DMM7510 Multimeter Specs & Accuracy
DMM7510 Multimeter Specs & Accuracy
Keithley Instruments
28775 Aurora Road
7-1/2 Digit Graphical Sampling Multimeter
Cleveland, Ohio 44139 Specifications
1-800-935-5595
www.tek.com/keithley
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Model DMM7510 7-1/2 Digit
Graphical Sampling Multimeter instrument. Specifications are the standards against which the DMM7510 is
tested. Upon leaving the factory, the DMM7510 meets these specifications. Supplemental and typical
values are nonwarranted, apply at 23 °C (73 °F), and are provided solely as useful information.
Measurement accuracies are specified at the DMM7510 terminals under these conditions:
• Temperature 23 °C ±5 °C, 5 % to 80 % relative humidity, noncondensing
• After a 90-minute warmup period
• 1 PLC or 5 PLC; for NPLC settings less than 1 PLC, add appropriate ppm of range for peak noise
uncertainty from the RMS noise table
• Autozero enabled unless otherwise noted
• Remote sense operation or properly zeroed local operation
• Calibration period: One year or two years (calibration period may vary depending on customer
requirements)
• TACAL = Ambient temperature of last automatic calibration
• TCAL = Ambient temperature of last external calibration; factory calibration performed at 23 °C ±1 °C
DC VOLTAGE
1
20 % overrange on all ranges except 1 % for 1000 V range.
2
Relative to calibration accuracy.
3
Add per degree from TCAL ± 5 oC.
4
When properly zeroed using the Rel function with external cables.
5
Specified within 30 days of autocalibration, TOPER ± 5 oC from TACAL.
6
For signal levels greater than 500 V, add 0.02 ppm/V to the ppm of the readings specification for measurements exceeding 500 V.
DC VOLTAGE RATIO
For input signals ≥ 1 % of the range, ±[[VINPUT ppm of reading + VINPUT ppm of range * (VINPUT range/VINPUT input)]
ratio accuracy = + [VSENSE ppm of reading + VSENSE ppm of range * (VSENSE range/VSENSE input)]]
DC VOLTAGE CHARACTERISTICS
ADC linearity 1.0 ppm of reading + 1.0 ppm of range
Input impedance 100 mV to 10 V ranges: Selectable > 10 GΩ ǁ < 400 pF (auto) or
10 MΩ ±1 % (10 MΩ)
100 V to 1000 V ranges: 10 MΩ ±1 %
Input bias current < 50 pA at 23 °C under the following conditions: Autozero off or input
impedance 10 MΩ
Common mode current < 2.1 µA peak-peak in 1 MHz bandwidth
< 100 nA peak-peak in 1 kHz bandwidth
Common mode voltage 500 Vpeak LO terminal to chassis maximum
DC voltage autozero off error For ±1 °C and ≤ 10 minutes, add ± (8 ppm of reading + 15 µV)
7
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. VRMS noise is typical. Additional peak noise is
guaranteed.
8
With line sync on.
9
Add per degree from TCAL ± 5 oC.
RESISTANCE
100.00000 MΩ 18
10 Ω 0.69 µA ǁ 10 MΩ 800 + 30 2000 + 30 2000 + 30 2000 + 30 385 + 1
1.0000000 GΩ 18
100 Ω 0.69 µA ǁ 10 MΩ 9000 + 100 9000 + 100 9000 + 100 9000 + 100 3000 + 1
10
Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ
measurements. 1 Ω range is 4-wire only. For 2-wire, with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test
leads, add 100 mΩ to ppm of range uncertainty.
11
20 % overrange on all ranges.
12
Test current with offset compensation off.
13
Relative to calibration accuracy.
14
Add per degree from TCAL ±5 oC.
15
Specifications are for external cable and load capacitance < 1 nF.
16
For offset compensation on, add 10 ppm uncertainty to ppm of reading.
17
For 4-wire 1 MΩ, open lead detector on, add 10 ppm uncertainty to ppm of reading.
