Ground tester
Thank you for purchasing a C.A 6416 or C.A 6417 ground tester. For best results from your instrument:
■ read these operating instructions carefully,
■ comply with the precautions for use.
Symbol Meaning
WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger symbol
appears.
Equipment protected throughout by double or reinforced insulation.
The product has been declared recyclable after analysis of its life cycle in accordance with the ISO14040
standard.
Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the
complete lifecycle has enabled us to control and optimize the effects of the product on the environment. In
particular this appliance exceeds regulation requirements with respect to recycling and reuse.
The CE marking indicates conformity with European directives, in particular LVD and EMC.
The rubbish bin with a line through it indicates that, in the European Union, the product must undergo
selective disposal in compliance with Directive WEEE 2002/96/EC. This equipment must not be treated as
household waste
■ The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be
taken in use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.
■ If you use this instrument other than as specified, the protection it provides may be compromised, thereby endangering
you.
■ Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
■ Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.
■ Before each use, check the condition of the insulation on housing. Any item of which the insulation is deteriorated (even
partially) must be set aside for repair or scrapping.
■ Use personal protection equipment systematically.
■ When handling the instrument, keep your fingers behind the physical guard.
■ All troubleshooting and metrological checks must be performed by competent and accredited personnel.
■ Avoid impacts on the measurement head, in particular the air gap.
■ Keep the surfaces of the air gap clean; even a little dirt can cause the clamp to malfunction.
®
Note: Bluetooth is a registered trade mark.
2
CONTENTS
1. GETTING STARTED ......................................................................................................................................................... 5
1.1 UNPACKING ................................................................................................................................................................ 5
1.2 INSERTING THE BATTERIES ..................................................................................................................................... 5
1.3 Setting the date and time.............................................................................................................................................. 5
1.4 EXAMPLE OF DISPLAY .............................................................................................................................................. 5
2. DESCRIPTION OF THE DEVICE ...................................................................................................................................... 7
2.1 FUNCTIONS OF THE DEVICE .................................................................................................................................... 7
2.2 FRONT PANEL ............................................................................................................................................................ 8
2.3 DEVICE–REAR PANEL ............................................................................................................................................... 9
2.4 DISPLAY UNIT ........................................................................................................................................................... 10
2.5 AUDIBLE SIGNALS.................................................................................................................................................... 11
3. MEASUREMENT PRINCIPLE ......................................................................................................................................... 12
4. USE.................................................................................................................................................................................. 13
4.1 INSERTING THE BATTERIES ................................................................................................................................... 13
4.2 COMMISSIONING THE DEVICE ............................................................................................................................... 13
4.3 SETTING THE INTERNAL CLOCK ............................................................................................................................ 13
4.4 STANDARD OR ADVANCED MODE ......................................................................................................................... 13
4.5 USE OF THE FUNCTIONS ........................................................................................................................................ 13
4.6 USE OF THE Hold KEY ............................................................................................................................................. 13
4.7 USE OF Pre-Hold ....................................................................................................................................................... 14
4.8 STORAGE OF THE DATA ......................................................................................................................................... 14
4.9 MANAGEMENT OF THE ALARMS ............................................................................................................................ 15
5. Ω+A POSITION ................................................................................................................................................................ 17
5.1 USE IN Standard MODE ............................................................................................................................................ 17
5.2 USE IN ADVANCED MODE ....................................................................................................................................... 17
5.3 COMPLEMENTARY INFORMATION ......................................................................................................................... 18
6. POSITION A .................................................................................................................................................................... 21
6.1 OBJECT ..................................................................................................................................................................... 21
6.2 PARAMETERIZING THE MEASUREMENT ............................................................................................................... 21
6.3 MEASUREMENT........................................................................................................................................................ 21
