DMM7510 Multimeter Specs & Accuracy

This document contains specifications and supplemental information for the Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter instrument. It provides detailed specifications for DC voltage, resistance, and other measurements over various time periods and temperature ranges.

Uploaded by

Santonio
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
68 views25 pages

DMM7510 Multimeter Specs & Accuracy

This document contains specifications and supplemental information for the Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter instrument. It provides detailed specifications for DC voltage, resistance, and other measurements over various time periods and temperature ranges.

Uploaded by

Santonio
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 25

Model DMM7510

Keithley Instruments
28775 Aurora Road
7-1/2 Digit Graphical Sampling Multimeter
Cleveland, Ohio 44139 Specifications
1-800-935-5595
www.tek.com/keithley

SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the Model DMM7510 7-1/2 Digit
Graphical Sampling Multimeter instrument. Specifications are the standards against which the DMM7510 is
tested. Upon leaving the factory, the DMM7510 meets these specifications. Supplemental and typical
values are nonwarranted, apply at 23 °C (73 °F), and are provided solely as useful information.
Measurement accuracies are specified at the DMM7510 terminals under these conditions:
• Temperature 23 °C ±5 °C, 5 % to 80 % relative humidity, noncondensing
• After a 90-minute warmup period
• 1 PLC or 5 PLC; for NPLC settings less than 1 PLC, add appropriate ppm of range for peak noise
uncertainty from the RMS noise table
• Autozero enabled unless otherwise noted
• Remote sense operation or properly zeroed local operation
• Calibration period: One year or two years (calibration period may vary depending on customer
requirements)
• TACAL = Ambient temperature of last automatic calibration
• TCAL = Ambient temperature of last external calibration; factory calibration performed at 23 °C ±1 °C

DC VOLTAGE

ACCURACY (INPUT IMPEDANCE AUTO)


Range 1 Resolution Input Accuracy ±[ppm of reading + ppm of range]
impedance
24 hour 90 day 1 year 2 year Temperature
TCAL ±1 °C 2 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 3
100.00000 mV 4 10 nV > 10 GΩ or 6+9 12 + 9 18 + 9 29 + 9 0.1 + 2.5
10 MΩ ±1 %
1.0000000 V4 100 nV > 10 GΩ or 4+1 9+2 15+ 2 26 + 2 0.1 + 0.5
10 MΩ ±1 %
10.000000 V4 1 µV > 10 GΩ or 2 + 0.7 9 + 1.2 14 + 1.2 22 + 1.2 0.1 + 0.05
10 MΩ ±1 %
100.00000 V4 10 µV 10 MΩ ±1 % 8+3 [18 + 5] 5 [22 + 5]5 [30 + 5]5 [0.15 + 0.05]5
35 + 5 40 + 5 45 + 5 2.0 + 0.5
4,6 5 5 5
1000.0000 V 100 µV 10 MΩ ±1 % 8+3 [19 + 5] [23 + 5] [31 + 5] [0.15 + 0.05]5
35 + 5 40 + 5 45 + 4 2.0 + 0.5

1
20 % overrange on all ranges except 1 % for 1000 V range.
2
Relative to calibration accuracy.
3
Add per degree from TCAL ± 5 oC.
4
When properly zeroed using the Rel function with external cables.
5
Specified within 30 days of autocalibration, TOPER ± 5 oC from TACAL.
6
For signal levels greater than 500 V, add 0.02 ppm/V to the ppm of the readings specification for measurements exceeding 500 V.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 1 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY) 7


 Applies to ± ppm of range
 Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
 Add peak noise uncertainty to measurements for < 1 PLC
 Input impedance set to auto
Examples:
 10 V at 0.006 PLC: 1.2 (from Accuracy table) + 11 (additional peak noise uncertainty) = 12.2 ppm of range
 10 V at 1 PLC: 1.2 + 0 = 1.2 ppm of range
NPLC Digits 100 mV 1V 10 V 100 V 1000 V
5 7½ 0.5 0.08 0.06 0.3 0.06
1 7½ 0.5 0.09 0.07 0.4 0.07
0.2 8 6½ 2 (10) 0.2 (1.6) 0.1 (1.1) 1.1 (9.4) 0.1 (1)
0.2 6½ 2 (12) 0.2 (1.6) 0.1 (1) 1.1 (8.9) 0.2 (1.1)
0.06 5½ 3 (17) 0.4 (2.7) 0.3 (2.1) 3 (17) 0.3 (2.4)
0.006 4½ 6 (42) 3 (18) 1 (11) 20 (100) 3 (18)
0.0005 3½ 30 (220) 20 (150) 20 (130) 120 (690) 20 (150)

DC VOLTAGE SENSE ACCURACY


Range Accuracy ±[ppm of reading + ppm of range]

24 hour 90 day 1 year 2 year Temperature


TCAL ±1 °C TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 9
100.00000 mV 6 + 14 12 + 14 18 + 14 29 + 14 0.1 + 2.5

1.0000000 V 4 + 1.5 9+3 15 + 3 26 + 3 0.1 + 0.5


10.000000 V 2 + 1.0 9 + 1.8 14 + 1.8 22 + 1.8 0.1 + 0.05

DC VOLTAGE RATIO
For input signals ≥ 1 % of the range, ±[[VINPUT ppm of reading + VINPUT ppm of range * (VINPUT range/VINPUT input)]
ratio accuracy = + [VSENSE ppm of reading + VSENSE ppm of range * (VSENSE range/VSENSE input)]]

DC VOLTAGE CHARACTERISTICS
ADC linearity 1.0 ppm of reading + 1.0 ppm of range
Input impedance 100 mV to 10 V ranges: Selectable > 10 GΩ ǁ < 400 pF (auto) or
10 MΩ ±1 % (10 MΩ)
100 V to 1000 V ranges: 10 MΩ ±1 %
Input bias current < 50 pA at 23 °C under the following conditions: Autozero off or input
impedance 10 MΩ
Common mode current < 2.1 µA peak-peak in 1 MHz bandwidth
< 100 nA peak-peak in 1 kHz bandwidth
Common mode voltage 500 Vpeak LO terminal to chassis maximum
DC voltage autozero off error For ±1 °C and ≤ 10 minutes, add ± (8 ppm of reading + 15 µV)

7
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. VRMS noise is typical. Additional peak noise is
guaranteed.
8
With line sync on.
9
Add per degree from TCAL ± 5 oC.

Specifications and characteristics are subject to change without notice.


