Specifications ProSoundF75
Specifications ProSoundF75
SPECIFICATIONS
for
Diagnostic Ultrasound System
1
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
How a diagnostic ultrasound system should be for image-constructing engine, and have taken a step
examinations to be more natural and with less closer to the ideal. The ProSound F75 stably delivers
inconvenience - We thoroughly thought over its easily diagnosable images, without being dependant
user’s, that the ideal size was determined based on the The color Doppler, first developed by Hitachi Aloka
reality of examinations, and all necessary functions and Medical, has continued to evolve owing to various
features were fit in without compromise. innovating technologies. The dynamic display of blood
What can a diagnostic ultrasound system do for smooth flow is one of the specialties of Hitachi Aloka Medical’s
routine examinations? The answers are provided with the diagnostic ultrasound system.
ProSound F75. Its features include natural and intuitive ProSound F75 will sensitively display blood flow in minute
ease of use, requiring less operation to obtain the blood vessels, regardless of its velocity.
required result.
The ProSound F75 learns how the user operates it and
will remember it, automatically optimizing itself to fit the
user’s style. It is a reliable system and easy to get used
to, enhancing efficiency in routine examinations.
What the ProSound F75 aims for is a diagnostic
ultrasound system that enables you to concentrate on
examinations without having to mind operating it.
Investment Return:
A wide range of clinical applications and envi-
ronmental features guarantee your investment
2
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
2
RT-3D (4D) mode* Trapezoidal Scan
2 3
* : Option: SOP-F75-4 Intermittent trigger mode *
Frame rate
Max. More than 610 frames/sec*
* Depends on probe and various settings
4
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
Display Gray Scale: 256 levels *: SCR (Strip Chart Recorder) Reduction,
Scanning area: Gain*: B-gain ±30 dB
100% to 25%, continuously variable Dynamic Range*: 34 dB-96 dB
Zoom AGC—Suppression of brightness saturation: 16 steps
Write zoom (magnification of real-time image): (including relief processing)
5
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
– Display patterns:
Cine Memory
Velocity (derived from mean Doppler frequency
Cine search and loop display (in B mode):
shift), Velocity + variance, Variance
ECG time phase display is possible
– Max. velocity range:±0.6cm/sec to ±199cm/sec Capacity
– Reference frequency: (Probe dependent) B mode: Max. 61,450 frames.
1.88, 2.14, 2.5, 3.0, 3.33, 3.75, 4.00, 4.29, 5.0, M and D modes: Max. approx. 813 seconds.
6.0 and 7.5MHz High-speed mutual data transfer between Cine
– Pulse repetition frequency: Memory and hard disk is possible.
0.5 to 10.0 kHz Note: The number of storable images in a loop
– Gradation: depends on probe type, scanning angle and other
±127 levels for velocity (red and blue) conditions.
16 levels for variance (green)
– Color Polarity: Normal, Invert
– Smoothing: 8 selections
– Flow filter: 6 selections
– Frame correlation: 16 selections
– Wall Motion Reduction: 16 selections
– Average: 5 levels
– Color base line shift:
Up to double velocity (±127 steps)
– Color coding (editable by color coding editor)
Abdomen : 5 kinds
PV : 5 kinds
Cardiology : 5 kinds
Other : 5 kinds
PowerFlow
– Display patterns: PowerFlow, Directional
PowerFlow
6
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
Sonographer
Data Management * Conforms to DICOM 3.0 standard
1. Image data
1-1. Format 4. Data storage media
Multiple-frame (moving) image Main unit hard disk
Raw data
Usable space: Up to about 120 GB
Image data (DICOM, M-JPEG, AVI, WMV)
* About 52,000 still images are storable.
Single-frame (still) image
DICOM (Raw, RGB, JPEG) USB memory
Tiff, Bmp, JPEG CD-R
1-2. Image acquisition mode DVD-RAM
Real-time multi-frame image acquisition (Raw, Image) Network interface: 10 BASE/T or 100 BASE/TX,
Post ECG: Max. 10cardiac cycles (R-R) (automatically switched)
Pre ECG: Max. 10 cardiac cycles (R-R)
Post Time: Max. 90 seconds 5. DICOM network communication*1
Pre Time: Max. 90 seconds Conformity to DICOM service class:
Manual: Ultrasound image storage SCU
Raw data: Max. 150 seconds Ultrasound multi-image storage SCU
Image data: Max. 210 seconds Storage media FSC/FSR
Cine loop high-speed data transfer (Raw, Image) Print management SCU
It is possible to selectively store data of arbitrary Modality worklist management SCU
(For details, please refer to the DICOM Conformance
section in the Cine Memory.
