GasAlertMicro5 - UserManual (D5615 5 EN)
GasAlertMicro5 - UserManual (D5615 5 EN)
D5615/5 [English]
© BW Technologies 2009. All rights reserved.
User Manual
Limited Warranty and Limitation Liability
BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of
shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase
price, repair or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the
purchase price actually paid by the buyer for the Product.
This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use;
b) any product which in BW’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of operation, handling or use;
c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on the product; or
The obligations set forth in this warranty are conditional on:
a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable recommendations of BW;
b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW until receipt by the buyer of
shipping instructions from BW; and
c) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER ’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR BASED ON CONTRACT, TORT OR RELIANCE
OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of
this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or
enforceability of any other provision.
v
GasAlertMicro 5/PID/IR
User Manual
Title Page
vi
List of Figures
Figure Title Page
vii
GasAlertMicro 5/PID/IR
User Manual
viii
List of Tables
Table Title Page
x
GasAlertMicro 5/PID/IR
ISO 9001
1
GasAlertMicro 5/PID/IR
User Manual
Oxygen (O2)
Safety Information - Read First
% volume
Use the detector only as specified in this user manual, otherwise the
a) percent of lower explosive protection provided by the detector may be impaired.
limit (%LEL)
Combustible gases (LEL) International symbols used on the detector and in this user manual are
b) percent by volume
defined in Table 3..
methane 0-5.0% v/v
Carbon monoxide (CO) parts per million (ppm) Read the Cautions on the following pages before using the detector.
2
GasAlertMicro 5/PID/IR
aCautions
aCautions
• Caution: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand
this user manual completely before operating or servicing.
• Do not use the detector if it is damaged. Inspect the detector before using. Look for cracks and/or missing parts.
• If the detector is damaged or parts are missing, contact BW Technologies by Honeywell immediately.
• Use only sensor(s) that are specifically designed for the GasAlertMicro 5/PID/IR detectors. Refer to Replacement Parts and
Accessories.
• Calibrate the detector before first-time use and then on a regular schedule, depending on use and sensor exposure to poisons
and contaminants. Sensors must be calibrated regularly and at least once every 180 days (6 months).
• BW recommends to bump test the sensors, before each day’s use, to confirm their ability to respond to gas by exposing the
detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible and visual alarms are activated.
Calibrate if the readings are not within the specified limits.
• BW recommends the combustible sensor be checked with a known concentration of calibration gas after any known exposure to
catalyst contaminants/poisons (sulfur compounds, silicon vapors, halogenated compounds, etc).
• The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in the % LEL range,
calibrate the sensor using the appropriate gas.
• Only the combustible gas detection portion of this instrument has been assessed for performance by CSA International.
• Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons. Although certain
organic vapors (such as leaded gasoline and halogenated hydrocarbons) may temporarily inhibit sensor performance, in most
cases, the sensor will recover after calibration.
• Before using common products around sensors, refer to Sensor Poisons and Contaminants.
3
GasAlertMicro 5/PID/IR
User Manual
aCautions
• Any rapid up-scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale
limit, which may be hazardous.
• Calibrate only in a safe area that is free of hazardous gas, in an atmosphere of 20.9% oxygen.
• Use only BW approved batteries for the GasAlertMicro 5/PID/IR detectors. Refer to Replacement Parts and Accessories.
• Charge the detector before first-time use. BW recommends the detector be charged after every workday.
• Charge the battery pack immediately when a low battery alarm occurs.
• Charge the GasAlertMicro 5/PID/IR batteries using the recommended charging adapter only. Do not use any other charging
adapter. Failure to adhere to this caution can lead to fire and/or explosion.
• Read and adhere to all instructions in the charger user manual. Failure to do so can result in fire, electrical shock, personal
injury, and/or property damage.
• Extended exposure of the GasAlertMicro 5/PID/IR detectors to certain concentrations of combustible gases and air may stress a
detector element, which can seriously affect its performance. If an alarm occurs due to high concentration of combustible gases,
recalibration should be performed, or if needed, the sensor replaced.
• When calibrating O3 and ClO2 sensors that are located in the Toxic 2 sensor position, a single gas calibration cap must be used
to ensure accurate calibration. For more information, refer to Single Gas Calibration Cap.
• Replace the CO2 sensor only in a safe area that is free of hazardous gas.
• The optional pump (M5-PUMP) is certified for use with the GasAlertMicro 5/PID/IR detectors only.
4
GasAlertMicro 5/PID/IR
aCautions
aCautions
• Do not test the combustible sensor’s response with a butane cigarette lighter; doing so will damage the sensor.
• Do not expose the detector to electrical shock or severe continuous mechanical shock.
• Do not attempt to disassemble, adjust, or service the detector unless instructions for that procedure are provided in the user
manual and/or that part is listed as a replacement part. Use only BW Technologies by Honeywell Replacement Parts and
Accessories.
• The detector warranty is void if customer, personnel, or third parties damage the detector during repair attempts. Repair
attempts made by non-BW Technologies by Honeywell repair/service personnel voids this warranty.
• Do not use any other lithium batteries with the GasAlertMicro 5/PID/IR detectors. Use of any other cell can cause fire and/or
explosion. To order and replace the M5-BAT08 battery, refer to Replacement Parts and Accessories.
• Warning: Lithium polymer cells exposed to heat at 266°F (130°C) for 10 minutes can cause fire and/or explosion.
• Dispose of used lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste
stream. Spent batteries must be disposed of by a qualified recycler or hazardous materials handler.
5
GasAlertMicro 5/PID/IR
User Manual
Sensor Poisons and Contaminants The following table lists common products to avoid using around
sensors.
Several cleaners, solvents, and lubricants can contaminate and cause
permanent damage to sensors. Before using cleaners, solvents, and Table 2. Sensor Poisons and Contaminants
lubricants in close proximity to the detector sensors, read and adhere
to the following caution and table. Cleaners and
Silicones Aerosols
Lubricants
a Caution Brake cleaners Silicone cleaners Bug repellents
Use only the following BW Technologies by Honeywell and protectants and sprays
recommended products and procedures: Lubricants Silicone based Lubricants
adhesives, seal-
• Use water based cleaners.
ants, and gels
• Use non-alcohol based cleaners.
Rust inhibitors Hand/body and Rust inhibitors
• Clean the exterior with a soft, damp cloth. medicinal creams
• Do not use soaps, polishes, or solvents. that contain silicone
Window and glass Tissues containing Window and
cleaners silicone glass cleaners
Dishsoaps Mold releasing
agents
Citrus based cleaners Polishes
Alcohol based
cleaners
Hand sanitizers
Anionic detergents
Methanol
(fuels and antifreezes)
6
GasAlertMicro 5/PID/IR
Sensor Poisons and Contaminants
Symbol Description
International Electrotechnical Commission Scheme for Certification to Standards for Electrical Equipment for
IECEx Explosive Atmospheres
7
GasAlertMicro 5/PID/IR
User Manual
Getting Started The detector is shipped with the sensors, and battery packs installed.
