SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

& ANSI/NCSL Z540-1-1994

DUNCAN AVIATION, INC.


3701 Aviation Rd
Lincoln, NE 68524
Scott Griess Phone: 402 479 4272

CALIBRATION

Valid To: November 30, 2023 Certificate Number: 3908.01

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 7:

I. Dimensional

Parameter/Equipment Range CMC2, 5 (±) Comments

Calipers3 Up to 6 in (290 + 2.6L) µin Gage blocks


(6 to 24) in (290 + 5.6L) µin

Micrometers3 – Up to 6 in (60 + 9.2L) µin Gage blocks

Dial and Test Indicators3

Up to 20 µin resolution Up to 4 in 14 µin Gage blocks


50 µin resolution 84 µin
100 µin resolution 98 µin

II. Electrical – DC/Low Frequency

Parameter/Equipment Range CMC2, 4 (±) Comments

DC Voltage – Generate3 (0 to 330) mV 22 μV/V + 1 μV Fluke 5520A/SC600


330 mV to 3.3 V 15 μV/V + 2 μV
(3.3 to 33) V 15 μV/V + 20 μV

(A2LA Cert. No. 3908.01) 01/20/2022 Page 1 of 14


Parameter/Equipment Range CMC2, 4, 5 (±) Comments

DC Voltage – Generate3 (33 to 330) V 18 μV/V + 150 μV Fluke 5520A/SC600


(cont) (330 to 1020) V 15 μV/V + 1500 μV

DC Voltage – Measure3 (0 to 100) mV 11 μV/V + 0.3 μV Agilent 3458A


(0.1 to 1) V 11 μV/V + 0.3 μV
(1 to 10) V 9.3 μV/V + 0.5 μV
(10 to 100) V 12 μV/V + 30 μV
(100 to 1000) V 12 μV/V + 100 μV

(1 to 20) kV 2.4 % Fluke 80K-40


(20 to 35) kV 1.2 %
(35 to 40) kV 2.4 %

DC Current – Generate3 (0 to 330) μA 0.012 % + 0.02 µA Fluke 5520A/SC600


330 µA to 3.3 mA 0.009 % + 0.05 µA
(3.3 to 33) mA 0.009 % + 0.25 µA
(33 to 330) mA 0.011 % + 2.5 µA
330 mA to 1.1 A 0.016 % + 40 µA
(1.1 to 3) A 0.031 % + 40 µA
(3 to 11) A 0.048 % + 500 µA
(11 to 20.5) A 0.096 % + 750 µA

Clamp-On Only (16.5 to 1025) A 0.66 % Fluke 5520A with


5500A coil

DC Current – Measure3 (0 to 100) nA 0.72 % + 0.04 nA Agilent 3458A


(0.1 to 1) µA 0.078 % + 0.04 nA
(1 to 10) µA 86 μA/A + 0.1 nA
(10 to 100) uA 25 μA/A + 0.8 nA
(0.1 to 1) mA 27 μA/A + 5 nA
(1 to 10) mA 25 μA/A + 50 nA
(10 to 100) mA 49 μA/A + 0.5 µA
(0.1 to 1) A 0.013 % + 10 µA

