B&G Hydra 2000
B&G Hydra 2000
USER MANUAL
Premier Way, Abbey Park
Romsey, Hampshire, S051 9AQ
England
HB-0844-02
Hydra 2000 User Manual
USER MANUAL
CONTENTS
Record of Amendments
PART 1 - INTRODUCTION
PART 3 - CALIBRATION
PART 5 - OPTIONS
HB-0844-02
Hydra 2000 User Manual
PRODUCT LIABILITY
Brookes and Gatehouse Limited accept no responsibility
for the use and/or operation of this equipment. It is the
user’s responsibility to ensure that under all circumstances
the equipment is used for the purposes for which it has
been designed.
WARNING - CALIBRATION
The safe operation of this equipment is dependent on
accurate and correct calibration. Incorrect calibration of
this equipment may lead to false and inaccurate
navigational readings placing the yacht into danger.
CAUTION
This equipment is designed for use with a power supply
source of 12v dc. The application of any other power supply
may result in permanent damage to the equipment.
HB-0844-02
Hydra 2000 User Manual
RECORD OF AMENDMENTS
HB-0844-02
Hydra 2000 User Manual
PART 1 - INTRODUCTION
CONTENTS
Para Page
ILLUSTRATIONS
Fig No Page
TABLES
Table No Page
HB-0844-02
Hydra 2000 User Manual
20/20
20/20
Mast
Halcyon Compass
Display FFD
Halcyon 000
Deck
GPS Plus FFD
000
Chart Table
NMEA FFD
0 00
Hydra 2000
Main Processor
EP
Network
Hull
Halcyon 2000
Compass Boat
Depth Speed
HB-0844-02
Hydra 2000 User Manual
PART 1 - INTRODUCTION
1.2 PROCESSORS
HB-0844-02
Hydra 2000 User Manual
Leeway
Heading Corrected for Leeway (Course)
True Wind Speed
True Wind Angle
True Wind Direction
Heading on Next Tack or Gybe
Pressure Trend
It also contains the battery backed memory that stores all the
calibration, damping and alarm settings whilst the power is OFF.
These are adjustable from any FFD.
1.3 SENSORS
The Masthead Unit senses apparent wind speed and wind angle.
The unit is light weight and sealed.
The Sonic Speed Unit provides highly accurate and stable boat
speed measurement. It does not rely on mechanical moving parts
that require constant attention to protect them from weed, etc. The
transducers are fitted virtually flush with the hull, creating almost
zero drag, and can be painted, faired or anti-fouled.
HB-0844-02
Hydra 2000 User Manual
The Halcyon 2000 Compass has the ability to ‘learn’ the magnetic
effect of the vessel on the compass and automatically apply
deviation correction.
1.4 DISPLAYS
This is the standard system display and every Hydra 2000 must
contain one NMEA FFD. The Standard FFDs and the other
display types, described in this Section, are options with which the
system can be expanded and made even more powerful.
The FFD's name gives an indication to the fact that they are really
much more than just a display. It is a terminal for the whole
system, allowing you to control everything from the functions
displayed on them (and at the 20/20’s) to the calibration of the
system.
HB-0844-02
Hydra 2000 User Manual
HB-0844-02
Hydra 2000 User Manual
1.5 CALIBRATION
1.6 DAMPING
Similarly the lower it is the bigger the jumps you will get in the
numbers but the response to any change will be quicker -
remember, high damping for rough weather, low damping for calm
weather.
Damping should not be confused with the update rate which is the
number of times each second that the function value is sent to the
display. It is fixed for all the functions.
HB-0844-02
Hydra 2000 User Manual
Note
The Functions available to the user are dependent on the range of
sensors fitted to the system.
HB-0844-02
Hydra 2000 User Manual
HB-0844-02
Hydra 2000 User Manual
HB-0844-02
Hydra 2000 User Manual
HB-0844-02
Hydra 2000 User Manual
HB-0844-02
Hydra 2000 User Manual
Part 2 - Operating Information
CONTENTS
Para Page
2.1 FULL FUNCTION DISPLAY (FFD) 2-4
2.1.1 The Keys 2-5
2.1.2 Power/Lights Key 2-5
2.1.3 Page Key 2-5
2.1.4 Scroll Keys 2-6
2.1.5 Enter/Reset Key 2-6
2.1.6 Speed/Depth (SPD/DEP) Key 2-7
2.1.7 Wind Key (WIND) 2-7
2.1.8 Navigation (NAV) Key 2-7
2.1.9 Remote Button Operation 2-8
HB-0844-02
2-1
Hydra 2000 User Manual
Part 2 - Operating Information
HB-0844-02
2-2
Hydra 2000 User Manual
Part 2 - Operating Information
CONTENTS (Contd.)
Para Page
ILLUSTRATIONS
Fig No Page
HB-0844-02
2-3
Hydra 2000 User Manual
Part 2 - Operating Information
The Hydra 2000 System is easily operated using the keys on any
one of the NMEA or Standard Full Function Displays (FFD).
HB-0844-02
2-4
Hydra 2000 User Manual
Part 2 - Operating Information
Power/Lights Key
Page Key
Scroll Up Key
Scroll Down Key
Enter/Reset Key
Speed/Depth Key
Wind Key
Navigation Key
This key controls the application of power to the system and the
level of illumination at all displays.
One short press of this key applies power to the Hydra System
and the display is activated. A second short press of the key
provides full background illumination on all system displays.
Further short presses of the key decrease the illumination in three
stages from full brightness to OFF.
The next press gives full illumination again. This operation on any
one FFD within the system, performs the same sequence on all
Hydra displays connected to the system. However, display lighting
can be localised so that the level can be adjusted for individual
displays (see Para 2.9 - LIGHTING CONTROL).
To switch OFF the system, press and hold down the key for three
seconds. After this time lapses the message POWER-OFF
appears in the upper text, release the key and after a further two
seconds the system switches OFF and is now in Standby Mode.
This key enables the user to select the User Defined Page
(Display) presented on the FFD.
