0% found this document useful (0 votes)
32 views92 pages

Fujifilm SonoSite X-Porte Ultrasound - Service Manual

Uploaded by

Cale
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
32 views92 pages

Fujifilm SonoSite X-Porte Ultrasound - Service Manual

Uploaded by

Cale
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 92

Service Manual

Stand and Desktop Systems

i
ii
Legal Notices

Manufacturer
FUJIFILM SonoSite, Inc.
21919 30th Drive SE
Bothell, WA 98021-3904
USA
T: (888) 482-9449 or (425) 951-1200
F: (425) 951-1201

EC Authorized Representative
FujiFilm SonoSite B.V.
Joop Geesinkweg 140
1114 AB Amsterdam, The Netherlands

Australia Sponsor
FUJIFILM SonoSite Australasia Pty Ltd
114 Old Pittwater RD
Brookvale, New South Wales 2100, Australia

Caution: United States federal law restricts this device to sale by or on the order of a physician.

SonoMB, SonoSite, Steep Needle Profiling, X-Porte, and the SonoSite logo are registered (in some jurisdictions) and unregistered trademarks
owned by FUJIFILM SonoSite, Inc. in various jurisdictions.

DICOM is the registered trademark of the National Electrical Manufacturers Association.

All other trademarks are the property of their respective owners.


Patents: US 8,216,146; US 8,213,467; US 8,147,408; US 8,137,278; US 8,088,071; US 8,066,642; US 8,052,606; US 7,819,807; US 7,804,970;
US 7,740,586; US 7,686,766; US 7,604,596; US 7,591,786; US 7,588,541; US 7,534,211; US 7,449,640; US 7,169,108; US 6,962,566; US
6,648,826; US 6,575,908; US 6,569,101; US 6,471,651; US 6,416,475; US 6,383,139; US 6,364,839; US 6,203,498; US 6,135,961; US
5,893,363 ; US 5,817,024; US 5,782,769; US 5,722,412; AU: 730822; AU: 727381; CA: 2,372,152; CA: 2,371,711; CN 98108973.9; CN:
98106133.8; CN: 97113678.5; DE: 69831698.3; DE: 69830539.6; DE: 69730563.5; DE: 602004027882.3; DE: 602004023816.3; DE:
60034670.6; DE: 60029777.2; EP: 1589878; EP: 1552792; EP: 1180971; EP: 0875203; EP: 0815793; EP 1180970; EP 0881492; ES: 2229318;
ES: 159878; ES: 1552792; ES: 0881492; FR: 158978; FR: 1552792; FR: 1180970; FR: 0881492; FR: 0875203; FR: 0815793; GB: 158978; GB:
1552792; GB: 1180971; GB: 1180970; GB: 0881492; GB: 0875203; GB: 0815793; IT: 1589878; IT: 1552792; IT: 0881492; IT: 0815793; JP:
4696150; KR: 532359; KR: 528102; NO: 326814; NO: 326202

Copyright © 2018 FUJIFILM SonoSite Inc, Inc. All rights reserved.

P16620-01

iii
iv
Table of Contents
Introduction ................................................................................. 1
Audience.................................................................................................................. 1
Contact Information ................................................................................................. 1
Terms and symbols ................................................................................................. 2
Labeling symbols ..................................................................................................... 2

Specifications .............................................................................. 3
Specifications........................................................................................................... 3
System Specifications ............................................................................................. 3
Electrical specifications ........................................................................................... 3
Environmental limits ................................................................................................ 4
Battery specifications............................................................................................... 4
Accessories and peripherals ................................................................................... 4

Safety .......................................................................................... 5
Electrical safety........................................................................................................ 5
Electrical safety classification .................................................................................. 7
Equipment safety ..................................................................................................... 7
DICOM standard...................................................................................................... 7
HIPAA standard....................................................................................................... 7
Transporting X-Porte Stand Systems ...................................................................... 7

System Overview ........................................................................ 9


About the System .................................................................................................... 9
Theory of Operation................................................................................................. 9
Description of Operating Modes .............................................................................. 10
Additional System Feature Performances ............................................................... 12
Front End Overview................................................................................................. 13
X-Porte Front End Board ......................................................................................... 14
PW Doppler Processing .......................................................................................... 15
CW Doppler Processing .......................................................................................... 16
Back End Overview ................................................................................................. 17
Control Subsystem .................................................................................................. 18
Power Supply and Control....................................................................................... 19
ECG Module ............................................................................................................ 20
DICOM..................................................................................................................... 21

Troubleshooting .......................................................................... 23
X-Porte Desktop and Stand Versions Major Component Diagrams ........................ 23
System and Subsystem Diagnosis .......................................................................... 23
System Repair ......................................................................................................... 24
Test Equipment........................................................................................................ 24
Failure (Assert) Codes............................................................................................. 24
Verifying a System Assert Code.............................................................................. 24
Clinical Monitor Display ........................................................................................... 25
DICOM..................................................................................................................... 26
SPMU ...................................................................................................................... 27
Touch Control Panel................................................................................................ 29
X-Porte Power-0 & Power-1 Procedural Steps........................................................ 29

v
Replacement Procedures............................................................ 33
Qualifying Service Personnel................................................................................... 34
Clinical Monitor Display ........................................................................................... 34
Required Parts......................................................................................................... 34
Required Tools ........................................................................................................ 34
Clinical Monitor Display Removal ............................................................................ 34
Clinical Monitor Display Replacement ..................................................................... 35
System (“Engine”).................................................................................................... 36
Required Part .......................................................................................................... 36
Required Tools ........................................................................................................ 36
System (“Engine”) Removal .................................................................................... 36
System (“Engine”) Replacement ............................................................................. 37
Clinical Monitor Stalk Cable..................................................................................... 37
Required Parts......................................................................................................... 37
Required Tools ........................................................................................................ 37
Clinical Monitor Stalk Cable Removal ..................................................................... 37
Clinical Monitor Stalk Cable Replacement .............................................................. 38
Touch Control Panel ................................................................................................ 38
Required Parts......................................................................................................... 38
Required Tools ........................................................................................................ 38
Control Panel Removal............................................................................................ 38
Control Panel Replacement..................................................................................... 40
Dock......................................................................................................................... 40
Required Part .......................................................................................................... 40
Required Tools ........................................................................................................ 40
Dock Removal ......................................................................................................... 40
Dock Replacement .................................................................................................. 41
Stand Power Management Unit (SPMU) ................................................................. 41
Required Part .......................................................................................................... 41
Required Tools ........................................................................................................ 41
Stand Power Management Unit (SPMU) Removal.................................................. 42
Stand Power Management Unit (SPMU) Replacement........................................... 45
Battery ..................................................................................................................... 46
Required Part .......................................................................................................... 46
Required Tools ........................................................................................................ 46
Battery Removal ...................................................................................................... 46
Battery Replacement ............................................................................................... 46
Triple Transducer Connect (TTC)............................................................................ 47
Required Part .......................................................................................................... 47
Required Tools ........................................................................................................ 47
Triple Transducer Connect (TTC) Removal ............................................................ 47
Triple Transducer Connect (TTC) Replacement ..................................................... 48
Optional Equipment - Stand System........................................................................ 49
PowerPark ............................................................................................................... 49
Required Part .......................................................................................................... 50
Required Tools ........................................................................................................ 50
PowerPark Stand Module Removal......................................................................... 50
PowerPark Stand Module Replacement.................................................................. 50
PowerPark Stand Module Removal (Alternate Method-Upright System) ................ 50
PowerPark Stand Module Replacement (Alternate Method-Upright System) ..... 50
Sony UP-D897MD Printer 50
Required Part .......................................................................................................... 51
Required Tools ........................................................................................................ 51

vi
Sony UP-D897MD Printer Removal ........................................................................ 51
Sony UP-D897MD Printer Replacement ................................................................. 52

Maintenance................................................................................ 51
Periodic Maintenance .............................................................................................. 51
Cleaning and Disinfecting ........................................................................................ 51

Performance Testing................................................................... 53
Overview.................................................................................................................. 53
Recommend Test Equipment .................................................................................. 53
Setting Up Performance Tests................................................................................. 53
Basic Operational Tests........................................................................................... 54
2D Performance Tests............................................................................................. 54
2D Performance / Image Quality ............................................................................. 54
Axial Measurement Accuracy .................................................................................. 55
Lateral Measurement Accuracy............................................................................... 55
Penetration .............................................................................................................. 56
Additional Performance Tests.................................................................................. 56
Color Doppler (Color) .............................................................................................. 56
Color Power Doppler (CPD) .................................................................................... 57
M Mode Imaging...................................................................................................... 57
Tissue Harmonic Imaging........................................................................................ 57
Pulsed Wave (PW) Doppler Imaging....................................................................... 58
Continuous Wave (CW) Doppler Imaging ............................................................... 58
Image Quality Verification Test/Livescan ................................................................ 58
Printer ...................................................................................................................... 59

Replacement Parts...................................................................... 61
Clinical Monitor ........................................................................................................ 62
Touch Control Panel ................................................................................................ 63
Dock......................................................................................................................... 64
“Ultrasound” Engine................................................................................................. 65
Stand Power Management Unit (SPMU & Batteries) .............................................. 66
Triple Transducer Connect (TTC)............................................................................ 67
Clinical Monitor Hardware........................................................................................ 68
Control Panel Hardware .......................................................................................... 68
Dock Hardware ........................................................................................................ 69
Stand Hardware Parts ............................................................................................. 69
Optional Equipment ................................................................................................. 70
PowerPark ............................................................................................................... 70
Sony UP-D897MD B&W Printer .............................................................................. 71
Desktop System Components ................................................................................. 72
EPS for Desktop System (Engine)........................................................................... 72
EPS for Desktop Clinical Monitor Display................................................................ 72
Other Desktop Components .................................................................................... 73
Ordering Replacement Parts ................................................................................... 73

Service Event Reporting ............................................................. 75


Service Event Report Form ..................................................................................... 76
Service Event Report Instructions ........................................................................... 77
Returning Products to FUJIFILM SonoSite.............................................................. 78

vii
Shipping Instructions .............................................................................................. 78

viii
Chapter 1: Introduction

Before servicing the X-Porte Ultrasound System, please read this manual.
The ultrasound system has multiple configurations and feature sets. All are described in this service manual
but not every option may apply to your system. System features depend on your system configuration,
transducer, and exam type.
Refer to the X-Porte Product Information & Safety Guide (P14646) for additional information regarding
safety, system controls, operation, capabilities, and specifications.

Audience
The intended audience of this manual are properly trained field and in-house service personnel. The
ultrasound system operators are expected to be able to replace all the components of the Desktop version
and the System “Engine” attached to the Stand version. With technical support guidance, a basic knowledge
of hand tools, and the ability to handle about 20 lb (9.1 kg), the system operators are expected to replace
the touch Control Panel and Clinical Monitor display. The Stand Power Management Unit (SPMU) and
batteries in the Stand version will be serviced by qualified field service personnel. The batteries used are
sealed Lithium-ion type which requires no maintenance and are not serviceable. Expected life for these
batteries are between 3-6 years depending on system usage and configuration of the Stand system.

Contact Information
Questions and comments are encouraged. FUJIFILM SonoSite is interested in your feedback regarding the
system and user documentation. Please call FUJIFILM SonoSite at 888-482-9449 in the U.S. Outside the
U.S., call the nearest FUJIFILM SonoSite representative.
For technical support, please contact FUJIFILM SonoSite as follows:

FUJIFILM SonoSite Technical Support

Phone 877-657-8118
(U.S. or Canada):

Phone 425-951-1330
(Outside U.S. and Or call your local representative.
Canada):

Fax: 425-951-6700

E-mail: [email protected]

Web site: www.sonosite.com

Europe FUJIFILM SonoSite - Amsterdam


Service Joop Geesinkweg 140
Center: 1114 AB Amsterdam
The Netherlands
Tel (Main): +31 20 751 2020
- English support: +44 14 6234 1151
- French support: +33 1 8288 0702
- German support: +49 69 8088 4030
- Italian support: +39 02 9475 3655
- Spanish support: +34 91 123 8451

Chapter 1: Introduction 1
FUJIFILM SonoSite Technical Support (Continued)

Asia e-mail:
Service [email protected]
Center: Tel: +65 6380-5581

Terms and symbols


The service manual follows these conventions:
• A WARNING describes precautions necessary to prevent injury or loss of life.
• A Caution describes precautions necessary to protect the products.
• Numbered steps in procedures must be performed in order.
• Items in bulleted lists do not require performance in sequence.

Labeling symbols
Please refer to Chapter 4: Labeling Symbols found in the X-Porte Product Information & Safety Guide
(P14646)

2 Chapter 1: Introduction
Chapter 2: Specifications

This chapter contains information regarding system specifications and accessory compatibility. The
information applies to the ultrasound system, transducers, accessories, and peripherals.

Specifications
System Specifications
Stand Dimensions:
• Length: 26.4 in. (67.1 cm)
• Width: 21.2 in. (53.8 cm)
• Height (max): 64 in. (162.6 cm)
• Height (min): 42.2 in. (107.2 cm)
• Height Adjustment: 9 in. (22.9 cm) travel

Touch Panel
• Capacitive Display Type
• Diagonal: 12.1 in. (30.7 cm)
• Multi-touch Gestures for System Controls
• Tilt Adjustment: 7.3 in. (18.5 cm)
• Side to Side Turning: +/- 90 degrees from center

Display (HD Monitor)


• Diagonal: 19 in. (48.3 cm)
• Screen Size: 1280 x 800
• Image Size: 800 x 600
• Monitor Tilt: 5° tilting forward from vertical, 20° tilting back from vertical

Electrical specifications
• Power Supply Input (Stand Version): 100-240V~ 6.0A Max, 50-60Hz
• Sony B/W Printer: 100-240V~ 1.5A-0.8A, 50-60Hz
• Power Supply Output 1 (Stand Version): 24 VDC, 11.5 A Max (Output not exceeding 275 watts)
• Power Supply Output 2 (Stand Version): 100-240V~ 3.0A - 1.5A, 50-60Hz
• Power Supply Input (Desktop Version): 100-240V~ 3.4A - 1.4A, 50-60Hz
• Power Supply Output (Desktop Version): 24 VDC, 6.25 A Max (Combined Output not exceeding 150
watts)
• Clinical Display Monitor Supply: 100-240V~ 2.7A, 50-60Hz
• Sony B/W Printer: 100-240V~ 1.5A-0.8A, 50-60Hz
• Batteries (Stand Version): 3 Lithium-ion batteries (385Wh total)
• Battery Use Time: 1.0 hour, 3 days on idle
• Battery Charge Time: 2.5 hours
• Battery Life: 3-6 years

Chapter 2: Specifications 3
Environmental limits
Note: The temperature, pressure, and humidity limits apply only to the ultrasound system, transducers,
and battery.

