Calibration
Dead-weight tester in compact design
Model CPB3800
WIKA data sheet CT 31.06
for further approvals
see page 9
Applications
■ Primary standard for calibrating the pressure scale in a
hydraulic range up to 1,200 bar [16,000 lb/in²]
■ Reference instrument for factory and calibration
laboratories for the testing, adjustment and calibration of
pressure measuring instruments
■ Complete, stand-alone system, also suitable for on-site
use
Special features
■ Total measurement uncertainty up to 0.025 % of reading
■ Upgradeable using CPS5800/CPM5800 to provide
increased accuracy to 0.006 %
■ Direct replacement of original DH-Budenberg 580 series
■ Factory calibration included as standard, traceable to
national standards, with UKAS calibration possible as an
option
■ Masses manufactured from stainless steel, can be
Dead-weight tester in compact design,
adjusted to local gravity model CPB3800
Description
Proven primary standard Basic principle
Pressure balances (dead-weight testers) are the most Pressure is defined as the quotient of force and area. The
accurate instruments available on the market for the core component of the CPB3800 is therefore a very precisely
calibration of electronic or mechanical pressure measuring manufactured piston-cylinder system, onto which a mass
instruments. The direct measurement of the pressure load is applied in order to generate the individual test points.
(p = F/A), as well as the use of high-quality materials enable
a very small measurement uncertainty, in conjunction with an The mass load is proportional to the target pressure and
excellent long-term stability. this is achieved through optimally graduated masses. As
The pressure balance (dead-weight tester) has therefore standard, these masses are manufactured to the standard
been used for years in factory and calibration laboratories in gravity (9.80665 m/s²), though they can be adjusted to
industry, national institutes and research laboratories. a specific location and calibrated according to UKAS
accreditation.
Stand-alone operation
Due to its integrated pressure generation and the purely
mechanical measuring principle, the model CPB3800 is ideal
for on-site use for maintenance and service.
WIKA data sheet CT 31.06 ∙ 08/2020 Page 1 of 16
Data sheets showing similar products and accessories:
Pneumatic pressure balance; model CPB5000; see data sheet CT 31.01
Hydraulic pressure balance; model CPB5800; see data sheet CT 31.11
Pressure balance for high pressure; model CPB5000HP; see data sheet CT 31.51
Pressure balance for differential pressure; model CPB5600DP; see data sheet CT 31.56
CalibratorUnit; model CPU6000; see data sheet CT 35.02
Easy operation
The integrated dual-area spindle pump enables rapid filling
of the test system and smooth pressure generation up to
1,200 bar [16,000 lb/in²]. At the same time, the precisely
adjustable spindle pump also enables fine pressure
adjustment. A control schematic for pressure generation on
the instrument base facilitates quick and easy operation.
As soon as the measuring system reaches equilibrium, there
is a balance of forces between the pressure and the masses.
The excellent quality of the system ensures that this pressure
remains stable over several minutes, so that the pressure
value for comparative measurements can be read without
any problems, or also so that more complex adjustments can
be carried out on the test item.
Compact instrument version
The CPB3800 is also particularly notable for its compact
dimensions, which are not altered during operation, since the Control schematic on the CPB3800 instrument base
spindle runs within the pump body.
With its compact dimensions, the exceptionally robust ABS
plastic case and the low weight associated with these, the
CPB3800 is also particularly suited to on-site applications.
CPB3800 with dual-range piston-cylinder
systems
Next to the standard piston-cylinder systems the CPB3800
instrument base can also be combined with the CPS5800
single- or dual-range piston-cylinder systems.
With this combination the former 580 series dead-weight
tester from DH-Budenberg is still available. This includes all
580L/580M/580DX and 580HX instruments and are identical
in form, fit, function, specification and quality.
Dead-weight tester CPB3800 with piston unit CPS5800
The table below shows WIKA's model code combinations to
the corresponding 580 series models.
Replacement instruments for DH-Budenberg 580 series
Former model Replacement model code combinations
designations Instrument Piston-cylinder Range 1) Accuracy
base system / mass set Standard Premium
580L CPB3800 CPS5800 / CPM5800 1 ... 120 bar [10 ... 1,600 lb/in²] 0.015 % 0.007 %
580M CPB3800 CPS5800 / CPM5800 2 ... 300 bar [30 ... 4,000 lb/in²] 0.015 % 0.006 %
580DX CPB3800 CPS5800 / CPM5800 1 ... 60 bar / 10 ... 700 bar 0.015 % 0.006 %
[10 ... 800 / 100 ... 10,000 lb/in²]
580HX CPB3800 CPS5800 / CPM5800 1 ... 60 bar / 20 ... 1,200 bar 0.015 % 0.007 %
[10 ... 800 / 200 ... 16,000 lb/in²]
1) Standard weights are dual marked in pressure units bar and kPa. Also available for kg/cm² and lb/in².