Page 2 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

NORMAL MODE REJECTION


For DC voltage, line frequency ±0.1 %
5 PLC 1 PLC ≤ 0.2 PLC ≤ 0.01 PLC
Line sync on 110 dB 90 dB 45 dB 
Line sync off 60 dB 60 dB  

COMMON MODE REJECTION


For DC voltage and 1 kΩ unbalanced in LO terminal; AC CMRR is 70 dB

NPLC 5 1 0.2 ≤ 0.2

Line sync On On On Off

CMRR 140 dB 140 dB 120 dB 80 dB

RESISTANCE

ENHANCED ACCURACY (WITHIN 30 DAYS OF AUTOCALIBRATION, TOPER ± 5 °C FROM TACAL) 10


Range 11 Resolution Test current 12 Accuracy ±[ppm of reading + ppm of range]
(± 5 %) 24 hour 90 day 1 year 2 year Temperature
TCAL ±1 °C 13 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 14

1.0000000 Ω 0.1 µΩ 10 mA 15 + 50 30 + 50 30 + 50 30 + 50 0.15 + 0.1

10.000000 Ω 1 µΩ 10 mA 15 + 5 30 + 5 30 + 5 30 + 5 0.15 + 0.1

100.00000 Ω 10 µΩ 1 mA 12 + 4 27 + 4 27 + 4 27 + 4 0.15 + 0.1

1.0000000 kΩ 100 µΩ 1 mA 12 + 3 24 + 3 24 + 3 24 + 3 0.15 + 0.1

10.000000 kΩ 15 1 mΩ 100 µA 13 + 3 30 + 3 30 + 3 30 + 3 0.15 + 0.1

100.00000 kΩ15,16 10 mΩ 10 µA 13 + 3 30 + 3 30 + 3 30 + 3 0.15 + 0.1

1.0000000 MΩ15, 17 100 mΩ 10 µA 14 + 3 30 + 4 30 + 4 30 + 4 0.15 + 0.1

10.000000 MΩ 18 1Ω 0.69 µA ǁ 10 MΩ 150 + 6 200 + 10 200 + 10 200 + 10 70 + 1

100.00000 MΩ 18
10 Ω 0.69 µA ǁ 10 MΩ 800 + 30 2000 + 30 2000 + 30 2000 + 30 385 + 1

1.0000000 GΩ 18
100 Ω 0.69 µA ǁ 10 MΩ 9000 + 100 9000 + 100 9000 + 100 9000 + 100 3000 + 1

10
Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ
measurements. 1 Ω range is 4-wire only. For 2-wire, with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test
leads, add 100 mΩ to ppm of range uncertainty.
11
20 % overrange on all ranges.
12
Test current with offset compensation off.
13
Relative to calibration accuracy.
14
Add per degree from TCAL ±5 oC.
15
Specifications are for external cable and load capacitance < 1 nF.
16
For offset compensation on, add 10 ppm uncertainty to ppm of reading.
17
For 4-wire 1 MΩ, open lead detector on, add 10 ppm uncertainty to ppm of reading.
18
Specified for < 10 % lead resistance mismatch in HI and LO.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 3 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

ACCURACY 19
Range 20 Resolution Test current 21 Accuracy ±[ppm of reading + ppm of range]
(±5 %) 24 hour 90 day 1 year 2 year Temperature
TCAL ±1 °C 22 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 23

1Ω 0.1 µΩ 10 mA 15 + 50 40 + 50 50 + 50 70 + 50 2.5 + 5

10 Ω 1 µΩ 10 mA 15 + 5 40 + 5 50 + 5 70 + 5 2.5 + 0.5

100 Ω 10 µΩ 1 mA 12 + 4 35 + 4 47 + 4 65 + 4 5 + 0.25

1 kΩ 100 µΩ 1 mA 12 + 3 30 + 3 41 + 3 65 + 3 5 + 0.25
24 100 µA 10 + 3 30 + 3 42 + 3 65 + 3 2.5 + 0.25
10 kΩ 1 mΩ

100 kΩ24,25 10 mΩ 10 µA 13 + 3 38 + 3 50 + 3 65 + 3 5+1

1 MΩ24, 26 100 mΩ 10 µA 14 + 3 38 + 5 50 + 5 65 + 5 5+1

10 MΩ 27 1Ω 0.69 µA ǁ 10 MΩ 150 + 6 200 + 10 400 + 10 600 + 12 70 + 1

100 MΩ27 10 Ω 0.69 µA ǁ 10 MΩ 800 + 30 2000 + 30 2000 + 30 2600 + 30 385 + 1

1 GΩ27 100 Ω 0.69 µA ǁ 10 MΩ 9000 + 200 9000 + 200 13000 + 200 14000 + 200 3000 + 1

RESISTANCE OPEN CIRCUIT DC VOLTAGE 28


Offset compensation off Offset compensation on
Range20 2-wire
4-wire 4-wire

1Ω – 9.2 V 9.5 V

10 Ω 9.2 V 9.2 V 9.5 V

100 Ω, 1 kΩ 14.0 V 14.2 V 14.3 V

10 kΩ 9.5 V 9.5 V 0.0 V

100 kΩ, 1 MΩ 12.7 V 14.3 V 0.0 V (100 kΩ range only)

10 MΩ to 1 GΩ 6.9 V 6.9 V –

19
Specifications are for 4-wire resistance, offset compensation on for ≤10 kΩ measurements, and offset compensation off for ≥10 kΩ
measurements. 1 Ω range is 4-wire only. For 2-wire, with Rel, add 50 mΩ to ppm of range uncertainty. Without Rel and with Model 1756 test
leads, add 100 mΩ to ppm of range uncertainty.
20
20 % overrange on all ranges.
21
Test current with offset compensation off.
22
Relative to calibration accuracy.
23
Add per degree from TCAL ±5 °C.
24
Specifications are for external cable and load capacitance < 1 nF.
25
For offset compensation on, add 10 ppm of uncertainty to ppm of reading.
26
For 4-wire, 1 MΩ, open lead detection on, add 10 ppm uncertainty to ppm of reading.
27
Specified for < 10 % lead resistance mismatch in HI and LO.
28
Open circuit voltage is typical, measured from input HI to LO, SHI and SLO open. For 1 Ω to 1 MΩ ranges using an external digital multimeter
(DMM) set to 10 MΩ input impedance; for 10 MΩ to 1 GΩ ranges, set external DMM to >10 GΩ input impedance.

Specifications and characteristics are subject to change without notice.


Page 4 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

4-WIRE OHMS (≤ 10 kΩ) OFFSET COMPENSATION ON


RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY) 29
 Applies to ± ppm of range
 Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
 Add peak noise uncertainty to measurements for < 1 PLC
Examples
 1 kΩ at 0.006 PLC: 3 (from Accuracy table) + 26 (additional peak noise uncertainty) = 29 ppm of range
 1 kΩ at 1 PLC: 3 + 0 = 3 ppm of range

NPLC Digits 1Ω 10 Ω 100 Ω 1 kΩ 10 kΩ


5 7½ 2.8 0.3 0.3 0.07 0.3
1 7½ 4.2 0.4 0.4 0.12 0.5
0.2 30 6½ 30 (160) 3 (13) 3 (13) 0.4 (2.6) 1.2 (8.2)
0.2 6½ 50 (250) 5 (22) 5 (22) 0.6 (3.2) 1.2 (8.3)
0.06 5½ 110 (490) 11 (47) 11 (46) 1.1 (6.6) 2 (16)
0.006 4½ 110 (710) 10 (70) 10 (70) 4 (26) 10 (60)
0.0005 3½ 520 (3420) 50 (340) 50 (340) 40 (220) 50 (300)

2-WIRE OHMS
RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY)29
 Applies to ± ppm of range
 Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
 Add peak noise uncertainty to measurements for < 1 PLC
Examples:
 10 kΩ at 0.006 PLC: 3 (from Accuracy table) + 5 (50 mΩ with Rel ) + 43 (additional peak noise uncertainty) = 51 ppm
of range
 10 kΩ at 1 PLC: 3 + 5 + 0 = 8 ppm of range