18
Specified for < 10 % lead resistance mismatch in HI and LO.
ACCURACY 19
Range 20 Resolution Test current 21 Accuracy ±[ppm of reading + ppm of range]
(±5 %) 24 hour 90 day 1 year 2 year Temperature
TCAL ±1 °C 22 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 23
1Ω 0.1 µΩ 10 mA 15 + 50 40 + 50 50 + 50 70 + 50 2.5 + 5
10 Ω 1 µΩ 10 mA 15 + 5 40 + 5 50 + 5 70 + 5 2.5 + 0.5
100 Ω 10 µΩ 1 mA 12 + 4 35 + 4 47 + 4 65 + 4 5 + 0.25
1 kΩ 100 µΩ 1 mA 12 + 3 30 + 3 41 + 3 65 + 3 5 + 0.25
24 100 µA 10 + 3 30 + 3 42 + 3 65 + 3 2.5 + 0.25
10 kΩ 1 mΩ
1 GΩ27 100 Ω 0.69 µA ǁ 10 MΩ 9000 + 200 9000 + 200 13000 + 200 14000 + 200 3000 + 1
1Ω – 9.2 V 9.5 V
10 MΩ to 1 GΩ 6.9 V 6.9 V –
19
Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ
measurements. 1 Ω range is 4-wire only. For 2-wire, with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test
leads, add 100 mΩ to ppm of range uncertainty.
20
20 % overrange on all ranges.
21
Test current with offset compensation off.
22
Relative to calibration accuracy.
23
Add per degree from TCAL ±5 °C.
24
Specifications are for external cable and load capacitance < 1 nF.
25
For offset compensation on, add 10 ppm of uncertainty to ppm of reading.
26
For 4-wire, 1 MΩ, open lead detection on, add 10 ppm uncertainty to ppm of reading.
27
Specified for < 10 % lead resistance mismatch in HI and LO.
28
Open circuit voltage is typical, measured from input HI to LO, SHI and SLO open. For 1 Ω to 1 MΩ ranges using an external digital multimeter
(DMM) set to 10 MΩ input impedance; for 10 MΩ to 1 GΩ ranges, set external DMM to >10 GΩ input impedance.
2-WIRE OHMS
RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY)29
Applies to ± ppm of range
Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
Add peak noise uncertainty to measurements for < 1 PLC
Examples:
10 kΩ at 0.006 PLC: 3 (from Accuracy table) + 5 (50 mΩ with Rel ) + 43 (additional peak noise uncertainty) = 51 ppm
of range
10 kΩ at 1 PLC: 3 + 5 + 0 = 8 ppm of range
29
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is
guaranteed.
30
With line sync on.
RESISTANCE CHARACTERISTICS
Maximum 4-wire ohms lead resistance 5 Ω per lead for 1 Ω range, 10 % of range per lead for 10 Ω to 1 kΩ
ranges; 1 kΩ per lead for all other ranges
Offset compensation Selectable on 4-wire, 1 Ω to 100 kΩ ranges
Open lead detector Default is off
Autozero off error For 2-wire ohms, ±1 °C and ≤ 10 minutes, add ±(8 ppm of reading)
and 1.5 mΩ for 10 Ω range, 15 mΩ for 100 Ω and 1 kΩ ranges,
150 mΩ for 10 kΩ range, 1.5 Ω for 100 kΩ range, and 15 Ω for all
other ranges
ACCURACY
Range31 Resolution Test Open circuit Accuracy ±[ppm of reading + ppm of range]
current 35 DUT
voltage32 24 hour 90 day 1 year 2 year Temperature
(±5 %) TCAL ±1 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient34
°C33
1.000000 Ω 1 µΩ 10 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 2.5 + 1
10.00000 Ω 10 µΩ 1 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 5+1
31
20 % overrange on all ranges, except 2.4 kΩ for the 10 K range.