Operating
System and Transducer
10–40°C (50–104°F), 15–95% R.H.
700 to 1060hPa (0.7 to 1.05 ATM)

Shipping and Storage


System and Transducer
-35-65°C (-31-149°F), 15-95% R.H.
500 to 1060hPa (0.5 to 1.05 ATM)
Battery
-20–60°C (-4–140°F), 15–95% R.H. (For storage longer than 30 days, store at or below room temperature.)
500 to 1060hPa (0.5 to 1.05 ATM)

Battery specifications
The X-Porte stand uses three lithium-ion batteries that are charged and controlled by the Stand Power
Management Unit (SPMU) that can be switched off via three rocker type switches. These switches can be
accessed by lifting up from the stand bottom a black plastic door located under the left rear caster leg.
Run time is up to one hour (up to 3days on idle) depending on imaging mode and display brightness. This
chapter contains electrical, and clinical safety information required by regulatory agencies.

Accessories and peripherals


The system supports various accessories and peripherals. See the X-Porte Product Information & Safety
Guide (P14646) for a list of compatible products.

4 Chapter 2: Specifications
Chapter 3: Safety

This chapter contains electrical and clinical safety information required by regulatory agencies. The
information applies to the ultrasound system, transducers, accessories, and peripherals.

Electrical safety
This system meets EN60601-1, Class I/internally-powered equipment requirements and Type BF
(transducers) and Type CF (ECG leads) isolated patient-applied parts safety requirements.
This system complies with the applicable medical equipment requirements published in the Canadian
Standards Association (CSA), European Norm Harmonized Standards, and Underwriters Laboratories (UL)
safety standards.
For maximum safety observe the following warnings and cautions..

WARNING: To avoid the risk of injury, do not operate the system in the presence of
flammable gasses or anesthetics. Explosion can result.
WARNING: To avoid the risk of electrical shock or injury, do not open the system
enclosures. All internal adjustments and replacements, including battery
replacement, must be made by a qualified technician.
WARNING: To avoid the risk of electrical shock:
• Use only properly grounded equipment. Shock hazards exist if the power
supply is not properly grounded. Grounding reliability can be achieved only
when equipment is connected to a receptacle marked “Hospital Only” or
“Hospital Grade” or equivalent. The grounding wire must not be removed or
defeated.
• This equipment must be connected only to supply mains with protective earth.
• Do not let any part of the system (including the bar code scanner, external
mouse, power supply, or power supply connector), except for the transducer
or ECG leads, touch the patient.
• Do not touch any of the following:
• The ungrounded signal input/output connectors on the back of the
ultrasound system
• The system battery contacts (Stand System Only)
• The system transducer connector when the transducer or Triple
Transducer Connect (TTC) is disconnected
• Any unused TTC transducer connector when the TTC is connected.
• Do not connect the system's power supply to an MPSO or extension cord.
• Before using the transducer, inspect the transducer face, housing, and cable.
Do not use the transducer if the transducer or cable is damaged.
• Always disconnect the power supply from the system before cleaning the
system.
• Do not use any transducer that has been immersed beyond the specified
cleaning or disinfection level. See “Cleaning and Disinfecting X-Porte
Products” P16529
• Use only accessories and peripherals recommended by FUJIFILM SonoSite,
including the power supply. Connection of accessories and peripherals not
recommended by FUJIFILM SonoSite could result in electrical shock. Contact
FUJIFILM SonoSite or your local representative for a list of accessories and
peripherals available from or recommended by FUJIFILM SonoSite.

Chapter 3: Safety 5
WARNING: To avoid the risk of electrical shock and fire hazard:
• Inspect the power supply, AC power cords, cables, and plugs on a regular
basis. Ensure that they are not damaged.
• The power cord set that connects the power supply of the ultrasound system
or to mains power must only be used with the power supply, and cannot be
used to connect other devices to mains power.
WARNING: To avoid the risk of electrical shock, burn, or fire, use only AC power supplies
specified by FUJIFILM SonoSite.
WARNING: To prevent injury to the operator/bystander, the transducer must be removed
from patient contact before the application of a high-voltage defibrillation pulse.
WARNING: To avoid possible electrical shock or electromagnetic interference, verify proper
operation and compliance with relevant safety standards for all equipment
before clinical use. Connecting additional equipment to the ultrasound system
constitutes configuring a medical system. FUJIFILM SonoSite recommends
verifying that the system, all combinations of equipment, and accessories
connected to the ultrasound system comply with JACHO installation
requirements and/or safety standards such as AAMI-ES1, NFPA 99 OR IEC
Standard 60601-1-1 and electromagnetic compatibility standard IEC 60601-1-2
(Electromagnetic compatibility), and are certified according to IEC Standard
60950 (Information Technology Equipment (ITE) or IEC 60601-1.).
WARNING: Due to the captive nature of the Stand version system batteries, any shipping
transport, service activities, or extended periods of non-operation with the
system unplugged from the AC wall source, will require that the battery
ON-OFF switch be set to the “•O” (OFF) position. Please refer to Figure 3.1 on
page 6.

Caution: Do not use the system if an error message appears on the display monitor:
note the error code; call FUJIFILM SonoSite or your local representative; turn
off the system by pressing and holding the power key until the system powers
down.
Caution: To avoid increasing the system and transducer connector temperature, do not
block the airflow to the rear fans and ventilation holes around the bottom and
side perimeter of the Engine.

6 Chapter 3: Safety
Clinical Monitor display in the down (horizontal)
transport position.
Touch Control Panel in the down (horizontal)
transport position.

Stand collapsed Access to the Stand battery switches can be gained by


(lowest height) lifting the black plastic panel and pulling outward. This
for transport. panel location is at the SPMU stand housing area near
the left rear caster leg.

Battery
Switches
“•O”= OFF

Figure 3.1 Stand System Battery Switches & Transport Positioning

Electrical safety classification

Equipment safety
Please refer to the X-Porte Product Information and Safety Guide (P14646) for the complete information
concerning all safety and standards applicable to the X-Porte Ultrasound System

DICOM standard
Please refer to the X-Porte Product Information and Safety Guide (P14646) for the complete information
concerning the DICOM standard applicable to the X-Porte Ultrasound System

HIPAA standard
Please refer to the X-Porte Product Information and Safety Guide (P14646) for the complete information
concerning HIPAA standard applicable to the X-Porte Ultrasound System

Transporting X-Porte Stand Systems


Please observe the prior Warnings and Cautions outlined in this chapter concerning the stand version
system for transport and servicing. Lower (collapse) the stand to the lowest height position, fold to the
horizontal position the Clinical Monitor display and touch Control Panel, and access the three battery
switches and ensure that all switches are in the “•O” (OFF) position. Many government agencies require
that no electronic equipment be shipped via air transport with live battery power attached to the equipment.
Refer to Figure 3.1 on page 6 as to how to position the system stand and disconnect the battery power for
transport and servicing.

Chapter 3: Safety 7
8 Chapter 3: Safety
Chapter 4: System Overview

About the System


The FUJIFILM SonoSite X-Porte high-resolution ultrasound system is a portable, full featured, general
purpose, software controlled, diagnostic ultrasound system using all digital architecture. The system is used
to acquire and display high-resolution, real-time ultrasound data in 2D, M Mode, Pulsed Wave (PW)
Doppler, Continuous Wave (CW) Doppler, Color Power Doppler (CPD), and color Doppler (Color) or in a
combination of these modes.
The system has an electrocardiography (ECG) display feature and supports a 3-lead ECG cable assembly
to collect data for M Mode and Doppler measurements. The system provides measurement capabilities for
anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The
system has a PW and CW Doppler audio output feature, cine review, image zoom, labeling, biopsy,
measurements and calculations, image storage and review, printing, and recording capabilities.
The system includes optional Digital Imaging and Communications (DICOM) capabilities as well as general
computer communication capabilities to provide the acceptance, transfer, display, storage, and digital
processing of ultrasound images and loops. Security support is also provided to facilitate HIPAA
compliance.
The system displays the current output level in terms of one of two bioeffects indices (“Mechanical Index
[MI]” and “Thermal Index [TI]”) in accordance with the AIUM/NEMA Standard for Real Time Display of
Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment.

Theory of Operation
The X-Porte ultrasound system has seven (7) major functional groups:
• Transducer
• Acquisition Subsystem
• Processing Subsystem
• Display Subsystem
• Control Subsystem
• User Interface Subsystem
• Power Subsystem
The following block diagram shows the relationship of the functional groups.

Chapter 4: System Overview 9


Figure 4.1 Ultrasound System Functional Group

The Transducer elements convert the pulser voltage to acoustic energy during the transmit portion of the
ultrasound acquisition cycle. The elements convert the acoustic echo to voltage in the receive portion of the
acquisition. The voltage developed on the transducer elements is sensed by the acquisition subsystem. The
system transducers have 64 to 256 elements.
The Acquisition Subsystem consists of the beamformer and interface to the transducer. The beamformer
controls the timing of the transmit pulses to focus the acoustic beam. The beamformer amplifies the
low-level received echos and controls the receive focusing. The system beamformer transmits on up to 128
elements and receives on 64 elements.
The Processing Subsystem includes capabilities for interfacing with the beamformer and performing high
speed processing. The processing subsystem demodulates, filters, detects, and compresses the signal
supplied by the beamformer into display information.
The Display Subsystem converts the detected ultrasound data into picture elements (pixels). The software
user interface graphics are combined with the ultrasound information and converted to a video stream. The
external video port supports NTSC and PAL format.
The Control Subsystem consists of the central processing unit, program and video memory, permanent
image storage and retrieval memory, external communication interface ports, and connection to the user
interface keys. The control software includes the acoustic power and intensity software subsystem, power
group monitors, and a beamformer monitor. This software guarantees a level of patient safety by ensuring
the system is operating within acoustic power and intensity limits.
The User Interface Subsystem represents the software interface and form factor. The software interface
is the interaction between the user and the screen layout components. The form factor is the type of physical
buttons, location, and grouping of the buttons and the device size, shape, and weight. Dedicated controls
are for high usage activities and grouped according to the user workflow.
The Power Subsystem provides the system power and protects the hardware from destructive and/or
unsafe conditions by detecting failures in the system through hardware and software monitors. Detection of
a fault results in disabling of the pulser supply, and signaling of an error to the Control Group. The power
subsystem includes the batteries (quantity of three lithium-ion) and battery charging electronics.

Description of Operating Modes

2D Mode 2D mode is a two dimensional image of the amplitude of the echo signal. It is used for
location and measurement of anatomical structures and for spatial orientation during
operation of other modes. In 2D, a two-dimensional cross-section of a 3-dimensional
soft tissue structure such as the heart is displayed in real time. Ultrasound echoes of
different intensities are mapped to different gray scale or color values in the display.
The outline of the 2D cross-section may be a rectangle, parallelogram, trapezoid,
sector, or a full circle, depending on the particular transducer used. 2D mode can be
used in combination with any other modes.

M Mode M Mode is also known as “T-M mode” or “time-motion” mode. It is used primarily for
cardiac measurements such as valve timing and septal wall thickness when accurate
timing information is required.
Ultrasound echoes of different intensities are mapped to different gray scale values in a
scrolling display. M Mode displays time motion information of the ultrasound data
derived from a stationary beam. Depth is arranged along the vertical axis with time
along the horizontal axis. M Mode can be used alone but is normally used in
conjunction with a 2D image for spatial reference. The 2D image has a graphical line
(M-line) superimposed on the 2D image indicating where the M Mode beam is located.

10 Chapter 4: System Overview


Color In color Doppler, a real-time, two-dimensional cross-section of blood flow is displayed.
Doppler The 2D cross-section may be presented as a rectangle, parallelogram, trapezoid,
(Color) sector, or a full circle, depending on the particular transducer used.
The 2D cross-section is presented as a full color display, with various colors being used
to represent the velocity, both positive and negative, of the blood flow echoes. Often, to
provide spatial orientation, the full color blood flow cross-section is overlaid on top of
the gray scale cross-section of soft tissue structure (2D echo). For each pixel in the
overlay, the decision of whether to display VCD, gray scale (echo) information or a
blended combination is based on the relative strength of echoes from the soft-tissue
structures and from the red blood battery.
A high pass filter (wall filter) is used to remove the signals from stationary or slowly
moving structures. Tissue motion is discriminated from blood flow by assuming that
blood is moving faster than the surrounding tissue, although additional parameters may
also be used to enhance the discrimination. The remaining signal after wall filtering
may be averaged over time (persistence) to present a steady state image of blood flow
distribution. Variance information may also be displayed to provide information when
large variance is observed in the velocity information.

Color Power In CPD, a real-time two-dimensional cross-section of blood flow is displayed. The 2D
Doppler cross-section may be presented as a rectangle, parallelogram, trapezoid, sector, or a
(CPD) full circle, depending on the particular transducer used.
The 2D cross-section is presented as a full color display, with various colors being used
to represent the power in blood flow echoes. Often, to provide spatial orientation, the
full color blood flow cross-section is overlaid on top of the gray scale cross-section of
soft tissue structure (2D echo). For each pixel in the overlay, the decision of whether to
display CPD, gray scale (echo) information or a blended combination is based on the
relative strength of echoes from the soft-tissue structures and from the red blood
battery.
A high pass filter (wall filter) is used to remove the signals from stationary or slowly
moving structures. Tissue motion is discriminated from blood flow by assuming that
blood is moving faster than the surrounding tissue, although additional parameters may
also be used to enhance the discrimination. The power in the remaining signal after
wall filtering may be averaged over time (persistence) to present a steady state image
of blood flow distribution.

