0% found this document useful (0 votes)
32 views32 pages

2147 01

Uploaded by

rp801113717
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
32 views32 pages

2147 01

Uploaded by

rp801113717
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 32

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

& ANSI/NCSL Z540-1-1994

ENI LABS
3120 Independence Dr.
Fort Wayne, IN 46808
Lynn Gagliano Phone: 260 471 6775

CALIBRATION

Valid To: June 30, 2022 Certificate Number: 2147.01

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 11:

I. Acoustical Quantities

Parameter/Equipment Range CMC2, 8 (±) Comments

Sound Level – Measuring 94 dB 0.28 dB Bruel & Kjaer 4231


Equipment, Fixed Points3 114 dB 0.28 dB

II. Chemical

Parameter/Equipment Range CMC2 (±) Comments

pH Meters, Fixed Points3 4 pH 0.017 pH Standard pH


7 pH 0.012 pH solutions
10 pH 0.019 pH

Conductivity Meters3 ≈10 µS/cm 4.4 µS/cm Standard


≈1413 µS/cm 9.8 µS/cm conductivity
≈10 mS/cm 0.15 mS/cm solutions
≈100.9 mS/cm 2.2 mS/cm

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 1 of 31


III. Dimensional

Parameter/Equipment Range CMC2, 6 (±) Comments

Bore Gauges3 Up to 6 in (47 + 26L) µin Master rings

Calipers3 – Dial, Digital,


Vernier –

Length Up to 18 in (290 + 7.3L) µin Gage blocks


(18 to 48) in (240 + 4.9L) µin

Parallelism Up to 2 in 30 µin

Depth Gauge3 – Dial, (0.001 to 48) in (59 + 7.2L) µin Gage blocks, surface plate
Digital, Vernier

Gage Blocks – Up to 10 in (3.2 + 0.9L) µin Pratt & Whitney


LabmasterTM LMU-175,
Flatness Up to 2 in 2.5 µin gauge blocks, optical flat

Height Gauges3 Up to 18 in (30 + 6.5L) µin Gage blocks


(18 to 36) in (16 + 7.3L) µin

Indicators – Dial & (0.000 01 to 2) in (11 + 8.9L) µin Gage blocks, surface plate
Digital3

Length Standards3 Up to 10 in (3.2 + 1.0L) µin Pratt & Whitney


(10 to 20) in (7.8 +15L) µin LabmasterTM LMU-175,
(20 to 96) in (45 + 17L) µin micrometer head, gage
blocks, P&W PC25

Measuring Rule & Tapes3 Up to 300 in (300 + 4.0L) µin Hexagon Optiv Classic
321

Micrometers3 –

OD Up to 48 in (30 + 13L) µin Gage blocks


ID Up to 24 in (30 + 13L) µin

Flatness Up to 2 in 2.5 µin Optical flat


Parallelism Up to 2 in 2.5 µin

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 2 of 31


Parameter/Equipment Range CMC2, 6 (±) Comments

Optical Comparators3 –

Angularity (0 to 180)° 0.79 arc sec Sine bar, gage blocks

Linearity X-Y Up to 12 in 160 µin Glass scale

Magnification Up to 50X 0.13 % of Magnification master, pin


Magnification gauge

Plain Gage3 –

Pins & Plugs Up to 1 in (7.8 + 9.1L) µin Pratt & Whitney LabmasterTM
(1 to 10) in (2.9 + 14L) µin LMU-175, gage blocks, P&W
(10 to 20) in (11 + 13L) µin PC25

Rings Up to 1 in (6.3 + 12L) µin


(1 to 10) in (2.4 + 16L) µin

Protractors3 Up to 90° 0.060° Surface plate, sine bar, gage


blocks

Sine Bars –

Length (5 to 10) in (76 + 4.6L) µin Hexagon Optiv Classic 321,


Hexagon Global S Chrome

Flatness (5 to 10) in 29 µin Electronic gauge head

Surface Plates3 –

Flatness Up to 170 in diagonal (7.1√D) µin Mahr Federal EMD-832-W3


electronic level

Repeatability (-0.002 to 0.002) in 28 µin Rahn Repeat-O-Meter

Thickness Gauge3 (0.001 to 1) in (8.6 + 30L) µin Pratt & Whitney LabmasterTM
(Feeler Type) LMU-175

Thickness Gauge3 (0.001 to 8) in 590 µin Gage Blocks


(Ultrasonic)

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 3 of 31


Parameter/Equipment Range CMC2, 9 (±) Comments

Thread Gauges –

Thread Ring, Up to 4 in Thread Plug Set using master plug


Adjustable3 tolerance (W) gages. ASME/ANSI B1.2-
1983 & ASME/ANSI B1.3-
Thread Plugs3 2007

Major Diameter Up to 3 in 79 µin


Pratt & Whitney
Pitch Diameter Up to 3 in| 120 µin LabmasterTM LMU-175,
(80 to 6) TPI gauge blocks, thread wires

Thread Wires (80 to 6) TPI 9.0 µin LabmasterTM LMU-175,


gauge blocks

Vision System3 – Linearity Up to 100 mm 0.0020 mm Standard glass scale

IV. Dimensional Testing/Calibration10

Parameter/Equipment Range CMC2, 6 (±) Comments

Dimensional Inspection –

Linear (1D) Up to 10) mm 0.0017 mm Hexagon Optiv Classic 321


(10 to 100) mm 0.0021 mm
(100 to 300) mm 0.0030 mm

Volumetric (300 x 200 x 150) mm 0.0042 mm Hexagon Global S Chrome

3D Fixtures & Artifacts (24 x 32 x 22) in (65 + 2.8L) µin

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 4 of 31


V. Electrical DC/Low Frequency

Parameter/Equipment Range CMC2, 7 (±) Comments

DC Voltage3 –

Fixed – Generate 10 V 0.5 µV/V Fluke 732B

Measure 0 V Short 0.62 µV/V Copper short

(0 to 100) mV 2.7 µV/V + 0.84 µV HP 3458A Opt. 002


(0.1 to 1) V 2.0 µV/V + 5.7 µV digital multimeter
(1 to 10) V 4.4 µV/V + 88 µV
(10 to 100) V 1.5 µV/V + 2.0 mV
(100 to 1000) V 5.3 µV/V + 2.0 mV

(1 to 10) kV 0.35 mV/V + 35 mV Vitrek 4700 high


(10 to 100) kV 0.58 mV/V + 0.50 V voltage multimeter,
Vitrek HVL-100 high
voltage probe

DC Voltage3 –

Generate (0 to 220) mV 9.6 µV/V + 0.63 µV Fluke 5700A


(0.22 to 2.2) V 4.3 µV/V + 4.4 µV
(2.2 to 11) V 4.0 µV/V + 4.5 µV
(11 to 22) V 4.1 µV/V + 4.9 µV
(22 to 220) V 5.7 µV/V + 98 µV
(220 to 1100) V 7.8 µV/V + 661 µV

Transfer (0 to 100) mV 0.98 µV Wavetek 4950


(0.1 to 1) V 6.0 µV
(1 to 10) V 0.037 mV
(10 to 100) V 0.77 mV
(100 to 1000) V 7.7 mV

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 5 of 31


Parameter/Equipment Range CMC2, 7 (±) Comments

DC Current3 –

Measure (0 to 100) µA 20 µA/A + 1.4 nA HP 3458A opt. 002


(0.1 to 1) mA 21 µA/A + 7.9 nA digital multimeter
(1 to 10) mA 21 µA/A + 81 nA
(10 to 100) mA 39 µA/A + 0.78 µA
(0.1 to 1) A 1.1 mA/A + 36 µA
(1 to 10) A 1.1 mA/A + 4.8 mA
(10 to 2000) A 6.1 mA/A + 19 µA

(2000 to 2800) A 18 mA/A + 12 mA Agilent 34461A


current shunt, HP
10 µA 7.9 nA 3458A opt. 002
100 µA 3.3 nA digital mutlimeter
1 mA 0.012 µA LEM PR2000 current
10 mA 0.18 µA clamp, HP 3458A
20 mA 0.15 µA opt. 002 digital
100 mA 1.8 µA multimeter
200 mA 1.8 µA HP 3458A opt. 002,
1A 24 µA Fluke A40B series
1.9 A 26 µA current shunts

Generate (0 to 220) µA 10 nA/A + 7.1 nA Fluke 5700A


(0.22 to 2.2) mA 8.7 nA/A + 13 nA
(2.2 to 22) mA 8.7 nA/A + 49 nA
(22 to 220) mA 1.1 nA/A + 0.84 µA
(0.22 to 2.2) A 2.0 nA/A + 31 µA

(1.1 to 3) A 402 µA/A + 1.5 mA Fluke 5520A


(1.2 (3 to 11) A 1 mA/A + 2.4 mA
(11 to 20.5) A 1.1 mA/A + 2.4 mA

Toroidal (20.5 to 1000) A 3.1 mA/A + 0.60 A Fluke 5520A w/ coil


Non-Toroidal (20.5 to 1000) A 6.0 mA/A + 0.60 A

DC Current3 – Transfer (0 to 100) µA 2.7 nA Wavetek 4950


(1.1 to 1) mA 0.020 µA
(1 to 10) mA 0.21 µA
(10 to 100) mA 2.4 µA
(0.1 to 1) A 0.035 mA

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 6 of 31


Parameter/Equipment Range CMC2, 7 (±) Comments

Resistance3 – Measure

Fixed 10 mΩ 0.89 mΩ Leeds & Northrop 4222


Resistor

100 mΩ 0.89 mΩ Guidline 9330 series resistors


1Ω 0.89 mΩ
10 kΩ 7.4 mΩ

10 MΩ 0.32 kΩ Metrohm P4085 series


1 GΩ 0.56 MΩ resistors
10 GΩ 0.61 MΩ
100 GΩ 0.12 GΩ
1 TΩ 2.6 GΩ

0Ω 47 µΩ Fluke 5720A
1Ω 110 µΩ
1.9 Ω 110 µΩ
10 Ω 35 µΩ
19 Ω 43 µΩ
100 Ω 0.15 mΩ
190 Ω 0.28 mΩ
1 kΩ 1.6 mΩ
1.9 kΩ 2.9 mΩ
10 kΩ 15 mΩ
19 kΩ 29 mΩ
100 kΩ 0.18 Ω
190 kΩ 0.30 Ω
1 MΩ 17 Ω
1.9 MΩ 19 Ω
10 MΩ 71 Ω
19 MΩ 250 Ω
100 MΩ 4.2 kΩ

(0 to 10) Ω 29 µΩ/Ω + 0.58 mΩ HP 3458A Opt. 002, digital


(10 to 100) Ω 12 µΩ/Ω + 0.58 mΩ multimeter
(100 to 1000) Ω 12 µΩ/Ω + 5.8 mΩ
(1 to 10) kΩ 12 µΩ/Ω + 58 mΩ
(10 to 100) kΩ 18 µΩ/Ω + 2.3 Ω
100 kΩ to 1 MΩ 58 µΩ/Ω + 0.12 kΩ
(1 to 100) MΩ 5.8 mΩ/Ω + 12 kΩ
100 MΩ to 1 GΩ 5.8 mΩ/Ω + 12 kΩ

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 7 of 31


Parameter/Equipment Range CMC2, 7 (±) Comments

Resistance3 – Generate (0 to 1) Ω 46 µΩ/Ω + 1.2 mΩ Fluke 5520A


(1 to 1.9) Ω 46 µΩ/Ω + 1.2 mΩ
(1.9 to 10) Ω 46 µΩ/Ω + 1.2 mΩ
(10 to 19) Ω 34 µΩ/Ω + 1.8 mΩ
(19-100) Ω 28 µΩ/Ω + 3.5 mΩ
(100 to 190) Ω 28 µΩ/Ω + 3.5 mΩ
(190 to 1000) Ω 29 µΩ/Ω + 6.3 mΩ
(1000 to 1900) Ω 27 µΩ/Ω + 4.0 mΩ
(1.9 to 10) kΩ 26 µΩ/Ω + 100 mΩ
(10 to 19) kΩ 27 µΩ/Ω + 40 mΩ
(19 to 100) kΩ 26 µΩ/Ω + 1 Ω
(100 to 190) kΩ 33 µΩ/Ω + 3.4 Ω
(0.19 to 1) MΩ 30 µΩ/Ω + 11 Ω
(1 to 1.9) MΩ 63 µΩ/Ω + 0.05 kΩ
(1.9 to 100) MΩ 0.52 mΩ/Ω + 10 kΩ

Resistance3 – Current Up to 80 mΩ 320 µΩ/Ω + 1.3 nΩ Fluke 5520A, HP


Shunt (80 to 800) mΩ 140 µΩ/Ω + 4.7 nΩ 3458A Opt.002
(0.8 to 1000) Ω 81 µΩ/Ω + 16 µΩ digital multimeter

Resistance3 – Transfer Up to 1 Ω 73 µΩ/Ω Wavetek 4950


(1 to 10) Ω 110 µΩ/Ω
(0.01 to 190) kΩ 11 µΩ/Ω
(0.19 to 1.9) MΩ 18 µΩ/Ω
(1.9 to 19) MΩ 26 µΩ/Ω
(19 to 100) MΩ 0.032 %

Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage3 – Measure

(0 to 2.2) mV (10 to 20) Hz 0.43 mV/V + 1.5 µV Fluke 5790A


(20 to 40) Hz 0.25 mV/V + 1.5 µV
(0.04 to 20) kHz 0.25 mV/V + 1.5 µV
(20 to 50) kHz 0.46 mV/V + 2.3 µV
(50 to 100) kHz 0.69 mV/V + 2.9 µV
(100 to 300) kHz 1.4 mV/V + 4.6 µV
(300 to 500) kHz 1.5 mV/V + 9.2 µV
(0.5 to 1) MHz 2.7 mV/V + 9.2 µV

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 8 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage3 – Measure
(cont)

(2.2 to 7) mV (10 to 20) Hz 0.22 mV/V + 1.5 µV Fluke 5790A


(20 to 40) Hz 0.13 mV/V + 1.5 µV
(0.04 to 20) kHz 0.13 mV/V + 1.5 µV
(20 to 50) kHz 0.24 mV/V + 2.3 µV
(50 to 100) kHz 0.36 mV/V + 2.9 µV
(100 to 300) kHz 0.95 mV/V + 4.6 µV
(300 to 500) kHz 1.0 µV/V + 9.2 µV
(0.5 to 1) MHz 2.0 mV/V + 9.2 µV

