Innovality (FDR3500) - Mammo QC
Innovality (FDR3500) - Mammo QC
Control
1
Quarterly Test
5
Semi-annual
Test
6
Annual Test
7
Assistance
Material
8
897N120636
ii FDR Mammography QC Program Guidebook 897N120636
Introduction
This quality control program primarily uses quantitative measurements and provides
the ability to see gradual changes in X-ray equipment performances, as these
changes are difficult to notice by visual checks alone.
In this Guidebook, the contents are based on the premise that the 1 Shot Phantom
M Plus 24 ×30 will be used. Different Phantom sizes cannot be used. If the Phantom
of different sizes has been used before, be sure to re-create the baseline values and
to re-examine criteria.
Exclusive Clauses
1. No part or all of this Guidebook (except Chapter 8) may be reproduced in any form without
prior permission.
2. The information contained in this Guidebook may be subject to change without prior notice.
3. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performed by other than dealers
specified by FUJIFILM Corporation.
4. FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM
Corporation products due to products of other manufacturers not supplied by FUJIFILM
Corporation.
5. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
remodeling, maintenance, and repair using repair parts other than those specified by FUJIFILM
Corporation.
6. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of precautions and operating methods contained in this Guidebook.
7. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range of using conditions for this product such as
power supply, installation environment, etc. contained in this Guidebook.
8. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from natural
disasters such as fires, earthquakes, floods, lightning, etc.
Trademark
FCR and FDR are trademarks or registered trademarks of FUJIFILM Corporation.
Chapter 2 Overview 2
This chapter describes the features of FDR Mammography QC Program
and the purpose of each quality control test.
Conventions
The names of the buttons in the window are shown as described below.
(Example) [Start study.]
The button name is shown in brackets ([ ]) regardless of
the feature.
Chapter 2 Overview
2.1 Product Outline................................................................................. 2-2
2.2 QC Test Items................................................................................... 2-5
2.2.1 Daily/Weekly Test................................................................ 2-7
2.2.2 Quarterly Test........................................................................ 2-11
2.2.3 Semi-annual Test................................................................. 2-11
2.2.4 Annual Test............................................................................ 2-12
2.3 Tools...................................................................................................... 2-13
2.4 QC Software Outline....................................................................... 2-17
2.5 Notes on Conducting the Program........................................... 2-20
Quality Control
Page
Quality control is defined as “operational techniques and activities that are used to fulfill requirements
for quality”.
1 Detailed instructions for carrying out quality control (e.g. when and who carries out quality control)
are established as a quality assurance program. In addition to quality control techniques, training for
Quality Control
providing adequate information on quality control is included so that any quality assurance program
may be effectively implemented.
For detailed information on quality control and quality assurance programs, see IEC 61223-1 “Evaluation
and routine testing in medical imaging departments”.
Tests for quality control are called performance tests. There are three types of performance tests,
acceptance test, constancy test and status test, depending on their purpose or implementation
frequency.
International standards and guidelines regarding the quality control of the mammography system are
as follows.
1
• European guidelines for quality assurance in breast cancer screening and diagnosis; Fourth
Quality Control
Edition (European Commission) (Hereafter abbreviated as EUREF Ed.4)
• IEC 61223-3-2 Ed.2 (International Electrotechnical Commission)
• ACR Quality Control Manual (American College of Radiology)
In this Program, test items and their implementation frequencies for the mammography system using
the FDR are determined based on IEC 61223-3-2 Ed.2. Test items consist of quantitive tests and visual
inspections, and the 1 Shot Phantom M Plus 24×30 for quantitive measurement and QC Software are
used for evaluation.
This program is only applicable to 2D images, and cannot be used for 3D images.
Chapter 3
Although performance verification tests are not defined in IEC, these tests are similar to acceptance
tests and status tests since absolute values are used for evaluation. However, performance
verification tests are distinguished from the two tests above since the implementation frequency is
different and these tests are not based on the specification.
Quality Control
measurement environment, the same test is executed multiple times and its average value is defined
as a baseline value, and the variation of measured values is defined as an established criteria. When
the measured value is within the acceptable range (baseline value plus or minus criteria), the test
result is regarded as “Pass”. If it is out of the acceptable range, the test result is regarded as “Fail”.
Fail
Fail
Fail
Upper Limit
Test Results
Pass
Pass
Test Results
Lower Limit
Fail
1
Quality Control
the FCR systems and can produce the image quality test results on 10 items by single exposure.
For the FDR systems as well, 1 Shot Phantom enables wide-ranging analysis of a system in a
relatively short time with high reproducibility.
• Exclusive Quality Control Software (Mammography QC Software)
Mammography QC Software (hereafter QC Software) is designed for performing periodic quality
control, data analysis, and QC test log management.
[Calculation functions]
• CNR.......Used for calculating the values in AEC (Automatic Exposure Control) function
adjustment or when determining the CNR baseline values.
• AGD.......Used for calculating the values in AEC.
Overview
X-ray equipment Viewing box
Chapter 3
1. Initial Settings
2
4. Checking Equipment Qualities Annual Test
at the Time of the Program
Installation
Semi-annual Test
Overview
Daily/Weekly Test
Daily/Weekly Test
Semi-annual Test
6 months
Quarterly Test
Annual Test
Semi-annual Test
18 months
Quarterly Test
NOTE
All tests are to be performed with grid set to [IN], unless specified otherwise.
NOTE
Use a standard compression plate, unless otherwise specified. If a small or flex type compression plate is
used, tests may not be conducted successfully.
2
[Daily/Weekly Test]
Overview
Test Items Exposure Menus Responsibility
Test with 1 Shot Phantom
1 Shot PhantomM
Good practice Technologist
Test with ACR Phantom ACR Phantom
NOTE
Conduct the Test with 1 Shot Phantom and Good practice in the Daily Test, and the Test with ACR Phantom
in the Weekly Test.
[Quarterly Test]
Test Items Exposure Menus Responsibility
Repeat analysis –– *1 Technologist
[Semi-annual Test]
Test Items Exposure Menus Responsibility
Image Basic Test
Compression device confirmation Semi-annual 1/1 Technologist
Viewing box maintenance
X-ray equipment
Annual 3/6
Collimation assessment performance
Radiation output
AEC density control function
AEC reproducibility Annual 4/6 through AEC system
CNR modes 1 to 3 Annual 6/6 performance
AGD modes 1 to 3
*2 The Missed tissue on chest wall edge (40 mm height) in Annual Test is a required test item to be conducted in some areas.
NOTE
The responsibility depends on local regulations where the equipment is to be used.
[Calculation functions]
Test Items Exposure Menus
CNR Calculation 1/2
AGD Calculation 2/2
When an exposure menu includes multiple test items/contents, make sure to conduct the test on all of
them.
(If any of the test items/contents remain untested, the QC Software malfunctions.)
An exposure menu can be selected on the following window of a FUJIFILM workstation. When an
exposure menu is selected, the relevant exposure submenu(s) is displayed (Example: 1 Shot PhantomM).
See the Operation Manual for FUJIFILM workstation for detailed FUJIFILM workstation operations.
Overview
<Judgment method>
• Auto-calculation based on
the image
• Visual inspection
<Judgment method>
• Density measurement
• Visual inspection
NOTE
In this Program, the 1 Shot Phantom M Plus 24 ×30
must be used.
Overview
Check that the 1 Shot Phantom sensitivity Check that the uniformity variation is within
variation is within the criteria. the criteria.
(4) Geometric distortion (7) Dynamic range
Horizontal
direction
Vertical
direction
Check that the image dimension variation Check that the dynamic range variation is
(geometric distortion) in horizontal/vertical within the criteria.
direction is within the criteria. (8) Spatial Resolution (SR)
Visually check that there is no distortion or
jitter in the image.
(5) System artifact evaluation
2
Overview
NOTE
The Repeat analysis is a required test item to be
conducted in some areas. Conduct the analysis
in compliance with local regulations where the
equipment is to be used.
Overview
2.2.3 Semi-annual Test
The Semi-annual Test includes the Image basic
test, Compression device confirmation and
Viewing box maintenance.
[1] Image basic test
Check that the relative sensitivity variation, Check if the viewing box luminance is
the degree of artifacts in the image are satisfactory, the equipment functions
within the criteria. correctly and optimum conditions for viewing
are provided.
2
(Ghost)
Check the temporal sensitivity [4] Visual and Functional test
change due to X-ray exposure. Check the X-ray equipment
Overview
NOTE
The Missed tissue on chest wall edge (40 mm height) in Annual Test is the required test item to be conducted
in some areas. Conduct the test in compliance with local regulations where the equipment is to be used.
...Required ...Optional
Daily/Weekly Test Semi-annual Calculation
Tools Annual Test
Daily Weekly Test Functions
[ A ] QC Software
[B] Worksheet (See Chapter 8)
[C] 1 Shot Phantom 2
[D] Viewing box (if film is used)
Overview
[ E ] Timer or watch/clock
ACR Phantom
[F] *1
(including an acrylic disk)
[G] Step Phantom *2
[H] Dosimeter
[ I ] Luminance meter
[ J ] kVp meter
[ K ] Force scale
[ L ] Illuminance meter
[M] Densitometer (if film is used)
PMMA phantoms (available, in
[N] combination, for 20, 40, 60 and 70 mm
in thickness)
Aluminum plate (0.2 mm) for CNR
[O]
measurement
Aluminum plates for half value layer
[ P ] measurement
(0.5 and 0.6 mm)
[Q] Scale
[R] Coins
[ S ] Jig for 1 Shot Phantom *3
[ T ] Lead sheet *4 *4
[B] Worksheet
Worksheets for recording or documenting QC test measurement and judgment results are contained in “8.3 Worksheets”.
Make a copy of those worksheets as necessary.
NOTE
Scratches, deformation, or dirt on the ACR or Step Phantom not only affects the image evaluation or
measured values but may damage the compression plate of the X-ray equipment. Make sure that there
are no scratches, deformation, or dirt on the Phantom before use.
NOTE
A Step Phantom is required only when conducting visual inspection of it.
