Coriolis Flow System Guide
Coriolis Flow System Guide
TI074D/06/ae/09.08
71079945
Proline Promass 80H, 83H
Table of contents
2 Endress+Hauser
Proline Promass 80H, 83H
The amplitude of the Coriolis force depends on the moving mass Δm, its velocity v in the system, and thus on
the mass flow. Instead of a constant angular velocity ω, the Promass sensor uses oscillation.
This causes the tube through which the fluid is flowing to oscillate. The Coriolis forces produced at the
measuring tubes cause a phase shift in the tube oscillations (see illustration):
• If there is zero flow, i.e. when the fluid stands still, the oscillation measured at points A and B has the same
phase, and thus there is no phase difference (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the outlet (3).
A A A
B B B
1 2 3
a0003383
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube
oscillations at the inlet and outlet.
For the Promass H, the system balance is created by a counterweight that runs parallel to the measuring tube.
This counterweight oscillates in antiphase to the measuring tubes and thus creates a balanced system. The
patented ITB™ (Intrinsic Tube Balance) system ensures balance and stability, thus providing accurate
measurements over a wide range of process and environmental conditions.
Therefore, the Promass H is just as easy to install as the familiar two-tube systems! Consequently, no special
measures for attachment are required in front of or behind the sensor.
The measuring principle operates independently of temperature, pressure, viscosity, conductivity and flow
profile.
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and thus the
density of the oscillating system (comprising the measuring tube and fluid) results in a corresponding, automatic
adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The
microprocessor utilizes this relationship to obtain a density signal.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor due to
temperature effects. This signal corresponds to the process temperature and is also available as an output.
Endress+Hauser 3
Proline Promass 80H, 83H
Measuring system The measuring system consists of a transmitter and a sensor. Two versions are available:
• Compact version: transmitter and sensor form a mechanical unit.
• Remote version: transmitter and sensor are mounted physically separate from one another.
Transmitter
Esc
- + E
a0003671
a0003672
Sensor
H • Single bent tube. Low pressure loss and chemically Documentation No.
resistant material TI 074D/06/en
• Nominal diameters DN 8 to 50 (3/8" to 2")
• Tube material: zirconium
a0003677
a0003679
a0002271
4 Endress+Hauser
Proline Promass 80H, 83H
a0003673
a0003675
a0003676
a0003678
a0006828
Endress+Hauser 5
Proline Promass 80H, 83H
Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Fluid density (proportional to resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)
Operable flow range Greater than 1000: 1. Flow rates above the preset full scale value do not overload the amplifier, i.e. the totalizer
values are registered correctly.
6 Endress+Hauser
Proline Promass 80H, 83H
Output
Output signal Promass 80
Current output:
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
selectable, temperature coefficient: typically 0.005% o.f.s./°C, resolution: 0.5 μA
• Active: 0/4 to 20 mA, RL < 700 Ω (for HART: RL ≥ 250 Ω)
• Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri ≥ 150 Ω
Pulse/frequency output:
Passive, open collector, 30 V DC, 250 mA, galvanically isolated.
• Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1, pulse width max. 2 s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.5 to 2000 ms)
PROFIBUS PA interface:
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Profile Version 3.0
• Current consumption: 11 mA
• Permitted supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic) = 0 mA
• Data transmission rate: 31.25 kBit/s
• Signal encoding: Manchester II
• Function blocks: 4 x Analog Input, 1 x Totalizer
• Output data: Mass flow, Volume flow, Density, Temperature, Totalizer
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
Promass 83
Current output:
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
selectable, temperature coefficient: typically 0.005% o.f.s./°C, resolution: 0.5 μA
• Active: 0/4 to 20 mA, RL < 700 Ω (for HART: RL ≥ 250 Ω)
• Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri ≥ 150 Ω
Pulse/frequency output:
Active/passive selectable, galvanically isolated
• Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Ω
• Passive: open collector, 30 V DC, 250 mA
• Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1, pulse width max.
