User Manual
The information and descriptions contained on this document, and the relevant standards are
specified for the use of this device. Exclusively MİXTA owns all of the information contained on
this document. It is prohibited to copy this document or to make quotations from it either by
showing the source or not, and to publish it elsewhere without permission. MİXTA Company
can change the content of this document. The new content that is replaced overrides the old
content on this document. You can learn the developments related to the changed content
from www.mixta.com.tr
As determine by the ministry, the expected life of this product is 10 years, and the guarantee
period is 1 years.
General View
This manual contains information about the installation and the usage of endoWash full-
automatic endoscopy washing and disinfecting machine.
Please act according to the instructions specified on the package of the product during freight,
transport, and shipping, please place the device according to the instructions on the package
move.
Warnings about Human and Environmental Health and Information about Misuses
* It is recommended that the device is used by the trained people in this regard.
* It is recommended that the installation is carried out in special rooms and places dedicated to this
or similar devices, if possible.
1
* If a failure occurs, do not open or dissemble the device, notify the failure by calling the authorized
service.
* In case of the electric shock risk, the device has a leakage current relays and fuses, the electricity
wiring in the environment should be according to the standards described in the guide.
* When the maintenance and repair are needed, please contact the authorized service.
* A soft and slightly damp cloth should be used for the interior and exterior cleaning of the device.
No other materials should be used.
Please keep the device clean according to the procedures described above. Apart from this, all
periodic maintenance and checks will be carried out by our authorized service. However, in order to
make your device last longer, recommended maintenance procedures are included in this manual.
(p. 73-75)
2
Connection and Installation Information
As the first connection and installation of the device is carried out by authorized technical personnel
of the company, it is recommended that no other unauthorized person intervene the installation and
execution of the device. Connection and installation information are provided below.
3
Contents
4
16.Program Air Settings ……..…..…………………………44
17.Emptying the Disinfectant Mixture ……..…..…………………………49
18.Manuel Operations ……..……..………………………51
19.Settings….(Time, date, staff, device, ext.) ……..……..………………………59
20.Auto Disinfection ……..……..………………………67
21.Emptying the Tanks ……..……..………………………68
22.How to use Leak Test Program ……..………..……………………69
5
1. Pre Information
Endowash is a device that allows washing and disinfecting the flexible endoscopes in a full
automatic way after being used.
Washing and disinfecting the flexible endoscopes after being used meet your daily needs
by virtue of Endowash’s special design and function.
By using the information about the installation and usage contained in this manual, you can
use Endowash device within a very short time.
This manual contains several important warnings and information. As the information and
the warnings may have an important performance factor in the usage of the device, the
relevant information and warnings should be taken into account.
You can call our help center for receiving information and advice about the operations
made in the device and any question that you cannot find answer in the manual. TEL:
+903123947215 or [email protected] for all questions.
6
2. Package Content
When you purchase your device, you receive all the parts in the list below that may start the
washing and disinfection.
Important Note: It is required to provide water supply from the water filter that makes water
purification at least 1 micron level and brings the microbiological protection by using
UltraViole in water inlet of the device.
Your device contains the following parts.
7
vııı. Tank evacuation hose (3 pcs)
ıx. User Manual
x. Regular Maintenance Documents
Endowash full-automatic washing and disinfection machine is available for your use after being
tested many times.
Before using the machine, the necessary training related to the usage of the machine should be
received. After the machine is delivered, the necessary training will be provided and the person or
organization receiving this training will be documented in writing by the main manufacturer. In order
to avoid certain problems that may affect the performance of the machine during the uncertified
usage, the necessary training should be received.
Machine located on panel description contains Turkish users at every stage in the machine what
steps are built, it is possible to see information about.
As the front panel of the machine contains descriptions in English, it is possible to receive
information about the operations in every stage.
The installation is described in this manual, and you can contact with MİXTA Help Desk about your
problems that may occur after the installation.
8
4. Installation
It is required to prepare a suitable area according to the sizes of the machine as well as the
preparation of electricity and water where the machine will be used.