Continuous CW provides a real-time representation of blood flow and is displayed as a


Wave (CW) velocity-versus-time sweeping output. Velocity (or frequency) is presented as the
Doppler vertical axis with time along the horizontal axis. The magnitude of the detected signal is
represented as different gray scale values.
CW Doppler mode provides the clinician with the ability to obtain blood flow velocities
focused about a user specified focal region. A continuous transmit waveform of
ultrasound energy with a known frequency is transmitted and focused by the system;
on the receive side, the transducer receive echoes are continuously amplified, focused
about the focal region and converted to a base band quadrature signal. The signal is
analyzed by a quadrature phase detector that establishes two receive channels to
allow detection of flow direction. These two channels are then analyzed by a fast
complex Fourier transform (FFT) circuit to establish the spectrum of frequencies
present in the echoes. The data are displayed as spectrum frequencies with respect to
time.
CW can be used alone but is normally used in conjunction with a 2D image for spatial
reference. The 2D image has a graphical line (D-line) superimposed on the 2D image
indicating where the M-mode beam is located.

Chapter 4: System Overview 11


Pulsed Wave PW provides a real-time representation of blood flow and is displayed as a
(PW) velocity-versus-time sweeping output. Velocity (or frequency) is presented as the
Doppler vertical axis with time along the horizontal axis. The magnitude of the detected signal is
represented as different gray scale values. The ultrasound data is derived from a single
area, the sample volume, on a stationary beam.
PW Doppler mode provides the clinician with the ability to obtain blood flow velocities
about a spatial sample volume. A burst of ultrasound with a known spectrum is
transmitted by the system; on the receive side, the transducer receive echoes are
amplified and range gated at the appropriate depth. The signal is analyzed by a
quadrature phase detector that establishes two receive channels to allow detection of
flow direction. These two channels are then analyzed by a fast complex Fourier
transform (FFT) circuit to establish the spectrum of frequencies present in the echoes.
The data are displayed as spectrum frequencies with respect to time.
PW can be used alone but is normally used in conjunction with a 2D image for spatial
reference. The 2D image has a graphical line (D-line) superimposed on the 2D image
indicating where the M-mode beam is located. The sample volume position (depth) and
size are also indicated on the D-Line.

Additional System Feature Performances

Broadband Imaging This ultrasound acquisition system uses high resolution broadband technology
in the transmit pulsers, transducer, and receivers. The receive path can capture
and process signals over a wide spectrum, from below 2.0 MHz to beyond 10
MHz. For each application, the transmit pulse is designed to produce an
appropriate bandwidth. For example, in 2D grayscale imaging, a wide band
pulse is used to support good axial resolution. For Doppler modes, a narrower
band pulse is used, which improves the spectral resolution of the detected
Doppler signal.
In addition to transmit pulse control, programmable digital signal processing is
used in the receive path to further refine the bandwidth used to produce the
final image. Digital filters are applied to the digitized received signal to limit and
shape the spectral bandwidth used to generate the displayed output.

Tissue Specific In this feature, parameters for signal and image processing are optimized to
Imaging maximize the image quality or to obtain the best compromise of resolution and
penetration for different specific clinical applications. These parameters
include: the order of received filters, the bandwidth, the dynamic range, the
compression curve, the gain setting and parameters for compounding
frequency band, etc. For example, different system parameter setups are used
for abdominal or peritoneal scanning. This feature is for ease of use for the
operator by automatically setting up system control parameters rather than
manually adjusting settings for best performance.

Biopsy Guidance The system can display a pair of biopsy guidelines that represent the
anticipated path of the biopsy needle. The image of an anatomical target,
biopsy guidelines, a scan plane marker, and a biopsy needle are displayed to
assist in guiding the biopsy needle to the target. The system also provides
needle guidance for vascular access procedures. For additional information,
see the biopsy user guides.

Measurement and The system offers a variety of measurements and calculations, specific to
Calculation exam type and transducer. A list of them, and author references, are in the
Capabilities system user guide. Measurement accuracy is also discussed.

12 Chapter 4: System Overview


Continuous Wave The system provides for audio output of the CW velocity information. This can
Doppler Audio be presented as stereo information, with flow moving towards the transducer
Output on one channel and flow away on the other, or as a mono output with the single
audio output representing the summation of the flow directions.

Pulsed Wave The system provides for audio output of the PW velocity information. This can
Doppler Audio be presented as stereo information, with flow moving towards the transducer
Output on one channel and flow away on the other, or as a mono output with the single
audio output representing the summation of the flow directions.

Electrocardiograph ECG is provided to measure the electrical signal generated by the heart. A
(ECG) Display three lead interface: Right Arm (RA), Left Arm (LA) and Left Leg (LL), is
provided on the system.
The ECG signal is displayed as an amplitude-versus-time sweeping output.
Amplitude is presented on the vertical axis with time along the horizontal axis.

Front End Overview


The Front End is designed to support various imaging modalities such as 2D, M-Mode, Spectral Doppler
and Color Doppler. From the Front End's perspective, all modes can be grouped into a few basic types:
Single mode, simultaneous modes and triggered modes. All these modes are built from similar, basic
transmit and receive sequences controlled within the Front End. A generic top level block diagram of a
typical Front End is shown below in Figure 4.2 on page 13

Transducer Receive Section


T/R Sw

A/D Delay X

TGC Amplifier Weight

Beamformed
Pulser Delay Data

Waveform

T/R Sw

A/D Delay X

TGC Amplifier Weight

Pulser Delay

Waveform Control

Transmit Section

Figure 4.2 Generic Front End Block Diagram

The transmit section consists of a waveform generator, delay block, and high power high voltage driver to
excite the transducer element. Multiple elements are driven with delays determined by the time of flight in
the medium from the elements to the point in space where the beam is to be focused. The longer the time
of flight is to the focal point the smaller the delay is for a given transmit element to allow all to arrive at the

Chapter 4: System Overview 13


focal point at the same time. The number of elements driven is determined by element sensitivity off axis
and depth of field considerations. The waveform is selected to drive the transducer at a certain center
frequency, bandwidth, and power and is optimized for the given mode.

The receive section consists of a transmit/receive switch to protect the receiver from the transmit voltage, a
variable gain receiver to amplify and condition the return echoes, an A/D to digitize the data, a delay block
to focus the return signals and a weight block to scale the return echoes for each channel. All the signals
are then summed together to generate the beam-formed receive data. The analog gain varies with depth
to compensate for signal attenuation through the medium. The delays and weights are independent for
each channel. The delay and weight for the receive channel can typically be changed dynamically to keep
the receive beam in continuous focus. The delay is simply set by the time of flight in the medium from the
point of interest to the element, which starts at skin-line and proceeds to the deepest depth of interest.
The control section drives the data to the various data path elements on a line by line basis, controls the
timing of the transmit and receive sections and controls the tagged information and timing of the data to the
rest of the system.

X-Porte Front End Board


The X-Porte ultrasound system can be divide into two board section groups: Front End and Back End.
Below is the block diagram linking the necessary functional groups of an ultrasound system as they apply
to the X-Porte system. Please refer to Figure 4.3 on page 14

Figure 4.3 X-Porte Front End Board Block Diagram

14 Chapter 4: System Overview


PW Doppler Processing
Doppler processing includes both audio processing which presents Doppler signal in the form of stereo
audio and spectral processing which generates data for display of Doppler spectrum in the form of a
scrolling spectrogram. Doppler power spectrum is estimated performing Discrete Fourier Transforms on
short, overlapped segments of wall filtered Doppler signal. Doppler audio data is generated from wall filtered
data by phase shifting the in-phase component. Refer to the PW Doppler Processing Function Block
Diagram Figure 4.4 on page 15

Figure 4.4 PW Doppler Processing Function Block Diagram

Chapter 4: System Overview 15


CW Doppler Processing
CW Doppler data will be presented to the signal processor as complex (I/Q) data from the analog front end
of the external DSP. The 20-bit data will be presented as consecutive samples at a data rate varying from
1.5 kHz to 64 kHz for the complex pair. Most of CW processing is similar to that of PW except for the QBP
function. In place of QBP will be a low pass decimating filter that operates on incoming I/Q data.

The Doppler Processing block must allow storage of 256 undetected I/Q pairs in to allow the system to
measure and correct for phase mismatch. Measuring and correcting will need to be accomplished in system
software. Refer to the CW Doppler Processing Function Block Diagram Figure 4.5 on page 16

Figure 4.5 CW Doppler Processing Function Block Diagram

16 Chapter 4: System Overview


Back End Overview
The Back End subsystem is responsible for the conversion of raw acquisition data into a raster image ready
for display. The Back End subsystem also contains the video data path that supports generation of video
comprising of the ultrasound image as well as graphics annotation. Video generation of both standard
composite interlaced video and progressive scan video is supported. Most functionality is within the ASIC
but the memory resources for acquisition memory, and display memory are found in external memory
components. The conversion from PC type video to TV type video is also performed externally.
Control is received initially from the CPU to setup each functional block and afterward the hardware is
completely data driven. This control takes the form of programming setup registers inside the blocks and
setting up scan conversion tables. Each block provides temporary storage as required to buffer data and
keep their respective processing pipeline full and operating. Also note that the block diagrams show only
the data path, but each block is responsible for generating any necessary memory addresses for their
respective input data stream.
The X-Porte Back End subsystem is shown in Figure 4.6 on page 17.

Figure 4.6 X-Porte System Back End Subsystem Block Diagram

The Back End Subsystem performs processing encompassing three main data domains, acquisition data,
raster data, and video data. Support for acquisition data includes the input buffer, flash suppression, frame
average, and external ACQ memory. Cine buffer management is performed by the acquisition controller.
Conversion from acquisition data to raster data is performed by the graphics overlay, scan conversion sys-
tem, sweeping system, and 3D system. Raster data is stored in an external DISPLAY memory. Also sup-
porting raster operations is the graphics support block that provides acceleration hardware for pixel
operations from the CPU and graphics overlay system. The progressive video path includes buffers, prior-
ity logic, and LUTs.

Chapter 4: System Overview 17


Control Subsystem
The X-Porte Control Subsystem is shown in the figure below.

Control/Processor Section

10/100/1G Ethernet Switch Ethernet RJ45


10/100/1G
Ethernet

10/100/1G Ethernet

I2C

Backlight enable and brightness

Video out (LVDS)


If M2/S2: Main Clinical display
If POC: Touch Screen

USB Host for Touch Screen

USB Host for Dock

PCI Express for Dock

SATA 2

Serial Port Serial Port

Headphones Headphones
HDA Speakers Speakers
Audio Codec/
Amp Line In Audio Line In
S/PDIF
GPIOs to Dock (TBD)
Dock Connector

SDVO 720P CH7308A LVDS (24 bit) HDMI/DVI


receiver
24 bit video HDMI in
TP401A or
equivalent
COMe Type 10

LVDS (24 bit)


24 bit Video to
VGA
ADV7123 or
Equivalent
S/PDIF
Display Control I/F HDMI/DVI
SATA Boot
SATA 2 Control FPGA Display Control I/F
Device transmitter
Connector 24 bit video AD9889B or
equivalent Note:
Processor Switch HDMI with
Audio

Main Clinical display for


POC. If M2/S2 to Dock for
Hard IP Block in FPGA Aux Monitor with Audio
Soft IP Block in FPGA

PCI Express
SATA2 SATA Connector
Interrupts from the Front
End,and FPGA.
GPIO

Some GPIOs
Serial Lite (x4), WakeFromFE, FE_Detect FE I/F Connector
tied to both
FPGA and
CPLD
Program

Hirose: Configuration
JTAG
(Debug only) Video Buffer Ram
FLASH/CPLD
DDR3 1 GByte

Mini PCIe
PCI Express for WiFi (Debug Only)
USB

USB Host WiFi


USB Module

USB Type A
(x3)
Connector USB Host x3
(Debug only)

JTAG

Figure 4.7 Control Subsystem Block Diagram

The core control subsystem contains the processor, the system bus, and the system memory resources of
RAM and solid state hard drives. These two solid state hard drive devices are responsible for the system
operating
system software, ultrasound program software, scanhead operation files, on-board training videos, and
storage for patient data (patient demographic information, images, and video clips)

18 Chapter 4: System Overview


The main operational solid state storage drive is a mSATA form factor device totalling 16 GB of storage
capacity. This device is responsible for handling the operational system software, the ultrasound program,
and the on-board training videos.

A secondary solid state storage drive of 80 GB with an 1.8 inch form factor is used to handle the scanhead
data files, other ultrasound system files, and 66 GB of storage capacity reserved for all patient data. It is
with these new solid state devices, the X-Porte architecture is similar to our past products utilizing CF and
SD storage cards, but with much improved data storage and throughput without the use of any mechanical
hard drive devices.

Power Supply and Control


The X-Porte ultrasound system utilizes one of two types of power supply arrangements depending on the
X-Porte system version, which are Desktop or Stand. The Desktop system uses two separate external
power supply (EPS) units. The 150 watt EPS supplies +24VDC to the main Ultrasound System Electronics
(Engine) and a separate 65 watt EPS is utilized to supply +24VDC for the Clinical Display Monitor. No
batteries can be utilized with this version of system. The electrical block diagram for the 150 watt EPS is
shown below in Figure 4.8 on page 19

Figure 4.8 Desktop External Power Supply (EPS)

The Stand version X-Porte system utilizes a power supply/power control module mounted under the stand
base. This System Power Management Unit (SPMU) supplies the +24VDC power to the Engine and Clinical
Display Monitor. This SPMU is also the mounting point for the three lithium-ion batteries, charges the
batteries, and is responsible for switching the system from AC power to battery power when needed. Please
refer to Figure 4.9 on page 20

Chapter 4: System Overview 19


Figure 4.9 Stand Power Management Unit (SPMU)

ECG Module
The ECG module allows a representation of the heart electrical activity to be displayed in real time with
ultrasound images acquired and displayed on the system video display.
The ECG module interfaces to the patient through three (3) ECG leads: Right Arm ECG lead (RA), Left Arm
ECG lead (LA), and Left Leg ECG lead (LL). The ECG received signal from the ECG electrodes are isolated,
amplified, and filtered by the ECG module before it is sent to the system for further processing and display.
The ECG module and cable are an integrated assembly. The module receives power from the system.
Patient isolation is provided by the ECG module, allowing the connection and signals to the system to be
system-ground referenced. The isolation between the patient and the system meets the requirements of IEC
601-1 for Type CF equipment.
The ECG function accepts input from an external serial A/D and performs gain, filter, and DC Offset
functions. The resultant data is output at either the 200Hz sample rate or decimates the data by 2 or 4 and
outputs the data into acquisition memory. The data is assumed to be signed. An interrupt is provided that
will interrupt the processor after a set delay from the detected level and slope. A simple block diagram is
shown below in Figure 4.10 on page 21

20 Chapter 4: System Overview


Figure 4.10 ECG Block Diagram

DICOM
The system features Digital Imaging and Communications (DICOM) capability to provide the acceptance,
transfer, display, storage, and digital processing of single ultrasound images as well as loops of ultrasound
images.