For detailed specifications see page 6, 8 and 9.
WIKA data sheet CT 31.06 ∙ 08/2020 Page 2 of 16
The standard piston-cylinder system
Both the piston and cylinder are manufactured from Force F
tungsten carbide. This pairing of materials, in comparison
to other materials, has very low pressure and temperature
coefficients of expansion, which results in a very good
linearity for the effective area of the piston and a very high
accuracy.
Piston and cylinder are very well protected against contact,
impacts or contamination from outside in a solid stainless-
steel case. At the same time, overpressure protection is
integrated, which prevents the piston from being forced out
vertically and avoids damage to the piston-cylinder system in 1
the event of masses removal under pressure.
The masses are stacked on mass carrier, which sits on the
piston shaft. The construction of the mass carrier provides
Pressure p
a very low centre of gravity for the stacked masses, which
minimises both the side thrust on the piston-cylinder system Piston-cylinder system with masses, without mass
and the friction. For smaller starting pressures, the masses carrier
can also be stacked directly onto the piston shaft, without
using the mass carrier.
The overall design of the piston-cylinder system and the Force F
very precise manufacturing of both the piston and the
cylinder, ensures excellent operating characteristics with
long free-rotation time and low sink rates. Thus a high
long-term stability is ensured. Therefore, the recommended
recalibration interval is two to five years dependent on the
conditions of use.
1 Effective area A
Pressure p
Piston-cylinder system with masses and mass carrier
WIKA data sheet CT 31.06 ∙ 08/2020 Page 3 of 16
The optional piston-cylinder system model CPS5800
The CPS5800 piston-cylinder systems are available in two Piston and cylinder are very well protected, against
fundamentally different designs, depending on the measuring contact, impacts or contamination from outside, in a solid
range: stainless-steel/hardened tool steel case. At the same time,
overpressure protection is integrated, which prevents the
■ Single-range piston-cylinder system for measuring ranges piston from being forced out vertically and avoids damage to
120 bar and 300 bar [1,600 and 4,000 lb/in²] the piston-cylinder system in the event of masses removal
■ Dual-range piston-cylinder system for measuring ranges under pressure.
700 bar and 1,200 bar [10,000 and 16,000 lb/in²]
The masses are stacked directly onto the piston shaft. This
High accuracy over a wide measuring range makes it easier to place the masses on and thus enables a
The dual-range piston-cylinder system offers two measuring lower start value.
ranges in one case with automatic measuring range
switching from low-pressure to high-pressure piston. This The overall design of the piston-cylinder system and the very
provides the user with an extremely flexible measuring precise manufacturing of both the piston and the cylinder,
instrument that can cover a wide measuring range with high ensure exceptionally low friction force, which results in
accuracy, with only one piston-cylinder system and one excellent operating characteristics with long free-rotation time
mass set. Additionally two test points can automatically be and low sink rates. Thus a high long-term stability is ensured.
achieved by loading just the one mass. Therefore, the recommended recalibration interval is two to
five years dependent on the conditions of use.
Both the piston and cylinder are manufactured from
hardened steel or tungsten carbide, respectively. This
pairing of materials has very low pressure and temperature 1 Effective area A
coefficients of expansion, which results in a very good 2 High-pressure piston
linearity for the effective area of the piston and a very high
3 Low-pressure piston = High-pressure cylinder
accuracy.
Force F Force F
2
1
Pressure p
Pressure p
Single-range piston-cylinder system, model CPS5800 Dual-range piston-cylinder system, model CPS5800
WIKA data sheet CT 31.06 ∙ 08/2020 Page 4 of 16
Tables of masses
The following tables show the number of masses within a For the measurement of the ambient conditions, the
mass set with their nominal mass values and the resulting CPU6000 CalibratorUnit can be used, see page 12.
nominal pressures for the respective measuring ranges.
The masses are manufactured, as standard, to the standard
Should the instrument not be operated under reference gravity (9.80665 m/s²) although they can be adjusted for any
conditions (ambient temperature 20 °C [68 °F], atmospheric local gravity.
pressure 1,013 mbar [14.69 lb/in²], relative humidity 40 %),
the measured values must be corrected arithmetically.