(7 to 22) mV (10 to 20) Hz 0.15 mV/V + 1.8 µV


(20 to 40) Hz 78 µV/V + 1.8 µV
(0.04 to 20) kHz 78 µV/V + 1.8 µV
(20 to 50) kHz 0.14 mV/V + 3.1 µV
(50 to 100) kHz 0.24 mV/V + 5.9 µV
(100 to 300) kHz 0.63 mV/V + 4.7 µV
(300 to 500) kHz 0.82 mV/V + 9.3 µV
(0.5 to 1) MHz 1.3 mV/V + 9.3 µV

(22 to 70) mV (10 to 20) Hz 99 µV/V + 1.9 µV


(20 to 40) Hz 45 µ/V + 1.9 µV
(0.04 to 20) kHz 45 µV/V + 1.9 µV
(20 to 50) kHz 70 µV/V + 4.1 µV
(50 to 100) kHz 0.14 mV/V + 9.8 µV
(100 to 300) kHz 0.31 mV/V + 5.0 µV
(300 to 500) kHz 0.45 mV/V + 9.5 µV
(0.5 to 1) MHz 1.2 mV/V + 9.6 µV

(70 to 220) mV (10 to 20) Hz 15 µV/V + 77 µV


(20 to 40) Hz 3.6 µV/V + 65 µV
(0.04 to 20) kHz 3.6 µV/V + 65 µV
(20 to 50) kHz 6.6 µV/V + 85 µV
(50 to 100) kHz 7.8 µV/V + 0.15 mV
(100 to 300) kHz 24 µV/V + 0.28 mV
(300 to 500) kHz 59 µV/V + 0.28 mV
(0.5 to 1) MHz 0.44 mV/V + 0.28 mV

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 9 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage3 – Measure
(cont)

(220 to 700) mV (10 to 20) Hz 12 µV/V + 0.17 mV Fluke 5790A


(20 to 40) Hz 3.4 µV/V + 83 µV
(0.04 to 20) kHz 3.4 µV/V + 84 µV
(20 to 50) kHz 24 µV/V + 36 µV
(50 to 100) kHz 39 µV/V + 49 µV
(100 to 300) kHz 44 µV/V + 0.34 mV
(300 to 500) kHz 92 µV/V + 0.35 mV
(0.5 to 1) MHz 0.68 µV/V + 0.36 mV

(0.7 to 2.2) V (10 to 20) Hz 28 µV/V + 0.22 mV


(20 to 40) Hz 15 µV/V + 45 µV
(0.04 to 20) kHz 15 µV/V + 45 µV
(20 to 50) kHz 52 µV/V + 16 µV
(50 to 100) kHz 83 µV/V + 26 µV
(100 to 300) kHz 0.22 mV/V + 50 µV
(300 to 500) kHz 0.13 mV/V + 1.7 mV
(0.5 to 1) MHz 0.89 mV/V + 1.5 mV

(2.2 to 7) V (10 to 20) Hz 70 µ/V + 80 µV


(20 to 40) Hz 25 µ/V + 59 µV
(0.04 to 20) kHz 25 µV/V + 59 µV
(20 to 50) kHz 49 µV/V + 59 µV
(50 to 100) kHz 87 µV/V + 52 µV
(100 to 300) kHz 0.23 mV/V + 40 µV
(300 to 500) kHz 0.45 mV/V + 0.19 mV
(0.5 to 1) MHz 1.3 mV/V + 0.43 mV

(7 to 22) V (10 to 20) Hz 76 µV/V + 0.12 mV


(20 to 40) Hz 34 µV/V + 0.10 mV
(0.04 to 20) kHz 34 µV/V + 0.10 mV
(20 to 50) kHz 62 µV/V + 0.10 mV
(50 to 100) kHz 100 µV/V + 0.30 mV
(100 to 300) kHz 0.20 mV/V + 0.86 mV
(300 to 500) kHz 0.45 mV/V + 0.75 mV
(0.5 to 1) MHz 1.4 mV/V + 0.58 mV

(22 to 70) V (10 to 20) Hz 75 µV/V+ 0.51 mV


(20 to 40) Hz 30 µV/V + 0.64 mV
(0.04 to 20) kHz 30 µV/V + 0.64 mV
(20 to 50) kHz 74 µV/V + 0.51 mV
(50 to 100) kHz 100 µV/V + 1.1 mV
(100 to 300) kHz 0.21 vmV/V + 2.8 mV
(300 to 500) kHz 0.57 mV/V + 1.0 mV

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 10 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage3 – Measure
(cont)

(70 to 220) V (10 to 20) Hz 80 µV/V + 0.51 mV Fluke 5790A


(20 to 40) Hz 35 µV/V + 0.64 mV
(0.04 to 20) kHz 35 µV/V + 0.64 mV
(20 to 50) kHz 78 µV/V + 0.51 mV
(50 to 100) kHz 0.11 mV/V + 1.1 mV
(100 to 300) kHz 0.23 mV/V + 2.8 mV
(300 to 500) kHz 0.58 mV/V + 1.0 mV

(220 to 700) V (10 to 20) Hz 0.11 mV/V + 4.2 mV


(20 to 40) Hz 43 µV/V + 3.4 mV
(0.04 to 20) kHz 43 µV/V + 3.4 mV
(20 to 50) kHz 90 µV/V + 64 mV
(50 to 100) kHz 0.31 mV/V + 0.27 V

(700 to 1000) V (10 to 20) Hz 0.11 mV/V + 4.2 mV


(20 to 40) Hz 44 µV/V + 3.4 mV
(0.04 to 20) kHz 44 µV/V + 3.4 mV
(20 to 50) kHz 0.11 mV/V + 64 mV
(50 to 100) kHz 0.37 mV/V +0.27 V

(1 to 10) kV (0.01 to 600) Hz 1.4 mV/V + 0.12 V Vitrek 4700 high


voltage multimeter

(10 to 75) kV (0.01 to 600) Hz 1.2 mV/V + 0.77 V Vitrek HVL-100


high voltage probe

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 11 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage – Wideband
Measure

(0 to 2.2) mV (10 to 30) Hz 7.9 mV/V + 1.9 µV Fluke 5790A w/


(30 to 120) Hz 7.5 mV/V + 1.9 µV wideband opt.
(0.12 to 1.2) kHz 7.5 mV/V + 1.9 µV
(1.2 to 120) kHz 7.5 mV/V + 1.9 µV
(120 to 500) kHz 7.7 mV/V + 3.0 µV

(2.2 to 7) mV (10 to 30) Hz 7.7 mV/V + 2.6 µV


(30 to 120) Hz 7.2 mV/V + 2.6 µV
(0.12 to 1.2) kHz 7.2 mV/V + 2.6 µV
(1.2 to 120) kHz 7.2 mV/V + 2.6 µV
(120 to 500) kHz 7.4 mV/V + 3.7 µV

(7 to 22) mV (10 to 30) Hz 7.4 mV/V + 4.9 µV


(30 to 120) Hz 6.9 mV/V + 4.9 µV
(0.12 to 1.2) kHz 6.9 mV/V + 4.9 µV
(1.2 to 120) kHz 6.9 mV/V + 4.9 µV
(120 to 500) kHz 7.0 mV/V + 6.6 µV

(22 to 70) mV (10 to 30) Hz 7.4 mV/V + 6.1 µV


(30 to 120) Hz 6.8 mV/V + 6.1 µV
(0.12 to 1.2) kHz 6.8 mV/V + 6.1 µV
(1.2 to 120) kHz 6.8 mV/V + 6.1 µV
(120 to 500) kHz 6.7 mV/V + 6.1 µV

(70 to 220) mV (10 to 30) Hz 7.0 mV/V + 35 µV


(30 to 120) Hz 6.4 mV/V + 35 µV
(0.12 to 1.2) kHz 6.4 mV/V + 35 µV
(1.2 to 120) kHz 6.4 mV/V + 35 µV
(120 to 500) kHz 6.4 mV/V + 35 µV

(220 to 700) mV (10 to 30) Hz 5.8 mV/V + 93 µV


(30 to 120) Hz 5.1 mV/V + 93 µV
(0.12 to 1.2) kHz 5.1 mV/V + 93 µV
(1.2 to 120) kHz 5.1 mV/V + 93 µV
(120 to 500) kHz 5.1 mV/V + 93 µV

(0.7 to 2.2) V (10 to 30) Hz 5.7 mV/V + 0.35 mV


(30 to 120) Hz 5.0 mV/V + 0.35 mV
(0.12 to 1.2) kHz 5.0 mV/V + 0.35 mV
(1.2 to 120) kHz 5.0 mV/V + 0.35 mV
(120 to 500) kHz 5.0 mV/V + 0.35 mV

(2.2 to 7) V (10 to 30) Hz 5.2 mV/V + 0.49 mV


(30 to 120) Hz 4.4 mV/V + 0.49 mV
(0.12 to 1.2) kHz 4.4 mV/V + 0.49 mV
(1.2 to 120) kHz 4.4 mV/V + 0.49 mV
(120 to 500) kHz 4.4 mV/V + 0.49 mV

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 12 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage3 – Generate

(0.005 to 2.2) mV (10 to 20) Hz 0.022 % + 5.2 µV Fluke 5700A


(20 to 40) Hz 0.011 % + 5.2 µV
(0.04 to 20) kHz 0.011 % + 5.2 µV
(20 to 50) kHz 0.038 % + 5.2 µV
(50 to 100) kHz 0.087 % + 8.1 µV
(100 to 300) kHz 0.12 %+ 15 µV
(300 to 500) kHz 0.17 % + 29 µV
(0.5 to 1) MHz 0.36 % + 29 µV

(2.2 to 220) mV (10 to 20) Hz 0.022 % + 9.5 µv


(20 to 40) Hz 0.011 % + 9.6 µv
(0.04 to 20) kHz 0.011 % + 9.4 µV
(20 to 50) kHz 0.018 % + 62 µV
(50 to 100) kHz 0.058 % + 130 µV
(100 to 300) kHz 0.11 % + 37 µV
(300 to 500) kHz 0.18 % + 41 µV
(0.5 to 1) MHz 0.37 % + 93 µV

(0.22 to 2.2) V (10 to 20) Hz 0.016 % + 48 µV


(20 to 40) Hz 0.074 % + 14 µV
(0.04 to 20) kHz 0.075 % + 11 µV
(20 to 50) kHz 29 µV/V + 0.70 mV
(50 to 100) kHz 0.016 % + 0.38 mV
(100 to 300) kHz 0.046 % + 0.15 mV
(300 to 500) kHz 0.11 % + 0.42 mV
(0.5 to 1) MHz 0.22 % + 1.0 mV

(2.2 to 22) V (10 to 20) Hz 0.016 % + 0.38 mV Fluke 5700A


(20 to 40) Hz 73 µV/V + 0.15 mV
(0.04 to 20) kHz 75 µV/V + 0.12 mV
(20 to 50) kHz 0.011 % + 0.82 mV
(50 to 100) kHz 0.024 % + 0.47 mV
(100 to 300) kHz 0.053 % + 1.7 mV
(300 to 500) kHz 0.14 % + 5.0 mV
(0.5 to 1) MHz 0.29 % + 1.0 mV

(22 to 220) V (10 to 20) Hz 0.16 % + 3.9 mV


(20 to 40) Hz 80 µV/V + 1.5 mV
(0.04 to 20) kHz 82 µV/V + 1.2 mV
(20 to 50) kHz 0.022 % + 4.2 mV
(50 to 100) kHz 0.53 % + 9.5 mV
(100 to 300) kHz 0.16 % + 0.1 V
(300 to 500) kHz 0.53 % + 0.1 V
(0.5 to 1) MHz 1.3 % + 0.2 V

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 13 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage3 – Generate

(220 to 750) V (15 to 50) Hz 0.042 % + 19 mV Wavetek 4808


(0.05 to 1) kHz 78 µV/V + 9.2 mV

(750 to 1000) V (10 to 31) Hz 110 µV/V + 340 µV


(32 to 330) Hz 66 µV/V + 210 µV
(0.3 to 10) kHz 56 µV/V + 110 µV
(10 to 33) kHz 65 µV/V + 210 µV
(30 to 100) kHz 110 µV/V + 340 µV
(100 to 330) kHz 620 µV/V + 3.5 mV
(0.3 to 1000) MHz 8.9 mV/V + 14 mV

(1000 to 1100) V (10 to 330) Hz 160 µV/V + 2.3 mV Fluke 5520A


(0.3 to 10) kHz 120 µV/V + 2.2 mV
(10 to 33) kHz 160 µV/V + 2.3 mV

(1 to 33) mV (10 to 45) Hz 99 µV/V + 19 µV


(0.045 to 10) kHz 99 µV/V + 19 µV
(10 to 20) kHz 9.3 µV/V + 250 µV
(20 to 50) kHz 48 µV/V + 140 µV
(50 to 100) kHz 0.045 % + 100 µV
(100 to 500) kHz 0.22 % + 87 µV

(33 to 330) mV (10 to 45) Hz 0.012 % + 0.12 mV


(0.045 to 10) kHz 0.012 % + 0.12 mV
(10 to 20) kHz 13 µV/V + 1.8 mV
(20 to 50) kHz 35 µV/V + 1.1 mV
(50 to 100) kHz 0.045 % + 0.43 mV
(100 to 500) kHz 0.26 % + 0.80 mV

(0.33 to 3.3) V (10 to 45) Hz 0.017 % + 0.71 mV


(0.045 to 10) kHz 0.017 % + 0.71 mV
(10 to 20) kHz 0.027 % + 0.72 mV
(20 to 50) kHz 0.040 % + 0.74 mV
(50 to 100) kHz 0.10 % + 1.9 mV
(100 to 500) kHz 0.28 % + 0.80 mV

(3.3 to 33) V (10 to 45) Hz 0.021 % + 3.0 mV


(0.045 to 10) kHz 0.022 % + 7.8 mV
(10 to 20) kHz 0.024 % + 10 mV
(20 to 50) kHz 0.019 % + 20 mV
(50 to 100) kHz 0.22 % + 62 mV

(330 to 1020) V (0.045 to 1) kHz 0.034 % + 19 mV


(1 to 5) kHz 0.028 % + 20 mV
(5 to 10) kHz 0.033 % + 32 mV

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 14 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Voltage3 – Transfer

1 mV 10 Hz to 30 kHz 3.0 µV Wavetek 4950


(30 to 50) kHz 3.0 µV
(50 to 100) kHz 3.1 µV
300 kHz to 1 MHz 5.0 µV

10 mV 10 Hz to 30 kHz 3.4 µV
(30 to 50) kHz 4.5 µV
(50 to 100) kHz 5.9 µV
300 kHz to 1 MHz 0.016 mV

100 mV 10 Hz to 30 kHz 7.0 µV


(30 to 50) kHz 0.014 mV
(50 to 100) kHz 0.016 mV
(300 to 500) kHz 0.074 mV
500 kHz to 1 MHz 0.082 mV

1V (10 to 40) Hz 0.021 mV


40 Hz to 30 kHz 0.021 mV
(30 to 50) kHz 0.050 mV
(50 to 100) kHz 0.059 mV
300 kHz to 1 MHz 0.43 mV

10 V (10 to 40) Hz 0.21 mV


40 Hz to 30 kHz 0.21 mV
(30 to 50) kHz 0.50 mV
(50 to 100) kHz 0.59 mV
300 kHz to 1 MHz 4.8 mV

19 V 1 kHz 0.51 mV

100 V (10 to 40) Hz 2.4 mV


40 Hz to 30 kHz 2.4 mV
(30 to 50) kHz 5.5 mV
(50 to 100) kHz 5.5 mV
(100 to 200) kHz 16 mV

1000 V (55 to 300) Hz 0.035 V


300 Hz to 1 kHz 0.032 V

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 15 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Current3 – Measure

Up to 100 µA (10 to 20) Hz 1.9 mA/A + 0.56 µA HP 3458 opt. 002


(20 to 45) Hz 0.37 mA/A + 0.58 µA digital multimeter
(45 to 100) Hz 91 µA/A + 0.58 µA
(0.1 to 1) kHz 84 µA/A + 0.57 µA

(1 to 100) mA (10 to 20) Hz 4.7 mA/A + 24 µA


(20 to 45) Hz 1.8 mA/A + 24 µA
(45 to 100) Hz 0.70 mA/A + 24 µA
(0.1 to 5) kHz 0.35 mA/A + 24 µA
(5 to 20) kHz 0.70 mA/A + 24 µA
(20 to 50) kHz 4.7 mA/A + 47 µA
(50 to 100) kHz 6.4 mA/A + 174 µA

(0.1 to 1) A (10 to 20) Hz 4.7 mA/A + 0.24 mA


(20 to 45) Hz 1.9 mA/A + 0.24 mA
(45 to 100) Hz 0.92 mA/A + 0.25 mA
(0.1 to 5) kHz 1.2 mA/A + 0.25 mA
(5 to 20) kHz 3.5 mA/A + 0.24 mA
(20 to 50) kHz 12 mA/A + 0.47 mA

(1 to 3) A 10 Hz to 3 kHz 1.7 mA/A + 5.8 mA Keithley 2015


(3 to 5) kHz 32 mA/A + 13 mA

(3 to 10) A 3 Hz to 10 kHz 33 mA/A + 13 mA Agilent 34461A

(10 to 2800) A 1 Hz to 10 kHz 0.22 mA/A + 1.3 A LEM PR2000 current


clamp, HP 3458A
opt. 002 digital
multimeter

AC Current3 – Measure
(cont)

1 kHz 10 µA 0.78 nA Fluke A40B series


100 µA 4.4 nA shunts, HP 3458A
1 mA 0.053 µA opt. 002 digital
10 mA 0.49 µA multimeter
20 mA 0.43 µA
100 mA 2.1 µA
200 mA 2.4 µA
1A 0.032 mA
1.9 A 0.047 mA

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 16 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Current3 – Generate

(0.02 to 220) µA (10 to 20) Hz 35 µA/A + 0.50 µA Fluke 5700A


(20 to 40) Hz 170 µA/A + 0.048 µA
(0.04 to 1) kHz 170 µA/A + 0.048 µA
(1 to 5) kHz 220 µA/A + 0.22 µA
(5 to 10) kHz 0.14 % + 1.0 µA

(0.22 to 2.2) mA (10 to 20) Hz 210 µA/A + 0.63 µA


(20 to 40) Hz 170 µA/A + 0.48 µA
(0.04 to 1) kHz 170 µA/A + 0.48 µA
(1 to 5) kHz 130 µA/A + 2.4 µA
(5 to 10) kHz 0.15 % + 7.2 µA

(2.2 to 22) mA (10 to 20) Hz 94 µA/A + 17 µA


(20 to 40) Hz 170 µA/A + 3.6 µA
(0.04 to 1) kHz 170 µA/A + 3.6 µA
(1 to 5) kHz 130 µA/A + 16 µA
(5 to 10) kHz 0.057 % + 77 µA

(22 to 220) mA (10 to 20) Hz 0.016 % + 0.20 mA Fluke 5520A


(20 to 40) Hz 0.037 % + 0.12 mA
(0.04 to 1) kHz 0.037 % + 0.12 mA
(1 to 5) kHz 0.044 % + 0.32 mA
(5 to 10) kHz 0.77 % + 0.73 mA

(0.22 to 2.2) A (0.02 to 1) kHz 110 µA/A + 3.9 mA


(1 to 5) kHz 0.034 % + 5.1 mA
(5 to 10) kHz 0.17 % + 12 mA

(1.1 to 3) A (10 to 45) Hz 0.045 % + 4.5 mA


(0.045 to 1) kHz 0.038 % + 5.6 mA
(1 to 5) kHz 0.69 % + 1.2 mA
(5 to 10) kHz 2.9 % + 5.8 mA

(3 to 11) A (45 to 100) Hz 0.16 % + 7.7 mA Fluke 5520A,


(0.1 to 1) kHz 0.16 % + 7.7 mA 5500A/coil
(1 to 5) kHz 3.4 % + 14 mA

(11 to 20.5) A (45 to 100) Hz 0.17 % + 7.7 mA


(0.1 to 1) kHz 0.17 % + 7.7 mA
(1 to 5) kHz 3.4 % + 14 mA

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 17 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

AC Current3 – Generate
(cont)

(20.5 to 150) A (45 to 65) Hz 5.8 mA/A + 0.53 A Fluke 5520A,


Toroidal (65 to 440) Hz 11 mA/A + 0.47 A 5500A/coil

Non-Toroidal (45 to 65) Hz 7.8 mA/A + 0.56 A


(65 to 440) Hz 12 mA/A + 0.51 A

(150 to 1000) A

Toroidal (45 to 65) Hz 7.0 mA/A + 0.59 A


(65 to 440) Hz 12 mA/A + 0.59 A

Non-Toroidal (45 to 65) Hz 8.6 mA/A + 0.59 A


(65 to 440) Hz 13 mA/A + 0.59 A

AC Current3 – Transfer

Up to 100 µA 10 Hz to 1 kHz 0.017 µA Wavetek 4950


(1 to 5) kHz 0.026 µA
(5 to 10) kHz 0.049 µA

100 µA to 1 mA 10 Hz to 1 kHz 0.16 µA


(1 to 5) kHz 0.21 µA
(5 to 10) kHz 0.53 µA

(1 to 10) mA 10 Hz to 1 kHz 1.6 µA


(1 to 5) kHz 2.4 µA
(5 to 10) kHz 6.8 µA

(10 to 100) mA (10 to 55) Hz 0.016 mA


(55 to 300) Hz 0.018 mA
300 Hz to 1 kHz 0.017 mA

(1 to 5) kHz 0.024 mA
(5 to 10) kHz 0.068 mA

100 mA to 1 A 10 Hz to 1 kHz 0.19 mA


(1 to 5) kHz 0.30 mA
(5 to 10) kHz 1.1 mA

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 18 of 31


Parameter/Equipment Range CMC2, 7 (±) Comments

Phase Angle3 –

Measure
(0 to 359.9999) ° 10 Hz to 1 kHz 0.012° North Atlantic PAV
(1 to 20) kHz 0.017° 2250A
(20 to 50) kHz 0.051°
(50 to 200) kHz 0.11°

Generate
(0 to 999.999) ° 1 Hz to 1 kHz 0.028° Clarke Hess 5000
(1 to 50) kHz 0.056°
(50 to 200) kHz 0.068°

Inductance3 –

Fixed 20 mH 0.012 mH General radio 1482


100 mH 0.054 mH series inductors
200 mH 0.12 mH
2H 1.3 mH

Measure (0.1 to 10) µH 0.022 µH HP 4284A


(0.01 to 10) mH 0.21 µH
0.01 to 100) mH 0.02 H
(0.1 to 10) kH 6.4 H
(10 to 100) kH 0.61 kH
General radio 1490D
Generate (1 to 10) mH 9.5 µH
(10 to 100) mH 0.81 mH
(0.1 to 1) H 0.011 H
(1 to 10) H 0.2 H

Capacitance3 – Fixed 1 nF 0.50 pF General radio 1409


2 nF 1.1 pF series standard
10 nF 5.4 pF capacitors
100 nF 7.4 pF
200 nF 0.10 nF
1 µF 0.19 nF

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 19 of 31


Parameter/Equipment Range CMC2, 7 (±) Comments

Capacitance3 –

Measure (0 to 1) nF 1.1 pF HP 4284A


(0.001 to 10) µF 5.4 nF
(0.001 to 10) mF 0.011 mF
(0.01 to 110) mF 0.28 mF

Generate (0.19 to 0.4) nF 0.26 % + 13 pF Fluke 5520A


(0.4 to 1.1) nF 0.56 % + 12 pF
(1.1 to 3.3) nF 0.27 % + 13 pF
(3.3 to 11) nF 0.29 % + 120 pF
(11 to 33) nF 0.27 % + 140 pF
(33 to 110) nF 0.27 % + 390 pF
(110 to 330) nF 0.27 % + 1.3 nF
(0.33 to 1.1) µF 0.27 % + 3.9 nF
(1.1 to 3.3) µF 0.25 % + 15 nF
(3.3 to 11) µF 0.26 % + 42 nF
(11 to 33) µF 0.38 % + 170 nF
(33 to 110) µF 0.45 % + 480 nF
(110 to 330) µF 0.50 %+ 1.4 µF
(0.33 to 1.1) mF 0.44 % + 4.9 µF
(1.1 to 3.3) mF 0.26 % + 36 µF
(3.3 to 11) mF 0.78 % + 49 µF
(11 to 33) mF 0.83 % + 360 µF
(33 to 100) mF 1.1 % + 360 µF

Thermocouple Simulation3 –

Type E (-250 to -100) °C 0.58 °C Fluke 5520A


(-100 to -25) °C 0.20 °C
(-25 to 350) °C 0.18 °C
(350 to 650) °C 0.20 °C
(650 to 1000) °C 0.26 °C

Type J (-210 to -30) ºC 0.30 ºC


(-30 to 50) ºC 0.29 ºC
(50 to 500) ºC 0.16 ºC
(500 to 1200) ºC 0.23 ºC

Type K (-200 to -100) ºC 0.35 ºC


(-100 to -30) ºC 0.19 ºC
(-30 to 120) ºC 0.17 ºC
(120 to 1050) ºC 0.15 ºC
(1050 to 1371.1) ºC 0.23 ºC

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 20 of 31


Parameter/Equipment Range CMC2 (±) Comments

Thermocouple Simulation3 –

Type N (-200 to -100) ºC 0.42 ºC Fluke 5520A


(-100 to -25) ºC 0.24 ºC
(-25 to 120) ºC 0.21 ºC
(120 to 410) ºC 0.20 ºC
(410 to 1300) ºC 0.28 ºC

Type R (0 to 250) ºC 0.62 ºC


(250 to 400) ºC 0.41 ºC
(400 to 1000) ºC 0.37 ºC
(1000 to 1767) ºC 0.43 ºC

Type S (0 to 250) ºC 0.53 ºC


(250 to 1000) ºC 0.41 ºC
(1000 to 1400) ºC 0.41 ºC
(1400 to 1767) ºC 0.49 ºC

Type T (-250 to -150) ºC 0.64 ºC


(-250 to -100) ºC 0.26 ºC
(-100 to 0) ºC 0.18 ºC

(0 to 400) ºC 0.15 ºC

RTD Simulation3 –

Cu427 100 to 260°C 0.34°C Fluke 5520A

Ni120 (-80 to 0) °C 0.085 °C


(0 to 100) °C 0.084 °C
(100 to 260) °C 0.15 °C

Pt 385, 100 Ω (-200 to -80) ºC 0.090 ºC


(-80 to 0) ºC 0.093 °C
(0 to 100) ºC 0.11 ºC
(100 to 300) ºC 0.13 ºC
(300 to 400) ºC 0.14 ºC
(400 to 630) ºC 0.15 ºC
(630 to 800) ºC 0.25 ºC

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 21 of 31


Parameter/Equipment Range CMC2 (±) Comments

RTD Simulation3 –

Pt 385, 200 Ω (-200 to 100) ºC 0.046 ºC Fluke 5520A


(100 to 260) ºC 0.055 ºC
(260 to 300) ºC 0.13 ºC
(300 to 400) ºC 0.14 ºC
(400 to 600) ºC 0.15 ºC
(600 to 630) ºC 0.17ºC

Pt 385, 500 Ω (-200 to -80) ºC 0.042 ºC


(-80 to 100) ºC 0.051 ºC
(100 to 300) ºC 0.061 ºC
(300 to 400) ºC 0.081 ºC
(400 to 600) ºC 0.092 ºC
(600 to 630) ºC 0.12 ºC

Pt 385, 1000 Ω (-200 to 0) ºC 0.031 ºC


(0 to 100) ºC 0.041 ºC
(100 to 260) ºC 0.052 ºC
(260 to 300) ºC 0.062 ºC
(300 to 600) ºC 0.072 ºC
(600 to 630) ºC 0.23 ºC

Pt 3916, 100 Ω (-200 to -190) ºC 0.28 ºC


(-190 to -80) ºC 0.058 ºC
(-80 to 0) ºC 0.067 ºC
(0 to 100) ºC 0.075 ºC
(100 to 260) ºC 0.093 ºC
(260 to 300) ºC 0.11 ºC
(300 to 400) ºC 0.11 ºC
(400 to 600) ºC 0.12 ºC
(600 to 630) ºC 0.24 ºC

Pt 3926, 100 Ω (-200 to 0) ºC 0.066 ºC


(0 to 100) ºC 0.083 ºC
(100 to 300) ºC 0.11 ºC
(300 to 400) ºC 0.12 ºC
(400 to 630) ºC 0.14 ºC

