525 DX
525 DX
Agilent Technologies
Notices
© Agilent Technologies, Inc .2004, 2005 Warranty Safety Notices
No part of this manual may be reproduced in The material contained in this docu-
any form or by any means (including elec- ment is provided “as is,” and is sub-
tronic storage and retrieval or translation ject to being changed, without notice, CAUTION
into a foreign language) without prior agree- in future editions. Further, to the max-
ment and written consent from Agilent imum extent permitted by applicable A CAUTION notice denotes a haz-
Technologies, Inc. as governed by United law, Agilent disclaims all warranties, ard. It calls attention to an operat-
States and international copyright laws. either express or implied, with regard ing procedure, practice, or the like
to this manual and any information
Manual Part Number contained herein, including but not that, if not correctly performed or
G1535-90328 limited to the implied warranties of adhered to, could result in damage
merchantability and fitness for a par- to the product or loss of important
Edition ticular purpose. Agilent shall not be data. Do not proceed beyond a
liable for errors or for incidental or
Fourth edition, May 2005 consequential damages in connec- CAUTION notice until the indicated
Third edition, April 2005 tion with the furnishing, use, or per- conditions are fully understood and
Second edition, August 2001 formance of this document or of any met.
First edition, May 2000 information contained herein. Should
Printed in USA Agilent and the user have a separate
written agreement with warranty
Agilent Technologies, Inc. terms covering the material in this
2850 Centerville Road
WA R N I N G
document that conflict with these
Wilmington, DE 19808-1610 USA terms, the warranty terms in the sep- A WARNING notice denotes a
arate agreement shall control.
hazard. It calls attention to an
operating procedure, practice, or
Acknowledgements the like that, if not correctly per-
Torx® is a U.S. registered trademark of formed or adhered to, could result
Textron, Inc. in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated condi-
tions are fully understood and
met.
3 Operation
General Rules 48
Starting up 48
Shutting down 48
Photomultiplier (PMT) protection 48
Lit Offset 48
High-purity gases 48
Installing Columns 50
Capillary columns 50
Packed columns, 1/8-inch 51
Packed columns, 1/4-inch 52
A Replacement Parts
Inert Transfer Line 84
FPD Ignitor and Heat Shield Assembly 86
FPD Lens Assembly 88
PMT and Bracket Assemblies 90
Covers, Flow Modules, and Electronics 92
1
Installation
Kit Contents 9
Required Tools 11
Introduction 12
Single-wavelength FPD 12
Dual-wavelength FPD 12
Detector construction 12
Typical Installation 15
Overview 16
Prepare the GC 17
Remove the Covers and Panels 18
Prepare the Detector Location 19
Install the Flow Module 21
Install the Detector 24
Update the Firmware 31
Replace the Covers and Panels 33
Return the GC to Operating Condition 34
Agilent Technologies 7
1 Installation
The G2333B and G2334B kits each contain a flame photometric detector
(FPD) for the Agilent 6890 Series gas chromatograph (GC). Table 1 describes
each kit further.
Kit Description
G2334B Contains a dual-wavelength FPD, which enables you to collect both sulfur and
phosphorus data at the same time.
Kit Contents
Table 2 lists the parts for both kits.
FPD assembly, 1
(single detector + flow module + board)
FPD assembly, 1
(dual detector + flow module + board)
Cable tie 1 1
Hex nut, M4 1 1
• Column nut 1 1
• Column tool 1 1
• Ferrule, No-hole 1 1
• Wrench, 3/8-inch × 7 mm 1 1
It is neither necessary nor advisable to separate detectors from their flow modules. Doing
CAUTION so can create leaks. Although handling the detector and flow module as a unit is awkward,
it can be managed.
Required Tools
• T-20 Torx® screwdriver
• Knife
• Chip removal tool (Amp 821903-1 or equivalent)
• Column measuring tool (for capillary columns)
• Isooctane
• Column cutter
• Grounded wrist strap
• Clean, lint-free gloves
• Cotton swabs on wooden sticks
• Lint-free cloth
• Diagonal cutters
• 1/4-inch open-end wrench
Introduction
The column effluent is burned in a hydrogen-rich flame. The excited species
decay in the cooler region above the flame and emit light. This is viewed by a
factory-calibrated photomultiplier tube (PMT). A shield blocks light from the
flame itself, while filters isolate the light from sulfur or phosphorus species.