32
Maximum clamp voltages are DC, typical accuracy is ±20 %. Add 20 % for offset compensation on.
33
Relative to calibration accuracy.
34
Add per degree from TCAL ±5 oC.
35
Test current with offset compensation off.
36
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is
guaranteed.
37
With line sync on.
DC CURRENT
ACCURACY
Maximum Accuracy ±[ppm of reading + ppm of range]
Range38 Resolution burden
voltage 24 hour 90 day 1 year 2 year Temperature
TCAL ±1 °C39 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient40
10.000000 µA 1 pA 15 mV 30 + 30 100 + 30 125 + 40 175 + 50 10 + 8
100.00000 µA 10 pA 15 mV 20 + 5 75 + 12 100 + 15 150 + 20 10 + 3
1.0000000 100 pA 15 mV 30 + 5 75 + 12 100 + 15 150 + 20 10 + 3
mA
10.000000 1 nA 20 mV 40 + 5 75 + 12 100 + 15 150 + 20 10 + 3
mA
100.00000 10 nA 200 mV 50 + 18 300 + 30 400 + 30 500 + 30 50 + 5
mA
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 450 + 50 500 + 50 10 + 10
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 450 + 40 500 + 40 10 + 10
41
10.000000 A 1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10
38
20 % overrange supported for all ranges except for 3 A and 10 A, which are 1 % supported.
39
Relative to calibration accuracy.
40
Add per degree from TCAL ±5 oC.
41
Rear input terminals only.
DC CURRENT CHARACTERISTICS
Range 10 µA 100 µA 1 mA 10 mA 100 mA 1A 3A 10 A43
Effective internal shunt value 45 1 kΩ 100 Ω 10 Ω 1Ω 0.1 Ω 0.1 Ω 0.1 Ω 0.005 Ω
Autozero off error: For ±1 °C and ≤ 10 150 pA 1.5 nA 15 nA 150 nA 15 µA 150 µA 150 µA 3 mA
minutes add ±(8 ppm of reading + range
error)
Overload recovery:
For each additional sustained amp beyond
±1.5 A, add the following initial ppm of range 15500 1800 150 150 6500 200
error until thermally settled after overload
recovery
42
Noise values are based on 1000 readings with autozero on and AMPS terminal open. RMS noise is typical. Additional peak noise is
guaranteed.
43
Rear input terminals only.
44
With line sync on.
45
Values are typical and guaranteed by design.
TEMPERATURE
THERMISTOR
Types: 2.252 kΩ, 5 kΩ, and 10 kΩ
Type Range Resolution Accuracy ± °C
THERMOCOUPLE
Types: B, E, J, K, N, R, S, T
Type Range Resolution Accuracy ± °C
46
Add per degree from TCAL ±5 oC; specifications without autocalibration.
47
For 3-wire RTD, accuracy is for < 0.1 Ω lead resistance mismatch for input HI and LO. Add 0.25 °C/ 0.1 Ω of HI-LO lead resistance mismatch.
48
Exclusive of cold-junction errors.
CONTINUITY
Range 49 Resolution Test Open circuit Accuracy ±[ppm of reading + ppm of range]
current voltage
2 year Temperature coefficient 50
TCAL ±5 °C
CONTINUITY CHARACTERISTICS
Continuity high limit User-selectable; default 10 Ω
CAPACITANCE
Accuracies specified for additional cable and stray capacitance properly zeroed with the Rel function.
