ACUSON Sequoia
ultrasound system
Version VA25 SW
siemens-healthineers.com/ultrasound
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
Table of Contents
System Overview 6
Formats7
2D-mode Features 8
Color Features 8
Power Features 8
M-mode Features 9
3D-mode Features 9
4D-mode Features 10
Cybersecurity10
Imaging Features 11
Harmonics11
Compounding11
Clarify Technology 11
Speed of Sound 11
Panoramic Imaging 11
Modality Compare 11
Biopsy12
Clips12
Protocols12
DICOM12
2 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
eSieCalcs Software 13
Mobility13
User-Accessible Connections 14
Monitor14
Footswitch Option 16
Printer Options 17
Transducer Options 17
siemens-healthineers.com/ultrasound 3
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
Service Options 18
Connect Platforms 18
System Requirements 19
Power Supply Requirements 19
Environmental Requirements 21
Ultrasound System 21
System Classifications 22
4 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
Standards Compliance 22
Quality Standards 22
Design Standards 22
CE Declaration 23
EU Authorized Representative 23
siemens-healthineers.com/ultrasound 5
ACUSON Sequoia ultrasound system –
See More. Know More. Do More.
Experts agree that a conventional one-size-fits-all approach
to patient care falls short of addressing the challenges of
demanding caseloads, patient diversity and user variability.
• 2D-mode
You need a better, smarter system with the ability to - 2D-mode with Harmonics imaging
scan more patients in less time all while delivering • Color flow Doppler
clearer results. - Color
The ACUSON Sequoia is designed for the best-in-class - Power
image quality and boasts advanced hardware and - Directional Power
components for greater depth and resolution. - Slow Flow State
• Pulsed Wave Doppler
- Pulsed Wave Doppler Tissue Imaging
- High Pulsed Repetition Frequency Pulsed Wave
Doppler
• Continuous Wave Doppler
- Steerable Continuous Wave Doppler for phased array
transducers
- Auxiliary Continuous Wave Doppler for pencil
transducers
• M-mode
- M-mode with Harmonics imaging
- Anatomical M-mode
6 siemens-healthineers.com/ultrasound
22“ OLED monitor
High resolution and
dynamic display
Integrated
storage shelf
Allows easy storage User-designed
of accessories
control panel
Ergonomically designed
for simplified workflow
Compact Pinless
Quiet operation Connector Ports
Less than a library Easy access transducer
connector ports and
integrated storage shelf
Mobility
Lightweight (125 kg)
and small foot print
design with integrated
central locking break
siemens-healthineers.com/ultrasound 7
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
8 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
• Optimization of scale, baseline, gain and/or dynamic • Sweep speed selections: 8 mm/s to 200 mm/s
range selected manually or upon entering freeze • User-selectable transmit frequencies: up to 15.0 MHz
• Pulsed wave Doppler tissue imaging available for all • User-selectable edge enhancement selections: up to 4
cardiac and fetal echo exams on vector and curved • Maps: up to 7
array transducers
• Tints: up to 10
• Anatomical M-mode for the cardiac exam supports
Continuous Wave Doppler Features
visualization of an M-mode sweep by rotating the
• Fast Fourier Transformation (FFT) processing: up to M-mode cursor off axis
256 points
• FFT speed: up to 2,880 FFTs per second 3D-mode Features
• User-selectable transmit frequencies per transducer: • Acquires three-dimensional images and maximizes
up to 2 resolution for assessing structures
• Simultaneous 2D-mode and Doppler display • User-adjustable region of interest and volume of
• User-selectable Doppler update mode interest during setup (ROI/VOI), for size, position and
• User-adjustable Doppler scale and baseline position curve (VOI only) and position
controls • Available in combination with 2D-mode
• User-adjustable Doppler UltraArt universal imaging • Requires the compatible transducers: 9VE4
and edge controls • Supported studies: Gynecology
• Angle correction in one degree increments: 0° to 89° • Sweep speed quality selections: up to 3
• Flow angle correction on real time or frozen images, • Layouts: 1-up, 2-up, 3-up and 4-up
with velocity readout update
• Sweep angle selections in increments of 10°: 20° to
• Transmit frequencies: 1.8 MHz to 5 MHz 140°, 145°
• Pulse repetition frequencies (PRF): 2,000 Hz to • VR Rotation: X (M knob), Y (PW/CW knob), and Z
50,000 Hz (C knob)
• Wall filter selections: 40 Hz to 1,000 Hz • VR Fixed rotation: 0°, 90°, 180° and 270°
• Doppler cine control feature: up to 30 seconds of • VR Flip: Up/Down, Front/Back, Left/Right, Down/Up,
Doppler data can be stored Back/Front, and Right/Left
• Sweep speed selections: up to 10 • VR Reset: All, Curve and Orientation
• Gain in one decibel increments: -30 dB to 30 dB • Volume slice selection using 3D/4D knob
• Dynamic range in five decibel increments: 10 dB to • MPR Maps: A to G
80 dB
• MPR Tints: 1 to 10
• Doppler gain, scale, baseline, spectral invert, sweep
• MPR DR: 10 to 80 (increments of 1)
speed, wall filter, edge, UltraArt universal imaging,
grayscale map, colorization (tint) map, flow angle • VR Tints: 1 to 7
correction, and dynamic range adjustment on real time • VR Contrast Settings: 25 to 95 (increments of 1)