NPLC Digits 10 Ω 100 Ω 1 kΩ 10 kΩ


5 7½ 1.1 0.8 0.1 0.2
1 7½ 0.6 0.6 0.09 0.4
0.230 6½ 2 (17) 2 (10) 0.2 (1.5) 0.8 (6.3)
0.2 6½ 2 (17) 2 (14) 0.3 (1.6) 0.8 (6.4)
0.06 5½ 3 (22) 3 (19) 0.4 (3.7) 2 (12)
0.006 4½ 6 (50) 6 (50) 3 (21) 6 (43)
0.0005 3½ 30 (300) 30 (230) 20 (150) 30 (210)

29
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is
guaranteed.
30
With line sync on.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 5 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

RESISTANCE CHARACTERISTICS
Maximum 4-wire ohms lead resistance 5 Ω per lead for 1 Ω range, 10 % of range per lead for 10 Ω to 1 kΩ
ranges; 1 kΩ per lead for all other ranges
Offset compensation Selectable on 4-wire, 1 Ω to 100 kΩ ranges
Open lead detector Default is off
Autozero off error For 2-wire ohms, ±1 °C and ≤ 10 minutes, add ±(8 ppm of reading)
and 1.5 mΩ for 10 Ω range, 15 mΩ for 100 Ω and 1 kΩ ranges,
150 mΩ for 10 kΩ range, 1.5 Ω for 100 kΩ range, and 15 Ω for all
other ranges

For 4-wire ohms, ±1 °C and ≤ 10 minutes, add ±(8 ppm of reading)


Input current limit For signals with a magnitude of +12 V to +40 V or -12 V to -40 V:
±13 mA source or sink, typical
For signals with a magnitude of greater than +40 V or -40 V:
±130 µA source or sink, typical

DRY CIRCUIT RESISTANCE

ENHANCED ACCURACY (WITHIN 30 DAYS OF AUTOCALIBRATION, TOPER ±5 ºC FROM TACAL)


Range 31 Resolution Test Open circuit Accuracy ±[ppm of reading + ppm of range]
current35 DUT
voltage 32 24 hour 90 day 1 year 2 year Temperature
(±5 %) TCAL ±1 °C 33 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 34

1.000000 Ω 1 µΩ 10 mA 25 mV 25 + 80 50 + 80 50 + 80 50 + 80 1.5 + 0.1

10.00000 Ω 10 µΩ 1 mA 25 mV 25 + 80 50 + 80 50 + 80 50 + 80 1.5 + 0.1

100.0000 Ω 100 µΩ 100 µA 25 mV 25 + 80 90 + 80 90 + 80 90 + 80 1.5 + 0.1

1.000000 kΩ 1 mΩ 10 µA 25 mV 25 + 80 180 + 80 180 + 80 180 + 80 1.5 + 0.1

10.00000 kΩ 10 mΩ 5 µA 25 mV 25 + 80 320 + 80 320 + 80 320 + 80 1.5 + 0.1

ACCURACY
Range31 Resolution Test Open circuit Accuracy ±[ppm of reading + ppm of range]
current 35 DUT
voltage32 24 hour 90 day 1 year 2 year Temperature
(±5 %) TCAL ±1 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient34
°C33

1.000000 Ω 1 µΩ 10 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 2.5 + 1

10.00000 Ω 10 µΩ 1 mA 25 mV 25 + 80 50 + 80 70 + 80 90 + 80 5+1

100.0000 Ω 100 µΩ 100 µA 25 mV 25 + 80 90 + 80 140 + 80 200 + 80 2.5 + 1

1.000000 kΩ 1 mΩ 10 µA 25 mV 25 + 80 180 + 80 400 + 80 600 + 80 5+1

10.00000 kΩ 10 mΩ 5 µA 25 mV 25 + 80 320 + 80 800 + 80 1300 + 80 8+1

31
20 % overrange on all ranges, except 2.4 kΩ for the 10 K range.
32
Maximum clamp voltages are DC, typical accuracy is ±20 %. Add 20 % for offset compensation on.
33
Relative to calibration accuracy.
34
Add per degree from TCAL ±5 oC.
35
Test current with offset compensation off.

Specifications and characteristics are subject to change without notice.


Page 6 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY) 36


 Applies to ± ppm of range
 Peak noise uncertainty is included in DC specifications for ≥ 1 PLC
 Add peak noise uncertainty to measurements when < 1 PLC
Examples:
 10 Ω at 0.2 PLC: 80 (from Accuracy table) + 230 (additional peak noise uncertainty) = 310 ppm of range
 10 Ω at 1 PLC: 80 + 0 = 80 ppm of range
NPLC Digits 1Ω 10 Ω 100 Ω 1 kΩ 10 kΩ
5 7½ 10 11 6 5 0.9
1 7½ 9 9 7 7 0.8
37
0.2 6½ 30 (130) 30 (120) 30 (120) 30 (120) 3 (16)
0.2 6½ 60 (220) 60 (230) 50 (190) 50 (190) 9 (35)
0.06 5½ 70 (350) 70 (350) 50 (290) 50 (280) 20 (90)
0.006 4½ 130 (750) 120 (830) 110 (700) 100 (690) 20 (110)
0.0005 3½ 520 (3550) 530 (3520) 530 (3380) 500 (3370) 100 (670)

DRY CIRCUIT RESISTANCE CHARACTERISTICS


Maximum 4-wire ohm lead resistance 0.5 Ω per lead for 1 Ω range
10 % of range per lead for 10 Ω to 100 Ω ranges
50 Ω per lead for 1 kΩ to 10 kΩ ranges
Input current limit For signals > [±20 mV], current limited, ±13 mA, typical
Offset compensation Selectable on 1 Ω to 10 kΩ ranges
Autozero off error For ±1 °C and ≤ 10 minutes, add ± 8 ppm of reading

36
Noise values are based on 1000 readings with autozero on and using low thermal 4-wire short. RMS noise is typical. Additional peak noise is
guaranteed.
37
With line sync on.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 7 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

DC CURRENT

ENHANCED ACCURACY (WITHIN 30 DAYS OF AUTOCALIBRATION, TOPER ±5 °C FROM TACAL)


Maximum Accuracy ±[ppm of reading + ppm of range]
Range 38 Resolution burden
voltage 24 hour 90 day 1 year 2 year Temperature
TCAL ±1 °C 39 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 40
10.000000 µA 1 pA 15 mV 30 + 30 75 + 30 75 + 30 75 + 30 0.15 + 0.1
100.00000 µA 10 pA 15 mV 20 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
1.0000000 mA 100 pA 15 mV 30 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
10.000000 mA 1 nA 20 mV 40 + 5 60 + 9 60 + 9 60 + 9 0.15 + 0.1
100.00000 mA 10 nA 200 mV 50 + 18 150 + 30 150 + 30 150 + 30 0.15 + 0.1
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 400 + 50 400 + 50 0.15 + 0.1
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 400 + 40 400 + 40 0.15 + 0.1
41
10.000000 A 1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10