To replace sensors, the pump, or the battery pack, refer to Replacement
The list below provides the standard items included with the detector.
Parts and Accessories.
If the detector is damaged or parts are missing, contact the place of
purchase immediately. To become oriented with the features and functions of the detector, refer
to the following figures and tables:
• Batteries: Three replaceable alkaline cells with battery pack, or
one rechargeable lithium battery pack • Figure 1. and Table 4. describe the detector’s components.
• Sensors: O2, combustible (LEL), toxic, H2S/CO (TwinTox sensor), • Figure 2. and Table 5. describe the detector’s screen elements.
PID, or CO2 • Table 6. describes the detector’s pushbutton.
• Calibration cap and hose
• Single gas calibration cap
• Screwdriver
• Quick reference guide
• Quick reference card
• User manual and training on CD-ROM
• Fleet Manager II (if applicable)
To order parts, refer to Replacement Parts and Accessories.
8
GasAlertMicro 5/PID/IR
Parts of the GasAlertMicro 5/PID/IR
Item Description
9
GasAlertMicro 5/PID/IR
User Manual
Item Description
1 Alarm condition
2 Automatically span sensor
3 Gas cylinder
4 Gas type
5 Battery life indicator
6 Passcode lock
7 Data transmission
8 Clock
9 Stealth mode
10 Pump indicator (optional)
11 MMC indicator (optional)
Alarm condition (low, high, TWA,
STEL, or multi alarm) or view TWA,
12
STEL and peak (MAX) gas expo-
Figure 2. Screen Elements sures
Note 13 Automatically zero sensor
If enabled, the backlight automatically activates for 8 seconds
when there is an alarm condition and whenever there is
insufficient light to view the LCD. Any pushbutton reactivates
the backlight in low light conditions.
10
GasAlertMicro 5/PID/IR
Pushbuttons
Pushbuttons
Table 6. Pushbutton
Pushbutton Description
• To enter the user options menu, press and hold G and H simultaneously until the countdown is complete.
G • To clear the TWA, STEL, and peak (MAX) gas exposure readings, press and hold C and G simultaneously
until the countdown is complete.
• To view the date and time, alarm setpoints (TWA, STEL, low, high) of all sensors, and the LEL/PID correction factor
(if applicable) press G.
• To view the TWA, STEL, and peak (MAX) gas exposure readings, press C.
C
• To acknowledge latched alarms, press C.
11
GasAlertMicro 5/PID/IR
User Manual
Note
Replace the batteries and reactivate the detector.
Only activate the detector in a safe atmosphere that is free of
1. All of the LCD elements display simultaneously as the detec-
hazardous gas in an atmosphere of 20.9% oxygen.
tor beeps, flashes, vibrates, and briefly activates the back-
To activate the detector, press A. light.
Self-Tests
When the detector is activated, it performs several self-tests. Confirm
the following tests occur.
Note
If an error message displays during the self-test, refer to
Troubleshooting.
12
GasAlertMicro 5/PID/IR
Activating the Detector
Datalogging (Optional)
4. If the detector is a datalogging unit, it determines if
13
GasAlertMicro 5/PID/IR
User Manual
Self-test Successful: If successful, the following screen displays. 7. The TWA, STEL, low, and high alarm setpoints then
display in the following order (left to right).
TWA STEL
6. If correction factors are set in the user options, the LEL or PID Note
(custom) correction factors display. The alarm setpoints may vary by region. Refer to Resetting
Gas Alarm Setpoints.
14
GasAlertMicro 5/PID/IR
Activating the Detector
Pump Test If the pump inlet is not blocked within 10 seconds or the pump
test fails, the following screens display.
8. If the pump module is attached to the detector, the following
screens display.
15
GasAlertMicro 5/PID/IR
User Manual
9. Unless disabled in user options, the oxygen (O2) sensor is cali- If any sensor is over due for calibration, the LCD displays the
brated automatically. name of the sensor and the number of days past due.
Note
If any sensor is overdue, Y displays continually until calibration
is performed.
16
GasAlertMicro 5/PID/IR
Activating the Detector
If C is not pressed to start calibration, the following screen dis- Apply the LEL gas and then apply the O2 (a higher or lower
plays and the detector deactivates. percentage than the default 20.9%, such as 18% O2).
17
GasAlertMicro 5/PID/IR
User Manual
Successful Bump Test: If the bump test passes, the following screens If additional sensors require a bump test but are not mandatory, the
display. following screens display.
The detector waits for the sensor(s) to clear (30 seconds) and then Press C Yes to accept and proceed to normal operation.
enters normal operation. Or
Unsuccessful Bump Test: If the bump test is unsuccessful or the bump If A No is pressed, or no buttons are pressed, the sensor(s) that is
test is not performed, the following screen displays and the detector past due displays as FAIL when the detector enters normal operation.
deactivates. In the following screen example, only the SO2, CO, and O2 sensors are
operational.
18
GasAlertMicro 5/PID/IR
Activating the Detector
19
GasAlertMicro 5/PID/IR
User Manual
Note
Tech mode is not visible in the user options menu. To access
this option, refer to Tech Mode.
20
GasAlertMicro 5/PID/IR
User Options Menu
1. To enter the user options menu, press and hold G and H Note
simultaneously as the detector beeps and flashes to the cor- If no pushbuttons are pressed within 20 seconds, the detector
responding countdown. returns to normal operation.
Options Menu
Each feature within the Options menu is enabled/disabled by pressing
C to toggle between the checkboxes.
Enabled
2. To scroll through the options, press H or G. When the cursor Disabled
displays beside the desired option, press C.
3. To return to the previous menu, scroll to Back and press C or
press A.
21
GasAlertMicro 5/PID/IR
User Manual
Backlight Due-Lock
The Backlght (backlight) option enables If the Due-lock (calibration user lockout)
the LCD backlight to activate automatically option is enabled and a sensor is overdue
in low-light conditions. for calibration upon startup, the passcode
If disabled, the backlight activates only must be entered to access normal opera-
when the detector is in alarm mode. tion.
Note Note
The detector is shipped with the Backlght The detector is shipped with the Due-lock
option enabled. option disabled.
Confidence Beep
The Confibeep (confidence beep) option
provides continuous confirmation that the Latched Alarms
detector is operating properly. When confi- If enabled, the Latch (latched alarms)
dence beep is enabled, the audible alarm option causes the low and high gas alarms
beeps once every 10 seconds. (audible, visual, and vibrator) to persist
Note until they are acknowledged. Press C to
acknowledge the alarm.