Please refer to the X-Porte Product Information & Safety Guide (P14646) for the DICOM standard and the
X-Porte User Guide (P14645) for DICOM setup

Chapter 4: System Overview 21


22 Chapter 4: System Overview
Chapter 5: Troubleshooting

X-Porte Desktop and Stand Versions Major Component Diagrams


Please use Figure 5.1 on page 23below as a guide to help identify the major component sections of the two
X-Porte system types, Desktop and Stand versions.

Figure 5.1 X-Porte Major Component Diagrams

System and Subsystem Diagnosis


This section covers basic diagnostic and troubleshooting procedures you may follow if the system does not
operate properly. To diagnose system failures, consult the referenced diagnostic figures that follow or
FUJIFILM SonoSite Technical Support.
Table 5.1: Troubleshooting Subassemblies and Diagnostic Figures

Subassemblies Diagnostic Figures or Table

Assert Codes Figure 5.2; Table 5.2

Battery - Stand System (Linked with SPMU) Figure 5.3; Table 5.6

Clinical Monitor Display - Stand Version System Figure 5.1; Figure 5.6; Table 5.3

Clinical Monitor Display - Desktop Version Figure 5.1; Table 5.4


System

DICOM Table 5.5

SPMU - Stand Version System Figure 5.3; Figure 5.4; Table 5.6

Touch Control Panel - Stand Version System Table 5.7

Chapter 5: Troubleshooting 23
Table 5.1: Troubleshooting Subassemblies and Diagnostic Figures (Continued)

Subassemblies Diagnostic Figures or Table

Ultrasound System - Stand Version System Figure 5.5

System Repair
The system is repairable through subassembly replacement or through replacement of parts as
recommended by FUJIFILM SonoSite. Component level repair of Printed Circuit Board Assemblies is
performed only at the FUJIFILM SonoSite repair facility. Replacement of board level components by
unauthorized service facilities voids the FUJIFILM SonoSite warranty.

Test Equipment
Test equipment is not required for this troubleshooting section.

Failure (Assert) Codes


The X-Porte system is capable in displaying an “assert screen-call for help” instructing the user to contact
SonoSite Technical Support for hardware and software issues related to failures. PCB failures typically
result in “assert codes” that are output to both the clinical display monitor and the touch panel display. If an
assert screen appears, note the assert code information and contact FUJIFILM SonoSite Technical Support
to clarify the fault or failure. Refer to Figure 5.2 for how this assert screen will be displayed and the
information Technical Support will be asking from this screen.

Record the following information


and report these values to FUJIFILM
SonoSite Technical Support.

P: xxxxx
C: xxxxx
D: x

Figure 5.2 Assert Screen-Call For Help

Verifying a System Assert Code

24 Chapter 5: Troubleshooting
System asserts are caused by hardware and/or software faults. Hardware asserts typically require a PCB
or hardware replacement. Software asserts can be reset and the system may recover. A simple method to
identify the cause of the assert is identified here:

Assert Cause 1 Record the assert code.


2 Press and release the Power button to power the system down.
3 Press the Power button again to power on the system.
• If the system powers on normally, it has recovered from the fault (software assert)
and you may use the system.
• If the assert condition remains, corrective action must be taken; usually
replacement of a PCB or hardware is required. Contact FUJIFILM SonoSite
Technical Support for assistance and to obtain repair parts.
If the Power button is not functional, all sources of power must be removed to allow
the system to power down. I.e., disconnect AC power and switch the three battery
switches to the “•O” (OFF) position. Refer to Figure 3.1 on page 6

Table 5.2: Assert Codes

Code Code Description Information Probable Cause Solution

16203 PS1 PCBA in Engine Replace Engine

17193 PS1 PCBA in Engine Replace Engine

18706 BE PCBA In Engine Replace Engine

19325 istouchinfopresent () Touch Control Panel cable Reconnect cable connection


connection to Dock Assembly

19361 BE PCBA in Engine Replace Engine

19466 Front-end not detected Internal cables/PCB Replace Engine


connections

Note: Assert Code Table will be updated in future X-Porte Service Manual Revisions

Clinical Monitor Display


Table 5.3: Clinical Monitor Display - Stand System

Symptom Possible Causes

No displayed video image, blank or black, no Check +24V is present from Dock to Clinical Display
screen saver ICON. Monitor. Replace Stalk Cable Assy. (P16773) or Dock
(P15164) to restore +24V. If +24V is present, then
replace Clinical Display Monitor (P14627).
Note: The Desktop Power Supply (P16776) can be
used to test the Clinical Monitor Display operation.

Chapter 5: Troubleshooting 25
Table 5.3: Clinical Monitor Display - Stand System

Symptom Possible Causes

No Ultrasound Image but screen-saver ICON Check Stalk Cable Assy. connections from the Dock to
is present. the Clinical Display Monitor. Replace Stalk Cable
Assy. (P16773) and/or Dock (P15164). Try another
video source plugged into the HDMI monitor port.
Replace Clinical Display Monitor (P14627) if the
monitor only displays the screen-saver ICON with
another video source. Replace X-Porte Engine
(P16055) if monitor functions with other video source.

Brighter background shading (blooming) at Replace Clinical Monitor Display (P14627).


any of the image corners.

Image background is purple, red, or green Re-seat HDMI cable connection. Replace Stalk Cable
Assy. (P16773). Replace Clinical Monitor Display
(P14627).

No Audio Access X-Porte audio controls via “More” on the Touch


Control Panel. Check “Mute” is not enabled, adjust
volume level higher. Re-seat HDMI cable from DOCK
to Clinical Display Monitor. Test with another HDMI
cable. Replace Stalk Cable Assy. (P16773), Clinical
Display Monitor (P14627), Dock (P15164), then
X-Porte Engine (P16055) in this order until audio is
restored.

Table 5.4: Clinical Display Monitor - Desktop System

Symptom Possible Causes

No displayed video image, blank or black, no Check +24V is present from external Power Supply to
screen saver ICON. Clinical Display Monitor. Replace Power Supply, 65W,
24V (P16776). Replace Clinical Display Monitor
(P14627).

No Ultrasound Image but screen-saver ICON Check the HDMI cable from the X-Porte Engine HDMI
is present. port to the Clinical Monitor Display. Check Monitor
Display with another HDMI cable and video source.
Replace Cable Assy, HDMI (P15334), replace X-Porte
Engine (P16055) if monitor is good, or replace Clinical
Display Monitor (P14627) if defective.

Brighter background shading (blooming) at Replace Clinical Monitor Display (P14627).


any of the image corners.

Image background is purple, red, or green Re-seat HDMI cable. Replace Cable Assy, HDMI
(P15334). Replace Clinical Monitor Display (P14627).

No Audio Access X-Porte audio controls via “More” using the


system mouse. Check “Mute” is not enabled, adjust
volume level higher. Re-seat HDMI cable from Engine
to Clinical Display Monitor. Test with another HDMI
cable. Replace Cable Assy, HDMI (P15334), Clinical
Display Monitor (P14627), then X-Porte Engine
(P16055) in this order until audio is restored.

26 Chapter 5: Troubleshooting
DICOM
Table 5.5: DICOM Troubleshooting

Error Message Tiller Error Code Cause Troubleshooting

Socket TSOCKET_CONNECT_FAIL Invalid network Using Ping, verify that the


communication URE configuration. Printer/Archiver is connected.
failed Wrong port • If Ping fails, check the
number. devices IP address, X-Porte
Application is not IP address, Subnet mask,
running. and Gateway IP address.
Printer is offline. • If Ping is OK, use Verify to
check if device is available.
If Verify fails:
a) Check the
Printer/Archiver’s Port
configuration on the X-Porte.
b) Ensure that the Printer is
online and the Archiver’s
application is running.

Archiver TDICARCH_OPEN_FAILUR Wrong Capture Verify that the Archiver supports


transaction E Type selected the selected Capture Type
failed setting, e.g., US Image, SC
Image or US-Ret Image.

Printer TDICPRNT_OPEN_FAILURE Wrong Image Verify that the printer supports


transaction settings the selected Image settings.
failed E.g,. Color (RGB) or Grayscale
(Monochrome)

DICOM TDNETWORK_OPEN_FAILU Device does not Verify that X-Porte AE Title or IP


network RE recognize address is correctly configured
communication X-Porte, rejects on the Printer/Archiver.
failed association Note: Some devices require
that the Imaging modality
(X-Porte) be recognized in
order to accept images. This
requires configuration on the
device.

Internal failure TDNETWORK_READ_FAILU Invalid DICOM Check X-Porte Printer DICOM


detected RE Attribute settings for correctness (e.g.,
film size, format)

SPMU
The SPMU module has three LEDs that will help to isolate AC power versus X-Porte hardware failures. On
the top-rear portion of the X-Porte stand base is a “Green” LED that is near the equipotential stud. A solid
“Green” LED signifies that the X-Porte system is connected to a live AC source. If this LED is flashing at a
1Hz repetitive rate, then the system is charging the system batteries. When these batteries are fully
charged, this LED will switch to a solid “Green” LED illumination. Refer to Figure 5.3

Chapter 5: Troubleshooting 27
A solid Green LED indicates the system is connected to a live AC
source. When batteries are installed, this LED will flash
at a 1 Hz rate when the batteries are charging and go to a solid
illumination state upon battery charge completion.

Figure 5.3 Upper Stand Base Power LED

Two more LEDs are visible under the stand and can shine onto a piece of white paper placed under the
stand base. The solid illuminated “Blue” LED indicates that the SPMU is producing the system’s +24 volts
DC necessary to power the system electronics and Clinical Display monitor. The solid illumination faint
“Green” LED is indicating that the system is connected to a live AC source. This LED is indicating the same
AC as the topside “Green” LED except, this LED does not indicate the battery charge operation and will
always be constantly lighted when connected to a live AC source. If the bottom “Green” LED is not visible
due to the ambient light conditions, then check for the “Green” topside LED and the bottom “Blue” LED.
These indicate that the SPMU is connected to a live AC source and that the SPMU is outputting +24 volts
DC. Refer to Figure 5.4

SPMU is outputting +24 volts DC. System is connected to a live


AC source. This LED does not
monitor the SPMU battery
charge function.

If only the Blue LED is lighted,


then the system is working from
the batteries.

Figure 5.4 Stand Baseplate LEDs

Refer to Table 5.6 below to verify the following normal conditions and configurations that the X-Porte system
will operate properly. Any LED conditions other than those outlined in this table that exist could indicate a
SPMU problem requiring replacement.

Note: If all three battery switches have been in the “•O” (OFF) position and all three are set to “ʘ” (ON),
AC power has to be applied to the SPMU to allow the system SPMU to develop the +24VDC from battery
operation. The lower Blue LED will remain off indicating no +24VDC until performing this procedure.

Table 5.6: SPMU LED Conditions - Stand SystemTable 5.7

Lower Green
Normal Operation Conditions Top Green LED Lower Blue LED
LED

Connected to AC, no batteries ON ON ON (+24Vdc)

28 Chapter 5: Troubleshooting
Table 5.6: SPMU LED Conditions - Stand SystemTable 5.7

Lower Green
Normal Operation Conditions Top Green LED Lower Blue LED
LED

Connected to AC, batteries charging 1 Hz flashing rate ON ON (+24Vdc)

Connected to AC, batteries fully charged ON ON ON (+24Vdc)

Connected to AC, battery switches “0” ON ON ON (+24Vdc)


(OFF)

No AC, operation on battery power OFF OFF * ON (+24Vdc)

* Requires the AC power to be restored to the SPMU after all three battery switches have been switched to
“•O” (OFF) after service or air transport shipping. This Blue LED will not illuminate on battery operation until
AC power is first restored.

Touch Control Panel

Table 5.7: Touch Control Panel - Stand System

Symptom Possible Causes

Assert Code 19325 with description: Reconnect Touch Control Panel cable to the Dock
istouchinfopresent () port. Replace Touch Control Panel Assy. (P14027)if
problems persist. Replace Dock Assy. (P15164) as
last option.

Blooming, jumpy, or jittery graphic display Remove the Engine and check the TTC Assy. has all
four roller guides in the Stand channel holes. Re-install
TTC if not properly installed to the Stand. Check the
Touch Control Panel Assy. cable connection to the
Dock Assy. and re-install Engine. Replace the Touch
Control Panel Assy. (P14027)if problem persists.

Touch Control Panel is black or not functional Replace the Touch Control Panel (P14027) first. If
problem persists, the replace the Dock Assy. (P15164)

X-Porte Power-0 & Power-1 Procedural Steps


The Power-0 and Power-1 functionality has been implemented in X-Porte differently than it was in Turbo
due to the lack of buttons.