Standard mass sets
Measuring range 1 ... 120 2.5 ... 300 5 ... 700 10 ... 1,200
[bar/kPa]
per piece
per piece
per piece
per piece
pressure
pressure
pressure
pressure
Quantity
Quantity
Quantity
Quantity
Nominal
Nominal
Nominal
Nominal
[bar] [bar] [bar] [bar]
Piston and make-up weight 1 1 1 2.5 1 5 1 10
Piston, mass carrier and 1 20 1 50 1 100 1 200
make-up weight for mass
carrier
Masses (stackable on mass 3 20 3 50 4 100 3 200
carrier)
Masses (stackable on piston) 1 20 1 50 1 100 1 200
1 10 1 25 1 50 1 100
2 4 2 10 2 20 2 40
1 2 1 5 1 10 1 20
1 1 1 2.5 1 5 1 10
Measuring range 10 ... 1,600 25 ... 4,000 50 ... 10,000 100 ... 16,000
[lb/in²]
per piece
per piece
per piece
per piece
pressure
pressure
pressure
pressure
Quantity
Quantity
Quantity
Quantity
Nominal
Nominal
Nominal
Nominal
[lb/in²] [lb/in²] [lb/in²] [lb/in²]
Piston 1 10 1 25 1 50 1 100
Piston, mass carrier and 1 190 1 475 1 950 1 1,900
make-up weight for mass
carrier
Masses (stackable on mass 5 200 5 500 7 1,000 5 2,000
carrier)
Masses (stackable on piston) 1 200 1 500 1 1,000 1 2,000
1 100 1 250 1 500 1 1,000
2 40 2 100 2 200 2 400
1 20 1 50 1 100 1 200
1 10 1 25 1 50 1 100
WIKA data sheet CT 31.06 ∙ 08/2020 Page 5 of 16
The mass sets model CPM5800 (option)
These mass sets are supplied in a flight cases with foam
inserts. This includes the masses listed in the table of
masses below, made from non-magnetic stainless steel, and
optimised for everyday use.
For finer increments and for a higher resolution, as an option,
the standard mass sets can be extended by a set of fine
increment masses.
If even smaller intermediate values need to be generated,
using one of the class M1 or F1 set of fine increment masses
from the accessories is recommended.
Mass set model CPM5800 (Sample picture)
Measuring range Single-piston measuring ranges Dual-piston measuring ranges
[bar] or [kg/cm²] 1 … 120 2 … 300 1 … 700 1 … 1,200
1 … 60 10 … 700 1 … 60 20 ... 1,200
per piece
per piece
per piece
per piece
per piece
per piece
pressure
pressure
pressure
pressure
pressure
pressure
Quantity
Quantity
Quantity
Quantity
Nominal
Nominal
Nominal
Nominal
Nominal
Nominal
[bar] or [bar] or [bar] or [bar] or [bar] or [bar] or
[kg/cm²] [kg/cm²] [kg/cm²] [kg/cm²] [kg/cm²] [kg/cm²]
Piston and make-up 1 1 1 2 1 1 10 1 1 20
weight
Standard mass set 4 20 4 50 5 10 100 4 10 200
1 18 1 45 1 9 90 1 9 180
1 10 1 25 1 5 50 1 5 100
2 4 2 10 2 2 20 2 2 40
1 2 1 5 1 1 10 1 1 20
2 1 1 3 1 0.5 5 1 0.5 10
1 0.5 1 2.5 -- -- -- -- -- --
Set of fine increment 1 0.4 2 1 2 0.2 2 2 0.2 4
masses
1 0.2 1 0.5 1 0.1 1 1 0.1 2
(optional)
1 0.1 1 0.25 1 0.05 0.5 1 0.05 1
2 0.04 2 0.1 2 0.02 0.2 2 0.02 0.4
1 0.02 1 0.05 1 0.01 0.1 1 0.01 0.2
WIKA data sheet CT 31.06 ∙ 08/2020 Page 6 of 16
Specifications
Model CPB3800
Piston-cylinder systems (standard)
Measuring range 1) 1 … 120 bar 2.5 ... 300 bar 5 ... 700 bar 10 ... 1,200 bar
Required masses 41 kg 50 kg 58 kg 50 kg
Smallest step 2) 1 bar 2.5 bar 5 bar 10 bar
(Standard mass set)
Nominal effective area of the piston 1/16 in² 1/40 in² 1/80 in² 1/160 in²
Measuring range 1) 10 … 1,600 lb/in² 25 ... 4,000 lb/in² 50 ... 10,000 lb/in² 100 ... 16,000 lb/in²
Required masses 37 kg 46 kg 58 kg 46 kg
Smallest step 2) 10 lb/in² 25 lb/in² 50 lb/in² 100 lb/in²
