sooner or later. The cherry on the cake is the requirement for a quality
management system discussed above (Clause 4.4.1). This will mean that
the walls of the carefully crafted legacy software silo may be breached at
will if any cracks appear during audits, inspections, vigilance, or customer
feedback.
For all the above reasons, we would not be content with reading the sub-
clause on legacy software and thinking we are ready to push such products
out onto the market. Every gap analysis should perhaps come with at least
a stopgap consideration of fallback positions and countermeasures should
you need to adjust your position later. Remember that IEC 62304 may only
be a semi-voluntary test report, but the software you put out in the context
of medical devices will also be a major point of discussion in any regula-
tory audits. Your business considerations may also change over time. It pays
to think ahead, but this consideration will also go a long way to hardening
your reasoning on any point, and that will have both internal and external
benefits.
Clause 4.4.3 only places three immediate requirements on the manufac-
turer regarding the gap analysis. You are required to:
General Requirements ◾ 103

◾ Assess available deliverables for continuing validity (Clause 4.4.3a)


◾ Evaluate the potential risk reduction gained by generating any missing
deliverables (incl. performing the associated activity) when gaps are
found (Clause 4.4.3b)
◾ Based on the above evaluation, determine deliverables and activities to
be run (Clause 4.4.3c)

The clause defines software system test records as the minimum level of
deliverables (see Clause 5.7.5) it expects to see. As a further non-binding
note, the standard also adds that the gap analysis should ensure any risk
controls implemented in the legacy software are included in the software
requirements.

4.4.3 Gap Closure Activities


You will probably have heard the viral Albert Einstein quote that if he had an
hour to save the world he would spend 55 minutes of that defining the prob-
lem and then the remaining 5 minutes solving the problem. This is all well
and good, but the quote says nothing about actually implementing the solu-
tion. It might thus be better to go out for one last pint than to leave the fate of
the world in the hands of a theoretical scientist, as brilliant as he was. Clause
4.4.4 will have none of this impractical grandstanding as it insists that the gap
analysis made above is also acted on. This is where after all the what-ifs and
daydreaming of the previous clause we get to go to the mattresses, to do the
work required. This is where after all that talk of melting the one ring in the
fires of Mordor we start putting one foot in front of the other.
The clause requires the manufacturer to establish and execute a plan to
act on the identified gaps. This plan may be combined with the software
maintenance plan (see Clause 6.1 in Section 6), as the standard suggests in
a non-binding note. The standard takes an admirably practical stance by
stating that missing deliverables may also be created based on objective
evidence, instead of running the associated activities retroactively (Clause
4.4.4a). What that objective evidence may be is not discussed by the stan-
dard, but, in practice, accumulated post-market surveillance data including
information on similar products, maintenance history, and customer feed-
back may provide answers here.
The plan made must address problem resolution for any problems
detected in the software or deliverables, and it must do this in accordance
104 ◾ Medical-Grade Software Development

with Clause 9 (see Section 9). Finally, any changes to the software must be
performed in accordance with Clause 6 (see Section 6). In other words, sail-
ing on a legacy yacht is a relatively smooth business until it’s not. As soon
as you are faced with issues or want to make changes you are again a few
steps closer to working with the full standard and all modern expectations.
Rightly so, but this will cause some friction in maintaining legacy software
out there on the seas.

4.4.4 Rationale for Use of Legacy Software


Finally, Clause 4.4.5 requires the manufacturer to document the version of
the software and a rationale, based on the outputs of Clause 4.4, for its con-
tinued use. A non-binding note clarifies that fulfilling the requirements of
Clause 4.4 enables further use of the software in accordance with IEC 62304.
Chapter 5

SOFTWARE DEVELOPMENT

Whatever your chosen software development model is – that c­ lassic


waterfall so strikingly depicted in The Last of the Mohicans or
Deliverance, some flavor of the now popular agile model as perhaps
neatly illustrated by the constantly mutating Fast & Furious saga, or
something in between – this is where the in-depth software-specific
requirements for it are located.
By far the biggest clause of the IEC 62304 standard, Clause 5 (36% of the
standard) dives deep into the actual bread and butter of the standard, the
development of software. This is the heart of the standard. Here the stan-
dard sketches out the life cycle of software development from requirements
to release – and even adds a preliminary step of development planning
which does not feature in the definition of the term development life cycle
earlier in the standard in Clause 3.24 but is very much a part of the software
business.
The topics covered in this clause are given in Table 5.1 below. The table
also notes which software safety classes each topic is applicable to and adds
a final section on the aspects not covered by IEC 62304 but expected by ISO
13485. Note that the numbering of the following subsections corresponds to
the top-level clause numbers in IEC 62304.
The clause immediately has apparent overlap with the ISO 13485 stan-
dard on quality-management systems, especially with the equally massive
Clause 7 of that standard on product realization (34% of that standard).
Clause 5 is also the first clause of the IEC 62304 standard where the soft-
ware safety classification introduced in Section 4.3 is used to set different
levels of requirements for low-risk and higher-risk software items. Note that

DOI: 10.4324/9781003454830-5 105


106 ◾ Medical-Grade Software Development

Table 5.1 Topics of Software Development


# Topic A B C
5.1 Development planning ☒ ☒ ☑
5.2 Requirements analysis ☒ ☑ ☑
5.3 Architectural design ☐ ☒ ☑
5.4 Detailed design ☐ ☒ ☑
5.5 Unit implementation and verification ☒ ☒ ☑
5.6 Integration and integration testing ☐ ☑ ☑
5.7 Software system testing ☑ ☑ ☑
5.8 Release ☒ ☑ ☑
5.9 The parts left out by IEC 62304 ☑ ☑ ☑
☑ = all apply, ☒ = some apply, ☐ = none apply

the safety classification may be applied to a software system, software item,


or a software unit, and all of these are to be considered software items (note
to Clause 4.3g).
Both standards talk of development stages in similar, compatible ways,
but where the more generic ISO 13485 talks of design and development
(D&D), thus remaining applicable to all sorts of products besides just soft-
ware, the IEC 62304 standard speaks of software development specifically.
IEC 62304 does occasionally refer to the wider system around the software
and asks that software development occurs in sync with the development of
the host device, but otherwise the details of developing non-software prod-
ucts are left outside the discussion.
Table 5.2 below shows a suggested mapping between the stages expected
by either standard. Note that the purpose of the table is to identify related
stages (and thus clauses) in the two standards, but not to suggest in any way
that one standard would cover the requirements of the other.
The complementary nature of the two standards is particularly visible
in the way IEC 62304 adds meat between D&D inputs (ISO 13485 Clause
7.3.3) and outputs (ISO 13485 Clause 7.3.4), and thus provides real structure
to the development activities of software products. In other words, the ISO
13485 standard does not instruct how you proceed from D&D inputs to
the outputs, it only wants to make sure the output makes sense consider-
ing the input, and that your reviews (including those based on your testing,
Software Development ◾ 107

Table 5.2 Alignment of ISO 13485 and IEC 62304 Clause 5


Stage ISO 13485 IEC 62304
Planning 7.1 Planning of product 5.1 Development planning
realization
7.3.2 D&D planning
Inputs 7.3.3 D&D inputs 5.2 Requirements analysis
Design and Surprise! The actual value- 5.3 Architectural design
implementation adding work that takes place 5.4 Detailed design
between D&D inputs and 5.5 Unit implementation and
outputs is not covered by the verification
ISO 13485 standard. 5.6 Integration and
integration testing
5.7 Software system testing
Outputs 7.3.4 D&D outputs Elements of Clauses 5.1–5.8
apply here, and vice versa.
Review 7.3.5 D&D review
This is explained in Sections
Verification 7.3.6 D&D verification 5.1–5.9 below.
Validation 7.3.7 D&D validation
Transfer 7.3.8 D&D transfer 5.8 Release

verification, and ultimately validation activities) confer this regardless of


whether these are single big reviews or a series of smaller assessments.
The verification and testing activities in IEC 62304, on the other hand, are
discussed iteratively from the level of software units all the way up to the
level of software systems. These stages are related to the review, verification,
and – perhaps to a certain extent – also the validation activities discussed
by ISO 13485, but as the focus in ISO 13485 is on the final product these
are best discussed separately. Remember that IEC 62304 does not actually
cover the validation and final release of the final product (medical device),
not even when the said product consists entirely of the software developed
(Clause 1.2).
Here, between D&D inputs and outputs, is where you can pull in the
experience from your past development work, maybe even from business
areas other than medical devices, and build a subprocess that you know
works for your organization and matches with the expectations of both stan-
dards. If you don’t like the waterfall model of development, you can design
your process here according to agile or whatever sound development model
you have. Some flavor of agile is the future, that much appears to be certain
108 ◾ Medical-Grade Software Development

in the development of software. ISO 13485 thus provides the bookends and
IEC 62304 the bookmarks in between but no specific development model is
forced on you in between by either standard, as depicted in Figure 5.1.
In the following, the discussion is structured around the clauses of the
IEC 62304 standard. The requirements of each standard, and the relation-
ship between these is discussed in detail. Each section then concludes with
a synthesis of the requirements. Note that unless a specific software safety
classification is stated in the following the requirements apply to all classes
of software and software items.

5.1 Development Planning
Before you embark on a software development project some plan-
ning is involved. That is unless you are John J. Rambo, and you
want to make up your way through that remote town as you go
– if you do decide that is you, be prepared for some unnecessary
fights, heated conversations, and even jail time along the way.
Development planning is when you will get to outline your path
from an idea to a finished product. Your horizons may never be as
wide as during the planning stage, as Figure 5.2 illustrates, so enjoy
this relative calm before the coming whirlwind.
As they stand, both standards acknowledge the need for planning. IEC 62304
sees planning as looking a little different for low-risk software than high-risk
software, but mostly the requirements placed on development planning here
are identical for all classes of software. ISO 13485, too, is compatible with the
notion that something high-risk requires more planning and checking than
something that is low-risk, but it does not provide a safety classification for
your use here in figuring out what requirements apply and how. Instead, ISO

Figure 5.1 Both standards look to provide bookends and bookmarks for your work.
Software Development ◾ 109

Figure 5.2 During planning your horizons are at their widest.

13485 talks of a risk-based approach that allows you to focus most of your
resources on what matters the most (see Juuso 2022).
The relevant clauses of both standards are:

◾ ISO 13485 Clause. 7.1 Planning of product realization


◾ ISO 13485 Clause 7.3.2 Design and development planning
◾ IEC 62304 Clause 5.1 Software development planning

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a solution to meet the requirements
of both. As the backdrop for your software development is necessary in the
QMS you have (Clause 4.1), we will start with ISO 13485 and then move to
IEC 62304 in the discussion. The following discussion will identify which
subclauses apply to which software safety classes.

5.1.1 Expectations from ISO 13485


5.1.1.1 Process Documentation
The baseline for software development planning is defined in Clause 7.1 of
ISO 13485. Here you are required to have at least one process for product
realization, and this is to also match consistently with the other processes of
your QMS.
You must also incorporate risk management into the process, but the use
of ISO 14971 is only suggested in a non-binding note to the clause here.
Your regulatory environment may, however, have stronger feelings about the
110 ◾ Medical-Grade Software Development

necessity of adopting the ISO 14971 standard in at least some of the opera-
tions. In general, the further away you get from the risk management for
your product – e.g., from product risks to product-realization process risks
to other process risks to general business risks – the lighter the weight of
the standard should be. This is not to make light of risk management, it is
certainly an essential part of everything you do regarding finished products,
but it is not all that you need to know to set up your pipeline for medical-
grade software development.

5.1.1.2 Planning Documentation
The name and format of your planning documentation is left up to you, but
you are instructed to determine, as appropriate, the following:

◾ Product requirements (e.g., intended use, applicable regulations, quality


objectives)
◾ Special resource needs (e.g., processes, documents, work environment,
or personnel competence)
◾ Product-specific activities (e.g., verification and validation, monitoring,
measurement, inspection, handling, storage, distribution, and traceabil-
ity) and product-acceptance criteria

ISO 13485 adds additional requirements in Clause 7.3.2 addressing design


and development planning. Here you are required to both plan and control
the design and development of your product, and to ensure that planning
documents are not only created but also maintained up to date throughout
the process. Your plans must address your D&D stages (incl. review, verifi-
cation, validation, and design-transfer activities), the overall D&D responsi-
bilities (incl. authorities), and methods for traceability from D&D outputs to
D&D inputs. This clause also calls for planning documentation to address
relevant resource needs (e.g., personnel competence).
Clause 7.3.6 further clarifies that your verification plans must include
methods, acceptance criteria, and as appropriate, statistical techniques with
a rationale for sample size. Clause 7.3.7 takes the exact same stance on
your validation plans. Remember that the standard here is also consider-
ing the manufacturing of physical products that are designed once and then
churned out time and again off the production line. Unlike with software,
with hardware products it will not be trivial to check whether each produc-
tion run is identical to the last. With software products acceptance has more
Software Development ◾ 111

in common with the acceptance of the build or the original binary than with
the acceptance of each clone known to be identical to its previous copies.
Note that both clauses, 7.3.6 and 7.3.7, stop short of requiring a docu-
mented process here and instead talk of working within your planned and
documented arrangements. You are required to have procedures for D&D,
but how much detail is in your SOPs may be debatable. In general, write
your SOP on the level of your product portfolio, but plan to add neces-
sary product- or project-level detail under that via planning documentation.
You may also consider splitting your SOP into sections or even individual
documents by the different types of products you make, e.g., software
versus hardware. Your choice here will depend on your product portfolio,
the number of each type of product-realization projects you regularly run,
and the number of adjustments you expect to perform from one project to
the next. In any case, keeping the SOP on an adequately high level should
help you in your everyday operations. Keep in mind though that something
too abstract will need to be translated into practice every time it is run,
whereas something too micromanaging will be easier to follow but may lead
to silently stifling the variation that occurs in real life. In other words, the
former may cause sweating during the everyday and the latter during the
last stages of development when hidden issues creep up. The sweet spot is
somewhere in between.
Finally, in reviewing and approving the planning documentation you
would do well to also ensure you meet the requirements of ISO 13485 for
the conducting of D&D reviews as per Clause 7.3.5 (see Section 5.9.1). If you
find that you need to revise some SOP that will be a process of its own (see
Juuso 2022).

5.1.2 Expectations from IEC 62304


5.1.2.1 Process Documentation
Compared with the ISO 13485 standard, IEC 62304 is a little less concerned
with a process definition and more concerned with a software development
plan (or plans) to ensure appropriate development planning takes place in
accordance with the requirements made in the whole of its Clause 5. You
should thus have a process in place for your product-realization activities,
but the IEC 62304 would be happy with this process being described in the
planning documentation itself (e.g., in the software development plan as
stated in the non-binding note 4 to Clause 5.1.1). IEC 62304 too appreciates
112 ◾ Medical-Grade Software Development

the existence of a wider quality management system (QMS) around your


software-development activities, and thus nothing prevents you from using
processes defined in your QMS when developing your planning documen-
tation. You are also required to map your software-development life-cycle
model to the expectations of the standard, as explained in its introduction
(see Section 3.2.2).
As a QMS is in fact required by the standard, and as that is likely to con-
form to ISO 13485, your best bet here is to ensure the standard operating
procedures (SOPs) you have written based on ISO 13485 also answer to the
requirements of IEC 62304 Clause 5.

5.1.2.2 Planning Documentation
IEC 62304 really wants to see a document called a software development
plan. The name of the document can be something different, but you
will be asked for such a document repeatedly so remember where you
have the equivalent information in case you call the document something
else.
This is basically your project plan for developing the software – although
you may also have a separate higher-level software-system development plan
if the two are not combined. The software development plan is principally
instructed in Clause 5.1.1 and elaborated on all throughout Clause 5.1. The
plan is given the following fundamental requirements.

◾ The plan must cover the activities of the clause in a way that is appro-
priate to the scope, magnitude, and safety classification of the software
being developed.
◾ The plan must fully define or reference the software-development life-
cycle model.

Furthermore, the plan must address the following aspects of the develop-
ment project

◾ Standards, methods, and tools C


The use of standards, methods, and tools associated with the devel-
opment of class C software items (Clause 5.1.4, class C only).
◾ Consideration of system requirements
The inclusion of system requirements as an input for software devel-
opment (Clause 5.1.3a), and the procedures for coordinating system and
Software Development ◾ 113

software development via, e.g., system integration, verification, and vali-


dation (Clause 5.1.3b). Note that the system requirements and software
requirements may in some cases be the same (e.g., standalone software,
software-only device).
◾ Development processes employed
The development processes to be used (Clause 5.1.1a). Note that
here a reference to your SOPs for product realization as per ISO 13485
should come in handy. Remember that you must also address the verifi-
cation tasks required for each life-cycle activity (Clause 5.1.6b).
◾ Deliverables expected and their verification
The deliverables (incl. documentation) of the activities and tasks
(Clause 5.1.1b). Furthermore, the plan must identify those deliverables
requiring verification, the milestones at which verification takes place,
and the acceptance criteria needed (Clause 5.1.6a, c–d).
◾ Documentation
The documentation produced during the development life cycle
(Clause 5.1.8). This must include a) the title, name, or naming conven-
tion; b) purpose; and c) procedures and responsibilities for develop-
ment, review, approval, and modification. Note that here a reference
to your SOPs for document control as per ISO 13485 should come in
handy.
◾ Traceability
Traceability between requirements (both system-level and software-
level), system tests, and risk-control measures implemented in software
(Clause 5.1.1c). Note that IEC 62304 Clause 3.32 defines traceability
quite narrowly as the degree to which a relationship can be established
between two or more products of the development process. The defini-
tion given by ISO 9000, and thus used by ISO 13485, is compatible but
also explicitly deeper: traceability refers to the ability to trace the his-
tory, application, or location of an object.
◾ Risk management
The conducting of risk-management activities and tasks, including
those related to SOUP components, in accordance with Clause 7 on risk
management (Clause 5.1.7).
◾ Configuration and change management
Configuration management and change management, also addressing
SOUP components and software used to support development (Clause
5.1.1d). Also see the discussion of Clauses 5.1.9, 5.1.10, and 5.1.11 below,
and Sections 6 and 8 of this book.
114 ◾ Medical-Grade Software Development

◾ Problem resolution
Problem resolution covering the medical-device software, deliver-
ables, and activities at each stage of the life cycle (Clause 5.1.1e). The
plan must also address when the problem-resolution process is to be
used (Clause 5.1.9f). Here too you might reference your SOPs as appro-
priate. Also see Section 9 of this book.
◾ Integration B C
A plan to integrate software items (incl. SOUP) and perform testing
during the integration (Clause 5.1.5, classes B and C only). Note that it is
acceptable to combine integration testing and system testing.
◾ Avoidance of common software defects B C
A procedure for the identification and avoidance of common software
defects (Clause 5.1.12, classes B and C only). This must include identifi-
cation of categories of defects introduced by the selected programming
technology and relevant to the system (Clause 5.1.12a), and evidence
of why these defects don’t contribute to an unacceptable risk (Clause
5.1.12b). The standard offers IEC TR 80002-1:2009 as further reading
on defects and hazardous situations. It may also make sense to invoke
the use of good programming style guides, Integrated Development
Environments (IDE), static code analysis tools, and perhaps even auto-
mated testing, as part of your answer to avoiding common defects. If
you work on the same platform repeatedly, or develop similar software
time and again, you may be able to reuse a previously made assessment
here via a reference. You should still assess whether the present project
poses any differences to those previous projects.

On the topic of traceability, note also that Clauses 5.2.6 and 7.3.3 discuss
traceability. The former requires software requirements to be traceable to
system requirements or another source (e.g., product specification, customer
requirements), and the latter adds requirements for document traceability
where the focus is on risk-management documents (see Section 7). The
takeaway here is that traceability is an extremely important topic in medical
devices and a relevant concern in just about all aspects of medical-software
development. Making use of modern development tools will help in main-
taining that traceability, but it is not enough to just think these will preserve
links between everything as is needed. This is somewhat analogous to ISO
13485 document control, where using a good system (version management
tool in that case) is expected, but you will still need to see to maintaining a
practical version history over changes. Here we are concerned with tracking
Software Development ◾ 115

not just changes but inputs and outputs of various types. Good traceability
is key to moving between such entities, going up and down, left and right,
as needed to get the evidence you are looking for in making any informed
decision, solving a nonconformity, or answering an audit question you hap-
pen to get. In the real world, traceability is often implemented well back-
ward, but issues may occasionally pop up when trying to move forwards,
or in the other direction less typically traveled. Think of this when planning
your approach to traceability.
The standard adds a few non-binding notes on Clause 5.1.1 to clarify that
your life-cycle model can specify different actions and deliverables for soft-
ware items based on their safety class, that the actions may overlap, interact,
and be performed iteratively or recursively, and that the actions may also be
referenced from outside the plan (e.g., the SOPs of your QMS). All of these
considerations will be useful to you in fashioning a lean, risk-based process
for software development.
Clause 5.1.2 then echoes the requirements of ISO 13485 to keep your
plan up to date as you progress. This does not, however, mean updating
time stamps for the sake of getting new versions of a document. When your
plans change you should update the planning documents, but to update
the documents unnecessarily is just madness – and madness that causes
more mayhem in your organization as people downstream try to figure out
whether there have been changes and what effects these may have on their
work.
Clause 5.1.9 goes into more detail on how the plan needs to address
configuration management. Information here must include, or reference, the
following.

◾ The classes, types, categories, or lists of software items to be controlled,


and when these items are to be placed under configuration control
(Clauses 5.1.9a,e).
◾ The activities and tasks needed in configuration management (Clauses
5.1.9b).
◾ The organizations responsible for performing activities, and their rela-
tionship with other organizations such as software development or
maintenance (Clauses 5.1.9c–d).
◾ When the problem resolution process is to be used (Clause 5.1.9f)

Clause 5.1.11 requires that for class B and C software, all configura-
tion items must be placed under configuration control before they are
116 ◾ Medical-Grade Software Development

verified. Note that the requirement for this control to be documented was
dropped in the 2015 amendment and that no requirement is thus made
here to include a statement in the software development plan or link to a
procedure.
Clause 5.1.10 requires that also those supporting items (e.g., tools,
items, or settings) used in the development of class B and C medical-grade
software must be controlled if these may impact the software. A further
non-binding note offers compiler versions, make files, and batch files as
examples of such items.
Note that most of the above requirements apply to all safety classes,
with only the few stated exceptions being made. The topics required to be
addressed may, in many cases, be addressed via discussion in the plan or by
referencing another plan or your SOPs as needed. To simplify your work in
defending the plan, and that of others in following it, it is generally a good
idea to briefly set the context or otherwise explain why a reference is made
and with what outcome instead of just providing a link. Beware of copy-
paste and micromanagement, though.

5.1.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. The quality-management standard is primarily concerned with
having processes adequately described in the standard operating proce-
dures (SOPs) of your QMS, and ensuring your D&D outputs match with
what you wanted to get when you specified the D&D inputs. The software
development standard relies more on a plan document to add a wealth of
information it sees as necessary to ensure the success of the development
project, although it too implies the existence of specific previously defined
processes. Thus, defining your processes well enough on an appropriately
high level in the SOPs will provide you with a good reference to cite in your
plans, and save you from unnecessarily repeating or reinventing overlapping
instruction in the plans. Similarly, the software development plan is your
go-to plan for development, but it may then be augmented by other lower-
level planning documents as is practical.
Table 5.3 provides a glance over the main requirements of both stan-
dards. Note the software safety classification used by IEC 62304 and remem-
ber that some clauses of ISO 13485 may be excluded with a valid rationale
(see Juuso 2022). Also note that the table is an overview of any qualifying
remarks presented in the previous discussion.
Software Development ◾ 117

Table 5.3 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Process for development planning which ✓ ✓ ✓ ✓
Matches with the other QMS processes ✓ - - -
Covers the entire development life cycle ✓ ✓ ✓ ✓
Software development plan (SDP), incl./ref. ✓ ✓ ✓ ✓
Product requirements ✓ - - -
Special resource needs ✓ - - -
Product-specific activities (incl. traceability) ✓ ✓ ✓ ✓
Product acceptance criteria ✓ - - -
Reference to system design and development - ✓ ✓ ✓
Standards, methods, and tools planning - - - ✓
Integration and integration testing planning - - ✓ ✓
Plan to integrate and test items (incl. SOUP) - - ✓ ✓
Verification planning ✓ ✓ ✓ ✓
Validation planning ✓ - - -
Risk management planning ✓ ✓ ✓ ✓
Documentation planning - ✓ ✓ ✓
Configuration management planning - ✓ ✓ ✓
Control of supporting items - - ✓ ✓
Configuration item control before verification - - ✓ ✓
Avoidance of common software defects - - ✓ ✓
Keep SDP updated ✓ ✓ ✓ ✓

Based on the combined requirements, a synthesis between the standards


may be negotiated based on the following items.

5.1.3.1 Standard Operating Procedure (SOP)


A documented process for development planning is required by ISO 13485
and vastly beneficial for consistently observing the requirements laid out
118 ◾ Medical-Grade Software Development

by IEC 62304. This process can be neatly defined in the SOPs of your QMS
– this is SOP-10 if you follow the QMS model set out in our previous book
(Juuso 2022). In writing the SOP for product realization your objective
here is to expand the earlier SOP to match the requirements of IEC 62304.
Remember to refer to your other related processes (e.g., supplier man-
agement, infrastructure management, software validation) as is relevant,
thereby avoiding the urge to repeat or reinvent overlapping instruction
here.
This SOP will house your processes for software development in accor-
dance with both standards. Here you should ensure the SOP instructs
software-planning activities in accordance with the stages expected by
both standards, including the key deliverables and reviews of each stage. In
particular, the SOP should instruct the use and high-level structure of the
software-development plan and allow the plan to add any further project-
specific information. Don’t plan to unnecessarily repeat information in the
plan, but instead refer to the SOP as setting up the grand lines and add addi-
tional detail as needed – and as allowed by your description in the SOP.
The contents of the product realization SOP should correspond to
Sections 5 through 9 of this book, specifically to what is discussed under
each subsection on the suggested synthesis. To write the SOP, take your ear-
lier version of SOP-10 (Juuso 2022), edit it to describe your software-devel-
opment life-cycle model, add the deliverables suggested in the synthesis
offered by this book, and add any further instruction you feel is warranted
for your organization. Our suggestion is to structure the instructions in the
SOP according to Sections 5 through 9 of this book and use each of these
sections to aid in the writing of the corresponding section of the SOP. You
may also split the instructions into several SOPs should this feel more practi-
cal to you, perhaps again along the lines of Sections 5 through 9. Remember
to also address the software safety classification and legacy software from
Section 4 as these are relevant to your operations.

5.1.3.2 System Development Plan


This is the highest-level development plan for an individual development
project that is to take place under your QMS. The objective here is to
describe the development of the whole system at a top level. The standard
implies that you are to develop system requirements in or based on the
plan, but this is poorly instructed in the standard. The system-level require-
ments are then referred to while developing the software under that system.
Software Development ◾ 119

The standard also states, via a non-binding note to Clause 5.1.1, that this
plan may be combined with the next-level development plan, the software-
development plan (see next), particularly if the software is standalone (note
to Clause 5.1.3b). This should be particularly appealing when the medi-
cal device to be developed is a software-only device, and thus the system
requirements and software requirements may be the same. The standard
suggests this combination of the plans to take place by including the soft-
ware development plan in the system development plan, which makes great
sense, but it then proceeds to place all the requirements on the software
development plan in its later clauses. You may thus wish to call the com-
bined plan by either name, or their combination, but you must be clear on
the setup yourself if you combine these plans.
The whole existence of a system development plan is taken for granted
by the IEC 62304 standard and not instructed well in any of its clauses.
The note to Clause 5.1.3b and the overall taxonomy set up in Clause 3 (i.e.,
system, software system, software item, software unit) make it seem self-evi-
dent that you should indeed have such a document, and this should some-
how mirror the content of the software development plan on a system-level.
This is, however, not explicitly required by either standard. You will do well
to consider the system-level beyond your software system (e.g., any hard-
ware, environment, and users for your software), but whether you decide to
do so in a separate plan or in the same plan you make for your software is
a choice you can make. The system-development plan is not discussed to
any great detail by either standard.

5.1.3.3 Software Development Plan (SDP)


Perhaps one of the most essential documents in your stack of documenta-
tion is the software development plan because it sets the foundations for all
that is to come. The SOP already described your overall approach and pro-
cess for software development so the objective here with the development
plan is to add appropriate project-specific information, if that wasn’t already
added by a separate system development plan.
The software development plan is a key document for running the soft-
ware-development activity. The plan is explicitly required by IEC 62304, and
it acts as a logical extension of your product realization SOP as required
by ISO 13485. As always, the purpose of a plan is to state what work you
tackle and how – if the properties of this change in some meaningful way
you should update the plan. No particular review of the plan is required
120 ◾ Medical-Grade Software Development

by either standard, but make sure you follow your own documented pro-
cess here to approve the plan and that you also follow ISO 13485 require-
ments for control of documents and records. If your plan is not approved at
all, or if it is approved by a summer intern, that will raise eyebrows during
any audit, but don’t feel like every plan needs to be approved by the whole
management team or the CEO either. Your management may greenlight a
project and assign a project manager or a steering group, and then take a
backseat until a later phase. The approval chain here will depend on you,
but make sure you can demonstrate top management commitment on an
appropriate level and that the chain of command is both clear and reflected
in the checkups you instruct. The involvement of top management may, for
example, be demonstrated through participation at some key reviews or
through device marketing decisions and the setting of fundamental policies
for topics such as risk management and cybersecurity.
The plan must be appropriate to the type of software being developed,
and the safety classification of that software. To satisfy ISO 13485 require-
ments it should determine project responsibilities, resource needs, product
requirements, and product-specific activities, as outlined above, unless those
are covered by a separate planning document (e.g., project plan, system
development plan).
The plan should address the entire development life cycle of the soft-
ware, which implies meeting requirements sprouting up all throughout
Clause 5. In doing so the plan should make use of the D&D stages expected
by both standards (see Table 5.2), and impose an appropriate level of control
over these, e.g., via stage reviews and the control of deliverables.
To satisfy IEC 62304 requirements the plan must address the topics listed
above in Section 5.1.2. Some topics call for information to be recorded in
or referenced by the plan, while on other topics you can answer the ques-
tions by referring to the processes defined in your SOPs – where the instruc-
tions are written on a high-enough level not to micromanage and yet to
adequately spell out your overall key activities, tasks, and deliverables. ISO
14971 risk management is also implicated for use in product realization by
both standards.
The final topic of joint interest in the context of development planning is
traceability. ISO 13485 expects traceability to be maintained throughout all
your operations, especially between the D&D outputs and D&D inputs. It
also expects you to address any product-specific traceability considerations
in the planning documentation. IEC 62304 asks to see traceability preserved
between requirements, system tests, and risk-control measures. To meet both
Software Development ◾ 121

expectations your SOP and your development plan must be on the same
page on traceability and together ensure that the necessary records can be
navigated back and forth as needed.
Note that the topics of referencing system development (Clause 5.1.3),
referencing standards, methods, and tools (Clause 5.1.4), integration and
integration testing (Clause 5.1.5), configuration management (Clause 5.1.9),
and avoiding common software defects (Clause 5.1.12) raised by IEC 62304,
and discussed above, are not directly discussed by ISO 13485. Similarly,
the topic of supporting items to control (Clause 5.1.10) is only discussed by
ISO 13485 via infrastructure management and tool validation (see SOP-3 in
Juuso 2022).
Don’t unnecessarily repeat information here from the SOP, but instead
refer to the process described in the SOP as setting up the grand lines and
add any further project-specific information here. Where you draw the line
between what is defined in the SOP and what is left for the plan is up to
you. The SOP should provide the overall framework for your operations in
a reusable way, but still leave enough room for adjustment on the planning
level should the specifics of the software product require this. The plan
should contain or reference all necessary information, and thus act as an
access point into the development of your device.
It is not always possible to know everything in advance when writing the
SOP, or even when writing the plan. To avoid a chicken-and-an-egg prob-
lem during planning, it is acceptable that not all of the information above
is written in the plan with an equal sense of finality from the start of the
development project. You may write in work items to investigate questions,
and you may update the plan when developments call for adjustments to
be made, but all the while you must make sure your plan is still up to date,
makes sense, and is appropriate for the work going forward. In determin-
ing whether the plan is still up to date or not, we would look at whether
the instructed course still holds true and not update the plan to record mile
markers on that course. We have heard arguments to update the plan after
a particular mile marker has been cleared, to record that event, but unless
the passed mile marker changes the plan, we would not court the additional
bureaucracy here. Update the plan when the plan needs to change, not as
it is being followed. You may, of course, have an appendix or another ref-
erenced document indexing the status of the various documents your plan
calls for. This way you can update that separate file without revising the
whole plan, and you will have a ready-to-use dossier of your key files when
you are compiling the final technical file.
122 ◾ Medical-Grade Software Development

5.1.3.4 Other Plans Mentioned


Clause 5.1 plays fast and loose with plans of different sorts, all of which are
required to either be contained or referenced from the software develop-
ment plan. It speaks of, e.g.,

◾ Tools planning C (Clause 5.1.4).


◾ Planning of integration and integration testing B C (Clause 5.1.5).
◾ Verification planning (Clause 5.1.6). Note here that ISO 13485 Clause
7.3.6 expects these plans to address methods, acceptance criteria, and
(as appropriate) statistical techniques with rationale for sample size. See
Section 5.9.2
◾ Validation planning (ISO 13486 Clause 7.3.7). Note here that ISO 13485
Clause 7.3.7 expects these plans to address methods, acceptance criteria,
and (as appropriate) statistical techniques with rationale for sample size.
See Section 5.9.3.
◾ Risk-management planning (Clause 5.1.7). See Section 7 and also SOP-8
in our previous book (Juuso 2022).
◾ Documentation planning (Clause 5.1.8). See Section 11 and also SOP-1
in our previous book (Juuso 2022).
◾ Configuration-management planning (Clause 5.1.9). See Section 8.

You might create individual plan documents for all these activities, but if
your software development plan can already cover these on an adequate
level why would you? You might also find that your SOPs already cover
everything on a good level and there is not that much project-specific detail
the plans need to add. You could thus more conveniently refer to the SOPs
from your SDP and save your staff some head scratching in figuring out if
the plan behind the copy-pasted text is the usual or something else. This
consolidated, cascaded approach to planning will also hopefully save you
from unnecessary work in maintaining multiple plans to all stay up to date.
The one exception to the above is possibly your validation plan which
holds the door for a critical and potentially costly activity that you do not
want questioned after it has been completed. Here you may want to look at
Section 12.2.3, and plan fastidiously before commencing with the validation.
If there is a need for the present software development project to devi-
ate from the instruction in the SOP in any way, you may acknowledge those
adjustments in the software development plan. Provided, of course, that the
deviations are acceptable and allowed by the SOP. If your SOP says that
Software Development ◾ 123

your SDP (or some document addressed by it) must be chiseled in triplicate
on stone tablets then that is what you must do; there is nothing the SDP can
say to change that.
We would thus suggest adopting the SOP as the baseline for planning,
then providing any known adjustments to that instruction for this particu-
lar project in that master plan, the software development plan. You might,
for example, provide further information to consider during some develop-
ment stage for this particular product, which you could then observe once
reaching the corresponding step. If you then find that you need to develop
a full-blown plan for some of the stages discussed, you should be free to do
so within the window to set up your SOP and SDP. Crucially, you will not
have tried to develop all these plans in-depth at the start when any small
change in your planned trajectory could have thrown all the hard work on
such other plans into the bin, and you won’t have finalized an SDP with big
gaping holes in it in terms of some of the above topics completely forgotten.
Just like a software item is understood by the standard, just about every-
thing is always an onion you can peel to reveal new layers and provide
more detailed instructions to guide the work as needed. Don’t try to peel
every onion in the software development plan, but know what onions you’ll
eventually need, and what you should perhaps consider when finally getting
ready to peel those.
See Section 11 for an overview of what documentation and information is
worked on during each stage of software development.

5.1.3.5 Use of Systems Instead of Documents


The IEC 62304 standard talks of the key types of information (e.g., require-
ments, architectural design, detailed design) to handle in a software devel-
opment project. In many cases it is natural to think of all this information
as leading to paper-like documents of different types (e.g., a specification, a
design document) but you could also meet the requirements of the standard
by maintaining this information in some database form inside a trusted prac-
tical tool. This way the information will be readily available to developers
and a lot of the friction involved in maintaining such information over time
will also be reduced.
Before opting for systems over documents, consider that you must still
be able to prove what information was reviewed, approved, and thus
cleared for use by activities (see Section 5.9.1). This may get particularly
cumbersome if the underlying entries, such as individual requirements, exist
124 ◾ Medical-Grade Software Development

in a free-flowing pool that may be added to at any time, but you are still
required to ensure consistency across the whole set and assess the whole
in some way for moving ahead to the next development stage. These issues
may be solved using appropriately defined workflows in the systems, but
also remember that you may then need to reproduce a particular snapshot
of the information later, perhaps even years down the road. You may, for
example, be asked to show the requirements or the architectural design
used for your software 10 versions, or a decade ago, and prove that these
formed a conformant set of information at the time. Retaining this informa-
tion in the systems is not impossible, but over time as the systems them-
selves go through updates and bug fixes it may also be a more complex
task than controlling documents might be. The sweet spot is perhaps in the
use of a hybrid approach that lets some information evolve in systems but
retains overarching assessments or key reviews in a more document-like
manner.

5.2 Requirements Analysis
In the excellent Clint Eastwood film, The Changeling, the mother
just wants her missing boy back. The police of 1930s Los Angeles
think it’s fine to respond by returning a boy almost matching
the profile, albeit 3 inches shorter than expected. The mother is
understandably distraught by the trade-in, as would any cus-
tomer handed software that falls short of expectations. The require-
ments should not be treated as an alphabet soup, as illustrated in
Figure 5.3.
Developing requirements for the software to be built is a fundamental and
profoundly important task. It is often claimed that most later issues detected
with software can be traced back to faults in the requirements. This is to
be expected as misunderstanding what it is you are to build, or losing track
of this over time, intuitively takes effort to avoid. You might fail to grasp all
the important aspects of the software at the start, or you might lose sight of
these as requirements evolve over the development project. In both cases,
luck would be your only life jacket to delivering a great piece of soft-
ware once you spiral out of control. Paying enough attention to specifying
requirements, on the other hand, may save you from floundering.
In practice, coming up with a solid list of input requirements here is both
a science and an art. It may also be an iterative process where you approach
Software Development ◾ 125

Figure 5.3 Requirements may be a bit of an octopus sushi to wrangle, but they
should never be treated as an alphabet soup.

and refine requirements multiple times from different points of view as your
understanding grows throughout the project. The two standards only pro-
vide inputs for you to carefully consider, and eliciting ready-to-use require-
ments from your customer is usually no small feat either. Much of your
success in developing solid, complete requirements that do not require
unnecessary changes later will come down to experience in your particular
line of business. Whether you have a real or an imaginary customer – a talk-
ative human being or a concept of a potential customer – to interview will
also make a difference here.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 7.3.3 Design and development inputs


◾ ISO 13485 Clause 7.2.1 Determination of requirements related to the
product
◾ ISO 13485 Clause 7.2.2 Review of requirements related to the product
◾ IEC 62304 Clause. 5.2 Software requirements analysis

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both. The following discussion will identify which subclauses apply to which
software safety classes.
126 ◾ Medical-Grade Software Development

5.2.1 Expectations from ISO 13485


The expectations for software requirements from ISO 13485 are primar-
ily discussed in Clause 7.3.3 on design and development inputs, but also
touched on in the customer-related Clauses 7.2.1 (Determination of require-
ments related to product) and 7.2.2 (Review of requirements related to
product).

5.2.1.1 Inputs for Requirements


You must identify and maintain records of the inputs relating to the require-
ments of your product (Clause 7.3.3). The inputs must include the following:

◾ The intended purpose/use (incl. the functional, performance, usability,


and safety requirements based on it)
◾ Applicable regulatory requirements
◾ Used standards
◾ Output of risk-management activities
◾ Experience from previous similar devices
◾ Any other requirements for the product or processes that you deem
essential

Clause 7.2.1 further requires your organization (manufacturer) to determine


requirements specified by the customer (incl. delivery and post-delivery
activities), any user training needed to ensure specified performance and
safe use, and any requirements not stated by the user but known to be nec-
essary for the specified intended use.

5.2.1.2 Finalizing Requirements
You must review the inputs for adequacy and approve them after ensur-
ing the requirements are complete, unambiguous, not in conflict with each
other, and amenable to verification and validation. The standard also sug-
gests the use of the IEC 62366-1 standard on usability as a further aid to
consult here.
Clause 7.2.2 further requires that the product requirements are reviewed
prior to the organization’s commitment to supply the product to the cus-
tomer (e.g., submission of tenders, acceptance of contracts/orders, or changes
to these). The review of the product requirements is to ensure the following:
Software Development ◾ 127

◾ Product requirements are defined and documented


◾ Any differences from previously expressed requirements are resolved
◾ Applicable regulatory requirements are met
◾ Any necessary user training is planned to be available
◾ The organization (manufacturer) can meet the defined requirements

You should also ensure that you meet the requirements of ISO 13485 for the
conducting of D&D reviews as per Clause 7.3.5 (see Section 5.9.1).

5.2.2 Expectations from IEC 62304


IEC 62304 sees the final medical device as something that goes above and
beyond the standard. The highest level it directly addresses is the software
system of the medical device, i.e., the software component of the released
medical device. The standard does at the same time concede in several
clauses that occasionally the software system may in fact be the medical
device (e.g., standalone software).
With the above mindset, Clause 5.2 requires the manufacturer to define
and document software-system requirements – for each software system of
the medical device – from the system-level requirements. The wider con-
struct of the system, as defined in Clause 3.30 (see Section 3), is the high-
level composite of the processes, hardware, software, facilities, and people
relevant to meeting some objective or stated need. In other words, the
system encompasses both the medical device and the context of its use in
a very broad sense. The objective here is to define software-system require-
ments, but to do so you must also consider the wider system around it.

5.2.2.1 Inputs for Requirements


Clause 5.2.2 details the types of requirements the standard expects to see,
provided these are relevant to the medical-device software. The standard
acknowledges, in a non-binding note, that not all the requirements may be
available at the start, and that the listed types of requirements (a–l) may
overlap. Below we have grouped the requirements into seven practical col-
lections for discussion.

◾ Functionality and the environment


Description of what the software needs to do and under what cir-
cumstances. The clauses primarily addressed here are functional and
128 ◾ Medical-Grade Software Development

capability requirements (Clause 5.2.2a), and requirements related to


IT-network aspects (Clause 5.2.2j). Requirements here may address,
e.g., performance (purpose of the software, timing, etc.), physical char-
acteristics and the intended computing environment (programming
language, platform, operating system, hardware, network environment,
etc.), compatibility needs (later upgrades, SOUP-components, other
devices, etc.), and network issues (protocols, alarms, and handling of
availability issues, etc.).
◾ Data and interfaces to other systems
Description of what inputs, data, and interfaces the software needs
to work with, what outputs it needs to produce, and how all this is
stored. The clauses primarily addressed here are software system
inputs and outputs (Clause 5.2.2b), interfaces with other systems (Clause
5.2.2c), and data definition and database requirements (Clause 5.2.2g).
Requirements here may address, e.g., data formats and characteristics
(value ranges, limits, defaults, etc.).
◾ User interface and alarms
How the user should interact or use the software, and when should
the software seek to interact with the user via alarms of different types.
The clauses primarily addressed here are user interface requirements
implemented by software (Clause 5.2.2f), and software-driven alarms,
warnings, and operator messages (Clause 5.2.2d). Requirements here
may address, e.g., the form of the user interface (full graphical, com-
mand line, API, etc.), and constraints on personnel. The standard
suggests IEC 62366-1 (see Section 12.5) and IEC 60601-1-6 as further
reading, both of which have been amended in 2020.
◾ Security
What security concerns should be addressed and how in the soft-
ware. The clause primarily addressed here is security requirements
(Clause 5.2.2e). Requirements here may address, e.g., authorization,
authentication, audit trails, system security/malware protection (incl.
firewalls), communication integrity (incl. hash checks), and the protec-
tion of sensitive information. In the modern world, cybersecurity is defi-
nitely a part of the equation here, especially if your device interfaces
with the outside world in some way.
◾ Installation and delivery acceptance
How the software should be installed and how this should be verified.
The clause primarily addressed here is installation and acceptance require-
ments of the delivered medical-device software at the operation and
Software Development ◾ 129

maintenance sites (Clause 5.2.2h). Requirements here may address, e.g.,


region settings for the keyboard, some database settings, time and date
format, or relevant security requirements to ensure the safety, security, and
essential performance of the functions performed by the software.
◾ Operation and maintenance
How the software should be used and maintained, also by the user
themself. The clause primarily addressed here is requirements related to
methods of operation and maintenance (Clause 5.2.2i), and user main-
tenance requirements (Clause 5.2.2k). Requirements here may address,
e.g., default presets for network protocols, procedures for adding new
devices to the system, default values for cybersecurity settings, recovery
in the event of disturbances or malfunction, and rollback requirements
for updates or upgrades (see Section 12.2 on IEC 82304-1).
◾ Regulatory requirements
What regulatory requirements are applicable here based on, e.g.,
the medical-device product being developed. The clause primarily
addressed here is regulatory requirements (Clause 5.2.2l). Requirements
here may address, e.g., the development of the software, the use of a
quality-management system, the conducting of risk-management activi-
ties, and what approvals are needed for what types of software features.

Clause 5.2.3 (classes B and C only) states that those risk-control measures
(mitigations) which are implemented in software are also included in the
requirements as appropriate to the medical-device software. The standard
acknowledges that these may not be known at the start of the develop-
ment project, and that these may change over time. This may in practice
be a cumbersome requirement to fulfill from the start as you may not yet
know all your risks and their mitigations, and the safety classification of the
affected software items may also still be veiled from you. Nonetheless, the
requirement holds firm for B and C class software items.

5.2.2.2 Finalizing Requirements
Once software requirements are nearing completion, Clause 5.2.6 instructs
how these are to be verified. No formal definition language is required here,
but the requirements are to be verified and documented so that these:

◾ Implement system requirements (incl. those related to risk control)


◾ Can be uniquely identified
130 ◾ Medical-Grade Software Development

◾ Are expressed in terms avoiding ambiguity


◾ Do not contradict each other
◾ Are expressed in terms permitting the establishment and performance
of tests (incl. acceptance criteria)
◾ Are traceable to system requirements or other sources

Clause 5.2.5 states that the manufacturer must also re-evaluate existing
requirements (incl. system requirements) and update these as appropri-
ate during the overall requirements analysis stage discussed here. Clause
5.2.4 further requires that the medical-device risk assessment is re-evalu-
ated when software requirements are established, and it too is updated as
appropriate.

5.2.3 Suggested Synthesis
The two standards are complementary. Table 5.4 provides a glance over the
main requirements of both standards. Note the software safety classification
used by IEC 62304 and remember that some clauses of ISO 13485 may be
excluded with a valid rationale (see Juuso 2022).
As your software-development process needs to conform to IEC 62304
Clause 5.2, ensure here that your SOP for product realization discusses the
development and review of requirements in accordance with the description
above. Your software development plan may then add any necessary addi-
tional information.
Based on the combined requirements, a synthesis between the standards
may be negotiated based on the following items.

5.2.3.1 Requirements Elicitation
IEC 62304 takes a fairly nuts-and-bolts view of requirements elicitation. It
puts the pressure of arriving at an unambiguous and cohesive set of require-
ments squarely on the manufacturer, as indeed does ISO 13485.
IEC 62304 is concerned with largely technical requirements including
data and interfaces, but it also acknowledges regulatory requirements face
on, and implies a need to consider the user and the use environment when
defining features, alarms, installation, and maintenance, for example ISO
13485, on the other hand, looks much more outside of the actual program
code ISO 13485 places greater emphasis on listening to and understanding
the customer.
Software Development ◾ 131

Table 5.4 Synthesis of Top-Level Requirements


IEC 62304

Combined requirements ISO 13485 A B C


Consider requirements from the ✓ - - -
intended use (purpose), including those
not mentioned by the customer
Consider use of standards ✓ - - -
Consider experience from previous ✓ - - -
similar devices
Consider requirements for user training ✓ - - -
Consider other customer requirements ✓ - - -
Consider requirements from system ✓ ✓ ✓ ✓
requirements
Software requirements content ✓ ✓ ✓ ✓
Include risk control measures in software ✓ - ✓ ✓
requirements
Re-evaluate medical device risk analysis - ✓ ✓ ✓
Update requirements ✓ ✓ ✓ ✓
Review/verify software requirements ✓ ✓ ✓ ✓

5.2.3.2 Requirements Review (i.e., Design Input Review)


The expectations of the two standards on requirements development and
review are highly compatible, but each has their own views on what aspects
should be considered during the development and review of software
requirements. You should review the requirements for completeness, unam-
biguity, traceability, and later verifiability via testing, but also in light of regu-
lations, changed expectations, and your ability to meet the requirements.
In essence, this is the first of your three big reviews: your Design Input
Review. The other two are covered in Section 5.8. Here you are making sure
the requirements, i.e., the inputs to your development activities warrant mov-
ing ahead to the actual design phase.
Once software requirements have been established, IEC 62304 expects
you to also re-evaluate and update higher-level system requirements and the
132 ◾ Medical-Grade Software Development

medical-device risk assessment. ISO 13485 too expects the risk-management


documentation to be up to date.