By pressing this key the user is able to choose any one of four
user defined pages accessing various functions of the Hydra
2000 System (2 functions per page).
HB-0844-02
2-5
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. If you become lost in the system, press the Page Key and you
will immediately return to the top level display without doing any
damage.
2. Successive operations of the Page Key will bring up each of
the user defined pages in rotation.
3. The four initial pages can be reconfigured to meet your specific
needs, as described in Para 2.5.2
4. Holding down the Page Key for two seconds initiates control of
20/20 displays (see Part 5 - Options).
Two scroll keys are provided: Scroll Up and Scroll Down. The
scroll keys have three functions:
HB-0844-02
2-6
Hydra 2000 User Manual
Part 2 - Operating Information
By pressing the SPD/DEP Key the user is able to select any one
of three factory set pages accessing the Speed and Depth
functions. Successive operation of the SPD/DEP Key will display
the following information in a fixed order:
Boat Speed/Depth
Boat Speed/Speed Over Ground (from NMEA Position Fix)
Boat Speed/Apparent Wind Angle
By pressing the Wind Key the user is able to select any one of
three factory set pages accessing the Wind Functions.
Successive presses of the Wind Key will display the following
information in a fixed order:
By pressing the NAV Key the user is able to select any one of
three factory set displays accessing the Navigation Functions.
Successive presses of the NAV Key will display the following
information in a fixed order:
HB-0844-02
2-7
Hydra 2000 User Manual
Part 2 - Operating Information
Note
The NAV key may be configured to show waypoint information in
either Great Circle or Rhumb modes. Refer to section 2.5.3 for
further details.
2.1.9 Remote Button Operation
Each FFD may be provided with an optional Remote Button to
allow basic control of the associated FFD from a remote position.
The Remote Button has two Functions: Page Selection and Fast
Reset.
Page Selection is achieved by successive pressing of the
Remote Push-Button. When the button is pressed each of the
four user defined pages followed by the first page of the factory
set Speed/Depth, Navigate and Wind pages are displayed in a
fixed order.
When any one of the trip functions is selected at the associated
FFD, pressing and holding down the Remote Push-Button for two
seconds will change the display to show RESET flashing, or if
TIMER is selected START 10 flashing. Pressing the Remote
Button again will Reset that function, or in the case of TIMER start
the countdown. After a reset the display will revert to normal page
set-up automatically.
2.2 20/20 DISPLAY
The 20/20 is a fully programmable, single function display,
incorporating large digits, which may be installed virtually
anywhere in the yacht.
Control of the 20/20 Display may be achieved via any FFD or from
an optional Remote Button, see Part 5 - Options.
2.3 SUPER HALCYON 3 AND HALCYON 2000 COMPASS
The Super Halcyon 3 and Halcyon 2000 Compasses are high
performance electronic transmitting compasses which interface
with your Hydra 2000 System. Heading and direction information
may be selected for display on any of the FFDs or on a Halcyon
Display. Additionally, analogue compass card displays may also
be added.
Software incorporated within the compass allows the unit to learn
the magnetic fields in the vessel that may cause deviation errors.
HB-0844-02
2-8
Hydra 2000 User Manual
Part 2 - Operating Information
HB-0844-02
2-9
Hydra 2000 User Manual
Part 2 - Operating Information
At this stage we have simply called up this function to view it, if the
Page Key is pressed then the configured page will return and
Stored Log will no longer be on view. If however you wish to keep
Stored Log on a page, then you can configure the page.
The Page Key allows the user to configure four pages per FFD
depending on its required use at that position.
Note
Scroll Up or Scroll Down can be used because we are
configuring the whole page, both upper and lower displays.
Our second example is configuring the NAV key. The NAV key
allows the user to select either Rhumb Line or Great Circle
navigation information to be displayed. To select the required
mode, proceed as follows:
HB-0844-02
2-10
Hydra 2000 User Manual
Part 2 - Operating Information
Damping control for any of the other functions that can be damped
is achieved in a similar manner.
HB-0844-02
2-11
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. SINGLE is the choice required if a single paddlewheel or sonic
sensor is fitted.
2. If two paddle-wheels or two sonic sensors are fitted, the Scroll
Up or Scroll Down Keys should be used to select PORT CAL
or STBD CAL as required.
HB-0844-02
2-12
Hydra 2000 User Manual
Part 2 - Operating Information
Note
If DATUM is referenced to the water line, the value is positive. If
DATUM is referenced to the keel line, the value is negative and
this is indicated by a minus sign before the left digit.
(7) Press Enter to input the new DATUM value into the
system.
(4) Press Enter and the current offset angle will flash.
Use Scroll Up or Scroll Down to set the new value.
HB-0844-02
2-13
Hydra 2000 User Manual
Part 2 - Operating Information
2.7 ALARMS
Example
HB-0844-02
2-14
Hydra 2000 User Manual
Part 2 - Operating Information
For example, when the SECTOR alarm is turned on, the alarm
reference heading is the current compass heading. If the
SECTOR alarm is set at 40° the sector value is the compass
heading +/- 20°. It is therefore important to switch OFF the
SECTOR alarm before carrying out a course alteration and
switching the alarm on again when settled on the new course
heading.
(3) Press Enter and the text shows ALL OFF, flashing.
HB-0844-02
2-15
Hydra 2000 User Manual
Part 2 - Operating Information
(3) Press Enter again, all alarms are turned OFF and
the normal page display will be restored.
WARNING - All alarms in the system will remain OFF until the
values are reset or the individual alarms are switched ON
again.
When any trip function is started, all other trip functions that have
been reset start with it, except when the timer countdown is
started. Under this condition, the other functions start again if
previously reset, when the countdown reaches zero. This is
designed for the start of races, so that you have DR, Resettable
Log and Timer running, without pressing buttons during the
moments of the start itself.
HB-0844-02
2-16
Hydra 2000 User Manual
Part 2 - Operating Information
The Hydra 2000 System incorporates a Fast Reset feature for all
trip functions (Trip Log, Timer and Barometric Pressure Trend).
