Air Liquid Monnal t75 - Rev 9a - 2021 06 - en
Air Liquid Monnal t75 - Rev 9a - 2021 06 - en
www.device.airliquidehealthcare.com
V3.6.xI9A
Contents
Before use...............................................................................................7
Definitions of user warnings..............................................................................................................7
Intended use...................................................................................................................................... 7
Brief description of the device.......................................................................................................... 7
Requisite knowledge.......................................................................................................................... 8
Training..................................................................................................................................................8
Symbols and markings on the device............................................................................................. 10
General safety instructions..............................................................................................................11
Description of the device.......................................................................15
Terminology used............................................................................................................................ 15
Front panel....................................................................................................................................... 15
Right-hand side................................................................................................................................ 16
Left-hand side.................................................................................................................................. 16
Rear panel........................................................................................................................................ 17
Installation and commissioning............................................................. 19
Unpacking.........................................................................................................................................19
Connections and commissioning.................................................................................................... 19
Power supply...................................................................................................................................... 19
Oxygen supply.................................................................................................................................... 19
Assembly of patient circuit and accessories.................................................................................. 19
CO2 Measurement Probe.................................................................................................................. 20
Humidifier............................................................................................................................................20
Nebulization........................................................................................................................................ 21
Applying power...................................................................................................................................22
Automatic tests.................................................................................................................................. 22
Transport on the rolling stand.......................................................................................................... 23
Use........................................................................................................ 26
Screen description........................................................................................................................... 27
Stand-by mode................................................................................................................................... 27
Ventilation mode................................................................................................................................ 28
Start/stop ventilation....................................................................................................................... 28
Shutting down the unit.................................................................................................................... 29
New patient / Category selection.................................................................................................... 29
Ventilator configuration................................................................................................................... 30
Language.............................................................................................................................................30
Date and time..................................................................................................................................... 30
Transfer............................................................................................................................................... 31
Brightness........................................................................................................................................... 31
Audible alarm......................................................................................................................................31
Ventilation modes............................................................................................................................ 32
Mode selection................................................................................................................................... 32
Ventilation set-points......................................................................................................................... 32
Description of ventilation modes.....................................................................................................33
VCV (Volume-Controlled Ventilation)................................................................................................33
PCV (Pressure-Controlled Ventilation)..............................................................................................33
PCV (Pressure-Controlled Ventilation)..............................................................................................34
SIMV (Synchronized Intermittent Mandatory Ventilation)...............................................................35
PSIMV (Synchronized Intermittent Mandatory Pressure Monitored Ventilation).......................... 36
PSV NIV (Non Invasive ventilation).................................................................................................. 37
CPAP (Continuous Positive Airway Pressure)................................................................................. 38
Duo levels (Alternation of two CPAP levels)....................................................................................38
APRV (Airway Presure Release Ventilation).....................................................................................39
PRVC (Pressure-Regulated Volume Controlled).............................................................................. 39
PS – PRO (interlocked barometric mode, intended for use in resuscitation)................................ 40
Oxygen therapy................................................................................................................................ 41
CO2 Option.......................................................................................................................................43
Introduction.........................................................................................................................................43
Installing the CO2 software option...................................................................................................43
Display of pressure and flow rate curves........................................................................................ 43
Setting up the CO2 measurement probe......................................................................................... 44
Adjusting the CO2 unit...................................................................................................................... 45
Monitored CO2 parameters...............................................................................................................45
Activation or deactivation of CO2 monitoring................................................................................. 47
Calibrating the CO2 probe................................................................................................................. 48
Recommendations for the use of the MASIMO IRMA™ CO2 measurement probe....................... 48
Display of pressure and flow rate curves........................................................................................49
Monitoring........................................................................................................................................ 51
Display of measurements..................................................................................................................51
Alarm settings.................................................................................................................................... 54
Automatic thresholds.........................................................................................................................56
Resetting............................................................................................................................................. 56
Alarm history...................................................................................................................................... 56
Measurements.......................................................................................58
Expiratory plateau............................................................................................................................ 58
Inspiratory plateau........................................................................................................................... 58
[R&C stat] key.................................................................................................................................. 58
P0.1 measurement key.................................................................................................................... 60
NIF (negative inspiratory force)....................................................................................................... 60
Manual Cycle....................................................................................................................................61
Work of Breathing (WOB)................................................................................................................ 62
Menu..................................................................................................... 63
Description....................................................................................................................................... 63
Menu structure.................................................................................................................................64
Apnea ventilation adjustment........................................................................................................... 64
TC (Tube Compensation).................................................................................................................. 65
O2 low pressure................................................................................................................................. 66
Sensors............................................................................................................................................... 67
Other functions................................................................................................................................ 67
Oxygenation function-suction............................................................................................................68
Screen lock key (lock)....................................................................................................................... 69
Inhibiting the spirometry sensor....................................................................................................... 69
Nebulization........................................................................................................................................ 70
Loop curves........................................................................................................................................ 72
Trends..................................................................................................................................................73
Administrator configuration...................................................................75
Access..............................................................................................................................................75
Presentation..................................................................................................................................... 75
Setting inspiratory time in VCV mode (60)......................................................................................76
Setting inspiratory time in PCV mode (61)......................................................................................76
Inspiratory time monitoring (62)......................................................................................................76
Monitoring etCO2 (63)..................................................................................................................... 76
Disabling the RRmini set-point (64).................................................................................................76
Altered O2 network (65).................................................................................................................. 76
Volume setting (66)......................................................................................................................... 77
Enabling Options (67)...................................................................................................................... 77
Saving the ventilation set-points (68)............................................................................................. 77
Pressure Support Setting (69)......................................................................................................... 78
Setting the ventilation set-points (70)............................................................................................. 78
Selecting ventilation modes (71).....................................................................................................78
Alarms...................................................................................................80
Parameters....................................................................................................................................... 80
Acknowledgement of alarms........................................................................................................... 81
Inhibition of alarms..........................................................................................................................81
List of alarms...................................................................................................................................81
Maintenance..........................................................................................92
Definitions........................................................................................................................................ 92
Routine maintenance....................................................................................................................... 92
Expiratory assembly: flow sensor + expiratory valve Monnal EVA.................................................. 93
Monnal Clean’in filter (Type filtre HEPA: High- Efficiency Particulate Air)...................................... 94
Air intake filter................................................................................................................................. 95
Bacteriological filter.........................................................................................................................95
Accessories........................................................................................... 96
Items included in the package........................................................................................................ 96
List of consumables........................................................................................................................ 96
List of accessories.......................................................................................................................... 97
List of available documents............................................................................................................ 98
Maintenance..........................................................................................99
By the user.......................................................................................................................................99
Battery life check............................................................................................................................... 99
By the technician........................................................................................................................... 100
O2 cell............................................................................................................................................ 100
Technical description.......................................................................... 102
Operation........................................................................................................................................102
Pneumatic system........................................................................................................................... 102
Ventilation operation........................................................................................................................103
Air/O2 mixture.................................................................................................................................. 104
Special characteristics of nebulization.......................................................................................... 105
CO2 monitoring................................................................................................................................ 105
Electrical power sources................................................................................................................105
Power supply management............................................................................................................ 106
Mains supply.................................................................................................................................... 106
Internal battery................................................................................................................................. 106
External battery................................................................................................................................ 107
Inputs and outputs.........................................................................................................................107
Alarm transfer...................................................................................................................................107
RS232 link.........................................................................................................................................108
Connectivity to hospital networks.................................................................................................. 108
Video output..................................................................................................................................... 110
Performances and characteristics.................................................................................................110
Regulatory requirements................................................................................................................. 110
Recovery of the components of the medical device.....................................................................110
Technical characteristics.................................................................................................................111
Settings tables..................................................................................................................................126
Testing the alarms......................................................................................................................... 133
Bibliography........................................................................................ 134
Appendix............................................................................................. 135
Checklist.........................................................................................................................................135
Expiratory assembly cleaning protocol..........................................................................................136
1 Before use
1 Before use
7
Monnal T75 user manual
Patient environment
Under normal use, the patient is in a hospital bed and the Monnal T75 equipment is placed nearby. All
parts of the medical electrical system are suitable for use in the patient environment.
User’s position
The equipment is positioned with the display facing towards the user allowing them to make the nec-
essary adjustments with the coder wheel and read the information displayed on the screen. The rec-
ommended distance depends on the environment, the ambient light and the user’s visual acuity. The
rear of the equipment remains, however, within the user’s reach.
Options
Monnal T75– reference KB033600 - comes with the option of purchasing CO2 monitoring.
1.4.1 Training
8
1 Before use
9
Monnal T75 user manual
YYYY-MM
10
1 Before use
IP3X Protection Index according to the This logo means that the equipment
EN 60529 standard must not be disposed of via ordinary
(if serial num- waste channels. It must receive ap-
ber below 3: protection from the penetration propriate end-of-life handling, in ac-
MT75-05000) of solid bodies of diameter ≥ 2.5 cordance with European Directive
mm. 2012/19/UE (WEEE).
IP31 1: protection against vertically This device was manufactured after
(for serial falling water drops the 13th of August 2005.
numbers from X: no particular protection from
MT75-05000) the penetration of liquids, but
compliant with EN 60601-1.
COM 2
11
Monnal T75 user manual
The ventilator does not directly administer nitric oxide but can be used concomitantly with a Nitric Ox-
ide administration system, provided that the manufacturer of the nitric oxide delivery system has vali-
dated its use.
Do not use the device with components contaminated by inflammable substances (e.g. grease, oil,
etc.).
The internal components of the device were degreased before delivery and use a type of grease which
is compatible with oxygen. Do not grease or lubricate any part of the device.
Medical quality oxygen, which is dust-free and dry, must be used (H2O < 20 mg/m3).
The supply pressure on the high pressure oxygen inlet must be between 280 kPa (2.8 bar) and 600 kPa
(6 bar).
The supply pressure on the low pressure oxygen inlet must be less than 150 kPa (1.5 bar).
Power supply
CAUTION: To avoid the risk of electric shock, this equipment must only be connected to a
grounded power outlet.
Check that the voltage in the mains socket used matches the electrical characteristics of the ventilator
(indicated on the rear panel of the power supply adaptor).
Electrical isolation is accomplished by disconnecting the equipment from the power outlet. According-
ly, do not place the equipment in a position that would make it difficult to use this disconnect device.
The ventilator has an internal battery. It must be plugged into the mains on a regular basis in order to
keep the internal battery sufficiently charged.
In the event of any doubt about the condition of the mains power supply cable, use the device on its in-
ternal battery.
Do not use antistatic or electrically conductive tubing.
The user must not touch the patient and the equipment enclosures at the same time.
Modifications to the EM system require an assessment of their compliance with the requirements of
standard IEC60601-1.
IP Protection
To ensure the IP protection level of the device is maintained during normal use, it is essential that the
power cable is plugged in and all removable components (air filter, expiratory assembly, O2 sensor
cover and the rear plastic panel) are fitted in place.
Electromagnetic compatibility
Monnal T75 complies with the EN 60601-1-2 standard on the electromagnetic compatibility of medical
devices.
This device requires specific precautions as regards EMC. It must be installed and put into operation
in accordance with the EMC information provided in this user manual.
The ventilator functioning can be impaired by the use of appliances in its immediate proximity, such as
diathermic devices, high frequency electro-surgery units, defibrillators, cell phones or, more generally,
by electromagnetic interference that exceeds the levels specified in the EN 60601-1-2 standard.
Monnal T75 should not be placed next to or on top of this equipment. If it is not possible to do other-
wise, Monnal T75 should be monitored to make sure that it operates correctly where placed.
Do not use this ventilator in a specifically magnetic environment (MRI, etc.).
12
1 Before use
Replacing a cable or internal component that is not provided by Air Liquide Medical Systems, may lead
to a rise in emissions or a decrease in the immunity of the device.
Connection to other electrical devices
Do not connect the device to other electrical appliances not mentioned in this user manual without
first consulting the manufacturers or technical support.
Devices connected to the inputs and signal outputs must comply with the 60601-1 Standard, Edition 2.
Commissioning
The device must not be put into service immediately after storage or transportation where the temper-
ature and humidity were different from the recommended operating conditions.
Before each use, check that the audible and visual alarms are working correctly and carry out the
checking process in the appendix (see Checklist on page 135).
Use
The manufacturer has provided solutions for most of the possible malfunctions of the ventilator, and
these are normally covered by the internal monitoring system. It is nevertheless recommended, in case
of complete patient dependence to provide an additional and fully autonomous system which can be
used to check the effectiveness of ventilation, as well as a back-up device, such as a suitable manual
insufflator. Lack of an alternative means of ventilation may result in patient death should the ventilator
fail.
If the accessories used are not compliant with the manufacturer’s recommendations, the manufactur-
er accepts no responsibility in the event of an incident.
Do not expose the device to direct sunlight.
Do not use the Monnal T75 in a hyperbaric chamber. It is not reccommended to push the pulmonary
ventilator by its top when fitted on its trolley. To move, push using the handle (at the base of the venti-
lator)
The device and its accessories (masks, circuits, etc.) are Latex Free in order to avoid the risk of allergic
reactions.
The air inlets behind and under the device must be completely unobstructed.
To operate the device from ambient air, a Monnal Clean’In (HEPA) filter must be used at the ventilator
inlet. This filter is recommended by Air Liquide Medical Systems.
Do not use the ventilator in an explosive or nicotine-laden atmosphere (cigarette smoke, fire, etc.).
Risk of cross-contamination
Reusing single-use accessories or consumables carries the risk of patient cross-contamination. This
risk also arises if reusable accessories or consumables are not sterilized between each use.
The endotracheal tube, mask, patient circuit, expiratory valve, humidification chamber, and the probe
or nebulizer adapters are part of the air pathway that may be contaminated under normal and sin-
gle-fault conditions by body fluids, secretions or gases exhaled by the patient.
13
Monnal T75 user manual
Maintenance
• The ventilator must be checked regularly. To schedule and record maintenance operations, refer to
the maintenance form in the appendix.