Endowash sizes:
Width: 80 cm, Height: 83 cm, Height: 106 cm,
Height when the cover is opened: 140 cm.
An area above the minimum sizes is needed to achieve the high-performance.
9
ii) Electrical and water connections of the machine
Electrical connection: Endowash machine works with 220 V electric voltages. There is 150 cm
grounded plug on the machine. Endowash should be at least 130 cm away from the grounded plug
on the wall. In addition to electrical and electronic circuits, there are many sensors. Therefore power
should provide the specified electric voltage and it should be free from parasites of the voltage and
be continuous.
As the disinfection is left half finished when the main voltage is cut off, the voltage continuance
should be maintained for a reliable disinfection.
Water Connection: Endowash machine needs water in washing and disinfection operations.
Machine works with water that is refined at 0,2 micron level, processed with UltraViole and cleared
from all kinds of organic or inorganic substances. Machinery should not connect to a water source
that does not contain these features. Water connection hose supplied with machine is 100 cm, one
end of which is connected to the machine and the other one is connected to the water filter. For this
connection, 3A female water connector is needed for the water filter. Water filter must be changed
in every 6 months.
10
iv) Other important issues about the machine area:
a. There should not be any other electromagnetic machine that may affect the machine around the
Endowash machine.
b. Machine should not be located in a very humid environment.
c. A suitable working environment should be provided according to the operation values of the
machine.
11
C. Water connection
Water inlet connection hose that is supplied with the machine is connected to
the place at the back of the machine, labeled as “water inlet”. As the water
pressure will occur at the water inlet (SG – Su Girişi), make sure
the connection is done thoroughly.
The other end of hose will be connected to water purification system that is
connected to the water source and contains the above-mentioned features.
12
D. Sewage connection
Waste fluid hose that is supplied with the machine is connected to the place at the back of the
machine, labeled as “Waste Fluid” (AS). As the water pressure will occur at the water inlet, make
sure the connection is done thoroughly.
The other end of the hose should be connect to the point where the waste fluid is sent and at the
place where the waste fluid is send, the air tightness between the hose and the point where the
hose is connected should be connected.
Waste fluid should not be sent to the sewer system directly. Due to some chemicals used in your
machine, waste fluid may damage the nature. Therefore, waste fluid should be connected to a
sewage system that will not damage the nature.
”It is required to have an outlet that is enough to meet the output flow of the machine at the
place where the waste fluid hose is connected.
E.Electrical connection
On the back of the machine, an electric plug is available to run the machine with electricity. (EG) It
is required to connect this plug to the voltage system. Electricity should be in accordance with the
standards of the machine. Remember that instant voltage changes may damage the machine.
Problems due to the electric supply may exclude your machine out of the warranty coverage.
13
Machine has a system that protects both user and machine. Machine is grounded in general. The
leakage current relay at the back of the machine should be in working condition to protect people
from death due to the leakage current. In addition, it prevents damages that may be occurred at two
separate main fuses. If your machine is not working although it is connected to the electric system,
first check the fuses and the leakage current relay. If your machine does not work despite your
controls, disconnect your machine from the main power supply immediately. Do not turn on the
covers on the machine. In case of problems, please call MİXTA help desk.
"The earthing of the electric system to which you connect your machine should be in perfect
condition."
"If your machine does not work although it is connected to the electricity, please call the
authorized service."
"Never open the side covers and back cover of your machine. To open the machine while the
authorized service personnel is not present may be life-threatening."
14
F. Locking the machine foots
There are two movable, reversible wheels at the front, and two movable, irreversible wheels at the
back of the machine. After the completion of the operations specified above, place your machine
where you will use it. Making sure you have place the machine thoroughly, lock the two reversible
wheels by stepping on the locking mechanism.
"Do not use your machine before locking the wheels of your machine."