Fast Reset can be initiated either by operation of a Remote Push-
Button (see Para 2.1.9) or by operation of the Enter Key on any
FFD.
Note
If two trip functions are displayed at the same time, the Fast Reset
facility will only reset the function shown in the upper display.
HB-0844-02
2-17
Hydra 2000 User Manual
Part 2 - Operating Information
(2) Press Scroll Down once and the lower text now
shows CONTROL flashing.
(4) Press the Enter key and the display now shows
RUN flashing.
(5) Press the Enter key again and the display shows the
trip log running.
Note
Every time you reset the TRIP LOG the AVERAGE SPEED
resets and begins its calculation again. The control facility for the
Dead Reckoned functions work in the same way; the bearing and
distance are linked, so that they are started simultaneously.
HB-0844-02
2-18
Hydra 2000 User Manual
Part 2 - Operating Information
Note
In order to control the lighting on analogue repeaters, at least one
FFD must be set to System Control.
HB-0844-02
2-19
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Requires a Temperature Sensor.
2. Audible high/low alarm available.
Notes
1. Alignment calibration.
2. Variable damping 0-99 seconds.
3. Analogue indicator available.
4. Audible sector alarm available
HB-0844-02
2-20
Hydra 2000 User Manual
Part 2 - Operating Information
The apparent wind angle is the angle between the boat's bow and
the wind blowing across deck (at mast height). It is a basic
number, useful in its own right like Boat Speed, Heading and
Apparent Wind Speed and provides the starting point for the
higher level, calculated functions.
Notes
1. Variable damping 0-99 seconds.
2. Audible, high/low alarm available.
3. Analogue indicator available.
Trip Log
HB-0844-02
2-21
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Requires pressure sensor.
2. Offset calibration available.
This is essential for the Offshore Sailor, giving not only the
instantaneous barometric value but also the important trends
towards higher or lower pressure. This helps position you in a
weather system and so predict the next change.
This shows the change in pressure over a period of time that can
be set for up to 24 hours. The period can be changed by
selecting a control option in the OPERATION Menu using the
following sequence:
HB-0844-02
2-22
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Audible, high/low alarm available.
2. Calibration.
This function measures the voltage that the batteries supply to the
system. This is especially useful for monitoring the supply voltage
offshore enabling you to optimise your engine running time.
Notes
1. Requires interfaced position fixing system.
2. Rhumb, Great Circle, Magnetic or True.
HB-0844-02
2-23
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Requires interfaced position fixing system.
2. This is the bearing from a waypoint to the next waypoint in the
position fixer’s route.
Notes
1. Port, starboard calibration with automatic facility.
2. Variable damping 0-99 seconds.
3. Acceleration, deceleration indicator.
4. Analogue indicators available.
5. Audible, high/low alarm available.
Boat speed is shown on the display with a small bar to the left of
the digits that indicates the increase or decrease of speed. The
amount of damping that is applied can be adjusted. This can be
increased to smooth out the display value in rough conditions or
conversely reduced to speed up the response in light conditions.
2.11.11 Course
HB-0844-02
2-24
Hydra 2000 User Manual
Part 2 - Operating Information
Note
Requires heel sensor.
Note
Requires interfaced position fixer.
Notes
1. Course and distance.
2. Corrected for leeway if heel sensor fitted.
3. Resettable as a trip function.
HB-0844-02
2-25
Hydra 2000 User Manual
Part 2 - Operating Information
HB-0844-02
2-26
Hydra 2000 User Manual
Part 2 - Operating Information
2.11.14 Depth
Menu Heading: DEPTH
Function Text: DEPTH
Update Rate: Once per second
Units: Metres, feet and fathoms
Notes
1. Audible, shallow/deep alarm available:
Shallow range 0-99.9m
Deep range unlimited
2. Analogue indicators available.
Depth is one of the most important functions on the boat, being an
essential navigational and safety tool. A datum adjustment allows
you to move the base point to give either depth under the keel or
actual water depth (see Para 2.6.2).
Depth sounder performance is dependent on many factors:
transducer type and installation, boat speed, electrical noise, sea
state, sea bed conditions, air and plankton in the water. But
however good the equipment, there will always be times when a
reliable measurement of depth is not possible. In these cases,
such as following in the wake of another boat, the display shows:
HB-0844-02
2-27
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Adjustable for horizontal sensor alignment.
2. Requires clinometer mounted fore and aft.
3. Variable damping 0 - 99 seconds.
Fore and Aft trim is an often ignored factor in Boat Speed, but it is
as important to trim the boat in this plane and to have a record of
the trim angles for various conditions.
2.11.17 Heading
Notes
1. Adjustable damping, 0-99 sec.
2. Sector alarm available.
3. Sensor alignment calibration.
4. Moving card analogue meter available.
5. Halcyon display available.
HB-0844-02
2-28
Hydra 2000 User Manual
Part 2 - Operating Information
This gives the heading the vessel would be sailing when on the
opposite tack with the same wind angle, and can be used in
conjunction with a hand bearing compass to tell you when you are
on the layline. It is calculated from Heading and True Wind angle.
Notes
1. Requires Halcyon Display.
2. Analogue indicator available.
Notes
1. Adjustable for horizontal sensor alignment.
2. Requires clinometer.
3. Variable damping 0 - 99 seconds.
HB-0844-02
2-29
Hydra 2000 User Manual
Part 2 - Operating Information
2.11.21 Leeway
Note
Requires clinometer for heel.
Leeway is the angle between the boats heading and its course
through the water. The difference is caused by the sideways slip
that the boat has when going upwind. For the Hydra to be able to
measure this it needs to know the heel angle, and hence a
clinometer must be fitted. Leeway is of great importance in the
calculation of Dead Reckoning, since leeway can considerably
affect the dead reckoned position.
Note
Requires NMEA 0183 interfaced position fixing system
transmitting the ZDL sentence.