Statement issued by Hitachi Aloka Medical, LTD.)
Simultaneous output to multiple media
Modality performed procedure step (MPPS) SCU
It is possible to output still image data to multiple of Storage: Possible to store patient information directly
storage media and printers at the touch of a button. to DICOM file server
1-3 Image data management tool Print: Possible to printout images with DICOM
Image viewer compatible printer directly
・Compatible with DICOM and PC-format images Work list management: Retrieval of patient and
Simultaneous display of stored and real-time reservation information from hospital information
images is possible (Compare mode) system (HIS)
・Thumbnail display of stored images (1-36 images) NOTE: The HIS needs to be compatible with DICOM standard
・Check mark is put on a transferred image supplement 10. The HIS network and the DICOM network
・Image zoom, rotation, inversion need to be linked.
・1:1 replay (main unit HDD or DICOM storage data) Router setting: possible
DVD-RAM Compatible with SR (Structured Report) for OB,
CD-R cardiology, vascular and abdominal measurements*2
USB memory Compatible with ED(Evidence Documents)
・Re-storing to media, transfer Profile
・Adjustment is possible on the reconstructed IHE (Integrated Healthcare Enterprise)
Raw-data image (gain, dynamic range, gamma SWF (Scheduled Work Flow)
curve type, and color coding in Color flow PIR (Patient Information Reconciliation)
mapping mode) *1 Option: SOP-F75-10
2. Measurement data *2 Option: SOP-F75-21
It is possible to store measurement data in the main 6. Security measures
unit hard disk User authentication function is available.
3 Types of user authority can be set.
3. Patient data
It is possible to set whether password is required
Displayed information*
or not at the start of operation.
Patient information
ID (up to 64 characters), Name (up to 64 Teaching file can be made.
characters, including middle name), Birthday,
Sex, Age, Height, Weight, Occupation
Study information
Procedure ID, Accession, Study ID, Study
Description Referring physician, Reporting Phys,
7
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
8
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
brachytherapy.
Abdominal measurements & calculations * Available when UST-678 is connected.
B mode
Gall bladder Physiological Signal Display*1
Common bile duct Displayed information: ECG, PCG, Pulse wave*2,
Liver breathing waveform
Pancreas
ECG synchronized display: Available for one phase
Kidney
Display position: Continuously variable (both in B and
Spleen
SOL (Space Occupying Lesion) M modes)
Vessel diameter (aorta, portal vein) Bar graph display for breathing waveform
Stenotic rate (diameter, area) *1 Option: PEU-F75
Doppler mode *2 Pulse wave transducer (TY-307A) is optional.
Abdominal aortic flow
Renal flow
Dual Doppler
Portal vein flow
This is a function which displays Doppler Spectrums of
Shunt flow
B/D mode two different sample points simultaneously.
Flow volume Supported combinations are PW/PW, TDI/TDI, and
PW/TDI.
This function is beneficial for functional evaluations of
Report Functions
heart failures and phase analysis. For example, E/e'
– Obstetrical report
measurements can be performed in a single scan.
– Gynecological report
– Cardiac function report
Probe: UST-9147, UST-9130, UST-9115, UST-567, ASU-1010,
– Vascular report
– IMT (Intima-Media Thickness) report UST-52105, UST-52124, UST-5296
– Urological report
– Abdominal measurement report
– Small parts report
Linear CW
It is possible to detect stenotic flows and the like in the
superficial vessels with a wide field of view keeping a high
image quality with no need to change the probe into a
sector one.
10
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
* Option: SOP-F75-43
*1 Option: SOP-F75-44
Probe: UST-9147, UST-9130, UST-9133, UST-9135P,
UST-9146T, UST-9146I, UST-52105, UST-5415,
ASU-1010,UST-9132I,UST-9132T,
,GF-UE160-AL5,GF-UE260-AL5,GF-UCT180,GF-UCT260,
UST-567,UST-5418,UST-5419
*2 Option: SOP-F75-4
12
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
Time, strain
Optional Analysis Functions M-mode
Velocity trace
eTRACKING (Echo Tracking) * Velocity, time, acceleration, ratio, velocity
It is possible to precisely measure displacement of difference
TDI-Myocardial Thickness (wall thickness)
blood vessel to obtain indexes of stiffness of the vessels
Distance, time, velocity
such as pressure-strain elastic modulus (Ep), stiffness Velocity Profile
parameter (β), arterial compliance (AC), one-point Velocity, distance
pulse wave velocity (PWVβ), and augmentation index CSV output of analyzed data is possible. CSV is a file
(AI). format that can be converted into Excel file directly.
