0% found this document useful (0 votes)
54 views2 pages

Refractometer Technical Data DR6000 Series en 5.7

This document provides technical specifications for several models of digital refractometers. The refractometers measure refractive index and Brix scale across defined measurement ranges with high accuracy. They feature integrated temperature control and compensation, automatic adjustment, and touchscreen control.

Uploaded by

Amro Ismail
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
54 views2 pages

Refractometer Technical Data DR6000 Series en 5.7

This document provides technical specifications for several models of digital refractometers. The refractometers measure refractive index and Brix scale across defined measurement ranges with high accuracy. They feature integrated temperature control and compensation, automatic adjustment, and touchscreen control.

Uploaded by

Amro Ismail
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

TECHNICAL DATA – DIGITAL REFRACTOMETERS

DR6000 DR6100 DR6000-T DR6100-T


Refractive index (nD)
concentration of sucrose, glucose,
SCALES fructose and invert sugar [%Brix],
User-defined

nD 1.3200–1.5800 nD 1.3200–1.7000 nD 1.3200–1.5800 nD 1.3200–1.7000


MEASUREMENT RANGE 0–95 %Brix 0–95 %Brix 0–95 %Brix 0–95 %Brix

nD ±0.0001 nD ±0.0001 nD ±0.0001


MEASUREMENT ACCURACY ±0.1 %Brix ±0.1 %Brix ±0.1 %Brix

nD 0.0001 nD 0.0001 nD 0.0001


RESOLUTION
0.1 %Brix 0.1 %Brix 0.1 %Brix

approx. 4 s
MEASUREMENT PERIOD

MEASUREMENT PRISM Sapphire

LIGHT SOURCE LED

WAVELENGTH 589 nm

TEMPERATURE CONTROL With integrated Peltier temperature control


Without a sample temperature control,
TEMP. CONTROL RANGE you can connect a thermostat for the tempe- 10–80 °C
rature control
ACCURACY OF TEMP. CONTROL ±0.1 °C

Can be activated
TEMPERATURE COMPENSATION
(ICUMSA or freely definable)

TEMPERATURE MEASUREMENT With integrated temperature sensor Pt100

TEMPERATURE MEASUREMENT RANGE 10–80 °C

TEMP. MEASUREMENT ACCURACY ±0.1 °C

TEMP. MEASUREMENT RESOLUTION 0.1 °C

METHODS A practically unlimited number of methods can be set

ADJUSTMENT Automatic (menu-guided)

CONTROL 7.0” - capacitive touch-screen, 800 x 480 Pixel

HOUSING Aluminum cast, powder-coated

INTERFACES USB, RS-232, Ethernet

OPERATING VOLTAGE 90–240 V, 47–63 Hz

POWER CONSUMPTION 25 W 40 W
(MEASUREMENT OPERATION)

POWER CONSUMPTION (MAX.) 75 W

DIMENSIONS (W X H X D) 215 mm x 150 mm x 345 mm

WEIGHT 5 kg
DR6200-T DR6300-T DR6000-TF DR6100-TF DR6200-TF DR6300-TF
Refractive index (nD) Refractive index (nD)
concentration of sucrose, glucose, concentration of sucrose, glucose,
fructose and invert sugar [%Brix], fructose and invert sugar [%Brix],
User-defined User-defined

nD 1.32000–1.58000 nD 1,32000–1,70000 nD 1.3200–1.5800 nD 1.3200–1.7000 nD 1.32000–1.58000 nD 1.32000–1.70000


0–95 %Brix 0–95 %Brix 0–95 %Brix 0–95 %Brix 0–95 %Brix 0–95 %Brix

nD ±0.00002 nD ±0.0001 nD ±0.00002


±0.02 %Brix ±0.1 %Brix ±0.02 %Brix

nD 0.00001 nD 0.0001 nD 0.00001


0.01 %Brix 0.1 %Brix 0.01 %Brix

approx. 4 s approx. 4 s

Sapphire Sapphire

LED LED

589 nm 589 nm

With integrated Peltier temperature control With integrated Peltier temperature control

10–80 °C 10–80 °C

±0.1 °C ±0.1 °C

Can be activated Can be activated


(ICUMSA or freely definable) (ICUMSA or freely definable)

With integrated temperature sensor Pt100 With integrated temperature sensor Pt100

10–80 °C 10–80 °C

±0.1 °C ±0.1 °C

0.1 °C 0.1 °C

A practically unlimited number of methods can be set A practically unlimited number of methods can be set

Automatic (menu-guided) Automatic (menu-guided)

7.0” - capacitive touch-screen, 800 x 480 Pixel 7.0” - capacitive touch-screen, 800 x 480 Pixel

Aluminum cast, powder-coated Aluminum cast, powder-coated

USB, RS-232, Ethernet USB, RS-232, Ethernet

90–240 V, 47–63 Hz 90–240 V, 47–63 Hz

40 W 45 W

75 W 75 W

215 mm x 150 mm x 345 mm 215 mm x 150 mm x 345 mm

5 kg 5.3 kg