The document provides technical information about Proline Promass 80H and 83H Coriolis mass flow measuring systems. The systems use a single-tube design and measure mass flow, density and temperature of liquids and gases. They can be used for flow measurement in various industrial processes and connect to common control systems.

Uploaded by

Ivan Garcia
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
42 views40 pages

Coriolis Flow System Guide

The document provides technical information about Proline Promass 80H and 83H Coriolis mass flow measuring systems. The systems use a single-tube design and measure mass flow, density and temperature of liquids and gases. They can be used for flow measurement in various industrial processes and connect to common control systems.

Uploaded by

Ivan Garcia
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 40

Technical Information

Proline Promass 80H, 83H


Coriolis Mass Flow Measuring System
The single-tube system with a "fit-and-forget" design:
does not harm the material being measured – chemical-resistant
materials

Application Your benefits


The Coriolis measuring principle operates independently The Promass measuring devices make it possible to
of physical fluid properties, such as viscosity and density. simultaneously record several process variables
(mass/density/temperature) for various process
• Extremely accurate measurement of liquids such as
conditions during measuring operation.
oils, lubricants, liquefied gases, paints, cleaning agents
and solvents The uniform Proline transmitter concept includes:
• Fluid temperatures up to +200 °C (+392 °F) • Modular device and operating concept resulting in a
• Process pressures up to 40 bar (580 psi) higher degree of efficiency
• Mass flow measurement up to 70 t/h (2570 lb/min) • Software options for batching and concentration
measurement for extended range of application
Approvals for hazardous area:
• Diagnostic ability and data back-up for increased
• ATEX, FM, CSA, TIIS, IECEx, NEPSI
process quality
Connection to all common process control systems:
The Promass sensors, tried and tested in over
• HART, PROFIBUS PA/DP, FOUNDATION Fieldbus,
100000 applications, offer:
MODBUS
• Multivariable flow measurement in compact design
Relevant safety aspects: • Insensitivity to vibrations thanks to balanced single-
• Secondary containment (up to 25 bar (362 psi)), tube measuring system
Pressure Equipment Directive, SIL-2 • Efficient protection against forces from piping thanks
to robust construction
• Easy installation without taking inlet and outlet runs
into account

TI074D/06/ae/09.08
71079945
Proline Promass 80H, 83H

Table of contents

Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 22


Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Design/dimensions in metric units . . . . . . . . . . . . . . . . . . . . . . . 22
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Design/dimensions in US units . . . . . . . . . . . . . . . . . . . . . . . . . 28
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Material load curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Unified control concept for both types of transmitter . . . . . . . . . . 36
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Language groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 37
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 C-Tick symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
FOUNDATION Fieldbus certification . . . . . . . . . . . . . . . . . . . . 37
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
PROFIBUS DP/PA certification . . . . . . . . . . . . . . . . . . . . . . . . . 37
Electrical connection Measuring unit . . . . . . . . . . . . . . . . . . . . . 10
MODBUS certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . 11
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Electrical connection Remote version . . . . . . . . . . . . . . . . . . . . . 12
Pressure Equipment Directive . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Functional safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Cable specification Remote version . . . . . . . . . . . . . . . . . . . . . . . 13
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 38
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

Performance characteristics. . . . . . . . . . . . . . . . . . . . 13 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39


Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . 39
Influence of fluid temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Influence of fluid pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Operating conditions: Installation . . . . . . . . . . . . . . . 16


Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Length of connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
System pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Operating conditions: Environment. . . . . . . . . . . . . . 19


Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Shock resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 19

Operating conditions: Process . . . . . . . . . . . . . . . . . . 20


Fluid temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Fluid pressure range (nominal pressure) . . . . . . . . . . . . . . . . . . . 20
Limiting flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Pressure loss (SI units) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Pressure loss (US units) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

2 Endress+Hauser
Proline Promass 80H, 83H

Function and system design


Measuring principle The measuring principle is based on the controlled generation of Coriolis forces. These forces are always present
when both translational and rotational movements are superimposed.
FC = 2 · Δm (v · ω)
FC = Coriolis force
Δm = moving mass
ω = rotational velocity
v = radial velocity in rotating or oscillating system

The amplitude of the Coriolis force depends on the moving mass Δm, its velocity v in the system, and thus on
the mass flow. Instead of a constant angular velocity ω, the Promass sensor uses oscillation.
This causes the tube through which the fluid is flowing to oscillate. The Coriolis forces produced at the
measuring tubes cause a phase shift in the tube oscillations (see illustration):
• If there is zero flow, i.e. when the fluid stands still, the oscillation measured at points A and B has the same
phase, and thus there is no phase difference (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the outlet (3).

A A A
B B B

1 2 3
a0003383

The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube
oscillations at the inlet and outlet.
For the Promass H, the system balance is created by a counterweight that runs parallel to the measuring tube.
This counterweight oscillates in antiphase to the measuring tubes and thus creates a balanced system. The
patented ITB™ (Intrinsic Tube Balance) system ensures balance and stability, thus providing accurate
measurements over a wide range of process and environmental conditions.
Therefore, the Promass H is just as easy to install as the familiar two-tube systems! Consequently, no special
measures for attachment are required in front of or behind the sensor.
The measuring principle operates independently of temperature, pressure, viscosity, conductivity and flow
profile.
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and thus the
density of the oscillating system (comprising the measuring tube and fluid) results in a corresponding, automatic
adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The
microprocessor utilizes this relationship to obtain a density signal.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor due to
temperature effects. This signal corresponds to the process temperature and is also available as an output.

Endress+Hauser 3
Proline Promass 80H, 83H

Measuring system The measuring system consists of a transmitter and a sensor. Two versions are available:
• Compact version: transmitter and sensor form a mechanical unit.
• Remote version: transmitter and sensor are mounted physically separate from one another.

Transmitter

Promass 80 • Two-line liquid-crystal display


• Operation with push buttons

Esc
- + E

a0003671

Promass 83 • Four-line liquid-crystal display


• Operation with "Touch control"
• Application-specific Quick Setup
• Mass flow, volume flow, density and temperature measurement as well as
calculated variables (e.g. fluid concentrations)
Esc
- + E

a0003672

Sensor

H • Single bent tube. Low pressure loss and chemically Documentation No.
resistant material TI 074D/06/en
• Nominal diameters DN 8 to 50 (3/8" to 2")
• Tube material: zirconium

a0003677

Other sensors can be found in the separate documentation

A • Single-tube system for highly accurate measurement of Documentation


very small flows No. TI 054D/06/en
• Nominal diameters DN 1 to 4 (1/24" to 1/8")
• Tube material: stainless steel or
Alloy C-22

a0003679

E • General purpose sensor, ideal replacement for volumetric Documentation No.


flowmeters. TI 061D/06/en
• Nominal diameters DN 8 to 50 (3/8" to 2")
• Tube material: stainless steel

a0002271

4 Endress+Hauser
Proline Promass 80H, 83H

F • Universal sensor for fluid temperatures Documentation No.


up to 200 °C (392 °F) TI 053D/06/en
• Nominal diameters DN 8 to 250 (1/12" to 10")
• Tube material: stainless steel or Alloy C-22

a0003673

F (High-temperature) • Universal high-temperature sensor for fluid temperatures


up to 350 °C. (662 °F)
• Nominal diameters DN 25, 50, 80 (1", 2", 3")
• Tube material: Alloy C-22

a0003675

M • Robust sensor for extreme process pressures, high


secondary containment requirements and fluid
temperatures up to 150 °C (302 °F)
• Nominal diameters DN 8 to 80 (1/12" to 3")
• Tube material: titanium

a0003676

I • Straight single-tube instrument. Minimal shear stress on Documentation No.


fluid, hygienic design, low pressure loss TI 075D/06/en
• Nominal diameters DN 8 to 80 (3/8" to 3")
• Tube material: titanium

a0003678

S • Single bent tube. Documentation No.