In accordance with IEC 60601-1 (Annex A, Subclause 7.9.2.6):
“The instructions for use can contain a statement saying that the Manufacturer, assembler, in-
staller or importer considers himself responsible for the effect on basic safety, reliability and per-
formance of the ME equipment or ME system only if:
• appropriately trained personnel carry out assembly operations, extensions, readjustments, modifi-
cations or repairs;
• the electrical installation of the relevant room complies with the appropriate requirements;
• the ME equipment or ME system is used in accordance with the instructions for use.»”
The qualified technician must use only Air Liquide Medical Systems spare parts when carrying out
routine maintenance on the device.
Do not use abrasive powders, alcohol, acetone or other easily flammable solvents.
The device must be disconnected from the mains during any procedure such as maintenance or
cleaning
Recommendations for the use of the CO2 measurement probe MASIMO IRMA™ (REF. KB033600)
See CO2 Option on page 43.
14
2 Description of the device
15
Monnal T75 user manual
11
12
16
2 Description of the device
13
14
15 16
17
Monnal T75 user manual
18
3 Installation and commissioning
3.1 Unpacking
Take the ventilator out of the packaging and place it on a table.
Unwrap the accessories supplied with the ventilator (see Items included in the package on page
96).
CAUTION: The ventilator weighs approximately 16 kg:
• Apply safe lifting procedures when installing it.
• For installation on a horizontal support, ensure that the horizontal support can bear the
weight of the unit. Check that the assembly is stable.
This ventilator may be used with various sources of electrical power (see Electrical power sources on
page 105):
• AC power supply;
• Power supply connected to an external DC source (external battery);
• Power supply connected to an internal DC source (internal battery).
To turn on the ventilator, plug it into a grounded AC power outlet.
CAUTION: Always check that the electrical network is compatible with the specifications in
this manual.
Check that the device preventing the mains plug from being pulled out is operative.
To supply FiO2 above 21%, connect the low or high pressure inlet of the ventilator oxygen to an avail-
able source, via an appropriate connection.
If this oxygen source is a cylinder, it must be equipped with a pressure reducer to suit the allowable
pressure range (2.8 to 6 bar).
CAUTION: Start by connecting the O2 connection hose to the ventilator before connecting it
to the oxygen network. Check the capacity of the oxygen cylinder before using the ventilator
for intra-hospital transport.
CAUTION: To prevent the low pressure O2 fitting becoming detached, ensure the input pres-
sure applied to the low pressure oxygen inlet is less than 150 kPa (1.5 bar).
Connect the patient circuit to the ventilator and the humidifier (if used):
19
Monnal T75 user manual
27 28
•
Connect the expiratory limb of the patient circuit to the expiratory valve of the ventilator: (27).
•
Connect the inspiratory limb of the patient circuit to the inspiratory outlet cone of the ventilator:
(28).
CAUTION: Make sure to limit dead space while installing patient circuit and accessories.
Air Liquide Medical Systems recommends the use of a bacteriological filter at the inspiratory outlet
of the ventilator or, preferably, at the Y-piece. It is also recommended to use patient circuits equipped
with water traps when using a humidifier.
When using the equipment on a patient for the first time, make sure that you follow the hospital’s hy-
giene protocol for new single-use equipment or adequately sterilized reusable equipment. The acces-
sories and consumables (patient circuit, masks, expiratory valves, adapters, nebulizer, etc.) are gener-
ally available in single-use and autoclavable versions.
CAUTION: Reusing single-use accessories or consumables carries the risk of patient cross-
contamination.
3.2.5 Humidifier
If the ventilator is used with a humidifier ensure that it is always placed lower than the ventilator and
the patient.
Empty the water traps regularly in order to limit condensation in the pipes.
20
3 Installation and commissioning
CAUTION: Ensure that water does not enter the unit during handling of the patient circuit
or the humidifier (if used). If this occurs, immediately stop using the device, and contact the
Technical Department.
CAUTION: Humidification may increase the resistance of the filters used in the patient cir-
cuit. The filters should be tested frequently to check for an increase in resistance or block-
age.
3.2.6 Nebulization
The ventilator operates with pneumatic nebulization. It supplies a reduction pressure of 1.2 bar.
The nebulization outlet is located at the front of the unit and identified by the following symbol:
To set up a nebulizer:
1 2
5 4
1. Install a connector (5) upstream from the Y-piece, in the inspiratory limb of the patient circuit
(1).
2. Connect the body of the nebulizer (3) to this connector (5).
3. Connect the nebulizer pipe (4) to the nebulizer output of the ventilator.
CAUTION: The nebulization output (front panel) and the low-pressure inlet (rear panel) are
similar. Beware of not confusing them.
CAUTION: Nebulization with the ventilator includes certain constraints which must be taken
into consideration:
• Do not conduct treatment sessions with inflammable agents,
• Y-piece respiratory filters can prevent medication from being effective: their use is there-
fore not recommended,
• The precision of the expired volume can be impaired: a protective filter can then be used
at the expiration end.
21
Monnal T75 user manual
CAUTION: Nebulization may increase the resistance of the filters placed in the patient cir-
cuit. The filters should be checked frequently for an increase in resistance or blockage.
Refer to Nebulization on page 70 Nebulization for a more detailed description of the nebulization
function and to Accessories on page 96 for different types of nebulizers available.
Turn on the unit using the switch button (green) at the back of the unit.
The [Automatic tests] button checks the integrity and correct operation of the unit’s internal compo-
nents. In particular, it calibrates certain sensors, including the expiratory flow sensor and the oxygen
cell. Without the automatic test, the precision of ventilation parameters and measurements cannot be
guaranteed.
Note: Air Liquide Medical Systems recommends to run the automatic test before each use of
the device on a patient.
22
3 Installation and commissioning
If the automatic tests fail with the message «Circuit resistance not evaluated»:
• check the consistency between the patient category selected and the patient circuit used,
• check that the patient circuit is properly connected to the device,
• check that the filters and other accessories used do not cause too high a resistance.
CAUTION: Make sure that the patient category selected corresponds to the patient circuit
and the accessories used (see New patient / Category selection on page 29).
For a full description of the automatic test, see Technical description on page 102.
When moving Monnal T75 Monnal T75 on its stand, place both hands on the handle to push it safely:
23
Monnal T75 user manual
CAUTION: When the floor is sloping, you must place both foot brakes on the same side: ei-
ther on the side sloping upwards or on the side sloping downwards. Placing one foot brake
on side sloping upwards and one brake on the side sloping downwards may imbalance Mon-
nal T75.
24
3 Installation and commissioning
25
Monnal T75 user manual
4 Use
The ventilator is controlled mainly via the touch screen and the control wheel.
CAUTION: Do not use any object that might scratch the screen.
26
4 Use
34 35 36 37 38
33 39
40
41
32
42
43
45
44
27
Monnal T75 user manual
49
51
46
47
48
50
28
4 Use
To stop ventilation in progress, press button [43]. A pop-up window appears telling the user whether
to cancel or validate by pressing the control wheel. In case of validation the unit then goes directly to
stand-by mode.
Note: If the unit was suddenly shut off during ventilation (e.g. battery exhausted), it automati-
cally resumes ventilation with the last parameters saved when restarted.
The choice of a patient category enables the prescriber to adapt each of the following elements to the
condition of the patient:
• Initial values of ventilation parameters and alarms,
• Ventilation parameter and alarm adjustment ranges.
Each of these values or ranges is given in the appendix of this manual.
CAUTION: For the safety of the patient and to optimize ventilation performance, the catego-
ry selected must be appropriate for the patient being ventilated. The diameter of the tubing in
the patient circuit must also be appropriate:
Adult 100-2000 22 mm
29
Monnal T75 user manual
Infant 20-75 12 mm
≥ 3kg
4.5.1 Language
30
4 Use
Note: If the date selected does not exist (e.g. 30 February), it will be displayed in red on the
screen and will not be taken into account. The previous date will be restored.
4.5.3 Transfer
This function can be reached from the stand-by screen. It allows the maintenance technician to trans-
fer data from the unit to a PC.
4.5.4 Brightness
The alarm volume adjustment is for the convenience of the doctor. The maximum distance and the
volume of the alarm must therefore be determined by the user according to the surroundings.
Access to the alarm list can only be made when the unit is in stand-by mode.
To choose an alarm:
• Press the [Alarm: …] button and turn the control wheel to select [Standard] or [Custom];
• Validate to confirm.
An audible display of the alarm is then performed so that the user is aware of the new setting.
Note: Whatever the operation done on the device (shutdown, software update…), every se-
lected alarm remains.
Note: It is possible to select standard or custom alarm melody in the Administrator Configu-
ration.
31
Monnal T75 user manual
The ventilation mode is selected from the ventilation mode display button in the upper left-hand part of
the screen (1).
To select or change the ventilation mode, press the display button, select the desired ventilation mode
(2), and press [Validate].
To adjust a set-point, touch it to select, set the desired value by turning the control wheel, and validate
by pressing the control wheel.
32
4 Use
Note: If the patient tries to inhale at a greater flow rate than the flow rate set point, the venti-
lator will deliver the desired flow rate to the patient but will switch to expiration when the vol-
ume has been delivered (TI will not be observed in this case). The “Patient demand greater
than peak flow rate setting” alarm will then be triggered. In this case, it is advisable to in-
crease the peak flow rate (or reduce the inspiration time) and/or increase the VT setting.
Note: In VCV, there is no apnea ventilation (or “backup ventilation”).
Safety is guaranteed by the lower limit on the frequency and tidal volume according to each
patient category.
Principle
PCV mode is used to control the pressure delivered to the patient, the inspiration time, and the fre-
quency of the respiratory cycles. During expiration, the ventilator regulates the pressure in order to
maintain the set PEEP level.
33
Monnal T75 user manual
The respiratory frequency can be increased as soon as the ventilator detects that the patient is making
a respiratory effort.
Note:
The value of PI corresponds to the pressure added to the current PEEP value.The PEEP value
is incorporated in the PI.
In PCV mode, unlike VCV, the pressure delivered to the patient is controlled but the tidal and minute
volumes are not.
Principle
A constant positive pressure is maintained above PEEP level in the patient circuit each time the patient
makes an inspiratory effort.
The switch to the expiratory phase can be triggered:
• If the flow rate falls below the set expiratory threshold (E. Trig),
• By an expiratory effort from the patient,
• If the maximum set insufflation time is reached (TImax).
If there is no inspiratory effort, the machine provides the minimum set frequency.
Ventilation set-points
34
4 Use
Note: Apnea ventilation can be activated. The apnea parameters should be suited to the con-
dition and requirements of the patient.
• Use of the “RR mini” parameter. In the absence of inspiratory demand for a time longer
than “1/RR mini”, the ventilator initiates a pressure support cycle. The RR low alarm can
be associated with this safety feature by setting it to a value above “RR mini”.
• Use of the “TImax” parameter. In the event of a leak in the circuit, the flow rate expirato-
ry trigger might not be activated. In this case, the limitation on inspiration time allows the
patient to enter the expiratory phase.
Principle
SIMV mode combines mandatory assisted controlled ventilation and pressure supports spontaneous
patient ventilation between the controlled cycles.
Ventilation set-points
Operation
In this ventilation mode, the “controlled” cycle (VCV) delivers a fixed volume at the set frequency
RRSIMV. A SIMV period arises from this frequency.
E.g. for a SIMV frequency set to 10 cycles per minute, the SIMV period between two “controlled” (VCV)
cycles is 6 seconds.
35
Monnal T75 user manual
In the event of absence of patient respiratory activity, this mode provides the guarantee of controlled
ventilation (VCV).
When the patient has regular and detectable respiratory activity, the unit responds by:
• supplying a “spontaneous” cycle (PSV) if the time elapsed since the last “controlled” cycle is less
than the SIMV period.
• supplying a “controlled” cycle (VCV) if the time elapsed since the last “controlled” cycle is greater
than the SIMV period. If patient respiratory activity declines again after a “spontaneous” cycle, the
unit waits for the SIMV period - set TiMax to expire before triggering a “controlled cycle” (VCV) it-
self.
Insufflation in a spontaneous cycle with pressure support ends:
• If the flow rate falls below the set expiratory threshold (E. Trig);
• If the patient makes an expiratory effort,
• or if the insufflation time reaches the maximum Ti setting (Timax).
Note: Apnea ventilation can be activated. The apnea parameters should be suited to the con-
dition and requirements of the patient.
Principle
PSIMV Mode combines mandatory assisted pressure-controlled ventilation and spontaneous patient
ventilation between the assisted controlled cycles.
Ventilation set-points
(c/min)
Operation
In this ventilation mode, the “controlled” cycle (PCV) delivers a fixed pressure at the set frequency
RRSIMV. A SIMV period arises from this frequency.
E.g. for a SIMV frequency set to 10 cycles per minute, the SIMV period between two “controlled” (PCV)
cycles is 6 seconds.
36
4 Use
In the event of absence of patient respiratory activity, this mode provides the guarantee of controlled
ventilation (PCV).
When the patient has regular and detectable respiratory activity, the unit responds by:
• supplying a “spontaneous” cycle (PSV) if the time elapsed since the last “controlled” cycle is less
than the SIMV period,
• supplying a “controlled” cycle (PCV) if the time elapsed since the last “controlled” cycle is greater
than the SIMV period.
If patient respiratory activity declines again after a “spontaneous” cycle, the unit waits for the SIMV pe-
riod - set TiMax to expire before triggering a “controlled” cycle (PCV) itself.
Insufflation in a spontaneous cycle with pressure support ends:
• If the flow rate falls below the set expiratory threshold (E. Trig);
• If the patient makes an expiratory effort,
• or if the insufflation time reaches the maximum Ti setting (Timax).
Note: Apnea ventilation can be activated; the apnea parameters should be suited to the con-
dition and requirements of the patient.
Ventilation set-points
37
Monnal T75 user manual
Principle
In CPAP mode, the ventilator regulates the pressure in the airways to the set CPAP value.
The patient breathes spontaneously through the device in this mode of operation.
Ventilation set-points
Note: Apnea ventilation can be activated; the apnea parameters should be suited to the con-
dition and requirements of the patient.
Note: Loops curves, inspiratory and expiratory plateau are deactivated in CPAP mode.
Principle
The Duo-Levels mode is characterized by controlled-pressure ventilation, combined with the patient
having the option of breathing spontaneously throughout the entire cycle.
