0% found this document useful (0 votes)
22 views4 pages

Ds 400 Chart Recorder Data Sheet-1

The document describes a chart recorder device for monitoring compressed air parameters. It has inputs for various sensors and outputs data via USB, Ethernet, or RS485 interfaces. It displays readings and provides alarms. Options include integrated data logging, additional sensor inputs, and mathematics or totalizer functions.

Uploaded by

gnitij1
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
22 views4 pages

Ds 400 Chart Recorder Data Sheet-1

The document describes a chart recorder device for monitoring compressed air parameters. It has inputs for various sensors and outputs data via USB, Ethernet, or RS485 interfaces. It displays readings and provides alarms. Options include integrated data logging, additional sensor inputs, and mathematics or totalizer functions.

Uploaded by

gnitij1
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Chart recorder

DS 400 - Chart recorder


for all relevant parameters of compressed air

Standard equipment:
• USB interface
• 3.5" graphic display with touch screen
• Integrated mains unit for supply of the sensors
• 4...20 mA analogue output of all connected active sensors
• Pulse output (for total consumption) in case of flow sen-
sors
• 2 alarm relays (pot.-free changeover contacts, max. 230 V,
3 A)

Software options:
• Integrated webserver
• Mathematics calculation function
• Totaliser function

Hardware options:
• Integrated data logger
• Ethernet / RS 485 interface
• Additional sensor inputs (digital or analogueue) selectable

The sensor inputs 1 and 2 and 3 and 4 can be selected according to the required sensors (see table pages 18 to 19):

Ana-
Digital Digital Digital Digital Digital Analogue Analogue Analogue
logue

m³/h, m³ °Ctd A, kWh bar A °C °C

4…20 mA
0…20 mA
0…10 V
MOD- Pulse
BUS Pt 100
Pt 1000

Current/ Third-party Third party


Dew point Clamp-on Temperature
Flow sensor effective sensors with Pressure sensor sensor ana-
sensor ammeter sensor
power meter RS 485 logue output

14 www.cs-instruments.com
Chart recorder

TECHNICAL DS 400
Dimensions: 118 x 115 x 98 mm
IP 54 (wall housing)
92 x 92 x 75 mm
(panel mounting)
Inputs: 2 digital inputs for FA 5xx
resp. VA 5xx
Interface: USB interface
Power supply: 100…240 VAC, 50-60 Hz
Accuracy: See sensor specifica-
tions
Alarm outputs: 2 relays, (pot.-free)
Options:
Panel mounting Data logger: 100 million measured
values start/stop time,
measuring rate freely
adjustable
2 additional sen- For connection of pres-
sor inputs: sure sensors, tempera-
ture sensors, clamp-on
ammeters, third-party
sensors with 4...20 mA,
0 to 10 V, Pt 100, Pt
1000

Back view

DESCRIPTION ORDER NO.


INPUT SIGNALS
Sensor input 1+2 Sensor input 3+4
Current signals (0…20 mA/4…20 mA)
Digital (Z500 4003) -------- 0500 4000 D
internal or external
DS 400 - Chart recorder Digital (Z500 4003) Digital (Z500 4003) 0500 4000 DD power supply
with graphic display and Measuring range
touch screen Digital (Z500 4003) Analogue (Z500 4001) 0500 4000 DA
Resolution 0…20 mA
Analogue (Z500 4001) -------- 0500 4000 A Accuracy 0.0001 mA
Analogue (Z500 4001) Analogue (Z500 4001) 0500 4000 AA Input resistance ± 0.03 mA ± 0.05 %
50 Ω
Options:
Voltage signal: (0…1 V)
Option: Integrated data logger for 100 million measured values Z500 4002
Measuring range 0…1 V
Option: Integrated Ethernet and RS 485 interface Z500 4004 Resolution 0.05 mV
Option: Integrated webserver Z500 4005 Accuracy ± 0.2 mV ± 0.05 %
Input resistance 100 kΩ
Option: “Mathematics calculation function” for 4 freely selectable channels, Z500 4007
(virtual channels): addition, subtraction, division, multiplication Voltage signal (0...10 V / 30 V)
Option: “Totaliser function for analogue signals” Z500 4006 Measuring range 0…10 V
Resolution 0.5 mV
External Gateway Profibus for RS 485 interface connection Z500 3008 Accuracy ± 2 mV ± 0.05 %
Input resistance 1 MΩ
Further accessories: RTD Pt 100
CS Basic – data evaluation graphically and in tabular form - reading of the 0554 8040 Measuring range -200…850 °C
measured data via USB or Ethernet, license for 2 workstations Resolution 0.1 °C
Accuracy ± 0.2 °C (-100…400 °C)
CS Network – energy monitoring with client/server solution (max. 20 mea- 0554 8041 ± 0.3 °C (further range)
sured values of different sensors/devices)
RTD Pt 1000
CS Network – energy monitoring with client/server solution (max. 50 mea- 0554 8042 Measuring range -200…850 °C
sured values of different sensors/devices) Resolution 0.1 °C
CS Network – energy monitoring with client/server solution (max. 100 mea- 0554 8043 Accuracy ± 0.2° (-100…400 °C)
sured values of different sensors/devices)
Pulse Min pulse length 500 µs
CS Network - Energy Monitoring with Client / Server Solution 0554 8044 Measuring range frequency 0...1 kHz
(max. 200 measured values of different sensors / devices) max. 30 VDC

www.cs-instruments.com 15
Chart recorder

DS 500 / DS 400
Easy operation via touchscreen:

Configuration of flow sensor


In the menu of the DS 500 / DS 400, the flow sensor
VA 5xx can be set to the respective pipe inside diame-
ter. Furthermore, the unit, the gas type and the reference
condition can be set. The meter reading can be set to
“zero” if necessary.

Graphic view
In the graphic view all measured values are indicated as
curves. It is possible to browse back on the time axis by a
slide of the finger (without data logger maximum 24 h, with
data logger back to the start of the measurement).

Data logger
With the option „integrated data logger“ the measured values
are stored in the DS 500 / DS 400. The time interval can be
freely set. Furthermore there is the possibility to fix the starting
time and the end time of the data recording. Read-out of the
measured data via USB interface or via the optional Ethernet
interface.

Selection of the language


DS 500 / DS 400 “speaks” several languages. The desired lan-
guage can be selected via the selection button.

All relevant parameters at a glance


In addition to the flow rate in m³ / h, the DS 500 / DS 400 also
displays other parameters such as total consumption in m³ and
speed in m/s.

16 www.cs-instruments.com
Chart recorder

Web server
The new webserver with substantially extended features for
the chart recorders DS 500 and DS 400 is available with
immediate effect. Users can thereby get direct access to their
measured data worldwide (current and historic ones) and dis-
play them on their smart phone, tablet or computer.

The new webserver can be ordered as an option with each


stationary DS 500/400, but also for their mobile devices. For
using the features of the webservers, the DS 500/400 must be
set up with it’s own IP address within the corporate network.

The web server in the DS 500/400 provides a website, which


displays the measured values. This website can be accessed
from smartphones, tablets and computers via the respectively
installed browser. Advantage: This is all possible without the
installation of any new or additional software.

View of the real time measured values View of the historic measured values as a single chart
(graphic table view) (time period freely selectable)

Access authorization
Different groups with different users/passwords can be assigned to different access levels.

Starting the data logger


In case of a stopped data logger the group operator or administrator can start the data logger remotely, via the web server.

PS: The new webserver can be retrofitted to any DS 500/DS 400 already in use.