5.3 Architectural Design B C
A classic in the disaster-movies genre, The Towering Inferno, opens
with grandiose images of the world’s tallest skyscraper designed by
architect Paul Newman. In the film, Newman’s character provides
the grand strokes of the glass vision, but the engineers and econo-
mists have then been left to fill in the rest of the details to less-than-
great results. The architectural plan sketching out that overall thing
to build is also the goal fueling Clause 5.3 of IEC 62304 or sketch-
ing out the software symbolized by the floppy disk in Figure 5.4.
Architectural design is where you translate those system and software
requirements you have now established into a good bird’s-eye view of the
software you are developing. Here the ISO 13485 standard is out to get cof-
fee and lets you and IEC 62304 iron out the details of how all this should
be done. ISO 13485 still expects to later understand the key reviews you
have performed here so make sure to document your reviews in a compat-
ible way (see Section 5.9). Note that IEC 62304 does not require architectural
design for class A software.
Clause 3.3 gives a formal definition for the term architecture as the orga-
nizational structure of a system or a component. This definition is quite
broad and by itself would not give much reason to introduce a new term
here. Architecture of software, or software architecture, is a common term,
however, and one which instantly brings up images of software components
organized neatly into grids like chocolates in a box – or like that dated
3D-animation at the end of Jurassic Park when the girl realizes that it’s a
Unix system.
The relevant clauses of both standards are:

◾ ISO 13485: None directly


◾ IEC 62304 Clause 5.3 Software architectural design

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.
Software Development ◾ 133

Figure 5.4 Floppies were once the modern equivalent to paper scrolls, but now they
too may be relegated to ancient ruins on our architectural map.

5.3.1 Expectations from ISO 13485


Architectural design is a topic that does not appear in the ISO 13485 stan-
dard, but instead silently sits between its Clauses 7.3.3 (inputs) and 7.3.4 (out-
puts). The bookends the standard provides are between the D&D inputs and
outputs so here you are in essence free to follow the IEC 62304 standard
and translate that to your use.

5.3.3.1 Developing the Architecture Design


No requirements.

5.3.3.2 Verifying the Architectural Design


Despite ISO 13485 not placing any specific requirements on your architec-
tural design or its review, you should still ensure that any review you make
134 ◾ Medical-Grade Software Development

for the design meets the requirements of ISO 13485 for the conducting of
D&D reviews in its clause 7.3.5 (see Sections 5.9.1 and 5.9.2). This much is
expected by ISO 13485 if you later want to pull up or lean on your archi-
tectural design in any context recognized by the standard (e.g., conformity
assessment of the final product).

5.3.2 Expectations from IEC 62304


The IEC 62304 expects you to design your software architecture according
to Clause 5.3 if, and only if, your software safety classification is B or C. The
standard does not expect architectural design to be completed for class A
software. The idea here is that you can prioritize the software, and those
software subsystems in your software that pose the highest risk and not
jump through perhaps unnecessary hoops in the development of low-risk
software items.
In practice, some organizations have adopted a one-size-fits-all model for
software development, and thus chosen to design their software architecture
for all software items using the same process they use for class C software.
The upside here is that you won’t have to train and maintain several differ-
ent paths through development based on the safety classification, and you
won’t perhaps have to retroactively produce documents if the classification
of some module changes. The downside is that you may be wasting some
resources by using a howitzer to shoot peas if you are not careful – you may
also be fooling yourself into thinking that a howitzer is a good instrument
for shooting peas.
The expectations of architecture design are laid out in six short sub-
clauses as discussed in the following. The top-level clause, Clause 5.1, has
no content of its own. Note that aside from software item segregation, which
is only required of class C software, all the other requirements apply to both
B and C classes.

5.3.2.1 Developing the Architecture Design


The specific requirements for an architectural design are as follows.

◾ The design transforms the previously established medical-device soft-


ware requirements into a documented software architecture that both
describes the structure of the software and identifies the software items
used in it (Clause 5.3.1). The architecture is also required to observe
Software Development ◾ 135

any software system requirements (see Clause 5.3.6 below). Note that
although the term software item technically may refer to software sys-
tems, software items, and software units – all of which are software
items according to a non-binding note to Clause 3.25 – the requirement
here does not extend down to the level of units. The requirement to
identify software units is discussed along with detailed design in Clause
5.4.
◾ The design describes the interfaces between the software items, and
also between them and any external hardware or software components
(Clause 5.3.2).

In case a software item is identified as software of unknown provenance


(SOUP; see Section 1.14), the manufacturer must also specify the following.

◾ The functional and performance requirements of the software item


in relation to its intended use (Clause 5.3.3). Examples of functional
requirements here might be the ability to parse XML messages over an
interface, the string lengths allowed, the numerical space expected, and
the supported character encoding when communicating over an inter-
face. Similarly, examples of performance requirements might be per-
forming some calculation in under 100ms or sorting data values at the
same time.
◾ The system hardware and software requirements for the proper opera-
tion of the software item (Clause 5.3.4). Typical examples of require-
ments that could be made here are, e.g., processor speed, memory use,
and input/output needs.

Some organizations have plunged in deep to define component-specific


requirements to the above while others have been content with more
generic performance-based requirements that are the same for all SOUP
components. The choice is yours, but it will also depend on the criticality of
the component and the risk associated with its function.
Finally, in case a software item is class C, the manufacturer is also
required to identify effective segregation between software items as neces-
sary for risk control (Clause 5.3.5). This also includes a statement on how
the effectiveness of the specified segregation is ensured. Examples of how
to ensure segregation may, for example, include running software items on
separate, unshared resources. Note that this is a new subclause added in the
2015 amendment.
136 ◾ Medical-Grade Software Development

5.3.2.2 Verifying the Architectural Design


Finally, Clause 5.3.6 instructs the verification of the architectural design. Here
the manufacturer must document verification that the architecture:

◾ Implements system and software requirements (incl. risk controls).


Here the standard notes that a traceability analysis between the archi-
tecture and the underlying software requirements may be a practi-
cal tool but note that the unbinding advice does not cover system
requirements
◾ Supports interfaces between software items, and between software
items and hardware. Note that this subclause does not mention external
software, but that was a part of the requirements in Clause 5.3.2 above
◾ Supports the proper operation of any SOUP items

The form of documenting this verification is up to the manufacturer. A


simple review document identifying the usual details of exactly what
was reviewed, by who, when, and to what effect is the basic expectation
here.
The third requirement on the above list, item C, is interesting in that it
may not be immediately clear what this entails. We are required to ensure
the architecture we designed supports the use of the SOUPs as designed. If
we think about our software as a battery-operated device and SOUPs as the
batteries we put inside that device this intuitively translates to placing the
batteries in correctly (with the right orientation of +/- poles) into the battery
compartment. If the compartment door doesn’t close, or if the batteries are
in any order (plus pole to plus, minus pole to minus pole between batteries)
there’s a good chance the mental architecture inside our brain needs a little
retuning. But what does this mean for actual SOUP items? The answer is
that the architecture should make use of SOUP components as is appropriate
for its goals and in agreement with how those components were intended
to be used. Don’t put an electric eel inside your boombox and expect good
things to come.
As a sidenote, remember that your software development plan is
expected by IEC 62304 to either contain or reference your verification plan-
ning. The most convenient way to address this on a universal level may be
to reference your overall approach to the verification of process deliverables
in your SOP. The verification work here for the architecture is a part of that
overall verification activity for the software system.
Software Development ◾ 137

5.3.3 Suggested Synthesis
As no specific requirements for architectural design are made by ISO 13485
the synthesis here is simple: follow the requirements made by IEC 62304.
You should, however, ensure that you meet the requirements of ISO 13485
for the conducting of D&D reviews (see Section 5.9.1) and D&D verification
(see Section 5.9.2).
Table 5.5 provides a glance over the main requirements of both standards.
Note the software safety classification used by IEC 62304 and remember
that some clauses of ISO 13485 may be excluded with a valid rationale (see
Juuso 2022).
Did you notice how the requirements in this clause felt a little lighter than
in the previous clauses? Here both standards seem to realize that you will
have to do some real soul searching and heavy lifting to come up with the
architecture of your new software, and thus give you some more room to
breathe. The IEC 62304 standard, in particular, gives you some parameters
for your work here, and the main properties you’ll need to verify at the end,
but otherwise you are given great leeway here to use the tools, models, and
visualization concepts you wish. This is both a welcome and a very practi-
cal approach by the standard to, for once, conform to your elected operation
models. In other words, you are free to do whatever you want as long as the
outcome is correct.

Table 5.5 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Transform software requirements into an architecture - - ✓ ✓
Develop an architecture for the interfaces of software - - ✓ ✓
items
Specify functional and performance requirements of - - ✓ ✓
soup item
Specify system hardware and software required by soup - - ✓ ✓
item
Identify segregation necessary for risk control - - - ✓
Verify software architecture - - ✓ ✓
Meet ISO 13485 requirements for verification review ✓ - - -
138 ◾ Medical-Grade Software Development

As your overall software-development process here needs to conform


to IEC 62304 Clause 5.3, ensure that your SOP for product realization
addresses the use of an architectural design on a high level. Your software
development plan may then add any necessary additional information on
developing and reviewing your design. If your design meets the require-
ments of IEC 62304 listed above, and your review of the design meets
both the specific requirements of IEC 62304 and the general D&D review
& verification requirements of ISO 13485, you are well-set for moving
ahead.

5.4 Detailed Design B C
If Paul Newman got the role of the architect in the previous sec-
tion, and the actual detailed design was left to monkeys behind
the stage, the role of the person faced with the devil in those details
was given to Steve McQueen as the fire chief battling that tower-
ing inferno. Similarly, here the aim is to dive inside that big pic-
ture and figure out what goes into it before it is all built. Figure 5.5
illustrates this.

Figure 5.5 Detailed design has us zooming into the architectural map we previously
created.
Software Development ◾ 139

This is the next depth for our diving bubble, or the next layer of the onion,
after establishing an architectural design for your software. Here you will
take your overall architecture and dive deeper into it to see all the way
down to the level of software units (class B) and below (class C) – or skip
this stage completely if you follow the class A route. In other words, note
here that the IEC 62304 does not require a detailed design for class A soft-
ware, and only requires a greatly simplified version of a detailed design for
class B software.
The relevant clauses of both standards are:

◾ ISO 13485: None directly


◾ IEC 62304 Clause 5.4 Software detailed design

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

5.4.1 Expectations from ISO 13485


Detailed design is a topic that does not appear in the ISO 13485 standard,
but instead silently sits between its Clauses 7.3.3 (inputs) and 7.3.4 (outputs).
The bookends the standard provides are between the D&D inputs and the
outputs so here you are in essence free to follow the IEC 62304 standard
and translate that to your use.

5.4.1.1 Developing the Detailed Design


No requirements.

5.4.1.2 Verifying the Detailed Design


Despite ISO 13485 not placing any specific requirements on your detailed
design or its review, you should still ensure that any review you make
for the design meets the requirements of ISO 13485 for the conducting of
D&D reviews in its clause 7.3.5 (see Sections 5.9.1 and 5.9.2). This much is
expected by ISO 13485 if you later want to pull up or lean on your detailed
design in any context recognized by the standard (e.g., conformity assess-
ment of the final product).
140 ◾ Medical-Grade Software Development

5.4.2 Expectations from IEC 62304


5.4.2.1 Developing the Detailed Design B C
IEC 62304 does not require a detailed design for class A software. For class
B software all that is required is that the manufacturer refines the software
architecture developed above by subdividing it into lower-level software
items and software units until these are all represented as software units
(Clause 5.4.1). Note that software units are by definition not divided fur-
ther, and the standard also adds a non-binding note to say encountered
software systems are not divided further here either. You should, how-
ever, address the encountered software systems via their own development
documentation.
For class C software the following additional requirements are made.

◾ Each software unit will also receive a detailed design (Clause 5.4.2).
This design is to be sufficiently detailed to allow the correct implemen-
tation of the unit.
◾ The detailed design for software units will document a design for any
interfaces between software units and external hardware or software
components. This design is to be sufficiently detailed to allow correct
implementation of the unit and its interfaces.

5.4.2.2 Verifying the Detailed Design C


For class C software only, the detailed design is to be verified and docu-
mented to:

◾ Implement the software architecture (Clause 5.4.4a)


◾ Be free from contradiction with the software architecture (Clause 5.4.4b)

As with the architectural design above, here too the standard offers a help-
ful note to say that a traceability analysis between the detailed design and
the underlying architecture may be a practical tool to meet the former
requirement.
Note also that your software development plan is expected by IEC 62304
to either contain or reference your verification planning. The most con-
venient way to address this on a universal level may be to reference your
overall approach to verification of process deliverables in your SOP. The
verification work here is a part of that overall activity.
Software Development ◾ 141

Table 5.6 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Refine software architecture into software units - - ✓ ✓
Develop detailed design for each software unit - - - ✓
Develop detailed design for interfaces - - - ✓
Verify detailed design - - - ✓
Meet ISO 13485 requirements for verification review ✓ - - -

5.4.3 Suggested Synthesis
As no specific requirements for detailed design are made by ISO 13485 the
synthesis here is simple: follow the requirements made by IEC 62304. You
should, however, ensure that you meet the requirements of ISO 13485 for
the conducting of D&D reviews (see Section 5.9.1) and D&D verification (see
Section 5.9.2).
Table 5.6 provides a glance over the main requirements of both stan-
dards. Note the software safety classification used by IEC 62304 and remem-
ber that some clauses of ISO 13485 may be excluded with a valid rationale
(see Juuso 2022).
In other words, ensure that your SOP for product realization addresses
the use of a detailed design on a high level and allows for more project-
specific details to be added by the software development plan. In particular,
pay attention to the software safety classification and how it affects it if and
how the requirements apply to the detailed design of your software. Finally,
review the detailed design if required.
After clearing the detailed design phase it’s time to pat ourselves on the
back and start implementing our designs. After one section of planning our
approach to development and then three sections looking at the design of
our software, we are now ready to roll up our sleeves and get to work in
actually writing the software. This is where the script is written and it’s time
to call your cast out of their trailers to start making some movie magic.

5.5 Unit Implementation and Verification


The unit in The Dirty Dozen may be put together from army rejects,
but after Lee Marvin is finished with them, they are shown to be
142 ◾ Medical-Grade Software Development

Figure 5.6 After all that designing, this is where we move to first doing and then
checking the results.
capable of outshining the best-trained commandos in the theater
of operations. Verification of this feat, too, is visible in the form of
a blown-up chateau in northern France. The aim in Clause 5.5 is
similar, only here the unit of study is the soldier being built to fit into
the army. Figure 5.6 illustrates this.
Software units are the atoms of your piece of software. From the point of
view of the standard, these are software items that do not merit any fur-
ther subdivision into software items. If you build these and integrate them
together the way you wanted, you will end up with the entire software sys-
tem you set out to develop. For class A all that is required is that you imple-
ment any units you intended to, but for class B and C, you must also verify
that you got what you intended to get.
The relevant clauses of both standards are:

◾ ISO 13485: None directly


◾ IEC 62304 Clause 5.5 Software unit implementation and verification

In the following we will look at the expectations of both standards, one stan-
dard at a time, and then synthesize a plan to meet the requirements of both.

5.5.1 Expectations from ISO 13485


5.5.1.1 Unit Implementation
Unit implementation is a topic that does not appear in the ISO 13485 stan-
dard, but instead silently sits between its Clauses 7.3.3 (inputs) and 7.3.4
Software Development ◾ 143

(outputs). The bookends the standard provides are between the D&D inputs
and the outputs so here you are in essence free to follow the IEC 62304
standard and translate that to your use.

5.5.1.2 Unit Verification
Unit verification is a topic that does not appear in the ISO 13485 standard
either. The topic of overall verification for your product, however, is some-
thing discussed in Clause 7.3.6 on design and development verification (see
Section 5.9.2). Also remember what Clause 7.3.5 wants to see in terms of any
D&D reviews made (see Section 5.9.1). You may choose to meet these require-
ments here head-on for software units but remember to then also address
integration verification in the next stage and finally to ensure that you have
covered your entire product through the body of your verification activities.
Alternatively, you might decide to treat software-unit verification as some-
thing separate from the D&D verification required by ISO 13485 and per-
form the latter separately. If you do so, realize that you may be forced to
redo some verification activities if you then run into issues leaning on the
earlier IEC 62304 verification from an ISO 13485 context. It therefore makes
all the sense in the world to make your unit verification compatible with
ISO 13485; the steps to do so are not steep. The few requirements set by
ISO 13485 here are discussed in Section 5.9.2. An example of a potentially
costly pitfall here is that ISO 13485 requires your verification to observe any
interfaces or connections to other devices described in your intended use. In
the case of software units, this will perhaps not matter much as you will be
ensuring such connections are observed during your integration verification
and software-system testing activities. It is nonetheless prudent to be aware
of this assertion here.
In addition to meeting these verification requirements, you should also
ensure that any unit-verification review you make meets the requirements
of ISO 13485 for the conducting of D&D reviews in Clause 7.3.5 (see Section
5.9.1). This much is expected by ISO 13485.

5.5.2 Expectations from IEC 62304


5.5.2.1 Unit Implementation
Software-unit implementation is half the title of the clause, but it is
only addressed by one short subclause: “the manufacturer is required
144 ◾ Medical-Grade Software Development

to implement each software unit” (Clause 5.5.1). In other words, all the
software units you intended to have must be implemented. In fact, the
requirement applies to all classes of software and thus is not limited by
the software units specified in the detailed design (class C software) or
the corresponding refined software architecture design (class B software).
For class A software, however, it may be difficult to check that you have
indeed implemented all the units you planned to have as neither of these
records is required to be available, but that is nonetheless the stated
requirement here. In practice, you probably wouldn’t start implementing
even the flimsiest of class A software systems without some sort of a plan,
software-item roster, or a stack of napkins. You should thus have some
record of the software items you planned to have, and here the objective
is to ensure you have not left any of those software items behind. Along
the way you may have decided that Private Ryan was no longer needed
for your campaign, and thus released him to go home, but if, on the other
hand, you had just misplaced him or forgotten about him you should now
take note and act accordingly. The purpose of this check is to ensure, in
part, that you have done all that you intended, which is ultimately going to
be a part of ensuring all of your activities are complete as ISO 13485 also
insists.
To play devil’s advocate here, note also that the term used here by the
standard is software unit and not software item. In other words, you are
required to implement all the units you planned, but you don’t have to
implement all the higher-level items. A unit is a type of an item, but an item
is not necessarily a unit, after all. You shouldn’t flirt with Dante’s inferno
though: always keep your documentation as up-to-date as you can.

5.5.2.2 Unit Verification B C


For classes B and C, the requirement here is to define a verification process
(Clause 5.5.2) and acceptance criteria (Clause 5.5.3), and then perform docu-
mented verification (Clause 5.5.4). Note that class A software is exempt from
all the verification requirements discussed in this section.
The process does not have to be described in an SOP – it could, for
example, be instructed in your software development plan – but the manu-
facturer is required to establish strategies, methods, and procedures for
verifying software units. If you perform verification by testing, you must also
evaluate the adequacy of the test procedures. This evaluation may be part of
the review and approval of your SOP (if appropriate), it may be recorded at
Software Development ◾ 145

the same time as approving your software development plan, or it may be a


separate review.
The acceptance criteria for software units must also be defined, and the
meeting of those criteria ensured, prior to integrating the units into any
larger items as appropriate (Clause 5.5.3). Note that the standard here stops
short of saying that the acceptance criteria should be set before the units are
ready and their acceptability can be assessed, although it is generally a good
approach to define what you want to get before you see what you got. The
acceptance criteria could be based on, e.g., whether risk controls are imple-
mented, if the unit matches the interface design set for it, and if the unit
meets the programming guidelines and conventions of your organization.
Clause 5.5.4 (class C only) further requires that if the following aspects are
relevant to the design, the manufacturer must include additional acceptance
criteria to cover them:

◾ Proper event sequence (Clause 5.5.4a)


◾ Data and control flow (Clause 5.5.4b)
◾ Planned resource allocation (Clause 5.5.4c)
◾ Fault handling (error definition, isolation, and recovery) (Clause 5.5.4d)
◾ Initialization of variables (Clause 5.5.4e)
◾ Self-diagnostics (Clause 5.5.4f)
◾ Memory management and memory overflows (Clause 5.5.4g)
◾ Boundary conditions. (Clause 5.5.4h)

Finally, Clause 5.5.5 requires that the manufacturer must perform software-
unit verification and document the results. Again, this requirement does not
apply to class A. Note also that your software development plan is expected
by IEC 62304 to either contain or reference your verification planning. The
most convenient way to address this on a universal level may be to reference
your overall approach to the verification of process deliverables in your SOP.
The verification work here is a part of that overall activity.

5.5.3 Suggested Synthesis
As no specific requirements for unit implementation and verification are
made by ISO 13485 the synthesis here is simple: follow the requirements
made by IEC 62304.
Table 5.7 provides a glance over the main requirements of both standards.
Note the software safety classification used by IEC 62304 and remember
146 ◾ Medical-Grade Software Development

Table 5.7 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Implement each software unit - - ✓ ✓
Establish software unit verification process - - ✓ ✓
Software unit acceptance criteria - - ✓ ✓
Additional software unit acceptance criteria - - - ✓
Software unit verification - - ✓ ✓
Meet ISO 13485 requirements for verification review ✓ - - -

that some clauses of ISO 13485 may be excluded with a valid rationale (see
Juuso 2022).
The process for your unit verification may conveniently be defined in
your product realization SOP, and then any project-specific details added in
the software development plan. You should address your overall approach
to D&D verification in your process, i.e., explain how your unit verifica-
tion activities fit into the verification of the entire product. Also ensure that
your verification process and unit-acceptance criteria meet the requirements
of IEC 62304 discussed above, and that your process includes some checks
to ensure you have implemented all software units appropriate to your
software safety class. As before, you should also ensure that you meet the
requirements of ISO 13485 for the conducting of any D&D verification and
reviews (see Sections 5.9.1 and 5.9.2).

5.6 Integration and Integration Testing B C


As a film it’s perhaps a guilty pleasure, but the Three Amigos!
starring Steve Martin et Co. is a serviceable example of uniting
the disparate villagers for a common purpose. Ridding the village
of a scourge may also be appropriate for a discussion of medical-
software development. A more street-credible reference here may be
Yul Brunner et Co. in The Magnificent Seven, or Akira Kurosawa’s
Seven Samurai, but the end result is the same. Here instead of inte-
grating oppressed plebs and pitchforks, you are assembling the whole
of your software. Figure 5.7 illustrates this.
Software Development ◾ 147

Figure 5.7 This is that “E pluribus unum” moment: out of many becomes one.
You may think of integration as stacking cargo containers when loading
them on to a ship but remember that any interconnecting locks or other
interfaces of those containers also need to be hooked up. What happens
inside the containers was already implemented and verified above in Section
5.5, but now you need to build up from those blocks to the level of the
whole you set out to develop and ensure that the pieces click together as
you planned. The verification of this may be a simple inspection, as the
standard suggests in a note, and the final testing of the whole is yet to
come. Here the focus is on correct assembly as per your plans; later you’ll
ensure that the whole holds water. Note that it is, in fact, acceptable to also
combine integration testing and system testing (see Section 5.7).
The relevant clauses of both standards are:

◾ ISO 13485: None directly


◾ IEC 62304 Clause 5.6 Software integration and integration testing

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

5.6.1 Expectations from ISO 13485


Software integration, verification of that integration, and then performing
integration testing are topics that do not appear in the ISO 13485 standard,
148 ◾ Medical-Grade Software Development

but instead silently sit between its Clauses 7.3.3 (D&D inputs) and 7.3.4
(D&D outputs). The bookends the standard provides are between the D&D
inputs and outputs so here you are in essence free to follow the IEC 62304
standard and translate that to your use.
You may choose to meet your overall verification needs here head-on for
the integration but remember that you may already have addressed unit veri-
fication during the previous stage, and that you will still need to ultimately
ensure that you have covered your entire product through the body of your
verification activities. Alternatively, you might decide to treat software verifica-
tion as something separate from the D&D verification required by ISO 13485,
and perform the latter separately, but realize that you may be forced to redo
some verification activities if you then run into issues leaning on the earlier
IEC 62304 verification from an ISO 13485 context. It therefore makes all the
sense in the world to make sure your integration verification is compatible
with ISO 13485. The steps to do so are not steep. The few requirements set
by ISO 13485 here are discussed in Section 5.9.2. An example of a potentially
costly pitfall here is that ISO 13485 requires your verification to observe any
interfaces or connections to other devices described in your intended use.
In addition to meeting these verification requirements, you should also
ensure that any integration verification review you make meets the require-
ments of ISO 13485 for the conducting of D&D reviews in its clause 7.3.5
(see Section 5.9.1). This much is expected by ISO 13485.

5.6.2 Expectations from IEC 62304


5.6.2.1 Integration Plan B C
Clause 5.1.5 (classes B and C only) calls for an integration plan to be
included or referenced from the software development plan, but nowhere
does the standard actually make requirements on the contents of this plan.
The minimum answer might therefore consist of what is instructed by your
SOP and what special considerations your software development plan may
bring up for this particular project. In terms of describing your planned
integration here, you should refer to the architectural design and detailed
design, provided that these exist. We would argue that this set of documen-
tation already addresses your planned whole and your arrangements for
integration sufficiently.
If, however, you find that a separate in-depth integration plan is needed,
you might develop that here before proceeding to the actual integration
Software Development ◾ 149

work and integration testing. If you find yourself writing an integration plan
at the same time as the integration and verification work, you should recon-
sider your process.

5.6.2.2 Integrating Software Units B C


Clause 5.6.1 requires the manufacturer to integrate software units in accor-
dance with the integration plan made in, or referenced from, the software
development plan. In other words, the standard wants to see the reality of
what you develop match what you planned. Should this not be the case you
may need to go back a few squares and revise your plans.

5.6.2.3 Integration Verification B C


To ensure that all the blocks have been put into place as planned, the
manufacturer is instructed to perform verification of the software integration
(Clause 5.6.2). Clause 5.6.2 also makes it clear that integration here addresses
all the blocks you described in your integration planning, whether the soft-
ware units build software items or the software system itself.
Clause 5.6.2a doubles down on the above message and requires the
manufacturer to verify that the software units have been integrated into
software items and the software system. Clause 5.6.2b applies even more
superglue around this requirement by stating that also the integration of
hardware items, software items, and support for manual operations of the
overall system is verified. By support for manual operations the standard
refers to, e.g., human-computer interface features such as voice control and
online help menus.
In a non-binding note to the clause, the standard offers an opinion that
the verification activity here most likely takes the form of some of inspec-
tion, and that the purpose of the verification is limited to ensuring the
integration has been done as planned. In other words, this clause is not
concerned with what happens inside the cargo containers, or even if the
connected set of cargo containers works as intended, only that the contain-
ers have been stacked and connected as planned. The test of the whole is
chiefly addressed in the next clause (see Section 5.7).
Whatever form the verification takes, there must be records of it, as
always. Note also that your software development plan is expected by IEC
62304 to either contain or reference your verification planning. The most
convenient way to address this on a universal level may be to reference your
150 ◾ Medical-Grade Software Development

overall approach to the verification of process deliverables in your SOP. The


verification work here is a part of that overall activity.

5.6.2.4 Integration Testing B C


If the verification of the integration instructed above was just a crude
inspection that the integration plan was implemented – that the building
instructions for your Lego set were followed – here you will get to make
sure the thing you built works as intended. If you built a paperweight
according to the instructions of a Lego digger this is where you should
notice your mistake. If the vast cargo ship you stacked full of containers gets
stuck in the Suez Canal that probably is not your fault, but if some of the
crates fall off because you neglected to observe the loading plan you will
feel the wrath of those looking on.
Clause 5.6.3 requires the manufacturer to test the integrated software
items in accordance with the integration plan and record the results. The
objective is to address whether the integrated software items perform as
intended (Clause 5.6.4). The standard provides a helpful list of aspects to
consider here in a non-binding and non-exhaustive note to the clause:

◾ The required functionality of the software is achieved


◾ Risk controls are implemented
◾ Specified timing and other behavior is reached
◾ Internal and external interfaces function as specified
◾ Testing under abnormal conditions (incl. foreseeable misuse)

A second, deceptively diminutive note to the clause then points out that
integration testing may in fact be combined with overall software-system
testing (see Section 5.7) into a single set of planned activities. This profound
statement should give you welcome flexibility in matching the expectations
of the standard with your development pipeline. Wherever you see your
integration testing take place, the odds are that you will apply some form of
continuous integration (CI) methodology and use automated tests conducted
as part of merge/pull requests.
Clause 5.6.5 requires that the manufacturer evaluates the integration test
procedures for adequacy. Appropriate regression tests are to be used to
demonstrate that defects have not been introduced into previously integrated
software when integrating software items (Clause 5.6.6). Note that regression
testing itself is defined in Clause 3.15 as the testing required to determine
Software Development ◾ 151

that a change to a system component has not adversely affected functional-


ity, reliability, or performance, and has not caused new defects.
The records kept of the integration testing must include the following
(Clause 5.6.7):

◾ The test result (pass/fail and a list of any anomalies discovered)


◾ Sufficient details to enable the test to be repeated later as needed, e.g.,
test-case specifications (incl. required actions and expected results) and
identification of the test environment (incl. equipment and software
used)
◾ The identity of the tester

Note that the term anomaly is given a formal definition in Clause 3.2. Here
the definition is quite loose referring to any condition that deviates from
the expectations set up by requirements specifications, design documents,
standards, or a person’s perceptions and expectations. In addition to testing,
anomalies are said to arise out of review, analysis, compilation, and the use
of the product or its documentation.
Finally, Clause 5.6.8 requires that any anomalies discovered during inte-
gration and integration testing are fed into a software-problem resolution
process (see Section 9).

5.6.3 Suggested Synthesis
As no specific requirements for integration, integration verification, and inte-
gration testing are made by ISO 13485 the synthesis here is simple: follow
the requirements made by IEC 62304. You should, however, ensure that you
meet the requirements of ISO 13485 for the conducting of D&D reviews (see
Section 5.9.1), and that your verification activities here are compatible with
the requirements for conducting D&D verification (see Section 5.9.2).
Table 5.8 provides a glance over the main requirements of both stan-
dards. Note the software safety classification used by IEC 62304 and remem-
ber that some clauses of ISO 13485 may be excluded with a valid rationale
(see Juuso 2022).
In other words, ensure your SOP for product realization addresses unit
integration, integration verification, and integration testing (incl. evaluation
of the test procedure) on a high level, and allows for more project-specific
details to be added by the software development plan. Your process should
then ensure that you have integrated all the software units, verified this
152 ◾ Medical-Grade Software Development

Table 5.8 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Integrate software units - - ✓ ✓
Verify software integration - - ✓ ✓
Software integration testing & testing content - - ✓ ✓
Evaluate software integration test procedures - - ✓ ✓
Conduct regression tests - - ✓ ✓
Integration test record contents - - ✓ ✓
Use software problem resolution process - - ✓ ✓
Meet ISO 13485 requirements for verification review ✓ - - -

according to the integration plan (included in or referenced from the soft-


ware development plan), performed integration testing, and evaluated the
test procedures for adequacy, as required.

5.7 Software-System Testing
This is where the marriage between the two standards gets inter-
esting. This is where Mrs Smith takes a sniper rifle to Mr Smith in
the desert, where Mr Smith rushes home to take stock of the pair’s
armory, and the two then battle out their world views as Brad Pitt
and Angelina Jolie did, first on the silver screen in that film and
then apparently more amicably at home. Figure 5.8 illustrates this.
The two standards look at the same thing – software in this case – from
two different angles and want to ensure it all makes sense according to
their points of view. IEC 62304 logically looks at the forest and the trees that
make up the forest. ISO 13485 equally logically looks at whether your for-
est meets the requirements you had for a forest and the expectations your
customer now has for a forest.
This inspection is to take place via testing on many levels as IEC 62304
sees it, but also for the purposes of verification (what you wanted) and
validation (what your customer needed) as ISO 13485 expects. See Section
1.14 for more discussion on the difference in the point of view between the
Software Development ◾ 153

Figure 5.8 Testing should lead to that “all systems are go” feeling.

standards. Also note that, according to IEC 62304, software system testing
was previously only required of higher software safety classes, but since
the 2015 amendment the requirements here apply equally to all classes of
software.
The relevant clauses of both standards are:

◾ ISO 13485: None directly


◾ IEC 62304 Clause 5.7 Software system testing

In the following we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

5.7.1 Expectations from ISO 13485


The requirements ISO 13485 places on the testing of your software are
addressed via the outputs, verification, and validation requirements pre-
sented in its Clauses 7.3.4 on D&D outputs, 7.3.6 on D&D verification, and
7.3.7 on D&D validation. The lens taken by the standard is markedly differ-
ent from the bifocal lens of a decomposed lower level (software units and
items) and an integrated top level (software system and system) taken by
IEC 62304. If you follow the structure of software development instructed
by IEC 62304 you will also have seen to many verification activities already.
It would be a shame to have to repeat work just because ISO 13485 has a
154 ◾ Medical-Grade Software Development

somewhat different line of sight. You may not have addressed the validation
activities to an equal level of care, but you may already have a good foun-
dation for performing that work if you have a customer in the loop and a
thought-out intended use for your planned product.
To make sure no work is unnecessarily replicated here, our suggestion
is to perform software-system testing as instructed by IEC 62304 and then,
once the software system is ready, perform any necessary additional checks
on the level of the final medical device to tie up your verification activities
and perform the important validation activities. In other words, look at what
you have accomplished on the parts-level, what you’ll still need to test on
the device-level in order to feel confident, and move on from there.
Aside from verification and validation activities, which we will sidestep
for the moment, the ISO 13485 standard may be left to the sidelines in
terms of software-system testing. As before, you should nonetheless ensure
that any review you make here meets the requirements of ISO 13485 for
the conducting of D&D reviews (see Section 5.9.1) and that you observe
the ISO 13485 requirements for conducting D&D verification (see Section
5.9.2) if the verification here is to feature in the verification of your overall
product.
After finishing with testing here we will move to software release in the
next section. The discussion there will remind you of the ISO 13485 require-
ments related to review, verification, and validation before release.

5.7.2 Expectations from IEC 62304


As noted above, the requirements for software-system testing apply equally
to all software safety classes since the 2015 amendment of IEC 62304. The
requirements are given in Clause 5.7 of the standard.

5.7.2.1 Establishing and Performing Testing


The manufacturer is required to establish and perform testing against all
software requirements (Clause 5.7.1a) and to evaluate the adequacy of the
verification strategies and test procedures (Clause 5.7.1b). Clause 5.7.1 also
provides a minimum set of information to establish for tests:

◾ Input stimuli
◾ Expected outcomes
◾ Pass/fail criteria and procedures
Software Development ◾ 155

A non-binding note to the clause suggests that in addition to separate tests


for each requirement, tests covering combinations of requirements may be
performed. Another particularly helpful note adds that software-system test-
ing may be combined with integration testing, and that it is even acceptable
to test software requirements in earlier phases. Adopting the latter advice
may speed your overall process greatly but remember to somehow still take
ownership of the whole and ensure that the whole system has been tested
appropriately.
Clause 5.7.2 instructs that any anomalies discovered during software-sys-
tem testing are fed into a software problem resolution process (see Section 9).
Finally, Clause 5.7.3 requires that when changes are made during testing
the following actions are taken:

◾ The effectiveness of the change in correcting the problem is verified


by repeating tests, and performing additional or modified tests (Clause
5.7.3a)
◾ Conduct testing to demonstrate that unintended side effects have not
been introduced (Clause 5.7.3b)
◾ Relevant risk-management activities are performed as defined in Clause
7.4 (Clause 5.7.3c). See Section 7.4

The minimum contents of software-system testing records are defined by


Clause 5.7.5 so that the repeatability of the tests is ensured. The information
to be covered includes:

◾ The version of software tested, the date of testing, and the identity of
the tester (i.e., the person responsible for executing the test and record-
ing the results)
◾ Identification of the test environment (incl. relevant hardware and soft-
ware test configurations, and relevant test tools)
◾ A reference to test case specifications (incl. test case procedures,
required actions, and expected results)
◾ The test result (pass/fail and a list of any anomalies discovered)

Note that the terms used here differ somewhat from those instructed above
in the clause for integration testing, but do not do so in any material way
in terms of content. The terms used by this clause are given in the brackets
above, but the more general terms are otherwise used for consistency with
the above clauses.
156 ◾ Medical-Grade Software Development

5.7.2.2 Verification of System Testing


Clause 5.7.4 requires the manufacturer to evaluate the appropriateness of
system-testing verification strategies and test procedures. The objective of
the verification here is to verify that:

◾ All software requirements have been tested or otherwise verified


(Clause 5.7.4a)
◾ Traceability between software requirements and tests (or other verifica-
tion) is recorded (Clause 5.7.4b)
◾ The test results meet the required pass/fail criteria (Clause 5.7.4c)

Note also that your software development plan is expected by IEC 62304 to
either contain or reference your verification planning. The most convenient
way to address this on a universal level may be to reference your overall
approach to the verification of process deliverables in your SOP. The verifi-
cation work here is a part of that overall activity.

5.7.3 Suggested Synthesis
As we have sidestepped ISO 13485 verification and validation for now, the
synthesis here is simple: follow the requirements made by IEC 62304. You
should, however, ensure that you meet the requirements of ISO 13485 for
the conducting of D&D reviews (see Section 5.9.1) and that your verification
activities here are compatible with the requirements for conducting D&D
verification (see Section 5.9.2).
Table 5.9 provides a glance over the main requirements of both standards.
Note the software safety classification used by IEC 62304 and remember
that some clauses of ISO 13485 may be excluded with a valid rationale (see
Juuso 2022).
In other words, regardless of the software safety classification you are to
establish tests, perform them, retest after changes, and link to software prob-
lem resolution (see Section 9) as appropriate.

5.8 Release
This is where we get ready to flip that plexiglas cover open and pull
the trigger on our software to see the world, or to conversely wrap
Software Development ◾ 157

Table 5.9 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Establish tests for software requirements - ✓ ✓ ✓
Use software problem resolution process - ✓ ✓ ✓
Retest after changes - ✓ ✓ ✓
Verify software system testing - ✓ ✓ ✓
Software system test record contents - ✓ ✓ ✓
Meet ISO 13485 requirements for verification review ✓ - - -

the software in some see-through plastic and ship it off. Whether it is


your management representative, quality manager, or someone else
punching the button you won’t get to jettison the software without
seeing to its support and maintenance from here on out. The lime-
light is on your product now, as illustrated in Figure 5.9.
IEC 62304 Clause 3.37 defines the term release as a particular version of a
configuration item made available for a specific purpose. Clause 3.12 intro-
duces the related term of “medical device software” which is noted to be
a software system developed for the purpose of being incorporated into a
medical device or intended for use as a medical device itself. Except for the
possibility that a release is full or incremental (see Section 6.3.2) the two
terms are highly compatible for a software product and may also include
documentation and data relevant to the software. Note that IEC 62304 Clause
3.11 also attempts to define the term “medical device”, but don’t take this
as a definition to override any regulatory requirements or definitions (see
Section 1.3).
The relevant clauses of both standards are:

◾ ISO 13485 Clause 7.3.4 Design and development outputs


◾ ISO 13485 Clause 7.3.6 Design and development verification
◾ ISO 13485 Clause 7.3.7 Design and development validation
◾ ISO 13485 Clause 7.3.8 Design and development transfer
◾ IEC 62304 Clause 5.8 Software release

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.
158 ◾ Medical-Grade Software Development

Figure 5.9 When the software is ready for the limelight, it’s time.

5.8.1 Expectations from ISO 13485


The requirements for product release itself are addressed in Clause 7.3.8 on
D&D transfer, but before you get to go there you must also take care of a
few additional considerations expected by ISO 13485 and not discussed by
IEC 62304. These stages cover the review of your D&D output, the wrap-up
of your verification activities, and the validation of your medical device. Each
of these steps is addressed in the following.

5.8.1.1 D&D Output Review


Earlier we handed the lead of our software-development activities to the
IEC 62304 standard after discussing the D&D input stage (ISO 13485 Clause
7.3.3) that we saw as corresponding to the software requirements analysis
stage in that standard. Therefore, it should be quite natural that once the
development work is essentially done, we return to check what require-
ments the ISO 13485 standard has for the D&D output stage. The appropri-
ate clause to return to is Clause 7.3.4 where the following is required of the
D&D outputs.
Software Development ◾ 159

◾ They meet the requirements set in D&D inputs (Clause 7.3.4a), i.e., we
have those deliverables available we expected to have available after
development.
◾ They are recorded in a form suitable for verification against the D&D
inputs (Clause 7.3.4), i.e., the deliverables are recognizable in light of the
requirements we had for them.
◾ They provide adequate information for purchasing, production, and ser-
vice provision (e.g., distribution, installation, and maintenance) (Clause
7.3.4b), i.e., we know what it takes to manufacture and deliver our
product.
◾ They instruct product acceptance criteria (Clause 7.3.4c), i.e., we believe
we know what an acceptable product looks like and how to tell it apart
from faulty products after the manufacturing line.
◾ They specify the essential characteristics for safe and proper use of the
product (Clause 7.3.4d), i.e., we know how the product is to be used
and we are ready to share this knowledge.
◾ They are approved prior to release (Clause 7.3.4), i.e., the deliverables
are approved to be used for the next stages before moving on.

The activity here should be straightforward to perform if your soft-


ware-development process has proceeded more or less as expected.
All that is required here is that you compare your outputs against the
inputs, record your assessment, and approve the outputs if they can be
approved.
The inputs you are comparing against may be objects such as require-
ments, specifications, and perhaps design drawings. The corresponding
outputs may then be, for example, lists of selected parts, assembly instruc-
tions, comprehensive instructions for the production line, and user instruc-
tions. In the case of a software product, the outputs may include the
software binary, the installation package, and the instructions for use, for
example.
In practice, the approval of the outputs may take the form of a design
output review documented in your QMS (see Section 5.9.1 for document-
ing reviews). The review may reference various documents and the reviews
conducted earlier during the software-development process. Here the objec-
tive is to ensure the outputs are fit for use by the later stages, i.e., to ensure
everything that was expected to be produced has been produced. The cor-
rectness of the outputs will then be addressed in the verification stage up
next.
160 ◾ Medical-Grade Software Development

5.8.1.2 Verification
The bulk of the work on performing D&D verification activities is discussed
in Section 5.9.2. You have already looked at this section earlier when per-
forming the verification steps expected in the previous stages by IEC 62304.
ISO 13485 did not loom large in the earlier work. Here, before product
release, the standard expects that your overall verification activities have
been performed in accordance with your documented arrangements to
ensure the D&D outputs meet the D&D inputs (Clause 7.3.6). This much is
achieved by the output review discussed in the previous section.
In addition, in a closer study of the ISO 13485 standard it becomes appar-
ent that it also wants to see the following verified:

◾ D&D outputs are verified as suitable for manufacturing (Clause 7.3.8)


◾ Purchased product that affects your product realization meets the pur-
chasing requirements, and this must be based on supplier evaluation
results and proportionate to the risks associated with the purchased
product (Clause 7.4.3)
◾ Acceptance criteria for verification of medical-device installation activi-
ties must exist (Clause 7.5.3)

Furthermore, ISO 13485 expects that you verify, both now as is necessary
and in the future as needs emerge, the following:

◾ D&D changes are verified before implementation (this is to be under-


stood as before the changes take effect and can have any effect outside
of their development)
◾ Production and service provision must be controlled, and these records
verified (Clause 7.5.1)
◾ Product must be monitored and measured to verify that product
requirements are being met (Clause 8.2.6)
◾ Monitoring and measuring equipment may need to be verified at speci-
fied intervals or prior to use against measurement standards and in
accordance with documented procedures (Clause 7.6)
◾ Any customer property provided to the organization for use or incorpo-
ration into the product must be verified (Clause 7.5.10)
◾ If rework is performed the product must be verified to meet applicable
acceptance criteria and regulatory requirements (Clause 8.3.4)
◾ If servicing is a specified requirement, instructions on verifying that
product requirements are met after servicing must exist (Clause 7.5.4)
Software Development ◾ 161

If the body of your verification work performed thus far falls short on any of
these aspects here is the place to correct the situation. If you are confident
that the verification work done meets all these needs, then congratulations:
you have finished your verification activities for now. You may document
this assessment in a verification review of its own, but a leaner approach is
to document it in the Design Transfer Review coming up in a short while.
Note that the verification of the above aspects and their acceptance
criteria, if specific to the product, should be instructed in any planning of
product realization (Clauses 7.1c and 7.3.2c). This information must be avail-
able from your software development plan, but the plan may reference it
from your SOP or another relevant document. Remember also that you must
document the interrelation of personnel who verify work affecting product
quality and ensure their independence and authority necessary to perform
these tasks (Clause 5.5.1) – these are best addressed in the quality manual of
your QMS (see Juuso 2022).

5.8.1.3 Validation
The validation work required for your product is addressed in Section 5.9.3.
In addition to requirements discussed in that section for the product itself,
ISO 13485 also wants to see that any production and service-provision pro-
cesses where the output can’t be or is not verified before the product is in
use or the service has been delivered must be validated (Clause 7.5.6). This
is primarily a topic for your QMS, but it is something worth pointing out
here as we near the launch of a product developed under that QMS.
Also note that if your validation work causes any changes to be made to
your earlier verification work, risk-management work, or other aspects of
your product and its documentation you should react accordingly here. If all
is as it should be you can move ahead to the release stage.

5.8.1.4 Production and Service Provision


Clause 7.5 provides a further list of items to consider before shipping the
product. Here you are required to address the following:

◾ Control of production and service provision


◾ Cleanliness of product (incl. sterile medical devices)
◾ Validation of processes for production and service provision (incl.
sterilization)
162 ◾ Medical-Grade Software Development

◾ Preservation of product
◾ Control of monitoring and measuring equipment
◾ Installation and servicing activities
◾ Identification and traceability
◾ Handling customer property

If you follow the QMS model set out in our earlier book (Juuso 2022), these
will be covered by SOPs on infrastructure (SOP-3), suppliers and distributors
(SOP-5), communications, sales, and post-market (SOP-7), and product realiza-
tion (SOP-10). Your objective here is to consider the relevant aspects and make
the provisions needed to ensure the right information and processes are in
place to move forward with the launch. See Juuso (2022) for more discussion.

5.8.1.5. Release
After the D&D outputs and all verification and validation work has been
approved it’s time to move on to software release together with the IEC
62304 standard. The requirements for this leg of the journey in ISO 13485
are in Clause 7.3.8 on D&D transfer.
Here the fundamental requirement is that your organization has proce-
dures for moving from D&D outputs to manufacturing. The principal require-
ment is now that D&D outputs are verified as suitable for manufacturing
before coming to final production specifications, and that production capa-
bility is ensured to be capable of meeting the product requirements (Clause
7.3.8). As always, the results and conclusions of the transfer must be recorded.
In practice, it should be an efficient approach to develop a Design
Transfer Review that accomplishes the above. This same document may also
conveniently record the outcome of all verification and validation activities
as it needs to comment on their outcome to provide clearance for moving
ahead to production.

