FOR REFERENCE ONLY.

ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

SPECIFICATIONS
for
Diagnostic Ultrasound System

MODEL: PROSOUND F75

1
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

ProSound F75 True Diagnostics:


What is necessary for an efficient and accurate
ProSound F75 was developed for conducting comfortable diagnosis?
and efficient examinations. It is beneficial for a wide
range of examination fields, and will deliver images with The demand for diagnostic imaging is not just a pretty

powerful diagnostic ability. picture, but an imaging function with powerful


diagnostic ability.

Facilitate Workflow: To achieve this, we have combined all of our


Make routine examinations more comfortable cutting-edge technologies, from probes to the

How a diagnostic ultrasound system should be for image-constructing engine, and have taken a step

examinations to be more natural and with less closer to the ideal. The ProSound F75 stably delivers

inconvenience - We thoroughly thought over its easily diagnosable images, without being dependant

conditions before developing. It is for the benefit of our on the patient.

user’s, that the ideal size was determined based on the The color Doppler, first developed by Hitachi Aloka
reality of examinations, and all necessary functions and Medical, has continued to evolve owing to various
features were fit in without compromise. innovating technologies. The dynamic display of blood

What can a diagnostic ultrasound system do for smooth flow is one of the specialties of Hitachi Aloka Medical’s

routine examinations? The answers are provided with the diagnostic ultrasound system.

ProSound F75. Its features include natural and intuitive ProSound F75 will sensitively display blood flow in minute

ease of use, requiring less operation to obtain the blood vessels, regardless of its velocity.

required result.
The ProSound F75 learns how the user operates it and
will remember it, automatically optimizing itself to fit the
user’s style. It is a reliable system and easy to get used
to, enhancing efficiency in routine examinations.
What the ProSound F75 aims for is a diagnostic
ultrasound system that enables you to concentrate on
examinations without having to mind operating it.

Investment Return:
A wide range of clinical applications and envi-
ronmental features guarantee your investment

There are three factors that help to expand the


boundaries of clinical applications:
- A wide variety of probes, tailored for various target
organs
- Measurements and reports for each clinical
application, necessary for diagnosis
- A compact size to fit in various places

By combining these 3 at high dimensions and pursuing


versatility, this new diagnostic ultrasound system ensures
cost effectiveness.
In order to use it for a long period of time, its total cost of
ownership must be considered, and not just the initial
purchase cost.
Various ways to minimize these costs have been devised
in all 3 phases; introduction, during use, and after use.

2
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Scanning Method  B(eFLOW) and D simultaneous real-time display


 Electronic Convex (Triplex mode)
 Electronic Linear  B and B(Flow) simultaneous real-time display (DDD:
 Electronic Phased Array Sector Dual Dynamic Display)
 B and B(PowerFlow) simultaneous real-time display
Operating Modes (DDD)
 B-mode  B and B(eFLOW) simultaneous real-time display
 M-mode (DDD)
 D: Spectral Doppler mode (PW, CW, HPRF-PW)  B(Flow), M(Flow), and D
 Dual Doppler mode  Dynamic Slow-motion Display
 Flow mode (Real-time image/Slow-motion image, side by side
 Power Flow mode (Directional Power Flow) display)
 eFLOW mode (Directional eFLOW)  EFV: Extended Field of View *
2

2
 RT-3D (4D) mode*  Trapezoidal Scan
2 3
* : Option: SOP-F75-4  Intermittent trigger mode *

Image Display Modes *1  Monitor mode *3 (Fundamental image/Intermittent CE


image, side by side display)
 B: gray-scale imaging
 TDI (Tissue Doppler Imaging)
 Dual B 4
 RT-3D (4D) mode *
 Quad B 5
 Real-time Tissue Elastography (RT-Elasto) *
 M
 B and M
Request function: In multi-mode display, it is possible to
 D: Spectral Doppler (PW, HPRF PW, and CW)
select one mode for full screen display.
 B and D
*1 Probe dependent.
 B, M, and D
*2 Option: SOP-F75-1
 B(Flow)
*3 Option: SOP-F75-44
 B(PowerFlow)
*4 Option: SOP-F75-4
 B(eFLOW)
*5 Option: SOP-F75-43
 Dual B(Flow)
 Quad B(Flow)
Beam former
 Dual B(PowerFlow)
Full-element transmission and reception
 Quad B(PowerFlow)
Full Aperture Apodization (FAA)
 Dual B( eFLOW)
Beam signal processing
 Quad B( eFLOW)
Dynamic 4D apodization
 M (Flow)
Transmission
 M (PowerFlow)
CPWG+ (Compound Pulse Wave Generator Plus)
 M (eFLOW)
Programmable waveform transmission
 B(Flow) and M (Flow)
Reception
 B(PowerFlow) and M(PowerFlow)
Multi processing high-speed digital beam former
 B(eFLOW) and M (eFLOW)
12-bit A/D converter (4096 gray levels)
 B(Flow) and D
Sampling Frequency: equivalent to 320 Hz
 B(Flow) and Dual
Parallel processing: Up to 4 directions
 B(PowerFlow) and D
Delay precision: 1/64λ at minimum in both
 B(PowerFlow) and Dual
transmission and reception
 B(eFLOW) and D
Transmission/Reception method of Harmonics
 B(eFLOW) and Dual
Broadband Harmonics (BbH) by Compound Impulse
 B(Flow) and D simultaneous real-time display (Triplex
Waveform Transmission and other technologies (Probe
mode)
dependent)
 B(PowerFlow) and D simultaneous real-time display
Reception frequency: 3.6-18.0 MHz*1
(Triplex mode)
3
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

*1 18MHz can be achieved under below conditions.


Probe: UST-5417
Image Frequency ”High”
Frequency Information ”Recieve”

Transmission/Reception method of pulse modulation


– Chirp Pulse modulation
Compound Broadband Pulse transmission
Tissue Adaptive Technology
Sound velocity adjustment: 26 steps
Automated sound velocity adjustment
Focusing
Lateral direction
Transmission: Multi-stage transmission focus of up
to 4 stages out of 16 stages
Reception: PixelFocusTM
Broad Focusing (CE mode)
Elevation direction
Compound Dual Focusing (when Compound Array
probe is used)

Frame rate
Max. More than 610 frames/sec*
* Depends on probe and various settings

4
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

B-mode 300, 200, 133, 100, 66.7, 50, 40, 25 mm/sec

 Display Gray Scale: 256 levels *: SCR (Strip Chart Recorder) Reduction,
 Scanning area:  Gain*: B-gain ±30 dB
100% to 25%, continuously variable  Dynamic Range*: 34 dB-96 dB
 Zoom  AGC—Suppression of brightness saturation: 16 steps
Write zoom (magnification of real-time image): (including relief processing)

Up to Maximum range of each probe  Relief: 4 steps


(Probe dependent)  FTC: On/Off
The maximum magnification is 7.5.  FAM (Free Angular M-mode) *
Read zoom (magnification of frozen image):  Color FAM
Max. Up to 0.5 cm display image *: Up to 3 M-mode cursors can be set omni-