Single-wavelength FPD
You can install one or two single-wavelength FPDs. If you install only one,
place it in the back location.
Each single-wavelength FPD has a signal board that mounts in a slot on the
right side of the GC. The board for the front detector must be in the front slot;
the board for the back detector must be in the back slot.
The signal, high voltage, and ignitor cables from the detector connect to its
signal board. The heater/sensor cables (two per detector; one short and one
long) plug into connectors on the right side of the GC. Details appear later in
this document.
The flow module mounts in the pneumatics chassis at the back-top part of the
GC. There are specific module locations for the front and back detector
locations, which appear later in this document.
Dual-wavelength FPD
A dual-wavelength FPD has one emissions block assembly and two PMT
assemblies. It must be mounted in the back location. It uses two signal boards,
one in each slot. The ignitor cable must connect to the board in the back slot. It
does not matter which board serves which PMT, but the high voltage and
signal cables for a given PMT must go to the same board. The heater/sensor
cables go to the positions for a back detector. Details appear later in this
document.
The flow module goes in the back detector position.
Detector construction
The FPD consists of a detector body, either single or dual, and one or two PMT
assemblies. See Figure 1.
Body
PMT assembly
PMT assembly
Detector body
The detector body consists of:
• A base that mounts to the GC
• A transfer line that conducts the column effluent to the emissions block
assembly
• A sheet metal chimney that encloses the transfer line
• An emissions block assembly with one (single-wavelength) or two
(dual-wavelength) windows and one or two support brackets for PMT
assemblies
• One or two lens and filter holders that mount over the emission block
windows and connect to the PMT assemblies
• A vent tube to remove gases from the detector and prevent room light from
entering the emissions block assembly
PMT assembly
A PMT assembly consists of:
• A long tube to place the temperature-sensitive photomultiplier tube well
away from the heated emissions block assembly
• An end cap that contains the photomultiplier tube and socket
Typical Installation
Figure 2 shows the parts for a single-wavelength FPD mounted in the back
location. For the front location, the flow module goes in the slot to the right
(when facing the back of the GC) of the one shown and the detector and board
are moved to the front location and slot.
FPD
Detector flow
module
Electronics
pan cover
Hot warning
label
Auxiliary heater
bracket and harness
(shown assembled)
Signal board
(1 or 2)
A dual FPD will have two of the long photomultiplier assemblies and mount in
the back location. There will be two signal boards.
Overview
Perform this installation only if you are experienced in GC maintenance. If not, trained
CAUTION
Agilent personnel can install the FPD.
Prepare the GC
1 Turn off the GC and unplug the power cord.
Hazardous voltages are present in the instrument whenever the power cord is
WA R N I N G
connected. Avoid a potentially dangerous shock hazard by disconnecting the power
cord before working on the instrument.
Hydrogen gas is flammable and potentially explosive. Before installing the detector,
WA R N I N G
turn off the hydrogen gas at the source.
2. Right-side panel
Loosen the two captive screws
along the top edge. Slide the
panel toward the back. At the top,
tilt outward, and lift.
It is neither necessary nor advisable to separate detectors from their flow modules.
CAUTION
Doing so can create leaks. Although handling the detector and flow module as a unit is
awkward, it can be managed.
To avoid damaging the printed circuit board electronics, use a grounded wrist strap
WA R N I N G
(part number 9300-1408) and connect it to a bare metal surface of the GC.
The insulation on the GC is made of refractory ceramic fibers. Ventilate your work
WA R N I N G
area. Wear long sleeves, gloves, safety glasses, and a disposable dust/mist
respirator. Dispose of insulation in a sealed plastic bag.
2 There are two layers of insulation on top of the oven. Use a sharp knife to
cut through the top layer—the soft one—using the hole in the metal as a
template. Remove the soft insulation to expose the hard insulation.