15
WASHING PHASES
SHORT PROGRAM
16
• Control panel
17
• Key Descrptions
18
Leak Test Start
Up
Leak Test Stop
Print Down
Setting
Cancel (Back space)
Program 1
OK. (Confirm)
Program 2
Px (P4 .. P11)
Stop (for the program operations)
Fill
Empty
19
Important! Please read until the end.
Dear User,
Your device's factory settings are programmed to use the most preferred mode, which is
“Periodical Manual (ready to use) disinfection”. This program consists of the following steps: 1.
Pre-Washing with water, 2. Washing with Enzymatic, 3. İntermediate Washing with water, 4.
Washing with Disinfectant, 5. Final Washing with water and 6. Filtration with air drying. For this
reason, the first pages of your manual explain how to operate your device accordingly.
You can change the washing times shown on the 16th page of your manual to be as long as you
want. You can also change any parameters in each stage.
If you want to install a different washing and disinfection program other than the factory settings,
you need an authorized account. How to create this account and a personal password is
described on this manual in detail.
If you feel that this manual is not explanatory enough, you can contact us on our Help Center
(+90 312 394 7215) or by e-mail at [email protected]
20
In order to clean up any blood or particles that have remained on the endoscope, you must
prewash your endoscope before putting it in the device. Biopsy channels should also be cleaned
with biopsy channel cleaners. (the biopsy cleaning kit is not supplied with the machine and you
need to obtain the relevant product from the endoscope vendors).
Place the pre-washed endoscope in the washing chamber of the machine. The parts that need
to be protected from water are closed with the connector. The channel connections on the
endoscope are connected to the washing channels in the machine.
The required washing and disinfection program is selected. As a result of the program, your
washed disinfected endoscope is ready for the next use.
If you encounter a problem after completing all the procedures described in the user manual
You can call the MİXTA Technical Support Line.
21
Starting The Machine
If you are confident that you have made your machine's water
connections and electrical connection perfect, you can open your
machine by pressing the on / off button on the front of your machine.
When you press the On / Off button, the model and program
information of your machine will appear on the LCD monitor screen in
the front panel within about 1 second.
If your machine has not been turned on despite the on / off button, call our technical
service is necessary.
never open the side covers of your machine except for authorized service stations in
order not to be exposed to vital danger.
22
In order to be able to wash and disinfect your device, you need disinfectant, enzyme and some
models of alcohol if you want to use it. If you do not have any of these three chemicals and the
program is not canceled, your device will give an error, warn you and will not work. When you
start the wash and disinfection program with your device, the device will automatically check
whether the chemicals (disinfectant, enzymatic-detergent or alcohol) contained in the selected
program are present in the tanks and will proceed with washing and disinfection. For this reason,
before going to wash and disinfection process, disinfectant, enzymes or alcohol should be filled
in again as described in this guide.
Your device has disposible disifectant, periodic automatic (concentrate) and periodic manual
(ready to use) disinfectant washing programs according to the disinfectant used. Since the
factory settings of your device are programmed according to the most preferred periodic
manual (Ready to Use) disinfectant type, it is explained how to use the disinfection
devices. Setting of other programs and filling of disinfectants have been promoted in the
following sections.
23
1.Filling the ready to use (periodic manuel) disinfectant
Be sure that "Tank I" or "Tank II" is empty by following the instructions on page 25,26
You have disinfectant mixes that are prepared outside or in ready-to-use containers 12-15 liters
in the MEYC Model until the scopes are completely covered with disinfectant. You should be
filled in to the empty and clean sink up to maximum point as in the picture
24
3.By using / keys, go to “Program 1” position.
(If your choice is Program1)
5. By using / keys, go to
“Empty” position.
25
6. By using OK key or directly
pressing key, empty the expired
disinfectant in tanks.
26
2. In LCD screen, go to “MAIN MENU”.
option.
4.By using OK key, enter to Other Main
Menu. .
27
7. Begin to fill enzymatic up to 1-5 liter. 9. When it is filled in full capacity, a new
audio warning will be heard: “Enz. is
8. You will hear an audio warning that full”. After this phase, filling operation
says “Enz. Filling operation has should CERTAINLY be stopped. Over
begun.” This is the first message of the filling could damage the device.