The maximum magnification is 60. directionally in real time at any position on a


 Depth range selections: B-mode image.
0.5/0.75/1.0/1.5/2.0/2.5/3.0/3.5/4.0/4.5/5.0, and M-mode images can be reconstructed from stored
1 cm intervals from 5.0 to 30 cm (probe dependent) Raw-data B-mode images.
 Longitudinal and lateral inversion Gain, Dynamic Range and sweep speed can be changed
 Rotation by 90 degrees (probe dependent), after freezing
 Frame rate: Up to 9 steps
 Gain*: 10 to 90 dB Spectral Doppler:
 STC (sensitivity time control) :  Display: Power spectrum
Gain versus depth curve control: 8 slide controls  Real-time Doppler Auto Trace
STC curve memory function  Doppler methods:
 Angle gain:
– PW (Pulsed Wave) Doppler
Gain versus angle curve control: 8 sectors (probe
– HPRF (High Pulse Repetition Frequency)
dependent)
PW Doppler
 Dynamic range: 34-96 dB
– Dual Doppler
 AGC—Suppression of brightness saturation and Edge
– CW (Continuous Wave) Doppler
Enhancement: 16 steps
 Reference frequencies (probe dependent):
 Relief: 4 steps
PW: 1.5, 1.88, 2.14, 2.5, 3.0, 3.33, 3.75,
 FTC: On/Off
4.0, 4.29, 5.0 and 6.0 MHz
 Frame correlation: 16 steps (Auto/Manual)
CW: 1.88, 2.0, 3.0, 3.75, 5.0 MHz
 Smoothing: 16 steps
 Analysis rate:
 Post Processing
PW: 0.06 to 20 kHz
Echo enhance curve: 5 kinds
CW: 1.3 to 41.67 kHz
Rejection: 64 steps
 Max. velocity range:
 View gamma: Linear + 16 kinds
PW/HPRF:±1.2 cm/sec to ±400 cm/sec
 Edge Optimizer: 8 steps
CW:±24.9 cm/sec to ±796.2 cm/sec
 Image Optimizer
 Base line shift: up to double velocity (changeable
Automated sound velocity adjustment included
 Spatial Compound Imaging after freezing)
(by selected linear and convex sector probes)  Steerable CW Doppler: probe dependent
Compound scan angle: 5-20 degrees  Steered linear scanning: Max. ±30 degrees
(probe dependent), 1 degree increment changeable at 5 degrees interval
 Trapezoidal scan (possible by selected linear probes)  Spectrum inversion
 Adaptive Image Processing (AIP) / HI REZ  Angle correction: Available up to 80 degrees
* Gain and Dynamic range can be changed after freezing (Pre-settable, changeable after freezing)
Auto-angle correction, Warning function available
M-mode:  Sample volume size for PW Doppler:
 Sweep method: Moving bar 0.5 – 20 mm, changeable in 0.5mm, 1.0 mm,
Sweep speed*: 2.0mm step

5
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

 Wall motion filter: – Gradation: 128 levels


Manual: 50, 100, 200, 400, 800 or 1600 Hz, – Color coding
Auto: PRF adaptive (12 level) Power: 5kinds
 Doppler gain: 0-50 dB (128 step) Other: 5 kinds
 Contrast: 16 steps (changeable after freezing) – Non-display of B/W image: Possible
 Black-and-white inversion: changeable after freezing – Smoothing: 8 levels
 Image Optimizer  eFLOW
(Optimization of velocity range and baseline shift ) One of the Color Flow imaging functions that can
 Adaptive CWD display blood flow information in a high spatial and
 Audio output: 2 channel temporal resolution.
– Display patterns:
Color Flow Mapping eFLOW, Directional eFLOW
 Color area size: Variable from 100% to 15% – Gradation: 128 levels
continuously – Color coding
 Steered linear scanning: eFLOW : 5 kinds
Max. ±30 degrees *, changeable at 5 degrees interval Other : 5 kinds
* Probe dependent – Non-display of B/W image: Possible
 Frame rate: Up to 9 steps – Advanced wall motion reduction
– Smoothing: 8 levels
 Velocity  TDI (Tissue Doppler Imaging)

– Display patterns:
Cine Memory
Velocity (derived from mean Doppler frequency
 Cine search and loop display (in B mode):
shift), Velocity + variance, Variance
ECG time phase display is possible
– Max. velocity range:±0.6cm/sec to ±199cm/sec  Capacity
– Reference frequency: (Probe dependent) B mode: Max. 61,450 frames.
1.88, 2.14, 2.5, 3.0, 3.33, 3.75, 4.00, 4.29, 5.0, M and D modes: Max. approx. 813 seconds.
6.0 and 7.5MHz  High-speed mutual data transfer between Cine
– Pulse repetition frequency: Memory and hard disk is possible.
0.5 to 10.0 kHz Note: The number of storable images in a loop
– Gradation: depends on probe type, scanning angle and other
±127 levels for velocity (red and blue) conditions.
16 levels for variance (green)
– Color Polarity: Normal, Invert
– Smoothing: 8 selections
– Flow filter: 6 selections
– Frame correlation: 16 selections
– Wall Motion Reduction: 16 selections
– Average: 5 levels
– Color base line shift:
Up to double velocity (±127 steps)
– Color coding (editable by color coding editor)
Abdomen : 5 kinds
PV : 5 kinds
Cardiology : 5 kinds
Other : 5 kinds
 PowerFlow
– Display patterns: PowerFlow, Directional
PowerFlow