5.8.2 Expectations from IEC 62304


5.8.2.1 D&D Output Review
IEC 62304 is oblivious to the existence of any D&D output stage as understood
by ISO 13485. Whether IEC 62304 is living in the Matrix, The Truman Show,
or just plain doesn’t care about such a mile marker on the way to a finished
product is immaterial: only the ISO 13485 standard places requirements here.
Software Development ◾ 163

5.8.2.2 Verification
Clause 5.8 for software release does not talk of any further verification
activities, but it does require that all verification activity is complete and that
all results have been evaluated. It therefore makes sense to now take stock
of everything we have done. As a reminder, you have already verified the
following according to the IEC 62304 standard:

◾ Software requirements (Section 5.2)


◾ Software architecture B C (Section 5.3)
◾ Detailed design C (Section 5.4)
◾ Implementation of software units B C (Section 5.5), and that any test pro-
cedures involved have been assessed to be adequate
◾ Software integration B C (Section 5.6), and that any test procedures
involved have been assessed to be adequate
◾ Software-system testing (Section 5.7), and that the appropriateness of
the verification strategies and test procedures is evaluated

In addition, the IEC 62304 requires that you verify, both now and in the
future as needs emerge, the following:

◾ The effectiveness of any changes made to the software (Section 5.7)


◾ Documented risk-control measures have been verified B C, and this
includes an assessment of whether the measures may lead to new haz-
ardous situations (Section 7)
◾ Resolved problems have been verified (Section 9)

In practice, the completion of all verification activities should also be


checked and stated in some review documents. Thus, you should verify that
you have verified everything you needed to verify. The practical place to do
this is addressed below along with the discussion on IEC 62304 expectations
on release.

5.8.2.3 Validation
The earlier D&D output review was something IEC 62304 was clueless
about, but the same can’t be said of validation. Here IEC 62304 is adamant
that it does not cover validation, not even when the software developed
constitutes the entirety of the medical device being released. The standard
164 ◾ Medical-Grade Software Development

offers up IEC 60601-1 and IEC 82304-1 as possible sources of instruction on


validation activities (see Section 12.2.3).
The standard does want to see the manufacturer synchronize software
development with the development of the wider system, including the
validation of that system, in some way in the software development plan
(Clause 5.1.3b). In other words, IEC 62304 knows there are perimeter fences
around it but rather than poke through them it is happy to graze on and
let other standards govern those outer parts of the world. If only the rap-
tors in Jurassic Park had been as complacent the humans on the other side
of the fences could have been saved a great deal of harm and hazardous
situations.

5.8.2.4 Production and Service Provision


IEC 62304 does not approach production and service provision explicitly.
The standard does expect processes for software maintenance, change
management, and problem resolution to be in place and functioning appro-
priately. It also speaks of post-production information and a necessity to
maintain risk-management activities in a few contexts and thus expects these
to be up and running for the life cycle of the device.

5.8.2.5 Release
Finally, Clause 5.8 sets the following requirements on the manufacturer for
software release to take place.

◾ Completion of planned activities


Ensure that all software-verification activities are complete, and their
results have been evaluated (Clause 5.8.1). In case of B- or C-class
software, you are also to ensure all software development plan (or
maintenance plan) activities and tasks are complete and the associated
documentation created (Clause 5.8.6).
◾ Establishment of delivery procedures
Ensure you have established procedures to establish reliable delivery
of the released medical-device software to the point of use without cor-
ruption or unauthorized changes (Clause 5.8.8). The procedures must
address the production and handling of media containing the software
product (including, as appropriate, replication, media labeling, packag-
ing, protection, storage, and delivery).
Software Development ◾ 165

◾ Documentation of residual anomalies


Ensure that all residual anomalies are documented (Clause 5.8.2), and
in the case of B- or C-class software, also evaluated to ensure they do
not contribute to an unacceptable risk (Clause 5.8.3).
◾ Documentation of version
Ensure that the version of the medical-device software release is
documented (Clause 5.8.4), and in the case of B- or C-class software,
the documentation also addresses how the software was created (i.e.,
the used procedure and environment is documented) (Clause 5.8.5).
◾ Archival of software, configuration items, and documentation
Ensure the medical-device software, configuration items, and docu-
mentation are archived (Clause 5.8.7a-b). The minimum time the
archived items are to be available is defined as the longer of a) the
lifetime of the medical device as defined by the manufacturer or b) the
time specified by relevant regulatory requirements.

Note that many of the above requirements previously only applied to B- and
C-class software, but since the 2015 amendment the large majority now also
applies to class A software. The exceptions to this are noted above.
The IEC 62304 does not require any document by a specific name here,
as indeed it doesn’t do elsewhere either, but here no specific document is
even implied to exist. In practice, recording the above assessments is neces-
sary, and the most convenient location to do so in your documentation is
discussed next in the suggested synthesis.

5.8.3 Suggested Synthesis
Table 5.10 provides a glance over the main requirements of both standards.
Note the software safety classification used by IEC 62304 and remember
that some clauses of ISO 13485 may be excluded with a valid rationale (see
Juuso 2022).
In general, the two standards can be quite neatly combined by fitting
most of IEC 62304 requirements between what is called the D&D input
stage and D&D output stage of ISO 13485. The quality-management system
standard is happy giving the larger framework for the work, and the soft-
ware-development standard is only too happy to dive into the details of the
actual development work. For the most part the requirements of ISO 13485
can be conveniently used to shade or augment the meaning of what a plan,
a review, or verification activity means.
166 ◾ Medical-Grade Software Development

Table 5.10 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Ensure D&D outputs meet D&D input requirements ✓ - - -
Ensure software verification is complete ✓ ✓ ✓ ✓
Ensure software validation is complete ✓ - - -
Ensure D&D transfer requirements are met ✓ - - -
Document known residual anomalies - ✓ ✓ ✓
Evaluate known residual anomalies - ✓ ✓
Document released versions ✓ ✓ ✓ ✓
Document how released software was created ✓ ✓ ✓
Ensure activities and tasks are complete ✓ ✓ ✓
Archive software ✓ ✓ ✓ ✓
Assure reliable delivery of released software ✓ ✓ ✓ ✓

There are, however, additional profoundly important requirements on


what the ISO 13485 wants to see take place before that thing you made
is released to the world. The microscope here is focused squarely on the
medical-device product you will make, and not the software units, items,
and systems IEC 62304 had its eye on. The work IEC 62304 had you do is
profoundly important, too, none of it will be wasted here, but as that stan-
dard was quite clear from the start that it didn’t address the validation or
final release of a medical device (see Clause 1.2 under Section 3.2), here is
where ISO 13485 steps in.
The earlier software development stages have already seen to most of
the heavy lifting involved in developing and testing your software. The
steps required by ISO 13485 on top of any requirements made by IEC 62304
stages are the D&D output review, D&D verification, and D&D validation.
These will be addressed in the following via the Design Output Review
(DOR), the Design Transfer Review (DTR), and the validation report.

5.8.3.1 Design Output Review (DOR)


As noted above, the Design Output Review is where you will take stock
of the results produced by your development activities. Here in the release
Software Development ◾ 167

stage, IEC 62304 sees all your verification activities as finished and it is time
to check that all the requirements placed on the activities by ISO 13485 are
also satisfied. In essence, what is required here is that you look at your IEC
62304 verification results from the earlier development stages and assess
whether these together meet the requirements of ISO 13485. See Section
5.9.2 for details.
By working on the implementation of your software, you will have
already performed several verification reviews that can now be referenced
here. These verification steps include both software-unit verification (see
Section 5.5) and software-integration verification (see Section 5.6) thereby
addressing both the trees and the forest of your software as IEC 62304 sees
it. You have also evaluated the adequacy of your verification strategies and
test procedures and checked that all software requirements have been veri-
fied when performing software-system testing (see Section 5.7 for both).
Here you will want to make use of these reviews instead of repeating work.
Refer to Section 5.8.1 for the contents of the review. Also remember to check
on the verification requirements from IEC 62304 detailed above in Section
5.8.2, naturally.
If everything has gone smoothly you are probably quite happy drawing
up the review here. If all is well, you can close the review and move to per-
forming or perhaps finishing your validation work.

5.8.3.2 Validation Report
Taking the waterfall model of software development, your validation activ-
ity will take place after the verification activity has been completed to the
desired effect. In principle the verification and validation activities (often just
called V&V) may be completed piecewise and may overlap, but as validation
may involve costly clinical studies you will want to make sure your soft-
ware works as you intended before sending it out for validation. In practice,
your clinical evaluation report (CER) will be a big piece of the puzzle here.
Validation is discussed in detail in Section 5.9.3.
After validation is in check you can move on to performing your third big
review, the Design Transfer Review.

5.8.3.3 Design Transfer Review (DTR)


The last of your three big software development reviews is the Design
Transfer Review (DTR). Here the objective is to complete any final
168 ◾ Medical-Grade Software Development

assessments for the release to take place as described above. You may also
take this opportunity to once more go over your product’s risk-management
activities to see if the validation work you performed should have any
impact on those activities.
A key factor here is that if you have defined product acceptance crite-
ria in your SOP or in the product-specific documentation (see DOR above)
those are satisfactorily met here. The ISO 13485 standard speaks of product-
acceptance criteria (Clauses 7.1, 7.3.4, and 8.2.6) in the sense that faulty
products coming from your production line can be detected, but as software
production is usually akin to cloning after the development is complete the
place to ensure product acceptance is here when making the transfer to pro-
duction. Therefore, you should ensure that any product-specific acceptance
criteria from the DOR are assessed here, and you may also choose to look at
all or some of the following.

◾ All work defined in the software development plan (SDP) is satisfacto-


rily complete. This includes, in particular, any testing, verification, and
validation activities.
◾ Product documentation meets the expectations of IEC 62304 (e.g.,
DOR supports moving forward) and any regulatory ­requirements
(e.g., TF, IFU, label, DoC, and UDI). Note that some final
­documents, such as the declaration of conformity (DoC) may not be
signed yet.
◾ The product and its documentation have been developed under your
QMS, and the status of that QMS is appropriate (e.g., ISO 13485 certifi-
cation is valid)
◾ The risk-management file of the product meets the requirements of
ISO 14971 and your QMS. The file is complete and its final benefit/risk
assessment supports releasing the product.
◾ The information expected for product and service provision is in place
(incl. installation, training, and servicing/maintenance) to the extent that
is defined as appropriate by your D&D documentation. Note that this
may have already been satisfactorily achieved by the DOR.
◾ The processes for product maintenance and post-market surveillance
are in place.
◾ Regulatory approval/clearance has been conducted in accordance with
the regulatory requirements or will be prior to the launch of the prod-
uct or is not appropriate for this type of product.
Software Development ◾ 169

◾ No obstacles, aside from any pending regulatory approvals/clearances,


are seen to exist for releasing the product and signing the declaration of
conformity

Note that ISO 13485 speaks of various types of acceptance criteria in many
clauses (e.g., 7.3.6 for verification, 7.3.7 for validation, 7.4.2 for purchases,
7.5.3 for installation, 7.5.6 for process validation, 8.3.2 and 8.3.4 for fixed
product), as does the IEC 62304 (Clause 5.5.3 for software units, and Clause
8.2.1 for changes), but these are not directly relevant to the topic of accep-
tance criteria for your final product. Both standards do, of course, expect
both your product and processes to meet the requirements of the standards,
the user, and any applicable regulatory requirements. The place you have to
be satisfied this really is the case is here, but the heavy lifting involved in
making the necessary checks has hopefully already been done when follow-
ing your instructed processes.
In practice, you may want to look at the above list in detail when get-
ting ready to release version 1.0.0.0 of your product but then employ a
streamlined approach when making subsequent releases. The streamlined
approach could focus on the changes made, the continued usability of your
previous DTR or its revision, and any other developments of import. The
exact chronological order of the DTR and your regulatory approvals/clear-
ances may also be debatable, but generally speaking you should not sign a
declaration of conformity before the product truly is fit for release. A typical
marching order here might be that you conduct a DTR, use those docu-
ments for communicating with the regulatory body (incl. an unsigned draft
of a DoC), and then sign the DoC once given a green light from your regula-
tory body.
Finally, after making sure any required regulatory approval activities have
been favorably completed you may proceed to launch your product in accor-
dance with your SOPs. This may entail passing a conformity assessment,
obtaining a UDI ID (if not yet done), completing your device label, signing a
declaration of conformity, and making registry entries (e.g., EUDAMED actor
and device registrations in Europe). These are all activities you should start
preparing for in advance, of course.
After all these steps are complete you may take a breath and let your
distributors and customers gulp up the merchandise. You will, of course,
already be thinking of the next release and you must be ready to react to
run your post-market surveillance activities and react to any support requests
170 ◾ Medical-Grade Software Development

and issues that pop up in the meantime. In any case, this is a good time to
crack open a case of champagne.

5.9 The Parts Left Out by IEC 62304


5.9.1 Conducting Reviews in D&D Stages
For IEC 62304 Clauses 5.3 through 5.6, the ISO 13485 standard does not
place requirements that talk directly to the activities, tasks, or deliverables
produced. These are topics the IEC 62304 standard conveniently adds
between the requirements of D&D inputs and D&D outputs as discussed in
ISO 13485. However, in the case you will want to claim that these stages are
performed and approved in conformance with the wider ISO 13485 standard
you must also ensure that the reviews performed during these stages comply
with the requirements this standard sets for D&D reviews in its Clause 7.3.5.
The chief requirement in Clause 7.3.5 is that you must perform systematic
reviews at suitable stages throughout your D&D activities. The reviews must
be performed in accordance with planned and documented arrangements
(i.e., observing SOPs, software development plans, and any other arrange-
ments made).
The objective of a review is to a) evaluate the ability of any result from
the stage to meet the requirements, and b) identify and propose any neces-
sary actions. The review must include participation from representatives of
the D&D stage concerned, as well as other specialists as needed. You may
also benefit from involving clinical expertise or an independent reviewer in
various reviews. In terms of records from the review, a simple review docu-
ment identifying the usual details of exactly what was reviewed, by whom,
when, and to what effect is the basic expectation here. Remember also to
comply with requirements for the control of records as defined by ISO 13485
in Clause 4.2.5. Your overall QMS will take care of the last mile here in
terms of documentation control.

5.9.2 D&D Verification
Both ISO 13485 and IEC 62304 lean on ISO 9000 for the definition of
the term verification and are thus in apparent agreement with what it
means. Verification, in general, is the confirmation through provision
of objective evidence that specified requirements have been met.
Software Development ◾ 171

This may entail measurement, but in many cases, verification is essentially


synonymous with a review of evidence. What is required here, from the
point of view of ISO 13485, is that we assess the output in light of the input
and determine whether what you got is what you wanted to get.
In other words, ISO 13485 sets the scene for how verification activities
are to be performed, IEC 62304 then gives us more specific requirements to
look at during some of those verification activities (as detailed in the previ-
ous sections of this book), and your documented arrangements should then
tell us at what milestones each verification activity takes place. Remember
that your software development plan is expected by IEC 62304 to either
contain or reference your verification planning. The most convenient way to
address this on a universal level may be to reference your overall approach
to the verification of process deliverables in your SOP. The verification work
here is a part of that overall activity.
Here, ISO 13485 Clause 7.3.6 builds on the concept of the D&D review
introduced above to require that you also perform verification – in accor-
dance with your SOP, plans, and other documented arrangements – of
the outputs meeting their input requirements. Note that unlike D&D
reviews, verification activities are not said to take place at suitable stages
during the development work. Instead, D&D verification as set out in the
standard sits as a stage of its own between the body of the development
work and the release of that work. Verification, however, does not have
to be a separate stage but can instead be built to make use of all earlier
work, including that performed during the unit and integration phases
presented by IEC 62304. This is also acknowledged by ISO 13485 Clause
7.3.2c which tasks D&D planning with identifying verification activities
that are appropriate at each D&D stage. Naturally, if there are any further
verification needs not met by these stages you will want to address those
needs too before release (i.e., D&D Transfer in ISO 13485 parlance) takes
place.
The standard does not require that you develop and describe a process
for this in an SOP, only that you perform verification appropriately and fol-
low any arrangements you have made. You must, however, have a plan for
verification that addresses the methods, acceptance criteria, and (if appro-
priate) statistical techniques including a rationale for sample size. A special
consideration here is that if your product is intended to be connected with
another device or interface this must also be covered in your verification. As
always, records will be kept, and these will include any conclusions made,
and necessary actions identified.
172 ◾ Medical-Grade Software Development

In practice, verification done according to ISO 13485, or according to IEC


62304 for that matter, is largely an internal activity ensuring that what you
obtained via the D&D process matches with what you set out to create. The
objective here is to compare the D&D output and the D&D input and assess
whether these match as expected. The result of verification may be conve-
niently documented as a review or a report summarizing the results and
approving the outcome. Here the overarching goal is to ensure you are con-
tent with the output so far and feel confident to proceed to the next stage. If
you are not, you should return to a previous step as is appropriate.
IEC 62304, as mentioned previously, sees verification as taking place as
part of the individual software-development stages. It is up to you how you
marry these expectations but performing piece-wise verification throughout
the D&D stages as expected by IEC 62304, making sure this verification is
compatible with the above requirements of ISO 13485, and then pulling the
whole body of verification work together before product release is a good
plan.

5.9.3 D&D Validation
The definition for validation used by ISO 13485 is derived from ISO 9000
Clause 3.8.13 where the term is defined as confirmation based on objec-
tive evidence that the requirements for a specific intended use or applica-
tion have been fulfilled. The point of comparison is thus the intended use
(intended purpose) of your medical device.
Similar to D&D verification, D&D validation sits as a stage of its own
preceding the release of the developed product (Clause 7.3.7). Unlike veri-
fication, though, validation is expected to take place using representative
product (e.g., initial production units) and the work is expected to ensure
the product is capable of meeting the requirements for its specified applica-
tion or intended use. The rationale for choosing the representative prod-
uct to use here must also be recorded. As with verification, here too you
must observe connections to other devices and interfaces implicated in the
intended purpose (intended use).
ISO 13485 Clause 7.3.2c does open the door to performing validation
activities throughout the D&D stages by tasking D&D planning with identi-
fying validation activities that are appropriate at each D&D stage. This is not
mentioned by the main clause on the subject, Clause 7.3.7, and may be diffi-
cult to implement given the above requirements, but it is thus possible. If the
above requirements can be met for some portion of the final device, and the
Software Development ◾ 173

validation work thus carried out in smaller meaningful segments, the over-
all validation work may then be tied together at the end – but you may run
into issues addressing the interplay of the parts and the entire device they
comprise.
Furthermore, validation is to include clinical evaluation or performance
evaluation of the product in accordance with applicable regulatory require-
ments. The two terms are defined in Clauses 3.3 and 3.17 as follows.

◾ Clinical evaluation (for medical devices)


Referring to the verification of the clinical safety and performance of
a medical device when used as intended by the manufacturer (Clause
3.3), this is based on the assessment and analysis of clinical data per-
taining to that device.
◾ Performance evaluation (for in-vitro diagnostic devices)
Referring to establishing or verifying the ability of a device to achieve
its intended use (Clause 3.13), this is based on the assessment and
analysis of some data.

Since the requirement for validation is from ISO 13485, and since IEC 62304
sits on its hands when it comes to validation, this is something required of
all medical-device software regardless of its software safety classification,
for example. In practice, the general risk-based approach, guiding virtually
all approaches to standardization and regulation, and the device classifica-
tion for your product (e.g., I–III) will dictate what is expected of your clinical
or performance evaluation here. You may find that performing a survey of
literature and public databases of devices, adverse events, and clinical tri-
als may get you far, but you may also discover that as the risk profile goes
up the larger the clinical studies you will have to perform get. You can, of
course, start work on many of these information sources before your final
device is ready.
In any case, this is a stage where you will again benefit from listening
to your customer. Understanding what your customers, users, and patients
expect and what they think of your product may be the only way of know-
ing that you all want to see the same species of trees as illustrated in Section
1.14. You may want to consider turning to the various testbeds run by hos-
pitals, medical centers, and other operators in your area to ensure an appro-
priate involvement of representative end users.
Remember also that you must have a validation plan that, like a verifica-
tion plan, addresses the methods, acceptance criteria, and (if appropriate)
174 ◾ Medical-Grade Software Development

statistical techniques including a rationale for sample size. All of your valida-
tion work must also be in accordance with your overall planned and docu-
mented arrangements. Your validation activities must then be completed
prior to the release of the product, but to avoid chasing your tail note that a
device used in the context of validation is not considered as released for use
to the customer.
Finally, records of the results, conclusions, and any necessary actions
must naturally be kept. In practice, a validation review or a report is a docu-
ment you will want to generate here as you will be referring to that docu-
ment in many different places and also for a long time to come. Also, IEC
82304-1 on health software expects both a validation plan and a report to
be established and sets requirements for both (see Section 12.2.3).
The activities in validation may be a vast exercise depending on your
product and affinity with clinical work, or it may be quite straightforward
to pull together building on from your existing customer processes, but it
is nonetheless a required part of manufacturing medical devices. Note that
due to the regulatory requirements on post-market clinical follow-up and
post-market surveillance this will not be a once-off thing either; instead you
will be monitoring and building on the clinical evaluation over the years to
come.
Chapter 6

SOFTWARE MAINTENANCE

This is the part that’s easy to neglect when first sketching out some
revolutionary new product on a napkin, but maintenance is a
key part of the proposition that allows your customers to reap the
benefits you envision, and for you to remain in business. It is also
required – not merely to maintain your products, but to also keep
an eye over them and churn out valuable post-market surveillance
information. In cinematic terms, it may not be enough to leave
an oblivious Tom Cruise behind on Earth to oversee those gigantic
water pumps – at least plan to call him up every once in a while to
see if he still thinks he is effective at maintenance, as indeed hap-
pens in the film.
Depending on your business model, you may be counting on maintenance
to bring in much of the revenue, or you may be hoping that nothing really
pops up for you to repair. In any case, you should not just be waiting with
a toy mallet for those monsters to pop out so you can squish them in record
time like in that carnival game. You should streamline your maintenance
processes so that they are effective to execute if and when you find an
urgent patch or a security update is needed. You must also keep an eye
open and use both active and passive methods to predict, pinpoint, and fix
any issues from here to eternity – or whatever you have defined as the sup-
ported lifetime of your device.
The topics covered in this clause are given in Table 6.1. Note that all the
requirements of Clause 6 apply to all software safety classes. The table also
adds a final section on the aspects not covered by IEC 62304 but expected
by ISO 13485.

DOI: 10.4324/9781003454830-6 175


176 ◾ Medical-Grade Software Development

Table 6.1 Topics of Software Maintenance


# Topic A B C
6.1 Software maintenance plan ☑ ☑ ☑
6.2 Problem and modification analysis ☑ ☑ ☑
6.3 Modification implementation ☑ ☑ ☑
6.4 The parts left out by IEC 62304 ☑ ☑ ☑
☑ = all apply, ☒ = some apply, ☐ = none apply

In general, ISO 13485 doesn’t discuss maintenance for the product in


any great detail even though maintenance and long-term support is part of
its overall worldview on medical devices. It expects your organization to
determine requirements related to, for example, delivery and post-delivery
activities, and the subsequent maintenance activities to take place, be docu-
mented, and even be a source for dredging up silent customer complaints
(see SOP-7, SOP-10 in Juuso 2022). Its Clause 0.1 even goes as far as to sug-
gest that the standard as a whole is also applicable to maintenance organiza-
tions, implying that many of the processes discussed may in fact be invoked
in the context of maintenance. IEC 62304, on the other hand, adds a great
detail of instruction on the maintenance of medical-device software.

6.1 Software Maintenance Plan


It is not enough to build it and let them come. You should also have
mechanisms in place for handling the issues and ideas for improve-
ment that may pop up once the first version of your product has
sailed. You don’t have to have that “James Bond will return in X”
signoff at the end of your credits, but you have to have an idea, a
process, and hopefully, all the necessary resources to run that main-
tenance process for the foreseeable future. Figure 6.1 illustrates this.
Software maintenance is the flip side of the coin when it comes to software
development. It is the afterlife of your development project, but at the same
time an integral part of manufacturing and releasing medical-device soft-
ware. The topic is addressed by both standards with compatible views but
different points of observation. ISO 13485 looks at everything from the point
of the status quo, your overall quality-management apparatus, and the users
Software Maintenance ◾ 177

Figure 6.1 Software maintenance must be on your roadmap.

in mind. IEC 62304 appreciates all of this but takes a more grassroots-level
approach to developing a plan for maintenance and then executing on ele-
ments of that plan almost as needed.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 7.2.1a Determination of requirements related to


product
◾ ISO 13485 Clause 7.2.3 Communication
◾ ISO 13485 Clause 7.5.1 Control of production and service provision
◾ ISO 13485 Clause 7.5.4. Servicing activities
◾ ISO 13485 Clause 7.3.9 Control of design and development changes
◾ IEC 62304 Clause 6.1 Establish software maintenance plan

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

6.1.1 Expectations from ISO 13485


The ISO 13485 doesn’t have a clause on a software maintenance plan, no
such document is in fact required by the standard, but a manufacturer is
required to have a documented process for controlling D&D changes (Clause
7.3.9). Technically controlling changes is not the same as being equipped
to do maintenance and actively staying on top of maintenance needs, but
in practice, the two are coupled. The requirement to monitor both produc-
tion and post-production information and act accordingly is implicated by
the introduction of the term post-market surveillance (Clause 3.14) and
178 ◾ Medical-Grade Software Development

stated, for example, in Clause 7.2.1 where you are expected to determine
the requirements related to the product including any post-delivery activi-
ties. Clause 7.5.1f then requires you to implement the determined activities.
Similarly, Clause 8.2.1 requires you to gather information from both produc-
tion and post-production activities, and any issues detected here should not
end up under the carpet.
Also note that Clause 7.5.4 directly discusses servicing activities related
to the product. Although the frame of thought here is first and foremost on
physical products shipped to the customer and requiring maintenance to
stay on-spec it may be worth comparing your situation to the requirements.
What is required is that you determine the servicing activities required
for your product and then define the processes and reference materials as
relevant. You are also required to keep records of maintenance activities
regardless of whether these are performed by you or your supplier and ana-
lyze the records for possible complaints. In the case of a SaaS product, you
may find that continued maintenance will also involve assessing updates
and revalidation needs of the tools and SOUP components your product
may utilize.
Similar to Clause 7.5.4, also note that as a small but potentially quite
important detail, Clause 6.3 discusses the need to document maintenance
activities for infrastructure, which may or may not be relevant here depend-
ing on the role of that infrastructure in the maintenance of your product and
its quality. You are, of course, required to maintain the ability of your entire
QMS to allow your processes to run as intended, and ensure resources and
competencies across your organization as needed. Given this is the case,
the link to your product may be more direct than is obvious at a first glance
to Clause 6.3, and thus you may want to make the appropriate provisions
in your software maintenance plan. If, for example, you maintain a server
backend or a cloud-based software system it may be critical to address
the maintenance of this part of your product adequately here. In the same
vein, note that Clause 7.5.1 sees control of production and service provision
as a continuous practice to ensure the product conforms to its specifica-
tions. Here too the emphasis is on the production environment instead of
the product out in the field, but for a software product the two may have a
strongly symbiotic relationship.
In addition, you will want to address how the various types of prob-
lems discussed in Section 9 will be linked to your software-maintenance
process. For example, how customer feedback and complaints, detected
nonconforming products or other nonconformities may factor in. Customer
Software Maintenance ◾ 179

communication (Clause 7.2.3) in particular is a topic invoked by IEC 62304 in


Clause 6.1, but also addressed in depth by ISO 13485 in a more general con-
text (see SOP-7 in Juuso 2022). The role of rework may be something you
wish to address or exclude in the plan or better yet in the higher-level SOP
(see SOP-10 in Juuso 2022).

6.1.2 Expectations from IEC 62304


The manufacturer is required to establish one or more maintenance plans
for conducting the tasks and activities of Clause 6. The plan must address
the following:

◾ Procedures for receiving, documenting, evaluating, resolving, and track-


ing feedback arising after the release of medical-device software (Clause
6.1a)
◾ Criteria for determining if the feedback constitutes a problem (Clause
6.1b)
◾ Use of the software risk-management process (Clause 6.1c; see Section
7)
◾ Use of the software problem-resolution process for analyzing and
resolving problems arising after the release (Clause 6.1d; see Section 9)
◾ Use of the software configuration-management process for managing
modifications to the existing software system (Clause 6.1e; see Section
8). Note that here the term “software system” is used instead of the term
“medical device software” used above
◾ Procedures to evaluate and implement upgrades, bug fixes, patches, and
obsolescence of SOUP components (Clause 6.1f)

Also keep in mind that IEC 62304 expects to see a maintenance plan cover-
ing work according to the maintenance requirements it makes in Clause 6,
not just matching the above list.

6.1.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 6.2 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
180 ◾ Medical-Grade Software Development

Table 6.2 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Documented process for controlling D&D changes ✓ - - -
Documented process for feedback handling ✓ ✓ ✓ ✓
Documented process for implementing upgrades, bug - ✓ ✓ ✓
fixes, patches, and obsolescence of SOUP components
Criteria for feedback constituting a problem (complaint) ✓ ✓ ✓ ✓
Determine servicing activities ✓
Use of risk management to advise planning - ✓ ✓ ✓
Use of problem-resolution process - ✓ ✓ ✓
Use of configuration management - ✓ ✓ ✓

Both standards expect a range of processes to be in place, although you


are perhaps allowed to define your own relationship with activities such as
servicing and control of the production environment to the extent it cor-
responds with the platform for deployment. Remember also that ISO 13485
lets you exclude certain clauses with a valid rationale, but this should be
done on a high level in your quality manual or SOPs, not in your mainte-
nance plan (see Juuso 2022).
In fact, all the necessary instructions could be contained in an appropri-
ate SOP and not your maintenance plan. In practice, though, you will want
to make product- or project-specific arrangements and resource allocations,
and a good place to document these may be in the maintenance plan. The
process for the actual change control is addressed in Section 8.2, but here
you are expected to address the possibility of changes and the overall pro-
cess for maintenance work.

6.2 Problem and Modification Analysis


Problem and modification analysis sounds like something Georg
Clooney’s suave character in Up in the Air might be into, both
professionally and in his personal life. This is where you land at the
realization that something might need to change, and you consider
Software Maintenance ◾ 181

all of the effects that change might have. This is not yet where
you run to Vera Farmiga’s front door and hope to give your life a
romantic root canal. Figure 6.2 illustrates another example of beg-
ging for a reaction.
Analyzing the need and the effects of change before diving into making
those changes in your software product is something both standards can
relate to. In general, software changes should not involve any amount of
MacGyvering. IEC 62304 goes out of its way to instruct the change-manage-
ment process here (see also Sections 6.3, 6.2 and 8.2), and ISO 13485 also
has expectations on the matter. A good change-management process with
an adequate analysis of the proposed change will be your key to figuring
out if what you are embarking on really is a quick fix or something closer
to a brand-new product development project. The changes you have made
to your product, and how you have validated these changes (see Section
12.2.3), are a hot topic for future conformity audits – and topics where you
don’t want to have to start second-guessing yourself when seated in the hot
seat.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 7.3.9 Control of D&D changes


◾ IEC 62304 Clause 6.2 Problem and modification analysis

Figure 6.2 My dog ate my software.


182 ◾ Medical-Grade Software Development

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

6.2.1 Expectations from ISO 13485


The clause of ISO 13485 of most immediate concern to the analysis of modi-
fications to the product through the provision of new releases is Clause 7.3.9
on the control of D&D changes. The topic is discussed in detail in our previ-
ous book (Juuso 2022), where SOP-10 contains the relevant guidance.
In terms of analyzing modifications, ISO 13485 calls for a proactive
approach where each contemplated change must be identified, reviewed,
verified, and validated (as appropriate) before implementation. Here verifica-
tion is to be understood as verification before the changes take effect and
can have any effect outside of their development.
Your organization is expected to determine the significance of the pro-
posed change to the function, usability, safety, and applicable regulatory
requirements for the medical device and its intended use. The review must
also look at the effect on constituent parts, products in the process or
already delivered, risk management (inputs and outputs), and the product-
realization processes. Records, as always, must be kept on the reviews and
any actions found necessary.

6.2.2 Expectations from IEC 62304


Clause 6.2 addresses your approach to problem and modification analysis,
i.e., analysis of the need for changes. The fundamental requirements here
are to monitor feedback, evaluate it, detect problems, use the problem-
resolution process (see Section 9), employ change requests, and communi-
cate with users and regulators as needed. Each of these requirements forms
a brief subclause under Clause 6.2, which may not be ideal for readability
as we have headings up to level four, but the requirements themselves are
simple. Incidentally, a change request as a term is set in Clause 3.4 by defin-
ing it as a documented specification of a change to be made in the medical-
device software. Thus, the term is not necessarily invoked for software that
is not a part of the medical device.
Clause 6.2.1 requires the manufacturer to monitor feedback on medical-
device software released for a specific intended use. The wording here raises
questions about the possibility of filtering the feedback based on the defined
Software Maintenance ◾ 183

intended use and leaves open your coverage of, for example, off-label use.
You must document the feedback and evaluate it for possible problems in
the product, though. If such problems are found, you are to record a prob-
lem report with actual or potential adverse events and deviations from the
specifications (see Section 9). Finally, you are to evaluate the reports to
determine whether product safety is affected (considering the intended use
also) and whether a change in the product is needed.
Clause 6.2.2 drives home the fact that you are to use your problem-reso-
lution process to address the problem reports. A nonbinding note highlights
the possibility that the problem could show that the software safety classifi-
cation of a software system or a software item may be incorrect. The prob-
lem-resolution process may therefore suggest changes to the classification,
and if so, this is to be documented and communicated after the completion
of the resolution process.
Clause 6.2.3 calls you to analyze change requests (which may follow from
the analysis of problem reports) to assess the effect of the change on your
organization, the product (considering the intended use also), and the sys-
tems it interfaces with. The object of it is left somewhat vague as it could
refer to interfaces between the change request and systems, or the product
and systems. Prior to the 2015 amendment this clause only applied to soft-
ware safety classes B and C, but since the amendment the clause applies to
all safety classes.
Clause 6.2.4 requires you to evaluate and approve change requests which
modify a released product. The wording of the requirement leaves it open
to interpretation whether what is meant are just changes when the modifi-
cations they make are released (at the point of release) or any changes to
a working copy where an as-of-yet-unreleased new version of a software
is modified. What is certain is that implementing change requests without
evaluation and approval will be a red flag for conformity and for the smart
use of your resources.
Clause 6.2.5 addresses the important topic of communicating with your
users and regulators. Here you are required to identify the change requests
that affect a released product. You must, as required by local regulation,
inform your users and regulators about:

a) Any problem in a released product and the consequences of continued


use unchanged
b) The nature of any available changes to the released product, and how
to obtain and install these changes
184 ◾ Medical-Grade Software Development

In practice, Clause 6.2.5a shoots itself in the head by requiring notifica-


tion of all detected problems. If this clause were to be followed the users
and the regulators would be inundated by benign reports of minor feature
and design changes made in response to detected problems. The recipients
would not have sufficient resources available to address and understand
what they are receiving, and far more serious problems would go unnoticed
under the barrage. The goal is a good one, of course, but some triage is
probably going to be called for in following this requirement. Note though
that the trigger for communication here is not the change request, but the
problem, and the problem is not trip-wired to the problem report. Thus, in
practice, you should be able to employ common sense here and develop a
process that ensures the right level of reporting for transparency and safety
to be preserved without shell shock at either end.
Also note that the topic of Clause 6.2 is problem and modification
analysis, which ought to imply that changes are possible also without the
occurrence of problems and that these ought not always lead to regulatory
reporting. In this sense, the clause appears somewhat old-fashioned and
less than optimally suited for making improvements to a product and ensur-
ing the smartest pipeline for facilitating the validation and release of those
improvements.

6.2.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 6.3 provides a glance over the main requirements of both

Table 6.3 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Monitor feedback on the product ✓ ✓ ✓ ✓
Detect problems and use the problem resolution process ✓ ✓ ✓
Use problem reports (incl. determine effect on safety) ✓ ✓ ✓
Use change requests (incl. analyze potential changes, ✓ ✓ ✓
approve appropriately)
Communicate with users and regulators (incl. problem, ✓ ✓ ✓ ✓
impact of continued use without fixes, availability of fixes)
Software Maintenance ◾ 185

standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
In terms of software-related problems and modification analysis call-
ing for a D&D change to be made, the above gives a good overview of
the requirements. In case your problem corresponds with a complaint, a
nonconformity or a nonconforming product, you will want to refer to the
related processes in your QMS, as noted above, to ensure all the appropriate
requirements are observed.

6.3 Modification Implementation
The cartoonish molding chamber from Streetfighter and the reani-
mation process from the Universal Soldier are for us the first images
conjured up by the topic of modification analysis. The business
of enacting changes on your software product should be much
more routine and much less sci-fi than these examples, but you
may find that designing the right process to meet your needs here
requires some deep thought. Figure 6.3 illustrates what the son of
Frankenstein might look like if you’re not careful.

Figure 6.3 Problem-resolution process may be akin to field surgery.


186 ◾ Medical-Grade Software Development

The topic of how to develop and release software updates is an object of


great fascination. The most bureaucratic process might be to treat every new
version as a completely new software-development project. Your staff, cus-
tomer, and regulators might not thank you for this. It would take far too long
and be far too costly to enact any changes to the software, and thus many
useful modifications would languish in the backlog. The simplest process
would be to just do agile and push new releases out at will, but this would
not meet the requirements. The sensible approach is somewhere in between.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 7.3.9 Control of D&D changes


◾ IEC 62304 Clause 6.3 Modification analysis

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

6.3.1 Expectations from ISO 13485


As ISO 13485 does not make any special concessions here regarding mainte-
nance updates, the basic expectation is that all of its D&D requirements, as
detailed in Clause 7.3, are met. This may be understood to mean the valida-
tion and release requirements discussed in Sections 5.8 and 5.9 are upheld.
The method of verifying this really is the case and the medium for recording
the assessment may vary, but it must take place for the maintenance release
to go ahead.
Remember also that ISO 13485 calls for changes to be identified,
reviewed, verified, and validated (as appropriate) before implementation
(i.e., before the changes take effect and can have any effect outside of their
development).

6.3.2 Expectations from IEC 62304


Clause 6.3.1 requires the manufacturer to use their established process to
implement the modifications assessed and approved in the previous section
(Clause 6.3.1). This includes the identification and implementation of any
Clause 5 activities (see Section 5) that need to be repeated because of the
modification. A nonbinding note refers to Clause 7.4 for requirements related
to risk management of software changes.
Software Maintenance ◾ 187

Clause 6.3.2 then requires the manufacturer to follow their process for
making releases as discussed in Section 5.8. A nonbinding note explains that
such releases may now take place as a) a full release of the software system
or b) an incremental batch comprising the changed software items and the
necessary tools to apply the modifications to the software system.

6.3.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 6.4 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
So, how do you design a safe, streamlined process for conducting soft-
ware maintenance and making the necessary releases? The internal main-
tenance process up to the point of the release is simple enough to instruct.
Where the beads of sweat start to appear on your forehead is in figuring
out what work from the previous stages (ISO 13485 D&D and IEC 62304
Software development) needs to be redone to move ahead with the release.
The great news here is that it depends on what has changed in the new ver-
sion of the product, and this is to be evaluated by you here.
Based on what has changed you can go through the technical documen-
tation and assess, even one document at a time, whether that document
still holds true or whether the document needs to be revised somehow. If
you come to the conclusion that some document needs to be revised, you

Table 6.4 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Observe D&D requirements the same way you would ✓ - - -
for a first release
Observe software development requirements the same - ✓ ✓ ✓
way you would for a first release
Repeat software development activities as needed - ✓ ✓ ✓
Use modification process to make changes - ✓ ✓ ✓
Use release process ✓ ✓ ✓ ✓
188 ◾ Medical-Grade Software Development

will then be able to assess how much of the work going into that document
needs to be redone. After having gone through your entire set of techni-
cal documentation you will have an excellent idea of what has changed
and what needs to still change as a result of allowing the new release to go
through. Make sure you do not compromise on the release conditions you
may have set in your processes, and that you are happy with the validation
status and risk management file also covering the new release.
Over time as you get more experience with running the maintenance
process you will start to have a better idea of what changes affect what
earlier work and documentation. You will then be able to pre-empt some
work while working on a maintenance release and pay less attention to
some areas of the documentation. Section 11 will give you our map of the
essential technical documentation, and the relationships between the vari-
ous documents. This discussion will hopefully help you in streamlining your
underground map from a change need to a released update.

6.4 The Parts Left Out By IEC 62304


There is both not a lot and a great deal that IEC 62304 does not cover
regarding product maintenance. ISO 13485 sets up the foundations of per-
forming the work from top management responsibility, resource allocation,
and personnel competence to listening to and reacting to various sources of
information and the events arising out of them as triggers to maintenance
and problem-resolution activities. These were all covered above, and are
discussed in greater detail in our previous book on ISO 13485 quality man-
agement (Juuso 2022). We are thus happy to wrap up software maintenance
here and move to a topic as integral to medical devices as quality manage-
ment itself: risk management. Both of these topics form an undercurrent to
every single clause of IEC 62304.
Chapter 7

RISK MANAGEMENT

We’ve all seen War Games, Terminator, and 2001. We can imagine
what an autonomous software system on a rampage might look
like. We also remember the AI from The Hitchhiker’s Guide to the
Galaxy that humanity placed its hopes on for generations only to
discover after all that time that the meaning of life was 42. With
great expectations come great fears, and already for that reason
alone it is to be expected that risk management features heavily in
any discussion of medical-grade software.
The deep end of the pool for medical-device risk management is instructed
by the international ISO 14971 standard. The intention here in Clause 7 of
the IEC 62304 standard is to make some further requirements to address the
special nature of software-specific risks, and to ensure that the risk manage-
ment for the software and any device using that software occurs in some
synchronized manner. The foundation for the use of both standards is, of
course, set by ISO 13485.
Safety is the operative word for all three standards. ISO 13485 grounds
risk as something pertaining to the safety or performance requirements of
the medical device or the meeting of applicable regulatory requirements
(Clause 0.2). The standard then does a bit of a doubletake to define the
applicable regulatory requirements as the requirements set for the quality
management system and the safety or performance of the device. To com-
plete the circle here, even that performance part is occasionally explained
by adopters of the standard to be examined via the safety of the device (e.g.,
the ability of the device to consistently function in a safe way). The top-level
takeaway that emerges here is that the safety of the device is always key,

DOI: 10.4324/9781003454830-7 189


190 ◾ Medical-Grade Software Development

and an appropriate QMS is seen as a key part of getting there. IEC 62304
takes a compatible view by placing great emphasis on the people coming
into touch with the device, and how any risks may affect the patient, the
operator, and any other relevant people. The goal is a device that is consis-
tently safe for people. ISO 14971 looks over the two other standards here
and nods approvingly on the course taken, but expects its provisions to be
the most important for going forward.
We will look at all the assertions made by IEC 62304 on risk management
next, and comment on the requirements made by the other two underly-
ing standards and regulations as relevant. Before that, it is prudent to briefly
introduce the cast of the last remaining concepts IEC 62304 has deemed
important enough to import from ISO 14971 between its covers. The main
concepts were already introduced above in Section 1.23 as hazardous situa-
tions, hazards, harms, and risks. Here we add a few more concepts to round
out the discussion offered by IEC 62304. The terms here relate to the pro-
cess phases and the risk management file itself. The new terms to consider
are as follows.

◾ Risk management
This is the top-level umbrella term (Clause 3.19) used to discuss the
systematic application of management policies and processes to analyz-
ing, evaluating, and controlling risk.
◾ Risk assessment (incl. risk analysis and risk evaluation)
Risk analysis (Clause 3.17) refers to the systematic use of available
information to identify hazards, as they are relevant within the context of
the intended use of the device (or its foreseeable misuse), and estimate
the risk involved. As part of this, risk estimation (Clause 3.39) is used as
the process to assign values to the probability of occurrence of harm and
the severity of that harm. Finally, risk evaluation (Clause 3.40) refers to the
process of comparing the estimated risk against criteria defined on the
acceptability of risk. Together all these steps comprise risk assessment,
which is a term used by ISO 14971 but not introduced by IEC 62304.
◾ Risk control
Risk control (Clause 3.18) refers to the process where risk-control
measures (mitigations) are defined to reduce or maintain risks within
specified levels.
◾ Evaluation of residual risk
The concept of residual risk (Clause 3.38) simply refers to the risk
remaining after risk-control measures have been assigned. The residual
Risk Management ◾ 191

risk may be considered for each risk by itself and for the entire set of
risks overall.
◾ Risk management review
The concept is not introduced by IEC 62304, but this is an integral
part of your risk-management activities where the overall status of the
activities is assessed and, usually, a decision is made on the benefit–risk
ratio of your product. The review might, for example, be conducted for
product-specific risk management, process-specific risk management, or
both.
◾ Production and post-production activities
Post-market surveillance and the continuous monitoring of the risks
related to your product, but also the benefit gained by using it, are
going to be a part of your follow-up activities for as long as the prod-
uct is expected to remain in use – which may be longer than just the
period it is sold. ISO 14971 and other standards and technical reports
will guide this work (see Juuso 2022), but here it is useful to know that
risk management might never actually end.

Finally, the term risk management file is introduced in Clause 3.20 as a set
of records and other documents that are produced by a risk-management
process. The clause interestingly enough points out that these records may
not necessarily be contiguous, which should be interpreted as records cre-
ated on a varying granularity and focus over time, but not as containing
gaps or omissions.
It may also be worthwhile to recall the definition of “safety” as free from
unacceptable risk (see Section 1.25) and to look at the definition of the intui-
tively related term of “security”. Clause 3.22 defines security as the protec-
tion of information and data so that they are protected from unauthorized
reading and modification but retains support for authorized access. Safety
thus refers to the safety of people, but security is understood from the point
of view of information.
The topics covered in Clause 7 are given in Table 7.1 below. The table
also notes which software safety classes each topic is applicable to, but
please don’t make the mistake of thinking that class A software is somehow
exempt from risk management as it is not implicated for Clauses 7.1–7.3. The
requirements from Clause 4 apply to all classes of software, and the expecta-
tions from regulations and other key standards will call for risk management
done with rigor even if not always with the same amount of work as would
be done for high-risk software.
192 ◾ Medical-Grade Software Development

Table 7.1 Topics of Risk Management


# Topic A B C
7.1 Analysis of software contributing to hazardous situations ☐ ☑ ☑
7.2 Risk-control measures ☐ ☑ ☑
7.3 Verification of risk-control measures ☐ ☑ ☑
7.4 Risk management of software changes ☒ ☑ ☑
7.5 The parts left out by IEC 62304 ☑ ☑ ☑
☑ = all apply, ☒ = some apply, ☐ = none apply

The need to adopt the ISO 14971 standard as a basis for your risk-man-
agement activities is, for now, an obligatory part of adopting IEC 62304 as
defined in Clause 4.2. The call to adopt ISO 13485 as the guide for quality
management is stated in the preceding clause, Clauses 4.1. Both of these
justly factor into how you do risk management in the context of medical-
software development.
There is, however, a third clause that leans in heavily on the implemen-
tation of risk management in our context here. That is Clause 4.3 which
ostensibly addresses only the assignment and use of the software safety
classification. The stipulations made in that clause are, however, routinely
used in all risk-management activities for medical software. This is fair to the
extent that the ABC safety classification does sit in the general requirements
section of IEC 62304 and does fundamentally curate how we choose which
requirements of the standard, including those of Clause 7, to heed in the
circumstances of a specific software item. Where the utility of some of these
assertions begins to break down a little is in the actual ISO 14971 risk man-
agement around a device, as we will discuss next. For a discussion of Clause
4 itself, refer back to Section 4.
The most striking assertions made by Clause 4.3 that often bleed over to
Clause 7 are as follows.