ACCURACY
Maximum Accuracy ±[ppm of reading + ppm of range]
Range38 Resolution burden
voltage 24 hour 90 day 1 year 2 year Temperature
TCAL ±1 °C39 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient40
10.000000 µA 1 pA 15 mV 30 + 30 100 + 30 125 + 40 175 + 50 10 + 8
100.00000 µA 10 pA 15 mV 20 + 5 75 + 12 100 + 15 150 + 20 10 + 3
1.0000000 100 pA 15 mV 30 + 5 75 + 12 100 + 15 150 + 20 10 + 3
mA
10.000000 1 nA 20 mV 40 + 5 75 + 12 100 + 15 150 + 20 10 + 3
mA
100.00000 10 nA 200 mV 50 + 18 300 + 30 400 + 30 500 + 30 50 + 5
mA
1.0000000 A 100 nA 400 mV 150 + 50 400 + 50 450 + 50 500 + 50 10 + 10
3.000000 A 1 µA 1300 mV 200 + 40 400 + 40 450 + 40 500 + 40 10 + 10
41
10.000000 A 1 µA 650 mV 700 + 275 800 + 275 1500 + 275 2000 + 275 50 + 10

38
20 % overrange supported for all ranges except for 3 A and 10 A, which are 1 % supported.
39
Relative to calibration accuracy.
40
Add per degree from TCAL ±5 oC.
41
Rear input terminals only.

Specifications and characteristics are subject to change without notice.


Page 8 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

RMS NOISE (ADDITIONAL PEAK NOISE UNCERTAINTY) 42


 Applies to ± ppm of range
 Peak noise uncertainty is included in DC Specifications for PLC ≥ 1
 Add peak noise uncertainty to measurements for PLC < 1
Examples:
 1 mA at 0.006 PLC: 9 (from Accuracy table) + 20 (additional peak noise uncertainty) = 29 ppm of range
 1 mA at 1 PLC: 9 + 0 = 9 ppm of range
NPLC Digits 10 µA 100 µA 1 mA 10 mA 100 mA 1A 3A 10 A 43
5 7½ 0.15 0.14 0.09 0.1 0.3 0.3 0.2 0.8
1 7½ 0.4 0.13 0.1 0.1 0.5 0.5 0.3 1.2
44
0.2 6½ 0 (220) 0 (23) 0.2 (3.4) 0.2 (1.6) 2 (10) 2 (11) 0.7 (4.6) 4 (32)
0.2 6½ 120 (260) 12 (26) 1.2 (3.8) 0.3 (1.8) 1.9 (9.8) 2 (10) 0.8 (5) 8 (37)
0.06 5½ 130 (280) 12 (29) 1.3 (5.6) 0.4 (3.9) 2 (14) 2 (14) 1.2 (7.7) 10 (59)
0.006 4½ 130 (350) 14 (42) 3 (20) 2 (20) 4 (30) 4 (31) 7 (51) 20 (110)
0.0005 3½ 260 (2110) 30 (300) 20 (150) 20 (160) 30 (190) 30 (190) 70 (510) 60 (420)

DC CURRENT CHARACTERISTICS
Range 10 µA 100 µA 1 mA 10 mA 100 mA 1A 3A 10 A43
Effective internal shunt value 45 1 kΩ 100 Ω 10 Ω 1Ω 0.1 Ω 0.1 Ω 0.1 Ω 0.005 Ω
Autozero off error: For ±1 °C and ≤ 10 150 pA 1.5 nA 15 nA 150 nA 15 µA 150 µA 150 µA 3 mA
minutes add ±(8 ppm of reading + range
error)
Overload recovery:
For each additional sustained amp beyond
±1.5 A, add the following initial ppm of range 15500 1800 150 150 6500 200  
error until thermally settled after overload
recovery

42
Noise values are based on 1000 readings with autozero on and AMPS terminal open. RMS noise is typical. Additional peak noise is
guaranteed.
43
Rear input terminals only.
44
With line sync on.
45
Values are typical and guaranteed by design.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 9 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

TEMPERATURE

4-WIRE RTD OR 3-WIRE RTD


Types: 100 Ω platinum PT100, D100, F100, PT385, PT3916; or user-configurable 0 Ω to 10 kΩ
Type Range Resolution Accuracy ± °C

2 year Temperature coefficient 46


TCAL ± 5 °C
4-Wire RTD -200 °C to 850 °C 0.01 °C 0.06 °C 0.003 °C /°C
3-Wire RTD 47 -200 °C to 850 °C 0.01 °C 0.75 °C 0.003 °C /°C

THERMISTOR
Types: 2.252 kΩ, 5 kΩ, and 10 kΩ
Type Range Resolution Accuracy ± °C

2 year Temperature coefficient46


TCAL ± 5 °C
Thermistor -80 °C to +150 °C 0.01 °C 0.08 °C 0.002 °C /°C

THERMOCOUPLE
Types: B, E, J, K, N, R, S, T
Type Range Resolution Accuracy ± °C

2 year Temperature coefficient46


48
TCAL ± 5 °C
Simulated reference junction
B 350 °C to +1820 °C 0.1 °C 0.6 °C 0.03 °C/°C

E −200 °C to +1000 °C 0.001 °C 0.2 °C 0.03 °C/°C

J −200 °C to +760 °C 0.001 °C 0.2 °C 0.03 °C/°C

K −200 °C to +1372 °C 0.001 °C 0.2 °C 0.03 °C/°C

N −200 °C to +1300 °C 0.001 °C 0.2 °C 0.03 °C/°C

R 0 °C to +1768 °C 0.1 °C 0.6 °C 0.03 °C/°C

S 0 °C to +1768 °C 0.1 °C 0.6 °C 0.03 °C/°C

T −100 °C to +400 °C 0.001 °C 0.2 °C 0.03 °C/°C

46
Add per degree from TCAL ±5 oC; specifications without autocalibration.
47
For 3-wire RTD, accuracy is for < 0.1 Ω lead resistance mismatch for input HI and LO. Add 0.25 °C/ 0.1 Ω of HI-LO lead resistance mismatch.
48
Exclusive of cold-junction errors.

Specifications and characteristics are subject to change without notice.


Page 10 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

CONTINUITY
Range 49 Resolution Test Open circuit Accuracy ±[ppm of reading + ppm of range]
current voltage
2 year Temperature coefficient 50
TCAL ±5 °C

1.0000 kΩ 100 mΩ 1 mA 14.0 V 100 + 100 2.5 + 1

CONTINUITY CHARACTERISTICS
Continuity high limit User-selectable; default 10 Ω

CAPACITANCE
Accuracies specified for additional cable and stray capacitance properly zeroed with the Rel function.