Duo-Levels mode enables the prescriber to maintain a constant level of pressure (PI) for a period of
time T high, then a lower level of pressure (PEEP) for a period of time T low. The duration of the high-
pressure phase can be set, as can the minimum frequency. However, in order to enable the patient’s
spontaneous breathing to properly adapt with the ventilator, there are trigger windows for synchroniz-
ing the inspiratory and expiratory phases:
• from low pressure to high pressure, the trigger window is 60% of T low, 4 seconds maximum,
• from high pressure to low pressure, the trigger window is 30% of T high, 2 seconds maximum. If at
the end of T high, inspiration is underway, inspiration is extended for up to 1 second.
Note: The value of PI corresponds to the pressure added to the current PEEP value.The PEEP
value is incorporated in the PI.
Ventilation set-points
38
4 Use
Principle
APRV mode is characterised by spontaneous ventilation, alternating between a constant high pressure
level and a short period at a lower pressure.
Ventilation set-points
Note: An absence of inspiratory demand during the apnea time (Tapnea) activates apnea
ventilation.
Note: The value of Phigh is absolute, meaning that the pressure on the high level will be
equal to Phigh, independent of the PEEP setting.
39
Monnal T75 user manual
• When the Vt Target is OFF, the delivered PI is the set PI. The mode is then equivalent to
PCV mode.
• If the Vt Target is disabled during ventilation, the set pressure takes the current regulation
value.
The PI value corresponds to the pressure added to the current PEEP value. The pressure applied to the
patient is then equal to PI + PEEP.
Whatever the circumstances, the delivered PI should never exceed the PI Max.
CAUTION: The “Vt target and regulated pressure incompatibility!!!” alarm is activated when:
• Monitored volume < 90% of the Vt Target and delivered PI = PI Max;
• Monitored volume > 110% of the Vt Target and delivered PI = PEEP + 5 cmH2O.
Furthermore if the Ppeak, Vti Max or patient disconnection alarm is set off, the adaptation of
the PI is inhibited until the alarm stops.
Ventilation set-points
CAUTION: The Vt Target should not be used in NIV. If there is a leak, the tidal volume moni-
tored by the ventilator no longer represents the tidal volume inhaled by the patient. Any adap-
tation of the pressure is therefore inappropriate.
Principle
The PS Pro mode is a PSV type pressure-regulated ventilation whose assistance frequency evolves au-
tomatically between two set frequencies: RR min and RR support (minimum frequency and maintained
frequency).
The aim of automatic adaptation of the frequency is to let the patient breathe autonomously when his
or her respiratory reflexes are active and allow the ventilator to take over when his or reflexes are inac-
tive.
Furthermore, the PS Pro mode can be combined with the Vt Target function. The inspiratory assis-
tance pressure is then adapted, cycle to cycle, so that the monitored Vt may converge towards the set
Vt Target.
If the patient spontaneously triggers above RR mini, each patient effort will be assisted. The set fre-
quency will then be the frequency of the patient.
40
4 Use
If the patient frequency reduces to the RR mini safety threshold, the ventilator will start the fixed cycles
to gradually reach the maintenance frequency, thus leaving the patient the opportunity of triggering a
new spontaneous cycle.
In the case of insufficient or no inspiratory effort (the patient’s respiratory frequency is lower than the
set minimum frequency), the ventilator guarantees an assistance frequency equal to the set mainte-
nance frequency.
A maintenance frequency equivalent to that set in VAC or PAC mode should therefore be set.
At the start of ventilation, the assistance frequency is equal to the maintenance frequency.
The maximum insufflation time (Ti max) is automatically set at 40% of the total time of a cycle with an
assistance frequency not exceeding 3.5 seconds.
Note: For operation of the Vt Target function, refer to PRVC mode.
CAUTION: The set minimum frequency is a safety threshold below which the ventilator pro-
vides controlled ventilation based on the set patient frequency.
If the patient is disconnected, the frequency is no longer adapted until the patient is recon-
nected.
Safety features of the Vt target function are applied when the function is activated.
Ventilation set-points
RR support maintenance frequency at which the patient is ventilated when he or she no longer
breathes spontaneously (c/min);
CAUTION: The Vt Target function should not be used in NIV. If there is a leak, the tidal vol-
ume monitored by the ventilator no longer represents the tidal volume inhalted by the patient.
Any adaptation of the pressure is therefore inappropriate.
41
Monnal T75 user manual
The oxygen therapy function is available for all categories of patients treated with Monnal T75 (adults,
children and infants) where they are connected to the high pressure O2 network (the function is dis-
abled in low pressure O2). It requires the use of a humidifier and nasal cannula (or tracheotomy tube)
specific to high throughput oxygen therapy.
Note: Please refer to the patient/machine interface manual.
Set-up
1. Install the humidifier on the Monnal T75 stand;
Note: Place the humidifier lower than the patient to prevent water flowing into the cir-
cuit.
2. Perform an automatic test with a dual branch circuit (see Automatic tests on page 22).
3. Disconnect the double-limb circuit and set up the required patient circuit on the inspiratory limb
of the ventilator (52) and on the humidifier.
52
4. Connect the patient/machine interface between the humidifier and the patient.
5. Check that the humidifier is in working order before switching it on.
6. Enable the Oxygen therapy function and set the desired flow rate.
CAUTION: Oxygen therapy is not a ventilation mode and must be used under supervision.
The patient breathes in a fully autonomous manner, this is why:
• Apnea ventilation is not ensured in the event of a patient’s respiratory arrest.
• No alarm is set off if the patient circuit is disconnected.
CAUTION: Monitored FiO2 is the FiO2 delivered by the ventilator. If the patient’s inspiratory
flow is greater than the flow set-point or if the nasal cannula are not properly positioned, then
monitored FiO2 is potentially higher than that inhaled by the patient.
42
4 Use
CO2 monitoring on the Monnal T75 (only for KB033600) is a fee-based software option.
After purchasing this option, Air Liquide Medical Systems provides:
• the activation code for the option,
• the CO2 probe for measuring the concentration of exhaled carbon dioxide (in accordance with ISO
80601-2-55),
• the necessary adapters.
The CO2 measurement probe is mounted on an airway adapter. This adapter is connected to patient
flow as close as possible to the intubation tube.
The IRMA™ CO2 measurement probe generates an infrared signal and transmits it to the receiver on
the opposite side, across the airway adapter. The measurement principle is based on the energy that
the CO2 absorbs. The probe sends the CO2 concentration measurement to the Monnal T75.
The Monnal T75 then displays the CO2 monitoring information that is received, in the form of a moni-
toring tile, time curve, trends curve and loop curve.
This information provides clinical support and can be used to assess the integrity of the patient’s air-
ways, as well as to check that accurate intubation has been successful.
CO2 monitoring requires installation of the software option with a code. If you would like this option,
contact your Air Liquide Medical Systems representative.
1. Go to the [Administrator configuration] menu.
2. Press the [Activate options] key.
3. Enter the code that your Air Liquide Medical Systems representative has provided.
4. Quit the [Administrator configuration] menu. The CO2 option is enabled.
The pressure and flow rate curves are displayed in real time by default. The volume and CO2 curves
may also be displayed (if the software option is present).
Note: Spontaneous patient activity is represented by the curves displayed on-screen chang-
ing color. When the respiratory cycle comes from a patient demand, the curves are green, but
a controlled cycle is displayed in yellow.
Changing curves
To change curves:
1. Press the curve to be replaced. A menu appears in the bar on the right-hand side of the screen.
2. Select the desired curve in the menu. The curve is displayed automatically.
43
Monnal T75 user manual
4. The LED flashes and then goes green. This means that the IRMA™ probe is ready for use (b).
5. Connect the IRMA™ probe, equipped with its adapter, to the Y-piece on the patient circuit (c).
6. Connect the IRMA™ probe to the patient’s endotracheal tube (d).
7. Position the IRMA™ probe (e) (see photo opposite).
Note: The probe sends information and alarms to Monnal T75. The probe has an LED indicat-
ing the following statuses:
44
4 Use
Continuous green OK
Note: For more information, please refer to the IRMA™ user manual from MASIMO.
The following CO2 parameters result from the interpretation of the temporal or volumetric capnogra-
phy curve. The volumetric capnography curve is the loop curve that gives the values of the exhaled
CO2 as a function of the exhaled volume.
45
Monnal T75 user manual
46
4 Use
When a CO2 probe is connected to the device, CO2 monitoring is automatically activated, and the CO2
curve is displayed on the screen.
Note: If the CO2 probe is connected and the user does not require monitoring, the user may
deactivate it.
When the CO2 probe is connected or activated, the [Calibration] key can be used to calibrate the probe.
47
Monnal T75 user manual
The IRMA™ probe must be calibrated when a significant shift in measurement is observed, or when the
[Calibrate the IRMA probe] alarm is triggered.
After the probe is connected to the device, you must wait approximately ten seconds before calibrat-
ing it.
If calibration is not an available option, the key is grayed out and unavailable.
To calibrate the CO2 probe:
1. Go to the [Menu].
2. Press the [Sensors] key.
3. Press the [Calibration] key.
The probe must be calibrated when using a new airway adapter on the sensor.
This must be disconnected from the Y-piece of the patient circuit and the patient. Then click on the
[Calibration] key to start calibrating the probe. The probe’s green indicator blinks for approximately
5 seconds during calibration. When the calibration is finished, the message [Calibration OK] is dis-
played.
Care must be taken to avoid breathing into the adapter when the probe is being calibrated. For the cali-
bration to succeed, it is extremely important for ambient air (21% O2 and 0% CO2) to be in the adapter/
probe combination.
Note: If the calibration is not done correctly, the measurement values will be skewed.
4.9.9 Recommendations for the use of the MASIMO IRMA™ CO2 measurement probe
• The IRMA™ probe may only be used by qualified, authorized medical personnel.
• The mainstream IRMA™ multi-gas probe is intended to be connected to a patient circuit to monitor
the gases inhaled/exhaled by adult or pediatric patients in emergency or intensive care situations.
• The probe must not be the sole patient monitoring device. It must always be used with other de-
vices to monitor vital signs and/or in addition to medical advice given by a qualified person. The
IRMA™ probe must be used by qualified, authorized medical personnel only.
• The probe must not be in direct contact with the patient.
• The IRMA™ probe must not be used with flammable anesthetic agents.
• IRMA™ airway adapters are single-use, and must not be re-used from one patient to another. Used
adapters must be disposed of via the appropriate hospital waste disposal channels. Reusing sin-
gle-use adapters can lead to patient cross-contamination.
• Used adapters must be disposed of via the appropriate hospital waste disposal channels.
• Do not use adapters intended for adult patients or children on infant patients, because such
adapters add 6 mL of dead volume to the patient circuit.
• Do not use adapters intended for infant patients on adult patients, because such adapters add 6
mL of dead volume to the patient circuit.
• The IRMA™ probe can give poor-quality measurements in the presence of devices that emit elec-
tromagnetic interference exceeding the levels mentioned in Standard 60601-1-2. Ensure that it is
used in a suitable environment.
• Only Air Liquide Medical Systems adapters may be used with the IRMA™ probe.
CAUTION: Use only IRMA™ probes sold by Air Liquide Medical Systems: Cat. No.:
KB020400.
48
4 Use
• Do not connect the IRMA™ at the outlet of an elbow connector, in order to avoid an accumulation
of patient secretions in the adapter and the obstruction of XTP™ windows.
• When using the IRMA™ probe, place the adapter so that its XTP™ windows are vertical, to avoid an
accumulation of patient secretions on these windows:
• When using the IRMA™ probe, position the adapter so that moisture and secretions do not accu-
mulate inside it due to the effect of gravity. Replace the adapter if it has been rained on or if it is
damp.
• Do not use any aerosol medication with the IRMA™ probe, to avoid a deterioration of IR transmis-
sion through the windows of the adapter.
• If the calibration is not done properly, the measurement values will be skewed.
• Never sterilize the IRMA™ probe or immerse it in liquid.
• Do not pull on the cable of the IRMA™ probe.
• Do not use the IRMA™ probe at a temperature below 0°C or above 40°C.
• During use, ensure that the IRMA™ probe is connected to the ventilator before you connect it to the
patient.
• The probe can be cleaned using a cloth dipped in alcohol (maximum 70% ethanol or 70% iso-
propanol).
• Remove the adapter before cleaning the probe.
• The adapters are not autoclavable.
• The gas measurements supplied by the probe must be checked regularly using a reference instru-
ment. We recommend that this check be performed annually.
49
Monnal T75 user manual
Note: Spontaneous patient activity is represented by the curves displayed on-screen chang-
ing color. When the respiratory cycle comes from a patient demand, the curves are green, but
a controlled cycle is displayed in yellow.
Changing curves
To change curves:
1. Press the curve to be replaced. A menu appears in the bar on the right-hand side of the screen
(1).
2. Select the desired curve in the menu. The curve is displayed automatically.
3. Press the [Back] key to quit the change curve menu.
Adjustment of scales
To adjust the time scale, press the [s] key (5).
To adjust the pressure scale, press the ordinate axis of the pressure curve or the [cmH2O] key (3).
To adjust the flow rate scale, press the ordinate axis of the flow rate curve or the [L/min] key (2).
To adjust the volume scale, press the y-axis of the volume curve or the [mL] key.
To adjust the CO2 scale, press the y-axis of the CO2 curve, the [mmHg] key, the [kPa] key or the [%] key.
Freezing the curves
Pressing the [Freeze] key immediately freezes the curves (4).
Two cursors (vertical lines) are then available in order to measure and display the values of each
curve. The cursor currently selected appears in green.
Use the control wheel to select, modify or move the cursor.
The symbol Δ indicates the pressure, flow rate, volume, CO2, and time interval between the two cur-
sors.
To exit from frozen curve mode, press the [Back] key.
Note: The digital display of patient parameters continues to be refreshed on-screen whilst
the frozen are being displayed.
50
4 Use
4.11 Monitoring
Eight respiratory parameters (1) are continuously displayed on the screen.