or frozen images • VR Brightness Settings: -100 to 100 (increments of 1)
• Adjustable audio volume with different levels and a • VR Smooth: 0 to 3
mute control: up to 21 levels
• VR Threshold: 0 to 255 (increments of 1)
• Automatic optimization of scale, baseline, gain and/or
• VR Opacity: 0 to 86 (increments of 2)
dynamic range selected manually or upon entering
freeze
siemens-healthineers.com/ultrasound 9
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
4D-mode Features
Cybersecurity
• Acquires and enables simultaneous viewing of three-
Your ultrasound system includes a security package to
dimensional images in real-time for assessing motion.
protect patient confidentiality and system security.
• User-adjustable region of interest and volume of
interest during setup (ROI/VOI), for size, position and
Patient Data Encryption Option
curve (VOI only).
• Available in combination with 2D-mode • Data storage encryption
- Protects patient information and system settings
• Requires the compatible transducers: 9VE4
stored on the ultrasound system by preventing
• Supported studies: Gynecology unauthorized access
• Sweep speed quality selections: up to 3 - Includes a USB storage device with the recovery key
• Layouts: 1-up, 2-up, 3-up and 4-up for recovering encrypted data
• Sweep angle selections in increments of 10°: 20°
to 80° System Operating Software
• VR Rotation: X (M knob), Y (PW/CW knob), and Z • Windows-based operating software
(C knob) • Multi-language capability
• VR Fixed rotation: 0°, 90°, 180° and 270° • User-modifiable configuration settings, including
• VR Flip: Up/Down, Front/Back, Left/Right, Down/Up, libraries of annotations and body markers
Back/Front, and Right/Left • User-defined exam, transducer, and image parameter
• VR Reset: All, Curve and Orientation settings
• VR slice selection using 3D/4D knob • Supports DICOM file format
• MPR Maps: A to G • Supports PC file format
• MPR Tints: 1 to 10 • Compatible with remote update handling for remote
• MPR DR: 10 to 80 (increments of 1) application support and remote troubleshooting
• VR Tints: 1 to 7
10 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
InFocus Coherent Image Formation Clarify can decrease artifacts in the 2D-mode image,
resulting in an improved view of anatomical structures.
InFocus coherent image formation focuses the image at Decreased artifacts can enhance definition of both
all depths and exploits high beamformer output capacity, tissue and vessel walls by increasing contrast resolution
which increases image uniformity. More information and improving boundary detection. Clarify uses flow
is harvested from the usual transmit sequence, using information to eliminate noise and reverberation from
massive overlapping multibeam groups rather than vessels.
individual or close parallel beam lines as in conventional
systems. This secondary beamforming enabled with
Speed of Sound
InFocus, physics-based delay, phase and amplitude
corrections can be made across transmit events to • Optimizes the 2D image by adjusting the speed of
significantly sharpen the image and improve spatial sound
resolution beyond what is typical for a given transducer • Available only with the following exams: Breast, MSK,
frequency. Abdomen Difficult
Modality Compare
siemens-healthineers.com/ultrasound 11
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
Available for transducers compatible with needle guide • 2D-mode and M-mode have unlimited sets of
attachments. measurement markers per image for distance
Please see transducer flyer for additional information. measurements
• Doppler has unlimited sets of measurement markers
Clips per image for velocity measurements
• Label then measure or measure then label workflow
• Prospective capture of motion image data
for individual labels
• Retrospective capture of motion image data
• Customizable reports
• Acoustic Rate Capture
• Institution logo, export to PDF
• Variable clip length (1 to 300 seconds)
• Clip formats General 2D-mode Measurements and Calculations
- Compressed JPEG Lossy
- Uncompressed Clips • Distance
- AVI and JPEG • Depth
- DICOM • Angle
• Supports storing of individual images within clips • Circumference (using a trace or ellipse measurement
tool)
Protocols • Area (using an ellipse or trace measurement tool)
A protocol is a predefined checklist that guides you • Volume and stenosis
through a clinical workflow. • Volume flow (using an area or diameter measurement
• Define protocol views that include an image or clip tool with a Doppler measurement)
with measurements and annotations • Ratio calculations (using two area, two distance
• Change the sequence of views during a protocol measurements)
• Import or export protocols using a USB storage device • Supports pediatric and adult use cases
DICOM 3.0 standard. • Supports alpha angle, alpha and beta angle, femoral
head coverage
• Sonometer report option
Measurements, Calculations,
and Reports General M-mode Measurements and Calculations
• Distance/depth
The measurement function is available during a patient • Heart rate
exam or with stored images. Each measurement label • Slope
supports up to 5 measurements. The ultrasound system • Time
copies only the five most recently labeled measurements
to the report.
