Field Master MS2070A-MS2080A
Field Master MS2070A-MS2080A
UPDATES
The latest service and sales contact information, and updated documents can be downloaded from the Anritsu
Website at: http://www.anritsu.com
1-1 Introduction
This manual provides the recommended test equipment and tools, replaceable parts list, verification
procedures, part replacement procedures, maintenance instructions, and test record templates for the
MS2070A/80A Anritsu Field Master.
Familiarity with the basic operation is assumed (for example, how to change measurement modes, preset the
instrument, and set up measurements).
This manual includes:
• General information including:
• Required Test Equipment and Tools listed in Table 1-2 on page 1-2 and Table 1-3 on page 1-3.
• Replaceable Parts listed in Table 1-4 on page 1-3.
• Performance verification procedures included in Chapter 2, “Spectrum Analyzer Verification”.
• Assembly replacement and troubleshooting procedures:
• Chapter 4, “Assembly Replacement”
• Chapter 5, “Troubleshooting”
• Blank test records are included in Appendix A.
• Copy the blank test records from Appendix A and use them to record measured values. Anritsu
recommends making a copy of the blank test records to document measurements each time a
performance verification is performed. Continuing to document this process each time provides a
detailed history of the instrument performance.
The MS2070A/80A performance verification procedures are provided in Chapter 2 and Chapter 4.
Test records templates are provided in Appendix A. Copy the test records templates and use them to
record the measured values obtained when verifying the performance of the MS2070A/80A. The
Note measurement values taken and documented provide a record of the performance of your instrument.
Anritsu recommends that you make a copy of the test records to document the measurements each
time a Performance Verification is performed. Continuing to document this process each time it is
performed provides a detailed history of instrument performance.
Additional Documentation
Table 1-1. Related Manuals
Updates, if any, can be downloaded from the Library tab on the Anritsu product page:
https://www.anritsu.com/en-us/test-measurement/products/ms2080a
https://www.anritsu.com/en-us/test-measurement/products/ms2070a
Tools Specification
Wrench 5/16”
Torx Screwdriver T10
Torx Screwdriver T 20
Phillips Screwdriver #1
Phillips Screwdriver #2
Nut Driver 5.5 mm
Hex Socket Driver 6 mm (1/4”)
2-1 Introduction
This chapter provides the verification procedures for the MS2070A/80A Field Master. Verification tests that
are associated with other instrument options are described in Chapter 4, “Assembly Replacement”. Record the
measurement results in the test record templates provided in Appendix A, “Test Records”.
This chapter includes the following performance verification procedures:
• Section 2-2 “Frequency Accuracy” on page 2-3
• Section 2-3 “Amplitude Accuracy” on page 2-5
• Section 2-4 “Single Side Band (SSB) Phase Noise” on page 2-8
• Section 2-5 “Second Harmonic Distortion” on page 2-10
• Section 2-6 “Displayed Average Noise Level (DANL)” on page 2-11
• Section 2-7 “Residual Spurs” on page 2-15
• Section 2-8 “Input Related Spurious” on page 2-19
+ Uncertainty
Measurement Point (Reading)
– Uncertainty
Measurement Data
Upper
Specification
Nominal
Lower
Specification
Upper
Specification
Nominal
Lower
Specification
The measurement uncertainty listed in each test record includes the best estimate of the errors contributed by
the measurement, test equipment, standards, and other correction factors (for example, calibration factors and
mismatch error) based on the suggested equipment, the equipment setup, and the prescribed test procedure.
Most of the uncertainties are type-B per ISO/IEC Guide 98-3, Guide to the Expression of Uncertainty in
Measurement (GUM).
