Catalogue Internet
Catalogue Internet
1
GENERAL
CATALOG
2023
HEMATOLOGY BIOCHEMISTRY IMMUNOLOGY
Justin MAFOUANA
2
HEMIX 3-60 (A0214)
WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%,
Parameters RBC, HGB, MCV, MCH, MCHC, RDW-CV, RDW-SD, HCT,
PLT, MPV, PDW, PCT, P-LCR, P-LCC
Measuring principles Impedance technology for blood cell counting and cyanide-free optical detection for
hemoglobin
0.30~1.00
Performance
Quality control Levey Jennings charts and calculation of Mean, SD, CV; up to 9 control files, 31 runs/ file.
Throughput 60 tests per hour
Venous ≤ 14 μl
Sample volume
Pre-diluted ≤ 20 μl
Memory storage 50 000 patients files - Multiple conditions query available - LIS transfer HL7 protocol
External monitor - RS232 x 1; USB port x 4 ; Ethernet x 1
Input / Output
Mouse, keyboard and barcode reader (optional)
Fast thermal printer, 57.5 mm wide paper, recording width 48 mm
Printout
Optional external printer, Ink jet or laser
Display 8'' TFT LCD touchscreen, high resolution 800 x 600, true colors
Interface Multilingual Linux® software highly user-friendly & intuitive
Temperature 18°C - 30°C
Operating
Humidity < 70
environment
Atmospheric pressure 70 - 106 kPa
Power requirements A.C. 110 / 220 V ± 10%; 50 - 60 Hz ; Input : ≤ 150VA - Fuse: A.C. 230V; 3.15 A
Dimensions 330 (W) x 430 (H) x 380 (D) mm ;
& weight 23 kg
NAME VOLUME REFERENCE
TYPE REFERENCE
7
BLOODTROL 16 - 3 TUBES N R021005
8
9
HEMIX 5-60 (A0210)
CBC + 5 DIFF mode
26 parameters: WBC, LYM#&%, MON #&%, NEU#&%, EOS#&%, BAS#&%, RBC, HGB,
Parameters
HCT, MCV, RDW (sd/cv), MCH, MCHC, PLT, PCT, MPV, PDW (sd/cv), P-LCR, P-LCC ; 2
histograms: RBC, PLT ; 2 scattergrams: 4 diff, BASO
Laser light scatter technology for 5-part WBC diferential
Measuring principles Impedence method for CBC: WBC(80μm), RBC & PLT (70μm)
Light absorbance for HGB measurement: cyanide free method
Samples Reagents
Volume: 100 μL of whole blood 3 cyanide-free reagents - 1x diluent, 2x
Reagents & sampling Closed & open tube sampling with optional lyse
system autosampler
Stat mode, specific control materials &
calibration
Quality control Separate QC database, unlimited number of QC files + Levey-Jennings graphs
Throughput 60 tests per hour
Capacity: 100 tubes - 10 racks with 10 sample tubes
Accept primary tubes, monovette, vacutainer, vacuette
Optional autosampler Built-in barcode reader / Built-in sampler mixer
Cap recognition (only capped tubes are mixed & processed)
Dimensions: 300 (W) x 300 (D) x 180 (H) mm Weight: 10 kg
Display User interface 600 x 800 color graphics LCD, touchscreen 10.4''
Printout External Microsoft Windows®compatible printers
Memory storage 100 000 records
Input / Output External keyboard - PS/2 or USB, RS232, USB, Ethernet
Windows® 8.1 embedded, multilingual, multiuser mode (multi level user modes with
Interface individual identification ex: username, password ...), software upgrade via USB port (USB
memory stick)
Operating Temperature 15°C - 35°C
environment Relative humidity max 80% (non-condensing)
A.C. 110 / 220 V ± 10%; 50 - 60 Hz
Power requirements
Input power 400 W
Dimensions 400 (W) x 450 (H) x 500 (D) mm ;
& weight 35 kg
NAME VOLUME REFERENCE
TYPE REFERENCE
10
BLOODTROL 22 - 12 x 3 TUBES LNH R024002
11
12
13
HEMIX 5-PRIME (SOON)
Whole Blood Mode: WBC, Neu%, Lym%, Mon%, Eos%, Bas%, Neu#, Lym#, Mon#, Eos#,
