SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
& ANSI/NCSL Z540-1-1994
GLOBAL CALIBRATION SERVICES LLC
502 South Lucile Street
Seattle, WA 98108
Dianne Rake Phone: 206 957 4479
CALIBRATION
Valid To: May 31, 2022 Certificate Number: 2367.01
In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 8:
I. Dimensional
Parameter/Equipment Range CMC2, 4 () Comments
Gage Blocks Up to 1 in (1.76L + 7.2) µin Mahr federal
Up to 4 in (3.0L + 6.0) µin comparator, master
gage blocks
Length Standards (1 to 10) in (5.7L + 70) µin Gage blocks,
SupermicrometerTM
(1 to 29) in (5L + 32) µin Gage blocks, gaging
head with amplifier
Diameter –
External3 Up to 10 in (5.7L + 70) µin SupermicrometerTM and
gage blocks
Up to 18 in (5L + 57) µin Electronic gaging amp,
gage block set, granite
surface plate
Internal Up to 8 in (2.3L + 100) µin Comparator and gage
blocks
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 1 of 18
Parameter/Equipment Range CMC2, 4 () Comments
Flatness3 – Optical Quality Up to 0.5 in diameter 5.6 µin Optical flats
Height Gages3 Up to 24 in (0.2L + 700) µin Gage blocks, surface
plate
Calipers3 – Up to 24 in (20L + 200) µin Gage blocks
Internal Measurement 0.00032 in
Micrometers Up to 1 in (-10L + 36) µin Gage blocks
(2 to 6) in (3L + 106) µin
(7 to 20) in (60L + 80) µin
Length Indicators3 –
Dial, Digital Up to 1 in 33 µin Gage blocks
Dial Test Up to 0.060 in 60 µin
Dial Indicator Calibrators3 Up to 1 in 43 µin Gage blocks
Gage Head/Amplifier (0.0001 to 0.2) in 20 µin Gage blocks
Optical Comparators –
X Axis 2 in, 4 in, 6 in 0.00013 in Magnification checker
Y Axis 1 in, 2 in, 3 in 0.00013 in
Protractor 90°, 180°, 270°, 360° 0.02°
Screen to Table X and Y 0.01° 0.1875 in sphere
Alignment
Thread Plug –
Lead Up to 2 in 0.0002 in Optical comparator
Flank Angle 60° 0.054°
Pitch Diameter (0.06 to 1) in (-18L + 104) µin Thread wires and
SupermicrometerTM
(UMM)
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 2 of 18
Parameter/Equipment Range CMC2, 4 () Comments
V Blocks –
V Central Up to (2 ¼ x 2 ¼ x 3) 58 µin Plain plug gage,
in electronic gaging amp
V Parallel 86 µin granite surface plate,
gaging amp, plain plug
gage
Precision Levels –
Concavity 0.0005 in/ft 17L µin Electronic gaging amp,
Left-Right Reading 6.1L µin granite surface plate,
gage blocks
Supermicrometers Linearity to 1 in (0.08L + 25) µin Gage blocks
Gage Block Comparator –
Measuring Range (0.1001 to 0.1005) in 5 µin Grade 00 gage blocks
Calibration Constant 1 in 7.3 µin Grade 00 gage blocks
Lower Contact Pressure 25 g 0.58 g
Upper Contact Pressure 85 g 1.4 g