(1 to 3) A 0.13 % + 0.6 mA Agilent 34410A

(3 to 20) A 0.024 % Current shunts

(A2LA Cert. No. 3908.01) 01/20/2022 Page 2 of 14


Parameter/Equipment Range CMC2, 4 (±) Comments

Resistance – Generate3 (0 to 11) Ω 72 μΩ/Ω + 1.0 mΩ Fluke 5520A/SC600


(11 to 33) Ω 43 μΩ/Ω + 1.5 mΩ
(33 to 110) Ω 27 μΩ/Ω + 1.4 mΩ
(110 to 330) Ω 24 μΩ/Ω + 2.0 mΩ
330 Ω to 1.1 kΩ 23 μΩ/Ω + 2.0 mΩ
(1.1 to 3.3) kΩ 24 μΩ/Ω + 0.02 Ω
(3.3 to 11) kΩ 24 μΩ/Ω + 0.02 Ω
(11 to 33) kΩ 24 μΩ/Ω + 0.2 Ω
(33 to 110) kΩ 24 μΩ/Ω + 0.2 Ω
(110 to 330) kΩ 27 μΩ/Ω + 2.0 Ω
330 kΩ to 1.1 MΩ 27 μΩ/Ω + 2.0 Ω
(1.1 to 3.3) MΩ 52 μΩ/Ω + 30 Ω
(3.3 to 11) MΩ 0.012 % + 50 Ω
(11 to 33) MΩ 0.029 % + 2.5 kΩ
(33 to 110) MΩ 0.041 % + 3.0 kΩ
(110 to 330) MΩ 0.24 % + 0.10 MΩ
(330 to 1100) MΩ 1.2 % + 0.50 MΩ

Resistance – Measure3 (0 to 10) Ω 36 μΩ/Ω + 0.05 mΩ Agilent 3458A


(10 to 100) Ω 20 μΩ/Ω + 0.5 mΩ
100 Ω to 1 kΩ 13 μΩ/Ω + 0.5 mΩ
(1 to 10) kΩ 12 μΩ/Ω + 5.0 mΩ
(10 to 100) kΩ 12 μΩ/Ω + 50 mΩ
100 kΩ to 1 MΩ 19 μΩ/Ω + 2 Ω
(1 to 10) MΩ 60 μΩ/Ω + 100 Ω
(10 to 100) MΩ 0.059 % + 1.0 kΩ
100 MΩ to 1 GΩ 0.58 % + 10 kΩ

Electrical Calibration
of Thermocouple
Indicators – Generate
and Measure3

Type E (-250 to -200) °C 0.30 °C Fluke 7526A


(-200 to -100) °C 0.15 °C
(-100 to 0) °C 0.11 °C
(0 to 600) °C 0.10 °C
(600 to 1000) °C 0.12 °C

Type J (-210 to 100) ºC 0.17 °C


(-100 to 800) ºC 0.11 °C
(800 to 1200) ºC 0.12 °C

(A2LA Cert. No. 3908.01) 01/20/2022 Page 3 of 14


Parameter/Equipment Range CMC2, 4 (±) Comments

Electrical Calibration
of Thermocouple
Indicators – Generate and
Measure3 (cont)

Type K (-250 to -200) ºC 0.56 °C Fluke 7526A


(-200 to -100) ºC 0.20 °C
(-100 to 500) ºC 0.12 °C
(500 to 800) ºC 0.12 °C
(800 to 1372) ºC 0.16 °C

Type N (-250 to -200) °C 0.88 °C


(-200 to -100) °C 0.28 °C
(-100 to 0) °C 0.15 °C
(0 to 100) °C 0.13 °C
(100 to 800) °C 0.12 °C
(800 to 1300) °C 0.15 °C

Type S (-50 to -25) °C 0.62 °C


(-25 to 0) °C 0.50 °C
(0 to 100) ºC 0.47 °C
(100 to 400) ºC 0.34 °C
(400 to 600) ºC 0.27 °C
(600 to 1000) ºC 0.26 °C
(1000 to 1600) ºC 0.26 °C
(1600 to 1767) ºC 0.31 °C

Type T (-250 to -200) ºC 0.43 °C


(-200 to -100) ºC 0.19 °C
(-100 to 0) ºC 0.13 °C
(0 to 200) ºC 0.11 °C
(200 to 400) ºC 0.11 °C

