OPENMARK VET

A specific veterinary MRI system


MRI System Specification

SHENZHEN ANKE HIGH-TECH CO., LTD.


Ver: MROM03VET-2305
Shenzhen Anke High-tech Co., Ltd.

I. Magnet system
Magnet Parameters

Magnet type Permanent

Magnetic field strength 0.3T ± 5%

Magnet appearance C-shape


Weight 12.8T ± 1.5%

Horizontal opening angle 320°

5 Gaussian fringe field X, Y, Z directions ≤ 2.5m

Shimming

Shimming methods: Passive Shimming + Active Shimming

Magnetic field homogeneity (VRMS measurement method, 24 points 24 sides)

40cm DSV (VRMS) ≤ 2.0ppm

www.anke.com 1
Shenzhen Anke High-tech Co., Ltd.

II. Gradient system


General features

Gradient type Ultrafast solid-state technology

Gradient amplifier cooling type Air cooling

Gradient coil cooling type Air cooling

Gradient performance

Max. amplitude (single axis) 28mT/m

Max. slew rate (single axis) 93T/m/s

Min. rise time 0.3ms

Resolution parameters

Max. FOV 400mm

Min. FOV 10mm

Min. 2D thickness 1.0mm

Min. 3D thickness 0.1mm

Max. Acquisition matrix 1024×1024

Min. TE (Spin Echo) 5ms

Min. TR (Spin Echo) 11ms

Min. TE (Gradient Echo) 2ms

Min. TR (Gradient Echo) 5ms

Max. b-value of DWI 1000

www.anke.com 2
Shenzhen Anke High-tech Co., Ltd.

III. RF (Radio Frequency) system


RF parameters

Type of RF system Full digital transmitting and receiving

RF amplifier maximum power 6 kW

Receiving channel number 4 channels

Receiving bandwidth 1.25 MHz


Standard PA (Phased Array) coils

L-Vet Coil Coil type Phased array receiving coil

Channels 4

External dimension 300mm × 264mm × 270mm (L×W×H)

• Head imaging
• Spine imaging
Clinical Applications • Abdomen imaging
• Pelvis imaging

M-Vet Coil Coil type Phased array receiving coil

Channels 2

External dimension 288mm × 246mm × 273mm (L×W×H)

• Head imaging
• Spine imaging
Clinical application
• Abdomen imaging
• Pelvis imaging
S-Vet Coil Coil type Phased array receiving coil

Channels 2

External dimension 200mm × 190mm × 190mm (L×W×H)

• Head imaging for small animals


Clinical application • Spine imaging for small animals
• Joints imaging for small animals

www.anke.com 3
Shenzhen Anke High-tech Co., Ltd.

Optional PA (Phased Array) coils 1)

Vet coil – Extra Large Coil type Phased array receiving coil

Channels 4

External dimension 300mm × 330mm × 280mm (L×W×H)

• Head imaging
• Spine imaging
Clinical application
• Abdomen imaging
• Pelvis imaging
Head coil Coil type Phased array receiving coil

Channels 4

External dimension 360mm × 340mm × 320mm (L×W×H)

Clinical application • Head imaging

Neck coil Coil type Phased array receiving coil

Channels 2

External dimension 420mm × 400mm × 290mm (L×W×H)

• Neck imaging
Clinical application
• Joints imaging
9-Inch Flex Coil Coil type • Single receiving coil

Channels • 1

External dimension • 60mm×250mm×250mm(L×W×H)

Neck imaging
Clinical application
• Joints imaging for large animals

6-Inch Flex Coil Coil type • Single receiving coil

Channels • 1

External dimension • 60mm×170mm×170mm(L×W×H)

• Neck imaging
Clinical application
• Joints imaging for small animals

1) All the Optional coils are not included in the standard offer, please contact to ANKE for future information about technology and
price.

www.anke.com 4
Shenzhen Anke High-tech Co., Ltd.

IV. Patient handling system


Patient table

Patient table type Integrated floating patient table

Operation mode Electric / manual dual-mode

Positioning type Laser cross cursor positioning


Position horizontal accuracy ± 0.5mm

Max. Patient weight load 200kg (440Ibs)

Patient table control Touch screen control

Patient table display Color LCD Monitor

Patient communication

Intercom system Two-way intercom, adjustable

Emergency call button Provided

Respiratory gating Provided

www.anke.com 5
Shenzhen Anke High-tech Co., Ltd.