(Standard mass set)
Nominal effective area of the piston 1/16 in² 1/40 in² 1/80 in² 1/160 in²
Accuracies
Standard 3) 4) 0.05 % of reading
Option 3) 4) 0.025 % of reading
Pressure transmission medium Hydraulic fluid based on VG22 mineral oil (0.5 l included in scope of delivery)
Material
Piston Tungsten carbide
Cylinder Tungsten carbide
Mass set Stainless steel, non-magnetic
Weight
Piston-cylinder system 2.4 kg [5.3 lbs]
bar mass set incl. mass carrier 41.5 kg [91.5 lbs] 50.5 kg [111.4 lbs] 58.5 kg [129.0 lbs] 50.5 kg [111.4 lbs]
lb/in² mass set incl. mass carrier 37 kg [81.4 lbs] 45.6 kg [100.5 lbs] 57 kg [125.5 lbs] 45.5 kg [100.5 lbs]
Storage case for mass set 11 kg [24.2 lbs]
(optional, 2 pieces required)
Dimensions (W x D x H)
Storage case for mass set (optional) 400 x 320 x 320 mm and 320 x 220 x 320 mm
[15.7 x 12.6 x 12.6 in] and [12.6 x 8.7 x 12.5 in]
1) Theoretical starting value; corresponds to the pressure value generated by the piston or the piston and its make-up weight (by their own weight). To optimise the operating
characteristics more masses should be loaded.
2) The smallest pressure change value that can be achieved based on the standard mass set. To reduce this, a set of fine increment masses is also available.
3) The accuracy from 10 % of the measuring range is based on the measured value. In the lower range, a fixed error based on 10 % of the range applies.
4) Measurement uncertainty assuming reference conditions (ambient temperature 20 °C [68 °F], atmospheric pressure 1,013 mbar [14.69 lb/in²], relative humidity 40 %). For operation
without a CalibratorUnit, corrections must be made if required.
WIKA data sheet CT 31.06 ∙ 08/2020 Page 7 of 16
Piston-cylinder systems model CPS5800 (option)
Version Single-piston measuring ranges Dual-piston measuring ranges
Measuring range 1) 1 … 120 bar 2 … 300 bar 1 … 60 bar / 1 … 60 bar /
10 ... 700 bar 20 … 1,200 bar
Required masses 49.7 kg 49.6 kg 57.4 kg 49.2 kg
Smallest step 2) 0.5 bar 2.5 bar 0.5 bar / 5.0 bar 0.5 bar / 10 bar
(Standard mass set)
Smallest step 3) 0.02 bar 0.05 bar 0.01 bar / 0.1 bar 0.01 bar / 0.2 bar
(Set of fine increment masses)
Nominal effective area of the piston 0.4032 cm² 0.1613 cm² 0.8065 cm² / 0.0807 cm² 0.8065 cm² / 0.0403 cm²
Measuring range 1) 10 … 1,600 lb/in² 30 … 4,000 lb/in² 10 … 800 / 10 … 800 /
100 … 10,000 lb/in² 200 … 16,000 lb/in²
Required masses 45.5 kg 45.3 kg 56.4 kg 45 kg
Smallest step 2) 5 lb/in² 20 lb/in² 5 lb/in² / 50 lb/in² 5 lb/in² / 100 lb/in²
(Standard mass set)
Smallest step 3) 0.2 lb/in² 0.5 lb/in² 0.1 lb/in² / 1 lb/in² 0.1 lb/in² / 2 lb/in²
(Set of fine increment masses)
Nominal effective area of the piston 0.4032 cm² 0.1613 cm² 0.8065 cm² / 0.0807 cm² 0.8065 cm² / 0.0403 cm²
Measuring range 1) 100 … 12,000 kPa 200 … 30,000 kPa 100 … 6,000 kPa / 100 … 6,000 kPa /
1,000 … 70,000 kPa 2,000 … 120,000 kPa
Required masses 49.7 kg 49.6 kg 57.4 kg 49.2 kg
Smallest step 2) 50 kPa 250 kPa 50 kPa / 500 kPa 50 kPa / 1,000 kPa
(Standard mass set)
Smallest step 3) 2 kPa 5 kPa 1 kPa / 10 kPa 1 kPa / 20 kPa
(Set of fine increment masses)
Nominal effective area of the piston 0.4032 cm² 0.1613 cm² 0.8065 cm² / 0.0807 cm² 0.8065 cm² / 0.0403 cm²
Accuracies
Standard 4) 5) 0.015 % of reading