Confidence beep is automatically
disabled during a low battery alarm. After the alarm is acknowledged, it reacti-
vates every 30 seconds until the gas
The detector is shipped with the concentration is below the setpoint.
Confibeep option disabled.
The detector is shipped with the Latch
option disabled.
22
GasAlertMicro 5/PID/IR
User Options Menu
23
GasAlertMicro 5/PID/IR
User Manual
Sensor Configuration 1. From the option menu screen, scroll to Sensors and press
C to access the following screen.
The Sensor option provides access to additional options and functions
for each sensor.
Depending upon the sensor that is selected, some or all of the following
options are available for configuration:
• Enabling/disabling a sensor
• Setting the span gas value
• Adjusting the STEL period (not applicable to LEL and O2 sensors)
• Selecting the TWA method (not applicable to LEL and O2 sensors) 2. Press G or H to scroll to the desired sensor. Press C to con-
firm and to access the menu options specific to the selected
• Resolution setting (not applicable to CO, LEL, O2, and CO2
sensor.
sensors)
For all sensor options, if a value is changed but not
• % vol CO2 (CO2 sensor only)
confirmed within 5 seconds, the detector emits an
• % vol CH4 (LEL sensor only) audible alarm and displays the following error message.
• Selecting the correction factor (LEL and PID sensors only)
• Automatic calibration (O2 sensor only)
24
GasAlertMicro 5/PID/IR
User Options Menu
Sensor Enable/Disable If disabled, the readings and the gas type for the sensor do not display
a Warning when in normal operation.
Disabling a sensor should only be performed with If a sensor is enabled but it is not installed in the detector, FAIL flashes
extreme caution. The disabled sensor cannot measure above the gas type of the missing sensor.
and alarm against the applicable gas.
Disabled Enabled/not installed
If a sensor fails, disabling the sensor deactivates the fail alarm. The
sensor should be replaced and enabled as soon as possible. The
detector will function normally with the remaining enabled sensors.
Note
Detectors that are configured for a 1, 2, 3, or 4 gases may con-
tain a dummy sensor in one of the four sensor locations.
After selecting the desired sensor, the following screen displays. If all the sensors are disabled, the following screens display.
Press C to toggle between enable/disable (sensor can be enabled at Enable one or more sensors to exit and access normal operation.
any time).
Enabled
Disabled
25
GasAlertMicro 5/PID/IR
User Manual
26
GasAlertMicro 5/PID/IR
User Options Menu
A check displays in the checkbox of the After selecting the desired sensor, press H to scroll to Resolution.
currently selected method. To select the Press C within 20 seconds to confirm.
other method, press H to move the check A check displays in the checkbox of the
to the other method. Press C to confirm the currently selected resolution. To select the
selection. other resolution, press H to move the
The detector is shipped with the OSHA check to other resolution. Press C to
method enabled. confirm the selection.
27
GasAlertMicro 5/PID/IR
User Manual
%Vol CO2 (CO2 Sensors Only) %Vol CH4 (LEL Sensors Only)
If the %vol CO2 is enabled, the detector If %vol CH4 is enabled, any currently
displays the carbon dioxide (CO2) enabled correction factor is ignored and
readings as %vol (0.0). the detector operates assuming a
methane (CH4) calibration.
From the Sensors option menu, select
CO2. From the Sensors menu, select LEL and
press C to confirm.
Next, press H to scroll to %vol CH4 and
Press H to scroll to %vol CO2. Press C
press C within 20 seconds to confirm.
to toggle between enable and disable.
Press C to toggle between enable and
Confirmation is not required. If no buttons
disable.
are pushed, after 20 seconds the detector
returns to the sensor selection screen. Confirmation is not required. If no buttons
The change is saved automatically. are pushed within 20 seconds, the
detector returns to the sensor selection
The detector is shipped with %vol CO2
screen. The change is saved
disabled.
automatically.
Enable Enable
Disable The detector is shipped with %vol CH4 Disable
Note disabled.
Correction factors are not applicable to the CO2 sensor. Correction Factor (CF)
Depending upon the selected sensor, refer to the following sections
LEL Sensor or PID Sensor for more information.
Note
Corrections factors are not applicable to CO2 sensors.
28
GasAlertMicro 5/PID/IR
User Options Menu
29
GasAlertMicro 5/PID/IR
User Manual
30
GasAlertMicro 5/PID/IR
User Options Menu
If C is not pressed within 5 seconds, the following screen displays. The screen displays showing the month highlighted indicating it is
selected to set.
The detector is shipped with the datalog sample time set to 5 seconds.
Press G or H to scroll to the desired month and press C within
Clock Option 20 seconds to confirm. To bypass and retain the current setting, press
C. Continue setting/bypassing the remaining options.
The Clock option is used to change the date and time.
From the user options menu, press H to scroll to Clock. Press C The date/time options are set in the following order:
within 20 seconds to confirm.
• month
• day
• year
• hour
• minutes
When the settings have been set or bypassed, the detector beeps twice
and returns to the user options menu.
Note
The detector is shipped with the date and time set to Mountain
Standard Time (MST).
31
GasAlertMicro 5/PID/IR
User Manual
Language Selection Wait for 20 seconds until the detector returns to the user options menu,
or press G to scroll to Back (English), Retour (French), Zurück
The detector is shipped with English selected as the default language.
(German), Regreso (Spanish), or Retornar (Portuguese).
The available languages are as follows:
All screens now display in the selected language.
• Français (French)
• Deutsch (German) Tech Mode
• Español (Spanish)
a Warning
• Prtuguês (Portuguese)
Tech mode should only be accessed by authorized
Press H to scroll to Language and press C within 20 seconds to personnel.
confirm. Tech mode can only be accessed from the Language option. Press H
to scroll to Language. Do not press C until instructed.
32
GasAlertMicro 5/PID/IR
User Options Menu
In the following order, press and continue to hold each button until Sensors
Tech mode displays below the Language option. a Caution
1. Press and hold H with the right index finger. To reconfigure the sensor type, physically change the
2. Press and hold G with the right middle finger. sensor prior to entering Tech mode.
3. Press and hold C with the left thumb. When a toxic sensor is physically removed and replaced by another toxic
sensor, the detector must be reconfigured to recognize the change.
Note
If a sensor is replaced, the detector will classify the sensor as
overdue for calibration. Calibrate the new sensor immediately.
• Sensors
• Initialize 2. Press G or H to scroll to Toxic 1
• Force Calibration (Force calibration) or Toxic 2 and press C within
20 seconds to confirm.
• Bump Daily (Bump test daily)
• Stealth Mode
• Sleep Mode
33
GasAlertMicro 5/PID/IR
User Manual
34
GasAlertMicro 5/PID/IR
User Options Menu
35
GasAlertMicro 5/PID/IR
User Manual
Note
Enabling the Sleep option may
require the battery be charged more
frequently.
