Using NanoVNA For SWR Measurement Rev1

This document provides step-by-step instructions for using a NanoVNA device to measure standing wave ratio (SWR) of an antenna. It details initial setup of display and frequency range, calibration procedure, and taking measurements. The entire process is described from start to finish, with screenshots and explanations of menu options.

Uploaded by

Gerry Dela Cruz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
91 views4 pages

Using NanoVNA For SWR Measurement Rev1

This document provides step-by-step instructions for using a NanoVNA device to measure standing wave ratio (SWR) of an antenna. It details initial setup of display and frequency range, calibration procedure, and taking measurements. The entire process is described from start to finish, with screenshots and explanations of menu options.

Uploaded by

Gerry Dela Cruz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

12-11-2021 Using NanoVNA for SWR Measurement – Step by Step KJ7ROX

This procedure is for the use of a NanoVNA for measuring SWR of an antenna directly on the NanoVNA device. There
are different hardware versions of the NanoVNA – The one used here is called the “NanoVNA-H 4” and it has a 4 inch
touchscreen display. The firmware is version 0.5.0 (as originally installed).

The NanoVNA kit came with most of what is needed to make SWR measurements. The only additional thing I had to buy
was a jumper cable to allow me to connect the PL-259 on the end of my antenna feed line to the SMA connector on the
NanoVNA.

Reading the characteristics of a typical ham antenna only requires one of the two inputs on the NanoVNA – CH0 also
described as S11 is used (upper connection). The other connector is not used in this procedure.

Initial Display Setup:

1. Turn on the NanoVNA


Your screen may look different than
this view (right). Out-of-the-box, it should
have the same range and may show four
color traces instead of the three shown.
The next steps will show you how to make
the necessary changes.

Page |1
12-11-2021 Using NanoVNA for SWR Measurement – Step by Step KJ7ROX

2. Click anywhere on screen with the (guitar pick) stylus to open the Main Menu

a. Select “Display”
b. Select “Trace” – 4 traces will be listed initially, you can de-select or select traces.
I turn off trace #3 as it is unnecessary.
If you momentarily click on “Trace 3” it
un-highlights it and turns it off.

We need to do one more thing in this


case – All 3 traces MUST be set to “CH0”
and as you can see, my “Trace 1” is set
to “CH1”

c. Momentarily select “Trace 1” to highlight it and then click “BACK” to the previous menu and then select
“CHANNEL” so you can set Trace 1 to “CH0 – REFLECT”.

Now all three traces should be set to “CH0”

Notice that Trace 1 is set to “LOGMAG”


We need to set it to “SWR”

d. Select “Back” to go back to the previous menu

e. Select “Format”
i. Select “SWR”
ii. Select “Back”, then “Back” to the
Main Menu

Now the screen should look like the figure


on the right. (except for the curves)

Set Desired Frequency Range:

f. Now select “Stimulus” from the main menu to set the Band Range (MHz)
i. Click on “Start” and enter the lower-end of the frequency range (ex: “7M”)
You must press either “M” (Megahertz) or “k” (Kilohertz) after the appropriate number.
ii. Click on “Stop” and enter the high-end of the frequency range (ex: “7.3M”)
iii. “Back” to previous menu

Page |2
12-11-2021 Using NanoVNA for SWR Measurement – Step by Step KJ7ROX

Calibrate for the desired frequency range:

g. Select “Cal” to go to Calibration Menu


i. Select “Calibrate”
ii. Select “RESET” (very important!) in order
to delete the previous calibration constants
and replace with specific calibration
values needed for your desired frequency
range.
iii. Select “CALIBRATE” to bring up the calibration menu

1. Screw on the OPEN calibration standard and Select “OPEN”


2. Screw on the SHORT calibration standard and Select “SHORT”
3. Screw on the LOAD calibration standard and Select “LOAD”
4. Select “DONE” and you have the option of saving the Stimulus Range and Calibration in
non-volatile memory locations S1-S4 – (Not required, but a good idea)
TIP: Do NOT use S0 because factory calibration data is stored there.

3. Now you are ready to analyze your antenna.

Before connecting your feed line, always use a conductive tool (screwdriver, etc) to short between your feed
line’s center connector and the outer shield of the connector so that you remove any static charge that may have
accumulated in the feedline which might damage your sensitive NanoVNA.

Page |3
12-11-2021 Using NanoVNA for SWR Measurement – Step by Step KJ7ROX

4. Now screw your feed line onto CH0 (S11) and you should see three traces: LogMag, Smith Chart and SWR. You
can move the marker along the SWR curve and read specific SWR values at points along the frequency axis by
using the multi-position switch on the top of the NanoVNA. You can also use the “MARKER” and then “SEARCH”
feature from the main menu to locate the MAXIMUM or MINIMUM values of the curve.

5. To save the graph, you can take a picture with your phone or there is software available which will allow you to
connect the NanoVNA to your computer via USB and record data as well. I have used NanoVNA-Saver software
on my Windows10 pc connected to the NanoVNA via a USB connection. NanoVNA-Saver has additional
functions for saving data and charts to your computer and you can put away the guitar pick, but the
fundamentals are the same.

6. To analyze another band, go back up to step# 2f above and repeat the process using the desired stimulus
frequency range and you must also re-calibrate for each new range. The combination of frequency range and
calibration data can be saved in memories 1-4 for future use. I always make it a habit to recalibrate every time I
make a new measurement. Once you get the hang of it, the time spent is a small cost for reliable
measurements.