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu MG36221A Synthesizer
• Anritsu 15NN50-1.5C RF Coaxial Cable
• Anritsu 34RKNF50 Ruggedized K(m) to N(f) Adapter
• 10 MHz Frequency Reference
• BNC(m) to BNC(m) Coaxial Cable
• SMA(m) to BNC(f) Adapter
Procedure
1. Connect the external 10 MHz frequency reference to the MG36221A.
Equipment Required
• Anritsu MG36221A Synthesizer
• 10 MHz Frequency Reference
• Anritsu 34RKNF50 Adapter
• Anritsu 15NN50-1.5C RF Coaxial Cable
• SMA(m) to BNC(f) Adapter
• BNC(m) to BNC(m) Coaxial Cable
• GPS antenna
Procedure
1. Connect the external 10 MHz Reference to the MG36221A.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu MG36221A Synthesizer
• Weinschel 1870A Power Splitter
• Weinshel 44-10 Attenuator
• Rohde and Schwarz NRP18T Power Sensor
• Anritsu 15NN50-1.5C RF Coaxial Cable
• Anritsu 34NN50A Adapter
• Anritsu 34RKNF50 Adapter
• SMA(m) to BNC(f) Adapter
• BNC(m) to BNC(m) Coaxial Cable
• 10 MHz Frequency Reference
Procedure
1. Connect the external 10 MHz Reference to both the MG36221A and MS2070A/80A.
2. Connect the MG36221A RF Out to the input of the splitter.
3. Connect the NRP18T directly to one output of the splitter.
4. Connect the 10 dB attenuator to the other output of the splitter.
5. Connect the RF input of the MS2070A/80A to the 10 dB attenuator.
6. Set the MS2070A/80A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
7. Set the MS2070A/80A parameters as follows:
• Span: 10 kHz
• Reference Level: 10 dBm
• Attenuation Level: 30 dB
• Preamplifier: Off
• RBW:1 kHz
• VBW: 10 Hz
• Detector Type: Peak
8. Using the Rohde and Schwarz Power Viewer software, set the signal frequency to 50 MHz.
9. On the MG36221A, press RF OFF.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu MG36221A Synthesizer
• SMA(m) to BNC(f) Adapter
• Weinschel 1870A Power Splitter
• Weinschel 44-10 Attenuator, Qty 2
• Weinschel 44-20 Attenuator
• Rohde and Schwarz NRP18T Power Sensor
• Anritsu 15NN50-1.5C RF Coaxial Cable
• Anritsu 34NN50A Adapter, N(m) to N(m)
• Ruggedized 34RKNF50 Adapter, K(m) to N(f)
• 10 MHz Frequency Reference
• BNC(m) to BNC(m) Coaxial Cable
Procedure
1. Connect the external 10 MHz Reference to both the MG36221A and MS2070A/80A.
2. Connect the MG36221A RF Out to the input of the splitter.
3. Connect the NRP18T directly to one output of the splitter.
4. Connect 40 dB of attenuation to the other output of the splitter.
5. Connect the RF input of the MS2070A/80A to the 40 dB attenuation.
6. Set the MS2070A/80A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
7. Set the MS2070A/80A parameters as follows:
• Span: 10 kHz
• Reference Level: –40 dBm
• Attenuation Level: 0 dB
• Preamplifier: On
• RBW: 1 kHz
• VBW: 10 Hz
• Detector Type: Peak
8. Using the Rohde and Schwarz Power Viewer software, set the signal frequency to 100 kHz. Ensure the
MG36221A is not outputting any power, and zero the NRP18T sensor.
9. On the MG36221A, set the frequency to 100 kHz and adjust the power level so the NRP18T
measures –10 dBm. (The MG36221A should be approximately –4 dBm.)
10. On the MS2070A/80A set the center frequency to 100 kHz and perform a peak search. Record the result
in Table A-5, “MS2070A/80A Power Level Accuracy (Preamplifier On)” on page A-2.
11. Repeat Step 8 through Step 10 for the remaining frequencies that apply to the unit being tested, in
Table A-5, “MS2070A/80A Power Level Accuracy (Preamplifier On)” on page A-2.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu MG36221A
• Ruggedized 34RKNF50 Adapter, K(m) to N(f)
• SMA(m) to BNC(f) Adapter
• Anritsu 15NN50-1.5C Coaxial Cable, N(m) to N(m)
• 10 MHz Frequency Reference
• BNC(m) to BNC(m) coaxial cable
Procedure
1. Connect the external 10 MHz Reference to both the MG36221A and MS2070A/80A.
2. Connect the MG36221A RF Out to the MS2070A/80A RF In.
3. Set the MG36221A to output frequency to 1.0 GHz.
4. Set the MG36221A to output power to –2 dBm.
5. Set the MS2070A/80A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
6. Set the MS2070A/80A parameters as follows:
• Reference Level: 0 dBm
• Input Attenuation: 15 dB
• Preamplifier: Off
• RBW: 1 kHz
• VBW: 3 Hz
• Detector Type: Peak
• Averages: 3
• Sweep Points: 601
7. 10 kHz and 100 kHz Offset Settings:
