Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 1 /22
Venu1717X Installation Guide
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 2 /22
Contents
1. Production Description...........................................................................................................3
2. Installation Requirement........................................................................................................4
2.1. Environment...................................................................................................................4
2.2. Power supply..................................................................................................................5
2.3. System Requirement...................................................................................................... 5
2.3.1. Computer Requirement....................................................................................... 5
2.4. Grid................................................................................................................................ 5
2.4.1. Grid direction...................................................................................................... 5
2.4.2. Grid Suggestion.................................................................................................. 6
3. Package Check........................................................................................................................ 6
3.1. Package List................................................................................................................... 6
3.2. Product Check................................................................................................................ 8
4. Detector Installation and Cable Connection........................................................................ 9
4.1. Cable Connection...........................................................................................................9
4.2. Software Environment Set up...................................................................................... 11
4.3. Network Configuration................................................................................................ 12
5. Correction and Calibration Template Generation.............................................................13
5.1. Gain Calibration Template Generation........................................................................ 14
5.1.1. Calibration.........................................................................................................14
5.2. Defect Correction.........................................................................................................17
5.3. Correction and Calibration Management.....................................................................19
5.4. DownLoadFile:Download templates from computer to panel..................................20
5.5. Hardware Correction and Management....................................................................... 21
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 3 /22
1. Production Description
Venu1717X is a cassette-size tethered X-ray flat panel detector based on amorphous silicon
thin-film transistor technology. It is designed to provide the high quality radiographic image
which contains an active matrix of 3072×3072 with 139um pixel pitch. The scintillator of
Venu1717X is CsI(Caesium Iodide). The technology of CsI direct growth reduces the exposure
dose and improves the image quality. Since Venu1717X supports multiple trigger modes, it
can satisfy both of the general DR system and retrofit DR system.
Item Specification
Model Venu1717X
Image Sensor a-Si (Amorphous Silicon) TFT
Scintillator CsI:Tl
Pixel Size 139um
Fill Factor 70%
Effective Array 3072x3072
Effective Area (H x V) 427mm×427mm
Spatial Resolution Min. 3.4 lp/mm
Image Transfer Gigabit Ethernet
Full Image Time 5s
Cycle Time 8s
Power Consumption 20W
Dimension (L × W × H) 460mmx460mmx15mm
Weight 4kg(without cable and control box)
X-ray Energy 40-150kV
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 4 /22
Panel protection IPX1
Software
Prep
Trigger Mode
Freesync
Inner
SID 90-180cm
2. Installation Requirement
2.1. Environment
Temperatur Temperature Atmospheri Pressure
Humidity
e change c Pressure Change
Operating 5~35℃ ≤0.5℃/min 30~80% RH ≤10 mbar/hour
700~1060m
bar
Storage 10~55℃ ≤1℃/min 10~90% RH ≤20 mbar/hour
The Venu1717X serial detectors shall operate at an altitude specified no more than
3000m.
Do not expose the equipment to high temperature and humidity, which will result
in equipment failure.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 5 /22
2.2. Power supply
Always connect the three-core power cord plug to a grounded AC power outlet.
To make it easy to disconnect the plug at any time, avoid putting any obstacles near the
outlet. Otherwise, it may not be possible to disconnect the plug in an emergency.
Be sure to ground the equipment to an indoor grounded connector. Also, be sure to connect
all the grounds for the system to a common ground.
Do not use any power source other than the one provided with this equipment.
Otherwise, fire or electric shock may be caused due to leakage.
2.3. System Requirement
2.3.1. Computer Requirement
Name Recommend
System Windows 7/8/10 64bit
CPU Intel Core i5 or more
Memory 8G or more
Storage 120G or more
Display 1280x1024 LCD
2.4. Grid
2.4.1. Grid direction
Direction of panel and grid need to be set as following picture
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 6 /22
2.4.2. Grid Suggestion
Detector Grid
Density
Scintillator Brand Grid ratio Size Material
(Lines/Inch)
CSI JPI 230 8:1, 10:1 460mm×460mm carbon fiber
GOS JPI 215 8:1, 10:1 460mm×460mm carbon fiber
CSI/GOS JPI
103 8:1, 10:1 460mm×460mm AL
Claymount
3. Package Check
Before open box, please ensure environment is fit the requirement in 2.1.
