Acceptance Testing for CT Image Displays
Acceptance Testing for CT Image Displays
ty
u
lo
a
c
d
M
s
n
m
ie
S
Quality Assurance
Image display devices
10351301
07740769
10045692
08874427
08872017
08379492
08377520
08098951
07734713
07543106
07543015
07407039
07407013
07393114
07114023
07005502
04806118
03815490
03813933
03812950
03810160
03810152
03810137
03810079
03804890
© Siemens AG 2006
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
Luminance meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Devices recommended: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Items to be tested . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Ambient lighting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Viewing conditions at the operating location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
11 ______ Characteristics and reference values for the constancy test ___________ 19
• On each page of the protocol, the report number has to be stated. The report number
comprises the order number and the date of the last test. This ensures explicit alloca-
tion of each protocol sheet. In place of the order number, you may want to provide the
KS number (master customer number).
• The completed protocol cannot contain empty fields. Fields that do not require an input
have to be identified with n.a. (not applicable) and/or crossed out.
• Do not fill out protocols using a pencil or use white-out to correct errors. Incorrect
entries need to be crossed out. The correct values have to be entered including the date
and the initials of the person performing the correction.
• The starting values for the constancy test determined and documented during the tests
have to be documented in a separate test protocol.
• The operator receives the original and a copy of the test protocol.
• The completed acceptance test protocol has to be signed on page 1-1 by the person
who performed the test (including name, date, and signature).
• Application category A These image display devices have to meet especially high
demands regarding image quality: Application category A: This application category
includes all image display devices for the diagnosis of digital images and for which high
spatial and contrast resolution are required. Image displaye devices of application cate-
gory A: have to display images in the diagnostically required format and with the diag-
nostically required number of image pixels, that is, at least 1000 x 1000 image pixels.
• Application category B Application category B includes all image display devices for
image diagnosis which are not classified in application category A and allocate the
image display devices for image viewing. The spatial and contrast resolution required
for individual application has to be ensured.
• Application category C: Image displaye devices of Application category C: cannot be
used for diagnosing or viewing medical images. They serve, for instance, as control
monitors for patient monitoring or as image display devices for displaying operating pro-
cedures.
• Prior to the test it needs to be determined which monitors are used for reporting (as dis-
cussed with customer). They have to be identified with a label.
Siemens AG
ATD TD LC Werkzeuglogistik
Siemensstr. 33
D 71254 Ditzingen
Tel. 0711-137-6510/6513
Fax 0711-137-6550/6531
E-mail: [email protected]
Devices recommended: 0
Test image
a) digital or analog test images
b) clinical reference images
Ambient lighting 0
It is recommended to allow for adjustment of the general lighting so that the values deter-
mined from table 1 can be obtained.. If this is not possible, please change the location of
the monitor or install diaphragms.
The ambient light influences the luminance density of the image display device.
The required minimum value of the relationship between white luminance density and veil-
ing luminance for application category A and B is listed in the table
Lmax/ Ls has to be larger than 40 (CT monitors belong to application class B).
¹ Lmax is the measured value of the white box of the test image. (Distance 40-60
cm).
¹ Lmin is the measured value of the black box of the test image (distance 40-60
cm).
¹ LS is the measured value in the center of the switched off monitor (distance
40-60 cm).
contrast 0
• The room lighting is adjusted to the operating conditions (it may be necessary to ask the
user) using a dimmer with a scale.
• When the image display device is switched off, the luminance density device is used to
measure and document the veiling luminance Ls in the center of the screen. The lumi-
nance density measurement device is located at the distance (40-60 cm) specified by
the manufacturer.
Maximum contrast 0
• While the monitor is switched on, the luminance density measurement device mea-
sures the minimum and maximum luminance.
• For test image 2, the two squares for the minimum gray value (L min.) and the maximum
gray value (L max, bright field) are used.
• For test image 6, the squares 0 % are used for L min. and 100 % for L max.
• The luminance density measurement device is kept at the conventional viewing dis-
tance (40-60 cm) from the image display device.
The ratio L max /L min is computed as the maximum contrast MK.
For image display devices of application category A, the ratio has to be L max./ L min. >
100.
For image display devices of application category B, the ratio has to be L max./ L min. > 40.
Lcenter
- Convergence =
When displaying grid lines on the
color monitor
NOTICE The display of the test image cannot show one of the listed errors
in diagnostically important areas.
• At least one clinical reference image should be provided for the acceptance test of the
imaging device. Only the use of a clinical reference image enables testing the image
quality of the clinical images required by the user.
• The clinical reference image serves for the visual test of the grayscale value display, of
the resolution and of the existence of artifacts (specially pseudo-contours).
If a clinical reference image is available, it will be displayed at the image display device
and the predefined display parameters will be taken into account. The user has to confirm
the quality of the clinical reference image/reference images. If the quality is not accepted,
the characteristic line may be adjusted to user-specific needs, if required. The final setting
or the look-up table used has to be documented. If the monitor settings are changed, the
respective test positions have to be checked again and documented, if necessary.