• Center Frequency: 1000.05 MHz
• Span: 110 kHz
8. In the Sweep menu, turn off Continuous Sweep, and choose Sweep to three and allow it to finish.
9. In the Marker menu, use the Peak Search menu to set Marker 1 to the peak at 1 GHz. Exit the Peak
Search menu and set the Marker Mode to Delta. Enter a frequency value of 10 kHz, so a marker is on the
peak at 1 GHz and the delta marker is 10 kHz away. Record the difference in level between the peak at
1 GHz and the noise level 10 kHz away at 1000.010 MHz in Table A-6, “MS2070A/80A Spectral Purity -
SSB Phase Noise Offset from 1 GHz” on page A-3. Then subtract another 30 dB and record this as the
10 kHz Offset Measured Value reading in Table A-6 on page A-3.
For example: –80 dBc measured – 30 dB = –110 dBc/Hz
10. Change the Delta Frequency from 10 kHz to 100 kHz and record the difference in level between the peak
at 1 GHz and the noise level 100 kHz away at 1000.100 MHz in Table A-6, “MS2070A/80A Spectral
Purity - SSB Phase Noise Offset from 1 GHz” on page A-3. Then subtract another 30 dB and record this
as the 100 kHz Offset Measured Value reading in Table A-6, “MS2070A/80A Spectral Purity - SSB Phase
Noise Offset from 1 GHz” on page A-3.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu MG36221A Synthesizer
• Min-Circuits NLP-50 Low Pass Filter
• Anritsu 15NN50-1.5C RF Cable
• 10 MHz Frequency Reference
• BNC(m) to BNC(m) Coaxial Cable
• Ruggedized 34RKNF50 Adapter, K(m) to N(f)
• SMA(m) to BNC(f) Adapter
Procedure
1. Connect the 10 MHz reference to the MG36221A and MS2070A/80A.
2. Connect the output of the MG36221A to the input of the low pass filter and the output of the low pass
filter to the RF In of the MS2070A/80A.
3. Set the MS2070A/80A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
4. Set the MS2070A/80A parameters as follows:
• Center frequency: 50.1 MHz
• Span: 100 kHz
• Reference Level: –20 dBm
• Attenuation Level: 0 dB
• Preamplifier: Off
• RBW: 1 kHz
• VBW: 10 Hz
5. Set the MG36221A frequency to 50.1 MHz and adjust the level so the MS2070A/80A shows –30 dBm at
50.1 MHz. Record this value in Table A-7, “MS2070A/80A Second Harmonic Distortion” on page A-3
6. Set the MS2070A/80A Center Frequency to 100.2 MHz, perform a peak search, and record this value in
Table A-7 on page A-3.
7. Subtract 50.1 MHz value from the 100.2 MHz value and record this as the Second Harmonic Distortion in
Table A-7 on page A-3.
For example, if the 50.1 MHz value is –30 dBm and the 100.2 MHz value is –90 dBm, then:
–90 dBm – (–30 dBm) = –60 dBc
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu 28N50-2 Termination
MS2070A Procedure
1. Attach the Termination to the RF input of the MS2070A/80A.
2. Set the MS2070A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2070A parameters as follows:
• Start Frequency: 10 MHz
• Stop Frequency: 2 GHz
• Reference Level: –20 dBm
• Input Attenuation: 0 dB
• Preamplifier: Off
• RBW: 1 MHz
• VBW: 1 kHz
• VBW Type: Logarithmic
• Detector Type: RMS/Avg
• Sweep Points: 551
• Continuous: Toggle Off
• Press Sweep Once.
4. Using the Marker menu, choose Peak Search, subtract 60 dB from the measurement, and record the
power level in Table A-8, “MS2070A Displayed Average Noise Level (Preamplifier Off)” on page A-3
5. Change the start frequency to 2 GHz and stop frequency to 3 GHz and repeat Step 3 and Step 4 to record
the power level in Table A-9, “MS2080A Displayed Average Noise Level (Preamplifier Off)” on page A-3.
MS2080A Procedure
1. Attach the Termination to the RF input of the MS2080A.
2. Set the MS2080A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2080A parameters as follows:
• Start Frequency: 10 MHz
• Stop Frequency: 2 GHz
• Reference Level: –20 dBm
• Input Attenuation: 0 dB
• Preamplifier: Off
• RBW: 1 MHz
• VBW: 1 kHz
• VBW Type: Logarithmic
• Detector Type: RMS/Avg
• Sweep Points: 551
• Continuous: Toggle Off
• Press Sweep Once.