The detector is shipped with the Sleep
option disabled.
36
GasAlertMicro 5/PID/IR
Alarms
Alarms
The following table describes the detector alarms and corresponding screens.
• During an alarm condition, the detector activates the backlight and displays the current ambient gas reading.
• If more than one type or level of alarm exists simultaneously, a multi alarm results.
• To change the factory-set alarm setpoints, refer to Calibration and Setting Alarm Setpoints.
Table 7. Alarms
37
GasAlertMicro 5/PID/IR
User Manual
Table 7. Alarms
• flashes
38
GasAlertMicro 5/PID/IR
Alarms
Table 7. Alarms
• One beep, one flash, and one • One beep every 5 seconds
vibrate every 10 seconds
• S flashes
Alarms Screen
Pump Alarm
• L and J flash
Note
If the Latch (latched alarms) option is activated, the audible, visual, and vibrator alarms persist until the alarm condition is acknowledged.
To acknowledge a latched alarm, press C. The alarms cannot be deactivated if an alarm condition exists.
If the Stealth option is enabled, the detector only vibrates during an alarm; the audible and visual alarms are disabled.
a Caution: If the detector emits a low battery alarm, charge the lithium battery pack immediately.
39
GasAlertMicro 5/PID/IR
User Manual
40
GasAlertMicro 5/PID/IR
Alarms
41
GasAlertMicro 5/PID/IR
User Manual
The time/date, TWA, STEL, low, and high alarm setpoint screens display Resetting Gas Alarm Setpoints
in the following order left to right:.
Note
Standard factory alarm setpoints may vary by region.
The following table lists the factory alarm setpoints according to the
Occupational Safety and Health Association (OSHA) settings.
Table 10. OSHA Sample Factory Alarm Setpoints
42
GasAlertMicro 5/PID/IR
Alarms
To change the factory-set alarm setpoints, refer to Calibration and Set- Sensor Alarm
ting Alarm Setpoints.
The detector tests for missing or defective sensors during the activation
Note self-test. If a sensor fails the self-test, FAIL flashes on the LCD above
To disable an alarm, set the alarm setpoint to 0 (zero). the failed sensor. Refer to Troubleshooting.
43
GasAlertMicro 5/PID/IR
User Manual
Refer to Pump Test for more information. If the pump test is successful, Calibration and Setting Alarm Setpoints
the detector returns to normal operation, otherwise the pump alarm con-
tinues. If the pump alarm persists, refer to the Pump Operation section in Guidelines
Troubleshooting.
When calibrating the detector, adhere to the following guidelines:
44
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
• Calibrate a new sensor before use. Install the sensor, activate Diagnostics Testing
the detector, and allow the sensor to stabilize before starting
calibration. The detector tests the ambient air (auto zero) and the test gas that is
Used sensor: 60 seconds / New sensor: 5 minutes applied (auto span) to ensure it meets expected values. Auto zero sets
the zero-gas level of the sensor. If ambient gas is present, the zero
• Calibrate the detector at least once every 180 days (every 90 days level will be incorrect.
for HCN sensors) depending upon use and sensor exposure to
poisons and contaminants. For a list of common products that can If excessive target gas is present, the detector displays an error
damage sensors, refer to Sensor Poisons and Contaminants. message and lists the affected sensor.
• Calibrate the detector if the ambient gas varies during startup.
• Calibrate the sensor before changing the alarm setpoints.
• Calibrate only in a safe area that is free of hazardous gas in an
atmosphere of 20.9% oxygen.
• To disable an alarm, set the alarm setpoint to 0 (zero).
• If the Auto cal option is enabled, the oxygen (O2) sensor calibrates
automatically every time the detector is activated. Activate the
detector in a safe area that is free of hazardous gas in an In auto span, if the target gas is not detected or does not meet expected
atmosphere of 20.9% oxygen. values, a message displays that the detector is exiting calibration mode.
The detector retains the previous set values.
• After activating the detector, allow it to stabilize for 1 minute before
performing a calibration or bump test.
• If a certified calibration is required, contact BW Technologies by
Honeywell.
Note
A generator must be used for O3 and ClO2 sensors.
45
GasAlertMicro 5/PID/IR
User Manual
Applying Gas to the Sensors Table 11. Applying Gas to the Sensors
The calibration cap, single gas calibration cap, and hose are shipped Item Description
with the detector. Refer to Figure 3. and Table 11. for installation.
1 Detector with calibration cap
Note
2 Calibration hose
The calibration cap and single gas calibration cap must only be 3 Regulator and gas cylinder
used during the calibration span process.
4 Toxic 2 sensor position
5 Single gas calibration cap
46
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
Table 12. Single Gas Calibration Cap Removing the Single Gas Calibration Cap
Using the thumb, push forward against both the inlet and the outlet
Item Description
simultaneously to remove the cap from the detector.
1 Intake inlet
2 Calibration hose
3 Gas flow direction arrow
4 Output outlet
Calibration Procedure
Figure 4. Single Gas Calibration Cap To calibrate the detector and set the alarm setpoints, perform the
following procedure.
Note
Note
The arrow on the cap indicates the direction of gas flow from
intake to outtake. To bypass a step during the calibration process (after auto
zero), press A.
47
GasAlertMicro 5/PID/IR
User Manual
Start Calibration
GasAlertMicro 5/PID GasAlertMicro 5 IR
Note
Verify that the calibration gas being used matches the span
concentration value(s) that are set for the detector. Refer to
Span Gas Value. Correction factors are not applied during cali-
bration. Correction factors that were set prior to calibration are
restored when the detector returns to normal operation.
a Caution
Calibrate only in a safe area that is free of hazardous gas Note
in an atmosphere of 20.9% oxygen. Do not apply calibration gas during this process, otherwise the
1. To enter calibration, press and hold C and H simultane- auto zero step will fail.
ously as the detector beeps, flashes, and vibrates to the
Zero CO2 (GasAlertMicro 5 IR only)
corresponding countdown.
If calibrating using the GasAlertMicro 5 or the GasAlertMicro PID, pro-
ceed to Auto Span step #5. If the detector is passcode protected, pro-
ceed to Passcode Protect step #4.
3. For GasAlertMicro 5 IR models, there is a second auto zero
that is performed for the CO2 sensor.
When auto zero for the other sensors is complete, the following
screen displays.
48
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
Press C Yes to zero the CO2 sensor. flashes whiles the detector zeros the CO2
Or sensor (approximately 30 seconds).
Press A No to bypass the CO2 zero and proceed to Auto Span Passcode Protect Activated
step #5. If the detector is passcode protected, refer to Pass- 4. When auto zero is complete and if the passcode protect option
code Protect Activated step #4. is enabled, the following screen displays.
If C Yes is pressed, the following screen displays.
