miniiSED Operator's Manual (English)
miniiSED Operator's Manual (English)
OPERATOR MANUAL
Cat. No. 112-01017
This page intentionally blank.
Dear miniiSED™ Customer,
ALCOR Scientific would like to welcome you into the world of fast, efficient and accurate
erythrocyte sedimentation rate (ESR) results. With our welcome, we have provided you a
packet of information to get you started. We hope this information will make using the miniiSED
even easier. Please find enclosed:
Warranty Card
The instrument is backed by a One (1) Year Warranty. To assure coverage, you must activate
your Warranty by filling out the warranty card included with your instrument and mailing it back
to ALCOR. The serial number label is located on the rear panel of the analyzer. Please see
the last page of the Operator Manual for more information and instructions.
ALCOR Scientific offers Technical Support Monday through Friday 8:30am-5:00pm EST
(excluding all US Federal Holidays) to help service you in the quickest way possible. The
Technical Support Team can be reached by any of the following:
Please do not hesitate to contact ALCOR or your authorized ALCOR distributor if you have any
questions regarding any of the information found in this manual.
We thank you for selecting ALCOR products and look forward to serving your laboratory!
Best Regards,
ALCOR Scientific Support Team
ii of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
SYMBOL REFERENCE
The following is a list of symbols and their meaning used on the instrument, consumables and accessory
labeling.
Symbol Meaning
Date of manufacture
Manufacturer
Serial Number
Product/Reference number
Fuse Rating (located on serial number label, replace with same value and type)
Consult instructions- refer operator to the instruction manual for additional information
iii of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Notes
NOTE: Highlights important facts, gives helpful information and tips and clarifies
procedures.
Cautions
Warnings
Biological Warnings
iv of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Please pay close attention to the instructions, notes and symbols as well as the standard
laboratory practices outlined by your facility and local regulatory agencies.
Always keep a distance of at least 4 inches (10 cm) between the rear of the instrument
and the wall to allow for proper ventilation.
Do not use power frequencies or voltage other than those specified in this document.
Connection to an inappropriate power source may cause injury or fire.
Do not disassemble or modify the instrument. Doing so may cause injury and /or
instrument malfunction and void the warranty.
Place the instrument on a stable and level surface free of vibration. Failure to do so may
cause injury or malfunction of the unit.
CAUTION: To reduce the risk of electrical shock, do not remove any panel or open
housing unless under the direction of qualified personnel.
WARNING: For continued protection against risk of fire and hazard, replace only with
the same type and rating fuse.
WARNING: The instrument’s main power switch is used as the main disconnect device.
v of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
TABLE OF CONTENTS
Notes, Precautions, Warnings and Biological Warnings ............................................iv
Precautions and Safety Information ............................................................................. v
1. Intended Use ......................................................................................................... 1
2. Methodology ......................................................................................................... 1
2.1.History ....................................................................................................................... 1
2.2.Comparison With Existing Methods............................................................................ 2
2.3.Method Limitations ..................................................................................................... 2
3. Principle of Procedure ......................................................................................... 3
4. General Information ............................................................................................. 3
4.1. For In Vitro Diagnostic Use Only ................................................................................ 3
4.2.Sample Requirements................................................................................................ 3
4.3.Tube Requirements ................................................................................................... 4
5. Instrument Overview ............................................................................................ 4
5.1.Consumables ............................................................................................................. 4
5.2.Parts Identification ..................................................................................................... 5
5.3.Continuous Operation Mode ...................................................................................... 5
6. Unpacking and Installation .................................................................................. 6
7. User Interface ..................................................................................................... 10
8. Calibration........................................................................................................... 17
9. Results ................................................................................................................ 18
9.1.Expected Values .......................................................................................................18
9.2.Performance .............................................................................................................18
10. Smart Cards ........................................................................................................ 18
10.1. Smart Card Menu ...............................................................................................19
11. Routine Maintenance ......................................................................................... 20
11.1. Replacing/Emptying the Waste Bottle .................................................................20
11.2. Replacing miniiWASH Bottle...............................................................................21
12. Safety Precautions ............................................................................................. 21
12.1. General Considerations ......................................................................................21
12.2. Biological Waste .................................................................................................22
13. Preventative Maintenance ................................................................................. 22
14. Technical Support .............................................................................................. 23
15. Technical Specifications ................................................................................... 24
16. Warranty Information ......................................................................................... 25
vi of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
1. Intended Use
The miniiSED Erythrocyte Sedimentation Rate analyzer, is an automated sedimentation rate analyzer which
reports sedimentation rate in mm/hr. It is a non-specific, quantitative result. Testing is done using
EDTA/whole blood samples, obtained by venipuncture or capillary blood collection. The instrument can be
used in laboratories authorized to perform test classified as Moderately Complex under CLIA categorization
by order of a physician to aid in assessing the general health status of a patient.