4. Using the Marker menu, choose Peak Search, subtract 60 dB from the measurement, and record the
power level in Table A-10, “MS2070A Displayed Average Noise Level (Preamplifier On)” on page A-4.
5. Change the start frequency to 3 GHz and stop frequency to 4 GHz and repeat Step 3 and Step 4 to record
the power level in Table A-11, “MS2080A Displayed Average Noise Level (Preamplifier On)” on page A-4.
DANL – Preamplifier On
Use the following procedure to verify the Displayed Average Noise Level (DANL) of the MS2070A/80A with the
Preamplifier On. Record the results in the test records provided in Appendix A.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu 28N50-2 Termination
MS2070A Procedure
1. Attach the Termination to the RF input of the MS2070A.
2. Set the MS2070A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2070A parameters as follows:
• Start Frequency: 10 MHz
• Stop Frequency: 2 GHz
• Reference Level: –50 dBm
• Attenuation Level: 0 dB
• Preamplifier: On
• RBW: 1 MHz
• VBW: 1 kHz
• VBW Type: Logarithmic
• Detector Type: RMS/Avg
• Sweep Points: 551
• Continuous: Toggle Off
• Press Sweep Once.
4. Using the Marker menu, choose Peak Search, subtract 60 dB from the measurement, and record the
power level in Table A-10, “MS2070A Displayed Average Noise Level (Preamplifier On)” on page A-4.
5. Change the start frequency to 2 GHz and stop frequency to 3 GHz and repeat Step 3 and Step 4 to record
the power level in Table A-10, “MS2070A Displayed Average Noise Level (Preamplifier On)” on page A-4.
MS2080A Procedure
1. Attach the Termination to the RF input of the MS2080A.
2. Set the MS2080A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2080A parameters as follows:
• Start Frequency: 10 MHz
• Stop Frequency: 2 GHz
• Reference Level: –50 dBm
• Attenuation Level: 0 dB
• Preamplifier: On
• RBW: 1 MHz
• VBW: 1 kHz
• VBW Type: Logarithmic
• Detector Type: RMS/Avg
• Sweep Points: 551
• Continuous: Toggle Off
• Press Sweep Once.
4. Using the Marker menu, choose Peak Search, subtract 60 dB from the measurement, and record the
power level in Table A-11, “MS2080A Displayed Average Noise Level (Preamplifier On)” on page A-4.
5. Change the start frequency to 2 GHz and stop frequency to 4 GHz and repeat Step 3 and Step 4 to record
the power level in Table A-11, “MS2080A Displayed Average Noise Level (Preamplifier On)” on page A-4.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu 28N50-2 Termination
MS2070A Procedure
1. Attach the Termination to the RF input of the MS2070A.
2. Set the MS2070A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2070A parameters as follows:
• Start Frequency: >10 MHz
• Stop Frequency: 500 MHz
• Reference Level: –50 dBm
• Attenuation Level: 0 dB
• Preamplifier: On
• RBW: 300 Hz
• VBW: 100 Hz
• Sweep Points: 601
• Continuous: Toggle off
• Press Sweep Once
4. Using the Marker menu, choose Peak Search and record the power level in Table A-12, “MS2070A
Residual Spurs (Preamplifier On)” on page A-4.
MS2080A Procedure
1. Attach the Termination to the RF input of the MS2080A.
2. Set the MS2080A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2080A parameters as follows:
• Start Frequency: >10 MHz
• Stop Frequency: 500 MHz
• Reference Level: –50 dBm
• Attenuation Level: 0 dB
• Preamplifier: On
• RBW: 300 Hz
• VBW: 100 Hz
• Sweep Points: 601
• Continuous: Toggle off
• Press Sweep Once.
4. Using the Marker menu, choose Peak Search and record the power level in Table A-13, “MS2080A
Residual Spurs (Preamplifier On)” on page A-5.
5. Refer to Table A-13a, “MS2080A Residual Spurs (Preamplifier On, for Rev1 Option 704 models only)”
on page A-5, for Rev1 Option 704 models.