6
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Sonographer
Data Management * Conforms to DICOM 3.0 standard
1. Image data
1-1. Format 4. Data storage media
Multiple-frame (moving) image  Main unit hard disk
Raw data
Usable space: Up to about 120 GB
Image data (DICOM, M-JPEG, AVI, WMV)
* About 52,000 still images are storable.
Single-frame (still) image
DICOM (Raw, RGB, JPEG)  USB memory
Tiff, Bmp, JPEG  CD-R
1-2. Image acquisition mode  DVD-RAM
 Real-time multi-frame image acquisition (Raw, Image)  Network interface: 10 BASE/T or 100 BASE/TX,
Post ECG: Max. 10cardiac cycles (R-R) (automatically switched)
Pre ECG: Max. 10 cardiac cycles (R-R)
Post Time: Max. 90 seconds 5. DICOM network communication*1
Pre Time: Max. 90 seconds  Conformity to DICOM service class:
Manual: Ultrasound image storage SCU
Raw data: Max. 150 seconds Ultrasound multi-image storage SCU
Image data: Max. 210 seconds Storage media FSC/FSR
 Cine loop high-speed data transfer (Raw, Image) Print management SCU
It is possible to selectively store data of arbitrary Modality worklist management SCU
(For details, please refer to the DICOM Conformance
section in the Cine Memory.
Statement issued by Hitachi Aloka Medical, LTD.)
 Simultaneous output to multiple media
Modality performed procedure step (MPPS) SCU
It is possible to output still image data to multiple of  Storage: Possible to store patient information directly
storage media and printers at the touch of a button. to DICOM file server
1-3 Image data management tool  Print: Possible to printout images with DICOM
Image viewer compatible printer directly
・Compatible with DICOM and PC-format images  Work list management: Retrieval of patient and
 Simultaneous display of stored and real-time reservation information from hospital information
images is possible (Compare mode) system (HIS)
・Thumbnail display of stored images (1-36 images) NOTE: The HIS needs to be compatible with DICOM standard
・Check mark is put on a transferred image supplement 10. The HIS network and the DICOM network
・Image zoom, rotation, inversion need to be linked.
・1:1 replay (main unit HDD or DICOM storage data)  Router setting: possible
 DVD-RAM  Compatible with SR (Structured Report) for OB,
 CD-R cardiology, vascular and abdominal measurements*2
 USB memory  Compatible with ED(Evidence Documents)
・Re-storing to media, transfer  Profile
・Adjustment is possible on the reconstructed  IHE (Integrated Healthcare Enterprise)
Raw-data image (gain, dynamic range, gamma SWF (Scheduled Work Flow)
curve type, and color coding in Color flow PIR (Patient Information Reconciliation)
mapping mode) *1 Option: SOP-F75-10
2. Measurement data *2 Option: SOP-F75-21
It is possible to store measurement data in the main 6. Security measures
unit hard disk User authentication function is available.
3 Types of user authority can be set.
3. Patient data
It is possible to set whether password is required
Displayed information*
or not at the start of operation.
Patient information
ID (up to 64 characters), Name (up to 64 Teaching file can be made.
characters, including middle name), Birthday,
Sex, Age, Height, Weight, Occupation
Study information
Procedure ID, Accession, Study ID, Study
Description Referring physician, Reporting Phys,

7
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Measurements and Analysis:


 Basic measurements
On B-mode image
Distance
Area and Circumference (Trace, Ellipse, Circle)
Volume (Spheroidal, Prolate, Area-length, BP Simpson,
SP Simpson)Automatic heart cavity trace is possible.
(3-point designation method)
Index (general purpose)
Histogram
Angle, Hip joint angle
On M-mode image
Velocity
Distance (amplitude)
Time interval
Heart rate
Index (general purpose)
On spectral Doppler
Velocity, Acceleration,
Mean flow velocity, Pressure gradient
RI: Resistance index,
PI: Pulsatility index
Pressure half time
Heart rate
Dop Caliper measurements
Index (general purpose)
Time interval
Stenotic flow measurements
Regurgitant flow measurements
D. Trace
Real-time Doppler auto trace
On B/D mode
Flow Volume
SV/CO
On B(Flow) mode
Flow Profile*
* Option: SOP-F75-7

8
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

CCA (common carotid artery)


Application measurements ICA (internal carotid artery)
ECA (external carotid artery)
 Obstetrical measurements & calculations BIFUR (Bifurcation of carotid artery)
Gestational age, Fetal weight VERT (Vertebral artery)
% Stenosis area
Fetal Doppler measurements
% Stenosis diameter
Fetal cardiac function measurements
IMT (Intima-media thickness) Measurements
AFI (Amniotic fluid index)
Automated IMT measurements*
Cervical length
* Option: SOP-F75-38
Supports multiple gestations
Measurements of arteries in extremities:
Growth analysis function (display of past measurement
Lower extremity artery flow
data)
Upper extremity artery flow
Automated NT Measurement*
Stenotic rate:
* Option: SOP-F75-42
% Stenosis area
 Gynecological measurements & calculations
% Stenosis diameter
Uterus measurements Transit time of Vessel Flow measurements(TVF)
Endometrial thickness measurements Measurements of veins in extremities:
Cervical measurements Lower extremity venous flow
Ovary measurements Upper extremity venous flow
Follicular measurements (Volume measurements by Trans-cranial blood flow measurements
3-axis measurements are possible.)
Urinary bladder measurements  Urological measurements & calculations
Uterine artery, Ovarian artery measurements Prostate volume: PSA volume, PRS Slice volume
 Cardiac analysis Bladder volume
B mode Seminal vesicle
LV Volume measurements Testicle volume
Area-length, BP-ellipse, Simpson (Disc), Modified Renal volume
Simpson*, Bullet, Pombo, Teichholz, Gibson Renal cortical thickness
*: Automatic heart cavity trace is possible. (3-point) Renal pelvis measurement
Valve area measurements (AVA, MVA) Renal artery Doppler measurements (pulsatility index,
LA/AO resistance index)
Ratio
Right ventricle measurements  Small parts measurements & calculations
LV myocardial mass B mode
IVC (inferior vena cava) measurements Lesion (Breast)
M mode D/W ratio
Pombo (wall), Teichholz (wall), Gibson (wall) NT distance
Mitral valve measurements Thyroid volume
LA/AO measurements Isthmus Thickness
Tricuspid valve measurements Doppler mode
Pulmonary valve measurements Breast Doppler flow
IVC (inferior vena cava) measurements Thyroid Doppler flow
Doppler mode B(Flow) mode
LVOT (left ventricle outflow tract) flow Flow Histogram
RVOT (right ventricle outflow tract) flow
Trans-mitral flow
Regurgitant flow (AR, PR, MR, TR)
Stenotic flow (AS, PS, MS, TS)
Pulmonary vein flow
Coronary flow
TDI PW
B(Flow)/D mode
PISA measurements
B TDI mode
Asynchrony analysis for CRT
 Vascular analysis
Carotid artery:
9
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

brachytherapy.
 Abdominal measurements & calculations * Available when UST-678 is connected.
B mode
Gall bladder Physiological Signal Display*1
Common bile duct  Displayed information: ECG, PCG, Pulse wave*2,
Liver breathing waveform
Pancreas
 ECG synchronized display: Available for one phase
Kidney
 Display position: Continuously variable (both in B and
Spleen
SOL (Space Occupying Lesion) M modes)
Vessel diameter (aorta, portal vein)  Bar graph display for breathing waveform
Stenotic rate (diameter, area) *1 Option: PEU-F75
Doppler mode *2 Pulse wave transducer (TY-307A) is optional.
Abdominal aortic flow
Renal flow
Dual Doppler
Portal vein flow
This is a function which displays Doppler Spectrums of
Shunt flow
B/D mode two different sample points simultaneously.
Flow volume Supported combinations are PW/PW, TDI/TDI, and
PW/TDI.
This function is beneficial for functional evaluations of
Report Functions
heart failures and phase analysis. For example, E/e'
– Obstetrical report
measurements can be performed in a single scan.
– Gynecological report
– Cardiac function report
Probe: UST-9147, UST-9130, UST-9115, UST-567, ASU-1010,
– Vascular report
– IMT (Intima-Media Thickness) report UST-52105, UST-52124, UST-5296
– Urological report
– Abdominal measurement report
– Small parts report

It is possible to recall past measurement reports.