Hygienic design, low pressure loss, for fluid temperatures TI 076D/06/en
up to 150 °C (+302 °F)
• Nominal diameters DN 8 to 50 (3/8" to 2")
• Tube material: stainless steel

a0006828

Endress+Hauser 5
Proline Promass 80H, 83H

Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Fluid density (proportional to resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)

Measuring range Measuring ranges for liquids

DN Range for full scale values (liquids) gmin(F) to gmax(F)

8 3/8" 0 to 2000 kg/h 0 to 73.5 lb/min

15 1/2" 0 to 6500 kg/h 0 to 238 lb/min

25 1" 0 to 18 000 kg/h 0 to 660 lb/min

40 1 ½" 0 to 45 000 kg/h 0 to 1 650 lb/min

50 2" 0 to 70 000 kg/h 0 to 2 570 lb/min

Operable flow range Greater than 1000: 1. Flow rates above the preset full scale value do not overload the amplifier, i.e. the totalizer
values are registered correctly.

Input signal Status input (auxiliary input):


U = 3 to 30 V DC, Ri = 5 kΩ, galvanically isolated.
Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start, batching
start/stop (optional).

Status input (auxiliary input) with PROFIBUS DP


U = 3 to 30 V DC, Ri = 3 kΩ, galvanically isolated.
Switch level: ±3…±30 V DC, independent of polarity.
Configurable for: positive zero return, error message reset, zero point adjustment start, batching start/stop
(optional), totalizer reset for batching (optional).

Status input (auxiliary input) with MODBUS RS485


U = 3 to 30 V DC, Ri = 3 kΩ, galvanically isolated.
Switch level: ±3…±30 V DC, independent of polarity.
Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start.

Current input (only Promass 83)


Active/passive selectable, galvanically isolated, resolution: 2 A
• Active: 4 to 20 mA, RL < 700 Ω, Uout = 24 V DC, short-circuit proof
• Passive: 0/4 to 20 mA, Ri = 150 Ω, Umax = 30 V DC

6 Endress+Hauser
Proline Promass 80H, 83H

Output
Output signal Promass 80

Current output:
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
selectable, temperature coefficient: typically 0.005% o.f.s./°C, resolution: 0.5 μA
• Active: 0/4 to 20 mA, RL < 700 Ω (for HART: RL ≥ 250 Ω)
• Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri ≥ 150 Ω

Pulse/frequency output:
Passive, open collector, 30 V DC, 250 mA, galvanically isolated.
• Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1, pulse width max. 2 s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.5 to 2000 ms)

PROFIBUS PA interface:
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Profile Version 3.0
• Current consumption: 11 mA
• Permitted supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic) = 0 mA
• Data transmission rate: 31.25 kBit/s
• Signal encoding: Manchester II
• Function blocks: 4 x Analog Input, 1 x Totalizer
• Output data: Mass flow, Volume flow, Density, Temperature, Totalizer
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)

Promass 83

Current output:
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
selectable, temperature coefficient: typically 0.005% o.f.s./°C, resolution: 0.5 μA
• Active: 0/4 to 20 mA, RL < 700 Ω (for HART: RL ≥ 250 Ω)
• Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri ≥ 150 Ω

Pulse/frequency output:
Active/passive selectable, galvanically isolated
• Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Ω
• Passive: open collector, 30 V DC, 250 mA
• Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1, pulse width max.
2s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms)

PROFIBUS DP interface:
• PROFIBUS DP in accordance with EN 50170 Volume 2
• Profile Version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding: NRZ Code
• Function blocks: 6 x Analog Input, 3 x Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizers 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
• Available output combination → Page 11

Endress+Hauser 7
Proline Promass 80H, 83H

PROFIBUS PA interface:
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Data transmission rate: 31.25 kBit/s
• Current consumption: 11 mA
• Permitted supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Signal encoding: Manchester II
• Function blocks: 6 x Analog Input, 3 x Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizers 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
• Available output combination → Page 11

MODBUS interface:
• MODBUS device type: slave
• Address range: 1 to 247
• Supported function codes: 03, 04, 06, 08, 16, 23
• Broadcast: supported with the function codes 06, 16, 23
• Physical interface: RS485 in accordance with EIA/TIA-485 standard
• Supported baud rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
• Transmission mode: RTU or ASCII
• Response times:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
• Possible output combinations → Page 11

FOUNDATION Fieldbus interface:


• FOUNDATION Fieldbus H1, IEC 61158-2, galvanically isolated
• Data transmission rate: 31.25 kBit/s
• Current consumption: 12 mA
• Permitted supply voltage: 9 to 32 V
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Bus connection with integrated reverse polarity protection
• Signal encoding: Manchester II
• ITK Version 4.01
• Function blocks: 7 x Analog Input, 1 x Digital Output, 1 x PID
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizers 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Reset totalizer
• Link Master (LM) function is supported

Signal on alarm Current output:


Failsafe mode selectable (e.g. in accordance with NAMUR Recommendation NE 43)
Pulse/frequency output:
Failsafe mode selectable

Status output (Promass 80):


"Nonconductive" in the event of a fault or if the power supply fails

Relay output (Promass 83):


"Dead" in the event of a fault or if the power supply fails

8 Endress+Hauser
Proline Promass 80H, 83H

Load see "Output signal"

Low flow cut off Switch points for low flow cut off are selectable.

Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.

Switching output Status output (Promass 80):


Open collector, max. 30 V DC / 250 mA, galvanically isolated.
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction,
limit values.

Relay output (Promass 83):


Normally closed (NC or break) or normally open (NO or make) contacts available (factory setting:
relay 1 = NO, relay 2 = NC), max. 30 V / 0.5 A AC; 60 V / 0.1 A DC, galvanically isolated.