Chapter 5: Troubleshooting 29
The functionality is simple, but the format must be exact. Simply places a text file called reset.txt on a USB
stick (P16967) then insert it into a USB slot of the system, remove all other USB sticks, and now power “ON”
the system. The text file contains a phrase to tell the system which action to perform. Only one action can
be performed at a time. The file will be deleted by the system after it is read to prevent inadvertent usage.
Therefore, it is a good idea to save the reset.txt file to some other name before inserting into the system
and powering up. The actual implementation of the action is performed during system initialization and is
complete when the system disk becomes ready.
The Power-1 Reset function is specified by placing the phrase "restore-factory-settings" in the reset.txt
file. The command causes the system to:
• "Remove Worklist files
• "Rebuild the patient database
• "Start a new patient
• "Reset the DICOM location
• "Reset the DICOM, user and system logs
• "Reset user preferences
• "Reset the system clock to January 1, 2003
• "Reset the Control Panel brightness and backlight
• "Records the action in the system log
The Power-0 Reset function is specified by placing the phrase "reformat-data-drive" in the reset.txt file.
This command causes the system to:
• "Format the patient data drive and deletes all patient data
• "Start a new patient
• "Add a system log message recording the action

Figure 5.5 Troubleshooting Flow Chart - No System Operation

30 Chapter 5: Troubleshooting
Figure 5.6 Troubleshooting Flow Chart - No Clinical Monitor Display Image

Chapter 5: Troubleshooting 31
32 Chapter 5: Troubleshooting
Chapter 6: Replacement Procedures

FUJIFILM SonoSite has determined several service activities that will be performed only by FUJIFILM
SonoSite service personnel and/or their representative qualified to perform the service duties. These are
the removal/replacement of the SPMU and the batteries that are in the X-Porte Stand system. Please
contact FUJIFILM SonoSite or your appropriate FUJIFILM SonoSite representative when service is
required. Refer to Table 6.1 on page 34 for a more complete list of subassemblies and the personnel that
should attempt service on the X-Porte systems.
Prior to performing any service to the X-Porte system, remove the A.C. power cord from the wall outlet.
This applies to both the X-Porte Stand and X-Porte Desktop system versions.

WARNING:
The stand version is equipped with three lithium-ion batteries and require all three battery switches to be
switched to their “•O” (OFF) position to cease the +24VDC developed by the SPMU. Access to these
switches are by lifting the black plastic panel outward from the bottom of the X-Porte stand SPMU housing
area near the left rear caster leg. Refer to Figure 3.1 on page 6 of Chapter 3 under “Safety”.

WARNING:
The SPMU removal/replacement as well as the replacement of the batteries in the X-Porte Stand system
requires the service personnel to be able to reposition an upright stand system onto its front caster legs
(about 65 lb of the system weight) and lay the complete system horizontally on the floor. It is advisable to
have a second person to help in repositioning the system if the 65 lb weight requirement is too great for
one person. This service should only be performed by trained and qualified FUJIFILM SonoSite service
personnel.

Caution:
Follow all steps for stand height, Clinical Monitor, and touch Control Panel positioning in the forthcoming
service procedures to ensure no damage to the system and/or possible injury to the servicing personnel.

Caution:
Always use correct ESD procedures. ESD damage is cumulative and may not be noticeable at first. Initial
ESD symptoms may be slightly degraded performance or image quality.

Caution:
All fasteners should be torque to 5.5-inch pounds except where noted. Two ranges of torque screwdrivers
will be required as torque requirements range from 3-inch pounds to 24-inch pounds. Generally the torque
screwdrivers ranges are 1.0-10.0 inch pounds (0.23-1.1 Newton meter) and 10.0-25.0 inch pounds
(1.1-2.8 Newton meter)

Note: If all three battery switches have been in the “•O” (OFF) position and all three are set to “ʘ” (ON),
AC power has to be applied to the SPMU to allow the system SPMU to develop the +24VDC from battery
operation. The lower Blue LED will remain off indicating no +24VDC until performing this procedure.

Chapter 6: Replacement Procedures 33


Qualifying Service Personnel
Table 6.1: Responsible Personnel for Subassemblies

X-Porte Desktop System X-Porte Stand System


Subassembly
Qualified Personnel Qualified Personnel

Batteries N/A FUJIFILM SonoSite


Representative

B/W Printer Customer Customer

Clinical Monitor Display Customer Customer

Touch Control Panel N/A Customer

Dock N/A Customer

System Customer Customer

External Power Supply (EPS) Customer N/A

Isolation Transformer Customer FUJIFILM SonoSite


Representative

Mouse Customer Customer

Stand Power Management Unit N/A FUJIFILM SonoSite


(SPMU) Representative

Transducers Customer Customer

Triple Transducer Connect (TTC) N/A Customer

Clinical Monitor Display


The weight of the Clinical Monitor is 16.8 lb (7.62kg) and will require the person removing this assembly to
be able to support this weight with one hand while using a screw driver with the other hand.

Required Parts
• P14627-N Service Assembly, Display Clinical Monitor, X-Porte
• P14627-W Warranty Service Assembly, Display Clinical Monitor, X-Porte
• V14627-N Vet Service Assembly, Display Clinical Monitor, X-Porte
• V14627-W Vet Warranty Service Assembly, Clinical Monitor Display, X-Porte

Required Tools
• #1 Phillips screwdriver
• Torque screwdriver, 2.0-10.0 inch pounds (0.23-1.1 Newton meter)

Clinical Monitor Display Removal


Clinical Monitor Display Removal
1 Adjust the stand height to a comfortable level as to allow the best working height to access the mounting
screws and cables.

34 Chapter 6: Replacement Procedures


2 Adjust the Clinical Monitor to the full vertical position and open the cable access door on the back.
3 Loosen the two screws and remove the HDMI video wire clamp and pull the cable connector downward
to disconnect from the Clinical Monitor. Refer to Figure 6.1 on page 35.

Back of Clinical
Monitor
HDMI Video
Cable
+24VDC Power Connector

HDMI
H Wire
Clamp (P16329) Cable Access Door, Monitor
Mount (P16390)

Figure 6.1 Clinical Monitor Cable Connections

4 Un-thread the outer ring of the +24VDC power connector and pull downward disconnecting from the
Clinical Monitor. Refer to Figure 6.1 on page 35.
5 Remove the six black Phillips pan head screws (4 short and 2 long) mounting the Clinical Monitor to the
Clinical Monitor Display Mount. Refer to Figure 6.2 on page 35.
Note: After removing five of the screws, support the monitor against the mount as the indentation pattern
of the back of the monitor will help hold the assembly against the mount. Be prepared to support 16.8 lb
(7.62 kg) with one hand/arm when the final screw is removed.
6 Remove the Clinical Monitor display with both hands from the mount.

(P16380) HDMI Wire Bracket, Clinical Monitor


P16369 M4-.7 x 12 black
screw (x4) short screws

P16859 Mount, Clinical


Monitor (P16639) HDMI Wire
Clamp, Clinical Monitor
(notched area up toward
HDMI connector)

(P15551P15551) M3-.5 x 6 Screw, Pan Head Phillips


P16371 M4-.7 x 35 black
screw (x2 long screws) *(P16390) Cable Door, Monitor Mount (not shown)

Figure 6.2 Clinical Monitor Mounting Hardware

Clinical Monitor Display Replacement


Clinical Monitor Display Replacement
1 Perform the Clinical Monitor Display Removal steps 1-6 in the reverse order.
2 Tighten all screws to 5.5-inch pounds.

Chapter 6: Replacement Procedures 35


System (“Engine”)
Caution:
Prior to unlatching the System from the stand, please check to ensure all USB connections on the right
stand side and the ECG cable on the left stand side are removed. Failure to remove any devices plugged
into the System ports will inhibit the System from unlatching from the stand as well as possible risk of
damage to the connection ports of the System. Check to ensure that the TTC connection has been
unlatched and the TTC main connector is pulled downward and away from the base of the System.

Required Part
• P16055-N Service Assembly, X-Porte Engine, CW
• P16055-W Warranty Service Assembly, X-Porte Engine, CW
• V16055-N Vet Service Assembly, X-Porte Engine, CW
• V16055-W Vet Warranty Service Assembly, X-Porte Engine, CW

Required Tools
• None

System (“Engine”) Removal


System Removal
1 Remove any USB devices connected into the right System/Stand side ports as well as the ECG cable
connected to the left System/Stand side ports. Refer to Figure 6.3 on page 36.
Note: Failure to remove connections into the System from either stand side will inhibit the release of the
System from the stand as well as cause damage to the System’s connector ports.
2 Unlatch the Triple Transducer Connector (TTC), if installed, or any transducers connected directly to the
System (Engine) and pull the TTC connector downward away from the base of the System. Refer to
Figure Figure 6.18 on page 48
3 Locate the two gray latch levers under the right and left side stand top platform near where the black
metal mates to the shiny metal handle. Refer to Figure 6.3 on page 36.

System USB Ports Right System Latch


Left System Latch System ECG Port

Left Stand Side Right Stand Side

Figure 6.3 System Latch Release & Ports

4 Simultaneously push each latch outward away from the System toward the stand outer side edges. If
all System connector ports are empty and the TTC main connector properly pulled away from the
bottom of the System, the System will eject upward from the stand.

36 Chapter 6: Replacement Procedures


5 Lift the System from the back edge and lift forward and up away from the top of the stand.

System (“Engine”) Replacement


System Replacement
1 Place the replacement X-Porte System (“Engine”) into the stand System head by placing the front edge
into the head first just under the two metal retaining tabs at the front edge of the stand’s System head.
2 Push down at the top back edge of the System and the stand latches should engage and hold the
System in place.
3 Reconnect the TTC, if equipped, by pushing upward on the TTC main connector and turning the latch
into the locked position.
4 Reconnect any USB cables and/or the ECG cable to the right and left sides of the System/Stand area.

Clinical Monitor Stalk Cable

Required Parts
• P16673--N Service Cable Assembly, Clinical Monitor Stalk Cable Set, X-Porte
• P16673-W Warranty Service Cable Assembly, Clinical Monitor Stalk Cable Set, X-Porte

Required Tools
• #1 Phillips screwdriver

Clinical Monitor Stalk Cable Removal


Clinical Monitor Stalk Cable Removal
Note: Prior to performing this procedure, remove the Clinical Monitor and Engine as directed in
procedures above. The HDMI cable and Clinical Monitor power cable should also be removed from the
back of the Dock as directed in the Dock Removal Procedure below.
1 Remove two screws from the top plate of the stalk assembly. Refer to Figure 6.4

Chapter 6: Replacement Procedures 37


2 Carefully pull Stalk Cable assembly upward to remove and stagger the power and HDMI cable
connectors as they go through the center of the mount so they do not get stuck.
Pull Stalk
Remove upward
Screws (x2)

Stagger
connectors

Figure 6.4 Cable Stalk Removal

Clinical Monitor Stalk Cable Replacement


Clinical Monitor Stalk Cable Replacement
1 Install Clinical Monitor Stalk Cable by reversing steps 1-2 above. Make sure the cables do not twist and
pass straight through the monitor mount.
2 Check to make sure the cable split sleeve is in proper position. Failure to do so may result in excessive
wear on the HDMI cable. Sleeve should be located as shown in Figure 6.5

Split Sleeve should not be visible from underneath. If it is, push


proper location back up as shown by the blue arrow.

Figure 6.5 Split Sleeve

Touch Control Panel

Required Parts
• P14027-N Service Assembly, Control Panel, X-Porte

38 Chapter 6: Replacement Procedures


• P14027-W Warranty Service Assembly, Control Panel, X-Porte
• V14027-N Vet Service Assembly, Control Panel, X-Porte
• V14027-W Vet Warranty Service Assembly, Control Panel, X-Porte

Required Tools
• #1 Phillips screwdriver
• 3mm Hex Allen Wrench
• Torque screwdriver, 10.0-25.0 inch pounds (1.1-2.8 Newton meter)

Control Panel Removal


Control Panel Removal
1 Adjust both the Clinical Monitor display and the Control Panel to their full vertical position.
2 Remove the X-Porte System (“Engine”) following the steps in the System (“Engine”) Removal section
below and refer to Figure 6.3 on page 36.
3 Remove the Control Panel cable connection at the Dock assembly by squeezing both sides of the
connector plug and pulling out from the Dock. Refer to Figure 6.6 on page 39

Dock Assembly

Control Panel Cable Plug


(squeeze the side tabs to
release)*

Cable Channel Retainer

*Note: (Not Pictured) a retainer


(P17671) has been added to the
latest build Dock Assy. to clamp
the Control Panel Cable Plug
straight into the Dock port.

Figure 6.6 Control Panel Cable Plug

4 Continue to pull the Control Panel cable out of the stand channel retainers toward the Control Panel

Chapter 6: Replacement Procedures 39


.
Backside of Control Panel

(P16375) M4 x 12, Socket


Cap Screw, Black (x4)
(P16374) M4 External
Tooth Washer, Silver (x4)
Rubber Grommet
Ferrite Iron Core

Figure 6.7 Control Panel Mount

5 At the vertical cable retention, carefully pry out the rubber grommet and the cylindrical ferrite iron core.
It may require some of the cable to be fed up from the bottom to loosen the cable. Refer to Figure 6.7
on page 40.
6 Using a 3mm Hex Allen Wrench, remove all four M4 x 12 socket cap (hex) screws in both Control Panel
hinges.
Note: Use care and do not lose the four M4 external tooth washers on the M4 x 12 socket cap screws.
7 The Control Panel is now unmounted from the stand.

Control Panel Replacement


Control Panel Replacement
1 Perform steps 1-7 in the reverse order from the Control Panel Removal section of this chapter.
2 Refer to Figure 6.6 on page 39 and Figure 6.7 on page 40.
3 Torque the four black M4 x 12 socket cap screws with a M4 external tooth washer at each of the two
hinges at 24-inch pounds.

Dock

Required Part
• P15164-N Service Assembly, Dock with DVR, X-Porte
• P15164-W Warranty Service Assembly, Dock with DVR, X-Porte
• V15164-N Vet Service Assembly, Dock with DVR, X-Porte
• V15164-W Vet Warranty Service Assembly, Dock with DVR, X-Porte
Note: The Latest version of the Dock Assy. will now incorporate additional hardware to secure the Touch
Control Panel cable and the Clinical Display Monitor HDMI cable not shown in this chapter’s figures. Two
pads will be located on the bottom of the Dock to help maintain the Dock level with the Engine connection
port. The extra hardware are as follows: (P17671) Retainer; (P16639) HDMI Wire Clamp; (P16894) HDMI

40 Chapter 6: Replacement Procedures


Cable Bracket; and (P18338) PORON Foot (x2).Please refer to Figure A.4 on page 65 for the latest image
and hardware comprising the Dock Assembly.