HB-0844-02
2-30
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Requires NMEA 0183 interfaced position fixing system
transmitting the ZLZ sentence.
2. Check that the Local Time Offset is entered into the position
fixer correctly.
The function normally shows the Local Time of Day in hours and
minutes as given by the position fixer. To reveal minutes and
seconds press the Enter Key once. A further press of the Enter
Key returns the display to hours and minutes.
Notes
1. Requires mast rotation sensor.
2. Offset calibration available.
Note
Requires NMEA 0183 interfaced position fixing system
transmitting the WDC sentence.
HB-0844-02
2-31
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Requires Halcyon Display.
2. Analogue Indicator available.
Note
Requires forestay loadcell.
HB-0844-02
2-32
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Requires rudder angle sensor.
2. Offset calibration available.
This function is used to indicate to the Trimmers how well the boat
is balanced.
Notes
1. Requires temperature sensor.
2. Audible high/low alarm available.
Note
Requires interfaced position fixing system.
HB-0844-02
2-33
Hydra 2000 User Manual
Part 2 - Operating Information
This also comes straight from the position fixing system, and is
available, provided it is sent, when the position fixer is connected.
Comparing the speed over ground to the speed you are doing
through the water is a key tactical tool. This is particularly so in
strong tidal waters and at night, when the only indication that the
tide has changed and you are no longer making progress is the
speed over ground.
2.11.31 Stored Log
HB-0844-02
2-34
Hydra 2000 User Manual
Part 2 - Operating Information
Note
If your position fixer sends magnetic bearing, check that the
variation is correctly entered into it.
2.11.33 Timer
Note
Individually resettable.
Pressing the Enter Key will toggle the display between min/secs
and hrs/mins as indicated on the display.
HB-0844-02
2-35
Hydra 2000 User Manual
Part 2 - Operating Information
Note
Requires NMEA 0183 interfaced position fixing system
transmitting the ZDL sentence.
Note
Note
Individually resettable.
HB-0844-02
2-36
Hydra 2000 User Manual
Part 2 - Operating Information
This is the resettable log for recording trip distance and reads
from the moment it is started, in nautical miles. It must be
remembered that this is the distance sailed through the water, not
over the ground. It also forms part of the calculation for Average
Speed. A useful feature is that if it is reset prior to the start of a
race, then when the Timer counts down to zero, the trip log (and
any other trip functions that have been reset) start automatically.
Notes
1. Relative to the boat’s heading.
2. Corrected for masthead and other errors via the Look-up Table
(see Part 3 - Calibration).
3. Variable damping 0-99 seconds.
The true wind is calculated from the vector triangle shown in Fig.
2.4. This uses the apparent wind speed, apparent wind angle and
the Boat Speed in the calculation. The results are then corrected
by the true wind correction tables, which are detailed in Part 3 -
Calibration.
Note
The true wind is the wind relative to the water, and is not the same
as the ground wind, unless there is zero tide or current. The true
wind angle is the angle between the boat's heading and the true
wind.
HB-0844-02
2-37
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Wind angle relative to magnetic compass point.
2. Corrected for masthead and other errors via a Look-Up Table
(see Part 3 - Calibration).
3. Damping control via true wind speed.
This is the tactician's greatest ally in the search for the right wind
shifts. It shows the compass direction of the wind regardless of
the boat's heading or speed. True wind direction is calculated
from the true wind angle and heading, and is corrected for
calibration errors by the true wind look-up table. A full explanation
is given in Part 3 - Calibration.
HB-0844-02
2-38
Hydra 2000 User Manual
Part 2 - Operating Information
Notes
1. Corrected for masthead and other errors via a Look-Up Table
(see Part 3 - Calibration).
2. Variable damping 0-99 seconds.
When the boat is sailing downwind, the air passing over the mast
is accelerated and tends to make the true wind speed over-read
downwind. The Hydra has a calibration to correct for this which is
explained in Part 3 - Calibration.
Note
Upwind and Downwind.
Velocity Made Good (VMG) is calculated from the true wind angle
and the Boat Speed as shown in Fig 2.5. VMG is a measure of
your performance upwind and downwind and is more important
than Boat Speed, since it takes into account how close you are
sailing to the wind. However, it is not possible for the helmsman to
sail to it directly because of the momentum of the boat.
As you sail closer to the wind the VMG will initially rise, because
the boat will hold its speed due to its momentum. So VMG
increases and the helmsman, seeing this, would be encouraged
to sail even higher to increase the VMG even more. Of course if
you carry on like this you will eventually end up head to wind, the
boat will stop dead and the VMG will plummet.
HB-0844-02
2-39
Hydra 2000 User Manual
Part 2 - Operating Information
Note
Requires mast rotation sensor if a rotating mast is installed.
HB-0844-02
2-40
Hydra 2000 User Manual
Part 3 - Calibration
PART 3 - CALIBRATION
CONTENTS
Para Page
3.1 INTRODUCTION 3-3
3.2 BOAT SPEED/LOG CALIBRATION 3-4
3.2.1 Principle of Log Calibration 3-4
3.2.2 Preparation for Log Calibration 3-4
3.2.3 Calibration Runs 3-4
3.2.4 Log AUTO CAL Facility 3-5
3.2.5 Log AUTO CAL Procedure 3-6
3.2.6 Manual Calibration 3-7
3.2.7 Boat Speed Calibration (ref. to a Known Value) 3-7
3.2.8 Boat Speed Calibration (Knots, MPH, KPH) 3-8
3.3 APPARENT WIND SPEED/ANGLE
CALIBRATION 3-9
3.3.1 Principles of Wind Speed/Angle Calibration 3-9
3.3.2 Apparent Wind Angle Calibration (AWA) 3-10
3.3.3 Apparent Wind Speed Calibration 3-11
3.4 COMPASS CALIBRATION 3-12
3.4.1 Principles of Compass Calibration
(Super Halcyon 3 & Halcyon 2000 Compasses) 3-12
3.4.2 Heading Node Selection 3-13
3.4.3 Super Halcyon 3 Compass Calibration Procedure3-14
3.4.4 Halcyon 2000 Calibration Procedure 3-15
3.5 HEEL ANGLE/LEEWAY CALIBRATION 3-16
3.5.1 Heel Angle Calibration 3-16
3.5.2 Leeway Calibration 3-17
3.6 TRUE WIND CORRECTION 3-18
3.6.1 True Wind Direction Calibration 3-18
3.6.2 True Wind Speed Calibration 3-21
3.7 DEPTH CALIBRATION 3-22
3.8 BATTERY VOLTS CALIBRATION 3-23
3.9 SEA TEMPERATURE CALIBRATION 3-23
3.9.1 Sea Temperature Offset Calibration 3-23
HB-0844-02
3-1
Hydra 2000 User Manual
Part 3 - Calibration
CONTENTS (Contd.)