Examination data history can be plotted on the report.
Direct printout of each report is possible with an optional
PC printer.
Output of measurement values in CSV file is possible.
 Hot Key function; It is possible to assign measuring
functions to the alphabet keys on the keyboard
 Measurements on VCR playback image: Possible
(manual calibration)
 User’s calculation
30 equations can be set for each application
User-assignable terms: 60 words possible
 Font size of measurement result: Possible to change in
3 kinds

Linear CW
It is possible to detect stenotic flows and the like in the
superficial vessels with a wide field of view keeping a high
image quality with no need to change the probe into a
sector one.

Brachytherapy grid display*


It is possible to display grid for prostate gland

10
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

* Option: SOP-F75-43

Optional Functions Probe: UST-5415, UST-5411, UST-567, UST-5417, UST-5419,


UST-9132T,UST-5418
PC printer
It is possible to printout OB/GYN, cardiology, PV, small Automated IMT Measurement*
parts and urology report screens including ultrasound It is possible to automatically extract max IMT, min IMT
and mean IMT by simply setting ROI (region of interest)
images directly with an external PC printer.
on a long-axis view of the vessel. In addition, the
thicknesses at 3 points, i.e. , the point at max IMT, and
RT-3D (4D)*1
the points at 1cm on the right and left of the max IMT,
Scanning rate: Up to 30 volumes/sec
can be automatically detected and averaged.
 It is possible to display 3 arbitrary sections simultaneously * Option: SOP-F75-38
 MPR (Multi-planar Reconstruction)
Automated NT Measurement*
 360 degrees omni-directional rotation
It is possible to automatically extract the nuchal
 5 kinds of rendering selectable
translucency thickness (NT) by simply setting ROI (region
 Combination of 2 kinds of rendering type of interest) . In addition, thicknesses at max NT, Mean NT
(Rendering Mode Mix) can be automatically calculated.
 Detail scan of the ROI (Region of interest) is possible * Option: SOP-F75-42
 Inversion Mode (black-and-white)
EFV (Extended Field of View)*
 B-mode measurements on an arbitrary plane possible
It is possible to display an image of an extensive range of
 Automated Volume Measurement (AVM) *2 on a still
the body by moving the probe. An area wider than the
image
scanning width of the probe can be displayed.
 Traced Volume Measurement (TVM) *2 on a still image
EFV with flow, power flow, and eFLOW is possible.
 Spatio-temporal Image Correlation (STIC)*3
* Option: SOP-F75-1
 Flow 3D*4
・ Multi Slice Imaging (MSI)
・ Adaptive Image Processing (AIP) / HI REZ
 Threshold Low
 Free Axis of MPR (FMPR)
 Freehand 3D*5
*1 Option:SOP-F75-4
*2 Option: SOP-F75-20B
*3 Option: SOP-F75-41
Probe: ASU-1010
4
* Option: SOP-F75-35
Probe: ASU-1010, ASU-1013, ASU-1012
*5 Option: SOP-F75-37
Probe: UST-9130, UST-9147, UST-5415, UST-567, UST-5411,
UST-9133, UST-9146I/T, UST-9120, UST-9115-5,
UST-9135P, UST-9132I/T, UST-5713T, UST-5417

Real-time Tissue Elastography*


This function is used to visualize the stiffness of a tissue in
real-time. The strain generated in a tissue on applying
pressure is represented by colors
(it is available to change Elastography Color Map)
• Strain Ratio measurement:
Calculation of ratio of strain between arbitrary 2
regions
• Strain Graph:
The time variation of distorted average value is
displayed on a graph in real time.
• Auto Select Frame:
The frame under stable pressure is chosen
automatically.
11
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

CE (Contrast Echo)*1 as Quick Setter, image freezing and various


Contrast agent generates abundant second harmonics measurements. * Option: RM-700
when disrupted, which eases detection by Harmonic Echo.
In Subtraction mode, difference from the reference image
is displayed to clarify the distribution of the contrast agent.
Setting for low-pressure type contrast agent is also
incorporated.
 Monitor mode
In the Monitor mode, images are available with a low
sound pressure during the intermission of high sound
pressure transmission.
Monitor Image
Display of low sound pressure imaging.
 3D Contrast Imaging *2 (Probe: ASU-1010)
 Color Agent Display
In display using low-pressure type contrast agent, it is
possible to superimpose contrast echo information as
color image onto tissue image.
 Capture Mode
It is possible to display running of fine blood vessels by
accumulating contrast echo information.
 Motion artifacts such as from patient breathing or
unsteady probe positioning can be minimized for
acquiring clearer capture mode images. (MC Level:
Motion Capture Level)
 Tissue Reduction Mode
It is possible to suppress the signals received from
tissues displaying contrast images more clearly.
 Possible with UST-9147, UST-9133, UST-9146I,
UST-9146T,UST-5415,UST-9132I,UST-9132T,GF-UE16
0-AL5,GF-UE260-AL5,GF-UCT180,GF-UCT260,UST-567,
UST-5418,UST-5419

*1 Option: SOP-F75-44
Probe: UST-9147, UST-9130, UST-9133, UST-9135P,
UST-9146T, UST-9146I, UST-52105, UST-5415,
ASU-1010,UST-9132I,UST-9132T,
,GF-UE160-AL5,GF-UE260-AL5,GF-UCT180,GF-UCT260,
UST-567,UST-5418,UST-5419
*2 Option: SOP-F75-4

Wireless Remote Controller*


- Compact and light enough to fit easily in your breast
pocket (40 mm x 90 mm).
- As the menu for the remote controller is displayed on
the main screen, the controller can be manipulated
while viewing images.
- The controller can control many functions including
display mode switching, image adjustments as well

12
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Time, strain
Optional Analysis Functions M-mode
Velocity trace
eTRACKING (Echo Tracking) * Velocity, time, acceleration, ratio, velocity
It is possible to precisely measure displacement of difference
TDI-Myocardial Thickness (wall thickness)
blood vessel to obtain indexes of stiffness of the vessels
Distance, time, velocity
such as pressure-strain elastic modulus (Ep), stiffness Velocity Profile
parameter (β), arterial compliance (AC), one-point Velocity, distance
pulse wave velocity (PWVβ), and augmentation index CSV output of analyzed data is possible. CSV is a file
(AI). format that can be converted into Excel file directly.

* Option: SOP-F75-11 (Physiological Signal Display unit * Option: SOP-F75-13

PEU-F75 is also necessary.)