[H] Dosimeter
A dosimeter is used for measuring air kerma (exposure). Some dosimeters may require correction of the measured value
according to the beam quality. See the operation manual for the dosimeter for details. The dosimeter should be calibrated using
low energy (at least 10 keV to 40 keV), allowing air kerma rate or exposure time measurement, in addition to air kerma (exposure)
measurement.
Overview
A 99.9% or higher purity aluminum plate of sufficient size for covering ROI is required. The plate can be also used as [P]
Aluminum plates for half value layer measurement. Multiple plates of 0.2 mm or less in thickness can be used in combination.
[P] Aluminum plates (0.5 and 0.6 mm) for half value layer measurement
A 99.9% or higher purity aluminum plate of sufficient size for covering the detection surface of the dosimeter is required. The
plate can also be used as [O] Aluminum plate for CNR measurement. Multiple plates of 0.3 mm or less in thickness can be used
in combination.
[Q] Scale
A millimeter scale is required.
[R] Coin
Some familiar-sized coins are to be used.
1 Shot Phantom
40 mm Jig
NOTE
This item is required only in the Missed tissue on chest wall edge (40 mm height) measurement performed
on the exposure table. Prepare the jig if necessary.
B X-ray equipment
An X-ray equipment to be used is the FDR MS-3500.
NOTE
In this Program, methods of exposure condition setting of an X-ray equipment are defined as follows:
2 • “Manu (manual)” mode
The mode allows manual setting of the kVp, mAs and target/filter of the X-ray equipment.
• “Semi (semi auto)” mode
Overview
The mode automatically specifies the mAs but allows manual setting of the kVp and target/filter of
the X-ray equipment.
• “Auto (full auto)” mode
The mode automatically specifies kVp, mAs and target/filter of the X-ray equipment depending on
the subject.
When conducting a QC test, turn off the Intelligent AEC.
C FUJIFILM workstation
A workstation to be used is the FDR-3000AWS.
Exposure
Overview
“QC window” Film output etc.
Auto-calculation
Judgment
Operations on the QC Software
Patient information
display field
Exposure
QC Software startup
2
Overview
“QC window”
The following functions can be executed on the “QC window”. The relevant button functions are described below.
Exposure condition display field
See the “Mammography QC Software Operation Manual” and/or operation manual for a FUJIFILM workstation for
detailed QC software operation.
The results of measurement and visual inspection can be entered in the window below (Example: Daily/Weekly
Test).
Overview
(6) Measurement execution
Each test result is auto-calculated based on the entered image and values, and the judgment (Pass/Fail) is
displayed in the window as shown below.
Judgment results
Pass/Fail
Calculated results
The past measurement and judgment results of each test item/content are displayed in the graph.
Log
2 (2) Dosimeter
Position the X-ray detector of the dosimeter 40 mm above the exposure table unless indicated
otherwise. Make sure of the X-ray detection position of the dosimeter and adjust the height as
Overview
specified by using an appropriate jig. In the X-ray reproducibility test, the dosimeter position differs
since the air kerma is measured under the AEC.
(3) Compression plate
Some test items/contents require the compression plate to be installed, while some do not. See the
test procedure for each item.
Use a normal compression plate for mammography unless otherwise specified. Be careful not to
scratch the compression plate when conducting a test.
(4) Solutions for failed test items
Actions required when a test item judged as “Fail” are described for each test item.
(5) Artifacts
Artifacts may be seen in the Spatial Resolution chart on the 1 Shot Phantom. This characteristic
does not affect clinical images since there are no such special structures in a human body. Auto-
calculation is also not affected.
If the image processing parameters are changed, streak artifacts may appear on the image due to
the resolution of digital data. It is recommended to perform visual inspection by using the default
parameters set to each exposure submenu.
(6) AGD
AGD (Average Glandular Dose) values measured in this Program may be different from those in the
test results displayed on a FUJIFILM workstation.
In this Program, AGD values are measured based on the assumption that a uniform PMMA phantom
is used and enable the user to evaluate the AEC (Automatic Exposure Control) performance of the
X-ray equipment quantitatively. On the other hand, AGD values on a workstation are calculated for
actual AGD values per mammogram. Due to this difference, the parameters used for calculations
are different. Conduct tests by following the procedures in this Program.
(7) Regular backup
Regularly back up the results of the test conducted by using this Program.
See “FDR-3000AWS Operation Manual” for details.
(8) AEC exposure modes
To conduct a test item which requires the AEC in this program, use a mode most frequently used
for normal mammography (H-mode, N-mode or L-mode) unless otherwise specified. Also, turn off
the Intelligent AEC. Use the same exposure mode every time the tests which require the AEC are
conducted.
equipment is used and measurements are performed. For these items, conduct the tests several
times when installing the Program and set the averages of the measured values as the baseline
values.
NOTE
When a new equipment or system has been installed or existing equipment has been remodeled,
perform the procedures described in “3.4 Baseline Value Settings” and later sections in Chapter 3.
Setting Procedure
7 Enter values for items requiring initial
1 Enter patient information and press [Next]. settings.
For details on each item, see the following
descriptions in (1) through (5).
3
(1) (5)
(2)
(3) (6)
(7)
Select [QC/TEST] in the exposure region (8)
2
list and [1 Shot PhantomM] in the exposure
menu list, and then press [Start study.].
(1) Dosimeter - Exposure table distance
• Measure the distance (mm) in the figure below
and enter the value. The recommended value
for h is 40 mm (This Guidebook assumes h
is 40 mm in the following explanation). Set
the distance l between the center of the
dosimeter and chest wall-side edge of the
exposure table as 60 mm.
Reference point
8 Press [OK].
The criteria should be specified on your own 3 The following exposure submenu is
responsibility based on the measurement results displayed.
obtained at the time of the Program installation. • 1 Shot PhantomM
Enter the percentage or the absolute value of
variations from the baseline value for each item.
The reference criteria are shown on the “List
of Reference Values for Criteria and Baseline
Values”. Specify the following and make an
The reference criteria are preset for each item. 4
exposure in “Manu” mode.
The criteria can be changed by following “Criteria 23 kV, 2 mAs and W/Rh
Change Procedure” below.
NOTE
Criteria Change Procedure
When changing the criteria following “3.2 Initial
It may take some time until you can press [QC]
after completing the exposure process. 3
5 Press [QC] to display the “QC window”.
Settings”, start from Step 6 .
*1 When these items are not checked, leave the preset values as they are. Do not delete the preset values.
n Semi-annual Test
3
[Items allowing criteria change]
Collimation assessment X-ray / Image receptor field gap (Left) [mm] SID X ___ or less
X-ray / Image receptor field gap (Right) [mm] SID X ___ or less
X-ray / Image receptor field gap (Nipple) [mm] SID X ___ or less
X-ray field / Exposure table gap [mm] ___ mm or less
Reproducibility Variation coefficient ___ or less
Air kerma rate [mGy/s] ___ mGy/s or more
Radiation output
Specific radiation output [μGy/mAs] ___ μGy/mAs or more
AEC reproducibility
Entrance air kerma reproducibility Variation coefficient ___ or less
CNR relative value 20mm [%] Baseline value ___ or more
CNR relative value 40mm [%] Baseline value ___ or more
CNR modes 1
CNR relative value 60mm [%] Baseline value ___ or more
CNR relative value 70mm [%] Baseline value ___ or more
CNR relative value 20mm [%] Baseline value ___ or more
CNR relative value 40mm [%] Baseline value ___ or more
CNR modes 2
CNR relative value 60mm [%] Baseline value ___ or more
CNR relative value 70mm [%] Baseline value ___ or more
CNR relative value 20mm [%] Baseline value ___ or more
CNR relative value 40mm [%] Baseline value ___ or more
CNR modes 3
CNR relative value 60mm [%] Baseline value ___ or more
CNR relative value 70mm [%] Baseline value ___ or more
AGD 20 mm [mGy] ___ mGy or less
AGD 40 mm [mGy] ___ mGy or less
AGD modes 1 to 3
AGD 60 mm [mGy] ___ mGy or less
AGD 70 mm [mGy] ___ mGy or less
TIP
• The criteria can be specified for each of the CNR modes (1 to 3).
• [CNR_H-mode], [CNR_N-mode] and [CNR_L-mode] in the window correspond to the CNR modes 1, 2
and 3, respectively.
• As in CNR modes, AGD modes 1 to 3 correspond to the respective AEC modes. However, a value is
commonly specified as the criteria for AGD modes (1 to 3).
In this Program, the judgment is made based on how much the measured values deviates from the
baseline values.
Since the measured values for some test items depend on the X-ray equipment and/or exposure
conditions, it is necessary to specify the baseline values for the test items.
3
CNR
1 Shot Phantom sensitivity constancy
Geometric distortion
Installation of FDR Mammography QC Program
[Points to be noted]
• Baseline values vary depending on the exposure environment.
• Measurements must be conducted several times under uniform conditions to specify the baseline
values. However, the operational procedure for the QC Software is different in the last measurement.
• Test items are categorized into an exposure menu as shown in the above table.
• Enter baseline values for all of the test items in an exposure menu, including the items not requiring
the baseline setting, to proceed with the operation in the QC software. It is recommended to enter
“0” in order not to mistake the entered value from the measured value when checking the setting
later.
• It is recommended to specify the baseline values in the order of Annual Test → Semi-annual Test →
Daily/Weekly Test, according to their potential influence on images.
• It is recommended to save image data based on which baseline values are determined to later
confirm that daily/weekly, semi-annual and annual tests have been conducted properly.
• When conducting a QC test, turn off the Intelligent AEC.
• If the Phantom size has been changed, be sure to re-create the baseline values.