Enzymatic Tank. Continiue to fill
enzymatic.
28
the alcohol tank. Continue to fill
alcohol
9. When it is filled in full capacity, a
new audio warning will be heard:
“Alcohol is full”. After this phase,
filling operation should CERTAINLY be
stopped. Over filling could damage
6. Open “Alcohol ” cover in the sink. the device.
7. Begin to fill alcohol. Washing and Disinfecting Operation
8. You will hear an audio warning that
says “Alcohol Filling operation has
begun.” This is the first message of
29
9.When the operation is finished, you
will receive both a verbal and written
8.Press the “START” key message.
on the right of control panel or by using 10.An audio warning will be heard in
every 5 second at the beginning.
/ keys, go to “Start” position and
11.The operation is completed with
press the OK key. either “Operation is successful”
or “Operation is unsuccessful”
message.
3. By using / keys, go to
“Administrator” position. 5. Press the OK key.
4.
30
6. By using / keys go to Sign in
position
7. Press the OK key
31
3. By using OK. Key enter to “P1
2. By using / keys, go to PROGRAM CYCLE SETTINGS”
“Setting I” position. menu
32
4. In “P1 PROGRAM CYCLE” menu, you can choose among 3 types washing operation
for your program. Choose a, b or c options described in below and continue.
33
iii. By using / keys, choose between 1-30 days.
34
v. By using / keys, go to “Tank I.” or “Tank II.” position.
35
v. By using / vii. By using
keys, choose
between 1-30 / keys, go to
days. “Tank I.” or
“Tank II.”
vi. By using OK position.
key, enter to
store selection viii. By using
menu. (As the OK key, enter to
EndoWash contains “P1 PROGRAM
two disinfectant tank, SETTING”
you can use the second menu.
tank for manual
washing.)
36
6. Filtration I (extra washing) ->Program Cycle Settings->Tank I or
7. Alcohol II->Number of days selection->P1
8. Filtration II (air-dried system) Program Settings->…Prewash, Det.
Enzymatc,
In each step, time setting is in the form of İntermediate washing, Disinfectant…
hour:minute:second (00:00:00).
37
e. By using / keys,
set the time “00:00:00”-
“59:59:59”
f. By using OK key, switch
to temperature setting.
2.Det.-Enzymatic
a.Go to “P1 PROGRAM SETTING”
menu. (See Program Setting, Stage 1,
Introduction to settings.)
38
e.By using OK key, switch
b.By using / keys,“Det.-
to temperature settings.
Enzymatic” position.
f.By using / keys, set the
temperature “20”- “40”
centigrade.
g.By using OK. Key, switch
to dosage setting.
c.By using OK. Key, enter to“P1 h.By using / keys, make
Det.-Enzymatic”menu. the “000”- “250” ml/Lt
setting.
ı.By using OK.key, save
these settings.
3.Intermediate washing
39
b.By using / keys d.By using / keys, set the
time; “00:00:00”- “59:59:59”
“Intermediate washing” position.
40
4.Disinfectant
d.By using / keys, set the
a.Go to “P1 PROGRAM SETTING” time; “00:00:00”- “59:59:59”
menu. (See Program Setting, Stage 1
Introduction to settings.) e.By using OK. Key, switch
to temperature setting.
b. By using / keys,
“Disinfectant” position. f.By using / keys, set the
temperature “20”- “75”
centigrade.
41
5.Final washing
42
6.Filtration I
43
7.Alcohol
d.By using / keys,
a.Go to “P1 PROGRAM SETTINGS” make the “000”- “250”
menu (See Program Setting, Stage 1, ml/Lt setting.
Introduction to settings.)
e.By using OK key, save
these settings.
b. By using / keys, go to
“Alcohol” position.
44
c.By using OK key, enter to
“P1 Filtration II” menu.