Internal Comparator A & B Scales (+ and -) 14 µin Grade 0 gage block set
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 3 of 18
II. Electrical – DC/Low Frequency
Parameter/Equipment Range CMC2, 5, 6, 7 (±) Comments
DC Voltage3 – (0 to 330) mV 55 µV/V + 1.2 µV Fluke 5520A/SC1100
Generate (0 to 3.3) V 20 µV/V + 2.4 µV
(3.3 to 33) V 21 µV/V + 24 µV
(33 to 330) V 24 µV/V + 180 µV
(330 to 1000) V 29 µV/V + 1.8 mV
(0 to 220) mV 14 µV/V + 0.4 µV Fluke 5720A
(0.22 to 2.2) V 7 µV/V + 0.7 µV
(2.2 to 22) V 5 µV/V + 2.5 µV
(22 to 220) V 7 µV/V + 40 µV
1000 V 8 µV/V + 400 µV
DC Voltage3 – (0 to 100) mV 26 µV/V + 0.3 µV HP 3458A
Measure 100.1 mV to 1 V 12 µV/V + 0.3 µV
(1.1 to 10) V 9.3 µV/V + 0.5 µV
(10.1 to 100) V 12 µV/V + 30 µV
(100.1 to 1000) V 15 µV/V + 0.1 mV
(1 to 5) kV 3.6 VDC Fluke 80E-10
DC Current3 – 330 µA to 3.3 mA 0.02 % + 0.024 µA Fluke 5520A/SC1100
Generate (3.3 to 33) mA 0.02 % + 0.024 µA multifunction
(33 to 330) mA 0.02 % + 0.024 µA calibrator
330 mA to 1.1 A 0.04 % + 47 µA
(1.1 to 3) A 0.06 % + 47 µA
(3 to 11) A 0.14 % + 580 µA
220 µA 84 µA/A + 6 nA Fluke 5720A
2 mA 80 µA/A + 7 nA
20 mA 65 µA/A + 40 nA
200 mA 73 µA/A + 0.7 nA
1A 0.012 % + 12 µA
DC Current3 – Up to 100 µA 0.04 mA/A + 0.8 nA HP 3458A multimeter
Measure 100.1 µA to 1 mA 0.07 mA/A + 5 nA
(1.1 to 10) mA 0.03 mA/A + 50 nA
(10.1 to 100) mA 0.05 mA/A + 0.5 µA
100.1 mA to 1 A 0.14 mA/A + 10 µA
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 4 of 18
Parameter/Equipment Range CMC2, 5, 6, 7 (±) Comments
DC Power3 –
33 mV to 1020 V 330 µA to 3.3 mA 0.02 % Fluke 5520A/SC1100
(3.3 to 33) mA 0.02 %
(33 to 330) mA 0.02 %
330 mA to 1.1 A 0.06 %
(1.1 to 3) A 0.08 %
(3 to 11) A 0.15 %
(11 to 20.5) A 0.25 %
Resistance3 – Generate (0 to 11) Ω 0.02 % + 0.0012 Ω Fluke 5520A/SC1100
(11 to 33) Ω 0.01 % + 0.0017 Ω
(33 to 110) Ω 0.01 % + 0.0016 Ω
(110 to 330) Ω 34 µΩ/Ω + 0.0023 Ω
330 Ω to 1.1 kΩ 40 µΩ/Ω + 0.0023 Ω
(1.1 to 3.3) kΩ 0.01 % + 0.024 Ω
(3.3 to 11) kΩ 41 µΩ/Ω + 0.024 Ω
(11 to 33) kΩ 0.01 % + 0.24 Ω
(33 to 110) kΩ 41 µΩ/Ω + 0.24 Ω
(110 to 330) kΩ 0.01 % + 2.6 Ω
330 kΩ to 1.1 MΩ 47 µΩ/Ω + 3 Ω
(1.1 to 3.3) MΩ 0.01 % + 37 Ω
(3.3 to 11) MΩ 0.02 % + 64 Ω
(11 to 33) MΩ 0.05 % + 3 kΩ
(33 to 110) MΩ 0.07 % + 3.6 kΩ
Fixed Points 1Ω 0.012 % Fluke 5720A
1.9 Ω 0.012 %
10 Ω 29 µΩ/Ω
19 Ω 31 µΩ/Ω
100 Ω 13 µΩ/Ω
190 Ω 14 µΩ/Ω
1 kΩ 12 µΩ/Ω
1.9 kΩ 10 µΩ/Ω
10 kΩ 11 µΩ/Ω
19 kΩ 10 µΩ/Ω
100 kΩ 13 µΩ/Ω
190 kΩ 13 µΩ/Ω
1 MΩ 24 µΩ/Ω
1.9 MΩ 25 µΩ/Ω
10 MΩ 47 µΩ/Ω
19 MΩ 56 µΩ/Ω
100 MΩ 0.013 %
Resistance3 – Measure Up to 10 Ω 31 µΩ/Ω + 50 µΩ HP 3458A
(10.1 to 100) Ω 22 µΩ/Ω + 0.05 mΩ