12 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
siemens-healthineers.com/ultrasound 13
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
Mobility
Operator Control Panel
Hibernate decreases the time to power on and off the
ultrasound system. The system is ready for use in • Backlit controls and keys
approximately 30 seconds. • Control panel adjustment for standing and sitting
• Compact and lightweight industrial design positions
- Left/right swivel: ± 90°
• Steering handle for portability and maneuvering
- Range of height: 74 cm to 97 cm (29.13 in to
• Rear handle for repositioning the ultrasound system 38.18 in)
• Four locking swivel wheels • Control panel text available in: English, German,
• Transducer holders and cable management French, Spanish, Italian
• Tilt down monitor • Control panel layout supports ambidextrous operation
• Locking arm for monitor • Touch screen
• Locking control panel - 39.624 cm (15.6 inch) diagonal widescreen
- Full high-definition video
- Variable tilt angle: 30° to 50°
User-Accessible Connections - Touch screen for use with gloved hands
• USB ports on the right side of the monitor for • Gel warmer
importing and exporting protocols and exams,
archiving, and serviceability (quantity 2) Monitor
• USB ports on the left side of touch screen for
• Full high-definition video display
importing and exporting protocols and exams,
- 54.61 cm to 55.88 cm (21.5 in to 22 in) diagonal
archiving, and serviceability (quantity 2)
- 16:9 widescreen format
• USB ports on the input/output panel for peripheral - 1920 × 1080 resolution
devices (quantity 4) - 1024 × 768 image screen capture for images and
• DC power sockets on the input/output panel for reports
on-board peripherals (quantity 2) • Organic light-emitting diode (OLED) technology
• DC power sockets on each side. the control panel - High contrast ratio
for the gel warmer (quantity 2) - Uniformity across a range of environmental lighting
and viewing angles
- Reduced glare in all working environments
14 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
For detailed transducer information please refer to the Cardiac Imaging Option
ACUSON Sequoia Transducer Flyer. • Cardiac imaging
• Lightweight, ergonomic transducer design • Pulsed Wave DTI (Doppler tissue imaging) capability
• Touch screen and gesture transducer selection • Continuos Wave (CW) capabilities
• Four array transducer ports • Cardiac measurements and reports
• Auxiliary continuous wave transducer port (option) • Compatible transducers: 5V1, 8V3, 10V4, CW2
• Compact-pinless connectors
• Integrated storage shelf for transducer connectors Fusion Imaging Option
• Transducer holders for all transducer designs and gel • Fusion imaging aligns reference data with a real-time
bottle storage ultrasound image for diagnosis and interventional
procedures
• Supported reference data: computed tomography,
magnetic resonance imaging
• Compatible transducers: 4V1, 5C1, DAX Supported
studies: Abdomen, General
• Compatible transducers: 10L4 Supported studies:
Musculoskeletal, General
• Fusion supports the following components
- Ascension Technologies Corporation driveBAY 2
electromagnetic tracking system
- Ascension Technologies Corporation Mid-range
Transmitter (for the driveBAY 2)
- GCX pole to support mid-range transmitter
- CIVCO General Purpose 8 mm sensors
siemens-healthineers.com/ultrasound 15
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
• Provides a qualitative representation of relative tissue • Supports the following contrast agent destruction
stiffness for the region of interest techniques: Burst, Flash Sequencing
• Strain Ratio provides a quantifiable method to compare • Compatible transducers: 7L2, 10L4, 14L5, 18L6, DAX,
the relative stiffness of tissue within two user- 5C1, 9C3, 18H6, 11M3, 4V1, 5V1, 8V3, 10V4, 9EC4,
selectable regions of interest 9VE4
• Compatible transducers: 10L4, 14L5, 18L6 • VueBox Support
Supported studies: Breast, Thyroid, General - Requires clips acquired during contrast agent
• Compatible transducers: 9EC4 imaging and stored in DICOM format
Supported studies: Prostate, General - Supports compatibility with the VueBox software
application from Bracco. The software is used to
perform qualitative and quantitative analysis of
Virtual Touch Point Shear Wave Option
linear data. For information regarding the software
• Measures tissue shear velocity (Vs) and elasticity (E) application, refer to the manufacturer’s instructions.