5. Filtration II
45
• Program Air Setting (Stage 1 Introduction to Program)
3. By using / keys, go to
“Program 1” position. (If your
choice is Program 1)
4. By using OK key, enter to “P1
PROGRAM MENU”.
46
• Program Air Setting (Stage 2. Air Settings)
47
2. Air Det.-Enzymatic 3. Air Intermediate washing
48
4. Air Disinfectant 5. Air Final washing
49
6. Air Filtration I 7. Air Alcohol
a. By using / keys, go to
a. By using / keys, go to
“Air Alcohol” position.
“Air Filtration I” position.
b. By using OK key, enter
b. By using OK key, enter
to “P1 Air Alcohol” menu.
to “P1 Air Filtration I”
menu.
50
• Emptying the disinfectant mixture
51
If you choose “Periodic manual dis.” the
expired disinfectant mixture can be
emptied manually. This operation should 5.You will see a message that tells you
be carried out from Tank I” or “Tank II”, the disinfectant mixture period is expired.
according to your program selection, as
follows:
6. By using / keys, “Empty” position.
1.Turn the power button of the device to
“I” position.
52
Manuel Operations
3. By using / keys, go to
“Program 1” position.(İf your
6. By using OK key, enter to
choice Program 1)
“MANUEL CONTROL” menu.
4. By using OK key, enter to “P1 giriniz.
PROGRAM MENU”.
7. By using / keys, “Wash with
water” position.
5. By using / keys, go to
“Manuel Control” position. 8. By using OK key, start the
operation.
53
* Manuel washing with enzymatic
3. By using / keys, go to
“Program 1” position.(İf your
6. By using key, enter to
choice Program 1)
“MANUEL CONTROL” menu
4. By using OK key, enter to “P1
PROGRAM MENU”. 7. By using / keys, “Wash with
Enz.” Position.
5. By using / keys,go to
“Manual Control” position
8. By using OK key, start the
operation.
54
* Manuel washing with disinfectant
3. By using / keys, go to
“Program 1” position.(İf your
6. By using OK key, enter to
choice Program 1)
“MANUEL CONTROL” menu.
4. By using OK key, enter to “P1
PROGRAM MENU”. 7. By using / keys, start “Wash
with Dis.” Position.
5. By using / keys, go to
“Manuel Control” position. 8. By using OK key, start the
operation.
55
* Manuel alcohol filling (For the Model with Alcohol Tank)
3. By using / keys, go to
“Program 1” position.(İf your choice
6.By using OK key,enter to
Program 1)
“MANUEL KONTROL” menu.
4.By using OK key, enter to “P1
PROGRAM MENU”. 7.By using / keys,go to “Pump
alcohol” position.
56
* Manuel air pumping
3. By using / keys, go to
“Program 1” position. (İf your
choice is Program 1)
6. By using OK key, enter to
4. By using OK key, enter to “P1 “MANUEL CONTROL” menu.
PROGRAM MENU”.
7. By using / keys, go to “Pump
Air” position.
5. By using / keys,“Manuel
Control” position. 8. By using OK key, start the
operation.
57
* Manuel filtration
3. By using / keys, go to
“Program 1” position. (If your
6. By using OK key, enter to
choice is Program1)
“MANUAL CONTROL” menu.
4. By using OK key, enter to “P1
PROGRAM MENU”. 7. By using / keys, go to
“Filtration” position.
58
* Manuel sink emptying
3. By using / keys, go to
“Program 1” position. (If your
6. By using OK key, enter to
choice is Program1)
“MANUAL CONTROL” menu.
4. By using OK key, enter to “P1
PROGRAM MENU”. 7. By using / keys, go to
“Empty sink” position.
5. By using / keys, go to
“Manual Control” position. 8. By using OK key, start the
operation.
59
* Manuel sewage disposal
3. By using / keys, go to
“Program 1” position. (If your
choice is Program1)
6. By using OK key, enter to
4. By using OK key, enter to “P1 “MANUAL CONTROL” menu.
PROGRAM MENU”.
7. By using / keys, “Empty the
sewage Tank” position.