Required Tools
• # 2 Phillips Screw Driver

Dock Removal
Dock Removal
1 Positional the Clinical Monitor display and touch Control Panel to their complete vertical positions.
2 Remove the X-Porte System (“Engine”) from the Stand using steps 1-5 under the System section of this
chapter. Refer to Figure 6.3 on page 36
3 Remove the (P16639) HDMI Wire Clamp from the (P16894) HDMI Cable Bracket. Refer to Figure A.4
on page 65 for the image of the hardware.
4 Disconnect the Control Panel cable connector from the Dock. Refer to Figure 6.6 on page 39. The new
Dock Assembly will have the (P17671) Retainer requiring it to be flipped upward before removing the
Control Panel cable. Please refer to Figure A.4 on page 65

Top View of Dock

Rear View of Dock


Dock

S (P16356) M4 Flathead Phillips Screw (x4)


Stand System head (P16355) M4 Washer, Stainless (x4)

Figure 6.8 Dock Cabling

5 Remove the four M4 flathead Phillips screws (P16356) and the four M4 stainless steel countersunk
washers (P16355) from the Dock mounting posts located inside the Stand System head area. Refer to
Figure 6.8 on page 41
6 Pull the Dock up and away from the four mounting posts.
7 Disconnect the remaining cables attached to the rear of the Dock. The Dock can now be moved out of
the stand.

Dock Replacement
Dock Replacement
1 Hold the Dock above the four mounting posts referencing Figure 6.8 and Figure 6.6.
2 Install the power cable for the Clinical Monitor.
3 Install the power cable from the SPMU.
4 Install the USB cable to the printer if present.
5 Place the Dock onto the four mounting posts and install the four phillips screws and washers.
6 Install the HDMI cable making sure a proper service loop is in place per Figure 6.9. Failure to do so may
cause excessive wear on the cable when the Clinical Monitor is folded down to collapse for transport.

Chapter 6: Replacement Procedures 41


7 Install the (P16639) HDMI Wire Clamp tighten the two screws..

Pull cable straight,


then push cable
back while making
one full clockwise
turn

Proper orientation
after installation

Figure 6.9 HDMI Cable Installation

Stand Power Management Unit (SPMU)


Service for the SPMU and batteries should only be performed by qualified FUJIFILM SonoSite service
personnel or their representative. Due to the lifting and positioning nature of performing the following steps,
please refer to the Warnings and Cautions printed at the beginning of this chapter.

Required Part
• P14522-N Service Assembly, Stand Power Management Unit, X-Porte
• P14522-W Warranty Service Assembly, Stand Power Management Unit, X-Porte
• V14522-N Vet Service Assembly, Stand Power Management Unit, X-Porte
• V14522-W Warranty Vet Service Assembly, Stand Power Management Unit, X-Porte

Required Tools
• #1 Phillips screwdriver
• 5mm Hex Allen Wrench
• Torque screwdriver, 10.0-25.0 inch pounds (1.1-2.8 Newton meter)

Stand Power Management Unit (SPMU) Removal


Stand Power Management Unit (SPMU) Removal
1 Removal all items from the stand rear storage basket.
2 Remove all cables, adapters, and any accessories plugged into either side of the System. Refer to
Figure 6.3 on page 36.
3 Unlatch the System via the two small gray unlock levers. Refer to Figure 6.3 on page 36.
4 Pull up, forward, and away from the stand to remove the System from the stand System head.
5 Position the touch Control Panel to the horizontal (shipping position). Refer to Figure 6.10 on page 43.
6 Fold the Clinical Monitor display to the horizontal (shipping position). Refer to Figure 6.10 on page 43.
7 Adjust the stand to the minimum height (collapsed-shipping position).
8 Position the two front casters under the stand caster legs and lock all four casters. Refer to Figure 6.10
on page 43.

42 Chapter 6: Replacement Procedures


Front Casters Tucked
& Locked Under The
Stand Legs

Stand In The
Service Or
Transport View of the Battery Switch Access
Mode Panel From The Cart Rear.

Figure 6.10 Stand Adjusted for Service or Transport

9 Access the three battery switches and set all switches to the “•O” (OFF) position. Refer to Figure 6.10
on page 43.
10 Tip the stand forward and continue towards the floor so that the stand rests with the front two caster legs
and the brushed metal handle rest horizontal on the floor. This will expose the bottom plate of the stand
base. Refer to Figure 6.11 on page 43.

Figure 6.11 Stand Tilted Forward-Bottom Plate

11 Remove the PowerPark Stand Module two mounting M5 x 20mm (P13629)Phillips screws. This is an
optional feature and may not be present on your stand. Refer to Figure 6.12 on page 44
12 Remove the six M5 x 8 (P16382) Phillip screws securing the bottom plate and remove the plate. Refer
to Figure 6.12 on page 44

Chapter 6: Replacement Procedures 43


(Optional) PowerPark
Stand Module (P12822)

(P13629) M5 x 20mm
Pan Head Steel Screw (x2)

(P16382) M5 x 8 Pan
Head Steel Screw (x6)

Figure 6.12 Stand Bottom Plate

13 Remove the three batteries mounted to the SPMU by loosing the four captive Phillips screws holding
each battery and pull each battery off of the SPMU. Refer to Figure 6.13 on page 44.

Screws (4x)
(Captive)

Figure 6.13 Batteries

14 Loosen the seven captive 5mm hex head screws securing the SPMU to the bottom of the stand and pull
away from the stand. Considerable effort is needed to break the adhesion caused by the gray thermal
pad sandwiched between the heat sink of the SPMU and the bottom of the stand. Refer to Figure 6.14
on page 45.

44 Chapter 6: Replacement Procedures


5mm Hex Screws 5mm Hex Screws
(Captive) (Captive)

Figure 6.14 SPMU Mount

15 Pull the SPMU away from the bottom of the stand and gain access to remove the AC power cord held
captive via a metal wire bail. Flip the wire bail to release the AC power cord.
16 Un-thread the retaining ring of the DC power cable plug and remove from the SPMU DC output socket.
The SPMU can now be removed from the stand. Refer to Figure 6.15 on page 45

SPMU +24VDC Output AC Power Cord to Optional


(Threaded Connector) Sony UP-D897MD Printer
(P15774)

Figure 6.15 SPMU Power Connectors

Chapter 6: Replacement Procedures 45


Stand Power Management Unit (SPMU) Replacement
Stand Power Management Unit (SPMU) Replacement
1 Remove the clear plastic sheath protecting the gray thermal pad on the replacement SPMU assembly
prior to installing in the Stand. Refer to Figure 6.16 on page 46.

Remove

Top View of SPMU

Remove the clear plastic


protective sheet from
the gray thermo-pad
prior to installation.

Figure 6.16 SPMU As Shipped

2 Follow the above SPMU Removal steps and perform in the reverse order for installation of the
replacement SPMU.
3 Tighten the seven SPMU 5mm hex head mounting screws, the captive battery screws, and the six M5
x 8 Phillips screws holding the bottom plate to 20-inch pounds.
Note: The Bottom Plate will mount inside out. When the Bottom Plate is incorrectly mounted, the
PowerPark will not connect to the stand. Please ensure that the Bottom Plate has the top opening under
the AC power cord IEC socket and that the two captive thread areas are not protruding out of the stand
base. Refer to Figure 6.17 on page 46

Rectangular
Opening
Stamped
Number
Captive Threaded
Correct Incorrect Nut (protruding)

Figure 6.17 Correct Bottom Plate Orientation

46 Chapter 6: Replacement Procedures


4 Ensure that the three battery switches on the SPMU are set to “ʘ” (ON) position after all service has
been performed (tested and batteries installed) and the system is back in the upright position.

Battery
The batteries in the X-Porte Stand are three Lithium-ion batteries and all three must be replaced at the same
time. Upon ordering these batteries, they are packaged as one set of three batteries. Since the batteries
mount directly to the SPMU module, the procedure for battery removal is explained during the SPMU
module removal section.

Required Part
• P14586 Battery, X-Porte Stand (Set of Qty. Three)

Required Tools
• #1 Phillips screwdriver
• Torque screwdriver, 10.0-25.0 inch pounds (1.1-2.8 Newton meter)

Battery Removal
Battery Removal
1 Please follow steps 1-12 of the “Stand Power Management Unit (SPMU) Removal” section of this
chapter.

Battery Replacement
Battery Replacement
1 Perform the battery replacement by following in reverse order steps 1-12 of the “Stand Power
Management Unit (SPMU) Removal” section of this chapter.
2 Tighten the captive battery retention screws and the six M5 x 8 Phillips screws for the bottom stand plate
to 20-inch pounds. Refer to Figure 6.13 on page 44 and Figure 6.17 on page 46

Triple Transducer Connect (TTC)

Required Part
• P14644-N Service Assembly, TTC Assembly, X-Porte
• P14644-W Warranty Service Assembly, TTC Assembly, X-Porte
• V14644-N Vet Service Assembly, TTC Assembly, X-Porte
• V14644-W Vet Warranty Service Assembly, TTC Assembly, X-Porte

Required Tools
• None

Triple Transducer Connect (TTC) Removal


Triple Transducer Connect (TTC) Removal
1 Reach under the stand head area and unlatch the main TTC connector from the bottom of the stand
head by flipping the latch handle forward and rotate counter clockwise. Refer to Figure 6.18 on page 48

Chapter 6: Replacement Procedures 47


TTC Latch Locked in Operate Mode TTC Latch Flipped Down to Unlock, Turn Counter
Clockwise to Release.

Figure 6.18 TTC Main Latch

2 When loose, pull the latch handle down and the connector should drop down several inches. The TTC
is free from the System. Refer to Figure 6.19 on page 48.

Left Underside
of Stand System
Head Assy.

T TC Stand Release

Note that the TTC Connector has Dropped Down & is Loose from the System.

Figure 6.19 TTC Stand Release

Note: Updated TTC ‘s now incorporate a secondary slide latch on the right side that requires the right
hand to activate this latch while the left hand pulls the blue stand release pin while pushing the TTC to the
left. This ensures that the TTC will be supported by both hands and not drop to the floor.
3 Place the right hand at the right bottom section of the TTC and the left hand under the TTC next to the
blue System Stand release pin. While pulling the blue pin downward and sliding the right secondary
latch to release, push the TTC sideways to the left.
4 When the TTC protrudes to the left beyond the left stand edge, it will drop free from the System Stand.
Be prepared to support the full weight of the TTC.

48 Chapter 6: Replacement Procedures


Triple Transducer Connect (TTC) Replacement
Triple Transducer Connect (TTC) Replacement
1 To mate the TTC to the stand, start by aligning the four stainless steel guide wheels on the top portion
of the TTC assembly which slide into the four TTC guide channels located at the bottom of the stand
System head.
2 Position the TTC approximately 3/4” farther to the left than the left stand edge to allow the TTC stainless
steel guide wheels to enter the stand TTC guide channels. Please refer to Figure 6.20 on page 49 for
the location for these attachment points.

Left Underside of Stand System head

TTC Guide Channels (x4)

TTC Guide Wheels (x4)


TTC Blue Locking Pin
(Part of TTC Stand Release)

Right Topside of TTC Assembly

Main TTC Connector

Figure 6.20 TTC Mounting Channels

3 Perform the TTC Removal steps 1-4 in the reverse order to complete the replacement operation.

Chapter 6: Replacement Procedures 49


Optional Equipment - Stand System

PowerPark
The PowerPark accessory is a cordless charging station for most FUJIFILM SonoSite system stands. It is
comprised of two portions with the PowerPark Stand Module (P12822) mounted to the underside of the
stand which then connects to the PowerPark Dock Module (P12834) that is connected to the wall AC and
sits on the floor. In the event that the stand module needs to be replaced on a X-Porte Stand System, please
perform the following procedures.

Required Part
• P12822 Assembly, PowerPark Stand Module

Required Tools
• #1 Phillips screwdriver
• #1 Phillips “stubby” screwdriver or ratchet type screwdriver (for repair without laying the stand on its side)

PowerPark Stand Module Removal


PowerPark Stand Module Removal
Note: The following will outline the replacement method by tipping the stand onto its side to gain full view
and access to the underside of the stand. If tipping the stand is not possible, then proceed to the “alternate”
methods.

1 Please refer to section in this chapter titled “Stand Power Management Unit (SPMU) Removal” and
follow steps 1 through 11. Please refer to Figure 6.10 on page 43, Figure 6.11 on page 43, and
Figure 6.12 on page 44
2 When the two M5 x 20mm Phillips (P13629) screws have been removed securing the PowerPark Stand
Module, slide the module out from stand IEC power cord port.

PowerPark Stand Module Replacement


PowerPark Stand Module Replacement
1 Perform the PowerPark Stand Module replacement following the prior removal steps 1 through 2 in the
reverse order.
2 Access the three battery switches and set all switches to the “ʘ” (ON) position. Refer to Figure 6.10 on
page 43

PowerPark Stand Module Removal (Alternate Method-Upright System)


PowerPark Stand Module Removal (Alternate Method)
1 Pull the X-Porte Stand System away from the PowerPark Dock Module or remove the AC power cord
connected to the PowerPark Stand Module.
2 Access the three battery switches and set all switches to the “•O” (OFF) position. Refer to Figure 6.10
on page 43.
3 While lying on the floor to gain access to the underside of the X-Porte Stand, use a “stubby” or rachet
type #1 Phillips screw driver to remove the two M5 x 20mm Phillips (P13629) screws.
4 Once the PowerPark Stand Module is loose, slide the module out away from the stand IEC power cord
port.

50 Chapter 6: Replacement Procedures


PowerPark Stand Module Replacement (Alternate Method-Upright System)
PowerPark Stand Module Replacement (Alternate Method)
1 Perform the PowerPark Stand Module replacement following the “alternate” removal steps 1 through 4
in the reverse order.
2 Access the three battery switches and set all switches to the “ʘ” (ON) position. Refer to Figure 6.10 on
page 43.

Sony UP-D897MD Printer


The Sony UP-D897 digital thermal printer is a monochrome black and white printer which uses the USB 2.0
port to integrate to the X-Porte Stand and Desktop systems. This is an optional device and no special
installation is necessary for the X-Porte Desktop system. The following instructions are for any replacement
needs if the Sony UP-D897MD has to be replaced which is mounted to the X-Porte Stand (Kiosk) system.