ACCURACY
Range 51 Resolution Charge Maximum Accuracy ±[% of reading + % of range]
current 52, 53 circuit
voltage 2 year Temperature coefficient50
TCAL ±5 °C
1.0000 nF 0.1 pF 1.1 µA 2.8 V 1 + 0.2 0.15 + 0.05
10.000 nF 1 pF 1.1 µA 2.8 V 1 + 0.1 0.15 + 0.01
100.00 nF 10 pF 10 µA 3V 0.4 + 0.1 0.01 + 0.01
1.0000 µF 0.1 nF 100 µA 3V 0.4 + 0.1 0.01 + 0.01
10.000 µF 1 nF 100 µA 3V 0.4 + 0.1 0.01 + 0.01
100.00 µF 10 nF 1 mA 3V 0.4 + 0.1 0.01 + 0.01
1000.0 µF 0.1 µF 10 mA 3V 0.5 + 0.1 0.01 + 0.01

DIODE
Voltage Resolution Bias level Accuracy ±[ppm of reading + ppm of range]
measure (selectable)
90 day 1 year 2 year Temperature
range51 TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient50
10.000000 V 1 µV 10 µA / 100 µA / 1 mA 20 + 5 30 + 5 45 + 5 2.5 + 1

49
Specifications exclude lead resistance.
50
Add per degree from TCAL ±5 oC; specifications without autocalibration.
51
20 % overrange on all ranges.
52
Charging current values are typical, guaranteed by design.
53
Discharge current limited to < 13 mA.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 11 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

DIGITIZE VOLTAGE

ACCURACY (INPUT IMPEDANCE AUTO)


Accuracy ±[ppm of reading + ppm of range]
Range 54, 55 Resolution 56 Input impedance 57 90 day 1 year 2 year Temperature
TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 58
100.000 mV 1 µV > 10 GΩ or 10 MΩ ±1 % 210 + 100 220 + 100 230 + 100 15 + 20

1.00000 V 10 µV > 10 GΩ or 10 MΩ ±1 % 110 + 75 120 + 75 130 + 75 15 + 20

10.0000 V 0.1 mV > 10 GΩ or 10 MΩ ±1 % 110 + 75 120 + 75 130 + 75 10 + 20

100.000 V 59 1 mV 10 MΩ ±1 % 110 + 75 120 + 75 130 + 75 15 + 20

1000.00 V 60 10 mV 10 MΩ ±1 % 110 + 75 120 + 75 130 + 75 10 + 20

SIGNAL CHARACTERISTICS 61,62,63

TYPICAL AC AND DC COUPLED

Analog Maximum THD DC-coupled AC-coupled AC-coupled AC coupling


bandwidth flatness error 20 kHz signal settling time filter fast filter slow low
Range
(-3 dB) 3 Hz to (-1 dB FS) 65 (0.5 %) settling time settling time frequency
20 kHz 64 (0.5 %) (0.5 %) (-3 dB) point 66
100.000 mV 600 kHz 0.015 dB 0.04 % 5 µs 80 ms 2.3 s 1 Hz
1.00000 V 600 kHz 0.01 dB 0.03 % 6 µs 80 ms 2.5 s 1 Hz
10.0000 V 600 kHz 0.01 dB 0.01 % 4 µs 80 ms 2.5 s 1 Hz

TYPICAL DC COUPLED

Analog Maximum flatness error Total harmonic distortion (THD) Settling time
Range bandwidth 3 Hz to 1 kHz64 1 kHz signal (0.5 %)
(-3 dB) (-1 dB FS)65
100.000 V 20 kHz 67 0.1 dB 1.3 % 160 µs
1000.00 V 20 kHz 0.1 dB 1.8 % 80 µs

54
For DC coupling, 20 % overrange for 100 mV to 100 V. For AC coupling, 500 % overrange 100 mV to 100 V. 1 % for 1000 V range DC and
AC coupling.
55
Accuracy with sample rate 1 k per second, aperture auto, and 100 reading buffer average.
56
Power up default is 4½ digits.
57
User-selectable.
58
Add per degree from TCAL ± 5 %.
59
For 100 V range, input impedance auto and without ACAL, add 100 ppm of range additional uncertainty and 15 ppm/°C additional uncertainty
for "of range" temperature coefficient for operation outside of TCAL ±5 °C.
60
For signal levels greater than 500 V, add 0.02 ppm/V to the ppm of the readings specification for measurements exceeding 500 V.
61
Accuracy with sample rate 1 M per second and aperture 1 µs.
62
Verified with sine wave input and DC content ≤ 3 % of range.
63
For AC coupling, maximum crest factor of 5.
64
For DC coupled, 0 dB reference frequency is 3 Hz. For AC coupled, 0 dB reference frequency is 1 kHz. For AC coupled operation below 1 kHz,
add 0.1 dB.
65
Exclusive of source input noise.
66
With AC coupling frequency = 3 Hz and AC coupling filter = Slow.
67
For input impedance auto, bandwidth is 6 kHz.

Specifications and characteristics are subject to change without notice.


Page 12 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

TYPICAL AC COUPLED

Analog Maximum flatness Filter Fast Filter Slow Low frequency


Range bandwidth error settling time settling time coupling point66
(-3 dB) 3 Hz to 20 kHz64 (0.5 %) (0.5 %) (-3 dB)
100.000 V 600 kHz 0.1 dB 80 ms 2.3 s 1 Hz
1000.00 V 600 kHz 0.1 dB 80 ms 2.3 s 1 Hz

DC-COUPLED ADDITIONAL NOISE UNCERTAINTY, TYPICAL 68

RMS noise, digitize voltage


1000

100
ppm of range

100mV
10 1V
10V
100V
1 1000V

0.1
1k 2k 5k 10k 20k 50k 100k 200k 500k 1M
Samples per second

68
Specified with aperture Auto and 4-wire short on input terminals. For 100 V range, input impedance 10 MΩ, multiply by 2.5. For all ranges and
sample rate > 1 k, add an additional 3× RMS noise uncertainty to ppm of range.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 13 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

DC-COUPLED TOTAL HARMONIC DISTORTION (THD), TYPICAL 69

Total harmonic distortion, digitize voltage


-50

-60

-70
dB of range

100mV
-80
1V
-90 10V
100V
-100
1000V

-110

-120
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second

DC-COUPLED EFFECTIVE NUMBER OF BITS (ENOB), TYPICAL 70

Effective number of bits (ENOB), digitize voltage


18

16

14
Bits

100mV

12 1V
10V

10

8
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second

69
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤ 5 k, and 1 kHz sine wave for sample rate ≥ 10 k. Distortion is calculated
using first five harmonics.
70
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤5 k, and 1 kHz sine wave for sample rate ≥10 k. For the 100 V and 1000 V
ranges, use the 1 V and 10 V range ENOB, respectively; guaranteed by design.

Specifications and characteristics are subject to change without notice.


Page 14 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

DIGITIZE CURRENT

DC ACCURACY 71
Accuracy ± [ppm of reading + ppm of range]
72 73
Burden
Range Resolution 90 day 1 year 2 year
voltage Temperature
TCAL ±5 °C TCAL ±5 °C TCAL ±5 °C coefficient 74
10.0000 0.1 nA 15 mV 150 + 75 160 + 75 170 + 75 30 + 15
µA
100.000 1 nA 15 mV 150 + 75 160 + 75 170 + 75 30 + 15
µA
1.00000 10 nA 15 mV 150 + 75 160 + 75 170 + 75 30 + 15
mA
10.0000 100 nA 20 mV 150 + 75 160 + 75 170 + 75 30 + 15
mA
100.000 1 µA 200 mV 340 + 100 450 + 100 560 + 100 50 + 20
mA
1.00000 A 10 µA 400 mV 400 + 110 500 + 110 600 + 110 50 + 25
3.00000 A 100 µA 1300 mV 650 + 150 900 + 150 900 + 150 50 + 25
10.0000 100 µA 650 mV 950 + 350 1500 + 350 2000 + 350 50 + 25
A 75