Bas#,
RBC, HGB, MCV, MCH, MCHC, RDW-CV, RDW-SD, HCT, PLT, MPV, PDW, PCT, P-LCC, P-
Parameters
LCR; 4 Research Parameters: ALY%, LIC%, ALY#, LIC#
3 Histograms: Baso , RBC, PLT
1 Scattergram for WBC Differential
Flow cytometry technology for 5-part WBC differential
Measuring principles Impedance method for CBC: WBC, RBC and PLT
Coloric method for HGB determination
Parameters Linearity Precision CV%
9
WBC 1 ~ 300 x 10 /L ≤ 2.0% (3.5 ~ 15.0)
RBC 0.3 ~ 7 x 1012/L ≤ 1.5% (3.5 ~ 6.0)
HGB 20 ~ 240 g/L ≤ 1.5% (110 ~ 180)
Performance
HCT 0 ~ 75 % PLT
10 ~ 7000 x 109/L ≤ 4.0% (100 ~ 500)
MCV ≤ 1.0% (70 ~ 120) fL
REAGENTS
SOON IN 2023
MAINTENANCE SOLUTION
TYPE REFERENCE
14
BLOODTROL 22 - 6 x 3 TUBES N R024007
15
ESR 3000 (A0301)
Photometric infrared reading principle
Principles
Automatic measurement of RBC level every 3 min
Reproducibility ‹ 3%; ± 2 mm
Automatic temperature
ESR results can be automatically adjusted to 18°C
correction
Weight 10 kg
TUBES ESR CITRATED TUBES VACUUM GLASS TUBES 1 000 units C030003
16
2 x 9 ml vials R030001
NORMAL &
SEDTROL 115 days 28 days
PATHOLOGICAL 4 x 9 ml vials R030002
17
18
COMPATIBLE REAGENTS
ABBOTT®
TYPE NAME REFERENCE VOLUME
CD 1600
DILUENT ISOTONIC SOLUTION LMG HAD320 20 L
CD 1700
LYSE LYSE LMG HAL302 2L
CD 1800
CLEANER ISOTONIC DETERGENT LMG HAG320 20 L
CD 2000
CLEANER FOR CD
ISOTONIC DETERGENT LMG HAG320 20 L
3500 / 3700
BECKMAN COULTER®
TYPE NAME REFERENCE VOLUME
STKS
DILUENT DILUTON 3 HCD320 20 L
MAXM
LYSE LYSOGLOBINE 3D HCL302 1L
GENS
CLEANER DILUCLAIR HCC302 5L
HMX
DIATRON®
TYPE NAME REFERENCE VOLUME
19
COMPATIBLE REAGENTS
HORIBA MEDICAL®
(ABX)®
TYPE NAME REFERENCE VOLUME
DILUCLAIR A HXC302 1L
CLEANER
HEMACLAIR HXC101 500 mL
20
COMPATIBLE REAGENTS
MEDONIC®
TYPE NAME REFERENCE VOLUME
CA 530
DILUENT DILUTON 3M HMD320 20L
CA 570
LYSE LYSOGLOBINE 3M HML305 5L
M16 / M20
CLEANER DILUTERGE M HMG305 5L
M-SERIES
MINDRAY®
TYPE NAME REFERENCE VOLUME
BC 2300
DILUENT DILUTON 3I HID320 20 L
BC 2800
BC 3000
LYSE LYSOGLOBINE 3I HIL301 500 mL
BC 3200
BC 3600
BC 5100 BC DILUTERGE 3I HIG320 20 L
5180
BC 5300 CLEANER DILUCLAIR I HIC301 60 mL
BC 5380
HEMACLAIR I HIC101 60 mL
NIHON KOHDEN®
TYPE NAME REFERENCE VOLUME
MEK 6318
LYSE LYSING REAGENT LMG HNKL301 1L
MEK 6400 ENZYMATIC CLEANER HNKG005 5L
CLEANER
CHLORINATED CLEANER HNKG301 1L
21
COMPATIBLE REAGENTS
ORPHÉE®
TYPE NAME REFERENCE VOLUME
PROKAN®
TYPE NAME REFERENCE VOLUME
CONCENTRATED
DILUCLAIR PE HKC301 60 mL
CLEANER
RAYTO® SIEMENS®
22
COMPATIBLE REAGENTS
ADVIA 60 & 70 Diluent, Lyse, Cleaner & Sheath -> Available upon simple request
SWELAB®
TYPE NAME REFERENCE VOLUME
SYSMEX®
TYPE NAME REFERENCE VOLUME
HYL305 500 mL
K 800 LYSOGLOBINE 3ST
HYL350 5L
K 1000 LYSE
HYL405 500 mL
LYSOGLOBINE ST Hb
K 4500 HYL450 5L
23
COMPATIBLE REAGENTS
24
26
27
BSA
3000
(A0101)
End point, Kinetic, two point Kinetic, Fixed time, Absorbance, Turbidimetry
Measuring method Reagent blank, Sample blank
Monochromatic & bichromatic method
Memory storage 3000 results per day available in memory at any time and transferable via LIS
28
Dimensions
445(W) x 190(H) x 420(D) mm 9kg
& weignt
STANDARD MODEL
29
1 position sample rack
MCI electrodes : Na+ & K+ Test
on primary tubes
30
IONI
X
(A04
08)
Ion Selective Electrode direct measurement
Principles
Manometric method for TCO2
Parameters 8 parameters: Na+, K+, Cl-, Ca2+, pH, Li+, TCO2, A.G.
