TE-171 Maintenance Guided For Service
TE-171 Maintenance Guided For Service
Applicable to:
Lot No.9803
1. GENERAL
This Service Manual has been prepared for BME maintenance service of the TERUFUSION Infusion
Pump (TE-171/172).
When informed of a pump unit malfunction from a doctor/nurse, please troubleshoot
according to the following procedures.
(1) Ensure that a thorough reading of both the Instruction and Service Manuals are
carried out before carrying out any repair or maintenance work.
(2) Ensure to verify with the user if operation of the equipment has been carried out
in accordance with the instruction manual.
(3) If any fault is found to be caused by improper operation by the doctor/nurse,
correct the cause of the problem and instruct the user how to operate the
equipment properly so that recurrence of the fault can be prevented.
(4) Before starting of actual repair of the equipment , be sure to confirm the fault that
the doctor/nurse has brought to your attention . After assessing the actual state of
the fault, only then should you start any repair work.
(5) If the equipment is found malfunction even when it is operated properly, use the
troubleshooting procedure in this manual to diagnose the faulty pump.
(6) If you are unable to correct the fault by using this manual, please contact your
local TERUMO distributor or TERUMO for corrective action and repair.
(7) Disinfect the items contaminated by patients body fluids before returning to
TERUMO distributor or TERUMO. Please include all items that have been
removed from the equipment during preliminary checks upon return of equipment.
If it is necessary to send back those items that have not been disinfected, seal
them completely and attach a memorandum indicating that the contents
unsanitary and to be handled with care.
(General precautions)
(1) Do not detach connectors or printed-circuit boards while the equipment is
supplied with power.
(2) Visually check the equipment under maintenance/repair for any open circuits,
short circuits, loose connectors and damaged parts. Also, ensure that checks for
abnormal sounds or foul smell(caused by burned components) are carried out.
(3) After completion of repair work, check that all switches are set in the respective
original settings. Clean the equipment and its periphery and, then, report the
result of repair work to the BME manager.
(4) When handling semiconductors, especially MOS-ICs, take special care not to cause
electrostatic breakage, etc.
Keep your body and soldering iron in constant connection with a ground point
through a 1M resistor. ( A statically insulted type of soldering iron does not
required grounding, but it must be used only after being checked for
completeness of insulation.)
When carrying ICs and printed-circuit boards, be sure to keep them in electrically
conductive containers or wrap them in metal sheets(such as aluminum foil).
1-1
SY10-9B002
Applicable to:
Lot No.9803
(5) When an abnormality on longer exists when a printed-circuit board (CPU) has
been replaced, try re-mounting the original board to verify that the previous
abnormality still recurs. If there is not recurrence, poor contact or another fault
could exist, continue to further troubleshoot.
* In this manual, the following units are in SI (System International d' Unites):
Quantity SI
Name of unit Symbol Conversion
Force Newton N 1 f 9.80665N
Pressure Pascal Pa 1 f/ 2 98066.5Pa
1-2
SY10-9B002
Applicable to:
Lot No.9803
2. SPECIFICATIONS
Product name TERUFUSION Infusion Pump TE-171 TERUFUSION Infusion Pump TE-172
Pumping Mechanism MIDPRESS peristaltic finger system (A revolutionary system used to provide
accurate infusion without completely pressing on the tubing)
Accuracy Within 3%: Conditions Infusion set :Specified TERUFUSION Infusion sets type E for pump
Solution : Water/Saline
Test time : Second hour
Delivery rate : higher than 1.0 mL/h
Temperature : 25 C
Within 5%: Conditions Infusion set : Specified TERUFUSION Infusion sets for pump
Solution : Water/Saline
Test time : Second hour
Delivery rate : higher than 1.0 mL/h
Temperature : 25 C
Delivery Rate range 0.1 - 1200 mL/h (0.1 - 99.9 mL/h: 0.1 mL/h step)
(100 - 1200 mL/h: 1 mL/h step)
Volume Delivered range 0.0 - 9999 mL (0.0 - 99.9 mL: 0.1 mL step)
(100 - 9999 mL: 1 mL step)
Occlusion Detection 13.3 - 133.3 kPa, 100 - 1000 mmHg, 0.1 - 1.4 kgf/cm2
Pressure range The detection pressure can be adjusted in 3 levels (L, M, H) between the above
range.
Delivery Complete function When the total volume delivered reaches the preset delivery limit, the indicator and
buzzer notify the operator.
Nurse Call Alarm function Outputs an alarm relay contact, (can be connected to the nurse call.)
Contact rating: DC12V 1.0A
2-1
SY10-9B002
Applicable to:
Lot No.9803
Special functions Keep-Vein-Open function (After the delivery limit has been reached, infusion is
continued at 1 mL/h for a set flow rate of 1 mL/h; or at the delivery rate
setting < 1 mL/h.
Tubing Clamp function (When the door is opened, the tubing is automatically
clamped).
Volume Delivered Clear function (When kept depressed for about 2 seconds,
this function is activated.)
Time remaining display (When kept depressed for about 2 seconds, this function
is activated.)
Battery Capacity indicator ( 3 setting levels)
Adjustable display brightness (toggle between 2 levels.)
Adjustable Buzzer Volume ( 3 setting levels)
Operation History function (Storage & display of operation history)
Standby function (Temporarily deactivates the start- reminder alarm)
Repeat Alarm function
Start -Reminder Alarm function
Volume Memory function
Delivery Rate display type selection(From 0.1 mL/h steps to 1 mL/h steps)
Types of Power Supply 24 VA (AC 100-120V) / 28 VA (AC 100- 240V), 11 W (DC power)
Electrical Classification Class I and Internal power supply: Type CF / IPX1 Continuous operation
Equipment Dimensions 100 (W) 215 (H) 208 (D) mm (excl. protrusions)
2-2
SY10-9B002
Applicable to:
Lot No.9803
TE-172
Body Weight Mode Dose rate : 0.01 - 99.99 g/kg/min. (0.01 g/kg/min. step)
(Dose Rate Calculation) Body weight : 0.1 - 300.0 kg (0.1 kg step)
Memory Function The latest five patterns of the above four parameters are automatically
saved and can be displayed.
Accessories
AC power cord : 1 unit
Pole clamp : 1 unit
Time lag fuse : 2 pcs.
Instruction manual : 1 copy
DC cable 05LN51
2-3
SY10-9B002
Applicable to:
Lot No.9803
Reorder number Voltage Plug type Body weight mode Computer interface
NOTE Separate set-up procedures are required for product with voltage-plug combination other than
listed above.
CE0197
2-4
SY10-9B002
Applicable to:
Lot No.9803
For adjustment
Description Specification
1) Digital volt-meter Voltage range 20 V, resistance range 1
2) Synchroscope Frequency range 500 kHz or more
3) TERUFUSION Infusion sets for pump
4) Occlusion sensor adjustment jig (2 kinds) 3.1 , 4.1 (See 6. Part List)
5) Pressure gauge Pressure range 0 970 kPa (0 9.9 kgf/cm2)
(See 6. Part List)
6) A set of standard tools
Pressure gauge
Occlusion sensor
adjustment jigs
3-1
SY10-9B002
Applicable to:
Lot No.9803
For inspection
Description Specification
1) Flow-rate measuring cup Nearly 200 cc
2) Measuring cylinder (50 mL) 0.5 mL graduation
3) Stopwatch
4) TERUFUSION Infusion set type E for pump
or TERUFUSION Infusion set for pump
5) Water
6) Empty container or infusion bag
7) Pressure gauge Pressure range 0 970 kPa (0 9.9 kgf/cm2)
See 6. Part List.