2s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms)
PROFIBUS DP interface:
• PROFIBUS DP in accordance with EN 50170 Volume 2
• Profile Version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding: NRZ Code
• Function blocks: 6 x Analog Input, 3 x Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizers 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
• Available output combination → Page 11
Endress+Hauser 7
Proline Promass 80H, 83H
PROFIBUS PA interface:
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Data transmission rate: 31.25 kBit/s
• Current consumption: 11 mA
• Permitted supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Signal encoding: Manchester II
• Function blocks: 6 x Analog Input, 3 x Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizers 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
• Available output combination → Page 11
MODBUS interface:
• MODBUS device type: slave
• Address range: 1 to 247
• Supported function codes: 03, 04, 06, 08, 16, 23
• Broadcast: supported with the function codes 06, 16, 23
• Physical interface: RS485 in accordance with EIA/TIA-485 standard
• Supported baud rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
• Transmission mode: RTU or ASCII
• Response times:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
• Possible output combinations → Page 11
8 Endress+Hauser
Proline Promass 80H, 83H
Low flow cut off Switch points for low flow cut off are selectable.
Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
Endress+Hauser 9
Proline Promass 80H, 83H
Power supply
Electrical connection
Measuring unit A B C
d d
g g
b b
a
a
a
b d/(g) (d)
– 27 PA(–)/FF(–) 27
+ 26 PA(+)/FF(+) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20
N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b
A (RxD/TxD-N) 27 A (RxD/TxD-N) 27
B (RxD/TxD-P) 26 B (RxD/TxD-P) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 g + 20 g
N (L-) 2 c N (L-) 2
c
L1 (L+) 1 L1 (L+) 1
b b
a0002441
10 Endress+Hauser
Proline Promass 80H, 83H
Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
80***-***********D Status input Status output Frequency output Current output, HART
80***-***********H - - - PROFIBUS PA
Promass 83
The inputs and outputs on the communication board can be either permanently assigned (fixed) or variable
(flexible), depending on the version ordered (see table). Replacements for modules which are defective or
which have to be replaced can be ordered as accessories.
Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
Current output
83***-***********A - - Frequency output
HART
Current output
83***-***********B Relay output Relay output Frequency output
HART
FOUNDATION
83***-***********G - - -
Fieldbus Ex i
83***-***********H - - - PROFIBUS PA
- FOUNDATION
83***-***********K - -
Fieldbus
Current output
83***-***********C Relay output 2 Relay output 1 Frequency output
HART
Current output
83***-***********D Status input Relay output Frequency output
HART
Current output 1
83***-***********E Status input Relay output Current output 2
HART
Current output
83***-***********L Status input Relay output 2 Relay output 1
HART
Endress+Hauser 11
Proline Promass 80H, 83H
Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
Current output
83***-***********M Status input Frequency output 2 Frequency output 1
HART
Current output 1
83***-***********W Relay output Current output 3 Current output 2
HART
Current output 1
83***-***********0 Status input Current output 3 Current output 2
HART
Current output 1
83***-***********2 Relay output Current output 2 Frequency output
HART
Electrical connection
Remote version S1 S1 S2 S2 GND TM TM TT TT
a b + + + +
4 5 6 7 8 9 10 11 12 41 42
d
d
d
e
c 4 5 6 7 8 9 10 11 12 41 42
+ + + +
S1 S1 S2 S2 GND TM TM TT TT
a0003681
12 Endress+Hauser
Proline Promass 80H, 83H
Cable specification • 6 x 0.38 mm2 (PVC cable with common shield and individually shielded cores
Remote version • Conductor resistance: ≤ 50 Ω/km
• Capacitance: core/shield: ≤ 420 pF/m
• Cable length: max. 20 m (65 ft)
• Permanent operating temperature: max. +105 °C (221 °F)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC
requirements of EN 61326/A1, and NAMUR recommendation NE 21/43.
Promass 83
Lasting min. 1 power cycle:
• EEPROM and T-DAT save the measuring system data if the power supply fails.
• HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial
number, calibration factor, zero point, etc.)
Performance characteristics
Reference operating Error limits following ISO/DIS 11631:
conditions • 20 to 30 °C (68 to 86 °F); 2 to 4 bar (30 to 60 psi)
• Accuracy based on accredited calibration rigs according to ISO 17025
• Zero point calibrated under operating conditions
• Field density calibrated (or special density calibration)
Maximum measured error The following values refer to the pulse/frequency output. Measured error at the current output is typically
±5 A.
o.r. = of reading
Promass 80:
±0.175% ± [(zero point stability / measured value) ⋅100]% o.r.
Promass 83:
±0.125% ± [(zero point stability / measured value) ⋅100]% o.r.