51
Monnal T75 user manual
AVCV, APCV, SIMV, PSIMV, AVCV, APCV, SIMV, PSIMV, PSV NIV, CPAP and Duo-
PRVC Modes PRVC Modes Levels Modes
Page 2 %leak MVe spont %leak MVe spont RR/VTe MVe spont
Pmean VTi Pmean VTi Ti/Ttot* VTi
RR spont % spont Pplat %spont Pmean %spont
peakl
peakl peakE peakE leak maxE
Inspiratory peak pressure of the Volume insufflated during the cycle (invasive
cycle ventilation)
52
4 Use
RR Respiratory frequency
53
Monnal T75 user manual
When the inspiratory plateau is not activated, the Pplat measurement is displayed as follows:
The static and dynamic resistance and compliance values are displayed in the same monitoring pan-
els and in the context of invasive ventilation only.
• When the RC stat function (see [R&C stat] key on page 58) is disabled, the dynamic resistance
and compliance values are displayed;
• When an RC stat measurement has been taken, the static values are displayed for a certain time
before the dynamic values reappear.
54
4 Use
58
59
• To set a threshold:
• Select the value to be adjusted: it is highlighted.
Adjust the desired value by turning the control wheel, and press the control wheel to validate.
When an alarm threshold has been exceeded, the block and the associated alarm threshold turn red
(58). This display is maintained as long as the alarm is present.
When the alarm disappears, the block resumes its original appearance, but the threshold concerned re-
mains red (59): this indicates that the alarm took place.
CAUTION: After a new patient selection, the alarm thresholds are automatically reset to their
standard value.
These threshold settings must be systematically re-assessed in order to ensure that they suit the pa-
tient and his ventilation.
The upper pressure threshold is especially important to protect the patient from excessive airway
pressure.
The setting of alarm thresholds to the extreme values in the adjustment ranges can render the alarm
system ineffective.
The alarm thresholds can also be adjusted automatically according to the values measured. Refer to
the next section.
55
Monnal T75 user manual
Note: During a total power outage (AC power, external and internal batteries) the alarm
threshold settings are retained in memory and may be displayed after the electrical power is
restored. The settings are saved by a backup battery with a life of approx. 2 years. An alarm
is issued when the battery life becomes too short. The battery must be replaced. A power
outage of more than 2 years will cause the settings to be lost. At the next startup, the alarm
thresholds will be automatically restored to their default values.
MVe low threshold settings
In PSV NIV mode, the user has the option of setting the lower threshold of the expiratory minute vol-
ume (MVe) in the OFF position.
This setting is used to inhibit the alarms associated with the low expiratory minute volume (MVe) par-
ticularly in the event of non-invasive ventilation with major leaks.
The [Auto set] button automatically configures the alarm thresholds according to the values measured
at the time when it is pressed.
To adjust the automatic thresholds, press on a monitoring pad and the [Auto. threshold] button will ap-
pear on the right of the screen
Press the [Auto setting of alarm thresholds?] button, then confirm by pressing the dial: all the alarm
thresholds will update automatically.
CAUTION: After pressing the [Auto. Set] button, verify that the thresholds obtained in this
way are appropriate for the clinical condition of the patient.
Note: The [Auto set] key is deactivated when the screen is locked.
4.11.4 Resetting
As soon as an alarm threshold has been exceeded, it is highlighted in red, and remains highlighted
even when the alarm disappears.
To reset all the thresholds highlighted in red, press the [Reset] button.
The alarm history shows the latest alarm-related events recorded by the ventilator.
56
4 Use
The [History] key opens a window which displays a chronological list of the last 200 alarms.
For each alarm there is a date, priority (color), time, activation or deactivation (ON or OFF), name, and
for physiological alarms, the alarm threshold setting at the time of activation.
Use the control wheel to browse the list.
To exit the alarm history, press the control wheel.
Note: When a complete power cut occurs (mains, external and internal batteries), the histo-
ry is preserved and can be displayed after the power supply has been restored. These data
are preserved thanks to a back-up battery with a working life of approximately two years. An
alarm is triggered when this battery runs low. The battery should then be replaced.
Note: When the log reaches its maximum size, the oldest events will be overwritten by in-
coming events.
57
Monnal T75 user manual
5 Measurements
Note: A curve freeze can be applied in order to measure the autoPEEP value via the cursors
(See Display of pressure and flow rate curves on page 49).
58
5 Measurements
Note: This window also displays the Pplat and Auto-PEEP results.
Measurement principle
The ventilator generates:
• end-expiratory occlusion;
• inspiration (consistent with the current mode), delivering a Vt;
• end-inspiratory occlusion.
Note: Inspiration between the two occlusions can be shortened to ensure that the flow rate
at end of inspiration is not zero. The measurement is interrupted in the event of Ppeak.
The Rstat and Cstat values result from the following calculations:
• Cstat = VT / (Pplat – total PEEP);
• Rstat = (Ppeak – Pplat) / end of inspiration flow rate.
Measurement accuracy is ± 20%.
Note: This function is only available in VCV, PCV and PRVC. In these modes, the patient is
sufficiently passive for the end-inspiratory and end-expiratory occlusion pressures to be sta-
ble.
59
Monnal T75 user manual
Note: The set periodicity remains in the memory when the function is disabled.
Note: P0.1 measurement is accessible in all modes in invasive ventilation except CPAP and
Duo-Levels. It is not available in non-invasive ventilation, where the risk of leakage does not
guarantee a reliable measurement.
Note: The set periodicity remains in the memory when the function is disabled.
60
5 Measurements
During the NIF operation, the patient circuit is blocked, and the low pressure generated by the patient
with respect to PEEP is measured. NIF is the patient’s maximum inspiratory effort. When the pressure
drop is less than -30 cmH2O, the probability of a successful extubation is higher.
The measurement result is written in the monitoring panel and in the trends, as well as a window
present during measurement.
The pressure curve is displayed on the upper part and the scale is changed so that it displays the neg-
ative pressures.
To activate the function, press the [Measurements] key and then press and hold the [NIF] key.
The LED lights up, and measurement begins, indicating the following message in the alarm panel: ‘NIF
measurement in progress’.
Note: As long as the key is pressed, the procedure takes place, the circuit is blocked, and the
pressure drop generated by the patient is measured.
Note: NIF measurement is accessible in all ventilation modes except CPAP, Duo-Levels and
APRV. It is not available in non-invasive ventilation, where the risk of leaks does not guarantee
reliable measurement.
Note: The procedure may be interrupted by the raising of alarms such as Ppeak.
61
Monnal T75 user manual
62
6 Menu
6 Menu
6.1 Description
The [Menu] key is on the right-hand side of the screen. It gives access to functions or commands via a
two-level structure.
The menu is accessible via the start-up, stand-by, and ventilation screens.
63
Monnal T75 user manual
Apnea ventilation VT
settings
RR
T apnea
TC Compensation level
Date/hour
Transfer
Brightness
Alarm
Calibration
Apnea ventilation uses an assisted volume-controlled mode with a frequency and tidal volume to be
adjusted. It keeps the PEEP and inspiratory trigger set in the current mode, Ti/Ttot of 33% and a de-
creasing flow rate form.
64
6 Menu
Apnea ventilation is triggered when no inspiration has been triggered for a time greater than T apnea.
It stops when:
• The patient triggers more than three consecutive cycles,
• The user presses the apnea ventilation Off key,
• The user changes ventilation modes.
The ventilation of apnoea may be deactivated in CPAP mode by setting the apnoea T set-point to OFF.
In such a case, an alarm sounds to confirm deactivation and should be acknowledged.
CAUTION: For safety reasons, the ventilation of apnoea should only be deactivated where
the clinical situation allows. Air Liquide Medical Systems recommends that apnoea ventila-
tion is activated.
Note: Ideally, the apnea ventilation adjustment should be performed before starting ventila-
tion.
To access the apnea ventilation adjustments, press the [Menu] key and then [Apnea ventilation set-
tings].
65
Monnal T75 user manual
Compensation level
Compensation level varies from ‘none’ to 100%. If compensation level is 100%, loss of tube pressure
will be fully compensated. Inversely, ‘none’ means that no compensation takes place and that the
Compensation Tube function is not enabled.
As an example, 50% compensation means that only half of loss of tube pressure will be compensated.
CAUTION: If a compensation level is set, an information message indicating that the func-
tion is enabled is displayed in the alarm banner: «intubation tube compensation enabled».
Tube type
There are 2 types: endotracheal or tracheostomy.
Tube diameter
Internal tube diameter varies from 2.5 mm to 11 mm. According to the patient category selected
(adult, child, infant), suitable minimum and maximum setting limits are available.
CAUTION: Users must check that the type but also the internal diameter of the tube match
that attached to the patient.
CAUTION: For tubes of a diameter smaller than 3 mm, compensation is deliberately un-
der-estimated to avoid barotrauma. Bottom pressure is actually lower than the value dis-
played.
CAUTION: Ensure there are no leaks in order to prevent barotrauma.
This function allows the device to enrich the O2 mix from a low-pressure source (typically, a concen-
trator).
This is different from a high-pressure source connection since it works on a lower scale (pressure gen-
erally less than 1 bar, flow rate less than 10 L/min):
• Depending on the nature of the source and the ventilation settings, certain FiO2 values cannot be
guaranteed. A chart is provided for information purposes in Technical description on page 102).
• The mechanisms for monitoring the presence/absence of the source must be adapted.
To activate the “Low Pressure” function, press [Menu], then [Ventilator configuration], then [Low Pres-
sure O2] then set the position to “ON”. The “Low Pressure FiO2” descriptive label then appears on the
alarm panel.
66
6 Menu
The enrichment of the O2 mix then takes place according to the FiO2 parameter, and the device then
delivers, as far as possible, the desired concentration. If the ventilation and low-pressure source set-
tings are incompatible, the FiO2 will not be reached and the “low FiO2” alarm will be generated. The
user will then be informed that these settings need to be modified.
CAUTION: To have efficient operation on a low O2 pressure source, the high pressure O2
connection hose must be disconnected from the device.
To prevent the low pressure O2 fitting becoming detached, ensure the pressure applied to the low
pressure oxygen inlet is less than 150 kPa (1.5 bar).
6.2.4 Sensors
CO2 monitoring
The [CO2Monitoring] key activates and deactivates CO2 monitoring.
See CO2 Option on page 43 for more information.
Calibration
The [CO2 sensor calibration] key calibrates the CO2 probe.
See CO2 Option on page 43 for more information.
67
Monnal T75 user manual
The [xx% O2 ↗] key is used to ensure the patient has oxygenation, at a FiO2 greater than the current
FiO2. It also makes it possible to inhibit the alarms and ventilation in order to carry out pulmonary aspi-
ration.
Note: If the oxygenation FiO2 is lower than the current FiO2 (see Settings), the button is dis-
abled.
During suction, the PEEP level is maintained to the extent possible, depending on the size of the leak,
in order to limit alveolar derecruitment.
Note: Suction may be carried out according to different methods (fully unplugging the circuit,
opening a connection from the respiratory circuit, or “closed system”).
2- Suction
During this step, the ventilator no longer generates ventilation-related alarms.
Additionally, ventilation is interrupted, thereby preventing the secretions contained in this circuit from
spreading. PEEP, however, is maintained, which is improved since the circuit’s leak is minimal.
Once aspiration is complete and the patient circuit has been reconnected, the ventilator automatically
detects the reconnection and activates the end-of-suction oxygenation phase.
68
6 Menu
Note: The suction time is limited to 40 seconds. Afterward, the end-of-aspiration oxygenation
phase (N°3) is automatically started.
Note: The user can also restart the ordinary ventilation by pressing the [Validate] button in
the dialogue box.
3- End-of-suction oxygenation
This step is used to re-oxygenate the patient after his or her disconnection during pulmonary suction.
A countdown indicates the time remaining for oxygenation.
At the end of this countdown, oxygenation shuts down and the [xx% O2 ↗] key turns off.
Note: It is possible to shut down the function at any time by pressing the [xx% O2 ↗] key.
Settings
The different oxygenation settings can be adjusted via a window of the [O2 ↗ Settings] [Menu].
This menu is used to set:
• Target O2: FiO2 applied to the whole oxygenation-suction period.
• Duration of oxygenation before suction (step 1 “Oxygenation”)
• Duration of oxygenation after suction (step 3 “End-of-aspiration oxygenation”)
For more information, refer to Checklist on page 135.
Whenever alarm #57 “Expiratory flow measurement inoperative” triggers, it appears as a high
priority (red color code).
69
Monnal T75 user manual
1. Touch the alarm banner. The “Choice of sensors” area appears on the right of the screen,
2. Disable the expiratory flow sensor by pressing the [Expi. Flow].
The active alarm disappears, then an informative message comes up: “Expiratory flow measure-
ment inactive, use an external monitor”.
This message stays displayed.
3. Press the [Back] button to return to the previous screen.
6.3.4 Nebulization
The ventilator can be used with a pneumatic nebulizer. It supplies a constant outlet pressure of 1.2 bar
(see Nebulization on page 21). The nebulization flow rate then depends on the choice of the nebulizer.
Because the only high-pressure source available in the ventilator is the high-pressure O2 inlet, nebu-
lization is possible only with the machine connected to the network (or a cylinder). The flow from the
nebulizer is then oxygen: FiO2 can therefore vary in the event of nebulization.
CAUTION: The accuracy of the Monnal T75 may be degraded by gases added during the use
of the nebulizer.
In order to limit these FiO2 variations and keep the ventilation parameters (such as tidal volume) con-
stant, the ventilator applies systematic compensation of the nebulization flow rate by reducing its own
outlet flow rate by an equivalent amount (and by prioritizing the reduction of its oxygen flow rate).
The FiO2 reading is displayed on screen; the value shown represents the concentration in the Y-piece
(i.e., it includes the nebulization flow).
Note: The display of the FiO2 value is an estimation of the gas concentration and cannot
substitute external monitor precision.
There are several mechanisms to alert the user to any changes in ventilation:
Oxygen concentration not observed
When the estimated FiO2 measurement in the Y-piece is more than 30% greater than the concentration
setting on the ventilator, an acknowledgeable HIGH alarm is triggered: “High FiO2!!! Nebuliza-
tion stopped”.
Nebulization is then shut off.
Nebulization flow greater than the machine output flow rate.