Equipment Required
• Anritsu 28N50-2 Termination
MS2070A Procedure
1. Attach the Termination to the RF input of the MS2070A.
2. Set the MS2070A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2070A parameters as follows:
• Start Frequency: 10 MHz
• Stop Frequency: 500 MHz
• Reference Level: –50 dBm
• Attenuation Level: 0 dB
• Preamplifier: Off
• RBW: 300 Hz
• VBW: 100 Hz
• Sweep Points: 601
• Continuous: Toggle off
• Press Sweep Once.
4. Using the Marker menu, choose Peak Search and record the power level in Table A-14, “MS2070A
Residual Spurs (Preamplifier Off)” on page A-5.
MS2080A Procedure
1. Attach the Termination to the RF input of the MS2080A.
2. Set the MS2080A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
3. Set the MS2080A parameters as follows:
• Start Frequency: 10 MHz
• Stop Frequency: 500 MHz
• Reference Level: –50 dBm
• Attenuation Level: 0 dB
• Preamplifier: Off
• RBW: 300 Hz
• VBW: 100 Hz
• Sweep Points: 601
• Continuous: Toggle off
• Press Sweep Once.
4. Using the Marker menu, choose Peak Search and record the power level in Table A-15, “MS2080A
Residual Spurs (Preamplifier Off)” on page A-6.
5. Change the Start Frequency to 3 GHz and Stop Frequency to 4 GHz.
6. Repeat Step 3 and Step 4 and record the power level in Table A-15, “MS2080A Residual Spurs
(Preamplifier Off)” on page A-6.
7. Refer to Table A-15a, “MS2080A Residual Spurs (Preamplifier Off, for Rev1 Option 704 models only)”
on page A-6 for Rev1 Option 704 models only.
Note Before continuing, allow a 30 minute warm-up for the internal circuitry to stabilize.
Equipment Required
• Anritsu MG36221A Synthesizer
• Anritsu 34RKNF50 Adapter
• Anritsu 15NNF50-1.5C RF Cable
• 10 MHz Frequency Reference
• BNC(m) to BNC(m) Coaxial Cable
• SMA(m) to BNC(f) Adapter
MS2070A Procedure
1. Connect the 10 MHz Reference to the MG36221A and MS2070A.
2. Connect the output of the MG36221A to the RF In of the MS2070A/80A.
3. Set the MS2070A/80A to spectrum analyzer mode and Preset the instrument (PRESET > Preset Mode).
4. Set the MS2070A parameters as follows:
• Center Frequency: 2086 MHz
• Span: 100 kHz
• Reference Level: –30 dBm
• Attenuation Level: 0 dB
• Preamplifier: Off
• RBW: 1 kHz
• VBW: 10 Hz
5. Set the MG36221A frequency to 2086 MHz, and set power level so –30 dBm is read on the MS2070A at
2086 MHz. Record the measured input signal level in Table A-16, “MS2070A Input Related Spurious”
on page A-6.
6. Change the center frequency of the MS2070A to 400.5 MHz and record the highest spur level in
Table A-16, “MS2070A Input Related Spurious” on page A-6.
MS2080A Procedure
1. Connect the 10 MHz Reference to the MG36221A and MS2080A.
2. Connect the output of the MG36221A to the RF In of the MS2080A.
3. Set the MS2080A to spectrum analyzer mode and preset the instrument (PRESET > Preset Mode).
4. Set the MS2080A parameters as follows:
• Center Frequency: 742 MHz
• Span: 100 kHz
• Reference Level: –30 dBm
• Attenuation Level: 0 dB
• Preamplifier: Off
• RBW: 1 kHz
• VBW: 10 Hz
5. Set the MG36221A frequency to 742 MHz, and set power level so –30 dBm is read on the MS2080A at
742 MHz. Record the highest spur level in Table A-17, “MS2080A Input Related Spurs” on page A-7.
6. Change the center frequency of the MS2080A to 3500.5 MHz and record the highest spur level in
Table A-17, “MS2080A Input Related Spurs” on page A-7.
When inserting the battery, the battery contacts should face up and slide in first. If the battery door
Note
does not close tightly, it means that the battery may have been inserted incorrectly.
6. With the battery access door removed, grasp the battery lanyard and pull the battery straight out of the
unit as illustrated in Figure 3-2.
4-1 Introduction
This chapter describes opening and closing the Field Master case along with basic parts replacement steps.