Endress+Hauser 13
Proline Promass 80H, 83H
Sample calculation
[%]
±1.0
±0.5
Promass 83
Promass 80
±0.2
0
0 1 2 4 6 8 10 12 14 16 18 t/h
a0004611
Max. measured error in % of measured value (example: Promass 80H, 83H/ DN 25)
Density (liquid)
1 g/cc = 1 kg/l
Standard calibration:
±0.02 g/cc
Special density calibration (optional), calibration range: 0.8 to 1.8 g/cc, 5 to 80 °C:
±0.002 g/cc
Temperature
±0.5 °C ±0.005 ⋅ T (T = fluid temperature in °C)
±1 °F ±0.003 ⋅ (T–32) (T = fluid temperature in °F)
14 Endress+Hauser
Proline Promass 80H, 83H
Temperature measurement
±0.25 °C ±0.0025 ⋅ T (T = fluid temperature in °C)
(±0.5 °F ±0.0015 ⋅ (T – 32), T = fluid temperature in °F)
Influence of fluid temperature When there is a difference between the temperature for zero point adjustment and the process temperature,
the typical measured error of the Promass sensor is ±0.0002% of the full scale value / °C (±0.0001% of the
full scale value / °F).
Influence of fluid pressure The table below shows the effect on accuracy of mass flow due to a difference between calibration pressure
and process pressure.
DN [% o.r./bar] [% o.r./psi]
Endress+Hauser 15
Proline Promass 80H, 83H
Mounting location
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
Therefore, avoid the following mounting locations in the pipe installation:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipeline.
a0003605
Mounting location
Notwithstanding the above, the installation proposal below permits installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the
sensor running empty while measurement is in progress.
5
a0003597
mm 8 15 25 40 50
DN
inch 3/8" 1/2" 1" 1 ½" 2"
mm 6 10 14 22 28
Ø Orifice plate, pipe restriction
inch 0.24" 0.40" 0.55" 0.87" 1.10"
16 Endress+Hauser
Proline Promass 80H, 83H
Orientation
Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of flow
(direction of fluid flow through the pipe.
Vertical (view V)
Recommended orientation with upward direction of flow. When fluid is not flowing, entrained solids will sink
down and gases will rise away from the measuring tube. The measuring tubes can be completely drained and
protected against solids buildup.
Horizontal
The transmitter can be installed in any orientation in a horizontal pipe run.
Standard, compact
Standard, remote
Fig. V:
ÃÃ ÃÃ
Vertical orientation
a0004572
Fig. H1:
Horizontal orientation ÃÃ ÃÃ
Transmitter head up
a0004576
Fig. H2:
ÃÃ ÃÃ
Horizontal orientation
m m
Transmitter head down
a0004580
ÃÃ = Recommended orientation
à = Orientation recommended in certain situations
✘ = Impermissible orientation
m = To ensure that the maximum permitted ambient temperature for the transmitter (–20 to +60 °C
(–4 to +140° F), optionally –40 to +60 °C (–40 to +140 °F)) is not exceeded, for low-temperature fluids,
we recommend the horizontal orientation with the transmitter head up (Fig. H1) or the vertical orientation
(Fig. V).
Endress+Hauser 17
Proline Promass 80H, 83H
" Caution!
The measuring tube of the Promass H is slightly curved. The position of the sensor, therefore, has to be matched
to the fluid properties when the sensor is installed horizontally.
1 2
a0004581
Heating
Some fluids require suitable measures to avoid heat transfer at the sensor. Heating can be electric, e.g. with
heated elements, or by means of hot water or steam pipes made of copper or heating jackets.
" Caution!
• Risk of electronics overheating! Make sure that the maximum permissible ambient temperature for the
transmitter is not exceeded. Consequently, make sure that the adapter between the sensor and transmitter
and the connection housing of the remote version always remain free of insulating material. Note that a
certain orientation might be required, depending on the fluid temperature. → Page 17
• With a fluid temperature between 200 °C to 350 °C the remote version of the high-temperature version is
preferable.
• If using an electric trace heating system whose heating is regulated via phase angle control or pulse packages,
influence on the measured values cannot be ruled out due to magnetic fields (i.e. for values that are greater
than the values approved by the EN standard (sine 30 A/m)). In such cases, the sensor must be magnetically
shielded.
The secondary containment can be shielded with tin plates or electric sheets without preferential direction
(e.g. V330-35A) with the following properties:
– Relative magnetic permeability μr ≥ 300
– Plate thickness d ≥ 0.35 mm (0.0011")
• Information on permitted temperature ranges → Page 20
Special heating jackets, which can be ordered separately from Endress+Hauser as an accessory, are available
for the sensors.
18 Endress+Hauser
Proline Promass 80H, 83H
Inlet and outlet runs There are no installation requirements regarding inlet and outlet runs.