(A2LA Cert. No. 3908.01) 01/20/2022 Page 4 of 14


Parameter/Equipment Range CMC2, 5 (±) Comments

Electrical Calibration
of RTD Indicators3 –

Pt 385, 100 Ω (-200 to 800) ºC 0.058 °C Fluke 7526A

Pt 3926, 100 Ω (-200 to 630) ºC 0.058 °C

Pt 3916, 100 Ω (-200 to 630) ºC 0.058 °C

Pt 385, 200 Ω (-200 to 400) ºC 0.47 °C


(400 to 630) ºC 0.58 °C

Pt 385, 500 Ω (-200 to 630) ºC 0.20 °C

Pt 385, 1000 Ω (-200 to 630) ºC 0.12 °C

PtNi 385, 120 Ω (-80 to 260) ºC 0.024 °C

Cu 427, 10 Ω (-100 to 260) ºC 0.45 °C

YSI 400 (15 to 50) ºC 0.032 °C

Oscilloscope3 –

Amplitude – DC Signal

50 Ω Load 1 mV to 6.599 V 0.32 % + 40 µV Fluke 5520A/SC600


1 MΩ Load 1 mV to 130 V 0.095 % + 40 µV

Square Wave Signal

50 Ω Load @1kHz 1 mV to 6.599 V 0.30 % + 40 µV


Source
1 MΩ Load @1kHz 1 mV to 130 V 0.32 % + 40 µV
Source

Level Sine Wave

Amplitude (50 kHz 50 kHz 2.2 % + 300 µV


Reference) 50 kHz to 100 MHz 4.0 % + 300 µV
(100 to 300) MHz 4.5 % + 300 µV
(300 to 600) MHz 6.8 % + 300 µV

Flatness 50 kHz to 100 MHz 2.0 % + 100 µV


(50 kHz Reference) (100 to 300) MHz 2.5 % + 100 µV
(300 to 600) MHz 4.8 % + 100 µV

(A2LA Cert. No. 3908.01) 01/20/2022 Page 5 of 14


Parameter/Equipment Range CMC2 (±) Comments

Oscilloscope3 (cont.) –

Time Markers:
Source and Period 5 s to 50 ms (42t ·1000) parts in 106 s t is the time in
Into a 50 Ω Load 20 ms to 2 ns 68 parts in 106 s seconds

Parameter/Range Frequency CMC2, 4 (±) Comments

AC Voltage – Generate3

(0 to 33) mV (10 to 45) Hz 0.071 % + 6 µV Fluke 5520A/SC600


45 Hz to 10 kHz 0.036 % + 6 µV
(10 to 20) kHz 0.038 % + 6 µV
(20 to 50) kHz 0.11 % + 6 µV
(50 to 100) kHz 0.29 % + 12 µV
(100 to 500) kHz 0.68 % + 50 µV

(33 to 330) mV (10 to 45) Hz 0.027 % + 8 µV


45 Hz to 10 kHz 0.016 % + 8 µV
(10 to 20) kHz 0.017 % + 8 µV
(20 to 50) kHz 0.032 % + 8 µV
(50 to 100) kHz 0.068 % + 32 µV
(100 to 500) kHz 0.18 % + 70 µV

(0.33 to 3.3) V (10 to 45) Hz 0.024 % + 50 µV


45 Hz to 10 kHz 0.013 % + 60 µV
(10 to 20) kHz 0.016 % + 60 µV
(20 to 50) kHz 0.024 % + 50 µV
(50 to 100) kHz 0.055 % + 130 µV
(100 to 500) kHz 0.20 % + 600 µV

(3.3 to 33) V (10 to 45) Hz 0.025 % + 650 µV


45 Hz to 10 kHz 0.017 % + 600 µV
(10 to 20) kHz 0.022 % + 600 µV
(20 to 50) kHz 0.030 % + 600 µV
(50 to 100) kHz 0.071 % + 1.6 mV

(33 to 330) V 45 Hz to 1 kHz 0.016 % + 2 mV


(1 to 10) kHz 0.017 % + 6 mV
(10 to 20) kHz 0.022 % + 6 mV
(20 to 50) kHz 0.028 % + 6 mV
(50 to 100) kHz 0.16 % + 50 mV