100.1 Ω to 1 kΩ 14 µΩ /Ω + 0.5 mΩ
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 5 of 18
Parameter/Equipment Range CMC2, 5, 6, 7 (±) Comments
Resistance3 – Measure (1.1 to 10) kΩ 14 µΩ /Ω + 5 mΩ HP 3458A
(cont) (10.1 to 100) kΩ 16 µΩ /Ω + 50 mΩ
100.1 kΩ to 1 MΩ 27 µΩ /Ω + 20 Ω
(1.1 to 10) MΩ 78 µΩ /Ω + 0.1 kΩ
(10.1 to 100) MΩ 600 µΩ /Ω + 1 kΩ
AC Power3
@ (45 to 65) Hz –
(33 to 330) mV (3.3 to 9) mA 0.2 % Fluke 5520A/SC1100
(9 to 33) mA 0.15 %
PF = 1 (33 to 90) mA 0.37 %
(90 to 330) mA 0.26 %
(330 to 900) mA 0.13 %
900 mA to 1.1 A 0.11 %
(1.1 to 3) A 0.1 %
(3 to 11) A 0.25 %
(11 to 20.5) A 0.31 %
330 mV to 3.3 V (3.3 to 9) mA 0.19 %
(9 to 33) mA 0.15 %
(33 to 90) mA 0.37 %
(90 to 330) mA 0.26 %
(330 to 900) mA 0.12 %
900 mA to 1.1 A 0.1 %
(1.1 to 3) A 0.1 %
(3 to 11) A 0.39 %
(11 to 20.5) A 0.3 %
(3.3 to 33) V (3.3 to 9) mA 0.19 %
(9 to 33) mA 0.15 %
(33 to 90) mA 0.37 %
(90 to 330) mA 0.26 %
(330 to 900) mA 0.12 %
900 mA to 1.1 A 0.1 %
(1.1 to 3) A 0.1 %
(3 to 11) A 0.25 %
(11 to 20.5) A 0.3 %
(33 to 1020) V (3.3 to 9) mA 1.2 %
(9 to 33) mA 1.2 %
(33 to 90) mA 1.3 %
(90 to 330) mA 1.2 %
(330 to 900) mA 1.2 %
900 mA to 1.1 A 1.2 %
(1.1 to 3) A 1.2 %
(3 to 11) A 1.2 %
(11 to 20.5) A 1.3 %
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 6 of 18
Parameter/Range Frequency CMC2, 5 (±) Comments
Capacitance3 – Generate
(0.19 to 0.4) nF 10 Hz to 10 kHz 4.5 % + 0.012 nF Fluke 5520A/SC1100
(0.4 to 1.1) nF 3.4 % + 0.012 nF multifunction
calibrator
(1.1 to 3.3) nF 10 Hz to 3 kHz 1.4 % + 0.012 nF
(3.3 to 11) nF 10 Hz to 1 kHz 0.57 % + 0.012 nF
(11 to 33) nF 1 % + 0.12 nF
(33 to 110) nF 0.47 % + 0.12 nF
(110 to 330) nF 0.54 % + 0.35 nF
330 nF to 1.1 µF (10 to 600) Hz 0.47 % + 1.2 nF
(1.1 to 3.3) µF 0.54 % + 3.5 nF
(3.3 to 11) µF (10 to 150) Hz 0.6 % + 12 nF
(11 to 33) µF (10 to 120) Hz 0.74 % + 35 nF
(33 to 110) µF (10 to 80) Hz 0.87 % + 120 nF
(110 to 330) µF Up to 50 Hz 0.82 % + 350 nF
AC Voltage3 – Generate
(1 to 33) mV (10 to 45) Hz 0.13 % + 7 µV Fluke 5520A/SC1100
45 Hz to 10 kHz 0.05 % + 7 µV
(10 to 20) kHz 0.05 % + 7 µV
(20 to 50) kHz 0.16 % + 7 µV
(50 to 100) kHz 0.52 % + 14 µV
(100 to 500) kHz 1.3 % + 58 µV
(33 to 330) mV (10 to 45) Hz 0.05 % + 9.3 µV
45 Hz to 10 kHz 0.03 % + 9.3 µV
(10 to 20) kHz 0.03 % + 9.3 µV
(20 to 50) kHz 0.06 % + 9.3 µV
(50 to 100) kHz 0.14 % + 37 µV
(100 to 500) kHz 0.33 % + 81 µV
330 mV to 3.3 V (10 to 45) Hz 0.04 % + 9.3 µV
45 Hz to 10 kHz 0.03 % + 9.3 µV
(10 to 20) kHz 0.03 % + 9.3 µV
(20 to 50) kHz 0.05 % + 9.3 µV
(50 to 100) kHz 0.11 % + 37 µV
(100 to 500) kHz 0.37 % + 81 µV