2. Methodology
2.1. History
The discoverer of Erythrocyte Sedimentation Rate (ESR) in 1897 was a Polish physician, Edmund
Faustyn Biernacki 1 (1866–1911). The most important conclusions from his observations were as
follows: the blood sedimentation rate is different in different individuals; blood with small amounts of
blood cells sediments faster; blood sedimentation rate depends on the level of plasma fibrinogen; in
febrile diseases (rheumatic fever included) with high levels of plasma fibrinogen the ESR is increased;
and in defibrinated blood the sedimentation process is slower. The findings presented by Biernacki
clearly showed the clinical significance of ESR.
In 1921, Swedish internist Alf Vilhelm Albertsson Westergren (1891–1968), presented a similar
description of the phenomenon of ESR 2 as those given by Biernacki and Swedish hematologist,
Robert Sanno Fåhraeus (1888–1968) 3. Westergren applied a blood sampling method to the ESR test
using sodium citrate as anticoagulant. Westergren also defined standards for the ESR test and to
which nearly all other automated ESR analyzers are referenced today. 4, 5
The miniiSED analyzer uses advanced Rheology Technology to measure the “earliest and most
critical phase” of the Erythrocyte Sedimentation, which is called rouleaux formation. The rouleaux
formation is the critical phase of ESR and the one that ultimately determines the rate at which the red
cells will settle in the Westergren tube.
The technical innovation of the miniiSED analyzers consists of “directly” measuring the aggregation
of the red blood cells, whereas the traditional ESR methods “indirectly” measure the aggregation of
the red blood cells by recording the length at which the red cells settle in a Westergren tube.
After measuring the aggregation directly, the miniiSED analyzer produces ESR results in mm/hr.
Utilizing EDTA blood from the primary tube, results are reported within seconds.
1 Biernacki E. Die spontane Blutsedimentirung als eine wissenschaftliche praktisch-klinische untersuchungsmethode. Dtsch Med
Wschr 1897; 23: 769–72.
2 Westergren A. Studies of the suspension stability of the blood in pulmonary tuberculosis. Acta Med Scand 1921; 54: 247–82.
3 Fåhraeus R. Über die Ursachen der verminderten Suspensionsstabilität der Blutkörperchen während der Schwangerschaft. Biochem
Z 1918; 89:355–64.
4 International Council for Standardization in Haematology (Expert Panel on Blood Rheology): ICSH recommendations for
measurement of erythrocyte sedimentation rate. J Clin Pathol 1993; 46:198-208.
5 Thomas RD, Westengard JC, Hay KL, et al: Calibration and validation for erythrocyte sedimentation tests. Arch Pathol Lab Med
1993; 117:719-72.
1 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
The miniiSED erythrocyte sedimentation rate analyzer is designed to sample directly from the primary
(lavender top) 13 x 75mm, EDTA blood collection tube, automatically withdrawing a test sample of
100µL volume and can produce a result in as little as 20 seconds, with appropriate prior
homogenization. The instrument’s micro-flow cell allows capture of the critical kinetics of red blood
cells aggregation in a highly controlled testing environment. This system eliminates handling and the
associated factors that can contribute to result variability. The miniiSED reports results which have
been correlated with the Westergren method.
Sample mixing is performed at the beginning of the analysis with the purpose of homogenizing the
sample. An inefficient homogenization can affect the results given by the instrument.
NOTE: ESR is a nonspecific reaction. It is highly recommended to perform other tests together with
ESR, since an ESR value is not enough to exclude that the patient is not affected by a pathology or
to diagnose a pathology.
6 CLSI. Procedures for the Erythrocyte Sedimentation Rate Test; Approved Standard-Fifth Edition. CLSI document H02-A5.
Wayne, PA: Clinical and Laboratory Standards Institute; 2011.