System pressure It is important to ensure that cavitation does not occur, because it would influence the oscillation of the
measuring tube. No special measures need to be taken for fluids which have properties similar to water under
normal conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is
important to ensure that pressure does not drop below the vapor pressure and that the liquid does not start to
boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects
can be prevented when system pressure is sufficiently high.
Therefore, the following locations should be preferred for installation:
• Downstream from pumps (no danger of vacuum)
• At the lowest point in a vertical pipe
Ambient temperature range Standard: –20 to +60 °C (-4 to +140°F) (sensor, transmitter)
Optional: –40 to +60 °C (-40 to +140°F) (sensor, transmitter)
! Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
Storage temperature –40 to +80 °C (–40 to +175 °F) (preferably +20 °C (+68 °F)
Endress+Hauser 19
Proline Promass 80H, 83H
# Warning!
In case a danger of measuring tube failure exists due to process characteristics, e.g. with corrosive process
fluids, we recommend the use of sensors whose secondary containment is equipped with special pressure
monitoring connections (ordering option). With the help of these connections, fluid collected in the secondary
containment in the event of tube failure can be bled off. Dimensions → Page 26 ff.
Pressure loss (SI units) Pressure loss depends on the fluid properties and on the flow rate. The following formulae can be used to
approximately calculate the pressure loss:
4·g
Reynolds number Re =
p·d·n·r
a0003381
Re < 2300 Dp = K1 · n · g + K3 r· g
a0004633
* To compute the pressure loss for gases, always use the formula for Re ≥ 2300.
20 Endress+Hauser
Proline Promass 80H, 83H
DN d [m] K K1 K3
Pressure loss data includes interface between measuring tube and piping
[mbar]
1000
DN 8 DN 15 DN 25 DN 40 DN 50
100
10
0.1
0.01 0.1 1 10 100 [t/h]
a0004607
Pressure loss (US units) Pressure loss ist dependent on fluid properties nominal diameter. Consult Endress+Hauser for Applicator PC
software to determine pressure loss in US units. All important instrument data is contained in the Applicator
software programm in order to optimize the design of measuring system. The software is used for following
calculations:
• Nominal diameter of the sensor with fluid characteristics such as viscosity, density, etc.
• Pressure loss downstream ot the measuring point.
• Converting mass flow to volume flow, etc.
• Simultaneous display of various meter size.
• Determining measuring ranges.
The Applicator runs on any IBM compatible PC with windows.
Endress+Hauser 21
Proline Promass 80H, 83H
Mechanical construction
Design/dimensions Dimensions:
in metric units
Field housing compact version, powder-coated die-cast aluminum → Page 22
A B
A* C
D
Esc
- +
E
E
G
F
+1.5
di L
–2.0
a0006962-en
A A* B C D
DN E F G L di
! Note!
Dimensions for transmitters II2G/Zone 1 → Page 24.
22 Endress+Hauser
Proline Promass 80H, 83H
A B
C
Esc
- + E
a0002245
A B C
A B
C
a0002516
DN A B C
Endress+Hauser 23
Proline Promass 80H, 83H
A
A*
B C
B* D
e r Sp annu
ht unt ng
Nic öffnen
Nicht-eigensichere
ove r tight wh
e
c
IP40-Abdeckung geschützt
p cover tight w
Non-intrinsically safe
circuits Ip40 protected
E
ep c
Esc
Boucles de courant - +
c
E
ve
sans sécurité intrinsèque
protégées par Ip40
sou s nsion l Ne
pas o t e pparei
uvrir l’a
M
L
J
K
F
G
a0002128
A A* B B* C D E
F G H J K L M
24 Endress+Hauser
Proline Promass 80H, 83H
D
Esc
- + E
B
C
F G
A E
H J K J
N
S S
M
R
L
O
P Q P
a0001150
A B C D E F G H J
K L M N O P Q R S
Endress+Hauser 25
Proline Promass 80H, 83H
N
Esc
- +
LK
G
U
E
+1.5
di L
–2.0
a0003287-en
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 μm
DN G L N S LK U di
1)
8 95 336 4 x Ø14 20 65 17.30 8.51
Flange according to ASME B16.5 / Cl 150: 1.4301/304, fluid wetted parts: zirconium 702
DN G L N S LK U di
26 Endress+Hauser
Proline Promass 80H, 83H
Flange according to ASME B16.5 / Cl 300: 1.4301/304, fluid wetted parts: zirconium 702
DN G L N S LK U di
Flange JIS B2238 / 20K: 1.4301/304, fluid wetted parts: zirconium 702
DN G L N S LK U di
" Caution!