8) Electronical balance Measuring resolution: about 0.01 g
3-2
SY10-9B002
Applicable to:
Lot No.9803
The Volume Memory", Repeat Alarm", Start-Reminder Alarm," Delivery Rate type selection" and
Self-Diagnosis" functions can be selected by using the respective dip-switches at the battery
compartment located below the pump unit.
Battery
Battery cover
Dip-switches
mounting screws
Display
Volume delivered/Time
mL
remaining display Indication appears
on display with
D. RATE, mL/h D. LIMIT, mL mode selection
Delivery rate/Delivery
limit display
4-1
SY10-9B002
Applicable to:
Lot No.9803
1-2. Press the [RATE/LIMIT] switch once again to go into Set item Year
the year adjusting mode, while month and date is [UP]switch extreme right
displayed. Year can be adjusted by pressing the [UP] or
[DOWN] switch. [DOWN]switch extreme right
2. While holding the [RATE/LIMIT] switch when the pump Set item Hour Minute
stops, press the [ ] switch to go into the time [UP]switch extreme right
adjusting mode. Time can be adjusted by pressing the
[UP] or [DOWN] switch. [DOWN]switch extreme right
NOTE: The adjusting mode ceases in about five seconds after the last switch is pressed.
4-2
SY10-9B002
Applicable to:
Lot No.9803
4-3
SY10-9B002
Applicable to:
Lot No.9803
Time Release
(ex. 18:30) [RATE/LIMIT]
D. RATE mL/h
4-4
SY10-9B002
Applicable to:
Lot No.9803
ON
TE-172C 12345678
1, 2, 3 : ON
D. RATE
Mode Dip-switch mL display Dip-switch: ON Dip-switch: OFF
display
Setting of ON When flow Decimal point display To be set to 1-step
decimal 12345678 rate<100mL/ mode (default) all over range for
point 1 : ON(default) h/ When 0.1-step: when flow both flow rate and
displayed volume rate < 100 mL/h Volume delivered.
(Dip switch OFF delivered ON **.* volume delivered <
#1) 1 2 3 4 5 6 7 8 <100 mL 100 mL.
1 : OFF ON **.* OFF ** (1-step: when flow
rate 100 mL/h
OFF ** and/or volume
delivered 100 mL)
Volume OFF OFF:Display Volume memory mode Default setting
memory 12345678 When power is
D. RATE
function 4 : switched on, D.
(Dip switch OFF(default) or . / D.
RATE , D. LIMIT and
#4) LIMIT
ON DOSE
12345678 or . RATE(TE-172C)
ON: D. RATE, stored in EEPROM
4 : ON D.LIMT and are indicated.
DOSE Display of flashing
RATE(TE-172 digits of D. RATE, D.
C) LIMIT and DOSE
RATE.(TE-172C)
4-5
SY10-9B002
Applicable to:
Lot No.9803
D. RATE
MODE Dip-switch mL display Dip-switch: ON Dip-switch: OFF
display
Repeat OFF
Deactivate repeat alarm Default setting
Alarm 12345678 Once an alarm is made When an alarm is
(Dip switch 5 : silent, buzzer no more made silent,
#5) OFF(default) sounds 2-minute later buzzer can
ON even if alarm state still resume sounding
12345678 remain. intermittently or
continuously
5 : ON 2-minute later if
alarm state still
remains.
Start-Remin OFF Deactivate Alarm (active)
der Alarm 12345678 If ready-to-start
start-reminder alarm
(Dip switch 6:OFF(default) pump is left
#6) Start-reminder alarm is
made invalid. stopped for 2
ON minutes without
12345678 operation any key,
6:ON a start-reminder
alarm causes
buzzer to sound
intermittently.
(Except when
another alarm
state exists or
standby function is
effective.)
Self-Diagno OFF See 4.3 Self-diagnosis Deactivate
sis Mode 12345678 mode.
self-diagnosis
(Dip switch 7:OFF(default)
#7) mode
ON
12345678
7:ON
Factory Use OFF No displayed Err 7 Check function of fail safe Default setting
(Dip-switch 1 2 3 4 5 6 7 8 is made circuit.
#8) 8:OFF(default)
Check
function of ON
fail safe 12345678
circuit 8:ON
4-6
SY10-9B002
Applicable to:
Lot No.9803
Contents
7-3
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) While pressing on both ends of the fuse holder, remove the holder off.
2) Take off fuse elements.
Fuse
Fuse holder
Body
Battery cover unit
Set-screws
2) Disconnect the battery lead from the connector CN4. Remove battery [24] .
Connector (CN4)
Battery housing
Battery
Body
Note: The main board inside the main body has an electric charge. Be sure to discharge it by
pushing [POWER] button on the body front after battery [24] is removed.
7-4
SY10-9B002(Rev.2)
Applicable to:
Lot No.9803
Set-screws
Note: Don't lay the main body on its side when removing the case unit. (To prevent the
brightness select switch from being damaged.)
Case unit
7-5
SY10-9B002
Applicable to:
Lot No.9803
1) If drip-proof cap (A) [7] has been damaged, remove it from the back of case unit.
Case unit
2) If drip-proof cap (B) [7] has been damaged, release the drip-proof cap (B) from three hooks and
detaches it.
Case unit
(as viewed from back)
Hooks
7-6
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Lay the main body on its side. Remove screws (3 - M3 × L8) and take off the FPC push plate (A) and FPC
pushers (A).
Screws
Flexible cable
FPC push plate (A)
FPC pushers (A)
2) Release the main board connector (CN7) from its lock. Disconnect flexible cable [8].
3) Disconnect main board connectors (CN6)(CN8)(CN10).
4) Remove 5 screws (M3 × L8) and immediately refer to No5.
Note: CN10 applies to only TE-171C and TE-172.
Screws
Screws
5) Slightly lift the main board and disconnect lead-wire assembly (3) from the connector (CN5) on the back of
main board. Now, detach main board [5].
Note: Place the detached main board [5] on the lithium-battery mounted side. Avoid placing the board on any
conducting or metallic surface.
Connector (CN5)
Main board
7-7
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) With door body unit [51] kept open unscrew the hinge screw* (M3 × L20) (1). Lift the door body unit [51] off
the main body and gently draw out the door with the flexible cable.
Main body
Door body unit
7-8
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Remove the downstream occlusion tubing press plate screw (M3 L8) and take off the plate.
2) Remove the door lock lever block screws (M3 L8) and lift the block.
Occlusion tubing
press plate
Screws
Screws Door lock lever block
3) Remove connecting pin. Slowly detach the door lock lever from its support block.
Note: When removing the connecting pin, ensure not to loose the wave washer.
Connecting pin
Door lock lever
block Door lock lever block
7-9
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Remove the tubing clamp block screws (M3 L8 and M3 L14). Detach the tubing clamp block.
2) Remove the tubing guide block screws (M3 L8). Detach the tubing guide block and door seal
rubber [26].
Note: There may be a washer t=0.5mm under the tubing guide block. Take care not to lose it.
3) Remove door case screws (flat-head screws M3 L10) and separate the door case[3] from the door
body.