8.
5. By using / keys, go to
“Manual Control” position. 9. By using OK key, start the
operation
60
Settings
• Time settings
7. By using / keys, go to
“Time Settings” position.
5. By usng / keys, go to
“Settings” position.
61
8. By using OK key, enter to
9. By using / keys, set the time
“Time Settings” menu.
between “00:00:00” and “23:59:59”
in the form of hour:minute:second
(hh:mm:ss).
Note: Check your device's clock regularly, and your device will not work if your
clock does not work
62
• Date settings
5. By using / keys, go to
“Settings” position.
8. By using OK key, enter to
“Date Settings” menu.
63
9. By using / keys, set the date
between “01/01/18” and
“31/12/99” in the form of
day/month/year (dd/mm/yy).
Note: The date setting of your device must be up to date to follow the disinfectant
usage period.
64
• Personnel settings
65
10. By using / keys go to
“Personnel 1” or other positions. 12. By using / keys, enter the
first letter of the name of your
11. By using OK key, enter to personnel.
“Personnel 1” menu. (If your 13. By using OK key, progress to
choice is Personnel 1) the next letter. After the last letter,
save the name by using OK
key.
66
• Device settings
7. By using / keys, go to
“Device Setting” position.
5. By using / keys, go to
“Settings” position.
67
8. By using OK key, enter to
“HOSPITAL NAME” menu.
68
• Auto disinfection
In order to clean and disinfect the all store and channels in your device periodically, follow
the steps below:
69
• Emptying the Tanks
This operation empties all the tanks (except the Concentrate Disinfectant tank) in your
device.
70
HOW TO USE LEAK TEST PROGRAM
71
Concentrated Disinfection, Enzymatic - You can empty the Concentrated or
and Alcohol Emptying (Manuel) Disposable disinfectant from the indicated
tap with manually.
You can empty the disposable (Con.)
disinfectant which is expired in your
machine as follows;
You must use a large container to empty the After the emptying operation, remove the
ready-to-use disinfectant from tanks I and II. hose and wash it and store it for further
(at least 15 liter) processing.
72
You should take the necessary -Internal and External Filters condition
precautions to protect yourself during (replace/continue)
emptying because the chemicals are -Top sink filter (clean/ replace)
harmful to the breathing. -Float sensor on top sink (clean) – this
sensor should be replaced after 2600 cycles
PPM procedure for All Endowash in case of using glutaraldehyde based
Models: disinfection solutions, and 3300 cycles for
acidic base disinfectant ( acetic acid)
1-During installation: -Sink pumps (check condition and sound) -
4200 working Hrs./ or damage
Please check all hoses are connected firmly -Solenoids, check functionality during one
to the appropriate outlet /inlet, water supply cycle, no over heat or noise sound during
should be around 1.5 bar, drain hose should operation, solenoids should be regularly
be firmly fixed to the drain sink of the checked and replaced when damage/ or
hospital to avoid cross infection! more than 3600 strokes achieved
Check plug earthling to avoid leakage -Check tanks for apparent damage
current, check current and voltage rating for -Check control panels and switches, replace
the specified machine. if necessary
73
-Outer surface free -Leakage test MUST be checked with
-Top cover scratches free scope, check rubber inflation, replace
-Filters condition (replace/continue) connector if necessary
-Top sink filter (clean/ replace)
-Float sensor on top sink (clean) – this
sensor should be replaced each 2600 cycles
in case of using glutaraldehyde-based
disinfection solutions, and 3300 cycles for
acidic base disinfectant (acetic acid)
-Sink pumps (check condition and sound) -
4200 working Hrs./ or damage
-Solenoids, check functionality during one
cycle, no over heat or noise sound during
operation, solenoids should be regularly
checked and replaced when damage/ or
more than 3600 strokes achieved
-Check tanks for apparent damage
-Check control panels and switches, replace
if necessary
-Check scope inlets connectors (suction/
biopsy , air/water and leakage test adaptor)
74
Manufacturer: