Moisture Analyser
USER MANUAL
I Instruction Tips:
1. Security 1. Warm-up the Meter 30 minutes and take the first time
Please reading the following before you use the Moisture Meter calibration;
a. The instrument can use for test the moisture of the sample,each 2. After the end of a water calibration experiments (see 3.6);
improper operation may result in personal injury and 3.Each time the switch machine, please interval of 5 seconds;
equipment damage 4 .Cover the heater, do not leave the weight on the pan, or
b. Please access the AC power accordance with the manual; halogen lamps will be crushed;
c. Please make sure the power cord does not interfere with people
action;
d. Do not use the instrument under the hazardous,wet or unstable
enviroment;
e. Unplug the power before you clean the Meter;
f. Please keep enough space around the instrument,and keep the top of the
space at least 1m;
g. During a special experiment, please wear protective equipment by youself;
h. Please do not change the instrument's components and any other aspects.
Maintenance of this equipment by the factory authorized professionals。
Moisture Meter use Halogen lamp.
i. Do not put the flammable materials around the instrument;
j. Do not touch the halogen lamp when it on working.
Special attention to the dangers of some sample;
k. Before you test the moisture of dangerous good,pleaes take Full
investigation
Inflammable and explosive materials:Heating the inflammable and
explosive materials will produce flammable gases or vapors, so
need to lower the temperature in a dry environment. In case of fire
or explosion.
Toxic and corrosive substance: Keep the samples of toxic or
corrosive ingredients in well-ventilated environment.
Above the sample test will be particularly careful, for any damage
caused by your own risk
1
Schedule: assembly drawing II.Installation
2.1 Standard Packing List
Please check whether the packaging and accessories is complete when you open it
Standard Quantity Note
Moisture Meter Mainframe 1 set
Wind Cover 1 piece
The sample pan holder 1 piece
Tray handle 1 piece
Power cord 1 piece
Aluminum sample pan 1 set 50pcs
100g standard weight 1 piece
Product Manual 1 piece
Product certification 1 piece
Warranty card 1 piece
2.2Operating Requirement
Operating table Temperature stable, good ventilation
should be stable dry, without strong vibration, of the environment
without strong magnetic field.
2.3 Installation of accessories
Please installation the instrument according the picture
Installation order from top to bottom :1 --2--3--4--5
1. Aluminum scale pan
2. The triangle scale frame
Place the wind cover, Place the sample pan Put the aluminum
3. Scale pan brackets
aligned holes. holder, rotate until sample pan on the
4. Wind cover
positioning. bracket.
5. Stainless steel plate
9
2
2.4 Adjust the level and connect the power 3.5 Calibration
There are two feet in front of the instrument,adjust it and keep the When the instrument error or improper operate. Calibration weights can be
instrument level. adjusted with accuracy.
Press "Reset" key and does not release until it display“-cal-”,
Then flashing “100.000”, put on the calibration weight, then it will display "= = =
Connect the instrument with power,put one side of power cord = = =",
until it display “100.000” remove the weight, it will display "= = = = = = "again,
Waiting it show" 0.000 ", the calibration is completed.
Please warm up more than 30minutes for keep it stable and accurate. IV Technical parameters
Mode 110MW 105MW 102MW 100MW
III Operation
Specification 110g/10mg 110g/5mg 110g/2mg 110g/1mg
3.1.Display Readability 10mg 5mg 2mg 1mg
Pan size Φ 90
Heating mode Halogen lamp
Display LCD
Temperature Setting 40℃~199℃
Time Setting 1~99 minutes
3.2.Icon Introdution ATRO 100%~999%
ATRO 0%~999%
----The percentage of dry weight ----The percentage of moisture
History store 15 set
Power supply 220V 50Hz
----Gram ----Temperature
---- Time set and heat time
----Heated mode
----Standard mode ----Rapid heating mode
---- Gentle heating mode
---- Stop heated mode
8
3
when flashing .Press"UP"or"Down"to choose the stop mode.and press ---- Automatic -----Manual stop mode
"ENTER" confirm , Parameters are defined, and go to the next parameter ----Timing stop mode
(temperature) setting.
3.5.4 Heating Temperature Setting ---- Heated state
The instrument temperature setting range is from 0-199 degrees. Press the 3.3. Keypad
"MENU" to enter the parameter setting mode, press the "ENTER" key
until the flashing , Press"UP"or"Down"to choose the
temperature.and press "ENTER" to go to next setting(time setting)
Icon will transform other temperature when its heating.