49
GasAlertMicro 5/PID/IR
User Manual
GasAlertMicro 5 and PID To ensure accurate calibration, a single gas calibration cap
must be used to calibrate O3 and ClO2 sensors.
50
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
K flashes as the detector initially detects the calibration gas. While the detector is spanning the sensor(s), a countdown of
GasAlertMicro 5 and PID GasAlertMicro 5 IR time remaining displays in the lower left of the screen.
GasAlertMicro 5 and PID GasAlertMicro 5 IR
LEL (combustibles) 30 seconds Successful Span: Proceed to Successful Span step #6.
Unsuccessful Span: Refer to Unsuccessful Span for possi-
PID gases 2 minutes
ble solutions.
Insufficient Level: If a sensor does not attain a sufficient level
of expected gas, it is cleared from the LCD and is not spanned.
51
GasAlertMicro 5/PID/IR
User Manual
Note
If a new due date is not entered, the detector automatically
resets to the previously entered number of days.
If a sensor fails to span, refer to Unsuccessful Span at the end of the
calibration section for possible solutions.
52
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
The calibration due dates are set in the following order: Note
Or
Press A to bypass a sensor and proceed to the next sensor.
53
GasAlertMicro 5/PID/IR
User Manual
Alarm Setpoints If a new setpoint is entered but not confirmed within 5 seconds
by pressing C, the following screen displays.
8. When all of the sensor due dates have been set or bypassed,
the alarm setpoints need to be set or bypassed. The following
screen displays.
54
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
Setting the TWA Alarm Setpoint Setting the Low Alarm Setpoint
The current TWA alarm setpoint displays for the selected sensor The current low alarm setpoint displays for the selected sensor.
(if applicable).
Press H or G to change the value for the low alarm setpoint. When the
Press H or G to change the value for the TWA alarm setpoint. When required value displays, press C to confirm.
the required value displays, press C to confirm. Setting the High Alarm Setpoint
Setting the STEL Alarm Setpoint The current high alarm setpoint displays for the selected sensor.
The current STEL alarm setpoint displays for the selected sensor
(if applicable).
Press H or G to change the value for the high alarm setpoint. When the
required value displays, press C to confirm.
Press H or G to change the value for the STEL alarm setpoint. When Proceed to step #9 Setting the Remaining Alarm Setpoints.
the required value displays, press C to confirm.
55
GasAlertMicro 5/PID/IR
User Manual
56
GasAlertMicro 5/PID/IR
Calibration and Setting Alarm Setpoints
Unsuccessful Span If all sensors fail the span, the following screen displays.
If the sensor(s) did not span successfully, refer to the following sections
for possible solutions:
• Failed Span
• No Gas Detected
• Large Span
1. Ensure the regulator valve is open and that the gas cylinder
Failed Span is not past the expiration date. Replace the gas cylinder and/
or regulator if required.
If a sensor fails the span, the following error message displays.
2. Replace the sensor(s) and sensor filter if required.
3. Press A to exit and then calibrate again in a safe area that
is free of hazardous gas in an atmosphere of 20.9% oxygen.
4. If the span fails a second time, reactivate the detector to test
the sensors.
If all sensors fail the span, the calibration due dates cannot be reset. The
following screens display.
If the sensor is not calibrated, it displays as FAIL in normal operation the
next time the detector is activated.
57
GasAlertMicro 5/PID/IR
User Manual
No Gas Detected Not reaching the target span can result from
If the detector does not detect any gas within 30 seconds, the following • a problem with the span gas,
screens display. • the gas cylinder being past the expiry date, or
• a problem with the sensor.
Accept Current Span: If the gas cylinder, regulator, and sensor
are operating correctly, press C to accept the current span.
Reject Current Span: Press A to reject. Verify the gas cylinder,
detector settings, and sensor are operating correctly. Calibrate the
sensor again.
Large Span
1. Ensure the regulator valve is open and the calibration hose is
If the span adjustment is unusually large (more than 15%), the following
free of debris.
screens display.
2. Replace the gas cylinder and/or regulator if required.
3. Replace the sensor(s) and sensor filter if required.
4. Press G to reapply gas or press H to end the span.
Did Not Reach Target Span
If the target span was not achieved, as defined in the Span Gas Value
option, the following screens display.
Ensure the correct gas cylinder being used and that the span concentra-
tion value(s) of the detector matches the span concentration value(s) of
the gas cylinder. Refer to Span Gas Value.
Adjustment Expected: If the calibration adjustment is expected,
press C to accept the span.
Adjustment Not Expected: If the calibration adjustment is not
expected or the span gas value does not match the gas cylinder,
press A to reject the span and calibrate that sensor again.
58
GasAlertMicro 5/PID/IR
Attaching the Accessories
Item Description
Installing the Pump Module
1 Pump module
The pump module can be used independently to measure gas or it can
2 Sensor filter
be used with the sample probe to test for gases in confined spaces. To
3 Detector
measure gas using the sample probe, refer to Attaching the Sample
Probe. 4 Machine screws (2)
To install the pump module, refer to Figure 6., Table 14., and the
following procedures.
59
GasAlertMicro 5/PID/IR
User Manual
To replace the pump filter, refer to Figure 7. and the following steps 1-8. 2. Remove the filter window screw.
If using the auxiliary filter on an older model pump module or 4. Remove the old filter from the filter window.
as an additional filtration system on the new pump module, Important! Ensure the filter cavity and filter window are clean
refer to Attaching the Auxiliary Filter. and free of debris.
5. Insert a new filter.
6. Replace the filter window and screw.
7. Activate the detector. The detector performs a pump test dur-
ing the startup self-tests to verify the pump module is operating
correctly.
8. To order additional filters, refer to Replacement Parts and
Accessories.
a Caution
Filters may need to be replaced more frequently in high
particulate areas. BW Technologies by Honeywell
recommends that the auxiliary filter be used as a
additional filtration in high particulate areas.
60
GasAlertMicro 5/PID/IR
Attaching the Accessories
Replacing the Pump Nozzle 3. Insert the new nozzle. Ensure the nozzle post inserts correctly
into the nozzle gasket.
To replace the pump nozzle, refer to Figure 8. and step 1-4.
4. Activate the detector. The detector performs a pump test dur-
ing the startup self-tests to verify the pump module is operating
correctly.
61
GasAlertMicro 5/PID/IR
User Manual
a Caution Note
The filter is designed to protect the pump. For older Attaching the filter cord to the alligator clip ensures the filter
model pumps, the filter must be connected when the remains with the detector when not in use.
pump is activated. Only remove the filter to perform
2. Pull the filter cord tight. Ensure the loop knot is secure.
calibrations, bump tests, and when using the sample
probe. 3. Align the pump quick connector with the filter quick connector
For new model pumps, the auxiliary filter can be used as and turn clockwise. Ensure the connection is secure.
an additional filtration system in high particulate areas. 4. Change the filter as required. Environments with heavy air-
To insert the auxiliary filter, complete the following: borne particulates require the filter to be changed
frequently.