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 7 of 18
Parameter/Range Frequency CMC2, 5 (±) Comments
AC Voltage3 – Generate
(cont)
(3.3 to 33) V (10 to 45) Hz 0.04 % + 760 µV Fluke 5520A/SC1100
45 Hz to 10 kHz 0.02 % + 700 µV
(10 to 20) kHz 0.04 % + 700 µV
(20 to 50) kHz 0.05 % + 700 µV
(50 to 100) kHz 0.13 % + 2.4 mV
(33 to 330) V 45 Hz to 1 kHz 0.03 % + 2.4 mV
(1 to 10) kHz 0.04 % + 7 mV
(10 to 20) kHz 0.04 % + 7 mV
(20 to 50) kHz 0.05 % + 7 mV
(50 to 100) kHz 0.32 % + 59 mV
(330 to 1020) V 45 Hz to 1 kHz 0.04 % + 12 mV
(1 to 5) kHz 0.04 % + 12 mV
(5 to 10) kHz 0.04 % + 12 mV
Fluke 5720A
2.2 mV 40 Hz to 20 kHz 1.3 % + 4 µV
(20 to 50) kHz 3.9 % + 4 µV
(50 to 100) kHz 1.1 % + 5 µV
22 mV 40 Hz to 20 kHz 0.05 % + 4 µV
(20 to 50) kHz 0.09 % + 4 µV
(50 to 100) kHz 0.15 % + 5 µV
(100 to 300) kHz 0.19 % + 10 µV
(300 to 500) kHz 0.3 % + 20 µV
200 mV (10 to 20) Hz 0.042 % + 12 µV
40 Hz to 20 kHz 0.018 % + 7 µV
(20 to 50) kHz 0.032 % + 7 µV
(50 to 100) kHz 0.1 % + 17 µV
(100 to 300) kHz 0.13 % + 20 µV
(300 to 500) kHz 0.2 % + 25 µV
2.2 V (10 to 20) Hz 31 µV/V + 40 µV
(20 to 40) Hz 0.013 % + 15 µV
40 Hz to 20 kHz 74 µV/V + 8 µV
(20 to 50) kHz 0.012 % + 10 µV
(50 to 100) kHz 0.019 % + 30 µV
(100 to 300) kHz 0.07 % + 80 µV
(300 to 500) kHz 0.15 % + 200 µV
22 V (10 to 20) Hz 0.034 % + 400 µV
(20 to 40) Hz 0.013 % + 150 µV
40 Hz to 20 kHz 0.007 % + 50 µV
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 8 of 18
Parameter/Range Frequency CMC2, 5, 6 () Comments
AC Voltage3 – Generate
(cont)
22 V (20 to 50) kHz 0.012 % + 100 µV Fluke 5720A
(50 to 100) kHz 0.017 % + 200 µV
(100 to 300) kHz 0.05 % + 600 µV
(300 to 500) kHz 0.15 % + 2 V
220 V (10 to 20) Hz 0.034 % + 4 mV
(20 to 40) Hz 0.013 % + 1.5 mV
40 Hz to 20 kHz 78 µV/V + 0.6 mV
(20 to 50) kHz 0.014 % + 1 mV
(50 to 100) kHz 0.023 % + 2.5 mV
1100 V (15 to 50) Hz 0.04 % + 16 mV
50 Hz to 1 kHz 95 µV/V + 3.5 mV
AC Voltage3 – Measure
Up to 10 mV (1 to 40) Hz 0.048 % + 3 µV HP 3458A
40.1 Hz to 1 kHz 0.046 % + 1.1 µV
(1.1 to 20) kHz 0.055 % + 1.1 µV
(20.1 to 50) kHz 0.59 % + 1.1 µV
(50.1 to 100) kHz 0.59 % + 1.1 µV
(100.1 to 300) kHz 4.6 % + 2 µV
(10 to 100) mV (1 to 40) Hz 0.011 % + 4 µV
40.1 Hz to 1 kHz 0.011 % + 2 µV
(1.1 to 20) kHz 0.019 % + 2 µV
(20.1 to 50) kHz 0.095 % + 2 µV
(50.1 to 100) kHz 0.095 % + 2 µV
(100.1 to 300) kHz 0.36 % + 10 µV
100 mV to 1 V (1 to 40) Hz 0.026 % + 40 µV
40.1 Hz to 1 kHz 0.011 % + 20 µV
(1.1 to 20) kHz 0.019 % + 20 µV
(20.1 to 50) kHz 0.037 % + 20 µV
(50.1 to 100) kHz 0.097 % + 20 µV
(100.1 to 300) kHz 0.36 % + 0.1 mV
(300 to 500) kHz 1.2 % + 0.1 mV