6.4 MEASUREMENT RESULT ........................................................................................................................................ 21
6.5 PRESENCE OF ALARMS .......................................................................................................................................... 21
6.6 MANAGEMENT OF THE ALARMS ............................................................................................................................ 21
7. MEMORY READ (MR) ..................................................................................................................................................... 22
7.1 OBJECT ..................................................................................................................................................................... 22
7.2 SELECTION OF THE memory read MODE ............................................................................................................... 22
7.3 DATA DISPLAYED ..................................................................................................................................................... 22
8. SET-UP ............................................................................................................................................................................ 25
8.1 OBJECT ..................................................................................................................................................................... 25
8.2 ACCESS TO THE SET-UP MENUS........................................................................................................................... 25
8.3 DISPLAY OF THE SET-UP MENUS .......................................................................................................................... 25
8.4 SELECTING A SPECIFIC MENU ............................................................................................................................... 25
8.5 DETAILS OF THE SET-UP MENUS .......................................................................................................................... 25
9. OFF POSITION ................................................................................................................................................................ 29
9.1 MANUAL SWITCHING OFF ....................................................................................................................................... 29
9.2 AUTOMATIC SWITCHING OFF ................................................................................................................................. 29
9.3 BACKUP OF THE CONFIGURATION........................................................................................................................ 29
9.4 PROLONGED STOP .................................................................................................................................................. 29
10. TECHNICAL CHARACTERISTICS ................................................................................................................................ 30
10.1 REFERENCE CONDITIONS .................................................................................................................................... 30
10.2 ELECTRICAL CHARACTERISTICS......................................................................................................................... 30
10.3 VARIATIONS IN THE DOMAIN OF USE ................................................................................................................. 31
10.4 POWER SUPPLY ..................................................................................................................................................... 31
10.5 ENVIRONMENTAL CONDITIONS ........................................................................................................................... 32
10.6 MECHANICAL CHARACTERISTICS ....................................................................................................................... 32
10.7 COMPLIANCE WITH INTERNATIONAL STANDARDS ........................................................................................... 32
3
10.8 Electromagnetic compatibility ................................................................................................................................... 32
11. SERVICING AND MAINTENANCE ................................................................................................................................ 33
11.1 Cleaning ................................................................................................................................................................... 33
11.2 Replacement of the batteries.................................................................................................................................... 33
11.3 CHECK OF ACCURACY .......................................................................................................................................... 33
11.4 ADJUSTMENT ......................................................................................................................................................... 33
12. WARRANTY .................................................................................................................................................................. 35
4
1. G E T T ING S TAR T E D
1.1 UNPACKING
Rep. Designation
1 Carrying case.
2 C.A. 6416 or C.A. 6417 earth clamp.
3 Set of 4 AA batteries (1.5V).
4 CD containing the GTC application and the operating instructions.
5 Verification certificate.
6 Multilingual safety data sheet
7 Multilingual quick start guide
Fig. 1
Accessories
CL1 calibration loop
DataView application software
Bluetooth modem
Replacement parts
Empty carrying case MLT 110
Set of 12 LR6 or AA batteries
Set of 24 LR6 or AA batteries
For the accessories and spares, consult our web site:
http://www.chauvin-arnoux.com
Fig. 2
5
The figure opposite shows a display, upon first use, with the device set to A.
The current measured is 30.0mA.
The buzzer is active and the memory is empty.
Fig. 3
6
2. DE S C R IP T ION OF T HE DE V IC E
The earth clamp is intended for testing the resistances of any conducting system having the characteristics of a conducting
loop. It is used to perform:
■ Earth resistance measurements if the earth is in series in a loop with its continuity conductor;
■ Other earth measurements: earth extended, for example by means of a guard wire between power transmission or
telecommunication poles;
■ Or the distributed earths of a single earth plane.
7
2.2 FRONT PANEL
Measurement head
1
Guard
Trigger
6
Fig. 4
8
Rep. Designation See §
1 Measurement head. -
2 Guard. The user's hand must be below this zone and not touch the measurement head (item 1). -
3 Function switch. 4.5
OFF: device off. 9
Ω+A: simultaneous selection of the Loop impedance measurement and the Leakage current 5
measurement.