Premium 4) 5) 0.007 % of reading 0.006 % of reading 0.007 % of reading
Pressure transmission medium
Standard Hydraulic fluid based on VG22 mineral oil 6)
Material
Piston Steel Steel Steel / tungsten carbide Steel / tungsten carbide
Cylinder Bronze Steel Steel / tungsten carbide Steel / tungsten carbide
Mass set Stainless steel, non-magnetic
Weight
Piston-cylinder system 1 kg [2.2 lbs] 0.8 kg [1.8 lbs] 2 kg [4.4 lbs] 2 kg [4.4 lbs]
Storage case for the piston-cylinder 3.1 kg [6.8 lbs]
system
bar standard mass sets 66.5 kg [146.3 lbs] 66.4 kg [146.1 lbs] 74.2 kg [163.2 lbs] 66 kg [145.2 lbs]
(in 2 flight cases)
lb/in² standard mass sets 62.3 kg [137.1 lbs] 62.1 kg [136.6 lbs] 73.2 kg [161 lbs] 61.8 kg [136 lbs]
(in 2 flight cases)
bar set of fine increment masses 0.33 kg [0.73 lbs] 0.5 kg [1.1 lbs] 0.5 kg [1.1 lbs] 0.5 kg [1.1 lbs]
lb/in² set of fine increment masses 0.23 kg [0.5 lbs] 0.34 kg [0.75 lbs] 0.34 kg [0.75 lbs] 0.34 kg [0.75 lbs]
1) Theoretical starting value; corresponds to the pressure value generated by the piston or the piston and its make-up weight (by their own weight). To optimise the operating
characteristics more masses should be loaded.
2) The smallest pressure change value that can be achieved based on the standard mass set. To reduce this, a set of fine increment masses is also available.
3) The smallest pressure change value that can be achieved based on the optional set of fine increment masses. For further reductions, an accessory of class M1 or F1 set of fine
increment masses is available to compensate for actual area of piston unit.
4) The accuracy from 10 % of the measuring range is based on the measured value. The standard accuracy of 0.02 % of reading or 10 % of measuring range is maintained without any
correction of actual area of the piston unit. In the lower range, the accuracy is 0.03 % of reading for the single-range piston-cylinder systems and 0.025 % of reading for dual-range
piston-cylinder systems.
5) Measurement uncertainty assuming reference conditions (ambient temperature 20 °C [68 °F], atmospheric pressure 1,013 mbar [14.69 lb/in²], relative humidity 40 %). For operation
without a CalibratorUnit, corrections must be made if required.
6) Other pressure transmission media on request.
WIKA data sheet CT 31.06 ∙ 08/2020 Page 8 of 16
Piston-cylinder systems model CPS5800 (option)
Dimensions (W x D x H)
Carrying case for standard mass 400 x 320 x 320 mm and 320 x 220 x 320 mm
sets [15.7 x 12.6 x 12.6 in] and [12.6 x 8.7 x 12.5 in]
Storage case for the piston-cylinder 380 x 160 x 170 mm [15 x 6.3 x 6.7 in]
system (optional)
Base
Connections
Connection for piston-cylinder system G ¾ B (male)
Test item connection G ½ female thread, loose union connection
Material
Wetted parts Austenitic stainless steel, high-tensile brass, nitrile rubber
Pressure transmission medium Hydraulic fluid based on VG22 mineral oil (0.5 l included in scope of delivery) 6)
Reservoir 170 cm³
Weight
Base 13.5 kg [29.8 lbs]
Storage case for the base (optional) 9 kg [19.8 lbs]
Permissible ambient conditions
Operating temperature 18 ... 28 °C [64 ... 82 °F]
Dimensions (W x D x H)
Base 401 x 397 x 155 mm [15.8 x 15.7 x 6.1 in], for details, see technical drawings
6) Other pressure transmission media on request.