The sections are as follows:
• “Battery Pack Removal and Replacement”
• “Bumpers Removal”
• “Rear Case Removal”
• “Touchscreen/LCD Removal”
• “SPA PCB Assembly Removal”
• “Main PCB Removal”
• “Fan Assembly Removal”
• “RF Input Adapter Removal”
• “Reassembling the Unit”
Electrostatic Discharge (ESD) can damage the highly sensitive circuits in the instrument. All work
should be performed in a static-safe work area.
Only qualified personnel should open the case and replace internal assemblies. Assemblies shown
in Table 1-4 are typically the only items that may be replaced. The assemblies are highly fragile,
items that must be soldered may not be replaced without specialized training.
Caution The Field Master contains components that can easily be damaged by electrostatic discharge
(ESD). An ESD safe work area and proper ESD handling procedures that conform to ANSI/ESD
S20.20-1999 or ANSI/ESD S20.20-2007 is mandatory to avoid ESD damage when handling
sub-assemblies or components found in the instrument.
Repair of damage that is found to be caused by electrostatic discharge is not covered under
warranty.
Removing the RF shield from PC boards or adjustment of screws on or near the shields may detune
Caution
sensitive RF circuits and will result in degraded performance.
Many of the procedures in this section are generic and apply to many similar instruments. Photos
Note
and illustrations used are representative and may show instruments other than the Field Master.
3. With the battery access door removed, grab the battery lanyard and pull the battery straight out of the
unit as illustrated in Figure 4-2.
2. Detach the bumpers completely by lifting them upwards to remove from the unit. See Figure 4-4.
4. With all screws removed, place the instrument face down on a stable flat work surface.
5. Gently pull the rear case apart from the front case.
4. Remove the Torx screw on the top connector panel using a T10 screw driver. See Figure 4-7.
5. Remove the three Torx screws securing the battery shroud using a T10 Torx screw driver. See Figure 4-8.
6. Disconnect the battery cable and remove the battery shroud. See Figure 4-9.
7. Carefully tilt the Main PCB/SPA PCB assembly to view the display cable. See Figure 4-10.
8. Release the locking connector by slowing lifting the metal flap upwards to disconnect the display cable.
See Figure 4-11.
9. Carefully tilt the Main PCB/SPA PCB assembly to remove the two Wi-Fi antennas. Release the Wi-Fi
Antennas by carefully prying the end slots using a straight screw driver. See Figure 4-12.
10. Carefully tilt the Main PCB/SPA PCB assembly to remove the keypad cable connector by slightly
pressing it at the ends and disconnecting it from the LCD top case assembly. See Figure 4-13.
11. The Touchscreen/LCD Display is permanently installed into the Top Case.
2. Remove the two Torx screws using T10 Torx driver from the brackets on the Main PCB. See Figure 4-15.
3. Remove the eight Torx screws securing the SPA PCB Assembly to the Main PCB using a T10 Torx driver.
See Figure 4-16.
2. Make sure that the label of the fan assembly faces outwards while fitting it into the LCD top case
assembly. See Figure 4-18.
3. Use caution as not to cripple the fan assembly harness in between while securing the Main PCB/SPA
assembly onto the LCD Top case assembly. See Figure 4-18.
5-1 Introduction
This chapter describes the primary troubleshooting operations that can be performed by all Anritsu service
centers. Perform the troubleshooting suggestions in the order they are listed.
Only qualified Anritsu personnel should replace internal assemblies. Major subassemblies shown in Table 1-4
are typically the items that may be replaced. Because they are highly fragile, items that must be soldered may
not be replaced without special training. Removal of RF shields from PC boards or adjustment of screws on or
near the shields will detune sensitive RF circuits and will result in degraded instrument performance.