1. Loop the filter cord through the ring on the alligator clip and
pull the filter through the loop of the filter cord. To order additional filters, refer to Replacement Parts and
Accessories.
62
GasAlertMicro 5/PID/IR
Attaching the Accessories
a Warning
Depending upon the length of the tubing and the type
of gas in the confined space, allow a minimum of
3 seconds per ft. of tubing to ensure the readings
stabilize before entering the area.
Example: 50 ft. = 2.5 minutes
63
GasAlertMicro 5/PID/IR
User Manual
64
GasAlertMicro 5/PID/IR
MMC/SD Card Troubleshooting
5. Activate the detector. The MMC/SD card is automatically for- MMC/SD Card Troubleshooting
matted during the startup self-test.
The MMC/SD card is not required for operation in detectors equipped
with datalogging. However, the following two screens display if the card
is not inserted during startup.
65
GasAlertMicro 5/PID/IR
User Manual
Restoring Datalog Files If the detector successfully restores the logfile, the
following screen displays and the startup tests continue.
If the MMC/SD card has been accidentally reformatted or erased by the
computer application, the following screens display when the card is
inserted into the detector.
2. From the computer, verify that the logfile has been restored.
When the normal operation screen displays, deactivate the
detector.
a Caution
3. Remove the MMC/SD card and insert it into the card reader.
Only erased data files can be restored using the detector.
Computer applications sometimes write data over erased 4. From the computer desktop, double-click My Computer to
files, and that erased data cannot be restored by the detector. view the list of drives.
Always create back up files on the computer.
5. Double-click the Removable Disk drive to access
To restore the logfile, complete the following: LOGFILE0.csv. Open the logfile and verify that the data has
1. From the detector, press H to restore the logfile. The been restored.
following screen displays. If LOGFILE0.csv does not display, ensure that the MMC/SD
card is inserted in the card reader correctly and that all connec-
tions are secure.
6. After verifying that the logfile has been restored, re-insert the
MMC/SD card into the detector.
66
GasAlertMicro 5/PID/IR
MMC/SD Card Troubleshooting
Reformatting the MMC/SD Card 7. Press G to format the MMC/SD card. The following screen dis-
plays.
To reformat the MMC/SD card, complete the following:
1. Insert the MMC/SD card into the card reader.
2. From the computer desktop, double-click My Computer to
view the list of drives.
3. Double-click the Removable Disk drive to access
LOGFILE0.csv.
4. Select LOGFILE0.csv and delete.
5. Insert the MMC/SD card into the detector. For any additional MMC/SD card errors, refer to
6. Activate the detector. The startup self-test begins and the fol- Troubleshooting.
lowing screens display.
67
GasAlertMicro 5/PID/IR
User Manual
Import Datalogs to Fleet Manager II Import to Fleet Manager II Using a Card Reader
Note To import a datalog file from the detector to Fleet Manager II, complete
Refer to the following minimum requirements before importing the following:
datalogs to Fleet Manager II. 1. Deactivate the detector.
2. Release the latch and remove the battery pack.
Minimum PC Requirements
3. Remove the MMC/SD card from the detector. Refer to Figure
• 500 MHz Pentium (or equivalent) 12. and 1.
• 100 MB free hard disk space 4. Connect the card reader to the USB port on the computer.
• Windows XP or Vista
5. Insert the MMC/SD card into the card reader (ensure the pins
• USB port face down).
6. Open Fleet Manager II. Click Start > All Programs > BW
Importing from MicroDock II to Fleet Manager II Technologies > Fleet Manager II > BW Technologies Fleet
If the detector is used with the MicroDock II Station to import datalogs to Manager II.
Fleet Manager II, refer to the MicroDock II User Manual for complete 7. From the top task bar, click Help > User Manual.
instructions.
8. Refer to section “Importing Data into Fleet Manager II”.
68
GasAlertMicro 5/PID/IR
View Datalog Files in Spreadsheets
69
GasAlertMicro 5/PID/IR
User Manual
Date Day Time Toxic1 Toxic2 Toxic3 LEL O2 Toxic 1 Toxic 1 Status Serial Unit
dd-mm-yy Mon=1 hh:mm:ss ppm ppm ppm %CH4 % TWA ppm STEL ppm Codes Number Config
%LEL
23-12-05 #4 9:54:25 5 10 35 -- -- 33------ S104-000001
23-12-05 #4 9:54:30 10 15 50 -- -- 44------ S104-000001
23-12-05 #4 9:54:35 5 10 35 10 19.5 -- -- 1111---- S104-000001
23-12-05 #4 9:54:40 10 15 200 20 23.5 -- -- 2222---- S104-000001
23-12-05 #4 9:54:45 0 0 0 24 20.9 -- -- -D-E---D S104-000001 FCEKNL
23-12-05 #4 9:54:50 0 0 0 24 20.9 0 0 -------- S104-000001 FCEKNL
23-12-05 #4 9:54:55 0 0 0 24 20.9 0 0 --L----L S104-000001 FCEKNL
23-12-05 #4 9:55:00 0 0 0 24 20.9 0 0 LLH----M S104-000001 FCEKNL
23-12-05 #4 9:55:05 5 10 35 0 0 LLH----M S104-000001
23-12-05 #4 9:55:10 10 15 50 0 0 LLL----M S104-000001
23-12-05 #4 9:55:15 5 10 35 10 19.5 0 0 -L-----L S104-000001
23-12-05 #4 9:55:20 10 15 200 20 23.5 0 0 -------- S104-000001
23-12-05 #4 9:55:25 0 0 0 24 20.9 0 0 -------B- S104-000001 FCEKNL
23-12-05 #4 9:55:30 0 0 0 24 20.9 0 0 ------B- S104-000001 FCEKNL
70
GasAlertMicro 5/PID/IR
View Datalog Files in Spreadsheets
Status Codes
⎯ Normal operation G Backlight is on
L Low alarm v STEL and high alarm (dual alarms) 1 Alarm setpoint 1 (low alarm)
H High alarm w TWA and STEL alarm (dual alarms) 2 Alarm setpoint 2 (high alarm)
T TWA alarm x TWA, STEL, and low (triple alarms) 3 Alarm setpoint 3 (TWA alarm)
U TWA and low alarm (dual alarms) y TWA, STEL, and high (triple alarms) 4 Alarm setpoint 4 (STEL alarm)
V TWA and high alarm (dual alarms) O Overload / sensor is over-ranged D Calibration due date (days)
s STEL alarm C Calibrating E Elapsed / last calibration (days)
u STEL and low alarm (dual alarms) F Failure - sensor failure Z Auto zeroing
f Fresh air delay I Time set t Testing
@ Zero CO2
Pump Codes
P Plugged (blocked) - pump alarm F Failure / pump failure
Battery Status Codes
⎯ Batteries OK B Low battery alarm C Confidence beep is active
Alarm Status Codes
L Low alarm M Multi alarm S Automatic shutdown
H High alarm C Calibration F Failure / self-test fail
T TWA alarm Q Off / quit / manual deactivation R RTCC / real-time clock failure
71
GasAlertMicro 5/PID/IR
User Manual
72
GasAlertMicro 5/PID/IR
Maintenance
• Replace the batteries immediately when the detector emits a • Dispose of used lithium cells immediately. Do not disassemble
low battery alarm. and do not dispose of in fire. Do not mix with the solid waste
stream. Spent batteries must be disposed of by a qualified
• Use only batteries that are recommended by BW Technologies recycler or hazardous materials handler.
by Honeywell. Refer to Specifications.