4) Release the flexible cable connector (CN1) on display board [4] from its lock. Disconnect the flexible
cable from the connector.
Screw
Door body
Flexible cable
Connector (CN1)
Display board
Door case(3)
5) Remove screws (M3 L8) and detach the FPC retainer plate (B).
6) Remove flexible cable unit [8] from the door body.
7-10
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Remove the display fastening screws (M3 L8) and remove the display board [4] from the door
case[3].
Note: Take off the display board toward direction, taking care not to damage the operation
indicator LED.
Screws
Door case[3]
Display board
1) *Remove the occlusion sensor cover screws (M3 L8) attaching the upper occlusion sensor cover
[10] and take off the cover. The cover is removed together with the occlusion plunger and occlusion
spring.
2) *Remove screws (M3 L8) attaching the lower occlusion sensor cover [10] and take off the cover.
The cover is removed together with the occlusion plunger and occlusion spring.
Occlusion sensor
cover (upper)
Occlusion sensor
cover (lower)
Occlusion Occlusion
plunger sensor
Screws cover
Occlusion
spring
7-11
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Remove the side plate screws (M3 × L8 and 1 - M3 × L10) and take off side plate [14].
Screws
Screws
Side plate
Screws Screws
1) Disconnect lead-wire assembly (4) from the rotation sensor board connector (CN1).
2) Remove rotation sensor screws (M3 × L8) and detach rotation sensor board [23].
Note: During
under the board, from falling out and go missing.
Screws
Rotation sensor
board
Connector (CN1)
7-12
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Disconnect the lead wire assembly (1) and (3) from the DC connector (CN2) and AC connector (CN1)
on the switch-mode power unit [21].
Connector (CN2)
Switch-Mode
power unit
Connector (CN1)
2) Remove 6 rear plate screws (M3 L8) and take off rear plate unit [13].
Screws
Lead-wire
assembly(3)
Rear plate unit
Screws
7-13
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Disconnect lead-wire assembly (6) from the communication board connector (CN2).
2) Remove the communication board screws (M3 × L8) and detach communication board unit [34] from rear plate
unit [13].
Communication board unit Connector (CN2)
Screws
1) Remove the battery case screws (M4 × L10) and detach bottom unit [12] from the main body.
Note: During the disassembly process, ensure to prevent the side-plate plate fixing nut from falling off the
battery case.
Bottom unit
Screw
Screw
7-14
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Remove the switching power unit screws (M3 × L8) and take off switching power unit [21] from the main body.
Note: During the disassembly process, ensure to check for loose parts or components e.g. screws, nuts....,
present on the switch-mode power case.
Switch-Mode
power unit
Screws
1) Loosen the motor unit screws (M3 × L10) and slide motor unit [15] toward the timing pulley side.
2) Remove the shading plate screw (M3 × L5) and take off the shading plate and timing belt. When removing the
fixing screw, use an extra timing belt (aside from the one used for repair) to secure the pulley (large), which
preventing it from turning.
Note: To avoid cutting your finger, carefully remove the pulley (large) with keeping the cam unit from
turning using a timing belt.
Screw
Timing belt
Timing Pulley Screw
Screw Timing belt
7-15-1
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
3) To remove the pulley (large), insert the flat screwdriver between the pulley (large) and motor mounting plate and
lift it up.
Timing Pulley
Flat screwdriver
4) Remove set screws (M3 × L10) and detach motor unit [15] from motor mounting plate [16].
7-15-2
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Disconnect the lead-wire assembly (4) from the air-in-line sensor board connector (CN2).
2) Remove the air-in-line sensor screws (M3 × L12). Detach the air-in-line sensor unit [17] and the air-in-line
sensor board [18] from the main body.
Main body
Screws
1) Disconnect the lead-wire assembly (4) from the occlusion (upper) board connector (CN1).
2) Remove the motor mounting plate screws (M4 × L12) and detach motor mounting plate [16] from the main
body.
Screws
Screws
Main body
1) Remove occlusion unit (upper) screws (M3 × L8) and detach occlusion unit (upper) [19] from the main body.
Screws
Main body
7-16
SY10-9B002
Applicable to:
Lot No.9803
1) Disconnect the lead-wire assembly (5) from the occlusion (lower) board connector (CN1).
2) Remove the occlusion unit (lower) screws (M3 L8) and detach occlusion unit (lower) [20] from the
main body. Screws
Connector (CN1)
Occlusion unit
(lower)
Main body
1) Disconnect the lead-wire assembly (5) from the tube clamp board connector (CN1).
2) Remove the tube clamp screws (M3 L8) and detach tube clamp unit [11] from the main body.
Screws
Main body
Tubing Clamp Unit
7-17
SY10-9B002
Applicable to:
Lot No.9803
1) Disconnect the lead-wire assembly (5) from the door open / temperature sensor board connector
(CN1).
2) Remove temperature sensor board screws (M3 L5) and detach the door open / temperature sensor
board from the main body.
Main body
Screws
7-18
SY10-9B002
Applicable to:
Lot No.9803
Contents
7-19
SY10-9B002
Applicable to:
Lot No.9803
1) Mount door open / temperature sensor board [22] with temperature sensor board screws (M3 L5).
2) Reconnect the lead-wire assembly (5) to door open / temperature sensor board connector (CN1).
Attaching screws
Connector (CN1)
1) Install the tubing clamp unit [11] on the main body with the tubing clamp screws (M3 L8).
2) Reconnect the lead-wire assembly (5) to the tubing clamp board connector (CN1).
Note: When reassembling, keep the tubing clamp in the open position and fix the unit at the left
most position.
Screws
7-20
SY10-9B002
Applicable to:
Lot No.9803
1) Install occlusion unit (lower) [20] on the main body with the occlusion unit (lower) screws (M3 L8).
For the installing procedure, first, apply the silicon adhesive(X-31-1286T, maker: Shinetsu Silicone) to
the connection parts between the main body and then the occlusion unit before fixing them.
2) Connect the lead-wire assembly (5) to the occlusion (lower) board connector (CN1).
Note: When reassembling, Occlusion unit (lower)[20], take note which direction the occlusion
board is facing; up or down. The unit must be placed properly with the board-side facing
downward (in the opposite direction of the cam shaft).
Screws
Connector (CN1)
Main body
1) Install occlusion unit (upper) [19] on the main body with the occlusion unit (upper) screws (M3 L8).
For the installing procedure, first, apply the silicon adhesive(X-31-1286T, maker: Shinetsu Silicone) to
the connection parts between the main body and then the occlusion unit before fixing them.
Note: Place the board properly at the top position with the board-side facing upwards ( in the
opposite direction of the cam shaft).
Connector (CN1)
7-21
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Fix motor mounting plate [16] to the main body with set screws ( M4 × L12).
2) Connect the lead-wire assembly (4) to the occlusion (upper) board connector (CN1).
Screws Screws
Main body
7-22
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Install air-in-line sensor unit [17] and air-in-line sensor board [18] onto the main body with the air-in-line sensor
screws(M3 × L12).
Notes: Apply silicone adhesive (X-31-1286T, maker: Shin-Etsu Silicones) to the outer-side of the air-in-line
sensor fastener section (refer to diagram for exact position). Next, place the sensor unit to the main
body.
Take note not to allow any silicone to slip in between the tubing groove in the air-in-line sensor case.