FMD (Flow Mediated Dilatation) analysis* Contrast Echo analysis*
It is possible to continually record and plot the vessel  Image Subtraction
Fixed Reference: Subtraction of reference frame from
diameter of the whole processes from baseline
all frames
through occlusion and vasodilatation to recovery.
Any 2 Frame: Subtraction between 2 selected frames
* Option: SOP-F75-16 (Physiological Signal Display unit Display modes: All images, arbitrary images
PEU-F75 is also necessary.) The SOP-F75-16 includes
eTRACKING software.  Time-Intensity Curve display
Optional Arm holder MP-AH0001, MP-AH0001-1 and - Intensity level graphing of average, median,
max and mode in the ROI.
Probe Fixing unit MP-PH0001 for FMD examination are
- Standard deviation (SD) graphing of intensity
available.
in the ROI .
WI (Wave Intensity)*
- Capture imaging reconstruction.
The heart and the arterial system are acting, - Color overlay function
constantly interfering with each other through forward - Inflow Time Mapping
traveling waves and reflected waves. WI is calculated  Color Mapping
based on changes in blood pressure and blood flow  Color Intermittent
 Graduation Map
speed obtained at an arbitrary point in a circulatory
 Color Overlay
system. WI is a new indicator of blood flow dynamics,
 Filtering
which is expected to help pave the way for analysis of  Blending
the interference between the heart and the arterial - Highlight display
system. The analysis comprises of contraction and - Measurement
dilatation characteristics, influence of reflected waves  Point to point
from peripherals, and an index related to time.  SD
 Raising time
* Option: SOP-F75-25 (Physiological Signal Display unit PEU-F75
 Fitting curve*
is also necessary.) SOP-F75-25 includes eTRACKING software.
 Distance
Blood-pressure gauge is necessary for this examination.
 Inflow Time Mapping Measurement
TDI analysis*
*: Parameters A, B and Beta can be calculated.
Tissue Doppler Imaging (TDI) analysis is an
Series: Graphic display in frame sequence or time
echocardiographic technique employing the Doppler
principle to measure the velocity of myocardial sequence
segments and other cardiac structures. Strain By Group: Graphic display with the time of one
information analysis is also available. sequence of intermittent acquisition as the horizontal
B-mode scale (Graphs of multiple sequences are overlapped.)
Temporal Velocity Profile Display mode: Image, Graph
Velocity, time, acceleration, ratio ROI type: Square, Draw, Arc, and Circle
Regional Velocity Profile CSV output of analyzed data is possible.
Velocity, distance
TDI-Myocardial Thickness (Wall thickness)
* Option: SOP-F75-14
Distance, time, velocity
Strain rate
Time, strain rate
Strain

13
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Stress Echo analysis* Twist (With Overlay function)


Image display modes in which image acquisition is - APEX: Apical long-axis view
possible: B, B (Flow)/TDI mode  Apex-S (Strain)
• Image acquisition methods: ・Longitudinal Strain and Strain Rate
ECG synchronized acquisition ・Transverse Strain and Strain Rate
• Compatible frame rate: Up to 75 Hz ・Global Strain and Strain Rate
• Recalled screen  Apex-V (Volume)
Playback speed: Selectable Volume for cavity
Image allocation: Possible
Single and Biplane EF
Scoring: Possible
(With Overlay Function)
Automatic registration: On/Off
• Protocol: Skip view function is available. • View
Exercise stress protocols: - SAX
- Exercise Stress Echo  Basal SAX
- Treadmill Exercise  Mid SAX
- Bicycle Exercise  Apical SAX
Pharmacological stress protocols: - APEX-S, -V
- DSE  2ch, 3ch, 4ch (Inversion view included)
- High-Dose DSE • Graph Type
- Low-Dose DSE
- Line Graph
- Arbutamine
- Color Graph
- Dipyridamole
User's protocol: - Line & Color Graph
The user can make a protocol within 8 views X - Bull’s eye (with Overlay function)
12 stages in 1 exam.
Full disclosure (Multi acquisition): 270 seconds • Measurements
• Scoring screen - Point to Point
Playback speed: Selectable
- Time to Peak
Comparison between different stages in the same
- 3 point: One-third point of diastole duration
view is possible
Image playback range is selectable - Distance
Systolic image acquisition * Option: SOP-F75-49
Bull’s eye display (16 or 17 segmentation selectable)
• Report screen
Display format
Chart/Stage overview/View overview
* Option: SOP-F75-15
2D Tissue Tracking (2DTT)*
 2DTT enables the evaluations of regional wall motion
abnormalities within your region of interest, such as the
myocardium and valves automatically by speckle tracking
method on a B-mode image. With only minimal angle
dependency, analysis form various cross sections have
become possible. Multiple analyses, including wall
thickness, various strains and rotation angle can be
performed.
• Analysis
- Free use
Distance and angular change between 2
points of your choice can be analyzed.
- SAX: Parasternal short-axis view
Circumferential Strain and Strain Rate
Radial Strain and Strain Rate
Angle and Angle Rate
14
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

General Specifications – Analog Video Input/Output*2


Output
Acoustic Power  Color composite (BNC): 1 channel
・ 0 to 100%, continuously adjustable  B/W composite (BNC): 1channel
 Y/C: 1 channels
Preset Function
Input
・ 45 kinds (selectable from 33 kinds of factory default
 Y/C: 1 channel
settings and user programmable settings)
*2Option: EU-9141
・ Preset contents storable in USB memory
– Others
・ Quick Setter
Audio (L/R): 1 channel
Image modifying parameters (e.g. Gain, frequency,
DVD control (RS-232C): 1 channel
depth) of your choice can be registered. (Up to 4 sets
per preset) Viewing Monitor
These parameters can immediately be registered and ・ 19-inch WSXGA(1600*1024 pixel) display
selected by touch panel during examinations. ・ Tilt and swivel are possible.

Characters and graphic displays ・ Height adjustment and swivel together with operation

 Character input area: panel

ID*, name, age, sex, retained text Safety Regulation


*: Can be corrected after exam ・ IEC 60601-1 Ed.3.0: 2005
 Input is possible with virtual keyboard on LCD panel Class I, Type BF
 Automatic Annotation Labeling:
Environmental Requirements
120 words or more (User registration is possible.)
In Operation
 Body mark:
・ Temperature: +10 to +40 degrees C
Default: 47 kinds
・ Relative Humidity: 30 to 75%
User definition: 24 kinds
(non condensing)
Body mark editor to create user’s body mark:
・ Atmospheric pressure: 700 to 1060 hPa
Available
・ Altitude: Up to 3000m
Probe mark: 4 kinds
In Storage/transportation
Display position: changeable
・ Temperature: -10 to +50 degrees C
Fetal body mark: rotatable
(0 to +50 degrees C for mechanical probes)
Menu control ・ Relative Humidity: 10 to 90%
10.4-inch color TFT LCD touch panel (non condensing)
・ Atmospheric pressure: 700 to 1060 hPa
Active Probe Ports
 For electronic scanning probes: 4 Power Requirement
 For independent probes*: 1 ・ 115/ 200 to 240V ±10%, 50 or 60 Hz,
*Option: EU-3044 Max. 1200 VA (with optional recorders connected)
Max. 800 VA (main unit only)
Input/Output Signals
Dimensions
– Data Input/Output
・ 59 cm (W) × 100 cm (D) ×120 - 168cm (H)
 USB2.0: 8 channel
 RS-232C: 1 channel Weight
– Digital Video Input/Output ・ Approx. 170 kg (main unit only)
 DVI-D digital (output only): 2 channel
Resolution: WSXGA(1600*1024 pixel)
for main display
SXGA(1280*1024 pixel) for DV-800
 DV_ITF(NTSC/PAL)*1: 1 channel
1
* Option: EU-9141 + PM-F75-H001