(330 to 1020) V 45 Hz to 1 kHz 0.024 % + 10 mV


(1 to 5) kHz 0.020 % + 10 mV
(5 to 10) kHz 0.024 % + 10 mV

(A2LA Cert. No. 3908.01) 01/20/2022 Page 6 of 14


Parameter/Range Frequency CMC2, 4 (±) Comments

AC Voltage – Measure3

10 mV 40 Hz to 1 kHz 0.024 % + 1.1 µV Agilent 3458A


(1 to 20) kHz 0.037 % + 1.1 µV
(20 to 50) kHz 0.12 % + 1.1 µV
(50 to 100) kHz 0.58 % + 1.1 µV
(100 to 300) kHz 4.7 % + 2 µV
300 kHz to 1 MHz 1.6 % + 5 µV
(1 to 2) MHz 3.8 % + 7 µV

100 mV 40 Hz to 1 kHz 0.011 % + 2 µV


(1 to 20) kHz 0.017 % + 2 µV
(20 to 50) kHz 0.035 % + 2 µV
(50 to 100) kHz 0.093 % + 2 µV
(100 to 300) kHz 0.35 % + 10 µV
300 kHz to 1 MHz 1.2 % + 10 µV
(1 to 2) MHz 1.7 % + 10 µV
(2 to 4) MHz 4.7 % + 70 µV
(4 to 8) MHz 4.9 % + 80 µV
(8 to 10) MHz 19 % + 100 µV

1V 40 Hz to 1 kHz 0.012 % + 20 µV
(1 to 20) kHz 0.017 % + 20 µV
(20 to 50) kHz 0.036 % + 20 µV
(50 to 100) kHz 0.093 % + 20 µV
(100 to 300) kHz 0.35 % + 0.10 mV
300 kHz to 1 MHz 1.2 % + 0.10 mV
(1 to 2) MHz 1.7 % + 0.10 mV
(2 to 4) MHz 4.6 % + 0.70 mV
(4 to 8) MHz 4.8 % + 0.80 mV
(8 to 10) MHz 19 % + 1.0 mV

10 V (1 to 40) Hz 0.012 % + 0.40 mV


40 Hz to 1 kHz 0.012 % + 0.20 mV
(1 to 20) kHz 0.018 % + 0.20 mV
(20 to 50) kHz 0.036 % + 0.20 mV
(50 to 100) kHz 0.093 % + 0.20 mV
(100 to 300) kHz 0.35 % + 1 mV
300 kHz to 1 MHz 1.2 % + 1 mV
(1 to 2) MHz 1.8 % + 1 mV
(2 to 4) MHz 4.5 % + 7 mV
(4 to 8) MHz 4.8 % + 8 mV
(8 to 10) MHz 19 % + 10 mV

(A2LA Cert. No. 3908.01) 01/20/2022 Page 7 of 14


Parameter/Range Frequency CMC2, 4, 5 (±) Comments

AC Voltage – Measure3
(cont)

100 V 40 Hz to 1 kHz 0.025 % + 2 mV Agilent 3458A


(1 to 20) kHz 0.027 % + 2 mV
(20 to 50) kHz 0.043 % + 2 mV
(50 to 100) kHz 0.14 % + 2 mV

1000 V 40 Hz to 1 kHz 0.047 % + 20 mV


(1 to 20) kHz 0.070 % + 20 mV

(1 to 40) kV 60 Hz 5.9 % Fluke 80K-40

AC Current – Generate3

(29 to 330) μA (10 to 20) Hz 0.17 % + 0.1 µA Fluke 5520A/SC600


(20 to 45) Hz 0.13 % + 0.1 µA
45 Hz to 1 kHz 0.11 % + 0.1 µA
(1 to 5) kHz 0.26 % + 0.15 µA
(5 to 10) kHz 0.65 % + 0.2 µA
(10 to 30) kHz 1.4 % + 0.4 µA