2) Connect the lead-wire assembly (4) to the air-in-line sensor board connector (CN2).
Lead-wire
assembly (4) Motor mounting plate
Main body
7-23
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Fix the motor unit [15] to the motor mounting plate [16] with the motor unit screws (M3 × L10), ensure motor
connector on the left side.
2) Install the shading plate (large) on the cam shaft so that D-cut is in alignment with each other. Fix the shading
plate (large) with the screw (M3 × L5). When fixing the shading plate with the fixing screw, use an extra timing
belt (aside from the one used for repair) to secure the pulley (large), which preventing it from turning.
Notes: When installing the pulley (large), be careful to avoid damage to the D-cut of the pulley (large).
Apply screw-lock to the threads before installing the screw.
To avoid cutting your finger, carefully install the pulley (large) with keeping the cam unit from
turning using a timing belt.
Screw
Timing belt
3) Install timing belt [25] on and around both the shading plates (large & small). Slide the motor in the direction of
Screw
Screw
Motor unit
7-24-1
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
Note: When you fasten the motor unit 15 , take care not to make the timing belt 25 too tight or too loose.
First, visually check that the timing belt 25 is stretched taut without a slack. Then adjust the timing
belt tension by depressing the belt lightly with a finger at the white arrowed point in the photo below
(the deflection recommended by the manufacturer: 0.8mm). The timing belt should not be too tight
or too loose.
Depress with a
Adjust the timing belt 25 tension while finger.
pulling the motor unit 15 toward the
back side of the main unit (red arrowed
direction).
1) Mount switching power unit [21] on the main body with the switching power unit screws (M3 × L8).
Note: During the assembly process, ensure to check for loose parts or components e.g. screws,, Nuts,...,
present on the switch-mode power case.
Screw
Screw
7-24-2
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Temporarily fit bottom unit [12] to the main body with the battery case unit (M4 × L10).
Bottom unit
Screws Screw
1) Fit communication board unit [34] to rear plate unit [13] with the communication board unit screws (M3 × L8).
2) Connect lead-wire assembly (6) to the communication board connector (CN2).
Communication
board unit
Screw
Connector (CN2)
7-25
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Temporarily fit rear plate unit [13] with the rear plate unit screws (M3 L8).
Screws
Screw
Screw
2) Connect the lead wire assembly (1) and (3) from the DC connector (CN2) and AC connector (CN1) on the
switch-mode power unit [21].
Note: Correctly arrange the lead-wire assembly (1) (AC input-side) and (3) (DC output-side) as illustrated
below.
Connector
(CN2)
Switch-Mode
Power unit
Connector
(CN1)
7-26
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Fit rotation sensor board [23] to motor mounting plate [16]with the rotation sensor board screws (M3 × L8).
2) Reconnect lead-wire assembly (4) to the rotation sensor board connector (CN1).
Note: Re-install the washers, if any were present on the underside of the board upon disassembly.
Screws
Rotation sensor
board
Connector (CN1)
1) Temporarily fit the side plate with the side plate screws (M3 × L8 and M3 × L10).
Screws
Screws
Side plate
Screws
Screws
7-27
SY10-9B002
Applicable to:
Lot No.9803
1) Retighten total 15 screws which fit Bottom unit [12](M4 L10), Rear plate unit [13](M3 L8), and
Side [14](M3 L8 and M3 L10).
1) Retighten the motor unit screws (M3 L12) so that specified timing-belt tension can be obtained.
Note: Too great a belt tension will cause the motor to halt. Too little a belt tension would cause
the belt to jump the grooves.
Take special care in ensuring the optimum belt tension all the time.
7-28
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Connect the lead-wire assembly (3) to the connector (CN5) on the back of main board.
Connector
(CN5)
Main board
Main board
Connector
(CN9)
Connector (CN10)
Connector
(CN6)
Connector
(CN8)
7-29
SY10-9B002(Rev.2)
Applicable to:
Lot No.9803
Note: CN11 on Main board [5] is set to ON in the service part when a service part is shipped from
the factory.
If CN8 is connected to the main board early, or if CN11 is set to ON, temporarily turn the
power supply to ON. If power ON/OFF is impossible, leave CN8 unconnected and CN11
in OFF until power ON/OFF becomes possible. After the power is turned on once, in rare
cases, Err1, Err6, or Err7 may occur during self-diagnosis. If this is the case, disconnect
CN8, CN11, the Ni-Cd battery, and the AC and DC powers to shut off all power. Then,
reconnect all the connectors and turn the power back on. If it starts up successfully,
there is no problem.
When main board [5] or pump body unit [2] is replaced with new one, create the
correction values in the main board in the self-diagnosis mode, after assembly, by using
those values stated on pump body unit [2].
CN11 CN11
OFF state ON state
7-30
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Install the upper occlusion sensor cover [10] with the occlusion sensor cover screws (M3 L8).
When installation set the spring, plunger and cover in that order.
2) Install the lower occlusion sensor cover [10] with the occlusion sensor cover screws (M3 L8).
When installation set the spring, plunger and cover in that order.
Notes: Insert the upper plunger with its magnet facing upwards and the lower plunger with its magnet
facing downwards.
Ensure the springs are resting properly in each of the plunger grooves.
Fix each occlusion sensor cover at near the center position.
After installing, check for smooth plunger movement. In addition, check that the plunger-to-cover
clearance is approximately uniform.
Do not apply any thread-lock to the occlusion cover set screws.
1) Install display board [4] in the door case with the display board screws (M3 L8).
Note: Carefully bring the display board in the arrow-marked direction so that LEDs are properly set in
the operation indicator cover.
Screws
Door case
Display board
7-31
SY10-9B002
Applicable to:
Lot No.9803
1) Insert the flexible cable unit into the hole in the door body opening in the arrow-marked direction.
2) Fix the flexible cable unit [8] with the FPC push plate (B) and the flexible cable unit screws (M3
L8).
Notes: Take note not to install the cable unit inside out. The cable with lettering (with the
conductive terminal) must be positioned on the top-side (display board side).
Ensure that the cable is in alignment with the rubber ring washer.
7-32
SY10-9B002
Applicable to:
Lot No.9803
1) Insert the flexible cable into the connector (CN1) on display board [4] and lock the connector.
Notes: Leave a half loop of the cable between the door body and the display board.
Make sure that the rubber seal is installed properly.
Door body
Flexible cable
Connector (CN1)
Door case
7-33
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
2) Fit the door case to the door body with the door body screws (flat-head screws M3 L10).
3) Fix the clamp block to the door body with the clamp block screws (M4 L8 and M4 L14).
4) Install the washer (t=0.5mm) onto the tubing guide block mounting boss. Fasten the tubing guide block 26
and door seal rubber with the tubing guide block screws (2 W sems screws M3 L8) to the door body.
Notes: Before fitting the door case to the door body, check that the seal rubber is properly set in the door
case groove.
Apply "thread lock" at the tip of set-screws when using them for assembling the door body.
7-34
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
1) Insert the door lock lever from the front side of Door case unit [3].
2) With the wave washer set on the door lock lever, insert the door lock lever into the door lock lever block, and put
in the connecting pin to the hole in the door lock lever block to fit the hook and wave washer into the block.
*Note: Apply thread-lock" to the tip of set-screws before using them to set the door lock lever block,
occlusion tubing press plate, and clamp block onto the door body.