3 The hard insulation has a precut hole that is filled with a plug. Push the
plug out into the oven and discard it
1 The tubing between the detector and flow module has been folded for
shipping. Unfold the tubing, being very careful to avoid kinks or tight bends.
2 Locate the correct slot for the flow module (Figure 6).
• If the detector is to be mounted in the back location, use the second slot
from the left (as viewed from the back of the GC). Note that a
dual-wavelength detector must be mounted in the back location.
• If the detector is to be mounted in the front location, use the third slot
from the left (as viewed from the back of the GC).
Hold the flow module by its support bracket to avoid damaging the components.
CAUTION
3 Route the ribbon cable behind the flow module assembly as shown in
Figure 4. Then slide the flow module and bracket assembly into the slot
until the bracket seats flush against the end of the rails (see Figure 5).
Back of
flow module
Ribbon cable
Tubing to detector
Bracket flush
with rail
4 Connect the ribbon cable to the mating connector on the pneumatics board.
The mating connector for the flow module is not the one directly under it.
The connectors are identified by raised lettering on the plastic immediately
above them.
Arrange the cable to keep it away from the valves on the right side and keep
it from being pinched against the flow module.
For the back detector, you may want to loosen the flow module and slide it
out of the carrier a few centimeters to connect the cable to the pneumatics
board. Then, reinstall the flow module.
a Secure the flow module in place using the T-20 Torx mounting screw on
the front of the pneumatics chassis.
b Use a pair of diagonal cutting pliers to remove the appropriate back-top
panel cutout. Also remove any cutouts needed to access other flow
modules or accessories installed in the GC. See Figure 6.
Insert tip of
pliers here
Front detector
Back detector
Detector body
1 Cut the shipping tie-wrap holding the PMT assembly to the bracket. Be
careful not to cut any wires. See Figure 7.
There are two tie-wraps. One is the cable tie-wrap that you do not want to cut.
CAUTION
2 The detector body has an extended bottom plate on one side. Place the
detector in the selected location with the extended plate to the right and
with the column fitting passing through the insulation into the oven.
3 The detector is held to the oven top by four screws, one at each corner.
Insert the two screws on the right but do not tighten them (see Figure 7).
4 Install the two screws on the left side through the chimney cover and the
detector base. Press the cover against the detector and tighten the screws.
5 Tighten the screws on the right side.
6 Bend the gas lines from the flow module downward to one of the T-slots in
the bracket on top of the oven. Pass the lines through the slot. Form excess
length into an S-curve on top of the oven.
7 The FPD produces considerable water vapor when the flame is on. Connect
the silicone tubing from the kit to the curved vent tube on top of the
detector. Run the other end to a suitable container or drain.
Heater/Sensor cables
The next steps require precautions against electrostatic discharge. Use the grounded
CAUTION
wrist strap (part number 9300-1408) and connect it to a bare metal surface of the GC.
Failure to heed this caution may result in damage to the instrument or to the printed
circuit board.
There are two heater/sensor cables for each detector; one for the emissions
block assembly and one for the transfer line. Both cables have woven glass
sleeves but one is much longer than the other.
Auxiliary heater
bracket
Wiring harness
3 Push the wiring harness ends into the auxiliary heater bracket from the
bottom as shown in Figure 11. The connectors are labeled.
A1 connector
A2 connector
4 Plug the large connector on the wiring harness into the P22 connector on
the main board (see Figure 9 and Figure 11).
5 Install the screws with the unthreaded shank into the bracket. The screws
will become captured by the bracket.
6 Attach the bracket to the GC frame.
7 Pass the long heater/sensor cable under the board and plug it into a
connector (see Figure 10) on the bracket as listed in Table 3.
Single-wavelength Front A1
Single-wavelength Back A2
Dual-wavelength Back A2
Signal boards
1 Install the detector signal board(s) on the right side of the instrument
behind the keyboard (see Figure 12).
Connector, bracket
on other side
2 Insert the board in the correct guide for the location used.
• For a single-wavelength detector, the board goes in the guide for the
detector position used.