SIGNAL CHARACTERISTICS, TYPICAL 76


Total harmonic
Analog DC-coupled
Maximum flatness error distortion (THD)
Range72 bandwidth settling time
3 Hz to 20 kHz 20 kHz signal
(-3 dB) (0.5 %)
(-1 dB FS)
10.0000 µA 0.15 dB 100 kHz 0.02 % 8 µs
100.000 µA 0.15 dB 100 kHz 0.01 % 7 µs
1.00000 mA 0.1 dB 100 kHz 0.01 % 3 µs
10.0000 mA 0.1 dB 100 kHz 0.01 % 8 µs
100.000 mA 0.1 dB 100 kHz 0.02 % 5 µs
77
1.00000 A 0.1 dB 100 kHz 0.02 % 6 µs
77
3.0000 A 0.1 dB 100 kHz 0.02 % 6 µs
75,77,78
10.0000 A 0.1 dB 100 kHz 0.02 % 6 µs

71
Accuracy with sample rate 1 k per second, aperture auto, and 100 reading buffer average.
72
20 % overrange on all ranges except 3.3 % for 3 A and 10 A ranges.
73
Power up default is 4½ digits.
74
Add per degree from TCAL ±5 oC.
75
Rear input terminals only.
76
Verified with sine wave input and DC content ≤ 3 % of range. For flatness error, 0 dB reference frequency is 3 Hz.
77
For the 1 A, 3 A, and 10 A ranges, use the 100 mA range accuracy; guaranteed by design.
78
10 A flatness verified to 10 kHz; 100 kHz guaranteed by design.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 15 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

ADDITIONAL NOISE UNCERTAINTY, TYPICAL 79

RMS noise, digitize current


1000

100 10uA
ppm of range

100uA
1mA
10mA
10 100mA, 1A
3A
10A

1
1k 2k 5k 10k 20k 50k 100k 200k 500k 1M
Samples per second

TOTAL HARMONIC DISTORTION (THD), TYPICAL 80

Total harmonic distortion, digitize current


-70

-80
dB of range

-90 10uA
100uA

-100 1mA
10mA

-110 100mA

-120
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second

79
Specified with aperture Auto and open input terminals. For all ranges and for ≥1 k sample rate, add an additional 3× RMS noise uncertainty to
ppm of range.
80
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤ 5 k, and 1 kHz sine wave for sample rate ≥ 10 k. Distortion is calculated
using first five harmonics. For the 1 A, 3 A, and 10 A ranges, use the 100 mA range accuracy; guaranteed by design.

Specifications and characteristics are subject to change without notice.


Page 16 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

EFFECTIVE NUMBER OF BITS (ENOB), TYPICAL 81

Effective number of bits (ENOB), digitize current


18

16

14 10uA
Bits

100uA

12 1mA
10mA
100mA
10

8
1k 2k 5k 10k 20k 50k 100k 200k
Samples per second

DIGITIZER CHARACTERISTICS
Maximum resolution 18 bits
Measurement input coupling DC or AC (voltage only)
82
Sampling rate Programmable 1 k through 1 million
Volatile sample memory with timestamp 27.5 million
Minimum record time 1 µs
Timestamp resolution 1 ns with standard or full buffer style
1 µs with compact buffer style
Timestamp accuracy With standard or full buffer style, 20 ns between adjacent readings,
with total buffer time < 2 s
With compact buffer style, 2 µs adjacent readings, with total buffer
buffer time < 2 s
Maximum record length 8 million

81
Specified with aperture Auto, 100 Hz sine wave for sample rate ≤5 k, and 1 kHz sine wave for sample rate ≥10 k. For the 1 A, 3 A, and 10 A
ranges, use the 100 mA ENOB; guaranteed by design.
82
Sample rate is not continuously adjustable. For valid discrete settings, see the Model DMM7510 Reference Manual.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 17 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

TRUE RMS AC VOLTAGE AND AC CURRENT


Function Range 83 Resolution 1 year accuracy: ± (% of reading + % of range) TCAL ± 5 °C

5 Hz to 10 Hz to 20 kHz to 50 kHz to 100 kHz to


3 Hz to 5 Hz 10 Hz 20 kHz 50 kHz 100 kHz 300 kHz

100.0000 mV 0.1 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5

1.000000 V 1 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5

Voltage 84
10.00000 V 10 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5

100.0000 V 100 µV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5

700.000 V 1 mV 1.0 + 0.03 0.30 + 0.03 0.06 + 0.03 0.14 + 0.05 0.6 + 0.08 4.0 + 0.5
Temperature – – 0.01 + 0.003 0.03 + 0.003 0.005 + 0.003 0.006 + 0.005 0.01 + 0.006 0.03 + 0.01
coefficient /
°C (all
ranges)

Function Resolution 1 year accuracy: ± (% of reading + % of range) TCAL ± 5 °C


Range83
3 Hz to 5 Hz 5 Hz to 10 Hz 10 Hz to 2 kHz 2 kHz to 5 kHz 5 kHz to 10 kHz
Current84 1.000000 mA 1 nA 1.0 + 0.04 0.30 + 0.04 0.08 + 0.03 0.09 + 0.03 0.09 + 0.03
10.00000 mA 10 nA 1.0 + 0.04 0.30 + 0.04 0.08 + 0.03 0.09 + 0.03 0.09 + 0.03
100.0000 mA 100 nA 1.0 + 0.04 0.30 + 0.04 0.08 + 0.03 0.09 + 0.03 0.09 + 0.03
1.000000 A 1 µA 1.0 + 0.04 0.30 + 0.04 0.20 + 0.04 0.88 + 0.04 2.0 + 0.04

3.000000 A 1 µA 1.0 + 0.05 0.30 + 0.05 0.20 + 0.05 0. 88 + 0.05 2.0 + 0.05
10.00000 A 85 10 µA 1.0 + 0.05 0.40 + 0.05 0.40 + 0.05 0. 88 + 0.05 2.0 + 0.05
Temperature – – 0.10 + 0.004 0.030 + 0.004 0.005 + 0.003 0.006 + 0.005 0.006 + 0.005
coefficient /
°C (all
ranges)

ADDITIONAL AC UNCERTAINTIES – LOW FREQUENCY UNCERTAINTY


Detector bandwidth (BW)
Additional uncertainty ±(% of reading),
lower frequency uncertainty 3 BW (3 Hz to 300 kHz) 30 BW (30 Hz to 300 kHz) 300 BW (300 Hz to 300 kHz)

20 Hz to 30 Hz 0 0.3 
30 Hz to 50 Hz 0 0 
50 Hz to 100 Hz 0 0 4.0
100 Hz to 200 Hz 0 0 0.72
200 Hz to 300 Hz 0 0 0.18
300 Hz to 500 Hz 0 0 0.07
> 500 Hz 0 0 0

83
20 % overrange on AC functions except 1 % on 700 V, 3.33 % on 3 A and 1 % on 10 A. Default resolution is 6½ digits.
84
Specifications are for detector bandwidth of 3 Hz and sine wave inputs > 5 % of range. Detector bandwidth of 3 Hz and 30 Hz are multisample
A/D conversions. Detector bandwidth of 300 Hz is a single A/D conversion, programmable from 0.0005 PLC to 15 PLC (60 Hz), 12 PLC
(50 Hz). Default condition set to 1 PLC.
85
Rear input terminals only.

Specifications and characteristics are subject to change without notice.


