SERVICE BULLETIN

Carestream Health, Inc., Rochester, NY 14608

MAY 2010
Supersedes
April 2010

KODAK DIRECTVIEW Classic/Elite CR Systems


Service Codes: 5258, 5259

Announcing Modification No. 17- Ethernet Isolation Board

IMPORTANT: When doing the procedures in this document, you must use safe work practices
and wear the correct Personal Protective Equipment (for example, Safety Eyewear) according to
your Company's Standard Operating Procedures.

Purpose

This bulletin is to announce the availability of Modification No. 17, Ethernet Isolation Board.

The purpose of the Ethernet Isolation Board is to allow a Classic/Elite Scanner to directly connect
to a WAIV console PC. Doing this, along with some minor IP address changes will allow both
the scanner and the PC to communicate directly off of the customer network and use a single
network connection for non scanner related communications to the customer network.

This does not change our current product requirements which require two network drops, one for
the WAIV Console PC and one for the Classic/Elite Scanner. The intention is to use this in cases
where the customer network causes difficulty with communication between the PC and the
Scanner.

Modification Instructions, 8H9936 are on the Service Portal.

NOTE: SP2G7266 consists of the circuit board, as well as the hardware necessary to mount it to
the Diamond Frame. The Ethernet Cable completes the connection from the IEB to the outer
network. The board itself is passive, and requires no additional connections, other than an
ethernet connection to the IEB, and another ethernet connection to the outer socket.

Page 1 Carestream Health, Inc. Confidential Pub No. SB8J3647


The Modification consists of 2 major parts:

NOTE: Each of these parts must be ordered separately.

Mark McKann, Service Engineer – Rochester, NY

For more information please contact:

*HCIS Address Only*

CARESTREAM HEALTH, INC. CARESTREAM HEALTH, INC.


Center of Excellence (TSC) Technical and Innovation Center (T&IC)
1049 West Ridge Rd 150 Verona Street
Rochester, NY 14615 Rochester, NY 14608
1-800-328-2910 (USA only) 1-800-328-2910 (USA only)
585-627-1864 (Outside the USA) 585-627-1864 (Outside the USA)

CARESTREAM and DIRECTVIEW are trademarks of Carestream Health, Inc.


Kodak is a trademark of Kodak used under license.

Printed in the USA


© Carestream Health, Inc., 2010
Rochester, NY 14608

Page 2 Carestream Health, Inc. Confidential Pub No. SB8J3647