◾ The software safety classification is conceptually associated


with the risks involved in the use of that software. All software
is to be assigned a software safety class, which by default is C, indicat-
ing a high risk. This classification is conceptually linked with the risks
related to the software, but no definite formula exists for mapping the
A, B, and C classes to the usual ISO 14971 risk classifications (e.g., 5x5
matrix of risk levels). Yet, you need to heed the impact of the safety
Risk Management ◾ 193

classification on your risk-management activities, as is evident from


Sections 7.1 to 7.4 below.
◾ All software will eventually fail. This is a steadfast expectation in
the field and one that even junior developers may be able to recite. It
is a handy assertion to make when figuring out the safety classification
for your software as it does not allow you to claim that software could
not fail just because it was programmed not to. As a carryover to then
performing the day-to-day risk management related to that software,
it is less practical and often circumvented by using both P1 and P2 as
explained in Section 4.3. The bottom line here is that while all software
will eventually fail (P1 is 100%), this failure may or may not lead to
harm (P2 is something between 0% and 100%). Today, even ISO 14971
talks of P1 and P2, and interprets these in agreement with the status
quo discussed above. Thus, to challenge the virtue of P1 = 100% would
be to run the risk of embarking on a quixotic campaign against the
windmills. In practice, you will see P1 in many risk matrices but its role
in the analysis may be debatable if it is always 100%.
◾ IEC 62304 occasionally only looks at harm in relation to people.
ISO 14971 understands harm as always relating to people, property,
or the environment but when assigning the IEC 62304 software safety
classification only consideration of people is required (Clause 4.3a).
That may be fine for assigning the software safety classification, but it is
not enough for conducting your subsequent body of risk-management
activities. Here, it would be unwise to forget about property and the
environment in your full risk management. In fact, there are a few more
considerations you should also observe in addition to the standards as
we will discuss in Section 7.5 in a short while.
◾ Risk controls should be external. Clause 4.3a suggests that exter-
nal risk controls may be used to lower an initial software safety clas-
sification. This may be mistakenly read as only risk controls external to
any software should be used, which is not what the standard expects.
Although only risk-control measures external to or independent from
the software itself may be considered when assigning the software
safety classification (Clause 4.3a), it is in fact all types of risk controls,
both internal and external, that should be considered when mitigating
risks. These can be anything from a try-catch expression used in the
source code to an elaborate check made by another module on the out-
put value provided by some function. Software-based risk controls thus
add a welcome solution to the arsenal available to software developers
194 ◾ Medical-Grade Software Development

and mean that hardware, environmental conditions, and procedural


instructions are not the only means of mitigating risks for software
products. Being able to rely on the independence of the software item
and the control (i.e., being able to prove segregation between the item
and another item controlling it) is important, though.
◾ Risk control may decrease both the probability and/or the
severity of a risk. Intuitively this makes a great deal of sense; after all,
you should be able to improve on both the probability and the sever-
ity of something going wrong if you put your mind to it. The matter
is occasionally debated, but the standard is quite clear on it in a non-
binding note to Figure 3 in Clause 4.3a. Note that the context here too
is software safety classification.

The above assertions represent some of the key philosophical pain points
of risk management and are no doubt debated endlessly on various expert
forums. Expect the whisky to run out before a complete consensus is
reached on any of these matters. The above discussion should, however,
help you navigate your way to meeting the requirements in practice.
One final requirement worth noting here before we dive into the sub-
clauses of Clause 7, is that Clause 5.2.4 requires the manufacturer to not
only re-evaluate risk assessment when software requirements are first estab-
lished but also later update this assessment as appropriate. This means,
ideally, that any change of software requirements – such as through change
management – should be accompanied by reconsideration of the risk
assessment. Don’t go overboard here and require an unnecessarily wide
overhaul just in case, but instead assess the situation and act accordingly
when making software changes.
In the following, we will look at the individual subclauses of Clause 7.
Note that the first three subclauses only apply to software safety classifica-
tions B and C, and significant leeway is given to class A also in terms of the
fourth subclause. ISO 14971 risk management, though, is required from all
classes of software by Clause 4.2.

7.1 Analysis of Software Contributing


to Hazardous Situations B C
IEC 62304 is not unhappy with the instruction given by its big
brother on risk management, the ISO 14971 standard, but it does
Risk Management ◾ 195

want some software-specific issues to be observed. Here in Clause


7.1, its goal is to have you thinking of the specific ways the nature
of software might contribute to the occurrence of harm. This is as if
Tony Soprano brought his Sicilian third cousin over from Italy and
we would be charged with predicting how he will fit into the New
Jersey family. The heavy in the equation will still be ISO 14971, but
IEC 62304 will not be a wallflower either.

IEC 62304 may have pushed the envelope when it was first published by
turning a keen eye on the software that may run in medical devices, or as
medical devices, but it did so by principally looking at how that piece of
software was put together and against what specifications. There is nothing
wrong with that, but in a world that has moved from F-14 Tomcats to F-35
stealth fighters, the standard seems more aged than Tom Cruise with the
same passage of time.
Today, it is IEC 82304-1 (see Section 12.2) that points out that it may
no longer be all that savvy to just look at your software and the hardware
surrounding it. Instead, you should look at the platforms your software
runs on and the stealthy cybersecurity threats coming at your software.
Anytime your medical-device software is connected to a network the
potential risks are catapulted to a new level. Manufacturers don’t always
pay adequate attention to what the addition of a network connection does
to the risk profile of the device and the risks associated with its features.
As a result, the risk analysis may be decidedly inadequate. Such risks can’t
be eliminated by turning a blind eye to IEC 82304-1 either, they will still
need to be on your radar via IEC 62304 Clause 7.1.2 (see below). Figure 7.1
below illustrates how the risk profile of the same functionality may have
an increased risk profile as you move from a standalone device to a device
connected to a private network, and finally, a device relying on a network
of services in the cloud. Your environment is relevant to your risk analy-
sis whether or not your standard has a word for the exact type of your
environment.
In practice, the analysis of software contributing to a hazardous situation
requires an in-depth understanding of not only the clinical procedure the
device is used in and basic software engineering, but also of the network
environment the software operates in. The risks may be related to the safety
or the security of data and the system, for example. ISO 14971 is applicable
to both safety and security risks, but it needs experience, insight, and judg-
ment to identify and appropriately manage those risks – and it may benefit
196 ◾ Medical-Grade Software Development

Figure 7.1 Standalone networks, private networks, and the cloud each come with
different risks associated with them.

from the lens offered by IEC 82304-1 in this, too. Note also that as such new
lenses are applied to identifying and managing risks, and as these are miti-
gated somehow, you may also end up affecting other risks or introducing
new risks. These may lead to new requirements, new verification needs, and
new test cases that then affect your overall design.
The use of a firewall may seem like a simple solution to countering many
of the risks brought on by the addition of a network connection. That too,
however, has more considerations than just a plug-and-play decision. You
should, for example, consider the following.

◾ Do you treat the firewall as a SOUP component? If so, does the manu-
facturer of the firewall provide lists of known vulnerabilities? How are
detected issues handled? How are updates and the need for updates
managed?
◾ Have you identified all the risks related to the use of the firewall?
◾ Have you addressed all risks related to how your software communi-
cates with the firewall?
◾ Does the use of the firewall place requirements on your software, e.g.,
in terms of how your software communicates with the firewall, or what
security features it implements? Does this affect your requirements for
any accompanying documentation?
◾ Does your validation adequately take the use of the firewall into
account?
Risk Management ◾ 197

The above list is not exhaustive, neither is it intended to scare you away
from using a firewall. The intention is only to point out that the situa-
tion is more complex when you have more components and more inter-
faces in play. This is also true when considering the interplay of devices,
both medical and non-medical, cohabiting in the same network. It may,
for example, transpire that some benign Zigbee devices intended for the
home market but used in a hospital environment will be overeager to
explore their surroundings and end up overwhelming their more limited
hospital friends. That is a true story, and an example of how the network
environment may mean that something thought too silly to happen previ-
ously (e.g., a janitor leaning against a light switch on the operating room
wall) can now easily happen when the distance grows and the physi-
cal connection between the source and the point of the harm becomes
more abstract (e.g., a janitor leaning against a wall switch in some remote
control room). A filmic reference to the Michael Crichton thriller Coma
comes to mind, but here the motive of the remotely controlled gadget
does not have to be sinister. The example is intended to be illustrative of
the unexpected complexity that may be introduced even when just tak-
ing previously available functionality into a new environment. It is up to
you to identify, prioritize, and control the risks related to the use of your
device.

7.1.1 Expectations from ISO 13485


As it happens, a similarly anointed Clause 7.1 in ISO 13485 expects the
manufacturer to implement one or more processes for risk management in
product realization, and thus an analysis of hazardous situations is required
for medical devices in general. However, no specific requirements for soft-
ware are made. The expectations for risk management in general are dis-
cussed in detail in our previous book (Juuso 2022).

7.1.2 Expectations from IEC 62304


For B- and C-class software, the manufacturer must identify software items
that could contribute to an identified hazardous situation (Clause 7.1.1).
Remember that software items may in fact be units, items, or systems fol-
lowing the definition of the term. In other words, you will have developed
a risk management file for your device where you identified hazardous
198 ◾ Medical-Grade Software Development

situations, and you are now required to pinpoint where in your software
those failures might originate from. A non-binding note instructs you to
consider not only software failures but also the failure of any software-based
risk-control measures you have defined.
Clause 7.1.2 offers a helping hand in identifying such sources by
instructing that you look at the following as potential causes for item
failure:

◾ Incorrect/incomplete functionality specification


◾ Software defects in the functionality of the item
◾ Failure or unexpected results from a SOUP component
◾ Hardware failures (resulting in unpredictable operation)
◾ Software defects resulting in unpredictable operation
◾ Reasonably foreseeable misuse

The cumbersome list above could be summarized as the faulty specification


for the item, faulty construction of the item, unexpected input to the item
(from SOUP, hardware, or other software), and reasonably foreseeable mis-
use. In effect, the clause requires you to not only identify the suspect items
but also consider the potential circumstances that could lead to the item
contributing to a hazardous situation.
Clause 7.1.3 examines SOUP components more closely. Provided that the
failure of a SOUP component, or the unexpected output of such a com-
ponent, is a potential cause for an item to contribute to a hazardous situ-
ation, you are required to take action to control SOUP components. This
must, at a minimum, entail the evaluation of any anomaly list published
by the SOUP supplier for the particular SOUP version used in the medical
device. The evaluation is to include a determination as to whether any of
the anomalies could lead to a sequence of events potentially leading to a
hazardous situation.
Finally, Clause 7.1.4 requires the potential causes behind an item contrib-
uting to a hazardous situation to be documented in the risk management
file. Note that the old subclause 7.1.5 has been deleted from the amended
version of the standard and you are no longer required to document
sequences of events potentially leading to hazardous situations in the risk
management file.
Please note that class A software gets a free pass in terms of clauses
7.1–7.3 discussed here, but it does not escape risk management altogether as
detailed in Section 4.2 above and Section 7.3.1 below.
Risk Management ◾ 199

7.1.3 Suggested Synthesis
As no software-specific requirements are made by ISO 13485 the synthe-
sis here is simple: follow the requirements made by IEC 62304 as outlined
above. Refer to our previous book for further discussion of risk management
in general (Juuso 2022).

7.2 Risk-Control Measures B C
This is where you identified certain software-specific risks involved
in the use of your device and you must now consider how to control
those risks. You might, for example, decide that the Italian cousin
should get an English-speaking sidekick before he goes out to make
his first collection rounds around the neighborhood.

7.2.1 Expectations from ISO 13485


In general, ISO 13485 is big on applying the risk-based approach and ensur-
ing that the amount of attention and effort is appropriate to the level of the
risk. Thus, it may be seen as fundamentally compatible with the use of the
software safety classification used by IEC 62304. No specific requirements
for software are made here.

7.2.2 Expectations from IEC 62304


For B- and C-class software, the manufacturer must document risk-control
measures (mitigations) for each identified risk where a software item could
contribute to a hazardous situation (Clause 7.2.1). This must be in accor-
dance with ISO 14971. Here a non-binding note reiterates that the risk-
control measures may be implemented in hardware, software, the working
environment, or user instruction.
Clause 7.2.2 specifically addresses risk-control measures implemented as
part of the functions of a software item, where it makes the following three
requirements beyond what is required by ISO 14971:

◾ The risk-control measure must be included in the software requirements


◾ Each software item contributing to the implementation of the risk-con-
trol measure must receive a software safety classification based on the
very risk the measure is controlling
200 ◾ Medical-Grade Software Development

◾ The software item must be developed in accordance with the require-


ments of Clause 5

7.2.3 Suggested Synthesis
As no software-specific requirements are made by ISO 13485 the synthe-
sis here is simple: follow the requirements made by IEC 62304 as outlined
above. Refer to our previous book for further discussion of risk management
in general (Juuso 2022).

7.3 Verification of Risk-Control Measures B C


After coming up with the perfect mitigation approach, you are also
expected to ensure that the plan is acted on. Remember that veri-
fication may represent a range of activities from a visual check to
an elaborate testing protocol. The risk itself will guide you through
addressing each risk control appropriately. In the previous example,
Tony might, for example, call up the sidekick to hear whether there
have been any issues. If he doesn’t like what he hears he might
even do spot checks or decide to talk with everyone his cousin has
recently interacted with.

7.3.1 Expectations from ISO 13485


No specific requirements for software are made by ISO 13485.

7.3.2 Expectations from IEC 62304


For B- and C-class software, Clause 7.3.1 then considers the verification of
the risk-control measures and requires that each identified risk-control mea-
sure is verified in a documented manner. The manufacturer is also required
to review the measure and determine if it could lead to a new hazardous
situation. Also remember that Clause 4.3a instructed the manufacturer to
evaluate the effectiveness of any risk-control measure used to lower the soft-
ware safety classification from B or C (see Section 4.3).
Clause 7.3.2 has been deleted from the amended version of IEC 62304
to reflect the deletion of 7.1.5 above and thus the removed requirement to
document sequences of events here.
Risk Management ◾ 201

Clause 7.3.3 addresses traceability via documentation of software hazards,


hazardous situations, software items, risk-control measures, and the verifica-
tion of the measures. Here the manufacturer is required to, as appropriate,
document traceability between a) the hazardous situation and the software
item, b) the software item and the specific software cause, c) the software
cause and the risk control measure, and d) the risk-control measure and
its verification. A somewhat cryptic non-binding note offers the ISO 14971
instruction on the risk-management report (its Clause 9 in the 2019 version)
as further reading and presumably a good place to ensure traceability has
been maintained adequately.

7.3.3 Suggested Synthesis
As no software-specific requirements are made by ISO 13485 the synthe-
sis here is simple: follow the requirements made by IEC 62304 as outlined
above. In general, though, ISO 13485 is big on a risk-based approach and
ensuring the outcome of actions is evaluated in a way commensurate with
the risks posed to the medical device. The link between the software safety
classification and the level of rigor applied to verifying risk controls is thus
justifiable from the point of view of both standards. Refer to our previous
book for further discussion of risk management in general (Juuso 2022).

7.4 Risk Management of Software Changes


Once you’ve built it, don’t batch it to smithereens. Both standards
realize that a change may not be the improvement it is intended to
be, and thus want you to pay adequate attention ahead of making
that change. If, for example, Tony decided to switch his cousin for
another henchman he thought might do a better job, he should con-
sider the risks involved in the personnel change from the first person
going ballistic on hearing he’s fired to the replacement then making
all the same mistakes.

7.4.1 Expectations from ISO 13485


ISO 13485 does not make any software-specific requirements here, but it
does speak of changes made to the product in general. See Section 8.2 for
in-depth coverage of these requirements.
202 ◾ Medical-Grade Software Development

7.4.2 Expectations from IEC 62304


For all classes of software, Clause 7.4.1 instructs the manufacturer to analyze
whether a change in the medical-device software (including any SOUP com-
ponents) a) introduces additional potential software causes contributing to
hazardous situations, and b) requires additional software risk-control mea-
sures. Note that the latter speaks of software risk-control measures, but it pre-
sumably means measures used for the software to control risks, not measures
implemented in software as also other forms of measures may be called for.
The clause then makes two additional requirements for class B and C
software. Clause 7.4.2 raises the stakes for what was instructed above by
Clause 7.4.1a and requires the manufacturer to also analyze whether the
changes (to the software or its SOUPs) could interfere with existing risk-con-
trol measures. Note that the language also changes from something going
wrong (does) to something possibly going wrong (could), and thus the scope
is also broader here. Finally, Clause 7.4.3 requires that the manufacturer
performs relevant risk-management activities defined in Clauses 7.1–7.3 above
based on these analyses.
In other words, the whole of Clause 7.4 compels you to consider any
changes made in your medical-device software in terms of the effect it will
have on the safety of that device. For all classes of software, this will include
an analysis of whether the change introduces new causes for software items
contributing to hazardous situations, and whether new risk-control mea-
sures are needed. For classes B and C this must also include an analysis of
how existing risk-control measures could be affected, and the analysis itself
should feed into other risk-management activities described in Clause 7.

7.4.3 Suggested Synthesis
As no software-specific requirements are made by ISO 13485 the synthe-
sis here is simple: follow the requirements made by IEC 62304 as outlined
above. Refer to our previous book for further discussion of risk management
in general (Juuso 2022).

7.5 The Parts Left Out by IEC 62304


As IEC 62304 in fact builds on the ISO 14971 standard for risk management,
and it is to be used on top of the foundations laid by ISO 13485 quality
Risk Management ◾ 203

management, there is not a lot the combination of these would not cover.
Here IEC 62304 is, however, quite silent on two important aspects: the types
of risks you are expected to address, and risk management during the post-
market phase of your product.
Addressing the first topic, the types of risks you should consider in the
risk management of your device go beyond just malfunctions of the soft-
ware. IEC 62304 Clause 4.3a alludes to risks beyond malfunctions when
it discusses software safety classification and the impact of software – not
just software failures – contributing to an undesired outcome, but it is not
too vocal on the matter. In practice, you will want to consider at least the
following:

◾ Risks associated with software failures


◾ Risks associated with the correct, intended use of software
◾ Risks associated with foreseeable misuse of the software

Note that only the first of these is readily addressed via techniques such as
failure modes and effects analysis (FMEA), but all three are to be considered
in conducting risk management for medical-device software. The require-
ment to do so is not made explicit in the IEC 62304, but conformity assess-
ment of the final medical device will raise questions regarding all three
types of risks.
On the second topic, IEC 62304 is happy to refer to both production and
post-production information in passing. This despite the fact that the IEC
62304 prides itself on being the software life cycle standard. Luckily, ISO
13485 is a little more articulated on the continual nature of risk manage-
ment. ISO 13485 wants to see risk management as the systematic application
of management policies, procedures, and practices to the tasks of analyzing,
evaluating, controlling, and monitoring risks (Clause 3.18). It thus opens the
lid on post-market surveillance of the device where the IEC 62304 is mostly
content with documenting the path to a ready software device and control-
ling the changes to it later.
IEC 62304, too, acknowledges that software changes may be an occasion
to reassess risk management, but it does not go into great detail on what
the relevant set of information to consider here might be. ISO 13485, and
ISO 14971 in particular, force you to keep your radar on and identify causes
for those changes. Similarly, the regulations out there such as the EU MDR
(2017/745) will insist you keep an eye on your creation also after it ships.
204 ◾ Medical-Grade Software Development

Note here that ISO 14971 also expects you to verify that any risk controls are
effective.
The adoption of the ISO 14971 standard is, for now, an obligatory part of
adopting IEC 62304. You must fashion your process according to that stan-
dard and also meet the expectations of ISO 13485 but do so while remain-
ing appreciative of the interpretations IEC 62304 wants to see in place.
Chapter 8

CONFIGURATION
MANAGEMENT

In some soap operas, the star of the show may change to another
actor mid-scene without much fanfare. In software development,
this would be considered a case of bait-and-switch. To prevent you
from accidentally using one set of deliverables to pass checks and
then delivering another set to the next stage, perhaps after some
haphazard changes to those deliverables to address failed checks,
you are required to pay attention to configuration management.

Configuration management is intuitively understood by people in the indus-


try, but many people will take a pause if you ask them to comprehensively
define what exactly a configuration is, and what parts in – and perhaps
around – your software should be considered in scope for configuration-
management activities.
The ancestral standard, ISO 9000, takes the responsibility of defining
what a configuration is in our context. In Clause 3.10.6 of the standard, the
term is defined as the interrelated functional and physical characteristics of
a product (or service) defined in the product configuration information. The
configuration information itself may be in the form of requirements or other
information (i.e., documentation of some sort) for product design, realiza-
tion, verification, operation, and support. Thus, the configuration is to be
understood as the grand constellation of your software – the planets, stars,
dust, and dark matter (i.e., SOUPs) that make up the whole of your soft-
ware. A distant cousin, the IEEE Standard Glossary of Software Engineering

DOI: 10.4324/9781003454830-8 205


206 ◾ Medical-Grade Software Development

Terminology, takes perhaps the most concrete view of a configuration item


by calling it an aggregation of hardware, software, or both that is treated as
a single entity in your scheme. In some contexts, you might thus even call
the Milky Way a configuration item if it made sense in your grand scheme
of things, but you would not expect it to just be one item if you dove deep
down into it. Imposing any configuration management over a celestial body
such as the Milky Way which moves at a velocity of approximately 1.3 mil-
lion miles per hour is probably a topic best left to astronomers, though.
ISO 13485 is oblivious to the whole concept of a configuration but
accepts the ISO 9000 definition at face value through a high-level norma-
tive reference made to the standard. IEC 62304 appears to take a live-and-let
live approach to this definition (i.e., no reference to it is made whatsoever)
and adopts somewhat of a politician’s approach to then defining the term
“configuration item” as some entity uniquely identifiable at a given refer-
ence point (Clause 3.5). Your whole software must be controlled as an entity,
and a version change on that level will require software configuration man-
agement action (note to Clause 3.34). The standard also requires that all
configuration items for B- and C-class software items are placed under con-
figuration control prior to verification activities – thus configuration manage-
ment is required to step in by the verification stage or just before release for
class A software.
The aim of configuration management is to understand what elements
you need to control, when, and how. The point of the exercise is to both
identify the constituent parts and to manage changes to the whole. In addi-
tion, the environments, platforms, tools, and equipment that touch your
software product may also be a part of your considerations here.
Both standards want to see changes being controlled by appropriate
reviews taking place before implementation and the corresponding records
being kept afterward. IEC 62304 is again more articulated in terms of what
this means in the context of software, but ISO 13485 does not sit silently
either. Now ISO 13485 has requirements on what it wants to see take place
in the context of changes being made, and it not only wants to see this evi-
dence for software but also for any other type of medical-device product.
The topics covered in this clause are given in Table 8.1 below. The table
also notes which software safety classes each topic is applicable to, although
in this case, all the requirements apply to all software safety classes. Note,
though, that the point during your software-development life cycle after
which configuration management must apply may vary based on your soft-
ware safety classification (see Section 8.1.2).
Configuration Management ◾ 207

Table 8.1 Topics of Configuration Management


# Topic A B C
8.1 Identification of configuration items ☑ ☑ ☑
8.2 Change control (incl. verification) ☑ ☑ ☑
8.3 History of controlled items ☑ ☑ ☑
8.4 The parts left out by IEC 62304 ☑ ☑ ☑
☑ = all apply, ☒ = some apply, ☐ = none apply

The clause has an immediate overlap with ISO 13485 Clause 7.3.9 on
control of design and development changes. In addition, some foundational
requirements related to configuration management are made by the stan-
dard’s Clauses 7.5.8 on identification and 7.5.9 on traceability. These require-
ments will be discussed next.

8.1 Identification of Configuration Items


Looking back, Michael J. Fox may be the only choice for the lead in
the Back to the Future trilogy, but he was not the first actor hired to
play the role. For a time, Sylvester Stallone, too, was set to walk in
Eddie Murphy’s star-sprinkled tennis shoes in Beverly Hills Cop –
hence the Stallone poster behind Rosewood’s door in the film. These
are not the only replacement players destined to walk the unlikely
paths they ended up taking. These examples, however, show that
casting matters, and in software development, this is known as the
management of the software-system configuration.

Identification of configuration items has, in effect, two relevant points of


view. IEC 62304 primarily looks at the version-controlled items that go into
your product, how these are identified, and what processes may be used
to enact any changes in them. The standard even goes as far as to define
the term “version” as an identified instance of a configuration item (Clause
3.34), thus tying a knot between the two concepts even if the term version
has broad use and is understandable based on even a dictionary definition.
A nonbinding note pulls the knot tighter by asking for the software configu-
ration process to be activated when modifying a version of medical-device
software.
208 ◾ Medical-Grade Software Development

ISO 13485 also appreciates the changes going into your product, in much
the same way as IEC 62304, but it perhaps steps back a little to address the
traceability of all your records and activities and the finished products you
ship. Both standards do want you to stay on top of both changes in the
software and the changed software going out. The former entails the use of
change requests and the latter some agreement with you, your customers,
your notified body, and your other regulatory authorities on what level of
changes require what types of actions.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 7.3.9 Control of design and development changes


◾ ISO 13485 Clause 7.5.8 Identification
◾ ISO 13485 Clause 7.5.9 Traceability
◾ IEC 62304 Clause 5.1.9 Software configuration management planning
◾ IEC 62304 Clause 8 Software configuration management process

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

8.1.1 Expectations from ISO 13485


As discussed above, the ISO 13485 standard does not speak of configura-
tions specifically although it does expect all D&D changes to be identified
(Clause 7.3.9). The general requirements it makes on maintaining the neces-
sary identification (Clause 7.5.8) and traceability (Clause 7.5.9) information are
defined on a high level in the standard. The requirements in these clauses
cut through all your operations, and thus apply equally to the documenta-
tion of your activities and to any medical devices placed on the market,
including any changes made to them. These themes are discussed in detail
in our ISO 13485 book (Juuso 2022).

8.1.2 Expectations from IEC 62304


IEC 62304 devotes the whole of its Clause 8 to the topic of configuration
management, and it also addresses the requirements for software configu-
ration-management planning earlier in Clause 5.1.9 (see Section 5.1). Clause
8.1.1 builds on this platform and requires you to establish a unique iden-
tification scheme for configuration items and their versions. This scheme
Configuration Management ◾ 209

must cover other software products, SOUP components, and documenta-


tion, and it is to match with the development and configuration planning
required by Clause 5.1. This requirement applies to all software safety
classes.
In terms of SOUP components (incl. standard libraries), you must docu-
ment at least the title and the manufacturer of the component (Clause 8.1.2a–
b). For uniquely identifying the component you may, for example, use the
software version, release date, patch number, or upgrade designation.
In terms of documentation, you are required to identify and document
the configuration items (incl. their versions) that comprise the software
system configuration (Clause 8.1.3). Also note that Clause 5.1.11 requires that
for class B and C software, all configuration items must be placed under
configuration control before they are verified. Finally, remember that Clause
6.2.5 (see Section 6.2) also requires you to communicate with your users and
regulatory authorities on problems and modifications.

8.1.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 8.2 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
The fundamental expectation by both standards is that you have a pro-
cess for configuration management. To make that process more practical, we
would suggest considering the following regarding how to implement ver-
sion numbering and what items to manage via configuration management.

Table 8.2 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
A process for identifying product and product ✓ ✓ ✓ ✓
components
Identify and document the configuration items (incl. - ✓ ✓ ✓
versions) that comprise the software-system
configuration
Place software under configuration control prior to - - ✓ ✓
verification
210 ◾ Medical-Grade Software Development

8.1.3.1 Version Numbering
The first question to solve in designing your overall approach to configura-
tions is how you intend to identify the versions of the individual software
modules that go into your software system, and how you designate the ver-
sions of the final software being released.
For the finished product, we would advise the use of a four-level ver-
sion numbering scheme that will allow you to address the magnitude of the
changes you are contemplating in terms easily understood by regulations.
This model might, for example, use a four-digit scheme (e.g., A.B.C.D.) to
observe the following triggers.

◾ A-level change
Any significant change to the intended use or indications of your
product. This will be a point of serious introspection where you con-
sider, among other relevant aspects, the validation, risks, and benefits
associated with your software.
◾ B-level change
A significant change to the risk management of your product. This
might, for example, involve new risk-control measures (mitigations) or
changes made to the software that affect the risks and benefits associ-
ated with it.
◾ C-level change
A feature update to the software within the confines of the present
intended use, indications, and risk management.
◾ D-level change
A security patch or a bug fix to the software.

The above scheme will speak directly to the regulatory requirements for
your product and will help you in placing the needed emphasis on the
major changes (A and B) while attending to more minor changes (C and D)
more efficiently. You may be able to use this granularity in conversing with
your regulatory body, e.g., on what level of changes need to be reported
ahead of time, and what changes may require new conformity assessments
(e.g., for IEC 62304 conformance).
Alternatively, you might, for example, employ a three-digit scheme based
on the widely popular semantic versioning scheme that is coupled with
changes on the level of major updates (A), minor updates (B), and bug fixes
(C). You might even use one scheme on the level of your finished product
Configuration Management ◾ 211

and another on the level of the software items comprising that product. This
way, a change on a lower level does not as directly translate to a change
of the same magnitude on the level of the final product. You must evaluate
the impact of the change in terms of the final device, but a major change
in some low-level component does not always indicate a major change on
the level of the integrated whole. Occasionally it may also be that a minor
change on a lower level leads to a major change on the top level depending
on the use of that component. For this reason, it may make sense to decou-
ple versioning on the product level and the component level. Also remember
that for any SOUP components you are, of course, more or less stuck with
the versioning scheme their manufacturer has chosen.
Neither of the above schemes, the three-digit or the four-digit scheme, is
required by either standard. You must have a way of identifying versions,
but how you do so is up to you. You may set up your versioning scheme as
you see fit. Remember also that you may later want to develop new versions
of the sold software both as descendants of any current version, but also as
forked versions to meet any number of emergent needs (e.g., bug fixes, fea-
ture tweaks, cheaper/more premium versions of the same software). Ideally,
your configuration management should support all these goals. Finally, if
possible, you should also ensure that you are able to differentiate between
changes that have not gone outside your door yet and those which have
been supplied to customers. This may be a register matter, but it may also
involve addon monikers such as “RC” for a release candidate.

8.1.3.2 What Items Should Be Tracked?


IEC 62304 Clause 8.1.1 requires you to establish a scheme for identifying
and controlling configuration items (incl. their versions), but it stops short of
specifying a closed list of what these items are. The objective here is not only
to know what the constituent parts of the medical-device software are, but
also to control any changes in them. In theory, Clause 8.1.1 leaves the door
open for the manufacturer to decide what is and what is not a configuration
item but arguing that something doesn’t need to be controlled may be an
uphill battle. In practice, you will want to control everything, but the level of
control you impose on any given type of item will depend on its risk profile.
From IEC 62304, it is quite clear that the software itself and any software
items it consists of, that is relevant objects of source code, should be iden-
tified on some meaningful and practical level. This also applies to entities
such as SOUP components and documentation.
212 ◾ Medical-Grade Software Development

In terms of documentation, both standards want to see control exercised


over the documents and records you have. Changes to the documentation
may be triggered by a change in the corresponding software or by some
identified need for a change in the documentation itself. Any changes to
the code must, of course, be weighed for their impact and any appropriate
action, including updating the corresponding documentation, must be taken.
If, on the other hand, there is a separate need to only change the accompa-
nying documentation, e.g., by fixing a small typo in it or by improving upon
the language somehow this should be managed via an appropriate review
and approval mechanism and versions maintained. You will want to record
the change using your configuration-management process, but it would be
difficult to argue why some documentation has evolved in a completely dif-
ferent way than it was originally created. Also, remember that the baseline
for revising documentation is given by your document and record control
procedures as instructed in your QMS (see Juuso 2022). In the case the
documentation in question is part of the product, IEC 62304 wants to see
changes recorded via your configuration-management processes. A note to
Clause 5.1.8 also points out that the whole of Clause 8 is relevant when con-
sidering configuration management of documentation.
Beyond code and documentation, the line becomes more subjective. The
IEEE 1042 guide to software configuration management, for example, offers
up a much more detailed list than what is part of either of the standards.
Here the entities to be considered cover, for example, the following:

◾ Documentation, including management plans, specifications (e.g.,


requirements and design), user documentation, test designs (incl. speci-
fied test data), and maintenance documentation
◾ Code, including source code, executable code (i.e., binaries), libraries
◾ Support software
◾ Data dictionaries, various cross-references, and databases (both
included and processed data)

Much of the above list is addressed by ISO 13485 quality management and
IEC 62304 software development (e.g., plans, documentation, code, libraries,
and support software), but notably data dictionaries, cross-references, and
databases may not be an instantly apparent part of the same picture. Test
data, too, may fall under this latter group, but we feel that test data is man-
aged by testing activities and their reviews, and thus is not a part that ships
with the configuration in any active form, although it too most certainly
Configuration Management ◾ 213

needs to be controlled. In the following, we address this group of three enti-


ties together as databases.

8.1.3.3 Configuration Management over Databases


One particularly complex class of components to consider here is the data-
bases used by your software. Depending on the nature of the databases,
they may have as much impact on your device as the code itself. The data-
bases may also evolve over time, and thus change the functioning of your
device over time. The impact of data and databases may be even more
pronounced if the device somehow uses machine learning or artificial intel-
ligence methods based on the data.
Databases may be a part of the medical device, or they may be placed
outside the medical device and used via some interface. The placement of
the database in terms of what is in and outside of the device is a critical
design choice here. Think of the architectural design of your software, but
also think about the intended use of your device and what the role of the
database is in achieving that use when deciding whether the database is in
or out. Of course, something placed outside of the device is not something
that could be regarded as out of sight, out of mind. Based on the use of the
database and its impact on the risk assessment of the device you will want
to address it differently.
Databases contained within the product should be controlled as configu-
ration items in a way commensurate with their risk profile: a database of log
entries may, for example, carry smaller risks than some databases used for
machine learning or deep learning activities and affect the performance of
the device directly. In the former case, the database may perhaps be consid-
ered on the level of its schema and allowed to be populated via the normal
use of the device. In the latter case a much more involved re-evaluation,
perhaps even including an updated clinical evaluation, may be called for. In
between the use of databases may be akin to the use of SOUP components.
Databases used via interfaces and not part of the device itself should
be acknowledged in the description of the device (perhaps even in the
intended use), and all care should be taken to assess and monitor their use,
but they are not a part of the configuration management of the device itself
in the same way as a constituent software component. In this case, the use
of the database may be part of the requirements analysis, where the require-
ments for using the interface (incl. identity, data formats, protocols, and qual-
ity of service parameters) are addressed carefully.
214 ◾ Medical-Grade Software Development

In figuring out whether a database, or indeed some other item, should


be brought into the domain of configuration management ask yourself the
following.

◾ Can a change in the functioning of the database affect the conformity


of the device with its requirements (incl. applicable regulations)?
◾ Is the functionality enabled by the database considered to be medical in
nature? In the EU, see the MDCG 2019-11 guidance, for example.
◾ Is the database related to achieving the intended use of the device?
◾ Is the database included inside the device in the architectural design?
◾ Is the database version controlled so that a version 1.0 and, for example,
a version 2.0 can be uniquely identified at a given reference point?
◾ Is the database used to modify the functionality of the device? For
example, via machine learning and artificial intelligence methods.

If your answer to any of the above is yes, you will more than likely be look-
ing at a database that should be brought into configuration management.
If your database is implicated in meeting the intended use of the device, is
architecturally located within the device, and can be uniquely identified in
terms of its version, you will want to address it via configuration manage-
ment. The last condition will no doubt sound sloppy, but it is how IEC 62304
defines a configuration item.
Regardless of whether the database is a part of your device, many
issues may potentially be caused by unexpected changes in a database,
and these may need to be reflected in your requirements analysis and risk
management. Examples here include a change in the format of a data-
base field (e.g., integer versus float), a change in the number or order of
database columns (e.g., items referred to by ID versus items referred to by
sequence), a change in the expected protocol of an interface, or a silent
change in how a database field is used. Particularly the last example may
be a troublesome source of errors because of a silent drift in how people
enter field values and some ambiguity exists in what exactly is expected.
The same may also be true for a database fed by web-connected sensors
or other IoT-devices: a new brand or model of a device may have different
characteristics from those used before and may thus provide measurements
that are not exactly in line with the measurements obtained previously.
Again, the above issues may be particularly problematic if the retrieved
data is used to guide the operation of the software, for example, via
machine learning methods.
Configuration Management ◾ 215

8.2 Change Control (Incl. Verification)


Hollywood types call their agents and lawyers when a change in
casting is needed, but for software, change control starts with a
softer concept: a change request. Here the objective is to ensure that
only evaluated changes are enacted and that the necessary records
are kept.
The previous clause went into detail on how to set up the GPS system for
identifying the parts, and the whole, of your software. If your software was
built once and never updated in any way that coordinate system would not
be needed for much. As your software does change, that system will, how-
ever, get its chance to affect the marquee signs. That is why the second part
of Clause 8 goes into detail on how you manage changes in your identified
configuration items.
The relevant clauses of both standards are:

◾ ISO 13485 Clause. 7.2.2 Review of requirements related to product


◾ ISO 13485 Clause 7.3.9 Control of design and development changes
◾ ISO 13485 Clause 7.3.10 Design and development files
◾ IEC 62304 Clause 8.2 Change control

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

8.2.1 Expectations from ISO 13485


The fundamental expectation from Clause 7.3.9 is that you have a docu-
mented process for controlling D&D changes. The trigger for the activity
here may be a minute change to the inner workings of a software module
but the point of study is placed squarely on how that affects the overall
product. Here you should apply your understanding of the regulatory land-
scape around your product and the expectations of your regulatory authori-
ties to arrive at a flowchart of what types of changes call for what types of
activities.
The significance of the proposed change to the function, performance,
usability, safety, and regulatory requirements applicable to the medical
device and its intended use are the prime concerns here. The changes
must be reviewed, verified (see the section below on how IEC 62304
216 ◾ Medical-Grade Software Development

understands verification in this context), validated (as appropriate), and


approved before implementation. The review must also include an evalu-
ation of effects on constituent parts, products in process, or already
delivered, and the inputs or outputs of risk management and product
realization. Records of the changes and their review (incl. any necessary
actions) must be kept.
Also, note that Clause 7.2.2 requires you to confirm customer require-
ments before acceptance if no documented requirements are provided by
the customer, and you must also ensure that when product requirements are
changed the relevant personnel are made aware. Remember that changing
requirements may also have a trickle-down effect throughout your software-
development activities, such as the revision of your risk management file.
Finally, Clause 7.3.10 expects that a D&D file is kept for each medical-
device type (or medical-device family) and that this includes or references
records generated to demonstrate conformity to D&D requirements and any
D&D changes. This short clause thus expects D&D changes to be on record.

8.2.2 Expectations from IEC 62304


8.2.2.1 Approval of Change Requests
Changes to the identified configuration items (see Section 8.1) may only
take place in response to approved change requests (Clause 8.2.1). For this
reason, it is impractical to drag out the approval of a change request form
in the hopes of recording absolutely everything from the implementation of
that change with the same form. This is occasionally attempted by organiza-
tions, but it flies in the face of the expectations of the standard. Don’t do it.
Note that the approval process may be defined as part of your change-
control process or elsewhere in your QMS processes. Different approval
processes may also be used, for example, depending on the stage of the
development (see Clauses 5.1.1d and 6.1e in Sections 5.1 and 6.1, respec-
tively). All that is fundamentally required here is that no changes are imple-
mented prior to approval and that you have some process for performing the
appropriate approval.

8.2.2.2 Implementation of Changes
The implementation of the approved change is to take place as prescribed in
the change request (Clause 8.2.2). You are required to identify and perform
Configuration Management ◾ 217

any activities that need to be repeated due to the change, including any
changes to the software safety classifications on any level (from software
systems to software units, all of which are software items). A note to the
clause emphasizes that the actual implementation process does not need
to be covered by the change-control process but can instead refer to your
other processes (see Clause 5.1.1e and 6.1e). There is no need to reinvent the
wheel here, and little point in maintaining multiple parallel implementation
processes.

8.2.2.3 Verification of Changes
Here the standard uses the term verification quite loosely: what is required
is that the changes are verified, not that some major verification activity is
necessarily performed. You are required to verify the change, and if some
earlier verification activity is rendered invalid by the change, redo that veri-
fication while taking Clauses 5.7.3 and 9.7 into account (Clause 8.2.3). What
is meant by verification here is thus not the same as your overall verifica-
tion activities unless the change made calls those activities into question
somehow.

8.2.2.4 Traceability over Changes


Finally, Clause 8.2.4 requires you to maintain traceable records of the rela-
tionships and dependencies between change requests (incl. any relevant
problem reports) and their approval. These are also to be available somehow
for later accounting as discussed in Section 8.3.

8.2.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 8.3 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
The fundamental expectation by both standards is that you have a pro-
cess for change control. IEC 62304 requires that changes to the software
product are only made based on documented specifications in the form
of change requests. ISO 13485 on the other hand requires you to have a
process for controlling all design and development changes. The result is
218 ◾ Medical-Grade Software Development

Table 8.3 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
A process for controlling D&D changes ✓ ✓ ✓ ✓
Control changes impacting the function, performance, ✓ - - -
usability, safety, and the regulatory requirements
applicable to the medical device and its intended use
Control changes over other software products, SOUP - ✓ ✓ ✓
components, and documentation

that you need to have a process using change requests in place, and the
place to define that process is in your SOPs (e.g., in SOP-11 as per Juuso
2022). The process should then ensure that any changes are assessed
before being enacted, verified after enacted and that appropriate records
are maintained.

8.3 History of Controlled Items


Sylvester Stallone does not list Beverly Hills Cop in his filmogra-
phy, but if the film was a medical device, he might have to live with
the fact that somewhere in its documentation he would be listed as
one of the actors that went through that particular version of police
academy. Similarly, with medical software, you are required to keep
track of its pedigree.
Here ISO 13485 is yawning and insisting that we already covered this
ground when setting up the whole of your quality management system. IEC
62304, on the other hand, thinks you may suffer from dementia and ham-
mers home the need to maintain traceable records.
The relevant clauses of both standards are:

◾ ISO 13485 cl. 7.3.9 Control of design and development changes


◾ IEC 62304 cl. 8.3 Configuration status accounting

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.
Configuration Management ◾ 219

8.3.1 Expectations from ISO 13485


As previously covered in Section 8.2, ISO 13485 expects records to be
kept of the changes and their review (incl. any necessary actions). In addi-
tion, the baseline for controlling documentation is given by your docu-
ment and record control procedures as instructed in your QMS (see Juuso
2022).

8.3.2 Expectations from IEC 62304


Clause 8.3 requires that you retain retrievable records of the history of con-
trolled configuration items (incl. system configuration). This requirement
applies to all software safety classes. Clause 8.3 is called Configuration Status
Accounting, which almost implies a need to audit the status of the con-
figuration occasionally, but what is actually meant here is that the records
enabling such an audit are maintained. No status review is required here,
only that you retain the appropriate records.
Expect to retain all the records you specified above in Sections 8.1 and
8.2, but note that no list is in fact given here. Clause 8.2.4 discussed above
did, however, name the change request, problem report, and approval of the
change request as records to maintain and link together.

8.3.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 8.4 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
Here the impact on your processes and records is nothing new: you
already addressed it all above.

Table 8.4 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Maintain traceability of the records ✓ ✓ ✓ ✓
220 ◾ Medical-Grade Software Development

8.4 The Parts Left Out by IEC 62304


Could it be that there’s really nothing more to add here? The above dis-
cussion already covered everything expected by the two standards in an
exhaustive manner. Perhaps the only comment to add here is that configu-
ration management is not an afterthought to meeting the standards, and
it is not something you should only see to once the cracks caused by rust
and icebergs alike start to affect the maneuvering of your ship. Keeping
track of those patches and dented parts of the hull is required all along. It
is, however, a fairly easy task to get done correctly: it is something which if
done correctly allows you to track your deliverables and the work done on
them efficiently, securely, and predictably. If your system works for you and
answers all of the needs discussed above, then you should be well-set.
Change control is something then built on top of configuration manage-
ment, to facilitate the change undertaken within that constellation. The first
objective here is naturally, as discussed above, to ensure that a seaworthy
ship stays seaworthy throughout any later changes you make. The change
documentation will then be viewed in any audits of the device to assess
how it has evolved and what types of pressures have been inflicted upon
it. The flip side of this is that your processes for change control are the very
mechanism that makes it possible to quickly evolve your device in response
to any identified issues, demands, and appealing new opportunities. Setting
up a lean, efficient, and safe process for making changes will be to your
advantage from both a business and a regulatory point of view. It will also
generate a significant part of the information you will exchange with your
regulatory authorities over the coming years, and as such is already the
object of much attention from the point of view of your regulators. Making
your change control processes unnecessarily burdensome at your organiza-
tion will prevent you from carrying out some changes and will mean you
misapply your attention during making the changes you do make – it may
also mean that your regulatory authorities drown under all of the noise
generated by the manufacturers under their control. Ensuring change control
achieves the objectives set out above in a lean, efficient, and effective man-
ner is in fact what both sides of the table want to enable safe, ever better
medical devices. You can thus expect more guidance, standards, and regula-
tion to come out in this space in the coming years.
Chapter 9

PROBLEM RESOLUTION

Houston, we have a problem. Apollo 13 is perhaps the best example


in popular culture about encountering problems, analyzing the
situation, coming up with a workable solution, and saving all the
human lives at risk. The Titanic hitting that iceberg may be another
example, but one with less success in any problem-resolution pro-
cess. Problem resolution is about what you do when more than hot
air hits that fan. It’s also that last failsafe lifejacket your auditors
look to when assessing your otherwise perfect operations – or the
preplanned startup sequence that ensures your spaceship can power
back up for Earth re-entry.
Problems will occur. What you then do makes the difference between
brushing the issues under the carpet and keeping them from becoming epi-
demics. It’s easy to think that a perfect system would make zero mistakes,
especially as you are starting out and especially if you are looking to focus
on the code itself and do as little of everything else around it as possible.
Not documenting issues is an issue. Not fixing issues is an issue. Not fixing
issues quickly, though, is less of an issue than not fixing issues correctly.
In recent years there have been some high-profile cases where regulatory
bodies have been moved to jolt even large manufacturers into faster action
as their recalls have slowed to a crawl despite having vast resources to deal
with the issues. With big resources usually comes big business and large
user bases, and as a result, even small problems may become behemoths in
this geographic space. For this reason, if you are a small manufacturer your
resources may be better for dealing with a problem than if you are a large
manufacturer, even if that seems counterintuitive at first.

DOI: 10.4324/9781003454830-9 221


222 ◾ Medical-Grade Software Development

But what is a problem? Neither standard defines the term, so the standard
dictionary definitions apply here. IEC 62304 does define the term problem
report as a record of actual or potential behavior of medical-device software
that a user or some other interested party believes to be unsafe, inappropriate
for the intended use, or contrary to the specification (Clause 3.13). Note that
the definition of software here includes programs, procedures, and associated
documentation and data. Thus, a problem may be seen as a mismatch between
what you got as a product and what was expected based on the intended use
and specification. The standard then makes nonbinding notes to explain that a
problem can concern a released product or a product still in development and
that in the former case, Clause 6 makes additional requirements on the han-
dling of the problem regarding regulatory actions (see Section 6.2).
Your problem reports, then, should not start with “Once upon a time…”
unless you are writing a fairytale that you don’t expect anyone to actu-
ally believe. Your problem-resolution process should not be about finding
a patsy to blame, nor should it always end with a statement that this was a
one-time occurrence that was a statistical anomaly in your otherwise per-
fect software. Your problem report should definitely not blame the user as a
go-to answer, and it should not say that you never thought of the software
as finished in the first place. Others have tried all these excuses, but as
excuses go, they are pitiful. The first step to fixing a problem is admitting
you have one, right?
This section will take you through the eight steps the IEC 62304 standard
sees as needed to resolve problems appropriately. The section also contrasts
these with the expectations of the ISO 13485 standard, as in earlier sections,
and provides a synthesis meeting the requirements from both. The top-
ics covered in this clause are given in Table 9.1 below. The table also notes
which software safety classes each topic is applicable to, although in this
instance all the requirements apply to all classes of software.
The requirements from IEC 62304 are thus set in a convenient chronologi-
cal order, whereas the related requirements from ISO 13485 are to be found
in a few different clauses of the standard. Let’s look at the steps involved in
problem resolution one at a time.

9.1 Prepare Problem Reports


After admitting you may have a problem, the first step is to record
that problem so that it can be analyzed, corrected, and all of the
Problem Resolution ◾ 223

Table 9.1 Topics of Problem Resolution


# Topic A B C
9.1 Prepare problem reports ☑ ☑ ☑
9.2 Investigate the problem ☑ ☑ ☑
9.3 Advise relevant parties ☑ ☑ ☑
9.4 Use change-control process ☑ ☑ ☑
9.5 Maintain records ☑ ☑ ☑
9.6 Analyze problems for trends ☑ ☑ ☑
9.7 Verify software problem resolution ☑ ☑ ☑
9.8 Test documentation contents ☑ ☑ ☑
9.9 The parts left out by IEC 62304 ☑ ☑ ☑
☑ = all apply, ☒ = some apply, ☐ = none apply

related parties can hear about the happy conclusion. This is where
you play Columbo and attempt to record all the facts after walking
into the room with the chalk outline on the floor.
Clause 9.1 in IEC 62304 takes a cautiously pragmatic approach to problem
resolution and instructs how the problem is to be recorded. First and fore-
most, it wants to see that any problems are in fact recorded by the manufac-
turer, but it does list some information it wants to see included in the report.
ISO 13485 chimes in to the extent that we may be talking about noncon-
forming products or processes.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 8.5 Improvement (incl. CAPA)


◾ ISO 13485 Clause 8.3 Control of nonconforming product
◾ ISO 13485 Clause 8.2.2. Complaint handling
◾ IEC 62304 Clause 9.1 Prepare problem reports

Also note that Clause 7.3.10 on D&D files instructs records on conformity
to D&D requirements and D&D changes to be maintained, which is some-
thing you may be acting on here either by investigating problems or making
changes. No specific requirements are made here on the kind of information
you will record. You could also find other related sections in ISO 13485, for
example, Clause 8.2.2. on complaint handling may describe a similar process
224 ◾ Medical-Grade Software Development

of recording a report, investigating it, and proceeding to fix any issues. If


comparing and contrasting with these processes somehow simplifies your
different approaches here go ahead but know that this is not required. Not
all customer feedback is on problems regarding your software. Not every
software problem is identified through customer feedback. We would sug-
gest that not every complaint represents a software problem needing to be
fixed, but those that do may then lead to even multiple CAPA actions. For
this reason, these clauses are listed here but covered only briefly below. You
will find additional discussion on the clauses in our previous book (Juuso
2022). In the following, we will look at the central expectations of both
standards, one standard at a time, and then synthesize a plan to meet the
requirements of both.