Required Part
• P16193-N Service Assembly, B&W Printer, X-Porte
• P16193-W Warranty Service Assembly, B&W Printer, X-Porte
• V16193-N Vet Service Assembly, B&W Printer, X-Porte
• V16193-W Vet Warranty Service Assembly, B&W Printer, X-Porte

Required Tools
• #1 Phillips screwdriver
• 3mm hex long T-handle driver (optional)

Sony UP-D897MD Printer Removal


Sony UP-D897 Printer Removal
1 Raise or lower the Stand to a comfortable height to gain access to the printer shelf hardware.
2 From the rear of the Stand, loosen the captive screw that may be tightened to holding the printer cable
shield closed. Loosen the screw and glide the screw through the channel to swing open the cover. Refer
to Figure 6.21 on page 51
3 Remove the AC power cord and USB digital video cable from the printer.

(P15551) M3-.5 x 8mm Screws (x4)

Mounting Bracket 3mm Hex Screw

Figure 6.21 Sony UP-D897MD Cable Cover

4 From the underside of the printer mounting shelf, remove the four M3-.5 x 8mm Pan Head Phillips
screws (P15551) threaded into the Sony UP-D897MD printer. Refer to Figure 6.22 on page 52

Chapter 6: Replacement Procedures 51


Note: Alternate method is to use a 3mm hex T-handle driver to loosen the hex screw in the clear mounting
bracket and remove the printer shelf from the stand column. This method allows the four (P15551) M3-.5 x
8 mm screws to be removed and the printer replaced on a table and not from under the mounted shelf.

(P15551) M3-.5 x 8mm Pan Head Phillips Mounting Screws (x4)

Note: Viewed with the printer shelf removed from the Stand.

Figure 6.22 Stand Printer Shelve & Mounting Screws

5 Remove the Sony UP-D897MD printer from the shelf.

Sony UP-D897MD Printer Replacement


Sony UP-D897 Printer Replacement
1 Perform the Sony printer replacement by following the Sony UP-D897MD removal steps 1 through 5 in
the reverse order.

52 Chapter 6: Replacement Procedures


Chapter 7: Maintenance

This chapter contains information to direct you to where information about care for the system, transducers,
and accessories can be obtained.

Periodic Maintenance
No periodic or preventive maintenance is required for the system, transducers, or accessories other than
cleaning and disinfecting the transducer after every use. (See the “Cleaning and Disinfecting X-Porte
Products” guide (P16529). There are no internal adjustments or alignments required and there are no
internal components that require periodic testing, calibration, adjustment, or alignment. Performance tests
are described in Chapter 8, “Performance Testing” of this manual. Performing maintenance procedures not
described in this manual may void the product warranty.
Local regulations may require electrical safety testing.
Contact FUJIFILM SonoSite Technical Support for any maintenance questions. (See “Contact Information”
on page 1.)

Cleaning and Disinfecting


Please refer to Warnings, Cautions, and Step by Step procedures outlined in the FUJIFILM SonoSite
“Cleaning and Disinfecting X-Porte Products” guide (P16529)

Chapter 7: Maintenance 53
54 Chapter 7: Maintenance
Chapter 8: Performance Testing

Overview
WARNING: Critical Test Function — A failure of the system functions tested in this section could affect
safety or effectiveness of the system adversely. While performing the steps in this section,
verify that the images on the system display and on the external monitor are acceptable.

To obtain 2D images, FUJIFILM SonoSite recommends using the Gammex 403GS Soft Tissue Phantom or
the Gammex 413A Multipurpose Phantom. A .7db/cm phantom is recommend but not required.
Some features and capabilities are optional and therefore may be unavailable to test.

Recommend Test Equipment


• FUJIFILM SonoSite ultrasound system under test
• C60xp/5-2 MHz transducer
• P21xp/5-1 MHz transducer
• Gammex 403 GS Multipurpose Phantom, 413A Soft Tissue Phantom, or equivalent.
• Acoustic gel

Setting Up Performance Tests

Set up 1 Attach the C60xp/5-2 MHz transducer to the system.


Performance 2 Select Gen for optimization and OB for exam type.
Tests 3 Couple the transducer to the phantom, adjusting gain settings and transducer for a
proper phantom image (e.g., pins are high-level echoes positioned in straight lines;
cysts are sonolucent, X-Portes are sharp, and graphite particles of the phantom are
mid-grays).

Chapter 8: Performance Testing 55


Basic Operational Tests

Basic System 1 Verify that the correct transducer name appears in the upper right corner of the system
Operation display.
Tests 2 Verify proper date and time.
3 Verify that the scan plane orientation mark in the image located near the skinline
corresponds to element #1 on the transducer. To test, put your finger on the probe and
run it across the transducer face. Your finger touching the transducer face should
appear at the orientation mark on the display image format.
4 Verify that all of the keyboard keys are functional. Verify that all controls operate
smoothly over their full range and that the system responds properly.
5 Verify that as the Gain controls are increased and decreased, there is a corresponding
increase and decrease in echo intensity.
6 Capture a Cineloop buffer. Exercise the Cineloop controls and verify proper operation.
7 Verify the airflow from the fan vents on the rear of the Engine flow outward.

2D Performance Tests

2D Performance / Image Quality

Test 2D 1 Use a C60xp/5-2 MHz transducer in 2D mode.


Performance 2 Adjust the position of the C60xp/5-2 MHz transducer on the phantom.
and Image 3 With the array pointing down and the orientation mark to the operator’s left, element
Quality #1 corresponds with the left side of the array.
4 Use the 2D system controls to obtain a clear image that shows both the horizontal and
vertical rows of pins.
5 Verify that the ultrasound image appears uniform in both the axial and lateral direction,
with no dropouts or intensity variations.
6 Verify that the cystic structure at the focal zone is clearly differentiated from the
surrounding tissue and is echo-free, while solid tissue with numerous echo sources,
appears solid.
7 Press the Freeze key and then save the image. Slide the Freeze Unlock key to return
to live imaging.

56 Chapter 8: Performance Testing


Axial Measurement Accuracy
Note: Measurements must be performed while the image is frozen.

Set Up Axial 1 Acquire the image.


Measurement 2 Press the Freeze key.
Accuracy 3 Press the Caliper key. The caliper appears on the image display. (See the X-Porte
Ultrasound System User Guide, if necessary, for caliper operation.)
4 Use the touchpad to position one of the calipers.
5 Press the Select key to fix the caliper and enable the other caliper.
6 Use the touchpad to move the other caliper. The results update as you move the
caliper, and the measurement is complete when you finish moving the calipers. (Press
the Select key to alternate the active caliper, and adjust the measurement with the
touchpad.)
Test Axial 1 Measure the distance, center to center, of any two pins that are 5-12 cm apart
Measurement vertically.
Accuracy 2 Verify that the distance measured is within the tolerance listed in Table 8.1.

Lateral Measurement Accuracy

Set Up Lateral Perform “Set Up Axial Measurement Accuracy” on page 56.


Measurement
Accuracy
Test Lateral 1 Measure the distance, center to center, of any two pins that are 4-10 cm apart
Measurement horizontally.
Accuracy 2 Verify that the distance measured is within the tolerance listed in Table 8.1.
3 Slide the Freeze Unlock to return the system to live 2D mode.

Table 8.1: System Measurement Accuracy

Measurements Tolerance

Axial Distance +/- 2%

Lateral Distance +/- 2%

Chapter 8: Performance Testing 57


Penetration
The penetration measurement is an integral part of the quality assurance program. Penetration is defined
as the deepest depth at which an ultrasound system can provide adequate image quality of small
anatomical structures.
Penetration measurements should be performed and the results retained for comparison to future
measurements. Penetration measurements should remain fairly consistent over time assuming use of the
same system settings and scanhead. Degradation of the penetration measurement in excess of 1cm may
indicate a transducer or system electronics issue.
Loss of measured penetration may also be caused by degradation (dessication) of the ultrasound phantom.
Ultrasound phantoms used for penetration measurements must also be part of a quality assurance program
to maintain their integrity. Follow all of the phantom manufacturer recommendations for use, storage, and
maintenance of the phantom.

Test 1 Use the same scanhead and system settings as previous measurements if possible.
Penetration 2 Adjust the system controls to obtain a clear image that shows the limits of echo
penetration.
3 Press the Freeze key and then save the image.
4 Measure from the center of the skinline to the deepest vertical position—where the
scatter echoes start to break up and tissue definition is lost.
5 Record and retain the results for future reference. Scanhead type and system settings
(exam type, depth, resolution mode, etc.) should also be recorded to ensure proper
comparison with future tests.
6 Slide the Freeze Unlock to return to live imaging.

Additional Performance Tests

Color Doppler (Color)

Test Color 1 Connect any transducer.


2 Press the Color key. “Color” should be annotated in the top left corner of the display.
3 A Region of Interest (ROI) box is displayed on top of the grayscale image. Use the
Touch Control Panel or Desktop System mouse to move the Color ROI. Verify that the
ROI moves to the new position on the display.
4 Adjust the Depth control for minimum depth in the image.
5 Adjust the Gain control so that color speckles just appear inside the ROI box.
6 Gently tap the face of the transducer and observe that the ROI box fills with color
information.
7 Press the Freeze key and then save the image. Slide the Freeze Unlock to return to
live imaging.

58 Chapter 8: Performance Testing


Color Power Doppler (CPD)

Test CPD 1 Connect any transducer.


2 Press the Color key. A Region of Interest (ROI) box is displayed on top of the
grayscale image.
3 Press the Color softkey to switch to CPD. “CPD” should be annotated in the lower right
corner of the display.
4 Adjust the Depth control for minimum depth in the image.
5 Adjust the Gain control so that color speckles just appear inside the ROI box.
6 Gently tap the face of the transducer and observe that the ROI box fills with color
information.

M Mode Imaging

Test M Mode 1 Attach a C60xp transducer and acquire an image.


Imaging 2 Press the M Mode key for the M Mode sample line.
3 Position the M Mode sample line over the image using the Touch Control Panel or the
Desktop System mouse.
4 Active the Start key to the right of the M Mode key to turn on M Mode.
5 Select the desired sweep speed from the on-screen menu (slow, med, or fast). The
on-screen menu will show the selected sweep speed.
6 Press the Freeze key to freeze the image. Save the image. Slide the Freeze Unlock
to return to live imaging.
7 Press the 2D key to return to 2D imaging.

Tissue Harmonic Imaging

Test THI 1 Attach the C60xp transducer and acquire an image.


Imaging 2 Set the depth to maximum and note the depth at which echo information is lost.
3 Press the THI key on the control panel so it displays THI on the display. Tissue
Harmonic Imaging in now active.
4 Observe a decrease in dot size and a significant loss in penetration due to the higher
frequency. Image resolution increases.
5 Press the Freeze key and then save the image. Slide the Freeze Unlock to return to
live imaging.
6 Press the THI key again to turn off Tissue Harmonic Imaging.

Chapter 8: Performance Testing 59


Pulsed Wave (PW) Doppler Imaging

Test PW 1 Attach the P21xp transducer.


Doppler 2 Press the Doppler key for the Doppler sample gate.
Imaging 3 Press the Doppler key again for the Doppler spectral trace.
4 Place a large drop of ultrasound gel on the transducer lens.
5 Adjust the Gain control as necessary and then gently tap the top of the gel and
observe a reflection on the spectral trace and the sound from the speakers.
6 Press the Freeze key and then save the image. Slide the Freeze Unlock to return to
live imaging.
7 Press the 2D key to return to 2D imaging.

Continuous Wave (CW) Doppler Imaging

Test CW 1 Attach the P21xp transducer.


Doppler 2 Press the Patient key.
Imaging 3 Select the Cardiac exam type.
4 Press the Done softkey.
5 Press the Doppler key for the Doppler sample gate.
6 Press the PW softkey to switch to CW Mode.
7 Press the Doppler key again for the Doppler spectral trace.
8 Place a large drop of ultrasound gel on the transducer lens.
9 Adjust the Gain control as necessary and then gently tap the top of the gel and
observe a reflection on the spectral trace and the sound from the speakers.
10 Press the Freeze key and then save the image. Slide the Freeze Unlock to return to
live imaging.
11 Press the 2D key to return to 2D imaging.

Image Quality Verification Test/Livescan


• Products with replaced subassemblies, or products that have been otherwise disassembled, must
undergo an Image Quality Verification Test/Livescan.
• The Image Quality Verification Test/Livescan should be performed after successfully completing all
applicable performance tests listed prior in this chapter.
• The test is completed before returning the system to service.
• A certified sonographer must perform the test.
• The Livescan test performed is at the discretion of the Sonographer and will represent their acceptance
of a successful service event.
• Review all saved images and verify that the images are displayed properly.

60 Chapter 8: Performance Testing


Printer
The printer test is an optional test that requires the Sony UP-D987MD USB Black & White Printer (P16193)
to be connected to the system under test. Skip this test if a printer is not available.

Test Printer 1 Press the print button and verify that the printer begins to print an image.
Operation 2 Verify the proper content of the printed image.

Chapter 8: Performance Testing 61


62 Chapter 8: Performance Testing
Appendix A: Replacement Parts

The following tables contain all the field-replaceable parts for the X-Porte ultrasound system. Quantities are
one unless otherwise noted. Please refer to Figure A.1 on page 62 to identify the major components of the
two X-Porte System variants which are the Desktop and Stand systems.