Endress+Hauser 9
Proline Promass 80H, 83H

Power supply
Electrical connection
Measuring unit A B C

d d
g g
b b

a
a
a
b d/(g) (d)

HART* PROFIBUS PA*


FOUNDATION Fieldbus*

– 27 PA(–)/FF(–) 27
+ 26 PA(+)/FF(+) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20

N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b

PROFIBUS DP* PROFIBUS DP**


MODBUS RS485**

A (RxD/TxD-N) 27 A (RxD/TxD-N) 27
B (RxD/TxD-P) 26 B (RxD/TxD-P) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 g + 20 g

N (L-) 2 c N (L-) 2
c
L1 (L+) 1 L1 (L+) 1
b b

a0002441

Connecting the transmitter, cable cross-section: max. 2.5 mm2


A View A (field housing)
B View B (stainless steel field housing)
C View C (wall-mount housing)
*) fixed communication board
**) flexible communication board
a Connection compartment cover
b Cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L– for DC
c Ground terminal for protective ground
d Signal cable: see Terminal assignment → Page 11
Fieldbus cable:
Terminal No. 26: DP / PA (+) / FF (+) / MODBUS RS485 /(PA, FF: with reverse polarity protection)
Terminal No. 27: DP / PA (–) / FF (–) / MODBUS RS485 / (PA, FF: with reverse polarity protection)
e Ground terminal for signal cable shield / fieldbus cable / RS485 line
f Service adapter for connecting service interface FXA193 (FieldCare)
g Signal cable: see Terminal assignment → Page 11
g Cable for external termination (only for PROFIBUS DP with permanent assignment communication board):
Terminal No. 24: +5 V
Terminal No. 25: DGND

10 Endress+Hauser
Proline Promass 80H, 83H

Electrical connection, Promass 80


terminal assignment
Terminal No. (inputs/outputs)

Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)

80***-***********A - - Frequency output Current output, HART

80***-***********D Status input Status output Frequency output Current output, HART

80***-***********H - - - PROFIBUS PA

Frequency output Current output Ex i


80***-***********S - -
Ex i, passive Active, HART

Frequency output Current output Ex i


80***-***********T - -
Ex i, passive Passive, HART

80***-***********8 Status input Frequency output Current output 2 Current output 1,


HART

Promass 83
The inputs and outputs on the communication board can be either permanently assigned (fixed) or variable
(flexible), depending on the version ordered (see table). Replacements for modules which are defective or
which have to be replaced can be ordered as accessories.

Terminal No. (inputs/outputs)

Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)

Fixed communication boards (permanent assignment)

Current output
83***-***********A - - Frequency output
HART

Current output
83***-***********B Relay output Relay output Frequency output
HART

83***-***********F - - - PROFIBUS PA, Ex i

FOUNDATION
83***-***********G - - -
Fieldbus Ex i

83***-***********H - - - PROFIBUS PA

83***-***********J - - +5V (ext. termination) PROFIBUS DP

- FOUNDATION
83***-***********K - -
Fieldbus

83***–***********Q - - Status input MODBUS RS485

83***-***********R - - Current output 2 Current output 1


Ex i, active Ex i active, HART

Frequency output Current output Ex i


83***-***********S - -
Ex i, passive Active, HART

Frequency output Current output Ex i


83***-***********T - -
Ex i, passive Passive, HART

83***-***********U - - Current output 2 Current output 1


Ex i, passive Ex i passive, HART

Flexible communication boards

Current output
83***-***********C Relay output 2 Relay output 1 Frequency output
HART

Current output
83***-***********D Status input Relay output Frequency output
HART

Current output 1
83***-***********E Status input Relay output Current output 2
HART

Current output
83***-***********L Status input Relay output 2 Relay output 1
HART

Endress+Hauser 11
Proline Promass 80H, 83H

Terminal No. (inputs/outputs)

Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)

Current output
83***-***********M Status input Frequency output 2 Frequency output 1
HART

83***–***********N Current output Frequency output Status input MODBUS RS485

83***–***********P Current output Frequency output Status input PROFIBUS DP

83***–***********V Relay output 2 Relay output 1 Status input PROFIBUS DP

Current output 1
83***-***********W Relay output Current output 3 Current output 2
HART

Current output 1
83***-***********0 Status input Current output 3 Current output 2
HART

Current output 1
83***-***********2 Relay output Current output 2 Frequency output
HART

83***-***********3 Current input Relay output Current output 2 Current output 1


HART

83***-***********4 Current input Relay output Frequency output Current output


HART

83***-***********5 Status input Current input Frequency output Current output


HART

83***-***********6 Status input Current input Current output 2 Current output


HART

83***–***********7 Relay output 2 Relay output 1 Status input MODBUS RS485

Electrical connection
Remote version S1 S1 S2 S2 GND TM TM TT TT
a b + + + +

4 5 6 7 8 9 10 11 12 41 42

d
d

d
e

c 4 5 6 7 8 9 10 11 12 41 42
+ + + +
S1 S1 S2 S2 GND TM TM TT TT
a0003681

Connecting the remote version


a Wall-mount housing: non-hazardous area and ATEX II3G / zone 2 → see separate "Ex documentation"
b Wall-mount housing: ATEX II2G / Zone 1 /FM/CSA → see separate "Ex documentation"
c Remote version, flanged version
Terminal No.: 4/5 = gray; 6/7 = green; 8 = yellow; 9/10 = pink; 11/12 = white; 41/42 = brown

Supply voltage 85 to 260 V AC, 45 to 65 Hz


20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC

12 Endress+Hauser
Proline Promass 80H, 83H

Cable entries Power-supply and signal cables (inputs/outputs):


• Cable entry M20 x 1.5 (8 to 12 mm) (0.31" to 0.47")
• Thread for cable entries, 1/2" NPT, G 1/2"
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8 to 12 mm) (0.31" to 0.47")
• Thread for cable entries, 1/2" NPT, G 1/2"

Cable specification • 6 x 0.38 mm2 (PVC cable with common shield and individually shielded cores
Remote version • Conductor resistance: ≤ 50 Ω/km
• Capacitance: core/shield: ≤ 420 pF/m
• Cable length: max. 20 m (65 ft)
• Permanent operating temperature: max. +105 °C (221 °F)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC
requirements of EN 61326/A1, and NAMUR recommendation NE 21/43.

Power consumption AC: <15 VA (including sensor)


DC: <15 W (including sensor)
Switch-on current:
• Max. 13.5 A (< 50 ms) at 24 V DC
• Max. 3 A (< 5 ms) at 260 V AC

Power supply failure Promass 80


Lasting min. 1 power cycle:
• EEPROM saves measuring system data if the power supply fails
• HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial
number, calibration factor, zero point, etc.)

Promass 83
Lasting min. 1 power cycle:
• EEPROM and T-DAT save the measuring system data if the power supply fails.
• HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial
number, calibration factor, zero point, etc.)

Potential equalization No measures necessary.

Performance characteristics
Reference operating Error limits following ISO/DIS 11631:
conditions • 20 to 30 °C (68 to 86 °F); 2 to 4 bar (30 to 60 psi)
• Accuracy based on accredited calibration rigs according to ISO 17025
• Zero point calibrated under operating conditions
• Field density calibrated (or special density calibration)

Maximum measured error The following values refer to the pulse/frequency output. Measured error at the current output is typically
±5 A.
o.r. = of reading

Mass flow (liquid):

Promass 80:
±0.175% ± [(zero point stability / measured value) ⋅100]% o.r.

