1 Configuration Changes the settings in the following items:
Language, Telemetry, Alarm Off Message, Battery
Insert Message, 12Lead Option, Pacing Option,
Freeze Function, Charge Time Clear, Default
Setting, Flash Save, Digital filter
2 Adjust AD Changes the A/D values for ECG waveform
acquisition, HV unit, and battery charge.
3 Check Hardware Checks the operation of the key switches, LEDs,
recorder, memories, sounds, etc
4 AD View Checks the A/D values that the main CPU
currently receives.
5 Operation Time Displays the count of internal and external
discharges and operation time.
6 Version Up You can update the system program and data
7 Debug mode Displays the characters used in this instrument
and checks the flash memory.
Configuration Language Selects the language or country.
Selection list: English, Japanese, American, Chinese, Spanish, French,
German, Italian, Norwegian, Finnish, Other 1 to 5
Telemetry(ZR) Select Export.
Selection list: Export (Other than Domes channel)
Domes (Channel A1000 to A6000)
Alarm Off Message Select On to display a message when any of the vital alarm settings
is set to off.
Selection list: On, Off
Battery Insert Message Select On to display a message when the battery is not in the
instrument.
Selection list: On, Off
12 Lead Option Select On when using the optional 12 lead module. This option is
only available for the TEC-7721/7731 series defibrillator.
Selection list: On, Off
Pacing Option On: For the TEC-7631/7731 series defibrillators.
Off: For the TEC-7621/7721 series defibrillators.
Freeze Function Select On to enable the freeze function. The ECG second trace can
be frozen when the Alarm setting key is pressed.
Selection list: On, Off
Charge Times Clear Deletes the count of internal and external discharges.
Selection list: Off, Exec
Default Setting Select Exec to return all settings (including the AD adjustment
settings, alarm settings) to the factory default settings. Before doing
this operation, write down the necessary settings for each item.
Selection list: Off, Exec
Flash Save Select Exec to save the current settings in the System Maintenance
screen in the flash ROM.
Selection list: Off, Exec
Digital filter Select On to enable the digital filter for ECG. This function is
available only when AC hum filter in the Setup screen is set to ON
Selection list: On, Off
Adjust AD Gain Adjusts the A/D value for the ECG sensitivity.
Adjust ECG AD Setting range: 0 to 31 in of 1
Offset Adjusts the A/D value for the ECG offset at gain 1/2, 1, 2 and 4.
Setting range: 0 to 50 in of 1
Adjust AD CHARGE AD Adjusts the A/D value to measure the charged energy in the HV
Adjust HV AD capacitor.
Setting range: -25 to +25 in of 1
TTR AD Adjusts the A/D value to measure the TTR (transthoracic resistance).
Setting range: -25 to +25 in of 1
DELIVERED ENERGY Adjusts the /AD value to measure the delivered energy from the HV
capacitor.
Setting range: -25 to +25 in of 1
Adjust AD BATTERY VOLTAGE Adjusts the A/D value of the voltage that is applied to the battery.
Adjust Battery AD Setting range: -25 to +25 in of 1
BATTERY CURRENT Adjusts the A/D value of the current that applied to the battery.
Setting range: -25 to +25 in of 1
The default setting of the following items may differ depending on the instrument:
Adjust ECG A/D, CHARGE AD, TTR AD, DELIVERED ENERGY, BATTERY
VOLTAGE, BATTERY CURRENT.
English, French, German, Def (defibrillation) MMM/DD/YYYY
Italian Spanish, Other
Chinese, Japanese Def (defibrillation) YYYY/MM/DD
American Sync (cardioversion) MMM/DD/YYYY
Language sets the language which is used in the instrument. The default setting
of Sync mode after CV (the mode after cardioversion) and Date format in the
System Setup screen changes according to the selected language.
2J ± 0.5 J
3J ±1J
5, 7, 10 J ±2J
20 J or more ± 10 %
2J ± 0.5 J
3J ±1J
5, 7, 10 J ±2J
20 J or more ± 10 %
Part Name Description
Pacing Rate Shows the status of the PACING RATE control on the top panel
(TEC-7631/7731 series only).
Pacing Start Highlights when the START/STOP key on the top panel is pressed
(TEC-7631/7731 series only).
SetUp Shows the status of the Energy/Mode Select control.
Record Highlights when the record key is pressed.
Event Highlights when the event key is pressed.
Card exist or No Card Shows whether a memory card is inserted into the memory card
slot.
Paper Exit or Out of Paper Shows whether the recording paper is set or not.
Battery or No Battery Shows whether the battery pack is installed or not.
AC AC: Displayed when the AC power cord is connected to the
instrument
DC: Displayed when the AC power cord is not connected to the
instrument.
Panel Charge Highlights when the CHARGE key on the front panel is pressed.
Panel Sync Highlights when the SYNC key on the front panel is pressed.
Panel Dis L Highlights when the left DISCHARGE key on the front panel is
pressed.
Panel Dis R Highlights when the right DISCHARGE key on the front panel is
pressed.
Paddle Charge Highlights when the CHARGE key on the apex external paddle is
pressed.
Panel Dis L Highlights when the left DISCHARGE key on the sternum external
panel is pressed.
Panel Dis R Highlights when the right DISCHARGE key on the sternum external
paddle is pressed.
External Shows the type of the connected paddle. Displays "No Connect"
when no paddle is connected.
DSI 1, DSI 2 or DSI 1 2 DSI 1, DSI 2: Displayed when SpO2 adapter or CO2 sensor kit is
connected to the left or right SpO2/CO2 connector.
DSI 1 2: Displayed when SpO2 adapter and CO2 sensor kit are
connected to the SpO2/CO2 connectors.
NIBP Displayed when the NIBP unit is installed.
Leads Highlights when the ECG lead key is pressed.
Sens Highlights when the ECG sensitivity key is pressed.
Suspend Highlights when the Silence alarms key is pressed.
Alarm Highlights when the alarms setting key is pressed, then the System
Maintenance - Check Hardware screen is displayed.
Program card Instrument < Version up card Versionup Start
Instrument > Version up card Versiondown Confirm
Data card ------- Versionup Start