7-35
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
1) Pass the flexible cable through the cable insert slot in the main body.
2) Put on the door body unit to the connecting pin on Pump body unit [2]
3) - , into the pine
hole for Hinge screw and then fix pump body unit with Hinge screw.
Hinge screw
4) Put the flexible cable, FPC pushers (A) and FPC pushers (B) together, in that order from the Door body unit [51]
side.
Align the pin holes on either side of the assembly with the corresponding screw hole by inserting a pin or the
like through the assembly. (It is good practice to keep pins or the like put into the respective screw holes and set
the assembly holes into alignment with those pins or the like.)
Pin or the like
Flexible cable
FPC pushers (A)
7-36
SY10-9B002
Applicable to:
Lot No.9803
5) Connect the flexible cable to the main-board connector (CN7) and lock it there.
6) Align the black line on the flexible cable with the side face of the FPC push plate (A) and fix the FPC
pushers (A) and FPC push plate (A) with push plate screws (M3 L8).
Note: When fixing the FPC push plate (A), leave an adequate slack of the flexible cable.
Push plate screws Case unit (as viewed
from back)
Main board
Connector (CN7)
Hooks
7-37
SY10-9B002(Rev.2)
Applicable to:
Lot No.9803
Notes: Take care not to break off the tact switch (BS1: Brightness selector switch) on Main
board [5].
Check that the rubber seal (body) is properly set in the main-body unit groove.
Case unit
Note: If the screw holes in case unit [1] cannot be aligned with those in Pump body unit [2],
temporarily loosen set screws and, then, try to align the holes again.
Attaching screw
Case unit
7-38
SY10-9B002
Applicable to:
Lot No.9803
Case unit
1) Mount battery [24] in place and connect its lead wire to the battery connector (CN4).
Notes: Place the lead wire as illustrated below to prevent the wire from touching the internal
dip-switches.
Ensure that the dip-switches are at their proper settings.
Battery
Internal dip-switches
CN4
Lead wire
7-39
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
2) Install the battery cover with set screws (flat-head screws with lock tight 4 - M4 L16)
Main body
Fuse
Fuse holder
7-40
SY10-9B002
Applicable to:
Lot No.9803
Every time a TERUFUSION Infusion Pump TE-171/172 is disassembled and re-assembled, be sure
to perform the necessary adjustments according to the section 7. Check List & Adjustments following
any repair/replacement work.
Should the necessary adjustments be not carried out the alarms and/or battery charge monitor will
not to function property.
Name Specification
1) Digital voltmeter Voltage range: 20 V, Resistance range: 1
2) Synchroscope Frequency range: 500 kHz or more
3) TERUFUSION Infusion set for pump
4) Occlusion sensor adjustment jig (2 types) 3.1 mm, 4.1 mm.
See 6. Parts List.
5) Pressure gauge Pressure range: 0 970kPa(0 9.9 f/ 2)
See 6. Parts List.
6) Standard tool set
7-41
SY10-9B002(Rev.3)
Applicable lots:
Lot No,9803
Revision date: Apr.2011
* Resetting the occlusion detection pressure cannot be made in the self-diagnosis mode. Select [M].
Set the self-diagnosis mode (DIPSW-7P: ON), turn on the power switch, and set the flow rate to [120]
mL/h.
* Delivery limit is automatically set to [----] (delivery limit free).
Press [RATE/LIMIT] button two times repeatedly to indicate [2000] on the volume delivered display.
With the door left open, adjust the VR1 on the occlusion sensor unit (lower) to set the A/D indication to
[155]. (This value is only for purpose of making a standard.)
Push the downstream occlusion plunger three times repeatedly with a jig (3.1 mm/4.1 mm) and calculate a
mean value of three measurements.
3.1 mm jig: AVG1 (a mean value of 3 measurements)
4.1 mm jig: AVG2 (a mean value of 3 measurements)
In this case, make sure that [51] (AVG1 - AVG2) [57].
When the value exceeds [57], turn VR1 clockwise to correct the excess of reading.
7-42
SY10-9B002
Applicable to:
Lot No.9803
Make sure that a value of detection falls in the range of 30 90 kPa (0.30 0.90 kgf/cm2).
Also, check that a variation is [90] or more when the plunger is pushed to the extreme end of
its stroke.
In case of a failure of entering the detection range, repeat adjustment, beginning at step .
Also, check for smoothness of plunger movement.
After adjustment, return to the ordinary mode and check for L, M and H occlusion detection
pressures. Readings must fall in the ranges below:
L 13.3 60kPa(0.1 0.6 f/ 2)
M 30 90kPa(0.3 0.9 f/ 2)
H 60 133.3kPa(0.6 1.4 f/ 2)
7-43
SY10-9B002(Rev.2)
Applicable to:
Lot No.9803
Press [RATE/LIMIT] button three times repeatedly to indicate [2001] on the volume delivered
display.
With the door left open, adjust the VR1 on the occlusion sensor unit (upper) to set the A/D
indication to [60 1].
Push the upstream occlusion plunger with the 3.1 mm jig and make sure that value of A/D is [70] or
larger. If [70] is not reached, replace the occlusion sensor unit (upstream) with new one.
Set the occlusion pressure test jig on the upstream side of the pump and push [START] button.
Make sure an occlusion alarm occurs in the range of -30 -80 kPa (-0.30 -0.80 kgf/cm2).
(Refer to 9-7)
7-44
SY10-9B002
Applicable to:
Lot No.9803
Adjust the Air-In-Line sensor according to the flow-chart on the next page.
Press [RATE/LIMIT] button four times repeatedly to indicate [3000] or [3001] on the volume
delivered display.
[3000]: A/D value of Air-In-Line sensor (lower) is indicated.
[3001]: A/D value of Air-In-Line sensor (upper) is indicated.
Initially set the VR1, VR2, VR3 at the center position (12 o'clock position). Turning a volume
counterclockwise causes A/D value to increase. Turning clockwise causes to decrease. Adjusting VR2
causes A/D values to vary for both [3000] and [3001]. Adjusting VR3 and VR1 causes A/D value to vary for
[3000] and [3001], respectively.
Note: VR1 and VR3 are mainly used to decrease the A/D value.
7-45
SY10-9B002
Applicable to:
Lot No.9803
START
Less
Set Volume delivered than 100
display at [3000]. A/D value for [3001] ?
100~140
Set water filled tube in position.
Adjust A/D value to approx. 120 with
VR2.
Less than
100 Over 140
A/D value ? Decrease A/D value for [3000] to
100 140 with VR3. Clockwise
rotation of VR3 is limited to 180
100~140 maximum.
Adjustment Replace
completed ? sensor.
7-46
SY10-9B002
Applicable to:
Lot No.9803
Decrease A/D value for [3001] to Set an empty tube in position and check
100 120 with VR2. for A/D values for [3000] and [3001].
Less
than 100 A/D value 10 or Replace
A/D value for [3000] ?
less ? sensor.
100~140
Adjustment Replace
completed ? sensor.
7-47
* [ ] represents service part number.
Operation Problem Possible causes Determining method Remedial method
8.2
1. Operate door lock lever Door fails to open and 1. Door Lock Lever 1. Visual inspection 1. Replace Crescent hook unit [9].
to open and close door. close. damaged. 2. Visual inspection 2. Replace Pump body unit [2].