ILLUSTRATIONS
Fig No Page
TABLES
Table No Page
HB-0844-02
3-2
Hydra 2000 User Manual
Part 3 - Calibration
PART 3 - CALIBRATION
3.1 INTRODUCTION
There are four basic inputs to your system that are fundamental to
its integrated approach - Boat Speed, Compass Heading,
Apparent Wind Angle and Apparent Wind Speed.
HB-0844-02
3-3
Hydra 2000 User Manual
Part 3 - Calibration
HB-0844-02
3-4
Hydra 2000 User Manual
Part 3 - Calibration
This facility enables the user to calibrate the yacht's log accurately
and simply as all calculations are performed internally by the Main
Processor. Referring to the Fig 3.1, A and B are the two markers
for each run and X is the actual distance for each run as
ascertained from the chart.
Note
The calibration process can be cancelled at any time during the
operation by pressing the Page Key if the operator is not satisfied
with the calibration runs underway, e.g., hampered by another
vessel or wash, etc.
HB-0844-02
3-5
Hydra 2000 User Manual
Part 3 - Calibration
HB-0844-02
3-6
Hydra 2000 User Manual
Part 3 - Calibration
The yacht's log is now calibrated and the new calibration value is
stored permanently in the computer’s memory.
Note
“ERR” will be displayed if the system encountered a problem
during the calibration run. For example, if the calibration distance
was too short.
3.2.6 Manual Calibration
The calibration values can be adjusted directly as shown in
'Examples of Calibration’ included in Part 2 - Operating
Information.
3.2.7 Boat Speed Calibration (ref. to a Known Value)
The following enhancement is only available when the Hydra
Depth Unit is used in conjunction with the Hydra 2000 FFD Display
version later than 5C.
To calibrate the Boat Speed by reference to a known value, e.g.,
another boat with an accurately calibrated log proceed as follows:
Select BOAT SPD KT on the upper half of the FFD Display.
(1) Press Scroll Down until the lower text shows
CALBRATE flashing.
(2) Press Enter, the lower text now shows AUTO CAL
flashing.
(3) Press Scroll Down until the lower text shows REF
CAL flashing.
(4) Press Enter, the lower text now shows REF CAL
along with current Boat Speed.
(5) Press Enter, the lower text now shows REF CAL
flashing and by use of the Scroll Up/Down the
reference speed may be adjusted.
(6) Press Enter, the new speed value is accepted and
the upper display will show the re-calibrated boat speed.
HB-0844-02
3-7
Hydra 2000 User Manual
Part 3 - Calibration
HB-0844-02
3-8
Hydra 2000 User Manual
Part 3 - Calibration
How do you know the shear is minimal? If you are finding it easier
to attain speed on one tack than the other, then there is shear. A
good look at the general weather conditions is also helpful. Do not
calibrate in building sea breezes!
What of wind gradient then? Well, this is the biggest culprit for
getting true wind speeds accused of gross inaccuracy. The
problem is that most people use the wind speed as a measure of
the pressure or force of the wind, which it is not. It is a measure of
the wind speed at the top of the mast, and that is all. If it is 12kts at
the top of the mast and only 4kts at the water, then the breeze will
feel a lot softer and provide less power for the rig than if the
breeze is twelve knots all the way down to the water.
This is why we recommend that the last thing you touch is the
Apparent Wind Speed. It is calibrated in the factory where wind
tunnel calibrated units are available, and apparent inaccuracies are
more likely attributable to effects such as wind gradient, rather
than to a basic calibration problem.
HB-0844-02
3-9
Hydra 2000 User Manual
Part 3 - Calibration
(3) Tack the vessel and sail at the optimum close haul
position as before.
HB-0844-02
3-10
Hydra 2000 User Manual
Part 3 - Calibration
and
HB-0844-02
3-11
Hydra 2000 User Manual
Part 3 - Calibration
(c) The rate of change of heading must not fall below 0.2
of a degree per second during the 540° turn - Super
Halcyon 3, i.e. the 540° turn must not take longer than 45
minutes.
HB-0844-02
3-12
Hydra 2000 User Manual
Part 3 - Calibration
The Hydra 2000 System can accept heading data from a variety
of different sources. These different sources are known as
Address Nodes and allow the system to identify which heading
devices are connected to the system. The list below shows the
various sources of heading available with its respective address
node:
Device Node
Super Halcyon 3 Compass 5
Halcyon 2000 Compass 16
Type PLC Pilot 17
Type ACP Pilot 18
NMEA Heading (NMEA FFD) 96
(c) Press the Scroll Up key again until the display shows
COURSE °M flashing.
HB-0844-02
3-13
Hydra 2000 User Manual
Part 3 - Calibration
Notes
(1) Hydra Pilots, Hercules Pilots, HS Pilots and Halcyon FFDs will
also require the Heading Node to be set to your desired choice.
Refer to the relevant user manual for further information.
(2) 20/20 displays will require Heading to be re-selected. Simply
re-select this function and refer to section 5.6.5 for more
information.
(3) Check for any large steel structures nearby that may
cause additional erratic deviations.
Note
If the displayed heading does not reverse, the calibration was
discarded because the conditions were not met. You must start
again, or try another day when wind and wave conditions are
calmer.