15
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

System Configuration
ProSound F75 main unit including a viewing monitor

Optional Software (License Key) Optional Recorders/Printers/Units/Probe connection kits

Contrast Echo (CE) Physiological Signal Display DVD recorder


unit
SOP-F75-44 PEU-F75 DV-800(C)
PM-F75-H002 is necessary for
CE analysis Pulse wave transducer: TY-307A connection of this recorder.
(DVI input is possible.)
SOP-F75-14 Independent probe
connection unit
Real-time Tissue
Analog video signal output unit
Elastography (RT-Elasto) EU-3044
EU-9141
SOP-F75-43 Footswitch
DV signal output unit
RT-3D (4D) MP-2345B, MP-2614B
PM-F75-H001
SOP-F75-4 Endo-cavity probe holder
3D Volume MP-PH-F75-2 DVD recorder
Measurements(AVM,TVM) MP-FX-F75-4-L is necessary DVO-1000MD (SONY)
PM-F75-H004 is necessary.
SOP-F75-20B Probe holder for ASU-1013
Flow 3D MP-PH7U B/W printer
MP-FX-F75-4-L is necessary
SOP-F75-35 SONY: UP-D897
Olympus ultrasonic endoscope
Freehand 3D MITSUBISHI: P95DW
connection kit
SOP-F75-37 See page 22 for details.
Color Printer
STIC
2D Tissue Tracking SONY: UP-D25MD
SOP-F75-41
SOP-F75-49 MITSUBISHI: CP30DW
Automated NT
Measurement Automated IMT Measurement Wireless Remote Controller
SOP-F75-42 SOP-F75-38 RM-700

Flow Profile Stress Echo


MotorizedTEEProbe(UST-52126)
SOP-F75-7 SOP-F75-15 connection kit

PM-F75-H005
DICOM communication
SOP-F75-10

Structured Report

SOP-F75-21

TDI/Strain Analysis*1 eTRACKING (ET)*1

SOP-F75-13 SOP-F75-11

Extended Field of View Wave Intensity (WI)*1*2


SOP-F75-1 SOP-F75-25

FMD analysis*1*2 Arm holder: MP-AH0001


MP-AH0001-1
SOP-F75-16 Probe Fixing unit: MP-PH0001

*1 Physiological Signal Display unit is a must to perform this function.


*2 The SOP-F75-16 (FMD analysis software) and SOP-F75-25 (Wave
16 Intensity software) include eTRACKING software.
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

OPTIONAL PROBES

Electronic convex sector probes T.H.E: Tissue Harmonic Echo, BbH: Broadband Harmonics,
CE: Contrast Echo, SCI: Spatial Compound Imaging,
Ultrasound Frequencies (MHz) Scanning Radius of
Application
Model angle curvature Optional accessories
(description) B and M modes Doppler/Flow (degrees) (mmR)
Abdomen, UST-9147 3.00/3.75/5.00/6.00 Flow: 71 51 -
OB/GYN BbH: 2.14/2.50 2.14/2.50/3.00/3.75
CE: PW:
1.33/1.50/1.88/2.50 2.14/2.50/3.00/3.75
3.00/3.75/4.00 CE: 1.88
Abdomen, UST-9130 3.00/3.75/5.00/6.00 Flow: 70 60 Puncture adapter:
OB/GYN BbH: 2.14/2.50 2.14/2.50/3.00/3.75 MP-2473
CE: PW:
1.50/1.88/2.14/2.50 2.14/2.50/3.00/3.75
CE: 1.88
Abdomen, UST-9115-5 3.75/5.00/6.00/7.50 Flow: 60 60 Puncture adapter:
OB/GYN BbH: 3.00/3.33/3.75 3.00/3.75/5.00/6.00 MP-2473
PW:
3.00/3.75/5.00/6.00
Endo-cavity UST-9118 3.75/5.00/6.00/7.50 Flow: 180 9 Puncture adapter set
BbH: 4.00/5.00 3.75/4.29/5.00/6.00 MP-2748-SET
PW:
3.75/4.29/5.00/6.00
Endo-cavity UST-675P 3.75/5.00/6.00/7.50 Flow: 180 9 Probe cover:
BbH: 4.00/5.00 3.75/4.29/5.00/6.00 RB-945BP-S *1
RB-945BP-NS *2
PW:
(Puncture adapter
3.75/4.29/5.00/6.00 MP-2452 is included
as standard.)
Intraoperative UST-9120 5.00/6.00/7.50/10.0 Flow: 70 20 Waterproof connector
Small parts BbH : 3.75/5.00/6.00/7.50 cover: MP-2790
Neonatal head 5.00/5.00/7.50/7.50 PW :
3.75/5.00/6.00/7.50

Intraoperative UST-9133 3.00/3.75/5.00/6.00 Flow: 82 20 Probe cover:


BbH: 2.14/2.50 2.14/2.50/3.00/3.75 RB-665P-S *1
Biopsy RB-665P-NS *2
CE: PW:
Puncture adapter :
1.50/1.88/2.50/3.00/ 2.14/2.50/3.00/3.75 MP-2781
3.75 CE: 1.88/2.50 MP-2781-5
MP-2781-25
Abdominal UST-9135P 3.00/3.75/5.00/6.00 Flow: 70 60 Puncture adapter:
BbH: 2.14/2.50 2.14/2.50/3.00/3.75 MP-2785
CE: PW:
1.50/1.88/2.14/2.50 2.14/2.50/3.00/3.75
CE: 1.88
Intraoperative UST-9132T 5.00/6.00/7.50/10.0 Flow: 70 20 Waterproof connector
UST-9132I CE:5.0/6.0/6.6 3.75/5.00/6.00/7.50 cover: MP-2790
/7.5/8.0/10.0 PW :
3.75/5.00/6.00/7.50

Intraoperative UST-9146T 3.00/3.75/5.00/6.00 Flow: 82 20 Waterproof connector


UST-9146I BbH: 2.14/2.50 2.14/2.50/3.00/3.75 cover: MP-2790
CE: PW:
1.50/1.88/2.50/3.00/ 2.14/2.50/3.00/3.75
3.75 CE: 1.88/2.50

*1 Sterilized (Note: Sterilized probe cover cannot be sold in EU member countries.) *2 Non-sterilized

17
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Electronic linear probes :Trapezoidal Scan


Application Ultrasound Frequencies (MHz) Scanning
Optional
(description) Model width
B and M modes Doppler, Flow (mm) accessories
Peripheral vessels/ UST-5415 5.00/7.50/10.0/13.0 Flow: 3.75/5.00/6.00/7.50 36 Probe stabilizer:
BbH: 5.00/7.50/10.0 PW: 3.75/4.29/5.00/6.00
Small parts CE: CE: 4.00 MP-2804
3.40/3.75/4.30/4.90/5.0 CW: 3.75
0/5.40/6.00/7.50
Peripheral vessels/ UST-5411 5.00/7.50/10.0/13.0 Flow: 3.75/5.00/6.00/7.50 36 -
Small parts BbH: 7.50/10.0 PW: 3.75/4.29/5.00/6.00
(Compound Array)