3.1. Package List
In box, it contents 3 independence boxes.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 7 /22
Detector box(Yellow)
Two Accessory boxes(Yellow)
Package list :
Number Item
1 Venu1717X Detector
2 Control Box
3 Medical Adapter
4 AC Power Cable
5 Gigabit Ethernet Cable
6 HVG Cable
7 CD
8 Syncbox(option)
10 OQC test report
11 Package list
12 OQC pass label
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 8 /22
Document in CD:
Number Document Describe
Venu1717X technical support
1 Venu1717X User Manual (CN)
document
Venu1717X technical support
2 Venu1717X User Manual
document
3 SDK Program guide
4 Detector+product type+CorrectFile Gain、Defect factory template
3.2. Product Check
Check product by package list in 3.1.
If it has sync box in order, check it in package.
Sync Box
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 9 /22
4. Detector Installation and Cable Connection
During installation, panel needs to be off power.
When connector cannot plug in, please do not force insertion. Please check direction and
needle is curved.
4.1. Cable Connection
Connect Power, HVG (if needed), and Gigabit
cables to the Control Box. Power to adaptor and
gigabit cables to computer
Connect the HVG cable to High Voltage
Generator
4.1.1 INTERFACE DEFINITION TO GENERATOR
Generator interface on control box show as following picture, it is DB15
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 10 /22
6
1 11
7
2 12 Enable_OUT_P
8
3 13
9
SY NC_IN_N 4 14 SY NC_IN_P
10
5 15
16 17
HGND HGND
GVG cable:
Pin Color Name I/O Description
1 White Reserved / Reserved(Do Not Connect)
2 Orange/White Reserved / Reserved(Do Not Connect)
3 Black Reserved / Reserved(Do Not Connect)
HVG generator signal inform FPD
4 Green Prep_IN_N IN
to start clear process
5 Gray Reserved / Reserved(Do Not Connect)
6 Brown Reserved / Reserved(Do Not Connect)
7 Brown/White Reserved / Reserve (Do Not Connect)
8 Blue Reserved / Reserved(Do Not Connect)
9 Yellow HVG_GND P Chassis ground
10 Black/White Reserved / Reserve (Do Not Connect)
11 Pink Reserved / Reserve (Do Not Connect)
FPD generator signal to HVG
12 Red Enable_OUT_P OUT indicate the clear process finished
and wait for exposure
13 Light green Reserved / Reserved(Do Not Connect)
HVG generator signal inform FPD
14 Purple Prep_IN_P IN
to start clear process
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 11 /22
15 Orange Reserved / Reserved(Do Not Connect)
Thermal
16 Shield P Earth Ground
casing
4.1.2 INTERFACE CIRCUIT OF CONTROL BOX
VDD_3V3 VDD_3V3 VISO_5V
1K FP D HV G
4 1
U13
1 8
2 VCCAVCCB 7 3 2
PREP 3 A1 B1 6 PREP_IN_N
A2 B2
3
4 5 TLP185
GND DIR PREP_IN_P
10K
1K 2N7002
1 39K
2
DGND DGND HGND HGND
VDD_3V3 VDD_3V3 VDD_3V3 VISO_5V
1 6
VDD
A
4.7K
U14 5
8 1 VO Enable_OUT_P
7 VCCBVCCA 2
6 B1 A1 3 3 C 4 HGND
OUT_EN 5 B2 A2 4 GND
DIR GND ACPL-M61L
DGND HGND
4.2. Software Environment Set up
Setup files and download url are included in SDK directory: Tools\env_setup
1. Please install Microsoft .NET Framework 4.5(Windows XP only can install V4.0 ). Download
from Microsoft web site, please.
2. Visual C++ redistributed package need to be installed: vcredist_x86_2013(or
vcredist_x64_vs2013).