V. Console system
Host computer

Computer operating system Windows professional (64-bit)

Operation software APEX intelligent operation platform

CPU ≥ 3.0GHz (Intel-i7, 8-core)

RAM ≥ 16GB

Video memory 2.0GB

Hard disk 1TB × 2

Media drives CD/DVD drive

External storage of image data DVD/USB

Image transmission interface DICOM 3.0

Image storage capacity (256×256 Matrix, uncompressed) Approx. 4,000,000 (256 × 256)

Image reconstruction speed (256x256, 100% FOV) Approx. 3,300 fps

Max. Image reconstruction matrix 1024 × 1024

Color LCD Monitor

Display size 23.8”

Display resolution 1920×1080

www.anke.com 6
Shenzhen Anke High-tech Co., Ltd.

VI. Sequence and scanning technology


Sequence

Spin Echo SE (Spin Echo)


sequences
FSE (Fast Spin Echo) 2D/3D, combined with fast recovery technology, provide a
better T2 contrast and reduce the TR time of FSE sequence.

PROP FSE, a kind of technology to eliminate motion artifact

SSFSE (Single Shot Fast Spin Echo) 2D/3D, combined with Half-Fourier Acquisition
technology, reduce the scanning time.

IR (Inversion Recovery)
STIR (Short Time Inversion Recovery) provide a good quality fat suppression imaging.
FLAIR (Fluid Attenuated Inversion Recovery) combined with fat suppression
technology, to provide a good quality T1 and T2 FLAIR imaging.

DIXON (Water and Fat Separation) technology to provide good quality of fat, water, in-
phase and out of-phase imaging.

Gradient Echo GRE (Gradient Echo) 2D/3D combined with Spoiled technology
sequences
GRSCOUT can provide single & multi slices three-dimensional positioning imaging.

GREDE (Gradient Echo with Dual Echo) 2D/3D, GREDESP (Gradient Echo with Dual
Echo with Shared Phase) 2D/3D for in-phase/out of-phase imaging

GREME (Gradient Echo with Multi Echo) 2D/3D, for high contrast T2 weighted
imaging, can effectively suppress flow artifacts of CSF and blood

TFE (Turbo Field Echo) 2D/3D/4D for abdominal imaging during free breathing, and
fast 3D/4D dynamic contrast imaging.

IRGRE (Inversion Recovery Gradient Echo) 2D/3D, can be used for neurology imaging
to increase the contrast between white matter and gray matter.

TOF (Time of Flight) 2D/3D

Echo Planar EPI (Echo Planar Imaging) with Single Shot and Multi Shot technology for high definition
sequences diffusion weighted imaging.

SEEPI (Spin Echo Planar Imaging)

GREEPI (Gradient Echo Planar Imaging)

DWEPI (Diffusion Weighted Echo Planar Imaging)

DWEPIMS (Diffusion Weighted Echo Planar Imaging with Multi Shot)

LSDW (Linear Shot Diffusion Weighted) imaging

www.anke.com 7
Shenzhen Anke High-tech Co., Ltd.

Scanning process

Automatic pre-scan technology Automatic RF correction

Automatic frequency locking

Automatic gain adjustment

Automatic coil identification

Automatic phase correction

Automatic shimming linear compensation

Protocol management Scanning scheme presetting and selection technology

Batch scanning technology (Scan Sequence Queued Waiting)

Interval scanning technology

GRSCOUT 3D positioning 3D single-slice rapid positioning scanning technology

3D multi-slice rapid positioning scanning technology

Multiple groups of arbitrary positioning technology

Fast imaging technology Rectangular FOV acquisition technology


Half-Fourier acquisition technology
Partial-echo acquisition technology
Key-hole imaging technology
K-space data sharing technology
Partial acquisition technology
Phase encode gradient optional technology
Ellipse-encoded acquisition imaging technology
Off-center Imaging
Arbitrary slice scanning technology
Optimization bandwidth acquisition technology

Artifact suppression technology

Pre-saturation technology RF saturation pulses to suppress flow and motion artifact.


• Multi-slice & multi-angle pre-saturation technology
• Pre-saturated automatic tracking technology

Flow compensation technology Can effectively compensate for image artifacts caused by
liquid flow.