Promass 83:
±0.125% ± [(zero point stability / measured value) ⋅100]% o.r.

Endress+Hauser 13
Proline Promass 80H, 83H

Volume flow (liquid)


±0.50% ± [(zero point stability / measured value) ⋅100]% o.r.

Zero point stability:

Max. full scale value Zero point stability

DN [kg/h] [lb/min] [kg/h] [lb/min]

8 3/8" 2000 73.5 0.200 0.007

15 1/2" 6500 238 0.650 0.024

25 1" 18000 660 1.80 0.066

40 1 ½" 45000 1 650 4.50 0.165

50 2" 70000 2 570 7.00 0.257

Sample calculation

[%]

±1.0

±0.5
Promass 83
Promass 80

±0.2

0
0 1 2 4 6 8 10 12 14 16 18 t/h
a0004611

Max. measured error in % of measured value (example: Promass 80H, 83H/ DN 25)

Calculation example (mass flow, liquid):


Given: Promass 83 H / DN 25, flow measured value = 8000 kg/h
Max. measured error: ±0.125% ± [(zero point stability / measured value) ⋅ x 100]% o.r.
Max. measured error: ±0.125% ±1.8 kg/h : 8000 kg/h ⋅ 100% = ±0.15%

Density (liquid)
1 g/cc = 1 kg/l

Standard calibration:
±0.02 g/cc

Special density calibration (optional), calibration range: 0.8 to 1.8 g/cc, 5 to 80 °C:
±0.002 g/cc

After field density calibration or under reference conditions:


±0.0010 g/cc

Temperature
±0.5 °C ±0.005 ⋅ T (T = fluid temperature in °C)
±1 °F ±0.003 ⋅ (T–32) (T = fluid temperature in °F)

14 Endress+Hauser
Proline Promass 80H, 83H

Repeatability Mass flow (liquid):


±0.05% ± [1/2 ⋅(zero point stability / measured value) ⋅100]% o.r.

Volume flow (liquid):


±0.20% ± [1/2 ⋅(zero point stability / measured value) ⋅100]% o.r.
o.r. = of reading
Zero point stability: see "Max. measured error" → Page 13 ff.
Calculation example (mass flow, liquid):
Given: Promass 83 H / DN 25, flow measured value = 8000 kg/h
Repeatability: ±0.05% ± [1/2 ⋅(zero point stability / measured value) ⋅ x 100]% o.r.
Repeatability: ±0.05% ±1/2 ⋅ 1.8 kg/h : 8000 kg/h ⋅ 100% = ±0.061%

Density measurement (liquid)


1 g/cc = 1 kg/l
±0.0005 g/cc

Temperature measurement
±0.25 °C ±0.0025 ⋅ T (T = fluid temperature in °C)
(±0.5 °F ±0.0015 ⋅ (T – 32), T = fluid temperature in °F)

Influence of fluid temperature When there is a difference between the temperature for zero point adjustment and the process temperature,
the typical measured error of the Promass sensor is ±0.0002% of the full scale value / °C (±0.0001% of the
full scale value / °F).

Influence of fluid pressure The table below shows the effect on accuracy of mass flow due to a difference between calibration pressure
and process pressure.

DN [% o.r./bar] [% o.r./psi]

8 3/8" –0.017 –0.0012

15 1/2" –0.021 –0.0014

25 1" –0.013 –0.0008

40 1 ½" –0.018 –0.0012

50 2" –0.020 –0.0014

Endress+Hauser 15
Proline Promass 80H, 83H

Operating conditions: Installation


Installation instructions Note the following points:
• No special measures such as supports are necessary. External forces are absorbed by the construction of the
instrument, for example the secondary containment.
• The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring
system is not influenced by pipe vibrations.
• No special precautions need to be taken for fittings which create turbulence (valves, elbows, T-pieces etc.),
as long as no cavitation occurs.
• For mechanical reasons and to protect the pipe, support is recommended for heavy sensors.

Mounting location
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
Therefore, avoid the following mounting locations in the pipe installation:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipeline.

a0003605

Mounting location

Notwithstanding the above, the installation proposal below permits installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the
sensor running empty while measurement is in progress.

5
a0003597

Installation in a down pipe (e.g. for batching applications)


1 = Supply tank, 2 = Sensor, 3 = Orifice plate, pipe restriction (see Table), 4 = Valve, 5 = Batching tank

mm 8 15 25 40 50
DN
inch 3/8" 1/2" 1" 1 ½" 2"

mm 6 10 14 22 28
Ø Orifice plate, pipe restriction
inch 0.24" 0.40" 0.55" 0.87" 1.10"

16 Endress+Hauser
Proline Promass 80H, 83H

Orientation
Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of flow
(direction of fluid flow through the pipe.

Vertical (view V)
Recommended orientation with upward direction of flow. When fluid is not flowing, entrained solids will sink
down and gases will rise away from the measuring tube. The measuring tubes can be completely drained and
protected against solids buildup.

Horizontal
The transmitter can be installed in any orientation in a horizontal pipe run.

Standard, compact

Standard, remote
Fig. V:
ÃÃ ÃÃ
Vertical orientation

a0004572

Fig. H1:
Horizontal orientation ÃÃ ÃÃ
Transmitter head up
a0004576

Fig. H2:
ÃÃ ÃÃ
Horizontal orientation
m m
Transmitter head down
a0004580

ÃÃ = Recommended orientation
à = Orientation recommended in certain situations
✘ = Impermissible orientation

m = To ensure that the maximum permitted ambient temperature for the transmitter (–20 to +60 °C
(–4 to +140° F), optionally –40 to +60 °C (–40 to +140 °F)) is not exceeded, for low-temperature fluids,
we recommend the horizontal orientation with the transmitter head up (Fig. H1) or the vertical orientation
(Fig. V).

Endress+Hauser 17
Proline Promass 80H, 83H

Special installation instructions for Promass H

" Caution!
The measuring tube of the Promass H is slightly curved. The position of the sensor, therefore, has to be matched
to the fluid properties when the sensor is installed horizontally.

1 2

a0004581

Horizontal installation with Promass H


1 Not suitable for fluids with entrained solids. Risk of solids accumulating.
2 Not suitable for outgassing fluids. Risk of air accumulating.

Heating
Some fluids require suitable measures to avoid heat transfer at the sensor. Heating can be electric, e.g. with
heated elements, or by means of hot water or steam pipes made of copper or heating jackets.