3. For Windows XP, full path should be used in file “bind.txt”.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 12 /22
4.3. Network Configuration
Select wired network adapter
that connected to the
detector.
Right click the network
adapter. Then select
properties.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 13 /22
Double click IPV4 item
Default IP settings:
IP address:192.168.8.188
Subnet mask: 255.255.255.0
5. Correction and Calibration Template
Generation
IRay recommends performing correction and calibration after installation or any major change on the system
settings and hardware configuration. On the other hand, it is also recommended to do the correction and
calibration in each 6 months.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 14 /22
5.1. Gain Calibration Template Generation
5.1.1. Calibration
Before Gain template generating, make sure SID1.2m, no copper is required.
Choose “Acquire” module, do
not choose any calibration
Choose Calibrate module, choose
PreOffset, click “Start Generate
Templates”.Make sure correction
and calibration template meet the
requirements.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 15 /22
In calibration page, choose one
calibration.
Click “Create Gain”,go to Gain
calibration page, click “Start”
According to the recommended
dosage set voltage and MAS, then
exposure, after get image, it will
show current value, and inform it is
below or above expect value.
Adjust mAs, ensure current value
40000 10%
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 16 /22
Click “Accept”,stage will increase
until to 5
After 5 image, click “Generate”
After generation, it will back “task
succeed:FinishGenerationProcess”
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 17 /22
5.2. Defect Correction
Defect map will generate in factory, and save in CD. It is located in following path road:\Tools\iDetector\x64
or x32\work_dir \Venu1717X\Correct\Default.
If it needs to do defect calibration again, Before Defect template generating, make sure SID1.2m, no
copper is required,
Click “ Create Defect ” enter
defect page, click “Start”
According to the recommended
dosage set voltage and MAS, then
exposure, after get image, it will
show current value, and inform it is
below or above expect value.
Adjust mAs, ensure current value
4000 10%
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 18 /22
Click “Accept”,stage number will
increase.according to new
recommended dosage set voltage to
40kv and except value to 4000,
adjust mas,ensure current value
4000 10%
Click “Accept”,according to new
recommended dosage set voltage to
120kv and except value to 16000,
adjust mas,ensure current value
16000 10%
Click “Accept”,according to new
recommended dosage set voltage to
70kv and except value to 40000,
adjust mas,ensure current value
40000 10%
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 19 /22
After 8 image, accept button can
not click,click “Generate”
Wait Map generation, message box
shows “ Task succeed :
FinishGenerationProcess”
Note:
1 please use software post offset mode.
2. Make sure your x ray dose is right, if your dose is out of the range, IDetector will remind you to adjust the
dose. Then you can click “start creating” and try again.
5.3. Correction and Calibration Management
Panel supports correction and calibration template storage. So template in panel could be uploaded to
Workstation, and template in Workstation could also be downloaded to panel.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 20 /22
5.4. DownLoadFile:Download templates from computer to panel
Make sure firewall is closed
Click Calibrate , click “ Start
Generate Templates”
In“Mode&Files”box,click“download
to FPD”
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 21 /22
Pop up a message box , click
“ Download” , all templates under
default path road download to panel
one time
5.5. Hardware Correction and Management
click ReadStatus button,check all
templates saved in panel. (Enable
or in use is valid ; disable is
invalid);
Choose one template and click
active;
It needs to be ensure that correction
templates are download to panel
success.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.
Doc ID 072-500-05
iRay 奕瑞 Version A0
Revision Date 2019-03-21
Venu1717X Installation Guide
Page 22 /22
Click “Acquire”;
Choose“HWPostOffset”,“HWGain”,
“HWDefect” one by one; and then
exposure, check image quality.
文件包含奕瑞公司机密资料,非经奕瑞正式书面授权不得复制、打印、披露及传播该文件的任何内容。
This document contains confidential and proprietary information of iRay, any reproduction,
disclosure, or use in whole or in part is strictly prohibited, except as may be specifically
authorized by prior written agreement or permission of iRay.