• For a dual-wavelength detector, boards go in both guides. The board in
the back slot becomes the master board, providing high voltage and
signal processing for one PMT assembly and ignitor power for the
detector emissions block assembly. It controls the other board.
• The board in the front guide provides high voltage and signal processing
for the second PMT assembly.
3 Press the board firmly into place so that the connector on the board
engages the matching connector on the GC main circuit board.
4 Secure the board with the captive screw in the bracket.
With a dual-wavelength FPD, it does not matter which of the two circuit boards is used
CAUTION
for which PMT (see Figure 12). However, the high voltage and signal cables from each
PMT must go to the same board. Do not “cross-wire.”
1 Connect the cables to the signal boards. Collect the cables and route them
through the metal clip on the side of the PMT assembly support.
Figure 13 locates the detector board connectors.
High-voltage connector
Signal connector
Ignitor connector
Signal cable
The signal cable is black with a BNC connector on the end (Figure 15).
Press it onto the signal connector on the board and turn the knurled ring to
lock it in place.
Ignitor cable
The ignitor cable is gray and ends in a 2-pin flat connector (Figure 16).
Press it into the ignitor connector on the board until it locks into place.
There is only one way that it can be connected.
If you are installing a dual-wavelength detector, there will be only one
ignitor cable. Connect it to the board in the back detector slot.
The GC displays the current firmware version when you first turn it on or press [STATUS]
NOTE followed by [Clear].
The chips that came in the kit are identified by a part number, a version
number, and a suffix of either “.0” or “.1”. They are referred to as the 0-chip
and the 1-chip, respectively.
Be sure to note which chip you are removing. You want to be sure to place the correct
CAUTION
replacement chip in the correct socket.
3 Orient the chip so the round dimple is at the right. When the chip is
properly aligned, snap it into place in the socket indicated in Figure 17
using firm pressure.
4 Repeat step 2 and step 3 for the other chip.
6890N
To download the latest 6890N firmware, visit the Agilent Web Site at
www.agilent.com/chem, login or register, and navigate to firmware.
Pan
Folded lip
Top cover
Ground strap
Front clip
2
Checkout Procedure
Overview 36
Creating Methods 38
Phosphorus method 38
Sulfur method 40
Using ChemStation macros 41
Running the Tests 42
Interpreting Results 46
When you have installed the FPD, you should test it to verify that it works
correctly. This chapter describes a test protocol for running sulfur and
phosphorus samples through the GC and examining the results.
The protocol will differ depending on the equipment you have. This chapter
describes the protocol in terms of a ChemStation, but contains sufficient
information so that the tests can be run with just a strip-chart recorder.
Agilent Technologies 35
2 Checkout Procedure
Overview
This procedure assumes that the tester is familiar with the safe operation and maintenance
NOTE of the gas chromatography system, including the automatic liquid sampler, GC, and data
handling device.
This is the recommended sequence of steps for running the checkout. Install
the column.
1 Create and save the phosphorus and sulfur methods.
2 Load the phosphorus method, turn off the flame and electrometer.
3 Install the phosphorus filter.
4 Set inlet, oven, and detector to 250 °C to bake out for 15 minutes.
5 Reload the phosphorus method. Monitor signal output. This output
typically runs between 40 and 55 but can be as high as 70.
6 When it levels off, run one sample. Interpret the results (page 46).
7 Turn off the electrometer. Install the sulfur filter and spacer.
8 Load the sulfur method. Monitor signal output. The output typically runs
between 50 and 60 but can be as high as 70.
9 When it levels off, run one sample. Interpret the results (page 46).
Materials needed
• Column: 5% Phenyl Methyl Siloxane Capillary 30.0 m × 0.32 mm × 0.25 µm
(part number 19091J-413).
For Japan only: HP-5MS 30 m × 0.32 mm × 0.50 µm (part number
19091S-113).
• Checkout sample, part number 5188-5953.
For Japan only, part number 5188-5245.