When the flow rate from the nebulizer is greater than the output flow rate of the machine, nebulization
is automatically shut off. This occurs, for example, in pressure-controlled cycles where the end-of-in-
spiration flow rate can be zero.
If nebulization is shut off for more than half the inspiratory cycle, an acknowledgeable MEDIUM alarm
is triggered: “Nebulization ineffective!!”.
Nebulization continues in spite of this.
CAUTION: Nebulization is deactivated for the pediatric use of the device.
70
6 Menu
71
Monnal T75 user manual
Manual adjustment
The device supplies a nebulization pressure of 1.2 bar. The user must refer to the nebulizer instruc-
tions to find out the correct nebulization flow rate for this pressure, and then adjust it accordingly on
screen.
Nebulization time
This defines the duration of the nebulization session.
Periodicity
This defines the frequency of the nebulization sessions.
Press the [Periodicity] button to set the frequency of nebulization (hours).
If necessary, you can activate just one nebulization session which will not be repeated periodically. To
do this, set the PERIODICITY parameter to “NONE”.
To go back to the previous screen, press [Back].
When the loop curve is frozen, two cursors appear and allow the value of the curve to be measured at
the point defined by the cursor.
To select the cursor, press the control wheel.
72
6 Menu
6.3.6 Trends
This function allows the user to monitor the development of the following ventilation settings over a
maximum 80 hours:
2
To change the top trend and display another parameter, press the current parameter button (top-left of
the graph) and choose the desired parameter.
To change the top trend and display another parameter, press the current parameter button (top-left of
the graph) and choose the desired parameter.
73
Monnal T75 user manual
The trend curves can be displayed every 2, 5, 10, 20 or 40 hours. To modify the time scales, use the
zoom + or - buttons (1).
To scroll the trends curves from left to right, use the arrows (2).
The user can display the precise values at a given moment (date and time) of the selected ventilation
parameters using the vertical line.
The user can thus move the line along the trends curve and read off the values of the two previously
defined parameters.
74
7 Administrator configuration
7 Administrator configuration
7.1 Access
The [Administrator config.] menu is accessible from the Stand-by mode. It is only accessible to autho-
rized persons.
After a special access procedure by authorized persons, press the [Administrator config.] button in the
right-hand bar to display the [Administrator config.] screen.
To close the [Administrator config.] menu, press the [Exit Administrator config.] button.
60 68
61
62
63
64 69
65
66 70
67
7.2 Presentation
The [Administrator config] menu enables the prescriber to customize the configuration of default ven-
tilation parameters (set-points, alarms, etc.)
The new parameters are applied whenever a new patient is selected.
75
Monnal T75 user manual
76
7 Administrator configuration
In order to enable optimum operation in these conditions, it must be indicated that Monnal T75 is con-
nected to such a network. Select the O2 network to do this:
• Go to the ‘Administrator configuration’ menu;
• Press the [O2 supply altered] button and select [Enabled];
• Quit the [Administrator configuration] menu.
Note: An information message on green background ‘Operating on altered network’ is dis-
played in the alarm display area when the device is in Standby mode.
Monnal T75 automatically measures the purity of the network, during the automatic test and adjusts
its algorithms accordingly.
Note: The automatic test must be launched before use on patients.
CAUTION: It must be ensured that the Monnal T75 mode of operation matches the actual
network oxygen concentration to prevent dysfunction (notably the appearance of technical
alarms)
Depending on oxygen concentration, the algorithms monitoring technical failures of the mixer may be
less sensitive.
Note: Depending on network concentration, certain set-points may not be available.
Note: The estimated FiO2 for nebulisation may be subject to error, up to the natural varia-
tions in network purity.
77
Monnal T75 user manual
Note: The operation must be repeated for each patient category (adult/child/infant).
Note: The [Retrieve default settings] key is used to return to the factory configuration.
To reactivate the modes: press the desired mode, turn the control wheel and validate by pressing the
control wheel again.
The reactivated modes are then displayed on the screen.
Configurations are saved when you press the [Exit Modes selection] key.
78
7 Administrator configuration
To restore default configuration mode, press the [Retrieve default settings] key.
79
Monnal T75 user manual
8 Alarms
Alarms are indicated by on-screen displays and the emission of audible sequences.
The header panel indicates the alarm status and level. Its color varies according to the importance:
red, yellow, cyan or green.
On the left of the alarm band is a software alarm indicator and an alarm silence button. The two but-
tons are displayed as follows:
Note: The software alarm signal flashes when an ULTRA, HIGH or MEDIUM alarm is raised.
The arrow means that there are several alarms at the same time.
8.1 Parameters
Alarms are ranked according to 3 priority levels, with different sounds and colors.
ULTRA alarms are high priority alarms referring to a device or power supply failure. HIGH alarms are
high priority alarms referring to a ventilation failure.
The INFO priority is displayed on a green background with the symbol .
80
8 Alarms
To indicated a technical error, the alarm also displays the symbol and a number so that the techni-
cal department can identify the exact source of the alarm.
If several alarms are triggered, a small arrow appears on the right-hand side of the panel, and only the
highest-priority alarm is visible. Press the panel to display the list of current alarms.
If multiple alarms are displayed in this list, the user can browse all alarms using the control wheel.
Note: The audible alarm is for a doctor working near the patient. The maximum distance
away and the volume of the alarm must therefore be determined by this person according to
the situation.
Press to cancel the alarm inhibition. The alarm message remains on the screen.
If this alarm is replaced by another audible alarm with a higher priority, the inhibit function is automati-
cally disabled and the new alarm begins to sound. The previous alarm is then automatically archived in
the history.
0 Shut off ventilation? Placing on STAND- Immediate triggering Validate using the
HIGH
BY mode control wheel to stop
[acknowledgeable] ventilation
1 Ventilator shutdown? Unit Off request Immediate triggering Validate using con-
HIGH
trol wheel to switch
[acknowledgeable] off the unit
81
Monnal T75 user manual
3-7 Machine out of service!!! Use a back- Electronics failure Immediate triggering
HIGH Use a different unit
up ventilator
and contact the tech-
(ULTRA) nical department
10 Patient disconnection!!! HIGH The patient is dis- Triggered within 15 s Check the level of
connected leakage
12 Error detected!!! Contact the tech. HIGH Electronics failure Immediate triggering Use a different unit
department and contact the tech-
nical department
14
16 Incorrect alarms settings!!! HIGH Electronics failure Immediate triggering Use a different unit
and contact the tech-
nical department
18 Low frequency!! MEDIUM Measured frequen- Immediate triggering Check the suitability
cy below the alarm of alarm levels with
threshold ventilation settings
Ventilation downgraded! Replace the HIGH Electronics failure Immediate triggering Use a different unit
19
machine ASAP. and contact the tech-
- nical department
20
82
8 Alarms
In case of opera-
tion on low pressure
source, check that
the high pressure O2
connection hose is
disconnected from
the device.
22 O2 supply pressure too high!!! HIGH O2 supply pressure Immediate triggering Check pressure in
> 7 bar wall network.
Safety Ventilation!!! recommend HIGH Electronics failure Immediate triggering Use a different unit
23
switching ventilator and contact the tech-
- nical department
24
26 High respiratory rate!! MEDIUM The frequency mea- Check the suitability
Triggered
sured by the ventila- of alarm levels with
tor is above the set after 5 s ventilation settings
threshold
Error detected!!! Contact the tech. HIGH Electronics failure Immediate triggering Use a different unit
29
department and contact the tech-
- nical department
30
83
Monnal T75 user manual
Error detected!!! Contact the tech. HIGH Electronics failure Immediate triggering Use a different unit
32
department and contact the tech-
- nical department
33
35 Internal battery inoperative!!! Connect The battery is empty. Immediate triggering Immediately connect
HIGH
AC power to mains power sup-
(ULTRA) ply.
36 Internal battery discharged!!! Connect Internal battery al- Immediate triggering Connect the unit to
HIGH
AC power most empty mains (battery will
[acknowledgeable] charge
Power Fault, connect to external HIGH Electronics failure Immediate triggering Use a different unit
38
source!!! Ventilation Effective and contact the tech-
- nical department
39
triggering
Safety Ventilation!!! recommend HIGH Electronics failure Immediate triggering Use a different unit
42
switching ventilator and contact the tech-
- nical department
45
46 Patient circuit leak detected during HIGH Patient circuit con- Triggering on exit Change the patient
automatic tests!!! nection problem from the automatic circuit, connect pa-
test tient circuit compo-
nents carefully, and
repeat the automatic
test
47 High FiO2 too!!! Nebulization stopped Incompatibility be- Immediate triggering Readjust the set-
HIGH
tween settings and points
[acknowledgeable] the nebulization
function
48 Safety Ventilation!!! recommend HIGH Electronics failure Immediate triggering Use a different unit
switching ventilator and contact the tech-
nical department
84
8 Alarms
49 Restart the automatic tests!!! HIGH The atmospher- Triggering when the Start the automatic
ic pressure has ventilator is started test in order to cali-
changed of more brate to the new at-
than 50 mB since the mospheric pressure
last automatic test
Safety Ventilation!!! recommend HIGH Electronics failure Immediate triggering Use a different unit
50
switching ventilator and contact the tech-
- nical department
52
Check O2 sensor
connection
54 Safety Ventilation!!! recommend HIGH Electronics failure Immediate triggering Use a different unit
switching ventilator and contact the tech-
nical department
55 Sounder failure !!! Ventilation Effec- HIGH Electronics failure Immediate triggering Use a different unit
tive and contact the tech-
nical department
56 Expiratory flow measurement inoper- HIGH Hot Wire Sensor bro- Immediate triggering Replace the expira-
ative!!! ken tion flow rate sensor
57 Safety Ventilation!!! recommend HIGH Electronics failure Immediate triggering Use a different unit
switching ventilator and contact the tech-
nical department
58 Safety Ventilation!!! recommend HIGH Electronics failure Immediate triggering Use a different unit
switching ventilator and contact the tech-
nical department
59 Low VTe!! MEDIUM Exsufflated volume < Triggered after 6 Check the suitability
set threshold consecutive ventila- of alarm levels with
tion cycles ventilation settings
60 High VTe!! MEDIUM Exsufflated volume > Triggered after 6 Check the suitability
set threshold consecutive ventila- of alarm levels with
tion cycles ventilation settings
61 Screen fault!!! Ventilation Effective HIGH Electronics failure Immediate triggering Use a different unit
and contact the tech-
nical department
62 High plateau pressure!! MEDIUM Lung pressure > Triggering at an in- Check the suitability
alarm threshold spiratory plateau of alarm levels with
ventilation settings
63 Low VTi!! MEDIUM Insufflated volume < Triggered after 3 Check the suitability
set threshold consecutive ventila- of alarm levels with
tion cycles ventilation settings
In pressure-control
mode, check leakage
level
85
Monnal T75 user manual
Safety Ventilation!! recommend MEDIUM Electronics failure Immediate triggering Use a different unit
67
switching ventilator and contact the tech-
- nical department
69
70 Screen fault!! Ventilation Effective MEDIUM Electronics failure Immediate triggering Use a different unit
and contact the tech-
nical department
71 Internal battery nearly discharged!! Battery low Immediate triggering Ventilation can con-
MEDIUM
tinue but the unit
[acknowledgeable] could stop at any
time. Immediately
connect the unit to
the mains power
supply
72 Error detected!! Contact the tech. de- MEDIUM Electronics failure Triggering when the Use a different unit
partment ventilator is started and contact the tech-
nical department
In case of opera-
tion on low pressure
source, check that
the high pressure O2
connection hose is
disconnected from
the device.
75 Low VTi! LOW Insufflated volume < Immediate triggering Check the suitability
set threshold of alarm levels with
ventilation settings
76 Low respiratory rate! LOW The measured fre- Immediate triggering Check the suitability
quency is below the of alarm levels with
set threshold ventilation settings
77 VT not reached! Check settings LOW The blower is on Immediate triggering Check the set-points
maximum and the
volume reached is
30% below the vol-
ume setting
78 Patient demand higher than set peak LOW Settings are below Triggered after 5 Adjust ventilation
flow! patient capacity consecutive ventila- settings to suit the
tion cycles patient
86
8 Alarms
In pressure-control
mode, check leakage
level
80 High respiratory rate! LOW The measured fre- Immediate triggering Check the suitability
quency is above the of alarm levels with
threshold ventilation settings
Non Critical error! Ventilation Effec- LOW Electronics failure Immediate triggering Use a different unit
81
tive and contact the tech-
- nical department
86
87 Ventilator operates from internal bat- Loss of mains power Immediate triggering
LOW Battery life (see Bat-
tery! supply
tery life check on
[acknowledgeable] page 99)
88 Ventilator operates from external bat- LOW [acknowledge- Loss of mains power Immediate triggering Monitor the remain-
tery! able] supply ing charge in the ex-
ternal battery and
connect the unit to
the mains if the bat-
tery is low
89 FiO2 sensor to be soon replaced! LOW Voltage below a Immediate triggering Replace the O2 sen-
threshold sor the next time you
put the unit in ser-
vice.
91 Did you start the automatic tests with Nebulization activat- Immediate triggering Run the automatic
LOW
the nebulizer? ed without having test
[acknowledgeable] run automatic test
with nebulizer con-
nected
96 Low pressure O2 activated INFO Press the “O2 low Immediate triggering
presure” button
87
Monnal T75 user manual
101 FiO2 sensor calibration: plan to Deviation from the Immediate triggering
LOW
restart the automatic tests! O2 sensor measure-
[acknowledgeable] ment
103 Oxygenation in progress – Suction INFO Press the “xx% Immediate triggering Reconnect the pa-
possible O2↗” button tient
105 Expi. flow measurement disabled, INFO Disable the hot wire Immediate triggering Use an external mon-
use an external monitor sensor itor
106 Low minute volume alarm disabled LOW Immediate triggering Set a low MVe alarm
threshold different
from OFF
107 Press detected on \nthe touch LOW Electronic fault Immediate triggering Change the ventila-
screen! tion and contact the
tech. department
108 Vt target and set pressure incompat- MEDIUM Exhaled volume < Vt Immediate triggering Check that settings
ibility!! target and Delivered of Vt target and
pressure = Ppeak Ppeak are consistent
+ 5 cmH2O)
109 Oxygen therapy enabled, Apnea venti- INFO Press the 'Oxygen Starts after pressing
lation disabled therapy' button the 'Start' button
110 Oxygen therapy branch may be ob- HIGH Pressure in pa- Starts after 4 s
Check the patient cir-
structed!!! tient circuit > alarm
cuit, the patient/ma-
threshold
chine interface and
the inspiratory limb
of the ventilator.