" Caution!
• Risk of electronics overheating! Make sure that the maximum permissible ambient temperature for the
transmitter is not exceeded. Consequently, make sure that the adapter between the sensor and transmitter
and the connection housing of the remote version always remain free of insulating material. Note that a
certain orientation might be required, depending on the fluid temperature. → Page 17
• With a fluid temperature between 200 °C to 350 °C the remote version of the high-temperature version is
preferable.
• If using an electric trace heating system whose heating is regulated via phase angle control or pulse packages,
influence on the measured values cannot be ruled out due to magnetic fields (i.e. for values that are greater
than the values approved by the EN standard (sine 30 A/m)). In such cases, the sensor must be magnetically
shielded.
The secondary containment can be shielded with tin plates or electric sheets without preferential direction
(e.g. V330-35A) with the following properties:
– Relative magnetic permeability μr ≥ 300
– Plate thickness d ≥ 0.35 mm (0.0011")
• Information on permitted temperature ranges → Page 20
Special heating jackets, which can be ordered separately from Endress+Hauser as an accessory, are available
for the sensors.

Zero point adjustment


All Promass devices are calibrated to state-of-the-art technology. The zero point determined in this way is
imprinted on the nameplate.
Calibration takes place under reference conditions. → Page 13 ff.
Promass therefore does not require zero point adjustment!
Experience shows that the zero point adjustment is advisable only in special cases:
• To achieve highest measuring accuracy also with very low flow rates
• Under extreme process or operating conditions (e.g. very high process temperatures or very high-viscosity
fluids).

18 Endress+Hauser
Proline Promass 80H, 83H

Inlet and outlet runs There are no installation requirements regarding inlet and outlet runs.

Length of connecting cable Max. 20 meters (66 ft) (remote version)

System pressure It is important to ensure that cavitation does not occur, because it would influence the oscillation of the
measuring tube. No special measures need to be taken for fluids which have properties similar to water under
normal conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is
important to ensure that pressure does not drop below the vapor pressure and that the liquid does not start to
boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects
can be prevented when system pressure is sufficiently high.
Therefore, the following locations should be preferred for installation:
• Downstream from pumps (no danger of vacuum)
• At the lowest point in a vertical pipe

Operating conditions: Environment

Ambient temperature range Standard: –20 to +60 °C (-4 to +140°F) (sensor, transmitter)
Optional: –40 to +60 °C (-40 to +140°F) (sensor, transmitter)

! Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.

Storage temperature –40 to +80 °C (–40 to +175 °F) (preferably +20 °C (+68 °F)

Degree of protection Standard: IP 67 (NEMA 4X) for transmitter and sensor

Shock resistance According to IEC 68-2-31

Vibration resistance Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6

Electromagnetic compatibility As per IEC/EN 61326 and NAMUR recommendation NE 21


(EMC)

Endress+Hauser 19
Proline Promass 80H, 83H

Operating conditions: Process

Fluid temperature range Sensor


–50 to +200 °C (–58 to +392 °F)

Fluid pressure range Flanges:


(nominal pressure)
According to DIN PN 40 / According to ASME B16.5 Cl 150, Cl 300 / JIS 10K, 20K

Pressure ranges of secondary containment:


DN 8 to 15 (3/8" to 1/2"): 25 bar (362 psi)
DN 25 to 50 (1" to 2"): 16 bar (232 psi)

# Warning!
In case a danger of measuring tube failure exists due to process characteristics, e.g. with corrosive process
fluids, we recommend the use of sensors whose secondary containment is equipped with special pressure
monitoring connections (ordering option). With the help of these connections, fluid collected in the secondary
containment in the event of tube failure can be bled off. Dimensions → Page 26 ff.

Limiting flow See information in the "Measuring range" section → Page 6


Select nominal diameter by optimizing between required flow range and permissible pressure loss. See the
"Measuring range" section for a list of maximum possible full scale values.
• The minimum recommended full scale value is approx. 1/20 of the max. full scale value.
• In most applications, 20 to 50% of the maximum full scale value can be considered ideal
• Select a lower full scale value for abrasive substances such as fluids with entrained solids (flow velocity
<1 m/s).

Pressure loss (SI units) Pressure loss depends on the fluid properties and on the flow rate. The following formulae can be used to
approximately calculate the pressure loss:

4·g
Reynolds number Re =
p·d·n·r
a0003381

Re ≥ 2300 * Dp = K · n0.25 · g1.75 · r –0.75 + K3 r· g


a0004631

Re < 2300 Dp = K1 · n · g + K3 r· g
a0004633

Δp = pressure loss [mbar] ρ = fluid density [kg/m3]


ν = kinematic viscosity [m2/s] d = inside diameter of measuring tubes [m]
m = mass flow [kg/s] K to K3 = constants (depending on nominal diameter)

* To compute the pressure loss for gases, always use the formula for Re ≥ 2300.

20 Endress+Hauser
Proline Promass 80H, 83H

Pressure loss coefficients for Promass H

DN d [m] K K1 K3

8 8.51 ⋅ 10–3 8.04 ⋅ 106 3.28 ⋅107 1.15 ⋅ 106

15 12.00 ⋅ 10–3 1.81 ⋅ 106 9.99 ⋅ 106 1.87 ⋅ 105

25 17.60 ⋅ 10–3 3.67 ⋅ 105 2.76 ⋅ 106 4.99 ⋅ 104

40 25.50 ⋅ 10–3 8.75 ⋅ 104 8.67 ⋅ 105 1.22 ⋅ 104

50 40.50 ⋅ 10–3 1.35 ⋅ 104 1.72 ⋅ 105 1.20 ⋅ 103

Pressure loss data includes interface between measuring tube and piping

[mbar]
1000
DN 8 DN 15 DN 25 DN 40 DN 50

100

10

0.1
0.01 0.1 1 10 100 [t/h]
a0004607

Pressure loss diagram for water

Pressure loss (US units) Pressure loss ist dependent on fluid properties nominal diameter. Consult Endress+Hauser for Applicator PC
software to determine pressure loss in US units. All important instrument data is contained in the Applicator
software programm in order to optimize the design of measuring system. The software is used for following
calculations:
• Nominal diameter of the sensor with fluid characteristics such as viscosity, density, etc.
• Pressure loss downstream ot the measuring point.
• Converting mass flow to volume flow, etc.
• Simultaneous display of various meter size.
• Determining measuring ranges.
The Applicator runs on any IBM compatible PC with windows.

Endress+Hauser 21
Proline Promass 80H, 83H

Mechanical construction
Design/dimensions Dimensions:
in metric units
Field housing compact version, powder-coated die-cast aluminum → Page 22

Transmitter compact version, stainless steel field housing → Page 23

Dimensions: Remote version → Page 23

Transmitter connection housing remote version (II2G/Zone 1) → Page 24

Transmitter wall-mount housing (non Ex-zone and II3G/Zone 2) → Page 25

Flange connections EN (DIN), ASME B16.5, JIS → Page 26

Promass purge connections / secondary containment monitoring

Purge connections / secondary containment monitoring → Page 27

Design/dimensions in US units → Page 28 ff.

Field housing compact version, powder-coated die-cast aluminum

A B
A* C

D
Esc
- +

E
E

G
F
+1.5
di L
–2.0

a0006962-en

A A* B C D

227 207 187 168 160

All dimensions in [mm];


* Blind version (without local display)

DN E F G L di

8 280 104 384 * *

15 280 105 385 * *

25 280 121 401 * *

40 304 171 475 * *

50 315 241 556 * *

All dimensions in [mm];


* dependent on respective process connection,
→ For dimensions, see the following pages

! Note!
Dimensions for transmitters II2G/Zone 1 → Page 24.