(1 to 10) V (1 to 40) Hz 0.011 % + 0.4 mV
40.1 Hz to 1 kHz 0.011 % + 0.2 mV
(1.1 to 20) kHz 0.019 % + 0.2 mV
(20.1 to 50) kHz 0.037 % + 0.2 mV
(50.1 to 100) kHz 0.095 % + 0.2 mV
(100.1 to 300) kHz 0.36 % + 1 mV
(300 to 500) kHz 1.2 % + 1 mV
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 9 of 18
Parameter/Range Frequency CMC2, 6 () Comments
AC Voltage3 – Measure
(cont)
(10 to 100) V (1 to 40) Hz 0.026 % + 4 mV HP 3458A
40.1 Hz to 1 kHz 0.026 % + 2 mV
(1.1 to 20) kHz 0.026 % + 2 mV
(20.1 to 50) kHz 0.043 % + 2 mV
(50.1 to 100) kHz 0.14 % + 2 mV
(100 to 700) V 1 Hz to 1 kHz 0.05 % + 0.02 V
AC Current3 – Generate
(29 to 330) µA 10 Hz 0.31 % + 0.12 µA Fluke 5520A/SC110
20 Hz 0.24 % + 0.12 µA
(20 to 45) Hz 0.24 % + 0.12 µA
45 Hz to 1 kHz 0.21 % + 0.12 µA
(1 to 5) kHz 0.47 % + 0.17 µA
(5 to 10) kHz 1.2 % + 0.23 µA
(10 to 30) kHz 2.4 % + 0.46 µA
330 µA to 3.3 mA 10 Hz 0.27 % + 0.17 µA
20 Hz 0.17 % + 0.17 µA
(20 to 45) Hz 0.14 % + 0.17 µA
45 Hz to 1 kHz 0.28 % + 0.23 µA
(1 to 5) kHz 0.68 % + 21 µA
(5 to 10) kHz 1.4 % + 0.69 µA
(10 to 30) kHz 1.4 % + 0.69 µA
(3.3 to 33) mA 10 Hz 0.25 % + 2.4 µA
20 Hz 0.13 % + 2.4 µA
(20 to 45) Hz 0.06 % + 2.4 µA
45 Hz to 1 kHz 0.12 % + 2.4 µA
(1 to 5) kHz 0.28 % + 3.5 µA
(5 to 10) kHz 0.55 % + 4.7 µA
(10 to 30) kHz 0.55 % + 4.7 µA
(33 to 330) mA 10 Hz 0.25 % + 24 µA
20 Hz 0.13 % + 24 µA
(20 to 45) Hz 0.06 % + 24 µA
45 Hz to 1 kHz 0.19 % + 58 µA
(1 to 5) kHz 0.31 % + 120 µA
(5 to 10) kHz 0.62 % + 240 µA
(10 to 30) kHz 0.62 % + 240 µA
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 10 of 18
Parameter/Range Frequency CMC2, 5 () Comments
AC Current3 – Generate
(cont)
330 mA to 1.1 A (10 to 45) Hz 0.25 % + 120 µA Fluke 5520A/SC110
45 Hz 0.08 % + 120 µA
45 Hz to 1 kHz 0.21 % + 1200 µA
(1 to 5) kHz 4 % + 5.8 mA
(5 to 10) kHz 4 % + 5.8 mA
(1.1 to 3) A (10 to 45) Hz 0.25 % + 120 µA
45 Hz 0.08 % + 120 µA
45 Hz to 1 kHz 0.85 % + 1.2 mA
(1 to 5) kHz 3.6 % + 5.8 mA
(5 to 10) kHz 3.6 % + 5.8 mA
(3 to 11) A 45 Hz 0.12 % + 2.4 mA Fluke 5520A/SC1100
(45 to 100) Hz 0.17 % + 2.4 mA
100 Hz to 1 kHz 4.1 % + 2.4 mA
(1 to 5) kHz 4.1 % + 2.4 mA
(11 to 20.5) A (45 to 100) Hz 0.25 % + 5.8 mA
100 Hz to 1 kHz 0.29 % + 5.8 mA
(1 to 5) kHz 4.1 % + 5.8 mA
(22 to 220) µA 40 Hz 0.026 % + 8 nA Fluke 5720A
1 kHz 0.048 % + 12 nA
5 kHz 0.21 % + 65 nA
(0.22 to 2.2) mA 40 Hz 99 µA/A + 35 nA
1 kHz 0.037 % + 110 nA
5 kHz 0.21 % + 650 nA
(2.3 to 22) mA 40 Hz 0.017 % + 350 nA
1 kHz 0.017 % + 350 nA
5 kHz 0.03 % + 550 nA
(22.1 to 220) mA 40 Hz 0.017 % + 3.5 nA
1 kHz 0.017 % + 2.5 nA
5 kHz 0.028 % + 3.5 nA
1A 40 Hz 0.037 % + 35 µA