2. C-Hook cradle 3. Visual inspection 3. Replace Pump body unit [2].
damaged. 4. Visual inspection 4. Replace Pump body unit [2].
3. Pump body deformed.
4. Door hinge failure.
2. Open door and check Buffer plate fails to 1. Malfunctioning due to 1. Visually inspect buffer plate. 1. Wipe off solution, if any, with a
Fault Finding
buffer plate for normal move. solution sticking cloth soaked with water or tepid
movement. water. If malfunctioning remains
even after wiping off sticking,
replace Pump body unit [2].
3. Move the occlusion Occlusion plunger fails 1. Foreign matter caught in 1. Visual inspection 1. Remove foreign matter.
detection plunger with to move smoothly. occlusion-plunger
finger force. moving part.
2. solution sticking 2. Visual inspection 2. Replace Occlusion sensor cover
[10].
4. Disengage the tubing Tube clamp cannot be 1. Tube clamp unit faulty. 1. Visually inspect tube clamp. 1. Replace Tube clamp unit [11].
clamp. released. 2. Solution sticking or 2. Visual inspection 2. Replace Tube clamp unit [11].
malfunctioning due to
foreign matter.
5. Visual Inspection of Handle, case and/or 1. Breakage due to fall, 1. Visual inspection 1. Replace following parts if any
8-11
Pump Main Body. body interior shows shock or aging. Check up by using history function. faulty:
abnormality. Confirm [Shock] indication in case Case unit [1]
of occurrence of a shock Pump body unit [2]
Door case unit [3]
Battery cover unit [6]
Drip-proof cap unit [7]
Crescent hook unit [9]
Bottom unit [12]
Rear plate unit [13]
Side plate [14]
Motor mounting plate [16]
Door seal rubber [26]
Door protection label [27]
Pole clamp [33]
Lot No.9803
SY10-9B002
Applicable to:
8-12
8. Err.2 is 1. ROM/RAM of sub CPU 1. Confirm Err.2 indication. 1. Replace Main board [5].
displayed. abnormal.
9. Err.3 is 1. Main board CPU faulty. 1. Suspected faulty if output pulse is not 1. Replace Main board [5].
displayed. detected at main board IC1-44P or input
pulse is detected at IC1-66, 67P is
detected without output pulse.
2. Rotation sensor board 2. Check for conductivity. 2. Replace Lead-wire assembly (4) [32].
relay cable
disconnected.
3. Rotation sensor board 3. Suspected faulty if no output pulse is 3. Replace Rotation sensor board [23].
faulty. detected at rotation sensor board
CN1-2P, -3P during motor rotation.
4. Motor faulty. 4. Suspected faulty if motor does not rotate 4. Replace Motor unit [15].
even when pulse output is detected at
CN5-1P 3P.
5. Motor drive circuit 5. Suspected faulty if pulse generation is 5. Replace Main board [5].
faulty.
Lot No.9803
3P.
6. Timing belt breaking or 6. Visual inspection. 6. Replace Timing belt [25] or correct
getting out of pulley. installed condition of it.
Issued in Jan., 1999
10. Err.4 is 1. Display CPU faulty. 1. Confirm indication of Err.4 or something 1. Replace Display board [4] or Flexible
displayed. else. cable unit [8].
Operation Problem Possible cause Determining method Remedial method
11. With AC power Err.5 is displayed. 1. EEPROM faulty. 1. Err.5 is displayed. 1. Replace Main board [5]
12. Err.6 or Err.7 is 1. Main CPU or sub CPU 1. Confirm indication of Err.6 or Err.7. 1. Turn on power switch again or
displayed. running away. replace Main board [5].
2. Flexible cable 2. Check flexible cable for conduction 2. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
13. Err.8 is displayed. Push buttons in self-diagnosis mode
[9000]. If [01] [12] are not
displayed in correspondence with
pushed buttons, proceed to
following:
1. Tact switch faulty. 1. Suspected faulty if [L] level signal is 1. Replace Display board [4].
detected from the first on display
board CN1-18 30P.
2. Flexible cable 2. Check flexible cable for conduction 2. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
3. Other than above. 3. Other than above. 3. Replace Main board [5]
14. Err.9 is displayed. 1. Malfunctioning of tube 1. Suspected faulty if locking cannot 1. Replace Tube clamp unit [11] or
clamp due to drops be released by door open/close clean it with cloth soaked with
sticking to it. operation and tube unclamp water or tepid water.
operation.
8-13
2. Tube-clamp board faulty. 2. Suspected faulty if tube 2. Replace Tube clamp unit [11] or
clamp/unclamp operation does not Lead-wire assembly (5) [32].
cause a change of voltage at
3. Main board faulty. main-board CN8-9P.
3. Suspected faulty if tube 3. Replace main board [5].
clamp/unclamp operation does not
cause a change of voltage at
main-board IC1-53P.
15. Err.A is displayed. 1. Communication failure 1. Confirm indication of Err.A. 1. Replace Main board [5]
between sub CPU and * However, Err.A may rarely occur
battery monitor IC. also with progress of battery
discharge.
Lot No.9803
SY10-9B002
Applicable to:
8-14
at CN6-5P.
18. Err.d is displayed. Mount water tube and make
checking in self-diagnosis mode
[2001]. If a value of 80 120 is not
displayed, proceed to the following:
1. Occlusion unit (upper) 1. Suspected faulty if any voltage 1. Replace Occlusion unit (upper)
faulty. exceeding approx. 3.5V is detected [19].
at occlusion sensor board (upper)
CN1-2P.
2. Main board faulty. 2. Suspected faulty if, with main board 2. Replace Main board [5]
CN6 disconnected, any voltage
exceeding approx. 3.5V is detected
at CN6-7P.
19. Err.E is displayed. Mount water tube and make
checking in self-diagnosis mode
[2000]. If a value of 80 120 is not
Lot No.9803
SY10-9B002
1. Occlusion unit (lower) 1. Suspected faulty if any voltage 1. Replace Occlusion unit (lower)
faulty. exceeding approx. 3.5V is detected [20].
at occlusion sensor board (lower)
Issued in Jan., 1999
CN1-2P.
2. Main board faulty. 2. Suspected faulty if, with main board 2. Replace Main board [5]
CN8 disconnected, any voltage
exceeding approx. 3.5V is detected
at CN8-2P.
Operation Problem Possible cause Determining method Remedial method
20. With AC power All of LED/LCD fail to 1. LED/LCD faulty. 1. Visual inspection (check for inverted 1. Replace Display board [4].
connected and door kept blink three times mounting, etc.)
open, depress repeatedly. 2. Flexible cable 2. Check flexible cable for conduction 2. Replace or re-connect Flexible
[POWER] button. disconnected or poorly or try to re-connect it. cable unit [8].
connected.
3. Other than above. 3. Other than above. 3. Replace Display board [4] or Main
board [5].
21. Finger fails to operate. 1. Pulley (large) idling. 1. Suspected faulty if camshaft does 1. Replace Pump body unit [2].
not rotate although pulley (large)
rotates.
2. Other than above. See 2. Other than above. See Determining 2. Other than above. See Remedial
Possible cause of No. 9. method of No.9. method of No.9.
22. [AC/DC] lamp fails to 1. Display board faulty. 1. Suspected faulty if signal [H] is not 1. Replace Display board [4].
light. detected at display board IC1-27P.