ACCURACY
Range 51 Resolution Charge Maximum Accuracy ±[% of reading + % of range]
current 52, 53 circuit
voltage 2 year Temperature coefficient50
TCAL ±5 °C
1.0000 nF 0.1 pF 1.1 µA 2.8 V 1 + 0.2 0.15 + 0.05
10.000 nF 1 pF 1.1 µA 2.8 V 1 + 0.1 0.15 + 0.01
100.00 nF 10 pF 10 µA 3V 0.4 + 0.1 0.01 + 0.01
1.0000 µF 0.1 nF 100 µA 3V 0.4 + 0.1 0.01 + 0.01
10.000 µF 1 nF 100 µA 3V 0.4 + 0.1 0.01 + 0.01
100.00 µF 10 nF 1 mA 3V 0.4 + 0.1 0.01 + 0.01
1000.0 µF 0.1 µF 10 mA 3V 0.5 + 0.1 0.01 + 0.01
DIODE
Voltage Resolution Bias level Accuracy ±[ppm of reading + ppm of range]
measure (selectable)
90 day 1 year 2 year Temperature
range51 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient50
10.000000 V 1 µV 10 µA / 100 µA / 1 mA 20 + 5 30 + 5 45 + 5 2.5 + 1
49
Specifications exclude lead resistance.
50
Add per degree from TCAL ±5 oC; specifications without autocalibration.
51
20 % overrange on all ranges.
52
Charging current values are typical, guaranteed by design.
53
Discharge current limited to < 13 mA.
DIGITIZE VOLTAGE
TYPICAL DC COUPLED
Analog Maximum flatness error Total harmonic distortion (THD) Settling time
Range bandwidth 3 Hz to 1 kHz64 1 kHz signal (0.5 %)
(-3 dB) (-1 dB FS)65
100.000 V 20 kHz 67 0.1 dB 1.3 % 160 µs
1000.00 V 20 kHz 0.1 dB 1.8 % 80 µs
54
For DC coupling, 20 % overrange for 100 mV to 100 V. For AC coupling, 500 % overrange 100 mV to 100 V. 1 % for 1000 V range DC and
AC coupling.
55
Accuracy with sample rate 1 k per second, aperture auto, and 100 reading buffer average.
56
Power up default is 4½ digits.
57
User-selectable.
58
Add per degree from TCAL ± 5 %.
59
For 100 V range, input impedance auto and without ACAL, add 100 ppm of range additional uncertainty and 15 ppm/°C additional uncertainty
for "of range" temperature coefficient for operation outside of TCAL ±5 °C.
60
For signal levels greater than 500 V, add 0.02 ppm/V to the ppm of the readings specification for measurements exceeding 500 V.
61
Accuracy with sample rate 1 M per second and aperture 1 µs.
62
Verified with sine wave input and DC content ≤ 3 % of range.
63
For AC coupling, maximum crest factor of 5.
64
For DC coupled, 0 dB reference frequency is 3 Hz. For AC coupled, 0 dB reference frequency is 1 kHz. For AC coupled operation below 1 kHz,
add 0.1 dB.
65
Exclusive of source input noise.
66
With AC coupling frequency = 3 Hz and AC coupling filter = Slow.
67
For input impedance auto, bandwidth is 6 kHz.
TYPICAL AC COUPLED
100
ppm of range
100mV
10 1V
10V
100V
1 1000V
0.1
1k 2k 5k 10k 20k 50k 100k 200k 500k 1M
Samples per second
68
Specified with aperture Auto and 4-wire short on input terminals. For 100 V range, input impedance 10 MΩ, multiply by 2.5. For all ranges and
sample rate > 1 k, add an additional 3× RMS noise uncertainty to ppm of range.
-60
-70
dB of range
100mV
-80
1V
-90 10V
100V
-100
1000V
-110
-120
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second
16
14
Bits
100mV
12 1V
10V
10
8
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second
69
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤ 5 k, and 1 kHz sine wave for sample rate ≥ 10 k. Distortion is calculated
using first five harmonics.
70
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤5 k, and 1 kHz sine wave for sample rate ≥10 k. For the 100 V and 1000 V
ranges, use the 1 V and 10 V range ENOB, respectively; guaranteed by design.
