RF Test Set for Telecom Engineers
RF Test Set for Telecom Engineers
com
Agilent 8920B
RF Communications Test Set
Technical Specifications
Test Features
• Call processing and on-call
parametric test interface.
• High-level GPIB commands
simplify call processing
programmability.
• DCCH (TIA/EIA-136) phone test
with 83206A TDMA Cellular
Adapter for 800 and 1900 MHz bands.
• Accurate digital power measure-
ments, with true-average TDMA
power measurements.
• VEE support with 8920B drivers.
• PCS (1710 to 1990 MHz) phone test
upgrade path with the 83236B PCS
interface.
8920B with 83206A TDMA Cellular Adapter
Improve throughput and quality
The 8920B is a full-function RF test set with accuracy,
speed, and flexibility for testing land mobile radios, cel-
lular telephones, and other communications systems
while improving throughput and quality in manufacturing.
3
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Spectral Purity Center Frequency Accuracy in DC FM Mode:
(External source impedance <1 kW): ±500 Hz (after DC FM zero),
Spurious Signals:
typically ±50 Hz.
Output port signal amplitude conditions
Fixed Attenuator to meet spurious specifications: output Supplemental Characteristics
Configuration signal must be ≤ to below value Ext. Mod. Input Impedance: 600 W nominal
Resolution: 50 Hz for <10 kHz deviation
Elect. Variable Duplex Port RF IN/OUT Port 500 Hz for ≥10 kHz deviation
4
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Option 055 (mechanical variable attenuator): Option 016: 2 dBm to 50 dBm (1.6 mW to 100 W continuous)
DUPLEX OUT: +4.0 dBm to –127 dBm 51 dBm or (125 W for 10 seconds per minute)
RF IN/OUT range by option:
Ant In: –36 dBm to +20 dBm in auto-tune mode
Standard: –22 dBm to –127 dBm
Option 016: –24 dBm to –127 dBm Accuracy: ±1 Hz plus timebase accuracy
Residual Error Vector Magnitude: <3.0% Supplemental Characteristics
Frequency Resolution: 1 Hz
Residual Phase Error: <2.0%
Residual Magnitude Error: <2.6% RF Power Measurements
Frequency Range:
IQ Origin Offset: <–30 dBc within 15 °C drift of last calibration
Standard: 10 MHz to 1 GHz
Frequency Error: ±4 Hz plus reference error Option 055 (mechanical variable attenuator): 400 kHz to 1 GHz
SWR: RF IN/OUT port <1.5:1
Audio Source Specifications
RF IN/OUT Measurement Range:
(Applicable to both internal sources) To achieve the specified accuracy when measuring power at the
RF In/Out port, the internal signal generator level must be 60 dB
Frequency below the measured power or less than –20 dBm at the Duplex
Range: dc to 25 kHz outport.
Accuracy: 0.025% of setting Standard: 1 mW to 60 W continuous (0 to 48 dBm)
Supplemental Characteristics or to 100 W for 10 seconds per minute (to 50 dBm)
Resolution: 0.1 Hz Accuracy: ±5% of reading ±0.01 mW ±1 count for temperature
range of 25 ±10 °C.
Output Level Accuracy: ±10% of reading for operating temperature range.
Range: 0.1 mVrms to 4 Vrms
Option 006: 50 µW to 10 W continuous (–13 to 40 dBm),
Maximum Output Current: 20 mA peak 15 W for 10 seconds per minute (to 42 dBm)
Frequency Range: 30 MHz to 1 GHz
Output Impedance: <1 Ω (≤1 kHz rate)
Accuracy: ±5% of reading ±0.5 µW ±1 count for tempera-
Accuracy: ±2% of setting plus resolution ture range of 25 ±10 °C.
Accuracy: ±10% of reading for operating temperature range.
Residual Distortion: 0.125%
(THD plus noise, for amplitudes >200 mVrms), for tones Option 007 (low power option): 40 µW to 2.4 W continuous
20 Hz to 25 kHz measured in an 80 kHz BW. (–14 to 34 dBm) to 4 W for 10 seconds per minute (36 dBm)
Accuracy: ±5% of reading ±400 nW ±1 count for tempera-
Supplemental Characteristics
ture range of 25 ±10 °C.
Resolution: Level ≤0.01 V: ±50 µV
Accuracy: ±10% of reading for operating temperature range.