3.5.5 Heating time setting
1.Just choose timing stop mode need set the stop time.
Press the "MENU" to enter the parameter setting mode, press the
"ENTER" key until the flashing , Press"UP"or"Down"to
choose the temperature.and press "ENTER" NAME Key features
2, Heating cycle:Just gentle heating mode need choose heating cycle
ON/OFF Power on / off, Just off the internal DC power supply
3.6 For example:
a. sample:water, START / STOP Start heated/ stop heated
b. Press the "MENU" to enter the parameter setting mode,choose the
TARE Weighing peeled, Settings exit
( ),temperature:120℃,Press "ENTER" go to weighing state MENU Enter the setting menu
c. Put a dry paper on the pan,and press"Tare" clean the weight. CONVERSION Switching measurement results
Put on the 3g water on the paper,until the readability stable. ENTER Confirm the current setting, go to next
Press" START / STOP " begin test. Return to the weighing mode (short press), calibration
Waiting for it sop.look the result. RESET
balance (long press)
The m% will be 100%.
DOWN Reduce
UP Add
Weighing error or sample weight deficiency can lead to the result of the
test a deviation
7
4
3.4 First testing
a. Standard heated mode:( )
1. First test, if not change the temperature and time settings, the moisture meter
defaults to factory defaults. This mode is the factory setting, suitable for most samples. Heating
2. Put empty sample pan on the holder (see section 2.3). temperature can be set, the sample is heated to a temperature, and keep to
3. Press the “TARE” to clean weight. the test ends automatically.
4.Put the sample on the pan (Sample weight should be heavy than 1g, the general b. Rapid heating mode: ( )
3-5g). This mode is suitable for samples with higher moisture content. You can
5. Close the heating cover, press the "Start / Stop" button to start the test. At this set the heating temperature, heating the temperature will exceeds the set
point, heated tube lights, fans turn on, heated symbol in the upper left flashing. value, the future will be reduced to the set temperature and hold until the
While rolling on the screen displays the current temperature, has been heating time, test ends automatically.
moisture content of the current sample. c. Gentle heating mode: ( )
6. End of the test, heated lamp automatic shut off, the buzzer alarm. Then the This mode is suitable for low moisture content, surface casing is easy to
sample is displayed on the moisture content, heating time, current temperature form, easy to burn the sample. From the heat began to rise to the long
and the parameter setting mode. time required to set temperature, heating up even slowly. Heating time
Note: Don't touch the sample plate,the temperature is too high now. user selectable.
7 Press the "Reset" button, the instrument returns to weighing mode. d. Press the "MENU" to enter the parameter setting mode, 0 to 1 ,press the
3.5 Parameters Setting "ENTER" until the heating mode, the icon flashes 。 Press
3.5.1 Heated parameter storage and recall "UP"or"Down"choose the mode you need.and press "ENTER"confirm.
a. The instrument can store a total of 16 0-F group of heated parameters. Each 3.5.3 Stop mode setting:( )
group can set the heating method, stop mode, time and time test parameters. a. Automatic stop mode:( )
b. In weighing mode, press the "Settings" button to enter the parameter setting Heating the sample within the specified time during the weight loss of
mode, then store the address number (the number show the highest bit) is less than 2mg will automatically stop when the test. This mode after 15
flashing. Press "down" or "up" key to select the desired address number, the seconds in the heating effect.
0 is the factory setting,1-F for the user store. b. Manual stop mode:( )
c. The memory address number 0 means factory settings, then click "ENTER" Sample heating process is the continuing until the press the "start/stop"
button to exit the settings back to weighing mode. At this point heated mode button to end testing.
is the factory defaults. c. Timing Stop Mode:( )
d. When the memory address range for the 1-F any one number, press "ENTER" Stop time set by the user, the sample heating process is continued until the
key, and enter the parameter setting after the completion of a parameter end of the heating time to set the value of the test.
(heated mode, stop mode, time mode) setting.Press the "TARE" key, the
address number of the test parameters have been stored. Press the "MENU" button to enter the parameter setting mode, press the
3.5.2 Heated mode and setting( ) "ENTER" key until the shutdown mode
5 6