DIGITIZE CURRENT
DC ACCURACY 71
Accuracy ± [ppm of reading + ppm of range]
72 73
Burden
Range Resolution 90 day 1 year 2 year
voltage Temperature
TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 74
10.0000 0.1 nA 15 mV 150 + 75 160 + 75 170 + 75 30 + 15
µA
100.000 1 nA 15 mV 150 + 75 160 + 75 170 + 75 30 + 15
µA
1.00000 10 nA 15 mV 150 + 75 160 + 75 170 + 75 30 + 15
mA
10.0000 100 nA 20 mV 150 + 75 160 + 75 170 + 75 30 + 15
mA
100.000 1 µA 200 mV 340 + 100 450 + 100 560 + 100 50 + 20
mA
1.00000 A 10 µA 400 mV 400 + 110 500 + 110 600 + 110 50 + 25
3.00000 A 100 µA 1300 mV 650 + 150 900 + 150 900 + 150 50 + 25
10.0000 100 µA 650 mV 950 + 350 1500 + 350 2000 + 350 50 + 25
A 75
71
Accuracy with sample rate 1 k per second, aperture auto, and 100 reading buffer average.
72
20 % overrange on all ranges except 3.3 % for 3 A and 10 A ranges.
73
Power up default is 4½ digits.
74
Add per degree from TCAL ±5 oC.
75
Rear input terminals only.
76
Verified with sine wave input and DC content ≤ 3 % of range. For flatness error, 0 dB reference frequency is 3 Hz.
77
For the 1 A, 3 A, and 10 A ranges, use the 100 mA range accuracy; guaranteed by design.
78
10 A flatness verified to 10 kHz; 100 kHz guaranteed by design.
100 10uA
ppm of range
100uA
1mA
10mA
10 100mA, 1A
3A
10A
1
1k 2k 5k 10k 20k 50k 100k 200k 500k 1M
Samples per second
-80
dB of range
-90 10uA
100uA
-100 1mA
10mA
-110 100mA
-120
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second
79
Specified with aperture Auto and open input terminals. For all ranges and for ≥1 k sample rate, add an additional 3× RMS noise uncertainty to
ppm of range.
80
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤ 5 k, and 1 kHz sine wave for sample rate ≥ 10 k. Distortion is calculated
using first five harmonics. For the 1 A, 3 A, and 10 A ranges, use the 100 mA range accuracy; guaranteed by design.
16
14 10uA
Bits
100uA
12 1mA
10mA
100mA
10
8
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second
DIGITIZER CHARACTERISTICS
Maximum resolution 18 bits
Measurement input coupling DC or AC (voltage only)
82
Sampling rate Programmable 1 k through 1 million
Volatile sample memory with timestamp 27.5 million
Minimum record time 1 µs
Timestamp resolution 1 ns with standard or full buffer style
1 µs with compact buffer style
Timestamp accuracy With standard or full buffer style, 20 ns between adjacent readings,
with total buffer time < 2 s
With compact buffer style, 2 µs adjacent readings, with total buffer
buffer time < 2 s
Maximum record length 8 million
81
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤5 k, and 1 kHz sine wave for sample rate ≥10 k. For the 1 A, 3 A, and 10 A
ranges, use the 100 mA ENOB; guaranteed by design.
82
Sample rate is not continuously adjustable. For valid discrete settings, see the Model DMM7510 Reference Manual.
100.0000 mV 0.1 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
1.000000 V 1 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
Voltage 84
10.00000 V 10 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
100.0000 V 100 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
700.000 V 1 mV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
Temperature – – 0.01 + 0.003 0.03 + 0.003 0.005 + 0.003 0.006 + 0.005 0.01 + 0.006 0.03 + 0.01
coefficient /
°C (all
ranges)
3.000000 A 1 µA 1.0 + 0.05 0.30 + 0.05 0.20 + 0.05 0. 88 + 0.05 2.0 + 0.05
10.00000 A 85 10 µA 1.0 + 0.05 0.40 + 0.05 0.40 + 0.05 0. 88 + 0.05 2.0 + 0.05
Temperature – – 0.10 + 0.004 0.030 + 0.004 0.005 + 0.003 0.006 + 0.005 0.006 + 0.005
coefficient /
°C (all
ranges)
20 Hz to 30 Hz 0 0.3
30 Hz to 50 Hz 0 0
50 Hz to 100 Hz 0 0 4.0
100 Hz to 200 Hz 0 0 0.72
200 Hz to 300 Hz 0 0 0.18
300 Hz to 500 Hz 0 0 0.07
> 500 Hz 0 0 0
83
20 % overrange on AC functions except 1 % on 700 V, 3.33 % on 3 A and 1 % on 10 A. Default resolution is 6½ digits.