• For cool on-column inlet use: 5-mm septa (part number 5183-4758, 50/pk),
5-μL on-column syringe for 320-µm column (part number 5182-0836), and
needle for 320-µm column (5182-0831).
• For split/splitless inlet use: 11-mm septa (part number 5183-4757 50/pk),
10-µL syringe (part number 5181-1267), or equivalent, and a liner. The liner
can be general purpose (part number 5183-4711) or single-taper
deactivated with (part number 5062-3587) or without (part number
5181-3316) glass wool.
For Japan only: 11-mm septa (part number 5183-4757 50/pk), 10-µL syringe
(part number 5181-1267), or equivalent, and a single-taper liner
deactivated without (part number 5181-3316) glass wool.
• Chromatographic-grade iso-octane for syringe wash solvent.
• Chromatographic-grade 99.9995% purity gas: helium as carrier, nitrogen as
makeup, hydrogen and air.
• 4-mL solvent and waste bottles or equivalent for autoinjector.
• 2-mL sample bottles or equivalent for sample.
Creating Methods
Create the test methods using these steps:
1 Create a phosphorus method (Table 4).
2 Save it using a name that identifies it as the phosphorus method (such as
P_Ckout.m).
3 Make a copy and modify the detector flow parameters in the copy to change
it to a sulfur method (page 40).
4 Save the copy using a distinctive name that identifies it as a sulfur method
(such as S_Ckout.m).
Phosphorus method
Table 4 contains the parameters for the phosphorus method. Give the method
a meaningful name that allows you to distinguish it from the sulfur method.
Sulfur method
Make a copy (Save As in ChemStation) of the phosphorus method. Rename the
copy to indicate that it is a sulfur method. Edit it as indicated in Table 6.
Parameter Value
H2 flow 50 mL/min
Air flow 60
These macros calculate the area counts, MDL, and noise, and print a report to
the printer setup in the ChemStation. The report includes the chromatogram,
the noise plot, and several calculations.
Place the macros in the appropriate subdirectory (Table 8) where
ChemStation can find them.
The macro can be set to run automatically as part of the method as a post-run
program, or can be run on a loaded data file from the ChemStation Data
Analysis view.
Methyl parathion
Isooctane
Methyl parathion
Isooctane
tert-Butyldisulfide
1-Dodecanethiol
Isooctane
Figure 21 Typical chromatogram for sulfur tests, sample 5188-5245 (Japan only)
Tributylphosphate
Isooctane
Figure 22 Typical chromatogram for phosphorus tests, sample 5188-5245 (Japan only)
If you receive a message like the following, check your printer setup and make any
CAUTION
corrections.
... has problems. , with page file:
C:\WINNT\TEMP\~P3D042A.TMP
Initial printing problem, 202.
System resources are low and/or device has problems.
Depending on the column length, the sulfur chromatograph may contain a contaminate
NOTE peak during the noise measurement. Reprocess the run by loading the signal and typing
macro mepck_s.mac,go,3
in the command line. This starts the noise measurement at 3 minutes instead of the default
3.8 minutes.
Interpreting Results
Compare the reported output against the values listed in Table 9 and Table 10.
Your value should be within the limits listed.
For phosphorus
For sulfur
Noise 2 to 4 ≤5
Half-width (min) 0.06 to 0.08 —
3
Operation
General Rules 48
Installing Columns 50
This chapter describes new or changed operating information (at the time of
this printing) that updates the information found in the 6890 user
documentation. Refer to the GC user information for any topics not covered
here.
Agilent Technologies 47
3 Operation
General Rules
Starting up
The FPD creates a great deal of water vapor when the flame is on. This could
condense in the vent tube on top of the detector and drop onto the flame,
possibly extinguishing it. To avoid this, turn the heaters on, wait 20 minutes
for the vent to heat up, and then ignite the flame. Water vapor will now make it
over the top of the vent tube before condensing.
Shutting down
For similar reasons, extinguish the flame before turning the heaters off.
Lit Offset
The default Lit Offset is 2.0 pA.