22 Endress+Hauser
Proline Promass 80H, 83H

Transmitter compact version, stainless steel field housing

A B

C
Esc
- + E

a0002245

A B C

225 153 168

All dimensions in [mm]

Dimensions: Remote version

A B

C
a0002516

DN A B C

8 118.5 137.5 127

15 118.5 137.5 127

25 118.5 137.5 127

40 118.5 137.5 151

50 118.5 137.5 162

Endress+Hauser 23
Proline Promass 80H, 83H

Transmitter connection housing remote version (II2G/Zone 1)

A
A*
B C
B* D

e r Sp annu
ht unt ng
Nic öffnen

Nicht-eigensichere

Ke ircuit s a re aliv ile

Kee ircuits are ali hile


Stromkreise durch

ove r tight wh
e

c
IP40-Abdeckung geschützt

p cover tight w
Non-intrinsically safe
circuits Ip40 protected

E
ep c
Esc
Boucles de courant - +

c
E

ve
sans sécurité intrinsèque
protégées par Ip40

sou s nsion l Ne
pas o t e pparei
uvrir l’a

M
L
J

K
F
G

a0002128

A A* B B* C D E

265 242 240 217 206 186 167

F G H J K L M

∅ 8.6 (M8) 100 123 100 133 188 355

* Blind version (without local display)


All dimensions in [mm]

24 Endress+Hauser
Proline Promass 80H, 83H

Transmitter wall-mount housing (non Ex-zone and II3G/Zone 2)

D
Esc
- + E

B
C
F G
A E

H J K J

N
S S

M
R

L
O
P Q P

a0001150

A B C D E F G H J

215 250 90.5 159.5 135 90 45 >50 81

K L M N O P Q R S

53 95 53 102 81.5 11.5 192 8xM5 20

All dimensions in [mm]

Endress+Hauser 25
Proline Promass 80H, 83H

Flange connections EN (DIN), ASME B16.5, JIS

N
Esc
- +

LK
G
U
E

+1.5
di L
–2.0

a0003287-en

Flange according to EN 1092-1 (DIN 2501 / DIN 2512N 1)) / PN 40:


1.4301/304, fluid wetted parts: zirconium 702

Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 μm

DN G L N S LK U di
1)
8 95 336 4 x Ø14 20 65 17.30 8.51

15 95 440 4 x Ø14 20 65 17.30 12.00

25 115 580 4 x Ø14 19 85 28.50 17.60

40 150 794 4 x Ø18 21.5 110 43.10 25.50

50 165 1071 4 x Ø18 23.5 125 54.50 40.50

All dimensions in [mm]; Further dimensions → Page 22 ff.


1)
DN 8 with DN 15 flanges as standard

Flange according to ASME B16.5 / Cl 150: 1.4301/304, fluid wetted parts: zirconium 702

Surface roughness (flange): Ra 3.2 to 6.3 μm

DN G L N S LK U di

8 1) 88.9 336 4 x Ø15.7 12.8 60.5 15.70 8.51

15 88.9 440 4 x Ø15.7 12.8 60.5 15.70 12.00

25 108.0 580 4 x Ø15.7 15.1 79.2 26.70 17.60

40 127.0 794 4 x Ø15.7 17.5 98.6 40.90 25.50

50 152.4 1071 4 x Ø19.1 23.6 120.7 52.60 40.50

All dimensions in [mm]; Further dimensions → Page 22 ff.


1)
DN 8 with DN 15 flanges as standard

26 Endress+Hauser
Proline Promass 80H, 83H

Flange according to ASME B16.5 / Cl 300: 1.4301/304, fluid wetted parts: zirconium 702

Surface roughness (flange): Ra 3.2 to 6.3 μm

DN G L N S LK U di

8 1) 95.2 336 4 x Ø15.7 14.2 66.5 15.70 8.51

15 95.2 440 4 x Ø15.7 14.2 66.5 15.70 12.00

25 124.0 580 4 x Ø19.1 17.5 88.9 26.70 17.60

40 155.4 794 4 x Ø22.3 20.6 114.3 40.90 25.50

50 165.1 1071 4 x Ø19.1 23.6 127 52.60 40.50

All dimensions in [mm]; Further dimensions → Page 22 ff.


1)DN 8 with DN 15 flanges as standard

Flange JIS B2238 / 20K: 1.4301/304, fluid wetted parts: zirconium 702

Surface roughness (flange): Ra 3.2 to 6.3 μm

DN G L N S LK U di

8 1) 95 336 4 x Ø15 14 70 15.00 8.51

15 95 440 4 x Ø15 14 70 15.00 12.00

25 125 580 4 x Ø19 16 90 25.00 17.60

40 140 794 4 x Ø19 18 105 40.00 25.50

50 165 1071 8 x Ø19 22 120 50.00 40.50

All dimensions in [mm]; Further dimensions → Page 22 ff.


1)
DN 8 with DN 15 flanges as standard

Purge connections / secondary containment monitoring

" Caution!
• The secondary containment is filled with dry nitrogen (N2). Do not open the purge connections unless the
containment can be filled immediately with a dry inert gas. Use only low gauge pressure to purge. Maximum
pressure: 5 bar.
• Purge connections or secondary containment monitoring can not be combined with separately available
heating jacket.

L L
H
G
A

a0003288

Endress+Hauser 27
Proline Promass 80H, 83H

DN L H A G

8 55 82 25 1/2" NPT

15 102 82 25 1/2" NPT

25 172 82 25 1/2" NPT

40 263 102 45 1/2" NPT

50 381.5 119.5 58 1/2" NPT

All dimensions in [mm]; Further dimensions → Page 22 ff.

Design/dimensions Dimensions:
in US units
Field housing compact version, powder-coated die-cast aluminum → Page 28

Transmitter compact version, stainless steel field housing → Page 29

Dimensions: Remote version → Page 29

Transmitter connection housing remote version (II2G/Zone 1) → Page 30

Transmitter wall-mount housing (non Ex-zone and II3G/Zone 2) → Page 31

Flange connections ASME B16.5 → Page 32

Promass purge connections / secondary containment monitoring

Purge connections / secondary containment monitoring → Page 33

Design/dimensions in metric units → Page 22 ff.

Field housing compact version, powder-coated die-cast aluminum

A B
A* C
D

Esc
- +
E

E
G
F

+1.5 (+0.05)
di L
–2.0 (–0.07)

mm (inch)
a0006962-ae

A A* B C D

9.08 8.28 7.48 6.72 6.40

All dimensions in [inch];


* Blind version (without local display)

28 Endress+Hauser
Proline Promass 80H, 83H

DN E F G L di

3/8" 11.02 4.09 15.12 * *

1/2" 11.02 4.13 15.16 * *

1" 11.02 4.76 15.79 * *

1 ½" 11.97 6.73 18.70 * *

2" 12.40 9.49 21.89 * *

All dimensions in [inch];


* dependent on respective process connection,
→ For dimensions, see the following pages

! Note!
Dimensions for transmitters II2G/Zone 1 → Page 24.