Small Parts UST-567 5.00/7.50/10.0/13.0 Flow: 3.75/5.00/6.00/7.50 50 -


BbH: 3.75/5.00 PW: 3.75/4.29/5.00/6.00
CE: CE:5.22
3.75/4.00/4.40/4.80
5.00/5.30/6.00/7.50
Intraoperative UST-5713T 5.00/6.00/7.50/10.0 Flow: 4.29/5.00/6.00/7.50 60 Puncture
PW: 4.29/5.00/6.00/7.50 adapter:
MP-2448

Peripheral vessels/ UST-5417 5.00/7.50/10.0/13.0 Flow: 5.00/6.00/7.50/8.00 38 Probe stabilizer:


BbH: 5.00/7.50/10.0 PW: 5.00/6.00/7.50/8.00
Small parts MP-2809

Peripheral vessels/ UST-5419 3.75/5.00/6.00/7.50 Flow: 3.00/3.75/4.29/5.00 38 Echo


BbH: PW: 3.00/3.75/4.29/5.00 Coupler(Slant):
Abdomen 3.00/3.75/4.00/5.00 CE:2.50
CE: EZU-PA3C1
2.50/3.00/3.53 Puncture
4.00/5.00/6.00
adapter:
EZU-PA3C1H
Echo
Coupler(Horizont
al):
EZU-PA3C2
Intraoperative UST-5418 5.00/7.50/10.0/13.0 Flow: 3.75/5.00/6.00/7.50 36
BbH: 7.5/10.0 PW: 3.75/4.29/5.00/6.00
CE:
3.75/4.00/4.62/5.00
5.22/6.00/7.50

18
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Electronic phased array sector probes


T.E.E.: Trans-esophageal Echocardiography
Ultrasound Frequencies (MHz) Scanning
Application Optional
Model angle
(description) B and M Doppler & Flow accessories
(degrees)
Cardiology UST-52105 2.50/3.00/3.75/5.00 Flow: 1.88/2.14/2.50/3.00 90 -
THE: 1.50/1.70/1.88 PW: 1.50/2.14/2.50/3.00
BbH: 1.50/1.70/1.88
CW: 1.88
CE: 1.50/1.25/1.50/1.88
CE: 1.88
Pediatric UST-52124 3.75/5.00/6.00/7.50 Flow: 3.33/3.75/4.29/5.00 90 -
cardiology THE: 3.00 PW: 3.33/3.75/4.29/5.00
CW: 3.00
Rotary plane UST-5293-5 3.75/5.00/6.00/7.50 Flow: 3.33/3.75/4.29/5.00 90 -
T.E.E. PW: 3.33/3.75/4.29/5.00
CW: 3.75
Rotary plane UST-52119S 3.75/5.00/6.00/7.50 Flow: 3.33/3.75/4.29/5.00 90 -
T.E.E for PW: 3.33/3.75/4.29/5.00
pediatrics CW: 3.75
Neonatal UST-52110S 3.75/5.00/6.00 Flow: 3.00/3.75/5.00/6.00 90
Cardiology PW: 3.00/3.75/5.00/6.00
CW: 3.75
Intraoperative UST-52114P 3.75/5.00/6.00/7.50 Flow: 90 Includes biopsy
3.75/4.29/5.00/6.00 adapter as
PW: standard
3.75/4.29/5.00/6.00
Neonatal UST-5296 5.00/6.00/7.50/10.0 Flow: 90 -
Cardiology 4.29/5.00/5.71/6.67
PW:
4.29/5.00/5.71/6.67
CW:3.75
Rotary plane UST-52126 3.75/5.00/6.00/7.50 Flow: 90 -
T.E.E. 3.00/3.33/3.75/4.29
PW:
3.00/3.33/3.75/4.29
CW:3.00
RT-3D (4D) Probe*
* SOP-F75-4 is necessary.
Ultrasound Frequencies (MHz) Radius of
Scanning Optional
Application Model range
curvature
B, M Doppler/Flow (mmR) accessories
Trans-abdominal ASU-1010 3.75/5.00/7.50/10.0 Flow: 2.14/2.50/3.00/3.75 70/60 40 —
scanning THE: 2.14 PW: 2.14/2.50/3.00/3.75 degrees
BbH: 2.14/2.50 CE: 1.88
CE: 1.50/1.88/2.14/2.50
Endovaginal ASU-1012 3.75/5.00/6.00/7.50 Flow: 3.00/3.75/5.00/6.00 140/84 10 —
Gynecological THE: 2.50/3.00 PW: 3.00/3.75/5.00/6.00 degrees
BbH: 3.75/5.00/6.00
Small parts ASU-1013 5.00/7.50/10.0/13.0 Flow: 3.75/5.00/6.00/7.50 36mm/
THE: 6.00/7.50/10.0 PW: 3.75/4.29/5.00/6.00
33deg.
BbH: 7.50/10.0 — —

19
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Bi-plane probe
Application Ultrasound Frequencies (MHz) Scanning Radius of
curvature Optional
(descriptio Model angle
B and M Doppler & Flow (mmR) accessories
n) (degrees)
Tramsrectal UST-678 Convex 3.75/5.00/6.00/ Flow: 120 9 Puncture adapter:
7.50 3.00/3.75/5.00/6.00 MP-2451
PW:
3.0/3.75/5.00/6.00 Probe cover:
Linear BL-664-S
5.00/6.00/7.50/ Flow: 60mm
(sterilized)*
10.0 3.00/3.75/5.00/6.00 BL-664-NS(Non-st

PW: erile)
3.00/3.75/5.00/6.00
Tramsrectal 3.75/5.00/6.00/ Flow: 178 9 Probe cover:
UST-677P Convex 7.50 3.75/4.29/5.00/6.00 RB-665P-S
2
BbH: 4.00/5.00 PW: (sterilized)*

3.75/4.29/5.00/6.00 RB-665P-NS(Non-s
terile)
Convex 3.75/5.00/6.00/ Flow: 178 9 Puncture adapter
7.50 3.75/4.29/5.00/6.00 is included as
BbH: 4.00/5.00 PW: standard.
3.75/4.29/5.00/6.00
* Sterilized probe cover cannot be sold in EU member countries.