High-purity gases
High-purity gases have a lower sulfur content. Standard purity gases have a
higher sulfur content which impairs sulfur detection in the compound being
studied. Instrument or Chromatographic grades work well.
Agilent recommends using helium carrier, nitrogen makeup gas, and air with
99.9995% purity or better. Use hydrocarbon, oxygen, and moisture traps.
Select traps to remove sulfur compounds from detector air and nitrogen gases.
A helium getter is also recommended.
Inlet liners
Compounds containing sulfur may adsorb to an inlet liner and degrade the
GC’s performance. Use deactivated, clean liners or a cool on-column inlet,
which injects directly onto the column.
For best results with splitless injection, use liner 5181-3316.
Installing Columns
Capillary columns
Wear safety glasses while handling, cutting, or installing fused silica capillary columns.
CAUTION
To prevent puncture wounds, use care when handling these columns.
The following procedure ensures that the FPD works efficiently. A special tool
is used to cut the column to the correct length.
Column nut
Ferrule Column measuring tool
3 Tighten the column nut until it grips the column. Use a pair of wrenches to
tighten it an additional 1/8 to 1/4 turn.
4 Use a wafer cutter at 45° to score the column.
5 Snap off the column end.
6 The column may protrude 1 to 2 mm beyond the end of the tool.
7 Remove the column, nut, and swaged ferrule from the tool.
FPD fitting
Capillary adapter
1/8-inch nut
2 Carefully thread the swaged column up into the adapter. Finger-tighten the
column nut, then tighten an additional 1/8-turn with a wrench.
FPD fitting
1/8-inch ferrule
1/8-inch nut
Adapter
2 Connect the 1/4-inch column to the adapter in the usual way. Use Type B
glass columns.
4
Maintenance and Troubleshooting
Changing the Wavelength Filter 54
Accessing Heaters and Ignitor 58
Replacing the Heater/Sensor Assemblies 62
Replacing the Ignitor 69
Replacing the Inert Transfer Line 73
Troubleshooting 76
Agilent Technologies 53
4 Maintenance and Troubleshooting
Do not touch the filters with your bare hands. For optimum performance and to avoid
CAUTION
scratches, use lint free gloves for assembling and inserting the filters into the
assembly.
Materials needed
• 1000-1437 Sulfur filter and 19256-20910 filter spacer
• 19256-80010 Phosphorus filter
• 5080-5400 Cotton swabs
• Toothpick or cotton swab
• Lens tissue
• 8650-0030 (large) or 8650-0029 (small) Nylon lint-free gloves
Procedure
1 Turn off the detector electrometer.
2 Disconnect the retaining spring that holds the PMT assembly to the bracket.
With a rotating motion, pull the assembly away from the filter housing
(Figure 26).
3 To prevent light from damaging the PMT, immediately cap the end or place
it face down (Figure 27).
4 Place a clean cloth under the filter housing to catch the filter.
Phosphorus filter: use the sharpened wooden tip of a toothpick or cotton
swab to dislodge the filter from the housing.
Sulfur filter: (Figure 28) use the wooden tip of the cotton swab to remove
the filter spacer. Then remove the filter as above.
Sulfur filter
Filter spacer
5 Use lens tissue to clean the new filter. Install it in the filter housing. If you
are installing a sulfur filter, add the filter spacer.
Do not use cleaning fluids. Cleaning fluids will damage lens coatings.
CAUTION
Filters are designed for the light from the flame to pass through in a specific direction.
CAUTION
On the edge of the phosphorus filter, there is a triangle. On the edge of the sulfur filter,
there is an arrow. These should face away from the flame and toward the PMT.
6 Replace the PMT assemblies and secure them with the springs.
7 Restore the operating conditions.
When turning the GC off, turn off the flame first to prevent condensation from dripping
CAUTION
into the jet and column.
You may wish to access this area for the following reasons:
• Replace the ignitor.
• Replace seals and O-rings.
• Replace the transfer line assembly.
• Replace or check the heater and PRT sensors.