111 R and C stat measurement enabled INFO Press the 'R&C stat' Immediate triggering
button
112 'P0.1 measurement enabled' INFO Press the 'P0.1' but- Immediate triggering
ton
88
8 Alarms
113 Intubation tube compensation en- INFO Press the 'TC' button Immediate triggering
abled
115 NIF measurement in progress INFO Press the NIF key Immediate triggering
116 Low pressure!!! HIGH Pressure in pa- Triggered after 3 Check the level of
tient circuit < alarm consecutive ventila- leakage
threshold tion cycles
117 Low pressure! LOW Pressure in pa- Immediate triggering Check the level of
tient circuit < alarm leakage
threshold
Cycle.M key
119 CO2 apnea!!! HIGH CO2 sensor: No pa- Immediate triggering A clinician must re-
tient breaths have evaluate the patient's
been detected for ventilation
more than 20 sec-
onds
120 Check IRMA probe (CO2) adapter! LOW CO2 sensor: UU Immediate triggering Check the UU
adapter is not prop- adapter
erly placed on the
probe
121 Replace IRMA probe (CO2) adapter! LOW CO2 sensor: UU Immediate triggering Change the adapter
adapter is obstruct-
ed or blocked
122 CO2 concentration is outside toler- LOW CO2 measured > Immediate triggering If CO2 is correctly
ance! 15% < 15%, calibrate the
probe
123 Calibrate the IRMA probe (CO2)! LOW CO2 sensor: Probe Immediate triggering Calibrate the probe
drift
124 IRMA probe error (CO2)! LOW CO2 sensor: probe Immediate triggering Disconnect and re-
hardware or soft- connect the probe. If
ware error the problem persists,
change probes.
125 CO2 measurement is inoperative! LOW CO2 sensor: Probe is Immediate triggering Connect the IRMA
not connected probe
126 IRMA probe (CO2): Internal tempera- LOW CO2 sensor: The in- Immediate triggering If the ambient tem-
ture outside tolerance! ternal probe temper- perature is normal,
ature exceeds the change probes
maximum measure-
ment threshold
127 IRMA probe (CO2 Ambient pressure LOW CO2 sensor: The at- Immediate triggering If the ambient pres-
is out of tolerance! mospheric pressure sure is normal,
measured by the change probes
probe exceeds the
maximum measure-
ment threshold
89
Monnal T75 user manual
130 Re-breathing detected!! MEDIUM re-breathing > thresh- Immediate triggering Reduce the dead
old (4 mmHg) space
131 VMalv high!! MEDIUM VMalv value is above Immediate triggering Check the suitability
the set threshold of the alarm levels
with the ventilation
settings.
132 VMalv low!! MEDIUM VMalv value is below Immediate triggering Check the suitability
the set threshold of the alarm levels
with the ventilation
settings.
133 VMCO2 high!! MEDIUM VMCO2 value is Immediate triggering Check the suitability
above the set thresh- of the alarm levels
old with the ventilation
settings.
134 VMCO2 low!! MEDIUM VMCO2 value is be- Immediate triggering Check the suitability
low the set threshold of the alarm levels
with the ventilation
settings.
136 Expiratory limb obstructed!!! Emer- HIGH Respiratory-tract Sounded after 5 s Check the condition
gency ventilation pressure during ex- in the event of low of the expiratory limb
halation > threshold frequency; otherwise and eliminate the
during exhalation after 2 cycles. cause of the obstruc-
tion. Backup ventila-
tion. Contact Techni-
cal Support.
Patient respiratory
contact with room
air as backup ventila-
tion.
90
8 Alarms
138 Apnoea ventilation deactivated! LOW The Tapnea set-point Immediate triggering The Tapnea set-point
switches OFF during switches to a value
CPAP ventilator, or other than OFF dur-
the mode starts up ing CPAP Ventilator
with Tapnea set to or ventilation other
OFF than CPAP.
139 Reduced battery power for transfer!! Battery capacity 50% Immediate triggering See battery (see Bat-
MEDIUM
tery life check on
[acknowledgeable] page 99)
140 Insufficient battery power for transfer Battery capacity 25% Immediate triggering
HIGH The battery has
reached its operating
[acknowledgeable] limit.
91
Monnal T75 user manual
9 Maintenance
Accessories can be re-usable (autoclavable) or single-use (disposable).
Re-usable elements must be regularly cleaned and sterilized to prevent infection.
This procedure, which is mandatory and extremely important, is the responsibility of the user.
9.1 Definitions
Cleaning
The act of removing all traces of soiling from a place, a surface, or an element.
Sterilization
Total destruction of all germ strains, viruses, and yeasts.
A sterilization or disinfection procedure is never possible on dirty or soiled elements.
A complete procedure includes:
1. Disassembly, pre-disinfection, rinsing and drying;
2. Cleaning, rinsing and drying;
3. Disinfection, rinsing and drying, or sterilization;
4. Reassembly and functional tests
CAUTION: Never use abrasive powders, acetone, or other powerful solvents.
Note: The following instructions have been approved by the manufacturer of the medical
equipment as ENABLING a medical device to be prepared for re-use.
It is always the responsibility of the department concerned to ensure that the sterilization procedure
actually implemented via the equipment, materials and personnel of the reprocessing facility achieves
the expected result.
This generally requires the validation and routine inspection of procedures.
92
9 Maintenance
Expiratory 134°
flow sensor
MAX 2
serial number part number
93
Monnal T75 user manual
After a cleaning / disinfection cycle (immersion, autoclave, etc.), the hot wire expiratory flow sensor
must be thoroughly dried before being reinstalled, connected to the expiratory valve Monnal EVA, and
connected to the ventilator.
Make a visual check to ensure that the various components are in good condition.
Sterilization of the reusable expiratory flow sensor and expiratory valve assembly before the first use
is recommended.
Preparation
Remove the expiratory assembly from its housing by pressing the eject pushbutton.
Separate the expiratory flow sensor (67) from the expiratory valve Monnal EVA, and then remove the
valve body (68) and the membrane (69).
Pre-disinfection / Cleaning
1. Immerse the expiratory assembly components (67,68,69) in a pre-disinfectant solution:
Air Liquide Medical Systems recommends the use of ANIOS products: ANIOSYME DD1,
SALVANIOS PH7, HEXANIOS G+R (obey product manufacturer instructions).
2. Shake the pre-disinfected components thoroughly in a cleaning solution*,
3. Rinse under running water*,
4. Dry the components fully.
*except the expiratory flow sensor
Sterilization
If necessary, sterilize the following parts of the expiratory assembly (compatible with sterilization for
18 min at 134°C and 1 bar above atmospheric pressure): expiratory flow sensor, valve body and mem-
brane.
Reassembly
Reassemble the expiratory assembly.
CAUTION: Incorrect reassembly of the expiratory assembly (incorrect positioning of the di-
aphragm in the valve body, omission of the silicone disc, placement of an additional silicone
disc, etc.) may result in degraded equipment performance.
9.4 Monnal Clean’in filter (Type filtre HEPA: High- Efficiency Particulate
Air)
The Monnal Clean’in filter is an air filter that is used to purify the air at the turbine’s intake.
It blocks 99.97% of all particles.
Air Liquide Medical Systems recommends checking this filter every three months.
A place on the filter’s label enables the user to indicate an installation date.
The Monnal Clean’in filter must be replaced at least once per year.
94
9 Maintenance
CAUTION: The Monnal Clean’in filter may not be disinfected, cleaned, or sterilized.
95
Monnal T75 user manual
10 Accessories
Air Liquide Medical Systems recommends using the following consumables and accessories with the
Monnal T75.
CAUTION: The use of non-recommended parts may result in degraded performance. It is the
user’s responsibility to verify the compatibility of the ventilator and all parts used to connect
the patient prior to using the ventilator.
Expiratory assembly
O2 Cell
Filters
96
10 Accessories
Consumables Reference
Patient circuits
Group of 20 single-use adult-patient circuits, 1.6 meter without any water-trap, KG020100
but with a smooth inside surface
Can be connected to a pneumatic nebulizer
Group of 20 single-use child-patient circuits, 1.6 meter without any water-trap, KG020200
but with a smooth inside surface
Can be connected to a pneumatic nebulizer
Group of 15 single-use adult-patient circuits, 1.6 meter with a water-trap on the VD315100
inspiratory circuit and a corrugated surface expiratory circuit
Cannot be connected to a pneumatic nebulizer
97
Monnal T75 user manual
Ultrasonic nebulizer
Humidifiers Reference
98
11 Maintenance
11 Maintenance
Certain maintenance operations are the responsibility of the user, whilst others must be performed by
a technician.
For any operation requiring the ventilator to be opened, call on a technician.
Air Liquide Medical Systems holds available to maintenance staff a maintenance manual containing
the circuit diagrams, component lists, descriptions, calibration instructions and all other information
relevant to maintenance staff.
In order to assist users to check battery life, an acknowledgeable alarm ‘Perform battery test!’ is trig-
gered every 6 months, and prompts users to perform a battery test.
This test is accessible in Standby mode by pressing the [Menu] and [Machine Configuration] keys.
Note: To access the test, the device must be connected to the AC power supply, and the bat-
tery must be fully charged. The test may last for more than 3 hours.
Once the test selected, a window is displayed, and users may start or cancel the test.
Note: Do not connect to other devices (patient circuit). Do not disconnect the device from
the AC power supply.
When the test starts, the device will automatically switch to its internal battery. Discharge it completely
then automatically switch to the AC power supply. A window displays a figure indicating the remaining
battery life, compared to a new battery (3 hours’ battery life on standard ventilation as indicated in the
technical description in this manual).
99
Monnal T75 user manual
In order to preserve a record of the test performed, the information is permanently displayed next to
the [Battery test] access key, and includes remaining battery life and test date.
Note: This test resizes the battery charge indicator displayed on-screen: for example, if the
battery is down to 50% of its initial charge, no more than three of the five pictogram cells can
be lit. This function keeps the user aware of the overall state of the battery.
11.3 O2 cell
Note: The working life of the O2 cell is approximately 5000 hours, but this can vary according
to the oxygen concentration used and the ambient temperature.
100
11 Maintenance
1.
To open the compartment, press the cover down slightly and turn towards the symbol.
2. Disconnect the used sensor and then unscrew it to remove it from its housing.
3. Replace it with a new sensor, referring to the instructions on the packaging of the new sensor if
necessary.
4. Tighten the sensor properly in order to prevent leakage, and then connect it.
5.
To close the cover, press it down slightly (with the arrow pointing to ) and turn it until the
arrow points to .
6. Run the sensor calibration automatic test.
Note: It is recommended that you calibrate the O2 sensor frequently via the automatic test.
101
Monnal T75 user manual
12 Technical description
12.1 Operation
O2 HP High-pressure O2 inlet
(network, cylinder)
C3 Non-return valve
F2 O2 inlet filter (oxygen concentrator) PS5 Airway inspiratory pressure sensor (redun-
dancy
FS4 Hot wire expiratory flow sensor V4 PEEP control on/off solenoid valve
102
12 Technical description
The blower (T) of the ventilator entrains ambient air via the microfilter (F1) and compresses it accord-
ing to the patient and the settings concerned. The compressed gas is then distributed via a pneumatic
network according to whether an inspiratory or expiratory phase is being delivered.
Inspiratory phase
The main component during the inspiratory phase is the solenoid valve (V6) which regulates the flow
rate via the flow sensor (FS3) when the selected mode is volume-controlled, or via the pressure sensor
(PS4) when the selected mode is pressure-controlled.
At the same time, the solenoid valves (V3), (V4) and (V5) are open, closed, and closed respectively
such that the blower pressure is applied to the membrane (M2) of the expiratory valve, thus forcing the
air sent via the solenoid valve (V6) to flow towards the patient only.
Expiratory phase
In this phase, the patient exhales the gases inhaled in the previous phase, and the unit is expected to
regulate to a pressure determined by the settings (PEEP).
With this in mind, the solenoid valve (V4) opens to depressurise the membrane (M2) and the PEEP
control proportional solenoid valve (V3) regulates the expiration pressure via the pressure sensor
(PS4).
At the same time, the inspiratory electromagnet regulates the flow via the flow sensor (FS3) for a rinse
flow of 3 L/min. This flow limits reinhalation and allows quick detection of an inspiratory demand.
Note: During ventilation with leakage, such as NIV, the solenoid valve (V6) is liable to in-
crease the rinse flow; it then enters an “on-demand valve” mode. The purpose of this function
is to compensate for leaks in order to maintain the PEEP in the circuit.
Obstruction management
Version A:
In normal operation, solenoid valve (V7) connects the turbine pressure to diaphragm (M1), which plugs
the safety ambient air intake. If the patient needs to be vented to atmosphere, solenoid valve (V7) con-
nects diaphragm (M1) to the patient. On each inspiratory effort by the patient, the safety ambient air
intake opens and the patient can breath freely through the device.
Version B:
In normal operation, solenoid valve (V7) connects the turbine pressure to diaphragm (M1), which plugs
the safety ambient air intake. If the patient needs to be vented to atmosphere, solenoid valve (V7) con-
nects diaphragm (M1) to the atmosphere, which opens the ambient air intake and depressurises the
patient’s airways. In the special case of an obstruction in the expiratory limb, once the pressure in the
airways reaches atmospheric pressure, ventilation continues in accordance with the mode selected by
the operator. This is the backup ventilation system: the patient exhales through the ambient air intake
valve. Patient exhalation periodically occurs through the expiratory valve, to check how it is functioning
and permit normal ventilation when there is no longer an obstruction.