1 kHz 0.063 % + 80 µA
5 kHz 0.063 % + 80 µA
1.5 A 40 Hz 0.034 % + 35 µA
1 kHz 0.059 % + 80 µA
5 kHz 0.059 % + 80 µA
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 11 of 18
Parameter/Range Frequency CMC2, 6 () Comments
AC Current3 – Measure
Up to 100 µA (10 to 20) Hz 0.5 % + 0.03 µA HP 3458A
(20.1 to 45) Hz 0.21 % + 0.03 µA
(45.1 to 100) Hz 0.11 % + 0.03 µA
100.1 Hz to 1 kHz 0.11 % + 0.03 µA
100 µA to 1 mA (10 to 20) Hz 0.49 % + 0.2 µA
(20.1 to 45) Hz 0.2 % + 0.2 µA
(45.1 to 100) Hz 0.09 % + 0.2 µA
100.1 Hz to 5 kHz 0.06 % + 0.2 µA
(5.1 to 20) kHz 0.07 % + 0.2 µA
(1 to 10) mA (10 to 20) Hz 0.49 % + 2 µA
(20.1 to 45) Hz 0.2 % + 2 µA
(45.1 to 100) Hz 0.09 % + 2 µA
100.1 Hz to 5 kHz 0.06 % + 2 µA
(5.1 to 20) kHz 0.09 % + 2 µA
(10 to 100) mA (10 to 20) Hz 0.49 % + 20 µA
(20.1 to 45) Hz 0.2 % + 20 µA
(45.1 to 100) Hz 0.09 % + 20 µA
100.1 Hz to 5 kHz 0.06 % + 20 µA
100 mA to 1 A (10 to 20) Hz 0.49 % + 0.2 mA
(20.1 to 45) Hz 0.21 % + 0.2 mA
(45.1 to 100) Hz 0.12 % + 0.2 mA
100.1 Hz to 5 kHz 0.14 % + 0.2 mA
(1 to 20) A (1 to 40) Hz 3.0 % + 2 mA w/Fluke Y5020 shunt
40.1 Hz to 1 kHz 4.1 % + 2 mA
Parameter/Equipment Range CMC2, 5, 7 () Comments
Oscilloscope
Calibration3 –
Squarewave Signal
50 at 1 kHz 100 mV 0.37 % + 40 µV Fluke 5520A/SC1100
1V 0.65 % + 40 µV
2V 0.54 % + 40 µV
5V 0.44 % + 40 µV
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 12 of 18
Parameter/Equipment Range CMC2, 5, 7 () Comments
Oscilloscope Calibration3 –
Squarewave Signal
1 M at 1 kHz 25 mV 0.48 % + 40 µV Fluke 5520A/SC1100
200 mV 0.087 % + 40 µV
2.2 V 0.29 % + 40 µV
10 V 0.14 % + 40 µV
Time Marker (50 Source 200 ns 0.0003 %
and Period) 1 µs 0.0003 %
50 µs 0.0003 %
200 µs 0.0003 %
1 ms 0.0003 %
10 ms 0.0003 %
Electrical Calibration of
RTD Indicators and
Indicating Systems3 –
Pt 395, 3926, 100 Ω (-200 to 0) ºC 0.035 °C Fluke 5520A/SC1100
(0.1 to 100) ºC 0.019 °C
(100.1 to 300) ºC 0.048 °C
(300.1 to 400) ºC 0.047 °C
(400.1 to 630) ºC 0.056 °C
Pt 395 Only (630 to 800) ºC 0.056 °C
Pt 3916, 100 Ω (-200 to -190) ºC 0.098 °C
(-189.9 to -80) ºC 0.042 °C
(-79.9 to 100) ºC 0.019 °C
(100.1 to 260) ºC 0.043 °C
(260.1 to 300) ºC 0.029 °C
(300.1 to 400) ºC 0.047 °C
(400.1 to 600) ºC 0.059 °C
(600.1 to 630) ºC 0.047 °C
Pt 385, 200 Ω (-200 to 100) ºC 0.042 °C
(100.1 to 260) ºC 0.037 °C
(260.1 to 300) ºC 0.048 °C
(300.1 to 400) ºC 0.068 °C
(400.1 to 600) ºC 0.056 °C
(600.1 to 630) ºC 0.059 °C
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 13 of 18
Parameter/Equipment Range CMC2, 5 () Comments
Electrical Calibration of
RTD Indicators and
Indicating Systems3 –
(cont.)