13
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
Characters and graphic displays ・ Height adjustment and swivel together with operation
15
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
System Configuration
ProSound F75 main unit including a viewing monitor
PM-F75-H005
DICOM communication
SOP-F75-10
Structured Report
SOP-F75-21
SOP-F75-13 SOP-F75-11
OPTIONAL PROBES
Electronic convex sector probes T.H.E: Tissue Harmonic Echo, BbH: Broadband Harmonics,
CE: Contrast Echo, SCI: Spatial Compound Imaging,
Ultrasound Frequencies (MHz) Scanning Radius of
Application
Model angle curvature Optional accessories
(description) B and M modes Doppler/Flow (degrees) (mmR)
Abdomen, UST-9147 3.00/3.75/5.00/6.00 Flow: 71 51 -
OB/GYN BbH: 2.14/2.50 2.14/2.50/3.00/3.75
CE: PW:
1.33/1.50/1.88/2.50 2.14/2.50/3.00/3.75
3.00/3.75/4.00 CE: 1.88
Abdomen, UST-9130 3.00/3.75/5.00/6.00 Flow: 70 60 Puncture adapter:
OB/GYN BbH: 2.14/2.50 2.14/2.50/3.00/3.75 MP-2473
CE: PW:
1.50/1.88/2.14/2.50 2.14/2.50/3.00/3.75
CE: 1.88
Abdomen, UST-9115-5 3.75/5.00/6.00/7.50 Flow: 60 60 Puncture adapter:
OB/GYN BbH: 3.00/3.33/3.75 3.00/3.75/5.00/6.00 MP-2473
PW:
3.00/3.75/5.00/6.00
Endo-cavity UST-9118 3.75/5.00/6.00/7.50 Flow: 180 9 Puncture adapter set
BbH: 4.00/5.00 3.75/4.29/5.00/6.00 MP-2748-SET
PW:
3.75/4.29/5.00/6.00
Endo-cavity UST-675P 3.75/5.00/6.00/7.50 Flow: 180 9 Probe cover:
BbH: 4.00/5.00 3.75/4.29/5.00/6.00 RB-945BP-S *1
RB-945BP-NS *2
PW:
(Puncture adapter
3.75/4.29/5.00/6.00 MP-2452 is included
as standard.)
Intraoperative UST-9120 5.00/6.00/7.50/10.0 Flow: 70 20 Waterproof connector
Small parts BbH : 3.75/5.00/6.00/7.50 cover: MP-2790
Neonatal head 5.00/5.00/7.50/7.50 PW :
3.75/5.00/6.00/7.50
*1 Sterilized (Note: Sterilized probe cover cannot be sold in EU member countries.) *2 Non-sterilized
17
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
18
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
19
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
Bi-plane probe
Application Ultrasound Frequencies (MHz) Scanning Radius of
curvature Optional
(descriptio Model angle
B and M Doppler & Flow (mmR) accessories
n) (degrees)
Tramsrectal UST-678 Convex 3.75/5.00/6.00/ Flow: 120 9 Puncture adapter:
7.50 3.00/3.75/5.00/6.00 MP-2451
PW:
3.0/3.75/5.00/6.00 Probe cover:
Linear BL-664-S
5.00/6.00/7.50/ Flow: 60mm
(sterilized)*
10.0 3.00/3.75/5.00/6.00 BL-664-NS(Non-st
—
PW: erile)
3.00/3.75/5.00/6.00
Tramsrectal 3.75/5.00/6.00/ Flow: 178 9 Probe cover:
UST-677P Convex 7.50 3.75/4.29/5.00/6.00 RB-665P-S
2
BbH: 4.00/5.00 PW: (sterilized)*
3.75/4.29/5.00/6.00 RB-665P-NS(Non-s
terile)
Convex 3.75/5.00/6.00/ Flow: 178 9 Puncture adapter
7.50 3.75/4.29/5.00/6.00 is included as
BbH: 4.00/5.00 PW: standard.
3.75/4.29/5.00/6.00
* Sterilized probe cover cannot be sold in EU member countries.
20
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
It is forbidden to use all probes in direct contact with the central nervous system in Europe.
21
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
22
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
23
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.
The standard components and optional items differ depending on the country.
Not all products are marketed in all countries.
Please contact your local Hitachi Aloka Medical distributors for details.
ProSound is a registered trademark or trademark of Hitachi Aloka Medical, Ltd. in Japan and
other countries.
Harmonic Echo is a registered trademark or trademark of Hitachi Aloka Medical, Ltd. in Japan
and other countries.
DICOM is the registered trademark of the National Electrical Manufacturers Association for its