A: selection of the Current measurement. 6
MR: (Memory Read) display of the data stored when MEM (item 8) was pressed. 7
SET-UP: access to configuration of the parameters and to erasure of the stored measurements. 8
4 HOLD key: freezes the measured and displayed values, at any time, along with the various functional 4.6
indications.
: Model C.A. 6417 only. When the function switch is set to MR or SET-UP, pressing this key
activates or deactivates the Bluetooth® connection.
5 OLED display unit. 2.4
6 Measurement head opening trigger. -
7 Dual-function key: -
(when device is set to Ω+A or A): increases the brightness of the OLED display unit; makes it -
easier to read the display unit in an environment with strong background illumination.
Highlighting activated for 30 seconds.
▲ (when device is set to SET-UP or MR): serves as up arrow when browsing in the menus and -
values. The brightness of the display unit does not change when device is set to SET-UP or
MR.
8 Dual-function key. -
MEM (when device is set to Ω+A or A): records the measured value. All of the data are recorded, in 4.8
the Standard or Advanced mode.
▼ (when device is set to SET-UP or MR): serves as down arrow when browsing in the menus and -
values.
9 ► Function that depends on the setting of the function switch, as follows:
When the device is set to Ω+A (Advanced Mode) 5.2.5
Short press: switches the display through the following 3 modes, in order:
■ Display of the impedance recalculated at the selected frequency.
■ Display of the contact voltage (product Z*I).
■ Display of R and L.
Long press: activates or deactivates the audible alarms. 2.5
when device is set to SET-UP -
Validation when browsing in the menus and values.
when device is set to MR (Advanced Mode)
Switches the display through the measurement screens and the measurement date/time.
Fig. 5
9
2.4 DISPLAY UNIT
Fig. 6
10
14 Signal indicating potentially hazardous voltage. Blinks when the contact voltage exceeds 50V. -
Rep. Designation See §
15 C.A. 6417: Remains displayed while the Bluetooth connection is being established. Blinks for the duration of -
the communication.
16 Memory index display unit. 4-digit digital display (0 to 9999 counts): -
■ Of the measurement sequence number, of the current memory in normal operation associated with the
Read (MR) or Storage (MEM) indications.
■ Of the date stamp (year) when the device is parameterized.
17 Memory read mode. 7
18 Data storage mode. 4.8
19 In the Advanced mode, these symbols identify the value displayed (resistance or impedance). 5.2
20 Main display unit: -
■ Measurement of the impedance or voltage.
■ Display of the date and time (month-day and hour-minute) in parameterizing and stored values read
mode.
21 In Advanced mode, indication displayed when the inductive component is negligible with respect to the 5.2.5
resistive component.
22 Indicates selection of the Advanced mode. 5.2
Remark: when switched on, the device performs a rapid self-test of the entire display unit. All available segments are
displayed briefly. During this stage, a sustained press on HOLD prolongs the display of all segments.
icon Meaning
Visible Buzzer activated; an alarm or a key press causes an audible signal to be emitted.
Missing No audible signal is emitted.
This programming is backed-up and retrieved at each reset. The audible alarm can be de-activated in the SET-UP menu (see
chapter 8, menu 2).
During a measurement, a long press on the ► key toggles between activation and deactivation of the buzzer.
Since the measurement frequency is audible, the operator hears a discontinuous audible signal (beep-beep). This is neither an
operating fault nor an alarm, and it cannot be eliminated. This audible signal is amplified by the presence of current in the loop.
11
3. ME AS UR E ME NT P R INC IP L E
The schematic diagram below illustrates the general case of measurement of the resistance of a loop comprising:
■ The earth electrode Rx;
■ The earth;
■ Several earth electrodes of resistance Ri;
■ A guard wire looping all of these earths, introducing an inductive component.
The clamp combines two functions in the measurement head:
■ The generator winding of the clamp emits an alternating voltage having a constant level E.