Pt 385, 500 Ω (-200 to -80) ºC 0.019 °C Fluke 5520A/SC1100
(-79.9 to 260) ºC 0.043 °C
(260.1 to 400) ºC 0.047 °C
(400.1 to 600) ºC 0.052 °C
(600.1 to 630) ºC 0.047 °C
Pt 385, 1000 Ω (-200 to 0) ºC 0.035 °C
(0.1 to 100) ºC 0.019 °C
(100.1 to 260) ºC 0.029 °C
(260.1 to 300) ºC 0.049 °C
(300.1 to 600) ºC 0.047 °C
(600.1 to 630) ºC 0.052 °C
PtNi 385, 120 Ω (-80 to 100) ºC 0.043 °C
(100.1 to 260) ºC 0.029 °C
Electrical Calibration of
Thermocouple Indicators
and Indicating Systems3 –
Type E (-250 to -100) ºC 0.076 °C Fluke 5520A/SC1100
(-99.9 to -25) ºC 0.073 °C
(-24.9 to 350) ºC 0.14 °C
(350.1 to 650) ºC 0.14 °C
(650.1 to 1000) ºC 0.14 °C
Type J (-210 to -100) ºC 0.081 °C
(-99.9 to -30) ºC 0.089 °C
(-29.9 to 150) ºC 0.088 °C
(150.1 to 760) ºC 0.17 °C
(760.1 to 1200) ºC 0.17 °C
Type K (-200 to -100) ºC 0.11 °C
(-99.9 to -25) ºC 0.092 °C
(-24.9 to 120) ºC 0.11 °C
(120.1 to 1000) ºC 0.21 °C
(1000.1 to 1372) ºC 0.21 °C
Type R (0 to 250) ºC 0.77 °C
(250.1 to 400) ºC 0.77 °C
(400.1 to 1000) ºC 1.6 °C
(1000.1 to 1767) ºC 1.6 °C
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 14 of 18
Parameter/Equipment Range CMC2, 5 () Comments
Electrical Calibration of
Thermocouple Indicators
and Indicating Systems3 –
(cont.)
Type S (0 to 250) ºC 0.75 °C Fluke 5520A/SC1100
(250.1 to 1000) ºC 1.6 °C
(1000.1 to 1400) ºC 1.6 °C
(1400.1 to 1767) ºC 1.6 °C
Type T (-250 to -150) ºC 0.12 °C
(-149.9 to 0) ºC 0.093 °C
(0.1 to 120) ºC 0.11 °C
(120.1 to 400) ºC 0.22 °C
III. Mechanical
Parameter/Equipment Range CMC2, 4, 7, 9 () Comments
Dynamometers (2001 to 4000) lb (0.0014 AF + 18) lb TSD1220 and TSD1200
(4001 to 6000) lb (0.0017AF + 25) lb load cell and readout
(6001 to 8000) lb (0.0021 AF + 34) lb
(8001 to 10 000) lb 0.0021 AF + 39) lb
Force – Measuring (0 to 40) lbf 0.31 % Dead weight
Equipment (40.1 to 200) lbf 0.04 % + 0.093 lbf
Scales and Balances3 (1 to 10) g 0.037 g OIML Class E2 weights
(10 to 500) g 0.016 % + 0.16 g
NIST Class F weights
210 g to 45 kg 0.008 % + 0.005 kg
(45.1 to 226) kg 0.002 % + 0.16 kg
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 15 of 18
Parameter/Equipment Range CMC2, 4, 9 () Comments
Torque3 – Measuring Up to 100 ftꞏlbf 0.35 ftꞏlbf AKO torque calibrator
Equipment transducer C
Up to 600 ftꞏlbf 3.7 ftꞏlbf AKO torque calibrator
transducer A
50 inꞏlbf 0.52 inꞏlbf AWS torque calibrator
(50.1 to 150) inꞏlbf 1.2 inꞏlbf transducer B
(150.1 to 250) inꞏlbf 1.6 inꞏlbf
(250.1 to 500) inꞏlbf 4.1 inꞏlbf
Up to 30 ftꞏlbf 0.45 ftꞏlbf AWS torque calibrator
(30.1 to 150) ftꞏlbf 1.1 ftꞏlbf transducer C
(150.1 to 250) ftꞏlbf 1.6 ftꞏlbf
Up to 150 ftꞏlbf 1.9 ftꞏlbf AWS torque calibrator
(150.1 to 360) ftꞏlbf 3.2 ftꞏlbf transducer D
(360.1 to 600) ftꞏlbf 6.6 ftꞏlbf
Pressure/Vacuum3 – (2 to 10) inwc (Inches 0.014 inwc Mensor CPC 8000
Measuring Equipment Water Column)
(10 to 100) psi 0.14 psi Fluke 700 P06
(500.1 to 1000) psi 0.62 psi Fluke 700 P08
(1000.1 to 1500) psi 0.58 psi Fluke 700 P09
(1000.1 to 3000) psi 2.3 psi Fluke 700 P29
(10 to 100) psi 0.11 psi Mensor CPC 8000
(100.1 to 700) psi 0.13 psi pressure calibrator
(700.1 to 1500) psi (0.0001P + 0.1) psi
(1500.1 to 3000) psi (0.0002P + 0.018) psi P = measured pressure
(3000.1 to 6015) psi (0.0002P + 0.003) psi
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 16 of 18
IV. Thermodynamics
Parameter/Equipment Range CMC2, 4, 9 () Comments
Temperature3 – Measure (-20 to 80) ºC 0.15 °C Guildline 9540A digital
thermometer