HB-0844-02
3-14
Hydra 2000 User Manual
Part 3 - Calibration
(3) Check for any large steel structures nearby, that may
cause additional, erratic deviations.
(7) Press the Enter key once and the ‘0’ starts to flash.
Use the Scroll Up key to change the value to ‘1’.
(8) Press the Enter key. The display now shows 000°.
HB-0844-02
3-15
Hydra 2000 User Manual
Part 3 - Calibration
Notes
1. The compass calibration swing may be aborted at any time.
To do this, press the Enter key once. Next, using the Scroll
Down key change the value from ‘1’ to ‘0’, and then press Enter
to store. The display will now show OFF.
2. The first time the system is switched on, or after a system
reset, the Heading will alternate with CAL. This is to indicate to the
user that the compass must be calibrated. This will disappear after
the compass has been calibrated.
HB-0844-02
3-16
Hydra 2000 User Manual
Part 3 - Calibration
L =KxH
Bs x Bs
HB-0844-02
3-17
Hydra 2000 User Manual
Part 3 - Calibration
The idea is to sail on a steady course and drop markers over the
stern at regular intervals, the angle between them and the
centreline of the yacht is measured with a hand-bearing compass
and hence leeway angle is measured. Whilst this is happening the
boat speed and heel angle should be noted at intervals and an
average calculated. These values can then be used to calculate
the leeway coefficient from the following expression:
K = L x Bs x Bs
H
Once the leeway coefficient 'K' is known it is entered into the
system under:
NAVIGATE → LEEWAY, CALBRATE → CAL VAL1
HB-0844-02
3-18
Hydra 2000 User Manual
Part 3 - Calibration
The reason for this is a variety of errors when calculating the true
wind angle from the apparent wind angle, one of the largest of
which is Upwash, an aerodynamic effect which is shown in Fig 3.5
The problem stems from the true wind direction 'tacking' as the
boat manoeuvres from tack to tack. We need to know the error
that the true wind suffers in any manoeuvre, be it a normal upwind
tack, a reach to reach tack, or a downwind gybe.
HB-0844-02
3-19
Hydra 2000 User Manual
Part 3 - Calibration
The true wind direction should read 205° on both tacks. To correct
the true wind angle so that the true wind direction reads 205° on
both tacks, 5° should be added to the true wind angle. As a
general rule it can be said:
HB-0844-02
3-20
Hydra 2000 User Manual
Part 3 - Calibration
HB-0844-02
3-21
Hydra 2000 User Manual
Part 3 - Calibration
HB-0844-02
3-22
Hydra 2000 User Manual
Part 3 - Calibration
HB-0844-02
3-23
Hydra 2000 User Manual
Part 3 - Calibration
(4) Press Scroll Down the lower text now shows CAL
VAL 2 flashing.
HB-0844-02
3-24
Hydra 2000 User Manual
Part 4 - Installation Information
CONTENTS
Para Page
HB-0844-02
4-1
Hydra 2000 User Manual
Part 4 - Installation Information
4.1 INTRODUCTION
Note
All cable screens should be connected under the metal clamp bar
located on the processor case.
HB-0844-02
4-2
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-3
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-4
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-5
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-6
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-7
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-8
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-9
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-10
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-11
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-12
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-13
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-14
Hydra 2000 User Manual
Part 4 - Installation Information
HB-0844-02
4-15
Hydra 2000 User Manual
Part 4 - Installation Information
INSTALLATION NOTES
• An NMEA FFD can be connected at any point on the system network. Typically an
NMEA FFD is mounted close to the NMEA device; for example at the chart table next to
the GPS, thus minimising the NMEA cable routes.
• NMEA FFD's can be used in combination with all other display types; that is standard
FFD's, Halcyon displays, 20/20 and Pilot displays.
SCREENED CABLES
• Shielded NMEA cables are supplied to provide protection against unwanted emissions
(EMC) and must be connected in accordance with these instructions.
• NMEA cable shields must be connected at the TRANSMITTING end.
• The NMEA output cable shield should be connected to the other cable shields in the
junction box supplied.
NETWORK TERMINATION
If the NMEA FFD is the last unit on the system network a network terminator MUST be fitted
across the network data wires, that is between the green and white wires. Only two network
terminators are required per system.
HB-0844-02
4-16
Hydra 2000 User Manual
Part 4 - Installation Information
SCREENED CABLES
• Shielded cables are supplied to provide protection against unwanted emissions (EMC) and
must be connected in accordance with these instructions.
NETWORK TERMINATION
• If the FFD is the last unit on the system network a network terminator MUST be fitted across
the network data wires, that is between the green and white wires.
HB-0844-02
4-17
Hydra 2000 User Manual
Part 5 - Options
PART 5 - OPTIONS
CONTENTS
Para Page
5.1 SYSTEM EXPANSION 5-3
5.2 SENSORS 5-3
5.2.1 Trim Angle Sensor 5-3
5.2.2 Mast Rotation Sensor 5-3
5.2.3 Heel Angle Sensor 5-3
5.2.4 Barometric Pressure Sensor 5-3
5.2.5 Rudder Angle Sensor 5-3
5.2.6 Sea Temperature Sensor 5-4
5.2.7 Air Temperature Sensor 5-4
5.2.8 Load Cells 5-4
5.2.9 Sensor Input Configuration 5-4
5.3 ANALOGUE INDICATORS 5-5
5.3.1 Options Available 5-5
5.3.2 Analogue Indicator Configuration 5-6
5.3.3 Meter Scaling 5-7
5.4 GRAVITY SWITCH 5-7
5.5 AUDIBLE ALARM 5-7
5.6 20/20 DISPLAY 5-8
5.6.1 The Display 5-8
5.6.2 Display Configuration 5-8
5.6.3 Function Selection - Remote Push-Button 5-9
5.6.4 Function Selection - FFD 5-9
5.6.5 Re-configuring 20/20 Display 5-10
5.7 HALCYON 2000 COMPASS 5-10
5.8 HALCYON DISPLAY 5-11
5.8.1 The Keys 5-12
5.8.2 Power/Lights Key 5-12
5.8.3 Page Key 5-12
5.8.4 Off Course Function 5-12
5.8.5 Stored Course Function 5-13
5.8.6 Head/Lift Trend Function 5-13
HB-0844-02
5-1
Hydra 2000 User Manual
Part 5 - Options
CONTENTS (Contd.)