9.1.1 Expectations from ISO 13485


In general, ISO 13485 wants to see any encountered problems weeded out,
whether these are detected in the products or the processes of the QMS. It
does not acknowledge a problem report per se, but it does speak profusely
about nonconformities (Clause 8.5) and nonconforming products (Clause
8.3). It also goes to great effort to point out that such problems may be
detected through several seemingly separate activities, such as customer
feedback and complaints (e.g., Clauses 7.2.3c and 8.2.2), lowly servicing
information (Clause 7.5.4a), design and development work, and the long-
term post-market surveillance activities. The area of most immediate com-
mon ground with IEC 62304 in terms of problem resolution is in the control
of nonconforming products (Clause 8.3) detected before or after delivery,
but equally a complaint report (Clause 8.2.2) could serve as the beginning
of a problem report.
The focus in Clause 8.3 is on faulty products that should first be seques-
tered and separated from sellable products and then investigated to find the
cause of the fault and an appropriate remedy if there is one. The clause does
not directly translate to a problem report as discussed in IEC 62304, but it is
nonetheless worthwhile to know what the standard expects to see regarding
problem handling when it involves faulty products. ISO 13485 Clause 8.3.1
further requires that you have a process for nonconforming products (incl.
controls, responsibilities, and authorities for their identification, documenta-
tion, segregation, evaluation, and disposition). This may factor into the con-
sideration of what information you want to record in your problem report
or other subsequent handling documentation. See SOP-2 in our earlier book
Problem Resolution ◾ 225

(Juuso 2022) for more on handling nonconforming products, or SOP-7 on


feedback and complaints.
If the problem represents a nonconformity to be addressed, you should
invoke your corrective actions and preventive actions process (CAPA; see
SOP-2 in Juuso 2022). If on the other hand, the problem does not meet your
definition of a nonconformity, e.g., the problem represents an issue in as-of-
yet unreleased software, a leaner process may suffice.

9.1.2 Expectations from IEC 62304


Clause 9.1 requires the manufacturer to prepare a problem report for
each problem detected in medical-device software. This must include the
following:

◾ Statement of criticality (e.g., effect on performance, safety, or security)


◾ Other information that may aid in the resolution (e.g., devices and
accessories affected)

Note that prior to the 2015 amendment, the clause also called for the type of
the problem report (e.g., corrective, preventive, or adaptive to new environ-
ment) and the scope of the problem (e.g., size of change, number of device
models/accessories affected, resources involved, or time to change) to be
recorded. Now these properties may be thought of as included in the cat-
egory of other information as needed.
A note to the clause points out that problems can be discovered before
or after release, and both inside and outside the manufacturer’s organization.
This corresponds nicely with the discussion of before and after delivery by
ISO 13485, although release and delivery are distinctly different events.

9.1.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.2 provides a glance over the main requirements of both
standards. In this case, the software safety classification used by IEC
62304 does not really change matters, and you might be hard-pressed
to argue that the clauses of ISO 13485 should be excluded (see Juuso
2022).
The requirements here are quite brief, and once again IEC 62304 would
be happy with the correct outcome even without a previously defined SOP.
226 ◾ Medical-Grade Software Development

Table 9.2 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
A process for nonconforming products ✓ - - -
A problem report - ✓ ✓ ✓
A statement of criticality - ✓ ✓ ✓
Other information that may aid in the resolution - ✓ ✓ ✓

Since you will have related SOPs from ISO 13485 you will do your best to
ensure their instruction meets both sets of requirements. Any instruction in
the SOP should address the problem report as discussed above.

9.2 Investigate the Problem


This is where you put your notepad back in the pocket of your
crumbled-up trench coat, absentmindedly thank your first inter-
viewees, and go off to massage those grey brain cells to figure out
what has happened. Unlike in daytime detective stories, the objective
here is not to outsmart the murderer but instead figure out what in
your product (or processes) should be fixed and how to avoid any
harm coming to your users.

After you have a problem report, it’s time to investigate it. If the report is
gibberish, as in clearly the result of a misunderstanding, or made in error
you should be able to close the case here before proceeding further. If it is a
valid report of a problem, you will be expected to look to your processes for
risk management, CAPA, nonconforming product, and act appropriately.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 8.5 Improvement (incl. CAPA)


◾ ISO 13485 Clause 8.3 Control of nonconforming product
◾ IEC 62304 Clause 9.2 Investigate the problem

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.
Problem Resolution ◾ 227

9.2.1 Expectations from ISO 13485


As with the previous clause, here too the most directly relevant requirements
for investigation of the problem are defined by ISO 13485 Clause 8.3.1,
which we also used in the previous section. The requirements added here
are as follows:

◾ You must determine the need for an investigation and notification of


any external parties responsible for the nonconformity
◾ You must record the nature of the nonconformity and any actions taken
(incl. investigation and rationale for decisions)

Note that these requirements are very similar to what is said about com-
plaint handling in Clause 8.2.2. Your processing of complaints is required to
be timely in accordance with applicable regulations, any resulting CAPAs are
to be recorded, and should you decide not to investigate a complaint you
must record a reason. The clause also gives you a minimum set of items to
record for the complaint: a) receiving/recording information, b) evaluation of
the feedback as a complaint, c) investigating complaints, d) determining the
need for reporting to regulatory authorities, e) handling of related product,
and f) determining the need for CAPA. If the problem arises out of a com-
plaint and represents a nonconformity to be addressed, you should follow
your communication processes and invoke your corrective actions and pre-
ventive actions process (see SOP-7 and SOP-2 in Juuso 2022).
All of the above may factor into the consideration of what information
you want to record in your problem report or other subsequent handling
documentation. You do not need to include all this information in a first-line
problem report, but you may want to record the information at an appropri-
ate place and time during your problem handling if a released product is
involved.

9.2.2 Expectations from IEC 62304


Clause 9.2 compels the manufacturer to a) investigate the problem and, if
possible, identify the causes, b) evaluate the relevance to safety using their
risk-management process, c) document the outcome of the investigation and
evaluation, and d) create change requests for actions needed to correct the
problem. In case no action is taken in d) the rationale for doing so must be
documented. A nonbinding note to the clause further explains that if the
228 ◾ Medical-Grade Software Development

problem is not relevant to safety, a correction is not required for comply-


ing with Clause 9. Similarly, a nonbinding note to the definition of the term
problem report in Clause 3.13 (see the beginning of Section 9 above) clari-
fies that a problem report may be rejected if it represents a misunderstand-
ing, error, or insignificant event.

9.2.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.3 provides a glance over the main requirements of both
standards. In this case, the software safety classification used by IEC 62304
does not really change matters, and you might be hard-pressed to argue that
the clauses of ISO 13485 should be excluded (see Juuso 2022).
Your process for investigating problems should most likely lean on your
already-defined CAPA process (see Juuso 2022) so as to not create multiple,
deceptively similar processes for similar purposes. The general approach
to CAPA with its emphasis on identifying the root cause should also work
here in the context of software. A quick glance over Table 9.3 confirms that
all safety classes of software are in fact treated the same here, and thus you
may now ensure that your general ISO 13485 CAPA process also suits your
needs in addressing software-related problems. A problem suspected in
high-risk software is going to be more worrisome than a similar problem
suspected in low-risk software, but in the eyes of the requirements stated
here by IEC 62304 there is no difference.

Table 9.3 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
A statement on the need for an investigation ✓ ✓ ✓ ✓
A statement on the notification of external parties ✓ - - -
Identify the causes (if possible) - ✓ ✓ ✓
Evaluate relevance to safety using risk management - ✓ ✓ ✓
Document the investigation/evaluation outcome ✓ ✓ ✓ ✓
Record of the nature of the nonconformity and actions ✓ - - -
taken
Create change requests for actions needed - ✓ ✓ ✓
Problem Resolution ◾ 229

Note that ISO 13485 seemingly allows for deliberation on the need for
an investigation, but IEC 62304 requires the manufacturer to investigate the
problem. The difference here appears to be semantics as at least a record
and analysis of the problem is in any case required by both standards.
Putting its fight face on, IEC 62304 now requires you to use your risk-
management process to investigate the problem’s relevance to safety. At a
minimum, we might interpret this to mean investigating the problem in light
of your risk management file for the device and reacting to any action and
revision needs from therein, but in the face of a starker problem, you might
even trigger a new ISO 14971 risk-assessment project as per your SOP. You
may, of course, find in the course of resolving the problem that you need to
also revise your risks or their mitigations somehow, which in turn might call
for communication with external parties, as covered next.

9.3 Advise Relevant Parties


All good detective stories end with all the interested parties couped
up at some final gathering where the master detective carefully
unwinds the story as he has put it together, and finally takes a vic-
tory lap around the room after naming the person who did it. The
standards don’t care about egos and dramatic storytelling, but they
do want you to make sure you don’t forget to tell people about the
problem and its solution.
This is a short clause pointing out a fairly obvious fact: you must advise the
relevant parties. Whether these are parties responsible for causing a product
nonconformity issue to arise, such as your suppliers, the users of your prod-
uct, or the regulatory authorities overseeing it all, this is where you make
sure they all know that a problem exists and that you have either solved it
or are solving it.
The clause may be short and self-evident to a conscientious manufac-
turer, but the implementation of it may leave you breathless and exasperated
already at the start line. The communication may, for example, call for user
manuals to be updated (also perhaps involving a new round of usability
studies), the material translated to an array of languages, and you may also
need to organize user training for your customers. This all depends on the
circumstances, of course, but it may be that walking the walk here is more
like a marathon than just the “one more thing” Columbo – and Steve Jobs
also – is famous for. Conversely, in some cases, the advice required of you
230 ◾ Medical-Grade Software Development

may simply be placed in the release notes of your next software release for
the same effect.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 8.3 Control of nonconforming product


◾ ISO 13485 Clause 8.2.2 Complaint handling
◾ ISO 13485 Clause 8.2.3 Reporting to regulatory authorities
◾ IEC 62304 Clause 9.3 Advise relevant parties

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

9.3.1 Expectations from ISO 13485


Clause 8.3.1 expects the manufacturer’s evaluation of the nonconformity to
include a determination of the need for notification of any external party
responsible for the nonconformity. The standard addresses nonconformities
discovered before delivery and after delivery separately. No instruction on
advising relevant parties is made in the case of the former, but in the case
of the latter Clause 8.2.3 requires the manufacturer to have processes in
place for issuing advisory notices in accordance with applicable regulations.
Similarly, in the case of a complaint, Clause 8.2.2 points out that if activi-
ties outside the organization contributed to the complaint somehow relevant
information will be exchanged with them. The definition of an external
party ought to be understood quite broadly here.

9.3.2 Expectations from IEC 62304


Clause 9.3 succinctly points out that the manufacturer must, as appropri-
ate, advise relevant parties of the existence of the problem. A non-binding
note clarifies that the manufacturer will identify the relevant parties and that
problems may be discovered before or after release, and inside or outside
the manufacturer’s organization.

9.3.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.4 provides a glance over the main requirements of both
Problem Resolution ◾ 231

Table 9.4 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Determine need for notification of any external party ✓ ✓ ✓ ✓
Process for issuing advisory notices ✓

standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
In practice, you will want to ensure your process instructs relevant third
parties to be notified. For issuing advisory notices and informing regulatory
authorities you may decide to refer to your SOP for communications (SOP-7
in Juuso 2022) instead of duplicating instructions here.

9.4 Use Change-Control Processes


Don’t reinvent the wheel for implementing changes. Instead use the
processes you have already defined to sort out any complaints, con-
formities, nonconforming product, and product changes the process
for problem investigation brings up.
After you have investigated the problem arising out of a complaint, a
detected nonconformity, or some discovered nonconforming product you
will know what you want to fix and how. The place to do that is here and
the process to follow in making the changes is something you get to pull
out from under the desk as something you made earlier. All television chefs
and how-to Youtubers will smile on approvingly.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 7.3.9 Control of design and development changes


◾ ISO 13485 Clause 8.2.2 Complaint handling
◾ ISO 13485 Clause 8.3 Control of nonconforming product (incl. rework)
◾ ISO 13485 Clause 8.5 Improvement (incl. CAPA)
◾ IEC 62304 Clause 9.4 Use change control process

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.
232 ◾ Medical-Grade Software Development

9.4.1 Expectations from ISO 13485


ISO 13485 provides distinct requirements for handling the sea monsters you
dredged up while investigating that initial problem report. For complaints,
you will have an instructed complaint-handling process (see SOP-7 in Juuso
2022). For nonconformities you will get to deploy your CAPA process (see
SOP-2 in Juuso 2022), and for nonconforming products, you may have a spe-
cial process that may also address the rework of already produced stock (see
SOP-2 in Juuso 2022). For making changes to your software product, you
will have a process stemming from product realization (see SOP-10 in Juuso
2022). You may invoke all or some of these processes based on exactly what
the problem report warranted as actions.

9.4.2 Expectations from IEC 62304


Clause 9.4 requires the manufacturer to approve and implement all change
requests while observing the requirements of the change-control process
defined in Section 8. This somewhat awkwardly worded requirement does
not, of course, mean that all change requests have to be approved, but it
does mean that all approved change orders have to be implemented and not
left in some limbo. It also means that problems are not fixed via some magic
du jour, but via the very process you defined for changes in Section 8 – note
though that any requirements for nonconformity handling and reporting to
authorities from your QMS must also be observed.

9.4.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.5 provides a glance over the main requirements of both

Table 9.5 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Handle complaints, nonconformities, nonconforming ✓ - - -
product (incl. rework), and D&D changes as instructed
Approve change orders ✓ ✓ ✓
Implement approved change orders ✓ ✓ ✓
Problem Resolution ◾ 233

standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
In practice, you will want to address how a complaint, a nonconformity,
a nonconforming product, or a D&D change may represent a problem, and
how you want the resolution of that problem to take place. The area of most
interest in the case of software is most likely going to be in making D&D
changes which we addressed in detail in Section 8.2.
As a final note here, don’t think that bug fixes get a dedicated path fly-
ing somehow under the radar here. Klingons may have cloaking devices,
Frodo may have had an invisibility cloak, and the great invisible man
may be see-through under those layers of gauze, but you don’t have such
tricks up your sleeve here. We have often heard it argued that bugs are
treated separately from everything else, that working on bugs can’t pos-
sibly affect the conformity of the device negatively, and that everyone
wants bugs fixed yesterday already. This may occasionally all be true, but
the standards are adamant that you use your change-management process
here. Bugs do not get priority access through the pre-flight security check,
even if both you and your regulators are eager for you to process them
quickly. Your change-management process may, of course, scale to the
type of change as is appropriate based on how fundamental or superficial
it is and how closely connected it is to the intended use of any medical
device.

9.5 Maintain Records
Show me the money, yelled Cuba Gooding Jr. in Jerry Maguire, and
Tom Cruise did his best to repeat after him in an effort to ensure
him of the promised results. In medical devices evidence always
matters, and without that black-and-white it’s hard to trust that
something happened the way you later explain. Therefore, it’s hardly
surprising that you are required to generate and maintain that
evidence.

Clause 9.5 is an exercise in pointing out the obvious, but the reason it takes
a Jack Nicklaus approach to driving home its message on retaining evidence
is that evidence does matter and may be accidentally forgotten in the heat of
the moment.
234 ◾ Medical-Grade Software Development

The relevant clauses of both standards are:

◾ ISO 13485 Clause 4.2.5 Control of records


◾ ISO 13485 Clause 7.3.10 D&D files
◾ IEC 62304 Clause 9.5 Maintain records

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

9.5.1 Expectations from ISO 13485


As noted above, a note that Clause 7.3.10 on D&D files instructs records
on conformity to D&D requirements and D&D changes to be maintained,
which is something you may be acting on here either by investigating prob-
lems or making changes. You should also make a note to use your process
for control of records (see SOP-1 in Juuso 2022).

9.5.2 Expectations from IEC 62304


Clause 9.5 requires the manufacturer to maintain records of problem reports
and their resolution (incl. verification). The risk management file must also
be updated as appropriate.

9.5.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.6 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and

Table 9.6 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Maintain records on conformity ✓ - - -
Maintain records on changes ✓ ✓ ✓ ✓
Use record control ✓ - - -
Problem Resolution ◾ 235

remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
In practice, you will want to think about what records you will need from
problem resolution, and ensure your SOP instructs their use in accordance
with your mechanisms for control of records (see SOP-1 in Juuso 2022).

9.6 Analyze Problems for Trends


You don’t have to be John Nash and see patterns in the movements
of pigeons, but you must keep an eye on the encountered problems
to identify points of further improvement for your product and
processes.
It’s not enough to fix something as you must also learn from your mistakes
to hopefully prevent them from occurring again. You can employ anything
from providence to artificial intelligence to analyze the data for trends, but
make sure you can defend your plan to yourself, and that you have a plan.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 8 Improvement


◾ IEC 62304 Clause 9.6 Analyze problems for trends

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

9.6.1 Expectations from ISO 13485


Clause 8.1 requires the organization to plan and implement monitoring, mea-
surement, analysis, and improvement processes needed to a) demonstrate
conformity of device, b) ensure conformity of the QMS, and c) maintain the
effectiveness of the QMS. This is explained to include the use of appropriate
methods such as statistical techniques.
It is technically a matter of definition whether problems are part of this
monitoring and analysis, as no such direct requirement is made. In practice,
claiming that problems, and the events implicated in problems, are not a
matter of conformity for the product and the QMS is akin to declaring crimi-
nal insanity. Thus, the requirement in Clause 8.1 is wider than IEC 62304
236 ◾ Medical-Grade Software Development

problem resolution, but it should very much apply to problem resolution.


Notice that the clause also requires you to strive for improvements based on
the captured information and trends, not just nonconformities themselves.
The time for planning and implementing the above is before the begin-
ning of time, and the place and time for analyzing the trends may, for
example, be in the context of an internal audit or a management review. In
any case, you usually do not need to analyze trends after each problem is
resolved unless there’s special cause to do so. For more discussion on audits
and management reviews please refer to SOP-6 in our previous book (Juuso
2022).

9.6.2 Expectations from IEC 62304


Clause 9.6 requires the manufacturer to perform analysis to detect trends in
problem reports. The timing of this trend analysis, or what the manufacturer
should then do, is not commented on.

9.6.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.7 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
In practice, you should not be content with having fixed some problem
presented to you. You should also look for patterns and reoccurring prob-
lems and improve your processes and operations accordingly. The intent is
the same in both standards, but ISO 13485 is more vocal on the how and
why as seen above. The classification and labeling of problems, the imple-
mentation of the database holding the data, and the available statistical tools
as well as the long-time scales sometimes required for problems to reoccur

Table 9.7 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Monitor, measure, and analyze trends in problems ✓ ✓ ✓ ✓
Improve based on detected trends and information ✓ - - -
Problem Resolution ◾ 237

will be factors in developing an effective process here. The choices to make


are not so much governed by the standards or the regulations as by your
knowledge of the subject matter, past trends, and your command of the vari-
ous tools involved. There is, for example, no standard or regulation that says
guest rooms must have their doors fully intact, but if you stayed as a guest
of Jack Nicholson at the Overlook Hotel and started to notice doors along
your corridor manifesting signs of axe blows one by one you might decide
to act.

9.7 Verify Software Problem Resolution


Good planning may be half the battle, but that also means plan-
ning will only get you halfway there. Here the standards want you
to ensure that whatever you saw as the remedy to the problem has
been implemented, has worked, and hasn’t led to new unhandled
problems. Think of Bruce Willis staying back on that asteroid to
ensure its detonation in “Armageddon”, but don’t feel verification
always has to be that fatalistic.

Verification of problem resolution for software products is quite a compact


clause in IEC 62304 but it has much longer, deeper roots in ISO 13485 even
if the term problem is not shared across the two domains. Don’t go over-
board here trying to realign the whole of ISO 13485 to the single topic of
problem resolution but consider how the key expectations of that standard
map to what IEC 62304 wants to see. The following will take a fairly broad
look at related ISO 13485 clauses, but keep your ship sighted at the expecta-
tions of IEC 62304 here.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 4.1.3


◾ ISO 13485 Clause 5.5.1 Management responsibility
◾ ISO 13485 Clause 8.5.2e Corrective action
◾ ISO 13485 Clause 8.5.3d Preventive action
◾ IEC 62304 Clause 9.7 Verify software problem resolution

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.
238 ◾ Medical-Grade Software Development

9.7.1 Expectations from ISO 13485


ISO 13485 does not talk of problems per se, but it does expect you to han-
dle complaints (see SOP-7 in Juuso 2022), nonconformities, nonconforming
product, and possible rework (see SOP-2 in Juuso 2022), as well as D&D
changes (see SOP-10 in Juuso 2022) as instructed in your QMS. Invoke the
correct verification activities here by referring to the instructions in the SOP
in question.
ISO 13485, in general, wants to see all of your operations achieve a high
level of existence where your entire quality management system runs along
like clockwork. Any issues that creep up should be handled efficiently and
with safety in mind, but instead of just instructing reactions to events the
standard wants to see processes and resources in place in order to ensure
a continuously well-oiled machine. On the top level, thus, verification of a
problem-free existence for your processes and products begins with Clause
5.5.1 which instructs the top management of your organization to document
and empower the personnel who verify work affecting quality. Verifying
problem resolution certainly qualifies as such work. Similarly, Clause 7.5.4
discussing verification of continually met product requirements in the con-
text of servicing activities may be seen as a related clause to ensuring the
continuance of that jovial existence. Clause 4.1.3, too, may be a part of
the equation by calling for monitoring, measuring, analyzing, and improv-
ing QMS processes in a way that ensures both their operation and control.
Clause 8.2.6 also chimes in here by requiring monitoring and measurement
of product, thereby hoping to detect any issues early on. The most reactive
clauses of the set here, Clauses 8.5.2e and 8.5.3d, then call for a process to
verify that the corrective and preventive actions you may have ended up
implementing as a response to a detected issue don’t adversely affect your
conformity. All of these process links may affect your instruction for soft-
ware realization here.
In case you had a problem report that then spawned nonconformities to
address it will be up to you to define when that problem report should be
closed: when all the child issues have been closed, or when they have been
left to run their acceptable, prescribed process. The quality of your overall
operations may not be best served by keeping all items open for extended
spans of time, but you will want to ensure that the actions you prescribed dur-
ing the investigation are carried out on the correct time scale. Even so, some
actions may be affected immediately and others necessarily play out over lon-
ger timescales. The definition of done is something to consider here too.
Problem Resolution ◾ 239

9.7.2 Expectations from IEC 62304


Clause 9.7 requires the manufacturer to verify the resolutions made to deter-
mine whether a) the problem has been resolved (incl. the closing of the
problem report), b) adverse trends have been reversed, c) change requests
have been implemented in the appropriate medical device software and
activities, and d) if additional problems have been introduced.
A particularly thorny requirement here is brought up by Clause 9.7b
which requires adverse trends to have been reversed. Requiring this as part
of the resolution of each individual problem, instead of as a part of some
long-term process and product-monitoring activity is potentially a massive
roadblock to a smooth problem-resolution process running predictably from
start to finish. In some cases, the true assessment of the reversal may require
months, or even years, worth of monitoring data to be available to capture
comparable time-series data. This would lead to problem reports staying
open for unjustifiably long periods of time, and each report then hogging
some sliver of resources indefinitely. In some other instances, it might be
possible to implement a fix, run a test sample, and see if the fix has worked
or not by inspecting the devices coming off the production line. For soft-
ware this is probably not going to be the case.
It is easy to see why trend analysis, or the identification of persistent
problems, is an important tool. In the 1986 disaster involving the space
shuttle Challenger a reoccurring fault with the O-ring seals was blamed
for the loss of the spacecraft and the seven lives onboard. The issues with
the seals were in fact detected long before the accident, but instead of fix-
ing the problems it appears that ineffective, perhaps cheaper workarounds
were devised (incl. limiting the weather conditions for launch, relying on
the existence of a second seal). The problem was recorded in a NASA data-
base, which in fact exasperated the issue further as it became an acknowl-
edged, accepted point of failure that sometimes raised its head but hadn’t
yet stopped any flights. Had the trend been taken seriously the disaster may
never have happened. Based on the story, it is easy to advocate that all
problem reports should be kept open forever and repeatedly studied after
the fact to detect any lingering concerns. With infinite resources, this would
work, too. This might, however, cause the organization’s resources to be
spread thin over time, and not enough attention be given to the problems
that truly were continuing concerns. Instead, it might be better to include a
review of problem databases for similar issues during the initial assessment
of a problem report. This way no reports are kept open unnecessarily and
240 ◾ Medical-Grade Software Development

all of the resources can be focused on meeting the present needs as well
as possible. In the Challenger example, the adverse trend could have been
reversed by noticing just once that the problem had actually occurred previ-
ously instead of just trying to circumvent the problem at hand. This would
not have needed an extra, after-the-fact round of analysis, just a realization
of what information was available when a problem was reported.
Our suggestion here with Clause 9.7 is to assess what the reversal of
trends translates to in the case of the problem and decide whether the prob-
lem report can be closed with or without some long-term follow-up moni-
toring. Ideally, the long-term monitoring should utilize ISO 13485 processes
for monitoring and improvement, and not leave unhelpful parallel threads
hanging here. Your circumstances may, however, vary and complacency can
come just as easily by having a catch-all monitoring process running behind
everything as with a pre-existing problem report in a database.

9.7.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.8 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
In practice, all of the requirements made by IEC 62304 are in line with
the expectations of ISO 13485, but the same sentiments are addressed under

Table 9.8 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Observe your processes for complaints, nonconformities, ✓ - - -
nonconforming product (incl. rework), and D&D changes
Verify that the problem has been resolved and the ✓ ✓ ✓
problem report closed
Verify that adverse trends have been reversed ✓ ✓ ✓
Verify that change requests have been implemented in ✓ ✓ ✓
the appropriate medical-device software and activities
Verify that additional problems have not been introduced ✓ ✓ ✓
(or are handled appropriately)
Problem Resolution ◾ 241

the discussion of, for example, complaints and nonconformities, instead


of an entity called a problem. You will want to ensure that the above IEC
62304 requirements are met by your processes when they concern a prob-
lem with your medical-device software product or its manufacturing. This
also includes programs, procedures, and associated documentation and data,
as pointed out by IEC 62304.

9.8 Test Documentation Contents


When the stuff has somehow hit the fan, and you are eager to get
the fix out, it may be tempting to forget about testing the fix in a
documented manner. Yet, the standards don’t want to see you forget
about retaining the necessary information.
Testing as part of the software-development processes was covered in
Clause 5 (see Section 5.7, for example). Here the focus is on testing as it
relates to detected problems, their verified resolutions, and the documenta-
tion to retain. By now you will have noticed that IEC 62304 is quite big on
testing and test records, which incidentally form the core of the documenta-
tion needed also for legacy software. It is therefore not surprising that test
documentation gets its own short section under problem resolution.
The relevant clauses of both standards are:

◾ ISO 13485 Clause 4.2.4 Control of documents


◾ ISO 13485 Clause 7.1 Planning of product realization
◾ ISO 13485 Clause 7.5.8 Identification
◾ ISO 13485 Clause 8.2.6 Monitoring and measuring of product
◾ IEC 62304 Clause 9.8 Test documentation contents

In the following, we will look at the expectations of both standards, one


standard at a time, and then synthesize a plan to meet the requirements of
both.

9.8.1 Expectations from ISO 13485


Clause 7.1 requires you to plan testing in planning product-realization. Here,
in this section, we are talking about reacting to detected problems and test-
ing after a fix has been applied instead of planning a new product-realiza-
tion process, but nonetheless your SOPs should instruct problem resolution,
242 ◾ Medical-Grade Software Development

and, for example, your software development plan or maintenance plan may
then expand on the arrangements and resources involved. Thus, Clause 7.1
may indirectly apply here.
The impact of Clause 7.5.8 is more direct as it requires the identifica-
tion of product status to be maintained throughout product realization to
ensure that only product having passed the required inspections and tests,
or released via an authorized concession, is put out. Clause 8.2.6 points out
that if test equipment is involved in the monitoring and measurement of
products the records must identify the equipment as appropriate, and, for
implantable devices, also the testing personnel must be identified (Clause
8.2.6).
Finally, note that Clause 4.2.4 requires documents to which medi-
cal devices have been tested to be available for at least the lifetime of the
device, but not less than the retention period for any resulting record, or as
dictated by applicable regulatory requirements. You will have defined the
retention period and device lifetime in your QMS (see Juuso 2022), but here
make sure that the documentation you produce meets this requirement. The
documentation you produce here should be classified as records, while the
instructions for creating those records will be in your documents (e.g., qual-
ity manual and SOPs).

9.8.2 Expectations from IEC 62304


The final clause of the standard, Clause 9.8, addresses what your test docu-
mentation for changes must include. The testing itself is explained to take
the form of testing, retesting, or regression of software items and systems
following a change. The instructed set of information here is as follows:
a) test results, b) anomalies found, c) version information of the software
tested, d) hardware and software configurations, e) test tools, f) testing date,
and g) identity of the tester. Items d) and e) are qualified with the call for
relevant information to be entered, and thus the answers to these fields may
be brief or verbose depending on your circumstances.

9.8.3 Suggested Synthesis
The requirements of the two standards are complementary through and
through. Table 9.9 provides a glance over the main requirements of both
standards. Note the software safety classification used by IEC 62304 and
Problem Resolution ◾ 243

Table 9.9 Synthesis of Top-Level Requirements


IEC 62304
Combined requirements ISO 13485 A B C
Ensure appropriate control of documents and records ✓ - - -
Record test results ✓ ✓ ✓ ✓
Record anomalies found - ✓ ✓ ✓
Record version information of the software tested - ✓ ✓ ✓
Record relevant hardware and software configurations - ✓ ✓ ✓
Record relevant test tools ✓ ✓ ✓ ✓
Record testing date - ✓ ✓ ✓
Record identity of the tester ✓ ✓ ✓ ✓

remember that some clauses of ISO 13485 may be excluded with a valid
rationale (see Juuso 2022).
IEC 62304 is more specific in its requirements here but in practice, all of
the expectations are the same between the standards. ISO 13485 expects
evidence to be on record for proving any activity and it places great impor-
tance on the immutability and auditability of that information. Thus dates,
identities of testers, and details of the testing environment become very
important for repeatability and traceability reasons. In practice, IEC 62304
only adds a few software-specific requirements.

9.9 The Parts Left Out by IEC 62304


In general, problem resolution for software products is a topic for IEC 62304
and is not covered as such in ISO 13485. However, considering the many
incarnations a problem can have in the QMS, the product-realization pro-
cesses defined within it, and the product itself, problem resolution does
have much longer, deeper roots in ISO 13485. A problem detected in the
IEC 62304 sense may have implications that, for example, take the form of
a complaint, nonconforming products, or a nonconformity in the processes.
All of these must be addressed in the appropriate ways meeting the expecta-
tions of both standards.
244 ◾ Medical-Grade Software Development

Don’t go overboard here trying to realign the whole of ISO 13485 to the
single topic of software problem resolution but consider how the key expec-
tations of that standard map to what IEC 62304 wants to see. In the above,
we took a fairly broad look at the related ISO 13485 clauses, but here at the
end of Clause 9 we thankfully find that all the key issues and requirements
from both standards have now been addressed.
Chapter 10

INTEGRATION WITH
YOUR QMS

Phew! That was a massive amount of detail on your software prod-


uct-realization activities dished out from two perpendicular points
of view. Each standard makes perfect sense by itself, but to marry
the two together and end up with something that makes sense in a
lean, practical way takes real effort. This is where we will start to
put those notes together and shape that symphony you need.
The master plan set out in this book is to negotiate together the require-
ments of the two foundational standards, ISO 13485 on quality management
and IEC 62304 on software development, in a way that has you meeting all
of the requirements but not adding unnecessary hoops to jump through.
ISO 14971 risk management, too, is important, but in our task here it can
be treated as an opinionated aunt popping by every once in a while. In
practice, risk management cuts through everything, and offers an alternate,
perpendicular view to what we are constructing here. This means that we
can, and should, start with the two other standards here to build our net-
work of processes. As the annexes to IEC 62304 are quick to point out, you
may already have a QMS in place prior to adopting the software life-cycle
standard. In this case, you will naturally perform some sort of a gap analysis
(e.g., internal assessment, internal audit) to assess your starting point first,
and then act as appropriate to align your processes with the discussion in
this book.
The key underlying notion here is that the foundations of your operations
are described by ISO 13485 quality management. The QMS you have built

DOI: 10.4324/9781003454830-10 245


246 ◾ Medical-Grade Software Development

according to this standard will give you the basic toolbox to run your orga-
nization and ensure that the products you place on the market meet their
requirements. This includes, among other activities, making sure that the
review, verification, and validation activities you run meet the expectations
set by the standard. In terms of the reviews and verifications you conduct
during software development, the meat will be in IEC 62304, but the pack-
aging required by ISO 13485 will help ensure that important points of view
are observed. The same is true for planning and requirements analysis also,
and especially your validation activities which are not covered by IEC 62304
at all.
Thus, forging a cohesive software-development pipeline with minimal
repetition, back-and-forth, and unnecessary overhead comes down to negoti-
ating the following pieces of the puzzle.

◾ Standard operating procedures (SOP)


The standard operating procedures of your QMS that describe your
product realization process and the other supporting processes.
◾ Planning documentation
The planning documentation for your software-development project,
incl. the software development plan (SDP).
◾ Execution of the project
The execution of your development project according to your docu-
mented arrangements (incl. the creation of the set deliverables).
◾ Execution of the post-release activities
The execution of your post-release activities.
◾ Continuous improvement of the QMS
The continued improvement of your QMS.

If you succeed in addressing all of the above in your software-development


project, you will meet the requirements of both standards. Each of the above
pieces of the solution is discussed in a section of its own in the following.
Note that Annex D in IEC 62304 attempts to give a rough checklist for
the adoption of the standard by small manufacturers who do not yet have a
certified QMS in place. In our view, this checklist is not all that helpful as it
boils down to a table of contents for the two main standards, but it is worth
knowing about its existence. In any case, software development according to
IEC 62304 can not survive for long, if at all, without ISO 13485 quality man-
agement and ISO 14971 risk management. Annex D does not change this
basic assumption of the standards.
Integration with Your QMS ◾ 247

10.1 Write Your Product Realization Processes in the QMS


This book has walked you to the edge of the reservation and pointed out
the important markers you should observe on your trek forward. However,
there is no magical shortcut to now putting one foot in front of the other,
or pen to paper, and plotting your course. Let this book be your Yoda on
the road ahead, but don’t expect the finished QMS to appear as a vision by
itself.
You must now make sure your processes match the expectations of both
IEC 62304 and ISO 13485 – and any regulations applicable to you. The
discussion in this book, and that given in our previous book focusing on
ISO 13485 QMS by itself (Juuso 2022), will help you in writing the neces-
sary processes one by one. Take the instructions given on the QMS as your
foundation here. Then add all the detail expected by IEC 62304 under the
specific software development (see Section 5), maintenance (see Section 6),
configuration management (see Section 8), and problem-resolution processes
(see Section 9) under product realization (see SOP-10 in Juuso 2022). Don’t
forget about risk management either (see Section 7).
You should also appreciate the influence of the customs and habits your
organization may have adopted over past software-development work, and
the properties of your chosen software-development model. These will
weigh in here as you bend the requirements to your will – or, perhaps, the
other way around.
In writing the SOPs also consider the documentation you will want to
create as part of performing work according to those processes. This will
include many of the various plans, reports, reviews, and other types of
records your SOPs want to conjure into being. You will want to create tem-
plates for some of these records, and thereby perhaps reduce the amount of
prose needed in the SOPs to list all of the relevant properties and fields of
content that are required for each. As for documents, we recommend that
you stick to defining documents as your top-level QMS documents: your
quality manual, your SOPs, and perhaps your working instructions. See
SOP-1 in our previous book (Juuso 2022) for more discussion on documents
and records.
Perhaps the greatest trick to designing a modern software-development
process while meeting the requirements of the standards is how you marry
your agile development cycle with the waterfall-like expectations of the
standards – in other words, how you turn that classic waterfall into a mod-
ern washing machine with the spin cycle your organization expects to have
248 ◾ Medical-Grade Software Development

at its disposal. Table 10.1 below sketches out one possible solution. Here
the overall structure consisting of four phases matches with the waterfall
model of the standards. The typical development tasks of planning, require-
ments elicitation, design, implementation, testing, release, and maintenance

Table 10.1 Basic Software Development Life Cycle


# Phase Key Deliverables
1 Product-Realization Planning Software Development Plan (SDP)
System Requirements
Risk Management Plan
DIR Design Input Review Review Minutes
2 Sprint Planning Sprint Management Records
Repeated Requirements Software Requirements
incrementally analysis Clinical Evaluation Plan (CEP)
Risk Management File (RMF)
Design Software Architecture Design B C
Software Detailed Design B C
Test Cases (incl. plan and procedures)
Implementation Software code
Testing Software Verification Report
Software Validation Report
Software System Test Report
Sprint Review Sprint management records
DOR Design Output Review Review Minutes
3 Release Software Maintenance Plan
Instructions for Use (IFU) and Label
Documentation of Residual Anomalies B C
Risk Management File (RMF)
Clinical Evaluation Report (CER)
Post-Market Surveillance Plan
Draft Declaration of Conformity
DTR Design Transfer Review Review Minutes
Final Declaration of Conformity
4 Maintenance Post-Market Surveillance Report or
Periodic Safety Update Report (PSUR)
Post-Market Clinical Follow-up Report
(PMCF report)
Integration with Your QMS ◾ 249

are punctuated by the three major reviews discussed earlier in this book:
DIR, DOR, and DTR. Given a set of requirements at the top of the waterfall
you could let gravity take over and end up with a finished piece of software
downstream. The magic of turning that age-old into something more agile
takes place in the second phase of development as depicted in Table 10.1.
Table 10.1 sketches out a four-phase approach to meeting both industry-
standard expectations for agile software development and the requirements
of the medical-device standards and regulations at the same time. The
motivation for the model here has been to allow for the typical sprint-based
approach to development while ensuring that the expected analysis and
reviews take place at the appropriate steps. The simple idea behind the
model is that the key documentation listed in the right-side column is avail-
able for each subsequent review, and once all of these can be approved
you may move to the next phase. Note that in phase 2, the four types of
activities – requirements analysis, design, implementation, and testing – are
performed iteratively. You may thus first run entire sprints to flesh out the
software requirements and design, and only after successive sprints finally
have an implementation to test. It is thus not required to perform activities
of all four types in every sprint and it is not required to finish all activities
of one type to proceed onto another type of activity. In practice, you may
choose a set of requirements for each sprint, use that sprint to work on their
realization, and then tie the full set of requirements (and their realization)
together in the design output review (DOR). Developers are used to tickets
and working on a few items of the whole at a time, but the standards want
to see you have a firm grip also on the whole that those items make, as dis-
cussed in Sections 5 through 9. Designing this model here is where you get
to overlay the expectation of the regulated medical-device environment over
the freedom software developers crave, and hopefully do so in a way that
doesn’t have the developers or the auditors fuming out of their ears.
Following the model introduced here you won’t be attempting to do
everything at the same time, you won’t be jumping too far ahead of yourself,
or waiting unnecessarily on some documentation that is actually needed
much later in the process. This is, of course, a rough overview of the overall
development process and you may choose to structure your development
life cycle slightly differently in your circumstances, but overall we feel this
generalization makes much sense.
In reality you will be updating a number of documents repeatedly along
the way, but exactly how often you do so will depend on how good your
initial documents are, and how you identify needs to update them.
250 ◾ Medical-Grade Software Development

10.2 Develop Your Planning Documentation


According to Your QMS
After you have instructed your processes in agreement with the standards,
ensure that they are also in agreement with your everyday practice. In other
words, it is not enough to write a beautiful, leather-bound work of fiction
for your QMS if you don’t then live by its tenets. Your SOPs should be clear
enough and comprehensive enough so that you – and everyone else at your
organization – can now follow them, perform their work as instructed, and
create the records requested.
Sections 5 through 9 of this book will introduce you to the required
records, and what is expected of each. The first record to consider here is
the software development plan (or system development plan) discussed in
Section 5.1. Also remember that your overall QMS will instruct other records
to consider here as well, including records on software tool validation, sup-
plier evaluation, and perhaps personnel qualification. You may not need
all such records at the start of preparing your planning documentation, but
these may soon become relevant as you execute that plan. Dodoism, or the
act of burying your head in the sand and hoping for the best, has no place
in any of these activities.

10.3 Execute on Your Plans as Expected


by Your Arrangements
The only thing left is to do the thing itself? Not quite, as in the world of
medical devices you are always required to also document the thing you
did. Nonetheless here is where you take off those training wheels – or the
unnecessary bandage around your injured arm if you are Chuck Norris –
and jump into action. Section 11 next should give you a good overview of
what records are expected of that action.
At the end of this phase, you will have developed your piece of software,
scrutinized it adequately, and even released the final product into the world
beyond your doors. You will have the finished product, all of its accompany-
ing documentation and supporting services, and you will have documenta-
tion covering its whole birth available at will. This documentation may be
requested by your notified body, regulatory authorities, or even your B2B
customer.
Integration with Your QMS ◾ 251

10.4 Execute on Your Post Release Activities


This is where you are Janine Melnitz hooking up the telephone line at the
brand-new Ghostbusters headquarters to open up for business, and hope
that no ghouls come over the telephone wires. In the past, it might have
been enough to just wait and listen, but today you are also required to
actively check up on how that software you created is fairing in the real
world. Your QMS will have defined a whole post-market surveillance pro-
cess for this (see Juuso 2022), and here we have added detailed processes
for addressing maintenance, changes, and any encountered problems. Once
again what remains is to follow through on those processes.

10.5 Keep Your QMS in Good Shape


Your company is up and running, and your product is changing the world
for the better. You might be thinking: why rock the boat? Maybe you
shouldn’t, but you have to be on the lookout for ways to improve your QMS.
ISO 13485 expects to see you embrace continuous improvement for all
your operations. Change for the sake of change is not the aim here, but
instead charting your position and considering ways to improve the sta-
tus quo always is. This topic is addressed in our previous book (Juuso
2022), and it is not specific to software but rather quality management in
the context of all medical devices. Here the strive for continuous improve-
ment is brought up purely as a kind of heads up to not sit on your laurels
after software release. In practice, a lot of the improvement activities around
your QMS will be led by functions other than software development at your
organization, but some of the best ideas for improvement may originate from
within the very function that first gave birth to the software. The notion of
the factory floor, or going right to the source, is an essential concept in qual-
ity management – even if the father of modern quality management, W. E.
Deming, famously attributed 95% of all quality issues to management. If you
are in management take that to heart, but if you are not, forget we said it.
Chapter 11

TECHNICAL
DOCUMENTATION

Did you skip ahead to here? We wouldn’t blame you if you did, but
after taking a dip here remember to go back and read all the detail
in the previous sections too. Developing technical documentation in
a lean and practical way that matches seamlessly with your chosen
life cycle model is a persistent engineering topic in the development
of medical devices today. It is perhaps not as flaming-hot as leverag-
ing new emerging technologies such as artificial intelligence, but
it may have some equally profound implications for the safety and
efficiency of your devices. The Holy Grail is a documentation model
that adds no overhead and meets with no friction while inching
its cuddly tentacles everywhere and at all times. That may not be
achievable, but you will get close enough if you fashion a develop-
ment model that neglects neither code nor documentation.
A stereotypical way to approach the medical-software business is to first
think about the code you want to write, the documentation you have to cre-
ate for it, the quality management system you have to have in place, and the
cool green cash you hope to make from it all. On this path, you may think
you have coding covered and the rest is just Styrofoam in the box. Not so.
As remarked on in Section 1.13, documentation is intended to be an
accurate reflection of the whole that is your medical-grade software. The
documentation must show what requirements the software is based on,
what design choices these led to, and how those were then implemented,
verified, validated, approved, and released. The risk management file must

252 DOI: 10.4324/9781003454830-11


Technical Documentation ◾ 253

then reflect this thread from beginning to end. The goal is that instead of
trying or retesting your software over an extensive period of time, the rel-
evant stakeholder, such as your auditor, can look into the wheelhouse by
referring to your documentation where they will see how it is all going. The
basic tenet of IEC 62304 is that mere testing is not enough to ensure safe
medical software. Instead, you must impose control over the entire develop-
ment and life cycle of that software. The technical documentation should
thus shadow your software as closely as possible, and if it falls behind or
out of touch with the code itself somehow, at any point, you may be in
trouble. If, however, both the code and the documentation match nicely,
and meet the applicable requirements, your house should be in good order.
The discussion in the earlier sections of this book has identified the key
types of documentation you will want to have whether or not you call the
documents and records by the exact same names or not. This section will
go over the expectations from both standards that are specific to your body
of technical documentation, and then slot in the various pieces of documen-
tation identified earlier to build a comprehensive and practical set of key
documentation. This set will then help you answer the needs your stake-
holders have both now and over time as your software evolves. Sections 5
through 9 of this book have already introduced the requirements for the
individual documents and records in detail, so here we will concentrate on
the overall architecture of the documentation.

11.1 What Is Technical Documentation Anyway?


Let’s first start by briefly considering the terminology involved in the defini-
tion of technical documentation. The terminology is simple, but some con-
fusion appears to exist among manufacturers, and partly this is caused by a
subtle difference in EU and US terminology. In general, the terms used are
as follows.

◾ Medical Device File (DMF)


This is the documentation set expected by ISO 13485 Clause 4.2.3
(see Juuso 2022). This set of information is kept for each medical device
or medical-device family and is expected to contain all necessary infor-
mation for demonstrating conformity with the standard and any appli-
cable regulations. In practice, this includes the intended use/purpose,
general description, labeling (incl. IFU), and specifications for the prod-
uct and its realization.
254 ◾ Medical-Grade Software Development

◾ Design and Development Files


This is the documentation set expected by ISO 13485 Clause 7.3.10
(see Juuso 2022). This set of information is kept for each medical device
or medical-device family and is expected to contain records demon-
strating conformity to D&D requirements, as well as any D&D changes
made. The implication here is that you need to maintain documentation
demonstrating conformity at the start, but also to ensure any changes
are reflected regardless of whether these take place a little later or a
lot later.
◾ Device Master Record (DMR)
This is a US term roughly corresponding with DMF in ISO 13485.
This refers to the set of documentation kept for a specific type of
device. The Code of Federal Regulations (820.181) sets requirements for
the contents of the documentation which are very much in line with the
expectations of ISO 13485.
◾ Device History Record (DHR)
This is a US term for documentation kept during the manufactur-
ing of a device on the level of batches, lots, or units. The Code of
Federal Regulations (820.184) sets requirements for the contents of the
documentation along the lines of manufacturing dates and quantities,
released quantities, acceptance records (in relation to the DMR), and
traceability.
◾ Quality System Record
This is a US term for documentation expected by the QMS but
not necessarily specific to any medical device. The Code of Federal
Regulations (820.186) sets requirements for the contents of the docu-
mentation. Note that complaint files are specifically called out by US
CFR 820.198.
◾ Technical Documentation (TD)
This is the general term used in the EU for technical documenta-
tion demonstrating the conformity of a device (or device family) to the
medical-device regulation. EU 2017/745 Annex II provides a detailed list
of the type of documentation expected as well as the overall hierarchy
it is expected to be made available under.

