AES Installation Guide en
AES Installation Guide en
WITHOUT DRYING
AES SERIES
CLASSIC LINE
INDEX
ELECTRICAL CONNECTION
The following table shows the plug configuration according to IEC and SCHUKO international standards for most countries
of the European Union. For customers who require other plugs and other electrical configurations, please contact our
technical team at [email protected].
1 2 3 4 5 6
16 A 16 A 32 A 32 A 16 A 32 A
The following table shows the plug configuration according to the NEMA standard for the United States and other
countries. For customers requiring other plugs and other electrical configurations, please contact our technical team at
[email protected].
1 2 3 4
CONNECTIONS GRAPH
1. RS-232 PORT
2. PC/PRINTER SELECTOR
3. POWER CABLE
(AES-110 and AES-150 models)
Cable length: 2 m
ATTENTION: If the
STERILIZATION CHAMBER
DRAIN AND PURGE
OUTLET (6) is used
shortly after the end of a
4. SAFETY THERMOSTAT FOR sterilization cycle, water
THE HEATING ELEMENTS OF THE may run out at a high
STERILIZATION CHAMBER temperature.
6. STERILIZATION CHAMBER
DRAIN AND PURGE OUTLET
COMPONENT INCLUDED
In addition to the accessories chosen at the time of purchase of the autoclave (baskets, tray supports, software kit, etc.), the
following component is always included:
1 stainless steel protective cover for the heating elements, with legs to place at the
bottom of the sterilization chamber.
MODELS DIAMETER
Ø
AES-28 290 mm
AES-50 290 mm
AES-75 390 mm
AES-110 390 mm
AES-150 490 mm
Purified water is necessary for the operation of the autoclave and is added
manually in the sterilization chamber until it covers completely the heating
elements cover.
The correct position of the drainage tap while filling the sterilization chamber and
during normal work with the autoclave is .
IMPORTANT NOTE:
PURIFIED WATER
The water used to feed the autoclave must be free of contaminants and meet the
following hardness and conductivity requirements:
- Hardness: ≤ 0.02 mmol/l
- Conductivity: between 5 µS/cm and 15 µS/cm
Multiple systems can be used to obtain water with these requirements: osmotized,
demineralized, decalcified, distilled water, etc*.
*Note: Take into account that distilled water that is too pure (conductivity less than 5 µS/cm) is not
recommended as it may cause long term corrosion on stainless steel.
DRAINAGE CONNECTION
To drain the sterilization chamber a container (drum, tray, etc.) must be placed
below the STERILIZATION CHAMBER DRAIN AND PURGE OUTLET (6), or a
hose (not included) should be connected from the outlet of the autoclave to a
drain.
Then, manually operate the drain tap to place it in position and let the water
out.
TANK-AE accessory
The TANK-AE accessory consists of a hose and a 12L bottle with a built-in tap to
collect the condensates during the purge phase, and also to drain the sterilization
chamber during autoclave cleaning operations. Especially useful in cases where there
is no access to a drainage connection near the autoclave.
Connect the supplied silicone hose (Ø5 x Ø8 mm, 1 m long) at exit 6 of the autoclave.
Position the bottle so that its base is lower than the autoclave drain outlet.
If you want to keep the bottle connected to the autoclave in order to collect
condensates during the purge phase, it is very important to remove the cap from the
condensate bottle to avoid overpressures inside the bottle.
For safety reasons, the distance between both sides of the autoclave and the wall or any other object must be 100 mm,
and between the autoclave and the rear wall must be at least 200 mm.
MODELS L L1 D H HL HD
LENGTH LENGTH DEPTH HEIGHT LOADING HEIGHT DRAIN OUTLET
with closed door with maximum door HEIGHT
opening
WARNING:
ELECTRICAL Respect the
OUTLET (3) recommended distances.
H
HL
MINIMUM
200 mm
MINIMUM
100 mm
D
L
L1
CONSIDER:
Height and position of the drain outlet.
(6)
*If you need to collect purges and empty the sterilization chamber into an
external container, usage of TANK-AE ACCESSORY is recommended (see
page 6). HD
ENVIRONMENTAL CONDITIONS
This autoclave can operate under the following maximum conditions:
• Ambient temperature: 30 ºC
• Humidity: 75%
• Altitude: 3,000 meters above sea level. Take into account that from 1,000 meters above
sea level an adjustment of the purge parameter must be made; consult with our technical
team.
CLICK!
ACCESS THE
AES SERIES
more info
VIDEO
REV 04.2023