for a selected region of interest using acoustic - Compatible transducers: 7L2, 10L4, 14L5, 18L6,
radiation force impulse (ARFI) technology of DAX, 5C1, 9C3, 18H6, 11M3, 4V1, 5V1, 8V3, 10V4,
diagnostic ultrasound to induce tissue displacement. 9EC4, 9VE4
• Provides measurement labels for sites, lesions, and
liver segments eSie OB Measurements Option
• Compatible transducers: 5C1, DAX, 4V1 • Provides system-generated measurements of crown
Supported studies: Abdomen rump length, biparietal diameter, head circumference,
• Compatible transducers: 10L4 abdominal circumference, femur length, humerus
Supported studies: Abdomen length, and occipital frontal diameter
• Supports biparietal diameter outer-to-outer or outer-to-
Virtual Touch Shear Wave Option inner measurements
* At the time of publication, the United States Food and Drug Administration has cleared ultrasound contrast agents only for use in left ventricular opacification
(LVO), liver, and vesicoureteral reflux (VUR). Check the current regulations for the country where you are using this ultrasound system for contrast agent
clearance.
16 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
Transducer Options
Refer to the following table to identify transducers
compatible with your ultrasound system.
siemens-healthineers.com/ultrasound 17
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
18 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
System Requirements
Power Supply Requirements
Mains Voltage 100 V ~ to 240 V ~
Maximum Current 5.4 to 13.0 amps
Frequency 50 to 60 Hz
Noise Level 35 to 38 dB
siemens-healthineers.com/ultrasound 19
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
Input and Output Signals for Audio, Video, and Data Transmission Connections
USB-A (four ports) Input/output panel Printer, Fusion electronics unit, footswitch, Bi-directional
Blu-ray/DVD/CD combination drive
DisplayPort Input/output panel External monitor Input
USB-A (two ports) Left side of the touch USB storage device, headset and camera for Bi-directional
screen virtual communication with a Siemens
service representative
USB-A (two ports) Right side of the monitor USB storage device, headset and camera for Bi-directional
virtual communication with a Siemens
service representative
ECG connector Physio panel ECG leads Input
The ultrasound system supports the following options for connection to wireless networks.
20 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
Environmental Requirements
siemens-healthineers.com/ultrasound 21
Data sheet · ACUSON Sequoia ultrasound system VA25 SW
• Type of protection against electrical shock: The diagnostic ultrasound system is in compliance
Class I, external powered with the following standards, including all applicable
• Degree of protection against electrical shock: amendments at the time of product release.
- Type BF applied part for endocavity, linear, curved,
and phased array transducers Quality Standards
- Type B applied part for Fusion transducer sensors
• FDA QSR 21 CFR Part 820
and needle tracking sensors
- Type CF defibrillation-proof applied part for ECG • EN ISO 13485 and ISO 13485
• Degree of protection against harmful ingress of water: • ISO 9001
Ordinary equipment
• Degree of safety of application in the presence of a Design Standards
flammable anesthetic material with air or with oxygen • ANSI/AAMI ES 60601-1
or nitrous oxide:
• CSA C22.2 No. 601.1
Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen • EN 60601-1 and IEC 60601-1
or nitrous oxide • EN 60601-1-2 and IEC 60601-1-2 (Class A)
• Mode of operation: • EN 60601-1-6 and IEC 60601-1-6
Continuous operation • EN 60601-2-18 and IEC 60601-2-18
• Ingress protection levels: • EN 60601-2-37 and IEC 60601-2-37
- Transducers: IPX8
• EN 62304 and IEC 62304
- Footswitch: IPX8
• EN 62366 and IEC 62366
• EN ISO 14971 and ISO 14971
22 siemens-healthineers.com/ultrasound
ACUSON Sequoia ultrasound system VA25 SW · Data sheet
CE Declaration
EU Authorized Representative
siemens-healthineers.com/ultrasound 23
The products/features mentioned in this document
may not be commercially available in all countries.
Due to regulatory reasons their future availability
cannot be guaranteed. Please contact your local
Siemens Healthineers organization for further details.
Published by Siemens Medical Solutions USA, Inc. · 9534 0920 online · © Siemens Medical Solutions USA, Inc., 2020