Level ≤0.1 V: ±0.5 mV
Level ≤1 V: ±5 mV Option 016 (high power option): 1.6 mW to 100 W continuous
Level >1 V: ±50 mV (2 to 50 dBm), 125 W for 10 seconds per minute (to 51 dBm)
Offset in DC Coupled Mode: <50 mV Accuracy: ±5% of reading ±0.01 mW ±1 count for tempera-
ture range of 25 ±10 °C.
RF Analyzer Specifications Typical accuracy: ±6% of reading for operating temperature
range.
RF Frequency Measurements
Supplemental Characteristics
Measurement Range: Resolution: P >10W: 10 mW, P <10W: 10 mW,
Standard: 10 MHz to 1 GHz P <100 mW: 0.1 mW, P <10 mW: 0.01 mW,
Option 055 (mechanical variable attenuator): P <100 W: 100 mW, P >100 W: 1 W
400 kHz to 1 GHz
Level Range (with or without Option 055 mechanical variable
attenuator):
Standard: 0 to 48 dBm (1 mW to 60 W continuous)
50 dBm (100 W for 10 seconds per minute)
Option 006: –13 to 40 dBm (50 µW to 10 W continuous)
42 dBm (to 7.5 W for 10 seconds per minute)
Option 007: –14 to 34 dBm (40 µW to 2.4 W continuous)
36 dBm (2 W for 10 seconds per minute)
5
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
FM Measurement SSB Measurement
Frequency Range: Standard Frequency Range: 300 MHz to 1 GHz
Standard, Options 006, 007, and 016: 30 MHz to 1 GHz
Bandwidth (3 dB): 20 Hz to 70 kHz supplemental characteristics
(usable to 400 kHz)
Option 055: 10 MHz to 1 GHz Distortion and Noise: <3% (at 1 kHz rate in a 0.3 to 3 kHz BW)
(usable to 5 MHz)
Deviation: 20 Hz to 75 kHz TDMA Analyzer Specifications
Sensitivity: 2 µV, typically <1 µV (15 kHz IF BW, high sensitivity 8920B Opt. 001, 004, 051, 800 ( 8920B + 83206A)
mode, 0.3 to 3 kHz BW, 12 dB SINAD for fc >10 MHz) Frequency Range: 824 MHz to 894 MHz
Accuracy: ±4% of reading plus residual FM and noise contribution Input Level Range:
(20 Hz to 25 kHz rates, deviation ≤25 kHz). RF IN/OUT:
Bandwidth (3 dB): 2 Hz to 70 kHz (DC FM measurements also Standard: 1 mW to 60 W continuous (0 to 48 dBm),
available) 100 W (to 50 dBm) for 10 seconds per minute
THD+Noise: <1% distortion rms (for deviation ≥5 kHz and at a Option 006: 50 µW to 10 W continuous (–13 to 40 dBm),
rate of 1 kHz in a 0.3 to 3 kHz BW) 15 W (42 dBm) for 10 seconds per minute
Input Level Range for Specified Accuracy: Option 007: 40 µW to 2.4 W continuous (–14 to +34 dBm),
ANT IN: –50 to +14 dBm (0.01 µW to 24 mW) 4 W (36 dBm) for 10 seconds per minute
RF IN/OUT: Option 016: 1.6 mW to 100 W continuous (2 to 50 dBm),
Standard: –14 to 48 dBm (40 µW to 60 W) 125 W (+51 dBm) for 10 seconds per minute
Option 006: –22 to 40 dBm (6.3 µW to 10 W)
Option 007: –28 to 34 dBm (1.6 µW to 2.4 W) Input Frequency Setting Error: ±1 kHz
Option 016: –12 to 50 dBm (6.3 µW to 100 W) RX DSP Level Setting Range: –23 to 0 dB full scale
Residual FM and Noise: <7 Hz (0.3 to 3 kHz, rms) Residual Error Vector Magnitude: <1.3%
Supplemental Characteristics Error Vector Magnitude Measurement Accuracy: ±0.4% plus 2%
Resolution: 1 Hz, f <10 kHz; 10 Hz, f ≥10 kHz of reading
AM Measurement Residual Phase Error: <1.0°
Frequency Range: Residual Magnitude Error: <0.9%
Standard, Option 016: 10 MHz to 1 GHz (usable to 400 kHz)
Options 006, 007: 30 MHz to 1 GHz I/Q Origin Offset Accuracy: ±0.5 dB for values to –40 dBc
Supplemental Characteristics Accuracy: ±3% rms of reading (20 Hz to 15 kHz, inputs ≥1 mV rms)
Resolution: 0.05% AM for 0 to 10% AM
0.5% AM for 10 to 100% AM
7
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Functions Implemented: SH1, AH1, T6, L4, SR1, RL1, LE0, TE0, Option 006, 10 W to 50 µW Power
PP0, DC1, DT1, C4, C11, E2 Measurement Range
RS-232: Two serial ports through RJ-11 connector used for serial Option 006 was designed to optimize the measurement range for
data in and out. cellular mobile station test. This option has a 6 dB attenuator at
Baud Rates: 150, 300, 600, 1200, 2400, 4800, 9600, and the RF IN/OUT port. It increases the dynamic range of the instru-
19200 Hz ment by 2 dB and enables true-average power measurements on
TDMA signals using the 83206A (8920B Option 800). Option 006 is
Memory Card Specifications recommended for TIA/EIA-136 testing. The RF frequency range for
Card Compatibility: Single industry standard PCMCIA slot accepts Option 006 power measurement is 10 MHz to 1GHz, and the RF
Type I or Type II SRAM and ROM memory cards. source range is 30 MHz to 1 GHz RF.