2 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
3. Principle of Procedure7
The ESR is a simple non-specific screening test that indirectly measures the presence of inflammation in
the body. It reflects the tendency of red blood cells to settle more rapidly in the face of some disease states,
usually because of increases in plasma fibrinogen, immunoglobulins, and other acute-phase reaction
proteins. Changes in red cell shape or numbers may also affect the ESR.
When anticoagulated whole blood is allowed to stand in a narrow vertical tube for a period of time, the
RBCs – under the influence of gravity - settle out from the plasma. The rate at which they settle is
measured as the number of millimeters of clear plasma present at the top of the column after one hour
(mm/hr). The RBCs sediment because their density is greater than that of plasma; this is particularly so,
when there is an alteration in the distribution of charges on the surface of the RBC (which normally keeps
them separate from each other) resulting in their coming together to form large aggregates known as
rouleaux. Rouleaux formation is determined largely by increased levels of plasma fibrinogen and globulins,
and so the ESR reflects mainly changes in the plasma proteins that accompany acute and chronic
infections, some tumors and degenerative diseases. In such situations, the ESR values are much greater
than 20mm/hr. Note that the ESR denotes merely the presence of tissue damage or disease, but not its
severity; it may be used to follow the progress of the diseased state or monitor the effectiveness of
treatment.
4. General Information
Read this manual carefully prior to operating the instrument.
This document is the operator’s manual for the instrument. It is intended to explain the instrument
operation in detail and can be used as a basis for training new operators. It is an information guide
and troubleshooting reference. Retain this manual for future use.
Sample volume for pediatric testing is 100 μL whole blood (400 μL dead volume)
Sample should be tested within 4 hours from venipuncture or within 24 hours if refrigerated
Sample must be brought to room temperature for at least fifteen (15) minutes (if refrigerated)
NOTE: The instrument requires no additional or special sample preparation. As with all anti-
coagulant collection tubes, the sample should be well mixed after collection to help avoid
clotting or other aggregates that may alter ESR test results.
3 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
5. Instrument Overview
The rate at which red blood cells aggregate in whole blood has a direct effect on the resulting sedimentation
rate. Sedimentation rate is therefore an indirect representation of the rate of aggregation. The miniiSED
erythrocyte sedimentation analyzer uses photometrical rheology to directly measure the aggregation of red
blood cells. Once the sample is automatically processed and in position, a sensitive optical detector in the
miniiSED follows the progress of aggregation over time. This produces a signal that is a direct
representation of the aggregation. The magnitude of time-dependent change is correlated to the
Westergren method.
5.1. Consumables
Consumables can be purchased by calling ALCOR Customer Service Department or your local
authorized ALCOR distributor.
NOTE: Only use consumables that are within their expiration date.
NOTE: Use of any other product could affect the performance of the instrument and void the
warranty.
4 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
1 1 Touch Screen
2
6 On/Off Switch
5
6
8
7
9
NOTE: The instrument is programmed to perform self-cleaning after being idle for fifteen
(15) minutes following the last sample tested. The process takes approximately one (1)
minute and utilizes approximately 2.5mL of miniiWASH for each wash cycle. Once
completed, testing can resume as normal.
5 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
CAUTION: The instrument weighs approximately 10 lbs. Use safe lifting techniques and
proper techniques when handling heavy objects. If necessary, obtain assistance to safely
lift the instrument.
CAUTION: If using a utility knife, extend the blade to appropriate length to avoid cutting
internal any components
6.1.1. Position the box upright and open the top flaps
(Fig. 1)
Fig. 1
Fig. 2
6 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Fig. 3
6.1.7. Remove the foam panels from the sides of the
instrument (Fig.4)
d
c
6.1.10. Save the box and foam pieces for future use e f
b a
Fig. 4
7 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
CAUTION: Always keep a distance of at least four (4) inches (10 cm)
between the rear of the instrument and the wall to allow for proper ventilation.
CAUTION: Place the instrument on a stable and level surface free of vibration.
Failure to do so may cause injury or malfunction of the unit.
8 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
6.4.3. Powering on
6.4.3.1. To power the unit on, press the On/Off switch located on the back of the instrument (Fig. 7)
6.4.3.2. Once the power button has been pressed the instrument will produce an audible beep, after
which the instrument is non-functional while the operating system boots up. This boot-up
process takes about 10-15 seconds.