(0.33 to 3.3) mA (10 to 20) Hz 0.17 % + 0.15 µA


(20 to 45) Hz 0.11 % + 0.15 µA
45 Hz to 1 kHz 0.089 % + 0.15 µA
(1 to 5) kHz 0.17 % + 0.2 µA
(5 to 10) kHz 0.41 % + 0.3 µA
(10 to 30) kHz 0.84 % + 0.6 µA

(3.3 to 33) mA (10 to 20) Hz 0.15 % + 2 µA


(20 to 45) Hz 0.084 % + 2 µA
45 Hz to 1 kHz 0.053 % + 2 µA
(1 to 5) kHz 0.082 % + 2 µA
(5 to 10) kHz 0.19 % + 3 µA
(10 to 30) kHz 0.39 % + 4 µA

(33 to 330) mA (10 to 20) Hz 0.15 % + 20 µA


(20 to 45) Hz 0.085 % + 20 µA
45 Hz to 1 kHz 0.055 % + 20 µA
(1 to 5) kHz 0.095 % + 50 µA
(5 to 10) kHz 0.19 % + 100 µA
(10 to 30) kHz 0.44 % + 200 µA

(A2LA Cert. No. 3908.01) 01/20/2022 Page 8 of 14


Parameter/Range Frequency CMC2, 4, 5 (±) Comments

AC Current – Generate3
(cont)

(0.33 to 1.1) A (10 to 45) Hz 0.14 % + 0.1 mA Fluke 5520A/SC600


45 Hz to 1 kHz 0.042 % + 0.1 mA
(1 to 5) kHz 0.47 % + 1 mA
(5 to 10) kHz 2.0 % + 5 mA

(1.1 to 3) A (10 to 45) Hz 0.15 % + 0.1 mA


45 Hz to 1 kHz 0.058 % + 0.1 mA
(1 to 5) kHz 0.47 % + 1 mA
(5 to 10) kHz 2.0 % + 5 mA

(3 to 11) A (45 to 100) Hz 0.18 % + 2 mA


100 Hz to 1 kHz 0.084 % + 2 mA
(1 to 5) kHz 2.4 % + 2 mA

(11 to 20.5) A (45 to 100) Hz 0.12 % + 5 mA


100 Hz to 1 kHz 0.13 % + 5 mA
(1 to 5) kHz 2.4 % + 5 mA

Clamp-On Only

(16.5 to 150) A (45 to 65) Hz 0.72 % Fluke 5520A with


(150 to 1025) A (45 to 65) Hz 0.87 % 5500A coil

(16.5 to 150) A (65 to 440) Hz 1.3 %


(150 to 1025) A (65 to 440) Hz 1.6 %

AC Current – Measure3

(0 to 100) µA (45 to 100) Hz 0.071 % + 0.03 µA Agilent 3458A


100 Hz to 5 kHz 0.072 % + 0.03 µA

100 µA to 1 mA (45 to 100) Hz 0.071 % + 0.2 µA


100 Hz to 5 kHz 0.037 % + 0.2 µA

(1 to 10) mA (45 to 100) Hz 0.071 % + 2 µA


100 Hz to 5 kHz 0.037 % + 2 µA

(A2LA Cert. No. 3908.01) 01/20/2022 Page 9 of 14


Parameter/Range Frequency CMC2, 4, 5 (±) Comments

AC Current – Measure3
(cont)