Storage Capability: Allows for the storage and retrieval of IBASIC Option 007, Low-level RF Power Measurements
program parameter and results data, input of new calibration data, Allows lower-level, higher sensitivity measurements. This option
and long-term storage of Store/Recall information. reduces the maximum continuous input power of the 8920B from
Firmware Upgrades: Accepts PCMCIA flash memory cards (4 MB) 60 watts to 2.4 watts. The RF frequency for Option 007 power
to allow automatic loading of new firmware for the host CPU from measurement is 10 MHz to 1 GHz and the RF source range is 30 MHz
the front panel. Upgrade time is about two minutes. to 1 GHz.
9
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
Absolute Level Measurements Option 800, Dual-mode TDMA Cellular Adapter –
Level: Results of absolute power in watts or dBm are calculated TIA/EIA-136 DCCH, IS-54 DAMPS
by adding the ACP ratio from the spectrum analyzer to the carrier The 8920B Option 800 includes the 83206A TDMA cellular adapter
power from the input section RF power detector. attached, tested, and calibrated with the 8920B communications
Level Range: test set.
Ant In: N/A Option 801, Dual-Mode Adapter and PCS Interface
RF IN/OUT:
Standard: 1 mW to 60 W continuous For TIA/EIA 136 testing in 800 and 1900 MHz bands. This option
or to 100 W for 10 seconds per minute. is for ordering convenience. It includes the 8920B with the
83206A attached and pretested together plus the additional unit
Option 006 (cellular power option): 50 µW to 83236B PCS interface shipped together.
10 W continuous, 15 W for 10 seconds per minute.
Option 007 (low power option): 40 µW to 2.4 W General Specifications
continuous, 4.0 W for 10 seconds per minute.
Dimensions
Option 016 (high power option): 1.6 mW to 100 W
continuous, 125 W for 10 seconds per minute. 8920B: 7.5 in.H x 13 in.W x 19 in.D (188 x 330 x 456 mm)
Option 055: 8920B Option 800 (83206A attached): 9.8 in.H x 13 in.W x 19 in.D
Level Range: –137 to –19 dBm into 50 W (250 x 330 x 456 mm)
Reverse Power: 60 watts continuous, 100 watts for Weight (fully optioned)
10 seconds per minute.
8920B: 37 lb net (16.8 kg)
W/Option 006 (cellular power option): Shipping Weight: 61 lb net (27.7 kg)
Level Range: –137 to –12.5 dBm into 50 W
Reverse Power: 5.0 watts continuous, 7.5 watts for 8920B Option 800: ( 83206A attached) 50 lb net (22.7 kg)
10 seconds per minute. Shipping Weight: 80 lb net (36.3 kg)
12.5 kHz 8.5 kHz –65 dBc Storage Temperature: –55 to +75 °C
20 kHz 14 kHz –68 dBc Calibration Interval: Two years
25 kHz 16 kHz –68 dBc
30 kHz 16 kHz –68 dBc Supplemental Characteristics
60 kHz 30 kHz –65 dBc Leakage: At signal generator output frequency and level <–40 dBm,
typical 8920B leakage is <1.0 µV (2.0 mV for 8920B Option 800)
Absolute Accuracy: Equals the sum of RF power measurement induced in a resonant dipole antenna one inch from any surface
accuracy found in the RF analyzer section and the ACP typical except the rear panel. Spurious leakage levels are typically <1 µV
relative accuracy of ±2.0 dB. in a resonant dipole antenna.
10