[Test flow]
5. Auto-calculation
7. QC Software shutdown
[Tools]
1 Shot Phantom
NOTE
If the area is out of the X-ray field while the
3 Phantom is positioned by pressing its corners
against the chest wall-side edge of the
Select [QC/TEST] in the exposure region magnification table, shift the Phantom to the
2
Installation of FDR Mammography QC Program
3
Installation of FDR Mammography QC Program
7. QC Software shutdown
6. Auto-calculation
8. QC Software shutdown
[Tools]
A dosimeter, scale, and lead sheet
[Baseline value setting procedure]
1 Determine the exposure conditions used in the test.
Remove the compression plate, place a lead sheet on the exposure table, and then position
the dosimeter at the lateral center of the exposure table, 60 mm away from the chest wall-
side edge, and also 40 mm above the exposure table.
Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list,
and then press [Start study.].
Specify “Manu” mode for the X-ray equipment.
Make an exposure with 25 kV and W/Rh, and record the mAs when the air kerma is closest
to 30 mR.
• Fill in the following item in the worksheet.
mAs: _____mAs
2 Specify “Manu” mode for the X-ray equipment. Specify the following and then measure the air
kerma 3 times with the mAs recorded in Step 1 .
25 kV and W/Rh
• Fill in the following item in the worksheet.
Entrance air kerma: _____mR µGy , _____µGy , _____µGy
mR mR
10 Press [User input], enter the measured 11 Press [Measurement execution]. The
entrance air kerma and visual inspection results are displayed at the bottom of the
results (Pass/Fail), and then press [OK]. window.
TIP
(4)
(1) If “No baseline value is input.” appears,
press [OK].
(2)
NOTE
(3) In the last time measurement, skip Steps 12
through 14 and go to Step 15 .
3
Installation of FDR Mammography QC Program
8. QC Software shutdown
Geometric distortion
3
System artifact evaluation
Dynamic range
6. Auto-calculation
8. QC Software shutdown
[Tools]
1 Shot Phantom
NOTE
Do not use the compression plate of the
following sizes: 24 × 30 cm (Small), 24 x 30 cm
(SSmall) or 18 x 24 cm (Small)
NOTE
NOTE
It may take some time until you can press [QC]
after completing the exposure process.
3. QC Software startup
TIP
If “No baseline value is input.” appears,
press [OK].
NOTE
Make sure to repeat the same procedure until
the test is finished. Do not add an exposure
submenu when repeating. Select [Study completed] to finish the
18
test.
15 When the result is displayed, press [Save]
and then [Test data log].
3
Installation of FDR Mammography QC Program
5. Image output
9. QC Software shutdown
[Tools]
ACR Phantom, an acrylic disk, a Step Phantom, and a densitometer
NOTE
The Step Phantom is required only when conducting visual inspection of it.
ACR Phantom
Mass
Thickness:
0.5 mm
Specks
Diameter:
0.2 mm
20 mm
Acrylic disk
5. Image output
15 Press [Measurement execution]. The 20 Select the results to be used for calculating
results are displayed at the bottom of the the baseline values among the measured
window. results, and then press [Baseline value
creation]. This completes the baseline
value setting. Press [Close].
NOTE
Make sure that the results for the Test with
TIP
ACR Phantom are selected. They are displayed
If “No baseline value is input.” appears, below those for the Test with 1 Shot Phantom.
press [OK].
NOTE
In the last time measurement, skip Steps 16
3
through 18 and go to Step 19 .
9. QC Software shutdown
18 Repeat Steps 1 through 17 several
21 Press [Save] and then [Close] to exit the
times (5 is recommended) to specify the QC Software.
averages of the multiple measured values
as the baseline values.
NOTE
Make sure to repeat the same procedure until
the test is finished. Do not add an exposure 22 Select [Study completed] to finish the
submenu when repeating. test.
Conduct all of the test items provided in this Program except for Repeat analysis to confirm
that the all judgment items result in [Pass].
This enables to find defects in the X-ray equipment or hardcopy/softcopy output equipment
when installed. In addition, this will help correct a test item judged as “Fail” in a future QC
test by providing the initial data for comparison. For this reason, it is recommended to
save the image data when equipment conditions are checked at the time of the Program
installation.
The specified baseline values or reference criteria are shown on the “List of Reference
Values for Criteria and Baseline Values”.
For visual inspections, it is recommended to clearly relate an output image and the judgment
3
result in the condition check performed when the Program has been installed.
Conducting these Tests complete the condition check at the time of the Program installation.
Every day/week
Daily/Weekly Test
The Daily/Weekly Test is comprised of the constancy test and the performance verification test of the
system.
The constancy test is designed for judging if variations of regularly measured values regarding system
performance are within the allowable range (criteria) based on the values measured at a certain point as
the baseline values. The performance test is intended to check if the values regarding system performance
satisfy the upper or lower limit of judgment criteria specified by the guidelines.
It is necessary to determine the criteria before conducting the Daily/Weekly Test.
NOTE
4
The reference criteria are shown on the “The List of Reference Vales for Criteria and Baseline Values”.
See “3.3 Criteria Confirmation and Determination” for details.
Daily/Weekly Test
When conducting the Daily/Weekly Test the first time after setting the criteria, specify the baseline
values to be used in the future Daily/Weekly Tests. In the second or later Daily/Weekly Test, check that
the variation from the specified baseline values is within the criteria.
See “3.4 Baseline Value Settings” for details.
1 Shot Phantom
Equipment use ACR Phantom
NO Judgment
Check that the variation from the specified baseline values is within the
criteria, or that the values indicating system performance satisfy the criteria.
If the criteria are satisfied, equipment can be used as is until the next
Daily/Weekly Test day. If the criteria are not satisfied, take corrective actions
by following "4.5.3 Solutions for Failed Test Items."
Items Auto-
calculated and Items Judged
Judged Based on [Pass/Fail]
Test Items/Contents Exposure Menus Exposure Submenus Input Data Based on Visual
Inspection by a
1. 2. Input User
Exposure Values
Missed tissue on chest
wall edge (40 mm height)
CNR
1 Shot Phantom
sensitivity constancy
Geometric distortion 3
4
Linearity/Beam quality
constancy
Good practice - 3
Test with ACR Phantom ACR Phantom ACR Phantom 3 3
Daily/Weekly Test
Three judgment methods:
1. Auto-calculation by the QC Software based on the image data acquired by
making an exposure.
2. Auto-calculation by the QC Software based on the input measurement data.
3. User selection of [Pass/Fail] based on visual inspection of the image.
Each Daily Test and Weekly Test item is tested by normally using an output image.
Most of the test results can be auto-calculated from the image data by the QC Software, though some
items require visual inspection.
NOTE
The Test with ACR Phantom is required to be conducted at a frequency defined by local regulations where
the equipment is to be used. If conducting the Test with ACR Phantom every day in addition to the Test with
1 Shot Phantom, select Exposure Menu [1 Shot/ACR].
In other cases, do not use the exposure menu. Otherwise, an incorrect test data log will be created.
Tools
[ A ] QC Software
[ B ] Worksheet (See Chapter 8)
[ C ] 1 Shot Phantom
[ D ] Viewing box (if film is used)
[ F ] ACR Phantom (including an acrylic disk)
[ G ] Step Phantom *1
[ M ] Densitometer (if film is used)
*1 Required only when conducting visual inspection of the Phantom.
4
Daily/Weekly Test
Daily/Weekly Test
Pixel Value (PV) ratio (Top center) [%] Baseline value ± ___%
Uniformity
SNR ratio (Top-Right) [%] Baseline value ± ___%
1 Shot PhantomM SNR ratio (Top-Left) [%] Baseline value ± ___%
SNR ratio (Bottom-Right) [%] Baseline value ± ___%
SNR ratio (Bottom-Left) [%] Baseline value ± ___%
SNR ratio (Top center) [%] Baseline value ± ___%
Dynamic range Average QL at thinnest step wedge [QL] Baseline value ± ___ QL
Low Contrast Detectability Low Contrast Detectability (Light) [%] Baseline value ± ___ %
(LCD) Low Contrast Detectability (Dark) [%] Baseline value ± ___ %
QL gap (1 step-2 step) [QL] Baseline value ± ___ QL
TIP
The criteria setting is not required for visual inspection.
*1 When not checking these items, leave the preset values as they are. Do not delete the preset values.
4.5.1 Procedure
[Test flow]
4. QC Software startup
4 6. Image output
9. Auto-calculation
TIP
When also conducting the Test with ACR Phantom every day, use Exposure Submenu [1 Shot/ACR] to
perform the “exposure” and “user input” steps of the Test with ACR Phantom immediately after performing
the respective steps of the Test with 1 Shot Phantom.
[Tools]
1 Shot Phantom
NOTE
Position the Phantom at the lateral center
of the exposure table by pressing the
corners against the chest wall-side edge of
2 Select [QC/TEST] in the exposure region the exposure table. If there are obstacles at
list and [1 Shot PhantomM] in the exposure the time of positioning, the test may not be
menu list, and then press [Start study.]. conducted accurately.
1 Shot Phantom 4
Daily/Weekly Test
3 The following exposure submenu is
displayed.
Exposure table
• 1 Shot PhantomM
Corner of the
1 Shot Phantom
8 Select Exposure Submenu [1 Shot 11 Enter the results of Good practice, and
PhantomM] from the exposure condition then press [OK].
display field and press [Film output]. For
performing image diagnosis on a softcopy
display, press [Network output] to output
the image to the network.
NOTE
Make sure that the result is selected for all
items. If there is an item left unentered, an error
Daily/Weekly Test
TIP
12 Press [Measurement execution]. The
For performing image diagnosis on a softcopy results are displayed at the bottom of the
display, check the items on the display.
window.
Daily/Weekly Test
If any of the above is not correct/appropriate, correct the problem.
Fail
4.6.1 Procedure
[Test flow]
3. QC Software startup
5. Image output
7. Auto-calculation
Daily/Weekly Test
9. QC Software shutdown
[Tools]
ACR Phantom, an acrylic disk, a Step
Phantom, and a densitometer
Daily/Weekly Test
the ACR Phantom (as shown in the figure
below).
Nipple side
ACR Phantom
Mass
Thickness:
0.5 mm
displayed.
0.2 mm
• ACR Phantom
20 mm
Acrylic disk
3. QC Software startup
Daily/Weekly Test
NOTE
If a viewing box was set not to be used in “3.2
Initial Settings”, the density input fields will be
grayed out.