(20 to 70) ºC 0.0036T + 0.19 ºC Vaisala HMI41/
HMP46
(-40 to 0) °C -0.020T + 0.90 °C
(1 to 150) °C 0.01T + 1.07 °C Omega CL-27
Temperature3 – Ice Point 0.030 °C SPRT, monitoring ice
Measuring Equipment bath
-5 ºC to 60 ºC 0.13 °C SPRT, monitoring Hart
60 ºC to 100 ºC 1.5 °C Scientific 7102 micro-
bath
Type K (-75 to 300) °C -0.00003T + 0.86 °C Omega CL-27
(301 to 1300) °C 0.0005T + 0.73 °C
Type J (-75 to 38) °C -0.002T + 0.46 °C
(39 to 1000) °C 0.0002T + 0.73 °C
Type E (-200 to 100) °C -0.0014T + 0.76 °C
(101 to 700) °C 0.0013T + 0.49 °C
Type T (-190 to 65) °C -0.0015T + 0.62 °C
(66 to 390) °C 0.00006T + 0.52 °C
Relative Humidity3 – 33.1 % RH 0.57 % RH Vaisala HMI41/
Measuring Equipment 75.5 % RH 1.0 % RH HMP46 with RH
reference salts
Relative Humidity3 – Up to 90 % RH 1.4 % RH Vaisala HMI41/
Measure HMP46
_______________________________________
1
This laboratory offers commercial calibration service and field calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 17 of 18
3
Field calibration service is available for this calibration and this laboratory meets A2LA R104 – General
Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations.
Please note the actual measurement uncertainties achievable on a customer's site can normally be
expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such
as the environment at the place of calibration and for other possible adverse effects such as those caused
by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced
by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result
in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.
4
In the statement of CMC, L is the length of the unit under test in inches, D is the diagonal length of the
unit under test in inches, M is the mass of the unit under test in grams, T is the temperature in ºC, and AF
is applied force in pounds.
5
The measurands stated are generated with the Fluke 5520A and 5720A series of instruments. This
capability is suitable for the calibration of the devices intended to measure the stated measurand in the
ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a fraction
of the reading plus a fixed floor specification.
6
The measurands stated are measured with the HP 3458A. This capability is suitable for the calibration of
the devices intended to generate the measurand in the ranges indicated. CMCs are expressed as either a
specific value that covers the full range or as a combination of the fraction of the reading/output plus a
range specification.
7
In the statement of CMC, the value is defined as the percentage of reading unless otherwise noted.
8
This scope meets A2LA’s P112 Flexible Scope Policy.
9
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges
indicated for the listed measurement parameter.
(A2LA Cert. No. 2367.01) Revised 04/28/2022 Page 18 of 18
Accredited Laboratory
A2LA has accredited
GLOBAL CALIBRATION SERVICES LLC
Seattle, WA
for technical competence in the field of
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General
requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of ANSI/NCSL
Z540-1-1994 and R205 – Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates
technical competence for a defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).
Presented this 13th day of February 2020.
_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 2367.01
Valid to May 31, 2022
Revised April 28, 2022
For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.