In practice, the above requirements can all be answered in the case of medi-
cal software via documentation corresponding to the expectations of the IEC
62304 and ISO 13485 standards. ISO 14971 risk-management documentation
will have a part in this as well. In practice, you are expected to maintain
Technical Documentation ◾ 255

relevant product and production information in a manner facilitating good


traceability between D&D inputs and outputs, and over the generations,
batches, lots, and units of devices you manufacture.
The information expected for your technical documentation will depend
on the type of medical device you have, what regional regulations are appli-
cable to your device, and which standards beyond ISO 13485 are applicable
to your activities. The documentation will likely also include subsets such as
the risk management file, the usability engineering file, and clinical evidence
for your device. It is up to you as the manufacturer to marry all of these
requirements together into a cohesive, navigable body of documentation
meeting all of the various stakeholder needs.
Figure 11.1 shows an example of a high-level structure for technical docu-
ments. The exact nature of documentation in your set will depend on your
type of device (e.g., hardware, software, or both), but the concept of provid-
ing a starting index is universal. This index will then quote and reference

Figure 11.1 Example structure for technical documentation.


256 ◾ Medical-Grade Software Development

other documents in your set as is needed to answer questions regarding


your medical device or medical-device family.

11.2 Process Documents
Much of the information listed above refers to what ISO 13485 understands
as records, i.e., records retained of work having taken place, but occasion-
ally you may also need to produce the documents instructing that work.
These are the documents, as ISO 13485 speaks of them, which set up your
processes through your quality manual, standard operating procedures, and
perhaps your working instructions. These documents may be thought of as
a part of your technical documentation, or as something existing on a layer
above it, but nonetheless, you are required to retain such documents for
at least the life of your records and produce them for reference when your
records are being assessed.

11.3 Development Records
Given the existence of separate process instructions, the objective of plan-
ning documentation is to add any necessary further detail on the execution
of those processes. After planning, all that remains, in theory, is to execute
those plans, processes, and other documented arrangements to obtain the
deliverables you expect. From a business point of view, here you are run-
ning your processes and doing the things you expected to do during soft-
ware development. From a regulatory point of view, you are performing the
planned activities, developing the deliverables, and collecting the required
evidence on the quality of this all.
In Section 10 we discussed a four-phase life cycle model for software
development. Table 10.1 (given in Section 10) sketched out the different
phases of development work, married the work with an agile approach,
listed typical deliverables, and presented the three principal reviews (DIR,
DOR, DTR) used to ensure the outcomes meet expectations. The two fig-
ures below, Figure 11.2 and Figure 11.3 provide a simplified view of the
technical documentation of most concern in each of the three reviews. The
set of documentation sketched out by the two figures is what the standards
expect from you in a more-or-less chronological order, and it is also the set
Technical Documentation ◾ 257

Figure 11.2 Technical documentation in DIR and DOR.

Figure 11.3 Technical documentation in DTR.

that anyone inspecting your operations from the outside in will expect to
find.
Figure 11.2 illustrates what records are typically viewed during a design
input review (DIR) and a design output review (DOR). In practice, you
may find that seldom are all of the requirements known at the start of a
258 ◾ Medical-Grade Software Development

development project, and these thus need to be updated perhaps through-


out the project. Similarly, risk management will evolve as the project
matures. Nonetheless, the figure here provides an overview of the ideal run
of things.
The design documentation (incl. architectural design, and detailed design)
may be developed in sequence using a waterfall-based approach or some-
thing more akin to agile methodology. The documentation must, though,
make sense throughout, on both the level of the system as a whole and the
individual software items that it comprises. Similarly, the body of verification
activities may be built up of smaller, incremental activities that address both
the components and their sum adequately. System testing and validation
activities must ensure that the sum of it all meets with what both you and
your customer expect to get at the end of it all.
Note that in addition to verification, a validation plan and a report are
also expected sooner or later. Requirements for both are given by the IEC
82304-1 standard on health software (see Section 12.2). You may decide
to have the plan available for DOR, but what is actually required by the
standards is that the plan is established before validation readiness can be
achieved (IEC 82304-1 Clause 6.2a), and thus before validation is carried out.
Figure 11.3 picks up after DOR and takes us to the finish line where a
design transfer review (DTR) takes us to the market.
The most important of the three reviews is the design transfer review
(DTR). Here you must be satisfied with launching the thing you created into
the real world. You must make sure you are happy with all of your principal
records, have kept them up to date, and have moved ahead in a rational,
defensible manner. If there’s any work to redo this is the place to notice it
and act accordingly. That said, remember that the standards take a sacred
view of defining what is acceptable before you actually see what you got,
mostly so that you avoid the temptation of saying what you got was actu-
ally what you wanted all along. No clairvoyance is called for, but you must
know what is good enough before getting there, and the place for the final
check on whether what you got is good enough is here before moving to
production.
After you know that you are good to go, and have received any neces-
sary approvals or clearances from your notified body and regulatory authori-
ties, you may finally sign the declaration of conformity (DoC) and step into
the market. You will, of course, then be running your maintenance, change
management, and post-market activities as promised – and churning out the
records you want to obtain from that work.
Technical Documentation ◾ 259

11.4 The Audit Package


Given the above broad definitions of what exactly is technical documenta-
tion, it may seem that your auditors will want to see it all. They may, in fact,
want to see it all, but you will probably not photocopy it all and messenger
it over. Instead, you are typically asked to provide a top-level index of your
technical documentation and some specific documents answering specific
topics.
The EU 2017/745 Annex II referenced above gives one possible way to
structure the index file for your technical documentation. Given this model,
the information discussed above might, for example, take the form of the
following documentation list:

◾ Device description and specification (incl. intended use/purpose)


◾ Information to be supplied by the manufacturer (incl. IFU, labeling)
◾ Design and manufacturing information (incl. IEC 62304 development
documentation, and specifications/procedures for product realization)
◾ General safety and performance requirements (GSPR), as applicable
◾ Risk-management documentation (incl. benefit–risk analysis, significant
residual risks, and anomalies)
◾ Clinical evidence

Much of the information listed above refers to what ISO 13485 understands
as records, i.e., the outcomes of work, but occasionally you may also need
to produce the documents instructing that work. Your index might therefore
reference both the instructions you give for work and the individual main
records kept of that work. All of your records should be under your record
control as instructed by your QMS (see Juuso 2022), but you do not have to
list all of your records in your index, and you do not have to ship copies of
the entire body of records or provide your auditors with remote access to
all of your records. You will need to produce any documents and records
requested by your auditors, though. You should also expect good traceabil-
ity, and the ability to move between individual records and documents in
any fashion facilitating an investigation of your instructions and actions, to
be required.
Chapter 12

SEEING INTO THE FUTURE

Being able to predict the plot points in films is occasionally fun, but
usually just a drag chipping away at the magic of cinema. Real life
is seldom about the suspension of disbelief or going along for the
ride, so having some inkling of what is to come ahead of time just
makes sense and lowers your blood pressure. This section will dis-
cuss the future challenges and opportunities for developing medical-
grade software. Some of these topics may be immediately relevant to
your type of software, while others may shape the standardization
and regulatory environment around them.

Ever since the microprocessor first revolutionized what medical devices


could do, we have been on a path of constantly improving upon what is
possible with software. The arrival of network connections between devices
decades ago has now given way to new levels of connectivity with cloud-
based services and hub-centric thinking first connecting hospital wards to
a central control room and then entire hospitals of different sizes together.
Most recently, the application of artificial technology has heightened the
need for safe solutions to understanding and controlling advanced technolo-
gies that are starting to take over tasks previously only possible for humans
to perform. Humans are in no danger of being replaced as the experts in
the loop – or indeed the patients themselves – but we are now getting to
a point where a small part of the doctor could be thought of as embedded
into the medical devices, e.g., in the form of AI-based analytics.
All this activity means that the standards and regulations, too, are under
a constant need to stay current with the times. The present version of IEC
62304 is outdated, but even a new version published today would not

260 DOI: 10.4324/9781003454830-12


Seeing into the Future ◾ 261

automatically stay current for any extended period of time. For this reason,
all standards have a routine maintenance cycle: all the key medical-device
standards discussed in this book are revised every five years by the stan-
dardization committees responsible for them. The committees don’t just
meet every five years; instead many of them run constant monitoring activi-
ties and forward-looking studies into what challenges might be ahead, and
how to best respond to such challenges.
This section looks at a few of the most current topics relating to the
development and maintenance of medical software we see being discussed
across the world. The topics discussed here are perhaps only implicitly
addressed by the present generation of key medical-software standards, but
they may have a profound impact on the software products of tomorrow. A
few of the topics, usability engineering in particular, are ones that are intui-
tively easy to approach but notoriously hard to master. Other topics, such
as edge intelligence, are novel new ideas that may grease the wheels in the
right places to unlock stalemates or catalyze new solutions to long-standing
problems such as data transfer, cybersecurity, the protection of personal
information, and also intellectual property.

12.1 Agile Software Development


In the film Armageddon a group of Texan oil drillers turn every-
thing they have learned upside down to go blow up a gigantic aster-
oid heading for Earth. No one would call this bunch nimble, but as
an exercise in agility, the move from an offshore oil rig to a space
station in a matter of weeks is worthy of entry into Cirque du Soleil.
The same sort of dexterity, nimbleness, and speedy project execution
is what we are talking about when considering the adoption of agile
methods in medical-software development.
This entire book has been written in a way that is as compatible with agile
software development as possible, yet we felt that a separate discussion of
the methodology itself in the context of medical devices is called for. That is
what this section aims to achieve.
We have been following the standardization of agile software devel-
opment for regulated industries over the past several years, including as
members of select international standardization committees and mirror
groups. Much work has been done on the standardization of agile, scrum,
regops, and many other related domains, but as of yet no single standout
262 ◾ Medical-Grade Software Development

standard for an agile approach to medical-device development has


emerged. To add fuel to the fire, Dr. Shuren, the director of FDA’s CDRH
even remarked recently at an industry forum that agile regulations might
also be a thing. This may or may not have been a slip of the tongue, but
we do not yet live in a world where agile nimbleness has outsmarted the
age-old waterfall.
It’s a cliché to say that developers want to run free and regulators want
them to sit quietly in the corner churning out perfect code to everyone’s
benefit. The agile world of the incremental, occasionally reactionary devel-
opment work revolving around tickets and sprints clashes with the perfectly
predictable, tediously planned document-centricity of the regulatory context.
The answer is neither anarchy nor paralysis, but something in between. The
Holy Grail here is a frictionless QMS and a maglev-levitated development
pipeline that accelerates any idea to a piece of safe software in less time
than it takes for the subsequent money orders to go through in the opposite
direction. Negotiating the everyday requirements of software development
in a business setting with the sometimes lofty expectations of the standards
and regulations is, however, a task to reckon with. In a world without mis-
takes, misinterpretations, imperfection, and friction this might be an easy
task, but in this one, and in a domain as critical as medical devices, a good
solution here takes real effort. The good news is that this goal is routinely
attempted, and passable or even good solutions are achieved. Our hope
in this book has been to point out the important design considerations
involved and even provide some of the smart solutions you might employ.
Nonetheless, agility and the safe mastery of agile methods in the medical
setting is a mission greater in scope than any book or standard written on
the topic. It is a lifelong undertaking, but one hopefully revolving around the
sort of considerations developed in this book.

12.1.1 What Is Agile Anyway?


Wizardry in a vacuum, the lone code guru as a prophet of their profession,
is not such a hot idea when the objective is to consistently create code that
fulfills some real need the users have. Error-free code doesn’t just appear
by itself, and a black-box approach to writing software even when you use
intelligence of the natural kind – and don’t just rely on AI or ChatGPT to
write your code – is a risky proposition. If you like hanging around in your
underwear all day and constantly sliding from one task to the next like Tom
Cruise in that famous scene from his non-ISO-14971 business film, you may
Seeing into the Future ◾ 263

decide that lifestyle is for you. For the rest of us, there are standards and
software-development life cycles to choose from.
The first development model that early mankind mastered after learn-
ing how to spark code may have been the waterfall model. Even then
man was thinking about the kinetics of the maglev development pipeline
we just brought up: if you push a boat down a series of waterfalls and it
makes it through all the trials you may have something worth putting out.
If that waterfall is Niagara Falls, and the thing you built travels better than a
wooden barrel you may even have ended up saving some daredevil lives a
hundred years ago. In any case, gravity worked just like magnetic forces to
remove any friction from the development path.
Today the waterfall model is recognized as a simple concept to grasp,
even a practical framework to use when describing requirements for devel-
opment work in standards. The model provides handles on the wizardry
and controls over its outcome, which is why it has been successfully used
at least since the 1950s. The model has its origins in the manufacturing and
construction industries and is generally well-suited for the project manage-
ment of undertakings where each subsequent step relies on the deliverables
of the previous steps. Think of Tetris with a greatly expanded field of vision.
If, however, the falling blocks start to get flakier and their specifications may
evolve in some incremental manner as they fall down your screen you might
want to reach for something a little faster-acting than the waterfall model.
Think of Schrodinger’s cat and how much more fun Tetris would be if the
blocks took after his offspring.
The waterfall model was good for setting checkpoints on the work per-
formed and setting criteria for moving ahead to the next phase. This, no
doubt, opened enough eyes and provided enough checks so that errors
were greatly reduced from any past state of the art. Viewed from a modern
lens, though, the waterfall model has started to seem too unresponsive and
also out of touch with the everyday of software-development work taking
place. A model that is viewed as a drain on resources and one which no
longer provides the required timely insight into the work and control over it
might need refreshing.
To put the granny on wheels, a new agile software-development life
cycle was developed. Not having to plan each possible path forward in
the Marvel multiverse of infinite possibilities, and at the same time retain-
ing a better understanding of the path that we actually were on at the
moment, probably led to the coining of the principal agile models we have
today. Part of the solution here was to narrow down the intervals between
264 ◾ Medical-Grade Software Development

checkpoints so that we might check up on the progress more frequently


and take smaller steps toward the right end goal. In other words, by using
an agile approach we would be building up the capabilities of the software
constantly and with more certainty of heading in the right direction. Now
we wouldn’t just have to wait for the development to reach its next stage,
but we could better remain in touch with the work that was going on all
the time.
The advantage of agile software-development models (e.g., XP, scrum,
kanban) is the increased velocity of implementing new features and func-
tionality for software. This, in theory, decreases the total expenses involved
and improves the transparency of the throughput. In a regulated industry
such as medical devices, however, velocity alone does not rule the day.
Here, the availability of a new feature takes a second seat to the reliability of
that feature as the stakes may be life and death. It thus becomes extremely
important to observe the standards and applicable regulations, which in turn
call for truly understanding the users, the risks involved, and the benefits
sought. Incorporating these aspects of medical device development appro-
priately into the agile model may feel like it sucks out the benefits of the
model.
The inclusion of the various planning, review, verification, validation, and
risk-management activities expected by regulations may feel like extra work
during development. These are a natural part of the waterfall model, but as
we start sprinting in cycles it may start to feel repetitive and unnecessary to
see to these activities each time, all the time. Indeed, performing and reper-
forming all of this for every sprint will be unnecessary, but forgetting to
perform them at all or in the right place in time will be unforgivable. Often
times the trick here will be to allow for the incremental, piece-wise working
model of agile, but also remembering to adequately address the whole that
is built from those incremental steps. A case in point is given by the require-
ments definition: here you may be tempted to hone in on the final set of
requirements over time, but don’t forget that you must also ensure the set of
requirements makes sense as a whole, covers all of the bases it is required
to, and none of the requirements contradict one another. This sounds simple
enough, but as you are adding to and removing from your set of require-
ments over sprints this will start to feel less simple. The more such loose
ends you find yourself leaving behind the more work you will find in front
of you as the release nears. Such neglected activities may even become a
showstopper for the release. Conformity to regulatory requirements is not an
afterthought activity.
Seeing into the Future ◾ 265

12.1.2 Development Life Cycle in a Regulated Environment


Medical-device regulations do not force any specific development life cycle
on you. You can meet the requirements using, for example, a waterfall-based
model or an agile development model. The regulations do, however, insist
on the relevant evidence being available to prove you and your device are
in conformity with the requirements. In the case of the EU MDR this means,
first and foremost, compliance with Annex I in EU 2017/745. Similar evi-
dence-oriented requirements exist in other regulatory contexts. ISO 13485,
therefore, requires that the manufacturer understands these requirements in
their target markets and appreciates their role in appropriately defining the
D&D inputs for the development project.
The regulations also expect you to have a good understanding of
the medical domain your software will be used in. Manufacturers usu-
ally remember to involve a clinical expert familiar with the domain when
developing the Clinical Evaluation Report (CER), but particularly smaller
manufacturers may forget to involve the same expertise during the require-
ments definition phase. Occasionally, too, we get the question of can a
manufacturer use a clinician they are familiar with to work on a product
that is targeting a very different area of medicine. In practice, this may be
problematic, and not an ideal choice as the goal is not to have a doctor sign
off on your analysis, but rather to understand the use of your device as well
as possible. In any case, involving a clinician early on in the design will
in all likelihood save you from discovering some critical issues late in the
development.
Occasionally. also the expected velocity of development, as well as
the notion of fixing any shortcomings in a later sprint, may become an
issue for regulatory compliance. It may, for example, happen that the
development of the next impending maintenance update always con-
sumes all of the manufacturer’s available resources and does not leave
much room for keeping up with the regulatory needs around the current
version.
In adapting your agile development life cycle to appropriately address
the regulated context you may want to consider the following questions.

◾ What is it that I value in agile development? Is it the velocity alone, or


do I feel that agile is the best way to control the development process?
◾ Are there requirements from the standards which I feel slow down my
agile model? Do I understand these requirements?
266 ◾ Medical-Grade Software Development

◾ What evidence does the management of the agile life cycle itself give
me? Which pieces of this evidence are valuable in the regulatory con-
text, and how?
◾ What deliverables are expected by the standards and my applicable
regulations?
◾ Do I know where each of the deliverables fit in my life-cycle model?
Do I know when each of these is intended to be ready? Do I know the
point after which these deliverables are unlikely to need revision?
◾ How do processes such as usability, risk management, and change
control affect my life-cycle model? At which points of the development
work do these fit in, or where is their impact the greatest?
◾ Am I happy with my control over both parallel and serial work?
◾ How do I monitor external requirements and changes in them? For
example, regulations and their guidance will evolve over time, perhaps
even over the development phase of the software.

Answers to the above questions, and to the follow-up questions you will run
up against in going through the above list, will be your way to figuring out
if and how you should adopt an agile software-development model from a
regulatory point of view. The more detailed you can get in the questions the
easier you will find it to fashion the required standard operating procedures
for your organization – and the better off you will be in meeting your regu-
latory expectations.
In practice, no life-cycle model available today will fill the regulatory
requirements for you. You will, in all likelihood, need to connect the dots
yourself in the way that best meets the needs of your organization and your
products. Every model has its pros and cons, there’s no use moping about
the issues you will face, but it is your responsibility to get it all done in a
practical, acceptable way. The discussion in this section was not intended to
say that you are lazy if you can’t handle the waterfall, or that you are stuck
in the 1950s if you don’t want to embrace the agile. Our aim here was to
introduce you to some of the topmost points of lamentation and exuber-
ance with either model. There are also more life-cycle models introduced in
literature than we covered here. We only chose two of the most prominent
examples at the opposite ends of the spectrum in terms of longsightedness
and flat-out clairvoyance.
At the end of the day, the three most important questions to answer in
shaping your development model, whether you start from a waterfall or a
washing machine, are as follows.
Seeing into the Future ◾ 267

◾ Does the model achieve what I want and get me the outcomes I expect?
◾ Does the model produce and maintain all the regulatory evidence I
need in my chosen market areas?
◾ Is the model cost-effective and reliable?

By analyzing the above questions during the development of your life-cycle


model you will stand the best chance of arriving at a model that not only
meets the regulatory expectations but does so reliably in the long term.
There are many more questions you will have to consider during setting up,
running, and later updating your development model, but many of them
should be traceable back to these fundamental questions.

12.2 IEC 82304-1 on Health Software


Films often get remakes after enough time has elapsed. Julia
Ormond stepped into the shoes of Audrey Hepburn to portray
Sabrina, Robert De Niro hung on to the bottom of a moving car to
follow Robert Mitchum to Cape Fear, and Timothée Chalamet wiped
away any traces of earlier versions of Dune. In all these cases, some-
one somewhere decided that it would be a good time to do a bit of a
do-over. In some respects, for IEC 62304 that remake is IEC 82304-
1. The new standard does not replace the old, but it does investigate
some of the points of criticism levied against its predecessor – even if
it appears to change the forum from the silver screen to YouTube in
the process.
The IEC 82304-1 standard on product safety requirements for health soft-
ware is technically beyond our scope here in this book, but as the standard
is routinely considered as an extension to IEC 62304 it is worth discussing
here in some detail. As a separate document it perhaps also demonstrates
some of the shortcomings identified in IEC 62304 during its prolonged stay
at the maintenance dock. It should, however, be noted that IEC 82304-1 has
been written with health software in mind, and this does not necessarily
imply the same deep end of the pool as medical-device software – depend-
ing on who you ask health software may be the whole pool in general or
just the kids’ end of the pool. The standard is intended to be compatible
with IEC 62304 through and through.
In recent years software has forged a definite beachhead into the moni-
toring and diagnosis of patient health. This foothold is only being enlarged
268 ◾ Medical-Grade Software Development

with each new feature and new device released into the global marketplace
– so much so that it is fair to say software has an increasingly important role
to play in how we detect and treat diseases today. The presence of soft-
ware also transcends the traditional boundaries of space as the automated
analysis of vital parameters and other forms of health data may, for example,
take place at the bedside, at an on-premises hospital hub, at an off-premises
central analysis hub, or even in a distributed fashion in the cloud. The deci-
sion on the treatment of the patient is still made by the healthcare profes-
sional, but the information they use may be prepared by a complex network
of software services. In addition to the analysis taking place in a distributed
fashion the data sources used may themselves be distributed across different
geographies and systems (e.g., electronic health records, laboratory systems,
imaging systems, and patient monitoring systems). Furthermore, the plat-
forms and operating systems such software runs on will be developed by
third parties and often for some general purpose beyond the intended use
of the medical software.
To address this modern-day diversity, IEC has published the IEC 82304-1
standard titled Health Software — Part 1: General Requirements for Product
Safety. In films any time you see a movie labeled as “part one” you know
that it’s not going to end satisfactorily until a few sequels later. The stan-
dard, though, is a fairly satisfactory extension of what has come before. This
standard is intended to build on the established IEC 62304 standard and set
requirements for stand-alone software that is intended to run on general
computation platforms. Appendix A of the standard lists examples of the
types of software it can be used for, but in practice all types of software
– including that complying with Article 2 of the EU MDR – can adopt the
standard on top of IEC 62304. However, medical electrical equipment, IVD
devices, and implantable devices are primarily left for other standards to
cover.
IEC 82304-1 is widely recognized in the medical-device world as a helpful
new kid on the block – someone who will help carry your groceries back
home, and even spot some key items you forgot to buy, but it will not speak
to the healthiness of the meal you will make out of the ingredients. In other
words, the standard is a helpful addition to refer to when trying to meet the
requirements of the standards and the regulations, but it doesn’t attempt to
offer anything close to a standalone solution.
The standard is recognized in many jurisdictions around the globe
(e.g., US FDA, Australian TGA) and routinely addressed in conformity
audits in one capacity or another. The standard does contain valuable
Seeing into the Future ◾ 269

elements that would otherwise be missing from the overall framework


set up by ISO 13485 and IEC 62304. The thinking offered by the standard
and the itemized lists of aspects it points out in relation to, for example,
identifying product requirements (including those regarding platforms,
security, usability, and accompanying documentation) and conducting
product validation, each serve to give added structure to the development
of medical devices. The thinking it offers is, however, best viewed as
guidance, not necessarily as requirements to hit during your D&D activi-
ties. Adopting the standard may help you to consider all the poignant
points of view as inputs for your D&D activities, and thus elevate the level
of your design activities.
The standard adds to the story told by IEC 62304 and ISO 13485 so far in
terms of the following five topics:

◾ Product requirements
◾ Software development process
◾ Validation
◾ Product documentation
◾ Post-market activities

Let’s look at each of these briefly in the following.

12.2.1 Product Requirements
The determination of product requirements now includes identifying risks
related to not just safety but also information security. The manufacturer
must also consider, as is applicable, those situations where their product is
configured or has interfaces with other products. Full ISO 14971 conformity
is not required here, but the manufacturer must determine the need for
risk controls (mitigations) for unacceptable risks. This risk assessment then
serves as an input to product requirements. IEC 82304-1 does not expect the
manufacturer to be able to adhere to ISO 14971 for each constituent part of
the software (e.g., proprietary components, non-healthcare components, and
legacy software), but they are expected to perform risk control around those
components where risks are found to be unacceptable.
The standard also goes into added detail on use requirement and system
requirements, and how the two are to be verified and updated (although
here no real new requirements are introduced). Here you are instructed to
consider the following new items for use requirements:
270 ◾ Medical-Grade Software Development

◾ Interfaces and user interfaces (e.g., What is to be achieved via the inter-
face? Is the user a human or another system?)
◾ Immunity from/susceptibility to interference from other software run-
ning in the same environment (e.g., concurrent memory use, EMC/ESD
interference)
◾ Privacy and security issues (e.g., authentication, malicious use)
◾ Accompanying documents (e.g., IFU)
◾ Software updates (e.g., upgrades, rollbacks, patches, distribution, dele-
tion, and retention of data)

In terms of system requirements, the standard points out the following


potentially new items:

◾ Technical user interface properties (e.g., display color, character size)


◾ Operating platform (e.g., operating system, software libraries)
◾ Inter-operability
◾ Localization and language support
◾ Risk controls implemented in the product (at the system level)
◾ Detection and handling of security compromises
◾ Protection of essential functions
◾ Product configuration, retention, and recovery

Glancing through the above abbreviated lists it becomes evident that the
IEC 82304-1 standard broadens the field of vision set up by the other two
standards, ISO 13485 and IEC 62304, from safety to security in quite distinct
ways. The consideration of the underlying platform is now more nuanced
than just some piece of hardware the manufacturer is expected to have full
control over as we are looking at general-purpose computing platforms.
They should, however, address the acceptable parameters of those plat-
forms and their own risk controls for working on top of those platforms.
Despite the throning of security on a level next to safety, we should not
lose focus of the user and the patient as the origin of our coordinate sys-
tem for risk management – IEC 62304 expects all risks to be measured
in relation to the patient and IEC 82304-1 says nothing to change this.
The Implication is that a business risk related to security is not necessar-
ily in scope, but a security risk with a harmful impact to the patient or the
user definitely is. News stories from the past years provide us with ample
examples of security vulnerabilities having led to leaked health records and
compromised patient privacy.
Seeing into the Future ◾ 271

12.2.2 Software Development Process


The IEC 82304-1 standard adopts IEC 62304 as the basis of all its require-
ments. In particular, it states that Clauses 4.2, 4.3, and 5 through 9 of IEC
62304 apply in addition to its own requirements. This constitutes the bulk
of IEC 62304, although technically the clauses on the QMS (Clause 4.1) and
legacy software (Clause 4.4) are omitted. IEC 82304-1 further states that the
system requirements identified are to be used as the primary input for the
life-cycle process of the software product.

12.2.3 Validation
IEC 82304-1 speaks of validation in far more detail than either of the two
principal standards discussed in this book. Here it sets requirements for how
the validation is performed as well as what an individual validation plan and
a report should achieve.
The plan must address all user requirements, the validation scope and
activities (incl. activity feasibility and criticality), and validation methods (incl.
inputs and acceptance criteria). It must also address the systems and services
needed in the validation (e.g., operating environments where the software
is to run), the qualification of personnel performing the validation, and the
independence of the validation team from the design team. The plan must
be established before validation is performed, or validation readiness can be
said to have been reached.
Before commencing with the validation, the validation team must be
sufficiently qualified, and the relevant IEC 62304 Clause 5 software-devel-
opment phases for the parts of the software being validated are to be
complete. The former stipulation is what ISO 13485 would also say on the
matter, but the latter stipulation is exciting as it opens the door to piecewise
validation of software. Remember that ISO 13485 is adamant that a repre-
sentative sample of the final product is used for validation, but here IEC
82304-1 appears content to validate parts of the eventual product. In prac-
tice, you will want to meet both expectations: if you validate parts of the
whole, you must be able to comfortably argue that also the validation of the
whole is satisfactorily met before product release. In practice, you will want
to carefully plan the interplay between your various verification and review
activities and explain how these are planned to be useful in your overall
validation activities – especially if you perform some validation during these
earlier development activities. The analysis of the whole body of work can
272 ◾ Medical-Grade Software Development

be made in the final validation report, but it is beneficial to know how the
various activities slot together from the start and to make sure your soft-
ware-development activities are complete enough to allow the correspond-
ing validation activities.
The validation itself must take place as planned and in the intended
operational environments. Any deviations from the plan must be justi-
fied in the report, and any anomalies discovered are to be handled via the
problem-resolution process discussed previously (see Section 9). Something
which is occasionally forgotten in today’s agile development environment
is that it is also necessary to assess the impact of the modification on the
other design choices already made, and possibly launch further change-
order processes to address these effects (see Section 8.2). The valida-
tion may then need to be repeated in whole or in part if the software is
modified.
In terms of reporting on the validation, IEC 82304-1 requires that the
validation team develops the report. This could be understood simply as
a report having to be developed, but the requirement is oddly specific in
requiring the validation team to be the ones developing it. In other words,
no stunt performers are accepted here. The report must record the valida-
tion conditions and activities, any deviations (incl. justification) or anoma-
lies, and the validation team (incl. name, affiliation, and function). It must
then provide a summary of the results, all the required evidence (i.e., the
traceability of results back to user requirements, the fulfillment of the user
requirements, and the acceptability of the residual risk), and a conclusion
that validation is now satisfactorily completed for the intended use based on
the user requirements.
Also, note that the standard discusses the need to re-validate the software
after changes in its clauses on post-market activities (see Section 12.2.5).

12.2.4 Product Documentation
As IEC 82304-1 expects your software to live a half-nomadic existence jump-
ing from one platform to the next, it quite rationally also has expectations
on the documentation that will travel along with it for the users to inspect.
You can think of this as the IEC 82304-1 passport for your product contain-
ing both the necessary identification information for it and the key docu-
mentation for its use.
In terms of identification, your software must be identifiable by your
name or trademark, product name or type reference, and a unique version
Seeing into the Future ◾ 273

identifier (e.g., version number, manufacturing date, UDI). The user must
have access to this information when using the software (e.g., via an “about”
box, on the opening screen).
In terms of the provided information, the standard expects there to
be both instructions written on a user-friendly level and a more technical
description of the product. The set of documentation provided must describe
any requirements placed on the user (e.g., training, skills, and knowledge),
the use environment or location, and the hardware or software to run the
software. This is all to be based on the intended use of the device. The
documentation must be user-friendly, or in other words match the user’s
education, training, and possible special considerations. The documenta-
tion may thus benefit from, or be required to undergo, usability-engineering
assessment (see Section 12.5).
The user-friendly instructions probably take the form of the now famil-
iar Instruction for Use (IFU). Here IEC 82304-1 expects all the topics rel-
evant to the installation and use of the software to be covered including the
following:

◾ The intended use, limitations for use, and known technical issues
◾ Installation and platform requirements
◾ Dependencies, interfaces, and configuration options
◾ Restrictions on IT network usage
◾ Starting up and shutting down the software
◾ Introduction and essential features
◾ Operating instructions and explanation of symbols, signals, warnings,
notices, and error messages
◾ Safe decommissioning

Note that the warnings and notices included for safety or security reasons
must be instructed adequately in the IFU – what is thought adequate is
determined via the manufacturer’s usability-engineering process while pay-
ing special attention to those warnings perhaps used to mitigate risks to an
acceptable level. IEC 82304-1 Clauses 7.2.2.1 to 7.2.2.10 give the full list of
the individual items to cover here.
The technical description, on the other hand, is to contain the following.

◾ All essential information (curiously called data by the standard) to safely


transport, store, and operate the software. This includes any necessary
information on environmental transport/storage conditions for media,
274 ◾ Medical-Grade Software Development

and any conditions, measures, restrictions, or special requirements nec-


essary for installation.
◾ Information on supported platforms and hardware/software system
requirements, and how to cope with hardware/software platform
changes.
◾ Information on configurable security options (incl. operational options),
and how to select platform settings to promote security.
◾ Information on required maintenance.
◾ Information on the effects of a security failure and its impact on patient
care, data, and clinical workflow.

In addition, the description is to contain or reference all the characteristics of


the software (incl. ranges, accuracy, and precision of the displayed values),
which may be a tall order. Clause 7.2.3.1 gives the full list of the individual
items to cover here but does so in a lamentably confused presentation.
The discussion of configurable security options grabs the attention from
the above list. This may, for example, include selected network ports, pass-
word options, and logging levels. Prescribing strict rules to promote high
security is a noble goal, but if this leads to poor usability, especially during
installation, the users may act loose when being faced with difficulties in
getting the software to work at all. Nevertheless, you should not leave doors
open if you don’t want visitors through those doors.
In the case of software intended to connect to an IT network not con-
trolled by the manufacturer, additional expectations are placed on the tech-
nical description. Firstly, the provided information must now also cover
descriptions of the intended information flow between the software and any
other software or the IT network (incl. characteristics, configuration, techni-
cal/security specifications, and malware protection). Secondly, any hazard-
ous situations arising from network failures are now to be listed and the
user of the software is to be cautioned that further unidentified risks may
arise to the user (also patient and other third party) when software is used
on a network. These risks should be appropriately managed, and it should
be remembered that any changes to the IT network (e.g., changed configu-
ration, and hardware/software updates or upgrades) may give rise to new
risks.
Finally, similar to the identification of your software, the accompanying
documentation is expected to reference the manufacturer, the manufacturer’s
contact information (incl. website address), product identification, and docu-
ment version. Don’t interpret this as a need to always stamp this information
Seeing into the Future ◾ 275

on each piece of the documentation, but to include the information some-


where practical within that body of documentation.

12.2.5 Post-Market Activities
The IEC 82304-1 standard intends to address the entire life cycle of software
all the way up to maintenance, decommissioning, and disposal, as stated in
Clause 1. Yet, in terms of assessing the compliance of post-market activities,
it draws the line at aspects related to product D&D (Clause 8.1). The inten-
tion here may be to limit the scope of information collection and analysis
needs to some practical sphere around D&D, but the wording is cumber-
some. Given that we know sensitive data has occasionally found its way into
scrapyards when some disk drives or memory media have not been wiped
appropriately, we should also concern ourselves with learning from such
incidents to better design our devices and instruct their end of life. Given
that users are not naturally giddy to analyze and install each new version
update put out by maintenance, we should think carefully about how we
want to instruct critical software maintenance. The standard does go a long
way to addressing all these aspects in its other clauses, but when it comes to
discussing them in the context of post-market activities it appears a little coy
here.
On maintenance, the standard invokes IEC 62304 with the disclaimers it
made earlier in its Clause 5 (e.g., IEC 62304-compatible development model,
ISO 14971-lite risk management). If maintenance is required, also re-valida-
tion must take place as is relevant to the modification and the supported
platforms, and the accompanying documentation may need to be updated.
The standard also calls for the validation plan to be updated accordingly
after re-validation, but whether that results in changes to your plan will
depend on your plan – only update plans when your plans change, not to
update timestamps or to just track their execution. Remember, too, that the
detected safety and security issues may also trigger regulatory reporting
needs and actions in revising the conformity assessment.
On product-related communication, the standard expects to see the
manufacturer take steps to inform users about a) security vulnerabilities, b)
changes in relevant regulatory requirements, c) available new versions of
the product, and d) a range of other information such as new features, fixed
issues, safety or security developments, and updates to software identifica-
tion or accompanying documentation. The decision by the user to install
an update should be based on the safety and security of the new version.
276 ◾ Medical-Grade Software Development

Appropriately, if the new update has a positive impact on safety and secu-
rity, the manufacturer may advise the users to replace their older versions
with it. The wording of the clause here is less than ideal, but it appears to
discuss both communications accompanying a new release and the commu-
nication taking place to, for example, enable the discovery of that release by
its intended users.
Finally, the standard addresses the skirted decommissioning and disposal
of products. Here the manufacturer is expected to consider applicable use
requirements and provide a safety function for disposing of the product.
This necessarily includes any measures needed to safeguard personal health
data and other security and privacy concerns.

12.3 Segregation in Software Architecture


When Riggs went berserk in “Lethal Weapon”, he had Murtagh to
talk him back from the final ledge. When the adults in “Jurassic
Park” were at a loss to know how to proceed, the girl pulled in a
trump card of knowing Unix. And when the reigning Phantom
kicked the dust, the next generation was again there to pick up the
slack just as promised by the old jungle saying. None of this would
have been possible if all the actors would have been handcuffed
together all along.

The term software architecture (see Section 5.3) refers to the composition
and internal structure of software, including how the various subcompo-
nents communicate with each other. The structure may, for example, be
defined along a functional axis (i.e., what components are involved in the
realization of what function or feature of the software) or an ontological axis
(i.e., what components are used by which other components). Occasionally
the difference between these two axes may become blurred, but what mat-
ters is that the developed architecture satisfies René Descartes to the extent
practical for realizing both the individual components and the whole of the
software.
As discussed in Section 1.5, in the eyes of IEC 62304 everything in the
software from units to the whole of the software system is fundamentally
a software item. Each item has its own part to play on the whole, but in
addition to their correct function, we must also address their possible neg-
ative effects on the whole. What happens if that item malfunctions? Can
Seeing into the Future ◾ 277

it cause harm to the users, patients, and environments around the medi-
cal device? This is where the concept of segregation becomes extremely
important. In other words, this is where we realize there’s a Fredo in the
architectural design and they are courting disaster so we must cut them
out of the family tree like Michael did in Godfather II. We already briefly
discussed the bulkheads onboard the Titanic and the role external risk
controls can play in addressing risks in Section 4.3, but let’s dive a little
deeper here.
Traditionally segregation has perhaps been thought of as a property of
physical components. Redundancy has been built into systems to ensure
that, for example, critical life support processes in hospitals or naturally
hostile environments such as space do not accidentally end on a glitch.
Redundancy may have meant adding parallel components such as mul-
tiple processors to a device or perhaps even just adding an uninterrupted
power supply to take over should a mains connection fail. Today, software
is responsible for many critical actions in medical devices, and a flaw that
goes unchecked in one component may cause serious harm depending on
the type of device. It is a common mistake to think that software errors
can only be caught by hardware controls or instructions to the user. Not so.
Software can, and in some cases should, watch over other software. This is
why segregation between software systems and possibly software items is
such an important topic today.
Standards such as IEC 60601-1 and IEC 62304 do speak of segregation.
The primary setting for this discussion is risk management, where the role
of reliable segregation is understandably heightened. When the risks regard-
ing one software item are mitigated (controlled) via expectations placed on
another software item it becomes extremely important for the second item to
be as immune as possible to the failure of the first. According to IEC 62304
Clause 5.3.5, the manufacturer must identify such segregation needs regard-
ing class C software items, and also ensure the effectiveness of the segrega-
tion. The standard gives us a simple, if non-binding and non-exhaustive,
example of running the two software items on separate processors that do
not share any resources as a reference to effective segregation. Figure 12.1
illustrates a possible breakdown of this solution where the processes from
each processor have access to the same memory space.
The example given by IEC 62304 is not exhaustive and it is not much
more satisfying than saying that two items separated by lightyears of space
between them are not likely to affect each other unless you believe in the
butterfly effect. Luckily, IEC 60101-1 is more forthcoming. Clause 14.8 of that
278 ◾ Medical-Grade Software Development

Figure 12.1 Segregation breakdown due to memory access.

standard lists the following design options for mitigating risks in the device
architecture:

◾ Using high-integrity components (e.g., components with high mechani-


cal durability)
◾ Employing fail-safe functions
◾ Building in redundancy
◾ Allowing for diversity
◾ Partitioning functionality
◾ Incorporating defensive design (i.e., countering harms via pre-emptive
design choices such as limiting electric currents and the physical move-
ment of actuators)

From the above list, it is apparent that IEC 60601-1 is considering physical
medical devices, not software, but the thinking it offers is valuable in both
contexts. For software, the list could, for example, refer to using validated
software tools and controlled SOUP components and building in fail-safe
functions, redundancies, and defensive design considerations. The take-
home message here is that it is entirely possible to improve the resilience
of software and reduce the risks in its use via designing for safety in case
someone still had their doubts on the matter. How effective, necessary, or
sufficient those methods are will depend on the type of medical device.
The IEC 60601-1 standard goes on to suggest that the following are
considered when designing the architecture: a) allocation of risk-control
Seeing into the Future ◾ 279

measures to subsystems and components such as sensors, actuators, and


interfaces, b) failure modes, c) common cause failures, d) systematic failures,
e) test interval duration and diagnostic coverage, f) maintainability, and g)
protection from reasonably foreseeable misuse.
Based on the above, good segregation is not limited to physical measures
(e.g., redundant processors and memory partitions). Despite the scarcity of
examples in IEC 62304, segregation measures include all appropriate mecha-
nisms that act to prevent one piece of software from negatively affecting
another. The adequacy of this segregation may then be determined based
on the associated risks, risk-management activities, and risk documenta-
tion. Even when relying on memory partition as a segregation method the
manufacturer must still consider all risks related to that function, including
the allocation of memory. An example here might be that a Task 1 running
on CPU1 is not able to access the memory space used by Task 2 on CPU2
but it is able to somehow affect the memory allocation criteria for CPU2. The
example here is not meant as fearmongering, only as a friendly reminder
that magic bullets do not exist, and it depends on the nature of your appli-
cation and how tight you should wind the tinfoil around your head.
The discussion in this section has brought up aspects to consider when
specifying segregation in your software. The lens we have used here is risk
management where segregation is naturally of the highest importance. We
don’t really care if Riggs and Murtagh share a slice of germ-laced pizza every
evening after work, or if Phantom number 35 had a happy childhood, but
we do care if the segregation we rely on is not really there in reality. The
point here is both that solid segregation may be necessary for reliable risk
control to take place, and that addressing any potentially significant leaks
in the segregation of components and subsystems is a part of risk manage-
ment. The place to analyze and set up controls for segregation is in develop-
ing the architectural design for your software, which also has to be in sync
with the requirements defined for it.
In the world of physical medical devices, segregation may have been
mostly about electric shock and EMC testing. In simple software, this may be
expanded to segregation between software items of varying degrees. In com-
plex, network-connected, and perhaps constantly evolving devices, though,
risks may also be introduced via changing or misbehaving datasets, not just
misbehaving code. When a decision support system, for example, bases its
assessment on available reference datasets, and these change over time, the
analysis offered by the software may change as well. If the dataset is no one
thing that changes through careful curation, but instead the reference dataset
280 ◾ Medical-Grade Software Development

or learning dataset consists of multiple information sources pooled together


the likelihood of issues propagating may also increase. Data quality criteria
may be part of the solution to working with such shifting foundations, but
so too may many of the segregation design considerations introduced above
for hardware devices. On other occasions, addressing cybersecurity issues in
some appropriate way may be key (see Section 12.6). In the long run, we will
also need a good standard for medical-device data governance and use to
augment the thinking offered by ISO 13485 and IEC 62304.

12.4 Risk Management Influenced by FMEA


Asteroids and natural disasters are probably more of a concern for
FEMA than for FMEA, but for analyzing risks in man-made con-
structs the latter has a very strong following. FMEA is no stranger in
an ISO 14971 setting, either, but in Tropic Thunder terms it is the
Tom Cruise of the show, not Ben Stiller or Robert Downey, Jr.
Failure modes and effects analysis (FMEA; see IEC 60812) is a method
developed in the 1950s to identify potential failure points in systems con-
sisting of multiple components and subsystems. The aim is to identify the
failure modes of each component and analyze their effect on the whole.
FMEA is a popular bottom-up method that analyzes each constituent part at
a time, and once a potential failure mode is identified, looks at the hazards
and harms that may arise from it, as well as the initiating cause behind the
failure. Each sequence is then evaluated based on the severity and probabil-
ity of the harm.
In medical devices, FMEA is often used just as it is used in other domains
of engineering, although more than one flavor of FMEA exists in each of
these domains. It is, however, crucial to note that FMEA does not necessar-
ily cover all the aspects of risk management expected in medical devices. As
discussed in Section 7.5, risk management in this context must consider at
least the following types of risks:

◾ Risks associated with software failures


◾ Risks associated with the correct, intended use of software
◾ Risks associated with foreseeable misuse of the software

Of the above types of risks, FMEA is most readily applicable to software


failures, of course, but it may miss important risks falling into the other two
Seeing into the Future ◾ 281

categories. In the context of medical devices, the base standard expected


of all risk-management activities is ISO 14971, which can be leveraged for
all three categories. Techniques such as FMEA may be used to help in this
overall analysis but are usually not comprehensive enough to tackle the
whole sphere of risks by themselves. It is up to the manufacturer to set up
their risk-management process so that it meets the combined expectations of
the standards discussed in this book (mainly ISO 14971, ISO 13485, and IEC
62304) and produces all the relevant information expected by the regula-
tions. In this, FMEA is not the magic bullet solution some appear to expect.
It is, however, a good tool to have in the complete toolkit.
Considering that an FMEA must meet the expectations of ISO 14971 to
pass muster in the medical-device context, we should consider what ele-
ments the two have in common. In practice only the starting point may be
different between the two: whereas ISO 14971 anticipates risks from a wide
field of sources, FMEA has its sights on the parts of the device. Both speak
of hazards, causes (in FMEA: initiating cause), harms, probabilities, severities,
and a score calculated for each risk (risk priority number or RPN for FMEA;
see discussion below). Thus, a practical way to combine the two worlds is to
allow for multiple different types of risks, one of which is component failure.
Figure 12.2 illustrates this. In the figure, failure modes identified for compo-
nents are one of the sources of identified risks, just like the intended use,
usability, foreseeable misuse, and cybersecurity of the device.
Note that the figure here omits requirements for the sake of clarity. The
intended use shown in the diagram should in many ways act as the ances-
tral root of all information on the developed device, including how it is

Figure 12.2 Hazards as a gateway to risk management.