7. Auto-calculation
4
Daily/Weekly Test
Pass
Make sure that the ACR The test is finished.
Phantom is correctly positioned
and then redo the test.
4
Fail
Daily/Weekly Test
Contact a FUJIFILM dealer.
Quarterly Test
Repeated images shall be evaluated quarterly. In order for the repeat rates to be meaningful, a patient
volume of at least 250 patients or 1,000 exposures is needed.
The Repeat analysis is the required test item to be conducted in some areas. Conduct the analysis in
compliance with local regulations where the equipment is to be used.
As described above, the Quarterly Test is neither a constancy test nor a performance test of the system.
Specify the criteria when conducting the test, not in advance.
The Retake Analysis software, an optional software for the FDR-3000AWS, is convenient for organizing
and managing the repeat analysis data.
Equipment use
Rejected image
5 Quarterly Test
analysis
Repeat analysis
Quarterly Test
NO
The Quarterly Test requires no settings since the QC Software is not used in the Test.
The Test also requires no tool.
Simply record the result of the Repeat analysis in the report form.
Quarterly Test
5.3.1 Procedure
[Test flow]
1. Repeat analysis (collecting rejected images)
Quarterly Test
for potential improvement.
5
Quarterly Test
Semi-annual Test
The Semi-annual Test is comprised of the constancy test and the performance verification test of the
system.
The constancy test is designed for judging if variations of regularly measured values regarding system
performance are within the allowable range (criteria) based on the values measured at a certain point
as the baseline values. The performance test is intended to check if the values regarding system
performance satisfy the upper or lower limit of judgment criteria specified by the guidelines.
It is necessary to determine the criteria before conducting the Semi-annual Test.
NOTE
The reference criteria are shown on the “List of Reference Values for Criteria and Baseline Values”.
See “3.3 Criteria Confirmation and Determination” for details.
When conducting the Semi-annual Test the first time after setting the criteria, specify the baseline values
to be used in the future Semi-annual Tests. In the second or later Semi-annual Test, check that the
variation from the specified baseline values is within the criteria.
See “3.4 Baseline Value Settings” for details.
6 value settings
measurement
Image basic test
Equipment use
Measurement
Semi-annual Test
Tools
[ A ] QC Software
[ B ] Worksheet (See Chapter 8)
[ D ] Viewing box (if film is used)
[ E ] Timer or watch/clock
[ H ] Dosimeter
[ I ] Luminance meter
[ K ] Force scale
[ L ] Illuminance Meter
[ N ] PMMA phantom
[ Q ] Scale
[ T ] Lead sheet
6
Semi-annual Test
TIP
The criteria setting is not required for visual inspection.
Semi-annual Test
6.5.1 Procedure
[Test flow]
1. Image basic test (measurement)
6. QC Software startup
8. Image output
12. Auto-calculation
Semi-annual Test
Exposure table
60 mm
Exposure table
60 mm 40 mm
Semi-annual Test
displayed.
Specify the kVp, mAs and target/filter
• Uniform Expo(Image Basic)
determined in “3.4.2 Baseline Values for
Semi-annual Test”, and measure the air
kerma 3 times in “Manu” mode.
• Fill in the following items in the worksheet.
kVp: _____kV
mAs: _____mAs
Target/Filter: _____/_____
Air kerma: _____mR µGy , _____µGy , _____µGy
mR mR
NOTE
Make sure to specify the same exposure
conditions as those used in the baseline
value setting.
If they are not certain, newly determine the
exposure conditions for the Semi-annual
Test by following “3.4.2 Baseline Values for
Semi-annual Test”.
4. Remove the dosimeter.
Select [Study completed] to finish the test.
6
measured value: _____mm _____N
Compression force measured value:
_____N
Semi-annual Test
Semi-annual Test
correctly.
• Fill in the following item in the worksheet.
Masking: Pass/Fail
6. Measure the illuminance near the viewing
box, as a sample illuminance in the
viewing condition, by using an illuminance
meter.
• Fill in the following item in the worksheet.
Interpretation room illuminance (Ambient
light): _____lx
NOTE
See the operation manuals for the viewing
box, luminance meter or illuminance meter
for how to measure the luminance or
illuminance.
NOTE
It may take some time until you can
press [QC] after completing the exposure
process.
White-out
NOTE
Make sure that the values are entered for all
items. If there is an item left unentered, an error
occurs.
12. Auto-calculation
Semi-annual Test
Pass
Redo the test. The test is finished.
Fail
6
n [Relative sensitivity] resulted in [Fail]
Semi-annual Test
Points to be checked
• The correct exposure submenu was selected.
• The X-ray equipment settings (target/filter, kVp) were correct.
• The air kerma used for the exposures was appropriate.
• The dosimeter was correctly positioned.
• The dosimeter was correctly used.
If any of the above is not correct/appropriate, correct the problem.
Pass
Redo the test. The test is finished.
Fail
The X-ray equipment or
dosimeter may be defective.
Pass
Redo the test. The test is finished.
Fail
Semi-annual Test
The X-ray equipment may be defective.
Pass
Redo the test. The test is finished.
Fail
Pass
Redo the test. The test is finished.
Fail
6
Semi-annual Test
Annual Test
The Annual Test is comprised of the constancy test and the performance verification test of the system.
The constancy test is designed for judging if variations of regularly measured values regarding system
performance are within the allowable range (criteria) based on the values measured at a certain point
as the baseline values. The performance test is intended to check if the values regarding system
performance satisfy the upper or lower limit of judgment criteria specified by the guidelines.
It is necessary to determine the criteria before conducting the Annual Test.
NOTE
The reference criteria are shown on the “List of Reference Values for Criteria and Baseline Values”.
See “3.3 Criteria Confirmation and Determination” for details.
When conducting the Annual Test the first time after setting the criteria, specify the baseline values to
be used in the future Annual Tests. In the second or later Annual Test, check that the variation from the
specified baseline values is within the criteria.
See “3.4 Baseline Value Settings” for details.
NO
Items Auto-
calculated and Items Judged
Judged Based on [Pass/Fail]
Test Types/Items Exposure Menus Exposure Submenus Input Data Based on Visual
Inspection by a
1. 2. Input User
Exposure Values
ACR Phantom1(Lag) 3
Additive lag effects (Lag)
Non Exposure(Lag) 3
Uniform Expo.1(Ghost) 3
ultiplicative lag effects
M
Image performance Annual 1/6 ACR Phantom2(Ghost) 3
(Ghost)
Uniform Expo.2(Ghost) 3
issed tissue on chest
M
1 Shot PhantomM(40mm) 3 3
wall edge (40 mm height)
Visual and Functional test - 3
Spatial Resolution
Spatial Resolution Annual 2/6 1 ShotM(Magnification) 3
(Magnification)
k Vp accuracy and
- 3
reproducibility
X-ray equipment Half Value Layer (HVL) - 3
Annual 3/6
performance
Collimation assessment Uniform Expo.3(Collimation) 3 3
Radiation output - 3
AEC reproducibility - 3
Mode1 through Mode3
CNR(PMMA20mm) 3
AGD modes 1 to 3 - 3
NOTE
The Missed tissue on chest wall edge (40 mm height) is the required test item to be conducted in some areas.
Conduct the analysis in compliance with local regulations where the equipment is to be used.
Tools
[ A ] QC Software
[ B ] Worksheet (See Chapter 8)
[ C ] 1 Shot Phantom
[ D ] Viewing box (if film is used)
[ F ] ACR Phantom
[ H ] Dosimeter
[ J ] kVp meter
[ N ] PMMA phantoms (available, in combination, for 20, 40, 60 and 70 mm in
thickness)
[ O ] Aluminum plate (0.2 mm) for CNR measurement
[ P ] Aluminum plates for half value layer measurement (0.5 and 0.6 mm)
[ Q ] Scale
[ R ] Coins
[ S ] Jig for 1 Shot Phantom*1
[ T ] Lead sheet
*1
The jig for 1 Shot Phantom is required only in the Missed tissue on chest wall edge
(40 mm height) measurement.
7
Annual Test
kVp accuracy and reproducibility Low volt. accuracy (SFS) [kV] Specified kVp ± ___ kV
X-ray / Light field gap (Right/Left) [mm] Total gap criteria = SID X ___ or less
X-ray / Light field gap (Chest/Nipple) [mm] Total gap criteria = SID X ___ or less
X-ray / Image receptor field gap (Left) [mm] SID X ___ or less
Collimation assessment
X-ray / Image receptor field gap (Right) [mm] SID X ___ or less
X-ray / Image receptor field gap (Nipple) [mm] SID X ___ or less
7
Reproducibility Variation coefficient ___ or less
TIP
• The criteria can be specified for each of the CNR modes (1 to 3).
The criteria setting is not required for visual inspection.
• [CNR_H-mode], [CNR_N-mode] and [CNR_L-mode] in the window correspond to the CNR modes 1, 2
and 3, respectively.
• As in CNR modes, AGD modes 1 to 3 correspond to the respective AEC modes. However, a value is
commonly specified as the criteria for AGD modes (1 to 3).
7
Annual Test
6. QC Software startup
11. Auto-calculation 7
12. Test result confirmation Annual Test
NOTE
The Missed tissue on chest wall edge (40 mm height) is a required test item to be conducted in some areas.
Conduct the analysis in compliance with local regulations where the equipment is to be used.
If not conducting the test item, the above Step “8. Missed tissue on chest wall edge (40 mm height)
(measurement)” can be skipped, but not Steps “4. Missed tissue on chest wall edge (40 mm height) (exposure)”
and “9. Missed tissue on chest wall edge (40 mm height) (user input)”. Perform operations in these Steps by
following the procedure without conducting the test item.
NOTE
Make exposures in Substeps 3 and 4
consecutively without much interval between
them to obtain high-quality results.
Exposure table
Nipple side
See “Tool Details” in “2.3 Tools” for details 7
on the jig for 1 Shot Phantom.