• The secondary containment is filled with dry nitrogen (N2). Do not open the purge connections unless the
containment can be filled immediately with a dry inert gas. Use only low gauge pressure to purge. Maximum
pressure: 5 bar.
• Purge connections or secondary containment monitoring can not be combined with separately available
heating jacket.
L L
H
G
A
a0003288
Endress+Hauser 27
Proline Promass 80H, 83H
DN L H A G
8 55 82 25 1/2" NPT
Design/dimensions Dimensions:
in US units
Field housing compact version, powder-coated die-cast aluminum → Page 28
A B
A* C
D
Esc
- +
E
E
G
F
+1.5 (+0.05)
di L
–2.0 (–0.07)
mm (inch)
a0006962-ae
A A* B C D
28 Endress+Hauser
Proline Promass 80H, 83H
DN E F G L di
! Note!
Dimensions for transmitters II2G/Zone 1 → Page 24.
A B
C
Esc
- + E
a0002245
A B C
A B
C
a0002516
T = dimension B in the compact version (with corresponding nominal diameter) minus 153 mm
DN A B C
Endress+Hauser 29
Proline Promass 80H, 83H
A
A*
B C
B* D
e r Sp annu
ht unt ng
Nic öffnen
Nicht-eigensichere
ove r tight wh
e
c
IP40-Abdeckung geschützt
p cover tight w
Non-intrinsically safe
circuits Ip40 protected
E
ep c
Esc
Boucles de courant - +
c
E
ve
sans sécurité intrinsèque
protégées par Ip40
sou s nsion l Ne
pas o t e pparei
uvrir l’a
M
L
J
K
F
G
a0002128
A A* B B* C D E
F G H J K L M
30 Endress+Hauser
Proline Promass 80H, 83H
D
Esc
- + E
B
C
F G
A E
H J K J
N
S S
M
R
L
O
P Q P
a0001150
A B C D E F G H J
K L M N O P Q R S
Endress+Hauser 31
Proline Promass 80H, 83H
N
Esc
- +
LK
G
U
E
+1.5 (+0.05)
di L
–2.0 (–0.07)
mm (inch)
a0003287-ae
Flange according to ASME B16.5 / Cl 150: 1.4301/304, fluid wetted parts: zirconium 702
DN G L N S LK U di
Flange according to ASME B16.5 / Cl 300: 1.4301/304, fluid wetted parts: zirconium 702
DN G L N S LK U di
32 Endress+Hauser
Proline Promass 80H, 83H
" Caution!
• The secondary containment is filled with dry nitrogen (N2). Do not open the purge connections unless the
containment can be filled immediately with a dry inert gas. Use only low gauge pressure to purge. Maximum
pressure: 5 bar.
• Purge connections or secondary containment monitioring can not be combined with separately available
heating jacket.
L L
H
G
A
a0003288
DN L H A G
DN [mm] 8 15 25 40 50
Compact version 13 15 21 43 80
Remote version 11 13 19 41 78
Endress+Hauser 33
Proline Promass 80H, 83H
Process connections
• Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / JIS B2220
→ stainless steel 1.4301/304, fluid wetted parts: zirconium 702
Measuring tubes:
• Zirconium 702/R 60702
Seals:
Welded process connections without internal seals
# Warning!
The following material loade curves refer to the entire sensor and not just the process conection.
[psi] [bar]
700 50
600 40
500 PN 40
30
400
300 20
200
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
[°F]
-80 -40 0 40 80 120 160 200 240 280 320 360 400
a0003289-ae
34 Endress+Hauser
Proline Promass 80H, 83H
[psi] [bar]
700 50
600 40 Class 300
500
30
400
300 20
200
Class 150
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
[°F]
-80 -40 0 40 80 120 160 200 240 280 320 360 400
a0003290-ae
[psi] [bar]
30
400
300 20
20K
200
10
100 10K
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
[°F]
-80 -40 0 40 80 120 160 200 240 280 320 360 400
a0003324-ae
Endress+Hauser 35
Proline Promass 80H, 83H
Human interface
Display elements • Liquid-crystal display: backlit, two lines (Promass 80) or four lines (Promass 83) with 16 characters per line
• Selectable display of different measured values and status variables
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
Promass 83:
• Local operation with three optical keys (S/O/F)
• Application-specific Quick Setup menus for straightforward commissioning
Language groups
! Note!