9 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
7. User Interface
Fig. 11
10 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Fig. 12
Fig. 13
11 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Fig. 14
Fig. 15
12 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
The instrument is touch screen controlled and all functions can be done by selecting or inputting
data on the following screens:
The Main Menu screen (Fig. 17) is From the Main Menu screen (Fig. 17)
accessed by touching this symbol in the navigation to the Result Log,
upper left-hand corner found on many of Maintenance and Settings menus can
the menu screens, see Fig 16, below: occur:
Fig. 16
Fig. 17
13 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
The Result Log sub-menu is organized starting with the most recent test results shown first. Fig.18,
displayed icons, listed left to right, top to bottom are:
Records can be exported 7 days at a time as shown in Figure 18 or individually as shown in Figure
19. In Either case, once the “Papers” icon is selected, it will bring up the Export Selection screen
(Fig. 20):
Fig. 19 Fig. 20
14 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
The Maintenance sub-menu provides a “dashboard” for the user related functions for the miniiSED.
The circular icons with the green check mark indicate that the listed category is within preset limits.
When a miniiSED category is approaching a preset limit, the circular icon will turn yellow indicating
attention is needed soon. Should the category limit be exceeded, the icon will turn red and require
action to be taken. In Figure 21, displayed icons, top to bottom are:
Fig. 21
The Settings sub-menu (Fig. 22) provides the user access to functions to customize the operation
of the miniiSED for their laboratory environment. Using the guided Setup process when the
instrument is first turned on will preconfigure some of the typical settings at the time of device
installation. There are two categories of settings: General and Advanced. The General Settings
selection is accessible to the user via the Admin PIN, when activated. Otherwise it is not password
protected. Any User can view the Advanced Settings level but changes to these parameters is only
allowed once the Advanced level PIN is entered.
Fig. 22
15 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
The General Settings sub-menu (Fig. 23) provides the user access to functions to customize the
operation of the miniiSED for their laboratory environment. Using the guided Setup process will
preconfigure some of the typical settings at the time of device installation. Protection of the General
Settings can be disabled by selecting the Access option.
Fig. 23
The Advanced Settings sub-menu provides the user “Read-only” status of various operational
sensors and device settings (Figs. 24, 25, and 26). The Advanced Settings level is only accessible
to Alcor Technicians.
16 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Barcoded Tubes: Patient samples are read and identified by the instrument’s internal barcode
reader automatically as they are loaded into the instrument. All common laboratory barcodes are
supported, including Code 39, UPC and Code 93 formats. Note barcode location range:
Instances when patient identification cannot be read by the internal barcode reader or there is no
barcode present the miniiSED will provide the user an option to enter the patient data manually or
to run the test without entering additional data. If the latter option is selected the miniiSED will
assign a unique number to the test sample.
8. Calibration
miniiSED instruments are factory calibrated utilizing samples which are compared with results from a
unique Reference Instrument. The Reference Instrument is correlated with the reference Westergren
method. The instrument range is from 1 to 130mm/hr. During normal operation, parameters affecting
calibration are constantly monitored and, if not within expected limits, a warning is given, and further
testing prevented.
17 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
9. Results
9.1. Expected Values
The reference values found in the table below are averages found in men, women, children and
newborns. An increase in these values can be a sign of multiple different health issues that
should be diagnosed by a physician or qualified individual.
The ranges provided are for reference only. All laboratories should follow their laboratory’s
protocol for verifying their own reference ranges.
9.2. Performance
Correlation:
227 samples have been tested on miniiSED and compared with results obtained from the
Westergren method. The comparisons demonstrate equivalence to the Westergren method.
See recommended miniiSED Correlation Protocol (1017-07-001) for detailed instructions.
• Sample result range 0 to 130 mm/hr
• N=227
• Slope=1.0625
• Intercept=-0.2500
• R=0.923
** These values do not indicate the expected performance of your miniiSED correlation
Repeatability/Stability:
See miniiSED Precision Test Protocol (1017-07-003) for detailed instructions.
Carry-over:
See miniiSED Inter-Sample Carry Over Test Protocol (1017-07-002) for detailed instructions.
8 Wheeler M, Thomas, MD, Chairman, Department of Pathology and Immunology, Baylor College of Medicine, February 15, 2012:
emedicine.medscape.com/article/2085201-overview.