2. Main board faulty. 2. Suspected faulty if signal [L] is not 2. Replace Main board [5].
detected at main board IC2-28P.
3. Flexible cable 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
23. Operate the Display of (D. 1. Tact switch faulty. 1. Check for conduction (Display 1. Replace Display board [4].
[RATE/LIMIT] button. RATE/D. LIMIT) fails board CN1-30P)
8-15
to alternate. 2. Display board faulty. 2. Suspected faulty if no signal is 2. Replace Display board [4].
detected at display board IC1-11
18P, 4 10P.
3. Flexible cable 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
4. Main board faulty. 4. Suspected faulty if switch operation 4. Replace Main board [5].
does not cause any change at main
board IC1-48P, IC2-16P.
24. Operate the (Volume Display of (Volume Push switch in self-diagnosis mode
delivered/Time delivered/Time [9000]. If [01] is not displayed,
remaining) button. remaining) fails to proceed to following:
alternate. 1. Tact switch faulty. 1. Check for conduction (Display 1. Replace Display board [4].
board CN1-29P)
2. Display board faulty. 2. Suspected faulty if no signal is 2. Replace Display board [4].
detected at display board IC1-11
18P, 4 10P.
Lot No.9803
SY10-9B002
Applicable to:
3. Flexible cable 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
Issued in Jan., 1999
4. Main board faulty. 4. Suspected faulty if switch operation 4. Replace Main board [5].
does not cause any change at main
board IC1-36P, IC2-15P.
Operation Problem Possible cause Determining method Remedial method
25. Using [value selector] Setting of flow rate Push switch in self-diagnosis mode
switch, set as follows: and delivery limit [9000]. If [02][05][07][10][12][13] are
Flow rate: 100 mL/h cannot be made. not displayed, proceed to following:
D. LIMIT: 9999 mL 1. Tact switch faulty. 1. Check for conduction (Display 1. Replace Display board [4].
board CN1-23 28P)
2. Display board faulty. 2. Suspected faulty if no signal is 2. Replace Display board [4].
detected at display board IC1-11
18P, 4 10P.
3. Flexible cable 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
4. Main board faulty. 4. Suspected faulty if switch operation 4. Replace Main board [5].
does not cause any change at main
board IC1-50, 55, 57 59, 73P.
26. With [RATE/LIMIT] Occlusion pressure 1. Tact switch faulty. 1. Check for conduction (Display 1. Replace Display board [4].
switch button pressed, set value LED fails to board CN1-23, 30P)
pushing [DOWN] (lower) change in sequence of 2. Display board faulty. 2. Suspected faulty if no signal is 2. Replace Display board [4].
button to display [PrES] [M] [H] [L] [M]. detected at display board IC1-42
and, further push 44P.
[DOWN] button three 3. Flexible cable 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
8-16
times repeatedly. disconnected or poorly or try to re-connect it. cable unit [8].
connected.
4. Main board faulty. 4. Suspected faulty if switch operation 4. Replace Main board [5].
does not cause any change at main
board IC1-48, 59P.
27. With [RATE/LIMIT] Indication on 1. Tact switch faulty. 1. Check for conduction (Display 1. Replace Display board [4].
switch button pressed, Flow-rate / board CN1-24, 30P)
push [DOWN] (lowest) Programmed-flow 2. Display board faulty. 2. Suspected faulty if no signal is 2. Replace Display board [4].
button three times display fails to change detected at display board IC1-11
repeatedly. in sequence [b-3] 18P, 4 10P.
[b-1 [b-2]. 3. Flexible cable 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
4. Main board faulty. 4. Suspected faulty if switch operation 4. Replace Main board [5].
does not cause any change at main
board IC1-48, 50P.
Lot No.9803
SY10-9B002
Applicable to:
Buzzer fails to sound 1. Main board CPU faulty. 1. Suspected fault if no pulse output is 1. Replace Main board [5].
or sound level cannot detected at main board IC1-6, 45P.
be adjusted. 2. Buzzer voltage 2. Suspected faulty if main board Q5 2. Replace Main board [5].
generation circuit faulty. emitter voltage (IC6-3P) fails to
Issued in Jan., 1999
change.
Operation Problem Possible cause Determining method Remedial method
28. Depress the [START] Pump fails to start. Push switch in self-diagnosis mode
button, after [9000]. If [03] is not displayed,
programming the proceed to following:
following; 1. Tact switch faulty. 1. Check for conduction (Display 1. Replace Display board [4].
Flow rate: 120 mL/h board CN1-21P)
2. Flexible cable 2. Check flexible cable for conduction 2. Replace or re-connect Flexible
D.LIMIT: 20 mL disconnected or poorly or try to re-connect it. cable unit [8].
connected.
3. Main board faulty. 3. Suspected faulty if switch operation 3. Replace Main board [5].
does not cause any change at main
board IC1-64P.
4. Motor drive circuit faulty. 4. Suspected fault if no pulse output is 4. Replace Main board [5].
detected at main board IC3-38
45P.
Suspected fault if no pulse input is
detected at motor connector
CN5-1 3P.
5. Motor faulty. 5. Suspected fault if motor fails to 5. Replace Motor unit [15].
rotate although no circuit failure is
found by foregoing steps 1 to 4.
29. Accuracy of 1. Pump unit or buffer plate 1. Check buffer plate for smooth 1. Replace Pump body unit [2].
10-minute flow-rate faulty. movement.
8-17
exceeds 20 mL 5%. Check for free-flow state.
2. Main-board 2. Check for correction value in 2. Replace Main board [5].
self-diagnosis correction self-diagnosis mode. If not within
value is set improperly. range of 0.80 1.20, main board
must be replaced.
3. Tube used is out of 3. Check for tube specification. 3. Re-test after replacing with a
specification. specified tube.
4. Pump is started without 4. Check for proper operation. 4. Re-start after operating [Volume
operating [Volume Delivered Clear] switch.
Delivered Clear] switch.
5. Other than above. 5. Other than above. 5. Replace Pump body unit [2].
Lot No.9803
SY10-9B002
Applicable to:
8-18
at main board CN8-6P. assembly (5) [32].
2. Suspected faulty if opening the door 2. Replace Main board [5].
2. Main-board door-open does not cause a change of signal
signal processing circuit at main board IC12-4P.
faulty. 3. Suspected faulty if opening the door 3. Replace display board [4].
3. Poor display-board does not cause any change of
operation indicator drive signal at main board IC1-47P.
circuit faulty. 4. Check flexible cable for conduction 4. Replace or re-connect Flexible
4. Flexible cable or try to re-connect it. cable unit [8].
disconnected or poorly
connected.
Lot No.9803
SY10-9B002
Applicable to:
8-19
1.6V or less (water tube) or approx.
0.2V or more (empty tube) when
connector CN6 is drawn off.
33. Set-up the occlusion Occlusion detection 1. Malfunctioning of 1. Visual inspection 1. Wipe off stuck solution with cloth
pressure measurement pressure is out of set occlusion plunger due to soaked with water or tepid water.
gauge at the range (M set mode: solution sticking to it or Remove foreign matter if any.
downstream side & start 0.3 0.9 kgf/cm2). foreign matter caught by
the pump. it.
2. Occlusion detecting part 2. Re-adjust with reference to 7.3.1 2. Re-adjust
out of adjustment. Adjustment Procedure.
