SPECIFICATION NO.

SHEET 1 OF 1
AQUAMON SMARTPro 8966 DATA SHEET pH ANALYSER REV 0
POLILYTE PRO 120
1 GENERAL
1.01 MANUFACTURERS Transmitter : Forbes Marshall Pvt. Ltd., India
Transmitter : Aquamon SMARTPro 8966
1.02 MODEL NUMBER
Electrode : POLILYTE PRO 120
1.03 INSTRUMENT RANGE 0-14 pH, Programmable Span
2 SENSORS
2.01 TYPE Combination electrode
2.02 ELECTRODE CONNECTION AS7 Type
2.03 TEMPERATURE RATING -10º C – 60º C
2.04 PRESSURE RATING 6 Bar
2.05 SHAFT MATERIAL Glass
2.06 DIAPHRAGM Single Pore - 1 nos.
2.07 TEMPERATURE SENSOR NA
2.08 REFERENCE ELECTROLYTE Polisolve
2.09 REFERENCE ELECTRODE EVEREF - B
2.10 ELECTRODE HEAD PG 13.5
2.11 CABLE Low noise Co-axial cable of 5 meters
2.12 SENSOR LENGTH 120 mm
3 TRANSMITTER
3.01 TYPE Microcontroller based, 2-wire transmitter
Range Resolution Accuracy
0 – 14 pH, Programmable +/- 0.01
3.02 pH 0.01 pH
Span pH
3.03 TEMPERATURE -20º C TO +200º C 0.01º C +/- 0.1 º C
3.04 SPAN 2pH – 14pH
2 Line Dot-matrix 10 characters per line LCD display,
3.05 DISPLAY
programmable along with conversion factor.
3 button keypad & magnetic pin programming for user
3.06 KEYPAD
friendly menu interface
3.07 COMMUNICATION PROTOCOL HART Communication Protocol
3.08 CURRENT OUTPUT 4-20 mA DC
3.09 CALIBRATION Single/Dual Point Calibration
3.10 TEMPERATURE COMPENSATION Automatic or Manual (PT 100/ PT 1000)
3.11 POWER REQUIREMENTS 12 – 36 VDC with maximum load 600 Ώ @ 24 VDC
3.12 POWER CONSUMPTION Less than 53 mW
Die-cast Aluminium (LM6) with epoxy paint (Non -
3.13 ENCLOSURE MATERIAL
corrosive)
1 x ½ “ NPT – Sensor
3.14 CABLE GLANDS
2 x M20 - Supply & Pulse Output
3.15 WEIGHT 4000 gms
3.16 MOUNTING TYPE 2" Pipe Type (Vertical)
3.18 PROTECTION CLASS IP 67
3.19 AMBIENT TEMPERATURE OPERATING RANGE -20˚C to + 70˚C
4 ELECTRODE HOLDER
4.01 HOLDER TYPE Flow thru chamber
4.02 MOC SS
4.03 HOLDER CONECTION Clamps / Flange