Using The Lab Instruments
Using The Lab Instruments
Nicole Hamilton
http://faculty.washington.edu/kd1uj/
Our lab instruments
Not all of our labs have the same instruments. Depending on
whether you’re working in Beardslee or Discovery, these are
the instruments you may encounter.
1. Tektronix DMM4020 multimeter
2. Keithley 2110 multimeter
3. Tektronix PWS 4205 programmable power supply
4. Keithley 2230-30-1 power supply
5. RSR HY3002-3 triple output power supply
6. Tektronix AFG3022B arbitrary function generator
7. Tektronix MSO3012 dual channel oscilloscope
Tektronix DMM4020 multimeter
DMM4020 multimeter features:
+
Continuity or diode test
+
If the buttons don’t work
The usual problem is that it’s set for remote operation.
Press the shift key to exit remote mode
Keithley 2110 multimeter
Keithley 2110 multimeter features:
Slave
Slave Slave
Tektronix AFG3022B arbitrary function generator
AFG3022B function generator features:
Ext Trig
+
XFG1 _
A B
+ _ + _
27kΩ
Here’s what I set.
RInternal
VS
RLoad Vout
The reason it’s 50 Ω
...
The signal propagates as power is passed from one stage to the next.
Think of a ripple on a pond.
This makes high quality coax a very low loss delay line, with some
characteristic impedance Z0.
For maximum power transfer, the impedances should be
matched: ZS = Z0 = ZL
ZS
Source ... ZL
Because it’s a delay line, if there’s an impedance mismatch between the coax and
the load, you get reflections.
The signal literally bounces back, the same way a ripple on a pond would if it hit a
rock.
This creates standing waves and higher peak voltages at the source than expected.
Power that was intended to be transferred to the load is instead dissipated as heat
in the transmitter, possibly damaging the transmitter.
To avoid reflections, both ends of the coax must be
terminated by 50 Ω loads.
It actually measures.
Tektronix MSO3012 oscilloscope
DC coupling
AC coupling
Adjustments for horizontal and vertical scales and vertical position.
Adjustments to set the trigger level for recognizing the start of the cycle.
To add an on-screen measurement, press Measure, select Add
Measurement, then scroll through the choices with the B knob.
To get good measurements, wait until the mean values settle down.
To add cursors, press Cursors, then use the A and B knobs to adjust
their positions.
The desired result: A good screenshot with big, clear waveforms,
helpful on-screen measurements and cursors and no menus.