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 22 of 31


Parameter/Equipment Range CMC2 (±) Comments

Oscilloscopes –

Amplitude – DC Signal3
50 Ω Load (0 to 6) V 29 mV/V + 0.58 mV 5520A-SC100
1 MΩ Load (0 to 130) V 590 µV/V + 75 µV multifunction
calibrator w/
Amplitude – Square Wave oscilloscope opt
50 Ω Load 1 mVp-p to 6.6 Vp-p; 29 mVp-p/Vp-p + 75 µVp-p
10 Hz to 10 kHz

1 MΩ Load 1 mVp-p to 130 Vp-p; 280 µVp-p/Vp-p + 75 µVp-p


10 Hz to 10 kHz

Oscilloscopes – Edge (200 to 300) ps 120 ps 5520A-SC100


Functions3 multifunction
calibrator w/
oscilloscope opt

Oscilloscopes –
Bandwidth3

5 mV to 5.5 Vp-p 50 kHz 58 mVp-p/Vp-p + 470 µVp-p 5520A-SC100


50 kHz to 100 MHz 58 mVp-p/Vp-p + 470 µVp-p multifunction
(100 to 300) MHz 70 mVp-p/Vp-p + 470 µVp-p calibrator w/
(300 to 600) MHz 120 mVp-p/Vp-p + 470 µVp-p oscilloscope opt

4 mV to 3.5 Vp-p (600 to 1100) MHz 140 mVp-p/Vp-p + 470 µVp-p

Leveled Sine Wave – 50 kHz to 1.1 GHz 0.58 kHz


Frequency3
2 ns to 20 ms 2.9 µs/s + 0.15 ps
Time Marker 50 ms to 5 s 1.2 ms/s + 58 µs

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 23 of 31


Parameter/Equipment Range CMC2 (±) Comments

Oscilloscopes –
Wave Functions - Square,
Sine, &Triangle3

50 Ω Load 1.8 mVp-p to 2.5 Vp-p; 35 mVp-p/Vp-p + 120 µVp-p 5520A-SC100


10 Hz to 10 kHz multifunction
calibrator w/
1 MΩ Load 1.8 mVp-p to 55 Vp-p; 35 mVp-p/Vp-p + 120 µVp-p oscilloscope
10 Hz to 10 kHz opt

Pulse Characteristics3 –

Pulse Width (4 to 500) ns 2.6 ns

Pulse Period 200 ns to 20 ms 150 ps

Rise Time – Generate 1 kHz to 2 MHz 120 ps


(200 to 300) ps

2 to 10 MHz 120 ps
(200 to 350) ps

IV. Electrical – RF/Microwave

Parameter/Range Frequency CMC2, 7 (±) Comments

RF Power3 – Measure

(-67 to +20) dB 10 MHz to 40 GHz 0.82 dB Anritsu MA2444D

(-35 to +20) dB DC to 100 MHz 0.061 dB Rohde & Schwarz


100 MHz to 4 GHz 0.077 dB NRP-Z55
(4 to 8) GHz 0.091 dB
(8 to 12.4) GHz 0.11 dB
(12.4 to 18) GHz 0.13 dB
(18 to 26.5) GHz 0.11 dB
(26.5 to 30) GHz 0.13 dB
(30 to 35) GHz 0.14 dB
(35 to 40) GHz 0.14 dB

(-30 to +20) dB 50 MHz to 50 GHz 0.11 dB HP 8487A, Agilent


E4418B

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 24 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

RF Power – Generate

(-100 to +24) dB 1 mHz to 4.024 GHz 0.60 dB Fluke 96270A


(-100 to +20) dB (1.4 to 20) GHz 1.2 dB leveled microwave
(-100 to +18) dB (20 to 26.5) GHz 1.2 dB output

Leveled Sine Wave –


Generate

(-130 to -94) dBm (10 to 128) MHz 0.74 dB Fluke 96270A


128 MHz to 3 GHz 1.6 dB

(-94 to -84) dBm 100 kHz to 10 MHz 0.52 dB


(10 to 128) MHz 0.35 dB
(128 to 300) MHz 0.51 dB

300 MHz to 3 GHz 1.1 dB

(-84 to -74) dBm 100 kHz to 10 MHz 0.51 dB


(10 to 128) MHz 0.13 dB
(128 to 300) MHz 0.33 dB
300 MHz to 1.4 GHz 0.54 dB
(1.4 to 4) GHz 1.1 dB

(-74 to -48) dBm 100 kHz to 10 MHz 0.21 dB


(10 to 128) MHz 0.11 dB
(128 to 300) MHz 0.11 dB
300 MHz to 1.4 GHz 0.43 dB
(1.4 to 4) GHz 0.58 dB

(-48 to -17) dBm (10 to 100) kHz 0.032 dB


100 kHz to 128 MHz 0.062 dB
(128 to 300) MHz 0.079 dB
300 MHz to 1.4 GHz 0.22 dB

(1.4 to 3) GHz 0.38 dB


(3 to 4) GHz 0.53 dB

(-17 to +14) dBm (10 to 100) kHz 0.031 dB


100 kHz to 128 MHz 0.059 dB
(128 to 300) MHz 0.079 dB
300 MHz to 1.4 GHz 0.22 dB
(1.4 to 4) GHz 0.33 dB

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 25 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

RF Power3 – Leveled Sine Wave –


Generate (cont)

(+14 to +20) dBm (10 to 100) kHz 0.051 dB Fluke 96270A


100 kHz to 128 MHz 0.058 dB
(128 to 300) MHz 0.11 dB
300 MHz to 1.4 GHz 0.26 dB

(+20 to +24) dBm (10 to 100) kHz 0.031 dB


100 kHz to 128 MHz 0.059 dB

Attenuation3, 5 – Measure

10 Hz to 128 MHz (0 to 55) dB 0.044 dB Fluke 96270A,


(55 to 64) dB 0.057 dB Rohde & Schwarz
(64 to 100) dB 0.17 dB FSP
(100 to 116) dB 0.23 dB

Amplitude Modulation3 –
Measure

Carrier: 150 kHz to 10 MHz


Rate: 50 Hz to 10 kHz Depth: (5 to 99) % 2.7 % HP 8901B
Rate: 20 Hz to 10 kHz Depth: (5 to 99) % 3.8 %

Carrier: 10 MHz to 1.3 GHz


Rate: 50 Hz to 50 kHz Depth: (5 to 99) % 1.9 %
Rate: 20 Hz to 100 kHz Depth: (5 to 99) % 3.8 %

Amplitude Modulation3 – Generate

Carrier: 50 kHz to 4 GHz Depth: (5 to 10) % 1.6 % Fluke 96270A


Rate: 1 Hz to 100 kHz Depth: (10 to 99) % 3.9 %

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 26 of 31


Parameter/Range Frequency CMC2, 7 (±) Comments

Frequency Modulation3 –
Measure

Carrier: 150 kHz to 10 MHz Deviation: 26 mHz/Hz + 310 mHz HP 8901B


Rate: 20 Hz to 10 kHz Up to 40 kHz

Carrier: 10 MHz to 1.3 GHz Deviation: 15 mHz/Hz + 320 mHz


Rate: 50 Hz to 100 kHz Up to 400 kHz

Carrier: 10 MHz to 1.3 GHz Deviation: 59 mHz/Hz + 290 mHz


Rate: 20 Hz to 200 kHz Up to 400 kHz

Frequency Modulation3 –
Generate

Carrier: 9 MHz to 31.25 MHz


Rate: 1 Hz to 50 kHz Deviation: 11 mHz/Hz Fluke
Rate: 50 Hz to 300 kHz 10 Hz to 300 kHz 60 mHz/Hz 96270A