(10 to 100) mA (45 to 100) Hz 0.071 % + 20 µA Agilent 3458A


100 Hz to 5 kHz 0.036 % + 20 µA

100 mA to 1A (45 to 100) Hz 0.094 % + 0.2 mA


100 Hz to 5 kHz 0.12 % + 0.2 mA

(1 to 3) A 3 Hz to 5 kHz 0.14 % + 1.2 mA Agilent 34410A

(3 to 30) A 45 Hz to 1 kHz 0.14 % Agilent 34410A with


current transformer

Capacitance – Generate3

(0.19 to 0.3999) nF 10 Hz to 10 kHz 0.46 % + 0.01 nF Fluke 5520A/SC600


(0.4 to 1.0999) nF 10 Hz to 10 kHz 0.43 % + 0.01 nF
(1.1 to 3.2999) nF 10 Hz to 3 kHz 0.41 % + 0.01 nF
(3.3 to 10.9999) nF 10 Hz to 1 kHz 0.21 % + 0.01 nF
(11 to 32.9999) nF 10 Hz to 1 kHz 0.21 % + 0.1 nF
(33 to 109.9999) nF 10 Hz to 1 kHz 0.21 % + 0.1 nF
(110 to 329.9999) nF 10 Hz to 1 kHz 0.21 % + 0.3 nF
(0.33 to 1.099 99) μF (10 to 600) Hz 0.21 % + 1 nF
(1.1 to 3.299 99) μF (10 to 300) Hz 0.21 % + 3 nF
(3.3 to 10.9999) μF (10 to 150) Hz 0.22 % + 10 nF
(11 to 32.9999) μF (10 to 120) Hz 0.33 % + 30 nF
(33 to 109.999) μF (10 to 80) Hz 0.38 % + 100 nF
(110 to 329.999) μF (0 to 50) Hz 0.37 % + 300 nF
(0.33 to 1.0999) mF (0 to 20) Hz 0.36 % + 1 µF
(1.1 to 3.2999) mF (0 to 6) Hz 0.35 % + 3 µF
(3.3 to 10.9999) mF (0 to 2) Hz 0.36 % + 10 µF
(11 to 32.9999) mF (0 to 0.6) Hz 0.58 % + 30 µF
(33 to 110) mF (0 to 0.2) Hz 0.84 % + 100 µF

(A2LA Cert. No. 3908.01) 01/20/2022 Page 10 of 14


Parameter/Range Frequency CMC2, 4, 5 (±) Comments

Capacitance – Measure3

Fixed Points

10 pF 300 kHz 0.058 % Keysight E4980AL


100 pF 300 kHz 0.058 %
1000 pF 300 kHz 0.058 %
0.1 µF 120 Hz 0.12 %
10 µF 120 Hz 0.10 %

Ranged

10 pF to <0.1 µF 1 kHz to 100 kHz 0.12 % Keysight E4980AL


(0.1 to 1) µF 120 Hz to 1 kHz 0.12 %
>1 µF to 1 mF 20 Hz to 120 Hz 0.29 %

Parameter/Equipment Range CMC2, 4, 5 (±) Comments

Inductance – Measure3 1 mH to 10 H 0.16 % Keysight E4980AL

III. Electrical – RF/Microwave

Parameter/Equipment Frequency CMC2, 4, 5 (±) Comments

RF Power – Measure

(-70 to -20) dBm 10 MHz to 18 GHz 3.2 % HP 437B/8481D


100 pW to 10 µW

(-30 to 20) dBm 100 kHz to 4.2 GHz 2.2 % HP 437B/8482A


1µW to 100 mW

(A2LA Cert. No. 3908.01) 01/20/2022 Page 11 of 14


IV. Mechanical

Parameter/Equipment Range CMC2, 5, 6 Comments

Pressure3 – Measure

Pneumatic (0.2 to 25) psi 0.0033 % Ruska 2465

Hydraulic (10 to 300) psi 0.030 % Pressurements


(>300 to 1500) psi 0.051 % M2200/3P
(>1500 to 3000) psi 0.051 %
(>3000 to 10 000) psi 0.10 %

Balances3 (0 to 100) g 0.004 % + 0.6R Class 1 weights


(100 to 5000) g 0.000 59 % + 0.6R

Scales3 (1 to 700) lb 0.019 % + 0.6R Class 6 and Class F


weights

Force – Tension and


Compression3 (0 to 5000) g 0.07 % + 0.6R Class 1 and Class 3
weights

(0 to 500) lb 0.07 % + 0.6R Class 6 and Class F


weights

Torque – Measuring (10 to 100) lbf·in 0.17 % Dead weights and torque
Equipment (100 to 600) lbf·in 0.15 % arms
(50 to 200) lbf·ft 0.12 %
(200 to 1000) lbf·ft 0.12 %