• Keep lithium cells away from children.
• Ensure the alkaline batteries are properly installed in the
detector battery pack. Charging the Battery
• Charge batteries using only a recommended BW charger. Do To charge the battery, refer to the GasAlertMicro 5/PID/IR Charger
not use any other charger. Failure to adhere to this caution Instruction Sheet.
can cause fire and/or explosion.
73
GasAlertMicro 5/PID/IR
User Manual
Replacing the Alkaline Batteries Table 18. Replacing the Alkaline Batteries
To replace the alkaline batteries, refer to Figure 13., Table 18., and the Item Description
following steps 1-6.
1 Detector
2 Latch
3 Battery pack
4 Battery tray
5 Captive screws (2)
6 Alkaline batteries (3)
7 Battery shell
74
GasAlertMicro 5/PID/IR
Maintenance
Replacing the Lithium Battery Pack 1. Open the latch on the bottom of the detector.
To replace the lithium battery pack, refer to Figure 14. and the following 2. Remove the battery pack by lifting the bottom of the pack
steps 1-4. upward from the detector.
a Warning
Do not disassemble lithium battery packs. Read and
adhere to the cautions in Battery Cautions.
75
GasAlertMicro 5/PID/IR
User Manual
To replace a sensor or sensor filter, refer to Figure 15., Table 19., and Table 19. Replacing a Sensor or Sensor Filter
the following procedures 1-7.
Item Description
1 Sensor cover
2 Sensor filter
3 Sensors
4 Detector
5 Machine screws (2)
76
GasAlertMicro 5/PID/IR
Maintenance
a Caution
To ensure proper maintenance and continued accurate
readings from the sensor, use only the PID Lamp
Cleaning Kit that is supplied by BW Technologies by
Honeywell.
Item Description
1 PID sensor
2 Diffusion barrier
3 Sensor cover
4 Electrode stack
5 Lamp
77
GasAlertMicro 5/PID/IR
User Manual
78
GasAlertMicro 5/PID/IR
Troubleshooting
Troubleshooting
If a problem occurs, refer to the solutions provided in Table 21.. If the problem persists, contact BW Technologies by Honeywell.
Table 21. Troubleshooting
Startup Troubleshooting
The detector does not No batteries Refer to Replacing the Alkaline Batteries or Replacing the Lithium Battery
activate. Pack.
Depleted batteries Refer to Replacing the Alkaline Batteries, or Replacing the Lithium Battery
Pack.
Damaged or defective detector Contact BW Technologies by Honeywell.
The detector immediately Sensor needs to stabilize Used sensor: wait 60 seconds.
enters alarm mode when New sensor: wait 5 minutes.
activated. Low battery alarm Refer to Battery Cautions, Replacing the Alkaline Batteries, Replacing the
Lithium Battery Pack.
Sensor alarm Refer to Replacing a Sensor or Sensor Filter.
Pump alarm If the sampling hose is attached, determine if it is obstructed. If not, clean
or replace the pump filter. If the pump alarm persists, refer to the “Pump
Operation” section in Troubleshooting.
The startup self-test fails. General fault Ensure that the sensors and battery pack are installed correctly and then
reactivate the detector.
If the fault persists, record the error message and contact BW Technologies
by Honeywell.
79
GasAlertMicro 5/PID/IR
User Manual
80
GasAlertMicro 5/PID/IR
Troubleshooting
The MMC/SD card is not Insert the MMC/SD card. Refer to Inserting the MMC/SD Card.
inserted.
81
GasAlertMicro 5/PID/IR
User Manual
Alarms Troubleshooting
Detector does not enter Alarm setpoint(s) are set Reset alarm setpoints. Refer to Calibration and Setting Alarm Setpoints.
alarm mode. incorrectly.
Alarm setpoint(s) are set to Reset alarm setpoints. Refer to Calibration and Setting Alarm Setpoints.
zero.
Detector requires calibration. Calibrate the sensors. Refer to Calibration and Setting Alarm Setpoints.
Detector intermittently Ambient gas levels are near Detector is operating normally. Use caution in suspect areas. Check MAX
enters alarm without any alarm setpoint or the sensor is gas exposure reading.
apparent reason. exposed to a puff of the target
gas.
Alarms setpoints are set Reset alarm setpoints. Refer to Calibration and Setting Alarm Setpoints.
incorrectly.
Missing or faulty sensor. Refer to Replacing a Sensor or Sensor Filter.
Detector enters alarm with Pump flow rate is set too high. Refer to the last solution in Pump Operation troubleshooting on the
LEL reading displaying as following page.
OL (over limit).
82
GasAlertMicro 5/PID/IR
Troubleshooting
There is an obstruction in the If using tubing that is attached to the sample probe, determine if it is
tubing. obstructed. If not, clean or replace the pump filter. Refer to Replacing the
Pump Filter.
Filter needs to be cleaned or Clean or replace the pump filter. Refer to Replacing the Pump Filter.
replaced.
Pump is drawing too much air For newer model pump modules, the pump may be drawing too much air
over the sensors. over the sensors. Attach the auxiliary filter to reduce the flow. Refer to
Attaching the Auxiliary Filter.
Detector enters alarm with Pump flow rate is set too high. 1. From Tech Mode, scroll to the Pump option and press C.
LEL reading displaying as 2. From the Pump Speed screen, press H to decrease the speed to 150.
over limit (OL).
83
GasAlertMicro 5/PID/IR
User Manual
Clock icon is flashing. The clock has failed. Contact BW Technologies by Honeywell.
The detector displays a General fault. Reactivate the detector. If the same error message displays, reset the clock
clock error message using in the user options menu. Reactivate the detector.
last recorded time. If the error message still displays, contact BW Technologies by Honeywell.