Carrier: 31.25 MHz to 125 MHz


Rate: 1 Hz to 50 kHz Deviation: 11 mHz/Hz
Rate: 50 Hz to 300 kHz 10 Hz to 750 kHz 36 mHz/Hz

Carrier: 125 MHz to 4 GHz


Rate: 1 Hz to 50 kHz Deviation: 11 mHz/Hz
Rate: 50 Hz to 300 kHz 10 Hz to 4.8 MHz 36 mHz/Hz

Phase Modulation3 –
Measure

Carrier: 150 kHz to 10 MHz Deviation: 0.050 rad/rad + 0.067 rad HP 8901B
Rate: 200 Hz to 10 kHz Up to 420 rad

Carrier: 10 MHz to 1.3 GHz Deviation: 0.040 rad/rad + 0.067 rad


Rate: 200 Hz to 20 kHz Up to 420 rad

Amplitude Modulation3 – Sine


Distortion

(10.0 to 89.0) % 50 kHz to 125 MHz -40 dBc Fluke


9640A

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 27 of 31


Parameter/Range Frequency CMC2, 6, 7 (±) Comments

Phase Modulation3 –
Generate

Carrier: 9 MHz to 4 GHz


Rate: 1 Hz to 50 kHz Deviation: 0.011 rad/rad Fluke 96270A
Rate: 50 kHz to 300 kHz Up to 1000 rad 0.036 rad/rad

Total Harmonic Distortion 47 Hz to 100 kHz 1.7 % North Atlantic PAV


2250A

VI. Mechanical

Parameter/Equipment Range CMC2, 6 (±) Comments

Accelerometers3 50 Hz to 2 kHz 3.4 % Modal shop 9100D


(2 to 10) kHz 5.0 % shaker, HP 3458A
opt. 002 digital
multimeter, Endevco
BF11 signal
conditioner

Scales & Balances3 (1 to 50) mg 0.017 mg Class 1 weights


(50 to 500) mg 0.027 mg
(0.5 to 1) g 0.055 mg
(1 to 5) g 0.055 mg
(5 to 50) g 0.16 mg
(50 to 100) g 0.64 mg
(100 to 200) g 0.77 mg
(200 to 2000) g 6.9 mg
(2000 to 6000) g 0.14 g

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 28 of 31


Parameter/Equipment Range CMC2, 6, 8 (±) Comments

Force – Measuring Up to 100 lb 0.052 % Rice Lake RL20000I-100


Equipment3 Rice Lake RL2000I-T10-
(50 to 500) lb 0.090 % 500

(0.5 to 5) klb 0.16 % Interface RL9000TWM-


5K

(10 to 100) klb 0.049 % Omega LCW-100K

Pneumatic Pressure – (-15 to 0) psig 0.0019 PSI Fluke 700PD7, Fluke 744
Measuring Equipment3
Up to 25 psig 0.004 % + 0.0017 PSI Fluke 7252i dual channel
(25 to 100) psig 0.0057 % + 0.0041 PSI pressure controller

Up to 625 psig 0.057 PSI


(625 to 2500) psig 0.70 PSI

Up to 3000 psig 0.33 PSI


(3000 to 10 000) psig 1.1 PSI Wika CPG 2500

Tachometers3 – (5 to 250 000) rpm 12 µRPM/RPM + 0.6R Starrett 7793


Optical

Torque – Measure3 (1 to 9) in·lbf 2.1 % Delta Regis DRBT-10


Sturtevant Richmont
(5 to 50) in·lbf 0.64 % system 8 w/ transducers
(50 to 300) in·lbf 0.38 %
(10 to 150) ft·lbf 0.35 %
(150 to 600) ft·lbf 0.31 %

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 29 of 31


V. Thermodynamics

Parameter/Equipment Range CMC2, 8 (±) Comments

Humidity3 – Measuring (5 to 95) % RH 0.84 % RH Rotronic HF53W-ID,


Equipment Rotronic HC2A probe

Temperature3 – Measuring
Equipment

Thermometers (-20 to 375) ºC 0.088 ºC WIKA CTP5000-450 probe


(375 to 600) °C 0.32 ºC

Infrared (-15 to 0) °C 0.96 °C Fluke 4180 wavelength


(0 to 50) °C 0.99 °C (8 to 14) µm emissivity
(50 to 100) °C 1.0 °C 0.95
(100 to 120) °C 1.2 °C

(35 to 100) °C 0.82 °C Fluke 4181 wavelength


(100 to 200) °C 1.4 °C (8 to 14) µm emissivity
(200 to 350) °C 2.0 °C 0.95
(350 to 500) °C 3.0 °C

Temperature – Measure (-200 to 0) °C 0.084 °C Fluke 5616-12 probe, Fluke


(0 to 100) °C 0.034 °C 1524 reference thermometer
(100 to 300) °C 0.037 °C
(300 to 420) °C 0.085 °C

VI. Time & Frequency

Parameter/Equipment Range CMC2, 8 (±) Comments

Frequency3 –

Measure Up to 20 GHz 0.085 µHz/Hz Agilent 53140

Generate Up to 10 kHz 0.091 µHz/Hz Fluke 96270A


10 kHz to 4 GHz 0.060 µHz/Hz
(4 to 26) GHz 0.058 µHz/Hz

GPS Clock 10 MHz 0.12 mHz Trimble Thunderbolt E

Timers/Stopwatches3 Up to 24 hr 0.065 s/day Timometer 4500

____________________________________________

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 30 of 31


1
This laboratory offers commercial calibration service and field calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3
Field calibration service is available for this calibration and this laboratory meets A2LA R104 – General
Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations.
Please note the actual measurement uncertainties achievable on a customer's site can normally be
expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such
as the environment at the place of calibration and for other possible adverse effects such as those caused
by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced
by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result
in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.
4
Specifications are typical below 10 MHz.
5
Relative to +16 dBm output.
6
In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches;
R is instrument resolution; D is diagonal length; percentages are percentage of reading, unless otherwise
indicated.
7
The measurands stated are generated using the indicated instrument (see Comments). This capability is
suitable for the calibration of the devices intended to measure the measurand in the ranges
indicated. CMC are expressed as either a specific value that covers the full range or as a fraction of the
reading plus a fixed floor specification.
8
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
9
Adjustable thread rings are set to applicable specifications using calibrated master set plug gages.
10
This laboratory meets R205 – Specific Requirements: Calibration Laboratory Accreditation Program for
the types of dimensional tests listed above and is considered equivalent to that of a calibration.
11
This scope meets A2LA’s P112 Flexible Scope Policy.

(A2LA Cert. No. 2147.01) Revised 04/21/2022 Page 31 of 31


Accredited Laboratory
A2LA has accredited

ENI LABS
Fort Wayne, IN
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General
requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of
ANSI/NCSL Z540-1-1994 and R205 – Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation
demonstrates technical competence for a defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 21st day of September 2020.

_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 2147.01
Valid to June 30, 2022
Revised April 21, 2022

For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.

You might also like