Transmitter compact version, stainless steel field housing

A B

C
Esc
- + E

a0002245

A B C

8.86 6.02 6.61

All dimensions in [inch]

Dimensions: Remote version

A B
C

a0002516

T = dimension B in the compact version (with corresponding nominal diameter) minus 153 mm

DN A B C

3/8" 4.67 5.41 4.45

1/2" 4.67 5.41 4.45

1" 4.67 5.41 4.45

1 ½" 4.67 5.41 4.65

2" 4.67 5.41 5.12

All dimensions in [inch]

Endress+Hauser 29
Proline Promass 80H, 83H

Transmitter connection housing remote version (II2G/Zone 1)

A
A*
B C
B* D

e r Sp annu
ht unt ng
Nic öffnen

Nicht-eigensichere

Ke ircuit s a re aliv ile

Kee ircuits are ali hile


Stromkreise durch

ove r tight wh
e

c
IP40-Abdeckung geschützt

p cover tight w
Non-intrinsically safe
circuits Ip40 protected

E
ep c
Esc
Boucles de courant - +

c
E

ve
sans sécurité intrinsèque
protégées par Ip40

sou s nsion l Ne
pas o t e pparei
uvrir l’a

M
L
J

K
F
G

a0002128

A A* B B* C D E

10.43 9.53 9.45 8.54 8.11 7.32 6.57

* Blind version (without local display)

F G H J K L M

∅ 0.34 (M8) 3.94 4.84 3.94 5.24 7.40 13.98

All dimensions in [inch]

30 Endress+Hauser
Proline Promass 80H, 83H

Transmitter wall-mount housing (non Ex-zone and II3G/Zone 2)

D
Esc
- + E

B
C
F G
A E

H J K J

N
S S

M
R

L
O
P Q P

a0001150

A B C D E F G H J

8.46 9.84 3.56 6.28 5.31 3.54 1.77 >1.97 3.91

K L M N O P Q R S

2.09 3.74 2.09 4.02 3.21 0.45 7.56 8xM5 0.79

All dimensions in [inch]

Endress+Hauser 31
Proline Promass 80H, 83H

Flange connections ASME B16.5

N
Esc
- +

LK
G
U
E

+1.5 (+0.05)
di L
–2.0 (–0.07)
mm (inch)
a0003287-ae

Flange according to ASME B16.5 / Cl 150: 1.4301/304, fluid wetted parts: zirconium 702

Surface roughness (flange): Ra 3.2 to 6.3 μm

DN G L N S LK U di

3/8" 1) 3.50 13.23 4 x Ø0.62 0.50 2.38 0.62 0.34

1/2" 3.50 17.32 4 x Ø0.62 0.50 2.38 0.62 0.47

1" 4.25 22.83 4 x Ø0.62 0.59 3.12 1.05 0.69

1 ½" 5.00 31.26 4 x Ø0.62 0.69 3.88 1.61 1.00

2" 6.00 42.17 4 x Ø0.75 0.93 4.75 2.07 1.59

All dimensions in [inch]; Further dimensions → Page 28 ff.


1) DN 3/8" with DN 1/2" flanges as standard

Flange according to ASME B16.5 / Cl 300: 1.4301/304, fluid wetted parts: zirconium 702

Surface roughness (flange): Ra 3.2 to 6.3 μm

DN G L N S LK U di

3/8" 1) 3.75 13.23 4 x Ø0.62 0.56 2.62 0.62 0.34

1/2" 3.75 17.32 4 x Ø0.62 0.56 2.62 0.62 0.47

1" 4.88 22.83 4 x Ø0.75 0.69 3.50 1.05 0.69

1 ½" 6.12 31.26 4 x Ø0.88 0.81 4.50 1.61 1.00

2" 6.50 42.17 4 x Ø0.75 0.93 5.00 2.07 1.59

All dimensions in [inch]; Further dimensions → Page 28 ff.


1) DN 3/8" with DN 1/2" flanges as standard

32 Endress+Hauser
Proline Promass 80H, 83H

Purge connections / secondary containment monitoring

" Caution!
• The secondary containment is filled with dry nitrogen (N2). Do not open the purge connections unless the
containment can be filled immediately with a dry inert gas. Use only low gauge pressure to purge. Maximum
pressure: 5 bar.
• Purge connections or secondary containment monitioring can not be combined with separately available
heating jacket.

L L
H
G

A
a0003288

DN L H A G

3/8" 2.17 3.23 0.98 1/2" NPT

1/2" 4.02 3.23 0.98 1/2" NPT

1" 6.77 3.23 0.98 1/2" NPT

1 ½" 10.35 4.02 1.77 1/2" NPT

2" 15.02 4.70 2.28 1/2" NPT

All dimensions in [inch]; Further dimensions → Page 28 ff.

Weight • Compact version: see table below


• Remote version
– Sensor: see table below
– Wall-mount housing: 5 kg (11 lb)

DN [mm] 8 15 25 40 50

Compact version 13 15 21 43 80

Remote version 11 13 19 41 78

DN [inch] 3/8" 1/2" 1 1 ½" 2"

Compact version 29 33 46 95 176

Remote version 24 29 42 90 172

All values (weight) refer to devices with EN/DIN PN 40 flanges.


Weight information in [kg] and [inch].

Endress+Hauser 33
Proline Promass 80H, 83H

Materials Transmitter housing:


• Compact housing: stainless steel 1.4301/304
• Compact housing: powder coated die-cast aluminum
• Wall-mount housing: powder coated die-cast aluminum
• Remote field housing: powder-coated die-cast aluminum

Sensor housing / containment:


• Acid and alkali-resistant outer surface
• Stainless steel 1.4301/304

Connection housing, sensor (remote version):


• Stainless steel 1.4301/304

Process connections
• Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / JIS B2220
→ stainless steel 1.4301/304, fluid wetted parts: zirconium 702

Measuring tubes:
• Zirconium 702/R 60702

Seals:
Welded process connections without internal seals

Material load curves

# Warning!
The following material loade curves refer to the entire sensor and not just the process conection.

Flange according to connection to EN 1092-1 (DIN 2501)


Flange material: 1.4301/301
Fluid wetted parts: Zirconium 702

[psi] [bar]
700 50
600 40
500 PN 40
30
400
300 20
200
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
[°F]
-80 -40 0 40 80 120 160 200 240 280 320 360 400
a0003289-ae

34 Endress+Hauser
Proline Promass 80H, 83H

Flange connection according to ASME B16.5


Flange material: 1.4301/304
Fluid wetted parts: Zirconium 702

[psi] [bar]
700 50
600 40 Class 300
500
30
400
300 20
200
Class 150
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
[°F]
-80 -40 0 40 80 120 160 200 240 280 320 360 400
a0003290-ae

Flange connection to JIS B2220


Flange material: 1.4301/304
Fluid wetted parts: Zirconium 702

[psi] [bar]
30
400
300 20
20K
200
10
100 10K
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
[°F]
-80 -40 0 40 80 120 160 200 240 280 320 360 400

a0003324-ae

Process connections Welded process connections:


• Flanges according to EN 1092-1 (DIN 2501), according to ASME B16.5, JIS B2220

Endress+Hauser 35
Proline Promass 80H, 83H

Human interface
Display elements • Liquid-crystal display: backlit, two lines (Promass 80) or four lines (Promass 83) with 16 characters per line
• Selectable display of different measured values and status variables
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.