84
GasAlertMicro 5/PID/IR
Replacement Parts and Accessories
S4-W04 Combustible (LEL) sensor M5PID-CLN-K1 Cleaning kit for PID sensor lamp 1
1
(with silicone protection filter) M5-SS Sensor filters (quad) kit of 2 2
Combustible (LEL) sensor Gas Cylinders
S4-W04-UF 1
(no silicone protection filter)
Quad calibration gas, CH4-2.5%, O2-
SR-X10 Oxygen (O2) sensor 1 CG-Q58-4 18.0%, H2S-25 ppm, CO-100 ppm, 1
PS-RM04 Carbon monoxide (CO) sensor 1 bal. N2 (58 l)
PS-RH04S Hydrogen sulfide (H2S) sensor 1 Quad calibration gas, CH4-2.5%, O2-
CG-Q34-4 18.0%, H2S-25 ppm, CO-100 ppm, 1
SR-P04 Phosphine (PH3) sensor 1
bal. N2 (34 l)
PS-RS04 Sulfur dioxide (SO2) sensor 1
Two gas calibration cylinder, 50% LEL
PS-RC10 Chlorine (Cl2) sensor 1 CG-T34 (CH4-2.5%) O2-20.9%, bal. N2 (34 l) 1
SR-A04 Ammonia (NH3) sensor 1 CG2-S-25-58 Calibration gas, SO2 25 ppm (58 l) 1
PS-RD04 Nitrogen dioxide (NO2) sensor 1 CG-BUMP-S25 SO2 bump test gas 1
PS-RZ10 Hydrogen cyanide (HCN) sensor 1 Bump alarm gas aerosol (CH4-2.5%,
CG-BUMP1 1
SR-V04 Chlorine dioxide (ClO2) sensor 1 O2-10%, H2S-40 ppm, CO-200 ppm)
85
GasAlertMicro 5/PID/IR
User Manual
CG2-Z-10-58 Calibration gas, HCN 10 ppm (58 l) 1 MMC128-D 128 MB Delkin SD card 1
86
GasAlertMicro 5/PID/IR
Specifications
Backlight: Automatically activates during an alarm condition and when Battery operating time:
there is insufficient light to view the LCD (if enabled in user options) Toxic, O2, and LEL sensors: Three alkaline cells or one lithium battery
Self-test: Initiated during activation pack at 20°C provides 20 hours operating runtime
Calibration: Automatic zero and automatic span Toxic, O2, LEL, and PID sensors: Three alkaline cells or one lithium
battery pack at 20°C provides 15 hours operating runtime
Oxygen sensor: Automatic span upon activation (enable/disable option)
Toxic, O2, LEL, and CO2 sensors: Three alkaline cells or one lithium
User options: Confidence beep, latching alarms, passcode protection, battery pack at 20°C provides 15 hours operating runtime
enable/disable safe display mode, combustible sensor measurement,
sensor disable, TWA and STEL, language selection, enable/disable Year of manufacture: The detector’s year of manufacture is determined
automatic oxygen calibration, define span concentration values, define from the serial number. The second and third number after the first letter
STEL calculation period, define TWA method, gas measurement resolu- determines the year of manufacture.
tion, enable/disable automatic backlight, adjust clock calendar, and E.g., H309-Y000001 = 2009 year of manufacture
define logging rate (datalogger models only), CO2 sensor measurement Approved batteries for GasAlertMicro 5/PID/IR models:
Alkaline (M5-BAT02) and lithium-ion polymer (M5-BAT08), as per
Approved MMC and SD cards for GasAlertMicro 5 and
standards EN 60079-11, EN 60079-0, UL913, CSA C22.2 No.157
GasAlertMicro 5 PID: 128 MB Delkin SD card and 64 MB Unigen SD
card Alkaline: Temperature Code
Approved MMC and SD cards for GasAlertMicro 5 IR: 128 MB Delkin Duracell MN1500 -20°C ≤ Ta ≤ +50°C T4 (129.9°C)
MMC, 128 MB Delkin SD card, 128 MB Transcend MMC, and 64 MB Energizer E91VP -20°C ≤ Ta ≤ +50°C T3C (135.3°C)
Unigen SD card
Lithium rechargeable:
M5-BAT08 -20°C ≤ Ta ≤ 50°C T4
Battery charger: GasAlertMicro 5/PID/IR battery charger
First-time charge: 6 hours for lithium battery pack
Normal charge: 6 hours for lithium battery pack
Warranty: 2 years including sensors
(1 year for NH3 sensor and PID lamp)
88
GasAlertMicro 5/PID/IR
Specifications
Approvals: This equipment has been tested and found to comply with the limits
GasAlertMicro 5 and GasAlertMicro 5 PID (Zone 0) for a Class B digital device, pursuant to Part 15 of the FCC Rules and
Approved by CSA to both U.S. and Canadian Standards ICES-003 Canadian EMI requirements. These limits are designed to
Standards: CAN/CSA C22.2 No. 157 and C22.2 No.152 provide reasonable protection against harmful interference in a residen-
ANSI/UL – 913 and ANSI/ISA – S12.13 Part 1 tial installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
CSA Class I, Division 1, Group A, B, C, and D
instructions, may cause harmful interference to radio communications.
Class 1, Zone 0, Group IIC
However, there is no guarantee that interference will not occur in a
ATEX CE 0539 g II 1 G Ga Ex ia IIC T4
KEMA 06 ATEX 0206X particular installation. If this equipment does cause harmful interference
EN 60079-0, EN 60079-11, EN 60079-26 to radio or television reception, which can be determined by turning
IECEx Ga Ex ia IIC IECEx CSA 06.0011X the equipment off and on, the user is encouraged to try to correct the
IEC 60079-0, IEC 60079-11, IEC 60079-26 interference by one of more of the following measures:
ABS Type Approved: VA-348169-X • Reorient or relocate the receiving antenna.
GasAlertMicro 5 IR (Zone 1): • Increase the separation between the equipment and receiver.
Approved by CSA to both U.S. and Canadian Standards • Connect the equipment into an outlet on a circuit different from that
Standards: CAN/CSA C22.2 No. 157 and C22.2 No.152 to which the receiver is connected.
ANSI/UL – 913 and ANSI/ISA – S12.13 Part 1
• Consult the dealer or an experienced radio/TV technician for help.
CSA Class I, Division 1, Group A, B, C, and D
Class 1, Zone 1, Group IIC
ATEX CE 0539 g II 2 G Ex d ia IIC
KEMA 06 ATEX 0206X
EN 60079-0, EN 60079-1, EN 60079-11
IECEx Ex d ia IIC IECEx CSA 06.0011X
IEC 60079-0, IEC 60079-1, IEC 60079-11
89
GasAlertMicro 5/PID/IR
User Manual
90
GasAlertMicro 5/PID/IR
PID Correction Factor (CF) Library
91
GasAlertMicro 5/PID/IR
User Manual
92
iERP: 128900
D5615/5 [English]
© BW Technologies 2009. All rights reserved.
User Manual