Torque – Measure (10 to 100) lbf·in 1.0 % A.K.O. torque system


(100 to 600) lbf·in 0.50 %
(50 to 200) lbf·ft 0.50 %
(200 to 1000) lbf·ft 1.6 %

(A2LA Cert. No. 3908.01) 01/20/2022 Page 12 of 14


V. Time & Frequency

Parameter/Equipment Range CMC2, 5, 6 (±) Comments

Timers/Stopwatches 1 s to 24 hr 0.19 s HP 53132A

Optical Tachometers (0 to 100 000) rpm 0.0018 % Comparison with


(100 000 to 200 000) 0.0088 % Keysight 33250A and
rpm LED

Frequency – Measure 10 Hz to 6 GHz 0.062 µHz/Hz HP 53132A


(> 6 to 12.5) GHz 0.28 µHz/Hz Fluke 910R

Frequency – 0.1 Hz to 80 MHz 2.3 x 10-6 Hz/Hz Keysight 33250A


Measuring Equipment
> 80 MHz to 6 GHz 1.8 x 10-6 Hz/Hz Agilent N5181A

(> 6 to 20) GHz 3.0 x 10-8 Hz/Hz Agilent E8257D

Fixed Point 10 MHz 2.0 x 10-5 Hz Fluke 910R

Time Interval – A to B 1 ns to 105 s 27 x 10-9 sec/sec HP 53132A

VI. Thermodynamics

Parameter/Equipment Range CMC2, 6 (±) Comments

Temperature – (0 to 100) °C 0.094 °C Fluke 7526A with Fluke


Liquid-in-Glass 5618B RTD probe in
Thermometers and temperature bath
Temperature Measuring
Devices3

Temperature – (-200 to 500) °C 0.13 °C Fluke 7526A with Fluke


Measure3 5618B RTD probe

(A2LA Cert. No. 3908.01) 01/20/2022 Page 13 of 14


Parameter/Equipment Range CMC2, 6 (±) Comments

Infrared Temperature 0 °C 1.2 °C Comparison to RTD


– Measuring Equipment3 probe in fixed point
water bath (0 °C)

(20 to 200) °C 3.1 °C Fluke 9132 monitored


(200 to 500) °C 8.1 °C with Fluke 7526A with
Fluke 5618B RTD probe

_________________________
1
This laboratory offers commercial calibration service and field calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMC’s
represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using
a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the
laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences
from the circumstances of the specific calibration.
3
Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA
Scope. Allowance must be made for aspects such as the environment at the place of calibration and for
other possible adverse effects such as those caused by transportation of the calibration equipment. The
usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must
also be considered and this, on its own, could result in the actual measurement uncertainty achievable on
a customer’s site being larger than the CMC.
4
The stated measured values are determined using the indicated instrument (see Comments). This capability
is suitable for the calibration of the devices intended to measure or generate the measured value in the
ranges indicated. CMCs are expressed as either a specific value that covers the full range or as a percent
or fraction of the reading plus a fixed floor specification.
5
In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches,
and R is the resolution of the unit under test and percentages are percentages of reading unless otherwise
noted.
6
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
7
This scope meets A2LA’s P112 Flexible Scope Policy.

(A2LA Cert. No. 3908.01) 01/20/2022 Page 14 of 14


Accredited Laboratory
A2LA has accredited

Duncan Aviation, Inc.


Lincoln, NE
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017 General
requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of ANSI/NCSL
Z540-1-1994 and R205 – Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates
technical competence for a defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 20th day of January 2022.

_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 3908.01
Valid to November 30, 2023