Hba1c Assay

Hitachi 917
Temperature 37°C
Analyze
1 Test Name [HbA1c] Sample type [whole blood]
2 Assay /Time/Point [2 point end][10][16][27][0][0]
3 Wave(sub/main) [800][700]
4 S.vol.(Normal) [25][0.0][0]
5 S.vol.(Decrease) [25][0.0][0]
6 S.vol.(Increase) [25][0.0][0]
7 Diluent [Water][0]
8 Reagent Vol.(R1) T1 [112][0][*][0]
9 Reagent Vol.(R2) T2 [48][0][*][0]
10 Reagent Vol.(R3) T3 [70][0][*][0]
11 Reagent Vol.(R4) T4 [0][0][*][0]
12 ABS Limit [32000][Increase] 2 Tests[ ]
13 Prozone limit [0][0] [Lower ]
14 Cell Detergent [Detergent 1]
Calibration
15 Calib Type [Linear][ ]
16 Point [2] Span point [2 ]
17 Weight [0]
Auto Calibration
Time out Chg. over
18 Blank [0] Change Lot[ ]
19 Span [0] Change Bottle[ ]
20 2 Point [168]
21 Full [0]
22 SD Limit [999]
23 Duplicate Limit [10]% [32000]Abs
24 Sensitivity Limit [-99999] [999999]
25 S1 ABS Range [-32000] [32000 ]
Range
26 Application code [*] Unit [ %]
27 Report Name [*] Data mode [ On board]
28 Control Interval [0]
29 Inst. Factor (Y=aX+b): a=[1.0] b=[0.0]
30 Technical Limit [4][12] Repeat limit [-99999][999999]
31 Expected value
32 Male [ ]
33 Female [ ]
Others
34 Standard Solution (1) (2) (3) (4) (5) (6)
35 Calibration code [*] [*] [0] [0] [0] [0]
35 Concentration [*.*] [*.*]
36 Position [*] [*]
37 Sample volume [25] [25] [0] [0] [0] [0]
38 Diluted S. volume [0] [0] [0] [0] [0] [0]
39 Diluent volume [0] [0] [0] [0] [0] [0]