Page 18 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

ADDITIONAL AC VOLTAGE CREST FACTOR UNCERTAINTIES 86


Additional uncertainty ± (% of reading)
Input signal Detector Crest factor
frequency bandwidth Maximum crest factor: 5 at range full scale
1 to 2 2 to 3 3 to 4 4 to 5
3 Hz to 5 Hz 3 Hz 1.00 4.00 4.80 5.00
5 Hz to 10 Hz 3 Hz 0.50 1.20 1.30 1.40
10 Hz to 30 Hz 3 Hz 0.20 0.30 0.60 0.90
5 Hz to 100 Hz 30 Hz 0.20 0.30 0.60 0.90
100 Hz to 300 Hz 30 Hz 0.05 0.15 0.30 0.40
100 Hz to 300 Hz 300 Hz 0.50 1.20 1.30 1.50
500 Hz to 10 kHz 300 Hz 0.05 0.15 0.30 1.20

AC VOLTAGE CHARACTERISTICS
Measurement method AC-coupled, true RMS
Input impedance 1 MΩ ± 2 % ǁ < 150 pF
Volt*Hertz product < 2.1 x 107 V*Hz verified; input frequency verified for < 300 kHz

AC CURRENT CHARACTERISTICS
Measurement method AC-coupled, true RMS

Range 1 mA 10 mA 100 mA 1A 3A 10 A 87
Burden voltage (RMS) < 16 mV < 20 mV < 0.2 V < 0.4 V < 1.3 V < 0.65 V
Overload recovery: For each
additional sustained ampere beyond 0.006 0.006 0.12 0.05  
±1.5 A, add the following initial % of
range error until thermally settled after
overload recovery

86
Applies for non-sine wave inputs, DC content ≤ 3 % of range, maximum crest factor ≤ 5.0.
87
Rear input terminals only.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 19 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

FREQUENCY AND PERIOD

MEASUREMENT ACCURACY 88
Aperture Measurement resolution Accuracy ±[ppm of reading + ppm of aperture time]
Frequency: 3 Hz to 500 kHz
Period: 333 ms to 2 µs
1 year 2 year
TCAL ±5 °C TCAL ±5 °C

250 ms 0.1 ppm 80 + 0.333 160 + 0.333


100 ms 0.1 ppm 80 + 3.33 160 + 3.33
10 ms 0.1 ppm 80 + 33.3 160 + 33.3

THRESHOLD LEVEL ACCURACY 89


Threshold range Threshold resolution Accuracy ±[% of reading]
2 year
TCAL ±5 °C
100 mV to 700 V 0.05 % 1.0 %

FREQUENCY AND PERIOD CHARACTERISTICS


Measurement method Reciprocal counting technique
Aperture 10 ms to 273 ms; default is 10 ms

88
Specified for square wave inputs. Input signal must be >10 % of ACV range. If input is<20 mV on the 100 mV range, then the frequency must
be >10 Hz. For sine wave inputs, frequency must be >100 Hz. For frequencies ≤100 Hz, threshold level ≤50 % of input signal and ≤7 Hz,
threshold level ≤3 % of range.
89
Threshold range is voltage RMS and threshold level voltage peak. Specified with 1 KHz square wave. 100 V and 700 V threshold ranges
guaranteed by design.

Specifications and characteristics are subject to change without notice.


Page 20 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

TYPICAL READING RATES, 60 Hz (50 Hz) OPERATION 90,91,92,93


Functions: Functions: Functions: Functions:
DC voltage (10 V) 4-wire ohms (≤ 1 kΩ) Thermistor Dry circuit (≤ 1 kΩ)
2-wire ohms (≤ 10 kΩ), 4-wire / 3-wire RTD
DC current (1 mA)
NPLC Digits Measure- Measure- Measure- Measure- Measure- Measure- Measure- Measure-
ments into ments into ments ments ments into ments into ments ments
buffer computer into into buffer computer into into
buffer computer buffer computer
1 7½ 59.8 (49.8) 58 (48) 29 (24) 28 (24) 57 (48) 57 (48) 27 (23) 26 (22)
0.2 6½ 295 (240) 250 (210) 128 (109) 119 (100) 230 (200) 230 (200) 100 (89) 96 (85)
0.06 5½ 965 (810) 950 (800) 310 (280) 315 (280) 900 (750) 900 (750) 190 (180) 190 (180)
0.006 4½ 7500 (6700) 7300 (6500) 750 (730) 740 (720) 6800 (6000) 6800 (6000) 295 (290) 295 (290)
0.0005 3½ 26000 (26000) 24000 (24000) 860 (860) 860 (860) 18000 (18000) 18000 310 (310) 310 (310)
(18000)

Functions:
AC voltage, AC current
Detector bandwidth (Hz) Digits Measurements into Measurements into
buffer computer
3 6½ 0.5 (0.5) 0.5 (0.5)
30 6½ 3.3 (3.3) 3.3 (3.3)
94
300 6½ 59.8 (49.8) 55 (46)
94
300 3½ 26200 (26200) 24500 (24500)

DIGITIZE, TYPICAL
Sampling rate Digits Resolution Measurements into computer93
10 kS/s 5½ 18 9700
20 kS/s 4½ 16 19000
50 kS/s 4½ 16 44400
100 kS/s 4½ 15 80000
1 MS/s 3½ 12 108000

90
Reading speeds for autozero off, fixed range, autodelay off. Offset compensation off and open lead detector off where applicable.
91
Buffer measurements: For < 0.2 PLC, multisample, single buffer transfer binary reading only.
92
PC measurements: For 1 and 0.2 PLC single reading and single transfer to computer (USB).
93
Reading rates using factory default operating conditions and autorange off, autodelay off. Speeds include measurement and data transfer out
of the USB. ≥1000 readings with binary transfer over USB.
94
For bandwidth 300 Hz, autozero off, 6½ digits at 1 PLC, 3½ digits at 0.0005 PLC.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 21 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

SYSTEM PERFORMANCE, TYPICAL


• Mode: 3½ digit, autozero off, 0.0005 PLC, excludes measurement time
• Time includes function change from DC voltage or 2-wire ohms to listed function

Function Function change (ms) Range change (ms)

DC voltage or 2-wire ohms (< 10 kΩ) 6 1.3

4-wire ohms (< 10 kΩ) 7 1.3

DC current 7 1.3
95
Frequency or period 7 1.3
AC voltage or AC current 7 1.3
Digitize voltage or current 7 1.3

RANGES FOR FUNCTION CHANGE TIMES


Function change times apply to the ranges listed in the table below.
Function Range
DC voltage 10 V
2-wire or 4-wire ohms 1 kΩ
DC current 1 mA
Dry-circuit ohms 10 Ω
Thermocouple Use DC voltage rates
Thermistor Use 2-wire ohms rates

AC current 1 mA
AC voltage 1V

Buffer transfer speed (binary) Measurements into computer (per second)


USB LAN GPIB
Average for 1000 readings 280000 270000 190000
Average for 1000 readings with timestamp 170000 140000 100000

95
For DC voltage or 2-wire ohms to frequency or period, 10 ms aperture. For AC current or AC voltage, detector bandwidth is 300 Hz.

Specifications and characteristics are subject to change without notice.