■ The receiver winding (current measurement) senses I=E/Zloop.
Knowing E imposed by the generator and I measured, the value Zloop can be deduced and displayed on the device. The
Advanced mode makes it possible to distinguish the resistive and inductive parts and to refer the impedance to the network
frequency.
More generally, this principle can be used to search for a defective earth. This is because the loop resistance comprises:
■ Rx (the value sought);
■ Zearth (a value that is normally very low, less than 1Ω);
■ R1//R2...// Rn (a negligible value: case of multiple earths in parallel);
■ Zguard wire (a value that is normally very low, less than 1Ω).
■ Rloop=Rx+Zearth+(R//R...//R )+Zguard wire;
As an approximation, Zloop can therefore be treated as equivalent to Rx.
If this value is very high, an inspection of this earth electrode is strongly recommended.
Z guard wire
Z earth
Fig. 7
12
4. US E
Set the date and time. The year, month, date, display mode (AM/PM - display from 01:00 to 12:00, symbol A. or P. or 24-hour,
symbol 24H), and time. Modify the blinking value with ▲ or ▼ and validate by ►. At the end of the procedure, the display unit
displays the screen associated with the function selected (Ω+A, A, MR or SET-UP).
The operator must change between standard and daylight saving time manually.
13
With the HOLD state active:
■ The ► key is active and can be used, in Advanced mode, to display the various measurement screens.
■ The MEM key is active and can be used to record the values displayed.
■ The HOLD key is used to exit from the HOLD state. The icon goes off and the device returns to the previous function.
4.8.1 CONDITIONS
Storage of the data is available in both the Ω+A and A measurement modes, provided that memory locations are free.
14
4.8.5 READING THE STORED DATA
The data can be looked up using the MR function. See chapter 7.
Fig. 8
Fig. 9
Fig. 10
15
4.9.4 IMPEDANCE ALARM
If there is no voltage alarm, no detection of NOISE, and no current alarm, an alarm on the impedance may be triggered. If the
buzzer is active, the corresponding audible signal is emitted.
Fig. 11
Fig. 12
If the impedance crosses the selected threshold, a low-pitched audible signal is emitted.
16
5. Ω+A P OS IT ION
Since the measurement frequency is audible, the operator hears a discontinuous audible signal (beep-beep). This is neither an
operating fault nor an alarm, and it cannot be eliminated. This audible signal is amplified by the presence of current in the loop.
5.1.1 OBJECT
In the Standard, mode, only one measurement screen is proposed. The clamp measures the loop impedance (Ω) at the fixed
frequency of 2083Hz and the leakage current.
5.1.3 MEASUREMENT
■ Place the conductor of the circuit to be measured in the clamp and close the clamp. If the clamp is incorrectly closed, the
icon is displayed.
■ If necessary, use the HOLD key to freeze the measurement. See §4.6,
■ If necessary, use the MEM key to store the measurement. See §4.8.2,
Remark:
If the measured impedance is less than 1Ω, the measurement display alternates between the value measured and the word
LOOP, in order to call the user's attention to the risk of measuring a local loop at the test point that does not include the
earthing.
Fig. 13
5.2.1 OBJECT
In this mode, 3 measurement screens are provided (impedance referred to the chosen frequency and leakage current, contact
voltage, display of R and L). The clamp measures the loop impedance (Ω) at the frequency of 2083 Hz. However, in addition to
what is done in the Standard, mode, the impedance is recalculated at the frequency defined by parameterizing.
5.2.2 SELECTION
The selection of this mode allows the display of additional measures only with switch on Ω+A position.
17
The selection of the Advanced mode is described in §8.5, menu 9.
The selection of the measurement frequency is described in §8.5, menu 10.
5.2.4 MEASUREMENT
■ Place the conducting wire of the circuit to be measured in the clamp and close the clamp. If the clamp is incorrectly closed,
the icon is displayed.