Para Page
ILLUSTRATIONS
Fig No Page
TABLE
Table No Page
HB-0844-02
5-2
Hydra 2000 User Manual
Part 5 - Options
PART 5 - OPTIONS
5.1 SYSTEM EXPANSION
The Hydra 2000 System may be expanded to provide a wider
range of facilities and features by the addition of further displays,
sensors, and interfaces. These Options are described in the
following Paragraphs.
5.2 SENSORS
There are four linear inputs into the Main Processor which enable
further sensors to be connected. In addition, the Expansion Unit
which has 12 extra inputs is also available.
5.2.1 Trim Angle Sensor
This is essential if your mast rotates, otherwise you lose all the
wind angle functions. The addition of this unit gives two new
functions, the Wind Angle to the Mast (W/A MAST) and Mast
Angle (MAST ANG)
By sensing Heel Angle and applying this value with that of the
functions of Course and Dead Reckoning, Leeway can be
calculated.
HB-0844-02
5-3
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-4
Hydra 2000 User Manual
Part 5 - Options
Note
Do not select the same number on more than one linear input,
otherwise the selection will be ignored (no function can be
connected to more than one input).
5.3. ANALOGUE INDICATORS
5.3.1 Options Available
Note
Items marked with an asterisk are only available with the Halcyon
Display.
HB-0844-02
5-5
Hydra 2000 User Manual
Part 5 - Options
(5) Press Enter, the lower text will flash. Use Scroll
Down to scroll through the options until the appropriate
one (e.g. DEPTH) is shown. Pressing Enter will select
that option, and the meter output will be configured to
drive a DEPTH meter.
Note
It is possible to configure more than one meter drive to the same
meter option (e.g. BOAT SPD on METER 1 and METER 2).
HB-0844-02
5-6
Hydra 2000 User Manual
Part 5 - Options
Meter scaling can also be varied for boat speed and wind speed.
For example if a 25 knot full scale Boat Speed Meter is required,
this can be done as follows:
HB-0844-02
5-7
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-8
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-9
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-10
Hydra 2000 User Manual
Part 5 - Options
Four rows of text are provided down the right-hand side of the
display. Each row is associated with the adjacent key and
indicates the selected operating function and the various settings.
HB-0844-02
5-11
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-12
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-13
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-14
Hydra 2000 User Manual
Part 5 - Options
The Head/Lift Trend may be reset for the current tack by selecting
CONTROL on the LIFT/HDR function and then RESET. This will
reset the trend on all displays and, if fitted, the analogue indicator.
Note
If the Halcyon Display is not in the Head/Lift Mode, then Head/Lift
trend will indicate OFF.
5.8.9 Head/Lift Trend - 20/20 Display
Head/Lift Trend information may be displayed on any 20/20
Display, see Para 5.8.6. An L or an H will be displayed adjacent to
the digits to signify Lift or Head.
5.8.10 Turn Rate Function
The Turn Rate Function displays the vessel's rate of turn in
degrees per second to port or starboard. There are two ways of
displaying this information, either as Mean Rate of Turn or as
Instantaneous Rate of Turn.
When the Turn Rate Function is selected by pressing the Enter
Key, the display defaults to Mean Rate of Turn and MEAN is
shown on the display.
When MEAN is selected, the present heading is compared with a
continually updated mean heading and displayed graphically and
numerically as degrees to port or starboard. This will give the
trend to right or left of the average heading. Hence, if the heading
fluctuates to port and starboard but with a bias to port, then the net
trend to port will be displayed. Similarly for starboard.
The Instantaneous Rate of Turn is selected by pressing the
Scroll Down Key when the Rate of Turn function is selected on
the Halcyon Display. When this function is selected the display will
change to show DEG/S indicating rate of turn in degrees per
second. The centre bar graph will now show the instantaneous
rate of turn to port or starboard with the numerical value displayed
below.
Note
This is a useful feature when calibrating the Halcyon 2000
Compass where it is necessary to maintain a steady rate of turn.
HB-0844-02
5-15
Hydra 2000 User Manual
Part 5 - Options
Note
Do not set different linear inputs to the same function (except 1 for
normal linear input) or the calibration value will be ignored. This
ensures that no function uses no more than one analogue input.
However, any one of the Linear 1 to 4 inputs from the main
HB-0844-02
5-16
Hydra 2000 User Manual
Part 5 - Options
Adjusting the MIN and MAX values allows the displayed value to
be scaled to the appropriate range for the sensor attached. Taking
a load cell for example, if the zero load output is 0V, MIN
VAL=000 and if maximum load is 650 kgF at 6.5V then MAX
VAL=650.
Notes
1. When calibration value 1 is changed to select a desired input
sensor, the linear value is no longer updated and a constant
value is displayed until the page key is pressed.
2. Extra functions selected, and then removed, remain in the
display menu but with no data shown until the system is
switched off and then back on again.
3. Linear functions will always be shown if selected by CAL VAL 4
on linear 5. They will show no data if the CAL VAL 1 setting is
configured to 1 or 2.
HB-0844-02
5-17
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-18
Hydra 2000 User Manual
Part 5 - Options
HB-0844-02
5-19
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-1
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-2
Hydra 2000 User Manual
Part 6 - Diagnostic Data
The display requests the user to press each key in turn as follows:
Key 1 Enter
Key 2 Scroll Down
Key 3 Scroll Up
Key 4 Page
Key 5 Power/Lights
Key 6 Navigation
Key 7 Wind
Key 8 Speed/Depth
If it does not detect the key being pressed in a short time then the
message TIME OUT is displayed.