Approvals
Logo Description Country
EU declaration of conformity European Union
Pressure equipment directive; PS > 1,000 bar, module A, pressure accessories
EAC (option) Eurasian Economic
■ EMC directive Community
■ Pressure equipment directive
■ Low voltage directive
■ Machinery directive
GOST (option) Russia
Metrology, measurement technology
UkrSEPRO (option) Ukraine
Metrology, measurement technology
- MTSCHS (option) Kazakhstan
Permission for commissioning
WIKA data sheet CT 31.06 ∙ 08/2020 Page 9 of 16
Certificates
Certificate
Calibration
CPB3800 Standard: Calibration certificate
Option: UKAS calibration certificate (pressure calibration with a mass set)
CPS5800 Standard: Calibration certificate
Option 1: UKAS calibration certificate (pressure calibration with a mass set)
Option 2: UKAS calibration certificate (area calibration)
CPM5800 Standard: without
Option 1: Calibration certificate
Option 2: UKAS calibration certificate (mass calibration)
Option 3: UKAS calibration certificate (pressure calibration with a piston-cylinder system)
Recommended recalibration interval 2 to 5 years (dependent on conditions of use)
Approvals and certificates, see website
Transport dimensions for complete instrument
The complete instrument, in its standard version and standard scope of delivery, consists of three packages on a single pallet.
The dimensions are 960 x 770 x 550 mm [37.8 x 27.6 x 21.6 in].
The overall weight is dependent on the measuring range.
Standard units CPB3800
Version Weight Version Weight
net gross net gross
1 … 120 bar 74 kg [163.2 lbs] 83 kg [183 lbs] 10 ... 1,600 lb/in² 70 kg [154.4 lbs] 79 kg [174.2 lbs]
2.5 … 300 bar 82 kg [180.8 lbs] 91 kg [201 lbs] 25 ... 4,000 lb/in² 79 kg [174.2 lbs] 88 kg [194 lbs]
5 … 700 bar 90 kg [198.5 lbs] 99 kg [218.3 lbs] 50 ... 10,000 lb/in² 90 kg [198.5 lbs] 99 kg [218.3 lbs]
10 … 1,200 bar 82 kg [180.8 lbs] 91 kg [201 lbs] 100 ... 16,000 lb/in² 79 kg [174.2 lbs] 88 kg [194 lbs]
Optional CPS/CPM5800 units with CPB3800 base
Version Weight Version Weight
net gross net gross
Single-piston measuring ranges Single-piston measuring ranges
1 … 120 bar 81.5 kg [179.7 lbs] 90.5 kg [199.6 lbs] 10 … 1,600 lb/in² 77 kg [169.8 lbs] 86 kg [189.6 lbs]
2 … 300 bar 77 kg [169.8 lbs] 99.5 kg [219.4 lbs] 30 … 4,000 lb/in² 77 kg [169.8 lbs] 86 kg [189.6 lbs]
Dual-piston measuring ranges Dual-piston measuring ranges
1 … 60 bar / 90.5 kg [199.6 lbs] 99.5 kg [219.4 lbs] 10 … 800 lb/in² / 89.5 kg [197.3 lbs] 98.5 kg [217.2 lbs]
10 … 700 bar 100 … 10,000 lb/in²
1 … 60 bar / 82.5 kg [181.9 lbs] 91.5 kg [201.8 lbs] 10 … 800 lb/in² / 82.5 kg [181.9 lbs] 87 kg [191.3 lbs]
20 … 1,200 bar 200 … 16,000 lb/in²
WIKA data sheet CT 31.06 ∙ 08/2020 Page 10 of 16
Dimensions in mm [in]
(without masses)
3 4
]
09
4,
259 [10,2]
[1
8
35
Ø
156 [6,14]
2
22 [0,87]
401 [15,79]
5 1 Test connection
108 [4,25]
2 Rotatable feet
6
3 Dual-area spindle pump with star handle
7 4 Piston-cylinder system
5 High-pressure shut-off valve
397 [15,63]
8 6 Reservoir with plug screw
7 Low-pressure shut-off valve
133 [5,24]
8 Pressure generation control schematic
83 [3,27]
9 Oil collecting tray
145 [5,71] 143 [5,63] 10 O-ring 8.9 x 1.8
93 [3,66]
11 Adapter, see scope of delivery
12 Sealing ring USIT 10.7 x 18 x 1.5
Standard connection piston-cylinder system Test connection
11
10
12
WIKA data sheet CT 31.06 ∙ 08/2020 Page 11 of 16
CalibratorUnit model CPU6000
The models of the CPU6000 series are compact tools for use
with a pressure balance (dead-weight tester). In particular
when highly accurate measuring values, with measurement
uncertainties of less than 0.025 %, are required, complicated
mathematical calculations and corrections are necessary.
With the CPU6000 in combination with the WIKA-Cal (PC
software) all critical ambient parameters can be registered
and automatically corrected.