Annual Test
Calculation Areas TIP
(20 mm x 20 mm) ACR Phantom
Use the same exposure conditions as those
used for the Phantom exposure in the Daily/
1 2 Weekly Test. See “3.4.3 Baseline Values for
20 mm Daily/Weekly Test” for details.
60 mm
NOTE
Beware that the jig does not overlap with
the measurement points of the Missed
tissue on chest wall edge.
3. Specify the exposure conditions closest
to those used for the 1 Shot Phantom
exposure in the Daily/Weekly Test in
7 “Manu” mode and make an exposure.
• Fill in the following items in the worksheet.
kVp: ______kV 6. QC Software startup
Annual Test
NOTE
It may take some time until you can press [QC]
after completing the exposure process.
10 mm
9 mm
8 mm
7 mm
7
6 mm
5 mm
4 mm
3 mm
2 mm Annual Test
1 mm
10. Visual and Functional test (user input) 17 Press [Save] and then [Close] to exit the
QC Software.
14 Select the results ([Pass/Fail]) and press
[OK].
NOTE
Make sure that the result is selected for all
items. If there is an item left unentered, an error
occurs.
Pass
Redo the test. The test is finished.
Fail
The Missed tissue on chest wall The Missed tissue on chest wall
edge (40 mm height) in the edge (40 mm height) in the Daily/Weekly
Daily/Weekly Test has been Test has been judged as “Pass” in
occasionally judged as “Fail”. every test.
Pass
Change the 1 Shot Phantom The test is finished.
position or its jig, and then redo
the test.
Fail
7
Annual Test
1. Procedure
[Test flow]
The spatial resolution test has only a single test item.
3. QC Software startup
5. Auto-calculation
7. QC Software shutdown
7
Annual Test
1 Enter patient information and press [Next]. 4 Follow the instructions below to conduct
the test item.
Make an exposure using the following exposure
submenu:
[1 ShotM(Magnification)]
[Tools]
1 Shot Phantom
NOTE
If the area in the broken line is out of the
X-ray field when the Phantom is positioned
by pressing its corners against the chest
wall-side edge of the magnification table,
shift the Phantom to the front. Make sure
that the shifted Phantom is parallel to the
chest wall-side edge.
NOTE
The Phantom position must be the same
for every exposure (the allowable positional 5. Auto-calculation
error is within ± 5 mm).
7 Press [Measurement execution]. The
• Fill in the following item in the worksheet.
results are displayed at the bottom of the
1 Shot Phantom position
window.
Lateral shift distance: _____mm
3. Specify “Manu” mode for the X-ray
equipment and make an exposure under
the same exposure conditions as those
used in “3.4.1 Baseline Values for Annual
Test”. 6. Test result confirmation
• Fill in the following items in the worksheet. Check the displayed judgment results.
kVp: ______kV 8
If there is a [Fail], see “7.5.2 3. Solutions
mAs: ______mAs
for Failed Test Items”.
Target/Filter: ______/______
Compressed breast thickness: _____mm
Compression force: _____N
7
NOTE Annual Test
It may take some time until you can press [QC] 7. QC Software shutdown
after completing the exposure process. Press [Save] and then [Close] to exit the
9
QC Software.
3. QC Software startup
The Spatial Resolution (SR) in the The Spatial Resolution (SR) in the
Daily/Weekly Test has been Daily/Weekly Test has been
occasionally judged as “Fail”. judged as “Pass” in every test.
Fail
7
Contact a FUJIFILM dealer.
Annual Test
1. Procedure
[Test flow]
The X-ray equipment performance test consists of the kVp accuracy and reproducibility, Half Value
Layer (HVL), Collimation assessment and Radiation output.
TIP
In this test, conduct the three test items not requiring an X-ray image first and then enter their measured
values all at once to save time and simplify QC Software operation.
3. Radiation output
6. QC Software startup
13. Auto-calculation
kVp meter
Magnification table
50 mm
NOTE
Wear gloves when using the aluminum
1. Remove the compression plate.
plate to keep it clean.
2. Place a lead sheet on the exposure table,
5. Position the 0.5 mm aluminum plate and
and position the dosimeter at the lateral
make an exposure under the condition
center of the table, 60 mm away from the
calculated in Substep 3 using each of
chest wall-side edge, and also 40 mm
the kVp and target/filter to measure the
above the table.
air kerma after the X-ray transmitted
through the plate.
• Fill in the following item in the worksheet.
Entrance air kerma – Al 0.5 mm: ____
7
use “Manu” mode for exposure.
4. Make an exposure and measure the air
kerma and mAs. The following setting is
Annual Test
recommended.
28 kV, 80 mAs and W/Rh 5 Select [QC/TEST] in the exposure region
In this Substep, measure air kerma list and [Annual 3/6] in the exposure menu
5 times under the same exposure list, and then press [Start study.].
conditions.
• Fill in the following items in the worksheet.
kVp: _____kV
mAs: _____mAs
Target/Filter: _____/_____
Entrance air kerma (1): _____mR µGy
7
Annual Test
• Fill in the following items in the worksheet. 2. Position 2 cassettes (Cassettes B1 and
Compression plate edge (Chest wall B2, hereafter) over the exposure table by
side as reference): Pass/Fail aligning their chest wall-side edges.
Visual inspection result: Pass (no gap)/
7 NOTE
Fail (with gap)
Cassette B1 Cassette B2
Go to Step 17 .
Annual Test
7
boundary of the cassettes.
5. Move the compression plate down onto
the Cassettes B1 and B2.
Annual Test
NOTE
Take care that the compression plate is not
scratched by the coins.
TIP
Check the following coin images.
Chest wall: Coins (a) and (e) Annual Test
IP
Assign the value recorded in Substeps
1 and 2 for each coin to A and B in the
Eg: 3 X=A-B corresponding formula.
A B
Eg: 4 X=B-A
A B
Y=(A-e)+{(B-e)-
Eg: 3 B-e
(A-a)} Coin (f)
A-e
A-a Coin (f)
Coin (a) Coin (e)
Coin (e)
• Fill in the following items in the worksheet.
TIP Compression plate edge (Chest wall
side as reference): Pass/Fail
When the X-ray field is inside of the image
receptor edge in the image read from the
Visual inspection result: Pass (no gap)/
FDR mammography system, as shown in Fail (with gap)
Eg: 1, the size of the gap can be determined
by measuring the distance (“y” in the figure) 9. kVp accuracy and reproducibility (user
from the image receptor edge to the X-ray input)
field.
17 Press [User input].
TIP
When the X-ray field is outside of the image
receptor edge in the image read from the
FDR mammography system as shown in
Egs: 2 and 3, the size of the gap can be Select [Annual test 3-1].
18
7
calculated. Measure the distance ”A-a” Enter the display values and measured
between the image receptor edge and values.
that of Coin (a) (on the chest wall-side
edge) and the distance “A-e” between
Annual Test
19 Select [Annual Test 3-2]. 21 Enter the measured entrance air kerma,
Enter the measured entrance air kerma mAs and air kerma rate, and then press
for each target/filter. [OK].
NOTE NOTE
Some dosimeters may require correction of the Some dosimeters may require correction of the
measured value according to the beam quality. measured value according to the beam quality.
TIP NOTE
Select the thickness of an aluminum plate Make sure that the values are entered for all
that was used if the thickness is other than items. If there is an item left unentered, an error
0.5 and 0.6 mm. occurs.
13. Auto-calculation
11. Collimation assessment (user input)
22 Press [Measurement execution]. The
20 Select [Annual test 3-3]. Enter the
results are displayed at the bottom of the
measured values and visual inspection
window.
result (Pass/Fail).
NOTE
Enter all items even if the test is conducted
without using cassettes. Although any value
can be entered (e.g. “0” for all items), a negative
7
value or significantly large one causes an error 14. Test result confirmation
Annual Test
to occur.
23 Check the displayed judgment results.
If there is a [Fail], see “7.5.3 3. Solutions
for Failed Test Items”.
7
Annual Test
Pass
Redo the test. The test is finished.
Fail
Pass
Redo the test. The test is finished.
Fail
Pass
Redo the test. The test is finished.
Fail
Pass
Redo the test. The test is finished.
Fail
7
Annual Test
3. QC Software startup
5. Auto-calculation
7. QC Software shutdown
9. AEC reproducibility
17. Auto-calculation
3. QC Software startup
5. Auto-calculation
7. QC Software shutdown
12. Auto-calculation
TIP
Select CNR mode 3 in Step 2 and 9 above when conducting the test in the mode. 7
Annual Test
7
Annual Test
[Tools]
PMMA phantom (40 mm) and aluminum
plate (0.2 mm) for CNR measurement
Calculation Areas
(20 mm x 20 mm) Aluminum plate
1 2
60 mm 20 mm
NOTE
Wear gloves when using the aluminum plate to
keep it clean.
NOTE
The “baseline values” above means the
reference CNR for recognizing the change in
image quality depending on the target thickness
(target thickness dependence).
7. QC Software shutdown
Press [Save] and then [Close] to exit the 9. AEC reproducibility
15
QC Software.
18 Follow the instructions below to conduct
the test item.
[Tools]
A dosimeter and PMMA phantom (40 mm)
16 Select [Study completed] to finish the test. 1. Position the PMMA phantom (40 mm) on
7 the exposure table.
NOTE
Annual Test
Lateral center
60 mm
Exposure table
60 mm 40 mm
NOTE
Wear gloves when using the aluminum
plate to keep it clean.
12. CNR mode 1 (exposure)
NOTE
Make sure that the aluminum plate covers Select [QC/TEST] in the exposure region
the detector of the dosimeter from the
21
list and [Annual 4/6] in the exposure menu
above. list, and then press [Start study.].
10. Make an exposure with the exposure
conditions used in Substep 7 and
measure the air kerma.
• Fill in the following item in the worksheet.