103
Monnal T75 user manual
Note: Because it is possible to exhale through the ambient air intake valve in backup ventila-
tion mode, Air Liquide Medical Systems recommends that a bacteria filter always be insert-
ed at the point of connection with the inspiratory limb to avoid contamination of the ventila-
tor. Since exhalation occurs through the inspiratory limb in backup mode, part of the expired
gas is reinhaled. The inspiratory trigger is disabled and a minimum backup frequency is main-
tained. Oversight and monitoring linked to the exhaled flow are altered.
Version B of patient venting applies from Monnal T75 device no. MT75-05000.
• A lower serial number identifies the device as version A.
• A higher serial number identifies the device as version B.
Note: For more information, contact Air Liquide Medical Systems.
The O2 concentration of the gases administered to the patient depends on the source connected to
the unit. A distinction is made between O2 network operation and concentrator operation.
Operation on an O2 network
For proper operation, the pressure at the O2 inlet terminals (O2 HP) must be between 2.8 and 6 bar.
The O2 is then filtered via F3.
The proportional solenoid valve (V1) enriches the mixture in oxygen by regulating the flow via the flow
sensor (FS1), where the set-point is proportional to the upstream flow rate (FS2) and depends on the
FiO2 setting.
Note: Operation at a network pressure between 1.5 and 2.8 bar or between 6 and 7 bar is
possible, but the quality of enrichment might be affected. If the pressure falls below 1.5 bar
or goes above 7 bar, the oxygen supply is cut off by the proportional solenoid valve (V1) and
an oxygen supply error alarm is triggered.
Operation with a concentrator
The ventilator is equipped with a “low pressure” connector for working with a flow meter (intake via the
filter (F2)). The principle of precise control of the O2 concentration applied at high pressure is main-
tained when operating with a low-pressure source, with the device constantly delivering a mix at the
correct FiO2.
104
12 Technical description
However, it may well be that the mix concentration cannot be achieved since it depends:
• on the type of concentrator and its settings (the O2 flow rate delivered by the concentrator is often
low and its O2 concentration variable [between 90 and 100%]);
• on the ventilation parameters.
The following graph shows the maximum concentrations that can be obtained with different concen-
trator flow rates (based on an O2 concentration of 100%) according to the minute volume of a patient
ventilated at the frequency of 15 bpm.
For example, for a concentrator delivering 5 L/min and a patient ventilated in volumetric mode with a
Vt of 0.5 L and a frequency of 15 bpm (i.e.: a minute volume of 0.5x15 = 7.5 L/min), the maximum con-
centration that may be obtained will be approximately 50%. The device will then be capable of deliver-
ing any required concentration below this value.
Note: Since the device only consumes the flow that it needs to ensure correct concentration
of the mixture, Air Liquide Medical Systems recommends that the low-pressure source be ad-
justed to its maximum flow rate. This will make it possible to obtain a wider range of possible
FiO2.
Whatever the function mode for the device, the oxygen sensor (O2S) ensures monitoring of the FiO2
concentration in the circuit.
This sensor is a chemical cell where an electrolytic reaction occurs. The current produced is propor-
tional to the partial pressure of oxygen.
However a compensation of this pressure is applied each time the automatic tests are lauched. Am-
bient humidity also affects the oxygen measurement at the rate of -0.03% per%Hr, at 25°C (see Mea-
sured parameters on page 122).
The O2 supply pressure (O2 HP) is reduced in the nebulization branch by the pressure reducer (PR1) at
a pressure of 1 bar.
The nebulization on/off solenoid valve (V2) provides nebulization via its open and closed positions.
The nebulization flow rate then depends on the nebulizer used.
The CO2 (CO2S) sensor ensures monitoring of the CO2 concentration during the expiration of the pa-
tient.
105
Monnal T75 user manual
The supply board manages the energy source changeover automatically according to the hierarchy as
follows:
1. Mains supply;
2. In the event of failure, power supply from external battery;
3. In the event of failure, power supply from internal battery.
Two pictograms indicate whether or not the mains power supply is present.
Mains present
Mains absent
In the event of absence of the mains and the external source, the switch to internal battery is indicated
by a low-priority acknowledgeable alarm “Ventilator operates from internal battery!”.
The charge status of the internal battery is then indicated in the upper left-hand part of the screen (ex-
ample based on a new battery):
When the unit is connected to the mains or an external battery, the unit’s internal battery recharges. A
special pictogram appears on the screen to show the progress of light segments (example based on a
new battery):
Battery discharged
Battery half-charged
106
12 Technical description
In the event of operation at a high ambient temperature, a safety device might limit the battery charge.
This can have a significant effect on the charging time and operating duration of the battery.
Note: The battery takes about 10 hours to charge (at ambient 25°C).
When the unit is switched off but connected to the mains or an external battery, the internal battery au-
tomatically recharges.
If the unit cannot recharge the battery because the battery is absent or defective, the following pic-
togram appears on screen:
Battery unavailable
Associated with a high-priority alarm ‘Internal battery unavailable!!! Check
battery’
The Monnal T75 medical electrical equipment may be connected to a separate external power supply
to create a medical electrical system.
Two icons indicates the absence or presence of an external source.
In case of a power-supply disconnection, the changeover to the external battery is indicated with a low-
priority acknowledgeable alarm “Ventilator operates from external battery”.
107
Monnal T75 user manual
3 2
The RS232 link of this ventilator allows data input ((1), software updates) and output ((2), sending of
machine data).
2
Software updates
The ventilator software is updated via the RS232 link using standard IT tools.
Technical features:
• Speed: 115200 bps
• Parity: none
• Format: 8 bits; 1 start bit; 1 stop bit
• Flow check: No
Note: With regard to the RS232 link, the use of an accessory that is not compliant with the
safety requirements of this ventilator can bring about a reduction in the safety level of the re-
sulting system.
The following factors should be considered when choosing the accessory:
• Its use near the patient
• Certainty that its safety certification was carried out in compliance with the local standards in
force and/or the EN 60601-1 standard as well as IEC 60950-1
• Do not apply any abnormal voltage to the RS232 connector (15 V max.).
This ventilator incorporates a communication protocol for sending data to data-collection software or
external monitoring systems.
108
12 Technical description
Choice of protocol
From the “Admin Config” menu, select the protocol which corresponds to the external device, using the
encoder wheel:
• OTP,
• Monnal Link,
• VueLink/Intellibridge EC10.
Contact the information system manager to select the appropriate protocol.
Connecting the communication module
Connect the cable supplied by the manufacturer of the communication module to RS232 connector
no. 1 on the Monnal T75 and then connect the other end of the cable to the external device.
Do not leave the cable connected to the Monnal T75 without connecting it to an external device, in or-
der to prevent electromagnetic interference.
Air Liquide Medical Systems protocols
The Monnal T75 includes the following Air Liquide Medical Systems communication protocols:
• OTP,
• Monnal Link.
These protocols are available upon request.
For more information, contact [email protected].
Information transmitted
The information transmitted is as follows:
• ventilation set-points,
• alarms,
• alarm thresholds,
• all measurements apart from time curves.
Compatible communication modules
• Capsuletech Datacaptor (OTP protocol),
• Philips monitors (VueLink/Intellibridge Protocol EC10),
• Mindray (OTP protocol).
To obtain a full list of the communication modules compatible with the Monnal T75, contact the manu-
facturer.
A hospital network failure can result in no data being transmitted (alarms and monitoring) to the exter-
nal device.
Connection to a hospital network which includes other apparatus could give rise to previously unidenti-
fied risks to patients, operators or third parties.
The user should identify, analyze, evaluate and control such risks. Subsequent modifications made to
the hospital network could introduce new risks and require additional analysis. Modifications to the
hospital network include:
• modifications to the configuration of the external device;
• the connection of additional units to the external device;
• the disconnection of units from the external device;
• the updating of devices connected to the external device;
109
Monnal T75 user manual
The video output is located on the back panel of the unit: a remote screen can be connected via this
output in order to display an image of the ventilator screen.
Directives
European Council Directive 93/42/CEE concerning medical devices.
Directive 2012/19/EU of the European Parliament and Council concerning Waste Electrical and Elec-
tronic Equipment (WEEE).
Standards
The compliance of Monnal T75 with the essential requirements of Directive 93/42 is based on the fol-
lowing standards:
EN ISO 14971 | Application of risk management to medical devices
EN ISO 60601-1 | Medical electrical equipment – Part 1: General requirements for basic safety and es-
sential performance
EN ISO 80601-2-12 | Medical electrical equipment – Part 2-12: Particular requirements for basic safety
and essential performance of critical-care ventilators.
EN ISO 80601-2-55 | Medical electrical equipment – Part 2-55: Particular requirements for the basic
safety and essential performance of respiratory gas monitors.
Device class: IIb
Service life of Monnal T75: 7 years.
Year of Monnal T75 (Ref. KB022600) CE marking: 2008
Year of Monnal T75 with CO2 option (Ref. KB033600) CE marking: 2013
110
12 Technical description
• Recycled paper
• Recyclable
Protective foam (1580 g):
• Polyethylene
• Recyclable
Method of recovery for consumable items
All used consumable items from use of this ventilator (the patient circuit, the bacteriological filters,
etc.) must be collected in the hospital’s appropriate disposal channels. The packaging waste for the
device and for its accessories must be disposed of in the hospital’s appropriate disposal channels.
Shipping equipment
Air Liquide Medical Systems recommends that the user keep the original packaging.
If the medical device is returned, always use the original packaging. Otherwise, contact your Air Liq-
uide Medical Systems representative to obtain replacement packaging.
Equipment modifications
CAUTION: It is prohibited to alter the Monnal T75 equipment.
Accessories
The accessories used with this ventilator must comply with the general requirements of European di-
rective 93/42/CEE, as well as EN 60601-1 and collateral standards.
Accessories from the Air Liquide Medical Systems catalog or included in the accessories kit delivered
with the ventilator comply with these requirements. The use of accessories not recommended by Air
Liquide Medical Systems releases Air Liquide Medical Systems from all liability in the event of an inci-
dent.
The user is responsible for ensuring that the use of accessories does not affect the safety and the ex-
pected performance of the ventilator.
The inspiratory and expiratory resistance measurements of this respiratory were taken using a stan-
dard adult patient circuit and a bacteriological filter (KV103300).
CAUTION: The addition of intermediate parts or other components or assemblies to the res-
piration system of the ventilator may change the pressure gradient of the system which may
have a negative impact on the performance of the ventilator.
The prescriber must ensure that the assembly thus obtained complies with IEC 80601-2-12.Items in-
cluded in the package.
Mass 16 kg
Operating conditions
111
Monnal T75 user manual
Protection
Protection index IP3X (if the serial number is less than MT75-05000)
IP31(if the serial number is greater than MT75-05000)
Applied parts The respiratory system of the Monnal T75 ventilator is a type B ap-
plied part.
The CO2 probe of the Monnal T75 is a type BF applied part
CAUTION: Ambient temperature warning: the temperature of the Monnal T75 delivery flow
can exceed that of the room. Exercise extreme caution when using the Monnal T75 at ambi-
ent temperatures above 38 C (100.4°F).
At an ambient temperature of 38 C or higher, gas temperature can climb as high as 43 C, de-
pending on the settings. The 43°C limit can never be exceeded. In such conditions, humidifi-
cation or nebulization may not be effective. When ambient temperature exceeds 38°C, Air Liq-
uide Medical Systems recommends use of HME filters.
Frequency 50 to 60 Hz
250 VA T 16 A
T 16 A 0,25 A FST
112
12 Technical description
Electrical class I
Type A
Protection following a loss of power Continuous audible alarm and patient vented to
atmosphere
Type NiMh
Rated voltage: 24 V
Rated capacity 2X4500 mAh
Isolation
Separation from the power supply To electrically separate the Monnal T75 from the
power supply, disconnect the flexible power ca-
ble.
113
Monnal T75 user manual
Note: The emission characteristics of this device allow it to be used in industrial areas and
in a hospital environment (class A as defined in CISPR 11). If used in a domestic environ-
ment (for which class B as defined in CISPR 11 is normally required), this equipment might
not offer adequate protection to radio-frequency communication services. The user may be
required to take mitigation measures, such as relocating or re-orienting the ventilator.
CAUTION: Exceptional electromagnetic disturbance may degrade ventilation performance.
114
12 Technical description
Electromagnetic emissions
The medical device is intended for use in an electromagnetic environment described in the table be-
low. The user and installer must ensure that the medical device is used in the environment described
below.
± 2 kV in common ± 2 kV in common
mode mode
115
Monnal T75 user manual
Voltage dips, short in- 0% UT for 0.5 cycles 0% UT for 0.5 cycles Professional health
terruptions and volt- care establishment en-
age variations (IEC A 0°, 45°, 90°, 135°, A 0°, 45°, 90°, 135°, vironment.
61000-4-11) 180°, 180°,
If use of the system re-
225°, 270° and 315° 225°, 270° and 315° quires continued oper-
ation during power sup-
0% UT for 1 cycle 0% UT for 1 cycle ply cuts, the use of a
separate power source
70% UT 70% UT to power the medical
for 25 cycles at 50 Hz for 25 cycles at 50 Hz device is recommend-
ed (UPS, etc.).
For 30 cycles at 60 Hz For 30 cycles at 60 Hz
Single phase: at 0° Single phase: at 0°
WARNING: RF hand-held communication devices should not be used (including peripherals such as
antenna cables and external cables) within a radius of 30 cm (12 inches) of any part of the Monnal
T75, including the cables specified by the manufacturer. Otherwise, the performance of this equip-
ment may be impaired.
116
12 Technical description
Conducted distur- 3V 3V
bances, induced by ra-
dio-frequency fields 150KHz to 80MHz 150KHz to 80MHz
(IEC 610004-6)
6 V ISM bandwidth and 6 V ISM bandwidth and
bandwidths between bandwidths between
0.15 MHZ and 80 MHZ, 0.15 MHZ and 80 MHZ,
including amateur ra- including amateur ra-
dio bandwidths dio bandwidths
80% MA to 1 KHz 80% MA to 1 KHz
Comment: These specifications may not apply in all situations. Electromagnetic propagation is affect-
ed by absorption and reflection from structures, objects and people.