Independent CW Doppler Probes*


* Independent probe connection unit EU-3044 is necessary.
Ultrasound Frequencies (MHz)
Application Model Optional accessories
B, M Doppler
CW Doppler (for heart) UST-2265-2 - CW: 2.00 -
CW Doppler (for peripheral vessels) UST-2266-5 - CW: 5.00 -

20
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Olympus Ultrasonic Endoscopes*


* Optional software SOP-F75-17, SOP-F75-18, SOP-F75-36, SOP-F75-39 or SOP-F75-40 is
necessary.
Frequency (MHz)
Model number Scanning Radius of
B/W B and M Remarks
(Olympus) Doppler/Flow angle curvature
mode
GF-UE160-AL5 5.0/6.0/7.5/10.0 Flow: 360 deg. 5 mm SOP-F75-18 is necessary.
GF-UE260-AL5 THE: 3.75/5.0 3.75/5.0/6.0/7.5 “UST-OLY-R5R1“ is displayed on
the screen.
BbH: PW:
3.75/5.0/6.0/7.5 3.75/5.0/6.0/7.5
CE:
3.40/4.00/4.30/4.
70/5.00/5.40/6.00
/7.50
BF-UC180F 5.0/7.5/10.0/12.0 Flow: 60 deg. 8 mm SOP-F75-36 is necessary.
BF-UC260FW THE: 3.75/5.0 3.75/5.0/6.0/7.5 “UST-OLY-R8C2“ is displayed on
the screen.
BbH: PW:
3.75/5.0/6.0/7.5 3.75/5.0/6.0/7.5
GF-UCT180 5.0/6.0/7.5/10.0 Flow: 180 deg. 5 mm SOP-F75-39 is necessary.
GF-UCT260 THE: 3.75/5.0 3.75/5.0/6.0/7.5 “UST-OLY-R5C4 “ is displayed on
the screen.
BbH: PW:
3.75/5.0/6.0/7.5 3.75/5.0/6.0/7.5
CE:
3.40/4.00/4.30/4.
70/5.00/5.40/6.00
/7.50

GF-UCT140-AL5 5.0/6.0/7.5/10.0 Flow: 180 deg. 5 mm SOP-F75-17 is necessary.


GF-UCT240-AL5 THE:3.75/5.0/6.0/ 4.29/5.0/6.0/7.5 “UST-OLY-R5C1“ is displayed on
7.5 PW: the screen.
BbH:3.75/5.0/6.0/ 4.29/5.0/6.0/7.5
7.5
TGF-UC260J 5.0/6.0/7.5/10.0 Flow: 90 deg 5 mm SOP-F75-40 is necessary.
TGF-UC180J THE: 3.75/5.0 3.75/5.0/6.0/7.5 “UST-OLY-R5C5 “ is displayed on
the screen.
BbH: PW:
3.75/5.0/6.0/7.5 3.75/5.0/6.0/7.5
Notes: The above endoscopes are not marketed in some countries and areas.
Marketable models differ according to the countries and areas.

It is forbidden to use all probes in direct contact with the central nervous system in Europe.

21
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Probes and available functions


Convex sector
Function Free-
Elasto- Dual
THE BbH CE TDI EFV AIP hand eFLOW SCI
graphy Doppler
Probe 3D
UST-9147 - YES YES YES YES YES YES - YES YES YES
UST-9130 - YES YES YES YES YES YES - YES YES YES
UST-9115-5 - YES - - YES YES YES - YES YES YES
UST-9118 - YES - - - YES - - YES - -
UST-675P - YES - - - YES - - YES - -
UST-9133 - YES YES - - YES YES - YES - -
UST-9135P - YES YES - YES YES YES - YES YES -
UST-9120 YES - - - YES YES - YES - -
UST-9132T - - YES - - YES YES YES YES - -
UST-9132I - - YES - - YES YES - YES - -
UST-9146T - YES YES - - YES YES - YES - -
UST-9146I - YES YES - - YES YES - YES - -
Linear
Function ET/ Free-
Elasto- Trape- Dual
THE BbH CE TDI EFV FMD/ hand eFLOW CW SCI
graphy zoid Doppler
Probe WI 3D
UST-5415 - YES YES YES YES YES YES YES YES YES YES YES -
UST-5411 - YES - YES YES YES YES YES YES - YES YES -
UST-567 - YES YES - YES - YES YES YES - YES YES YES
UST-5713T - - - - - - YES - YES - YES YES -
UST-5417 - YES - YES YES YES YES YES YES - YES YES -
UST-5419 - YES YES YES YES YES YES YES YES - YES YES -
UST-5418 - YES YES YES - - - YES YES - YES YES -
Phased array sector
Function ET/ Free-
Elasto- Dual
THE BbH CE M/PW TDI EFV FMD/ hand eFLOW CW SCI
graphy Doppler
Probe WI 3D
UST-52105 YES YES YES YES YES - - - - YES YES - YES
UST-52124 YES - - YES YES - - - - YES YES - YES
UST-5293-5 - - - YES YES - - - - - YES - -
UST-52119S - - - YES YES - - - - - YES - -
UST-52110S - - - YES YES - - - - - YES - -
UST-52114P - - - YES - - - - - YES - - -
UST-5296 - - - YES YES - - - - YES YES - YES
UST-52126 - - - YES YES - - - - - YES - -

RT-3D (4D)/ Independent


Function ET/ Free-
Elasto- Trape Dual
THE BbH CE TDI EFV FMD/ hand eFLOW CW SCI
graphy zoid Doppler
Probe WI 3D
ASU-1010 YES YES YES - - - - - YES - YES - YES
ASU-1012 YES YES - - - - - - YES - - - -
ASU-1013 YES YES - - - - - YES - YES YES -
UST-2265-2 - - - - - - - - - YES - - -
UST-2266-5 - - - - - - - - - YES - - -
Bi-plane
Function Free-
Elastogra Trape-
THE BbH CE TDI EFV AIP hand eFLOW SCI
phy zoid
Probe 3D
UST-678 - - - - - YES - - YES - -
UST-677P - YES - - - YES - - YES - -

22
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

Olympus Ultrasonic Endoscopes


Function ET/ Free-
Elasto- Trape
THE BbH CE M/PW TDI EFV FMD/ hand eFLOW CW SCI
graphy zoid
Probe WI 3D
GF-UE160-AL5
YES YES YES - - - - - - YES - - -
GF-UE260-AL5
BF-UC180F YES YES
- - - - - - - YES - - -
BF-UC260FW
GF-UCT180
YES YES YES - - - - - - YES - - -
GF-UCT260
GF-UCT140-AL5
YES YES - - - - - - - YES - - -
GF-UCT240-AL5
TGF-UC180J
YES YES - - - - - - - YES - - -
TGF-UC260J

23
FOR REFERENCE ONLY.
ACQUIRE CURRENT LABELING FROM THE MANUFACTURER.

 The specifications are subject to change without notice.

 The standard components and optional items differ depending on the country.
Not all products are marketed in all countries.
Please contact your local Hitachi Aloka Medical distributors for details.

 Real-time Tissue Elastography, HI REZ are registered trademarks or trademarks of Hitachi


Medical Corporation in Japan and other countries.

 ProSound is a registered trademark or trademark of Hitachi Aloka Medical, Ltd. in Japan and
other countries.

 Harmonic Echo is a registered trademark or trademark of Hitachi Aloka Medical, Ltd. in Japan
and other countries.

 DICOM is the registered trademark of the National Electrical Manufacturers Association for its

standards publications relating to digital communications of medical information.

6-22-1 Mure, Mitaka-shi, Tokyo 181-8622 Japan


Phone: +81 422 45 6049, Fax: +81 422 45 4058
www.hitachi-aloka.com

Printed in Japan, 2013-03


SP-F75-V31-01
24