■ If necessary, use the HOLD key to freeze the measurement. See §4.6.
■ If necessary, use the MEM key to store the measurement. See §4.8.2.
Fig. 14
Second screen
Press ► to display the 2nd screen, which indicates the contact voltage (product
ZxI).
Fig. 15
Third screen
Press ► to display the 3rd screen, which indicates the values of R and L.
■ The loop inductance and loop resistance are displayed.
Fig. 16
■ When the inductive component is negligible (*) with respect to the resistive
component, the symbol R=Z is displayed, and only the impedance is
displayed; the inductance is replaced by dashes.
5
(*) R > 25Ω or R[Ω] / L[H] > 10 .
Fig. 17
18
5.3.1 PRODUCT ZXI GREATER THAN 50V
In this case:
■ The blinking Noise symbol is displayed.
■ The impedance blinks.
■ The hazardous voltage symbol blinks.
Fig. 18
Fig. 19
Fig. 20
Fig. 21
19
If the current is greater than 40A, the current display indicates O.R (Over
Range).
Fig. 22
20
6. P OS IT ION A
6.1 OBJECT
In this mode, the clamp measures the electric current, independently of any earth measurement.
6.3 MEASUREMENT
■ Place the conducting wire of the circuit in which the current is to be measured, in the clamp and close the clamp. If the
clamp is incorrectly closed, the icon is displayed.
■ If necessary, use the HOLD key to freeze the measurement. See §4.6.
■ If necessary, use the MEM key to store the measurement. See §4.8.2
Fig. 23
Fig. 24
21
7. ME MOR Y R E AD (MR )
7.1 OBJECT
The MR position (Memory Read, display of stored data) is used to display measurements previously stored by pressing the
MEM key.
Fig. 25
The stored values are displayed as they were when recorded, namely, same
display range, alarm states, NOISE signal, battery condition, etc.
However, the audible alarms are not reproduced; only the AL icon and the alarm
threshold blink.
The figure opposite illustrates a current measurement (A setting).
Fig. 26
Press ► to display the measurement storage date-time screen.
To exit from the memory read mode, set the rotary switch to the desired mode.
Fig. 27
22
7.3.2 DATA DISPLAYED IN ADVANCED MODE
The icon indicates use of the Advanced mode; the user then disposes of 4 distinct screens.
Screen no. 1
The last measurement is displayed, namely the impedance referred to the
chosen frequency.
The MR memory read symbol and the sequence number of the record being
read are also displayed.
The figure opposite illustrates an impedance and current measurement.
Press ► to display the next screen.
Fig. 28
Screen no. 2
The figure opposite illustrates a contact voltage measurement (product ZxI).
Press ► to display the next screen.
Fig. 29
Screen no. 3
The figure opposite illustrates a resistance and impedance measurement
(switch set to Ω+A).
Press ► to display the next screen.
Fig. 30
Screen no. 4
The figure opposite illustrates the date and time of the measurement (switch set
to Ω+A), namely:
■ 12.30: December 30.
■ 15:39: 1539 hours.
Press ► to return to screen no. 1.
To exit from the memory read mode, set the rotary switch to the desired mode.
Fig. 31
23
7.3.4 ERASURE OF THE STORED DATA
Refer to §8.5, menu 1.
Once the model has been identified, it appears in the list of paired peripherals.
3) Activation of the wait for connection of the clamp:
■ Set the switch on the clamp to MR or SETUP and press the HOLD button.
■ On the screen of the clamp, the Bluetooth icon lights. Steady lighting indicates waiting for connection to the
PC.
4) The connection between GTC and the clamp
At start-up, GTC requests selection of the peripheral to be connected. Depending on which Bluetooth adapter is
installed, it will propose either the serial port associated with the peripheral (e.g. Com40) or the name of the Bluetooth
peripheral (e.g. GT -CA6417_).
Once connected, the Bluetooth icon flashes on the screen of the clamp. GTC displays the status of the clamp
(model, serial no., firmware version) and the date and time of the earth clamp:
Refer to the on-line help of the software for a description of its operation.