Materials needed
• 8710-0803 9/16-inch wrench
• 8710-1807 T-20 Torx driver
Procedure
4 Use a wrench to loosen and remove the vent tube assembly (Figure 29).
6 Dual-wavelength detector only. Remove the four screws securing the Dual
FPD chimney front (with the attached dual main bracket and PMT
assembly). Remove the entire left part of the detector. Follow the order of
operations in Figure 32. Cover the open end of the PMT assembly to protect
the tube from light.
3. Remove PMT
assembly and
attached parts.
When turning the GC off, turn off the flame first to prevent condensation from dripping
CAUTION
into the jet and column.
You may wish to replace or check the FPD heater/sensor assemblies for the
following reasons:
• One or both of the heaters or sensors are defective.
• The actual temperature reading on the display of the heaters is cycling more
than 1 °C.
Materials needed
• G1535-60610 Heater/Sensor assembly with short lead for the emissions
block assembly.
• G1535-60620 Heater/Sensor assembly with long lead for the transfer line.
• ESD protection: For example, a wrist strap
• Wrenches: 8710-0803 (9/16-inch), 8720-0010 (5/8-inch), 1/4-inch × 3/8-inch
• Torx drivers: 8710-2140 T-10 and 8710-1807 T-20.
Procedure
1 Turn off the flame.
2 Turn off the GC.
3 Remove the vent assembly and cover. Refer to “Accessing Heaters and
Ignitor” on page 58.
4 Put on an ESD wrist strap. Remove the right-side cover to access the GC
electronics. Disconnect the heater/sensor leads from the auxiliary heater
board and the connectors above the main board.
5 Carefully pull the cables up onto the top of the GC.
Transfer line
1 Use a Torx T-10 driver to remove the screw and retainer clip holding the
lower heater/sensor assembly (Figure 33). Remove the heater and sensor
from the transfer line.
Retainer clip
3 Insert the heater and sensor into the transfer line. Make sure the sensor is
seated at the bottom of the hole. If not, the AUX temperature will wander
above and below the detector setpoint.
4 Position the retainer clip over the heater/sensor assembly and install the
screw.
3 Install the upper heater and sensor in the emissions block assembly.
Closing up
1 Route the heater/sensor cables out of the bracket (Figure 37).
2 Replace the cover, vent assembly, and, on the dual-wavelength detector, the
left PMT assembly. Refer to “Accessing Heaters and Ignitor” on page 58.
3 Carefully thread the heater/sensor cables into the electronics compartment.
4 Put on an ESD wrist strap. Connect the short cable as shown in Figure 6 on
page 23. Connect the long cable to the auxiliary heater board (Table 11).
Single-wavelength Front A1
Single-wavelength Back A2
Dual-wavelength Back A2
Materials needed
• 19256-60800 Ignitor replacement kit
• Wrenches: 1/4-inch × 5/16-inch (8710-0510)
• Torx driver: T-10 (8710-2140)
• 8650-0030 (large) or 8650-0029 (small) Nylon lint-free gloves
The FPD Ignitor replacement kit includes a glow plug, a spacer, and an O-ring.
Do not use the copper ring that ships with the glow plug.
Procedure
1 Turn off the flame.
2 Turn off the GC.
3 Remove the vent assembly, cover, and, on the dual-wavelength detector, the
left PMT assembly. Refer to “Accessing Heaters and Ignitor” on page 58.
4 Use a Torx T-10 driver to loosen the collar screw holding the cable assembly
to the ignitor (Figure 38). Remove the collar and cable assembly. Some
FPDs have a version of the collar with two screws.
5 Use a wrench to loosen and remove the glow plug. Use tweezers to remove
the O-ring.
Glow plug
Spacer
O-ring
9 Replace the cover, the vent tube assembly, and, on the dual-wavelength
detector, the left PMT assembly.
10 Turn on the GC. Confirm that the flame is off.
11 Restore the operating conditions.
12 Wait 20 minutes for the detector to heat up, then ignite the flame.
Mounting nuts
To avoid damaging the photomultiplier tube, always turn the GC or electrometer off
CAUTION
before removing the PMT assembly.