The CPU6000 series is made up of three
instruments
Weather station, model CPU6000-W
The CPU6000-W provides measured values such as
atmospheric pressure, relative humidity and the ambient CPU6000 series
temperature of the laboratory environment.
Pressure balance sensor box, model CPU6000-S
The CPU6000-S measures the piston temperature and
displays the floating position of the masses.
Digital multimeter, model CPU6000-M
The CPU6000-M fulfils the function of a digital multimeter
and power supply unit when electronic pressure transmitters
must be calibrated.
Typical application
WIKA-Cal PC software - Weight calculator
With the demo version of the WIKA-Cal software and a CPB
series pressure balance (dead-weight tester), the masses to
be applied and the corresponding reference pressure can be
determined. The pressure balance data (dead-weight tester
data) can be entered into the database manually or imported Model CPU6000-W, CPU6000-S, CPB5800 and PC with
automatically via an XML file available online. WIKA-Cal software
All ambient parameters and piston temperature can be
entered manually into WIKA-Cal or can be measured
automatically with the CPU6000 series, so that the highest
accuracy can be achieved. WIKA-Cal demo version can be
downloaded free of charge from the WIKA website.
Further specifications on the CPU6000 series can be found
in data sheet CT 35.02.
For details of the WIKA-Cal calibration software see data
sheet CT 95.10.
WIKA-Cal PC software - Weight calculator
WIKA data sheet CT 31.06 ∙ 08/2020 Page 12 of 16
Further pressure balances (dead-weight testers) within our calibration technology
programme
Hydraulic pressure balance, model CPB5800
Measuring ranges:
Hydraulic Single-piston measuring ranges:
1 ... 120 to 2 ... 300 bar
[10 ... 1,600 to 30 ... 4,000 lb/in²]
Dual-piston measuring ranges:
1 ... 60 bar / 10 ... 700 bar to
1 ... 60 bar / 20 ... 1,400 bar
[10 ... 800 lb/in²/ 100 ... 10,000 lb/in² to
10 ... 800 lb/in² / 200 ... 20,000 lb/in²]
Accuracy: 0.015 % of reading
up to 0.006 % of reading (optional)
For specifications see data sheet CT 31.11 Hydraulic pressure balance, model CPB5800
Pneumatic pressure balance, model CPB5000
Measuring ranges:
Hydraulic -0.03 ... -1 to +0.4 ... +100 bar
[-0.435 … -14 to +5.8 … +1,500 lb/in²]
Accuracy: 0.015 % of reading
0.008 % of reading (optional)
For specifications see data sheet CT 31.01 Pneumatic pressure balance, model CPB5000
Pressure balance for high pressure, model CPB5000HP
Measuring ranges:
Hydraulic 25 ... 2,500, 25 ... 4,000 or 40 … 6,000 bar
[350 ... 40,000, 350 ... 60,000 or
400 … 90,000 lb/in²]
Accuracy: 0.025 % of reading
0.02 % of reading (optional)
Pressure balance for high pressure,
For specifications see data sheet CT 31.51 model CPB5000HP
Pressure balance for differential pressure,
model CPB5600DP
Measuring ranges (= static pressure + differential
pressure):
Pneumatic 0.03 ... 2 to 0.4 … 100 bar
[0.435 ... 30 to 5.8 … 1,500 lb/in²]
Hydraulic 0.2 ... 60 to 25 … 1,600 bar
[2.9 ... 1,000 to 350 … 23,200 lb/in²]
Accuracy: 0.015 % of reading
0.008 % of reading (optional)
Pressure balance for differential pressure,
For specifications see data sheet CT 31.56 model CPB5600DP
WIKA data sheet CT 31.06 ∙ 08/2020 Page 13 of 16
Accessories
Set of fine increment masses M1 and F1
The masses included in the standard mass set are ideally
suited for everyday use. If even smaller intermediate values
need to be generated, using one of the class M1 or F1
set of fine increment masses with the following masses is
recommended:
1 x 50 g, 2 x 20 g, 1 x 10 g, 1 x 5 g, 2 x 2 g, 1 x 1 g,
1 x 500 mg, 2 x 200 mg, 1 x 100 mg, 1 x 50 mg, 2 x 20 mg, Set of fine increment masses
1 x 10 mg, 1 x 5 mg, 2 x 2 mg, 1 x 1 mg
Test connections
With the existing standard test item connection, test items
with radial connection points can be mounted. For units with
rear connection points, a 90° angled connection piece is
available.
