4 - HPLC - Part 1
4 - HPLC - Part 1
High Performance
Liquid Chromatography (HPLC) - 1
2
Introduction
Introduction
Qualitative HPLC: Separation and Identification of the sample composition
(identification of the compounds using reference standards ) using analytical column
(µg)
Quantitative HPLC: Separation and Quantification of the sample components
using analytical (µg) (using calibration curve), semi-preparative (mg) or
preparative columns (g).
Strengths:
Precise
Unlike GC, no heat is required (i.e. less sample degradation)
Various types of columns (analytical, preparatory and semi
preparatory)
Sensitive (traces of the analytes can be detected)
Automated
Limitations:
Requires large amount of organic solvents
Requires specific detectors that monitor compounds that lack
3
a chromophore
Instrumentation
tR
Peak
Signal
h
A
Retention Time
Controller /
Data processing
Degasser
Waste or
Fraction collector
Solvent reservoirs
Instrumentation
1. Solvent reservoirs.
contain the solvents used to carry the sample
through the system. The solvent should be filtered
with an inlet solvent filter to remove any particles
that could potentially damage the system's
sensitive components.
Instrumentation
2. Degasser.
HPLC degassing of solvents, is very essential for proper operation and
retrieval of uniform data.
Degassing removes entrapped air bubbles from the solvents. Improper
degassing leads to many problems like fluctuation in pressure,
baseline fluctuation, decrease in sensitivity, improper peaks, improper
detection, clogging of column etc.
Vacuum chamber
To Pump
Mobile Phase
Pump head
Motor & Cam
Check valve
Plunger
Plunger seal
Mobile Phase
Instrumentation
Isocratic mode - delivers constant mobile phase composition
(composition remains constant with time)
• solvent must be pre-mixed
• lowest cost pump
• best for simple separations
LOAD INJECT
Instrumentation
from Pump
to Column
LOAD
from Pump
to Column
INJECT
Instrumentation
Autosampler:
1. User loads vials filled with sample solution into the autosampler
tray (100 samples)
2. and the autosampler automatically
measures the appropriate sample volume
injects the sample,
then flushes the injector to be ready for the next sample, etc.,
until all sample vials are processed …
… for unattended automatic operation
Instrumentation
From pump
to the column
waiste
Step 1 Step 2
Step 3
Instrumentation
5. HPLC Column:
Within the Column is where separation occurs.
Key Point – Proper choice of column is critical for success in HPLC
Stainless steel columns packed with porous materials
Instrumentation
Columns are packed with small diameter porous particles.
– The most popular sizes are: 5-μm, 3.5- μm and 1.8-μm
• Columns are packed using high-pressure to ensure that they are
stable during use (most users purchase pre-packed columns to
use in their liquid chromatographs)
• These porous particles in the column usually have a chemically
bonded phase on their surface which interacts with the sample
components to separate them from one another
Instrumentation
Optional:
Column Oven can control and maintain the constant
column temperature.
Tube
Aspirator
UV-Visible detectors
Photodiode Array Detectors (PDA)
Fluorescence detector
Electrochemical detector
Noise
this equation gives us an idea of what S/N is required for a desired method
precision.
For example, if the %RSD for a method's precision can be no larger than
2%, this means that S/N = 25 is required (this assumes that S/N is the
primary contribution to %RSD, which might or might not be true).
For bioanalytical methods, S/N = 2.5 would translate into %RSD ≈ 20%.
Repeatability: Closeness between the results of measurements of the
same sample carried out using following conditions:
The same measurement procedure
The same observer
The same measuring instrument used under the same conditions
The same location
Repetition over a short period of time
Blank samples: analyses the solvent/matrix without the analyte and ensures that
glassware/instruments/equipment are free from contamination.
Known samples/open control: uses a standard with known concentration in order
to validate the method and ensure accuracy.
Calibration check: Different concentrations of an analyte are analyzed and a
calibration curve is plotted. This curve has to be validated regularly using a known
analyte concentration.
Replicates: A subsample aliquot (taken from the original sample) analyzed to
determine the mean and standard deviation (%RSD).
Duplicates: Two original analyte samples are obtained and analyzed.
Spikes: A known concentration of an analyte is added to a sample matrix and then
analyzed to determine the percentage recovery. Often, deuterated compounds can
be used as spikes as they can be distinguished from the target analyte via MS
analysis.
Instrumentation
Qualitative Information
Chlortoluron Atrazine
? ?
200
58
215
44
68 172
ne
e
en
re
ne
py
th
yle
ne ne
ne
an
d)
er
e
re
or
-c
i)p
yr
py
flu
23
Pe e)p
e
gh
a)
k)
(1
en
ne
o(
o(
o(
o(
no
le
ys
nz
nz
nz
nz
ry
re
de
hr
Be
Be
Be
Be
Py
In
C
UV-signal
248/411
270/388
241/394
247/504
302/420
WL
WL
WL
Fluorescence
WL
WL
PAH's extracted from soil;
Flecainide in
Sup.LC-PAH 150x4.6mm; Serum
Solv.: H2O/CH3OH= 10:90
UV signal
FL signal
Cell C : Concentration
Iin Iout
D2 / W Lamps
l
Cell
Grating
D2 / W Lamps
1 nm / element
512…… photodiodes
Detectors
PDA detector Spectrum
Chromatogram
Absorbance
Retention time
Detectors
Fluorescence detector
Compared to UV-Vis detectors fluorescence detectors offer a higher sensitivity and
selectivity that allows to quantify and identify compounds and impurities in complex
matrices at extremely low concentration levels (trace level analysis).
• Fluorescence detectors sense only those substances that fluoresce
Excitation wavelength
+ hn1 *
* hn2 +
Emission wavelength
Excitation state
Quasi-excitation state
hn1
hn2
Fluorescence
Ground state
Detectors
Refractive Index (RI) detector
The ability of a compound or solvent to deflect light
• The RI is a measure of molecule’s ability to deflect light in a flowing mobile
phase in a flow cell relative to a static mobile phase contained in a reference
flow cell.
• The amount of deflection is proportional to concentration.
• The RI detector is considered to be a universal detector but it is not very
sensitive.
Photodiode
Reference
W Lamp
Sample
Detectors
Refractive Index (RI) detector
Reference cell
Reference cell
containing
containing
flowing solvent Sample cell
flowing solvent Sample cell
(mobile phase) containing only
(mobile phase) containing
mobile phase
mobile phase
with analyte
Calculate the resolution of the compounds in red
Calculate the retention factor for the last peak (the hold-up time (to) is
indicated by the vertical ‘tick mark’ on the baseline of the chromatogram).
Calculate the plate number of the column (from the last peak eluted).
to = 1.64 min