3 Inside the oven, remove the column and adapter (if installed) to the FPD.
4 Locate the gray ignitor cable attached to the side of the detector. Trace the
cable to the signal board and disconnect it there.
9 Grasp the emissions block assembly and carefully lift it straight up. The
transfer line will come with it. Twist and pull to separate the two parts.
If the emissions block assembly/transfer line will not lift free of the detector
bracket, check for tubing or cables catching on the sheet metal.
10 If you will be installing a new transfer line assembly, trace the gas tubing to
the manifold block on the flow module. Remove the single screw that holds
the fitting. Install new O-rings.
Reassembly
1 Insert the transfer line into the bottom of the emissions block assembly.
2 Slide the combined parts into the detector body. The semicircular plates
that the three screws connect to go into the gap between the two parts of
the bracket.
3 Install the two mounting nuts that hold the transfer line.
4 Check that the emissions block assembly is fully seated on the transfer line
and that all of the cables and tubing are properly placed.
5 Tighten the screws that secure the lens holders.
6 Install the PMT assemblies).
7 Reconnect the ignitor cable.
8 Install the cover and the vent tube.
9 If necessary, connect the gas lines to the flow module.
10 Restore operating conditions.
11 Wait 20 minutes for the vent tube to heat up, then ignite the flame.
Troubleshooting
Outputs
Gas flows optimized for With sulfur filter With phosphorus filter
Sulfur 30 to 50 10 to 12 (low)
Besides having the wrong filter for a particular set of gas flows, also consider:
• If the Lit Offset is 0.5 to 3.0, check that the flame is ON.
• If the Lit Offset is 0, check if the electrometer is turned OFF or the signal
cable is disconnected.
• If the Lit Offset <30, the flame may be in the wrong position. Check detector
flows, column flow, and column position.
S filter 3.8 to 5 ≤6
The MDL is dependent on the peak area and the detector noise. The
relationship is: the smaller the peak area and the larger the noise
measurement, the larger the MDL. If the MDL is high, check under the “High
noise level” on page 77 and “Small peak area” on page 77.
S filter 2 to 4 ≤5
If the AUX temperature for the FPD is cycling or wandering, view the PID
thermal diagnostic for the AUX heater by:
1 On the keypad, press Options.
2 Scroll to and select Diagnostics.
3 Scroll to and select Thermal PID control.
4 Scroll to and select AUX 1 or 2. This allows you to view the temperature
range and duty cycle. Check the table above for normal ranges.
Clipped peaks
If you have an application at the upper limit of the dynamic range (especially
with sulfur), you may have to desensitize your instrument. Replace the
1000-1437 sulfur filter with filter part number 19256-80000. Than set the
detector gas flows to the values used in the phophorus checkout method. This
raises the baseline but with some loss in the signal-to-noise ratio.
There should be no problems with either solution and the use of hydrogen.
However, please note the following warning.
Hydrogen gas is flammable and potentially explosive. Be sure to turn off the
WA R N I N G
hydrogen gas at the source until connections are made. Also, leak test connections,
lines and valves before operating or servicing the instrument.
A
Replacement Parts
Inert Transfer Line 84
FPD Ignitor and Heat Shield Assembly 86
FPD Lens Assembly 88
PMT and Bracket Assemblies 90
Covers, Flow Modules, and Electronics 92
Agilent Technologies 83
A Replacement Parts
9 Screw, M3 × 12 mm 0515-1084
1 3
9
10
7
4
5
8
Table 19 FPD ignitor and heat shield assembly parts (Figure 44)
12 Collar 19256-20690
3
4
6
7
8
A
13 * B 9
12 C
1 Clamp 19256-00090
3
4
5
4 Bracket/Support G1535-00010
6 Filters
Sulfur 1000-1437
Phosphorus 19256-80010
7 Filter spacer (used only with sulfur filter) 19256-20910
7
2
5
3
10
4
5b
11
12
1 FPD assembly
Single wavelength G1535-61555
Dual wavelength (not shown) G1535-61565
2 Top cover
Single wavelength G1535-80555
Dual wavelength G1535-80560
3
2
6C
6A
6B