Fig. right: 90° angle connector
Separators
The separators (with diaphragm) have been specifically
designed for measuring instruments, which should not
come into contact with the medium of the pressure balance
(dead-weight tester) or to protect the pressure balance
(dead-weight tester) against contamination from the test
items.
Fig. left: Separator (with diaphragm)
700 bar [10,000 lb/in²]
Fig. right: Separator (with diaphragm)
1,200 bar [16,000 lb/in²]
Accessories
Special features Order code
CPB-A-BB-
Set of fine increment masses -4-
1 mg up to 50 g, class F1
1 mg up to 50 g, class M1 -5-
Carrying case -1-
for CPB3800 instrument base and piston cylinder system
Set of 2 carrying cases -D-
for bar mass set
Set of 2 carrying cases -E-
for lb/in² mass set
WIKA data sheet CT 31.06 ∙ 08/2020 Page 14 of 16
Special features Order code
CPB-A-BB-
Set of 3 storage cases -3-
for mass set, instrument base and piston-cylinder system
Adapter set -B-
“BSP” for G ½ B male test item connector to G ⅛, G ¼, G ⅜ and G ½ female
Adapter set -N-
“NPT” for test item connection G ½ B male on ⅛ NPT, ¼ NPT, ⅜ NPT and ½ NPT female
Adapter set -M-
“metric” for test item connection G ½ B male on M12 x 1.5, M16 x 1.5 and M20 x 1.5 female
90° angle connection -6-
for test items with back mounting connection
Threaded insert G ½ (½" BSP)
Test item connection -9-
G ¾ female to G ½ female, rotating
Separator -J-
to separate two liquid media by a diaphragm, max. 700 bar [10,000 lb/in²]
Separator -K-
to separate two liquid media by a diaphragm, max. 1,200 bar [16,000 lb/in²]
Sealing set -7-
for CPB3800 instrument base
Operating fluid -8-
for CPB series up to max. 4,000 bar [60,000 lb/in²], 0,5 litre
Electrical piston rotation unit -I-
AC 110 V for industrial power plug, 3 pole
only for measuring ranges 700 bar and 1,200 bar [10,000 lb/in² and 16,000 lb/in²]
Electrical piston rotation unit -F-
AC 230 V for industrial power plug, 3 pole
only for measuring ranges 700 bar and 1,200 bar [10,000 lb/in² and 16,000 lb/in²]
Tool set -H-
consisting of: open-ended spanner, BSP adapter, sealings, pointer remover and pointer
punch
Ordering information for your request
1. Order code: CPB-A-BB ⇓
2. Option: [ ]
WIKA data sheet CT 31.06 ∙ 08/2020 Page 15 of 16
Scope of delivery Options
■ Base ■ Systems with increased accuracy to 0.006 %
■ Dual-area spindle pump for filling, pressure generation ■ Further pressure transmission media
and fine pressure adjustment ■ Further pressure units
■ Piston connection with G ¾ B male thread ■ Set of fine increment masses (CPS/CPM5800 option only)
■ Test connection with G ½ female thread, loose union ■ Masses, manufactured to local gravity
connection ■ UKAS calibration certificate
■ Adapter set for test item connection consisting of:
Adapter set “BSP” with G ½ male on G ⅛, G ¼, G ⅜ and
G ½ female
■ Piston-cylinder system
■ Masses manufactured to standard gravity
(standard value: 9.80665 m/s²)
■ VG22 mineral oil (0.5 litre)
■ Tool and maintenance set consisting of:
1 hexagon wrench key 3 mm A/F
2 x 30 mm A/F open-ended spanners
1 spirit level
4 level plates
1 set of sealing rings
1 Pointer punch
1 Pointer remover
■ Operating instructions
■ Factory calibration certificate
Ordering information
Instrument base
CPB3800 / Instrument version / Accuracy / Gravity value g / Standard adapter set / Storage case / Calibration for pressure
balance / Additional ordering information
Piston-cylinder system
CPS5800 / Accuracy / Gravity value g / Measuring range / Connection of piston-cylinder system / Storage case for piston-
cylinder system / Calibration for piston-cylinder system / Additional ordering information
Mass set
CPM5800 / Pressure unit / Gravity value g / Standard mass set / Set of fine increment masses / Calibration for standard mass
set / Calibration for set of fine increment masses / Additional ordering information
© 03/2012 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specifications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modifications to the specifications and materials.
WIKA data sheet CT 31.06 ∙ 08/2020 Page 16 of 16
08/2020 EN
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30
63911 Klingenberg/Germany
Tel. +49 9372 132-0
Fax +49 9372 132-406
[email protected] www.wika.de