The language group is changed using the "FieldCare" operating program.
Language groups available for operation in different countries:
• Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch and Portuguese
• Eastern Europe/Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish and Czech
• South and Eastern Asia (SEA):
English, Japanese, Indonesian
Only Promass 83:
• China (CN):
English, Chinese
Promass 83:
Remote operation via HART, PROFIBUS PA/DP, FOUNDATION fieldbus
36 Endress+Hauser
Proline Promass 80H, 83H
C-Tick symbol The measuring system complies with the EMC requirements of the "Australian Communications and Media
Authority (ACMA)"
Ex approval Information about currently available Ex versions (ATEX, FM, CSA, IECEx, NEPSI etc.) can be supplied by your
Endress+Hauser Sales Center on request. All information relevant to explosion protection is available in
separate Ex documents that you can order as necessary.
FOUNDATION Fieldbus The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus FOUNDATION. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1.
• Interoperability Test Kit (ITK), revision status 4.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus FOUNDATION
PROFIBUS DP/PA The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS User Organization). The device thus meets all the requirements of the following
specifications:
• Certified in accordance with PROFIBUS Profile Version 3.0 (device certification number: available on
request)
• The device can also be operated with certified devices of other manufacturers (interoperability)
MODBUS certification The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
"MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed all the
test procedures carried out and is certified by the "MODBUS/TCP Conformance Test Laboratory" of the
University of Michigan.
Pressure Equipment Directive Measuring devices with a nominal diameter smaller than or equal to DN 25 correspond to Article 3(3) of the
EC Directive 97/23/EC (Pressure Equipment Directive) and have been designed and manufactured according
to good engineering practice. For larger nominal diameters, optional approvals according to Cat. II/III are
available when required (depends on fluid and process pressure).
Endress+Hauser 37
Proline Promass 80H, 83H
Promass 80
Promass80***-************A
Promass80***-************D
Promass80***-************S
Promass80***-************T
Promass80***-************8
Promass 83
Promass83***-************A Promass83***-************M Promass83***-************Ø
Promass83***-************B Promass83***-************R Promass83***-************2
Promass83***-************C Promass83***-************S Promass83***-************3
Promass83***-************D Promass83***-************T Promass83***-************4
Promass83***-************E Promass83***-************U Promass83***-************5
Promass83***-************L Promass83***-************W Promass83***-************6
Ordering information
The Endress +Hauser service organization can provide detailed ordering information and information on the
order codes on request.
Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor.
38 Endress+Hauser
Proline Promass 80H, 83H
Documentation
• Flow measurement (FA005D/06/en)
• Technical Information Promass 80F, 80M, 83F, 83M (TI053D/06/en)
• Technical Information Promass 80E, 83E (TI061D/06/en)
• Technical Information Promass 80A, 83A (TI054D/06/en)
• Technical Information Promass 80I, 83I (TI075D/06/en)
• Technical Information Promass 80S, 83S (TI076D/06/en)
• Technical Information Promass 80P, 83P (TI078D/06/en)
• Operating Instructions Promass 80 (BA057D/06/en)
• Operating Instructions Promass 80 PROFIBUS PA (BA072D/06/en)
• Operating Instructions Promass 83 (BA059D/06/en)
• Operating Instructions Promass 83 FOUNDATION Fieldbus (BA065D/06/en)
• Operating Instructions Promass 83 PROFIBUS DP/PA(BA063D/06/en)
• Operating Instructions Promass 83 MODBUS (BA107D/06/en)
• Description of Device Functions Promass 80 (BA058D/06/en)
• Description of Device Functions Promass 80 PROFIBUS PA (BA073D/06/en)
• Description of Device Functions Promass 83 (BA060D/06/en)
• Description of Device Functions Promass 83 FOUNDATION Fieldbus (BA066D/06/en)
• Description of Device Functions Promass 83, PROFIBUS DP/PA (BA064D/06/en)
• Description of Device Functions Promass 83 MODBUS (BA108D/06/en)
• Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx, NEPSI
• Functional safety manual Promass 80, 83 (SD077D/06/en)
Registered trademarks
KALREZ® and VITON®
Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP ®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
SWAGELOK ®
Registered trademark of Swagelok & Co., Solon, USA
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
MODBUS®
Registered trademark of the MODBUS Organization
HistoROM™, S-DAT®, T-DAT™, F-CHIP®, FieldCare®, FieldCare® Fieldcheck®, Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
Endress+Hauser 39
USA Canada México Instruments International
TI074D/06/ae/09.08
71079945