Page 22 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

TRIGGERING
Time base accuracy 25 ppm
Trigger source Analog DC voltage, DC current, or any system trigger
Trigger coupling DC or AC (DC voltage function only)
Input trigger latency96, 97, 98 < 225 ns
Input trigger jitter96,97 < 50 ns
Sample period jitter96,97 < 1 ns

DMM REAR-PANEL TRIGGERS


EXT TRIG IN and OUT 0 V to 5 V logic signal input and output, TTL-compatible
EXT trigger latency (IN and OUT) < 400 ns
EXT trigger latency (IN or OUT) < 200 ns (guaranteed by design)

ANALOG TRIGGERING 99

ANALOG LEVEL, EDGE, OR WINDOW TRIGGER TYPES 100

Trigger characteristics Voltage input Current input


Input range 100 mV to 1000 V 10 µA to 10 A
Resolution 0.05 % 0.05 %

Basic accuracy (TACAL ±5 °C) 101, 102 1% 1%

ANALOG TRIGGER LATENCIES

Digital I/O External


Positive logic 800 ns + 40 ns jitter 930 ns + 40 ns jitter
Negative logic 800 ns + 40 ns jitter 840 ns + 40 ns jitter

WINDOW FILTER AND MEMORY (BUFFER)


Window filter size 0 to 10 % of reading, where 0 averages all readings
Memory Up to 27.5 million timestamped readings with the compact buffer
style, with additional memory available using an external USB flash
drive
Maximum Internal memory (buffer) 27.5 million readings with the compact buffer style (6½-digit without
formatting); 11 million readings with the standard or full buffer styles

96
Guaranteed by design; for digital I/O only.
97
Stimulus command required to meet specifications.
98
If using trigger model, add 200 ns uncertainty.
99
For DC or AC coupled, the trigger level can be set up to 100 % of measure range.
100
Rising or falling edge triggering supported. Window trigger requires setting two independent levels.
101
Trigger event occurs after the threshold crossing at a time determined by total trigger latencies.
102
Accuracy specifications require user ACAL and are verified with level trigger amplitude set to 50 % of range with a 100 Hz sine wave at 100 %
full scale of range. High frequency rejection is off. NPLC 0.0005 (DC voltage/DC current) or aperture 1 µs for digitize voltage or digitize
current. Specified for fixed range, autozero off. For digitized DC voltage AC coupled, add 0.5 %. For DC current and digitized DC current 3 A
or 10 A ranges, add an additional 2 %.

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 23 of 25
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

GENERAL INSTRUMENT SPECIFICATIONS

SPECIFICATION CONDITIONS
This document contains specifications and supplemental information for the DMM7510 Precision Sampling
Digital Multimeter instrument. Specifications are the standards against which the DMM7510 is tested. Upon
leaving the factory, the DMM7510 meets these specifications. Supplemental, typical, and characteristic
values are non-warranted, apply at 23 °C, and are provided solely as useful information. All specifications
apply to front or rear terminal inputs, except 10 A specifications (rear-terminals only).
Input protection 1010 V DC (715 VRMS V AC) all ranges and functions on HI and LO
terminals; 350 V all ranges and functions on sense HI, sense LO
terminals; 250 V rated current input terminal; fused 3 A and 10 A
ranges; current input terminals protected to 1 kV
3 A input fuse protection 3.5 A, 1 kV fast blow type; Keithley part number
DMM7510-FUSE-3A
10 A input fuse protection 11 A, 1 kV fast blow type; Keithley part number
DMM7510-FUSE-10A
AC voltage input Maximum DC voltage: 1000 V on any AC voltage range
Common mode isolation 500 V DC or AC Vpeak LO to chassis
All terminals > 10 GΩ, < 350 pF any terminal to chassis
Power line Universal input, 100 V to 240 V
Line frequency 50 Hz or 60 Hz, automatically sensed at power-up
Power consumption 60 VA
Operating environment Specified for 0 °C to 50 °C, ≤ 80 % relative humidity at 35 °C,
altitude up to 2000 meters
Storage environment -30 °C to 70 °C
Real time clock Lithium battery backup (3+ years battery life)
EMC Conforms to European Union EMC Directive
Safety NRTL listed to UL61010-1, and CSA C22.2 No 61010-1; conforms
with European Union Low Voltage Directive
Vibration MIL-PRF-28800F Class 3, Random
Warm-up 90 minutes to rated accuracy
Input signal connections Front and rear safety banana jacks
Cooling Forced air, fixed speed
Dimensions Without handle and bumpers: 88 mm high × 213 mm wide ×
410 mm deep (3.46 in. × 8.39 in. × 16.13 in.)
With handle and bumpers (bench configuration): 106 mm high ×
255 mm wide × 425 mm deep (4.18 in. × 10.05 in. × 16.75 in.)
Shipping weight (with bumpers and handle) 4.08 kg (9.0 lb)
Shipping weight (without bumpers and handle) 3.63 kg (8.0 lb)

Specifications and characteristics are subject to change without notice.


Page 24 of 25 SPEC-DMM7510 Rev. B / October 2016
Model DMM7510 7-1/2 Digit Graphical Sampling Multimeter Specifications

Digital I/O Connector 9-pin female D


5 V power supply pin Limited to 500 mA at > 4 V
(solid-state fuse protected)
Lines Six input/output, user-defined, for
digital I/O or triggering
Input signal levels 0.7 V (maximum logic low)
3.7 V (minimum logic high)
Input voltage limits -0.25 V (absolute minimum)
+5.25 V (absolute maximum)
Maximum source current +2.0 mA at > 2.7 V (per pin)
Maximum sink current -50 mA at 0.7 V (per pin, solid-state
fuse protected)
Handler User-defined start of test, end of test,
four category bits
Math functions Rel, dB, Limit Test, Percentage, 1/x, and mX + b
Remote interface LAN: RJ-45 connector, 10/100BT; Virtual Front Panel
GPIB: IEEE-488.1 compliant. Supports IEEE-488.2 common
commands and status model topology
USB device (rear panel, type B): 2.0 full speed, USBTMC compliant
USB host (front panel, type A): USB 2.0, support for flash drives,
FAT 32
LXI compliance LXI version 1.4 Core 2011
Language Embedded Test Script Processor (TSP) accessible from any host
interface; responds to high-speed test scripts comprised of remote
commands and statements (for example, branching, looping, math);
able to execute high-speed test scripts stored in memory without
host intervention; also SCPI (default command set)
Accessories supplied Product Information CD-ROM, Model DMM7510 Quick Start Guide,
Kickstart Software Quick Start Guide, power cord, 1 m USB cable
(type A to type B), 3 m LAN cable, and Model 1756 Standard Test
Lead Kit
Accessories available (Calibration / Data / ISO 17025), software IVI/VISA drivers for
Microsoft® Visual Basic®, Visual C/C++®, National Instruments
(NI™) LabVIEW™, Keithley Test Script Builder, Keithley KickStart,
and NI LabWindows™/CVI
Display Five-inch capacitive touch, color thin-film-transistor (TFT) WVGA
(800 x 480) with LED backlight
Password protection 30 characters
Expansion interface The TSP-Link® expansion interface allows TSP-enabled
instruments to trigger and communicate with each other
IP configuration Static or DHCP (manual or automatic)

Specifications are subject to change without notice.


SPEC-DMM7510 Rev. B / October 2016 Page 25 of 25