(a) The electromagnetic field strengths of fixed radio-frequency transmitters, such as base stations for
mobile phones (cellular/wireless), mobile radios, amateur radio operators, AM/FM radio broadcasts
and TV broadcasts cannot be determined accurately through theory. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be carried out. If the
intensity level of the radio-frequency field in the immediate environment in which the product is being
117
Monnal T75 user manual
used exceeds the radio-frequency compliance level specified above, the product performance must be
tested in order to check whether or not it is compliant with the specifications. In the case of any abnor-
mal performance of the device, additional measures may be necessary, such as reorienting or relocat-
ing the product.
Required maximum flow rate (at atmospheric 105 L/min at 2.8 bar / 130 L/min at 6 bar (HP)
pressure)
85 L/min at 1.5 bar (LP)
Connectors
Pressure
(P lim max)
(P w max)
118
12 Technical description
Compatibility of accessories (parts of the respiration system which can be removed and assem-
bled by the operator)
Compliance 6 mL/cmH2O
Ventilation modes
119
Monnal T75 user manual
During any spontaneous cycle, the switch to expiration occurs as soon as one of the following crite-
ria is encountered:
- expiratory flow trigger (Trig. E); Expiration as soon as the inspiratory flow rate reaches x% of peak
inspiratory flow rate
- expiratory pressure trigger (not adjustable): expiration as soon as an overpressure of 3 cmH2O is
detected on the inspiratory
pressure signal
- expiratory time trigger (TI max.): expiration as soon as the insufflation time reaches the maximum
Ti setting (Timax).
Automatic Test
120
12 Technical description
Curves
121
Monnal T75 user manual
Curves
Monitoring
Data storage
Alarm log List of alarms triggered during the use of the unit
(4000 events recorded (1))
(1) The unit records 4000 events. An event is represented by a set-point or an alarm threshold when an
adjustment is made and an alarm if triggered. Because the alarm log reports only alarm-related events,
the maximum number of events recorded can be below 4000.
122
12 Technical description
123
Monnal T75 user manual
124
12 Technical description
125
Monnal T75 user manual
Invasive ventilation
TI/Ttot % 5 80 1 33 5 60 1 33 5 60 1 33
I/E - 1/19 1/0.3 1/0.1 1/2t 1/19 1/0.7 1/0.1 1/2 1/19 1/0.7 1/0.1 1/2
Tplat % 0 60 5 20 0 60 5 0 0 40 5 0
cmH2O OFF, 3 0,2 0,6 OFF, 3 0,2 0,6 OFF, 3 0,2 0,6
0,2* 0,2* 0,2*
PEEP cmH2O 0 50 1 4 0 50 1 4 0 50 1 4
Tins s 0.2 10.0 0.1 1.3 0.2 8.0 0.1 0.8 0.2 3.0 0.1 0.5
PS cmH2O 2 40 1 10 2 40 1 10 2 40 1 10
PI cmH2O 2 99 1 10 2 99 1 10 2 99 1 10
126
12 Technical description
Timax s 0.3 5 0.1 1.5 0.3 5.0 0.1 1.5 0.2 2.5 0.1 1
E. Trig % 5 90 5 30 0 90 5 30 0 90 5 30
T high s 0,3 30 0,1 1,5 0,2 30 0,1 1,5 0,2 30 0,1 0,5
T low s 0,1 30 0,1 0,6 0,1 30 0,1 0,5 0,1 30 0,1 0,3
P high cmH2O 2 99 1 15 2 99 1 15 2 99 1 15
P low cmH2O 0 50 1 0 0 50 1 0 0 50 1 0
Min. Max. Incr. Default Min. Max. Incr. Default Min. Max. Incr. Default
Low mL OFF 2900 10 OFF OFF 1900 10 OFF OFF 500 10 OFF
VTi
High mL 100 3000 10 3000 100 2000 10 1000 100 800 10 100
VTi
Low L/min 0.5 39.0 0.5 3.0 0.5 39.0 0.5 1.5 0.1 39.0 0.1 1
MVe
High L/min 1.5 30.0 0.5 25.0 1.5 40.0 0.5 10.0 1.0 40.0 0.1 5.0
MVe
FiO2 % 1,8 95 1 55 18 95 1 45 18 95 1 30
low
127
Monnal T75 user manual
Min. Max. Incr. Default Min. Max. Incr. Default Min. Max. Incr. Default
T ap- s 15 60 1 20 4 60 1 20 2 30 1 10
nea
etCO2 mmHg 1 99 1 49 1 99 1 49 1 99 1 49
haut
% 1 15 1 6 1 15 1 6 1 15 1 6
VM- mL/ OFF 990 10 OFF OFF 990 10 OFF OFF 990 10 OFF
CO2 min
high
VMalv L/min OFF 39 0,5 OFF OFF 39 0,5 OFF OFF 39 0,5 OFF
low
VCRP PI cmH2O 1 1 1
Insuf- AUTO AUTO AUTO AUTO AUTO AUTO AUTO AUTO AUTO
fla-
tion 2 99 15 2 99 15 2 99 15
pres-
PS AI
sure AUTO AUTO AUTO AUTO AUTO AUTO
PRO
2 40 2 40 2 40
Pres- Pente cmH2O/ AUTO AUTO 10 150 20 200 10 150 20 200 10 150
PRVC
sure s
In- PS
crease PRO
slope
128
12 Technical description
Expi- PS Trig. E % 5 90 5 30 5 90 5 30 0 90 5 30
ratory PRO
trig-
ger
PI PRVC PI cmH2O 2 99 1 20 2 99 1 20 2 99 1 20
max max
PS 2 99 2 99 2 99
PRO
Non-invasive ventilation
PS cmH2O 2 25 1 15 2 25 1 15 2 25 1 15
PEP cmH2O 0 15 1 5 0 15 1 5 0 15 1 5
Timax s 0.3 5.0 0.1 1,3 0.3 5.0 0.1 1.2 0.2 2.5 0.1 0.5
I. Trig L/min 1 15 1 5 1 15 1 5 1 15 1 5
E. Trig % 10 90 5 50 10 90 5 50 5 90 5 50
CPAP cmH2O 2 50 1 5 2 25 1 5 2 25 1 5
Pi cmH2O 2 99 1 15 2 99 1 15 2 99 1 15
T high s 0.3 30 0.1 1.5 0.2 30 0.1 1.5 0.2 30 0.1 0.5
T low s 0,1 30 0,1 0,6 0,1 30 0,1 0,5 0,1 30 0,1 0,3
P high cmH2O 2 99 1 15 2 99 1 15 2 99 1 15
129
Monnal T75 user manual
P low cmH2O 0 50 1 0 0 50 1 0 0 50 1 0
Min Max Incr. Default Min Max Incr. Default Min Max Incr. Default
Low L/min OFF, 59.0 0.1 2.0 OFF, 59.0 0.1 1.5 OFF, 39 0.1 1
MVe 0.5 0.1 0.1
High L/min 1.0 60.0 0.1 30.0 1.0 60.0 0.1 10.0 0,9 40 0.1 5
MVe
FiO2 % 18 95 1 55 18 95 1 45 18 95 1 30*
low
T ap- s 15 60 1 20 4 60 1 20 2 30 1 10
nea
etCO2 mmHg 1 99 1 49 1 99 1 49 1 99 1 49
high
% 1 15 1 6 1 15 1 6 1 15 1 6
VM- mL/ OFF 990 10 OFF OFF 990 10 OFF OFF 990 10 OFF
CO2 min
low
VMalv L/min OFF 39 0,5 OFF OFF 39 0,5 OFF OFF 39 0,5 OFF
low
130
12 Technical description
Suction settings
Nebulization settings
Flow L/min 1.0 20.0 0.5 6.0 Inactive function Inactive function
min 1 60 1 10
Nebu-
lization
dura-
tion
Fre- h 1 24 / 1 none
quency none
Min Max Incr. Default Min Max Incr. Default Min Max Incr. Default
Flow L/min 2 80 1 60 2 60 1 25 2 60 1 15
rate
131
Monnal T75 user manual
VCV VPC SIMV PSIMV PSV PSV CPAP Duo PRVC PS PRO
NIV APRV Levels
(20%, 5)
132
12 Technical description
Low exhaled volume alarm Temporarily change the threshold to activate the
corresponding alarm.
133
Monnal T75 user manual
13 Bibliography
• Martin J. Tobin, Principles and Practice of Mechanical Ventilation. Second edition, McGraw-Hill
Inc., 2006.
• Minaret G. ; Richard J.-C. ; Ventilation mécanique : modes de référence. ITBM-RBM Volume 26, Is-
sue 1, January 2005, Pages 14-18.
• Chatburn RL. Classification of ventilator modes : update and proposal for implementation. Respir
Care. 2007 Mar;52(3):301-23.
• Esteban A, Anzueto A, Alía I, Gordo F, Apezteguía C, Pálizas F, Cide D, Goldwaser R, Soto L, Buge-
do G, Rodrigo C, Pimentel J, Raimondi G, Tobin MJ. How is mechanical ventilation employed
in the intensive care unit? An international utilization review. Am J Respir Crit Care Med. 2000
May;161(5):1450-8.
• Esteban A, Anzueto A, Frutos F, Alía I, Brochard L, Stewart TE, Benito S, Epstein
SK, Apezteguía C, Nightingale P, Arroliga AC, Tobin MJ; Mechanical Ventilation International Study
Group. Characteristics and outcomes in adult patients receiving mechanical ventilation : a 28-day
international study. JAMA. 2002 Jan 16;287(3):345-55.
• S. Jaber, M. Sebbane, D. Verzilli, P. Courouble, G. Chanques, J.J. Eledjam, Les nouveaux modes
ventilatoires : avancée ou gadget ? Journées Méditerranéennes d’Anesthésie Réanimation et Ur-
gences 2006.
• CoutantG., Réanimation-La ventilation artificielle. http://www.infirmiers.com/etudiants- en-if-
si/cours/cours-reanimation-la-ventilation-artificielle.html
• Laurent Brochard, Alain Mercat, Jean-Christophe M. Richard. Ventilation artificielle : De la physiolo-
gie à la pratique. Collection Manuel d’anesthésie, de réanimation et d’urgences. Elsevier Masson,
2008.
• J.Roeseler, J-B Michotte , A.F.De Ranter, S. Delaere, P.F.Laterre, M.Reynaert, Différentes techniques
de ventilation. Cliniques universitaires Saint-Luc Soins intensifs C. p. 66-110
• Fernandez R. Blanch L. Ventilation assistée contrôlée intermittente. In: L. Brochard and J. Mance-
bo, Editors, Ventilation artificielle principes et applications, Arnette Ed, Paris (1994), pp. 121–132.
• Corinne Taniguchi, Raquel C Eid, Cilene Saghabi, Rogério Souza, Eliezer Silva, Elias Knobel, Ângela
T Paes, and Carmen S Barbas. Automatic versus manual pressure support reduction in the wean-
ing of post-operative patients: a randomised controlled trial. Crit Care. 2009; 13(1): R6.
• Fourcade, Modes AI VT et AV APS en ventilation non invasive. 17ème journée JARCA 2008
• Storre JH, Seuthe B, Fiechter R, Milioglou S, Dreher M, Sorichter S, Windisch W. Average volume-as-
sured pressure support in obesity hypoventilation: A randomized crossover trial. Chest. 2006
Sep;130(3):815-21.
• Brochard L., Isabey D. La ventilation non invasive. In: L. Brochard and J. Mancebo, Editors, Ventila-
tion artificielle principes et applications, Arnette Ed, Paris (1994), pp. 241–259.
134
14 Appendix
14 Appendix
14.1 Checklist
The following actions must be applied when commissioning a unit after every maintenance operation:
Done
Install the patient circuit on the device and connect a test lung.
Connect the unit to the mains power supply and check that the indicator light on the
front panel lights up.
Apply power to the unit pressing the On/Off button at the back of the device.
You should hear a BEEP and the ventilator screen should light up.
Select [New Patient], “Adult” type, in the stand-by screen, and start ventilation.
Check there is no technical alarm after 1 minute of ventilation.
Power supply
During ventilation on a test lung, disconnect the electrical connection
Check that the ventilator ventilates on its internal battery and announces this fact on
screen.
Check the battery is charged enough (3 squares at least) if an intra-hospital trans-
port is programmed. Reconnect the AC power supply. (See also Testing the alarms on
page 133)
Seal off the Y-piece and run the automatic test sequence; the message “test success-
fully completed” appears, with a compliance value.
Select [New Patient] and the appropriate type of patient (ADULT or CHILD or INFANT)
from the stand-by screen.
135
Monnal T75 user manual
• Disassemble the patient circuit by removing its components: pipes, fittings, water traps, and Y-
piece.
• Disassemble the expiratory assembly according to the diagram below.
• Remove the expiratory assembly from its housing by pressing the eject pushbutton.
• Remove the expiratory flow hot wire sensor (1), the valve body (2), the membrane (3), and the sili-
cone disc (4). Do not remove the two seals from the valve body (2).
CAUTION: The flow sensor requires special precautions:
• Avoid inserting any object into the flow sensor,
• Avoid exposing it to a jet of water or air,
• Avoid impacts or dropping it.
Pre-disinfection / cleaning
Immerse the expiratory assembly components in a pre-disinfectant solution: Air Liquide Medical Sys-
tems recommends the use of ANIOS products: ANIOSYME DD1, SALVANIOS PH7, HEXANIOS G+R
(obey the instructions of the product manufacturer).
Rinse the parts under running water, except for the expiratory flow sensor (1), which should only be
dipped in water.
Leave the components to dry on absorbent paper.
Sterilisation: prion cycle 134°/18 min/ 1 bar above atmospheric pressure
CAUTION: Sterilisation must be performed with care by certified personnel.
Condition the components removed from the expiratory assembly before the procedure.
The expiratory assembly can tolerate 50 sterilisation cycles. A unique serial number on the expiratory
flow sensor (1) and the valve body (2) gives the date of manufacture of the components and can be
used to track the number of cycles they have undergone.
136
14 Appendix
Reassembly
Reassemble the expiratory assembly (wear sterile gloves to do this):
- install the silicone disc in the membrane,
- reposition the membrane in the valve body.
CAUTION: Before any further use of the expiratory assembly, check that it was correctly re-
assembled by connecting it to the ventilator and performing the automatic tests.
137
V3.6.xI9A
UM000103
www.device.airliquidehealthcare.com