Oversampling technology Effectively avoid image artifacts.

PROP (Periodically Rotated Overlapping Improves image quality by rotated data filling in K-space to
Parallel) data filling technology correct the effects of motion during MR sequence

www.anke.com 8
Shenzhen Anke High-tech Co., Ltd.

acquisition.

Gating trigger technology Use the Respiratory gating to perform triggered scanning,
can accurately suppress image artifacts caused by
physiological motion.

Breath-hold scanning technology K-space share technology combined with patient breath-
holding to achieve fast breath-hold scanning, which can
effectively improve the success rate of examination.

www.anke.com 9
Shenzhen Anke High-tech Co., Ltd.

VII. Clinical application packages


Neurology Imaging suite

• Fast 2D&3D imaging based on SE, FSE, GRE sequence


• Diffusion imaging with multiple b-values, ADC-map, eADC-map
• LSDW (Linear Shot Diffusion Weighted) imaging
• T1-FLAIR, T2-FLAIR combined with water suppression technology
• STIR fat suppression technology
• High-definition inner ear imaging
• T2*-GREME 2D&3D to avoid CSF and blood flow artifacts
• Water and fat separation technology
• High definition myelography

Body Imaging suite

• 2D&3D imaging based on SE, FSE, GRE sequence


• Diffusion imaging with high b-value
• STIR fat suppression technology
• In-phase & out-phase imaging
• Free breath scanning
• Ultra-fast breath hold scanning
• Triggered scanning
• MRCP 2D&3D
• MRU 2D&3D

Orthopedics Imaging suite

• High resolution 2D&3D imaging based on SE, FSE, GRE sequence


• Off-center positions imaging
• STIR fat suppression technology
• Water and fat separation technology
• Water excitation with fat suppression imaging technology
• High-definition 3D imaging for small joints

Oncology Imaging suit

• Diffusion imaging with high b-value


• STIR fat suppression technology
• Water and fat separation technology
• Contrast enhancement imaging

www.anke.com 10
Shenzhen Anke High-tech Co., Ltd.

Angiography Imaging suite

• TOF 3D MR angiography
• TOF 2D MR Venography
• MTC (Magnetization Transfer Contrast) technology and TONE (Tilted
Optimized Non-saturation Excitation) pulse to improved Contrast to Noise
Ratio of images
• MIP, MinIP and MPR reconstruction software

www.anke.com 11
Shenzhen Anke High-tech Co., Ltd.

VIII. Installation
Radio frequency shielding

For shielding the examination room from external RF sources

RF attenuation factor >100dB

Frequency range 1−80MHz

Magnetic shielding5)

For additional reduction of the magnetic fringe field, suitable iron shielding can be installed in the walls of
the examination room. The room shielding can be used to create a magnetic shielding enclosure

Power Requirements

Line voltage 3N-380VAC

Stability tolerances ±10%

Line frequency 50/60 Hz, ±1 Hz

Maximum input value 15 kVA

Power consumption

Stand-by ≤1 kW

Typical measurement ≤12kW

Recommended site dimension

Room Recommended Dimension: L×W×H Smallest Dimension: L×W×H

Scanning Room 5.5m×4.5m×3.2m 5m×4.2m×2.8m

Equipment Room 2.0m×3.0m×3.2m 2.0m×3.0m×2.8m

Operation Room 3.8m×4.5m×3.2m 3.0m×2.0m×2.5m

5) The implementation plan of magnetic shielding depends on the condition of customer’s site, for more information, please refer to
the site planning manual.

www.anke.com 12
Shenzhen Anke High-tech Co., Ltd.

IX. Site Planning

Equipment room
Scanning room

Operation room

www.anke.com 13
Shenzhen Anke High-tech Co., Ltd.
Address: Block B, LingYa Industrial Zone, Tangtou No.1 Road, Bao'an Twitter: ANKE (@ankeint)
District, Shenzhen, 518108, P.R. China
Facebook: ANKE (@ankemri)
Tel: +86-755-21622518 26688889
Fax: +86-755-26695307 26685908 Instagram: ANKE (@ankemarcom)
Email: [email protected] Linkedin: Shenzhen Anke High-tech Co., Ltd.
Website: www.anke.com
Specifications and appearance of the equipment may change without prior notice. ANKE reserves the right of final interpretation.