Setup is accessed to configure the elements accessible in the clamp setup. It also serves to personalize the name of the
Bluetooth peripheral when the user has several earth clamps.
24
8. S E T-UP
8.1 OBJECT
The SET-UP position gives access to the following menus:
No. Function
1 Erasure of the memory.
2 Activation/Deactivation of the buzzer.
3 Activation/Deactivation of automatic switching off.
4 Setting of the impedance alarm threshold (Ω).
5 Setting of the voltage alarm threshold (V).
6 Setting of the current alarm threshold (I).
7 Setting of the date.
8 Setting of the time.
9 Selection of the Standard or Advanced operating mode.
10 Choice of transposition frequency for the impedance.
11 Activation/Deactivation of the Pre-Hold mode.
12 Display of the version number.
- Access to the 2 adjustment procedures (menus 13 to 14) and to the restoration procedure (menu 15).
13 Impedance measurement adjustment procedure.
14 Current measurement adjustment procedure.
15 Restoration of factory values.
Fig. 32
Key Action
▲ Move up in the menu tree.
▼ Move down in the menu tree.
► Select the menu displayed or return to the previous menu.
When changes have been made in one of the SET-UP menus (other than deletion), the changes can be cancelled by turning
the function switch to a position other than SET-UP, provided that there has not been a return to the main menu (press on ►).
In order to facilitate working with these menus, the procedure for access to each menu is systematically reproduced.
25
Menu no. Indication Object and use
1 CLr Erasure of the memory
■ Enter this menu by ►. Clr blinks.
■ Press ▲ and ▼ simultaneously for 6 seconds. The recorded data are all erased. The meter
indicates MEM 0.
■ Return to the previous menu by ►.
26
Menu no. Indication Object and use
8 HOUR Setting the time
■ Enter the menu by ►. HOUR blinks.
■ Press ▲ or ▼ to select the AM/PM (A. or P.) or 24H display mode (24H), which blinks.
Validate by ►.
■ Press ▲ or ▼ to select the hour, which blinks. Validate by ►.
■ Press ▲ or ▼ to select the minutes, which blink.
■ Validate by ►. The return to the previous menu is immediate.
27
Menu no. Indication Object and use
- End CAL.R FAIL: invalid measurement loop calibration.
■ Storage and return to CAL R menu 13 by ►.
14 CAL. Current measurement adjustment procedure
I Additional equipment necessary
■ A source of stabilized 0.1 and 10A current.
Adjustment of the current measurement
■ Press ► for 3 seconds to validate CAL. I function 14.
The sensitivity of the channels used in the current measurement will be recalculated on 2
current values.
■ PreS rt is displayed followed by 100.0mA Set.
■ Open the clamp and in it place a conductor carrying a current between 50mA and 150mA
from the current source.
■ Use the ▲ and ▼ keys to set the value displayed to match the value of the source.
■ Press ► to validate the value.
■ The display unit indicates run CAL.I for approximately 15 seconds.
■ PreS rt is displayed followed by 10.00A Set.
■ Set the current source to between 9A and 10.5A.
■ Use the ▲ and ▼ keys to set the value displayed to match the value of the source.
■ Press ► to validate the value.
■ The display unit indicates run CAL.I for approximately 15 seconds.
■ The result of the procedure is displayed:
- End CAL.I PASS: valid measurement loop calibration.
- End CAL.I FAIL: invalid measurement loop calibration.
■ Storage and return to CAL I menu 14 by ►.
15 CAL. Restoration of the factory values
dFLt ■ Press ► for 3 seconds to validate CAL. dFLt function 15.
The settings of the device will all be returned to the factory values. However, the stored
configuration and measurements are not erased.
■ PreS rt is displayed. Press ►.
■ End dFLt PASS is displayed:
■ Storage and return to CAL dFLt menu 15 by ►.