Desktop System

1. Clinical Monitor Display


1 2. Monitor EPS (not shown)
3. system EPS (power supply)
4. system Assembly
5. Mouse

2 3

Stand System

Figure A.1 X-Porte Major Components

Appendix A: Replacement Parts 63


Clinical Monitor

Figure A.2 Clinical Monitor Display

Table A.1: Clinical Monitor Display

Part Number Description

P14627-N Service Assembly, Display, Clinical Monitor, X-Porte

P14627-W Warranty Service Assembly, Display, Clinical Monitor, X-Porte

V14627-N Vet Service Assembly, Display, Clinical Monitor, X-Porte

V14627-W Vet Warranty Service Assembly, Display, Clinical Monitor, X-Porte

Table A.2: Clinical Monitor Display Associated Hardware

Part Number Quantity Description

P16380 1 HDMI Wire Bracket, Clinical Monitor

P16639 1 HDMI Wire Clamp, Clinical Monitor

P15551 2 M3-.5 x 8mm Screw, Pan Head Phillips (used to mount P16380 to
P14627)

P16390 1 Cable Door, Monitor Mount

P16859 1 Mount, Clinical Monitor

P16369 4 M4-.7 x 12 Black Screw, Short (used to mount P14627 to P16859)

P16371 2 M4-.7 x 35 Black Screw, Long (used to mount P14627 to P16859)

64 Appendix A: Replacement Parts


Table A.2: Clinical Monitor Display Associated Hardware

Part Number Quantity Description

P16773-N 1 Service Cable Assembly, Clinical Monitor Stalk Cable Set, M2

Touch Control Panel

Figure A.3 Touch Control Panel

Table A.3: Touch Control Panel

Part Number Description

P14027-N Service Assembly, Control Panel, X-Porte

P14027-W Warranty Service Assembly, Control Panel, X-Porte

V14027-N Vet Service Assembly, Control Panel, X-Porte

V14027-W Vet Warranty Service Assembly, Control Panel, X-Porte

Table A.4: Touch Control Panel Associated Hardware

Part Number Quantity Description

P16374 4 M4 External Tooth Washer, Silver (used with P16375 to mount P14027 to
Stand)

P16375 4 M4 x 12, Socket Cap Screw, Black (used with P16374 to mount P14027 to
Stand)

Appendix A: Replacement Parts 65


Dock

P16894

P17671
P16639
P18338
P16894 (x2)

Figure A.4 Dock

Table A.5: Dock

Part Number Description

P15164-N Service Assembly, Dock with DVR, X-Porte

P15164-W Warranty Service Assembly, Dock with DVR, X-Porte

V15164-N Vet Service Assembly, Dock with DVR, X-Porte

V15164-W Vet Warranty Service Assembly, Dock with DVR, X-Porte

Table A.6: Dock Associated Hardware

Part Number Quantity Description

P16355 4 M4 Washer, Stainless Steel

P16356 4 M4 Flathead Phillips Screw, Stainless Steel

P16639 1 **HDMI Wire Clamp, Clinical Monitor (used with P16894)

P16894 1 **HDMI Cable Bracket, Dock (used with P16639)

P17671 1 **Retainer, Wire, Cable Connector, M2 Control Panel

P18338 2 **Dock Stabilization Foot, PORON, M2

Note: **Recent production change to the Dock Assembly. Replacement Assembly may require these
additional parts to be installed to the Dock Assembly prior to installation in the X-Porte Stand.

66 Appendix A: Replacement Parts


“Ultrasound” Engine

Figure A.5 Engine

Table A.7: “Ultrasound” Engine

Part Number Description

P16055-N Service Assembly, X-Porte Engine, CW

P16055-W Warranty Service Assembly, X-Porte Engine, CW

V16055-N Vet Service Assembly, X-Porte Engine, CW

V16055-W Vet Warranty Service Assembly, X-Porte Engine, CW

Appendix A: Replacement Parts 67


Stand Power Management Unit (SPMU & Batteries)

*Batteries Only Ordered in Sets of 3


(All 3 Batteries Must Be Replaced as a Set)

Figure A.6 SPMU & Batteries

Table A.8: Stand Power Module Unit (SPMU & Batteries)

Part Number Description

P14522-N Service Assembly, Stand Power Management Unit, X-Porte

P14522-W Warranty Service Assembly, Stand Power Management Unit, X-Porte

V14522-N Vet Service Assembly, Stand Power Management Unit, X-Porte

V14522-W Vet Warranty Service Assembly, Stand Power Management Unit,


X-Porte

P14586 Battery, X-Porte Stand (Set of Qty. Three)

68 Appendix A: Replacement Parts


Triple Transducer Connect (TTC)

Figure A.7 Triple Transducer Connect (TTC)

Table A.9: Triple Transducer Connect (TTC)

Part Number Description

P14644-N Service Assembly, TTC Assembly, X-Porte

P14644-W Warranty Service Assembly, TTC Assembly, X-Porte

V14644-N Vet Service Assembly, TTC Assembly, X-Porte

V14644-W Vet Warranty Service Assembly, TTC Assembly, X-Porte

Appendix A: Replacement Parts 69


Clinical Monitor Hardware
Item 2 Item 3
Item 4

Item 6
Item 1
Item 5

Figure A.8 Clinical Monitor Hardware

Table A.10: Clinical Monitor Hardware

Item Part Number Description

1 P16390 Cable Door, Monitor Mount

2 P16371 M4-.7 x 35 Black Screw (x2)

3 P16369 M4-.7 x 12 Black Screw (x4)

4 P16380 HDMI Wire Bracket, Clinical Monitor

5 P15551 M3-.5 x 8mm Screw, Pan Head Phillips (x2)

6 P16639 HDMI Wire Clamp, Clinical Monitor

Figure 6.2 P16859 Mount, Clinical Display

Control Panel Hardware

Item 1

Item 2

Figure A.9 Control Panel Hardware

Table A.11: Control Panel Hardware

Item Part Number Description

1 P16375 M4 x 12, Socket Cap Screw, Black (x4)

2 P16374 M4 External Tooth Washer, Silver (x4)

70 Appendix A: Replacement Parts


Dock Hardware

Item 2

Item 1

Figure A.10 Dock Hardware

Table A.12: Dock Hardware

Item Part Number Description

1 P16355 M4 Washer, Stainless (x4)

2 P16355 M4 Flathead Phillips Screw (x4)

Figure A.4 P16639 HDMI Wire Clamp, Clinical Monitor

Figure A.4 P16894 HDMI Cable Bracket, Dock

Figure A.4 P17671 Retainer, Wire, Cable Connector, M2 Control Panel

Figure A.4 P18338 Dock Stabilization Foot, PORON, M2

Stand Hardware Parts


Item 1

Item 2

Figure A.11 Stand Hardware Parts

Table A.13: Stand Part

Item Part Number Description

1 P16733 X-Porte, Stand, Cable Baffle Plate

2 P16799 X-Porte, Stand, Screw, M4x5 (x3)

Figure 6.12 P16382 Screw, M5 X 8, Pan Head Phillips (x6) - (Bottom Stand Plate)

Figure 6.12 P13629 Screw, Pan Head, M5 X 20mm (x2) - (Power Park Stand Module)

Appendix A: Replacement Parts 71


Optional Equipment

PowerPark

(P12834) Assembly, PowerPark Dock Module

(P12822) Assembly, PowerPark Stand Module

*Note: Shown with the H-Universal Stand

Figure A.12 PowerPark

Table A.14: PowerPark

Part Number Description

P12822 Assembly, PowerPark Stand Module (Mounts to Stand)

P12834 Assembly, PowerPark Dock Module (Floor Portion Connected to AC


Source)

P13629 Screw, Pan Head M5 x 20mm, Steel) (Mounts P12822 to Stand Bottom
Plate)

72 Appendix A: Replacement Parts


Sony UP-D897MD B&W Printer

Figure A.13 Sony UP-D897MD B&W Printer

Table A.15: Sony UP-D897MD B&W Printer & Hardware

Part Number Description

P16193-N Service Assembly, B&W Printer, X-Porte

P16193-W Warranty Service Assembly, B&W Printer, X-Porte

V16193-N Vet Service Assembly, B&W Printer, X-Porte

V16193-W Vet Warranty Service Assembly, B&W Printer, X-Porte

P15551 Screw, Pan Head, Phillips, M3-.5 x 8mm, Steel (x4)

P15333 Cable Assembly, USB, Dock To Printer, M2

P15774 Cable, M2, AC, SPMU To Printer

Appendix A: Replacement Parts 73


Desktop System Components

EPS for Desktop System (Engine)

Figure A.14 EPS for Desktop System (Engine)

Table A.16: EPS for Desktop System (Engine)

Part Number Description

P14521 AC Power Supply, X-Porte, 150 W

EPS for Desktop Clinical Monitor Display

Figure A.15 EPS for Desktop Clinical Monitor Display

Table A.17: EPS for Desktop Clinical Monitor Display

Part Number Description

P16776 Power Supply, 65W, 24V, X-Porte, Clinical Monitor

74 Appendix A: Replacement Parts


Other Desktop Components

Item 1
Item 2

Item 3

Figure A.16 Other Components

Table A.18: Other Components

Item Part Number Description

1 P02296 Transformer, Isolation

2 P16786 Pedestal Mount, X-Porte Monitor, 75 mm X 75 mm VESA

3 P14451 Mouse, Optical, 5 Button, Black, Washable, 6 ft. Cable

Not Shown P15334 Cable Assembly, HDMI, Dock To Clinical Monitor, M2

Not Shown P16967 USB Memory Stick, 16GB, X-Porte (Used for both Desktop & Stand
Systems)

Ordering Replacement Parts


To order parts, contact FUJIFILM SonoSite Technical Support as indicated in “Contact Information” on
page 1.

Appendix A: Replacement Parts 75


76 Appendix A: Replacement Parts
Appendix B: Service Event Reporting

The Service Event Report provides information about product failures to the manufacturer and to authorized
service facilities, which provide approved warranty services for FUJIFILM SonoSite products. For all repairs
completed, complete the form and email a copy of it to [email protected] or mail to the following
address:
FUJIFILM SonoSite, Inc.
Technical Support
21919 30th Drive SE
Bothell, Washington 98021
USA

To contact FUJIFILM SonoSite Technical Support, see“Contact Information” on page 1.

Appendix B: Service Event Reporting 77


Service Event Report Form

Service Event Report Choose a Service Type


Choose a Part Status
Instructions on reverse

Service Type (check one) Parts Status (check one) For SonoSite Use O
No parts necessary for this repair. Service
† Out of Box Failure † Event Report for your information. Service Request
I need parts for this repair (list the parts below
† Warranty Service † and attach Purchase Order) Order Number
I need parts to replenish my stock (list the
† Out of Warranty Service † parts used below and attach Purchase Order) RMA Number
Will not replenish stock. Please give me a
† RMA for the return of the faulty parts. Work Order
Enter complete
No parts necessary. Please issue a RMA for
contact information † repair at SonoSite.

Service Provider
Name: Provider Reference:
Company: Date Reported:
Address:
Phone Number: Fax Number:
E-mail address:
Enter product information for
Device Description the system being repaired
Ref Number: Serial Number:
Name: Lot Number:
ARM/SHDB Version: Configuration:

Problem Found
Put the system configuration here.
You can find this on the System
Enter DETAILED problem Information Page
Service Performed description here.

Performed By: Date:

Parts Removed
Part Name Part Number Serial Number Lot Number Rev Replace

Enter details for parts being Enter details for parts being
removed from the system. installed into the system.
Parts Installed
Part Name Part Number Serial Number Lot Number Rev Replace

Tests Performed (attach test data)


Test: Test:

78 Appendix B: Service Event Reporting


Service Event Report Instructions

Instructions for completing the Service Event Report


Sections highlighted in yellow must be completed for SonoSite to accept the Service Event Report. If additional information is
required for certain circumstances you will be advised.
Forward the completed form to:
Email: [email protected]
Fax: +1-425-951-6700
Service Type
x Out of Box Failure: the item has arrived from SonoSite with failures.
x Warranty Service: the item has failed after arrival and is covered by either the included warranty or a valid extended
warranty.
x Out of Warranty Service: the item has failed and it is no longer covered by a warranty.

Parts Status
x Check One.

Service Provider
x Name: the name of the technician performing the work.
x Provider Reference: a unique number used by the Provider to track Service Event Reports. Any format is acceptable.
x Company: the name of the Distributor or authorized repair facility.
x Address: the address replacement parts will be shipped to.
x Date Reported: the date the failure was reported to SonoSite.
x Phone Number: the phone number to contact the service technician.
x Fax Number: the fax number to contact the service technician.
x Email Address: the email address to contact the service technician.

Device Description:
x Name: the description of the failed product.
x Ref Number: the reference number from the part number label of the failed product.
x Serial Number: the serial number from the part number label of the failed product.
x Lot Number: if applicable, the Lot Number from the device identification label.
x ARM/SHDB Version: the software level of the failed device. Typically found on the system information screen.
x Configuration: for configurable devices, the optional features enabled.

Event Description
x A description of the problem in the words of the user. Typically what the user reports to the repair facility.

Diagnosis
x A description of what the repair technician found. Include a list of the suspect parts.

Service Performed
x A description of the work performed to repair the system. Typically only completed if it is repaired from stock repair parts.

Parts Removed
x Part Name: the name of the failed/suspect part to be replaced.
x Part Number: the part number of the failed/suspect part.
x Serial Number: the serial number from the failed/suspect part.
x Lot Number: the lot number if applicable.
x Rev: the revision of the failed/suspect part if available.
x Replaced By: the person replacing the part.

Parts Installed
x The same information as the Parts Removed except from the parts installed if work has already been performed. If you are
waiting for parts to be ordered, leave this section blank.

Tests Performed
x The results of any testing performed, if testing has already been performed.

Appendix B: Service Event Reporting 79


Returning Products to FUJIFILM SonoSite
You will be asked to provide the following information:
• Contact name and phone number
• Product name
• Serial number
• Description of the problem

Shipping Instructions
Please contact FUJIFILM SonoSite to get a return material authorization number (RMA). Contact FUJIFILM
SonoSite before returning any product.
The shipping address for all returned products is:
FUJIFILM SonoSite, Inc.
Attn: Technical Support RMA ___________________
21919 30th Drive SE
Bothell, Washington 98021
USA

80 Appendix B: Service Event Reporting


INDEX
Numerics
2D performance tests
axial measurement accuracy 55
image quality 55
lateral measurement accuracy 55
penetration 56
A
assert code 24
assistance, customer 1
B
battery
safety 7
specifications 4
C
cautions, definition 2
E
electrical
safety 5
specifications 3
equipment safety 7
error message 6
F
Failure Reporting 75
H
humidity limits 4
I
image
quality verification test 58
review 59
M
main PCBA
failures 24
P
performance tests
2D 54
CPD 57
CW 58
M-Mode 57
overview 53
printer 59
PW 58
THI 57
Velocity Color 56
periodic maintenance 51
pressure limits 4
printer
test 59
Product Failures 75
R
replacement parts

81
list 61
ordering 73
Return Material Authorization number (RMA) 78
Returning Products 78
S
safety
battery 7
electrical 5
equipment 7
Service Event Report 75
Shipping Instructions 78
shipping specifications 4
storage specifications
equipment 4
subassembly replacement 23
system
measurement accuracy 55
overview 9
T
temperature limits 4
theory of operation 9
U
user guide, conventions used 2
W
warnings, definition 2

82
83
P16620-01
*P16620-01*

84

You might also like