Entrance air kerma – Al 0.3 mm:
_____mR
µGy
[Tools]
PMMA phantoms (20, 40, 60, 70 mm)
and aluminum plate (0.2 mm) for CNR
measurement
Aluminum plate
TIP
In CNR mode 1, use the same H-mode of
the AEC as that used in AGD mode 1. 4. Specify the exposure conditions for
If multiple exposure modes are used in PMMA phantom 20 mm determined
clinical practice, conduct the test in all in Substep 3 of AGD mode 1 test
of the modes used. Beware that the test procedure. Then make an exposure in
procedure for CNR mode 2 or later differs “Manu” mode.
from that for CNR mode 1.
• Fill in the following items in the worksheet.
kVp: _____kV
NOTE mAs: _____mAs
Wear gloves when using the aluminum plate to Target/Filter: _____/_____
keep it clean. Compressed breast thickness: _____mm
Compression force: _____N
1. Position the PMMA phantom (20 mm) on
the exposure table. NOTE
The “Semi” and “Auto” modes may not
2. Position the 0.2 mm aluminum plate
work correctly for this measurement. Make
for CNR measurement as shown in the sure to use “Manu” mode for exposure.
figure below.
5. Using PMMA phantoms 40 mm, 60
Nipple side
mm and 70 mm, repeat the operation
PMMA described in Substeps 1 to 4. Use the
7 Calculation Areas
Aluminum plate
exposure conditions for the PMMA
phantom of each thickness determined
(20 mm x 20 mm)
in Substep 4 of the AGD mode 1 test
Annual Test
procedure.
1 2
60 mm 20 mm
NOTE
Make sure that the aluminum plate covers
the Calculation Area 2 without protruding
into the left-hand half of the phantom. NOTE
It may take some time until you can press
[QC] after completing the exposure process.
7
Annual Test
30 Select [Annual test 4-2]. Enter the 32 Check the displayed judgment results.
measured entrance air kerma and target/ If there is a [Fail], see “7.5.4 3. Solutions
filter for each, and then press [OK]. for Failed Test Items”.
NOTE
Some dosimeters may require correction of the
measured value according to the beam quality.
TIP
Select the thickness of an aluminum plate
that was used if the thickness is other than
0.5 and 0.6 mm.
Pass
Redo the test. The test is finished.
Fail
7
• The unit of air kerma (mR or µGy) was correctly used.
If any of the above is not correct/appropriate, correct the problem.
Annual Test
Pass
Redo the test. The test is finished.
Fail
Fail
7
Annual Test
Pass
The X-ray equipment may be defective.
Pass
Redo the test. The test is finished.
7
Annual Test
8 1
20 mm
2
60 mm
Assistance Material
NOTE
Wear gloves when using the aluminum plate to
keep the plate clean.
Aluminum plate
Specify the exposure conditions for the 11 Press [Save] and then [Close] to exit the
6 QC Software.
X-ray equipment and make an exposure
in “Manu” mode.
NOTE
The “Semi” and “Auto” modes may not work
correctly for this measurement. Make sure to
use “Manu” mode for exposure. 12 Select [Study completed] to finish the
test.
7 Repeat Steps 4 to 6 as many times as
necessary to capture the desired number
of images to be calculated (max. 32
images).
NOTE
It may take some time until you can press [QC]
after completing the exposure process.
8
3 Specify the desired exposure conditions
Assistance Material
NOTE
It may take some time until you can press [QC]
after completing the exposure process.
19 Select [Study completed] to finish the
13 Press [QC] to display the “QC window”. test.
n Basic Procedure
1 Starting FUJIFILM workstation
Start a FUJIFILM workstation and enter
patient information, and then press [Next].
2 Starting QC test
Select [QC/TEST] in the exposure region (5) Quits the
QC Software.
list and an exposure submenu displayed (1) Outputs an image (2) Displays the user input
in the exposure menu list, and then press for visual
inspection.
screen for the measurement
and visual inspection results.
[Start study.].
6 Finishing the test
Press [Study completed] to finish the test.
When finishing the test, the window in Step 1
is redisplayed. To conduct another test, repeat
the procedure.
3 Exposure
Make an exposure using all of the
8
displayed exposure submenus.
4 Starting QC Software
Press [QC].
Assistance Material
Switches information
displayed in the test
item field between
8
measured values and
judgment results.
Displays exposure
conditions.
Creates baseline
values.
8
Assistance Material
2 Air kerma mR
µGy
mR
µGy
mR
µGy Result input ( 10 )
[1] Image Basic Test
kVp: kV
Exposure conditions for Exposure condition
7 Exposure Submenu [Uniform mAs:mAs
confirmation ( 9 )
Expo(Image Basic)]
Target/Filter: /
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
[1]
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
[2] [1]
8
Assistance Material
[1] Test
with 1 Shot Distortion Pass Fail
Phantom 9 Result input ( 10 )
Artifact Pass Fail
kVp: kV
mAs:mAs
Exposure conditions for
Exposure condition
4 -5 Exposure Submenu Target/Filter: /
confirmation ( 7 )
[ACR Phantom]
Compressed breast thickness: mm
Compression force: N
Density at center of Phantom image
Density inside the disk
[3] Test
with ACR
Phantom Density outside the disk
Fibers (ACR Phantom) points
Specks (ACR Phantom) points
8 Result input ( 10 )
Masses (ACR Phantom) points
Visible step wedge (Step
steps
Phantom)
Specks (Step Phantom) steps
Masses (Step Phantom) steps
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
[3] [1]
[2]
Air kerma mR
µGy
mR
µGy
mR
µGy Result input ( 13 )
[1] Image Basic Test kVp: kV
Exposure conditions for Exposure condition
5 Exposure Submenu [Uniform mAs:mAs
confirmation ( 9 )
Expo(Image Basic)]
Target/Filter: /
7 -2 Luminance cd/m2
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
[1]
[3]
[2]
8
Assistance Material
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
[3]
[4]
kVp: kV
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
8
Assistance Material
Air kerma
kVp mAs Al 0.3 mm Al 0.5 mm
w/o Al
[7] Half
Value Layer 2 ( ) ( )
(HVL) Target/Filter ( / ) kV mAs mR mR mR Result input ( 19 )
-3, 4, 5 µGy µGy µGy
kVp: kV
Exposure conditions mAs:mAs
Target/Filter: /
mR mR mR
[9] Radiation output 3 -4 Entrance air kerma
(1) µGy (2) µGy (3) µGy Result input ( 21 )
mR mR
(4) µGy (5) µGy
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
8
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
EC reproducibility
[11] A Compressed breast thickness: mm
Compression force: N
mR mR mR
(1) µGy (2) µGy (3) µGy
17 -6 Entrance air kerma
mR mR Result input ( 26 )
(4) µGy (5) µGy
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
[10]
8
Assistance Material
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
[12] [12]
20 mm kV mAs / mm N
Exposure Mode 1 40 mm /
kV mAs mm N
60 mm kV mAs / mm N
70 mm kV mAs / mm N
Compressed
PMMA Target/ Compression
kVp mAs breast
thickness Filter thickness
force
20 mm kV mAs / mm N
[11] CNR modes Exposure condition
1 to 3 22 -4 Exposure Mode 2 40 mm kV mAs / mm N confirmation ( 29 )
60 mm kV mAs / mm N
70 mm kV mAs / mm N
Compressed
PMMA Target/ Compression
kVp mAs breast
thickness Filter thickness
force
20 mm kV mAs / mm N
Exposure Mode 3 40 mm /
kV mAs mm N
60 mm kV mAs / mm N
70 mm kV mAs / mm N
White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8.
Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired
automatically.)