84
Specifications are for detector bandwidth of 3 Hz and sine wave inputs > 5 % of range. Detector bandwidth of 3 Hz and 30 Hz are multisample
A/D conversions. Detector bandwidth of 300 Hz is a single A/D conversion, programmable from 0.0005 PLC to 15 PLC (60 Hz), 12 PLC
(50 Hz). Default condition set to 1 PLC.
85
Rear input terminals only.
AC VOLTAGE CHARACTERISTICS
Measurement method AC-coupled, true RMS
Input impedance 1 MΩ ± 2 % ǁ < 150 pF
Volt*Hertz product < 2.1 x 107 V*Hz verified; input frequency verified for < 300 kHz
AC CURRENT CHARACTERISTICS
Measurement method AC-coupled, true RMS
Range 1 mA 10 mA 100 mA 1A 3A 10 A 87
Burden voltage (RMS) < 16 mV < 20 mV < 0.2 V < 0.4 V < 1.3 V < 0.65 V
Overload recovery: For each
additional sustained ampere beyond 0.006 0.006 0.12 0.05
±1.5 A, add the following initial % of
range error until thermally settled after
overload recovery
86
Applies for non-sine wave inputs, DC content ≤ 3 % of range, maximum crest factor ≤ 5.0.
87
Rear input terminals only.
MEASUREMENT ACCURACY 88
Aperture Measurement resolution Accuracy ±[ppm of reading + ppm of aperture time]
Frequency: 3 Hz to 500 kHz
Period: 333 ms to 2 µs
1 year 2 year
TCAL ±5 °C TCAL ±5 °C
88
Specified for square wave inputs. Input signal must be >10 % of ACV range. If input is<20 mV on the 100 mV range, then the frequency must
be >10 Hz. For sine wave inputs, frequency must be >100 Hz. For frequencies ≤100 Hz, threshold level ≤50 % of input signal and ≤7 Hz,
threshold level ≤3 % of range.
89
Threshold range is voltage RMS and threshold level voltage peak. Specified with 1 KHz square wave. 100 V and 700 V threshold ranges
guaranteed by design.
Functions:
AC voltage, AC current
Detector bandwidth (Hz) Digits Measurements into Measurements into
buffer computer
3 6½ 0.5 (0.5) 0.5 (0.5)
30 6½ 3.3 (3.3) 3.3 (3.3)
94
300 6½ 59.8 (49.8) 55 (46)
94
300 3½ 26200 (26200) 24500 (24500)
DIGITIZE, TYPICAL
Sampling rate Digits Resolution Measurements into computer93
10 kS/s 5½ 18 9700
20 kS/s 4½ 16 19000
50 kS/s 4½ 16 44400
100 kS/s 4½ 15 80000
1 MS/s 3½ 12 108000
90
Reading speeds for autozero off, fixed range, autodelay off. Offset compensation off and open lead detector off where applicable.
91
Buffer measurements: For < 0.2 PLC, multisample, single buffer transfer binary reading only.
92
PC measurements: For 1 and 0.2 PLC single reading and single transfer to computer (USB).
93
Reading rates using factory default operating conditions and autorange off, autodelay off. Speeds include measurement and data transfer out
of the USB. ≥1000 readings with binary transfer over USB.
94
For bandwidth 300 Hz, autozero off, 6½ digits at 1 PLC, 3½ digits at 0.0005 PLC.