Date: ___________
MS2070A/80A Firmware Rev: ______________ Operator: ____________________
_
Serial Number: _______________ Options: __________________________________________________
Measurement
Frequency Measured Value Error Uncertainty Specification
Measurement
Frequency Measured Value Error Uncertainty Specification
Measurement
Frequency Measured Value Error Uncertainty Specification
Date: ___________
MS2070A/80A Firmware Rev: ______________ Operator: ____________________
_
Serial Number: _______________ Options: __________________________________________________
MS2070A/8
0A
NRP18T Expected MS2070A/80A
Input Reference Power Power Measured Measurement
Frequency Attenuation Level Level Level Power Level Uncertainty Specification
50 MHz 20 dB 0 dBm +0 dBm –10 dBm dBm 0.30 dB –10 dBm ±1.0 dB
50 MHz 20 dB 0 dBm –4 dBm –14 dBm dBm 0.30 dB –14 dBm ±1.0 dB
50 MHz 20 dB 10 dBm –8 dBm –18 dBm dBm 0.30 dB –18 dBm ±1.0 dB
50 MHz 10 dB –10 dBm –10 dBm –20 dBm dBm 0.30 dB –20 dBm ±1.0 dB
50 MHz 10 dB –10 dBm –14 dBm –24 dBm dBm 0.30 dB –24 dBm ±1.0 dB
50 MHz 10 dB –10 dBm –18 dBm –28 dBm dBm 0.30 dB –28 dBm ±1.0 dB
50 MHz 0 dB –20 dBm –20 dBm –30 dBm dBm 0.30 dB –30 dBm ±1.0 dB
NRP18T MS2070A/80A
Power Expected MS2070A/80A Measured Power Measurement
Frequency Level Power Level Level Uncertainty Specification
100 kHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
10 MHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
150 MHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
1 GHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
2.99 GHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
MS2080A-0704
3.99 GHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
MS2080A-0706
5 GHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
5.99 GHz –10 dBm –50 dBm dBm 0.30 dB –50 dBm ± 1.0 dB
Date: ___________
MS2070A/80A Firmware Rev: ______________ Operator: ____________________
_
Serial Number: _______________ Options: __________________________________________________
Table A-6. MS2070A/80A Spectral Purity - SSB Phase Noise Offset from 1 GHz
Date: ___________
MS2070A/80A Firmware Rev: ______________ Operator: ____________________
_
Serial Number: _______________ Options: __________________________________________________
10 MHz 500 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
500 MHz 1000 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
1000 MHz 1500 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
1500 MHz 2500 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
2500 MHz 3000 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
Date: ___________
MS2070A/80A Firmware Rev: ______________ Operator: ____________________
_
Serial Number: _______________ Options: __________________________________________________
10 MHz 500 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
500 MHz 1000 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
1000 MHz 1500 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
1500 MHz 2500 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
2500 MHz 3000 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
3000 MHz 4000 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
4000 MHz 5000 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
5000 MHz 6000 MHz 300 Hz 100 Hz dBm 1.4 dB <–110 dBm, max
Table A-13a.MS2080A Residual Spurs (Preamplifier On, for Rev1 Option 704 models only)
10 MHz 500 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
500 MHz 1000 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
1000 MHz 1500 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
1500 MHz 2500 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
2500 MHz 3000 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
3000 MHz 4000 MHz 300 Hz 100 Hz dBm 1.4 dB <–105 dBm, max
10 MHz 500 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
500 MHz 1000 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
1000 MHz 1500 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
1500 MHz 2500 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
2500 MHz 3000 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
Date: ___________
MS2070A/80A Firmware Rev: ______________ Operator: ____________________
_
Serial Number: _______________ Options: __________________________________________________
10 MHz 500 GHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
500 MHz 1000 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
1000 MHz 1500 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
1500 MHz 2500 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
2500 MHz 3000 MHz 300 Hz 100 Hz dBm 1.4 dB <–95 dBm, max
3000 MHz 4000 MHz 300 Hz 100 Hz dBm 1.4 dB <-95 dBm, max
4000 MHz 5000 MHz 300 Hz 100 Hz dBm 1.4 dB <–90 dBm, max
5000 MHz 6000 MHz 300 Hz 100 Hz dBm 1.4 dB <–90 dBm, max
Table A-15a.MS2080A Residual Spurs (Preamplifier Off, for Rev1 Option 704 models only)
10 MHz 500 GHz 300 Hz 100 Hz dBm 1.4 dB <–94 dBm, max
500 MHz 1000 MHz 300 Hz 100 Hz dBm 1.4 dB <–94 dBm, max
1000 MHz 1500 MHz 300 Hz 100 Hz dBm 1.4 dB <–94 dBm, max
1500 MHz 2500 MHz 300 Hz 100 Hz dBm 1.4 dB <–94 dBm, max
2500 MHz 3000 MHz 300 Hz 100 Hz dBm 1.4 dB <–94 dBm, max
3000 MHz 4000 MHz 300 Hz 100 Hz dBm 1.4 dB <-88 dBm, max
Date: ___________
MS2070A/80A Firmware Rev: ______________ Operator: ____________________
_
Serial Number: _______________ Options: __________________________________________________