8
Assistance Material
Facility Information
Client Date
Exposure Room Time
Operator Department Name
Calibration Expiration
Manufacturer Model S/N
Date
1 Shot Phantom
FUJIFILM
M
ACR Phantom
Step Phantom
Densitometer
Density control
strip
Dosimeter
Force scale
Luminance
meter
Illuminance
meter
kVp meter
Signature
1/3
Exposure Conditions
Compressed
Compression
kVp mAs Target/Filter Breast S Value
Force
Thickness
1 Shot
kV mm N
PhantomM
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Missed tissue on
chest wall edge mm - PASS FAIL
Missed tissue on chest wall (Right)
edge Missed tissue on
chest wall edge mm - PASS FAIL
(Left)
CNR CNR PASS FAIL
1 Shot Phantom sensitivity
System sensitivity PASS FAIL
constancy
Horizontal direction mm PASS FAIL
Geometric distortion Vertical direction mm PASS FAIL
Geometric distortion - - PASS FAIL
System artifact evaluation Artifacts - - PASS FAIL
Pixel Value (PV)
% PASS FAIL
ratio (Top-Right)
Pixel Value (PV)
% PASS FAIL
ratio (Top-Left)
Pixel Value (PV)
% PASS FAIL
ratio (Bottom-Right)
Pixel Value (PV)
% PASS FAIL
ratio (Bottom-Left)
Pixel Value (PV)
% PASS FAIL
ratio (Top center)
Uniformity
SNR ratio
% PASS FAIL
(Top-Right)
SNR ratio
% PASS FAIL
(Top-Left)
SNR ratio
% PASS FAIL
(Bottom-Right)
SNR ratio
% PASS FAIL
(Bottom-Left)
SNR ratio
% PASS FAIL
(Top center)
Average QL at
Dynamic range thinnest step QL PASS FAIL
wedge
2 lp/mm % PASS FAIL
Spatial Resolution (SR)
4 lp/mm % PASS FAIL
2/3
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Low Contrast
% PASS FAIL
Low Contrast Detectability Detectability (Light)
(LCD) Low Contrast
% PASS FAIL
Detectability (Dark)
QL gap (1 step-2
QL PASS FAIL
step)
QL gap (2 step-3
QL PASS FAIL
Linearity/Beam quality step)
constancy QL gap (3 step-4
QL PASS FAIL
step)
QL gap (4 step-5
QL PASS FAIL
step)
Remarks
Signature
3/3
Good practice
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
X-ray equipment cleanliness - - PASS FAIL
Screen cleanliness - -
Viewing box cleanliness - -
Remarks
Signature
1/3
Exposure Conditions
Compressed
Compression
kVp mAs Target/Filter Breast S Value
Force
Thickness
ACR Phantom kV mm N
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Density at center of Phantom
PASS FAIL
image
Density difference
PASS FAIL
(disk/outside)
Fibers (ACR Phantom) - PASS FAIL
Specks (ACR Phantom) - PASS FAIL
Masses (ACR Phantom) - PASS FAIL
Visible step wedge (Step
- PASS FAIL
Phantom)
Specks (Step Phantom) - PASS FAIL
Masses (Step Phantom) - PASS FAIL
2/3
Density Difference
+0.05 +0.05
Baseline
-0.05 -0.05
Background Density
+0.20 +0.20
+0.15 +0.15
Baseline
-0.15 -0.15
-0.20 -0.20
No.Visible
Fibers
+0.5 +0.5
- 0.5 - 0.5
Specks
+0.5 +0.5
- 0.5 - 0.5
Masses
+0.5 +0.5
- 0.5 - 0.5
mAs (optional)
+15% +15%
Baseline
-15% -15%
3/3
No.Visible
Fibers
+0.5 +0.5
- 0.5 - 0.5
Specks
+0.5 +0.5
- 0.5 - 0.5
Masses
+0.5 +0.5
- 0.5 - 0.5
S Value
+20% +20%
Baseline
-20% -20%
mAs (optional)
+15% +15%
Baseline
-15% -15%
Remarks
Signature
1/1
Repeat analysis
PROJECTION
REASON FOR REJECT TOTAL % of
CC MLO ML or LM AXILLARY OTHER
REPEATS REPEATS
POSITIONING
PATIENT MOTION
COMPRESSION
ARTIFACTS
X-RAY EQUIP
MALFUNCTION
SOFTWARE
MALFUNCTION
AEC
MISPLACEMENT
UNDER
EXPOSURE
OVER
EXPOSURE
INCORRECT
PATIENT ID
WASTE
SUB-TOTAL
GRAND TOTAL
REPEATS
REPEAT RATE = REPEAT RATE = %
TOTAL FILMS INCLUDING REPEATS
Signature
1/3
Exposure Conditions
Compressed Compression
kVp mAs Target/Filter S Value
Breast Thickness Force
S Value kV mm N
Visual
kV mm N
Inspection
Measured Values
Entrance air
mR (μC/kg)
kerma 1
Entrance air
mR (μC/kg)
kerma 2
Entrance air
mR (μC/kg)
kerma 3
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Relative sensitivity (S value) PASS FAIL
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Artifact - - PASS FAIL
White out - - PASS FAIL
Remarks
Signature
2/3
Measured Values
Compressed breast thickness
mm
- Display value
Compressed breast thickness
mm
- Measured value
Compression force - Display
N
value
Compression force -
N
Measured value
Maximum compression force N
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Compressed breast thickness
mm PASS FAIL
accuracy
Compression force accuracy N PASS FAIL
Maximum compression force N PASS FAIL
Compression force sustainability - - PASS FAIL
Remarks
Signature
3/3
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Luminance cd/m2 - PASS FAIL
Viewing box uniformity - - PASS FAIL
Luminance adjustment function - - PASS FAIL
Masking - - PASS FAIL
Interpretation room
lx - PASS FAIL
illuminance (Ambient light)
Remarks
Signature
1/13
Exposure Conditions
Compressed Compression
kVp mAs Target/Filter S Value
Breast Thickness Force
ACR Phantom1(Lag) kV mm N
Non
kV mm N
Exposure(Lag)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Lag Factor QL - PASS FAIL
Remarks
Signature
2/13
Exposure Conditions
Compressed Compression
kVp mAs Target/Filter S Value
Breast Thickness Force
Uniform
kV mm N
Expo.1(Ghost)
ACR Phantom2(Ghost) kV mm N
Uniform
kV mm N
Expo.2(Ghost)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Ghost Factor - PASS FAIL
Remarks
Signature
3/13
Exposure Conditions
Compressed Compression
kVp mAs Target/Filter S Value
Breast Thickness Force
1 Shot Phantom
kV mm N
M(40 mm)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Missed tissue on chest wall
- - PASS FAIL
edge (Right)
Missed tissue on chest wall
- - PASS FAIL
edge (Left)
Remarks
Signature
4/13
Test Result
Judgment Item Judgment Result
Verify mechanical stability PASS FAIL
Verify that all moving parts move smoothly PASS FAIL
All locks and detents work properly PASS FAIL
No sharp, rough edges or other hazards including electrical hazards PASS FAIL
Operator technique control charts are posted PASS FAIL
Operator protected during exposure by adequate radiation shielding PASS FAIL
All indicator lights working properly PASS FAIL
Auto decompression can be overridden and status displayed PASS FAIL
Compression can be manually released / in any failure PASS FAIL
Remarks
Signature
5/13
Exposure Conditions
Compressed Compression
kVp mAs Target/Filter S Value
Breast Thickness Force
1 ShotM
kV mm N
(Magnification)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
MTF (4lp/mm) % PASS FAIL
MTF (8lp/mm) % PASS FAIL
Remarks
Signature
6/13
Measurement Conditions
Low kV
kVp Display Values Mid. kV
High kV
Measured Values
Low kV
Mid.-1 kV
Mid.-2 kV
Large Focal Spot
Mid.-3 kV
(LFS)
Mid.-4 kV
Mid.-5 kV
High kV
Low kV
Small Focal Spot
Mid. kV
(SFS)
High kV
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Low volt. accuracy (LFS) kV PASS FAIL
Mid. volt. accuracy (LFS) kV PASS FAIL
High volt. accuracy (LFS) kV PASS FAIL
Low volt. accuracy (SFS) kV PASS FAIL
Mid. volt. accuracy (SFS) kV PASS FAIL
High volt. accuracy (SFS) kV PASS FAIL
Mid. volt. reproducibility (LFS) kV PASS FAIL
Remarks
Signature
7/13
Measured Values
Al 0.3 mm Al 0.5 mm
w/o Al
( ) ( )
Measured HVL (W/Rh) mR(μC/kg) mR(μC/kg) mR(μC/kg)
Measured HVL (W/Al) mR(μC/kg) mR(μC/kg) mR(μC/kg)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
HVL (W/Rh) mmAl - PASS FAIL
HVL (W/Al) mmAl - PASS FAIL
Remarks
Signature
8/13
Collimation assessment
Exposure Conditions
Compressed Compression
kVp mAs Target/Filter S Value
Breast Thickness Force
Uniform Expo.3
kV mm N
(Collimation)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
X-ray / Light field gap
mm - PASS FAIL
(Right/Left)
X-ray / Light field gap
mm - PASS FAIL
(Chest/Nipple)
X-ray / Image receptor field
mm - PASS FAIL
gap (Left)
X-ray / Image receptor field
mm - PASS FAIL
gap (Right)
X-ray / Image receptor field
mm - PASS FAIL
gap (Nipple)
X-ray field / Exposure table
mm - PASS FAIL
gap
Compression paddle edge (Chest wall) - - PASS FAIL
Remarks
Signature
9/13
Radiation output
Measured Values
mAs
Entrance air
mR(μC/kg)
kerma 1
Entrance air
mR(μC/kg)
kerma 2
Entrance air
mR(μC/kg)
kerma 3
Entrance air
mR(μC/kg)
kerma 4
Entrance air
mR(μC/kg)
kerma 5
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Reproducibility - PASS FAIL
Air kerma rate mGy/s - PASS FAIL
Specific radiation output μGy/mAs - PASS FAIL
Remarks
Signature
10/13
Measured Values
±0 +1 +2 +3 -1 -2 -3
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
mAs variation (±0 to +1) % PASS FAIL
mAs variation (+1 to +2) % PASS FAIL
mAs variation (+2 to +3) % PASS FAIL
mAs variation (-1 to ±0) % PASS FAIL
mAs variation (-2 to -1) % PASS FAIL
mAs variation (-3 to -2) % PASS FAIL
Remarks
Signature
11/13
AEC reproducibility
Measured Values
mAs
Entrance air
mR(μC/kg)
kerma 1
Entrance air
mR(μC/kg)
kerma 2
Entrance air
mR(μC/kg)
kerma 3
Entrance air
mR(μC/kg)
kerma 4
Entrance air
mR(μC/kg)
kerma 5
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
Entrance air kerma accuracy % PASS FAIL
Entrance air kerma
PASS FAIL
reproducibility
Remarks
Signature
12/13
CNR modes 1 to 3
Exposure Conditions
Exposure Mode
Compressed Compression
kVp mAs Target/Filter S Value
Breast Thickness Force
Mode n CNR
kV mm N
(PMMA20mm)
Mode n CNR
kV mm N
(PMMA40mm)
Mode n CNR
kV mm N
(PMMA60mm)
Mode n CNR
kV mm N
(PMMA70mm)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
CNR relative value 20 mm % - PASS FAIL
CNR relative value 40 mm % - PASS FAIL
CNR relative value 60 mm % - PASS FAIL
CNR relative value 70 mm % - PASS FAIL
Remarks
Signature
13/13
AGD modes 1 to 3
Measured Values
Exposure Mode
Al 0.3 mm Al 0.5 mm
w/o Al
( ) ( )
Entrance air kerma (20 mm) mR(μC/kg) mR(μC/kg) mR(μC/kg)
Entrance air kerma (40mm) mR(μC/kg) mR(μC/kg) mR(μC/kg)
Entrance air kerma (60mm) mR(μC/kg) mR(μC/kg) mR(μC/kg)
Entrance air kerma (70mm) mR(μC/kg) mR(μC/kg) mR(μC/kg)
Test Result
Criteria
Judgment Item Judgment Result
Lower Limit Upper Limit
AGD 20mm mGy - PASS FAIL
AGD 40mm mGy - PASS FAIL
AGD 60mm mGy - PASS FAIL
AGD 70mm mGy - PASS FAIL
Remarks
Signature
Chapter 9
Technical Information
Page
9
Technical Information
*1: DANCE, DR et al. Additional factors for the estimation of mean glandular breast
dose using the UK mammography dosimetry protocol. Phys. Med. Biol., 2000,
Vol. 45, 3225-3240
6 Dynamic range
In IEC 61223-3-2 Ed.2*2, visual evaluation on how many thin steps can be seen at the
high dose side is executed using a stepped PMMA phantom. In this Program, based
on the principle of IEC, a region for direct line has been reserved in the phantom and
dynamic range is quantified by calculating the average pixel value in this region. If this
value is the maximum value (in the case of 14 bits: 16384), there is a possibility of
image saturation.