DC current 7 1.3
95
Frequency or period 7 1.3
AC voltage or AC current 7 1.3
Digitize voltage or current 7 1.3
AC current 1 mA
AC voltage 1V
95
For DC voltage or 2-wire ohms to frequency or period, 10 ms aperture. For AC current or AC voltage, detector bandwidth is 300 Hz.
TRIGGERING
Time base accuracy 25 ppm
Trigger source Analog DC voltage, DC current, or any system trigger
Trigger coupling DC or AC (DC voltage function only)
Input trigger latency96, 97, 98 < 225 ns
Input trigger jitter96,97 < 50 ns
Sample period jitter96,97 < 1 ns
ANALOG TRIGGERING 99
96
Guaranteed by design; for digital I/O only.
97
Stimulus command required to meet specifications.
98
If using trigger model, add 200 ns uncertainty.
99
For DC or AC coupled, the trigger level can be set up to 100 % of measure range.
100
Rising or falling edge triggering supported. Window trigger requires setting two independent levels.
101
Trigger event occurs after the threshold crossing at a time determined by total trigger latencies.
102
Accuracy specifications require user ACAL and are verified with level trigger amplitude set to 50 % of range with a 100 Hz sine wave at 100 %
full scale of range. High frequency rejection is off. NPLC 0.0005 (DC voltage/DC current) or aperture 1 µs for digitize voltage or digitize
current. Specified for fixed range, autozero off. For digitized DC voltage AC coupled, add 0.5 %. For DC current and digitized DC current 3 A
or 10 A ranges, add an additional 2 %.
SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the DMM7510 Precision Sampling
Digital Multimeter instrument. Specifications are the standards against which the DMM7510 is tested. Upon
leaving the factory, the DMM7510 meets these specifications. Supplemental, typical, and characteristic
values are non-warranted, apply at 23 °C, and are provided solely as useful information. All specifications
apply to front or rear terminal inputs, except 10 A specifications (rear-terminals only).
Input protection 1010 V DC (715 VRMS V AC) all ranges and functions on HI and LO
terminals; 350 V all ranges and functions on sense HI, sense LO
terminals; 250 V rated current input terminal; fused 3 A and 10 A
ranges; current input terminals protected to 1 kV
3 A input fuse protection 3.5 A, 1 kV fast blow type; Keithley part number
DMM7510-FUSE-3A
10 A input fuse protection 11 A, 1 kV fast blow type; Keithley part number
DMM7510-FUSE-10A
AC voltage input Maximum DC voltage: 1000 V on any AC voltage range
Common mode isolation 500 V DC or AC Vpeak LO to chassis
All terminals > 10 GΩ, < 350 pF any terminal to chassis
Power line Universal input, 100 V to 240 V
Line frequency 50 Hz or 60 Hz, automatically sensed at power-up
Power consumption 60 VA
Operating environment Specified for 0 °C to 50 °C, ≤ 80 % relative humidity at 35 °C,
altitude up to 2000 meters
Storage environment -30 °C to 70 °C
Real time clock Lithium battery backup (3+ years battery life)
EMC Conforms to European Union EMC Directive
Safety NRTL listed to UL61010-1, and CSA C22.2 No 61010-1; conforms
with European Union Low Voltage Directive
Vibration MIL-PRF-28800F Class 3, Random
Warm-up 90 minutes to rated accuracy
Input signal connections Front and rear safety banana jacks
Cooling Forced air, fixed speed
Dimensions Without handle and bumpers: 88 mm high × 213 mm wide ×
410 mm deep (3.46 in. × 8.39 in. × 16.13 in.)
With handle and bumpers (bench configuration): 106 mm high ×
255 mm wide × 425 mm deep (4.18 in. × 10.05 in. × 16.75 in.)
Shipping weight (with bumpers and handle) 4.08 kg (9.0 lb)
Shipping weight (without bumpers and handle) 3.63 kg (8.0 lb)