6.1.3 Display
6.1.4 Network
6.1.5 RAM
This tests the correct operation of the FFD's internal RAM. If the
test passes RAM PASS OK is displayed. If the test fails then
RAM FAIL is displayed and correct operation of the unit cannot be
guaranteed.
HB-0844-02
6-3
Hydra 2000 User Manual
Part 6 - Diagnostic Data
6.1.6 PROM
This tests the correct operation of the FFD's program memory
and calculates its checksum by adding all the individual bytes of
memory together. The total is displayed in hexadecimal in the
bottom half of the display and the last two digits are the software
version number. If this test fails then if possible, FATAL ER 02 is
displayed, and further operation of the FFD cannot be accurately
defined. The upper half of the display shows the FFD's network
node number.
6.1.7 EEPROM
This tests the correct operation of the FFD’s non-volatile memory
where the display's settings are stored when power is switched
OFF. If this fails FATAL ER 03 or FATAL ER 04 will be displayed.
The display's page configuration and possibly the text for some
functions will be lost.
6.1.8 Lighting
This test checks the correct operation of the display lighting
control cycles through the red and green lighting at the three
illumination levels.
6.1.9 Debug
6.1.10 RES-SYS
HB-0844-02
6-4
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Note
When the system is reset all calibration, damping and alarm
values will be set back to default settings, all log values and trip
functions will be reset to zero. All display units will reset to their
default page settings. After a delay of about 20 seconds the
system should be switched OFF and then back ON twice to
complete the reset procedure.
6.1.11 Versions
This option allows the user to obtain the software version numbers
for the FFD (DISPLAY), Depth board (DEPTH), Wind board
(WIND), Expansion Unit (EXP UNIT) and Pilot, if fitted. With
'VERSIONS' flashing on the display, press Enter. Then use the
Scroll Up Key to select the option whose version number is
required and press Enter to display the software checksum of the
appropriate unit, in the bottom display. The last two digits of the
checksum are the software version number.
6.1.12 Errors
This test is used chiefly to interrogate the Hydra 2000 network and
is designed mainly for use by service technicians to ascertain
levels of interference that may be present. For example inter-
ference may be induced by an SSB transmitter or radar. The
display shows the number of messages that have not been
transmitted correctly first time. Between 10 and 20 errors per
minute is quite normal. A higher number of errors can also be an
indication that the network terminators are not correctly fitted.
6.1.13 Remote
This facility allows the user to invoke internal RAM, PROM and
basic network checks on any remote processor node on the
network, and display the appropriate error messages on the FFD.
Node numbers are allocated when the system is switched ON for
the first time or after a system reset.
In the Hydra 2000 System the numbers will be:
Depth board 1
Wind board 5
In a system with additional processor nodes it may be necessary
to determine the individual node numbers by disconnecting nodes
in turn and running the remote PROM test. If no pass or fail
HB-0844-02
6-5
Hydra 2000 User Manual
Part 6 - Diagnostic Data
The main processor contains two circuit boards: the wind board
and the depth board. The depth board is responsible for boat
speed, sea temperature and depth measurement. The wind board
is the main sensor responsible for wind functions and also drives
the analogue meters. It has special inputs for the Masthead Unit,
Super Halcyon compass, air temperature, battery voltage and four
linear inputs that can be connected to a variety of other sensors
(see Part 4 - Installation Instructions).
Note
The Masthead Unit's bearings should not be oiled as they are of a
sealed pre-lubricated type and additional oil may cause chemical
breakdown of the existing lubricant.
HB-0844-02
6-6
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Possible Causes:
HB-0844-02
6-7
Hydra 2000 User Manual
Part 6 - Diagnostic Data
(e) The gain of the receiver has been set too low. It is
possible to adjust the maximum gain via CAL VAL1 on
gain. This is normally set to 30 and should not be
adjusted.
-
- -
-
Possible Causes:
HB-0844-02
6-8
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Possible causes:
HB-0844-02
6-9
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Possible Causes:
HB-0844-02
6-10
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-11
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Possible Causes:
Possible Cause:
Possible Cause:
HB-0844-02
6-12
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Possible Cause:
The Pilot has not been set to use the Halcyon 2000 as its
heading source. Pilot Heading Node must be set to Node
16. If the Halcyon 2000 is not to be the source of heading
then it must be unplugged from the network.
Possible Cause:
The Main Processor has not been set to use the Halcyon
2000 as the heading source. Set
NAVIGATE→COURSE, CAL VAL 1 (HDG NODE) to 16.
HB-0844-02
6-13
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Notes
1. Function repeated when no B&G fluxgate connected to system.
Not used by Autopilot.
2. Displayed when no B&G fluxgate connected to system. Can be
used by Autopilot.
3. Re-transmitted only. Not used by Hydra 2000.
HB-0844-02
6-14
Hydra 2000 User Manual
Part 6 - Diagnostic Data
Notes
1. NMEA output data is subject to the appropriate sensor or NMEA
sentence being connected to the system.
HB-0844-02
6-15
Hydra 2000 User Manual
Part 6 - Diagnostic Data
NMEA Sentences
HB-0844-02
6-16
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-17
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-18
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-19
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-20
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-21
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-22
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-23
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-24
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-25
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-26
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-27
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-28
Hydra 2000 User Manual
Part 6 - Diagnostic Data
HB-0844-02
6-29
Hydra 2000 User Manual
Appendix 1
A1 HYDRA 2000 CALIBRATION DATA
A1.1 True Wind Speed Correction Table
PORT High
BOAT SPD Hz/Kt
STD LOG STBD Low
Hz/Kt
HB-0844-02
A-1
Hydra 2000 User Manual
Appendix 1
Hydra 2000 Settings Chart Continued
Function Calibration Calibration Damping Alarm Alarm
Type / Unit Value Value Value
Datum High
DEPTH Offset
m, ft or fms Low
HEEL Alignment°
LEEWAY Alignment°
RUDDER Alignment°
HB-0844-02
A-2