7 Ghost Factor
The ghost factor is an index for representing the degree of change in sensitivity of the
9
flat panel detector (FPD) caused by the last X-ray exposure. The greater the degree
Technical Information
9
Technical Information
12bit
Digit
4 Digits
Figure 1 Relationship between QL Value and Digit (before EDR Normalization)
15 Relative Sensitivity
The relative sensitivity is a relative value calculated by using the average pixel value in
the calculation region exposed under a certain condition, based on the medial value
calculated by using the average pixel value in the specified calculation region of the
flat panel detector (FPD) to which a standard amount of dose is exposed.
18 Spatial Resolution
Spatial resolution is an index for representing the ability of the imaging system to
clearly define the shape of objects within an image.
19 System Sensitivity 9
The system sensitivity is equivalent to the median value of the histogram of the
Technical Information
digital image at the time of converting the dose on the flat panel detector (FPD) to a
digital value. The system sensitivity is defined depending on the type of FPD. In this
Guidebook, the high-resolution FPD is used. Accordingly, the system sensitivity is
defined as follows.
When uniform exposure is executed under the condition of 25 kVp and 0.03 mm Mo
using X-ray equipment with a molybdenum target, if the dose on FPD is specified as
y (mR) (1 (mR) ≈ 8.7 (µGy)):
S Value = 2400/y
9
Comment
Study time
Operator
Technical Information
Software version
Baseline
Baseline value-used flag
Exposure condition
X-ray source individual identification
data
Reader unit individual identification
data
Exposure menu
DetectorID
S Value
Tube voltage (kV)
mAs
Exposure1 Target / Filter
Compressed breast thickness [mm]
Compression force [N]
Exposure Mode
Dose mode
D-Tap
Exposure menu
DetectorID
S Value
Tube voltage (kV)
mAs
Exposure2 Target / Filter
Compressed breast thickness [mm]
D-Tap
Exposure condition
X-ray source individual identification
data
Reader unit individual identification
data
Exposure menu
DetectorID
S Value
Tube voltage (kV)
mAs
Exposure1 Target / Filter
Compressed breast thickness [mm]
Compression force [N]
Exposure Mode
Dose mode
D-Tap
9
Technical Information
9
Technical Information
9
including electrical hazards
Operator technique control charts are posted
X-ray equipment test 2 - Visual and
Operator protected during exposure by
functional test
Technical Information
9
Entrance air kerma(1)
Entrance air kerma(2)
AEC system performance - AEC
Entrance air kerma(3)
reproducibility
Technical Information
mAs
Exposure4 Target / Filter
Compressed breast thickness [mm]
Compression force [N]
Exposure Mode
Dose mode
D-Tap
9
Technical Information
9
Viewing box cleanliness
Baseline value creation
Study time
Technical Information
Operator
Software version
<Baseline_AnnualTest>
Test contents Judgment items
Lag Lag factor[QL]
Ghost Ghost factor
Missed tissue on chest wall edge Missed tissue on chest wall edge (R) [mm]
(40mm height) Missed tissue on chest wall edge (L) [mm]
4lp/mm[%]
Spatial Resolution (Magnification)
8lp/mm[%]
Verify mechanical stability
Verify that all moving parts more smoothly
All locks and detents work properly
No sharp, rough edges or other hazards
including electrical hazards
Operator technique control charts are posted
Visual and functional test Operator protected during exposure by
adequate radiation shielding
All indicator lights working properly
Auto decompression can be overridden and
status displayed
Compression can be manually released in
9
case of failure
Low volt. accuracy (LFS) [kV]
Mid. volt. accuracy (LFS) [kV]
Technical Information
9
CNR ratio 70mm_L - mode[%]
AGD 20mm_L - mode[mGy]
AGD 40mm_L - mode[mGy]
AGD L - mode
Technical Information
9
Technical Information
Page
1 3 4 5
Study Menu
MPM Code
AP/PA
Menu Description
2 Exposure Menu
S Value
L Value
PRIEF
the digital image.
Type
Type
L value : A logarithmic value showing the range of X-ray dose when making an exposure.
1 EDR Mode
EDR mode consists of AUTO mode, SEMI AUTO mode, and FIX mode.
A (AUTO mode) : A mode which adjusts density and contrast automatically.
(S and L values are dependent on this mode.)
S (SEMI AUTO mode) : A mode where the dynamic range (L value) of X-ray dose to be recorded as an
image has been determined, and the center point (S value) used for the purpose of
image recording is decided based on the average X-ray dose that enters the preset
photometric area so that the density is adjusted automatically.
F (FIX mode) : A mode where the range of X-ray dose to be recorded as an image has been
determined.
* Default settings for this mode can be changed using the User Utility. It is also possible to change the mode type currently being
used.
2 Auto - Parameters used in AUTO mode -
PRIEF (Pattern Recognizer for Iris of Exposure Field) .....
This is a generic denomination of processing that recognizes split exposures and irradiated field automatically.
PRIEF includes the following technique types:
- : Does not recognize split exposures and irradiated field. (SEMI AUTO mode and type IV described below.)
1 : Does not recognize split exposures and judges a rectangular area as an irradiated field.
1S : Recognizes split exposures and judges individually recognized areas as rectangular irradiated fields.
2 : Recognizes irradiated field of a breast.
4 : Does not recognize split exposures and judges a protrusive area as an irradiated field.
4S : Recognizes split exposures and judges individually recognized areas as protrusive irradiated field. (normal
mode)
4* : Judges a protrusive area as an irradiated field by split areas determined.
AN : Auto neck algorithm
SP : Activates AUTO mode based on a specially determined area, irrespective of the image size specified by DR
equipment.
TYPE ..... A type of technique for histogram or neuro analysis subjected after PRIEF processing.
I : A mode that captures regions covering from the skin to the bone in an image (Note that direct X-rays are
needed to activate this mode.)
II : A mode that is activated in a stable manner even if there are no direct X-rays.
III : A mode applied to contrast exposure.
IV : A mode that attaches importance to improved representation of soft tissue.
V : A mode that attaches importance to improved representation of areas where X-rays are difficult to be
penetrated.
VI : Neuro analysis mode applied when variations become large on the shape of a histogram.
VII : Neuro analysis mode applied when the position of a region of interest changes on a histogram.
3 Semi Fix - Parameters used in SEMI AUTO mode and FIX mode -
TYPE ..... Determines layout and size of a photometric area preset in SEMI AUTO mode.
I : A 10cm square located at the center of an image exposed by DR equipment.
II : A 7cm square located at the center of an image exposed by DR equipment.
III : A 5cm square located at the center of an image exposed by DR equipment
III' : A 5cm square located at a position other than the center of an image when it is divided into nine portions
up-and-down and right-and-left.
IV : A special area determined for the chest.
L value........ A logarithmic value (L value) representing the width of an X-ray dose to be recorded as an image in
SEMI AUTO and FIX modes.
S value....... A
center point (S value) pre-determined in FIX mode for recording as an Image.
* Default settings for this mode can be changed using the User Utility. It is also possible to change default settings by the
currently used mode.
2
1 Image Format
Monitor : Image display suitable for reading on the monitor is set (monitor display parameters).
Film : Conventional image display or equivalent is set (film output parameters).
1.2
0.8
Response
0.6 A B C D
0.4 E F G
0.2
0
0.001 0.01 0.1 1 10
Frequency (cy/mm)
MPM Code
EDR mode
AP/PA
S Value
L Value
PRIEF
Menu Description
Type
Type
Exposure Menu
Calculation 1/2
CNR
7302 AP [Calculation] For checkup of mammography QC CNR S - V I 1 200
Calculation 2/2
AGD
7302 AP [Calculation] For checkup of mammography QC AGD S - V I 1 200
1Shot PhantomM
1Shot PhantomM [Daily/Weekly] For checkup of mammography QC
7375 AP 1Shot Phantom M S - V III’ 2 80
ACR Phantom
ACR Phantom [Daily/Weekly] For checkup of mammography QC
7326 AP ACR Phantom S - V III’ 2 80
1Shot/ACR
1Shot PhantomM [Daily/Weekly] For checkup of mammography QC
7375 AP 1Shot Phantom M S - V III’ 2 80
Semi-annual 1/1
Uniform Expo [Semi-annual] For checkup of mammography QC
(Image Basic) 7302 AP Uniform Exposure (Image Basic) S - V I 1 200
Annual 1/6
ACR Phantom1 (Lag)
7313 AP [Annual] For checkup of mammography QC Lag1 F - V I 2 200
* High luminance is recommended in the environment where high-density film is used and normal luminance in the environment where a monitor, which does
not support high-density film, is used.
MPM Code
EDR mode
AP/PA
S Value
L Value
PRIEF
Menu Description
Type
Type
Exposure Menu
Annual 2/6
1ShotM (Magnification) [Annual] For checkup of mammography QC
7375 AP Spatial Resolution (Magnification) S - V III’ 2 80
Annual 3/6
Uniform Expo.3
(Collimation) 7303 AP [Annual] For checkup of mammography QC Collimation S - V I 2 200
Annual 4/6
H-mode CNR [Annual] For checkup of mammography QC
(PMMA20mm) 7302 AP H-mode CNR (20 mm height) S - V I 1 200
Annual 5/6
N-mode CNR [Annual] For checkup of mammography QC
(PMMA20mm) 7302 AP N-mode CNR (20 mm height) S - V I 1 200
Annual 6/6
L-mode CNR [Annual] For checkup of mammography QC
(PMMA20mm) 7302 AP L-mode CNR (20 mm height) S - V I 1 200
* High luminance is recommended in the environment where high-density film is used and normal luminance in the environment where a monitor, which does
not support high-density film, is used.