282 ◾ Medical-Grade Software Development

ultimately to be qualified and classified as a medical device under the appli-


cable regulations. The intended use would then spell out requirements for
the device, which in turn would lead to identified failure modes and which
could then contribute to hazards. At the same time, the intended use would
advise the preparation of a use specification (See Section 12.5), which in
turn could lead to the identification of new hazards. Thus, the figure here is
a simplification drawn up to illustrate the general role these types of infor-
mation play in risk management.
The risk score (Sc), as expected by ISO 14971, is often a simple multipli-
cation of the severity (S) and the probability (P) of each risk. Both severity
and probability may be expressed as percentages (0.0–1.0 or 0–100%) or on
a chosen practical scale (e.g., 1–5). The score is then a multiplication of these
leading to a combined classification on a scale from 0 to 1 (0 to 100%) or
the square of the scales used originally (e.g., 1–25). The risk priority num-
ber (RPN) introduced by FMEA is similar but not the same as this simple
score. The chief motivation behind RPN is to differentiate between some
risk that is severe but rare (e.g., severity 5 and probability 1) and a risk that
is negligible in effect but likely (e.g., severity 1 and probability 5). Plotted in
a matrix of severity and probability these two risks would occupy opposite
corners of the matrix, but as numbers they would look the same. To tackle
any confusion, the RPN scheme assigns user-defined scaling factors to either
a) the severity classification alone or b) both the severity and the probability
classifications. This way the RPN scheme separates such risks at a glance.
For example, by assigning a multiplier of 5–25 to the severity classification,
the combined score for a severe but rare risk becomes 125 (25*5*1) and the
score for a negligible but likely risk becomes 25 (5*1*5). For comparison, the
RPN score for a risk that is both severe and likely would be 625 (25*5*5).
Figure 12.3 illustrates the calculations involved in the determination of the
RPN score.
Similarly, we could also assign arbitrary multipliers to the probability
classification if this provided some practical separation for us in risk man-
agement. Also, remember what we discussed in Section 4.3 about software
always failing, and if you decided to extend the P1=100% assertion beyond
simply assigning the software safety classification act accordingly here.
In practice, FMEA may capture much more than just a third of the medi-
cal-device risks expected to be managed. FMEA may also be the best, most
practical, and best-known method for tackling risks related to software fail-
ures. In such cases, the gap between FMEA and ISO 14971 risk management
may also be assessed via a summative risk-management report that addresses
Seeing into the Future ◾ 283

Figure 12.3 RPN risk scores scaled to emphasize the role of severity.

any FMEA shortcomings when viewed from an ISO 14971 perspective. The
report must then demonstrate that the overall risk management (incl. residual
risks and the overall residual risk) is at an acceptable level.
In assessing how, and to what extent, FMEA can be used to meet med-
ical-device risk management needs it is important to consider the types of
devices manufactured, their intended uses, their features, and the technolo-
gies they incorporate when attempting to identify the risks involved in their
use. Component failures are a part of the picture, but as illustrated by the
list above this may not be the whole picture. Herein lies the principal short-
coming of FMEA in the medical-device domain: how can we reliably capture
all the relevant risks and harms that may befall the patient, the user, and the
environment via analyzing component failures? And if we are too tuned into
component failures, will we be working overtime on risks that are not actu-
ally medical-device risks in that no harm can come to people? Both of these
issues can, of course, be solved with FMEA, but neither may be a simple
task when the device and the context of its use are as complex as is to be
expected in the clinical setting.
One potential hurdle to cross when adopting FMEA is, therefore, the
need to align activities and outcomes with the ISO 14971 standard which
sets the overall framework of expectation in medical devices. At the same
time, more than one flavor of FMEA exists which means the manufacturer
must choose from multiple analysis matrices and perhaps still modify these
for their own use. This can all be done, and it may lead to the best solution
for risk management, but FMEA is not the be-all and end-all of medical-
device risk management. For more discussion of FMEA see IEC 60812.
284 ◾ Medical-Grade Software Development

As popular as FMEA is, it is only one possible method to use. Another


popular option is fault tree analysis (FTA; see IEC 61025). This is a top-down
method for identifying risks resulting in harm to patients, users, bystanders,
and the environment. The analysis is performed via so-called top events,
which are an abstraction of faults that can then be linked with hazards, haz-
ardous situations, and harms. For more discussion of FMEA see IEC 61025.
Both FMEA and FTA are standardized methods and are described in
detail in their associated standards and guidance. Both can be leveraged
in the context of medical devices, but it is worth remembering that in the
end whatever method is chosen for use in medical-device risk management
must meet the expectations of ISO 14971 and the applicable regulations. As
a friendly warning, also note that not all standards agree on the definition of
risk: business risks, for example, do not generally feature in medical-device
risk management unless they may cause harm to the patients and the users
of the device, but such risks may be given as the primary reason for the
existence of some other standards developed on other domains.
The discussion in this section should not be interpreted as an attack on
FMEA, only as a friendly warning that reaching the Rockies does not really
mean you have driven the length of Route 66 no matter which end you
started from. ISO 14971 gives you the yardstick you will need to use here in
each and every case, but methods like FMEA and FTA may help you cross
the finish line in style – just like Tom Cruise helped wrap up Tropic Thunder
with a bang.

12.5 Usability Engineering
Good usability in medical devices means safety over snazziness. It is
the med bay Noomi Rapace climbs onto to single-handedly perform
an abortion of the alien lifeform, not the futuristic floating displays
Tom Cruise commands like he’s the conductor of a great symphony
orchestra. Equating coolness and usability is an easy mistake to
make, particularly when entering medical devices from another
industry.
The usability of medical software is a topic often addressed from the point
of view of risk management, and rightly so. The different jurisdictions and
market areas across the world may each impose requirements on the usabil-
ity of medical devices. Such requirements are usually not set from the point
Seeing into the Future ◾ 285

of view of intuitivity or ease-of-use, but with a view to avoiding use errors


stemming from bad design (e.g., Clause 5 in EU MDR Annex I). In medical
devices the terms “usability” and “performance” both actually refer to safety,
one from the outside in and the other from the inside out in reference to the
device.
In medical devices, usability is primarily about understanding the interac-
tion between the device and its user or user group. At this interface a simple
erroneous action performed on the part of the user may affect the device in
unwanted ways, and similarly the user’s incorrect understanding of what the
device is saying may misdirect subsequent actions in the care of a patient.
Both errors may lead to a hazardous situation or a right-out use error.
Neither error can be chalked up to mere user error, which is why the term
use error is used instead and the software is also on the hook as a source of
contributing factors.
A user interface thus acts in two opposite directions. It is the medium
through which the device presents information to the user and at the same
time the means for the user to manipulate the device. In terms of usability,
it is both important to ensure that the information is presented in a clear,
appropriate, and readily interpretable way, and to make sure the actions
offered to the user also make sense to them before these are invoked. Clarity
here means, for example, that numerical display can’t be misinterpreted from
an off-axis angle (as has famously not always been the case with recessed
segment displays), and appropriateness in tuning the offered information to
the skills, education, and mental context of the user. It is in the best inter-
est of both the user and the device to make this interaction as predictable as
possible, especially in terms of what is going to happen after each particu-
lar user input or action. The science of usability engineering – whether that
takes the form of human–computer interaction, service design, or some other
appropriate method – is how we formally tackle this interface and attempt to
arrive at clear design requirements in the form of, e.g., use specification, UI
specification, and hazardous use scenarios. Methods we can use in this work
include task analysis, simulation, function analysis, contextual inquiry and
observation, surveys, interviews, expert reviews, and advisory panel reviews.
The methods we might use each have different characteristics and may
vary in their suitability to the different stages of a usability study. Out of
all the methods, task analysis is perhaps the one best suited to modeling
the way users interact with the devices. In task analysis, we investigate this
interaction via the use of scenarios that attempt to capture the tasks the user
wants to perform and help the manufacturer design matching user interfaces
286 ◾ Medical-Grade Software Development

and identify the risks involved. As issues are detected that may lead to use
errors these are analyzed further. In practice, the manufacturer may choose
the methods that are best suited for their type of product, type of organiza-
tion, skillset, resources, and past experience. It is, however, important to be
able to demonstrate adequate competence in the use of the method down to
the personnel-level and to maintain sufficient documentation.
Typical examples of areas where shortcomings are identified in the
usability of software include the following:

◾ Labeling (incl. visible and audible prompts for interaction)


◾ Navigation (incl. drop-down menus)
◾ Alarms (incl. alarm fatigue)
◾ Presets for presumed user groups
◾ Configuration performed by the user or maintenance staff

Many of the above types of issues are shared by both software and hard-
ware, and it would be wrong to assume software would be immune to any
of the issues. IEC 62304 itself is adamant that all software fails (see Section
4.3), which may or may not be the case, but it certainly further heightens the
need to ensure good usability, transparency, and understanding between the
device and its user. These characteristics will only become more important
as we adopt artificial intelligence and other technologies enabling devices to
evolve while in use.
A particular subtopic for usability engineering is the use of labeling as
part of the overall risk-management activities for a device. If labeling is used
to implement information for a safety-type risk control, it must be analyzed
as part of usability engineering. ISO 14971 Clause 7.1 and IEC 62366-1 Clause
4.1.3 both agree on the matter.
The IEC 62366-1 standard on the application of usability engineering to
medical devices is a cornerstone for ensuring safety through usability. The
associated technical report, IEC/TR 62366-2, provides further discussion on
the standard, its terminology and requirements, and the usability-engineer-
ing methods to use in the various stages of work. IEC 62366-1 sets the fol-
lowing requirements for addressing medical-device usability.

◾ The manufacturer must have a usability engineering process in place.


This covers the use of the device when used according to its accompa-
nying documentation. The process is to be implemented by qualified
personnel.
Seeing into the Future ◾ 287

◾ Any risk control is performed in accordance with ISO 14971 Clause 7.1
and should consider, for example, inherent safety, protective measures,
and information for safety as possible controls.
◾ If information for safety is used as a risk control this is verified to be
perceivable, understandable, and to support the correct use of the
device (incl. by the correct user groups and in the correct environment).
◾ The conducting of usability-engineering activities is planned, and
appropriate for the stage of the device or the magnitude of the change
(e.g., a new device versus a small UI update to an existing device).
◾ The appropriate documentation is maintained in a way that facilitates
the use of the usability-engineering information in the technical file of
the device and as required in the applicable jurisdictions.

When a manufacturer fires up the usability-engineering process for their


device, the first step expected is the preparation of a use specification for
the device (IEC 62366-1 Clause 5.1). This specification is to include the
intended use, intended patient population, intended part of the body or
tissue type interacted with (if any), intended user profile, intended use
environment, and the operating principle of the device. Note that this
information is also needed in the technical file of the device for regulatory
purposes.
The specification is then used to identify a) the characteristics related to
safe use of the device and any potential use errors (Clause 5.2), b) known
or foreseeable hazards and hazardous situations (Clause 5.3), and c) the use
scenarios involved in the hazards (Clause 5.4). Figure 12.4 illustrates the gen-
eral relationship between the intended use, use specification, and the task
descriptions used in task analysis. In the figure, analysis of the intended use
of the device first leads to identified characteristics for the use of the device

Figure 12.4 Deriving use specifications and task descriptions from the intended use.
288 ◾ Medical-Grade Software Development

(defined as a specification and individual tasks) and may then lead to the
identification of benefits and risks for the device.
Task analysis is a good method for developing task descriptions. This
analysis does not just develop a valuable picture of the user and their con-
text, but it also facilitates the subsequent development of an appropriate user
interface design and lays the foundations for a solid analysis of risks and
benefits around the use of the device. Task analysis thus builds a view of
what the user groups for the device are, how they interact with the device,
and what features they will use. This may also include the identification of
physical and social elements of the environment as well as its soundscape, if
relevant.
After appropriate modeling of the use scenarios, the scenarios now ear-
marked for inclusion in the final summative evaluation are selected before
moving into the initial formative evaluation phase (Clause 5.5). This selec-
tion may include all of the scenarios or just a subset based on some sound,
documented criteria (e.g., those scenarios which can lead to a user error and
harm). Next, the corresponding user interface specification and evaluation
plan are developed, the user interface designed and implemented, and the
initial formative evaluation performed (Clause 5.8). Finally, the ultimate sum-
mative evaluation is performed (Clause 5.9).
The two evaluations performed, the formative and the summative evalua-
tion, may be singular activities or consist of several activities each. Formative
evaluation refers to the assessment done while working on the user inter-
face design. The objective here is to identify the strengths and weaknesses
of the design, as well as any unexpected use errors. In all likelihood, this
is an ongoing iterative process that takes place throughout the development
phase. Summative evaluation, on the other hand, takes place after the user
interface has been developed. The aim here is to create objective evidence
on the safe use of the interface. This final summative evaluation must also
identify any potential difficulties and near-miss situations if these may result
in a use error and harm. All of the activities here must be documented, and
the relevant documentation linked to the technical documentation of the
device. If the summative evaluation concludes that issues exist and redesign
activities are required, summative evaluation will be appropriately updated
after these activities are ready.
In practice, it is possible to meet the usability-engineering expectations
in one of two ways. Firstly, if the device in question is new, the user inter-
face and the user interaction around it are assessed in accordance with the
body of IEC 62366-1 (see Clauses 5.1 and 5.2). Secondly, if the device has
Seeing into the Future ◾ 289

already been on the market for quite some time, the concept of usability
of unknown provenance (Clause 5.10) may be employed as described in
Appendix C of IEC 62366-1. In the latter case, the preparation of the use
specification is followed by a review of available post-production informa-
tion, and the assessment of the hazards, hazardous situations, risk analysis,
risk control verification, and benefit–risk analysis based on all of the avail-
able information.
Regardless of the chosen path to conformity with the standard, revision
of the manufacturer’s standard operating procedures is likely to be needed
to ensure appropriate attention to usability engineering. Addressing usability
will lead to specific activities in requirements specification and design, risk
management, and documentation (incl. the technical documentation of a
device). Usability may also add considerations for compiling the accompany-
ing documentation for a device as this affects, for example, the installation,
use, and maintenance of the device. By the same token, having a clearly
instructed usability engineering process may sharpen and even streamline
the design process as, for example, task analysis is used to model user inter-
action with the device in a more structured way. This way it may be both
faster and safer to develop the correct requirements and designs.
Usability engineering is a fundamental activity with implications for sev-
eral of the software development life-cycle phases discussed elsewhere in
this book. The magnitude of this impact is perhaps self-evident when we
consider that usability issues will factor into at least the requirements speci-
fication (Section 5.2), design (Sections 5.3 and 5.4), and risk-management
(Section 7) activities performed during software development – and really
should be fused into the instruction of the processes and the planning of
the product development project itself. Similarly, usability engineering has
clear links with post-market surveillance and change management activities
as well.
Usability engineering, however, should not be seen as something extra to
do. The fundamental task of usability engineering is to understand the user,
how they will interact with the device, and for what purpose. As a goal,
this is 100% in line with what ISO 13485 and IEC 62304 expect to see the
manufacturer strive towards. Usability engineering methods can, in fact, help
us model the use context better, smarter, safer, and faster than what might
otherwise be possible. Methods such as task analysis may become the key
solutions to understanding not only the risks but also the benefits surround-
ing the use of our software in the real environment. As such, these methods
may give us real rigor in answering not just the risk assessment but also
290 ◾ Medical-Grade Software Development

the benefit assessment required of us as manufacturers and making the ISO


14971 benefit–risk ratio assessment in a solid, defensible manner. Usability is
not a tick-box activity, not really – not when it can offer us real insight into
the way our products should work and how they are in fact used out there
in the real world.

12.6 Cybersecurity
Johnny Mnemonic showed us a nihilistic future where cybercrime
had taken us to extreme measures in protecting information.
Minority Report took us to another extreme where crime was abol-
ished by treading on the privacy and free will of people by jump-
ing on a crime before the gun was ever picked up. Neither of these
futures has yet become a reality, but the topic of cybersecurity con-
tinues to both fascinate and scare us. For medical devices the stakes
are high, and the effects are possibly immediate.
Cybersecurity is one of the hottest topics in medical devices today and per-
haps only second to the use of artificial intelligence. In the short term cyber-
security will probably have a much greater impact on medical devices than
artificial intelligence as it touches on just about all medical-device software
whereas artificial intelligence is still in many ways more of a promise than a
practice. Both are, however, shaping our environment today and will con-
tinue to do so to increasing effects in the future.
Cybersecurity for a standalone, disconnected device may be a non-issue,
but when that same device is connected to a network – be it a local network
or the world-wide-web – the risk of cyber-related issues becomes a distinct
possibility. These in turn lead to new requirements being imposed on, for
example, the design, implementation, testing, accompanying information,
and maintenance of that device. Cybersecurity, much like usability engineer-
ing, is not an optional extra to the development of your medical device. It is
part of the picture for developing that medical device in a way that allows
you to rely on the benefits and risks assessment you have made and the
value promise you make to your customers.
Risks arising out of compromised cybersecurity may be diverse and range
from financial risks (e.g., damage to the brand), to privacy risks (e.g., loss
of sensitive patient data), to possible patient-safety risks (e.g., hacked pace-
maker software). All of these risks are serious and may lead to very measur-
able harm coming to both people and property, but out of all the risks, the
Seeing into the Future ◾ 291

most disconcerting ones regard situations where serious, life-threatening


harm may rapidly come to patients. These are cybersecurity problems
leading to the loss of some critical systems during their use, for example,
a BitLocker-lock of some control software during a surgical operation, or
breakdowns in life support devices at hospital ICUs. In these environments,
a vulnerability may affect an obvious, essential system that has direct influ-
ence over the life or death of patients (e.g., respirators, vital signs monitors),
or it may concern a less-obvious support system where breakdown may only
cause annoyances or disruptions in the environment (e.g., the unavailability
of some patient records), but which may end up being fatal due to the very
nature of the critical environment.
If even a small cybersecurity vulnerability can be exploited for great loss,
and if some cybersecurity issues may even be unfortunate accidents instead
of malicious attacks (e.g., congestion and interference caused by general-
purpose IoT devices in a medical environment), it becomes self-evident that
cybersecurity is a very real topic to address. The goal here should be one
of prioritizing risks, of course, but also preventing all cybersecurity issues to
the fullest extent possible. Small fissures may become large crevasses over
time, particularly when some malicious intent drives the raptors to try the
cages.

12.6.1 Cybersecurity Over Product Life Cycle


In practice, addressing cybersecurity calls for actions during the design, pro-
duction, and maintenance of devices – decommissioning too unless we want
to see troves of sensitive information end up in landfills, which has been
known to happen.
During design, all relevant sources of information including the prevalent
state of the art are considered in defining the requirements for the device
and designing both the device and its maintenance over the coming years.
The developing software architecture and the technologies used should be
analyzed from the point of view of safety and security from an early point
on as the different choices made here (e.g., defensive design) set the scene
for what is to come later. Techniques such as penetration testing and threat
analysis may be useful here, and the basic risk-management framework
instructed by ISO 14971 is a must. Clause 7 of ISO 13485 will do its part to
ensure a sound design is reached, but also standards such as IEC 82304-1
(see Section 12.2) may be applicable based on the type of the device.
Understanding the user, the patient, and any operating environment will be
292 ◾ Medical-Grade Software Development

crucial here, and usability engineering may provide a part of the necessary
picture (see Section 12.5). The term cybersecurity may not always leap at
you from the standards and regulations written on medical devices, but you
can expect it to always be a part of any consideration of risks around medi-
cal devices.
The production phase of the device should be simpler to handle than
the design phase, but from here to the basic cleanliness requirements set
in ISO 13485 should be interpreted as a call to ensure the software ships
without malware and viruses. This may be as easy as cloning a binary or
as complicated as assessing all of the new versions of the constituent SOUP
components. Note also that the installation phase of a device may see dras-
tic changes in expected settings if the configuration is allowed. Some criti-
cal security settings may be relaxed (e.g., unnecessary ports left open) just
because it is easier for the installer to get it all working that way. Similarly,
once something is up and running, it may be all too easy to forget about
watching over it or tending to security updates later. All of this may, at least
in part, be countered with the appropriate accompanying information and
training provided to the user along with simple mechanisms for checking on
the availability of updates.
That brings us to the next phase in the life cycle of software: mainte-
nance. There is no discernible wear and tear involved in software, yet main-
tenance is just as important as for physical devices whenever that software is
exposed to the elements of a network environment. In an environment that
changes constantly the software itself may need to change to stay up-to-date
and safe. Maintenance here refers to, first and foremost, keeping the device
safe by reacting to issues quickly and providing reliable updates to the
device. Providing feature updates to devices only comes second, although
a manufacturer might internally view maintenance work from exactly the
opposite frame of mind: as something leading to new features, not shoring
up old features.
It may be tempting to approach cybersecurity as a later-stage activity
performed toward the completion of a piece of software. In some ways, this
is because the post-market phase is definitely implicated in cybersecurity
activities, and the closer we are to that phase when we perform the final
pre-release assessments the more up-to-date this work is likely to be when
the product ships. It may, however, be the case that some threats could have
already been thwarted in the requirements specification or architectural
design phase of the software. By the same token, issues discovered near
the launch may require adding new requirements, new test cases, and new
Seeing into the Future ◾ 293

verification or validation activities. If software development was a game of


snakes and ladders this would definitely be a vicious snake to encounter.
Instead, you should look to modularize your device and rely on already
completed cybersecurity assessments available for some of those modules as
the ladders to improving your responsiveness in tackling any newly detected
issues.
Reducing the level of complexity to some manageable level will likely be
key to maintaining cybersecurity activities on a practical level. This means
modularizing your software in some smart way so that you can reuse mod-
ules and tend to their maintenance on different clock cycles as needed. This
means utilizing your resources so that threat modeling for your device, for
example, is not your only source of information on current cyber threats in
your environment. Cybersecurity will be part of your post-market activities
going forward, and like those post-market activities, it should not just be up
to the original development team to keep abreast of all relevant develop-
ments. Some tasks here will be up to the engineers, but some other tasks
are more akin to market intelligence activities.
Involving a sufficiently broad base of experts with appropriate back-
grounds may also be important for spotting cybersecurity threats surround-
ing your software. Expertise in the clinical setting, usability engineering,
software engineering, risk management, and cybersecurity management may
all be called for at various stages during the life cycle of the device.

12.6.2 Cybersecurity Across Jurisdictions


Cybersecurity has been a hotbed of activity around the world over the
past few years. A wealth of thought has been published on the topic by,
for example, the FDA in the US, the Medical Device Coordination Group
(MDCG), and the joint organization of European notified bodies, Team-NB.
All of this published information, as well as the underlying regulations them-
selves, will be of interest to you as you look to address cybersecurity in the
jurisdictions. In Europe, the new Cybersecurity Act, too, will demand atten-
tion on this front.
The international standard on risk management for medical devices,
ISO 14971, has already seen the elevation of data and systems security
to a level on par with the human-centric safety focus it had previously.
National guidance and proposed rules are following suit and beginning to
set direct requirements for medical-device manufacturers. It should also be
noted that regulations beyond those directly discussing medical devices and
294 ◾ Medical-Grade Software Development

cybersecurity issues may add related requirements. A case in point here is


given by the EU GDPR and the US HIPAA legislation.
The good news here is that international standards are seen as facilitators
of global harmonization. A position paper on cybersecurity by Team-NB, for
example, refers to IEC TR 60601-4-5 for determining product requirements and
IEC 81001-5-1 for both cybersecurity-related product requirements and life-cycle
activities. The former document discusses both hardware and software to pro-
vide further guidance on the design of medical devices for use in medical IT
networks while the latter standard is fast becoming a reference for addressing
cybersecurity. Let’s take a closer look at the latter standard in the following.

12.6.3 IEC 81001-5-1
The IEC 81001-5-1 standard on security activities over the life cycle of health
software is currently gathering momentum as a reference for not just health
devices but also medical devices. The 50-page document sets requirements
for both the development and maintenance of health software but does not
speak to the contents of the accompanying documentation for a device. The
standard does expect all activity to take place under a quality management
system implemented according to ISO 13485 (or equivalent).
IEC 81001-5-1 has been organized so that it matches with the clause
structure of IEC 62304 as closely as possible, but it does not conform to the
same use of the software safety classification. The requirements given by
IEC 81001-5-1 are said to apply to all classes of software.
The impact of IEC 81001-5-1 on the IEC 62304 development of software
depends on the type of product in question and the extent to which you
have already addressed cybersecurity in your software development pro-
cesses. In general, the standard does not revolutionize the approach to soft-
ware development, but it does act as a good reference calling attention to
sources of cybersecurity risks, and the appropriate way to handle these.
For the software-development process (see Section 5), the standard per-
forms a comprehensive sweep touching on the determination and review of
requirements, the development of the software design, the performed testing
(incl. threat mitigation testing, vulnerability testing, and penetration testing),
release, and decommissioning. Interestingly the standard also specifically
speaks of managing conflicts between testers and developers, ensuring file
integrity, and controlling private keys, among other security-related topics.
For the software-maintenance process (see Section 6), the standard
describes its own expectations for the process, including the timely delivery
Seeing into the Future ◾ 295

of security updates to the software. On the agenda are also, for example,
monitoring public incident reports, update verification, update documenta-
tion, and update delivery. Note that in terms of documentation, the instruc-
tion here is brief and in line with what IEC 82304-1 expects.
For the risk-management process (see Section 7), the standard naturally
brings a list of new risks to the table. The effects here may be profound
if cybersecurity was not already a part of the picture for risk-management
activities based on ISO 14971. Here the standard attempts to help you iden-
tify, estimate, control, and monitor the security-related risks surrounding
your software. The discussion is written with a view to pointing out relevant
aspects, but it does not go into too much detail on how to address these in
a systematic manner. Out of all the sections of the standard this may be the
clause with the biggest impact on your activities due to the omnipresence
and importance of risk management in medical-device development, par-
ticularly for a network-connected software-based device.
For the configuration-management process (see Section 8), the discussion
in the standard is light. The requirements set here boil down to the abil-
ity to produce a list of external components used to the extent these could
become susceptible to vulnerabilities. The discussion here is only five lines
long and in line with the expectations already given by other standards.
Finally, for the problem-resolution process, the standard wants to see you
anticipate receiving notifications on vulnerabilities, reviewing and analyzing
these, and addressing any issues accordingly.
Integrating all of the requirements of the IEC 81001-5-1 standard into the
various standard operating procedures of your QMS is not going to happen
with the snap of your fingers. The good news here is that you may not need
to claim conformity with the standard as such, but instead you are free to
use the standard to harden your approach to cybersecurity issues. You may
end up taking a piecewise approach to the standard where you first identify
it as relevant guidance, perform a gap analysis comparing your existing pro-
cesses to the expectations of the standard, and only then attempt to write in
the exact requirements. The requirements given by the standard do act as
checklists that you can use in your software development activities.

12.6.4 ISO/IEC 27001
In the above, we mostly discussed product-related security, and did so
with a definite affinity for patient safety. In this context, information secu-
rity exerted most of its weight via possible harm to people, even if the
296 ◾ Medical-Grade Software Development

immediate object in the crosshairs was information and not human beings.
If we are willing to broaden our scope from the medical-device focus on
patient safety and look at the management of information security on an
organizational level, we may also consider the ISO/IEC 27001 standard on
information security, cybersecurity, and privacy protection.
This standard does not replace our ISO 134865 standard on quality-manage-
ment systems or any of the other key standards for medical devices discussed
above, but it may provide elements to incorporate on the organizational level to
ensure cybersecurity and information security are adequately addressed.
The point here is that a single supreme developer with an exceptional
talent for producing safe, secure code will be wasted in the war on cyber
threats if the whole company isn’t ready to do its part. Other developers
and staff acting on their own may sink the ship unless the organization
itself takes a firm hold of cybersecurity and sees to the relevant resources
and guidance being in place, and the staff being adequately trained for
their roles in security. The call to have appropriate training programs and
to demonstrate security expertise for personnel appropriate to their roles is
also echoed by IEC 81001-5-1 in its Clause 4.1.4. Ideally, this training and
the documentation of education, skills, and expertise should be part of your
overall personnel management in your QMS (see Juuso 2022).
It is similarly important to acknowledge the role of cybersecurity and
information security in the design activities for your software. It all starts
with having a good security policy for the organization, having appropri-
ate SOPs in place, and having a solid, practical organizational plan for how
the various security processes are assigned at the organization. Ensuring
adequate staff competence for the activities expected of them is just as
important under ISO 27001 as it is under ISO 13485. Having solid, secure
organizational foundations here will give you the best launchpad for creat-
ing secure software, training, or instructing your users as is appropriate, and
ensuring also the long-term maintenance of your software is on a secure
footing. As with the other parts of your QMS, expect cybersecurity to also
be a part of your internal auditing activities going forward, even if you don’t
go for separate certifications according to ISO 27001.

12.7 Artificial Intelligence
Artificial intelligence and the multitude of technologies this
label encompasses act as the modern-day touchstone for any
Seeing into the Future ◾ 297

software-based medical device. Software that doesn’t brandish


some form of an AI/ML flag will seem old-fashioned. The hopes and
dreams of the entire industry seem to be loaded onto this band-
wagon, but it remains a subject of keen interest and much debate
about what sort of a driver’s license should be required of the
devices at the helm.

Great effort and considerable hype presently float around any mention of
an AI-enabled medical device. Both regulators around the world and the
big international standardization organizations are scratching their heads to
come up with the best practices to develop, apply, maintain, and regulate AI
technologies. New regulations are drafted, existing work in standardization
scoured, and new thinking is published in the form of position papers, draft
guidance, and white papers. To be honest, an overwhelming amount of
work is presently taking place and the sheer volume of the output from this,
as well as the occasionally unharmonious terminology used, is fast becom-
ing a mountain to scale. Information overload and the problem of separat-
ing the wheat from the chaff or understanding how the different schools of
thought agree or disagree, are constant issues in this fertile field.
Given the above, it is somewhat encouraging to learn that the present
legislation and standards have built a solid foundation against which we can
evaluate each new AI product proposed and assess its benefits and risks to
patients. One of the cornerstones of this foundation is set by the Declaration
of Helsinki from 1964. The declaration developed by the World Medical
Association (WMA) as a statement of ethical principles for medicine involv-
ing human subjects is a brief 37-point document. The document primarily
looks at medical research but ends up setting the scene for how physicians
are to address the best interest of patients, especially in the context of inves-
tigating new treatments or interventions.
The layer of global legislation written on top of this declaration appears
to adhere to these stipulations and objectives. The call for safe drugs has
been around since at least 1906 when the Pure Food and Drug Act was
passed in the US as a response to encountered fatal issues in substances
available to people. In 1938 this was extended to also address medical
devices, and a few decades later efficacy of the devices became a part of
the picture. In Europe, the story is the same, and the advent of the new
European Medical Device Regulation (EU 2017/745) was meant as a new step
along this path to regulating the constantly evolving and increasingly com-
plex devices. The great news here is that we already know what the large
298 ◾ Medical-Grade Software Development

goals are in both safety and efficacy, and addressing the best interests of
patients, even if we don’t yet understand what AI technologies, best prac-
tices, and exact configuration items these correspond to. The same is true
for the key medical-device standards we have today: they may not speak of
artificial intelligence per se, but they show us the technology-agnostic goal-
posts to compare our devices and operations against.
Many organizations around the world are working on translating the gen-
eral goal posts to concrete design choices and control options for artificial
intelligence. Everyone and their uncles want the benefits AI can afford us,
but no one wants to flirt with the risks that may also come at the same time.
For this reason, it is understandable that much of the conversation in a regu-
latory or clinical context at the moment concentrates on the topic of trans-
parency, i.e., understanding what it is that AI might be telling us when it is
telling us something. On the highest level, this conversation can be distilled
into a comparison between a black-box and a white-box approach to AI, as
illustrated in Figure 12.5.
In the figure, a nihilistic example of two AI devices is given. One of the
AI models is a black box, meaning it has learned some patterns by itself
and we don’t actually know what phenomena it bases its judgment on.

Figure 12.5 The difference between black-box and white-box AI.


Seeing into the Future ◾ 299

The second AI model is a white box, indicating that we do have an under-


standing of how that model works. In the figure, both models give us the
same answer to the meaning of life as given in the Hitchhiker’s Guide to the
Galaxy, but only with the white-box model do we understand what led to
that particular answer.
The figure appears to frown upon the idea of using a black-box model,
and many regulators have taken a very stern view towards such models in
the past but given large amounts of high-quality data it is a feasible model
for AI. Techniques also exist to lighten the black box into some shade of
gray so that we could understand what is happening inside the box. At a
recent FDA event, the agency gave a number of cautionary examples of
using black-box models where, for example, an AI had latched onto the
doctor’s annotations made on radiological images to detect the phenomena
already labeled by the doctor. Needless to say, the model would have been
utterly useless without the doctor first doing the work. Nonetheless, AI holds
great promise for the future of medical devices and the health of patients.
The work on enabling artificial intelligence applications in healthcare is
rooted in the ongoing work to regulate increasingly complex software in
the field of medical devices. The newly adopted European Medical Device
Regulation (EU 2017/745), the EU Artificial Intelligence Act currently tak-
ing shape, and programs such as the Pre-Cert initiative run by the US Food
and Drug Administration (FDA) all look at the intersection of software and
healthcare with a keen eye towards the future. At the same time, cloud-
based products and services are increasing their foothold in the sector, par-
ticularly in the domain of health and wellness applications where the path
to user adoption is generally not hindered by considerations such as instal-
lation and maintenance inside hospital firewalls. These connected platforms
will improve the availability of health data and provide a fertile field for the
application of AI to tackle health issues of many sorts.
It is noteworthy that not all software is a medical device, and not all soft-
ware modules in a medical device are to be considered medical in nature,
either. The intended use specified for the device by the manufacturer is the
key to assessing whether a piece of software, or some part of it, is a medical
device or not. In Europe, the Medical Device Coordination Group (MDCG)
of the European Commission has published helpful guidance on how to
classify software (see MDCG 2019-11). According to the guidance, e.g.,
software used for the storage and transfer of medical data is not a medical
device unless it alters the data somehow. Conversely, any analysis performed
for a medical purpose will likely lead to designation as a medical device.
300 ◾ Medical-Grade Software Development

When the product or service is to be considered a medical device, a num-


ber of international standards come into play to govern the development of
those devices and the organizations behind the manufacturing and distribu-
tion of the devices. Chief among the standards are the ISO 13485 standard
on quality management, ISO 14971 on risk management in medical devices
and the IEC 62304 standard on software life-cycle processes, all of which
we have discussed in this book. These three standards form the basis of
much of the work behind developing medical devices, but also other stan-
dards factor in to address clinical evaluation and usability engineering, for
example. The key standards are stable at the moment, but due to the great
interest in AI and other emerging technologies that are shaping the future of
software in healthcare it can be expected to evolve over the next 5–10 years.
If artificial intelligence plays a part in your future plans, as it is likely to do
sooner or later, this is all development to watch.

12.8 Cloud Computing
The iconography for cloud services invokes vistas of perfect blue
skies and bubbly cumulus clouds. An equally apt metaphor might be
Spider-Man laying on an interconnected patchwork of spider webs
stretched out across the sewer system of New York City, but somehow
cloak-computing doesn’t quite have the same ring to it. Nonetheless,
clouds are here to stay every bit as much as Stan Lee’s creation.
All forms of cloud computing, including the proliferation of so-called
Software-as-a-Service (SaaS) tools and services, have taken over the world
of computation. No one remembers the mail-order software catalogs used
for software delivery in the decades past – these days a floppy disk is only
really good as a coaster for the cold drink on the programmer’s desk. If that.
There are still critical use cases for non-cloud software, of course, but even
in the world of medical computation, cloud computing has come to stay.
The cloud is used for storing data, performing calculations, and building
bridges between what used to be distant silos of creating and using data
across industrial domains. The cloud also scales to increasing stress from
growing pools of users and the ever-new features requested in a way that
puts personal workstations to shame. Adding new resources to a cloud ser-
vice doesn’t usually require a screwdriver and is occasionally as quick to do
as the click of a mouse. It is thus not hard to see why cloud computing is an
attractive proposition.
Seeing into the Future ◾ 301

Cloud computing combined with an Internet of Things (IoT) mindset


means that physical devices are mostly just sensors or drill bits pumping out
data which is then streamed to the cloud for processing. The link to the cloud
can reside within the sensor, or more often for now, with a mobile device, tab-
let, or laptop computer relaying the data from the bedside to the “bitside”. All
of the user’s interaction with the data then takes place on the cloud.
The computational prowess available on the cloud for a fee, the plethora
of available solutions for data management, and even the ready-made librar-
ies of analysis algorithms and tools all mean the temptation to migrate to
the cloud is great. There are no free lunches and even AI can be a mirage
of sorts, but the promise of a plug-n-play environment for analyzing patient
data and investigating the appropriate diagnoses is a tantalizing prospect. To
some extent the cloud delivers here: once all of the data flows to a unified
repository and queries are put at the fingertips of the users, there is little
reason to doubt that new knowledge would not be distilled. The code might
work the same on a personal workstation, but the data probably would not
find its way there, and so the proposition of the cloud is quite different from
the standalone. The potential here is limitless.
The cloud can be used as a solution to expedite software development,
delivery, and adoption as long as we observe the following caveats:

◾ Risk management must keep up with the particulars of the new envi-
ronment and the new features it enables – the old risk analysis con-
ducted for a standalone solution does not hold water when the solution
is taken to the cloud
◾ The physical location of the data and its transmission path matter
greatly and are usually subject to several different regulations – redun-
dancy and load distribution may have geographic dimensions that also
need to be addressed
◾ The stacks of software components living on the cloud don’t just give
you neat features out of the box but also create expectations for appro-
priately addressing, e.g., SOUP validation and platform updates
◾ The user of network functionality must be addressed in the Instruction
for Use (IFU) as per IEC 62304, IEC 82304-1, and the applicable
regulations
◾ Defining what is critical performance for the device in a worst-case sce-
nario will get to be increasingly difficult as the number of moving parts
involved increases and the parts are on possibly differing update cycles
– you may need to interact with the suppliers of each component
302 ◾ Medical-Grade Software Development

◾ The needs to appropriately address cybersecurity and information secu-


rity are only heightened when you use the cloud to process and store
your crown jewels

In addressing the above, you will find that your architectural design is a key
piece of the information you’ll need and also the context where you will be
looking at your risk-management activities repeatedly. As you’d expect, if
you are building a critical medical-device service on the cloud, even when
using all the usual suspects for technology, your task will be much more
complex than following the typical assembling instructions for a flatpack
piece of furniture. One way of looking at the cloud is that it brings new
types of interactions into the mix, and possibly opens the door to new mali-
cious and inadvertent sources of harm. The man in the middle won’t be
your customer, but they might be one unwanted type of user to address in
your plans.
There are two sure ways of failing in the adoption of cloud dimensions
for your medical software. Firstly, you might fail by thinking that whatever
worked in the standalone world would be enough. Secondly, you might be
discouraged by the increased complexity and not look deep enough into
what the cloud could mean for your application. Don’t fall victim to either
bear trap, or you’ll miss out on the opportunities that cloud computing can
bring.

12.9 Edge Intelligence
When Tom Cruise finds himself at the edge of tomorrow, he has a
bit of a groundhog day, but ultimately cracks the code and beats
the artificial intelligence at its own game. That is the vague promise
we see in a technology called edge intelligence – not that it would
dethrone AI, but that it might offer an interesting methodology to
consider in the context of medical AI.
Edge intelligence, or Edge AI, refers to a brand of artificial intelligence
where the intelligence is not located in a standalone device or in the cloud,
but on the edge of the network. Many of the same concepts apply to Edge
AI as to cloud computing, edge computing, and artificial intelligence. Yet,
the technology has particular appeal in the context of medical devices
where the dataset can be debilitatingly large, and both the data and the
Seeing into the Future ◾ 303

proprietary algorithms may need protecting – as well as the human lives


connected to the algorithms.
Work here is rooted in the larger cross-disciplinary investigations into
enabling AI in healthcare and is also related to the regulation of increas-
ingly complex software in the domain of medical devices. The EU MDR, EU
Artificial Intelligence Act, and programs such as FDA’s Pre-Cert initiative all
have a bearing here. In this context, edge intelligence could offer a novel
practical solution to overcoming traditional issues faced by cloud solutions
(e.g., issues of moving large amounts of sensitive data to servers outside the
hospitals, or alternatively installing and maintaining cloud installations inside
hospital firewalls) and stand-alone software solutions (e.g., protection of
proprietary software onsite, limitations on computational power) by offering
a new low-latency and high-power environment for computation with a high
scalability factor. The edge-intelligence environment could offer an appealing
single-use environment, the software equivalent to a cleanroom, that allows
the high-volume data, the proprietary analytics algorithms, and the doctor to
meet in an environment that not only minimizes data transfer requirements
but also does so in an inherently cybersecure way due to its transient on-
demand nature.
As a flavor of AI and an exception to the now-traditional classification of
devices along an axis from standalone to cloud, Edge AI may prove to be an
interesting environment to consider in designing medical software. In this
book, we bring the technology up as food for thought, as an outlier perhaps
challenging our notions of what it means to be online or offline.
Chapter 13

CONFORMITY ASSESSMENT

After all the hard work of developing software and deciphering the
applicable requirements, the definitive test for your software may
come in the form of an official IEC 62304 conformity assessment.
You may think of this as the redcoats and the bluecoats assembled
across a misty field at dawn, it may feel like that after all the effort
expended by both sides in preparation, but the point of the assess-
ment is to ensure your development process and its deliverables
meet the expectations. Unlike at a battlefield, the expectation here
is not that you will die, but the goal is that no one else will suffer
either.
As remarked earlier in this book, the role of software in medical devices has
grown constantly over the past few decades. Not only is it increasingly com-
mon to find some form of software in many medical devices, but also the
type of functionality enabled by that software and the role of such function-
ality plays in forming medical diagnoses and guiding treatment is continu-
ally growing. This trend is only set to continue as increasingly sophisticated
technologies, such as advanced machine learning and artificial intelligence,
are brought to bear on the medical issues we face.
The result of this development is that software is increasingly the subject
of scrutiny both in terms of safety and in terms of conformity to applicable
requirements. Both of these are assessed against the context set by the
intended use of your device, particularly if this involves the introduction of
some new functionality into the inherently complex and diverse care envi-
ronment. This heightened focus will also invariably lead to the addition of
new software-specific requirements into both medical-device regulations and

304 DOI: 10.4324/9781003454830-13


Conformity Assessment ◾ 305

standards. Some of the requirements will speak directly to software and oth-
ers will indirectly affect the development, maintenance, and use of software
– or the management of the organizations involved (e.g., ISO 13485, ISO/IEC
27001).
The translation of general medical-device requirements to the specific
case of software, and the interpretation of these requirements in terms relat-
able to software and readily understood in the context of software develop-
ment, is already a challenge. The complexity of software, the distributed
nature of some software, and the effects software can have on other devices
over interfaces and network connections often require in-depth analysis to
assess. This is particularly the case when instead of a single piece of soft-
ware the whole consists of several software systems spread over multiple
servers, the interaction of which may vary. This may be akin to a surgeon’s
scalpel not having a blade to cut with during some denial-of-service (DoS)
attack, but it may also be the difference between having a dull rusty blade
or a pristine sharp blade as IT enables ever newer advances to take place.
With great power comes great responsibility, as one webby superhero of a
comic-book movie might say. That responsibility weighs on the manufac-
turer, but it also weighs on the assessor.
Note that conformity with ISO 13485 and the regulatory requirements
related to it are usually the topic of another conformity assessment, such as
the assessment preceding CE-marking and the release of a medical device.

13.1 Requirements Placed on the Assessor


The assessment of the safety, performance, and regulatory conformity of
software is often a markedly more complex task than in the case of some
simple physical what-you-see-is-what-you-get medical device. In such situ-
ations, the assessor is expected to have a good command of standards and
regulations but to also be in a position to interpret the requirements based
on their experience with similar devices and knowledge of commonly
accepted software technologies. In many cases, the assessment calls for a
multidisciplinary skillset to weigh several factors at the same time and to
be able to identify relevant design features for close scrutiny. Occasionally
it may even be necessary to form assessment teams to tap into all of the
required skillsets.
The assessment of software thus calls for the following needs to be ful-
filled by the assessor or the assessment team:
306 ◾ Medical-Grade Software Development

◾ Appreciation of the regulatory requirements that are applicable to the


software in question (e.g., with respect to EU MDR, harmonized stan-
dards, the concept of the State of the Art, IEC 60601-1 Clause 14, IEC
62304, and IEC 82304-1)
◾ Appreciation of risk-management processes and documentation which
form a cornerstone of achieving the necessary level of safety and per-
formance for a device (e.g., in respect to EU MDR, IEC 60601-1 Clause
14, and IEC 62304)
◾ Appreciation of software-development life cycle models (e.g., the clas-
sic waterfall model, agile models) and the deliverables produced via the
individual phases of these models (e.g., requirements analysis, verifica-
tions, test cases, and review processes)
◾ Appreciation of software technologies including those related to pro-
gramming languages, software architectures, and cybersecurity models

All of the above are naturally also a challenge for the manufacturer of the
software. For each of the above aspects, the manufacturer must be able to
point to an activity, a process instruction, or result document, and the accep-
tance criteria according to which the software may be judged to be in con-
formity with the relevant requirements.
The assessment of software is also impacted by the networked architec-
ture of some of the current medical-software devices. Whether the software
and its host device form a standalone device or are part of a larger network
of connected devices and software each contributing to the intended use
somehow are markedly different situations. Understanding all of the mov-
ing parts in the latter, and their role in the clinical environment for meet-
ing the intended use may become quite complex. It was once enough to
understand hardware and firmware, but now that is no longer necessarily
the case.
In case of assessing medical-device software that is not connected to a
network the following is demanded of a competent assessor:

◾ Ability to identify the applicable regulations and standards, and knowl-


edge of these
◾ Ability to identify the appropriate assessment criteria (e.g., IEC 60601-1
Clause 14, IEC 62304), and knowledge in applying the criteria to real-
world cases
◾ Knowledge of software-development life cycles and the deliverables
involved at each stage
Conformity Assessment ◾ 307

◾ Knowledge of software development and the ability to read code


◾ Experience in software architectures and requirements (e.g., IEC 60601-1
Clause 14.7, IEC 62304 Clause 5.2.2, and possibly even IEEE 830)
◾ Ability to understand documentation requirements and traceability in
the context of risk control or mitigation (e.g., IEC 60601-1 Clauses 14.7,
14.10, IEC 62304 Clauses 5.2.3, 5.2.6, 7.2.2)
◾ Experience in working with software tools (e.g., interpreters, libraries,
requirement-management software, and integrated development envi-
ronments of various sorts)
◾ Command of essential concepts such as basic safety, essential perfor-
mance, and risk control measure, and their application to software
assessment
◾ Command of ISO 14971 risk-management principles and requirements,
and practical experience in their use in assessments

If the device to be assessed is connected to a network the assessor must


also meet at least the following requirements:

◾ Expertise in cybersecurity and information security principles, solutions,


and technologies (note that several product standards already make
requirements on security issues)
◾ Expertise in usability engineering and risk management in a complex
network of human–computer interaction
◾ Expertise in network topology and its impact on safety and
performance
◾ Knowledge of cloud technologies including the typical differences in
their implementation (e.g., private cloud, public cloud, hybrid cloud)
and the risk scenarios involved in each
◾ Command of a wider set of programming languages and programming
tools in simultaneous use

Based on the above it is apparent that the level of complexity increases sig-
nificantly as we move from the assessment of a single standalone device to
a networked device consisting of diverse components each perhaps working
under different parameters. As the intended behavior becomes more elabo-
rate, also the sheer number of software components used in a device often
increases. The demands placed on the assessor’s command of risk manage-
ment, in particular, are heightened at the same time (see IEC 60601-1 Clause
14.6 and IEC 62304 Clauses 4.2 and 7).
308 ◾ Medical-Grade Software Development

13.2 The Assessment
The IEC 62304 assessment may take several forms, but the objective is
always the same: assess how and to what extent a manufacturer’s software-
development process and documentation meet the requirements of the
standard. Based on our personal experience we have found two alternative
models of conducting the assessment to work well in practice.
The first model is a typical audit-type assessment. Here the assessor goes
through all the requirements one by one and assesses how the documenta-
tion answers to the requirements of the standard while also observing any
applicable regulatory expectations. Any observations are recorded using an
auditing form or the IEC 62304 test report template.
The second option is to ask the manufacturer to prefill a test report form
answering the same questions and then provide this information as well
as any invoked documentation to the assessor. The assessor will then go
through the answers and evidence to provide the final conformity assess-
ment. In practice, this model has often been the most cost-effective imple-
mentation for the assessment as the correct documentation is identified
ahead of time by the manufacturer and a large portion, if not all, of the
assessment work may be performed offsite.
The choice between the assessment model is up to the manufacturer and
the assessor. Experience has shown that some form of a preliminary assess-
ment is often a practical step to take. During this preliminary assessment
the manufacturer and the assessor both get to calibrate their expectations–
the manufacturer regarding what documentation is in fact needed, and the
assessor regarding the manufacturer’s overall setup and the status of pro-
cesses and documentation. When the manufacturer’s set of documentation
is assessed to all be there ready and waiting, the assessment itself is begun.
Occasionally the preliminary assessment may lead to the conclusion that the
documentation is so woefully lacking that no assessment should in fact be
performed, in which case both time and money are saved.
The tables below, Table 13.1 and Table 13.2, show excerpts from the
type of test reports filled in for medical devices during IEC 60601-1 and IEC
62304 assessments, respectively.
On the test reports, the first column references the clause of the stan-
dard, the second column contains an explanation of the requirements (note
that there may be multiple requirements under a single clause number), the
third column describes the manufacturer’s evidence for meeting the require-
ment, and the fourth column provides the final assessment of conformity to
Conformity Assessment ◾ 309

Table 13.1 Excerpt From IEC 60601-1 Clause 14.6 Test Report
Pass/
Clause Requirement Manufacturer’s evidence Fail
14.6 Risk management process
14.6.1 Manufacturer has considered See below. P
software- and hardware-
related HAZARDS when
incorporating PEMS into and
IT network.
The RISK MANAGEMENT FILE The RMF references specific P
(RMF) includes foreseeable hazards. The required items
HAZARDS, IT network appear in the risk analysis.
characteristics, 3rd party Real examples are viewed
components, and legacy from the product-specific
subsystems. (ISO 14971 Cl. 5.3) risk analysis files.

