0% found this document useful (0 votes)
127 views2 pages

2.HDG502 Thermal Resistance Temperature Isolation Transmitter

HDG

Uploaded by

phildec73000
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
127 views2 pages

2.HDG502 Thermal Resistance Temperature Isolation Transmitter

HDG

Uploaded by

phildec73000
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

HDG502 Thermal resistance temperature isolation transmitter

The input thermal resistance signal is converted into a current / voltage


signal output after isolation. The three ends of the input end and the
power end are photoelectrically isolated. It has functions such as online
fault self-diagnosis, and has the advantages of fast response speed, low
power consumption, and good temperature characteristics. Is an
intelligent temperature transmitter product.

 It can be used in conjunction with modular instruments and DCS, PLC


and other systems to provide signal isolation, signal conversion, signal
distribution, and signal processing functions for field instruments, which
enhances the anti-interference ability of the automatic control system in
the industrial production process, ensuring the system's Stability and
reliability.

Technical Parameters
Isolated
transmission ±0.1%F.S(25℃±2℃)
accuracy
Response time ≤500ms
Temperature drift ≤100 ppm/℃
Ripple <1 mV
Frequency range 0~300HZ
current≤60Ω;
Input resistance
voltage:0~5V:≥1MΩ;0~10V;≥2MΩ
DC current: 4~20mA:≤600Ω;
Output load DC voltage:0(1)~5V:≥1MΩ; Product Model List
capacity 0(2)~10V:≥2MΩ;
HDG502 X X X X
Power supply 18~32VDC(typical value 24V DC) 1 1 in 1 out
range 85~265VAC/DC Number
DC Full load
1 input 2 2 input 2
of 2 1 in 2 out
power 1input 1 output
output output channels 5 2 in 2 out
consumption
At 20mA output 1W 1.5W 2W P1 Pt100(-200~+850℃)
Insulation
≥100MΩ, (between input / output / power) Input signal C5 Cu50(-50~+150℃)
resistance
Insulation ≥1500 V AC (input / output / power, leakage current
strength 1mA, test time 1 minute) S Other signal
Ambient temperature during
-20℃~+60℃ 1 4~20mA
work temperature during work
2 0~20mA

Output signal 4 0~5V


Panel indicator
6 0~10

PWR: Power indicator (green). OUT1, OUT2: Channel input S Other signal
signal status indicator (red), corresponding to channel 1 and DC 24V default
channel 2 respectively. This indicator is off during normal
working status; This indicator flashes when the input signal Power supply
of the corresponding channel fails; When the input signal of A 85-265VAC/DC
the corresponding channel is out of range, this indicator is
Note: The customer needs to determine the input signal form and
on. output signal form when ordering, and can be customized if there
are special needs.

Product selection example:

Example: HDG502-1P11 / 0-100, input signal PT100 (0-100


degrees), output 4-20mA, one input and one output, 24VDC power
[键入文字] supply

Huaibei Huadian Automation Technology Co.,Ltd Http://www.hd-a.cn Mob:18605613933


HDG502 Thermal resistance temperature isolation transmitter

Wiring diagram

[键入文字]

Huaibei Huadian Automation Technology Co.,Ltd Http://www.hd-a.cn Mob:18605613933

You might also like