18 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Fig. 27 Fig. 28
Once a Test Credit card (Fig. 29) is inserted, the next screen will show: the number of credits
currently on the instrument, the serial number of the inserted Test Credit card and the number of
credits on the Test Credit card (Fig. 30). If the user taps the “Confirm” bubble, all of the available
credits will be transferred to the instrument. The total number of credits is then displayed and the
user is instructed to discard the the Test Credit card (Fig. 31).
19 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
NOTE: A washing cycle should be run prior to replacing the waste bottle.
WARNING: Dispose liquid waste container contents in a manner consistent with local
regulations and laboratory procedures.
3. Remove the waste bottle from the back of the instrument and
dispose according to your laboratory biologic waste protocol
5. Press Waste Container Empty reset button in the Maintenance sub-menu to reset
(Fig. 32).
Fig. 32
20 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
3. Remove the empty miniiWASH bottle, unscrew the cap and replace
it with a new miniiWASH bottle
4. Place the new miniiWASH bottle at the rear of the instrument and
firmly reconnect the LUER connector on the plastic screw
cap
5. Press Cleaning Fluid Replace reset button in the Maintenance sub-menu to reset (Fig. 33).
Fig. 33
21 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
WARNING: All other biohazardous waste should be deposited into a biohazard bag.
WARNING: Biohazard bags will be placed into a Medical Waste Management bin for
pick up.
WARNING: Dispose liquid waste container contents in a manner consistent with local
regulations and laboratory procedures.
22 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
WARNING: In the event that the instrument must be returned for service,
REMOVE ALL FLUID CONTAINERS BEFORE SHIPPING.
Any instrument containing accumulated blood must be cleaned prior to shipment to the
manufacturer. This decontamination is required by Federal Law (Title 48 and 49 of the Federal
Regulations) in accordance with the Environmental Protected Agency’s Regulations for Biohazard
Waste Management.
23 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Barcode Internal
Printer Accessory
Frequency 50/60 Hz
24 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A
Manufacturer’s Warranty
ALCOR Scientific Inc., warrants that this product is free from defects in materials and workmanship for
a period of One (1) Year from the date of original purchase (except as noted below). During the stated
one (1) year period, ALCOR Scientific shall, at its sole discretion, repair or replace at no charge to the
original end use purchaser or person receiving the product, any product found to be defective due to
material or workmanship. In the case of replacement, a new or reconditioned product may be provided
at ALCOR’s option. This warranty is limited to the repair or replacement due to defects in parts or
workmanship and shall not include any maintenance and repairs or replacement of parts due to normal
wear and tear. Parts required which were not defective shall be replaced at additional costs, and
ALCOR Scientific shall not be required to make any repairs or replace any parts which are necessitated
by abuse, accident, alteration, misuse, neglect, maintenance by other than ALCOR Scientific, an
authorized ALCOR service agent, or failure to operate the instrument in accordance with instructions.
Further, ALCOR Scientific extends no warranty for malfunction or damage to its products resulting from
improper or unreasonable use or maintenance; failure to follow operating instructions; connections to
improper voltage supply; unauthorized alteration or modification of original condition; damages caused
by inadequate packing or shipping procedures; loss of, damage to or corruption of stored data; and any
damage due to use of operating supplies other than those manufactured or recommended by ALCOR
Scientific. ALCOR Scientific reserves the right to make changes in design or software of this instrument
without obligation to incorporate such changes into previously manufactured instruments.
Disclaimer of Warranties
THIS WARRANTY IS EXPRESSLY MADE IN LIEU OF ANY AND ALL OTHER WARRANTIES
EXPRESS OR IMPLIED INCLUDING THE WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR USE.
This warranty is void if the label bearing the serial number has been removed or defaced.
Limitations of Liability
In no event shall ALCOR Scientific be liable for indirect, special or consequential damages, even if
ALCOR Scientific has been advised of the possibility of such damages. Any transport costs and risks
are not covered by this warranty. In the event the instrument shall be returned to ALCOR Scientific for
servicing, replacement or for other reasons, it must be shipped and received in original packaging.
Otherwise, additional charges may be incurred. Proof of purchase from an authorized ALCOR
distributor and proof of delivery may be required.
25 of 25
miniiSED™ | OPERATOR’S MANUAL
1017-09-001 Revision A