Table 13.2 Excerpt From IEC 62304 Clause 5.4 Test Report
Pass/
Clause Requirement Evidence Fail
5.4 Software detailed design
5.4.1 [B, C] The MANUFACTURER has The detailed design extends P
subdivided the software until it the architectural design
is represented by SOFTWARE down to the level of software
UNITS units. The detailed design is
given in document
DOC-341144.
5.4.2 [C] The MANUFACTURER has The detailed design meets P
documented the design with the expected level of detail.
adequate detail to allow The detailed design is given
correct implementation of in document DOC-341144.
SOFTWARE UNITs.

the requirement. Note that the final assessment might also be “N/A” if the
requirement is not applicable to the type of device in question.
The assessment is then done when the manufacturer has provided sat-
isfactory responses to all requirements. If the point of the assessment is
only to gauge the status of operations and identify points of improvement
some of the answers may be missing or fall short of the expectations of
the standard. Similarly, if the assessment is to be utilized in the notified
310 ◾ Medical-Grade Software Development

body context to ensure appropriate operations as the basis of CE-marking


a device some fails in certain requirements may be acceptable for the final
device conformity assessment to still proceed. This is a topic of conversa-
tion between the manufacturer and the notified body before the assessment
takes place. If, however, the objective of the assessment is to get a clean bill
of health in the certification body context, and under the CB scheme (see
Section 1.24), all the requirements must be passed for the assessment to go
through. Even a single “fail” on a requirement will torpedo the bill of health
for your software.
It is important to remember that the scope of the assessment, and the
extent of required documentation, depend on the specified software safety
classification (see Section 4.3). Here the ABC classification corresponds with
the risk profile of the software, from low (A) to high (C). The classification is
based on the risks involved, and the reasoning for a particular classification
must be proved using documentation – and this should also demonstrate the
incorporation of appropriate clinical expertise. Should the classification of
the device change in the future, a new assessment and new evidence may
be required.
After a passed assessment the assessor will provide a final test report
which is then useful as a bill of health for your software and software-devel-
opment processes according to the IEC 62304 standard. This report may be
of great interest to your notified body for moving ahead with the conformity
assessment of the device and your quality management system.

13.3 Typical Shortcomings in Assessments


Three of the most typical kinds of shortcomings encountered in IEC 62304
conformity assessments are:

◾ Inadequate definition of requirements


(IEC 62304 Clause 5.2.2, IEC 60601-1 Clause 14.7)
◾ Insufficient or too narrow risk analysis
(IEC 62304 Clause 7, IEC 60601-1 Clause 14.6)
◾ Insufficient verification and validation activities, lack of
evidence
(IEC 62304 Clause 5, IEC 60601-1 Clauses 14.10 and 14.11)
◾ Errors in change management over time
(IEC 62304 Clauses 8 and 9)
Conformity Assessment ◾ 311

Requirements elicitation is notoriously difficult and at the same time criti-


cal for the outcome of the software-development project (see Section 5.2).
Taking all of the relevant inputs into consideration from the standards,
regulations, and all of the involved stakeholders is no simple feat. Many
software project failures can be traced back to incomplete or incorrect
requirements elicitation. Thus, it is not surprising that requirements also
frequently come up in IEC 62304 assessments. Misinterpretations, overly
simplistic approaches, and flat-out blindness to meeting the require-
ments of the standard are the usual culprits here. Typical challenges here
include:

◾ Identification of the software version each requirement applies to


◾ Identification of the software platform each requirement applies to
◾ Determination of requirements regarding essential performance (see
below), usability, and risk management
◾ The interfaces and communication between different software systems
are not covered (this creates further issues for the validity of test cases
and the conformity to several IEC 62304 requirements)
◾ Incomplete, conflicted, or unrealistic system requirements are accepted,
which causes issues for deriving software requirements
◾ Incomplete, conflicted, or unrealistic software requirements are
accepted
◾ Planning documentation is insufficient and may lack a description of
the software-development life cycle
◾ Traceability between design documentation breaks down (IEC 62304
Clauses 5.1.1c, 5.3.6, 5.4.4, 5.7.4b, 7.3.3, 8.2.4)
◾ The operation of the software and its operating environment has not
been adequately described (this creates further issues for relying on the
derived requirements and subsequent deliverables)

Risk analysis, too, is an area where manufacturers may struggle. Identifying


risks may be as difficult for some manufacturers as identifying require-
ments, and the two are often related to one another. Both overly simplistic
and overly complicated risk-classification schemes are occasionally used,
and the linking of risk probabilities to real-world measurement data may be
problematic. What matters here most is that the identified matrix of risks is
sufficiently comprehensive, the prioritization of risks to put under the micro-
scope is sound, and that all risk control and follow-up actions are appropri-
ately seen to. Typical challenges here include:
312 ◾ Medical-Grade Software Development

◾ Risk management during software development is non-existent or lack-


ing, e.g., no risk management plan is available, risk identification is
done by the developer themselves in a vacuum, no risk-management
expertise is present, no clinical specialist or domain expert is involved
◾ Information security risks are inadequately addressed, particularly in the
context of network-connected or cloud-based devices

Verification and validation activities may also be a bar of soap with which
to grapple for some manufacturers. Here the principal issue appears to be
that not all designers understand the concepts of basic safety and essential
performance, especially when it comes to determining appropriate validation
criteria for software. Let’s thus take a moment to discuss these concepts.
Basic safety (IEC 60601-1 Clause 3.10) refers to those structural properties
of the device which are intended to prevent the device from becoming dan-
gerous under normal conditions and so-called single-fault conditions. In the
case of software, this translates to those aspects of the medical device which
involve software in, for example, monitoring segregation interfaces, handling
power management, and reacting to overheating issues.
Essential performance (IEC 60601-1 Clause 3.27; see Section 3.1.2) then
refers to the performance of some clinical function to a level below which
unacceptable risk might occur. The level may be specified as limits by the
manufacturer. In practice, the clinical function is often directly linked with
the intended use of the device. The collateral (IEC 60601-1-x) and particular
(IEC 60601/80601-2-x) standards of the IEC 60601 family of standards place
direct requirements on essential performance. These requirements may, for
example, address requirements analysis, design, test-case definition, verifica-
tion, validation, and risk management. For software, essential performance
often translates to the ability of the employed algorithms to detect, identify,
and segment some medical features from biosignals.
IEC 62304 does not reference either of the two concepts, but both are
relevant, practical tools for appropriate requirements definition (IEC 62304
Clause 5.2.2) and risk analysis (IEC 62304 Clause 7).
Typical challenges here include:

◾ Insufficient verification plans and reports


◾ Insufficient validation plans and reports
◾ Missing validation reports (note that this is not required by IEC 62304,
but is expected by regulations and standards such as IEC 60601-1, IEC
82304-1, and ISO 13485)
Conformity Assessment ◾ 313

Change management is another typical area of mea culpas for software


manufacturers. Once a software is finished, assessed, and taken through all
the necessary regulatory hoops it may be all too easy to forget that also its
maintenance is important and regulated. Every once in a while you hear
developers make statements like bug fixes should have an express pathway
under the radar, or small fixes don’t require any impact assessment, which
is why changes are occasionally inappropriately handled. The stress sur-
rounding some urgent bug fix may also be conducive to cutting corners.
Nonetheless, recording and appropriately handling proposed changes is
a big part of IEC 62304 and what is, in fact, expected in medical devices.
Typical challenges here include:

◾ Performed changes are not recorded


◾ Description of changes does not cover what has been changed, why
it has been changed, or how conformity after the change has been
assessed
◾ Impact of the change on documentation has not been assessed
◾ Changes in regulations prompting software changes have been missed

The shortcomings discussed above, and others like them, cause delays and
increased expenses when discovered during conformity assessments. The
worse the detected issues are the longer it will take for the manufacturer
to correct them, and the longer it will be before the assessment is passed
and the device can be put on the market. Issues in information security,
for example, may lead to new test cases being generated, risk management
being revised, accompanying documents being updated, and more likely
than not also new programming work taking place. Unless a manufacturer
is well-versed in IEC 62304, a pre-assessment meeting with the certifica-
tion body will help prevent these issues, calibrate expectations, and set the
appropriate schedule for work leading up to the conformity assessment.

13.4 What Happens Afterward?


After you have the IEC 62304 test report in your hands you can breathe a
sigh of relief. The test report as well as any other testing reports and certifi-
cates you may have obtained for your product form a part of its technical
documentation. This body of documentation may then be used to demon-
strate the conformity of the device to applicable regulations and standards.
314 ◾ Medical-Grade Software Development

This documentation may be requested by the notified body and your regula-
tory authorities from time to time.
You will need to maintain the documentation up to date, and also
observe any revision needs brought on by changes in regulations and stan-
dards, or by changes you make to the product or its intended use. It is
important to remember that some of these changes may trigger a need to
reassess the conformity of your software in part or in full. The regulations
and any contracts you may have signed with your certification providers will
specify such triggers, but when in doubt consult your notified body or a
friendly regulatory consultant.
In general, undertaking IEC 62304 testing does not commit you to per-
form annual retesting, as is the case with maintaining ISO 13485 certificates
of your QMS. However, the IEC 62304 test report will have to be redrawn
if changes are made or new functionality is added beyond the previously
agreed parameters. The test report is always written for a specific software
version or a range of software versions allowing for certain types of con-
trolled changes to be made (see Section 8). If you move beyond this range,
a revised or completely new test report may be needed to preserve confor-
mity. Note also that IEC 62304 Clause 6.2.5 sets requirements for making
notifications of software changes (see Section 6.2). In practice, any signifi-
cant change to the medical device will also interest the notified body.
Chapter 14

REGULATORY APPROVAL

You have built it and your certification body has given you their
nod of approval on your hard work. There’s just one more check left
and that is obtaining the final regulatory approval, or clearance, for
your device. This is where you tie on that bandana, strap on your
test reports, and march towards that troop carrier hoping the ser-
geant at the door agrees to let you pass without too much groundho-
gery. The training montages are now over.
After jumping through all of the hoops discussed in the previous sections of
this book, and now feeling confident that you meet IEC 62304 requirements,
you will still probably need to obtain regulatory approvals or clearances for
your device. The kids in the back seat – or perhaps the investors and your
management team – might already be murmuring, “why we are not there
yet”? So, why did we go through all that trouble of passing the IEC 62304
conformity assessment if it doesn’t get us through the finish line?

14.1 Benefits of the IEC 62304 Test Report


At the beginning of this book, we discussed how the IEC 62304 standard
does not actually give us the tools to validate your software or release the
final medical device that either is the software you wrote or makes use of
that piece of software. Throughout the discussion between these covers, we
relied on ISO 13485, and a few other standards, to add the necessary layers
on top of IEC 62304. We also discussed how for medical devices both ISO
13485 quality management (see Section 4.1) and ISO 14971 risk management

DOI: 10.4324/9781003454830-14 315


316 ◾ Medical-Grade Software Development

(see Section 4.2) are part of the fabric, and something expected of the foun-
dations of your software-development pipeline.
The IEC 62304 conformity assessment done by the certification body
(see Section 13) and the medical-device conformity assessment done by
your regulatory authorities, or the notified body, are similar and hopefully
always compatible, but they do have slightly different points of view. Chief
among the differences here is the assessment of the clinical performance of
the device, which doesn’t get much attention during the IEC 62304 assess-
ment but is the 800-pound gorilla in the room during any medical-device
assessment.
During the IEC 62304 assessment, great focus was placed on the defi-
nition of requirements, the verification activities performed, and the risk
management conducted. During a regulatory assessment, all of the above
still have a role to play, but now much greater emphasis is placed on, for
example, the intended use, the clinical claims made, the medically sound
demonstration of clinical performance, and the handling of any possible
adverse effects.
The IEC 62304 assessment you have in your back pocket allows you to
provide evidence of your house being in order, which is not only benefi-
cial for signing off on the health of some required processes, but it will also
allow both you and your auditor to rely on the documentation produced by
those processes and your product performing as described by the documen-
tation. Being able to quote an IEC 62304 test report will not be the same as
having a Monopoly card in your pocket to thwart any software-related ques-
tion that might come up during an audit, but it might just give you the sense
of calm in knowing your own pockets as you head into the medical-device
audit process. The folks on the other side of the table may also want to see
that test report for their sense of calm in knowing your pockets.
In practice, the IEC 62304 conformity assessment is a useful and com-
monly expected health check of any medical device incorporating or con-
sisting of software. In theory, the assessment can be performed by the
manufacturer themselves, but demonstrating adequate competence and the
required objectivity may be problematic if done this way. This is also true in
the case of unaccredited or otherwise untrusted testing providers who in the
eyes of the regulatory authority are unknown quantities.
If, however, the IEC 62304 test report can be trusted, the regulatory
authorities can more easily accept that software development occurs in
accordance with the expected standards and that the documentation pro-
duced throughout the development life cycle will meet their expectations.
Regulatory Approval ◾ 317

Knowing this, the remaining audit can address the regulatory expectations
with less ambiguity over how the requirements should be interpreted in
this specific context. In other words, even if an IEC 62304 assessment is
not required of you in your circumstances, you would still benefit from it
by going through your activities in a more easygoing setting with genuine
software experts and getting a report on it all that allows you to anticipate
the sort of questions you will get regarding your software-development
activities, and either feel confident in your setup or work on the detected
issues before the medical-device auditors ride into town. You will be able
to strike up a conversation with your IEC 62304 experts in a way that your
regulatory authorities will not be allowed to interact in. The breadth of the
medical-device audits will be much wider than just your software-develop-
ment activities, but having an IEC 62304 assessment done first should mean
that you don’t need to go as deep into software during your medical-device
audits as you might otherwise.

14.2 ISO 13485 Certification


As part of ensuring that the foundations for your software-development
activities are solid, you will also be asked to show evidence for your overall
quality management meeting the expectations set for it. Increasingly this will
involve the use of the ISO 13485 standard, which was discussed briefly in
Sections 2 and 4.1 of this book and covered in-depth in our previous book
(Juuso 2022). This will also include ISO 14971 risk management (see Section
4.2), which is generally built into the QMS manufacturer’s setup.
Whether you have already obtained certification for your QMS, will do so
in concert with your first device certification, or are technically exempt from
certification but still expected to adhere to the standard, it should be a part
of your approach to conformity with regulatory requirements. Check your
regulations on what is expected of you based on the type of devices you
work with but expect ISO 13485 to come knocking sooner or later.
As discussed in Section 2, IEC 62304 mainly fits in Clause 7 of ISO 13485,
but it does shade the interpretation of all requirements of the standard if you
work with software products. If you have not done a separate IEC 62304
assessment you can expect the full weight of that standard to be exacted on
your operations here, along the lines discussed in Sections 4 through 9 of
this book. In other words, any ISO 13485 requirement will be linked to IEC
62304 requirements by the auditor and your process instructions, body of
318 ◾ Medical-Grade Software Development

development documentation, and other records will be analyzed as evidence


of meeting the expectations or falling short of them.
It is also possible to obtain a voluntary ISO 13485 certificate with a note
on IEC 62304 compliance for the software-development activities. Such a
certificate may be particularly practical for software subcontractors who pro-
duce software used as part of a medical device or with medical devices.

14.3 Medical-Device Conformity Assessment


After your software is ready, and it has passed the IEC 62304 conformity
assessment you subjected it to (see Section 13), it is time to proceed toward
that final regulatory gate with a sense of calm. In the EU, you will now be
talking with your notified body. In the US, you may be looking at arranging
a pre-submission meeting with the FDA or perhaps confidently putting the
submission together. In other parts of the world, you will be staring at letter-
heads with acronyms like TGA, NMPA (formerly CFDA), Anvisa, and the like.
When inspectors from these organizations come by you can expect some
poker faces and tough questions. Rightly so as the hopes and fears of every
patient, doctor, and consumer out there weigh on their shoulders as much
as these should be felt by your shoulders too. The one piece of advice we
would part with in this context is to breathe. Breathe when they walk in,
breathe when they ask you a question, and breathe while you give them the
answers they ask. This way you will not only have oxygen to formulate the
answers in your brain, but also the time to hear what they are actually ask-
ing for.
Assuming your auditors know of the IEC 62304 test report, and trust its
source, the types of questions you can now expect are as follows.

◾ What is the intended use of your device?


◾ What is your device intended for?
◾ What clinical claims do you make for your device? How has its clinical
performance been assessed? What clinical evidence do you have?
◾ How has risk management (incl. cybersecurity issues) been addressed
in the development of the product? How are these issues and activities
addressed over the life cycle of the product?

Your answers on all of the above will make heavy use of the technical file
of your product, including the documentation created in response to IEC
Regulatory Approval ◾ 319

62304 requirements. Many other questions regarding specific regulatory


requirements and quality-management processes will also be asked, includ-
ing staff competence, but in terms of software the above list represents the
most frequently reoccurring topics in audits. If you have gone over the pre-
vious sections of this book methodically you should be in a great place to
answer all these questions, and even occasionally politely discuss differences
of opinion in the interpretation of some requirements.

14.4 Changes
We will discuss life after approvals in the next section, but here it is per-
haps pertinent to point out that a regulatory approval is not a one-time
thing. This has been brought up repeatedly between the covers of this
book, but once you have gone through all the trouble of getting a product
and a development pipeline set up according to all of the many expecta-
tions it would be silly to go about changing that machinery in any haphaz-
ard way.
The need to revise your QMS or your product may be a good sign or
a bad sign. You may have identified some crack needing repairs, you may
have discovered a new improved way of doing something, or it may just be
that your product actually has customers and is being used so it also needs
a little seeing to from time to time. Changes and popularity may, after all, go
hand in hand: a product not used does not lead to feedback, but a product
getting a lot of use may see a steady stream of change requests.
Regardless of why a product has been changed, the technical file of the
product must reflect the changes (see Section 11). Changes in the QMS,
changes in the standards and regulations, or changes in the product and its
intended use all cause revisions to be considered in the documentation of
the product and the processes around it. With certain types of changes, it
may also become necessary to revisit the conformity assessments performed.
In general, the types of changes and their impact on regulatory approval
may be summarized as follows.

◾ A change in the QMS


This should not immediately affect product conformity, but unless the
change is trivial the notified body may be interested to hear about it. In
response, a new assessment may be initiated, or the status assessed as
part of, for example, the next scheduled surveillance audit. The change
320 ◾ Medical-Grade Software Development

here may also be a change in key personnel (e.g., management repre-


sentative, quality manager, PRRC, risk manager).
◾ A change in the key standards
Every once in a while, the standards do evolve, and it is your job to
stay on top of upcoming changes (see Section 15). With most if not all
changes you will be given a reasonable transition time to adopt the new
standards or realign to meet new guidance. The adoption of a standard
is generally optional, but to miss changes in the adopted standards is
suspicious and potentially dangerous.
◾ A change in the regulations
The effect of the change on the QMS, the product, and the possible
other activities of the manufacturer will be assessed separately. The
impact of revising any regulatory approvals will be a part of this assess-
ment. Anyone who has been following the EU MDR transition will
know that it can all be complicated, and both the timelines and effects
involved will vary.
◾ A change in the product
Your configuration management will have set up some form of ver-
sion control over changes to the product (see Section 8). This scheme
will have been the target of much interest from your certification pro-
viders, and you will have made promises to them on which types of
changes you will ask their permission on. This scheme will now dictate
how you address changes in the products, and how those changes can
affect product conformity. Any time the change may touch on the clini-
cal performance of your product you can expect some serious ques-
tions on the continued validity of your evidence.

Often the changes will have implications for reassessing product conformity.
If in doubt, you can reach out to your trusted consultants or even your regu-
latory authorities despite popular wisdom suggesting that it may be easier to
ask for forgiveness rather than permission. After all, you will be more easily
forgiven for asking an unnecessary question repeatedly than given permis-
sion to do as you will. In practice, your QMS processes will instruct what
changes are to be communicated and how, and these will have been audited
by your overseers.
Chapter 15

BUSINESS AS USUAL

You built it and they came? Your product is now flying off the shelf
and you are busy coming up with the next cool version that will
again totally revolutionize your target market. But what should you
be keeping your eye on, and what should you be doing now that it
is all built for the time being?
When you first have a big new idea for a medical-software product, you
probably have a gauntlet to run to get it funded, built, approved, and into
the hands of paying customers. The long-term support and maintenance of
that software is probably a distant blip on your radar, something you will “of
course” see to once you get there and have made it big. The medical-device
standards and regulations, however, want to see you well-prepared for that
long game, and for you to have built that bridge before you get to it. The
long game, for them, doesn’t end until your software is safely decommis-
sioned, and even then, you will be required to keep the lights on for several
more years to come in case issues pop up.
The standards and regulations speak of production information, mainte-
nance, and – most of all – post-market surveillance. The bottom line in all
this work is not your financial bottom line, but customer satisfaction, patient
safety, and whatever else your intended use and the regulations may pre-
scribe. All of this is a far cry from the fail-fast and mutate-to-win mindset
that may get you far in software development in other domains, such as the
development of mobile games. In medical-device software, the concept of
a customer relationship involves a much bigger commitment than just pick-
ing out car keys at a swingers’ party – or doing your initial elevator pitch
at some free-for-all investor event. In medical devices, you may not have

DOI: 10.4324/9781003454830-15 321


322 ◾ Medical-Grade Software Development

to commit to putting the offspring through college, but you may need to at
least know about the future educational update needs of your software and
its users. The long play is the end game.

15.1 The Everyday Processes to Now Run


Colorful language aside, the long-term existence of your software should
be on your mind when you design, develop, and release medical soft-
ware. You will want to, at least, address the following before placing
software on the market and develop the implicated processes, plans,
and arrangements as appropriate. The considerations for “business as
usual” after the release of medical software thus include, for example, the
following:

◾ Production monitoring
◾ Maintenance
◾ Post-market surveillance
◾ Regulatory reporting
◾ Regulatory monitoring (incl. the use of international standards)

Production monitoring is primarily a topic for churning out copies of a


physical product over a long period of time. Production line equipment may
experience wear and tear, the quality and other characteristics of raw materi-
als may fluctuate, and you may introduce other changes to the printing press
which may affect the conformity of the products you make. For software,
such fluctuation may be nonexistent, but nonetheless, components such as
SOUP modules and changes to the software platform you deploy your prod-
uct on will be on your mind as time goes by.
Maintenance is going to be a big factor for software-based products. As
mentioned in the preface to this book, software without updates will get to
be an oddity over any extended period of time. It will depend on the issues
you encounter, the features you want to revise or add, the nature of your
product, and how soon you will be putting out maintenance updates. You
have already made instructions and plans for maintenance activities before
releasing your product, so the first thing to see to here is that you live by
those promises.
Post-market surveillance is an increasingly important part of life after
product release. You must keep an eye on how your product – and any
Business as Usual ◾ 323

similar devices released by you or your competitors – is fairing out there


in the real world, how the science and technology around your device is
evolving, and how the regulatory and standardization landscape is chang-
ing. It all starts with sales and use figures, but your field of vision should be
wider here than just the mail inbox you might receive part of the feedback
through. Also, remember that ISO 13485 requires you to look at servicing
information as a source of customer feedback and complaints even when
your customer is only indirectly involved.
Regulatory reporting may also be an activity you need to have
running constantly or jumping into action at specific times or
after specific events. You may need to report on your post-market
­surveillance activities periodically. Similarly, some of the issues you
uncover during your post-market activities, or otherwise, may warrant
reporting to your notified body and regulatory authorities. The timeline here
will depend on the class of medical device you have, the type of event in
question, and the regulatory requirements applicable to you in the target
jurisdictions.
Finally, regulatory monitoring and the use of international standards
will be a domain to keep a watchful eye over in the coming years. It may
be that not much will change in the first few months after your software
is released, but over time significant changes will occur and it is impor-
tant to stay on top of such change. You will see official guidance on leg-
islation and standards being published all the time, new regulations and
standards will appear gradually, and the state-of-the-art – which in medi-
cal devices doesn’t even refer to the cutting edge, only to the generally
accepted good practice – will be constantly evolving. The only constant in
life is change, as countless wise people have no doubt said over the last
millennia.
It is, of course, impossible to keep track of everything at the same time.
Our recommendation is to get involved in industry forums, follow relevant
industry and scientific publishing to whatever extent you can, and perhaps
utilize a good regulatory intelligence service. A good consultant will also be
able to help you here.
IEC 62304 discusses many activities relevant during the post-market
phase of your product (e.g., change of management and problem resolu-
tion), but many of the other ongoing activities you will need to run around
your software are instructed by ISO 13485. You can find an in-depth discus-
sion of these in our first book Developing an ISO 13485-Certified Quality
Management System (Juuso 2022).
324 ◾ Medical-Grade Software Development

15.2 The Future of IEC 62304


The IEC 62304 standard is for now a way of life for medical software. The
standard has been declared stable for the next few years, and it comes with
such a following that it would not just be foolish, but dangerous, to feed it to
a shredder without first mapping out the way forward. The needs addressed
in the standard through its 9 clauses of requirements and administered
through its risk-based software safety classification are very real and are not
rendered obsolete by the newer challenges of agile development, artificial
intelligence, and whatever else soup du jour the standard is dipped in. The
same requirements will still need answers in the future.
The reality is that challenges have been identified relating to the use of
the standard in the environment of today. Even with a standard this old,
smart guidance tying it to modern needs and clarifying the requirements in
a way that doesn’t just rely on the assessor’s subjective assessment across the
board is clearly needed. Many of these challenges have led to the introduc-
tion of guidance or standards such as those discussed in Section 12. These
documents can be used together with the aging IEC 62304 standard to give
it a new lease on life. In this book, we have sketched out a joint existence
for IEC 62304 and ISO 13485, both of which set up the most significant
blocks of the Stonehenge that is medical-software development. In doing
so we have shown that the two really are not at odds with each other, can
coexist, and can be utilized to create a better environment for software
development and for the long-term supply of medical-device software.
In the coming years, the form the IEC 62304 standard takes may even see
dramatic turns, but whatever the metamorphosis, the DNA of the standard
will still be recognizably IEC 62304 – or the standardization organizations
may have missed their goal of fostering systematic, consistent, and predict-
able processes. Looking to the darkness of the clouds ahead, Sarah Connor
scratched out the words “No fate” on that wooden tabletop in Terminator 2.
In terms of IEC 62304 that may also be true, but the standard is still state-of-
the-art, and as such, is as necessary as the aging T-800 in that film. To shred
the IEC 62304 framework and leave the present adopters behind is not pos-
sible, smart, or safe.
Should the end of the world come, though, you should be well-positioned
as this book has guided you toward meeting first the requirements of ISO
13485 quality management, and from this foundation, meeting the present-
day requirements for software development. In the future, you can look for
guidance and early warning signs of how the requirements around both
Business as Usual ◾ 325

of these evolve, but at present no change is imminent, and no change will


happen overnight unless the values of standardization itself are put to the
shredder.

15.3 The Future of ISO 13485


The ISO 13485 standard is for all intents and purposes a bedrock standard.
One should never say never, but in the present stage of its use and budding
adoption across the globe, it should be more likely that dinosaurs return to
the Earth than the standard is reworked in any upsetting way.
We take an active role in following and taking part in the development
of the standard and have done so for a number of years. As committee
members on the ISO technical committee and its national mirror groups,
we have watched with delight how this standard is living up to its moni-
ker as an international standard. The challenges put to the standard are
largely the same as those put to IEC 62304, but no one in their right mind
wants to rock the stability and trans-continental compatibility found in the
standard at the moment. A wealth of related guidance is seen as benefi-
cial, and this will come out in due course in one form or another, but no
reshuffling or seismic changes to the standard itself are to be expected. In
this book we chose ISO 13485 as the backdrop to everything, and with
good reason. The standard is the foundation of medical device develop-
ment and realization.

15.4 The Joy of Compliance


The standards are written by experts wanting to point out the important fac-
ets of doing something, and all the various information sources and factors
you will want to consider while climbing toward that elusive peak of perfec-
tion. In this book, we have been careful not to make this sound like you
must know everything to make a start with anything. We have, on many
occasions, pointed out things, both great and small, that you will want to
at least consider on your climb. At the same time, we have wanted to distill
each of the requirements to their core and show you the big motivations and
driving considerations behind them. We hope we have succeeded in this
and shown you the axes around which many of the moving parts revolve,
and around which you can expect all future additions to also revolve.
326 ◾ Medical-Grade Software Development

As changes emerge, you will face transition activities of different mag-


nitudes. Many changes in the regulations and standards may feel like they
only call for minor revisions to be made in your documentation. This may
even be the case for some changes. In the case of other changes, you will
need to rethink some subprocess, or add some activity, and revise your
instructions and documents in turn. In a more dire situation, you might
find that the clinical evidence for your legacy product is called into ques-
tion or deemed insufficient due to changing regulations, in which case
conducting new testing, gearing up risk-management activities, and obtain-
ing more expansive data on your clinical performance will perhaps get to
be an expensive proposition. The better you understand why you have
been asked to shape a process a certain way, perform some testing or risk-
management activity, or provide some particular piece of documentation the
better you will be prepared for reacting to – or perhaps even anticipating –
some change in the external requirements applicable to your products and
organization.
Shaping your operations according to commonly accepted international
standards, such as those discussed in this book, will not only give you a
solid, defensible way of developing software but also the most straightfor-
ward platform for addressing any future challenges. In the end, regulatory
compliance will be key but maintaining compliance with the right standards
will be your best bet to also remaining in compliance with the applicable
regulations both domestically and internationally too.
Chapter 16

CONCLUSIONS

At the climax of the fourth Indiana Jones film, it all ends in a vortex
of all the knowledge mankind has ever accumulated – and this is
then comically flushed down into the depths of the jungle. We hope
this has not been, and now is not, your experience of trying to fuse
together the expectations of medical-software development. If it is,
try to count the number of sprint cycles the fulcrum goes through in
that scene, and after recovering from that distraction, reconsider.
Reading the history of medical devices, you will come across inspirational
stories from the recent past that identified needs in the treatment of patients,
coupled these with the necessary engineering skills, and made the world an
undeniably better place. In these stories, the spark for a new medical device
may have come from a fortuitous meeting of minds, like today, but the path
to testing the prototypes out with real patients was occasionally significantly
shorter than that expected today – perhaps only a matter of days from the
initial concept. In this regard, no example is perhaps as inspiring as the
origin story of the Minnesota-based medical-device mammoth Medtronic.
It was in 1957 when a University of Minnesota heart surgeon, Dr C. Walton
Lillehei, and an enterprising electrical engineer, Earl Bakken, dreamt up a
wearable transistorized pacemaker and only days after attached a proto-
type to the first young patient. A year later the device was on the market,
and decades later an implantable version of this same pacemaker is now
commonplace.
Today this timeline may sound fantastical, and certainly, the risk-man-
agement activities involved and the investigational procedures called for
are now more evolved, but the drive to meet new unmet needs and to also

DOI: 10.4324/9781003454830-16 327


328 ◾ Medical-Grade Software Development

better meet those previously addressed is forever present. Those needs are
increasingly being met by software – software embedded in medical devices
and software acting as medical devices. Just as the invention of the transis-
tor, and more specifically the use of those transistors in a metronome cir-
cuit, allowed the creation of the first heart pacemaker, the ongoing surge of
software holds the promise of ushering in the next generations of medical
devices. Not all software has to be dubbed “artificial intelligence” for it to
change the world in some discernible increments, either.
It is, therefore, of great importance that the development and mainte-
nance of that software can take place with as little friction and overhead
as possible to support innovation and patient health. There are vastly more
software developers out there who could contribute to medical software
than those software developers who are willing to learn and live by med-
ical-device standards and regulations. That is not to say that we should do
away with standards and regulations, most certainly we should not. It does,
however, suggest that we should make it as simple as possible to know what
is good practice and what is required in the field so that the platform for
developing – and maintaining – medical software makes as much sense as
possible. This also has implications in academia where a great number of
research projects are set up, but the fruits of those endeavors may rot away
after the project ends because it would be too much work to redo every-
thing according to what a medical project launch would require. Knowing
enough about the science or the clinical setting may not be a problem in
such a context, but recording that information and considering enough of
the potential big showstoppers from standards and regulations may drop a
wrench-shaped anchor on the proceedings.
It was this drive to figure out the requirements and the best practices and
to engineer them into some cohesive, practical framework that drove our
efforts to write this book. In the discussion within these covers, we have
attempted to align the two major standards, ISO 13485 and IEC 62304, in
a practical and purpose-driven way. That purpose was to meet all of the
universal expectations placed on medical software in a way that also makes
sense for the developers. We firmly believe that understanding the reason-
ing behind such expectations and requirements does not only mean that
the demands are met, but also met better, safer, and with less friction than
what could be expected if they were viewed just as something required, for
whatever reason.
Writing this book was almost a two-year project that involved diving
deep into the two principal standards, and a good number of their brethren,
Conclusions ◾ 329

as well as the applicable regulatory requirements that these occasionally


reference. We spent countless hours in discussion on the various aspects
involved and reached out to other expert voices on many of the topics. In
the discussion here we hope we have given a good overview of the com-
bined requirements, identified the various linkages, and pointed out all of
the key design considerations and questions you may want to address in
developing and streamlining your software activities.
The discussion here in this book is the most developed and most com-
prehensively considered approach to a consolidated, unified model of medi-
cal software development we are aware of. Still, the solutions offered here
are based on our own experience and insight into the development and use
of international standards in medical devices. Your own solutions may differ
in places and still make as much sense as our suggestions here. Considering
your position on our suggestion will, however, give you both new ideas and
the support you want in your choices. As we pointed out at the beginning
of this book, there are more needles than just one in that haystack, but we
hope that by reading the discussion here you will have gained confidence in
knowing that haystack.
The last thought we want to leave you with here is that, if you are
already accustomed to developing software on some other domain, you
shouldn’t feel like you have to reinvent everything for the medical-device
world. If you have a solid development process that has worked for you
well in the past, the chances are that it – or many elements from it at least –
will also work for you in medical devices. ISO 13485 leaves great room for
you to develop software between its clauses on D&D inputs and outputs.
IEC 62304 adds many more markers to hit in doing so, but even it doesn’t
micromanage your day-to-day. In practice, just about all of the best practices
and software development models you may be accustomed to from other
domains and disciplines will also be translatable to medical-software devel-
opment – but translation is almost certainly required.
The fundamental requirements for the medical-software development
process are that all work occurs in a planned-out fashion, outcomes are
approved by appropriate functions, the whole is somehow approved as well
as the incremental parts, long-term maintenance is part of the big picture,
effects are analyzed before executing changes, and throughout all records
are maintained as appropriate. Any development model that meets these
fundamentals can be mapped to the requirements of the standards and
regulations. In practice, the reality of designing an appropriate develop-
ment process is, of course, a little bit more complex than just fulfilling this
330 ◾ Medical-Grade Software Development

handful of requirements, as this book has perhaps shown, but the grandest
of strokes are accurately set by them. Use this realization to look at where
you are today, and then think carefully about where you want to be and you
should get there. This is what Deming would also advise you to do.
Similarly, the world never stops. There will be new standards and regula-
tions to comply with, some perhaps prompted by new advanced technolo-
gies and others by increased calls to interact with your users over the life
cycle of your software. Although the requirements will evolve, the funda-
mental needs behind each requirement will remain much more stable. We
will always be asked for better, cheaper, and more powerful medical devices
providing the greatest of benefits without any risks to anyone. There is no
such thing as future-proof, of course, but knowing the fundamental goals
behind each essential requirement, and ensuring you have lean processes in
place for meeting these goals, will give you the best platform for any future
developments too.
That’s it for now. We searched high and low for the different kinds of
needles in the haystack, and hopefully gave you new ideas and perspectives
on developing medical software in the context of ISO 13485 quality man-
agement. If you agree or disagree with any of the discussion in this book,
or have further comments, questions, anecdotes, or other needles from the
haystacks to share please contact us. You can reach us both via LinkedIn
and also by writing to sw​@theqmsbook​.c​om.
Thank you for reading our book! Like James Bond, we may return.
References

Juuso, I. (2022). Developing an ISO 13485-Certified Quality Management System


– An Implementation Guide for the Medical-Device Industry. Routledge, New
York.
MDCG. (2019-11). Guidance on Qualification and Classification of Software in
Regulation (EU) 2017/745 – MDR and Regulation (EU) 2017/746 – IVDR (2019).
Medical Device Coordination Group.
Pure Food and Drug Act. (1906). United States Statutes at Large (59th Cong., Sess. I,
Chp. 3915, p. 768-772; cited as 34 Stat. 768).

331
Index

A Center for Devices and Radiological Health


(CDRH), 56
ABC classification, 11, 12, 92, 99, 310 CER, see Clinical evaluation report
Agile software development Certification body (CB), 5, 52, 61, 75, 310,
advantage, 264 313, 316
ChatGPT, 262 Change control
Analyze data, trends IEC 62304 standard, 216–217
IEC 62304 standard, 236 ISO 13485 standard, 215–216
ISO 13485 standard, 235–236 synthesis, 217–218
synthesis, 236–237 use, 231–233
Applicable regulations, 20–22 Clinical evaluation report (CER), 167, 265
Architectural design, 132–133 Cloudbased services, 260
IEC 62304, 134–136 Cloud computing, 300–302
ISO 13485, 133–134 Code vs. documentation, 26–27
synthesis, 137–138 Configuration management
Artificial intelligence, 16, 25, 213, 286, 290, aim, 206
296–300, 328 A-level change, 210
Audit package, 259 B-level change, 210
C-level change, 210
B databases, 213–214
D-level change, 210
B2B customers, 27, 32 identification, 207–208
Business as usual IEC 62304 standard, 208–209
compliance, 325–326 ISO 13485 standard, 208
everyday processes, 322–323 synthesis, 209
IEC 62304 standard, 324–325 tracking, 211–213
ISO 13485 standard, 325 version numbering, 210–211
Configuration status accounting, 218, 219
C Conformity assessment
assessor/assessment team, 305–307
CAPA, see Corrective and preventive actions documentation, 313–314
CB, see Certification body IEC 60601-1, 309

332
Index ◾ 333

IEC 62304 assessment, 308–310 synthesis


regulatory approval, 318–319 plans, 122–123
shortcomings, 310–313 SDP, 119–121
Controlled items, history SOP, 117–118
IEC 62304 standard, 219 system development plan, 118–119
ISO 13485 standard, 219 use of systems, 123–124
synthesis, 219 Development records, 256–258
Corrective and preventive actions (CAPA), Device History Record (DHR), 254
56, 223–228, 231, 232 Device Master Record (DMR), 254
Cybersecurity, 290–291 DHR, see Device History Record
IEC 81001-5-1 standard, 294–295 DIR, see Design input review
ISO/IEC 27001 standard, 295–296 DMR, see Device Master Record
jurisdictions, 293–294 Documentation Level rating, 12
product life cycle, 291–293 DOR, see Design output review
DoS, see Denial-of-service
D DRR, see Design Readiness Review
DTR, see Design Transfer Review
D&D, see Design and development
D&D deliverables, 47 E
D&D input stage, 158, 165
D&D output stage, 158, 162, 165 Edge AI, see Edge intelligence
Denial-of-service (DoS), 305 Edge intelligence, 261, 302–303
Design and development (D&D), 13, 33, 57, Electronic health-record systems, 9
106, 126, 207, 224, 254 External risk controls, 96, 193–194, 277
Design input review (DIR), 29, 31–33, 131, 257
Design output review (DOR), 33, 159, F
166–167, 249, 257
Design readiness review (DRR), 33 Failure modes and effects analysis (FMEA),
Design transfer review (DTR), 33–34, 161, 98, 203
167–170, 258 hazards, 281
Detailed design ISO 14971 standard, 283
architectural map, 138 risk score, 282
class B software, 139 RPN risk scores, 283
IEC 62304 standard, 140 types of risks, 280–281
ISO 13485 standard, 139 Fault tree analysis (FTA), 284
synthesis, 141 FMEA, see Failure modes and effects
Development life cycle, 15, 60, 105, 113, analysis
120, 248, 265–267 Foundational requirements, 89, 207
Development planning FTA, see Fault tree analysis
aspects, 112–115
configuration management, 115–116 G
IEC 62304
planning documentation, 110–111 Gap analysis
process documentation, 111–112 closure activities, 103–104
ISO 13485 rationale, 104
planning documentation, 110–111 Global medical-device ecosystem, 56
process documentation, 109–110 Green-coding practices, 94
334 ◾ Index

H modification implementation, 186–187


nature of the software, 74
Hazardous situation, 94, 194–197 non-binding notes, 75
IEC 62304 standard, 197–198 normative references, 76–77
ISO 13485 standard, 197 primordial ooze, 62–65
synthesis, 199 probability, 97–98
problem and modification analysis,
I 182–184
problem investigation, 227–228
IEC, see International Electrotechnical problem reports, 225
Committee process, 71
IEC 60601-1, 9, 37, 62–66, 86, 128, product maintenance, 188
277, 306 QMS, 90
IEC 62304 standard, 6–7, 10, 14, 17 release
activity, 71 D&D Output Review, 162
alignment, 107 production and service provision,
analyze data, trends, 236 164–165
architectural design, 134–136 validation, 163–164
business as usual, 324–325 verification, 163
change control, 216–217 relevant parties, advise, 230
conducting reviews in D&D stages, 170 requirements, 84–85
configuration management, 208–209 requirements analysis
controlled items, history, 219 finalizing requirements, 129–130
D&D validation, 172–174 inputs for requirements, 126–129
D&D verification, 170–172 risk-control measures
detailed design, 140 software changes, 202
developing, 22–23 verification, 200–201
development planning risk management, 91–92
planning documentation, 110–111 safety classification, 92–94
process documentation, 111–112 scope, 73–76
documentation set, 19–20 software configuration management
external risk controls, 96–97 process, 82–83
final remarks, 99–100 software development process, 80
foreword, 69–70 software maintenance plan, 179
future, 67–68 software maintenance process, 80–82
general requirements, 79–80 software problem resolution, 239–240
guidance on the provisions, 85–88 software problem resolution process,
harm in relation to people, 193 83–84
hazardous situation, 94, 197–198 software risk-management process, 82
implementation, 88 software-system testing, 154–156
integration testing, 148–151 software will fail, 95–96
introduction, 70–71 specific process structure, 18
legacy software, 48–49 task, 71–73
maintain records, 234 terms and definitions
mode of execution, 74 software item, 78
modern times, 65–67 software system, 77–78
Index ◾ 335

software unit, 78 release


system, 77 D&D Output Review, 158–159
test documentation contents, 242 production and service provision,
testing, 36 161–162
type of delivery, 74 validation, 161
type of storage, 74 verification, 160–161
unit implementation and verification, relevant parties, advise, 330
143–145 requirements analysis
verification and testing activities, 107 finalizing requirements, 126–127
word cloud, 74, 76, 80, 83 inputs for requirements, 126
IEC 81001-5-1 standard, 294–295 risk-control measures
IEC 82304-1, 267–269 software changes, 201–202
post-market activities, 275–276 verification, 200
product documentation, 272–275 software maintenance plan, 177–179
product requirements, 269–270 software problem resolution, 238
software development process, 271 software-system testing, 153–154
validation, 271–272 test documentation contents, 241–242
IEEE 610.12:1990 standard, 14 testing, 36
Integration testing unit implementation and verification,
IEC 62304 standard, 148–151 142–143
ISO 13485 standard, 147–148 vs. IEC 62304, 59–60
synthesis, 151–152 ISO 14971 standard, 7, 50, 78, 283
International Electrotechnical Committee
(IEC), 63, 69 L
ISO/IEC 27001 standard, 295–296
ISO 13485 standard, 7, 10, 16, 17 Legacy concept, 65
alignment, 107 Legacy software, 48–49
analyze data, trends, 235–236 gap analysis, 101–104
architectural design, 133–134 risk-management activities, 100–101
business as usual, 325 Level of concern (LoC), 12, 76, 99
change control, 215–216
configuration management, 208 M
controlled items, history, 219
detailed design, 139 Maintain records, 126, 223, 233–235
development planning MDCG, see Medical Device Coordination
planning documentation, 110–111 Group
process documentation, 109–110 MDR, see Medical Device Regulation
hazardous situation, 197 MDSAP, see Medical Device Single Audit
integration testing, 147–148 Program
maintain records, 234 Medical Device Coordination Group
modification implementation, 186 (MDCG), 9, 293, 299
problem and modification analysis, 182 Medical Device Regulation (MDR), 9, 66–68,
problem investigation, 227 254, 265, 297, 304
problem reports, 224–225 Medical Device Single Audit Program
product realization, 57–60 (MDSAP), 56
quality management system, 56–57 Microchips, 62
336 ◾ Index

Minnesota-based medical-device, 327 Q


Modification implementation
IEC 62304 standard, 186–187 QMS, see Quality management system;
ISO 13485 standard, 186 Quality management systems
problem-resolution process, 185 Quality management system (QMS), 17
synthesis, 187–188 best ideas, 251
execute plans, 250
planning documentation, 250
N
realization processes, 247–249
National certification body (NCB), 52 release activities, 251
Non-cloud software, 300 Quality System Record, 254
Not applicable (N/A), 39
R
O
Regulation
Object-oriented programming language, applicable, 20–22
13 maintaining trust, 24–25
Regulatory approval
changes, 319–320
P
IEC 62304 test report, 315–317
PEMS, see Programmable medical electrical ISO 13485 certification, 317–318
systems medical-device conformity assessment,
PESS, see Programmable electronic 318–319
subsystems Release
Plan-Do-Check-Act model, 31 IEC 62304 standard
Post-production activity, 191 D&D output review, 162
Pre-Cert program set, 16 production and service provision,
Prerequisites, 89 164–165
Problem and modification analysis, validation, 163–164
180–182 verification, 163
IEC 62304 standard, 182–184 ISO 13485 standard
ISO 13485 standard, 182 D&D output review, 158–159
synthesis, 184–185 production and service provision,
Problem investigation 161–162
IEC 62304 standard, 227–228 validation, 161
ISO 13485 standard, 227 verification, 160–161
synthesis, 228–229 synthesis, 165–166
Problem reports, 221–224 DOR, 166–167
IEC 62304 standard, 225 DTR, 167–170
ISO 13485 standard, 224–225 validation report, 167
synthesis, 225–226 Relevant parties, advise
Process documents, 256 IEC 62304 standard, 230
Production activity, 191 ISO 13485 standard, 330
Programmable electronic subsystems synthesis, 230–231
(PESS), 64 Requirements analysis, 124–125
Programmable medical electrical systems IEC 62304
(PEMS), 64 finalizing requirements, 129–130
Index ◾ 337

inputs for requirements, 126–129 risk management, 49–51


ISO 13485 safety, 53–54
finalizing requirements, 126–127 shortcomings, 41
inputs for requirements, 126 testing, 34–38
synthesis validation, 30–31
requirements elicitation, 130–131 verification, 29
requirements review, 131–132 Software development plan (SDP), 112,
Residual risk, 190–191, 272, 283 119–121, 246
Risk assessment, 130, 132, 168, 190, 194, Software in a Medical Device (SiMD), 9, 13
213, 229, 269, 289 Software items, 13, 14, 78
Risk control, 190 Software maintenance plan
Risk-control measures IEC 62304 standard, 179
IEC 62304 standard, 199–200 ISO 13485 standard, 177–179
ISO 13485 standard, 199 synthesis, 179–180
software changes Software of Unknown Provenance (SOUP),
IEC 62304 standard, 202 45–46
ISO 13485 standard, 201–202 Software problem resolution
synthesis, 202 IEC 62304 standard, 239–240
synthesis, 200 ISO 13485 standard, 238
verification synthesis, 240–241
IEC 62304 standard, 200–201 Software safety classification, 92, 192–193
ISO 13485 standard, 200 Software-system testing, 152–153
synthesis, 201 IEC 62304 standard, 154–156
ISO 13485 standard, 153–154
S synthesis, 156
Software unit, 13
SaaS, see Software-as-a-Service Standard operating procedure (SOP),
SaMD, see Software as a Medical Device 117–118, 246
SDP, see Software development plan State of the art, 7
Segregation, 276–280 Stockholm syndrome, 55
SiMD, see Software in a Medical Device System begets, 12
Single fault conditions, 65, 312
Software architecture, 134, 276–280 T
Software as a Medical Device (SaMD), 9, 12, 13
Software-as-a-Service (SaaS), 300 Technical documentation, 187, 252–255,
Software-development 289, 313
CB, 51–53 design and development files, 254
configuration items, 46–48 medical-device family, 253
continuous improvement, 43–45 Terminological reductionism, 17
controlling changes, 46–48 Test documentation contents
evaluation, 29 IEC 62304 standard, 242
fail, 40 ISO 13485 standard, 241–242
forward momentum, 41–43 synthesis, 242–243
life cycle, 14–16 Three-level safety classification, 11
not applicable (N/A), 39 Top events, 284
pass, 40 Transfer Readiness Review (TRR), 33
review, 29, 31–33 TRR, see Transfer Readiness Review
338 ◾ Index

U W
Unit implementation and verification Waterfall model, 25, 107, 167, 248,
designing, 142 263, 264, 306
IEC 62304 standard, 143–145
ISO 13485 standard, 142–143 Z
synthesis, 145–146
Usability engineering, 255, 261, 273, Z-annexes, 7
284–290, 292, 293, 300, 307 Zero emissions goals, 94