3. Main board faulty. 3. Suspected faulty if occlusion alarm 3. Replace Main board [5].
does not occur when main board
connector CN8 is drawn off.
4. Other than above. See 4. Other than above. See Determining 4. Other than above. See Remedial
Possible cause of No. method of No.19. method of No.19.
19.
Lot No.9803
SY10-9B002
Applicable to:
8-20
faulty. does not cause pulse output to stop
at IC1-45P.
36. Open the door. Door lamp fails to 1. Door-open sensor 1. Suspected faulty if opening the 1. Replace Door-open / Temp.
light. faulty. door does not cause a change of sensor board [22] or Lead-wire
signal at main board CN6-6P. assembly (5) [32].
Suspected faulty if opening the Replace Main board [5].
door does not cause a change of
signal at main board IC1-56P.
2. Check for damage to door lamp 2. Replace display board [4].
2. Door lamp LED or LED or failure of display board.
display board faulty. 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
3. Flexible cable unit poorly or try to re-connect it. cable unit [8].
replaced or
re-connected.
Lot No.9803
SY10-9B002
Applicable to:
8-21
direct-current signal to pulse signal
at display board IC1-45, 46P.
39. Depress the [History] History is not 1. [History] switch faulty. 1. Visually check for conduction (Main 1. Replace Main board [5].
button. displayed. board BS2).
2. Main board faulty. 2. Suspected faulty if switch operation 2. Replace Main board [5].
does not cause a change at main
board IC1-1P.
3. Flexible cable 3. Check flexible cable for conduction 3. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
4. Display board faulty. 4. Other than above. 4. Replace display board [4].
40. Connect the nurse-call Short circuit cannot be 1. Main board faulty. 1. Suspected faulty if alarm fails to 1. Replace Main board [5].
Lot No.9803
cable to the pump and made between cable short nurse-call relay (RY1).
SY10-9B002
Applicable to:
generate an alarm. terminals. 2. Nurse-call cable faulty. 2. Check for conduction. 2. Replace nurse-call cable.
Issued in Jan., 1999
Operation Problem Possible cause Determining method Remedial method
41. Connect drip sensor , No. of selected 1. [IV set] switch faulty. 1. Suspected faulty if select-switch 1. Replace Main board [5].
set the desired number infusion-set drops is operation does not cause a change
of drips for infusion set not displayed. of output at main-board drop select
drops with the switch at 2. Flexible cable switch SSI-3P (center).
the rear of the pump. Do disconnected or poorly 2. Check flexible cable for conduction 2. Replace or re-connect Flexible
not mount the drip connected. or try to re-connect it. cable unit [8].
sensor onto the drip 3. Drip-detect probe poorly
chamber, start the pump. connected. 3. Check for proper connection. 3. Reconnect drip-detect probe and
switch on power or replace probe.
4. Display board faulty. 4. Replace Display board [4].
42. Flow error alarm is not 1. Main board faulty. 1. Suspected faulty if no alarm is 1. Replace Main board [5].
made. made though pulse output is
detected at IC13-3P.
43. Mount the drip sensor on Drop detect LED fails 1. Drip sensor faulty. 1. Connect another known drip sensor 1. Replace drip sensor.
drip chamber and start to blink. and test. If LED blinks normally,
the pump. doubted sensor is suspected faulty.
2. Suspected faulty if no pulse output
2. Main board faulty. detected at IC13-3P or if no signal 2. Replace Main board [5].
changes detected at CN1-3P when
drops fall.
8-22
44. Using the [ DISPLAY Body weight mode Push switch in self-diagnosis mode
ON/OFF ], [ ITEM cannot be set. [9000]. If [08] [11] are not displayed,
SELECT ] & [ Value proceed to following:
Selector ] buttons, 1. Tact switch faulty. 1. Check for conduction (Display 1. Replace display board [4].
attempt to operate this board CN1-20, 22P)
function. 2. Flexible cable 2. Check flexible cable for conduction 2. Replace or re-connect Flexible
disconnected or poorly or try to re-connect it. cable unit [8].
connected.
3. Main board faulty. 3. Suspected faulty if [Display 3. Replace Main board [5].
ON/OFF] and [Item Select] switch
operations do not cause any
change of signals at main board
IC1-61, 62P.
Lot No.9803
SY10-9B003
47. Allow the pump to Two or more battery 1. Battery life has expired. 1. Expiration suspected if three LEDs 1. Replace battery [24].
operate till shutdown LEDs fail to light. flicker when power is switched on.
8-23
occurs, then re-charge 2. Battery recharging 2. Suspected faulty if recharging 2. Replace main board [5].
for another 15 hours. circuit faulty. current is not approx. 130 mA.
3. Battery monitor IC reset. 3. Reset suspected if three LEDs 3. Recharge battery for 15 hours
flicker when power is switched on. according to 9. Maintenance and
Cleaning.
48 With AC power Err.F is displayed. 1. The power was not 1. 1. Turn the power off properly with
connected and door kept turned off properly. [POWER] button.
open, depress [POWER]
button. 2. Main board faulty. 2. Suspected faulty if, with main board 2. Clean off the . If it does not
CN6 disconnected, 4.2V or above is solve the problem, replace main
detected at CN6-4P. board [5]
Lot No.9803
Applicable to:
SY10-9B002(Rev.1)
When any error conditions occurs such as those listed below, take the necessary action/s as stated in
the Instruction Manual, External Communication Protocol, and Section 8 Troubleshooting.
8.3.1 Alarms
Low battery When the internal Ni-Cad battery charge falls below 20% available
operational time, the [BATTERY] indicator lamp blinks with a buzzer alarm
indication. An intermittent beep will sound first followed by a continuous
beep sound as time elapses. The [ Operation Indicator ] does not show any
change.
Flow error An alarm is activated when an inconsistent flow-rate is detected (free-flow,
(when the DRIP empty drip chamber or flow error).
Sensor is used) when the pump is in [STOP] operation and a continuous flow of
solution drops are detected. This is regarded as [Free Flow] condition.
When the pump is in [START] operation and no flow of solution drops
are detected.
When the pump is in [START] operation and a continuous flow
of solution drops are detected to be out-of-time with the specified
drop selection. This is regarded as [Flow Error] condition.
8-24
SY10-9B002(Rev.1)
Applicable to:
Lot No.9803
Issued in Feb., 2007
8-25
SY10-9B002 (Rev.4)
Applicable to:
Lot No.9803
Inspection during commissioning (every time before a new pump is put into initial use)
1 Was there any damage/fluid adhesion to the case and/or the pole clamp? Pass Fail
2 Was there normal operation when the pump was operated with battery
Pass Fail
power?
3 Was the normal self-check function in operation when power has turned on? Pass Fail
4 Were the [BATTERY] indictors lit when AC power was connected? Pass Fail
5 Was the [AC/DC] indicator lit when AC power was turned on? Pass Fail
6 With the door open, did the MIDPRESS finger section move momentarily? Pass Fail
7 With the tubing clamp disengaged, close and open the door, Was the tubing
Pass Fail
clamp is engaged?
8 Does the [START] & [STOP/SILENCE] buttons function normally? Pass Fail
WARNING Seize all pump operation when any malfunction has been discovered.
9-15
SY10-9B002 (Rev.4)
Applicable to:
Lot No.9803
9-16
SY10-9B002 (Rev.4)
Applicable to:
Lot No.9803
9-17