Note: if an adjustment operation fails (FAIL message), check that there is nothing interfering with the closing of the clamp,
then repeat the operation. If the problem persists, the clamp must be sent back for repair (see §11.6).
28
9. OF F P OS IT ION
The device can be switched off manually or automatically.
29
10. T E C HNIC AL C HAR AC T E R IS T IC S
30
10.2.4 CURRENT MEASUREMENT
Measurement range:
- Ammeter function: 0.2mA to 40A. 4000-counts display.
(1)
: Ω designates the quantities R, L, and Z.
(2)
: Offset on the current measurement.
31
10.5 ENVIRONMENTAL CONDITIONS
Fig. 33
1. Domain of reference (IEC160).
2. Domain of operation.
3. Domain of storage (without batteries). According to IEC359 category II (equipment for indoor and outdoor use).
32
11. S E R V IC ING AND MAINT E NANC E
Except for the batteries, the instrument contains no parts likely to be replaced by personnel who are not specially
trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may gravely impair
safety.
11.1 CLEANING
Disconnect everything connected to the device and set the switch to OFF.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry rapidly with a dry cloth or forced air. Do not use
alcohol, solvents, or hydrocarbons.
Keep the air gap of the clamp perfectly clean.
11.2.1 PROCEDURE
■ Disconnect everything connected to the device and set the switch to OFF.
■ Use a cross-headed or plain screwdriver to unscrew the 2 attachment screws (Fig. 5, item 10) and remove the battery
compartment cover (Fig. 5, item 11).
■ Withdraw the old batteries and replace them with 4 new batteries having identical characteristics (LR6, AA, 1.5V); take
care with the polarities.
Note: the alkaline batteries can be replaced by Ni-Mh rechargeable batteries (AA, 1.2V) having similar characteristics.
However, the time between the low battery message and the automatic switching off of the device will be shorter with the
rechargeable batteries.
Spent batteries must not be treated as ordinary household waste. Take them to the appropriate recycling
collection point.
■ Close the battery compartment cover and screw the two screws back in.
■ Check the proper operation of the device.
11.3.2 EQUIPMENT
Standard loop resistance available as an accessory. Simulates 5 loop resistance values.
11.3.3 PROCEDURE
Place the calibration loop in the jaws of the clamp. Set the function switch of the device to Ω+A, then compare the
measurement displayed with the value marked on the segment placed in the clamp. Repeat this process for each standard
value of the calibration loop.
Depending on the measurement errors found, you may decide that your clamp must be calibrated. The first step is to carry out
the adjustment procedure described in §11.4, before getting in touch with your supplier.
■ Standard loop values: 7.9Ω / 12.4Ω / 22Ω / 49.5Ω / 198Ω.
■ Accuracy of these values 0.3% typical, 0.5% max.
Remark: the accuracy of the device must be added to the accuracy of the standard values.
11.4 ADJUSTMENT
11.4.1 OBJECT AND EQUIPMENT NECESSARY
Regular adjustment is required; the more intensively the device is used, the more often it will have to be adjusted.
33
The user can perform 2 adjustment operations in addition to restoring the factory settings, directly on the clamp, in the SET-
UP position.
11.4.2 EQUIPMENT
Standard loop resistance available as an accessory. Simulates 5 loop resistance values.
11.4.3 PROCEDURE
Refer to §8.5, menu no. 13, 14, 15.
34
12. WAR R ANT Y
Except as otherwise stated, our warranty is valid for twelve months starting from the date on which the equipment was sold.
Extract from our General Conditions of Sale, provided on request.
The warranty does not apply in the following cases:
■ Inappropriate use of the equipment or use with incompatible equipment.
■ Modifications made to the equipment without the explicit permission of the manufacturer's technical staff.
■ Work done on the device by a person not approved by the manufacturer.
■ Adaptation to a particular application not anticipated in the definition of the equipment or not indicated in the user's manual.
■ Damage caused by shocks, falls, or floods.