Neuviz Ace Slice CT Scanner
Neuviz Ace Slice CT Scanner
4.Patient Table
Surview
Length: 1500 mm;
Views: A.P., Lateral, Dual;
Real-time surview: Yes.
Sequence Acquisition
Sequence Acquisition Modes:
32x0.8 mm, 16x0.8 mm, 8x0.8 mm, 4x0.8 mm, 2x0.5 mm;
Reconstruction Slice Thickness:
0.5 mm, 0.8 mm, 1.0 mm, 1.6 mm, 3.2 mm, 6.4 mm, 12.8
mm;
Max. Scan Length: 1570 mm.
Spiral Acquisition
Spiral Acquisition Modes:
32x0.8 mm, 16x0.8 mm, 8x0.8 mm, 4x0.8 mm, 2x0.5 mm;
Reconstruction Slice Thickness:
0.8 mm, 1.0 mm, 1.25 mm, 1.5 mm, 2.0 mm, 3.0 mm, 4.0
mm, 5.0 mm, 10.0 mm;
Max. Scan Length: 1540 mm;
Pitch: 0.3-1.5(continuous).
8.Image Reconstruction 2.SAS
Real-time Display: real-time image display during Spiral Auto Start integrates the injector with the
spiral acquisition; scanner, allowing the technologist to monitor the
Recon Field: 50445 mm; contrast injection to check for extravasation and to
Recon Time: up to 20 images/s with full cone beam initiate and stop the scan (with the pre-determined
reconstruction; delay) while in the scan room.
Recon Matrix: 512x512, 768x768,1024x1024;
HU Scale: 32,768 to +32,767.
3.MAR+ (Metal Artifact Reduction)
9.CINE Display
Changes the algorithms in the abnormal area with
high pixel value by creating the human mode, which
Display of image sequences automatic or interactive
removes the highlighted artifacts and returns it to the
with mouse control;
real tissue structure.
Max. Image Rate: 30 frames/s.
10.Filming
4.Real-time MPR
Film application is mainly used for image receipt,
image viewing and management, layout setting, print Real-time helical reconstruction mode makes it
preview and print output. possible to observe the images being scanned in real
time. This mode shows any shift in the slice position in
11.Image Transfer/Networking real time and allows the operator to check the scan
field on the image, contrast study timing, patient body
Exam can be selected and moved between NeuViz motion, etc. The patient can therefore be released
ACE SP and any imaging system supporting the immediately after scanning.
DICOM 3.0 standard;
DICOM Storage (Send/Receive);
DICOM Query/Retrieve; 5.CCT *
DICOM Basic print;
DICOM Get Worklist (HIS/RIS); CCT is a scanning mode that allows the physician to
DICOM MPPS; perform extended, low-dosescans while performing a
DICOM Storage Commitment; biopsy. You can control the scan by pressing the foot-
DICOM Viewer on CD. pedal switch in the scanning room. The resulting
images display on a remote monitor in the scanning
room,providing near-real-time visual feedback during
Software for Application the biopsy.
1.Bolus Tracking
6.Prism Imaging*
Bolus Tracking is an automated injection planning
technique that permits the user to monitor actual Prism Imaging allows back to back axial and helical
contrast enhancement and initiate scanning at a pre- acquisition of the same anatomy at two different X-ray
determined enhancement level. Combine with the energies (kVps). To further improve registration
SAS option for full automation and efficacy. accuracy, patient immobilization may be utilized.
7.Low Dose Solution
60kV: 60kV offers clinical breakthroughs in low dose scanning with the most advanced image reconstruction
algorithm. Lower radiation dose is achieved without sacrificing image quality.
240 Degree Exposure: Controlled by foot pedal, 240 degree exposure prevents direct exposure to operators
during interventional or biopsy procedures, thereby decreasing operational risks, and giving more care to
medical personnel.
Organ Safe: Organs such as eyes, thyroid, thymus, breast, small intestine and gonads are more sensitive to
radiation. Sensitive organs are usually covered with lead apron. However, when sensitive organs are included
in the scanning range, Organ-Safe function can reduce radiation dose to sensitive areas with unaffected image
quality.
Pediatric Protocols: Children are 10 times more sensitive to radiation effects than adults. The oncogenic effect of
radiation may have a long latent period (up to decades). Dose is relatively higher for a child's smaller cross-
section compared to that of an adult. Age and weight-based infant and pediatric protocols to deliver the best
clinical results with minimal dose.
New Detector Design: Compared to conventional detectors, modularization detector design with new material
increases the signal-to-noise ratio, providing an ideal afterglow. Compact design reduces electron-noise while
increases geometric efficiency. The new detector design employs Matric 3D shield to reduce scattering effect,
largely eliminating noise and objectively improving image quality.
ClearView: Iterative processing in both projection and image space delivering clear image even with low dose.
Dose Check: Multiple built-in checks to ensure that scanning dose does not exceed preset values. All preset
dose could be re-evaluated before scanning based on estimated CTDI and DLP.
3D Dose Modulation: A low-dose technology based on different patient body sizes, anatomy, tube exposure
angles and scanning positions. Automatically recommending the optimal mAs with the lowest possible dose for
guaranteed image quality.
Clinical Applications
3D visualization software With the help of contrast agent Vessel Analysis (VA) offers a
provides unique simultaneous or air, the virtual endoscope can set of tools for general vascular
visualization of vasculature, soft display the normal anatomical analysis including bone
tissue and bone. Support structure and abnormal removal, vessel extraction and
viewing through and beyond pathological changes in the measurements, with various
surrounding structures. cavity of organs such as large review modes, such as MIP or
blood vessels, trachea, colon VR.
and other organs in a roaming
manner. It is safe to operate and
distinguish.
Clinical Applications
Tumor
Tu Assessment* Prism Viewer* Nerve System DSA*
Evaluates the tumor Prism Viewer allows users to Nerve System DSA is used to
characteristics, whether it’s view images of 101 energy remove bone structure in Nerve
benign or malignant, provides levels with a variety of System scanning. The
additional information including parameters and visual tools to structure of Nerve system
location, quantity and size as assist in accurate lesion blood vessel can be displayed
reference for diagnosis, detection. clearly and intuitively through
surgery planning, and/or adaptive registration with or
tracking treatment response for without contrast agent
prognosis. sequence, one-key bone-
removing and various display
modes including VR, MIP.
Power requirements:
Power capacity: 50 KVA;
Input voltage: 380/400 VAC;
Voltage variation: ±10%; Min. Area of Room:
3-phase imbalance: less than 5%; Scanning Room: 4200 mm x 3800 mm;
Frequency: 50/60 Hz ± 1 Hz; Operating Room: 2000 mm x 3800 mm.
Ground resistance: ≤4 Ω (specialized grounding),
≤1 Ω (grounding system).
Environment requirements