With SFRI controls : 3 levels with unlimited results per level and Levey Jenning
Quality control
graphs Any other control accepted
Infinate - depends on the SD card (8Go offered) - patients files with results & log journal &
Memory storage calibration
Display Large backlit LCD 7’’ color touch screen ; High resolution 800 x 480
Operating Temperature: 18 to 32°C (64. °F to 90°F) ; Humidity: 80% maximum at 32°C (90°F)
environment Reagent storage temperature: 10°C to 40°C (50,0°F to 104,0°F)
31
Power requirements A.C.: 100-240 V, 2.0 A, 50-60 Hz
1 easily installed module for quick & precise dosage of bicarbonates & Anion Gap
TCO2 module
measurement
1 reference electrode +
6 electrodes available 2 standard electrodes: Na+, K+
4 optional electrodes: Cl-, Ca2+, pH, Li+
32
33
ISE SERIES
(A0400-05)
Ion Selective Electrode direct measurement
Principles
Manometric method for TCO2
Serum, plasma, urine and whole blood 150 μL
Sample Specific modes for urine and lipemic serum
Throughput 60 tests / hour - without TCO2 ; 45 tests / hour - with TCO2
Urine
K+ 50 - 250 mmol / L(pre-diluted) Na+
10 - 350 mmol / L
Cl- 10 - 350 mmol / L
Accuracy Reproducibility
K+ ±0.2 mmol/L < 1.0%
Na+ ±2.0 mmol/L < 1.0%
Cl- ±2.5 mmol/L < 1.0%
Accuracy &
Li+ ±0.1 mmol/L < 3.0%
reproducibility
Ca2+ ±0.1 mmol/L* < 2.0%
pH ±0.1 < 1.0%
TCO2 ±2.0 mmol/L* < 3.0%
* for plasma only
Quality control Two levels with optional autosampler ; one level without autosampler
Calibration Automatic: 2 points every 4h or on demande
35
36
COMPATIBLE REAGENTS
Total Protein
Albumine
Alanine Aminotransferase (ALT)
Aspartate Aminotransferase (AST) * BUN: Blood Urea Nitrogen
37
COMPATIBLE REAGENTS
C G11CC4001 C4 R1 1 x 40 mL + R2 1 x 10 mL Turbidimetric
P G11FIN001 Ferritin R1 2 x 50 mL + R2 3 x 7 mL Turbidimetric
R G11IGA001 IgA R1 1 x 40 mL + R2 1 x 10 mL Turbidimetric
38
COMPATIBLE REAGENTS
39
COMPATIBLE REAGENTS
40
COMPATIBLE REAGENTS
G13CTL003 CK MB control 1 x 2 mL CK MB
41
COMPATIBLE REAGENTS
42
COMPATIBLE REAGENTS
43
[Tapez ici]
[Tapez ici]
[Tapez ici]
IRE 96
Principle Absorbance technology with semiconductor photodiode
Plate type 96-well plate with U, V or flat bottom
Measuring System 8-channel manifold - Monochromatic and bichromatic
Spectral range: 400 - 750 nm
4 built-in filters: 405 nm, 450 nm, 492 nm and 630 nm
Optics Filter wheel can hold up to 8 filters
Tungsten halogen light source: 12 V/ 20 W
Optical linearity: ‹ 2.5%
Shaking 3 speed linear shaking (low, medium, high)
Reading range 0.001-3.500 ABS
Range: 0.001-2.500 ABS
Measurement Precision: ± 1%; ± 0.001 ABS
Speed: 12 s (monochromatic) or 30 s (bichromatic) per 96-well plate
Repeatability CV ‹ 1%
Memory storage 42 user-programmable protocols - 2,500 results per day available at any time
Quality control 2 QC files
Calibration Up to 8 points
Input / output Parallel printer interface - RS232 port for mono-directional LIS
Printout Optional external printer
Display 5.7” LCD touch screen - Mono-color display
Operating Temperature 10°C - 40°C - Humidity 15% - 85%
environment
Power requirements A.C. 110/220 V ± 10%; 50 – 60 Hz
Dimensions & weight 480(W) x 200(H) x 440(D) mm - 9.6 kg
IW 96
Plate type 96-well plate with U, V or flat bottom
Manifold 8-channel or 12-channel
Volume 50 - 450 μl in 5 μl increments
Liquid dispensing
Precision ‹ 5% at 3000 μl
N° of wash cycles 1-9
Wash modes Row or plate
Washing, soaking & Residual volume ‹ 2 μl for U and V bottom plates
shaking ‹ 3 μl for flat bottom plates
Aspiration strength Programmable simple or double
Soaking & shaking time 0 - 99 seconds
4 bottles vol. 3.5 L
Liquid containers
2 washing solutions, 1 distilled water, 1 waste with volume sensor probe
[Tapez ici]