Unified control concept for Promass 80:


both types of transmitter
• Local operation with three keys (–, +, E)
• Quick Setup menus for straightforward commissioning

Promass 83:
• Local operation with three optical keys (S/O/F)
• Application-specific Quick Setup menus for straightforward commissioning

Language groups

! Note!
The language group is changed using the "FieldCare" operating program.
Language groups available for operation in different countries:
• Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch and Portuguese
• Eastern Europe/Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish and Czech
• South and Eastern Asia (SEA):
English, Japanese, Indonesian
Only Promass 83:
• China (CN):
English, Chinese

Remote operation Promass 80:


Remote operation via HART, PROFIBUS PA

Promass 83:
Remote operation via HART, PROFIBUS PA/DP, FOUNDATION fieldbus

36 Endress+Hauser
Proline Promass 80H, 83H

Certificates and approvals


CE mark The measuring system is in conformity with the statutory requirements of the EC Directives.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.

C-Tick symbol The measuring system complies with the EMC requirements of the "Australian Communications and Media
Authority (ACMA)"

Ex approval Information about currently available Ex versions (ATEX, FM, CSA, IECEx, NEPSI etc.) can be supplied by your
Endress+Hauser Sales Center on request. All information relevant to explosion protection is available in
separate Ex documents that you can order as necessary.

FOUNDATION Fieldbus The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus FOUNDATION. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1.
• Interoperability Test Kit (ITK), revision status 4.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus FOUNDATION

PROFIBUS DP/PA The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS User Organization). The device thus meets all the requirements of the following
specifications:
• Certified in accordance with PROFIBUS Profile Version 3.0 (device certification number: available on
request)
• The device can also be operated with certified devices of other manufacturers (interoperability)

MODBUS certification The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
"MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed all the
test procedures carried out and is certified by the "MODBUS/TCP Conformance Test Laboratory" of the
University of Michigan.

Other standards and • EN 60529


guidelines Degrees of protection by housing (IP code)
• EN 61010-1
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory
Procedures.
• IEC/EN 61326
"Emission in accordance with Class A requirements".
Electromagnetic compatibility (EMC requirements)
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with analog output
signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics

Pressure Equipment Directive Measuring devices with a nominal diameter smaller than or equal to DN 25 correspond to Article 3(3) of the
EC Directive 97/23/EC (Pressure Equipment Directive) and have been designed and manufactured according
to good engineering practice. For larger nominal diameters, optional approvals according to Cat. II/III are
available when required (depends on fluid and process pressure).

Endress+Hauser 37
Proline Promass 80H, 83H

Functional safety SIL -2:


accordance IEC 61508/IEC 61511-1 (FDIS)
"4–20 mA HART" output according to the following order code:

Promass 80
Promass80***-************A
Promass80***-************D
Promass80***-************S
Promass80***-************T
Promass80***-************8

Promass 83
Promass83***-************A Promass83***-************M Promass83***-************Ø
Promass83***-************B Promass83***-************R Promass83***-************2
Promass83***-************C Promass83***-************S Promass83***-************3
Promass83***-************D Promass83***-************T Promass83***-************4
Promass83***-************E Promass83***-************U Promass83***-************5
Promass83***-************L Promass83***-************W Promass83***-************6

Ordering information
The Endress +Hauser service organization can provide detailed ordering information and information on the
order codes on request.

Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor.

38 Endress+Hauser
Proline Promass 80H, 83H

Documentation
• Flow measurement (FA005D/06/en)
• Technical Information Promass 80F, 80M, 83F, 83M (TI053D/06/en)
• Technical Information Promass 80E, 83E (TI061D/06/en)
• Technical Information Promass 80A, 83A (TI054D/06/en)
• Technical Information Promass 80I, 83I (TI075D/06/en)
• Technical Information Promass 80S, 83S (TI076D/06/en)
• Technical Information Promass 80P, 83P (TI078D/06/en)
• Operating Instructions Promass 80 (BA057D/06/en)
• Operating Instructions Promass 80 PROFIBUS PA (BA072D/06/en)
• Operating Instructions Promass 83 (BA059D/06/en)
• Operating Instructions Promass 83 FOUNDATION Fieldbus (BA065D/06/en)
• Operating Instructions Promass 83 PROFIBUS DP/PA(BA063D/06/en)
• Operating Instructions Promass 83 MODBUS (BA107D/06/en)
• Description of Device Functions Promass 80 (BA058D/06/en)
• Description of Device Functions Promass 80 PROFIBUS PA (BA073D/06/en)
• Description of Device Functions Promass 83 (BA060D/06/en)
• Description of Device Functions Promass 83 FOUNDATION Fieldbus (BA066D/06/en)
• Description of Device Functions Promass 83, PROFIBUS DP/PA (BA064D/06/en)
• Description of Device Functions Promass 83 MODBUS (BA108D/06/en)
• Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx, NEPSI
• Functional safety manual Promass 80, 83 (SD077D/06/en)

Registered trademarks
KALREZ® and VITON®
Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP ®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
SWAGELOK ®
Registered trademark of Swagelok & Co., Solon, USA
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
MODBUS®
Registered trademark of the MODBUS Organization
HistoROM™, S-DAT®, T-DAT™, F-CHIP®, FieldCare®, FieldCare® Fieldcheck®, Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH

Endress+Hauser 39
USA Canada México Instruments International

Endress+Hauser, Inc. Endress+Hauser Canada Endress+Hauser, México, S.A. de C.V.F Endress+Hauser


2350 Endress Place 1075 Sutton Drive Fernando Montes de Oca 21 Edificio A Piso 3 Instruments International AG
Greenwood, IN 46143 Burlington, ON L7L 5Z8 Fracc. Industrial San Nicolas Kaegenstrasse 2
USA Canada 54030. Tlalnepantla de Baz 4153 Reinach
Estado de México Switzerland
Tel. 317-535-7138 Tel. 905-681-9292
México
Fax 317-535-8498 800-668-3199 Tel. +41 61 715 81 00
Sales 888-ENDRESS Fax 905-681-9444 Tel. +52 55 5321 2080 Fax +41 61 715 25 00
Service 800-642-8737 www.ca.endress.com Fax +52 55 5321 2099 www.endress.com
[email protected] [email protected] [email protected]
www.us.endress.com www.mx.endress.com